Operation Manual Multi-parameter Transmitter M200

Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Operation Manual Multi-parameter Transmitter M200

Page 2: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 2

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Page 3: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 3

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Operation Manual Multi-parameter Transmitter M200

Page 4: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 4

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Content

1 Introduction ___________________________________________________________________________________________ 72 Safety instructions ______________________________________________________________________________________ 7

2.1 DefinitionofEquipmentandDocumentationSymbolsandDesignations _______________________________________ 72.2 CorrectDisposaloftheUnit __________________________________________________________________________ 8

3 Unit Overview __________________________________________________________________________________________ 93.1 Overview 1/4DIN __________________________________________________________________________________ 93.2 Overview 1/2DIN __________________________________________________________________________________ 93.3 Control/NavigationKeys ___________________________________________________________________________ 10

3.3.1 MenuStructure ___________________________________________________________________________ 103.3.2 NavigationKeys __________________________________________________________________________ 10

3.3.2.1 NavigatingtheMenuTree _________________________________________________________ 103.3.2.2 Escape _______________________________________________________________________ 113.3.2.3 Enter _________________________________________________________________________ 113.3.2.4 Menu _________________________________________________________________________ 113.3.2.5 Calibration Mode ________________________________________________________________ 113.3.2.6 Info Mode _____________________________________________________________________ 11

3.3.3 NavigationofDataEntryFields ______________________________________________________________ 113.3.4 EntryofDataValues,SelectionofDataEntryOptions _____________________________________________ 113.3.5 NavigationwithuinDisplay ________________________________________________________________ 123.3.6 ”Savechanges”Dialog ____________________________________________________________________ 123.3.7 SecurityPasswords _______________________________________________________________________ 12

3.4 Display ________________________________________________________________________________________ 124 Installation instruction __________________________________________________________________________________ 13

4.1 UnpackingandInspectionofEquipment _______________________________________________________________ 134.2 Installation – 1/4DIN Models _______________________________________________________________________ 13

4.2.1 1/4DINVersion–DimensionDrawings ________________________________________________________ 134.2.2 InstallationProcedure–1/4DINModels _______________________________________________________ 14

4.3 Installation – 1/2DIN Models _______________________________________________________________________ 154.3.1 1/2DINVersion–DimensionDrawings ________________________________________________________ 154.3.2 1/2DINVersion–PipeMounting _____________________________________________________________ 154.3.3 InstallationProcedure–1/2DINModels _______________________________________________________ 16

4.4 ConnectionofPowerSupply ________________________________________________________________________ 184.4.1 1/4DINHousing(PanelMount) ______________________________________________________________ 184.4.2 1/2DINHousing(WallMount)_______________________________________________________________ 19

4.5 ConnectorTerminalDefinition _______________________________________________________________________ 204.5.1 TB1andTB2for1/2DINand1/4DINVersions __________________________________________________ 204.5.2 TB3/TB4*–pH,ORP,DissolvedOxygen,Ozoneand4-ElectrodeConductivitySensor ____________________ 204.5.3 TB3/TB4–2-electrodeConductivitySensor _____________________________________________________ 21

4.6 AssemblingofSensorandCable ____________________________________________________________________ 224.6.1 ConnectionofSensorsforpH,ORP,DissolvedOxygen,Ozoneand4-electrodeConductivity _______________ 224.6.2 AK9CableAssignment ____________________________________________________________________ 22

5 Placing transmitter in, or Out of Service ____________________________________________________________________ 235.1 PlacingTransmitterinService _______________________________________________________________________ 235.2 PlacingTransmitterOutofService ___________________________________________________________________ 23

6 Quick Setup __________________________________________________________________________________________ 247 Sensor Calibration _____________________________________________________________________________________ 25

7.1 EnterCalibrationMode ____________________________________________________________________________ 257.2 Conductivity/ResistivityCalibration ___________________________________________________________________ 25

7.2.1 One-point Sensor Calibration _______________________________________________________________ 267.2.2 Two-pointSensorCalibration(4-electrodeSensorsonly) __________________________________________ 267.2.3 ProcessCalibration _______________________________________________________________________ 27

7.3 OxygenCalibration _______________________________________________________________________________ 287.3.1 One-Point Sensor Calibration ________________________________________________________________ 287.3.2 ProcessCalibration _______________________________________________________________________ 28

7.4 pH Calibration ___________________________________________________________________________________ 297.4.1 One-Point Calibration _____________________________________________________________________ 29

7.4.1.1 Auto Mode _____________________________________________________________________ 297.4.1.2 Manual mode __________________________________________________________________ 30

7.4.2 Two-Point Calibration _____________________________________________________________________ 307.4.2.1 Auto Mode _____________________________________________________________________ 307.4.2.2 Manual Mode __________________________________________________________________ 31

7.4.3 ProcessCalibration _______________________________________________________________________ 317.5 ORP Calibration _________________________________________________________________________________ 32

7.5.1 One-Point Calibration ______________________________________________________________________ 32

Page 5: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 5

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

7.6 Ozone Calibration ________________________________________________________________________________ 327.6.1 One-Point ZeroPt Calibration ________________________________________________________________ 327.6.2 ProcessCalibration _______________________________________________________________________ 33

7.7 SensorVerification ________________________________________________________________________________ 338 Configuration _________________________________________________________________________________________ 34

8.1 EnterConfigurationMode __________________________________________________________________________ 348.2 Measurement ___________________________________________________________________________________ 34

8.2.1 ChannelSetup ___________________________________________________________________________ 348.2.2 Derived Measurements ____________________________________________________________________ 35

8.2.2.1 %Rejectionmeasurement ________________________________________________________ 358.2.2.2 CalculatedpH(PowerPlantApplicationsonly) ________________________________________ 368.2.2.3 CalculatedCO2(PowerPlantApplicationsonly) ________________________________________ 36

8.2.3 ParameterRelatedSettings _________________________________________________________________ 368.2.3.1 Conductivity/TemperatureCompensation _____________________________________________ 378.2.3.2 pH Parameters _________________________________________________________________ 388.2.3.3 DissolvedOxygenParameters _____________________________________________________ 38

8.2.4 SetAveraging ____________________________________________________________________________ 398.3 AnalogOutputs __________________________________________________________________________________ 408.4 Setpoints ______________________________________________________________________________________ 418.5 Alarm/Clean ____________________________________________________________________________________ 42

8.5.1 Alarm __________________________________________________________________________________ 428.5.2 Clean __________________________________________________________________________________ 43

8.6 Display ________________________________________________________________________________________ 438.6.1 Measurement ____________________________________________________________________________ 438.6.2 Resolution ______________________________________________________________________________ 448.6.3 Backlight _______________________________________________________________________________ 448.6.4 Name __________________________________________________________________________________ 44

8.7 HoldAnalogOutputs ______________________________________________________________________________ 459 System ______________________________________________________________________________________________ 46

9.1 SetLanguage ___________________________________________________________________________________ 469.2 USB ___________________________________________________________________________________________ 469.3 Passwords _____________________________________________________________________________________ 47

9.3.1 ChangingPasswords ______________________________________________________________________ 479.3.2 ConfiguringMenuAccessforOperator _________________________________________________________ 47

9.4 Set/ClearLockout ________________________________________________________________________________ 489.5 Reset __________________________________________________________________________________________ 48

9.5.1 ResetSystem ____________________________________________________________________________ 489.5.2 ResetAnalogCalibration ___________________________________________________________________ 48

10 Service ______________________________________________________________________________________________ 4910.1 Diagnostics _____________________________________________________________________________________ 49

10.1.1 Model/Software Revision ___________________________________________________________________ 4910.1.2 DigitalInput _____________________________________________________________________________ 4910.1.3 Display ________________________________________________________________________________ 5010.1.4 Keypad ________________________________________________________________________________ 5010.1.5 Memory ________________________________________________________________________________ 5010.1.6 SetRelay _______________________________________________________________________________ 5010.1.7 ReadRelays _____________________________________________________________________________ 5110.1.8 SetAnalogOutputs _______________________________________________________________________ 5110.1.9 ReadAnalogOutputs ______________________________________________________________________ 51

10.2 Calibrate _______________________________________________________________________________________ 5110.2.1 CalibrateAnalog _________________________________________________________________________ 5210.2.2 CalibrateUnlock __________________________________________________________________________ 52

10.3 TechService ____________________________________________________________________________________ 5211 Info _________________________________________________________________________________________________ 53

11.1 Messages ______________________________________________________________________________________ 5311.2 Calibration Data _________________________________________________________________________________ 5311.3 Model/Software Revision ___________________________________________________________________________ 5411.4 Sensor Info _____________________________________________________________________________________ 54

12 Maintenance __________________________________________________________________________________________ 5512.1 FrontPanelCleaning ______________________________________________________________________________ 55

13 Troubleshooting _______________________________________________________________________________________ 5613.1 Cond(Resistivity)ErrorMessages/Warning-andAlarmList ______________________________________________ 5713.2 OxygenErrorMessages/Warning-andAlarmList _______________________________________________________ 5713.3 pHErrorMessages/Warning-andAlarmList __________________________________________________________ 5713.4 ORPErrorMessages/Warning-andAlarmList _________________________________________________________ 58

Page 6: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 6

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

13.5 Warning-andAlarmIndicationontheDisplay __________________________________________________________ 5813.5.1 WarningIndication________________________________________________________________________ 5813.5.2 AlarmIndication __________________________________________________________________________ 58

14 Accessories and Spare Parts _____________________________________________________________________________ 5915 Specifications _________________________________________________________________________________________ 60

15.1 GeneralSpecifications _____________________________________________________________________________ 6015.2 ElectricalSpecifications ____________________________________________________________________________ 6215.3 MechanicalSpecifications __________________________________________________________________________ 62

15.3.1 MechanicalSpecificationsfor1/2DINVersion ___________________________________________________ 6215.3.2 MechanicalSpecificationsfor1/4DINVersion ___________________________________________________ 63

15.4 EnvironmentalSpecifications _______________________________________________________________________ 6316 Default Tables _________________________________________________________________________________________ 64

16.1 M200(1-channelVersion) _________________________________________________________________________ 6416.2 M200(2-channelVersion) _________________________________________________________________________ 6516.3 ParameterRelatedValues __________________________________________________________________________ 67

16.3.1 Conductivity _____________________________________________________________________________ 6716.3.2 Oxygen ________________________________________________________________________________ 6816.3.3 pH ____________________________________________________________________________________ 6916.3.4 ORP ___________________________________________________________________________________ 7016.3.5 Ozone _________________________________________________________________________________ 70

17 Warranty _____________________________________________________________________________________________ 7118 Certificate ____________________________________________________________________________________________ 7219 Buffer Tables __________________________________________________________________________________________ 73

19.1 Mettler-9 _______________________________________________________________________________________ 7319.2 Mettler-10 ______________________________________________________________________________________ 7319.3 NISTTechnicalBuffers _____________________________________________________________________________ 7419.4 NISTStandardBuffers(DIN19266:2000–01) _________________________________________________________ 7419.5 HachBuffers ___________________________________________________________________________________ 7519.6 Ciba(94)Buffers ________________________________________________________________________________ 7519.7 MerckTitrisole,Riedel-de-HaënFixanale ______________________________________________________________ 7619.8 WTWBuffers ____________________________________________________________________________________ 76

Page 7: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 7

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

1 IntroductionStatementofIntendedUse–TheM200multi-parametertransmitterisa1-or2-channelonlineprocessinstrumentformeasuringvariouspropertiesoffluids.TheseincludeConductivity/Resistivity,DissolvedOxygen,pH,ORPandOzone.ItwillinterfacewithavarietyofdifferentMettler-Toledosensors,whichconnecttothetransmitterusingcablesofvariedlengths.

AlargefourlinebacklitLiquidCrystalDisplayconveysmeasuringdataandsetupinformation.Themenustructureallowstheoperatortomodifyalloperationalparametersbyusingkeysonthefrontpanel.Amenu-lockoutfeature,withpasswordprotection,isavailabletopreventtheunauthorizeduseofthemeter.TheM200multi-parametertransmittercanbeconfiguredtouseits2(4on2-channelversion)analogand/or2relayoutputsforprocesscontrol.

TheM200multi-parametertransmitterisequippedwithaUSBcommunicationinterface.Thisinterfaceprovidesreal-timedataoutputandcompleteinstrumentconfigurationcapabilitiesforcentralmonitoringviaPersonalComputer(PC).

ThismanualappliesforallavailableM200transmittersasfollows:– Multi-parameter2-channelversion– Multi-parameter1-channelversion

Theprintscreenimagesinthismanualhaveageneralexplainingcharacterandcandifferfromtherealdisplayinyourtransmitter.

2 Safety instructionsThismanualincludessafetyinformationwiththefollowingdesignationsandformats.

2.1 DefinitionofEquipmentandDocumentationSymbolsand Designations

a WARNING:POTENTIALFORPERSONALINJURY.

a CAUTION:possibleinstrumentdamageormalfunction.

h NOTE:Importantoperatinginformation.

a On the transmitter or in this manual text indicates:Cautionand/orotherpossiblehazardincludingriskofelectricshock(refertoaccompanyingdocuments).

Page 8: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 8

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Thefollowingisalistofgeneralsafetyinstructionsandwarnings.Failuretoadheretotheseinstructionscanresultindamagetotheequipmentand/orpersonalinjurytotheoperator.

– TheM200transmittershouldbeinstalledandoperatedonlybypersonnelfamiliarwith thetransmitterandwhoarequalifiedforsuchwork.

– TheM200transmittermustonlybeoperatedunderthespecifiedoperatingconditions (seesection15“Specifications”).

– RepairoftheM200transmittermustbeperformedbyauthorized,trainedpersonnelonly.– Withtheexceptionofroutinemaintenance,cleaningproceduresorfusereplacement,as

describedinthismanual,theM200transmittermustnotbetamperedwithoralteredin anymanner.

– Mettler-Toledoacceptsnoresponsibilityfordamagecausedbyunauthorizedmodifications tothetransmitter.

– Followallwarnings,cautions,andinstructionsindicatedonandsuppliedwiththisproduct.– Installequipmentasspecifiedinthisinstructionmanual.Followappropriatelocaland

nationalcodes.– Protectivecoversmustbeinplaceatalltimesduringnormaloperation.– Ifthisequipmentisusedinamannernotspecifiedbythemanufacturer,theprotection

providedbyitagainsthazardsmaybeimpaired.

WARNINGS:Installationofcableconnectionsandservicingofthisproductrequireaccesstoshockhazardvoltagelevels.Mainpowerandrelaycontactswiredtoseparatepowersourcemustbedisconnectedbeforeservicing.SwitchorcircuitbreakershallbeincloseproximitytotheequipmentandwithineasyreachoftheOPERATOR;itshallbemarkedasthedisconnectingdevicefortheequipment.Mainpowermustemployaswitchorcircuitbreakerasthedisconnectingdevicefortheequipment.ElectricalinstallationmustbeinaccordancewiththeNationalElectricalCodeand/oranyotherapplicablenationalorlocalcodes.

h NOTE:RELAYCONTROLACTION:theM200transmitterrelayswillalwaysde-energizeonlossofpower,equivalenttonormalstate,regardlessofrelaystatesettingforpoweredoperation.Configureanycontrolsystemusingtheserelayswithfail-safelogicaccordingly.

h NOTE:PROCESSUPSETS:Becauseprocessandsafetyconditionsmaydependonconsistentoperationofthistransmitter,provideappropriatemeanstomaintainoperationduringsensorcleaning,replacementorsensororinstrumentcalibration.

2.2 Correct Disposal of the UnitWhenthetransmitterisfinallyremovedfromservice,observealllocalenvironmentalregulationsforproperdisposal.

Page 9: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 9

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

3 Unit OverviewM200modelsareavailableinbotha1/4DINand1/2DINcasesize.The1/4DINisapanel-mountonlydesignandthe1/2DINmodelsprovideanintegralIP65housingforwall-,orpipe-mount.

3.1 Overview 1/4DIN

4.01"102 mm

4.01

"10

2 m

m

EnterInfo

MET

TLER

TO

LED

O

ESC

Menu Cal

M200

1

2

34

56

7

8

1–HardPolycarbonatecase2–FiveTactile-FeedbackNavigationKeys3–Four-lineLCDisplay4–PowerSupplyTerminals

5–USBInterfacePort6–RelayOutputTerminals7–AnalogOutput/DigitalInputTerminals8 – Sensor Input Terminals

3.2 Overview 1/2DIN

EnterInfoCalMenu

M200

1

2

3

ESC

5.90"150 mm

5.90

"15

0 m

m

4

5

6

7 8

1–HardPolycarbonatecase2–FiveTactile-FeedbackNavigationKeys3–Four-lineLCDisplay4–PowerSupplyTerminals

5–USBInterfacePort6–RelayOutputTerminals7–AnalogOutput/DigitalInputTerminals8 – Sensor Input Terminals

Page 10: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 10

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

3.3 Control/Navigation Keys

3.3.1 Menu Structure

BelowisthestructureoftheM200menutree:

MeasurementMode M200

CalMenu Info

Quick Setup Configure System Service

Measurement

Analog Outputs

Set Points

Alarm/Clean

Display

Calibrate

Tech Service

Diagnostics

Hold Outputs

Set Language

USB

Passwords

Set/ClearLockout

Reset

Messages Sensor Info Calibration Data Model/SoftwareRevision

Channel Setup

Analog Outputs

Set Points

3.3.2 Navigation Keys

Menu Cal Info Enter

ESC

3.3.2.1 Navigating the Menu Tree

EnterthedesiredmainMenubranchwiththe c or .keys.Usethe and .keystonavigatethroughtheselectedMenubranch.

h NOTE:Inordertobackuponemenupage,withoutescapingtothemeasurementmode, movethecursorundertheUPArrowcharacter(u) atthebottomrightofthedisplayscreenandpress[Enter].

Page 11: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 11

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

3.3.2.2 Escape

Pressthe and ckeysimultaneously(escape)toreturntotheMeasurementmode.

3.3.2.3 Enter

Usetheekeytoconfirmactionorselections.

3.3.2.4 Menu

PressthekeytoaccessthemainMenu.

3.3.2.5 Calibration Mode

PresstheckeytoenterCalibrationMode.

3.3.2.6 Info Mode

Pressthe.keytoenterInfoMode

3.3.3 NavigationofDataEntryFields

Usetheckeytonavigateforwardorthekeytonavigatebackwardswithinthechangeabledataentryfieldsofthedisplay.

3.3.4 EntryofDataValues,SelectionofDataEntryOptions

Usethekeytoincreaseorthe.keytodecreaseadigit.Usethesamekeystonavigatewithinaselectionofvaluesoroptionsofadataentryfield.

h NOTE:Somescreensrequireconfiguringmultiplevaluesviathesamedatafield(ex:configuringmultiplesetpoints).Besuretousethec or keytoreturntotheprimaryfieldandthe or .keytotogglebetweenallconfigurationoptionsbeforeenteringtothenextdisplayscreen.

Page 12: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 12

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

3.3.5 Navigation with u in Display

If a uisdisplayedonthebottomrighthandcornerofthedisplay,youcanusethecorthe keytonavigatetoit.Ifyouclick[ENTER]youwillnavigatebackwardsthroughthemenu (gobackonescreen).Thiscanbeaveryusefuloptiontomovebackupthemenutreewithouthavingtoexitintothemeasuringmodeandre-enterthemenu.

3.3.6 ”Save changes” Dialog

Threeoptionsarepossibleforthe”Savechanges”dialog:Yes&Exit(Savechangesandexittomeasuringmode),”Yes&u”(Savechangesandgobackonescreen)and”No&Exit”(Don’tsavechangesandexittomeasuringmode).The”Yes&u”optionisveryusefulifyouwanttocontinueconfiguringwithouthavingtore-enterthemenu.

3.3.7 Security Passwords

TheM200transmitterallowsasecuritylock-outofvariousmenus.Ifthesecuritylock-out featureofthetransmitterhasbeenenabled,asecuritypasswordmustbeenteredtoallowaccesstothemenu.Seesection9.3”System/Passwords”formoreinformation.

3.4 Display

h NOTE:IntheeventofanalarmorothererrorconditiontheM200transmitterwilldisplayaflashing intheupperrightcornerofthedisplay.Thissymbolwillremainuntiltheconditionthatcausedithasbeencleared.

h NOTE:Duringcalibrations,clean,DigitalInwithAnalogOutput/Relay/USBinHoldstate, aflashingHwillappearintheupperleftcornerofthedisplay.Thissymbolwillremainfor 20secondsuntilafterthecalibrationorcleaniscompleted.ThissymbolwillalsodisappearwhenDigitalInisdeactivated.

Page 13: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 13

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4 Installation instruction

4.1 UnpackingandInspectionofEquipmentInspecttheshippingcontainer.Ifitisdamaged,contacttheshipperimmediatelyforinstructions.Donotdiscardthebox.

Ifthereisnoapparentdamage,unpackthecontainer.Besureallitemsshownonthepackinglist are present. Ifitemsaremissing,notifyyourMETTLERTOLEDOrepresentativeimmediately.

4.2 Installation – 1/4DIN Models

4.2.1 1/4DINVersion–DimensionDrawings110 mm / 4.33"

102 mm / 4.02"

90 mm / 3.54"

102

mm

/ 4.

02"

Power

102 mm / 4.02"

126

mm

/ 4.

96"

125

mm

/ 4.

92"

90 mm / 3.54"

110 mm / 4.33"

TB1A

TB1BUSB TB2

TB4 TB3

Fuse

1 2

45 mm / 1.77"

max

. 3 m

m

0.11

8"

2

3

92 mm

3.62" +0.02"

+0.5 mm

92 m

m

3.62

"+0.

02"

+0.

5 m

m

1

4

1–Dimensionsforpanelcutout2–Mountingscrews,2pieces3–Flatgasket,1piece4–Mountingbracket,2pieces

Page 14: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 14

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.2.2 Installation Procedure – 1/4DIN Models

1/4DINModeltransmittersaredesignedforpanel-mountinstallationonly.Eachtransmitterissuppliedwithmountinghardwaretoprovidefastandsimpleinstallationtoaflatpanelorflatenclosuredoor.ToinsureagoodsealandmaintainIP65integrityofinstallation,thepanelordoormustbeflatandhaveasmoothfinish.

Thesuppliedhardwareconsistsof:Twopiecessnap-onmountingbracketsOnepiecemountinggasketseal

– Makecutoutinpanel.Fordimensionsreferto4.2.1“1/4DINVersion–DimensionDrawings”.– Besuresurfacesurroundingcutoutisclean,smoothandfreeofburrs.– Slidefacegasketaroundtransmitterfromthebackoftheunit.– Placetransmitterintocutouthole.Besuretherearenogapsbetweenthetransmitter

andpanelsurface.– Placethetwomountingbracketsoneithersideofthetransmitterasshown.– Whileholdingtransmitterfirmlyintothecutouthole,pushthemountingbrackets

towardthebacksideofpanel.– Oncesecure,useascrewdrivertotightenthebracketsagainstthepanel.Inordertoprovide

IP65environmentalenclosurerating,thetwoclampsprovidedshallbesecurelytightenedtocreateanadequatesealbetweenthepanelenclosureandM200frontface.

– Facegasketwillcompressbetweentransmitterandpanel.

a CAUTION:Donotovertightenbrackets.

Page 15: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 15

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.3 Installation – 1/2DIN Models

4.3.1 1/2DINVersion–DimensionDrawings

136

mm

/5.3

5"

116 mm/4.57"

29 mm

1.14"

29 mm1.14"

84 mm3.31"

42 mm1.65"

150

mm

/5.9

"

6.5

mm

0.25

6"

150 mm/5.9"

M200

90 mm/3.54"

80 mm/3.15"

75 m

m/2

.95"

35 mm/1.38"

6 m

m/

0.23

6"

90 m

m/3

.54"

137 mm

5.39" +0.02"

+0.5 mm

137

mm

5.39

"+0.

02"

+0.

5 m

m

1

1–Dimensionsforpanelcutout

4.3.2 1/2DINVersion–PipeMounting

�40 ... �60 mm�1.57... �2.36"

Page 16: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 16

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.3.3 Installation Procedure – 1/2DIN Models

1/2DINModeltransmittersaredesignedforthefollowinginstallationversions:panalmount,wallmountorpipemount.Forwallmounttheintegralrearcoverisused.Optionalhardwareaccessoriesareavailablethatallowforpanel-orpipe-mount.Refertosection14“AccessoriesandSpareParts”.

Assembly:

1

2

3

1 3piecesM20cableglands2 2piecesplasticplugs3 4piecesscrews

General:– Orientthetransmittersothatthecablegripsfacedownward.– Wiringroutedthroughthecablegripsshallbesuitableforuseinwetlocations.– InordertoprovideIP65enclosureratings,allcableglandsmustbeinplace.Eachcable

glandmustbefilledusingacable,orsuitablecableglandholeseal.

ForPanelMount:Toinsureagoodseal,thepanelordoormustbeflatandhaveasmoothfinish.Texturedorroughsurfacesarenotrecommendedandmaylimittheeffectivenessofthegasketsealprovided.– Makecutoutinpanel.Fordimensionsreferto4.3.1“1/2DINVersion–DimensionDrawings”.– Besuresurfacesurroundingcutoutisclean,smoothandfreeofburrs.– Slidefacegasketaroundtransmitterfromthebackoftheunit.– Placetransmitterintocutouthole.Besuretherearenogapsbetweenthetransmitter

andpanelsurface.– Placethetwomountingbracketsoneithersideofthetransmitterasshown.– Whileholdingtransmitterfirmlyintothecutouthole,pushthemountingbrackets

towardthebacksideofpanel.– Oncesecure,useascrewdrivertotightenthebracketsagainstthepanel.Inordertoprovide

IP65environmentalenclosurerating,thetwoclampsprovidedshallbesecurelytightenedtocreateanadequatesealbetweenthepanelenclosureandM200frontface.

– Facegasketwillcompressbetweentransmitterandpanel.

Page 17: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 17

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

ForWallMount:– Removerearcoverfromfronthousing.– Startbyunscrewingthefourscrewslocatedonthefaceofthetransmitter,ineachcorner.

Thisallowsthefrontcovertoswingawayfromtherearhousing.– Removethehinge-pinbysqueezingthepinfromeachend.

Thisallowsthefronthousingtoberemovedfromtherearhousing.– Mountrearhousingtowallusingonlymanufacturer-suppliedmountingkit.Securemounting

kittotheM200accordingtothesuppliedinstructions.Attachtowallusingappropriatemountinghardwareforwallsurface.Besureitislevelandsecurelyfastenedandtheinstallationadherestoanyandallclearancedimensionsrequiredfortransmitterserviceandmaintenance.Orientthetransmittersothatthecablegripsarefacingdownward.

– Replacethefronthousingtotherearhousing.Securelytightentherear-coverscrewstoensurethatIP65enclosureenvironmentalratingisprovided.Theunitisreadytobewired.

ForPipeMount:– Useonlymanufacturer-suppliedcomponentsforpipe-mountingtheM200transmitterand

installperthesuppliedinstructions.Seesection14“AccessoriesandSpareParts” for orderinginformation.

Page 18: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 18

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.4 Connection of Power SupplyAllconnectionstothetransmitteraremadeontherearpanelofallmodels.

a Besurepowertoallwiresisturnedoffbeforeproceedingwiththeinstallation.Highvoltagemaybepresentontheinputpowerwiresandrelaywires.

Atwo-terminalconnectorontherearpanelofallM200modelsisprovidedforpowerconnection.AllM200modelsaredesignedtooperatefroma20–30VDCora100to240VACpowersource.Refertospecificationsforpowerrequirementsandratingsandsizepowerwiringaccordingly.

Theterminalblockforpowerconnectionsislabeled”Power”ontherearpanelofthetransmitter. One terminal is labeled – NfortheNeutralwireandtheother+LfortheLine(orLoad)wire.Thereisnoearthgroundterminalonthetransmitter.Forthisreasontheinternalpowerwiringwithinthetransmitterisdoubleinsulatedandtheproductlabeldesignatesthisusingthedsymbol.

4.4.1 1/4DIN Housing (Panel Mount)

1

2

1: Connectionofpowersupply2: Terminalforsensors

Page 19: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 19

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.4.2 1/2DIN Housing (Wall Mount)

1

2

1: Connectionofpowersupply2: Terminalforsensors

Page 20: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 20

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.5 ConnectorTerminalDefinition

4.5.1 TB1andTB2for1/2DINand1/4DINVersions

Power connections are labeled –N for Neutral and +L for Line, for 100 to 240 VAC or 20–30 VDC.

1⁄4 DIN 1⁄2 DIN

* Dual channel only

TB1A for 1⁄4 DIN1 NO22 COM23 NC24 –5 –6 –7 –

TB1B for 1⁄4 DIN1 NO12 COM13 NC14 –5 –6 –7 –

TB2 for 1⁄4 DIN1 AO1+2 AO1–/AO2–3 AO2+4 AO3+*5 AO3–/AO4–*6 AO4+*7 DI1+8 DI1–/DI2–*9 DI2+*

TB2 for 1⁄2 DIN1 AO1+2 AO1–/AO2–3 AO2+4 AO3+*5 AO3–/AO4–*6 AO4+*7 DI1+8 DI1–/DI2–*9 DI2+*

1 NO12 COM13 NC14 NO25 COM26 NC27 –

8 –9 –

10 –11 –12 –13 –14 –

TB4

TB2

TB3

TB1A

TB1B1 7

1 7

9

1

9

1

9

1TB2 TB3 TB4

TB1 1 14

1 9 1 9 1 9

TB1 for 1⁄2 DIN

NO: normallyopen(contactopenifun-actuated) AO:AnalogOutputNC: normallyclosed(contactclosedifun-actuated) DI:DigitalInput

4.5.2 TB3/TB4* – pH, ORP, Dissolved Oxygen, Ozone and 4-ElectrodeConductivitySensor

ThewiringofthesensorsforpH,oxygen,ozoneand4-electrodeconductivitytoTB3resp.TB4is:

Terminal Sensor wire color Function1 – –2 – –3 Cablecore(transparent) 1-Wire4 Shield(red) GND(5VDC)5 – –6 – GND(5VDC)7 – RS485–B8 – RS485–A9 – 5VDC

*Onlyon2-channelversion.

Page 21: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 21

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.5.3 TB3/TB4 – 2-electrode Conductivity Sensor

Thewiringofthesensorsfor2-electrodeconductivitytoTB3resp.TB4is:

Terminal Sensor wire color* FunctioneasySense UniCond

1 – –2 – –3 – 1-Wire4 – GND(5VDC)5 – –6 Green White GND(5VDC)7 Orange Black RS485–B8 White/Orange Red RS485–A9 White/Green Blue 5VDC

*Barewirenotconnected.

Page 22: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 22

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

4.6 Assembling of Sensor and Cable

4.6.1 Connection of Sensors for pH, ORP, Dissolved Oxygen, Ozone and 4-electrode Conductivity

A B

h NOTE:Connectthesensorandscrewtheplugheadclockwise(handtight).

4.6.2 AK9 Cable Assignment

A:1-wiredata(transparent)B:Ground/shield(red)

Page 23: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 23

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

5 Placing transmitter in, or Out of Service

5.1 Placing Transmitter in Service

a Afterconnectingthetransmittertopowersupplycircuit,itwillbeactiveassoonasthecircuitispowered.

5.2 Placing Transmitter Out of Service Firstdisconnecttheunitfromthemainpowersource,thendisconnectallremainingelectricalconnections.Removetheunitfromthewall/panel.Usetheinstallationinstructioninthismanualasreferencefordis-assemblingmountinghardware.

Page 24: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 24

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

6 Quick Setup(PATH:Menu/QuickSetup)

SelectQuickSetupandpressthe[ENTER]key.Enterthesecuritycodeifnecessary (seesection9.3 “Passwords”).

h Note:PleasefindthecompletedescriptionoftheQuickSetuproutinedescribedintheseparatebooklet”QuickSetupGuideforTransmitterM200”enclosedinthebox.

h Note:Refertosection 3.3“Control/NavigationKeys” for informationonmenunavigation.

Page 25: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 25

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

7 Sensor Calibration(PATH:Cal)

Thecalibrationkey[CAL]allowstheuserone-touchaccesstoSensorcalibrationandverificationfeatures.TheM200alsoallowsaccesstoAnalogOutputcalibrationiftheaccesshasbeenpreviouslyunlocked(seesection10.2 “Calibrate”).

h NOTE:DuringCalibration,aflashing”H”intheupperleftcornerofthedisplayindicatesacalibrationisinprocesswithaHoldconditionactive.(Theholdoutputfunctionneedtobeactivated.)

7.1 EnterCalibrationModeWhileinMeasurementmodepressthekey[CAL].Ifthedisplaypromptsyoutoenterthecalibrationsecuritycode,pressthe or .keytosetthecalibrationsecuritycode,thenpressthe[ENTER]keytoconfirmthecalibrationsecuritycode.

ForMulti-channeldevices:Usingthe or .keyonthe”ChannelA”fieldletstheuserchangethechanneltobecalibrated.Thenusetheckeytomovetothecalibrationfield.

Selectthedesiredsensorcalibrationtask.Thechoicesforeachsensortypeare:Conductivity=Conductivity,Resistivity,VerifyOxygen =Oxygen,VerifypH/ORP =pH,VerifyORP =ORP,VerifyOzone =Ozone,Verify

Press[ENTER].

7.2 Conductivity/Resistivity CalibrationThisfeatureprovidestheabilitytoperformaone-point,two-pointorprocessconductivityorResistivity”Sensor”calibration.Theproceduredescribedbelowworksforbothtypesofcalibrations.Thereisnoreasontoperformatwo-pointcalibrationonatwo-electrodeconductivitysensor.Fourelectrodesensorsdorequireatwo-pointcalibration.Itisalsonotpracticaltocalibrateresistivitysensorsusing(lowconductivity)referencesolutions.Itisrecommendedthatresistivitysensorsbesentbacktothefactoryforcalibration.Consultfactoryforassistance.

h NOTE:Whenperformingcalibrationonaconductivityorresistivitysensor,resultswillvarydependingonthemethods,calibrationapparatusand/orqualityofreferencestandardsusedtoperformthecalibration.

Enterconductivitysensorcalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”.

Page 26: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 26

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Afterselectingthedesiredsensorcalibrationandpressing[ENTER],thenextscreenwillasktoselectthetypeoftemperaturecompensationmodedesiredduringthecalibrationprocess. Thechoicesare”Standard”,”Light84”,”Std75ºC”,”Lin20°C=02.0%/ºC”(userselectablevalue),”Lin25°C=02.0%/ºC”(userselectablevalue),”Glycol.5”,”Glycol1”,”Alcohol”and”NatH2O”.

Press[ENTER].

7.2.1 One-point Sensor Calibration

Enterconductivitysensorcalibrationmodeasdescribedinsection7.1“EnterCalibrationMode” andchooseoneofthecompensationmodes(seesection7.2“Conductivity/ResistivityCalibration”).

(Displayreflectstypicalconductivitysensorcalibration)

Select1pointCalibrationandpress[ENTER].

h NOTE:Rinsesensorswithahigh-puritywatersolutionbeforeeverycalibrationtopreventcontaminationofthereferencesolutions.

Placethesensorintothereferencesolution.

EntertheValueofcalibrationPoint1andthenpressthe[ENTER]keytostartcalibration. Thevalueinthe2ndtextlineistheactualmeasuredvaluefromthesensorpriortocalibration.

AfterthecalibrationtheMultiplierorslopecalibrationfactor”M”andtheAdderoroffsetcalibrationfactor”A”aredisplayed.

SelectYestosavethecalibrationvaluesandtheSuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressEnter”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.2.2 Two-point Sensor Calibration (4-electrode Sensors only)

Enterconductivitysensorcalibrationmodeasdescribedinsection7.1“EnterCalibrationMode” andchooseoneofthecompensationmodes(seesection7.2“Conductivity/ResistivityCalibration”).

Select2pointCalibrationandpress[ENTER].

h NOTE: Rinsesensorswithahigh-puritywatersolutionbetweencalibrationpointstopreventcontaminationofthereferencesolutions.

Placethesensorintothefirstreferencesolution.

Page 27: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 27

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

EntertheValueofPoint1andpressthe[ENTER]key.Placethesensorintothesecondreferencesolution.

EntertheValueofPoint2andpressthe[ENTER]keytostartthecalibration.

AfterthecalibrationtheMultiplierorslopecalibrationfactor”M”andtheAdderoroffsetcalibrationfactor”A”aredisplayed.

SelectYestosavethecalibrationvaluesandtheSuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressEnter”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.2.3 Process Calibration

Enterconductivitysensorcalibrationmodeasdescribedinsection7.1“EnterCalibrationMode” andchooseoneofthecompensationmodes(seesection7.2“Conductivity/ResistivityCalibration”).

SelectProcessCalibrationandpress[ENTER].

Takeasampleandpressthe[ENTER]keyagaintostorethecurrentmeasuringvalue.

Duringtheongoingcalibrationprocess,theletterofthechannel,whichisconcernedbythecalibration,”A”or“B”isblinkinginthedisplay.

Afterdeterminingtheconductivityvalueofthesample,pressthe[CAL]keyagaintoproceedwiththecalibration.

Entertheconductivityvalueofthesamplethenpressthe[ENTER]keytostartthecalculationofcalibrationresults.

AfterthecalibrationtheMultiplierorslopecalibrationfactor”M”andtheAdderoroffsetcalibrationfactor”A”aredisplayed.

SelectYestosavethecalibrationvaluesandtheSuccessfulCalibrationisconfirmedonthedisplay.

Page 28: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 28

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

7.3 Oxygen CalibrationDissolvedOxygencalibrationisperformedaseitheraone-pointorprocesscalibration.

7.3.1 One-Point Sensor Calibration

Beforeaircalibration,forhighestaccuracy,enterthebarometricpressureandrelativehumidityasinsection8.2.3.3”Configuration/Measurement/ParameterRelatedSettings/DissolvedOxygenParamaters”described.

EnterOxygenCalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”.

ADOsensorcalibrationisalwayseitheraone-pointAir(Slope)oraZero(Offset)calibration.Aone-pointslopecalibrationisdoneinairandaone-pointoffsetcalibrationisdoneat0ppbDO.Aone-pointzerodissolvedoxygencalibrationisavailablebutnotnormallyrecommendedsincezeroDOisveryhardtoachieve.

Select1pointfollowedbyeitherSlopeorZeroPtasthecalibrationtype.Press[ENTER].

EnterthevalueforPoint1includingadecimalpointandunits.Thevalueinthesecondtextlineisthevaluebeingmeasuredbythetransmitterandsensorintheunitsselectedbytheuser.Press[ENTER]whenthisvalueisstabletoperformthecalibration.

AfterthecalibrationtheslopecalibrationfactorSandtheoffsetcalibrationfactorZaredisplayed.

SelectYestosavethecalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressENTER”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.3.2 Process Calibration

EnterOxygenCalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”.

SelectProcessfollowedbyeitherSlopeorZeroPtasthecalibrationtype.Press[ENTER]

Takeasampleandpressthe[ENTER]keyagaintostorethecurrentmeasuringValue.Toshowtheongoingcalibrationprocess,AorB(dependingonthechannel)isblinkinginthedisplay.

Page 29: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 29

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

AfterdeterminingtheO2Valueofthesamplepressthe[CAL]keyagaintoproceedwiththecalibration.EntertheO2valueofthesamplethenpressthe[ENTER]keytostartcalibration.

AfterthecalibrationtheslopecalibrationfactorSandtheoffsetcalibrationfactorZaredisplayed.SelectYestosavethenewcalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

7.4 pH CalibrationForpHsensors,theM200transmitterfeaturesone-point,two-point(AutoorManualmode) orprocesscalibrationwith8presetbuffersetsormanualbufferentry.Buffervaluesreferto 25°C.Tocalibratetheinstrumentwithautomaticbufferrecognition,youneedastandardpHbuffersolutionthatmatchesoneofthesevalues.(Seesection8.2.3.2 “pH Parameters” for configuringmodes).Pleaseselectthecorrectbuffertablebeforeusingautomaticcalibration(seechapter19 “Buffer Tables”).

EnterpHCalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”.

7.4.1 One-Point Calibration

Select1pointCalibration.

DependingontheparameterizedDriftcontrol(seesection8.2.3.2 “pH Parameters”)oneofthetwofollowingmodesisactive.

7.4.1.1 Auto Mode

Placetheelectrodeinthebuffersolutionandpressthe[ENTER]keytostartthecalibration.

Thedisplayshowsthebufferthetransmitterhasrecognized(Point1)andthemeasuredvalue.

AssoonasthedriftconditionshavestabilizedthedisplaychangestoshowtheslopecalibrationfactorSandtheoffsetcalibrationfactorZ.

SelectYestosavethecalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressEnter”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

Page 30: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 30

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

7.4.1.2 Manual mode

Placetheelectrodeinthebuffersolution.Thedisplayshowsthebufferthetransmitterhasrecognized(Point1)andthemeasuredvalue.Press[ENTER]toproceed.

ThedisplayshowsnowtheslopecalibrationfactorSandtheoffsetcalibrationfactorZ.

SelectYestosavethecalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressEnter”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.4.2 Two-Point Calibration

Select2pointCalibration.

DependingontheparameterizedDriftcontrol(seesection8.2.3.2 “pH Parameters”)oneofthetwofollowingmodesisactive.

7.4.2.1 Auto Mode

Placetheelectrodeinthefirstbuffersolutionandthenpressthe[ENTER]key.

Thedisplayshowsthebufferthetransmitterhasrecognized(Point1)andthemeasuredvalue.

Assoonasthedriftconditionshavestabilized,thedisplaychangesandpromptsyoutoplacetheelectrodeinthesecondbuffer.

Placetheelectrodeinthesecondbuffersolutionandpressthe[ENTER]keytogoonwiththecalibration.

Thedisplayshowsthesecondbufferthetransmitterhasrecognized(Point2)andthemeasured value.

Page 31: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 31

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

AssoonasthedriftconditionshavestabilizedthedisplaychangestoshowtheslopecalibrationfactorSandtheoffsetcalibrationfactorZ.

SelectYestosavethecalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressEnter”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.4.2.2 Manual Mode

Placetheelectrodeinthefirstbuffersolution.Thedisplayshowsthebufferthetransmitterhasrecognized(Point1)andthemeasuredvalue.Press[ENTER]toproceed.

Placethetransmitterinthesecondbuffersolution.Thedisplayshowsthebufferthetransmitterhasrecognized(Point2)andthemeasuredvalue.Press[ENTER]toproceed.

ThedisplayshowstheslopecalibrationfactorSandtheoffsetcalibrationfactorZ.

SelectYestosavethecalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressEnter”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.4.3 Process Calibration

SelectProcessCalibration.

Takeasampleandpressthe[ENTER]keyagaintostorethecurrentmeasuringValue.Toshowtheongoingcalibrationprocess,AorB(dependingonthechannel)isblinkinginthedisplay.

AfterdeterminingthepHValueofthesample,pressthe[CAL]keyagaintoproceedwiththecalibration.

EnterthepHvalueofthesamplethenpressthe[ENTER]keytostartcalibration.

AfterthecalibrationtheslopecalibrationfactorSandtheoffsetcalibrationfactorZaredisplayed.SelectYestosavethenewcalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

Page 32: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 32

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

7.5 ORP CalibrationForORPsensors,theM200featuresone-pointcalibration.EnterORPCalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”.

7.5.1 One-Point Calibration

TheM200automaticallyperforms1-pointcalibrationfortheparameterORP.

EnterthevalueofcalibrationPoint1andthenpressthe[ENTER]keytostartcalibration.

Thevalueinthe2ndtextlineistheactualmeasuredvaluefromthesensorpriortocalibration.

ThedisplayshowstheslopecalibrationsfactorS,whichisalways1.00000andtheoffsetcalibrationfactorZ.

SelectYestosavethecalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.

Theusergetsthemessage”Re-installsensor”and”PressEnter”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.6 Ozone CalibrationOzonecalibrationisperformedaseitheraone-pointorprocesscalibration.

7.6.1 One-Point ZeroPt Calibration

EnterOzonecalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”.

Select1pointZeroPtasthecalibrationtype. Press[ENTER].

EnterthevalueforPoint1includingadecimalpoint.Ozoneisthevaluebeingmeasuredbythetransmitterandsensorintheunitssetbytheuser.Press[ENTER]whenthisvalueisstabletoperformthecalibration.

Assoonasthestabilizationcriteriahavebeenfulfilledthedisplaychanges.Thedisplayshowsthecalibrationresultforslope“S”andoffsetvalue”Z”.

Theusergetsthemessage”Re-installsensor”and”PressENTER”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

Page 33: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 33

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

7.6.2 Process Calibration

EnterOzonecalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”.AnozonesensorProcesscalibrationiseitheraSlopeorZeroPtcalibration.TheSlopecalibrationisalwaysobtainedfromacomparisoninstrumentorcolorimetrictestkit.TheZeroPtcalibrationisdoneinairorinozone-freewater.

SelectProcessfollowedbyeitherSlopeorZeroPtasthecalibrationtype.Press[ENTER].

Takeasampleandpressthe[ENTER]keyagaintostorethecurrentmeasuringvalue.

AfterdeterminingtheO3valueofthesamplepresstheckeyagaintoproceedwiththecalibration.

EntertheO3valueofthesample.Pressthe[ENTER]keytostartthecalculationofthecalibrationresults.

Afterthecalibrationtheslope”S”andtheoffsetvalue”Z”aredisplayed.SelectYestosavethecalibrationvaluesandthesuccessfulCalibrationisconfirmedonthedisplay.Theusergetsthemessage”Re-installsensor”and”PressENTER”onthedisplay.Afterpressing[ENTER]theM200returnstothemeasuringmode.

7.7 SensorVerificationEnterCalibrationmodeasdescribedinsection7.1“EnterCalibrationMode”andselectVerify.

Themeasuredsignaloftheprimaryandthesecondarymeasurementinelectricalunits areshown.

Usethe or .keytotogglebetweenChannelAandB*.

*Onlyon2-channelversion.

Page 34: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 34

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

8 Configuration(PATH:Menu/Configure)

Configure

Measurement Analog Outputs Set Points Alarm/Clean Display Hold Outputs

8.1 EnterConfigurationMode

WhileinMeasurementmode,pressthe[MENU]key.Pressthe or .keytonavigatetotheConfigure–Menuandpress[ENTER].

8.2 Measurement(PATH:Menu/Configure/Measurement)

Enterconfigurationmodeasdescribedinsection8.1“EnterConfigurationMode”.

Pressthe[ENTER]keytoselectthisMenu.Thefollowingsubmenuscannowbeselected:ChannelSetup,Comp/pH/O2andSetAveraging.

8.2.1 Channel Setup

Pressthe[ENTER]keytoselectthe”ChannelSetup”Menu.

SelectSensorTypeandpress[ENTER].

pH/ORP =pH/ORPmeasurementO2hi =Dissolvedoxygen(ppm)Cond(2) =2electrodeconductivityCond(4) =4electrodeconductivityORP =ORPmeasurementOzone =OzonemeasurementAuto: =Thetransmitterautomaticallyrecognizestheconnectedsensor

Ifyouselectaspecificparameterinsteadofauto,thetransmitteronlyacceptstheselectedparametertype.

Page 35: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 35

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

The4linesofthedisplaycannowbeconfiguredwithsensorchannel”A”or”B”foreachline ofthedisplayaswellasmeasurementsandunitmultipliers.Pressingthe[ENTER]keywilldisplaytheselectionforlinescandd.

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowill discardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswill savechangesmade.

8.2.2 Derived Measurements

Therearethreederivedmeasurementsavailableforconfigurationwithtwoconductivitysensors:%Rej(%Rejection),pHCal(CalculatedpH)andCO2Cal(CalculatedCO2).Tosetupanyof thederivedmeasurements,firstsetupthetwoprimaryconductivitymeasurements,whichwillbeusedtocalculatethederivedmeasurement.Definetheprimarymeasurementsasiftheywerestand-alonereadings.Thenthederivedmeasurementcanbedefined.

h NOTE:Itisimportanttousethesameunitsforbothmeasurements.

8.2.2.1 % Rejection measurement

Forreverseosmosis(RO)applications,percentrejectionismeasuredwithconductivitytodeterminetheratioofimpuritiesremovedfromproductorpermeatewatertothetotalimpuritiesintheincomingfeedwater.TheformulaforobtainingPercentRejectionis:[1–(Product/Feed)]X100=%RejectionWhereProductandFeedaretheconductivityvaluesmeasuredbytherespectivesensors. Figure4.1showsadiagramofanROinstallationwithsensorsinstalledforPercentRejection.

Reverse Osmosis Membrane

Feed Product

Reject

ConductivitySensor

B

ConductivitySensor

A

Figure4.1:%Rejection

h NOTE:Theproductmonitoringsensormustbeonthechannelthatwillmeasurepercentrejection.IftheproductconductivitysensorisinstalledinchannelA,thenpercentrejectionmustbemeasuredinchannelA.

Page 36: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 36

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

8.2.2.2 Calculated pH (Power Plant Applications only)

CalculatedpHmaybeobtainedveryaccuratelyfromspecificandcationconductivityvaluesonpowerplantsampleswhenthepHisbetween7.5and10.5duetoammoniaoraminesandwhenthespecificconductivityissignificantlygreaterthanthecationconductivity.Thiscalculationisnotsuitablewheresignificantlevelsofphosphatesarepresent.TheM200usesthisalgorithmwhenpHCALisselectedasameasurement.

ThecalculatedpHmustbeconfiguredonthesamechannelasspecificconductivity.Forexample,setupmeasurement”a”onchannelAtobespecificconductivity,measurement”b”onChannelBtobecationconductivity,measurement”c”onchannelAtobecalculatedpHandmeasurement”d”onchannelAtobetemperature.Setthetemperaturecompensationmodeto”Ammonia” for measurement ”a” and to ”Cation” for measurement ”b”.

h NOTE:Ifoperationgoesoutsidetherecommendedconditions,aglasselectrodepHmeasurementisneededtoobtainanaccuratevalue.Ontheotherhand,whensampleconditionsarewithintherangesnotedabove,thecalculatedpHprovidesanaccuratestandardforone-pointtrimcalibrationoftheelectrodepHmeasurement.

8.2.2.3 Calculated CO2 (Power Plant Applications only)

Carbon dioxide maybecalculatedfromcationconductivityanddegassedcationconductivitymeasurementsonpowerplantsamplesusingtablesfromASTMStandardD4519.TheM200hasthesetablesstoredinmemory,whichituseswhenunitsofCO2CALareselected.

ThecalculatedCO2measurementmustbeconfiguredtothesamechannelascationconductivity.Forexample,setupmeasurement”a”onchannelAtobecationconductivity,measurement”b”onchannelBtobedegassedcationconductivity,measurement”c”on channelAtobecalculatedCO2andmeasurement”d”onchannelBtobetemperature. Setthetemperaturecompensationmodeto”Cation”forbothconductivitymeasurements.

8.2.3 Parameter Related Settings

Additionalmeasurementandcalibrationparameterscanbesetforeachparameter;conductivity,pHandO2.

EnterConfigurationModeasdescribedinsection8.1”EnterConfigurationmode”andselectthemenuMeasurement(seesection8.2”Configuration/Measurement”).

For2-channeldevices:ThemenuComp/pH/O2canbeselectedbyusingthe or .key.Thenusetheckeytomovetothenextlineandselecttheparameter.ThechoicesareResistivity(forconductivitymeasurement),pHandO2.Press[ENTER]

For1-channeldevices:Dependingontheconnectedsensorthefollowingparameterisshowninthedisplay:Resistivity(forconductivitymeasurement),pHorO2.Press[ENTER]

Formoredetails,pleaseseethefollowingexplanationsdependingontheselectedparameter.

Page 37: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 37

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

8.2.3.1 Conductivity/Temperature Compensation

SelectResistivityandpress[ENTER].

Thetemperaturecompensationmodeforanyofthefourmeasurementlinescanbeselected.Temperaturecompensationshouldbematchedtothecharacteristicsoftheapplication. Choicesare”Standard”,”Light84”,”Std75ºC”,”Lin20°C”,”Lin25°C”,”NatH2O”,”Glycol.5”,”Glycol1”,”Cation”,”Alcohol”and”Ammonia”.

Ifcompensationmode”Lin25°C”or”Lin20°C”hasbeenchosen,thefactorfortheadjustmentofthereadingcanbemodifiedafterpressing[ENTER](Ifworkingatmeasurementline1or2press[ENTER]twice).

Pressing[ENTER]willbringuptheSaveChangesdialog.SelectingNowilldiscardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswillsavechangesmade.

Standardcompensationincludescompensationfornon-linearhighpurityeffectsaswellasconventionalneutralsaltimpuritiesandconformstoASTMstandardsD1125andD5391.

Std75°CcompensationistheStandardcompensationalgorithmreferencedto75°C.ThiscompensationmaybepreferredwhenmeasuringUltrapureWateratanelevatedtemperature.(Resistivityofultrapurewatercompensatedto75°Cis2.4818Mohm-cm.)

Lin20°Ccompensationadjuststhereadingbyafactorexpressedasa”%per°C”(deviationfrom20°C).Useonlyifthesolutionhasawell-characterizedlineartemperaturecoefficient.Thefactorydefaultsettingis2.0%/°C.

Nat H2Ocompensationincludesthecompensationto25°CaccordingtoEN27888fornaturalwater.

Lin25°Ccompensationadjuststhereadingbyafactorexpressedasa”%per°C”(deviationfrom25°C).Useonlyifthesamplehasawell-characterizedlineartemperaturecoefficient.Thefactorydefaultsettingis2.0%/°C.

Glycol.5compensationmatchesthetemperaturecharacteristicsof50%ethyleneglycolin water.Compensatedmeasurementsusingthissolutionmaygoabove18Mohm-cm.

Glycol1compensationmatchesthetemperaturecharacteristicsof100%ethyleneglycol.Compensatedmeasurementsmaygowellabove18Mohm-cm.

Cationcompensationisusedinpowerindustryapplicationsmeasuringthesampleaftera cationexchanger.Ittakesintoaccounttheeffectsoftemperatureonthedissociationofpurewaterinthepresenceofacids.

Alcoholcompensationprovidesforthetemperaturecharacteristicsofa75%solutionofisopropylalcoholinpurewater.Compensatedmeasurementsusingthissolutionmaygo above18Mohm-cm.

Light84compensationmatchesthehighpuritywaterresearchresultsofDr.T.S.Light publishedin1984.Useonlyifyourinstitutionhasstandardizedonthatwork.

Ammoniacompensationisusedinpowerindustryapplicationsforspecificconductivitymeasuredonsamplesusingammoniaand/orETA(ethanolamine)watertreatment.Ittakes intoaccounttheeffectsoftemperatureonthedissociationofpurewaterinthepresenceof thesebases.

Page 38: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 38

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

8.2.3.2 pH Parameters

SelectpHandpress[ENTER].

SelecttheDriftcontrolforcalibrationasAuto(driftandtimecriteriahavetobefulfilled)orManual(Theusercandecidewhenasignalisstableenoughtocompletecalibration) followedbytherelevantbuffertablefortheautomaticbufferrecognition.Ifthedriftrateislessthan0.8mVovera20secondintervalthenthereadingisstableandthecalibrationisdoneusingthelastreading.Ifthedriftcriteriaisnotmetwithin300secondsthenthecalibration timesoutandthemessage”Calibrationnotdone”isdisplayed.

Forautomaticbufferrecognitionduringcalibration,selectthebuffersolutionsetthatwillbeused:Mettler-9,Mettler-10,NISTTech,NISTStd,HACH,CIBA,MERCK,WTWorNone.Seesection19”Buffertables”forbuffervalues.Iftheautobufferfeaturewillnotbeusedoriftheavailablebuffersaredifferentfromthoseabove,selectNone.

STCisthesolutiontemperaturecoefficientinunitsofpH/°Creferencedto25°C(Default=0.000formostapplications).Forpurewaters,asettingof0.016pH/°Cshouldbeused.Forlowconductivitypowerplantsamplesnear9pH,asettingof0.033pH/°Cshouldbeused.ThesepositivecoefficientscompensateforthenegativetemperatureinfluenceonthepHofthesesamples.

IPistheisothermalpointvalue(Default=7.000formostapplications).Forspecificcompensationrequirementsornonstandardinnerbuffervalue,thisvaluecanbechanged.

”Fixed”allowsaspecifictemperaturevaluetobeentered.Selecting”No”meansthetemperaturegivenbythedigitalsensorconnectedtothechannelwillbeusedfortheCalibration.

Pressing[ENTER]againwillbringuptheSaveChangesdialog.SelectingNowilldiscardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswillsavechangesmade.

8.2.3.3 Dissolved Oxygen Parameters

SelectO2andpress[ENTER]

EntertheCalibrationpressure.ThedefaultvalueforCalPresis759.8andthedefaultunitismmHg.

Page 39: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 39

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

EntertheProcessPressure.TheunitsforProcPresandCalPresdonothavetobethesame.

Forthealgorithmoftheprocesscalibrationtheappliedpressure(ProcCalPres)hastobedefined.Thevalueoftheprocesspressure(ProcPres)orthecalibrationpressure(CalPres) canbeused.Chosethepressure,thatappliesduringtheprocesscalibration,resp.shouldbeusedforthealgorithmandpress[ENTER].

Thesalinityofthemeasuredsolutionandtherelativehumidityofthecalibrationgascanalsobeentered.TheallowedvaluesforRelativeHumidityareintherange0%to100%.

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowilldiscardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswillsavechangesmade.

8.2.4 Set Averaging

Pressthe[ENTER]keytoselectthisMenu.Theaveragingmethod(noisefilter)foreachmeasurementlinecannowbeselected.TheoptionsareSpecial(Default),None,Low,MediumandHigh:

None =noaveragingorfilteringLow =equivalenttoa3pointmovingaverageMedium=equivalenttoa6pointmovingaverageHigh =equivalenttoa10pointmovingaverageSpecial =averagingdependingonsignalchange(normallyHighaveragingbutLowaveraging

forlargechangesininputsignal)

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowilldiscardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswillsavechangesmade.

Page 40: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 40

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

8.3 Analog Outputs(PATH:Menu/Configure/AnalogOutputs)

Enterconfigurationmodeasdescribedinsection8.1“EnterConfigurationMode”.

GotothemenuAnalogOutputbyusingthe or .key.Pressthe[ENTER]keytoselectthisMenu,whichletsyouconfigurethe2(4for2-channelversion)AnalogOutputs.Onceanalogoutputshavebeenselected,usethe and ckeystonavigatebetweenconfigurableparameters.Onceaparameterisselected,itssettingcanbeselectedperthefollowingtable:

WhenanAlarmValueisselected,theanalogoutputwillgotothisvalueifanyalarmconditionoccurs.

Parameter SelectableValuesAout: 1,2,3*or4*(defaultis1)Measurement: a,b,c,dorblank(none)(defaultisblank)AlarmValue: 3.6mA,22.0mAorOff(defaultisoff)*Onlyon2-channelversion.

Therangecanbe4–20mAor0–20mA.

EntertheminimumandmaximumValueofAout.

IfAuto-rangewasselectedthenAoutmax1canbeconfigured.Aoutmax1isthemaximum valueforthefirstrangeonAuto-Range.ThemaximumvalueforthesecondrangeonAuto-Rangewassetinthepreviousmenu.IfLogarithmicRangewasselected,itwillalsoprompt forthenumberofdecadesas”Aout1#ofDecades=2”.

ThevaluefortheHoldmodecanbeconfiguredtoholdtheLastvalueorcanbesettoaFixedvalue.

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowill discardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswill savechangesmade.

Page 41: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 41

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

8.4 Setpoints (PATH:Menu/Configure/Setpoints)

Enterconfigurationmodeasdescribedinsection8.1“EnterConfigurationMode”.

GotothemenuSetPointsbyusingthe or .key.

Pressthe[ENTER]keytoselectthisMenu.

4(6for2-channelversion)Setpointscanbeconfiguredonanyofthemeasurements(athrud).ThepossibleSetpointtypesareOff,High,Low,Outside,Between,USP(%safetymarginbelowU.S.Pharmacopoeialimits),EPPW(%safetymarginbelowEuropeanPharmacopeialimitsforPurifiedWater)andEPWFI(%safetymarginbelowEuropeanPharmacopeialimitsforWaterforInjection).

An”Outside”Setpointwillcauseanalarmconditionwheneverthemeasurementgoesaboveitshighlimitorbelowitslowlimit.A”Between”Setpointwillcauseanalarmconditiontooccurwheneverthemeasurementisbetweenitshighandlowlimits.

Enterthedesiredvalue(s)fortheSetpointandpress[ENTER]

Thisscreenprovidestheoptiontoconfigureasetpointtobeactiveonanoverrangecondition.Selectthesetpointand”Yes”or”No”.Selectthedesiredrelaythatwillactivatewhenthesetpointalarmconditionisreached.

OverRangeOnceconfigured,theselectedrelaywillbeactivatedifasensorover-rangeconditionisdetectedontheassignedinputchannel.

DelayEnterthedelaytimeinseconds.Atimedelayrequiresthesetpointtobeexceededcontinuouslyforthespecifiedlengthoftimebeforeactivatingtherelay.Iftheconditiondisappearsbeforethedelayperiodisover,therelaywillnotbeactivated.

HysteresisEnterthehysteresisasapercentage-value.Ahysteresisvaluerequiresthemeasurementtoreturnwithinthesetpointvaluebyaspecifiedpercentagebeforetherelayisdeactivated.

Forahighsetpoint,themeasurementmustdecreasemorethantheindicatedpercentage belowthesetpointvaluebeforetherelayisdeactivated.Withalowsetpoint,themeasurementmustriseatleastthispercentageabovethesetpointvaluebeforetherelayisdeactivated.Forexample,withahighsetpointof100,whenthisvalueisexceeded,themeasurementmustfallbelow90beforetherelayisdeactivated.

HoldEntertheRelayHoldStatusof”Last”,”On”or”Off”.ThisisthestatetheRelaywillgotoduringaHold status.

State Relaycontactsareinnormalstateuntiltheassociatedsetpointisexceeded,thentherelayisactivatedandthecontactstateschange.

Select”Inverted”toreversethenormaloperatingstateoftherelay(i.e.Normallyopencontactsareinaclosedstate,andnormallyclosedcontactsareinanopenstate,untilthesetpointisexceeded).”Inverted”relayoperationisfunctionalwhenpowerisappliedtotheM200transmitter.

Page 42: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 42

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowilldiscardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswill savechangesmade.

8.5 Alarm/Clean (PATH:Menu/Configure/Alarm/Clean)

Enterconfigurationmodeasdescribedinsection8.1“EnterConfigurationMode”.

GotothemenuAlarm/Cleanbyusingthe or .key.

Pressthe[ENTER]keytoselectthismenu.

ThisMenuallowstheconfigurationofAlarmandCleanfunctionality.

8.5.1 Alarm

Toselect”SetupAlarm”,pressthe or .keysothat”Alarm”isflashing.

Usingthe and cbuttons,navigateto”UseRelay#”.Usingthe or .keys,selectarelay tobeusedfortheAlarmandpress[ENTER].

Oneofthefollowingeventsmaybealarmed:1.PowerFailure2.SoftwareFailure3.RgDiagnostics–pHglassmembraneresistance4.ChannelAdisconnected5.ChannelBdisconnected(onlyfor2-channelversion)

IfanyofthesecriteriaaresettoYesandtheconditionsforanalarmaregiven,theflashingsymbolwillbeshowninthedisplay,analarmmessagewillberecorded(seesection11.1 “Messages”;PATH:Info/Messages)andtheselectedrelaywillbeactivated.Furthermoreanalarmcanbeindicatedbythecurrentoutputifthishasbeenparameterized(seesection8.4 “Setpoints”;PATH:Menu/Configure/AnalogOutputs).

Theconditionsforalarmsare:1.thereisapowerfailureorpowercycling2.thesoftwarewatchdogperformsareset3.Rgisoutoftolerance–forexample,brokenmeasuringelectrode(pHonly)4.IfnosensorisconnectedonchannelA5.IfnosensorisconnectedonchannelB(onlyfor2-channelversion)

For1and2thealarmindicatorwillbeturnedoffwhenthealarmmessageiscleared.Itwillreappearifthepowerisconstantlycyclingorifthewatchdogisrepeatedlyresettingthesystem.

Pleasenote,thatthereareadditionalalarms,whichwillbeindicatedinthedisplay.Seesection13“Troubleshooting”.

Only for pH sensorsFor3thealarmindicatorwillgooffifthemessageisclearedandthesensorhasbeenreplacedorrepairedsothattheRgvalueiswithinspecification.IftheRgmessageisclearedandRgisstilloutoftolerancethenthealarmwillstayonandthemessagewillreappear.TheRgalarmcanbeturnedoffbygoingintothismenuandsettingRgDiagnosticstoNo.ThemessagecanthenbeclearedandthealarmindicatorwillbeoffeventhoughRgisoutoftolerance.

Page 43: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 43

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

EachAlarmRelaycanbeconfiguredineitheraNormalorInvertedstate.Inaddition,aDelay fortheactivationcanbeset.Formoreinformation,refertosection8.4 “Setpoints”.

Ifpowerfailureisturnedon,onlyinvertedstateispossibleandcannotbechanged.

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowill discardtheenteredvalues,selectingYeswillmaketheenteredvaluesthecurrentones.

8.5.2 Clean

ConfiguretheRelaytobeusedforthecleaningcycle.TheDefaultvalueisRelay1.

TheCleaningIntervalcanbesetfrom0.000to999.9hours.Settingitto0turnstheclean cycleoff.Thecleaningtimecanbe0to9999secondsandmustbesmallerthantheCleaningInterval.

SelectthedesiredRelaystate:NormalorInverted.

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowill discardtheenteredvaluesandreturntothemeasurementdisplayscreen,selectingYeswill savechangesmade.

8.6 Display(PATH:Menu/Configure/Display)

Enterconfigurationmodeasdescribedinsection8.1“EnterConfigurationMode”.

ThisMenuallowsfortheconfigurationofthevaluestobedisplayedandalsotheconfigurationoftheDisplayitself.

8.6.1 Measurement

TheDisplayhas4lines.Line1ontopandLine4onthebottom.

Selectthevalues(Measurementa,b,cord)tobedisplayedoneachlineofthedisplay.

Theselectionofthevaluesfora,b,c,dneedstobedoneunderConfiguration/Measurement/ChannelSetup.

Selectthe”ErrorDisplay”mode.Ifthisissetto”On”whenanalarmhasoccurred,themessage”Failure–PressEnter”willbedisplayedonLine4whenanalarmoccursinthenormalMeasurement mode.

Page 44: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 44

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Pressingthe[ENTER]keyagainwillbringuptheSaveChangesdialog.SelectingNowill discardtheenteredvalues,selectingYeswillmaketheenteredvaluesthecurrentones.

8.6.2 Resolution

Thismenuallowsthesettingoftheresolutionofeachdisplayedvalue.

Possiblesettingsare1,0.1,0.01,0.001orAuto.

Pressingthe[ENTER]keywillbringuptheSaveChangesdialog.

8.6.3 Backlight

ThisMenuallowsthesettingofthebacklightoptionsofthedisplay.

PossiblesettingsareOn,On50%orAutoOff50%.IfAutoOff50%isselectedthenthebacklightwillgoto50%after4minuteswithnokeypadactivity.Thebacklightwillautomaticallycomebackonifakeyispressed.

Pressingthe[ENTER]keyWillbringuptheSaveChangesdialog.

8.6.4 Name

Thismenuallowsfortheconfigurationofanalpha-numericnamewhichisdisplayedinthefirst9charactersonLines3and4oftheDisplay.Thedefaultisnothing(blank).

Ifanameisenteredonline3and/or4ameasurementcanbestilldisplayedonthesameline.

Usethe and ckeystonavigatebetweendigitstobealtered.Usingthe and .keystochangethecharactertobedisplayed.Oncealldigitsofbothdisplaychannelshavebeenentered,press[ENTER]tobringuptheSaveChangesdialog.

TheresultingdisplayinthemeasurementmodeappearsonLines3and4aheadofthemeasurements.

Page 45: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 45

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

8.7 Hold Analog Outputs(PATH:Menu/Configure/HoldOutputs)

Enterconfigurationmodeasdescribedinsection8.1“EnterConfigurationMode”.

The”Hold outputs” functionappliesduringthecalibrationprocess.Ifset”Holdoutputs”toYes,duringcalibrationprocesstheanalogoutput,theoutputrelayandUSBouptutwillbeatholdstate.Theholdstatedependsonthesetting.Forthepossibleholdsettings,seethelistbelow.Thefollowingoptionsarepossible:

HoldOutputs? Yes/No

The”DigitalIn”functionappliesallthetime.Assoonasasignalisactiveonthedigitalinputthetransmittergoestoholdmodeandthevaluesontheanalogoutput,theoutputrelaysandtheUSBoutputwillbeatholdstate.

DigitalIn1/2* State=Off/Low/High

h NOTE:DigitalIn1istoholdchannelA DigitalIn2istoholdchannelB*

*Onlyon2-channelversion.

PossibleHoldstates:Outputrelays: On/Off (Configuration/Setpoint)AnalogOutput: Last/Fixed (Configuration/Analogoutput)USB: Last/Off (System/USB)

Page 46: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 46

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

9 System(PATH:Menu/System)

System

Set Language USB Passwords Set/ClearLockout

Reset

WhileinMeasurementmodepressthekey.Pressthe. or keytonavigateto”System”–Menuandpress[ENTER].

9.1 SetLanguage(PATH:Menu/System/SetLanguage)

ThisMenuallowstheconfigurationoftheDisplaylanguage.

Thefollowingselectionsarepossible: English,French,German,Italian,Spanish,Russian,PortugueseandJapanese.Pressingthe[ENTER]keywillbringuptheSaveChangesdialog.

9.2 USB(PATH:Menu/System/USB)

ThismenuallowsconfigurationoftheUSBholdfunction.

USBHoldmaybesettoeitherOfforLastValues.Anexternalhostdevicemaypollthe M200fordata.IftheUSBHoldissettoOff,currentvaluesarereturned.IftheUSBHoldissettoLastValues,thevaluespresentatthetimetheholdconditionwasestablishedarereturned.

Press[ENTER]tobringuptheSaveChangesdialog.

Page 47: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 47

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

9.3 Passwords(PATH:Menu/System/Passwords)

ThisMenuallowsfortheconfigurationofOperatorandAdministratorPasswords,aswellassettingupaListofallowedMenusfortheOperator.TheAdministratorhasrightstoaccessallMenus. All default passwords for new transmitters are ”00000”.

ThePasswordsMenuisprotected:EntertheAdministratorPasswordtoentertheMenu.

9.3.1 Changing Passwords

Seesection9.3 “Passwords”onhowtoenterthePasswordsMenu.SelectChangeAdministratororChangeOperatorandsetthenewPassword.

Pressthe[ENTER]keyandconfirmthenewpassword.Press[ENTER]againtobringuptheSaveChangeddialog.

9.3.2 ConfiguringMenuAccessforOperator

Seesection9.3 “Passwords”onhowtoenterthePasswordsMenu.SelectConfigureOperatortoconfiguretheAccesslistfortheOperator.Itispossibletoassign/denyrightstothefollowingMenus:CalKey,QuickSetup,Configuration,SystemandService.

ChooseeitherYesorNotogive/denyaccesstotheaboveMenusandpress[ENTER]toadvancetothenextitems.Pressingthe[ENTER]keyafterconfiguringallmenuswillbringuptheSaveChangesdialog.SelectingNowilldiscardtheenteredvalues,selectingYeswillmaketheenteredvaluesthecurrentones.

Page 48: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 48

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

9.4 Set/ClearLockout(PATH:Menu/System/Set/ClearLockout)

Thismenuenables/disablestheLockoutfunctionalityofthetransmitter.TheuserwillbeaskedforapasswordbeforebeingallowedintoanymenusiftheLockoutfunctionalityisenabled.

TheLockout–Menuisprotected:EntertheAdministratorPasswordandselectYEStoenableorNOtodisabletheLockoutfunctionality.Pressingthe[ENTER]keyaftertheselectionwillbringuptheSaveChangesdialog.SelectingNowilldiscardtheenteredvalue,selectingYeswillmaketheenteredvaluethecurrentone.

9.5 Reset(PATH:Menu/System/Reset)

ThisMenuallowsaccesstothefollowingoptions:ResetSystem,ResetAnalogCal.

9.5.1 Reset System

ThisMenuallowstheresetofthemetertothefactorydefaultsettings(Setpointsoff,analogoutputsoff,etc.).Themetercalibrationandtheanalogoutputcalibrationarenotaffected.

Pressingthe[ENTER]keyaftertheselectionwillbringupaconfirmationscreen.SelectingNowillreturntheusertotheMeasurementmodewithnochanges.SelectingYeswillresetthemeter.

9.5.2 Reset Analog Calibration

ThisMenuallowsresetoftheAnalogOutputcalibrationfactorstothelastfactorycalibrationvalues.

Pressingthe[ENTER]keyaftertheselectionwillbringupaconfirmationscreen.SelectingNowillreturntheusertotheMeasurementmodewithnochanges.SelectingYeswillresettheAnalogOutputcalibration.

Page 49: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 49

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

10 Service (PATH:Menu/Service)

Service

Tech ServiceCalibrate Diagnostics

WhileinMeasurementmodepressthekey.Pressthe or .keytonavigatetothe”Service”Menuandpress[ENTER].Theavailablesystemconfigurationoptionsaredetailedbelow

10.1 Diagnostics(PATH:Menu/Service/Diagnostics)

ThisMenuisavaluabletoolfortroubleshootingandprovidesdiagnosticfunctionalityforthefollowingitems:Model/SoftwareRevision,DigitalInput,Display,Keypad,Memory,SetRelays,ReadRelays,SetAnalogOutputs,ReadAnalogOutputs.

10.1.1 Model/Software Revision

EssentialinformationforeveryServicecallisthemodelandsoftwarerevisionnumber.ThisMenushowsthetransmitterpartnumber,serialnumberandsoftwareversionnumber.

Byusingthe.keyitispossibletonavigateforwardthroughthissubmenuandgetadditionalinformationlikethecurrentversionofsoftwareimplementedonthetransmitter: MasterV_XXXXandCommV_XXXX);andtheversionofthesensorfirmware(FWV_XXX)andsensorhardware(HWXXXX).

Press[ENTER]toexitfromthisdisplay.

10.1.2 Digital Input

ThedigitalInputmenushowsthestateofthedigitalinputs.Press[ENTER]toexitfromthisdisplay.

Page 50: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 50

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

10.1.3 Display

Allpixelsofthedisplaywillbelitfor15secondstoallowtroubleshootingofthedisplay.After 15secondsthetransmitterwillreturntothenormalMeasuringmodeorpress[ENTER]toexitsooner.

10.1.4 Keypad

Forkeypaddiagnostics,thedisplaywillindicatewhichkeyispressed.Pressing[ENTER]willreturnthetransmittertothenormalMeasuringmode.

10.1.5 Memory

IfMemoryisselectedthenthetransmitterwillperformaRAMandROMmemorytest.TestpatternswillbewrittentoandreadfromallRAMmemorylocations.TheROMchecksumwill berecalculatedandcomparedtothevaluestoredintheROM.

10.1.6 Set Relay

TheSetRelaysdiagnosticmenuallowsfortheactivation/deactivationofeachRelay.

0=Normal(normallyopencontactsareopen)1=Inverted(normallyopencontactsareclosed)

Press[ENTER]toreturntoMeasurementmode.

Page 51: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 51

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

10.1.7 Read Relays

TheReadRelaysdiagnosticmenushowsthestateofeachRelayasdefinedbelow.Press[ENTER]toexitfromthisdisplay.

0=Normal1=Inverted.

10.1.8 Set Analog Outputs

ThismenuenablestheusertosetallanalogoutputstoanymAvaluewithinthe0–22mArange.Press[ENTER]toexitfromthisdisplay.

10.1.9 Read Analog Outputs

ThismenushowsthemAvalueoftheanalogOutputs.Press[ENTER]toexitfromthisdisplay.

10.2 Calibrate(PATH:Menu/Service/Calibrate)

Thismenuhastheoptionstocalibrateandtheanalogoutputsandalsoallowstheunlockingofcalibrationfunctionality.

Page 52: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 52

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

10.2.1 Calibrate Analog

SelecttheAnalogOutputyouwishtocalibrate.EachAnalogoutputcanbecalibratedat4and20 mA.

ConnectanaccuratemilliampmetertotheAnalogoutputterminalsandthenadjustthefive digitnumberinthedisplayuntilthemilliampmeterreads4.00mAandrepeatfor20.00mA.

Asthefivedigitnumberisincreasedtheoutputcurrentincreasesandasthenumberisdecreasedtheoutputcurrentdecreases.Thuscoarsechangesintheoutputcurrentcanbe madebychangingthethousandsorhundredsdigitsandfinechangescanbemadebychangingthetensoronesdigits.

Pressingthe[ENTER]keyafterenteringbothvalueswillbringupaconfirmationscreen.SelectingNowilldiscardtheenteredvalues,selectingYeswillmaketheenteredvaluesthecurrentones.

10.2.2 Calibrate Unlock

SelectthisMenutoconfiguretheCALMenu(seesection7 “Sensor Calibration”).

SelectingYesmeansthatMeterandAnalogOutputcalibrationMenuswillbeselectableunder theCALMenu.SelectingNomeansthatonlytheSensorcalibrationisavailableundertheCALMenu.Press[ENTER]aftertheselectiontodisplayaconfirmationscreen.

10.3 Tech Service(PATH:Menu/TechService)

h Note:ThisMenuisforMettlerToledoServicepersonneluseonly.

Page 53: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 53

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

11 Info(PATH:Info)

Info

Calibration Data

Messages Model/SoftwareRevision

Sensor Info

Pressingthe.keywilldisplaytheInfoMenuwiththeoptionsMessages,CalibrationDataandModel/Software Revision.

11.1 Messages(PATH:Info/Messages)

Themostrecentmessageisdisplayed.Theupanddownarrowkeysallowscrollingthrough thelastfourmessagesthathaveoccurred.

ClearMessagesclearsallthemessages.Messagesareaddedtothemessagelistwhentheconditionthatgeneratesthemessagefirstoccurs.Ifallmessagesareclearedandamessageconditionstillexistsandstartedbeforetheclearthenitwillnotappearinthelist.Forthismessagetore-occurinthelisttheconditionmustgoawayandthenreappear.

11.2 Calibration Data(PATH:Info/CalibrationData)

SelectingCalibrationDatadisplaysthecalibrationconstantsforeachsensor.Usetheupanddownarrowkeystotogglebetweenchannels”A”and”B”.

P=calibrationconstantsfortheprimarymeasurementS=calibrationconstantsforthesecondarymeasurement

Press[ENTER]toexitfromthisdisplay.

Page 54: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 54

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

11.3 Model/Software RevisionSelectingModel/SoftwareRevisionwilldisplaythepartnumber,modelandtheserialnumberofthetransmitterandinformationabouttheconnectedsensor(s).

Byusingthe.keyitispossibletonavigateforwardthroughthissubmenuandgetadditionalinformationlikethecurrentversionofsoftwareimplementedonthetransmitter: MasterV_XXXXandCommV_XXXX;andtheversionofthesensorfirmware(FWV_XXX)andsensorhardware(HWXXXX).

ThedisplayedinformationisimportantforanyServicecall.Press[ENTER]toreturntothenormal measurement mode.

11.4 Sensor InfoAfterplugginginasensor,thefollowinginformationaboutthesensorwillbeshowninthismenu.Useupanddownarrowstoscrollinthemenu.

Type: TypeofsensorCalDate*: DateofthelastadjustmentSerial-No.: SerialnumberoftheconnectedSensorPart-No.: PartnumberoftheconnectedSensor

*AfterconnectinganISMsensor

Page 55: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 55

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

12 Maintenance

12.1 FrontPanelCleaningCleanthefrontpanelwithadampsoftcloth(wateronly,nosolvents).Gentlywipethesurfaceanddrywithasoftcloth.

Page 56: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 56

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

13 TroubleshootingIftheequipmentisusedinamannernotspecifiedbyMettler-Toledo,theprotectionprovidedbytheequipmentmaybeimpaired.Reviewthetablebelowforpossiblecausesofcommonproblems:

Problem Possible Cause

Displayisblank.

– No power to M200.– Blown fuse.– LCDdisplaycontrastsetincorrectly.– Hardware failure.

Incorrectmeasurementreadings.

– Sensorimproperlyinstalled.– Incorrectunitsmultiplierentered.– Temperaturecompensationincorrectlysetor

disabled.– Sensorneedscalibration.– Sensororpatchcorddefectiveorexceeds

recommendedmaximumlength.– Hardware failure.

Measurementreadingsnotstable.

– Sensorsorcablesinstalledtooclosetoequipmentthatgenerateshighlevelofelectricalnoise.

– Recommendedcablelengthexceeded.– Averagingsettoolow.– Sensororpatchcorddefective.

Displayed isflashing.– Setpointisinalarmcondition(setpointexceeded).– Alarmhasbeenselected

(seechapter8.5Alarm/Clean)andoccurred.

Cannotchangemenusettings. – Userlockedoutforsecurityreasons.

Page 57: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 57

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

13.1 Cond (Resistivity) ErrorMessages/ Warning-andAlarmList

Alarms DescriptionWatchdogtime-out SW/Systemfault

13.2 OxygenErrorMessages/Warning-andAlarmList

Warnings DescriptionWarningO2 Slope < –90 nA SlopetoobigWarningO2 Slope > –35 nA Slope too smallWarningO2 ZeroPt > 0.3 nA ZerooffsettoobigWarningO2 ZeroPt < –0.3 nA Zero offset too small

Alarms DescriptionWatchdogtime-out SW/SystemfaultErrorO2 Slope < –110 nA SlopetoobigErrorO2 Slope > – 30 nA Slope too smallErrorO2 ZeroPt > 0.6 nA ZerooffsettoobigErrorO2 ZeroPt < – 0.6 nA Zero offset too small

13.3 pHErrorMessages/Warning-andAlarmList

Warnings DescriptionWarningpHslope>102% SlopetoobigWarningpHSlope<90% Slope too smallWarningpHZero±0.5pH OutofrangeWarningpHGlschange<0.3 Glasselectroderesistancechangedbylessthanfactor0.3WarningpHGlschange>3 Glasselectroderesistancechangedbymorethanfactor3

Alarms DescriptionWatchdogtime-out SW/SystemfaultErrorpHSlope>103% SlopetoobigErrorpHSlope<80% Slope too smallErrorpHZero±1.0pH OutofrangeErrorpHGlsRes>2000MΩ Glasselectroderesistancetoobig(break)ErrorpHGlsRes<5MΩ Glasselectroderesistancetoosmall(short)

Page 58: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 58

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

13.4 ORPErrorMessages/Warning-andAlarmList

Warnings DescriptionWarningORPZeroPt>30mV ZerooffsettoobigWarningORPZeroPt<–30mV Zero offset too small

Alarms DescriptionWatchdogtime-out SW/SystemfaultErrorORPZeroPt>60mV ZerooffsettoobigErrorORPZeroPt<–60mV Zero offset too small

13.5 Warning- and Alarm Indication on the Display

13.5.1 Warning Indication

Ifthereareconditions,whichgenerateawarning,themessagewillberecordedandcanbeselectedthroughthemenupointMessages(PATH:Info/Messages;seealsochapterMessages).Accordingtotheparameterisationofthetransmitterthehint”Failure–PressEnter”willbeshownatline4ofthedisplayifawarningoralarmhasoccurred(seealsosection8.6 “Display”;PATH:Menu/Configure/Display/Measurement).

13.5.2 Alarm Indication

AlarmswillbeshowninthedisplaybyaflashingsymbolandrecordedthroughthemenupointMessages(PATH:Info/Messages;seealsosection11.1“Messages”).

Furthermorethedetectionofsomealarmscanbeactivatedordeactivated(PATH:Menu/Configure/Alarm/Clean)foranindicationonthedisplay.Ifoneofthesealarmsoccursandthedetectionhasbeenactivated,alsoaflashingsymbolwillbeshownonthedisplayandthemessagewillberecordedthroughthemenupointMessages (PATH:Info/Messages;seealsosection11.1“Messages”).

Alarms,whicharecausedbyaviolationofthelimitationofasetpointortherange(seechapter8.4”Setpoints”;PATH:Menu/Configure/Setpoint)willalsobeshownbyaflashingsymbolandrecordedthroughthemenupointMessages (PATH:Info/Messages;seealsosection11.1“Messages”).

Accordingtotheparameterizationofthetransmitterthehint”Failure–PressEnter”willbeshownatline4ofthedisplayifawarningoralarmhasoccurred(seealsosection8.6 “Display”;PATH:Menu/Configure/Display/Measurement).

Page 59: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 59

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

14 Accessories and Spare PartsPleasecontactyourlocalMettler-ToledoSalesofficeorrepresentativefordetailsforadditionalaccessoriesandspareparts.

ForM200

Description Order no.PipeMountKitfor1/2DIN models 52 500 212PanelMountKitfor1/2DIN models 52 500 213ProtectiveHoodfor1/2DIN models 52 500 214TerminalblocksforM200,M300,M400 52 121 504

Page 60: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 60

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

15 Specifications

15.1 GeneralSpecifications

pH/ORP (incl. pH/pNa)Measurement parameters pH,mVandtemperaturepHdisplayrange – 2.00 to + 16.00 pHpH resolution Auto/0.001/0.01/0.1/1(canbeselected)pHaccuracy1) ±1digitmVrange –1500to+1500mVmVresolution Auto/0.001/0.01/0.1/1mV(canbeselected)mVaccuracy1) ±1digitTemperaturemeasuringrange –30to130°C(–22to266°F)Temperature resolution Auto/0.001/0.01/0.1/1(canbeselected)Temperatureaccuracy1) ±1digitTemperaturecompensation Automatic/ManualMax.sensorcablelength 80m(260ft)Calibration • pH:1-point(offset),2-point(slopeandoffset)or

process(offset)• ORP:1-point(offset)

1) ISMinputsignalcausesnoadditionalerror.

Amperometric oxygenMeasurement parameters Dissolvedoxygen(DO)saturationorconcentrationandtem-

peratureMeasuringcurrentrange 0 to 900 nA

DOdisplayranges • Saturation:0to500%air,0to200%O2 sat• Concentration:0.0ppb(µg/L)to50.00ppm(mg/L)

DOaccuracy ±1digitDO Resolution Auto/0.001/0.01/0.1/1(canbeselected)Polarizationvoltage –674mVTemperaturemeasuringrange –10to+80°C(+14to+176°F)Temperature resolution Auto/0.001/0.01/0.1/1(canbeselected)Temperatureaccuracy ±1digitTemperaturerepeatability ±1digitTemperaturecompensation AutomaticMax.sensorcablelength 80m(260ft)Calibration 1-point(slopeandoffset)orprocess(slopeandoffset)

Page 61: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 61

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Dissolved ozoneMeasurement parameters ConcentrationandtemperatureDisplayrangeforcurrent 0 to – 900 nAOzone displayrange Concentration0.1ppb(µg/L)to5.00ppm(mg/L)O3

Ozoneaccuracy ±1digitResolutioncurrent ±1digitTemperaturecompensation AutomaticTemperature displayrange –30to+150°C(–22to+302°F)Temperature resolution Auto/0.001/0.01/0.1/1(canbeselected)Temperatureaccuracy ±1digitMax.sensorcablelength 80 mCalibration 1-point(offset)orprocess(slopeoroffset)

Conductivity 2-e / 4-eMeasurement parameters Conductivity/resistivityandtemperatureConductivitydisplayranges 2-electrodesensor

C=0.01 0.002to500µS/cm(2000Ωxcmto500MΩxcm)

C=0.1 0.02to50,000µS/cm(20Ωxcmto50MΩxcm)Conductivitydisplayranges 4-electrodesensor

0.01 to 1000mS/cm(1.0Ωxcmto0.1MΩxcm)

Chemicalconcentrationcurves • NaCl: 0–26%@0°Cto0–28%@+100°C

• NaOH: 0–12%@0°Cto0–16%@+40°Cto0–6%@+100°C

• HCl: 0–18%@–20°Cto0–18%@0°Cto0–5%@+50°C

• HNO3: 0–30%@–20°Cto0–30%@0°Cto0–8%@+50°C

• H2SO4: 0–26%@–12°Cto0–26%@+5°Cto0–9%@+100°C

• H3PO4: 0–35%@+5°Cto+80°C

TDSranges NaCl and CaCO3

Cond/Resaccuracy1) ±1digitCond/Resrepeatability1) ±1digitCond / Res resolution Auto/0.001/0.01/0.1/1(canbeselected)Temperature displayrange –40to+200°C(–40to+392°F)Temperature resolution Auto/0.001/0.01/0.1/1(canbeselected)Temperatureaccuracy ±1digitMax.sensorcablelength • 2-esensors:90m(300ft)

• 4-esensors:80m(260ft)Calibration 1-point(offset),2-point(slopeandoffset)orprocess(slope)

Page 62: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 62

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

15.2 ElectricalSpecifications

Supplyvoltage • 100to240VAC,50to60Hz,10VA• 20to30VDC,10VA

Connectionterminal Detachablescrewterminals,appropriateforwirecrosssection0.2 to 1.5 mm2(AWG16–24)

Mains fuse 1.0Aslowblow,typeFCNumberofanalogoutputs • 4for2-channelversion

• 2for1-channelversionAnalogoutputsignals 0/4to20 mA,22mAalarm,galvanicallyisolatedfrominput

andfromearth/groundMeasurementerrorthrough analogoutputs

<±0.05mAover1to22mArange <±0.10mAover0to1mArange

Analogoutputconfiguration LinearLoad Max.500ΩHoldinput/Alarmcontact Yes/YesAlarmoutputdelay 0 to 999 sRelays 2SPDT,mechanical,ratedat250VAC,3AmpsDigitalinput • 2for2-channelversion

• 1for1-channelversionGalvanicallyisolatedfromoutputandground/earth

Display BacklitLCD,4linesKeypad 5tactilefeedbackkeysLanguages 8(English,German,French,Italian,Spanish,Portuguese,

RussianandJapanese)Digitalcommunication USB,TypeBconnector

15.3 MechanicalSpecifications

15.3.1 MechanicalSpecificationsfor1/2DINVersion

Dimensions Housing– HeightxWidthxDepth

144 x 144 x 116 mm (5.7x5.7x4.6inch)

Frontbezel– HeightxWidth

150 x 150 mm (5.9x5.9inch)

Max.depth– panel mounted

87 mm (excludesplug-inconnectors)

Weight 0.95kg(2lb)Material ABS/PolycarbonateEnclosurerating IP 65

Page 63: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 63

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

15.3.2 MechanicalSpecificationsfor1/4DINVersion

Dimensions Housing– HeightxWidthxDepth

90 x 90 x 126 mm (3.54x3.54x4.96)

Frontbezel– HeightxWidth

102 x 102 mm (4.02x4.02inch)

Max.depth– panel mounted

126 mm (excludesplug-inconnectors)

Weight 0.6kg(1.5lb)Material ABS/PolycarbonateEnclosurerating IP65(front)/IP20(housing)

15.4 EnvironmentalSpecifications

Storagetemperature –40to+70°C(–40to+158°F)Ambient temperature operatingrange

–10to+50°C(+14to+122°F)

Relativehumidity 0to95%non-condensingEMC CompliantwithEN 61326-1:2013(Industrialenvironment)

Emission:Class B,Immunity:Class AUL Installation(overvoltage)CategoryIICEmark Themeasuringsystemisinconformitywiththestatutory

requirementsoftheECDirectives.METTLERTOLEDOconfirmssuccessfultestingofthedevicebyaffixingtoittheCEmark.

Page 64: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 64

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

16 Default Tables

16.1 M200(1-channelVersion)

Parameter Sub parameter Value Unit

Alarm

relay 2diagnostics Nopower failure Nosoftware failure NoDisconnectChA NoHold mode* Lastdelay 1 sechysteresis 0state inverted

Clean

relay 1holdmode* LastInterval 0 Hrscleantime 0 Secstate normaldelay 0hysteresis 0

Language English

Passwordsadministrator 00000operator 00000

Lockout Yes/No No

AnalogOut1 a2 b

Allanalogout

mode 4–20 mAtype normalalarm offholdmode last value

Set point 1

measurement atype offhigh/lowvalue 0Relay 2

Set point 2

measurement b**type offhigh/lowvalue 0Relay 2

Set point 3

measurement _(none)type offhigh/lowvalue 0Relay _(none)

Set point 4

measurement _(none)type offhigh/lowvalue 0Relay _(none)

Relay1 clean

Relay2alarm, set point 1, set point 2

*foranalogoutputsignalifrelayisswitched **_(none)ifORPsensorisconnected

Page 65: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 65

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

16.2 M200(2-channelVersion)

Parameter Sub parameter Value Unit

Alarm

relay 2diagnostics Nopower failure Nosoftware failure NoDisconnectChA NoDisconnectChB Noholdmode* Lastdelay 1 sechysteresis 0state inverted

Clean

relay 1holdmode* LastInterval 0 Hrscleantime 0 Secstate normaldelay 0hysteresis 0

Language English

Passwordsadministrator 00000operator 00000

Lockout Yes/No No

AnalogOut

1 a2 b**3 c4 d**

Allanalogout

mode 4–20 mAtype normalalarm offholdmode last value

Set point 1

measurement atype offhigh/lowvalue 0relay 2

Set point 2

measurement ctype offhigh/lowvalue 0relay 2

Set point 3

Measurement _(none)Type offhigh/lowvalue 0relay _(none)

Set point 4

measurement _(none)type offhigh/lowvalue 0relay _(none)

Page 66: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 66

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Parameter Sub parameter Value Unit

Set point 5

measurement _(none)type offhigh/lowvalue 0relay _(none)

Set point 6

measurement _(none)type offhigh/lowvalue 0relay _(none)

Relay1 CleanRelay2 Alarm, set point 2

*foranalogoutputsignalifrelayisswitched **_(none)ifORPsensorisconnected

Page 67: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 67

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

16.3 ParameterRelatedValuesThetransmitterrecognizestheconnecteddigitalsensor,andloadsdifferentdefaultvalues,dependingonthetypeofdigitalsensor.InthischapterthedefaultvaluesarelistedifasensorisconnectedtochannelA.Ifnototherwisementioned,forthesecondchannel(2-channeldevices)thevaluesarealsovalid.

16.3.1 Conductivity

Parameter Sub parameter Value Unit

AnalogOut1

a–conductivity (resistivity)

µS/cm(MV-cm)

2 a – temperature °C

Allanalogout

mode 4–20 mAtype normalalarm offholdmode last value

Conductivityvalue 4 mA 0.1(10) µS/cm(MV-cm)

value 20 mA 10(20) µS/cm(MV-cm)

Temperaturevalue 4 mA 0 °Cvalue 20 mA 100 °C

Set point 1

measurement atype off

highvalue 00 µS/cm(MV-cm)

low value 00 µS/cm(MV-cm)

relay 1

Set point 2

measurement b(2ndchannel:c)type off

highvalue 0(0)°C2ndchannel: µS/cm(MV-cm)

low value 0(0)°C2ndchannel: µS/cm(MV-cm)

relay 1

Resolutionconductivity(resistivity)

0.01(0.01) µS/cm(MV-cm)

temperature 0.1 °C

Valuesinparentheses:Defaultvaluesifresistivityinsteadofconductivityischosen.

Page 68: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 68

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

16.3.2 Oxygen

Parameter Sub parameter Value Unit

AnalogOut1 a–oxygen % air2 a – temperature °C

Allanalogout

mode 4–20 mAtype normalalarm offholdmode last value

Oxygenvalue 4 mA 0 % airvalue 20 mA 100 % air

Temperaturevalue 4 mA 0 °Cvalue 20 mA 100 °C

Set point 1

measurement atype offhighvalue 50 % airlow value 0 % airrelay 1

Set point 2

measurement b(2ndchannel:c)type off

highvalue 0(2ndchannel:50)°C(2ndchannel: %air)

low value 0(2ndchannel:0)°C(2ndchannel: %air)

relay 1

ResolutionOxygen auto % sat

1.0 ppbTemperature 0.1 °C

Vpolarization* + 675 mVCalPres 759.8 mmHgProcPres 759.8 mmHgProcCalPres CalPresSalinity 0.0 g/KgHumidity 100 %

* not adjustable

Page 69: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 69

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

16.3.3 pH

Parameter Sub parameter Value UnitpH Buffer Mettler-9

AnalogOut1 a – pH2 a – temperature °C

Allanalogout

Mode 4–20 mAtype normalalarm offholdmode last value

pHvalue 4 mA 2 pHvalue 20 mA 12 pH

Temperaturevalue 4 mA 0 °Cvalue 20 mA 100 °C

Set point 1

measurement atype offhighvalue 12 pHlow value 0 pHrelay 1

Set point 2

measurement b(2ndchannel:c)type offhighvalue 0(2ndchannel:12) °C(2ndchannel:pH)low value 0(2ndchannel:0) °C(2ndchannel:pH)relay 1

ResolutionpH 0.01 pHTemperature 0.1 °C

Driftcontrol AutoIP 7.0 pHSTC 0.000 pH/°CFixCalTemp No

Page 70: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 70

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

16.3.4 ORP

Parameter Sub parameter Value Unit

AnalogOut1 a – ORP mVORP

2 a – none

Allanalogout

mode 4–20 mAtype normalalarm offholdmode last value

ORPvalue 4 mA –500 mV

value 20 mA +500 mV

Set point 1

measurement atype offhighvalue +500 mVlow value –500 mVrelay 2

Set point 2

measurement none(2ndchannel:c)

type offhighvalue none(2ndchannel:+500) (2ndchannel:mV)

low value none(2ndchannel:+500) (2ndchannel:mV)

relay 2Resolution ORP auto mV

16.3.5 Ozone

Parameter Sub parameter Value Unit

AnalogOut1 a – O3 ppm O3

2 a – temperature °CCalconstants read from sensor

Allanalogout

mode 4–20 mAtype normalalarm offholdmode last value

O3value 4 mA 0.00 ppb

value 20 mA 20.00 ppm

Temperaturevalue 4 mA 0 °Cvalue 20 mA 100 °C

Set point 1measurement atype offrelay 1

Set point 2measurement b(2ndchannel:c)

type offrelay 1

ResolutionO3 0.1 ppm

temperature 0.1 °C

Page 71: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 71

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

17 WarrantyMETTLERTOLEDOwarrantsthisproducttobefreefromsignificantdeviationsinmaterialandworkmanshipforaperiodofoneyearfromthedateofpurchase.Ifrepairisnecessaryandnottheresultofabuseormisusewithinthewarrantyperiod,pleasereturnbyfreightpre-paidandamendmentwillbemadewithoutanycharge.METTLERTOLEDO”sCustomerServiceDept.willdetermineiftheproductproblemisduetodeviationsorcustomerabuse.Out-of-warrantyproductswillberepairedonanexchangebasisatcost.

TheabovewarrantyistheonlywarrantymadebyMETTLERTOLEDOandislieuofallotherwarranties,expressedorimplied,including,withoutlimitation,impliedwarrantiesofmerchantabilityandfitnessforaparticularpurpose.METTLERTOLEDOshallnotbeliablefor anyloss,claim,expenseordamagecausedby,contributedtoorarisingoutoftheactsoromissionsoftheBuyerorThirdParties,whethernegligentorotherwise.InnoeventshallMETTLERTOLEDO’sliabilityforanycauseofactionwhatsoeverexceedthecostoftheitemgivingrisetotheclaim,whetherbasedincontract,warranty,indemnity,ortort(includingnegligence).

Page 72: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 72

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

18 CertificateMETTLERTOLEDOThornton,Inc.,900MiddlesexTurnpike,Building8,Billerica,MA01821USAhasobtainedUnderwritersLaboratories’listingforM200ModelTransmitters.TheybearthecULusListedmark,signifyingthattheproductshavebeenevaluatedtotheapplicableANSI/ULandCSAStandardsforuseintheU.S.andCanada.

Page 73: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 73

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

19 Buffer TablesM200transmittershavetheabilitytodoautomaticpHbufferrecognition.Thefollowingtablesshowdifferentstandardbuffersthatareautomaticallyrecognized.

19.1 Mettler-9

Temp (°C) pH of buffer solutions0 2.03 4.01 7.12 9.525 2.02 4.01 7.09 9.4510 2.01 4.00 7.06 9.3815 2.00 4.00 7.04 9.3220 2.00 4.00 7.02 9.2625 2.00 4.01 7.00 9.2130 1.99 4.01 6.99 9.1635 1.99 4.02 6.98 9.1140 1.98 4.03 6.97 9.0645 1.98 4.04 6.97 9.0350 1.98 4.06 6.97 8.9955 1.98 4.08 6.98 8.9660 1.98 4.10 6.98 8.9365 1.99 4.13 6.99 8.9070 1.99 4.16 7.00 8.8875 2.00 4.19 7.02 8.8580 2.00 4.22 7.04 8.8385 2.00 4.26 7.06 8.8190 2.00 4.30 7.09 8.7995 2.00 4.35 7.12 8.77

19.2 Mettler-10

Temp (°C) pH of buffer solutions0 2.03 4.01 7.12 10.325 2.02 4.01 7.09 10.2510 2.01 4.00 7.06 10.1815 2.00 4.00 7.04 10.1220 2.00 4.00 7.02 10.0625 2.00 4.01 7.00 10.0130 1.99 4.01 6.99 9.9735 1.99 4.02 6.98 9.9340 1.98 4.03 6.97 9.8945 1.98 4.04 6.97 9.8650 1.98 4.06 6.97 9.8355 1.98 4.08 6.98 9.8360 1.98 4.10 6.98 9.8365 1.99 4.13 6.99 9.8370 1.99 4.16 7.00 9.8375 2.00 4.19 7.02 9.8380 2.00 4.22 7.04 9.8385 2.00 4.26 7.06 9.8390 2.00 4.30 7.09 9.8395 2.00 4.35 7.12 9.83

Page 74: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 74

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

19.3 NIST Technical Buffers

Temp (°C) pH of buffer solutions0 1.67 4.00 7.115 10.32 13.425 1.67 4.00 7.085 10.25 13.2110 1.67 4.00 7.06 10.18 13.0115 1.67 4.00 7.04 10.12 12.8020 1.675 4.00 7.015 10.06 12.6425 1.68 4.005 7.00 10.01 12.4630 1.68 4.015 6.985 9.97 12.3037 1.69 4.025 6.98 9.93 12.1340 1.69 4.03 6.975 9.89 11.9945 1.70 4.045 6.975 9.86 11.8450 1.705 4.06 6.97 9.83 11.7155 1.715 4.075 6.97 9.83* 11.5760 1.72 4.085 6.97 9.83* 11.4565 1.73 4.10 6.98 9.83* 11.45*70 1.74 4.13 6.99 9.83* 11.45*75 1.75 4.14 7.01 9.83* 11.45*80 1.765 4.16 7.03 9.83* 11.45*85 1.78 4.18 7.05 9.83* 11.45*90 1.79 4.21 7.08 9.83* 11.45*95 1.805 4.23 7.11 9.83* 11.45*

*Extrapolated

19.4 NIST Standard Buffers (DIN 19266: 2000–01)

Temp (°C) pH of buffer solutions05 1.668 4.004 6.950 9.39210 1.670 4.001 6.922 9.33115 1.672 4.001 6.900 9.27720 1.676 4.003 6.880 9.22825 1.680 4.008 6.865 9.18430 1.685 4.015 6.853 9.14435 1.694 4.028 6.841 9.09540 1.697 4.036 6.837 9.07645 1.704 4.049 6.834 9.04650 1.712 4.064 6.833 9.01855 1.715 4.075 6.834 8.98560 1.723 4.091 6.836 8.96270 1.743 4.126 6.845 8.92180 1.766 4.164 6.859 8.88590 1.792 4.205 6.877 8.85095 1.806 4.227 6.886 8.833

h NOTE:ThepH(S)valuesoftheindividualchargesofthesecondaryreferencematerialsaredocumentedinacertificateofanaccreditedlaboratory.Thiscertificateissuppliedwiththerespectivebuffermaterials.OnlythesepH(S)valuesshallbeusedasstandardvaluesforthesecondaryreferencebuffermaterials.Correspondingly,thisstandarddoesnotincludeatablewithstandardpHvaluesforpracticaluse.ThetableaboveonlyprovidesexamplesofpH(PS)values for orientation.

Page 75: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 75

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

19.5 Hach Buffers Buffervaluesupto60°CasspecifiedbyBergmann&BevingProcessAB.

Temp (°C) pH of buffer solutions0 4.00 7.14 10.305 4.00 7.10 10.2310 4.00 7.04 10.1115 4.00 7.04 10.1120 4.00 7.02 10.0525 4.01 7.00 10.0030 4.01 6.99 9.9635 4.02 6.98 9.9240 4.03 6.98 9.8845 4.05 6.98 9.8550 4.06 6.98 9.8255 4.07 6.98 9.7960 4.09 6.99 9.7665 4.09* 6.99* 9.76*70 4.09* 6.99* 9.76*75 4.09* 6.99* 9.76*80 4.09* 6.99* 9.76*85 4.09* 6.99* 9.76*90 4.09* 6.99* 9.76*95 4.09* 6.99* 9.76*

*Valuescomplemented

19.6 Ciba (94) Buffers

Temp (°C) pH of buffer solutions0 2.04 4.00 7.10 10.305 2.09 4.02 7.08 10.2110 2.07 4.00 7.05 10.1415 2.08 4.00 7.02 10.0620 2.09 4.01 6.98 9.9925 2.08 4.02 6.98 9.9530 2.06 4.00 6.96 9.8935 2.06 4.01 6.95 9.8540 2.07 4.02 6.94 9.8145 2.06 4.03 6.93 9.7750 2.06 4.04 6.93 9.7355 2.05 4.05 6.91 9.6860 2.08 4.10 6.93 9.6665 2.07* 4.10* 6.92* 9.61*70 2.07 4.11 6.92 9.5775 2.04* 4.13* 6.92* 9.54*80 2.02 4.15 6.93 9.5285 2.03* 4.17* 6.95* 9.47*90 2.04 4.20 6.97 9.4395 2.05* 4.22* 6.99* 9.38*

*Extrapolated

Page 76: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 76

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

19.7 MerckTitrisole,Riedel-de-HaënFixanale

Temp (°C) pH of buffer solutions0 2.01 4.05 7.13 9.24 12.585 2.01 4.05 7.07 9.16 12.4110 2.01 4.02 7.05 9.11 12.2615 2.00 4.01 7.02 9.05 12.1020 2.00 4.00 7.00 9.00 12.0025 2.00 4.01 6.98 8.95 11.8830 2.00 4.01 6.98 8.91 11.7235 2.00 4.01 6.96 8.88 11.6740 2.00 4.01 6.95 8.85 11.5445 2.00 4.01 6.95 8.82 11.4450 2.00 4.00 6.95 8.79 11.3355 2.00 4.00 6.95 8.76 11.1960 2.00 4.00 6.96 8.73 11.0465 2.00 4.00 6.96 8.72 10.9770 2.01 4.00 6.96 8.70 10.9075 2.01 4.00 6.96 8.68 10.8080 2.01 4.00 6.97 8.66 10.7085 2.01 4.00 6.98 8.65 10.5990 2.01 4.00 7.00 8.64 10.4895 2.01 4.00 7.02 8.64 10.37

19.8 WTW Buffers

Temp (°C) pH of buffer solutions0 2.03 4.01 7.12 10.655 2.02 4.01 7.09 10.5210 2.01 4.00 7.06 10.3915 2.00 4.00 7.04 10.2620 2.00 4.00 7.02 10.1325 2.00 4.01 7.00 10.0030 1.99 4.01 6.99 9.8735 1.99 4.02 6.98 9.7440 1.98 4.03 6.97 9.6145 1.98 4.04 6.97 9.4850 1.98 4.06 6.97 9.3555 1.98 4.08 6.9860 1.98 4.10 6.9865 1.99 4.13 6.9970 2.00 4.16 7.0075 2.00 4.19 7.0280 2.00 4.22 7.0485 2.00 4.26 7.0690 2.00 4.30 7.0995 2.00 4.35 7.12

Page 77: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

Transmitter M200 77

© 03/2016 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter Multi-parameter M200Printed in Switzerland 30 300 227

Page 78: Operation Manual Multi-parameter Transmitter M200€¦ · 7.4 pH Calibration_____ 29 7.4.1 One-Point Calibration ... On the transmitter or in this manual text indicates: Caution and/or

www.mt.com/pro

METTLER TOLEDO Market Organizations

Subject to technical changes.XX/ 2015. © Mettler-Toledo GmbHPrinted in Switzerland. 52 XXX XXX

Management Systemcertified according toISO 9001 / ISO 14001

9001certified

ISO 14001certified

ISO

Sales and Service:

AustraliaMettler-Toledo Limited220 Turner StreetPort Melbourne, VIC 3207AustraliaPhone +61 1300 659 761e-mail [email protected]

AustriaMettler-Toledo Ges.m.b.H.Laxenburger Str. 252/2AT-1230 WienPhone +43 1 607 4356e-mail [email protected]

BrazilMettler-Toledo Ind. e Com. Ltda.Avenida Tamboré, 418TamboréBR-06460-000 Barueri /SPPhone +55 11 4166 7400e-mail [email protected]

CanadaMettler-Toledo Inc.2915 Argentia Rd #6CA-ON L5N 8G6 MississaugaPhone +1 800 638 8537e-mail [email protected]

ChinaMettler-Toledo International Trading(Shanghai) Co. Ltd.589 Gui Ping RoadCao He JingCN-200233 ShanghaiPhone +86 21 64 85 04 35e-mail [email protected]

CroatiaMettler-Toledo d.o.o.Mandlova 3HR-10000 ZagrebPhone +385 1 292 06 33e-mail [email protected]

Czech RepublicMettler-Toledo s.r.o.Trebohosticka 2283/2CZ-100 00 Praha 10 Phone +420 2 72 123 150e-mail [email protected]

DenmarkMettler-Toledo A /SNaverland 8DK-2600 GlostrupPhone +45 43 27 08 00e-mail [email protected]

FranceMettler-ToledoAnalyse Industrielle S.A.S.30, Boulevard de DouaumontFR-75017 ParisPhone +33 1 47 37 06 00e-mail [email protected]

GermanyMettler-Toledo GmbHProzeßanalytikOckerweg 3DE-35396 GießenPhone +49 641 507 444e-mail [email protected]

Great BritainMettler-Toledo LTD64 Boston Road, Beaumont LeysGB-Leicester LE4 1AW Phone +44 116 235 7070e-mail [email protected]

HungaryMettler-Toledo Kereskedelmi KFTTeve u. 41HU-1139 BudapestPhone +36 1 288 40 40e-mail [email protected]

IndiaMettler-Toledo India Private LimitedAmar Hill, Saki Vihar RoadPowaiIN-400 072 MumbaiPhone +91 22 2857 0808e-mail [email protected]

IndonesiaPT. Mettler-Toledo IndonesiaGRHA PERSADA 3rd FloorJl. KH. Noer Ali No.3A, Kayuringin JayaKalimalang, Bekasi 17144, IDPhone +62 21 294 [email protected]

ItalyMettler-Toledo S.p.A.Via Vialba 42IT -20026 Novate MilanesePhone +39 02 333 [email protected]

JapanMettler-Toledo K.K.Process Division6F Ikenohata Nisshoku Bldg.2-9-7, IkenohataTaito-kuJP-110-0008 TokyoPhone +81 3 5815 5606e-mail [email protected]

MalaysiaMettler-Toledo (M) Sdn BhdBangunan Electroscon Holding, U 1-01Lot 8 Jalan Astaka U8/84Seksyen U8, Bukit JelutongMY-40150 Shah Alam SelangorPhone +60 3 78 44 58 88 [email protected]

MexicoMettler-Toledo S.A. de C.V.Ejército Nacional #340Polanco V SecciónC.P. 11560MX-México D.F.Phone +52 55 1946 0900e-mail [email protected]

NorwayMettler-Toledo ASUlvenveien 92BNO-0581 Oslo NorwayPhone +47 22 30 44 90e-mail [email protected]

PolandMettler-Toledo (Poland) Sp.z.o.o.ul. Poleczki 21PL-02-822 WarszawaPhone +48 22 545 06 80e-mail [email protected]

RussiaMettler-Toledo Vostok ZAOSretenskij Bulvar 6/1Office 6RU-101000 MoscowPhone +7 495 621 56 66e-mail [email protected]

SingaporeMettler-Toledo (S) Pte. Ltd.Block 28Ayer Rajah Crescent #05-01SG-139959 SingaporePhone +65 6890 00 [email protected]

SlovakiaMettler-Toledo s.r.o.Hattalova 12/ASK-83103 BratislavaPhone +421 2 4444 12 20-2e-mail [email protected]

SloveniaMettler-Toledo d.o.o.Pot heroja Trtnika 26SI -1261 Ljubljana-DobrunjePhone +386 1 530 80 50e-mail [email protected]

South KoreaMettler-Toledo (Korea) Ltd.1&4F, Yeil Building 21Yangjaecheon- ro 19-gilSeoCho-GuSeoul 06753 KoreaPhone +82 2 3498 3500e-mail [email protected]

SpainMettler-Toledo S.A.E.C /Miguel Hernández, 69-71ES-08908 L’Hospitalet de Llobregat(Barcelona)Phone +34 902 32 00 23e-mail [email protected]

SwedenMettler-Toledo ABVirkesvägen 10Box 92161SE-12008 StockholmPhone +46 8 702 50 00e-mail [email protected]

SwitzerlandMettler-Toledo (Schweiz) GmbHIm Langacher, PostfachCH-8606 GreifenseePhone +41 44 944 47 60e-mail [email protected]

ThailandMettler-Toledo (Thailand) Ltd.272 Soi Soonvijai 4Rama 9 Rd., BangkapiHuay KwangTH-10320 BangkokPhone +66 2 723 03 [email protected]

TurkeyMettler-Toledo TürkiyeHaluk Türksoy Sokak No: 6 Zemin ve 1.Bodrum Kat 34662 Üsküdar-Istanbul, TRPhone +90 216 400 20 20e-mail [email protected]

USAMETTLER TOLEDOProcess Analytics900 Middlesex Turnpike, Bld. 8Billerica, MA 01821, USAPhone +1 781 301 8800Freephone +1 800 352 8763e-mail [email protected]

VietnamMettler-Toledo (Vietnam) LLC29A Hoang Hoa Tham Street, Ward 6Binh Thanh DistrictHo Chi Minh City, VietnamPhone +84 8 [email protected]

Mettler-Toledo GmbH, Process AnalyticsIm Hackacker 15, CH-8902 Urdorf, SwitzerlandPhone + 41 44 729 62 11, Fax +41 44 729 66 36

Subjecttotechnicalchanges.©Mettler-ToledoGmbH,ProcessAnalytics03/2016 Printed in Switzerland. 30 300 227