PAHCARDIACANAESTHESIA+PERFUSION|OxygenFailure2017
OXYGENFAILURESignsofO2failure:
- Darkarterialblood- M3arterialsatsalarm- M3orB-CareLowvenoussatsalarm- OxygenanalyserlowFiO2alarmsounding- LowPaO2onABG- O2failurealarmsounding
Check:
- O2linescrewedintothewallport- O2linetapopen- O2connectedtotheoxygenatorgasinlet(greenfilter)- Rotameterflowturnedon- Blenderturnedto100%oxygen- Oxygenatorfreeofclots
IfOxygennotflowing:
1. Askforgreenpatientoxygentubingtobeconnectedtotheanaestheticmachineoxygenoutlet.(foundinemergencybox)
2. ConnecttheotherendtotheOxygenatorgasinlet;youwillneedtoinsertaspigotintotheinletinordertoconnecttheoxygen(foundinemergencyboxtapedtooxygentubing)orhavestronghandstoattachtubingdirectly
3. Controlthesweepspeedfromtheanaestheticmachine
4. Intheeventofpipedoxygenfailure,oxygencanstillbedeliveredfromthebottleontheanaestheticmachine.UrgentlyrequestanadditionalOxygenbottle