Operating Manual SunCool for SUNTEST® CPS+

- 2 -

CopyrightThis operating manual is protected by copyright. These rights, especially reprinting, photomechanical or digital processing or reproduction even in part are only allowed with written permission of Atlas Material Testing Technology GmbH. This condition does not cover the reproduction of the manual for internal use. The contents of the operating manual are subject to change at any time without notice. The German version of this operating manual is binding for translations into foreign languages.Atlas Material Testing Technology GmbH • 63589 Linsengericht • Germany

TrademarksSUNTEST® CPS+ is a registered trademark of Atlas Material Testing Technology GmbH. All other trademarks used inthis operating manual are the exclusive property of the manufacturers concerned.

Page 3: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 3 -

Operating Manual SunCool for SUNTEST® CPS+

Contents Page 1 Instructions for safe operation ...................................................................................................................... 51.1 Explanation of the symbols .......................................................................................................................... 71.2 General safety instructions ........................................................................................................................... 81.3 Safety instructions for coolants .................................................................................................................. 10

2 Delivery of the instrument ............................................................................................................................ 122.1 Packaging .................................................................................................................................................. 122.2 Scope of delivery ........................................................................................................................................ 12

3 Requirements for the installation site ......................................................................................................... 133.1 Room climate ............................................................................................................................................. 133.2 Room ventilation ........................................................................................................................................ 133.3 Installation .................................................................................................................................................. 133.4 Heat sources .............................................................................................................................................. 143.5 Transport .................................................................................................................................................... 143.6 Effects of vibration ...................................................................................................................................... 143.7 Installation and space requirements .......................................................................................................... 15

4 Description of the instrument ...................................................................................................................... 164.1 SunCool Components ................................................................................................................................ 16

5 Functional description .................................................................................................................................. 175.1 SunCool funktionality ................................................................................................................................. 175.2 Setting the test parameters ........................................................................................................................ 17

6 SunCool start up ........................................................................................................................................... 186.1 Preparatory measures ................................................................................................................................ 186.2 Starting for the first time ............................................................................................................................. 196.3 Checking the instrument components ........................................................................................................ 196.4 Closing the coolant circuit .......................................................................................................................... 206.5 Indoor installation of the liquefier unit ......................................................................................................... 216.6 Outdoor installation of the liquefier unit ...................................................................................................... 226.7 Main power supply ..................................................................................................................................... 25

7 Operation ....................................................................................................................................................... 267.1 Menu structure of the device controller ...................................................................................................... 267.2 Setting the SunCool accessory .................................................................................................................. 267.3 Starting the cooling process ....................................................................................................................... 27

8 Taking out of operation ................................................................................................................................. 288.1 Switching off the instrument ....................................................................................................................... 288.2 Switching off the instrument in an emergency ........................................................................................... 288.3 Decommissioning of instrument ................................................................................................................. 28

9 Troubleshooting ............................................................................................................................................ 309.1 Error messages and troubleshooting ......................................................................................................... 30

10 Maintenance .................................................................................................................................................. 3010.1 Inspection ................................................................................................................................................... 3010.2 Repair ......................................................................................................................................................... 3010.3 Maintenance ............................................................................................................................................... 3110.4 Cleaning ..................................................................................................................................................... 31

11 Technical Data ............................................................................................................................................... 32

12 Accessories .................................................................................................................................................. 33

13 Declaration of conformity ............................................................................................................................ 34

14. Notes ............................................................................................................................................................. 35

Page 4: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 4 -

List of figures Page

Fig. 1 Installation and space requirements .................................................................................................. 15Fig. 2 Installation and space requirements .................................................................................................. 15Fig. 3 SunCool components ........................................................................................................................... 16Fig. 4 Test chamber and black standard temperatures ............................................................................... 17Fig. 5 Preparatory measures .......................................................................................................................... 18Fig. 6.1 Closing the coolant circuit ................................................................................................................... 20Fig. 6.2 Connecting the waste water pump ..................................................................................................... 20Fig. 6.3 Inside installation ................................................................................................................................. 21Fig. 6.4 Outdoor installation .............................................................................................................................. 22Fig. 6.5 Pipe connections .................................................................................................................................. 23Fig. 6.6 Opening the coolant circuit ................................................................................................................. 24Fig. 6.7 Incoming power connection ................................................................................................................ 25Fig. 7 Air temperature ..................................................................................................................................... 27Fig. 8 Cleaning the air filters .......................................................................................................................... 31

Page 5: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 5 -

Operating Manual SunCool for SUNTEST® CPS+

Instructions for the user:

This operating manual describes the SunCool accessory (chiller) for the SUNTEST® CPS+ weathering instrument. Please note that certain work may only be performed by appropriately qualified personnel.

The following persons can be considered appropriately qualified:• Persons who have acquired their knowledge in special training courses.• Persons who have been trained in the operation and operating possibilities of the SunCool and the SUNTEST CPS+ instrument based on this operating instructions.• Persons who are capable of assessing the activity they perform based on their working experience and instruction in the relevant safety conditions and of recognizing possible hazards at work.Service:• Work on the electrical components may only be carried out by qualified personnel.

Cleaning:• Cleaning work on the accessory may only be carried out by trained personnel who have been instructed based on these operating instructions.

Please read these operating instructions carefully before using the SunCool for the first time in order to fullyutilize the equipment and avoid possible damage to the SunCool.

Should a particular problem arise which you feel is not adequately addressed in these operating instructions,we urge you to contact your supplier for your own safety.

1Instructions for safe operation

Page 6: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 6 -

1Instructions for safe operation

Instructions for the user:

The SunCool is constructed according to state-of-the-art technology and is safe to operate. However, this instrumentcould be hazardous if operated by untrained personnel or utilized in a manner for which it was not intended.

Commissioning and maintenance work:• Commissioningmay only be carried out by an authorized and qualified air conditioning technician• Electrical components may only be changed by a qualified electrician

Briefing and prevention of accidents:• For personnel who work on and with this instrument, the owner must create written instructions in a compre-

hensive form and in the language of the employees based on this operating instruction (FRG: Directive for the Prevention of Accidents, UVV BGV A1).

• The work health and safety regulations of the trade cooperative associations (BGV A 3) must be observed.• Instruct the operating and cleaning personnel in the operation and care of the instrument based on these instruc-

tions.• Unauthorized modifications or conversions to the instrument are not permitted for safety reasons.

Warranty:Atlas MTT GmbH guarantees the safety and functional capability of the instrument only on the conditions that:• only original spare parts or accessories approved by Atlas MTT are used.• inspections and maintenance work are performed according to the given time intervals.

Validity of the contents of the manual:• The contents of these operating instructions are subject to change at any time without notice.• The German version of these operating instructions is binding for translations into foreign languages.

Keep these operating instructions in a safe place near to the instrument in order to consult safety instructions and important operating information at all times.

Atlas Material Testing Technology GmbHVogelsbergstr. 2263589 Linsengericht / Germany(p) + 49-6051-707-140(f) + 49-6051-707-149email: [email protected]

Page 7: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 7 -

Operating Manual SunCool for SUNTEST® CPS+

1Instructions for safe operation

Symbols in the operating manual: WARNING!

Disregarding this information can result in serious injury or death.

CAUTION!

Disregarding this information can result in slight to severe injury and property damage.

NOTE!

Provides application tips and useful information.

Warns against hazardous solvents.

Warns against hot surfaces / burn hazards

Warns against electrical shock

Warns against toxic materials.

Symbols on the instrument:

WARNING AGAINST A DANGEROUS PLACE!

Caution! See the operating manual.

CE conformity mark

REFERENCE TO DISPOSAL DIRECTIVE (WEEE)!

The disposal of this product must comply with the EC directive 2012/19/EU (updated version) with regard to the used electrical and electronic equipment (WEEE).

1.1 Explanation of the symbols

Page 8: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 8 -

1Instructions for safe operation

1.2 General safety instructions

Proper use:• The SunCool instrument serves to supply the SUNTEST CPS+ with a constantly controlled cooling air.• The instrument is suitable for continuous operation.• The SunCool instrument is tested for electromagnetic compatibility and suitable for installation in an industrial

environment.• The SunCool can be operated both in closed rooms and outdoor installations.• In order to keep a constant room temperature for operation in closed rooms, an air extractor (technical ventila-

tion) with a suction capacity of approx. 700 W / 500 nfVh can be installed in the room.

Improper use:• The SunCool instrument is unsuitable for operation in areas where there is a risk of fire or explosion.• The SunCool may not be employed to air condition working or waiting areas or to refrigerate foods or tempera-

ture-sensitive substances.

The device conforms to the safety requirements set forth in:

• DIN EN 378, Part 1 DIN EN ISO 12100-1 and 2 / Safety of Machinery 2004-04• DIN EN 14121-1 Safety of Machinery, Risk Analysis, 2007-12BetrSichV, DIN VDE 0100, DIN EN 378 Part1 EMC

Law

Page 9: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 9 -

Operating Manual SunCool for SUNTEST® CPS+

1Instructions for safe operation

1.2 General safety instructions

Main power supply connection:The SunCool refrigeration device can be operated from a power supply of:• 230 V ± 10%, 50 / 60 Hz

The instrument is prepared for the main power supply according to the order on delivery.

Disposal:

OBSERVE THE REGULATIONS FOR DISPOSAL!

The manufacturers‘ obligations to take back the equipment in accordance with the respective national version of the EC directive 2012/19/EU apply since March 24, 2006.

RohS conformity: The SunCool meets the requirements of the 2011/65/EU RohS directive (restriction of the use of certain hazardous substances in electrical and electronic equipment) for equipment group 9.

This directive does not apply for equipment which was designed exclusively for research and development purposes and is only provided at intra-company level.

Therefore, the SUNTEST CPS+ instrument does not come under the validity of the above mentioned directive.

Disposal of the packaging:Please dispose of the packaging materials according to the valid disposal regulations. A list of used packaging materials can be found in chapter 2, par. 2.1 “Packaging.”

CAUTION – special waste

The device is operated with the R410A coolant. This coolant may not be poured down the drain or filled into containers and thrown in with domestic waste. The coolant must be disposed of by a company specializing in coolants!

Page 10: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 10 -

1Instructions for safe operation

1.3 Safety instructions for coolant

Danger specifications:The R410A coolant is classified as a chemical with low toxicity. The maximum workplace concentration (MAK value) should not exceed the following limit values:

Difluormethane CH2F2 and pentafluorethane C2HF5: max. 1000 ppm and 3500 mg/cm³.

Handling of R410A may be dangerous in the case of:• Direct contact with the skin • Prolonged inhalation of coolant fumes • Reaction of the chemical with intense heat and open flames

CAUTION – danger of burns!

Escaping coolant cools down rapidly. Contact with the R410A can lead to cold burns and frostbite. Protective overalls, gloves, face mask and glasses must be worn for all work on the chiller.

WARNING – Intoxication

If large amounts of R410A escape into the room atmosphere, fumes are produced that can lead to intoxication or disturbance of the heart rate which could be fatal. Take immediate safety precautions in case of heavy leakage from the coolant circuit.• Air the room immediately, if safe to do so • Leave the room and secure the entrances to prevent anyone else from entering the room • Post signs alerting people to the possible danger of fumes! • Inform the safety service or fire department!

CAUTION – dangerous chemical reactions!

R410A reacts with open lights or when in contact with very hot surfaces and will produce toxic or irritative substances.Do not use open lights in the vicinity of the SunCool! Do not expose the SunCool to excessive heat!

Page 11: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 11 -

Operating Manual SunCool for SUNTEST® CPS+

1Instructions for safe operation

1.3 Safety instructions for coolant

First aid measures:The first aid measure is only an emergency measure or immediate help in the case of minor injuries; Victims should be placed in a stable side position in case of unconsciousness.

A physician should always be consulted in case of prolonged contact with the coolant. Inform the physician that the injured person was exposed to R410A coolant (the physician may not administer any drugs of the adrenaline-ephedrine group).

Contaminated clothing: Remove contaminated clothing immediately if they are not stuck to the skin. Clothing which is stuck to the skin may only be removed under medical supervision.

Skin contact with the coolant:Thaw out contact points (burns) on the skin. Wash the affected area of the skin with generous amounts of warm water.

Retina contact with the coolant:Rinse eye carefully with large amounts of water. Keep eyelids open for at least 10 minutes. Consult a physician.

Inhalation of coolant fumes:Inform an emergency doctor immediately. Get the victim into a well-heated room and supply with fresh air; have a trained person treat the victim with oxygen if he has inhaled large amounts of fumes. In case of intoxication symptoms have a trained person respirate artificially or with oxygen and apply heart massage.

Ingestion of coolant:If the victim is conscious, rinse out their mouth and throat and give them water or a warm drink.Inform an emergency doctor immediately. Get the victim into a stable side position to reduce the risk of vomiting. Do not induce vomiting.

Clearing up small amounts of R410A:Put on protective clothing and seal the leak in the coolant circuit if possible. Escaping coolant turns into a gas which gathers near the floor as it is heavier than air. Therefore, ventilate the operations room well. Open several doors or exits to the outside if possible to create a draft.

Page 12: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 12 -

2Delivery of the instrument

2.1 Packaging

The SunCool instrument is delivered in a stable packag-ing crate. All packing materials can be separated and are recyclable.

• Corrugated cardboard outer packaging • Polyethylene foil (PE) • Polyethylene foam (PE) • Plastic hooping bands (PP) • Wooden pallet

2.2 Scope of delivery

The SunCool instrument is not delivered in a ready-to-operate condition. The components must be assembled before operating.

The scope of delivery includes:

Accessories:• Liquefier unit • Evaporator unit• Hose connection coolant supply (2.5 m)• Hose connection coolant return (2.5 m)• Data connection cable• Dummy plug for operation with Suntest CPS+

Technical documentation:• Operating manual

NOTE – hose connections

The coolant circuit is a closed system.A bending radius of at least 20 cm must be kept when positioning the coolant hoses. Kinked hoses will impair operation of the SunCool unit.

Page 13: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 13 -

Operating Manual SunCool for SUNTEST® CPS+

3Requirements for the installation site

3.1 Room climate

Climatic requirements for the operating room:In continuous operation together with the SUNTEST CPS+ there is a constant change in the room climate due to the hot air emitted by the radiator cooling and the liquefier unit as well as the cold air emitted by the evaporator unit. Therefore, the device may only be installed in adequately aired, dust-free rooms or must be connected to a suitable technical ventilation unit.Ambient conditions:• Room temperatures of 18 °C to 25 °C • Relative humidity of 50% max.

3.2 Room ventilation

The coolant is processed in the cooling system in different aggregate states. The coolant is heavier than air in vapor form. The room must therefore have sufficient ventilation facilities near the floor. The doors of the room must have ventilation slits near to the floor.

3.3 Installation

• Evaporator dimensions (W x D x H): 35 cm x 32 cm x 33 cm• Evaporator weight: approx. 7 kg• Liquefier unit dimensions (W x D x H): 70 cm x 60 cm x 35 cm• Liquefier unit weight: 30 kg• Place the evaporator unit on a stable laboratory bench

with a non-flammable surface and align the lab bench horizontally

NOTE –

minimum load capacity of the lab bench

The total dead weight of the SUNTEST® CPS+ and a SunCool is approx. 40 kg. Please note the minimum load capacity of your lab bench.

CAUTION – overheating of the device!

If the air feed and discharge (of the SUNTEST CPS+) is disturbed, the device overheats and can be damaged! Keep the following minimum distances from adjacent walls or objects when installing:

Front: 70 cmRear: 50 cmTop: 70 cmSides: 50 cm

Page 14: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 14 -

3.5 Transport

NOTE – Transport!

The instrument may only be transported by experienced service personnel who have suitable tools.

3.6 Effects of vibrations

The compressor produces vibrations during the refriger-ant liquefaction process. The refrigeration device should therefore not be set up near machinery or equipment that is sensitive to vibrations.

3.4 Heat sources

Heat sources employing open flames or which utilize extremely hot surfaces may not be employed to heat the installation area.

CAUTION – dangerous chemical reactions!

Leaking coolant can produce toxic or irritant substances in reaction with open lights or hot surfaces.

The SunCool should not be installed in the immediate vicinity of heat sources or exposed to direct sunlight be-cause the energy consumption of the cooling system is increased by the effect of heat.

3Requirements for the installation site

Page 15: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 15 -

Operating Manual SunCool for SUNTEST® CPS+

3.7 Installation and space requirements

NOTE – load at the installation site

If several devices are installed in one place, the static load capacity of the installation site must be ensured.

Installing the device (Fig. 1):Place the SunCool either on the floor or on a stable, non-flammable lab bench and align horizontally.

The liquefier unit of the chiller must be installed at the following distances from adjacent surfaces regardless of whether it is installed inside or outdoors (the stands shown in the bottom of Figure 1 are suggested for mounting in an outdoor installation): Fig. 2:

A (to the rear) 15 cmB (to the front) 60 cmC (to the left) 15 cmD (to the right, valve side) 40 cmE (to the top) min. 2 m

Weights and dimensions of the chiller: Liquefier unit: 30 kgEvaporator unit: approx. 7 kg

Dimensions (W x D x H in cm):Liquefier unit 70 x 60 x 35Evaporator unit: 35 x 32 x 33

3Requirements for the installation site

De

uts

ch

Air extractor at least 5 cm

Air emissionat least5 cm at least

20 cm

Valve sideat least 25 cm

at least2 m

at least2 m

Floorat least 5 cmAir extractor

Obstruction above

Air e

mis

sio

n

Air extractor

Concrete orsimilarmaterial

approx. 10 cm

Anchor bolts(4 pcs.)

approx. 40 cm

Obstr

uction

appro

x. 10 c

m

A

BC

D

E

Fig. 1

Fig. 2

Page 16: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 16 -

4Description of the instrument

4.1 SunCool components

The SunCool instrument equipment is pre-assembled and fully functional for installation indoors and consists of the following components:

• Evaporator unit 2 • Air inlet aperture with dust filter 1 for heat exchanger • Air inlet aperture 3 for cooled air, connection side for

SUNTEST® CPS+weathering instruments • Drain 7 for condensation • Spacer 6

• Liquefier unit B • Air inlet aperture A for fan • Mains power supply 9 • Coolant hose connections 8 (supply and return) • Two control cables 4 for connecting the liquefier unit

and the evaporator unit • Interface cable 5 for connecting the SunCool to the

SUNTEST weathering instrument (only for CPS+ ver-sion)

Fig. 3

9

43 5

6

B A 8

7

21

Page 17: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 17 -

Operating Manual SunCool for SUNTEST® CPS+

5.1 SunCool functionality

The SunCool unit allows the Suntest® CPS+ instrument to supply cooled air to the test chamber. To do this, the evaporator unit 2 (Fig. 3, page 16) takes in room air and feeds it into the test chamber at the temperature reduced by the heat exchanger.

NOTE – freezing!

Freezing is improbable in the system butplease contact your technical service representative in the event that it does occur. The SunCool instrument is switched off by a thermostat to avoid freezing and switches back on after thawing. The values for the test chamber temperature may therefore deviate from the set values during freezing. Fig. 4

5Functional description

0

5

10

15

20

25

30

35

200 300 400 500 600 700 800

E [W/m²]

BST/CHT Values Suntest CPS+ (II) with SunCool[Daylight red. IR Filter]

BST

CHT

0

5

10

15

20

25

30

35

200 300 400 500 600 700 800

E [W/m²]

BST/CHT Values Suntest CPS+ (II) with SunCool[Windowglass red. IRFilter]

BST

CHT

CHT,BST [°C]

CHT,BST [°C]

5.2 Setting the test parameters

The minimum test chamber temperature (CHT) and black standard temperature (BST) in the SUNTEST are directly dependent on the ambient temperature. The higher the ambient temperature, the higher the blow-in temperature of the chiller and, as a result, the minimum CHT and BST.

The diagram below (Fig. 4) shows the approximately achievable test chamber and black standard temperatures.

Page 18: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 18 -

6.1 Preparatory measures

The following steps are necessary to start up the equip-ment combination of the SunCool instrument with the SUNTEST instrument:1. Remove the filter mat from the air inlet aperture on the

SUNTEST instrument.2. Insulate the housing surface of the SUNTEST CPS+

instrument (optional).3. Connect the drain hose to the evaporator unit. Connect the micro-pump for the waste water to the

connection nozzle of the evaporator unit optionally.4. Install the liquefier unit. Install the liquefier unit optionally

outside and position the electric cable for the power supply as well as the connecting cables between the liquefier unit and the evaporator unit.

5. Install the SUNTEST CPS+ instrument and place the evaporator unit on the SUNTEST instrument.

6. Connect the liquefier unit and the SUNTEST instrument using the interface cable.

7. Connect the drain hose of the evaporator unit to the waste water tank or sewer drain.

8. Connect the drain hose of the micro-pump to the waste water tank or sewer drain.

9. Connect the liquefier unit to the power supply.10. Connect the SUNTEST instrument to the power supply

and start.

Insulating the surfaces of the SUNTEST instrument(Fig. 5)The three insulating pads prevent condensation from forming on the housing surfaces in the areas:• air inlet aperture 2 of the SUNTEST CPS+ 1

NOTE – adhesion:

The insulating pads are strongly adhesive and, once applied, are very difficult to remove.Therefore, mark the position of the insulating pads before applying them to the housing surfaces.

Attaching insulating pads:

1. Clean the housing surface with a commercially available cleaning agent. The adhesion areas must be dry and free from grease and dust.

2. Mark the position on the right and rear sides of the instrument and on the inside top of the housing, pull off the backing foil, position the insulating pads and press firmly.

Fig. 5

6SunCool start up

1

2

Page 19: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 19 -

Operating Manual SunCool for SUNTEST® CPS+

6.2 Starting for the first time

NOTE – starting for the first time

The instrument may only be started for the first time by an Atlas technical service representative or authorized Atlas service provider.

6.3 Checking the instrument components

6SunCool start up

WARNING – electrical shock!

Touching live parts can lead to life threaten-ing electrical shocks. Disconnect the device from the power supply before testing! • Pull the main power supply plug out of the

receptacle and secure against reconnecting • Check whether the instrument is voltage

free

Only instrument parts which are in perfect condition ensure the functional reliability of the SunCool and the SUNTEST® CPS+ device. Check the instrument compo-nents and parts listed below for damage and operation before starting or before every new test run. Damaged or worn parts should not be used.

Checklist:The following parts must be checked before starting:• Seat and tightness of the quick-release coupling • Condition of the coolant lines • Condition of plugs and power cables

Page 20: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 20 -

6SunCool start up

6.4 Closing the coolant circuit

Connecting the condensation drain of the evaporator unit (Fig. 6.1):The condensation produced in the evaporator unit in the heat exchanger is discharged at the drain spout 1.The condensation can be collected in a waste water tank 3 via a hose connection 2. Alternatively, the conden-sation drain can be connected to a sewer drain on site.This is provided that the evaporator unit is installed at a height which allows drainage by gravity.

Connecting the waste water pump (Fig. 6.2):If it is necessary to pump off the condensation, a micro pump (delivery volume 12 l/h) can be connected between the connection spouts and the drain hose. The micro-pump is equipped with an internal floating contact and only switches on when there is a sufficient amount of condensation.1. Connect the micro-pump 2 to the hose nozzle of the

evaporator unit with the length of hose 1 provided.2. Lay the drain hose 3 of the micro-pump into the sewer

drain or waste water tank.3. Insert the plug 4 of the micro-pump into a properly

grounded and fused receptacle.4. Make sure that no tensile and tractive forces act on

the power cable.

WARNING – electrical shock!

Touching live parts can lead to life threatening electrical shock.Check plugs and power cables of the micro-pump for damage before connecting to the main power supply.Damaged parts must not be used for main power supply connection!

Fig. 6.1

Fig. 6.2

2

1

3

1

3

2

4

Page 21: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 21 -

Operating Manual SunCool for SUNTEST® CPS+

The liquefier unit can be installed separately on the floor or in combination with the SUNTEST® CPS+ on a workbench or sub frame with shelves. If the equipment is placed on a device, no other equipment which is sensitive to vibrations should be placed on this surface. The compressor of the liquefier unit vibrates during liquefaction of the coolant and could have a damaging effect on equipment which is sensitive to vibrations.

NOTE – condensation

Colder indoor temperatures may result in small amounts of condensation on the liquefier unit when the chiller starts. Please bare in mind when choosing the installation site that dripping condensation can cause damage. The SUNTEST CPS+ and the evaporator trough should, therefore, always be placed on a surface above the liquefier unit.

Installing the liquefier unit (Fig. 6.3)1. Install the SUNTEST CPS+ instrument 2.2. Install the liquefier unit 5 of the chiller, taking into

account that the length of the hose/cable connections 4 to the evaporator unit 1 is 2.5 m. Assure proper distances are kept from adjacent surfaces as noted in Figure 3. Align the liquefier unit horizontally and check the stability of the unit.

3. Remove the filter mat 3 from the air inlet aperture on the SUNTEST instrument. Position the evaporator unit of the chiller close to the air inlet aperture of the weathering instrument. The instrument connection requires no other connecting elements.

4. Lay the hose connections 4 between the units so that a bending radius of at least 20 cm is kept at bends.

5. Make sure that the hoses or connecting cables are not exposed to tensile or tractive forces.

NOTE – return coolant

If the liquefier unit was tilted or turned upside down during transport or installation, it must remain in its correct posi-tion for two hours prior to starting the cooler to allow the coolant to collect completely in the liquefier unit.

6SunCool start up

2

31

5 4

Fig. 6.36.5 Indoor installation of the liquefier unit

Page 22: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 22 -

6SunCool start up

6.6 Outdoor installation of the liquefier unit

NOTE – installation work

The outdoor installation of the liquefier unit, the position-ing of supply and return cables for the coolant and the positioning of the power and connecting cables should only be performed by authorized and qualified cooling and air conditioning technicians and electricians!

WARNING – malfunction!

When installing outdoors, the chiller connec-tion must be protected from rain and splash-ing water to avoid malfunctions or electrical shocks!

Assembly position:In outdoor installation, the liquefier unit should be mounted in a place protected from the sun and rain. For protection against collecting water or snow, the mounting position should be approx. 20 to 30 cm above the ground or the floor or the liquefier unit should be placed on an appro-priately high base. The safety distances (chap. 3.7) from adjacent surfaces should be maintained on all sides.

Condensation forms when operating the liquefier unit at cold outdoor temperatures. Therefore, an appropriate as-sembly position must be utilized in order to avoid dripping condensation causing damage.The difference in height between the assembly position of the liquefier unit and the installation site of the evaporator unit/SUNTEST instrument should not exceed 5 m. See section 6.4 for installation of these units.

NOTE – winter operation

The liquefier unit is equipped with a winter control; no additional condensation pressure regulation is required for outdoor installation.

Pipe material (Fig. 6.4): For outdoor installation, the supply and return pipes (see fig. 6.4) must be installed using flexible copper pipe de-signed for the transfer of refrigerant. The pipe should not exceed a maximum length of 15 m. Under no circum-stances should the enclosed standard hose connections be used. Supply and return pipes should be installed with as few bends as possible; the inside diameter of the pipes must not be compromised by crushed pipe material when laying bends. Then bending radius must be at least 20 cm.

Fig. 6.4

12 3

Page 23: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 23 -

Operating Manual SunCool for SUNTEST® CPS+

6SunCool start up

6.6 Outdoor installation of the liquefier unit

WARNING – danger of intoxication!

If large amounts of R410A escape into the room atmosphere, fumes are produced which can lead to intoxication or irregular heartbeat which could be fatal. Take immediate safety precautions in case of heavy leakage from the coolant circuit!

Pipe connections (liquefier unit): The pipe connections on the liquefier unit consist of two valves 1 and 2 (see fig. 6.5). The valve 2 of the sup-ply line has an additional maintenance connection 6 with Schrader valve. The pipes 3 to be laid by the customer must be fitted with a union nut 4 and equipped with a flanged end 5.

NOTE – nature of the flanged end!

The wall length of the socket should be evenly flanged. Make sure that no chips, dust or other foreign bodies fall inside the pipes because the coolant circuit could otherwise be blocked at the level of the hair pipe. This would result in blockage of the system or seizing of the compressor of the liquefier unit.

Coolant volume: The coolant volume of 1050 g is sufficient to supply a piping system (supply and return line) of a total length of 10 m. If the piping system exceeds this dimension, an additional 15 g of coolant (R410A) for every further meter of pipe laid must be added into the circuit. The coolant is added at the maintenance connection 6 through the Schrader valve.

Checking for leaks:To check for leaks, pressure is applied to the supply line (liquid gas pipe) and to the return line (suction gas pipe) for approx. 30 minutes.

Continued on page 24

Fig. 6.5

4

1

2

6

7

5

3

Page 24: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 24 -

6SunCool start up

6.6 Outdoor installation of the liquefier unit

NOTE – closing state of the valves!

To check for leaks, the stop taps of the valves on the liquefier unit must not be open (see Fig. 6.5, Page 23):1. Connect pipes to the valves of the liquefier unit and

the pipe with the quick-release couplings. 2. Connect the vacuum pump to the maintenance aperture 6 of the return valve of the liquefier unit.

3. Drain the pipes empty until an absolute pressure of 0.1 bar is reached.

4. Switch off the vacuum pump. Measure the pressure again after 30 minutes. The pipe must hold a constant pressure of 0.1 bar.

5. Remove the connecting hose of the vacuum pump from the maintenance aperture of the stop tap and seal the maintenance aperture with the stopper 7.

Opening the coolant circuit (Fig. 6.6): 1. Unscrew the stoppers 1 and 4 from the valves. 2. Fully open the stop taps 3 and 5 of the two valves

with the allen wrench 2. Turn the stop tap carefully counterclockwise to the stop.

3. Unscrew the stoppers 1 and 4 on the valves and tighten until the torque increases. Then tighten the stopper by a quarter turn.

NOTE – leak test!

After the tightness test and opening of the coolant circuit, the entire pipe system must be checked for leaks with a leak tester.

Checking the fill volume of the coolant:The inspection can be carried out with a portable pressure gauge for freon with a pressurized connection of ¼ SAE. The vacuum pump is connected to the valve‘s maintenance connection. The read pressure must correspond to the saturation pressure of the respective room temperature (20 °C = 14.4 bar / 0 °C = 7.98 bar).

Fig. 6.6

1

4

23

5

Page 25: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 25 -

Operating Manual SunCool for SUNTEST® CPS+

6SunCool start up

6.7 Main power supply

WARNING – electrical shock!

Touching live parts can lead to life threaten-ing electrical shocks. Check plugs and power cables for damage before connecting to the main power supply. Damaged parts may not be used for electrical connection!

Connecting interface cables (Fig. 6.7):The liquefier unit and SUNTEST CPS+ are connected by the interface cable (4-pin plug coupling).1. Insert the plug coupling 3 with the union nut into the

receptacle 2 on the back of the SUNTEST CPS+ and secure with the union nut.

2. Make sure that the interface cable is not subject to excessive stress or strain.

Connecting SunCool (Fig. 6.7):1. Before connecting to the main power supply, check

whether the values of the power supply network of the room of installation match the data on the rating plate on the back of the instrument.

2. If the data for voltage (V) and maximum current (A) do not match, the instrument should not be connected.

3. Plug one end of the PE contact plug 1 (on the instru-ment side) into the light equipment socket and the other end into a properly grounded and fused socket. The SunCool instrument will then be ready for operation and the cooling process can begin.

4. Make sure that no tensile or tractive forces act on the power cables.

CAUTION – outdoor installation!

In case of outdoor installation of the liquefier unit, the light equipment plug connection must be protected against rain and splashing water.

3

2

2

1

Fig. 6.7

Page 26: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 26 -

7.1 Menu structure of the instrument control

An overview of the menu structure of the program modules and the corresponding sub-menus can be found in the software documentation manual.

7Operation

7.2 Setting the SunCool accessory

The SunCool accessory must be selected on the in-strument control panel to be able to run a test program with SunCool support:

1. Select the „Settings“ item in the main menu.

2. Select the „Accessory“ menu item.

3. Select the „SunCool“ accessory.

4. Press the ENTER key.

Page 27: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 27 -

Operating Manual SunCool for SUNTEST® CPS+

7Operation

7.3 Starting the cooling process

A short safety check must be performed on the chiller every time it is started:• Check whether the coolant hoses have been installed

with sufficient bending radius and are not kinked.• Check whether the coolant is leaking from the coolant

hoses or the instrument connections.• Make sure that the condensation drain of the evaporator

unit is fed into a sewer drain or waste water tank.

NOTE – delayed cooling start

The cooling process does not start until shortly after swit-ching on the chiller. If this start delay should present a problem for any temperature-sensitive samples, a delayed program start can be selected in the instrument control of the test instrument.

NOTE – software documentation

The operation of the instrument control and the program-ming of the test programs are explained in the separate software documentation manual.

NOTE – air temperature (Fig. 7)

Constantly cooled air is fed into the test chamber. By altering the distance between the evaporator unit and the air inlet aperture of the SUNTEST CPS+, the air tempe-rature can be adjusted upwards. Since the temperature value depends on the climatic conditions of the room of installation, the alteration value of the air temperature must be determined by experimentation. However, in order to achieve full cooling power, the evaporator unit must be placed directly in contact with the SUNTEST CPS+ unit.

Fig. 7

-°C

+°C

Page 28: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 28 -

8Taking out of operation

8.1 Switching off the instrument

The device can be turned off at the end of or interruption of a test program.

Switch off at the end of the test program: • Pull the plug (light equipment plug) out of the recep-

tacle or disconnect from the power supply at a fixed connection.

• If the condensation is pumped off with a micro-pump, this must also be switched off. To disconnect, pull the electrical plug of the micro-pump out of the receptacle.

• Drain the condensation from the drain pipe, empty the waste water vessel.

Interrupt the test program, switch off the instrument:• Interrupt the program run on the SUNTEST® CPS+

by pressing the “STOP” key.• Pull the plug (light equipment plug) out of the recep-

tacle or disconnect from the power supply at a fixed connection.

• If the condensation is pumped off with a micro-pump this must also be switched off. To disconnect, pull the plug of the micro-pump out of the receptacle.

8.2 Switching off the instrument in an emergency

• Pull the plug (light equipment plug) out of the recep-tacle.

8.3 Decommissioning of instrument

Atlas instruments are classified according to the EU di-rective 2012/19/EU (updated version) of the European Parliament and Commission on Waste Electrical and Electronic Equipment (WEEE) as „monitoring and control instruments for exclusively commercial use“ (Category 9) and must NOT be decommissioned/discarded through public disposal channels.

The instruments carry the symbol (crossed out trash can on wheels and bars) to identify electrical and electronic equipment which came onto the EU market after August 13, 2005 and must be disposed of separately in accordance with EU directive 2012/19/EU (up-dated version) (WEEE) on waste electrical and electronic equipment.

Notify the dealer from whom you purchased the instrument when it is ready for decommissioning so that they can organize the acceptance and disposal in accordance with the Electrical and Electronic Equipment Act (ElektroG) of the German Civil Code I S. 762.

CAUTION – Violation of applicable law

Do not hand in Atlas instruments to public collection points

Used Atlas instruments are dismantled and sorted into separate materials by certified companies in accordance with EU directive 2012/19/EU (updated version). The in-struments must be cleaned and free from test materials to rule out health hazards for employers of the disposal companies.

NOTE

The user of the instrument is responsible for ensuring that the instrument is free from health hazardous materials before handing it over to a disposal company.Prior to collection, the disposal companies require a safety declaration which can be obtained from Atlas.

WARNING

Contamination of the instrument with toxic, infectious or radioactive material.Danger of intoxicationDanger of infection

NEVER pass on the instrument for recycling in accordance with EU directive 2012/19/EU (updated version) when it is soiled by toxic substances or sources of infection.Dispose of instruments with irremovable toxic substances or sources of infection as special waste in accordance with national regulations.

Page 29: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 29 -

Operating Manual SunCool for SUNTEST® CPS+

8Taking out of operation

8.3 Decommissioning of instrument

Disposal of the instrument in EU countries outside the Federal Republic of Germany:

Atlas instruments are classified according to the EU di-rective 2012/19/EU (updated version) of the European Parliament and Commission on Waste Electrical and Electronic Equipment (WEEE) as „monitoring and control instruments for exclusively commercial use“ (Category 9) and may NOT be handed into public collection points.

The instruments carry the symbol (crossed out trash can on wheels and bars) to identify electrical and electronic equipment which came onto the EU market after August 13, 2005 and must be disposed of separately in accordance with EU directive 2012/19/EU (up-dated version) (WEEE) on waste electrical and electronic equipment.

Notify the dealer from whom you purchased the instrument when you have finished using it so that he takes back and disposes of it in accordance with the EU directive 2012/19/EU (updated version) on waste electrical and electronic equipment.

Disposal of the instrument in non-EU countries:

CAUTION – environmental damage!

Observe the pertinent public disposal regulations for protection of the environment.

Page 30: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 30 -

10Maintenance

10.1 Inspection

The instrument must be inspected annually. The test scope of the inspection includes: • Checking the fill level of the coolant• Checking the condition of the coolant lines • Function and safety of the electrical components

NOTE – warranty!

The manufacturer warrants the safety and operation only on the condition that:• Inspection intervals are maintained • All inspections are carried out by appropriately trained

and qualified personnel or by an Atlas technical service representative

• The instructions in the operating manual are observed

NOTE – condition of the hoses!

If a coolant hose is crushed or kinked, should not be used when operating the cooler and must be replaced.

10.2 Repair

Repairs may only be carried out by appropriately trained and qualified personnel or by an Atlas technical service representative. Only original spare parts should be used when installing or changing instrument components. Atlas will accept no liability for instrument components or parts from third party manufacturers.

9Troubleshooting

9.1 Error messages and troubleshooting

A detailed list of the error messages and remedies meas-ures can be found in the software manual for XenoSmart.

Page 31: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 31 -

Operating Manual SunCool for SUNTEST® CPS+

10Maintenance

10.3 Maintenance

WARNING – electrical shock!

Touching live parts can lead to life threaten-ing electrical shocks. Disconnect the instru-ment from the power supply before beginning maintenance work! • Pull the electrical plug out of the receptacle

and secure against reconnecting • Check whether the instrument is voltage

free

Cleaning the air inlet of the compressor unit:Residue dust and dirt on the air inlet and air outlet apertu-res of the liquefier unit must be removed every six months.

Cleaning the air inlet of the evaporator unit:The air filter of the evaporator unit must be checked week-ly. If dust and dirt residue are clearly visible, it must be cleaned. The air filter must be cleaned at least every six months in the course of the maintenance intervals.

All other instrument components of the SunCool system are maintenance-free.

10.4 Cleaning

WARNING – electrical shock!

Touching live parts can lead to life threaten-ing electrical shocks. Disconnect the instru-ment from the power supply before beginning maintenance work! • Pull the electrical plug out of the receptacle

and secure against reconnecting • Check whether the instrument is voltage

free

Cleaning the instrument surface: CAUTION – plastic surfaces!

Parts of the instrument surface are made of plastic. Solvents can soften plastics and make them useless. Do not clean the instrument surface with solvents containing hydrocarbons..

Wipe off the instrument surface with warm water and a mild detergent and then dry with a clean cloth.

Cleaning the air filters (Fig. 8):The air filter mat sits in the air inlet aperture of the eva-porator unit.1. Remove the filter mat 1 from the air inlet aperture 2.2. Remove any visible. dust or dirt from the filter.3. Wipe off the surface of the heat exchanger 3.4. Wash out the filter mat in warm water with a mild de-

tergent and dry well.5. Place the filter mat back in the air inlet aperture.

2 3

1

Fig. 8

Page 32: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 32 -

11Technical Data

11 Technical Data

Incoming power:Incoming power connection: (1P/N/PE) AC (2P/PE) AC Light equipment plug, PE contact plug

Nominal voltage/frequency: 230 V ± 10% 50 / 60 HzNominal current: 4ANominal power: 1 kW

Cooling capacity: 2.6 kW

Coolant: Type: R410AFilling volume: 650 gMaximum line length: 20 m

Dimensions and weights: Evaporator unitDimensions in cm: 35 x 32 x 33 (W x D x H)Liquefier unit:Dimensions in cm: 70 x 60 x 35 (W x D x H)

Empty weights:Evaporator unit approx. 7 kgLiquefier unit 30 kg(without connections)

Hose connections: Length: 2,5 mTemperature range: -20 °C – 100 °CSmallest bending radius: 20 cm

Ambient conditions:Room temperature: max. 18 °C – 25 °CRelative humidity: max. 50%

Page 33: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 33 -

Operating Manual SunCool for SUNTEST® CPS+

12Accessories

12 Accessories

Pos. Accessories Description Ident-Nr.:

1 Waste water pump 56078463

2 Isolation pads for SUNTEST CPS+ 56077904

3 Air filter forSunCool CPS+ 56055110

4Assembly kit (for external installation of the liquefier unit) 50/60 Hz

56078465

Page 34: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 34 -

13Declaration of conformity

Page 35: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

- 35 -

Operating Manual SunCool for SUNTEST® CPS+

14Notes

14 Notes

Page 36: Operating Manual SunCool for SUNTEST CPS+atlasmtt.net/data/ger/manuals/UM_suncool_cps+(II)_2013.pdf- 9 - Operating Manual SunCool for SUNTEST® CPS+ 1Instructions for safe operation

Operating Manual SunCool for SUNTEST® CPS+

- 36 -

Atlas Material Testing Technology GmbH Vogelsbergstr. 22 63589 Linsengericht / Germany (p) + 49-6051-707-140 (f) + 49-6051-707-149 email: [email protected] www.atlas-mts.com Id.-Nr.: 56352634 01/13