XS-Ex2 and X-Ex2 Models – Part 1 Operating Instructions Excellence Precision Balances

Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

XS-Ex2 and X-Ex2 Models – Part 1O

pera

ting

Inst

ruct

ions Excellence Precision Balances

Page 2: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located
Page 3: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Table of Contents

Introduction1 5

Symbols and presentations used1.1 6

Safety Information2 7

Definition of signal warnings and symbols2.1 7Product specific safety notes2.2 7

Overview Balances and Weighing Platforms for Ex-Zone 23 11

Overview XS-Ex2 "S" and "M" weighing platform3.1 11Overview XS-Ex2 "L" weighing platform3.2 12Overview X-Ex2 "S" and "M" weighing platform3.3 13Overview X-Ex2 "L" weighing platform3.4 14

Setting up the Balance4 15

Unpacking and checking the delivered items4.1 15Scope of delivery4.1.1 15Unpacking the Draft Shield "S" Weighing Platform (Model-dependent)4.1.2 17Assembling the balance4.2 17Installing the draft shield and weighing pan4.2.1 17Placing the terminal on the "L" weighing platform4.2.2 20Selecting the location and level of the balance4.3 21Selecting the location4.3.1 21Leveling the balance4.3.2 21Installation of the AC adapter PSX24.4 22Installation in the Ex hazardous area (zone 2)4.4.1 22Installation outside the Ex hazardous area4.4.2 22Connection to the balance/weighing platform4.4.3 23Dismantling4.4.4 24Marking the balance/weighing platform and the AC adapter4.5 24

Connection of Peripheral Devices and System Integration5 26

Usage of data interface "RS232" in the Ex hazardous area (zone 2)5.1 26The peripheral device «P» is located inside the Ex hazardous area5.1.1 26The peripheral device «P» (e.g. PC or printer) is located outside the Exhazardous area

5.1.2 27

Usage of the optional data interface "Bluetooth" in the Ex hazardous area5.2 27The peripheral device «P» is located inside the Ex hazardous area5.2.1 27The peripheral device «P» (e.g. printer) is located outside the Ex haz­ardous area

5.2.2 28

Maintenance6 29

Cleaning6.1 29Cleaning the draft shield (0.1 mg and 1 mg models)6.2 30Disposal6.3 31

Technical Data7 32

General data7.1 32

Table of Contents 3

Page 4: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Model-specif data of the Excellence XS-Ex2 precision balances7.2 33XS-Ex2 balances with readability of 0.1 mg / 1 mg, "S" platform with draftshield

7.2.1 33

XS-Ex2 balances with readability of 0.1 mg, "S" platform7.2.2 35XS-Ex2 balances with readability of 10 mg / 0.1 g, "S" platform7.2.3 36XS-Ex2 balances with readability of 10 mg / 0.1 g, "M" platform7.2.4 37XS-Ex2 balances with readability of 1 mg / 5 mg, "L" platform7.2.5 38XS-Ex2 balances with readability of 0.1 g / 1 g, "L" platform7.2.6 39Model-specif data of the Excellence X-Ex2 precision balances7.3 40X-Ex2 balances with readability of 0.1 mg / 1 mg, "S" platform with draftshield

7.3.1 40

X-Ex2 balances with readability of 10 mg / 0.1 g, "S" platform7.3.2 42X-Ex2 balances with readability of 10 mg / 0.1 g, "M" platform7.3.3 43X-Ex2 balances with readability of 0.1 g / 1 g, "L" platform7.3.4 43Dimensions of the Excellence XS-Ex2 balances7.4 45XS-Ex2 balances with readability of 0.1 mg, "S" platform with draft shield7.4.1 45XS-Ex2 balances with readability of 1 mg, "S" platform with draft shield7.4.2 46XS-Ex2 balances with readability of 10 mg, "S" platform with draft shieldelement

7.4.3 47

XS-Ex2 balances with readability of 0.1 mg, "S" platform7.4.4 48XS-Ex2 balances with readability of 0.1 g, "S" platform7.4.5 49XS-Ex2 balances with readability of 10 mg / 0.1 g, "M" platform7.4.6 50XS-Ex2 balances with readability of 1 mg / 5 mg, "L" platform7.4.7 51XS-Ex2 balances with readability of 0.1 g / 1 g, "L" platform7.4.8 52Dimensions of the Excellence X-Ex2 weighing platforms7.5 53X-Ex2 weighing platforms with readability of 0.1 mg, "S" platform withdraft shield

7.5.1 53

X-Ex2 weighing platforms with readability of 1 mg, "S" platform with draftshield

7.5.2 54

X-Ex2 weighing platforms with readability of 10 mg, "S" platform with draftshield element

7.5.3 55

X-Ex2 weighing platforms with readability of 0.1 g, "S" platform7.5.4 56X-Ex2 weighing platforms with readability of 0.1 g, "M" platform7.5.5 57X-Ex2 weighing platforms with readability of 0.1 g, "L" platform7.5.6 58Dimensions of the AC adapter PSX27.6 59Interfaces7.7 60Specifications of RS232C7.7.1 60

Accessories and Spare Parts8 61

Accessories8.1 61Spare parts8.2 66

Certificate of the Issuing Institution9 68

Power supply PSX29.1 68Balances XS-Ex2 and platforms X-Ex29.2 70

Appendix10 73

MT-SICS interface commands and functions10.1 73

Index11 74

Table of Contents4

Page 5: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

1 IntroductionThank you for choosing a METTLER TOLEDO balance.

Balances and platforms for use in zone 2 hazardous areas combine a host of weighing and setting possibilitieswith uncommon operating comfort.

In this chapter you will be given basic information about your balance. Please read right through this chaptercarefully even if you already have experience with METTLER TOLEDO balances. Please pay special attention tothe safety warnings!

The different models have different characteristics regarding equipment and performance. Special notes in thetext indicate where this makes a difference to operation.

The XS-Ex2/X-Ex2 family comprises a range of balances which differ from each other in relation to their weigh­ing range and resolution.

The following features are common to all models of the XS lines:

Type of protection: Ex nL for use in zone 2 hazardous areas.

Fully automatic adjustment "FACT" using internal weight.

Built-in applications for normal weighing, statistics, formulation, density, percent weighing, piece counting,dynamic weighing and LabX Client (only XS-Ex2 models).

Integral RS232C interface.

Slot for second interface (optional).

Touch-sensitive graphics terminal ("Touch screen") for easy, convenient operation (only XS-Ex2 models).

A brief word about standards, guidelines, and methods of quality assurance: The balances comply with usualstandards and guidelines. They support standard procedures, specifications, working methods, and reportsaccording to GLP (Good Laboratory Practice). In this connection, records of working procedures and adjust­ments become very important; for this purpose we recommend you to use a printer from the METTLER TOLEDOrange, since these are optimally adapted to your balance. The balances conform to the applicable standardsand guidelines and possess a EC declaration of conformity. METTLER TOLEDO is certified as manufactureraccording to ISO 9001 and ISO 14001.

The operating instructions for the XS-Ex2/X-Ex2 balances consist of 3 separate documents, whose con­tents are listed below.

Part 1, this documentContents

Introduction Safety notes Setting up the balance/weighing platform Leveling the balance/weighing platform Installation of the AC adapter PSX2 Connection of peripheral devices and system integration Cleaning and service Technical data Accessories Spare parts Certificate of the issuing institution Interface commands and MT-SICS functions

5Introduction

Page 6: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

6 Introduction

Part 2, separate document (only XS-Ex2 balances)Contents: Terminal, System, and Applications

Basic principles for using the terminal and the firmware System settings Applications Firmware (Software) updates Error and status messages Conversion table for weight units Recommended printer settings

Part 3, separate document (only XS-Ex2 balances)Contents: Adjustments, and Tests

Adjustments Tests

Finding more informationInternet http://www.mt.com/excellence

1.1 Symbols and presentations usedThe following conventions apply to the operating instructions: part 1, part 2 and part 3.

Key designations are indicated by a picture or text in double angular parentheses (e.g. [ ] or [On/Off]).

This symbol indicates press key briefly (less than 1.5 s).

This symbol indicates press and hold key down (longer than 1.5 s).

These symbols indicate an instruction:prerequisites

1

2 …

steps

results

Page 7: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

2 Safety Information

2.1 Definition of signal warnings and symbolsSafety notes are indicated by signal words and warning symbols and contain warnings and information aboutsafety issues. Ignoring safety notes can lead to personal injury, damage to the instrument, malfunctions anderroneous results.

Signal words

WARNING for a hazardous situation with medium risk, possibly resulting in severeinjuries or death if not avoided.

CAUTION for a hazardous situation with low risk, resulting in damage to the device orthe property or in loss of data or minor or medium injuries if not avoided.

Attention (no symbol)for important information about the product.

Note (no symbol)for useful information about the product.

Warning symbols

General hazard Electrical shock

Heavy object

2.2 Product specific safety notesApproved for use in zone 2 hazardous areas

The AC adapter PSX2 supplies the power to the XS-Ex2 precision balances and X-Ex2 preci­sion weighing platforms for their intended use in the Ex hazardous area classified as zone 2.In this zone, ignitable concentrations of flammable gases, vapors or liquids are not likely tooccur in normal operation, and if they do occur, will exist for only a short period. The ACadapter can be installed inside or outside the Ex hazardous area.

Use in zone 1 and 0 is not permitted.

General safety informationYour instrument meets the state of the art technology and complies with all recognized safety rules, however,certain hazards may arise in extraneous cirmumstances. Do not open the housing of the instrument: It does notcontain any parts which can be maintained, repaired or replaced by the user. If you ever have problems withyour instrument, contact your authorized METTLER TOLEDO dealer or service representative.

Always operate and use your balance only in accordance with the operating instructions part 1, part 2 andpart 3.

The instructions for setting up your new balance/weighing platform must be strictly observed.

If the instrument is not used according to the manufacturer’s operating instructions (part 1, part 2 andpart 3), protection of the instrument may be impaired.

7Safety Information

Page 8: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

8 Safety Information

Staff safetyIn order to use the instrument, you must have read and understood the operating instructions. Keep the operat­ing instructions for further reference.

Never make any modifications or constructional alterations to the instrument and use only original spare partsand optional equipment from METTLER TOLEDO.

Safety notes CAUTION

Damage of device

– For use only in dry interior rooms.

– Do not use sharply pointed objects to operate the keyboard! Although your balance is veryruggedly constructed, it is nevertheless a precision instrument. Treat it with correspondingcare.

CAUTION

Risk of electric shock

Use only the AC adapter PSX2 delivered with your balance/weighing platform, and check thatthe voltage printed on its the same as your local power supply voltage.

WARNING

Heavy object

– Single person lift could cause injury.

– Do not move or lift this equipment without assistance.

Page 9: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Safety instructions regarding installation and operation in the Ex hazardous area, zone 2 CAUTION

Safety instructions

– The installation may be performed only by a qualified electrician as described in thisinstallation information.

– Check whether special clothing or special tools are needed for work in the hazardous areaof the end-user and then wear or use these if required.

– Find out whether the use of certain electronic devices (cell phones, computers, etc.) is pro­hibited in the hazardous area of the end-user, and then follow these regulations.

– Establishing and separating the connections to the power supply are exclusively the task ofthe electrician(s) of the end user.

– Any modifications to the balance/weighing platform and AC adapter, as well as repairs tothe assemblies, are prohibited. These actions endanger the safety of the system, result inloss of the Ex Approval, and nullify any warranty and product liability claims.

– Service work and repairs may be performed only by personnel authorized by METTLERTOLEDO.

– The classification as zone 2 must be confirmed by the regulatory body of your profession.

– Check that the balance and AC adapter are in perfect condition before putting into operationfor the first time and at least every 3 years.

– It is important to observe the European Directive on Equipment Intended for Use in Poten­tially Explosive Atmospheres (94/9/CE)!

– Only accessories which are explicitly mentioned in this document may be used, and theymust be installed and operated as described here.

– Installation work on the balance/weighing platform must be carried out outside the haz­ardous area.

Safety measures during operation in the Ex hazardous area, zone 2 CAUTION

Damage of device

– AC adapter and balance may be operated only when the housing is closed.

– The instrument must only be cleaned with a damp cloth.Note– According to Directive 94/9/EU (ATEX 95), XS-Ex2 precision balances, X-Ex2 precisionweighing platforms, and the AC adapter PSX2 are Group II, Category 3G devices that,according to Directive 99/92/EU (ATEX 137), can be used in zone 2, as well as in GasGroups IIA, IIB and IIC, which are potentially ignitable by combustible substances in therange of Temperature Classes T1 to T5.

– For use/installation, the requirements of EN 60079-14 must be observed.

9Safety Information

Page 10: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

10 Safety Information

Special conditions X for the AC adapter PSX2 CAUTION

Damage of device

– The power supply cable of the METTLER TOLEDO Type AC adapter PSX2 must be installedwith protection against mechanical damage.

– Only the special plug-connector of the output circuit may be plugged into, or unpluggedfrom, a power-consuming device such as a balance while it is under voltage.

– The internal inductance (including the connecting cable) of any power-consuming deviceconnected to the AC adapter must not exceed ≤ 30 µH, and its internal capacity must notexceed ≤ 4.9 µF.

Special condition X for the balance/weighing platform CAUTION

Damage of device

– XS-Ex2 precision balances and X-Ex2 precision weighing platforms can only be used inclosed and clean interior spaces.

– The instrument must only be cleaned using a damp cloth.

Technical data of the precision balances in the XS/X-Ex2 line and the AC adapter PSX2.

See Technical Data (page 32).

Page 11: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

3 Overview Balances and Weighing Platforms for Ex-Zone 2

3.1 Overview XS-Ex2 "S" and "M" weighing platform

1mg

Power Supply Aux 1 Aux 2 RS 232 C

1

2

3

4

56

7

7

9

8

8

10

7

7

7

10

10

1111

6

11

12

10

11

131415161718

7

11

10

XS-Ex2 View S and M Platform

1 Terminal (details see Operating Instruction –Part 2)

2 Display (Touch-sensitive "Touch screen")

3 Operating keys 4 Type name5 Level indicator 6 Draft shield element7 Weighing pan 8 Handle for operation of the draft-shield door9 Glass draft shield 10 Foot screws11 Safety feet (10 mg, 0.1 g + 1 g models) 12 Fastening point for anti-theft device13 RS232C serial interface 14 Aux 1 (connection for "ErgoSens")15 Aux 2 (connection for "ErgoSens") 16 Fastening for auxiliary display stand or termi­

nal stand (optional)17 Socket for AC adapter 18 Slot for second interface (optional)

11Overview Balances and Weighing Platforms for Ex-Zone 2

Page 12: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

12 Overview Balances and Weighing Platforms for Ex-Zone 2

3.2 Overview XS-Ex2 "L" weighing platform

8

1

2

4

5

5

9

676

3

10

10

11 12

13

15

14

17 161819

XS-Ex2 View L Platform

1 Terminal (details see Operating Instruction –Part 2)

2 Display (Touch-sensitive "Touch screen")

3 Operating keys 4 Type name5 Weighing pan 6 Foot screws7 Cover 8 Fastening point for anti-theft device9 Draft shield 10 Points of attachment for terminal or cover11 Fastening for terminal stand (optional) 12 Level indicator13 Aux 1 (connection for "ErgoSens") 14 Aux 2 (connection for "ErgoSens")15 RS232C serial interface 16 Socket for AC adapter17 Connector for terminal cable 18 Slot for second interface (optional)19 Cover plate for below-the balance weighing

(hook optional)

Page 13: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

3.3 Overview X-Ex2 "S" and "M" weighing platform

1mg

7

1

1

4

2

3

4

5

6

58

8

8

8

1

1

4

4

3

7

7

7

Power Supply Aux 1 Aux 2 RS 232 C

9

101112131415

X-Ex2 View S and M Platform

1 Type name 2 Level indicator3 Draft shield element 4 Weighing pan5 Handle for operation of the draft-shield door 6 Glass draft shield7 Safety feet (10 mg, 0.1 g + 1 g models) 8 Foot screws9 Fastening point for anti-theft device 10 RS232C serial interface11 Aux 1 (connection for "ErgoSens") 12 Aux 2 (connection for "ErgoSens")13 Fastening for auxiliary display stand or termi­

nal stand (optional)14 Socket for AC adapter

15 Slot for second interface (optional)

13Overview Balances and Weighing Platforms for Ex-Zone 2

Page 14: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

14 Overview Balances and Weighing Platforms for Ex-Zone 2

3.4 Overview X-Ex2 "L" weighing platform

2

1

5

343

6

6

7 8

9

11

10

13 121415

X-Ex2 View L Platform

1 Type name 2 Weighing pan3 Foot screws 4 Cover5 Fastening point for anti-theft device 6 Points of attachment for terminal or cover7 Fastening for terminal stand (optional) 8 Level indicator9 Aux 1 10 Aux 211 RS232C serial interface 12 Socket for AC adapter13 Connector for terminal cable 14 Slot for second interface (optional)15 Cover plate for below-the balance weighing

(hook optional)

Page 15: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

4 Setting up the BalanceThis chapter describes how to unpack your new balance, set it up and prepare it for operation. On completionof the steps described in this chapter, your balance is ready for operation.

CAUTION

Risk of electric shock

– The balance/weighing platform must be disconnected from the power supply before anyset-up or installation work.

– All unpacking, set-up, installation or disassembly work must be performed outside thehazardous area.

4.1 Unpacking and checking the delivered itemsOpen the packaging and carefully remove all components.

4.1.1 Scope of delivery

NoteThe AC adapter PSX2 is not included in the scope of delivery for the balance/weighing platform and must beordered separately (11132730).

Components S platform0.1 mg XS2004SX

0.1 mg1 mg 10 mg 0.1 g

XS-Ex2 balance or X-Ex2 weighing plat­formDraft shield "Pro" with bottom plate – – –

Draft shield – – – – –Draft shield element – –

Ring seal – – – –

Pan support –

ø 90 mm – – – –

ø 258 mm – – – – –127 x 127 mm – – –

170 x 205 mm – – – –

190 x 223 mm – – – –

237 x 237 mm – – – – –

Weighing pan

280 x 360 mm – – – – –Operating Instructions Part 1 (this docu­ment)Operating Instructions Part 2 and Part 3(only XS-Ex2 balances)Production certificate

CE declaration of conformity

Components M platform * L platform10 mg 0.1 g 1 g 0.1 g 1 g

15Setting up the Balance

Page 16: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

16 Setting up the Balance

Components M platform * L platformXS-Ex2 balance or X-Ex2 weighing plat­formDraft shield "Pro" with bottom plate – – – – –Draft shield – – – – –Draft shield element – –

Ring seal – – – – –Pan support – –

ø 90 mm – – – – –ø 258 mm – – – – –127 x 127 mm – – – – –170 x 205 mm – – – – –190 x 223 mm – – – – –237 x 237 mm – –

Weighing pan

280 x 360 mm – – –

Operating Instructions Part 1 (this docu­ment)Operating Instructions Part 2 and Part 3(only XS-Ex2 balances)Production certificate

CE declaration of conformity

* The precision balances with M platform are no longer available.

Components L platform1 mg 5 mg

XS-Ex2 balance or X-Ex2 weighing plat­formDraft shield "Pro" with bottom plate – –Draft shield

Draft shield element – –Ring seal – –

ø 90 mm – –ø 258 mm

127 x 127 mm – –170 x 205 mm – –190 x 223 mm – –237 x 237 mm – –

Weighing pan

280 x 360 mm – –Operating Instructions Part 1 (this docu­ment)Operating Instructions Part 2 and Part 3(only XS-Ex2 balances)Production certificate

CE declaration of conformity

Page 17: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

4.1.2 Unpacking the Draft Shield "S" Weighing Platform (Model-dependent)

1 Place the draft shield on a clean surface.

2 Turn so the cover (1) is vertically on top.

3 Lift the carton (2) off over the handle and the pull the carton offtoward the back.

4 Hold the glass panels firmly so they do not get pulled off and fall tothe floor.

5 Close the cover (1) again.

6 Push all the glass panels as far as they will go to the back.

Balance with 0.1 mg readability

1 Place the ring seal (3) into the draft shield base (4) from above.

2 Push the complete ring seal (3) through the opening in the draftshield base.

3 Pull the top edge upward through the opening until the edge liesflat and tight all the way round.

4 Check by running a finger all the way round that the seal ring (3)is fitted tightly into the opening in the draft shield base (4).

5 Insert the bottom plate (5).

Balance with 1 mg readability

– Insert the bottom plate (5).

1

2

3 5 4

4 3

4.2 Assembling the balanceThe size of the weighing pan depends on the readability and maximum capacity of the balance.

4.2.1 Installing the draft shield and weighing pan

Balances with readability of 0.1 mg, "S" platform with draft shield

– Place the following components on the balance in the specifiedorder:

AttentionPush the side glass back as far as it will go. Grasp the draft shieldwith both hands on the bars at the top.

Draft shield (1) with ring seal inserted, see Unpacking the DraftShield "S" Weighing Platform (Model-dependent) (page 17).

Bottom plate (2), if not already inserted, see Unpacking the DraftShield "S" Weighing Platform (Model-dependent) (page 17)!

Draft shield element (3). Weighing pan (4).

NoteCleaning the draft shield, see Cleaning the Draft Shield (0.1 mg and1 mg Models) (page 30).

1

2

4

3

17Setting up the Balance

Page 18: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

18 Setting up the Balance

Balances with readability of 1 mg, "S" platform with draft shield

– Place the following components on the balance in the specifiedorder:

AttentionPush the side glass back as far as it will go. Grasp the draft shieldwith both hands on the bars at the top.

Draft shield (1) with ring seal inserted. Bottom plate (2), if not already inserted, see Unpacking the Draft

Shield "S" Weighing Platform (Model-dependent) (page 17)! Pan support (3). Weighing pan (4).

NoteCleaning the draft shield, see Cleaning the Draft Shield (0.1 mg and1 mg Models) (page 30).

1

2

4

3

Balances with readability of 0.1 mg, "S" platform

– Place the following components on the balance in the specifiedorder:

Draft shield element (1). Pan support (2). Weighing pan (3).

1

2

3

Balances with readability of 10 mg, "S" platform with draft shieldelement

– Place the following components on the balance in the specifiedorder:

Pan support (1). Weighing pan (2). Draft shield element (3).

NoteIt is also possible to work without the draft shield element (3). Howev­er, depending on the ambient conditions, the display of the resultsmay be slightly more unstable.

1

2

3

Page 19: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Balances with readability of 10 mg ("M" platform), 0.1 g ("S" and"M" platform) and 1 g ("M" platform)

– Place the following components on the balance in the specifiedorder:

Pan support (1). Weighing pan (2).

1

2

Balances with readability of 1 mg and 5 mg ("L" platform)

AttentionRemove the cells of the transport lock screws!

– Place the following components on the balance in the specifiedorder:

Weighing pan (1). Draft shield (2). 1

2

19Setting up the Balance

Page 20: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

20 Setting up the Balance

4.2.2 Placing the terminal on the "L" weighing platform

The terminal can be attached to the long or short side of the balance.

1 Mount the weighing pan.

2 Carefully turn the weighing platform over onto the weighing pan.

Attaching the terminal to the long side

1

221 Dismantle the cover (1) by removing the 2 screws (2).

2 Attach the terminal to the terminal support (5) as shown, using thescrews (2) from the cover that was removed.

3 Insert the terminal cable (3) into the cable channel, as shown.

4 Screw the connector of the terminal cable into the socket (4).

5 Turn the balance back over into position.

22

5

3 4

Attaching the terminal to the short side

2

2

1

1 Dismantle the cover (1) by removing the 2 screws (2).

2 Attach the terminal with terminal support to the points of attachment(6) with the screws (2).

3 Insert the terminal cable (3) into the cable channel, as shown.

4 Screw the connector of the terminal cable into the socket (4).

5 Turn the balance back over into position.

3 46 6

Page 21: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

4.3 Selecting the location and level of the balanceYour balance is a precision instrument and will thank you for an optimum location with high accuracy anddependability.

4.3.1 Selecting the location

Choose a position which is stable, free from vibration, and as nearlyhorizontal as possible. The supporting surface must be able to bearthe weight of the fully loaded balance safely.

Avoid the following:

Direct sunlight Draft (e.g. from fans or air conditioning) Excessive fluctuations in temperature.

Further information can be found in Weighing the Right Way.

Observe ambient conditions, see General data (page 32).

4.3.2 Leveling the balance

Balances with readability of 1 mg "S" platform and balances with "L" platform

1 Align the balance horizontally.

2 Turning the leveling screws of the housing until the air bubble is inthe inner circle of the level indicator.

The position of the air bubble illustrates which leveling screwyou need to turn (L = left leveling screw, R = right levelingscrew) and in which direction so that the air bubble moves tothe center.

ExampleIn this example, turn the left leveling screw counterclockwise.

Balances with readability of 10 mg, 0.1 g and 1 g ("S" and "M" platforms)

1 Remove the clamps (1) for the safety feet by turning them out­wards.AttentionTurn the clamps (1) outwards as far as they will go (~ 90°), sothat the safety feet can move freely.

2 Now level the balance by turning both leveling screws (2) until theair bubble is in the inner circle of the level indicator.

3 Secure the safety feet by turning the clamps (1) inwards as far asthey will go.

NoteThe balance must be leveled and adjusted each time it is moved to anew location.

11

22

21Setting up the Balance

Page 22: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

22 Setting up the Balance

4.4 Installation of the AC adapter PSX2

WARNING

Risk of electric shock

– Your weighing platform has been delivered with a country-specific AC adapter. Checkwhether the local power supply voltage is compatible with the AC adapter. If this is not thecase, on no account connect the AC adapter to the power supply but contact your responsi­ble METTLER TOLEDO dealer.

– The installation may be performed only by a qualified electrician.

4.4.1 Installation in the Ex hazardous area (zone 2)

If the AC adapter is installed in zone 2, the power supply cable must either be connected to the screw terminalsor a connection box outside the hazardous area, or to a connection box of suitable protection type within thehazardous area of the building.

An all-pole supply isolation device must beconnected on the line side of the screw termi­nals of the building installation.

1 Install AC adapter in a permanent position sothat it is protected against mechanical dam­age.

2 Attach power cable to the screw terminals ofthe building installation as follows:Core end yellow/green = groundCore end black No. 1 = phaseCore end black No. 2 = neutral conductor

3 Route power cable so that it is protectedagainst mechanical damage.

Zone 2

Country-specific 115/230 V power supply

connection

4.4.2 Installation outside the Ex hazardous area

If the AC adapter is installed outside the hazardous area and only the balance/platform is in the hazardousarea, the AC adapter can also be connected to the building installation using a commercial plug connection. Insuch a case the power plug is used as a supply isolation device.

NoteThe power plug is not included in the standard equipment.

Page 23: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

1 Install AC adapter in a permanent position sothat it is protected against mechanical dam­age.

2 Connect 3-pin power plug to power cable:Core end yellow/green = groundCore end black No. 1 = phaseCore end black No. 2 = neutral conductor

3 Plug power plug into receptacle outlet of thebuilding installation.

4 Route power cable so that it is protectedagainst mechanical damage.

Zone 2 Outside

hazardous

area

Country-specific 115/230 V power supply

connection

4.4.3 Connection to the balance/weighing platform

1 Connect the AC adapter to the connector socketon the backside/underside of your bal­ance/weighing platform.

2 Secure the connection to the balance by screw­ing the plug tightly.

3 Install the cable in such a way that it will notbe damaged and will not hinder day-to-daywork.

After the balance/weighing platform hasbeen connected to the power supply, it car­ries out a self-test and is then ready foroperation.

"S" and "M" platform

"L" platform

23Setting up the Balance

Page 24: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

24 Setting up the Balance

4.4.4 Dismantling

WARNING

Risk of electric shock

Disconnect AC adapter from power supply

1 Disconnect the power supply cable from the screw terminals of the in-house power supply in the followingsequence:Core end black No. 2 = neutral conductorCore end black No. 1 = phaseCore end yellow/green = ground

2 Unscrew the power supply unit.

3 Remove the balance and power supply cable.

4.5 Marking the balance/weighing platform and the AC adapterInstruments for use in hazardous areas must be marked with their type of protection and with warnings. Theappropriate stickers are affixed to the balance.

– Check to make sure model plate with the Ex symbol (1) and the "Electrostatic" warning label (2) has beenapplied to the balance/weighing platform.

Model plate balance/weighing platform

Page 25: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

– Check to make sure that the model plate with the Ex symbol (1) has been applied to the AC adapter.

Model plate AC adapter

25Setting up the Balance

Page 26: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

26 Setting up the Balance

5 Connection of Peripheral Devices and System IntegrationMany of the balances in current use, and especially weighing platforms, are normally integrated into complexcomputer or weighing systems. The integrated RS232 or an optional data interface can be used to captureweighing results and to control or adjust the weighing platform. Standard commands are available for this pur­pose.

System integration

5.1 Usage of data interface "RS232" in the Ex hazardous area (zone 2)AttentionThe installation must be performed by a professional in accordance with the applicable regulations.

If peripheral devices are connected to the RS232 data interface of the balance in the Ex hazardous area, thereare two possible cases.

5.1.1 The peripheral device «P» is located inside the Ex hazardous area

The peripheral device «P» which is used must be approved for zone 2.

The peripheral device «P» does not conform to protection type Ex ic.

– In order for the energy flow via the data cable to be limited between the balance and the peripheral device«P», an energy limiter with corresponding Ex certification «B» must be used.

The peripheral device «P» does conform to protection type Ex ic.

– The data cable does not need a special energy limiter.

«B» = Standard energy limiter in housing with pro­tection type (Ex ic IIC T5).(If «P» ≠ Ex ic).

Ex hazardous area, zone 2

P

B

P ≠ Ex ic

Page 27: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

If the peripheral device «P» conforms to protectiontype Ex ic, no special energy limiter is necessaryRS = standard RS cable, use for circuit [Ex ic] IIC.

Ex hazardous area, zone 2

P

RS

P = Ex ic

5.1.2 The peripheral device «P» (e.g. PC or printer) is located outside the Ex hazardous area

The peripheral device «P» does not need special approval with regard to Ex zone 2.

– In order for the energy flow via the data cable to be limited between the balance and the peripheral device«P», an energy limiter with corresponding Ex certification «B» must be used.

«B» = Standard energy limiter in housing with pro­tection type (Ex ic IIC T5).(If «P» ≠ Ex ic).

Ex hazardous area, zone 2

P

B

Outside hazardous

area

5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous areaIf peripheral devices are connected to the Bluetooth data interface of the balance in the Ex hazardous area, thereare two possible cases.

5.2.1 The peripheral device «P» is located inside the Ex hazardous area

The peripheral device «P» must be approved foruse in zone 2.

Ex hazardous area, zone 2

P

27Connection of Peripheral Devices and System Integration

Page 28: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

28 Connection of Peripheral Devices and System Integration

5.2.2 The peripheral device «P» (e.g. printer) is located outside the Ex hazardous area

The peripheral device «P» does not need specialapproval for use inzone 2.

Ex hazardous area, zone 2

P

Outside hazardous

area

Page 29: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

6 Maintenance

6.1 CleaningPeriodically clean the weighing pan, draft shield element, draft shield (depending on the model), housing andterminal of your balance using a damp cloth. The maintenance interval depends on your standard operatingprocedure (SOP).

Please observe the following notes

WARNING

Risk of electric shock

- The balance must be disconnected from the power supply.

- Balances located in the hazardous area must be damp-cleaned to prevent the risk of sparkformation through dry rubbing on plastic parts.

- Make sure that no liquid penetrates the balance/weighing platform.

- Never open the balance/weighing platform or AC adapter. They do not contain any partsthat can be cleaned, repaired or replaced by the user.

CAUTION

Damage of balance

On no account use cleaning agents which contain solvents or abrasive ingredients, as thiscan result in damage to the terminal overlay.

Cleaning

Your balance is made from high quality, resistant materials and can therefore be cleaned with a commerciallyavailable, mild cleaning agent.

1 Remove the draft shield for thoroughly cleaning the draft shield glasses.

2 When you replace these parts, make sure they are in the correct position.

NotePlease contact your METTLER TOLEDO dealer for details of the available service options. Regular servicing byan authorized service engineer ensures constant accuracy for years to come and prolongs the service life ofyour balance.

29Maintenance

Page 30: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

30 Maintenance

6.2 Cleaning the draft shield (0.1 mg and 1 mg models)

1 Remove the weighing pan, draft-shield element(0.1 mg models), pan support (1 mg models).

2 Lift the draft shield off the balance and place it ona clean surface.

3 Remove the bottom plate.

4 Push all the glasses back as far as they will go.

5 Turn the cover (1) to the front.

6 Pull the top glass toward the back and off.

7 Pull the side glasses toward the back and off.AttentionAlways hold the 2 parallel guided glasses (sideglasses and top glasses) together with one hand.

8 Clean all parts then reassemble the draft shield inthe reverse order.

1

2

2

Insert glasses

AttentionAlways hold the 2 parallel guided glasses (sideglasses and top glasses) together and parallel withone hand. The side glasses must not be placed oust­side the projections (2).

Page 31: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

6.3 Disposal

In conformance with the European Directive 2002/96/EC on Waste Electrical and ElectronicEquipment (WEEE) this device may not be disposed of in domestic waste. This also appliesto countries outside the EU, per their specific requirements.

Please dispose of this product in accordance with local regulations at the collecting pointspecified for electrical and electronic equipment. If you have any questions, please contactthe responsible authority or the distributor from which you purchased this device. Should thisdevice be passed on to other parties (for private or professional use), the content of this regu­lation must also be related.

Thank you for your contribution to environmental protection.

31Maintenance

Page 32: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

32 Maintenance

7 Technical Data

7.1 General data

Technical data AC adapter PSX2, 11132730 (external)Dimensions: 233 x 80 x 60 mmHole distance: 215 x 62 mm (center mounting holes)Weight: 1.42 kgLine voltage: 230 V ±10%, 0.1 A, 50/60 Hz / 115 V (±10%), 0.2 A,

50/60 Hz (country-specific)Fuses: Primary: T125L250V (230 V Version) / T250L250V (115 V Ver­

sion)Secondary: T500L250V

Electrical data: U0 ≤ 13.5 V, I0 ≤ 1.2 A, P0 ≤ 6 WC0 ≤ 4.9 µF, L0 ≤ 30 µH, rectangular characteristics

Applied standards: IEC EN 61010-1, IEC EN 60950-1, CAN/CSA-C22.2 No.61010-1, UL Std No. 61010A-1,EN 61326+A1+A2+A3 (Class B + Industrial environments),FCC Part 15 (Class A),AS/NZS CISPR 22, AS/NZS 61000.4.3ATEX: EN 60079-0 (IEC 60079-0),EN 60079-15 (IEC 60079-15), EN 60079-11 (IEC 60079-11)

Classification: II 3G Ex [ic] nA IIC T5

Application range: • For use only in closed interior rooms• Ex hazardous area, Zone 2

Overvoltage category: Class IIDegree of pollution: 2Degree of protection housing: IP66High above mean sea level: Up to 4000 mAmbient temperature: -10 to +40 °CRelative air humidity: Max. 80% at 31 °C, linearly decreasing to 50% at 40 C, non­

condensingHousing materials: Die-cast aluminium, laquered

Technical data balance / weighing platform

Protection and standardsOvervoltage categoriy: Class IIDegree of pollution: 2Electrical data: Ui ≤ 13.5 V, Ii ≤ 1.2 A, Pi ≤ 6 W, Ci = 0 nF, Li = 0 µHApplied standards: IEC EN 61010-1, CAN/CSA-C22.2 No. 61010-1,

UL Std No. 61010A-1,EN 61326+A1+A2+A3 (Class B + Industrial environments),FCC Part 15 (Class A),AS/NZS 4251.1, AS/NZS 61000 4252.1ATEX: EN 60079-0 (IEC 60079-0),EN 60079-11 (IEC 60079-11)

Classification: II 3G Ex ic IIC T5

Degree of protection: IP44Application range: • For use only in closed and clean interior rooms

• Ex hazardous area, Zone 2

Page 33: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Environmental conditionsHeight above mean sea level: Up to 4000 mAmbient temperature: 5-40 °CRelative air humidity: Max. 80% at 31 °C, linearly decreasing to 50% at 40 °C, non­

condensing

Environmental conditions XS2004SX, XS26003LX, XS64003LXHigh above mean sea level: Up to 4000 mAmbient temperature: 10 - 30 °C ±0.3 °C / 1 h, resp. ±0.5 °C / 12 hRelative air humidity: 40 to 60% ±5% / 4 hWarm-up time: At least 12 hours after connecting the balance to the power sup­

ply.The balance should not be switched into standby mode.

MaterialsHousing: Die-cast aluminium, laquered, plastic and chrome steelTerminal: Die-cast zinc, chromed and plasticsWeighing pan: Chrome-nickel steel ("S" and "M" platform: X2CrNiMo-17-12-2,

"L" platform: X5CrNi18-10)Draft shield: Aluminium, plastic, chrome steel and glassDraft shield element: Die-cast zinc, chromed (10 mg models, "S" platform and

XS2004SX 0.1 mg, "S" platform), Chrome steel X2 Cr Ni Mo17 13 2 (0.1 mg models)

7.2 Model-specif data of the Excellence XS-Ex2 precision balances

7.2.1 XS-Ex2 balances with readability of 0.1 mg / 1 mg, "S" platform with draft shieldXS204SX XS603SX

Limit valuesMaximum capacity 220 g 610 gReadability 0.1 mg 1 mgTare range (from…to) 0…220 g 0…610 gRepeatability (at nominal load) sd 0.1 mg (200 g) 0.9 mgLinearity deviation 0.2 mg 2 mgEccentricity deviation (test load) 0.3 mg (100 g) 3 mg (200 g)Sensitivity offset (test weight) 1 mg (200 g) 4.5 mg (600 g)Sensitivity temperature drift 1) 0.00015%/°C 0.0002%/°CSensitivity stability 0.0002%/a 0.001%/aTypical valuesRepeatability sd 0.04 mg 0.5 mgLinearity deviation 0.13 mg 0.7 mgEccentric deviation (test load) 0.16 mg (100 g) 0.6 mg (200 g)Sensitivity offset (test load) 0.8 mg (200 g) 3 mg (600 g)Minimum sample weight (according to USP) 120 mg 1500 mgMinimum sample weight(U=1%, k=2)

8 mg 100 mg

Settling time 1.5 s 1.5 sDimensionsBalance dimensions (WxDxH) 263x453x322 mm 198x366x276 mmWeighing pan dimensions 78x73 mm (WxD) 127x127 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 0.04mg

+0.000015%·Rgr0.5mg+0.00005%·Rgr

Differential linearity deviation sd √(20pg·Rnt) √(200pg·Rnt)

33Technical Data

Page 34: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

34 Technical Data

XS204SX XS603SXDifferential eccentric load deviation sd 0.00008%·Rnt 0.00015%·RntSensitivity offset sd 0.0002%·Rnt 0.00025%·RntMinimum sample weight (according to USP) 120mg+0.045%·Rgr 1500mg+0.15%·RgrMinimum sample weight(U=1%, k=2)

8mg+0.003%·Rgr 100mg+0.01%·Rgr

Interface update rate 23 /s 23 /sUsable height of draft shield 235 mm 175 mmWeight of balance 9.1 kg 7.6 kgNumber of built-in reference weights 2 1Weights for routine testingOIML Single Weights

Small

10 g F1

#11119528

20 g F1

#11119529

OIML Single Weights

Large

200 g F2

#11118202

500 g F2

#11118203

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

XS1003SX XS5003SXDRLimit valuesMaximum capacity 1010 g 5.1 kgReadability 1 mg 10 mgMaximum capacity, fine range – 1 kgReadability, fine range – 1 mgTare range (from…to) 0…1010 g 0…5.1 kgRepeatability (at nominal load) sd 0.8 mg 6 mgRepeatability, fine range (at nominal load) sd – 1 mgLinearity deviation 2 mg 6 mgEccentricity deviation (test load) 3 mg (500 g) 10 mg (2 kg)Sensitivity offset (test weight) 5 mg (1000 g) 20 mg (5 kg)Sensitivity temperature drift 1) 0.0002%/°C 0.0003%/°CSensitivity stability 0.001%/a 0.0015%/aTypical valuesRepeatability sd 0.4 mg 4 mgRepeatability, fine range sd – 0.6 mgLinearity deviation 0.7 mg 1 mgEccentric deviation (test load) 0.6 mg (500 g) 0.6 mg (2 kg)Sensitivity offset (test load) 3 mg (1000 g) 10 mg (5 kg)Minimum sample weight (according to USP) 1200 mg 12 gMinimum sample weight(U=1%, k=2)

80 mg 800 mg

Minimum sample weight (according to USP),fine range

– 1.8 g

Minimum sample weight(U=1%, k=2), fine range

– 120 mg

Settling time 1.5 s 2 sDimensionsBalance dimensions (WxDxH) 198x366x276 mm 214x366x363 mmWeighing pan dimensions 127x127 mm (WxD) 127x127 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 0.4mg+0.00002%·Rgr 4mg+0.00002%·RgrRepeatability, fine range sd – 0.6mg+0.00002%·RgrDifferential linearity deviation sd √(120pg·Rnt) √(50pg·Rnt)Differential eccentric load deviation sd 0.00006%·Rnt 0.000015%·RntSensitivity offset sd 0.00015%·Rnt 0.0001%·RntMinimum sample weight (according to USP) 1200mg+0.06%·Rgr 12g+0.06%·Rgr

Page 35: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

XS1003SX XS5003SXDRMinimum sample weight (according to USP),fine range

– 1.8g+0.06%·Rgr

Minimum sample weight(U=1%, k=2)

80mg+0.004%·Rgr 800mg+0.004%·Rgr

Minimum sample weight(U=1%, k=2), fine range

– 120mg+0.004%·Rgr

Interface update rate 23 /s 23 /sUsable height of draft shield 175 mm –Weight of balance 7.6 kg 8.1 kgNumber of built-in reference weights 1 1Weights for routine testingOIML Single Weights

Small

50 g F2

#11118196

200 g F2

#11118202

OIML Single Weights

Large

1000 g F2

#11118204

5000 g F2

#11118206

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

7.2.2 XS-Ex2 balances with readability of 0.1 mg, "S" platformXS2004SX

Limit valuesMaximum capacity 2.3 kgReadability 0.1 mgTare range (from…to) 0…2.3 kgRepeatability at nominal load (ABA, measuredat) 2)

0.1 mg (2 kg)

Repeatability at low load (ABA, measurered at)2)

0.08 mg (100 g)

Repeatability (at nominal load) sd 0.35 mgLinearity deviation 1 mgEccentricity deviation (test load) 1.5 mg (1 kg)Sensitivity offset (test weight) 10 mg (2 kg)Sensitivity temperature drift 1) 0.0005%/°CSensitivity stability 0.0025%/aTypical valuesRepeatability sd 0.2 mgRepeatability ABA typical 2) sd 0.06 mgLinearity deviation 0.7 mgEccentric deviation (test load) 0.3 mg (1 kg)Sensitivity offset (test load) 1.6 mg (2 kg)Minimum sample weight (according to USP) 0.6 gMinimum sample weight(U=1%, k=2)

0.04 g

Settling time 10 sDimensionsBalance dimensions (WxDxH) 214x366x96 mmWeighing pan dimensions 127x127 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 0.2mg

+0.0000045%·RgrRepeatability ABA typical 2) sd 0.06mg

+0.0000009%·RgrDifferential linearity deviation sd √(60pg·Rnt)Differential eccentric load deviation sd 0.00003%·RntSensitivity offset sd 0.00008%·Rnt

35Technical Data

Page 36: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

36 Technical Data

XS2004SXMinimum sample weight (according to USP) 600mg+0.0135%·RgrMinimum sample weight(U=1%, k=2)

40mg+0.0009%·Rgr

Interface update rate 23 /sWeight of balance 6.9 kgNumber of built-in reference weights 1Weights for routine testingOIML CarePac

Weights

100 g F2#11118201

2 kg M1#11117722

ASTM CarePac

Weights

100 g ASTM 3#11123576

2 kg ASTM 3#11123580

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

2) Value out of 5 ABA measurements according to OIML R111

7.2.3 XS-Ex2 balances with readability of 10 mg / 0.1 g, "S" platformXS4002SX XS6002SX XS4001SX

Limit valuesMaximum capacity 4.1 kg 6.1 kg 4.1 kgReadability 10 mg 10 mg 100 mgTare range (from…to) 0…4.1 kg 0…6.1 kg 0…4.1 kgRepeatability (at nominal load) sd 8 mg 8 mg 80 mgLinearity deviation 20 mg 20 mg 60 mgEccentricity deviation (test load) 30 mg (2 kg) 30 mg (2 kg) 200 mg (2 kg)Sensitivity offset (test weight) 60 mg (4 kg) 60 mg (6 kg) 240 mg (4 kg)Sensitivity temperature drift 1) 0.0003%/°C 0.0003%/°C 0.0015%/°CSensitivity stability 0.0015%/a 0.0015%/a 0.005%/aTypical valuesRepeatability sd 4 mg 4 mg 40 mgLinearity deviation 7 mg 7 mg 20 mgEccentric deviation (test load) 6 mg (2 kg) 6 mg (2 kg) 32 mg (2 kg)Sensitivity offset (test load) 32 mg (4 kg) 30 mg (6 kg) 160 mg (4 kg)Minimum sample weight (according to USP) 12 g 12 g 120 gMinimum sample weight(U=1%, k=2)

800 mg 800 mg 8 g

Settling time 1.2 s 1.2 s 0.8 sDimensionsBalance dimensions (WxDxH) 194x366x96 mm 194x366x96 mm 194x366x96 mmWeighing pan dimensions 170x205 mm (WxD) 170x205 mm (WxD) 190x223 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 4mg+0.00005%·Rgr 4mg+0.00003%·Rgr 40mg+0.0005%·RgrDifferential linearity deviation sd √(3ng·Rnt) √(2ng·Rnt) √(25ng·Rnt)Differential eccentric load deviation sd 0.00015%·Rnt 0.00015%·Rnt 0.0008%·RntSensitivity offset sd 0.0004%·Rnt 0.00025%·Rnt 0.002%·RntMinimum sample weight (according to USP) 12g+0.15%·Rgr 12g+0.09%·Rgr 120g+1.5%·RgrMinimum sample weight(U=1%, k=2)

800mg+0.01%·Rgr 800mg+0.006%·Rgr 8g+0.1%·Rgr

Interface update rate 23 /s 23 /s 23 /sWeight of balance 6.9 kg 6.9 kg 6.4 kgNumber of built-in reference weights 1 1 1

Page 37: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

XS4002SX XS6002SX XS4001SXWeights for routine testingOIML Single Weights

Small

200 g F2

#11118202

200 g F2

#11118202

200 g F2

#11118202

OIML Single Weights

Large

2000 g F2

#11118205

5000 g F2

#11118206

2000 g F2

#11118205

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

7.2.4 XS-Ex2 balances with readability of 10 mg / 0.1 g, "M" platformXS6002MXDR XS6001MX XS12001MX

Limit valuesMaximum capacity 6.1 kg 6.1 kg 12.1 kgReadability 100 mg 100 mg 1000 mgMaximum capacity, fine range 1.2 kg – 2.4 kgReadability, fine range 10 mg – 100 mgTare range (from…to) 0…6.1 kg 0…6.1 kg 0…12.1 kgRepeatability (at nominal load) sd 60 mg 80 mg 600 mgRepeatability, fine range (at nominal load) sd 10 mg – 80 mgLinearity deviation 60 mg 60 mg 600 mgEccentricity deviation (test load) 100 mg (2 kg) 200 mg (2 kg) 1000 mg (5 kg)Sensitivity offset (test weight) 150 mg (6 kg) 240 mg (6 kg) 600 mg (12 kg)Sensitivity temperature drift 1) 0.0003%/°C 0.0015%/°C 0.0015%/°CSensitivity stability 0.0015%/a 0.005%/a 0.005%/aTypical valuesRepeatability sd 40 mg 40 mg 400 mgRepeatability, fine range sd 6 mg – 40 mgLinearity deviation 7 mg 19 mg 34 mgEccentric deviation (test load) 10 mg (2 kg) 32 mg (2 kg) 30 mg (5 kg)Sensitivity offset (test load) 60 mg (6 kg) 180 mg (6 kg) 290 mg (12 kg)Minimum sample weight (according to USP) 120 g 120 g 1200 gMinimum sample weight(U=1%, k=2)

8 g 8 g 80 g

Minimum sample weight (according to USP),fine range

18 g – 120 g

Minimum sample weight(U=1%, k=2), fine range

1.2 g – 8 g

Settling time 1.5 s 1 s 1 sDimensionsBalance dimensions (WxDxH) 240x393x110 mm 240x393x110 mm 240x393x110 mmWeighing pan dimensions 237x237 mm (WxD) 237x237 mm (WxD) 237x237 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 40mg+0.00015%·Rgr 40mg+0.0003%·Rgr 400mg+0.0008%·RgrRepeatability, fine range sd 6mg+0.00015%·Rgr – 40mg+0.0008%·RgrDifferential linearity deviation sd √(2ng·Rnt) √(15ng·Rnt) √(25ng·Rnt)Differential eccentric load deviation sd 0.00025%·Rnt 0.0008%·Rnt 0.0003%·RntSensitivity offset sd 0.0005%·Rnt 0.0015%·Rnt 0.0012%·RntMinimum sample weight (according to USP) 120g+0.45%·Rgr 120g+0.9%·Rgr 1200g+2.4%·RgrMinimum sample weight (according to USP),fine range

18g+0.45%·Rgr – 120g+2.4%·Rgr

Minimum sample weight(U=1%, k=2)

8g+0.03%·Rgr 8g+0.06%·Rgr 80g+0.16%·Rgr

Minimum sample weight(U=1%, k=2), fine range

1.2g+0.03%·Rgr – 8g+0.16%·Rgr

Interface update rate 23 /s 23 /s 23 /sWeight of balance 8 kg 8 kg 8 kg

37Technical Data

Page 38: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

38 Technical Data

XS6002MXDR XS6001MX XS12001MXNumber of built-in reference weights 1 1 1Weights for routine testingOIML Single Weights

Small

200 g F2

#11118202

200 g F2

#11118202

500 g F2

#11118203

OIML Single Weights

Large

5000 g F2

#11118206

5000 g F2

#11118206

10000 g F2

#11118211

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

The precision balances with "M" platform are no longer available.

7.2.5 XS-Ex2 balances with readability of 1 mg / 5 mg, "L" platformXS26003LX XS64003LX

Limit valuesMaximum capacity 26.1 kg 64.1 kgReadability 1 mg 5 mgTare range (from…to) 0…26.1 kg 0…64.4 kgRepeatability at nominal load (ABA, measuredat) 2)

3 mg (26 kg) 10 mg (60 kg)

Repeatability at low load (ABA, measurered at)2)

2.5 mg (1 kg) 6 mg (5 kg)

Repeatability (at nominal load) sd 5 mg 40 mgLinearity deviation 25 mg 50 mgEccentricity deviation (test load) 0.2 g (8 kg) 0.35 g (24 kg)Sensitivity offset (test weight) 240 mg (24 kg) 960 mg (64 kg)Sensitivity temperature drift 1) 0.0003%/°C 0.0003%/°CSensitivity stability 0.0015%/a 0.0015%/aTypical valuesRepeatability sd 2 mg 4 mgLinearity deviation 17.5 mg 35 mgEccentric deviation (test load) 40 mg (8 kg) 70 mg (24 kg)Sensitivity offset (test load) 48 mg (24 kg) 260 mg (64 kg)Minimum sample weight (according to USP) 6 g 12 gMinimum sample weight(U=1%, k=2)

400 mg 800 mg

Settling time 10 s 10 sDimensionsBalance dimensions (WxDxH) 360x410x147 mm 360x410x147 mmWeighing pan dimensions 258 mm (Ø) 258 mm (Ø)Typical uncertainties and supplementary dataRepeatability sd 2mg+0.000006%·Rgr 4mg+0.000023%·RgrRepeatability ABA typical 2) sd 1.5mg+0.000002%·Rgr 3.5mg

+0.0000094%·RgrDifferential linearity deviation sd √(5pg·Rnt) √(4pg·Rnt)Differential eccentric load deviation sd 0.0001%·Rnt 0.00015%·RntSensitivity offset sd 0.0001%·Rnt 0.0002%·RntMinimum sample weight (according to USP) 6g+0.018%·Rgr 12g+0.069%·RgrMinimum sample weight(U=1%, k=2)

400mg+0.0012%·Rgr 800mg+0.0046%·Rgr

Interface update rate 23 /s 23 /sWeight of balance 15.7 kg 15.7 kgNumber of built-in reference weights 2 2Weights for routine testing

Page 39: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

XS26003LX XS64003LXOIML CarePac

Weights

1 kg F2#11118204

20 kg M1#30024249

2 kg F2#11118205

50 kg M1#11125499

ASTM CarePac

Weights

1 kg ASTM 3#11123579

20 kg ASTM 3#11123583

2 kg ASTM 3#11123580

50 kg ASTM 3#00158741

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

2) Value out of 5 ABA measurements according to OIML R111

7.2.6 XS-Ex2 balances with readability of 0.1 g / 1 g, "L" platformXS32001LX XS32000LX

Limit valuesMaximum capacity 32.1 kg 32.1 kgReadability 100 mg 1000 mgTare range (from…to) 0…32.1 kg 0…32.1 kgRepeatability (at nominal load) sd 80 mg 600 mgLinearity deviation 300 mg 600 mgEccentricity deviation (test load) 300 mg (10 kg) 1000 mg (10 kg)Sensitivity offset (test weight) 960 mg (32 kg) 1920 mg (32 kg)Sensitivity temperature drift 1) 0.001%/°C 0.0015%/°CSensitivity stability 0.003%/a 0.005%/aTypical valuesRepeatability sd 40 mg 400 mgLinearity deviation 200 mg 400 mgEccentric deviation (test load) 200 mg (10 kg) 600 mg (10 kg)Sensitivity offset (test load) 320 mg (32 kg) 650 mg (32 kg)Minimum sample weight (according to USP) 120 g 1200 gMinimum sample weight(U=1%, k=2)

8 g 80 g

Settling time 1.5 s 1.2 sDimensionsBalance dimensions (WxDxH) 360x404x130 mm 360x404x130 mm Weighing pan dimensions 360x280 mm (WxD) 360x280 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 40mg+0.00006%·Rgr 400mg+0.0003%·RgrDifferential linearity deviation sd √(300ng·Rnt) √(1.2ug·Rnt)Differential eccentric load deviation sd 0.001%·Rnt 0.003%·RntSensitivity offset sd 0.0005%·Rnt 0.001%·RntMinimum sample weight (according to USP) 120g+0.18%·Rgr 1200g+0.9%·RgrMinimum sample weight(U=1%, k=2)

8g+0.012%·Rgr 80g+0.06%·Rgr

Interface update rate 23 /s 23 /sWeight of balance 12.4 kg 12.4 kgNumber of built-in reference weights 2 2Weights for routine testingOIML Single Weights

Small

1000 g F2

#11118204

1000 g F2

#11118204

OIML Single Weights

Large

20000 g F2

#11118212

20000 g F2

#11118212

sd = Standard deviation Rnt = Net weight (sample weight)

39Technical Data

Page 40: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

40 Technical Data

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

XS64001LX XS64000LXLimit valuesMaximum capacity 64.1 kg 64.1 kgReadability 100 mg 1000 mgTare range (from…to) 0…64.1 kg 0…64.1 kgRepeatability (at nominal load) sd 100 mg 600 mgLinearity deviation 500 mg 600 mgEccentricity deviation (test load) 500 mg (20 kg) 1000 mg (20 kg)Sensitivity offset (test weight) 1280 mg (64 kg) 1920 mg (64 kg)Sensitivity temperature drift 1) 0.001%/°C 0.0015%/°CSensitivity stability 0.005%/a 0.003%/aTypical valuesRepeatability sd 40 mg 400 mgLinearity deviation 320 mg 400 mgEccentric deviation (test load) 320 mg (20 kg) 600 mg (20 kg)Sensitivity offset (test load) 380 mg (64 kg) 650 mg (64 kg)Minimum sample weight (according to USP) 120 g 1200 gMinimum sample weight(U=1%, k=2)

8 g 80 g

Settling time 1.8 s 1.5 sDimensionsBalance dimensions (WxDxH) 362x404x131 mm 362x404x131 mm Weighing pan dimensions 362x282 mm (WxD) 362x282 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 40mg+0.00006%·Rgr 400mg+0.0003%·RgrDifferential linearity deviation sd √(400ng·Rnt) √(600ng·Rnt)Differential eccentric load deviation sd 0.0008%·Rnt 0.0015%·RntSensitivity offset sd 0.0003%·Rnt 0.0005%·RntMinimum sample weight (according to USP) 120g+0.18%·Rgr 1200g+0.9%·RgrMinimum sample weight(U=1%, k=2)

8g+0.012%·Rgr 80g+0.06%·Rgr

Interface update rate 23 /s 23 /sWeight of balance 14.1 kg 14.1 kgNumber of built-in reference weights 2 2Weights for routine testingOIML Single Weights

Small

2000 g F2

#11118205

2000 g F2

#11118205

OIML Single Weights

Large

50000 g F2

#11116659

50000 g F2

#11116659

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

7.3 Model-specif data of the Excellence X-Ex2 precision balances

7.3.1 X-Ex2 balances with readability of 0.1 mg / 1 mg, "S" platform with draft shieldX204SX X603SX

Limit valuesMaximum capacity 210 g 610 gReadability 0.1 mg 1 mgTare range (from…to) 0…210 g 0…610 gRepeatability (at nominal load) sd 0.2 mg 0.9 mgLinearity deviation 0.2 mg 2 mg

Page 41: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

X204SX X603SXEccentricity deviation (test load) 0.3 mg (100 g) 3 mg (200 g)Sensitivity offset (test weight) 1 mg (200 g) 4.5 mg (600 g)Sensitivity temperature drift 1) 0.00015%/°C 0.0002%/°CSensitivity stability 0.00025%/a 0.001%/aTypical valuesRepeatability sd 0.12 mg 0.5 mgLinearity deviation 0.07 mg 0.7 mgEccentric deviation (test load) 0.08 mg (100 g) 0.8 mg (200 g)Sensitivity offset (test load) 0.4 mg (200 g) 2.4 mg (600 g)Minimum sample weight (according to USP) 360 mg 1500 mgMinimum sample weight(U=1%, k=2)

24 mg 100 mg

Settling time 2 s 1.5 sDimensionsBalance dimensions (WxDxH) 214x260x363 mm 198x366x276 mmWeighing pan dimensions 90 mm (Ø) 127x127 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 0.12mg

+0.000015%·Rgr0.5mg+0.000025%·Rgr

Differential linearity deviation sd √(6pg·Rnt) √(200pg·Rnt)Differential eccentric load deviation sd 0.00004%·Rnt 0.0002%·RntSensitivity offset sd 0.0001%·Rnt 0.0002%·RntMinimum sample weight (according to USP) 360mg+0.045%·Rgr 1500mg+0.075%·RgrMinimum sample weight(U=1%, k=2)

24mg+0.003%·Rgr 100mg+0.005%·Rgr

Interface update rate 23 /s 23 /sUsable height of draft shield 248 mm 175 mmWeight of balance 7 kg 6.5 kgNumber of built-in reference weights 1 1

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

X1003SX X5003SXDRLimit valuesMaximum capacity 1010 g 5.1 kgReadability 1 mg 10 mgMaximum capacity, fine range – 1 kgReadability, fine range – 1 mgTare range (from…to) 0…1010 g 0…5.1 kgRepeatability (at nominal load) sd 0.8 mg 6 mgRepeatability, fine range (at nominal load) sd – 1 mgLinearity deviation 2 mg 6 mgEccentricity deviation (test load) 3 mg (500 g) 10 mg (2 kg)Sensitivity offset (test weight) 5 mg (1000 g) 20 mg (5 kg)Sensitivity temperature drift 1) 0.0002%/°C 0.0003%/°CSensitivity stability 0.001%/a 0.0015%/aTypical valuesRepeatability sd 0.4 mg 4 mgRepeatability, fine range sd – 0.6 mgLinearity deviation 0.7 mg 1 mgEccentric deviation (test load) 0.6 mg (500 g) 0.6 mg (2 kg)Sensitivity offset (test load) 3 mg (1000 g) 10 mg (5 kg)Minimum sample weight (according to USP) 1200 mg 12 gMinimum sample weight(U=1%, k=2)

80 mg 800 mg

Minimum sample weight (according to USP),fine range

– 1.8 g

41Technical Data

Page 42: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

42 Technical Data

X1003SX X5003SXDRMinimum sample weight(U=1%, k=2), fine range

– 120 mg

Settling time 1.5 s 2 sDimensionsBalance dimensions (WxDxH) 214x260x363 mm 198257x276 mmWeighing pan dimensions 127x127 mm (WxD) 127x127 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 0.4mg+0.00002%·Rgr 4mg+0.00002%·RgrRepeatability, fine range sd – 0.6mg+0.00002%·RgrDifferential linearity deviation sd √(120pg·Rnt) √(50pg·Rnt)Differential eccentric load deviation sd 0.00006%·Rnt 0.000015%·RntSensitivity offset sd 0.00015%·Rnt 0.0001%·RntMinimum sample weight (according to USP) 1200mg+0.06%·Rgr 12g+0.06%·RgrMinimum sample weight (according to USP),fine range

– 1.8g+0.06%·Rgr

Minimum sample weight(U=1%, k=2)

80mg+0.004%·Rgr 800mg+0.004%·Rgr

Minimum sample weight(U=1%, k=2), fine range

– 120mg+0.004%·Rgr

Interface update rate 23 /s 23 /sUsable height of draft shield 248 mm 175 mmWeight of balance 7 kg 6.8 kgNumber of built-in reference weights 1 1

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

7.3.2 X-Ex2 balances with readability of 10 mg / 0.1 g, "S" platformX4002SX X6002SX X4001SX

Limit valuesMaximum capacity 4.1 kg 6.1 kg 4.1 kgReadability 10 mg 10 mg 100 mgTare range (from…to) 0…4.1 kg 0…6.1 kg 0…4.1 kgRepeatability (at nominal load) sd 8 mg 8 mg 80 mgLinearity deviation 20 mg 20 mg 60 mgEccentricity deviation (test load) 30 mg (2 kg) 30 mg (2 kg) 200 mg (2 kg)Sensitivity offset (test weight) 60 mg (4 kg) 60 mg (6 kg) 240 mg (4 kg)Sensitivity temperature drift 1) 0.0003%/°C 0.0003%/°C 0.0015%/°CSensitivity stability 0.0015%/a 0.0015%/a 0.005%/aTypical valuesRepeatability sd 4 mg 4 mg 40 mgLinearity deviation 7 mg 7 mg 20 mgEccentric deviation (test load) 8 mg (2 kg) 8 mg (2 kg) 32 mg (2 kg)Sensitivity offset (test load) 32 mg (4 kg) 30 mg (6 kg) 160 mg (4 kg)Minimum sample weight (according to USP) 12 g 12 g 120 gMinimum sample weight(U=1%, k=2)

800 mg 800 mg 8 g

Settling time 1.2 s 1.2 s 0.8 sDimensionsBalance dimensions (WxDxH) 194x257x96 mm 194x257x96 mm 194x257x96 mmWeighing pan dimensions 170x205 mm (WxD) 170x205 mm (WxD) 190x223 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 4mg+0.00005%·Rgr 4mg+0.00003%·Rgr 40mg+0.0005%·RgrDifferential linearity deviation sd √(3ng·Rnt) √(2ng·Rnt) √(25ng·Rnt)Differential eccentric load deviation sd 0.0002%·Rnt 0.0002%·Rnt 0.0008%·RntSensitivity offset sd 0.0004%·Rnt 0.00025%·Rnt 0.002%·RntMinimum sample weight (according to USP) 12g+0.15%·Rgr 12g+0.09%·Rgr 120g+1.5%·Rgr

Page 43: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

X4002SX X6002SX X4001SXMinimum sample weight(U=1%, k=2)

800mg+0.01%·Rgr 800mg+0.006%·Rgr 8g+0.1%·Rgr

Interface update rate 23 /s 23 /s 23 /sWeight of balance 5.4 kg 5.4 kg 5.4 kgNumber of built-in reference weights 1 1 1

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

7.3.3 X-Ex2 balances with readability of 10 mg / 0.1 g, "M" platformX6001MX X12001MX

Limit valuesMaximum capacity 6.1 kg 12.1 kgReadability 100 mg 100 mgTare range (from…to) 0…6.1 kg 0…12.1 kgRepeatability (at nominal load) sd 80 mg 80 mgLinearity deviation 60 mg 100 mgEccentricity deviation (test load) 200 mg (2 kg) 200 mg (5 kg)Sensitivity offset (test weight) 240 mg (6 kg) 600 mg (12 kg)Sensitivity temperature drift 1) 0.0015%/°C 0.0015%/°CSensitivity stability 0.005%/a 0.005%/aTypical valuesRepeatability sd 40 mg 40 mgLinearity deviation 19 mg 34 mgEccentric deviation (test load) 32 mg (2 kg) 30 mg (5 kg)Sensitivity offset (test load) 180 mg (6 kg) 290 mg (12 kg)Minimum sample weight (according to USP) 120 g 120 gMinimum sample weight(U=1%, k=2)

8 g 8 g

Settling time 1 s 1.2 sDimensionsBalance dimensions (WxDxH) 240x278x110 mm 240x278x110 mmWeighing pan dimensions 237x237 mm (WxD) 237x237 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 40mg+0.0003%·Rgr 40mg+0.00015%·RgrDifferential linearity deviation sd √(15ng·Rnt) √(25ng·Rnt)Differential eccentric load deviation sd 0.0008%·Rnt 0.0003%·RntSensitivity offset sd 0.0015%·Rnt 0.0012%·RntMinimum sample weight (according to USP) 120g+0.9%·Rgr 120g+0.45%·RgrMinimum sample weight(U=1%, k=2)

8g+0.06%·Rgr 8g+0.03%·Rgr

Interface update rate 23 /s 23 /sWeight of balance 6.9 kg 6.9 kgNumber of built-in reference weights 1 1

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

The precision balances with "M" platform are no longer available.

7.3.4 X-Ex2 balances with readability of 0.1 g / 1 g, "L" platformX32001LX X64001LX

Limit valuesMaximum capacity 32.1 kg 64.1 kgReadability 100 mg 100 mg

43Technical Data

Page 44: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

44 Technical Data

X32001LX X64001LXTare range (from…to) 0…32.1 kg 0…64.1 kgRepeatability (at nominal load) sd 80 mg 100 mgLinearity deviation 300 mg 500 mgEccentricity deviation (test load) 300 mg (10 kg) 500 mg (20 kg)Sensitivity offset (test weight) 960 mg (32 kg) 1280 mg (64 kg)Sensitivity temperature drift 1) 0.001%/°C 0.001%/°CSensitivity stability 0.003%/a 0.005%/aTypical valuesRepeatability sd 40 mg 40 mgLinearity deviation 200 mg 320 mgEccentric deviation (test load) 200 mg (10 kg) 320 mg (20 kg)Sensitivity offset (test load) 320 mg (32 kg) 380 mg (64 kg)Minimum sample weight (according to USP) 120 g 120 gMinimum sample weight(U=1%, k=2)

8 g 8 g

Settling time 1.5 s 1.8 sDimensionsBalance dimensions (WxDxH) 360x280x130 mm 360x280x130 mmWeighing pan dimensions 360x280 mm (WxD) 360x282 mm (WxD)Typical uncertainties and supplementary dataRepeatability sd 40mg+0.00006%·Rgr 40mg+0.00006%·RgrDifferential linearity deviation sd √(300ng·Rnt) √(400ng·Rnt)Differential eccentric load deviation sd 0.001%·Rnt 0.0008%·RntSensitivity offset sd 0.0005%·Rnt 0.0003%·RntMinimum sample weight (according to USP) 120g+0.18%·Rgr 120g+0.18%·RgrMinimum sample weight(U=1%, k=2)

8g+0.012%·Rgr 8g+0.012%·Rgr

Interface update rate 23 /s 23 /sWeight of balance 12.4 kg 14.1 kgNumber of built-in reference weights 2 2

sd = Standard deviation Rnt = Net weight (sample weight)

Rgr = Gross weight a = Year (annum)

1) In the temperature range 10…30 °C

Page 45: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.4 Dimensions of the Excellence XS-Ex2 balances

7.4.1 XS-Ex2 balances with readability of 0.1 mg, "S" platform with draft shield

17

8

15

6

19

9

21

4

17

2.5

257

95

251

177

156

72

197128

366

58

77 9

5

24

8

35

6

36

3

3

115

Ø 90

Models: XS204SX

45Technical Data

Page 46: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

46 Technical Data

7.4.2 XS-Ex2 balances with readability of 1 mg, "S" platform with draft shield

17

8

127

127

156

199

214

172

.5

257

95

251

177

156

72

197128

366

58

77 9

5

248

356

363

3

115

Models: XS603SX, XS1003SX, XS5003SXDR

Page 47: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.4.3 XS-Ex2 balances with readability of 10 mg, "S" platform with draft shield element

170

156

194

257

226

72

22197128

115

366

178

263

168

58

77 88 96

205

251

25

Models: XS4002SX, XS6002SX

47Technical Data

Page 48: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

48 Technical Data

7.4.4 XS-Ex2 balances with readability of 0.1 mg, "S" platform

127

156

194

257

226

72

22197128

115

366

178

263

168

58

77 88 96

127

251

25

Models: XS2004SX

Page 49: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.4.5 XS-Ex2 balances with readability of 0.1 g, "S" platform

190

156

194

226

72

22197128

115

366

178

263

168

58

77 96

223

257

251

25

Models: XS4001SX

49Technical Data

Page 50: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

50 Technical Data

7.4.6 XS-Ex2 balances with readability of 10 mg / 0.1 g, "M" platform

214

77.8

178

58

77

110

202

240

278

393

237

206.5135 34

237

115

309

32

Models: XS6001MX, XS6002MXDE, XS12001MX

Page 51: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.4.7 XS-Ex2 balances with readability of 1 mg / 5 mg, "L" platform

280124

410

296

194 15

360

45

142

46

216156

124

77

174

147

258

Models: XS26003LX, XS64003LX

51Technical Data

Page 52: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

52 Technical Data

7.4.8 XS-Ex2 balances with readability of 0.1 g / 1 g, "L" platform

280124

404

296

194 15

360

68

130

46

216156

124

77

174

Models: XS32001LX, XS64001LX, XS32000LX, XS64000LX

Page 53: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.5 Dimensions of the Excellence X-Ex2 weighing platforms

7.5.1 X-Ex2 weighing platforms with readability of 0.1 mg, "S" platform with draft shield

15

6

19

9

21

4

17

2.5

257 3

95

251

177

156

Ø 90

72

197

58

95

24

8

35

6

36

3

Models: X204SX

53Technical Data

Page 54: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

54 Technical Data

7.5.2 X-Ex2 weighing platforms with readability of 1 mg, "S" platform with draft shield

127

127

156

199

214

172.5

257 3

95

251

177

156

72

197

58

95

248

356

363

Models: X603SX, X1003SX, X5003SXDR

Page 55: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.5.3 X-Ex2 weighing platforms with readability of 10 mg, "S" platform with draft shield ele­ment

170

156

194

257

226

72

22197

263

168

58

88 96

205

25

Models: X4002SX, X6002SX

55Technical Data

Page 56: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

56 Technical Data

7.5.4 X-Ex2 weighing platforms with readability of 0.1 g, "S" platform

190

156

194

257

226

72

22197

263

168

58

96

223

25

Models: X4001SX

Page 57: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.5.5 X-Ex2 weighing platforms with readability of 0.1 g, "M" platform

214

77.8

58

110

20

2

240

309

278

237 x 237

206.5 34

32

Models: X6001MX, X12001MX

57Technical Data

Page 58: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

58 Technical Data

7.5.6 X-Ex2 weighing platforms with readability of 0.1 g, "L" platform

280

29615

360

68

130

46

216

Models: X32001LX, X64001LX

Page 59: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

7.6 Dimensions of the AC adapter PSX2Dimensions in mm

2.5

25

60

62

215

233

ø 5.8

9

80

AC Adapter PSX2

59Technical Data

Page 60: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

60 Technical Data

7.7 Interfaces

7.7.1 Specifications of RS232CInterface type: Voltage interface according to EIA RS-232C/DIN 66020 (CCITT V24/V.28)Max. cable length: 15 mSignal level: Outputs:

+5 V ... +15 V (RL = 3 – 7 kΩ)

–5 V ... –15 V (RL = 3 – 7 kΩ)

Inputs:

+3 V ... 25 V

–3 V ... 25 V

Connector: Sub-D, 9-pole, femaleOperating mode: Full duplexTransmission mode: Bit-serial, asynchronousTransmission code: ASCIIBaud rates: 600, 1200, 2400, 4800, 9600, 19200, 384001) (firmware selectable)Bits/parity: 7-bit/even, 7-bit/odd, 7-bit/none, 8-bit/none (firmware selectable)Stop bits: 1 stop bitHandshake: None, XON/XOFF, RTS/CTS (firmware selectable)End-of-line: <CR><LF>, <CR>, <LF> (firmware selectable)

6

15

9

DataGND

Handshake

Pin 2: Balance transmit line (TxD)

Pin 3: Balance receive line (RxD)

Pin 5: Ground signal (GND)

Pin 7: Clear to send (hardware handshake) (CTS)

Pin 8: Request to send (hardware handshake) (RTS)

1) 38400 baud is only possible in special cases, such as:• Weighing platform without terminal, or• Weighing platform with terminal, only via the optional RS232C interface.

Page 61: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

8 Accessories and Spare Parts

8.1 AccessoriesYou can increase the functionality of your balance with accessories from the METTLER TOLEDO range. The fol­lowing options are available:

Description Part No.

PrintersIt is not permitted to use these units in Ex hazadous areas!

BT-P42 printer with Bluetooth connection to instrument 11132540Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388

Printer

17.375 g

19.319 g

8.003 g

7.773 g

6.554 g

10.506 g

8.097 g

5.876 g

3.205 g

1.098 g

METTLER TOLEDO

Ribbon cartridge, black, set of 2 pcs 00065975

RS-P42 printer with RS232C connection to instrument 00229265Paper roll, set of 5 pcs 00072456Paper roll, self-adhesive, set of 3 pcs 11600388

Printer

17.375 g

19.319 g

8.003 g

7.773 g

6.554 g

10.506 g

8.097 g

5.876 g

3.205 g

1.098 g

METTLER TOLEDO

Ribbon cartridge, black, set of 2 pcs 00065975

Optional interfaces

Second RS232C Interface 11132500

Ethernet Interface for connection to an Ethernet network 11132515

BT option: Bluetooth Interface for multipoint connection for upto 6 Bluetooth devices (XS-Ex2 models only)

11132530

BTS option: Bluetooth Interface, single-point connection 11132535

61Accessories and Spare Parts

Page 62: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

62 Accessories and Spare Parts

Auxiliary displaysIt is not permitted to use these units in Ex hazadous areas!

Displays only the weight value and unit, if defined.

BT-BLD Bluetooth auxiliary display for table mounting,168 mm, LCD display with backlighting

11132555

LC/RS-BLD auxiliary display on bench stand, backlit (incl. RScable and separate AC adapter)

00224200

RS/LC-BLDS auxiliary display for table or balance mounting,480 mm, LCD display with backlighting

11132630

Sensors

ErgoSens, optical sensor for hands-free operation 11132601

Density determination

Kit for density determination of solids and liquids (for 1 mgand 0.1 mg models)

11132680

Sinker for density of liquids in conjunction with Density Kit 00210260Calibrated (sinker + certificate) 00210672Recalibrated (new certificate) 00210674

Calibrated thermometer with certificate 11132685

Page 63: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Anti-theft devices

Steel cable 11600361

AC adapter PSX2

AC adapter PSX2 (not included in the scope of delivery for thebalance/weighing platform)

11132730

Transport cases

Transport case for precision balances for 10 mg and 0.1 gmodels, "S" platform

11132595

Software

LabX software for one click™ weighing solutionsEnables you to perform One Click™ Standard Preparation, OneClick™ Loss on Drying, One Click™ Sieve Analysis and manyother applications.Simply start the method with the One Click™ shortcut on thebalance touchscreen. LabX guides you step-by-step throughthe SOP on the balance, performs your calculations automati­cally, and takes care of saving all your data. The completesolution can be tailored to match your process requirements.Visit www.mt.com/one-click-weighing for more information

on request

Production certificate

Production Certificate "PRO" 11106895

Weighing pans

MPS (Magnetic Protection Shield) weighing pan for 0.1 gmodels 190 x 223 mm, "S" platform

11132625

63Accessories and Spare Parts

Page 64: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

64 Accessories and Spare Parts

MPS (Magnetic Protection Shield) weighing pan for 10 mgmodels 170 x 205 mm, "S" platform

11132626

Weighing pan 190 x 223 mm, incl. pan support, "S" platform 11132655

Weighing pan 170 x 205 mm, incl. pan support and and draftshield element, "S" platform

11132660

Draft shields

"MagicCube" Draft shield glass free for 0.1 mg and 1 mg mod­els, usable height 175 mm, "S" platform

11131650

Simple Draft shield glass free for 0.1 g and 10 mg models,usable height 175 mm, "S" platform

11131653

For 0.1 g model the weighing pan set must be ordered addi­tionally

11132660

Draft shield to cover the entire balance "XP-W12"300 x 450 x 450 mm (W x D x H), "S" and "M" platform

11134430

Draft shield to cover the entire balance "XP-W64"550 x 470 x 580 mm (W x D x H), "M" and "L" platform

11134470

Weighing below the balance

Hook for the hanger (depending on the model), "M" and "L"platform

11132565

Page 65: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Various

Terminal stand for placement of the terminal 30 cm above theweighing pan, "S" and "M" platform

11132636

Terminal stand for placement of the terminal 30 cm above theweighing pan, "L" platform

11132653

65Accessories and Spare Parts

Page 66: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

66 Accessories and Spare Parts

8.2 Spare partsDraft shield, "S" platform (0.1 mg + 1 mg)

Pos Description Part No.1 Door front left 111330802 Door back left 111330793 Door top front 111330824 Door top back 111330815 Door back right 111330776 Door front right 111330787 Draft shield element, 0.1 mg 11131531

1

2

3

4

5

6

7

8

8 Ring seal, 0.1 mg 11131551

VariousPos Description Part No.Weighing pans, "S" platform

1 0.1 mg Weighing pan 90 mm 111330642 1 mg Weighing pan

127 x 127 mm11131022

3 10 mg Weighing pan170 x 205 mm

11131030

3 0.1 g Weighing pan190 x 223 mm

11131031

4 10 mg Draft shield element 11131040Weighing pans, "M" platform

3 all Weighing pan237 x 237 mm

11131173

Weighing pans, "L" platform 3 ≤ 32 kg Weighing pan

280 x 360 mm00239105

3 64 kg Weighing pan280 x 360 mm

11102124

Various5 ≥ 10 mg Pan support, "S" and "M"

platform11131029

5 Pan support, "L" platform 00239104

12

3

4

5

6

6 Foot screw, "S" and "M"platform

11106323

Page 67: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

PackagingPos Description Part No."S" Weighing platform, Type X/XSxx4SX (0.1 mg)

Packaging 11133052Export box 11132834Packaging Draft shield 11133054Export carton Draft shield 11132867

"S" Weighing platform, Type X/XSxx3SX (1 mg)Packaging 11133053Export box 11132834Packaging Draft shield 11133054Export carton Draft shield 11132867

"S" Weighing platform, Type X/XSxx2SX (10 mg)Packaging 11133050Export box 11132839

"S" Weighing platform, Type X/XSxx1SX (0.1 g)Packaging 11133051Export box 11132839

"M" Weighing platform, Type X/XSxxxMXPackaging 11133056Export box 11132879

"L" Weighing platform, Type X/XSxxxxLXPackaging 11133057Export box 11132912

67Accessories and Spare Parts

Page 68: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

68 Accessories and Spare Parts

9 Certificate of the Issuing Institution

9.1 Power supply PSX2

Page 69: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

69Certificate of the Issuing Institution

Page 70: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

70 Certificate of the Issuing Institution

9.2 Balances XS-Ex2 and platforms X-Ex2

Page 71: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

71Certificate of the Issuing Institution

Page 72: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

72 Certificate of the Issuing Institution

Page 73: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

10 Appendix

10.1 MT-SICS interface commands and functionsMany of the instruments and balances used have to be capable of integration in a complex computer or dataacquisition system.

To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full,most balance functions are also available as appropriate commands via the data interface.

All new METTLER TOLEDO balances launched on the market support the standardized command set "METTLERTOLEDO Standard Interface Command Set" (MT-SICS). The commands available depending on the functionalityof the balance.

For further information please refer to the Reference Manual MT-SICS downloadable from the Internet under

u www.mt.com/xs-ex

73Appendix

Page 74: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

11 Index

AAC adapter PSX2 7, 32Accessories 61

BBluetooth 27

CCertificate 68, 70Cleaning 29Cleaning the draft shield 30Connection of peripheral devices 26Conventions 6Country-specific AC adapter 22

DDimensions 45, 53, 59Dismantling 24Disposal 31

EEnvironmental conditions 33

GGeneral data 32General safety information 7GLP 5Good Laboratory Practice 5

HHazardous areas 7

IInstallation of the AC AdapterPSX2

22

Installing the draft shield andweighing pan

17

InterfaceMT-SICS 73

ISO 14001 5ISO 9001 5

LLevel indicator 21Leveling 21Location 21

MMarking the balance/weighingplatform

24

Materials 33Model-specif data 33MT-SICS 73

PPeripheral devices 26Placing the terminal 20

Power supply PSX2 68Power supply voltage 8, 22Protection and standards 32

RRS232C interface 60

SSafety information

General 7Staff safety 8

safety information 7Safety instructions regardinginstallation

9

Safety measures during opera­tion

9

Scope of delivery 15Setting up 15Spare parts 67Special conditions 10Staff safety 8System integration 26

TTechnical data 32

UUnpacking 17Unpacking the draft shield 17Usage of data interface "RS232" 26Usage of the optional data inter­face "Bluetooth"

27

WWeighing platform 17

Index74

Page 75: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located
Page 76: Operating Instructions Excellence Precision Balances XS ...5.2 Usage of the optional data interface "Bluetooth" in the Ex hazardous area 27 5.2.1 The peripheral device «P» is located

Mettler-Toledo AG, Laboratory & Weighing TechnologiesCH-8606 Greifensee, SwitzerlandTel. +41 (0)44 944 22 11Fax +41 (0)44 944 30 60www.mt.com

Subject to technical changes.© Mettler-Toledo AG 05/201311781111D en