Yasmine Calisesi. Ozone Radiometer at 142 GHz: Calibration. The Stratospheric Ozone Monitoring Radiometer SOMORA. Developed by the IAP for MeteoSwiss Tuned to Ozone at 142.175 GHz Improved version of the GROMOS instrument Operated at the IAP since January 2000 Moved to Payerne in June 2002. - PowerPoint PPT Presentation

Citation preview

Page 1: Ozone Radiometer at 142 GHz: Calibration

Ozone Radiometer at 142 GHz:Calibration

Yasmine Calisesi

Page 2: Ozone Radiometer at 142 GHz: Calibration

The Stratospheric Ozone Monitoring Radiometer SOMORA

●Developed by the IAP for MeteoSwiss●Tuned to Ozone at 142.175 GHz●Improved version of the GROMOS instrument●Operated at the IAP since January 2000●Moved to Payerne in June 2002

Page 3: Ozone Radiometer at 142 GHz: Calibration

Somora location in Payerne

Page 4: Ozone Radiometer at 142 GHz: Calibration

SOMORA Quasi-Optics

Rotating Planar MirrorHot Load

ColdLoad

Eccosorb CV-3

IF 7.1 Ghz

T = 310 K

T=77K

Page 5: Ozone Radiometer at 142 GHz: Calibration

Fans PT 100HeatingElement

EccosorbCV-3

PT 100

PolystyreneWindow

SOMORA Hot Load

Page 6: Ozone Radiometer at 142 GHz: Calibration

EccosorbCV-3

SOMORA Cold Load

Dewar Vessel

Packaging Foam Material

Page 7: Ozone Radiometer at 142 GHz: Calibration

Liquid N2

InletGaz-phase N2 exhaust

ExpansionVesselP = 0.4 Bar

Page 8: Ozone Radiometer at 142 GHz: Calibration

Total Power Calibration

coldcoldhotcoldhot

coldskysky TTT

PP

PPT

)(

,1

y

yTTT coldhotrec

0

0

PP

PPy

cold

hot

Page 9: Ozone Radiometer at 142 GHz: Calibration

Temperature Definitions

● Physical Temperature● „Planck“ Brightness Temp.:

T :

TP :

)2

1ln(

1

2

3

Pc

hk

hTP

● In the RJ limit (hkT):

P ÷ T P can be replaced by T in RT model

RJP Tk

PcT :

2 2

2

Page 10: Ozone Radiometer at 142 GHz: Calibration

Example: Planck and RJ emission curves at various frequencies

Brightness Temperature [K]

Page 11: Ozone Radiometer at 142 GHz: Calibration

Conclusions:

● SOMORA: Total power radiometer

● P ÷ T in considered T range Total power calibration formula can be used

● TRJ ≠ T

P Radiative transfer

computed in the units of P