For standard applications in process andenvironmental technology, with dirt-repellentPTFE diaphragm, built-in temperature sensor(optional for analog sensor)
Application
• Long-term monitoring and limit value monitoring of processes with stable processconditions– Chemical industry: strong acids/bases, plastic, pulp and paper industry– Power plants (e.g. flue gas cleaning), oil and gas– Incinerator plants
• Water/wastewater treatment– Boiler feedwater and cooling water– Well water and drinking water– All industrial and municipal treatment plants
With ATEX, IECEx, FM, CSA, TIIS and NEPSI approval for use in hazardous areas
Your benefits
• Low-maintenance and robust thanks to large PTFE ring junction• Can be used at pressures up to 17 bar abs. (246 psi)• Process glass also for very alkaline applications (BA and BT versions)• Process glass for applications in media containing hydrofluoric acid (FA version)• For media with low conductivity (AS version)• Integrated NTC30K temperature sensor (Memosens) for effective temperature
compensation; Pt100 or Pt1000 for analog sensors• Optional: Poison-resistant reference with ion trap
Other advantages of Memosens technology
• Maximum process safety• Data security thanks to digital data transmission• Very easy to use as sensor data saved in the sensor• Recording of sensor load data in the sensor enables predictive maintenance with
the Memobase Plus CYZ71D
Products Solutions Services
Technical InformationOrbisint CPS11D and CPS11pH electrodes, analog or with digital Memosenstechnology
TI00028C/07/EN/14.1571299911
Orbisint CPS11D and CPS11
2 Endress+Hauser
Function and system design
Measuring principle pH measurementThe pH value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. Themembrane glass of the electrode supplies an electrochemical potential which is dependent upon thepH value of the medium. This potential is generated by the selective penetration of H+ ions throughthe outer layer of the membrane. An electrochemical boundary layer with an electric potential formsat this point. An integrated Ag/AgCl reference system serves as the required reference electrode.The transmitter converts the measured voltage into the corresponding pH value using the Nernstequation.
Measuring system A complete measuring system comprises at least the following components:• pH electrode CPS11D or CPS11• Transmitter, e.g. Liquiline CM42, CM44x, Mycom S CPM153, Liquisys M CPM2x3• Memosens data cable CYK10 for Memosens sensors or CPK9 for analog sensors• Immersion, flow or retractable assembly, e.g. Cleanfit CPA871
3
4
1
2
A0025757
1 Example of a measuring system for pH measurement
1 Retractable assembly Cleanfit CPA8712 pH electrode CPS11D3 Memosens data cable CYK104 Liquiline M CM42 two-wire transmitter for hazardous area
Orbisint CPS11D and CPS11
Endress+Hauser 3
CPS11D communication and data processingCommunication with the transmitter
Always connect digital sensors with Memosens technology to a transmitter with Memosenstechnology. Data transmission to a transmitter for analog sensors is not possible.
Digital sensors can store measuring system data in the sensor. These include the following:• Manufacturer data
– Serial number– Order code– Date of manufacture
• Calibration data– Calibration date– Slope at 25 °C (77 °F)– Zero point at 25 °C (77 °F)– Temperature offset– Number of calibrations– Serial number of the transmitter used to perform the last calibration
• Operating data– Temperature application range– pH application range– Date of initial commissioning– Maximum temperature value– Hours of operation under extreme conditions– Number of sterilizations– Resistance of glass membrane
You can display the abovementioned data using the Liquiline CM44x, CM42 and Memobase PlusCYZ71D.
Dependability
Reliability Easy handlingSensors with Memosens technology have an integrated electronics unit that stores calibration dataand other information (e.g. total operating hours and operating hours under extreme measuringconditions). Once the sensor has been connected, the sensor data are transferred automatically tothe transmitter and used to calculate the current measured value. As the calibration data are storedin the sensor, the sensor can be calibrated and adjusted independently of the measuring point. Theresult:• Easy calibration in the measuring lab under optimum external conditions increases the quality of
the calibration.• Pre-calibrated sensors can be replaced quickly and easily, resulting in a dramatic increase in the
availability of the measuring point .• Maintenance intervals can be defined based on all stored sensor load and calibration data and
predictive maintenance is possible.• The sensor history can be documented using external storage media and evaluation programs, e.g.
Memobase Plus CYZ71D. Thus, the current application of the sensors can be made to depend ontheir previous history.
Integrity Data security thanks to digital data transmission
Memosens technology digitizes the measured values in the sensor and transmits the data to thetransmitter using a non-contact connection that is free from potential interference. The result:• Automatic error message if sensor fails or connection between sensor and transmitter is
interrupted• Immediate error detection increases measuring point availability
Orbisint CPS11D and CPS11
4 Endress+Hauser
Safety Maximum process safetyWith inductive transmission of the measured value using a non-contact connection, Memosensguarantees maximum process safety and offers the following benefits:• All problems caused by moisture are eliminated:
– Plug-in connection free from corrosion– Measured values cannot be distorted by moisture.– Can even be connected under water
• The transmitter is galvanically decoupled from the medium. Issues concerning "symmetrical high-impedance" or "asymmetry" or an impedance converter are a thing of the past.
• EMC safety is guaranteed by screening measures for the digital transmission of measured values.• Intrinsically safe electronics mean operation in hazardous areas is not a problem.
Input
Measured variables pH value
Temperature
Measuring range Electrode version AA (for water / wastewater), AS (for boiler feedwater):pH: 1 to 12Temperature: -15 to 80 °C (5 to 176 °F)Electrode version BA (for process):pH: 0 to 14Temperature: 0 to 135 °C (32 to 275 °F)Electrode version FA (for hydrofluoric acid process):pH: 0 to 10Temperature: 0 to 70 °C (32 to 158 °F)Electrode version BT with ion trap (for chemicals, scrubbers, pulp and paper):pH: 0 to 14Temperature: 0 to 135 °C (32 to 275 °F)
Please note the process operating conditions.
Orbisint CPS11D and CPS11
Endress+Hauser 5
Installation
Installation instructions Do not install the electrodes upside down. The angle of inclination must be at least 15° from thehorizontal. A smaller inclination angle is not permitted as it could cause an air bubble to form in theglass sphere and prevent the inner electrolyte from completely wetting the pH diaphragm.
NOTICEBefore screwing in the electrode, make sure the threaded connection of the assembly is cleanand runs smoothly.‣ Screw in the electrode finger-tight (3 Nm)! (Data apply only if installing with Endress+Hauser
assemblies.)‣ Make sure to follow the installation instructions in the Operating Instructions of the used
assembly.
15° 15°
A
B
A0024316
2 Electrode installation; installation angle at least 15° from the horizontal
A Permitted orientationB Forbidden orientation
Environment
Ambient temperature NOTICERisk of damage due to frost‣ The sensor must not be used if the temperature drops below –15 °C (5 °F).
Storage temperature 0 to 50 °C (32 to 122 °F)
Degree of protection IP 68: Memosens plug-in head, (10 m (33 ft) water column, 25 °C (77 °F), 45 days, 1 M KCl)IP 68: TOP68 plug-in head, autoclavable up to 135 °C (275 °F), (1 m (3.3 ft) water column,
50 °C (122 °F), 168 h)IP 67: GSA plug-in head (with closed connector system)
Orbisint CPS11D and CPS11
6 Endress+Hauser
Process
Process temperature Version AA, AS: -15 to 80 °C (5 to 176 °F)Version BA, BT: 0 to 135 °C (32 to 275 °F)Version FA: 0 to 70 °C (32 to 158 °F)
Process pressure (absolute) Version AA, AS, FA: 1 to 7 bar (15 to 101 psi)Version BA, BT: 1 to 17 bar (15 to 246 psi)
LCAUTIONPressurization of sensor due to prolonged use under increased process pressureRisk of injury from glass breakage‣ Avoid excessive heating of such sensors if using them under reduced process pressure or under
atmospheric pressure.‣ When handling such sensors, wear protective goggles and suitable gloves.
Pressure-temperatureratings (absolute)
A C
10
12
14
16
18 B
50 100 200 250275
3005
0
40
80
120
160
200
246
[psi] p [bar]
T [°F]400 20 60 80 100
0
4
6
8
2
T [°C]
120 135-15
7
115
17
101
D
32176
158
70
A0025761
3 Pressure-temperature ratings
A Version AA, ASB Version BA, BTC Version FAD Atmospheric pressure
Minimum conductivity Version AA, BA, BT, FA: Min. 50 μS/cm (minimum flow; pressure and temperature must bestable)
Version AS: Min. 0.1 μS/cm (stainless steel flow assembly with grounding;stable and minimum flow; pressure and temperature must bestable)
pH range Version AA, AS: 1 to 12 pHVersion BA, BT: 0 to 14 pHVersion FA: 0 to 10 pH
NOTICERisk of damage to electrode‣ Never use the electrode outside of the listed specifications!
Orbisint CPS11D and CPS11
Endress+Hauser 7
Mechanical construction
Design, dimensions ofCPS11D
3
4
5
6
1
2
7
812(0.47)
120, 225, 360
(4.7
2, 8.8
6, 14.1
7)
50 (
1.9
7)
A0025726
4 CPS11D with Memosens plug-in head, temperaturesensor
1 Memosens plug-in head, Pg 13.52 Viton O-ring with thrust collar3 Ag/AgCl reference lead - reference4 "Advanced Gel" electrolyte5 Ag/AgCl reference lead - pH6 PTFE diaphragm7 pH glass membrane8 Temperature sensor NTC30K
3
4
5
6
1
2
7
8
9
12(0.47)
120, 225, 360
(4.7
2, 8.8
6, 14.1
7)
50 (
1.9
7)
A0025725
5 CPS11D-7BTxx, with ion trap
1 Memosens plug-in head, Pg 13.52 Viton O-ring with thrust collar3 Ag/AgCl reference lead - reference4 "Advanced Gel" electrolyte5 Ion trap6 Ag/AgCl reference lead - pH7 PTFE diaphragm8 pH glass membrane9 Temperature sensor NTC30K
Orbisint CPS11D and CPS11
8 Endress+Hauser
Design, dimensions of CPS11
1
2
3
4
6
5
7
120, 225, 360, 425
(4.7
2, 8.8
6, 14.1
7, 16.7
3)
12 /0.47
A0025733
6 CPS11 with GSA plug-in head
1 GSA plug-in head, Pg 13.52 Viton O-ring with thrust collar3 Ag/AgCl reference lead - reference4 "Advanced Gel" electrolyte5 Ag/AgCl reference lead - pH6 PTFE diaphragm7 pH glass membrane
3
4
5
6
1
2
7
8
120, 225, 360, 425
(4.7
2, 8.8
6, 14.1
7, 16.7
3)
12 /0.47
A0025729
7 CPS11 with TOP68 plug-in head, temperature sensor
1 TOP68 plug-in head, Pg 13.52 Viton O-ring with thrust collar3 Ag/AgCl reference lead - reference4 "Advanced Gel" electrolyte5 Ag/AgCl reference lead - pH6 PTFE diaphragm7 pH glass membrane8 Temperature sensor Pt100
Weight 0.1 kg (0.2 lbs)
Materials Electrode shaft: Glass to suit processpH membrane glasses: Type A, B, FMetal lead: Ag/AgClDiaphragm: Ring-shaped Teflon® diaphragm, sterilizable, not cytotoxic
Process connection Pg 13.5
Temperature sensor CPS11D: NTC30KCPS11: Pt100, Pt1000
Plug-in heads CPS11D: Memosens plug-in head for digital, non-contact data transmission
CPS11:ESA: Threaded plug-in head Pg 13.5, TOP68 for electrodes with or without
temperature sensor, 17 bar abs. (246 psi) overpressure protection (threefold),Ex
GSA: Threaded plug-in head Pg 13.5 for electrodes without temperature sensor
Orbisint CPS11D and CPS11
Endress+Hauser 9
Reference system Version AA, BA, FA: Ag/AgCl reference lead with Advanced Gel 3M KCl, AgCl-freeVersion AS: Ag/AgCl reference lead with Advanced Gel, saturated KCl (> 3M KCl) with salt
rings, AgCl-freeThe following are indicative of used salt rings (fixed supply of KCl) underconstant process conditions (e.g. stable temperature and flow):• a continuous upward trend in the pH value (to alkaline pH values)• a continuous downward trend in the zero point (to acidic pH values)
following adjustment during calibrationBT version: Ag/AgCl reference lead with ion trap and Advanced Gel 3M KCl
Certificates and approvals
Ex approval for CPS11D • ATEX II 1G Ex ia IIC T3/T4/T6 Ga• FM / CSA Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters
Hazardous area versions of the digital sensors with Memosens technology are marked by a red-orange ring in the plug-in head.
Ex approval for CPS11(TOP68)
• ATEX II 1G Ex ia IIC T3/T4/T6 Ga• FM Class I Div. 2, in conjunction with Liquiline M CM42 and Mycom S CPM153 transmitters
Biocompatibility Cytotoxicity verified in accordance with:USP 2009, chapter <88> (USP Class VI) for diaphragm
TÜV certificate forMemosens plug-in head
Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
TÜV certificate for TOP68plug-in head
Pressure resistance 16 bar rel. (232 psi), minimum three times the safety pressure
CPS11D electromagneticcompatibility
Interference emission and interference immunity as per EN 61326: 2012
Ordering information
Product page www.endress.com/cps11d
www.endress.com/cps11
Product Configurator The navigation area is located on the right of the product page.1. Under "Device support" click "Configure your selected product".
The Configurator opens in a separate window.2. Select all the options to configure the device in line with your requirements.
In this way, you receive a valid and complete order code for the device.3. Export the order code as a PDF or Excel file. To do so, click the appropriate button at the top of
the screen.
Scope of delivery The scope of delivery includes:• Sensor in the version ordered• Technical Information
Orbisint CPS11D and CPS11
10 Endress+Hauser
AccessoriesThe following are the most important accessories available at the time this documentation wasissued. For accessories not listed here, please contact your service or sales office.
Assemblies Cleanfit CPA871• Flexible process retractable assembly for water, wastewater and the chemical industry• For applications with standard 12mm sensors• Product Configurator on the product page: www.endress.com/cpa871
Technical Information TI01191C
Cleanfit CPA875• Retractable process assembly for sterile and hygienic applications• For in-line measurement with standard 12 mm sensors for parameters such as pH, ORP and
oxygen• Product Configurator on the product page: www.endress.com/cpa875
Technical Information TI01168C
Cleanfit CPA472D• Robust retractable assembly for pH, ORP and other industrial sensors• Heavy-duty version made of durable materials• For manual or pneumatic, remote-controlled operation• Product Configurator on the product page: www.endress.com/cpa472d
Technical Information TI00403C
Cleanfit CPA450• Manual retractable assembly for installing 120 mm sensors in tanks and pipes• Product Configurator on the product page: www.endress.com/cpa450
Technical Information TI00183C
Cleanfit CPA471• Compact stainless steel retractable assembly for installation in tanks and pipes, for manual or
pneumatically remote-controlled operation• Product Configurator on the product page: www.endress.com/cpa471
Technical Information TI00217C
Cleanfit CPA472• Compact plastic retractable assembly for installation in tanks and pipes• For manual or pneumatic, remote-controlled operation• Product Configurator on the product page: www.endress.com/cpa472
Technical Information TI00223C
Cleanfit CPA473• Stainless steel process retractable assembly with ball valve shutoff for particularly reliable
separation of the medium from the environment• Product Configurator on the product page: www.endress.com/cpa473
Technical Information TI00344C
Cleanfit CPA474• Plastic process retractable assembly with ball valve shutoff for particularly reliable separation of
the medium from the environment• Product Configurator on the product page: www.endress.com/cpa474
Technical Information TI00345C
Unifit CPA442• Installation assembly for food, biotechnology and pharmaceutics• With EHEDG and 3A certificate• Product Configurator on the product page: www.endress.com/cpa442
Technical Information TI00306C
Orbisint CPS11D and CPS11
Endress+Hauser 11
Dipfit CPA111• Immersion and installation assembly made of plastic for open and closed vessels• Product Configurator on the product page: www.endress.com/cpa111
Technical Information TI00112C
Dipfit CPA140• pH/ORP immersion assembly with flange connection for very demanding processes• Product Configurator on the product page: www.endress.com/cpa140
Technical Information TI00178C
Flowfit CPA240• pH/ORP flow assembly for processes with stringent requirements• Product Configurator on the product page: www.endress.com/cpa240
Technical Information TI00179C
Flowfit CPA250• Flow assembly for pH/ORP measurement• Product Configurator on the product page: www.endress.com/cpa250
Technical Information TI00041C
Ecofit CPA640• Set comprising adapter for 120 mm pH/ORP electrodes and sensor cable with TOP68 coupling• Product Configurator on the product page: www.endress.com/cpa640
Technical Information TI00246C
Flexdip CYA112• Immersion assembly for water and wastewater• Modular assembly system for sensors in open basins, channels and tanks• Product Configurator on the product page: www.endress.com/cya112
Technical Information TI00432C
Buffer solutions High-quality buffer solutions from Endress+Hauser - CPY20The secondary buffer solutions have been referenced to primary reference material of the PTB(German Federal Physico-technical Institute) and to standard reference material of NIST (NationalInstitute of Standards and Technology) according to DIN 19266 by a DKD (German CalibrationService) accredited laboratory.Product Configurator on the product page: www.endress.com/cpy20
Measuring cables CYK10 Memosens data cable• For digital sensors with Memosens technology• Product Configurator on the product page: www.endress.com/cyk10
Technical Information TI00118C
CPK9• Terminated measuring cable for connecting analog sensors with TOP68 plug-in head• Selection in accordance with product structure
For more information and to order, please contact your sales office.
CPK1For pH/ORP electrodes with GSA plug-in head
Ordering information is available from your sales office or at www.endress.com.
www.addresses.endress.com