57
1 Invasive Cardiac Invasive Cardiac Pacing Pacing (Pacemaker) (Pacemaker) Prepared by: Ahmad Khalil Al-Sadi (RN, MSN, CNS)

Pacemaker Invasive Cardiac Pacing

Embed Size (px)

Citation preview

Page 1: Pacemaker Invasive Cardiac Pacing

1

Invasive Cardiac PacingInvasive Cardiac Pacing(Pacemaker)(Pacemaker)

Prepared by: Ahmad Khalil Al-Sadi (RN, MSN, CNS)

Page 2: Pacemaker Invasive Cardiac Pacing

2

OutlineOutline Definition.Definition. Indications for invasive cardiac pacing.Indications for invasive cardiac pacing. Types of pacemakers.Types of pacemakers. Components of Pacing. Pacemaker concepts:1) Connection1) Connection2) Output (rate, amount, chamber)2) Output (rate, amount, chamber)3) Capture3) Capture4) Sensetivity4) Sensetivity

Page 3: Pacemaker Invasive Cardiac Pacing

3

Pacemaker Is an artificial electronic device that provide

electrical stimuli to cause cardiac contraction during periods when intrinsic cardiac electrical activity is inappropriately slow or absent.

Used to initiate myocardial contractions when 1) Intrinsic stimulation is insufficient, 2) The native impulses are not being conducted,3) The heart rate is too slow to maintain an adequate

cardiac output.

Page 4: Pacemaker Invasive Cardiac Pacing

4

Care of Patients With PacemakersCare of Patients With Pacemakers Complications Related to Pacemakers.Complications Related to Pacemakers. Patient education regarding permanent Patient education regarding permanent

pacemaker insertion procedure.pacemaker insertion procedure.

Page 5: Pacemaker Invasive Cardiac Pacing

5

Indications for temporary Indications for temporary invasive cardiac pacinginvasive cardiac pacing

1) Sick sinus syndrome:1) Sick sinus syndrome: Symptomatic sinus arrest.Symptomatic sinus arrest. Suppression of ventricular ectopy resulting from Suppression of ventricular ectopy resulting from

bradycardiabradycardia Atrial fibrillationAtrial fibrillation Bradycardia/tachycardia syndromeBradycardia/tachycardia syndrome Symptomatic sinus bradycardiaSymptomatic sinus bradycardia2) Heart blocks:2) Heart blocks: Type I (occasionally) and type II second-degree Type I (occasionally) and type II second-degree

atrioventricular blockatrioventricular block Acute bifascicular or trifascicular blockAcute bifascicular or trifascicular block Complete atrioventricular blockComplete atrioventricular block Cardiac arrest with ventricular asystoleCardiac arrest with ventricular asystole

Page 6: Pacemaker Invasive Cardiac Pacing

6

3) Drug-refractory dysrhythmia:3) Drug-refractory dysrhythmia: Overdrive ventricular pacing to suppress or Overdrive ventricular pacing to suppress or

prevent ventricular ectopic activityprevent ventricular ectopic activity Overdrive atrial pacing to “break” Overdrive atrial pacing to “break”

supraventricular tachycardia or atrial fluttersupraventricular tachycardia or atrial flutter4) Cardiovascular surgery:4) Cardiovascular surgery: Prophylactic use during anesthesia and Prophylactic use during anesthesia and

surgery in patients with a history of acute surgery in patients with a history of acute coronary syndrome or cardiac dysrhythmiascoronary syndrome or cardiac dysrhythmias

Treatment for complete heart block Treatment for complete heart block developed during or after surgerydeveloped during or after surgery

Cardiac output augmentation postoperativelyCardiac output augmentation postoperatively

Page 7: Pacemaker Invasive Cardiac Pacing

7

5) Recent myocardial infarction: Treatment with β-blocking agents Srious bradydysrhythmias often occur

even without drug therapy. Temporary pacing after myocardial

infarction can be a lifesaving intervention, allowing patients to enjoy the benefits of β-blocker therapy while simultaneously maintaining an adequate heart rate and cardiac output.

Page 8: Pacemaker Invasive Cardiac Pacing

8

Types of pacemakers Types of pacemakers Permanent/ long-term pacemaker:Permanent/ long-term pacemaker: An incision (cut) is made in the neck or chest. The An incision (cut) is made in the neck or chest. The

leads are guided through a vein into the heart, and leads are guided through a vein into the heart, and attached to the generator.attached to the generator.

The generator is placed in a pocket under the skin.The generator is placed in a pocket under the skin. Transvenous/ Temporary pacemaker :Transvenous/ Temporary pacemaker : The leads for the pacemaker are guided into the The leads for the pacemaker are guided into the

heart through an incision in neck, chest, or leg. heart through an incision in neck, chest, or leg. The generator is worn outside the body in a small The generator is worn outside the body in a small

pouch.pouch.

Page 9: Pacemaker Invasive Cardiac Pacing

9

Components of Pacing

1) The battery-powered generator that initiates electrical stimulation.

Three basic types of pulse generators are available: 1. single-chamber atrial pacers, are adequate for enhancing heart rate in patients with an intact cardiac conduction system.

2. Single-chamber ventricular pacers, designed for patients with normal contractility who require an increase in heart rate.

3. Atrioventricular sequential pacers, pace both atrial and ventricular chambers.

Provide synchronization between the atria and the ventricles for optimal cardiac output.

Page 10: Pacemaker Invasive Cardiac Pacing

10

Permanent pacemaker pulse Permanent pacemaker pulse generatorsgenerators are usually powered by are usually powered by lithium batteries that function for lithium batteries that function for five to eight years before they need five to eight years before they need to be replacedto be replaced..

Replacing the generator usually Replacing the generator usually requires a simple procedure in which requires a simple procedure in which a a repeat incisionrepeat incision is made, the old is made, the old generator is removed, and a new generator is removed, and a new generator is implanted and joined with generator is implanted and joined with the existing leads.the existing leads.

Page 11: Pacemaker Invasive Cardiac Pacing

11

2) The pacing generator is connected to pacing wires/leads and may or may not require a connecting cable.

3) The patient’s own cardiac tissue: Failure to pace may be due to a

malfunction or disruption of mechanical components.

May also be due to poor myocardial function associated with electrolyte disturbances, myocardial scarring, or any factors that affect impulse conduction or cardiac contractility.

Page 12: Pacemaker Invasive Cardiac Pacing

12

Pacemaker Functioning Concepts

(Connection, output, capture, sensetivity)1) The Connection between the pacemaker

generator and the heart:A) Unipolar electrode wire, only the negative

electrode is in direct contact with the heart.B) Bipolar electrode wires, both negative and

positive electrodes lie within the heart. Pacemakers can be either unipolar or bipolar. Distinguishing between the negative and

positive electrodes is important so that the wires are connected appropriately to the pulse generator.

Page 13: Pacemaker Invasive Cardiac Pacing

13

ConnectionConnection Types of invasive temporary pacing:

1) The Transvenous category, involves a pulse generator externally connected to 2 electrode wires, threaded through a large vein into either the right atrium or the right ventricle. These wires directly contact the endocardium within the heart.

Page 14: Pacemaker Invasive Cardiac Pacing

14

ConnectionConnection2) Directly stimulating the epicardium: Initiated after cardiac surgery. Electrodes are lightly

sutured to the epicardium before the thorax is closed.

These pacing wires are pulled through the skin and secured to the external chest wall, ready for attachment to a temporary pacing generator as needed.

A patient may have a single set or a double set of electrodes, but each set of electrodes includes 2 wires that protrude from a stab incision in the chest wall.

Ventricular pacing wires exit through the left side of the sternum; atrial pairs are placed on the right side.

Having identifying labels on the wires is helpful. If no labels are present, ascertaining which wire is which (and marking the wires accordingly) can save precious time later in emergency situations.

Page 15: Pacemaker Invasive Cardiac Pacing

15

Temporary pulse generator for a dual-chamber pacemakerOn the top of the generator of each temporary atrioventricular sequential pacemaker are connectors labeled Atrial +/-, and Ventricular +/-. Single-chamber pacemakers have only 2 connector ports, positive and negative; the wires must be connected and secured to the correct port.

Page 16: Pacemaker Invasive Cardiac Pacing

16

Page 17: Pacemaker Invasive Cardiac Pacing

17

2) 2) Output (rate, amount, and chamber) The sole function of the generator is to supply

sufficient energy to the heart muscle to stimulate a contraction.

A) The Rate: determines the number of stimulations to be delivered per minute.

The original rate setting depends on both the patient’s condition and the reason for pacing.

Rates for a surgical patient can start as high as 90 to 110 beats/min.

In medical patients, therapy is generally started at 70 to 90 beats/min.

In patients who have had cardiac arrest, the initial rate is 80 beats/min.

The heart rate on a patient’s rhythm strip should never be lower than the patient’s set pacemaker rate.

Page 18: Pacemaker Invasive Cardiac Pacing

18

B) The Output Amount: Is the level of energy delivered by the pulse

generator to the heart to initiate depolarization. Output is measured in milliamperes (mA). The usual starting point is 10 mA in nonurgent

situations. Output is then slowly increased until capture is

obtained and the “pacing threshold” is defined. This level is not constant; it fluctuates over time

as an endothelial sheath forms around the tips of the electrodes. Therefore, to prevent loss of capture, the output is set 1.5 to 3 times higher than the identified pacing threshold.

In emergent circumstances, starting with a high output (15-20 mA) is recommended.

Page 19: Pacemaker Invasive Cardiac Pacing

19

C) The chamber: The location in the heart to which the

energy is delivered. The atrial, ventricular, or both chambers

can be paced. Pulse generators for dual-chamber

pacemakers have separate atrial and ventricular output controls.

The settings for the 2 chambers may be different or identical.

Page 20: Pacemaker Invasive Cardiac Pacing

20

The ability of the electrical impulse to initiate a cardiac responsecardiac response, is detected by examining an electrocardiogram.

Capture is both an electrical & a mechanical event. Electrical capture is indicated by a pacer spike

followed by a corresponding P wave or QRS complex, depending on which chamber is being paced (F).

If the atrium is paced, the spike appears before the P wave.

If the ventricle is paced, the spike occurs before the QRS complex.

Because the pacemaker causes the heart to depolarize in an artificial fashion, the path of depolarization is abnormal, resulting in widened P waves and QRS complexes.

3) Electrical Capture

Page 21: Pacemaker Invasive Cardiac Pacing

21

Normal electrical capture, atrial pacer. Normal atrial electrical capture is demonstrated by a pacer spike (Ap) followed by a corresponding P wave.

Abbreviations: Ap, atrial pacing; As, atrial stimulation (intrinsic).

Page 22: Pacemaker Invasive Cardiac Pacing

22

Normal electrical capture, ventricular pacer. Normal ventricular electrical capture is demonstrated by a pacer spike (Vp) followed by a corresponding, widened QRS complex.Abbreviations: Vp, ventricular pacing; Vs, ventricular stimulation (intrinsic).

Page 23: Pacemaker Invasive Cardiac Pacing

23

A pacer spike without a corresponding P wave or QRS complex indicates failure to capture (F).

If loss of capture occurs, the patient is assessed first and then connections and settings are checked to detect disconnections, broken wires, or other mechanical issues.

If the patient’s condition is stable, threshold testing is done.

If hemodynamic compromise exists, the milliamperae/amount is quickly increased until capture occurs.

Page 24: Pacemaker Invasive Cardiac Pacing

24

Turning the patient onto his/her left side may also improve capture by increasing contact between the electrode and myocardial tissue.

Electrical capture alone is inadequate. Adequacy of mechanical capture is assessed by feeling for a pulse or checking blood pressure.

Mechanical capture exists when the pacer spike and its corresponding QRS complex are followed by a cardiac contraction.

Page 25: Pacemaker Invasive Cardiac Pacing

25

Failure to capture, dual-chamber pacemaker. Failure to capture occurs when a pacer spike is present but is not followed by a corresponding waveform (P wave or QRS complex). Arrow indicates electrical stimulus without ventricular capture.

Page 26: Pacemaker Invasive Cardiac Pacing

26

4) Sensitivity The ability of the generator to detect &

recognize the impulses the myocardial tissue is generating on its own (Intrinsic cardiac activity ).

Intrinsic activity stimulates better contractions, & more effective forward blood flow, from the chambers of the heart, especially when the atria and ventricles are intrinsically beating synchronously.

30% of normal C.O. is due to the atrial “kick” or atrial systole, that occurs during ventricular filling when the 2 chambers perform in sync.

Page 27: Pacemaker Invasive Cardiac Pacing

27

When atrial beats are unsynchronized or absent, as in atrial fibrillation, atrioventricular block, or sinus arrest, cardiac output decreases because less blood than usual is ejected from the atria during ventricular diastole.

Intrinsic beats must be supplemented when the rate is insufficient or conduction of the beats does not generate ventricular contractions.

For maximum effectiveness, paced beats and intrinsic beats must be synchronized.

To synchronize the beats, the generator first analyzes the intrinsic rhythm and then stimulates the heart only as needed.

Page 28: Pacemaker Invasive Cardiac Pacing

28

Adjusting the sensing level sets the pacer to “look” for intrinsic beats, & provide pacing on demand.

Pacemaker generators can be programmed to deliver an impulse to the ventricle each time an atrial beat is sensed, or they can be set to stimulate only when no intrinsic beat has been detected during a predetermined interval.

The sensitivity setting is measured in millivolts (mV) and is initially set at about 2 to 5 mV.

Failure to sense occurs when the generator does not recognize the heart’s intrinsic impulses.

Page 29: Pacemaker Invasive Cardiac Pacing

29

Failure to sense, single-chamber atrial pacemaker. Failure to sense occurs when the generator does not detect intrinsic beats and initiates an inappropriate impulse that may or may not capture. Arrow indicates an atrial beat not sensed by the pulse generator, which is followed by a pacer spike.

Page 30: Pacemaker Invasive Cardiac Pacing

30

Factors that influence capture and sensing

Page 31: Pacemaker Invasive Cardiac Pacing

31

The most common cause of failure to sense is displacement of the electrode.

Repositioning the patient on his or her left side may improve contact between the electrode and the myocardium.

If the response is still inadequate, then the sensitivity must be increased by turning down the millivoltage, allowing the generator to detect beats that occur at lower millivolt levels.

If the pacemaker is detecting beats that are not actually occurring (inappropriate sensing), then the sensitivity threshold must be increased to block out artifact by turning up the millivoltage.

Page 32: Pacemaker Invasive Cardiac Pacing

32

Care of Patients With Pacemakers1) Electrical safety practices, to prevent

microshocks which are associated with ventricular dysrhythmias.

Always wear gloves when handling electrodes

Cover unused transthoracic wires2) The pulse generator must be placed in a

location where it will not be dropped on the floor.

3) Lengthy cables attached to the pacemaker need careful attention so that wires are not inadvertently dislodged.

Page 33: Pacemaker Invasive Cardiac Pacing

33

4) Routine nursing care of regular cardiovascular assessment to evaluate pulses, level of consciousness, heart rhythm, pacer activity, and hemodynamic response.

5) The wires must be secure and not at risk for dislodgement.

check to see if the wires are intact and note the number & location.

6) Care for a transvenous wire site: Daily assessment of the insertion site. dressing are changed as prescribed cleaned with povidone-iodine/another

antimicrobial agent A sterile occlusive dressing is applied, with

changes every 48 to 72 hours.

Page 34: Pacemaker Invasive Cardiac Pacing

34

7) After cardiac surgery & once dressings are removed, care of the epicardial site should be performed daily whether the site is left open to air or covered with a light dressing.

The area is cleaned with isotonic sodium chloride solution and assessed for redness or drainage.

The wires must be taped securely to the skin to prevent accidental dislodgement.

Page 35: Pacemaker Invasive Cardiac Pacing

35

8) Capture and sensitivity threshold testing should be performed every 12 to 24 hours to determine the best settings for the generator.

As time passes, both transvenous and epicardial wires acquire endothelial sheaths around their tips and thus require more milliamperes to capture and fewer millivolts to sense.

Threshold testing is generally contraindicated when patients are being paced more than 90% of the time because of the risk of losing capture while the patients are dependent on generator-initiated beats.

Page 36: Pacemaker Invasive Cardiac Pacing

36

Complications Related to Complications Related to PacemakersPacemakers

Page 37: Pacemaker Invasive Cardiac Pacing

37

Complication treatment depends on the Complication treatment depends on the etiologyetiology

PneumothoraxPneumothorax, being dependent on size, may , being dependent on size, may require require chest thoracostomychest thoracostomy..

ErosionErosion of the pacer through the of the pacer through the skinskin, while rare, , while rare, requires requires pacer replacement and systemic antibioticspacer replacement and systemic antibiotics..

HematomasHematomas may require may require drainagedrainage. . Lead dislodgmentLead dislodgment usually occurs within 2 days usually occurs within 2 days

following implantation of a permanent pacer and following implantation of a permanent pacer and may be may be seen on chest radiographyseen on chest radiography. If the lead is . If the lead is floating freely in the ventricle, malignant floating freely in the ventricle, malignant arrhythmias may develop.arrhythmias may develop.

Page 38: Pacemaker Invasive Cardiac Pacing

38

ThrombosisThrombosis is rare and usually presents as is rare and usually presents as unilateral arm edemaunilateral arm edema. Treatment includes . Treatment includes arm arm elevation and anticoagulation. elevation and anticoagulation.

Advanced life support protocolsAdvanced life support protocols, including , including defibrillationdefibrillation, may safely be executed in , may safely be executed in patients with pacemakers in place. patients with pacemakers in place. Sternal Sternal paddles are placed at a safe distance (10 cm) paddles are placed at a safe distance (10 cm) from the pulse generatorfrom the pulse generator. .

Page 39: Pacemaker Invasive Cardiac Pacing

39

Patient education regarding Patient education regarding permanent pacemaker insertion permanent pacemaker insertion

procedureprocedure

Page 40: Pacemaker Invasive Cardiac Pacing

40

Before the ProcedureBefore the Procedure You will be asked to sign a You will be asked to sign a consent form.consent form. If you are pregnant or suspect that you are If you are pregnant or suspect that you are pregnantpregnant, ,

you should notify your physician. you should notify your physician. Notify your physician if you have Notify your physician if you have a history of a history of

bleeding disorders or if you are taking any bleeding disorders or if you are taking any anticoagulant (blood-thinning) medicationsanticoagulant (blood-thinning) medications. It may . It may be necessary for you to be necessary for you to stopstop some of these some of these medications prior to the procedure.medications prior to the procedure.

Your physician may request a Your physician may request a blood testblood test prior to the prior to the procedure.procedure.

Remove any Remove any jewelry or other objectsjewelry or other objects that may that may interfere with the procedure. interfere with the procedure.

Page 41: Pacemaker Invasive Cardiac Pacing

41

Not eat or drink after midnight the nightNot eat or drink after midnight the night before before the procedure.the procedure.

Tell your doctor and nurses if you are Tell your doctor and nurses if you are allergic allergic to any medicines or Iodine.to any medicines or Iodine.

Empty your bladderEmpty your bladder before the procedure. before the procedure. Your Your chest area may be shavedchest area may be shaved, and will be , and will be

prepared for the procedure.prepared for the procedure. You may be given You may be given medicine to help you relax.medicine to help you relax. An intravenous line (IV line)An intravenous line (IV line) will be inserted will be inserted

into your arm.into your arm.

Page 42: Pacemaker Invasive Cardiac Pacing

42

During the ProcedureDuring the Procedure You will be placed on You will be placed on a thin, hard tablea thin, hard table.. A A large drapelarge drape will be placed on the chest area where will be placed on the chest area where

the pacemaker will be inserted. the pacemaker will be inserted. You will be given an You will be given an antibiotic in your IV.antibiotic in your IV. The doctor will insert the pacing leads (wires) into The doctor will insert the pacing leads (wires) into

the heart, and then attach the leads to the pacemaker the heart, and then attach the leads to the pacemaker or generator.or generator.

You may be You may be awakeawake during the procedure, and you during the procedure, and you may be may be asked to take deep breaths.asked to take deep breaths.

Page 43: Pacemaker Invasive Cardiac Pacing

43

There will be a There will be a large cameralarge camera that rotates and that rotates and takes pictures of your heart as the leads are takes pictures of your heart as the leads are being positioned by the doctor. being positioned by the doctor.

It is very important to It is very important to tell the doctor or nursetell the doctor or nurse if if you have any you have any discomfort or chest pain.discomfort or chest pain.

A A large bandagelarge bandage that applies pressure will that applies pressure will have been placed on the have been placed on the incision siteincision site (the area (the area where your stitches are).where your stitches are).

Page 44: Pacemaker Invasive Cardiac Pacing

44

Page 45: Pacemaker Invasive Cardiac Pacing

45

After the Procedure:After the Procedure: Not raise the arm on the same side as Not raise the arm on the same side as the incision site above your shoulder for the incision site above your shoulder for three weeks.three weeks.Arrange to have someone drive you Arrange to have someone drive you home from the hospital following the home from the hospital following the procedure.procedure.

Page 46: Pacemaker Invasive Cardiac Pacing

46

Notify your physician &/or nurseNotify your physician &/or nurse to report to report any of the following:any of the following:

Fever and/or chills Fever and/or chills Pain, redness, swelling, or bleeding or Pain, redness, swelling, or bleeding or

other drainage from the insertion site.other drainage from the insertion site. Chest pain/pressure, nausea and/or Chest pain/pressure, nausea and/or

vomiting, profuse sweating, dizziness vomiting, profuse sweating, dizziness and/or fainting.and/or fainting.

Swelling at the pacemaker site.Swelling at the pacemaker site. Shortness of breathShortness of breath Palpitations (a feeling of irregular Palpitations (a feeling of irregular

heartbeat).heartbeat).

Page 47: Pacemaker Invasive Cardiac Pacing

47

At HomeAt Home

Not do any Not do any lifting or pullinglifting or pulling on anything for a on anything for a few weeks.few weeks.

Keep the Keep the insertion site clean and dry.insertion site clean and dry.

Page 48: Pacemaker Invasive Cardiac Pacing

48

Pacemaker Precautions/ Pacemaker Precautions/ Follow-up care Follow-up care

Always carry an Always carry an ID cardID card that states you are wearing that states you are wearing a pacemaker. a pacemaker.

Tell your Tell your surgeon or dentistsurgeon or dentist that you have a that you have a pacemaker; so that pacemaker; so that electrocauteryelectrocautery will not be used to will not be used to control bleeding (the electrocautery device can control bleeding (the electrocautery device can change the pacemaker settings). change the pacemaker settings).

Always Always consult your physicianconsult your physician if you have any if you have any questions concerning the use of certain questions concerning the use of certain equipment equipment near your pacemaker.near your pacemaker.

Always consult your physician when you Always consult your physician when you feel ill feel ill after an activityafter an activity, or when you have questions about , or when you have questions about beginning a new activitybeginning a new activity..

Page 49: Pacemaker Invasive Cardiac Pacing

49

A follow-up visit at the physician’s office is usually scheduled every six to 12 months, to ensure that the pacemaker is working properly by programming testing and If needed battery replacement is necessary

Page 50: Pacemaker Invasive Cardiac Pacing

50

Permanent Pacemaker Permanent Pacemaker programmingprogramming

Can be performed Can be performed noninvasivelynoninvasively by an by an electrophysiologist or cardiologist.electrophysiologist or cardiologist.

Patients should carry a Patients should carry a cardcard with them with them providing information about their providing information about their particular model. This information is particular model. This information is crucial when communicating with the crucial when communicating with the cardiologist about a pacer problem. cardiologist about a pacer problem.

Most pacemaker generators have an Most pacemaker generators have an x-x-ray coderay code that can be seen on a standard that can be seen on a standard chest x-ray.chest x-ray.

Page 51: Pacemaker Invasive Cardiac Pacing

51

Page 52: Pacemaker Invasive Cardiac Pacing

52

AVOIDING POSSIBLE AVOIDING POSSIBLE ELECTROMAGNETIC ELECTROMAGNETIC

INTERFERENCE INTERFERENCE

Page 53: Pacemaker Invasive Cardiac Pacing

53

Patients with artificial Patients with artificial pacemakers should be aware of pacemakers should be aware of

the following:the following: Household appliances — microwave energy.Household appliances — microwave energy. Cellular phonesCellular phones— Avoid placing a cell phone — Avoid placing a cell phone over the over the

pacemakerpacemaker (especially the antenna of the phone), and should (especially the antenna of the phone), and should not carry it in a pocket close to or over the pacemakernot carry it in a pocket close to or over the pacemaker while while the phone is on.the phone is on.

Electromagnetic Security SystemsElectromagnetic Security Systems — found in or near the — found in or near the workplace, at airports, in shopping malls, at courthouses, or workplace, at airports, in shopping malls, at courthouses, or in other in other high-security areas.high-security areas.

Be aware of the Be aware of the location of security systemslocation of security systems and move and move through them at a normal pace. through them at a normal pace.

Avoid Avoid sitting or standing closesitting or standing close to a security system. to a security system.

Page 54: Pacemaker Invasive Cardiac Pacing

54

External electrical equipmentExternal electrical equipment — In places that — In places that contain contain welding equipment or motor-generator welding equipment or motor-generator systems.systems.

Remain Remain at least two feetat least two feet from external from external electrical equipment,electrical equipment,

Verify that the Verify that the equipment is properly equipment is properly groundedgrounded,,

Wear insulated glovesWear insulated gloves when using electrical when using electrical devices, devices,

Leave the work area immediatelyLeave the work area immediately if they if they experience experience lightheadedness or other lightheadedness or other concerning symptoms.concerning symptoms.

Page 55: Pacemaker Invasive Cardiac Pacing

55

Diagnostic or therapeutic procedures:Diagnostic or therapeutic procedures:Certain diagnostic or treatment procedures may Certain diagnostic or treatment procedures may interfere with pacemakers. Thus, clinician interfere with pacemakers. Thus, clinician recommend avoiding these procedures or using recommend avoiding these procedures or using special precautions, such as reprogramming of the special precautions, such as reprogramming of the pacemaker. Such procedures include:pacemaker. Such procedures include:

Magnetic resonance imaging (MRI),Magnetic resonance imaging (MRI), a noninvasive a noninvasive diagnostic imaging procedure that uses a strong diagnostic imaging procedure that uses a strong magnetic field that is pulsed on and off at a rapid rate.magnetic field that is pulsed on and off at a rapid rate.

Page 56: Pacemaker Invasive Cardiac Pacing

56

Transcutaneous electrical nerve/muscle stimulatorsTranscutaneous electrical nerve/muscle stimulators, , a method of pain control involving the application a method of pain control involving the application of electrical impulses to muscle or nerve via of electrical impulses to muscle or nerve via electrodes placed on the skinelectrodes placed on the skin

DiathermyDiathermy, the production of heat in body tissues , the production of heat in body tissues for therapeutic purposes through high-frequency for therapeutic purposes through high-frequency electromagnetic radiation or microwaveselectromagnetic radiation or microwaves

Extracorporeal shock wave lithotripsyExtracorporeal shock wave lithotripsy, the use of , the use of sound waves to break up stones (calculi) in the sound waves to break up stones (calculi) in the urinary tract or gallbladderurinary tract or gallbladder

Therapeutic radiation for cancer or tumorsTherapeutic radiation for cancer or tumors, which , which can cause permanent pacemaker damage from the can cause permanent pacemaker damage from the radiationradiation

Page 57: Pacemaker Invasive Cardiac Pacing

57

The EndThe End

Thank YouThank You