Page 1 Datasheet - detector accessory Evaluation kit for Fabry-Pérot detectors InfraTec GmbH Infrarotsensorik und Messtechnik Dresden / GERMANY www.InfraTec.de The evaluation kit supports customer needs for an initial test of InfraTec´s Fabry-Pérot detectors (FP) without having to develop test circuitry and software themselves. It allows easy control of detector and IR source with customized software to analyze the detector signals. With the help of this kit a quick and easy configuration of a simple FP spectrometer is possible. Overview The kit includes hardware for a first test setup, a test and measurement software and a manual. Mechanical parts, like gas cell or holder of detector/IR source, define the IR absorption in an application specific way. Thus, these parts have to be designed by the user and are not part of the kit. Hardware features Requirements IR source control with constant current operation, up to 800 mA DC to 500 Hz (square, 0-100 % duty cycle) detector control Signal acquisition with 12 bit, 2 kHz PC with Windows 2000, XP, Vista, 7 or 10 USB 2.0 Interface Power supply 12 V DC, stabilized (800 mA minimum)

Page 1 Datasheet - detector accessory Evaluation kit for ......Please, find more details in the evaluation kit manual on the attached CD-ROM and in our Fabry-Pérot detector application

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Page 1 Datasheet - detector accessory Evaluation kit for ......Please, find more details in the evaluation kit manual on the attached CD-ROM and in our Fabry-Pérot detector application

Page 1 Datasheet - detector accessory Evaluation kit for Fabry-Pérot detectors

InfraTec GmbH Infrarotsensorik und Messtechnik Dresden / GERMANY www.InfraTec.de

The evaluation kit supports customer needs for an initial test of InfraTec´s Fabry-Pérot detectors (FP) without having to develop test circuitry and software themselves. It allows easy control of detector and IR source with customized software to analyze the detector signals. With the help of this kit a quick and easy configuration of a simple FP spectrometer is possible. Overview

The kit includes hardware for a first test setup, a test and measurement software and a manual. Mechanical parts, like gas cell or holder of detector/IR source, define the IR absorption in an application specific way. Thus, these parts have to be designed by the user and are not part of the kit. Hardware features Requirements � IR source control with constant current operation, up to 800 mA DC to 500 Hz (square, 0-100 % duty cycle) � detector control � Signal acquisition with 12 bit, 2 kHz � PC with Windows 2000, XP, Vista, 7 or 10

� USB 2.0 Interface � Power supply 12 V DC, stabilized (800 mA minimum)

Page 2: Page 1 Datasheet - detector accessory Evaluation kit for ......Please, find more details in the evaluation kit manual on the attached CD-ROM and in our Fabry-Pérot detector application

Page 2 Datasheet - detector accessory Evaluation kit for Fabry-Pérot detectors

Rev.0 , © InfraTec GmbH

Software features � IR source control

� modulation frequency � duty cycle � driving current

� detector control � detector information � read out of FP filter calibration data � detector temperature

� signal evaluation with FFT technique � mapping of spectra

User interface of evaluation software; signal spectrum of ambient air measured in scan mode Operation modes � Basic operation

� adjust filter wavelength manually � measurements can be recorded to log file

� Sequence mode � for comparing with conventional multispectral solutions (filter wheel, multi color detectors) � adjusting filter periodically to set of wavelengths

� Scan mode � obtain complete spectra for desired wavelength range and step size � display or save as raw signal spectra or transmittance spectra (after recording background spectrum)

� Sweep mode � can accelerate spectra recording remarkably � IR source DC-driven, FP-filter scanned continuously

� spectral information used for IR modulation � adapt waveform, tuning speed and wavelength range for best suiting application specific configuration � spectra can be recorded for further external analysis

Signal voltage of LFP detector @ 10 Hz modulation frequency and constant control voltage

Sweep mode measurement of a polystyrene foil; detector signal and control voltage Please, find more details in the evaluation kit manual on the attached CD-ROM and in our Fabry-Pérot detector application notes on our website. Please, be aware the Evaluation kit is not intended for usage beyond testing and evaluation of InfraTec’s FP-detectors.