15
OPTICAL MEASUREMENT OF PHOTOSENSITIZER CONCENTRATION IN VIVO MARTIN R. AUSTWICK * , JOSEPHINE H. WOODHAMS * ,z , VADZIM CHALAU * , CHARLES A. MOSSE * , CAROLINE ELIOT * , LAURENCE LOVAT * , ALEXANDER J. MACROBERT * , IRVING J. BIGIO y and STEPHEN G. BOWN * * National Medical Laser Centre Charles Bell House, 67-73 Riding House Street University College London London W1W 7EJ, UK y Boston University, BME, 44 Cummington Street Boston, MA 02215, USA z [email protected] Most techniques for measuring tissue concentrations of drugs are invasive, time-consuming, and often require the removal of tissue or body °uids. Optical pharmacokinetics (OP) is a minimally invasive alternative giving an immediate result. Pulses of white light are directed at the tissue of interest using a ¯ber optic probe. Scattered light is detected by a second ¯ber immediately adjacent to the ¯rst in the same probe (separation 1.7 mm). Using the photosensitizer dis- ulfonated aluminium phthalocyanine (AlS 2 Pc), OP measurements were made in phantoms and on the mouth, stomach, colon, skin, and liver of normal rats 1 and 24 h after intravenous AlS 2 Pc administration. AlS 2 Pc concentration was determined by calculating the area under the curve (AUC) in the spectral region around the peak drug absorption or measuring the height of the peak. Spectral baseline interpolation removed the need for pre-drug, control optical measure- ments. OP measurements correlated well with values from alkali chemical extraction (CE) of the corresponding tissues, (R 2 0.870.97). OP measurements in the mouth also correlated with CE of less accessible internal organs (R 2 0.770.88). In phantoms, the lowest detectable concentration was 0.1 "g/g. In vivo, results were limited by the lower accuracy in the CE measurements but were almost certainly comparable. An incidental ¯nding was a 1215 nm red shifted component in the spectra observed 1 h after drug administration, suggesting partitioning of the drug in di®erent microenvironment compartments, which could prove to be of considerable interest in future studies. In conclusion, OP shows promise for real-time measurement of concentrations of drugs with suitable absorption peaks. Keywords : Optical pharmacokinetics; chemometrics; photodynamic therapy; noninvasive measurement. 1. Introduction Most current methods for quantifying drug concen- trations in tissue are invasive and usually require tis- sue extraction, making them unsuitable for real-time dosimetry. An important potential application of optical pharmacokinetics (OP) is for the dosimetry of photosensitizers for photodynamic therapy (PDT). This paper addresses the use of OP, a noninvasive Journal of Innovative Optical Health Sciences Vol. 4, No. 2 (2011) 97111 # . c World Scienti¯c Publishing Company DOI: 10.1142/S1793545811001460 97

optical measurement of photosensitizer concentration in vivo

  • Upload
    lamtram

  • View
    221

  • Download
    2

Embed Size (px)

Citation preview

Page 1: optical measurement of photosensitizer concentration in vivo

OPTICAL MEASUREMENT OFPHOTOSENSITIZER CONCENTRATION IN VIVO

MARTIN R. AUSTWICK*, JOSEPHINE H. WOODHAMS*,z,VADZIM CHALAU*, CHARLES A. MOSSE*, CAROLINE ELIOT*,

LAURENCE LOVAT*, ALEXANDER J. MACROBERT*,IRVING J. BIGIOy and STEPHEN G. BOWN*

*National Medical Laser CentreCharles Bell House, 67-73 Riding House Street

University College LondonLondon W1W 7EJ, UK

yBoston University, BME, 44 Cummington StreetBoston, MA 02215, [email protected]

Most techniques for measuring tissue concentrations of drugs are invasive, time-consuming, andoften require the removal of tissue or body °uids. Optical pharmacokinetics (OP) is a minimallyinvasive alternative giving an immediate result. Pulses of white light are directed at the tissue ofinterest using a ¯ber optic probe. Scattered light is detected by a second ¯ber immediatelyadjacent to the ¯rst in the same probe (separation 1.7 mm). Using the photosensitizer dis-ulfonated aluminium phthalocyanine (AlS2Pc), OP measurements were made in phantoms andon the mouth, stomach, colon, skin, and liver of normal rats 1 and 24 h after intravenous AlS2Pcadministration. AlS2Pc concentration was determined by calculating the area under the curve(AUC) in the spectral region around the peak drug absorption or measuring the height of thepeak. Spectral baseline interpolation removed the need for pre-drug, control optical measure-ments. OP measurements correlated well with values from alkali chemical extraction (CE) of thecorresponding tissues, (R2 0.87�0.97). OP measurements in the mouth also correlated with CE ofless accessible internal organs (R2 0.77�0.88). In phantoms, the lowest detectable concentrationwas 0.1�g/g. In vivo, results were limited by the lower accuracy in the CE measurements butwere almost certainly comparable. An incidental ¯nding was a 12�15 nm red shifted componentin the spectra observed 1 h after drug administration, suggesting partitioning of the drug indi®erent microenvironment compartments, which could prove to be of considerable interest infuture studies. In conclusion, OP shows promise for real-time measurement of concentrations ofdrugs with suitable absorption peaks.

Keywords: Optical pharmacokinetics; chemometrics; photodynamic therapy; noninvasivemeasurement.

1. Introduction

Most current methods for quantifying drug concen-trations in tissue are invasive and usually require tis-sue extraction, making them unsuitable for real-time

dosimetry. An important potential application of

optical pharmacokinetics (OP) is for the dosimetry of

photosensitizers for photodynamic therapy (PDT).

This paper addresses the use of OP, a noninvasive

Journal of Innovative Optical Health SciencesVol. 4, No. 2 (2011) 97�111#.c World Scienti¯c Publishing CompanyDOI: 10.1142/S1793545811001460

97

Page 2: optical measurement of photosensitizer concentration in vivo

white-light spectroscopic measurement collected via a¯ber-optic probe, for the quantitative monitoring ofin vivo tissue concentrations of PDT drugs thatabsorb in the visible (vis) and near infrared (NIR)region.1 Previous work focused on chemotherapyagents1,2 and topical photodynamic agents.3 In thispaper, we describe novel applications of the OPmethod in assessing the concentration of a systemi-cally administered photosensitizer and the correlationbetween chemical measurements in less-accessibleinternal organs and OP measurements in a moreaccessible organ (the mouth).

For very small source-detector separations of¯ber-optic probes (<500 microns), in the techniqueknown as elastic scattering spectroscopy (ESS),diagnostic signatures have been shown in a number ofdisease/organ systems4�6 and can be used to extractthe intrinsic optical properties of the super¯cial lay-ers.7 For larger source-detector separations (>4mm),methods based on the di®usion equation can extractthe optical properties, but these methods provideaverage values for larger tissue volumes, thus lackingsite speci¯city. Also, most systems for doing this arelimited to the red and NIR because absorption bywater and hemoglobin is too strong at otherwavelengths.8�10 For intermediate source-detectorseparations, the OP method o®ers the opportunityfor the quantitative measurement of absorption, dueto speci¯c chromophore absorption characteristics,enabling greater sensitivity thanESS, while retaininghighly localized site-speci¯city. If the ¯ber separationis in the range 1.6�2.5mm, the geometry is such thatthe e®ective photon path length is almost insensitiveto the (reduced) scattering coe±cient of the tissue,� 0s. The path length is hence independent of mor-

phological changes within the tissue, yet is longenough for sensitive detection of small concentrationsof strongly absorbing chromophores.1�11 This facili-tates the use of Beer's law for extraction of theabsorption coe±cient, independent of scatteringproperties. Within this range of separations, thedepth of sensitivity, as well as the sensitivity to smalldrug concentrations, increases for larger separations.The optimum separation may be chosen to best suitthe thickness of the tissue being sampled. This scat-tering-independent path length means that the con-centration can be extracted in a straightforwardfashion. In tissue, the path length is approximately5 mm and the sampling volume is presumed to be

close to 1mm3 for this ¯ber geometry andwavelengthrange (650�700 nm). In this study, a ¯ber separationof 1.7mm was chosen. Other ¯ber separations werenot studied.

PDT is a chemotherapeutic technique usedto treat a range of cancerous and pre-cancerousconditions.12 The patient is administered a photo-sensitizer (PS) either topically or systemically,and the targeted area is illuminated, usually withlaser light. The PS is excited to a chemically activestate, which combines with oxygen in the tissue tocreate cytotoxic species (e.g., singlet oxygen, super-oxide), causing local cellular damage whilst typicallyleaving connective tissue intact. E®ective therapydepends on the concentration of photosensitizer, ofoxygen, and photoactivating light at every point inthe target tissue. A real-time measurement (mini-mally invasive or noninvasive) of PS concentration inthe target region would thus enable more predictablecytotoxicity, assuming available oxygen and light.PS biodistribution, even within a healthy organ,can be markedly heterogeneous, and site-speci¯cmeasurements can provide valuable information.

The OP technique could be particularly valuablewhere the PDT is to be carried out interstitiallyunder image guidance13�15 on an inaccessibleorgan16; if OP were extended to correlate measure-ments in the mouth (or other accessible organ) withPDT e®ects in a less-accessible organ, the techniquecould improve the e±cacy of therapy even further.17

This paper addresses these possibilities with studiesusing the photosensitizer aluminium disulfonatedphthalocyanine (AlS2Pc), chosen for its stability,solubility, ease of administration, and low toxicity inthe absence of light. AlS2Pc displays a strong, well-de¯ned absorption band around 670�675 nm, with aline width of about 10 nm.18 Its extinction coe±cientat this wavelength reaches 2� 105 M�1cm�1, com-pared to 3� 103 M�1cm�1 for por¯mer sodium(Photofrin) and 3� 104 M�1cm�1 for mTHPC(meso-tetra(hydroxyphenyl)chlorine. Foscan), otherPS drugs in clinical use. Even adjusting for thetypical molar concentrations used clinically, AlS2Pcgives a much stronger optical signal, and our grouphas extensive pre-clinical experience with thisPS.19,20 Work with other phthalocyanines has begunrecently in the treatment of cutaneous t-cell lym-phoma,21 and many aspects of our study can begeneralized to a variety of agents.

98 M. R. Austwick et al.

Page 3: optical measurement of photosensitizer concentration in vivo

2. Methods and Materials

2.1. Photosensitizer

AlS2Pc powder (Frontier Scienti¯c Ltd., UK) wasused to prepare the required concentration forintravenous injection by dissolving 1mg into 50�Lof 0.1M sodium hydroxide (Sigma-Aldrich, UK)and then adding phosphate bu®ered saline (Sigma-Aldrich, UK) to make the required volume.

2.2. Phantom measurements

To test the limits of sensitivity of our equipmentand examine any path length e®ects, we preparedoptical tissue phantoms using 37mL PhosphateBu®ered Saline (PBS), 2mL of 20% Intralipidr,and varying quantities (0.01, 0.025, 0.05, 0.1, 0.25,0.5, 1, 2.5, 5, 10, 25, 50, and 75�g/mL) of AlS2Pc.The peak height of the negative-log-ratio to a linearbaseline interpolation (described below) was exam-ined. Previous work1 has estimated the path length-dependence on absorption coe±cient �a in thisgeometry, using the expression for the path length L

Lð�aÞ ¼ x0 þ x1 expð��ax2Þ; ð1Þand so the phantom data absorbance A can be ¯ttedusing the function

Að�aÞ ¼ C þ �a½x0 þ x1 expð��ax2Þ�; ð2Þwhere C is a constant that accounts for inadequaciesin the baseline ¯t, which lead to nonzero absorbancemeasures in the absence of added absorbers. Pre-viously,1 parameters x0; x1, and x2 were measured as0.6, 1.3, and 3.57 cm, respectively. Another way tocapture this deviation from linearity is to use a power-law ¯t, where Að�aÞ ¼ mð�aÞn, as utilized in Ref. 2.

2.3. Animal model

Wistar rats (180�220 g, Harlan UK Ltd., Black-thorn, Oxfordshire, UK) were housed under diurnallighting conditions and allowed food and tap waterad libitum. Rats were anesthetized with 2% halo-thane (Merial Animal Health Ltd., Harlow, UK) inoxygen for induction and maintenance throughoutsurgery, with the animals breathing spontaneously.Analgesia was administered subcutaneously 1 h priorto surgery (Vetergesicr, buprenorphine hydrochlo-ride, Reckitt Benckiser Healthcare Ltd., Hull, UK).

All procedures were carried out under the authorityof project and personal licences granted by the UKHome O±ce and the UKCCCR Guidelines (2010).

2.4. OP measurements on tissues

Rats were sensitized with dosages of 0 (controlanimals), 0.1, 0.2, 0.25, 0.5, 1.0, or 2.5mg/kg ofAlS2Pc as a bolus injection into the tail vein. At 1 or24 h later (immediately for control animals), OPmeasurements were performed on shaved abdomi-nal skin, thigh muscle with skin removed, oralmucosa and, after laparotomy, on liver, glandularstomach, and colon. Three rats were used for eachdrug dose and each time interval; the intervals werechosen to see whether di®erent biodistributionsa®ected the quality of the results. At 1 h, the drugwill still have a signi¯cant proportion within thevasculature, whereas by 24 h it will reside mainlywithin the organ parenchyma.22 The OP ¯ber-opticprobe (described below) was positioned by means ofa micromanipulator, so that it was just touching thesurface of the organ under examination. On eachorgan, 10 separate positions were chosen, with 4 OPmeasurements being taken at each position (i.e., 40spectra per organ). When only a small thickness oftissue was being investigated, a piece of matt blackcard was placed under the tissue to minimize thee®ects of light that passed through the tissue andout the far side, and might be scattered back fromother adjacent tissues.

Immediately after the OP measurements werecompleted, the animals were killed, and the OPmeasurements were repeated post mortem in thesame organs. Tissue samples were collected fromeach animal and stored at �20�C for alkalineextraction of AlS2Pc.

20

2.5. Alkaline extraction of AlS2Pc

from tissue

Finely-cut thawed tissue samples, including controlsof each organ from unsensitized animals, wereadded to 0.1M sodium hydroxide (NaOH, Sigma-Aldrich, UK) at a ratio of 0.1 g into 10mL NaOH.From each animal, four samples of each organ wereprocessed. The tissues in NaOH were incubated in awater bath at 50�C for 6 h in darkness to allow cellbreakdown. The samples were shaken approximately

In vivo Optical Pharmacokinetics 99

Page 4: optical measurement of photosensitizer concentration in vivo

halfway through the incubation period to ensureadequate breakdown of cell structures to release theAlS2Pc. AlS2Pc concentrations in the resulting sol-ution were measured by spectro°uorimetry (Perkin-Elmer LS50 Luminescence Spectrophotometer). Thesolutions were loaded into 96-well plates, and °uor-escence was measured using a Perkin-Elmer LS50luminescence spectrometer at excitation and emis-sion wavelengths of 605 and 675 nm, respectively. Aslit width of 10 nm and an integration time of 1 wereset for both excitation and emission. The emissionwas long-pass ¯ltered to eliminate scattered lightat wavelengths shorter than 645 nm (RG645 Schottlong-pass ¯lter). The concentration ofAlS2Pc in eachspecimen could then be calculated against a standardcurve. A standard curve was prepared each time achemical extraction (CE) was undertaken using0.1M NaOH solutions with known concentrations ofAlS2Pc. Calibration curves were plotted for threeranges of concentration: 0 to 0.5, 0 to 0.1 and 0 to0.01�g/mL; for the liver, in which higher concen-trations of drug accrued, a larger-range calibrationcurve was constructed, covering 0 to 100�g/mL.A molecular weight of 735.08 g/mol was used forcalculations of AlS2Pc concentrations.

2.6. OP system

The OP collection apparatus consists of a lightsource, delivery and collection ¯bers, a spectrometer,and a laptop computer, see Fig. 1. OP data werecollected from tissues using a pair of ¯bers inside acatheter. The tip of the catheter was positioned tobe just touching the tissues to be interrogated and,at the tip, the ¯ber centers were 1.7mm apart, which

was chosen on the basis of previous publications asdiscussed in Sec. 1. The light from a pulsed Xenonarc lamp (model LS1130/FX1100, Perkin ElmerOptoelectronics) was delivered using a 400-�m core,vis/NIR ¯ber and the light scattered back was col-lected with a 200-�m core, vis/NIR ¯ber connectedto a spectrometer (model S2000, Ocean Optics,Inc.). Before each experiment the system was cali-brated using a reference spectrum collected fromwhite Spectralonr (Labsphere, NH, USA), a spec-trally °at di®use-re°ecting polymer. For eachmeasurement, ambient light was collected immedi-ately before the (lamp-illuminated) measurementand subtracted from the signal. The tissue spectrum[Irawð�Þ] and Spectralonr reference spectrum [Irefð�Þ]were smoothed using a seven-point (�3 nm) moving-window average, and the tissue spectrumwas dividedby the reference spectrum to provide a spectrum thatcould be calibrated for system response.

Ið�Þ ¼ Irawð�Þ=Irefð�Þ: ð3Þ

2.6.1. OP spectral analysis

If Að�Þ is the signal attenuation, the Beer�Lambertlaw states

Irawð�Þ ¼ I0ð�Þ10�Að�Þ; ð4Þwhere I0ð�Þ represents the spectral output of thelight source modi¯ed by the system response(detector, transfer optics, etc.), and can then begiven (within a scaling factor) by I0ð�Þ ¼ Irefð�Þ.Inserting this into Eq. (4) gives

Ið�Þ ¼ 10�Að�Þ: ð5ÞWe can further split Að�Þ into photosensitizer andtissue components

Að�Þ ¼ APSð�Þ þAtissueð�Þ: ð6ÞThe tissue attenuation Atissue will include tissueabsorption and scattering, in particular the e®ectsof (oxyhemoglobin (HbO) and deoxyhemoglobin(Hb)). From Eqs. (5) and (6)

APSð�Þ ¼ Að�Þ �Atissueð�Þ¼ �ðlog10½Ið�Þ� � log10½Itissueð�Þ�Þ ð7Þ

hLicPS�½PS� � � log10

Ið�ÞItissueð�Þ

� �; ð8Þ

Tissue

Afferent fiber(400 µm)

Efferent fiber(200 µm)

Sampling depth

1.7 mm fiberseparation

DetectorXe light source

Spectral analysis

Fig. 1. Schematic diagram of the Optical Pharmacokineticsystem.

100 M. R. Austwick et al.

Page 5: optical measurement of photosensitizer concentration in vivo

where cPS is the photosensitizer concentration, hLiis the e®ective mean photon path length, and�½PS�ð�Þ is the extinction coe±cient of the photo-sensitizer. Our OP measure will be based on thequantity APS (see below).

If it were necessary to take baseline opticalmeasurements in di±cult-to-reach sites prior todrug administration, this would limit the clinicaluse of the technique in some clinical applications,and even when such a measurement is possible, timevariation of the tissue properties and errors inco-registration mean that the tissue component willnot be completely removed from the spectrum bythe process. It is therefore desirable to use methodsto estimate the tissue contribution from spectrataken from sensitized animals, without the need fora reference spectrum from an unsensitized animal.

2.6.2. Optical pharmacokinetic metrics

In this study, the OP metrics we considered tomeasure the drug absorption were AUC*, AUC,y andpeak-¯tting metrics (A1 and A2). Area-under-curve(AUC) methods (AUC* and AUCy) calculate thearea under a peak over a speci¯ed wavelength range,which inherently incorporates some noise-averaging.Each AUC technique used a di®erent method forbaseline removal, detailed below. Finally, a peak-¯tting method was tried, which uses aqueous AlS2Pcspectra to ¯t to animal data; this should accountfor any spectrally shifted (but not broadened orotherwise modi¯ed) components in the in vivo drugspectrum, but is more analytically complex. Theseapproaches are detailed below.

The absorption feature of interest in the AlS2Pcspectrum is the strong peak at 672 nm.23 The tissueabsorption at these wavelengths can be estimatedby linear interpolation of the spectrum between twopoints (650�700 nm), shown in Fig. 2. We call thisfunction Ilinearð�Þ. In this range we expect thebaseline to be smooth but not linear, so we wouldexpect a ¯nite (positive or negative) area, even forzero AlS2Pc. We de¯ne

A�ð�Þ ¼ � log10

Ið�ÞIlinearð�Þ

� �: ð9Þ

We use an area under curve measure, AUC*,de¯ned as:

AUC� ¼def

Z 700 nm

650 nm

d�A�ð�Þ : ð10Þ

These limits lie within the full width of the drugpeak, chosen in order to capture the local baselineand any possible wavelength shifts and to sample a¯xed proportion of the drug absorption peak.Choosing wider limits would mean modulations inthe baseline nonlocal to the central drug peak,which could distort the interpolation of the localbaseline. Outlier points [points over three standarddeviations (SDs) from the animal mean] wereiteratively removed.

2.6.3. Control animals and hemoglobinspectrum removal

Ratioing spectral data against a characteristicspectrum of the same organ from unsensitized ani-mals was carried out to remove a large proportion ofthe baseline features common to all animals. Anaverage was formed over ¯ve control animals (10sites per organ, four measurements per site) foreach organ to produce the characteristic spectra.Spectra from AlS2Pc sensitized animals could thenbe divided by this spectrum, and the negative log-arithm taken to give the control-ratio-log (CRL)attenuation

ACRLð�Þ ¼ � log10

Ið�ÞIcontrolð�Þ

� �: ð11Þ

The features present in ACRLð�Þ will be due todi®erences between the control animal and thesensitized animal. This will include changes in the

Fig. 2. Linear interpolation of OP data (taken from the liver).Inset shows negative log10 ratio of data and linear interpolation.

In vivo Optical Pharmacokinetics 101

Page 6: optical measurement of photosensitizer concentration in vivo

hemoglobin saturation and perfusion, the drug sig-nal, and to a less signi¯cant extent, in absorptionand scattering due to other e®ects. Oxyhemoglobin(HbO) and deoxyhemoglobin (Hb), the largestcontributors to absorption changes, were ¯tted toACRLð�Þ using spectral Hb/HbO extinction coe±-cients ("HbO=HbÞ, with �CHbO and �CHb as ¯ttingparameters, representing changes in concentrationof HbO and Hb (and a constant o®set � to com-pensate for any changes in intensity of the rawspectra). The hemoglobin extinction coe±cients,both HBO and Hb were taken from the OregonMedical Laser Centre website.24

Afitð�Þ ¼ �CHbO"HbOð�Þ þ�CHb"Hbð�Þ þ �: ð12Þ

This ¯t was carried out in Matlab using theirstandard curve-¯tting toolbox which utilizes aLevenberg�Marquardt approach. The hemoglobin-¯t region (520�600 nm and 750�900 nm) was setoutside the AlS2Pc absorption peak. The resulting¯t was subtracted from the data to leave a residualspectrum. This was interpolated from 625 to725 nm, and the line was subtracted to remove anylinear baseline remaining (from scatter and anyother chromophores). The remaining spectrum,

Ayð�Þ, was integrated to give

AUCy ¼defZ 700 nm

650 nm

d�Ayð�Þ: ð13Þ

2.6.4. Peak position and line shape

There were variations in both the line shape and inthe peak absorption wavelength of the measureddrug. Suspecting that this might be due to a chemicalshift, we attempted to ¯t the observed absorption,A�with the function

A�fitð�Þ ¼A1Pthð�� �1Þ þA2Pthð�� �2Þ

þ �0 þ �1�þ �2�2 ; ð14Þ

wherePthð�Þ is the aqueousAlS2Pc spectrum and �n

is a factor shifting the central wavelength of theAlS2Pc peak. The �n terms represent a local baseline,due to tissue absorption and scattering e®ects (anattempt was made to ¯t this baseline with hemo-globin extinction coe±cients, but did not provide anaccurate or as general a ¯t). A1 and A2 are the con-tribution from each peak. Figure 3 shows data takenfrom the colon and ¯tted using this method. One

(a) (b)

Fig. 3. Sample colon data showing shifted Phthalocyanine signals at: (a) 1 h, and (b) 24 h.

102 M. R. Austwick et al.

Page 7: optical measurement of photosensitizer concentration in vivo

peak is typically blue-shifted by 0�5 nm, and theother red-shifted by 12�15 nm. Similar results havebeen reported elsewhere25 and this will be covered inmore detail in the discussion.

Because the line width of each peak did not appearto change, the quantitiesA1 andA2 can be combinedto give a measure of total drug concentration, A12:

A12 ¼ A1 þA2: ð15ÞNote that the e®ects of wavelength-dependent linewidth changes are not explicitly incorporated,although the baseline-¯t might be expected to com-pensate for this to some degree.

3. Results

3.1. Phantom study

The results are shown in Fig. 4, where the absorbancesignal of AlS2Pc is signi¯cant and well-separatedabove 0.1�g/mL, with an s=n ratio of >10. We wereunable to reproduce the path length parametersproduced in Ref. 1, and these parameters exhibitedsensitivity to the concentration range considered. Inpart, this is due to aggregation e®ects of AlS2Pc, sowe constrained our concentration range to typicalphysiological values for the majority of organs.Con¯ning ourselves to the 0�5mg/kg range yieldsx0 ¼ 0:38 cm, x1 ¼ 0:38 cm, and x2 ¼ 2:47 cm. Inthis spectral region, hemoglobin contributes anabsorption baseline of �0:5 cm�1, giving a baselinepath length of �0.5 cm. 5mg/kg of AlS2Pc results ina total �a of approximately 2 cm�1 and a path lengthreduction to �0.4 cm, a 20% decrease in path lengthfrom baseline; which would result in a commensurate20% absorbance reduction. Fitting this relationshipwith a power law over the same region, followingRef. 2, gave a power of 0.75.

3.2. Tissue measurements

Figure 5 shows the correlation between theOPAUC*andCEmeasurements for colon, skin, muscle,mouth,and stomach. At moderate AlS2Pc concentrations,there are strong correlations between CE and OP inmost organs at 1 and 24h (Table 1). The correlationis well-described by a linear regression, passing closeto the origin — this suggests little zero-point error isintroduced by using a linear baseline interpolation.Fitting this relationship using power-law functions

proved problematic. Data from the muscle, skin andstomach gave power coe±cients greater than or equalto 1, i.e., the ¯t converged to a linear function, or gavenonphysical results (see discussion). The mouth andcolon showed slightly improved ¯ts when a power lawwas applied; for 1 h, 24 h and all data categoriestogether, the colon exhibited powers of 0.8, 1.2 and0.6, respectively; and the mouth 0.6, 1.3 and 0.7;clearly the 24 h data is not giving a physical ¯t to thedata.

Figure 6 shows both measures, A1 þA2, from thepeak-¯tting function, expected to cover all drugenvironments. The colon and the mouth alone dis-play a power-law-like behavior, and so power-law¯ts are included. The powers for the colon (for 1 h,24 h, and all times together) are 0.7, 1.06, and 0.6.For the mouth, the power coe±cients are 0.6, 1.0,and 0.65. For the other organs, these coe±cientsapproached or exceeded unity.

Table 1 summarizes the data correlating OPmeasurements with CE using AUC*, AUCy, andA1 þA2, and the correlation of AUC* in the mouthwith CE in di®erent organs within the same animal.1 and 24 h data are shown separately (as AlS2Pcmay have di®erent biodistributions at these times)and together (all data points, not grouped by time).The results using AUC* and AUCy were almostidentical. Similarly, the OP results from post-mortemversus (alive) in vivo tissues showed close similarity(data not shown), with R2 values >0.9 being typicalfor correlations between the two.

One of the aims of this project was to makemeasurements in an easily accessible organ like themouth to quantify drug levels in less accessible organslike the stomach. This is illustrated in Fig. 5(f), whichcorrelates CE in the stomach with OPAlS2Pc signalsin the mouth. Table 1 summarizes these correlationsbetween measurements in the mouth with those inthe range of organs studied. The liver data showsrelatively weak correlation, but the other organs haveR2 > 0:75 at 1 h. At 24 h, the correlation is low in themuscle (possibly due to low drug levels in that organ),but with R2 > 0:8 for all others.

In the liver, much higher AlS2Pc concentrationsare seen compared with the other organs. The lowOP values documented 1 h after AlS2Pc adminis-tration are probably due to the exceptionally highdrug concentration at this time causing shadowingand dimerization (see discussion). Figure 7 showshow poorly the extinction coe±cient method cor-relates for the liver, due to these e®ects.

In vivo Optical Pharmacokinetics 103

Page 8: optical measurement of photosensitizer concentration in vivo

(a)

(b)

Fig. 4. OP measures taken on Phosphate Bu®ered Saline/Intralipid phantoms, plotted against the known absorption at the drugpeak. The phantoms were composed of 38mL PBS and 2mL of 20% Intralipidr, plus varying amounts of phthalocyanine; (a) showsthe full concentration range, and (b) shows the typical physiological range with an exponential path length function ¯tted to thedata (see text).

104 M. R. Austwick et al.

Page 9: optical measurement of photosensitizer concentration in vivo

(a) Colon (b) Mouth

(c) Muscle (d) Skin

(e) Stomach (f) Stomach ce vs AUC* in mouth

Fig. 5. AUC* data versus Chemical Extraction for all organs except liver, which is shown on di®erent scale in Figs. 7. (a)�(e)correlate AUC* with Chemical Extraction in individual organs, (f) correlates CE in the stomach with AUC* in the mouth.

In vivo Optical Pharmacokinetics 105

Page 10: optical measurement of photosensitizer concentration in vivo

Tab

le1.

R2values

forOP-C

Ecorrelation,includingcorrelationof

CE

ineach

orga

nwithAUC*measuredin

themou

th.R

2values

for¯ts

undertaken

witha

pow

er-law

mod

elareshow

nin

parentheses

—thepow

erparam

eter

isstated

inthetext.

Colon

Liver

Mou

thMuscle

Skin

Stomach

1h

24h

all

1h

24h

all

1h

24h

all

1h

24h

all

1h

24h

all

1h

24h

all

AUC*R

20.93

0.95

0.84

0.57

0.97

0.82

0.77

0.96

0.79

0.87

0.49

0.83

0.94

0.92

0.91

0.94

0.92

0.95

(0.94)

(0.96)

(0.88)

(0.82)

(0.96)

(0.83)

AUC

yR

20.91

0.95

0.87

0.56

0.95

0.82

0.81

0.95

0.84

0.89

0.58

0.86

0.94

0.91

0.94

0.95

0.87

0.94

A1þA

2R

20.92

0.94

0.87

——

—0.80

0.91

0.82

0.85

0.14

0.79

0.91

0.76

0.88

0.91

0.60

0.90

(0.93)

(0.94)

(0.91)

(0.84)

(0.93)

(0.82)

AUC*R

2(vsOP

inmou

th)

0.77

0.94

0.79

0.36

0.82

0.27

——

—0.86

0.45

0.82

0.79

0.95

0.81

0.81

0.88

0.82

106 M. R. Austwick et al.

Page 11: optical measurement of photosensitizer concentration in vivo

(a) Colon (b) Mouth

(c) Muscle (d) Skin

(e) Stomach

Fig. 6. A1 þA2 versus Chemical Extraction for all organs except liver. Note that each graph is on a di®erent scale for readability.Colon and mouth have power-law regression plotted.

In vivo Optical Pharmacokinetics 107

Page 12: optical measurement of photosensitizer concentration in vivo

4. Discussion

4.1. Chemical extraction — OP

correlation

One of our aims was to determine whether OpticalPharmacokinetics could be used as a minimallyinvasive technique to measure the concentration ofphotodynamically active drugs, photosensitizers, invivo in real time, with a particular emphasis onestimating the level of photosensitizers in targettissues at the time of light delivery for PDT. TheOP technique has proven successful in phantommeasurements, where OP can comfortably detectdrug at 0.1�g/g (100 ng/g), but the in vivo situationproved to be more complicated. The uncertaintyinherent in the Chemical Extraction measurementsis at least 0.1�g/g, giving a lower limit of quanti-tation (LLQ) of at least 0.3�g/g; the uncertainty ineach AUC* measurement is of order 0.01�g/g, butthe statistical scatter about the line of best ¯t issomewhat larger.

At larger concentrations, as measured by CE, thevariance can be extremely large (residuals> 1 �g/g),and even at the lowest values, CE gives a variance oftypically 0.1�0.3�g/g. However, because this issimilar to the absolute value, it is di±cult to place aLLQ on the in vivo OP measurements based on theCE data. It appears the variance within individualOP sites is a factor of 10 lower than the CE measure,and the scatter about the line of best ¯t is similar inmagnitude to the error in the CE measure, implying

substantial variation between individual measure-ment sites within organs. To fully test the LLQ sen-sitivity of OP, it will be necessary to improve thesensitivity of CE by at least a factor of 10. This couldbe achieved following recent improvements in thetechnique, such as using the SolvableTM (PerkinElmer) Chemical Extraction medium, which typi-cally requires ¯ve times less solvent (increasing signal¯vefold)26 and more sensitive °uorimeters. Alter-natively, either taking measurements using cuvettesor photon-counting detectors would signi¯cantlyimprove the sensitivity of the assay, but wouldincrease the collection time and cost, respectively.The method of HPLC utilized by other researchers(for example, Ref. 2) is extremely sensitive (drugconcentration as low as 0.03�M) and would provideanother alternative. Unfortunately, we did not haveaccess to such equipment.

Phantom measurements showed a nonlinear OP/concentration curve, indicative of the absorptiondependence of the path length. While previousstudies1,2 show a similar nonlinear behavior, theparameters we extracted from our measurementswere somewhat di®erent to those previously repor-ted; for example Mourant et al., 19991 used a muchsmaller range of drug concentrations than ourselves;we were able to observe a strong concentration-dependence of the path length-¯tting parameters.

For the majority of the in vivo data, power-law¯ts provide no ¯tting advantage, and in many casesutilize a power coe±cient �1, or even worse, >1;this case predicts an increasing path length with

(a) AUC* (b) A12

Fig. 7. (a) AUC* versus Chemical Extraction for liver. (b) A12 versus Chemical Extraction for liver. Poor correlation in the curve-¯tting method may be due to dimerization.

108 M. R. Austwick et al.

Page 13: optical measurement of photosensitizer concentration in vivo

increasing absorption, and can be interpreted as a¯tting artefact due to large statistical scatter in thedataset. This can also be ascribed to inter- andintra-animal variations. Based on our phantommeasurements, a tissue absorption baseline intowhich 5�g/g of AlS2Pc is introduced will result in a20% reduction in signal; the other sources of var-iance and uncertainty at present dwarf this change,explaining why the nonlinearity of the ¯t is not well-pronounced in much of the data.

At high concentrations, one must considerdimerization, where two AlS2Pc molecules areweakly bound together; the resulting organic com-plex has a broader, weaker extinction coe±cient(reported as a feature centered around 670 nm witha full width half maximum in excess of 100 nm18,23).The CE process ensures that any AlS2Pc is fullymonomerized at the time of the °uorescence mea-surement, giving a measure of the total AlS2Pc, butin vivo dimerization will mean that the OP AUC*integral will not be an accurate re°ection of the totaldrug concentration. The dimer form is signi¯cantlyless active in photodynamic therapy,18 so it may bebene¯cial that OP preferentially detects the mainphotoactive component. However the situation iscomplex as there is also evidence that the PDTprocess can cause remonomerization, so one cannotcompletely ignore the nonmonomeric forms.27

Analysis of the line shape of the AlS2Pc absorp-tion peak in vivo showed it to be composed of twospectral components, one blue-shifted by 0�5 nmand one red-shifted by 12�15 nm from the spectrallocations of the single peak of AlS2Pc seen in aqu-eous solution. The 0�5 nm blue-shifted componentis very similar to the AlS2Pc spectrum we see in acuvette for an aqueous solution (where small shiftscan be induced by di®erent saline and pH environ-ments). The component that is red-shifted is con-sistent with a similar shift seen by previous workersin murine blood.25,28 This points to the drug exist-ing in two distinct chemical environments, andpassing from the red-shifted to the slightly blue-shifted compartment between the 1 and 24 hmeasurements (the red-shifted peak is much moredominant at the 1 h time point in all organs exceptfor the liver). However, there are a number of fac-tors pointing to this e®ect not being solely due tolocalization of AlS2Pc in blood: a phantom thatutilized whole blood exhibited no red-shifted com-ponent (not shown); and previous workers saw onlysmall red shifts when adding an excess of human

serum albumin to phthalocyanine in solution.29 Inwork carried out on murine ascites, a concentration-dependent red shift was seen, which was attributedto binding to nonspeci¯c biological substrates.25 Itwas noted that intracellular AlS2Pc does not show aspectral shift of this nature, although encapsulationin lysosomes signi¯cantly increases the local con-centration, creating dimerized species.30 This com-partmentalization measurement o®ers a promisingfuture avenue for research, especially in the ¯eld ofPhotochemical Internalization.27

5. Conclusions

In conclusion, OP shows promise for measuringtissue concentration of photosensitizers in real timeand so for better real-time prediction of the extentof necrosis in photodynamic therapy.

Acknowledgments

We thank Dr Stephen Chad Kanick and Dr RobertParker of the University of Pittsburgh for discus-sions regarding analytical approaches. The USDepartment of Health and Human ServicesNational Institutes of Health (NIH) are gratefullyacknowledged for their funding of this project(project No. 5 U54 CA 104677-05), which was partof the NTROI programme (Network for Transla-tional Research on Optical Imaging). This work wasundertaken at UCLH/UCL who received a pro-portion of funding from the Department of Health'sNIHR Biomedical Research Centres fundingscheme. The views expressed in this publicationare those of the authors and not necessarily thoseof the Department of Health. This work wassupported by the Experimental Cancer MedicineCentre, University College London.

References

1. J. R. Mourant, T. M. Johnson, G. Los, I. J. Bigio,\Non-invasive measurement of chemotherapy drugconcentrations in tissue: Preliminary demon-strations of in vivomeasurements," Phys. Med. Biol.44, 1397�1417 (1999).

2. S. C. Kanick, J. L. Eiseman, R. S. Parker,\Pharmacokinetic modeling of motexa¯n gadoli-nium disposition in mouse tissues using opticalpharmacokinetic system measurements," Photo-diagnosis. Photodyn. Ther. 5, 276�284 (2008).

In vivo Optical Pharmacokinetics 109

Page 14: optical measurement of photosensitizer concentration in vivo

3. A. Johansson, J. Svensson, N. Bendsoe, K. Svanberg,E. Alexandratou, M. Kyriazi, D. Yova, S. Grafe,T. Trebst, S. Andersson-Engels, \Fluorescence andabsorption assessment of a lipid mTHPC formu-lation following topical application in a non-melanotic skin tumor model," J. Biomed. Opt. 12,034026 (2007).

4. K. S. Johnson, D. W. Chicken, D. C. Pickard, A. C.Lee, G. Briggs, M. Falzon, I. J. Bigio, M. R.Keshtgar, S. G. Bown, \Elastic scattering spec-troscopy for intraoperative determination of sentinellymph node status in the breast," J. Biomed. Opt. 9,1122�1128 (2004).

5. L. B. Lovat, K. Johnson, G. D. Mackenzie, B. R.Clark, M. R. Novelli, S. Davies, M. O'Donovan,C. Selvasekar, S.M.Thorpe,D. Pickard, R. Fitzgerald,T. Fearn, I. J. Bigio, S. G. Bown, \Elastic scatteringspectroscopy accurately detects high grade dysplasiaand cancer in Barrett's oesophagus," Gut 55,1078�1083 (2006).

6. A. Sharwani, W. Jerjes, V. Salih, B. Swinson, I. J.Bigio, M. El-Maaytah, C. Hopper, \Assessment oforal premalignancy using elastic scattering spectro-scopy," Oral Oncol. 42, 343�349 (2006).

7. R. Reif, O. A'amar, I. J. Bigio, \Analytical model oflight re°ectance for extraction of the optical prop-erties in small volumes of turbid media," Appl. Opt.46, 7317�7328 (2007).

8. R. A. Weersink, J. E. Hayward, K. R. Diamond, M.S. Patterson, \Accuracy of noninvasive in vivomeasurements of photosensitizer uptake based on adi®usion model of re°ectance spectroscopy," Pho-tochem. Photobiol. 66, 326�335 (1997).

9. A. Kienle, M. S. Patterson, \Improved solutions ofthe steady-state and the time-resolved di®usionequations for re°ectance from a semi-in¯nite turbidmedium," J. Opt. Soc. Am. A Opt. Image Sci. Vis.14, 246�254 (1997).

10. R. M. Doornbos, R. Lang, M. C. Aalders, F. W.Cross, H. J. Sterenborg, \The determination ofin vivo human tissue optical properties and absolutechromophore concentrations using spatially resolvedsteady-state di®use re°ectance spectroscopy," Phys.Med. Biol. 44, 967�981 (1999).

11. J. R. Mourant, I. J. Bigio, D. A. Jack, T. M. John-son, H. D. Miller, \Measuring absorption coe±cientsin small volumes of highly scattering media: Source-detector separations for which path lengths do notdepend on scattering properties," Appl. Opt. 36,5655�5661 (1997).

12. T. J. Dougherty, C. J. Gomer, B. W. Henderson,G. Jori, D. Kessel, M. Korbelik, J. Moan, Q. Peng,\Photodynamic therapy," J. Natl. Cancer Inst. 90,889�905 (1998).

13. H. U. Ahmed, D. Pendse, R. Illing, C. Allen, J. H.van der Meulen, M. Emberton, \Will focal therapybecome a standard of care for men with localizedprostate cancer?," Nat. Clin. Pract. Oncol. 4,632�642 (2007).

14. L. Ayaru, J. Wittmann, A. J. Macrobert, M. Novelli,S. G. Bown, S. P. Pereira, \Photodynamic therapyusing vertepor¯n photosensitization in the pancreasand surrounding tissues in the Syrian goldenhamster," Pancreatology 7, 20�27 (2007).

15. H. R. Jager, M. N. Taylor, T. Theodossy, C. Hopper,\MR imaging-guided interstitial photodynamic lasertherapy for advanced head and neck tumors," AJNRAm. J. Neuroradiol. 26, 1193�1200 (2005).

16. S. P. Pereira, L. Ayaru, A. Rogowska, A. Mosse,A. R. Hat¯eld, S. G. Bown, \Photodynamic therapyof malignant biliary structures using meso-tetrahy-droxyphenylchlorin,"Eur. J.Gastroenterol. Hepatol.19, 479�485 (2007).

17. M. Zellweger, P. Grosjean, P. Monnier, H. van denBergh, G. Wagnieres, \Stability of the °uorescencemeasurement of Foscan in the normal human oralcavity as an indicator of its content in early cancersof the esophagus and the bronchi," Photochem.Photobiol. 69, 605�610 (1999).

18. D. Phillips, D. Dhami, R. Ostler, Z. Petrasek, \Thedimerisation of phthalocyanines," Progress inReaction Kinetics and Mechanism 28, 299�420(2003).

19. P. T. Chatlani, P. J. Nuutinen, N. Toda, H. Barr,A. J. Macrobert, J. Bedwell, S. G. Bown, \Selectivenecrosis in hamster pancreatic tumours usingphotodynamic therapy with phthalocyanine photo-sensitisation," Br. J. Surg. 79, 786�790 (1992).

20. J. H. Woodhams, L. Kunz, S. G. Bown, A. J.Macrobert, \Correlation of real-time haemoglobinoxygen saturation monitoring during photodynamictherapy with microvascular e®ects and tissuenecrosis in normal rat liver," Br. J. Cancer 91,788�794 (2004).

21. T. K. Lee, E. D. Baron, T. H. Foster, \MonitoringPc 4 photodynamic therapy in clinical trials ofcutaneous T-cell lymphoma using noninvasivespectroscopy," J. Biomed. Opt. 13, 030507 (2008).

22. C. J. Tralau, H. Barr, D. R. Sandeman, T. Barton,M. R. Lewin, S. G. Bown, \Aluminum sulfonatedphthalocyanine distribution in rodent tumors of thecolon, brain and pancreas," Photochem. Photobiol.46, 777�781 (1987).

23. H. Barr, C. J. Tralau, A. J. Macrobert, N. Krasner,P. B. Boulos, C. G. Clark, S. G. Bown, \Photo-dynamic therapy in the normal rat colon withphthalocyanine sensitisation," Br. J. Cancer 56,111�118 (1987).

110 M. R. Austwick et al.

Page 15: optical measurement of photosensitizer concentration in vivo

24. S. A. Prahl, \Hemoglobin spectra," Oregon MedicalLaser Center, Oregon, USA (1999).

25. A. J. Welch and M. J. van Gemert, Optical-ThermalResponse of Laser-Irradiated Tissue, Plenum, 2005.

26. S. Kascakova, B. Kruijt, H. S. de Bruijn, van der Ploeg-van den Heuvel , D. J. Robinson, H. J. Sterenborg,A. Amelink, \Ex vivo quanti¯cation of mTHPC con-centration in tissue: In°uence of chemical extraction onthe optical properties," J. Photochem. Photobiol. B 91,99�107 (2008).

27. K. Berg, J. C. Bommer, J. Moan, \Evaluation of sul-fonated aluminum phthalocyanines for use in photo-chemotherapy. A study on the relative e±ciencies ofphotoinactivation," Photochem. Photobiol. 49, 587�594 (1989).

28. R. Cubeddu, G. Canti, M. Musolino, A. Pi®eri,P. Taroni, G. Valentini, \In vivo absorption spec-trum of disulphonated aluminium phthalocyanine ina murine tumour model," J. Photochem. Photobiol.B 34, 229�235 (1996).

29. L. Kunz, J. P. Connelly, J. H. Woodhams,A. J. Macrobert, \Photodynamic modi¯cation ofdisulfonated aluminium phthalocyanine °uorescencein a macrophage cell line," Photochem. Photobiol.Sci. 6, 940�948 (2007).

30. P. Taroni, A. Pi®eri, A. Torricelli, D. Comelli,R. Cubeddu, \In vivo absorption and scatteringspectroscopy of biological tissues," Photochem.Photobiol. Sci. 2, 124�129 (2003).

In vivo Optical Pharmacokinetics 111