Operating Instructions
DULCOMETER® D1CPart 1: Mounting and installation instructions for wall-mounted and
control panel-mounted devices
ProM
inen
t®
T. Nr./P. No.: 987725 ProMinent Dosiertechnik GmbH · 69123 Heidelberg · Germany BA DM 123 04/03 G/GB/F/E
Identcode
D1C A ___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___ ___
pH
7.43 pH
START
STOP
Temp.: 24.7 ° CDULCOMETER
pH
7.20pH
START
STOP
DULCOMETER
DED1C1S001 DED1C1W001
G/GB/F/E
Typ/type D
2
Betriebsanleitung in deutschvon Seite 1 bis 21
Operating Instructions in Englishfrom Page 23 to Page 41
D
GB
E
23
1 Contents / General User Information
Please completely read through these operating instructions. Do not discard!The warranty shall be invalidated by damage caused by operating errors!
Page
General User Information ............................................................................................................................ 23
Device Identification / Identity Code ........................................................................................................... 24
Device Overview / Controls ......................................................................................................................... 25
Functional Description ................................................................................................................................. 26
Mounting / Installation .................................................................................................................................. 27
Safety information ............................................................................................................................... 27
Technical Data .............................................................................................................................................. 33
Maintenance / Repair ................................................................................................................................... 37
Applicable Types of Enclosure / Standards ................................................................................................ 39
Spare Parts / Accessories ............................................................................................................................ 40
Used Part Disposal ....................................................................................................................................... 40
EC Declaration of Conformity ...................................................................................................................... 41
Overview of terminal arrangement ............................................................................................................... 42
Terminal diagrams ........................................................................................................................................ 44
General User InformationPlease read the following information carefully and thoroughly. Knowledge of this information will greatlyincrease the benefit you gain from the operating instructions.Particular attention is drawn to:• Lists
Instructions
Setting menus
NOTEThe information provided in a note is intended to make your work easier.
and safety information:
WARNINGThis symbol draws attention to possible hazardous situations. Disregard of this informationmay result in the direct threat to life and serious injuries.
CAUTIONThis symbol draws attention to a possibly dangerous situation. Disregard of thisinformation may result in serious injuries or damage to property.
IMPORTANTThis symbol is used to draw attention to possible damaging situations. Disregard of thisinformation may result in damage to property.
SAFETY INFORMATION
CAUTION• Please observe the parts of these operating instructions applicable to your specific type
of equipment! Applicable parts are listed in the device identification/ID code list!• Correct measurement and metering is possible only with the probe in perfect working
order! The probe must be calibrated/checked at regular intervals!Probe failure may cause the uncontrolled metering of chemicals.
24
2 Device Identification / Identity Code
D1C A
Please enter the identity code of your device here.
D1C DULCOMETER® Controller Series D1C
W Wall mountingD Control panel installation, 96 x 96 mm
Operating voltage0 230 V 50/60 Hz1 115 V 50/60 Hz2 200 V 50/60 Hz (control panel installation only)3 100 V 50/60 Hz (control panel installation only)4 24 V AC/DC
Measured variableB Bromine (0.1...13 mg/l) (3rd quarter 2001)C Chlorine (0...0.5/2/10/20 mg/l)L Conductivity (0…20/200/2000 µS/cm; 0…20/200 mS/cm)P pH (0...14 pH)R Redox (-1000...1000 mV)H Hydrogen peroxide (1…20; 10…200; 100…2000 mg/l)S Standard signal (0/4-20 mA)A Peracetic acid (10…200; 100…2000 mg/l)D Chlorine dioxide (0...0.5/2/10/20 mg/l)Z Ozone (0...2 mg/l)X Dissolved Oxygen (0.1...10/20 ppm)T Temperature (0...100 °C)
Connection of measured variable1 Terminal, standard signal 0/4-20 mA (signal converters necessary for controllers
with measured variable connection, standard signal 0/4-20 mA2 SN6 connector for P or R3 Terminal for L4 Terminal PT 100 for T5 Terminal mV for P or R
Correction variable0 None1 pH for chlorine2 Temperature for P, D, H, A or L via terminal3 Temperature for P, D, H, A or L via standard signal 0/4-20 mA4 manual temperature entry for P, H, A or L
Feedforward control0 None1 Flow as standard signal 0/4-20 mA2 Flow as frequency 0 - 500 Hz3 Flow as frequency 0 - 10 Hz
Control input0 None1 Pause
Signal output0 None1 Standard signal 0/4-20 mA measured value2 Standard signal 0/4-20 mA controlled variable3 Standard signal 0/4-20 mA correction variable4 2 standard signal outputs 0/4-20 mA, freely programmable (not for H, A)
Power controlG Alarm and 2 limit value relaysM Alarm and 2 solenoid valve relays (pulse length control)R Alarm relay and actuator with feedback
Pump control0 None2 Two pumps
Control characteristic0 None1 Proportional control2 PID control
Log output0 None
LanguageD German (E, F, N)E English (D, F, N)F French (D, E, N)I Italian (D, F, S)N Dutch (D, E, F)S Spanish (D, I, F)P Polish (D, E, A)A Swedish (D, E, P)B Portuguese (E, S, F)U Hungarian (P, A, E)G Czech
25
®
DULCOMETER®
STOPSTART
Display fieldmeasuredvariable
Graphic display
START/STOPButton
ENTERButton
UPButton
CHANGEButton
DOWNButton
RETURNButton
D1C
CHANGE button
To change over within a menu leveland to change from one variable toanother within a menu point.
START/STOP button
Start/stop of control and meteringfunction
ENTER button
To accept, confirm or save a displayedvalue or status. For alarm acknowl-edgement.
3 Device Overview / Controls
D1C2-Cl-002-GB
STOPSTART
UP button
To increase a displayed numericalvalue and to change variables (flashingdisplay).
RETURN button
To exit operating menu (back to startof relevant setting menu).
DOWN button
To decrease a displayed numericalvalue and to change variables (flashingdisplay).
26
4.1 Brief functional description
The DULCOMETER® D1C is a device designed for measuring, displaying and controlling measuresvariables. With the corresponding expansion stage it can also process disturbance variables.
The measured variables to be processed are:- pH, ORP- Standard signal, Temperature- Dissolved Oxygen- Chlorine, ClO2, Br, O3
- Conductivity- H2O2 , peracetic acid
4.2 Mechanical design
The DULCOMETER® D1C is supplied in versions suitable for control panel installation and wall mounting.
4.2.1 Control panel installation in accordance with DIN 43700 (96 x 96 mm)
The DULCOMETER® D1C is a device suitable for control panel installation in accordance with DIN 43700with the format 96 x 96 mm, depth 140 mm. In this installation arrangement, the device is completelyintegrated in the control panel or installed in a housing. The device is electrically connected directly viaterminals at the rear.
The terminals as well as an SN6 socket for pH or redox input project beyond the rear panel.
Retaining brackets for mounting the device in the control panel are provided on the housing.
The display pc-board with the graphic display is located at the front. It features 6 operating buttons and atransparent display window.
4.2.2 Wall mounting
The DULCOMETER® D1C W is suitable both for wall mounting as well as for installation in a control panel(control panel installation kit, Part No. 792908). The durable plastic housing is made up of an upper sectionand lower section. The graphic display and transparent display window are accommodated in the uppersection while the lower section houses the processor, power supply and options board. The connection tothe display is made by means of a ribbon cable. Electrical connection is made via the originally closed,punch-out cable leadthroughs on the underside of the housing. On devices with an SN6 input (dependenton identity code), the standard SN6 socket is located on the left-hand side. A wall mounting bracket isprovided at the rear of the device to facilitate simple wall mounting.
4.3 Electrical design
The device processes an input signal while taking into consideration disturbance variables and operatorinputs. The result is displayed and made available to other devices via a standard signal or a serial interface.
When equipped with corresponding actuators, the device can undertake control functions. It is designed toactivate metering pumps, solenoid valves, servo motors with feedback as well as mA standard signal. Theactivation variable is recalculated every second.
The controller does not feature a separate power switch. It is therefore immediately ready for operationafter being connected to the power supply.
The devices correspond to relevant requirements concerning electrical operating equipment. For thispurpose, the following standards are complied with:
• Supply voltage in accordance with - DIN IEC 38• Electrical safety in accordance with - EN 61010-1• Emitted electromagnetic interference in accordance with - EN 55011 Gr. 1/Cl. A
4 Functional Description
27
5.1 Safety information
WARNING
The device is suitable for installation in a control panel or in a corresponding housing (seeaccessories). The device must not be placed into operation if not installed as intended!
CAUTION
• The generally applicable safety precautions must be observed for installation.Corresponding national regulations must be complied with!
• The operating instructions must be read through carefully before starting any installationand start-up procedures!
• Only specially trained and qualified personnel are permitted to carry out electricalinstallation of the device!
• The power ratings specified on the device must agree with those of the supply voltage!
• The power connection line and the data lines must not be installed together withinterference-prone lines! If low electrical disturbance cannot be guaranteed in theworking environment, special interference suppression measures must be implemented!Severe disturbances can cause malfunctions through to irreparable damage of thedevice!
5.2 Mounting description, mechanical
Please remove the protective film from the display!
5.2.1 Mounting DULCOMETER® D1C D (control panel installation)
This device is designed specifically for installation in a control panel. The housing corresponds to DIN43700.
The aperture in the control panel for installing the device is defined in DIN 43700.
We recommend a smaller aperture. In this way, the device is held more securely in place (reduced lateralplay) and the seal is pressed more evenly.
5 Mounting / Installation
92 +0,6
92 +
0,6
90 +0,5
90 +
0,5
DED1C1S003
Aperture to ProMinentrecommendation
Aperture toDIN 43700
28
To make aperture:
As an installation aid, a drill/punch template at a scale of 1:1 is provided with the device for the purpose ofoptimally positioning the device on the control panel.
With the aid of a spirit level, align the template in the corresponding position on the control panel andsecure in this position. Mark the corner points with a centre punch and drill out with a 6 mm Ø twist drill.Then saw out the intermediate web with a compass saw. Neatly rework the surfaces until the dimensionsare within the specified tolerances.
Cleanly deburr edges.
Before fitting the device into the control panelaperture, check the position of the seal (must reston front shoulder). Working from the outside, fit thedevice in the aperture, attach the retaining bracketsand slide back as far as it will go. All four retainingbrackets must be fitted otherwise the enclosureclass IP54 will not be complied with.
Using a suitable screwdriver, screw the threadedpins forward until the seal is evenly pressed allround.
Check once again that the seal is fitted correctly, ifnecessary, release the threaded pins and correctthe position.
5.2.2 Mounting DULCOMETER® D1C W (wall mounting)
NOTE
The device should be mounted in such a position as to facilitate easy read-off andoperation (at eye-level where possible).
The device can be screwed directly onto the wall with the aid of the wall mounting bracket provided (drillingtemplate enclosed).
Securing material for wall mounting:
• Item (1) 3x button head screws 5x45• Item (2) 3x U-washer 5.3• Item (3) 3x plastic wall plug d8
The wall mounting bracket (4) can also be used as a drilling aid. For this purpose, set up the mountingbracket in the corresponding position on the wall.
NOTE
Take particular care when setting up to ensure that sufficient space is available for thepurpose of installing the cables. A space of approx. 120 mm must be left at the top for the“park position”.
Mark and drill the holes. Insert wall plugs (3) and secure mounting bracket with screws (1) and U-washers(2). Mount device from above on mounting bracket, press slightly against the wall and slide upward byapprox. 4 mm until it is heard to engage in position.
DED1C1S004
®
STOPSTART
DULCOMETER®
D1C
Mounting / Installation
29
5.2.3 Control panel installation D1C W
A 4 mm wide flange acting as the stop for the control panel together with an all-round groove for a seal isprovided on the perimeter of the device. When mounted in a control panel, the entire front sectionprotrudes by approx. 35 mm from the control panel. The device is mounted from the outside in a preparedaperture in the control panel. The device can be secured to the control panel from the inside with thesecuring material provided.
Securing material for control panel installation:
• Item (1) 1x cellular rubber seal d3• Item (2) 6x retaining bracket, galvanised steel• Item (3) 6x PT self-tapping screw, galvanised
Procedure
Using the cut-out template provided, first mark the exact position of the device on the control panel. Wherepossible, it should be positioned at eye-level. Ensure sufficient space is left at the top to allow for the “parkposition”. Mark and drill the corners. Hole diameter 12 - 13 mm.
IMPORTANT
Dimensional deviations may occur as the result of photocopying the cut-out template.
Now make the aperture as specified in the drawing with the aid of a punch or compass saw. The controlpanel should be 2 - 3 mm thick. Before fitting the device in the aperture, evenly press the seal into thegroove on the outside of the housing. With the aid of a second person if necessary, the controller can thenbe firmly secured with the retaining brackets and screws to the control panel.
3 4 2 1 D1C1(W) 003 D
Ø 8 x 50 deep drilled hole
Mounting / Installation
30
Mounting with SN6 socket (depending on identity code)
Corresponding to the order, an SN6 input socket may bepreassembled on the device. This socket must first be removedin order to facilitate installation in the control panel. For thispurpose, open device as described under Point 5.3.
By pressing the orange-coloured levers simultaneously,disconnect connection cable from terminals No. 11 and 12.
Unscrew SN 6 socket (WAF 22) complete with O-ring.
After installing the controller in the control panel, the SN6socket can be re-fitted together with the coaxial cable andO-ring.
IMPORTANT
Ensure the O-ring is fitted correctly!
Reconnect cable to terminals 11 and 12.
5.3 Electrical installation
WARNING
• Electrical connection must not be carried out before the device has been installed in thecontrol panel!
• The supply voltage must be disconnected when removing the device together with theconnection lines!
• Generally, the device may only be opened by qualified personnel.
• Particular care must be taken before opening the device for the purpose of possibleservice work to ensure that no voltage is applied to the device and the voltage supplycannot be switched on while carrying out service work.
IMPORTANT
Tie together stranded wires with cable ties at a distance of 30 mm from the terminals! Ifthey work loose in the event of a defect, stranded wires carrying mains voltage must beprevented from coming in contact with low voltage terminals!
5.3.1 Electrical installation D1C D (control panel installation)
Terminals are provided at the rear of the device to facilitate electrical connection. The number and type ofterminals depend on the type of device.
Not all devices are equipped with terminal row X1 (left).
Among other things, terminal row X2 (centre) serves the purpose of connecting the measured variable. Themeasured variables pH and redox can be connected directly to the terminals with SN6 connectors or, asother measured variables, via a mA signal. The jumper must be connected at terminals 9 and 10 of terminalrow X2 if pH or redox probes are connected to SN6 connectors or to the terminal without connection ofequipotential bonding!
Terminals rows XR1, XR2 and XR3 (right) serve the purpose of connecting power relays. The mains powerconnection XP is located at the bottom right. An earthing cable is not necessary.
The connectors XHK are provided for service purposes.
Refer to the terminal connection diagrams for the connection terminals (see Page 42-53).
3 2 1D1C1(W) 004 D
Control panel
Mounting / Installation
31
a) b) c)
d)
e)
321
4
Cable ties
16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
123456789101112 2 1 2 1 N2 1 3 2 1 L1
pH
®
pH
DULCOMETER®
STOPSTART
1) 2) 5)
3) 4)
D1C1(W) 005 D
a) b)
c) d)
6
5
DED1C1W 006 DED1C1W 007
Fig. 1: Rear row Fig. 2: Front row
5.3.2 Electrical installation D1C W (wall mounting)
Opening the housing
The device should only be opened when it is mounted on a wall or installed in a control panel.
To open the housing, initially, the four captive countersunk screws must be released.
The upper section is additionally locked to the bottom section by means of snap hooks. The housing canbe opened by pulling the upper section forward thus releasing the snap hooks.
With the aid of the two guide rails, the upper section can be moved to the approx. 100 mm higher “parkposition”. The fuse and all connection terminals are freely accessible in the park position.
Electrical installation for wall mounting
Initially, threaded holes must be broken out corresponding to the number of cables.
• Punch aids are provided to break out the individual threads.
Mounting / Installation
32
• Use following tools to punch out holes:
Rear row (Fig. 1): Screwdriver DIN 5262-B, size 1 (Ø 4.5 mm)Front row (Fig. 2): Screwdriver DIN 5262-B, size 0 (Ø 3.0 mm)
Remove cable sheathing over a sufficient length (corresponding to length of terminals). Fit screwedgland (1), thrust ring (2) and seal (3) over cable and insert in threaded hole.
Screw in screwed gland and firmly tighten with WAF 19 spanner. Shorten stranded wires to the exactoverall length, then strip approx. 7 mm insulation and connect to terminals corresponding to the electricalconnection diagram.
Core sleeves may be used for the stranded wires. If too many threaded holes are punched out, they canbe closed off again with the dummy washers PG 11 (4) supplied with the device.
If the controller features an SN6 input (corresponding to identity code), the corresponding input socket islocated in the rear row on the left side in a PG 11 threaded hole. Any ProMinent cable combination coaxSN6 can be connected to this input.
Packing list, cable screwed glands
5x screwed gland PG 11 Item (1)5x thrust ring PG 11 Item (2)5x seal PG 11 inside Ø 9 mm Item (3)3x seal PG 11 inside Ø 7 mm Item (3)3x seal PG 11 inside Ø 5 mm Item (3)2x seal PG 11 inside Ø 4 mm Item (3)5x seal PG 11 2 x Ø 5 mm Item (3)2x seal PG 11 2 x Ø 4 mm Item (3)3x dummy washer PG 11 Item (4)
Additionally for expansion version
4 Stck. 4x screwed gland PG 7 Item (5)4x lock nut PG 7 brass, nickel-plated Item (6)
The four additional cut-outs in the front row can be used for PG 7 screwed glands. All commerciallyavailable PG 7 screwed glands (suitable for type of enclosure IP65) with lock nut (brass, galvanised, nickel-plated) can be used as cable screw fittings.
Available from ProMinent under:
1x cable screwed gland PG 7, black Part No. 7038961x lock nut PG 7 brass, nickel-plated Part No. 703819
Fit lock nut PG 7 (6) on the inside and mount PG 7 screwed gland (5) from outside and firmly tighten(WAF 15).
Electrical connection for control panel installation
Normally, only the rear row of threaded holes should be used on devices mounted in a control panel. Thefront row (PG 7 cut-outs) are located outside the control panel. The cable screw fittings supplied with thedevice are not required for control panel installation. In this case, the individual stranded wires (without pull-relief and seal) are routed directly through the holes and connected to terminals corresponding to theelectrical connection diagram. The holes are broken out as described above.
Mounting / Installation
33
Temperature specifications D1C D Control panel Wall mountingPermissible ambient temperature installation
Basic version: 0 °C...50 °C -5 °C...45 °CExpansion version: with position feedback
or with correction value via mAor with disturbance variable via mA 0 °C...45 °C -5 °C...40 °C
D1C WPermissible ambient temperatureBasic version: -5 °C...50 °CExpansion version: with position feedback
or with correction value via mAor with d isturbance variable via mA -5 °C...40 °C
Permissible storage temperature: -10 °C...70 °C
Material specifications / chemical resistancePart MaterialHousing and frame D1C D PPO GF 10Housing D1C W PPE GF 10Rear panel D1C D PPE GF 20Membrane keypad Polyester film PETSeal Cellular rubber CRInner seal, D1C D Silicon-based sealing compoundRetaining bracket and screws Galvanised, zinc-plated steelM5 screws A2
Chemical resistance:The device is resistant to normal atmospheres in installation rooms.
Dimensions and weightsD1C D96 x 96 mm in accordance with DIN 43700, depth 140 mmDevice weight without packing: approx. 850 gGross weight of device with packing: approx. 1200 g
D1C W198 x 200 x 76 mm (W x H x D) Wall mounting198 x 200 x 35 mm (W x H x D) Control panel installation, external198 x 200 x 38 mm (W x H x D) Control panel installation, internal
Device weight without packing: approx. 1.2 kgNet weight of device with packing: approx. 2.0 kg
Electrical dataProbe input via terminals(X2.12 ... X2.9): Input resistance: > 5 x 1011 Ω
Input resistance of reference electrode with respect tochassis ground: <1 kΩInput range: ±1 VAccuracy: ±0.5 % of input rangeResolution: 0.0625 % of input rangeConnection option for one equipotential bonding electrode(alternatively, two connection terminals are to be connectedby a wire jumper).
6 Technical Data
34
Probe input viaSN6 socket (X2.12 ... X2.9): Input resistance: 1012 Ω
Other data same as for “Probe input via terminals”
Standard signal input(all measured variables)(X2.12 ... X2.9): Input range: 0/4…20 mA (programmable)
Input resistance: 50 ΩAccuracy: 0.5 % of input rangeResolution: 0.014/0.012 mASupply voltage and current for external electronics:
19 V ±1.5 V, 20 mAConductivity input viaterminals (X2.12 ... X2.9): 2 electrode measuring cells via 2-conductors
2 electrode measuring cells via 4-conductors4 electrode measuring cells via 4-conductorsMeasuring range: 20, 200, 2000 µS/cm
20, 200 µS/cmMeasuring span: 1 : 100Cell constant: 0.006…12.0 cm-1 depending on measuring rangeProbe activation: Sinusoidal 56 Hz or 2.7 kHz
depending on measuring range
Pt 100 input (X2.8, X2.7): Input range: 0 °C…100 °CAccuracy: ±0.5 °CResolution: 0.1 °C
Frequency outputs(2 reed relays)for pump activation(X2.6, X2.5 and X2.4, X2.3): Type of contact: n/o contact noise-suppressed with Varistors
Load capacity: 25 V peak, 0.100 A switching currentContact lifespan: >50 x 106 switching operations at
contact load 10 V, 10 mAMax. frequency: 8.33 Hz (500 strokes/min)Closing time: 100 ms
Standard signal output mA(X2.2, X2.1): Galvanically isolated from remaining inputs and outputs
Insulation voltage: 500 VOutput range: 0/4...20 mA (programmable)Max. load: 600 Ω current output 1
400 Ω current output 2Accuracy: 0.5 % of output range referred to
displayed value
Technical Data
35
Standard signal inputfor correction variable ordisturbance variable mA(X1.16 ... X1.14): Galvanically isolated from other inputs and outputs
Insulation voltage: 500 VInput range: 0/4…20 mA (programmable)Input resistance: 50 ΩAccuracy: 0.5 % of input rangeResolution: 0.014/0.012 mASupply voltage and current for external electronics:
22 V ±1.5 V, 20 mA
Standard signal output mA(X1.13, X1.12): Galvanically isolated from other inputs and outputs
Insulation voltage: 500 VOutput range: 0/4…20 mA (programmable)Max. load: 600 Ω current output 1
400 Ω current output 2Accuracy: 0.5 % of output range referred to displayed value
Digital input (X1.10, X1.9 Common reference potential and with RS interface but galvanicallyand X1.7, X1.6): isolated from remaining inputs and outputs
Insulation voltage: 500 V- Pause- Disturbance variable up to 10 Hz or up to 500 Hz(according to identity code/programmable)
Position feedback input(X1.3 ... X1.1): Galvanically isolated from other inputs and outputs
Insulation voltage: 500 VPotentiometer to be connected: 900 Ω...10 kΩAccuracy (not including potentiometer error):
1 % of input rangeResolution: 0.5 % of input rangeActuating time: min.: 25 s
max.: 180 s
Power relay output forcontrolled variable output (M, R)or limit value signalling (G)(XR1 and XR2): Type of contact: Changeover contact, noise-suppressed
with varistorsLoad capacity: 250 V AC, 3 A, 700 VAContact lifespan: >20 x 106 switching operations
IMPORTANT
The supply voltage applied at relays XR1-XR3 must be identical to the supply voltageof XP.
Technical Data
36
Power relay outputfor alarm triggering (XR3): Type of contact: Changeover contact, noise-suppressed
with varistorsLoad capacity: 250 V AC, 3 A, 700 VAContact lifespan: >20 x 106 switching operations
IMPORTANT
The supply voltage applied at relays XR1-XR3 must be identical to the supply voltageof XP.
Rated voltage (XP): 100/200 V AC, 50/60 Hz (D1C D only)Maximum current intake: 150 mA at 100 V AC
75 mA at 200 V ACInternal fuse protection with: 5 x 20 mm miniature fuse
160 mA, 250 V slow-blowRated voltage (XP): 115/230 V AC, 50/60 HzMaximum current intake: 140 mA at 115 V AC
70 mA at 230 V ACInternal fuse protection with: 5 x 20 mm miniature fuse
160 mA, 250 V slow-blowRated voltage (XP): 24 V DC or 24 V AC, 50/60 Hz
(operation with safety extra-low voltage only!)Internal fuse protection with: 5 x 20 mm miniature fuse
315 mA, 250 V slow-blow
Technical Data
37
Input /OutputterminalsPower supply
pc-board on rail 1
Fuse
Rear cover
Terminal block withflange screw fitting
Pin strip
SN6-connector
Safety information
WARNING
• The device or system must be disconnected from the power supply before starting anymaintenance work. The DULCOMETER® D1C does not feature a separate power switch!The power supply must therefore be interrupted by means of an external master switchor by a main fuse. General safety regulations are applicable at all times!
Even when the voltage supply is switched off, mains power may still be applied atterminals XR 1-3!
• Only use fuses of the specified manufacturer!Only use 5 x 20 mm miniature fuses!Fuse rating at mains voltage 100…240 V: 0.160 A slow-blow, Part No. 712048
24 V: 0.315 A slow-blow, Part No. 712026
Fuse change D1C D:
The device can be opened from the rear. For this purpose:
Release flange screw connections of plug-in terminal strips
Unplug terminal strips together with electrical connections
Release SN6 connector (if fitted) and unplug from socket
Unscrew rear panel and remove; the captive screws are located on the rear cover
The power board is located on the right-hand side (viewed from rear); pull out towards rear
If necessary, remove fuse and replace by new fuse
Take particular care when re-inserting the card to ensure that it is fitted in the same rail otherwiseelectrical connection to the display board will not be established
Press card fully forward onto plug contacts
Reinstall rear housing cover and firmly tighten
Reconnect terminal strips with electrical lines to corresponding pin strips and firmly tighten.
Maintenance / Repair
DED1C1S005
38
Fuse change D1C W:
• The above-specified safety measures must be implemented (disconnection from mains!) before replacingthe device fuse:
The mains power fuse is located in a closed fuse holder (6) in the terminal box.
Open device and set in “park position”
Release bayonet catches of fuse holder
Remove fuse and replace by new fuse
Lock bayonet catch and close housing
Maintenance / Repair
6534
2
1
......
D1C-..
FW-_._
DED1C1W010
Upper section of housinginside
Lower section of housinginside
Sectionalside view
Item (1) Electrical assembly, display Item (4) Ribbon cableItem (2) Ribbon cable Item (5) Electrical assembly, processorItem (3) Electrical assembly I/O/S Item (6) Fuse holder
39
8.1 Electric shock and moisture protection (IP)
D1C D: Device in installed condition: Type of enclosure DIN 40050 - IP 54
D1C W: Device in sealed housing type of enclosure IP 65 in accordance with DIN VDE 0470 correspondingto EN 60529 and IEC 529outer seal (control panel installation): type of enclosure IP 54 in accordance with DIN VDE 0470corresponding to EN 60529 and IEC 529.
8.2 Electrical safety/interference suppression
EC low voltage guideline (73/23/EEC) consequently 93/44/EECEC EMC guideline (89/336/EEC) consequently 92/31/EEC being preparedSupply voltage in accordance with DIN IEC 38Electrical safety in accordance with EN 61010-1Electromagnetic emitted interference in accordance with EN 55011 Gr. 1/Cl. BInterference immunity in accordance with IEC 801-2, -3, -4 or DIN VDE 0843, Part 2, Part 3, Part 4 or EN50082-2
EN 60335-1 Safety of electrical devices for household useEN 50081-1 EMC, emitted interference, residential areasEN 50082-2 EMC, emitted interference, industrial areasEN 60555-2 EMC, reactions in power supply networks, harmonicsEN 60555-3 EMC, reactions in power supply networks, voltage fluctuations
8.3 Test in moist changeable climate
D1C D: Permissible relative humidity: max. 80 % (condensation not permitted)
WARNING
The device is not suitable for use in a moist changeable climate as there is no sealedhousing and constant exchange of air!
D1C W: Moist changeable climate in accordance with FW DIN 50016.
8 Applicable Types of Enclosure / Standards
40
9 Spare Parts / Accessories
- Housing for wall mounting, including securing material, Order No. 790235
- Mounting kit for control unit installation, Order No. 792908
- Sensors, instrument leads, see product catalogue, chap. 6
- Measuring transducers, see product catalogue, chap. 5
10 Used Part Disposal
NOTE
Plastics and scrapped electronic components are special waste and must be recycled!
Used electronic components are accepted by municipal collection points set up by townsand municipal districts or ProMinent branches!
With the exception of the electrical assemblies, the design of the device comprised fewmechanical parts. They are relatively easy to separate into specific materials; e.g. nometallic thread inserts were used in the housing. With the exception of the membranekeypad, these parts can be re-introduced to the material recycling system (see underChap. 6 “Material Specifications”)!
The membrane keypad is to be classified and disposed of in compliance with applicablemunicipal guidelines!
9 Spare Parts / Accessories10 Used Part Disposal
41
EC Declaration of Conformity
42
Übersicht Klemmenanordnung / Overview of terminal arrangement
KlemmenanordnungSchalttafelgerätTerminal orderswitchboard mounting
X1 X2XR1
XP
XR2
XR3
X2
43
KlemmenanordnungWandgerätTerminal orderwall mounting
X1
X2
XR1 XR3XR2 XP
ACHTUNG
Nicht die Klemmenbezeichnungen von X2 und X1verwechseln!
IMPORTANT
Do not confuse the terminal designations X2 and X1!
44
pH/Redox über Klemme / pH/ORP via terminals
Pt 1
00 (
tem
pera
ture
)***
Mains power Mains powerNetz
Ext
ern
Pum
pe 2
sen
ken
Ext
ern
Pum
pe 1
heb
en
Normsignal-Ausgang 10/4-20mA
Standard signal output 10/4-20mA
Netz
externalExtern
(pot
entia
lfrei
)
(pot
entia
lfrei
)
with without
Liqu
id r
efer
ence
pot
entia
l
mV-input terminal
Terminal connection for pH / ORP via terminals
mV
Ref
.
Pt 1
00 (
Tem
pera
tur)
***
8 7 6 5 4 3 2 1
Ref
.
mV
Pot
entia
laus
glei
ch
pH/R
edox
(O
RP
)
pH/R
edox
(O
RP
)
mV-Eingang-Klemme
ohne/mit/
PotentialausgleichLiquid reference potential
Klemmenanschluss für pH / Redox über Klemme
X2
12111091211109+-
pH/Redox (ORP)
**
Temperature input (Pt 100)
Temperatur-Eingang (Pt 100)
Frequency outputs (pumps)
Frequenz-Ausgänge(Pumpen)
i. e.
rec
orde
r
z. B
. Sch
reib
er
internalIntern
Ext
erna
l pum
p 2
decr
ease
(po
tenc
ial f
ree)
Ext
erna
l pum
p 1
incr
ease
(po
tenc
ial f
ree)
*** Correction variable for pH (see also p. 53)
*** Korrekturgröße für pH (vgl. auch S. 53)
** IMPORTANT: When connecting the electrode without liquid reference potential
clamp terminal X2.9 and X2.10 have to be closed.
** ACHTUNG: Bei Anschluss der Sonde ohne Potenzialausgleichdie Klemmen X2.9 u. X2.10 brücken.
45
pH/Redox über SN6-Eingang / pH/ORP via SN6 socket
Pt 1
00 (
tem
pera
ture
)***
Mains power Mains powerNetz
Ext
ern
Pum
pe 2
sen
ken
Ext
ern
Pum
pe 1
heb
en
Normsignal-Ausgang 10/4-20mA
Standard signal output 10/4-20mA
Netz
externalExtern
(pot
entia
lfrei
)
(pot
entia
lfrei
)
with without
SN6-input
Pt 1
00 (
Tem
pera
tur)
***
8 7 6 5 4 3 2 1
pH/R
edox
(O
RP
)
SN6-Eingang
ohne/mit/
PotentialausgleichLiquid reference potential
Klemmenanschluss für pH / Redox über SN6-EingangTerminal connection for pH / ORP via SN6 socket
X2
12111091211109+-
pH/Redox (ORP)
**
Temperature input (Pt 100)
Temperatur-Eingang (Pt 100)
Frequency outputs (pumps)
Frequenz-Ausgänge(Pumpen)
i. e.
rec
orde
r
z. B
. Sch
reib
er
internalIntern
Liqu
id r
efer
ence
pot
entia
l
SN6
Pot
entia
laus
glei
ch
SN6
pH/R
edox
(O
RP
)
* *
clamp terminals X2.12 and X2.11 are connected internaly.
* Bei Version D (Schalttafeleinbau) sind die Klemmen X2.12 und X2.11 intern verdrahtet.
* Version D (Control panel installation) only:
Ext
erna
l pum
p 2
decr
ease
(po
tenc
ial f
ree)
Ext
erna
l pum
p 1
incr
ease
(po
tenc
ial f
ree)
*** Correction variable for pH (see also p. 53)
*** Korrekturgröße für pH (vgl. auch S. 53)** ACHTUNG: Bei Anschluss der Sonde ohne Potenzialausgleichdie Klemmen X2.9 und X2.10 brücken.
** IMPORTANT: When connecting the electrode without liquid reference potentialclamp terminal X2.9 and X2.10 have to be closed.
46
F, Br2, Cl2, ClO2, O2, O3, mA, mS/cm, pH*, Redox / ORP*, ºC*
Pt 1
00 (
tem
pera
ture
)***
Mains power Mains power
** See also D1C, part 2 "Measured value setting"-
"range adjustment" and operating instructions DMT!
Netz
Ext
ern
Pum
pe 2
sen
ken
Ext
ern
Pum
pe 1
heb
en
Normsignal-Ausgang 10/4-20 mA
Standard signal output 10/4-20 mA
Netz
external
** Siehe auch D1C, Teil 2 "Messbereich einstellen""Bereichzuordnung" und Betriebsanleitung DMT!
Extern
(pot
entia
lfrei
)
(pot
entia
lfrei
)
Assignment variants
Measured variableStandard signal input (mA)
Pt 1
00 (
Tem
pera
tur)
***
8 7 6 5 4 3 2 1
Normsignal-Eingang (mA)
Belegungsvarianten
Klemmenanschluss fürTerminal connection for
F, Br2, Cl2, ClO2, O2, O3, mA, mS/cm, pH*, Redox*, ˚C*
X2
12111091211109+-
Messgröße
Temperature input (Pt 100)
Temperatur-Eingang (Pt 100)
Frequency outputs (pumps)
Frequenz-Ausgänge(Pumpen)
***Correction variable for pH, ClO2, mS/cm (see also p. 53)
*** Korrekturgröße für pH, ClO2, mS/cm (vgl. auch S. 53)
i. e.
rec
orde
r
z. B
. Sch
reib
er
internalIntern
Ext
erna
l cur
rent
sou
rce
1211109
Pro
Min
ent t
rans
duce
r
2 1
Pro
Min
ent U
mfo
rmer
+- +- +-
+-+- +-
Pro
Min
ent D
MT
8 7**
pass
ives
Nor
msi
gnal
4 -
20
mA
aktiv
es N
orm
sign
al 0
/4 -
20
mA
Ext
erne
Str
omqu
elle
activ
e st
anda
rd s
igna
l 0/4
- 2
0 m
A
pass
ive
stan
dard
sig
nal 4
- 2
0 m
A
Ext
erna
l pum
p 2
decr
ease
(po
tenc
ial f
ree)
Ext
erna
l pum
p 1
incr
ease
(po
tenc
ial f
ree)
X1
* Nur mit Umformer* Only with transducer
47
Leitfähigkeit / conductivity
Normsignal-Ausgang 10/4-20mA
Standard signal output 10/4-20mA
Ext
erna
l pum
p 2
decr
ease
(po
tenc
ial f
ree)
Ext
erna
l pum
p 1
incr
ease
(po
tenc
ial f
ree)
Pt 1
00 (
tem
pera
ture
)***
Mains power Mains powerNetz
Ext
ern
Pum
pe 2
sen
ken
Ext
ern
Pum
pe 1
heb
en
Netz
externalExtern
(pot
entia
lfrei
)
(pot
entia
lfrei
)
Conductivity input
Terminal connection for conductivity
Pt 1
00 (
Tem
pera
tur)
***
8 7 6 5 4 3 2 1
Leitfähigkeits-Eingang
Klemmenanschluss für Leitfähigkeit
X2
1211109+-
Temperature input
Temperatur-Eingang
Frequency outputs (pumps)
Frequenzausgänge(Pumpen)
***Correction variable for mS/cm (see also p. 53)
*** Korrekturgröße für mS/cm (vgl. auch S. 53)
i. e.
rec
orde
r
z. B
. Sch
reib
er
internalIntern
2-E
lekt
rode
n-M
essz
elle
2-E
lekt
rode
n-M
essz
elle
,4-
Leite
r-A
nsch
luss
4-E
lekt
rode
n-M
essz
elle
curr
ent b
curr
ent a
volta
ge b
volta
ge a
Str
om b
/S
trom
a /
Spa
nnun
g b
/
Spa
nnun
g a
/
four
wire
con
nect
ion
two
elec
trod
esm
easu
ring
cell,
mea
surin
g ce
lltw
o el
ectr
odes
mea
surin
g ce
llfo
ur e
lect
rode
s
48
mit Grenzwertrelais / with limit value relays
Klemmenanschluss mit Grenzwertrelais
Leistungsrelais-Ausgänge
Power relay outputfor alarm triggering
Leistungsrelais-Ausgang zurAlarmgabe Internal
power supply
Interne Spannungs-versorgung
XR2XR1 XPXR3
Terminal connection with limit value relay
Power relayoutput
1212 L1N123
externalExtern
internalIntern
Hor
nH
upe
limit
rela
y 1
limit
rela
y 2
Gre
nzw
ertr
elai
s 1
Gre
nzw
ertr
elai
s 2
Speise-spannung*
Speise-spannung**
Speise-spannung**
Speise-spannung**
HINWEIS: zu Relais XR3: Das Gerät ist in betriebs-bereitem Zustand und es steht kein Alarm an.
NOTE: to relay XR3: The device is shown ready and no alarm is present.
feedvoltage**
feedvoltage**
feedvoltage*
feedvoltage**
is not relevant !* With 24V DC polarity on feed voltage
Speisespannung ohne Bedeutung !Bei 24 V DC ist die Polarität der*
ACHTUNG: Die Speisespannung, die andie Relais XR1 - XR3 gelegt wird, mussidentisch mit der Speisespannung von XP sein!
** IMPORTANT: The feed voltage to be connected tothe relays XR1 - XR3, has to be identicallywith the feed voltage of XP.
**
49
mit Magnetventilen / with solenoid valves
Klemmenanschluss mit Magnetventilen
Interne Spannungs-versorgung
Terminal connection with solenoid valves
Leistungsrelais-Ausgänge
Leistungsrelais-Ausgang zurAlarmgabePower relay
output Power relay outputfor alarm triggering
Internalpower supply
XR2XR1 XPXR3
1212 L1N123
externalExtern
internalIntern
Hor
nH
upe
sole
noid
val
ve 1
sole
noid
val
ve 2
Mag
netv
entil
1
Mag
netv
entil
2
Speise-spannung*
Speise-spannung**
Speise-spannung**
feedvoltage**
feedvoltage**
feedvoltage*
Speise-spannung**
feedvoltage**
ACHTUNG: Die Speisespannung, die an die RelaisXR1 - XR3 gelegt wird, muss identischmit der Speisespannung von XP sein!
is not relevant !* With 24V DC polarity on feed voltage
Speisespannung ohne Bedeutung !Bei 24 V DC ist die Polarität der*
**
NOTE: to relay XR3: The device is shown ready and noalarm is present.
HINWEIS: zu Relais XR3: Das Gerät ist in betriebs-bereitem Zustand und es steht kein Alarm an.
** IMPORTANT: The feed voltage to be connected tothe relays XR1 - XR3, has to be identicallywith the feed voltage of XP.
50
mit Stellmotor / with servomotor
Klemmenanschluss mit Stellmotor
Leistungsrelais-AusgängePower relay
output
Leistungsrelais-Ausgang zurAlarmgabe
Power relay outputfor alarm triggering
Interne Spannungs-versorgung
Internalpower supply
Terminal connection with servomotor
Stellungsrückmelde-Eingang
Position feedbackinput
XR2XR1 XPXR3X1
1212 L1N1233 2 1
- +S
externalExtern
internalIntern
S...
Sch
leife
r / S
...sl
ider
gesc
hlos
sen
/ clo
sed
offe
n / o
pen
Öffn
er /
to o
pen
Sch
ließ
er /
to c
lose
S
Act
uato
r m
otor
Ste
llmot
or
Hor
nH
upe
Speise-spannung*
Speise-spannung**
feedvoltage**
feedvoltage*
Speise-spannung**
feedvoltage**
M
is not relevant !* With 24V DC polarity on feed voltage
Speisespannung ohne Bedeutung !Bei 24 V DC ist die Polarität der*
NOTE: to relay XR3: The device is shown ready and noalarm is present.
HINWEIS: zu Relais XR3: Das Gerät ist in betriebs-bereitem Zustand und es steht kein Alarm an.
ACHTUNG: Die Speisespannung, die an die RelaisXR1 - XR3 gelegt wird, muss identischmit der Speisespannung von XP sein!
**
** IMPORTANT: The feed voltage to be connected tothe relays XR1 - XR3, has to be identicallywith the feed voltage of XP.
51
mit Normsignal-Ausgang 2, mit Pause / with standard signaloutput 2, with pause
Standard signaloutput 2
0/4-20mA
Digital-Eingang
Anschlussdetails Normsignalausgang 2 / Digitaleingang PauseConnection details Standard signal output 2 / Digital input pause
X1
Intern
Normsignal-Ausgang 2
Digital input
-
13
+
12 10 9 8
externalExtern
internal
- +
PotenzialfreierKontakt nötig!
potential free connectionnecessary.
i. e.
rec
orde
rz.
B. S
chre
iber
Pau
se- +
52
mit Störgröße / with disturbance variable
Anschlussdetails Störgröße
Digital-Eingang
Digital input
Normsignal-Eingangfür Korrekturgröße oder Störgröße (mA)
Connection details disturbance variable
X1
externalExtern
internalIntern
Standard signal inputcorrection variable or disturbance variable (mA)
Assignment variantsBelegungsvarianten
Disturbance variable (mA) i. e. flow signal from:
7 6 5
- +
16 15 14 16 15 14
Ext
erna
l cur
rent
sou
rce
Mea
surin
g tr
ansd
ucer
2 1
Mes
sum
form
er
+- +-
+-+-
pass
ives
Nor
msi
gnal
4 -
20
mA
aktiv
es N
orm
sign
al 0
/4 -
20
mA
Ext
erne
Str
omqu
elle
activ
e st
anda
rd s
igna
l 0/4
- 2
0 m
A
pass
ive
stan
dard
sig
nal 4
- 2
0 m
A
- +
dist
urba
nce
varia
ble
freq
uenc
y
Stö
rgrö
ße
Fre
quen
z
X1
Störgröße (mA) z.B. Durchfluss-Signal von:
potential free connectionnecessary.
PotenzialfreierKontakt nötig!
Electromagneticflow meters
Magnetisch-induktivemWassermesser Contact water meter*
Kontaktwassermesser *
* IMPORTANT: Note maximum frequency.
* ACHTUNG: Maximalfrequenz beachten!
53
mit Korrekturgröße / with correction variable
Anschlussdetails KorrekturgrößeConnection details correction variable
Normsignal-Eingangfür Korrekturgröße oder Störgröße (mA)
Temperature input (Pt 100)
Temperatur-Eingang (Pt 100)
X1
externalExtern
internalIntern
Standard signal inputcorrection variable or disturbance variable (mA)
Assignment variantsBelegungsvarianten
8 7
X2
161514 161514 161514
+-+- +-
Pro
Min
ent t
rans
duce
r
2 1
+- +- +-
Pro
Min
ent D
MT
aktiv
es N
orm
sign
al 0
/4 -
20
mA
activ
e st
anda
rd s
igna
l 0/4
- 2
0 m
A
pass
ive
stan
dard
sig
nal 4
- 2
0 m
A
X1
Pro
Min
ent U
mfo
rmer
8 7**
pass
ives
Nor
msi
gnal
4 -
20
mA
Ext
erne
Str
omqu
elle
Ext
erna
l cur
rent
sou
rce
Pt 1
00 (
tem
pera
ture
)
Pt 1
00 (
Tem
pera
tur)
** Siehe auch D1C, Teil 2 "Messbereich einstellen""Bereichzuordnung" und Betriebsanleitung DULCOMETER® DMT!
** See also D1C, part 2 "Measured value setting"-
"range adjustment" and operating instructions DULCOMETER® DMT!
54
108