76
安科 Permanent-magnet MRI System (OPENMARK 5000) User’s Guide Vol.Equipment Manual Version: 1.0 Shenzhen Anke High-tech Co., Ltd. Add: 26 Yanshan Road, Shekou, Shenzhen, China Tel: (86) (755) 2668 8889 Fax: (86) (755) 2669 5307 Website: ww))w.anketech.com Post Code: 518067

Openmark 5000 User Manual

Embed Size (px)

DESCRIPTION

manual de Usuario de Ecografo

Citation preview

Page 1: Openmark 5000 User Manual

安科 Permanent-magnet MRI System (OPENMARK 5000) User’s Guide Vol.Ⅰ

Equipment Manual Version: 1.0

Shenzhen Anke High-tech Co., Ltd. Add: 26 Yanshan Road, Shekou, Shenzhen, China Tel: (86) (755) 2668 8889 Fax: (86) (755) 2669 5307 Website: ww))w.anketech.com Post Code: 518067

Page 2: Openmark 5000 User Manual

Equipment Manual

2

Contents

Foreword 3

Chapter 1 Safety 6

1.1 Warning Signs and Forbidden Signs 6

1.2 External Signs 7

1.3 Safety Prompts Used in This Manual 9

1.4 Safety Precautions 10

1.5 Alarm System 24

1.6 Manufactuerer’s Responsibility 24

Chapter 2 Structure and Components of the Equipment 25

2.1 Structure of the Equipment 25

2.2 Components of the Equipment 28

2.3 Brief Introduction of Operating Principle 38

2.4 Transporting and Storing of the Components 39

2.5 Key Performance Index 39

Chapter 3 Routine Operation Guide 41

3.1 Routine Scan Procedure Reference 41

3.2 Startup and Shutdown 41

3.3 Console Operation 43

3.4 Patient Table Operation 44

3.5 Usage of Coils 50

Chapter 4 Equipment Maintenance Guide 56

4.1 Monitoring Equipment Status 56

4.2 Maintaining Equipment 57

Chapter 5 Troubleshooting 61

Chapter 6 Guidance and Manufacturer’s Declaration– IEC 60601-1-2 65

Appendix A: Guide for Auxiliary Power Supply 69

Appendix B: Guide for Ankle Joint Supporter 72

Page 3: Openmark 5000 User Manual

Equipment Manual

3

Foreword

Thank you for using OPENMARK 5000 double column Type Permanent-magnet MRI System made by

Anke. This system can image whole human body with various views, obtaining anatomic, physiological

and biological information in diagnostic medicine. It is mainly used to examine and diagnose for pathologic

changes of tumors, blood vessels, infections, surgical trauma, congenital malformation, as well as

retrograde affection.

For operating this system safely and reliably, please read the manuals carefully and thoroughly before

operation.

Manufacturer:

Full Name of Manufacturer: Shenzhen Anke High-Tech Co., Ltd.

Abbreviation of Manufacturer: Anke

Address of Manufacturer: 26 Yanshan Road, Shekou, Shenzhen, Guangdong 518067, P.R.China

Post Code: 518067

Tel: (86) (755) 2668 8889

Fax: (86) (755) 2669 5307

Website: www.anketech.com

Page 4: Openmark 5000 User Manual

Equipment Manual

4

Recording Number of Name, Type, Product Registration Number and Product Standard of this System

Name: MRI System

Type: ASM-050P (OPENMARK 5000)

Product Registration Number:

Product Standard Number:

The meaning of CE mark (only applicable to EC market)

“CE” mark originates from the abbreviation of European Community (Commeunante Euorpeenne in

French). It indicates that the device meets the essential requirements of the medical device directives of

European Community and is allowed to be legally launched to the market of European Community. In

addition, it also indicates that the device has passed the conformity audit in accordance with the procedures

of the European Community Medical Device Directives. The four digits after CE mark are the code of the

Notified Body.

CE mark certification channel selected by Anke (only applicable to EC market)

In accordance with MDD 93/42/EEC, Annex II.3 for medical devices of Europe Community, the

certification is done through overall quality assurance system (ISO 13485+ISO 9001) and audit and

supervision by the Notified Body.

Product Classification (only applicable to EC market)

In accordance with MDD 93/42/EEC: Annex XI Clause 10 for medical devices of Europe Community, the

diagnostic device intended to provide energy for human body to absorb but excluding those visible light

radiating human body is classified to be Class IIa device.

Anke’s representative in Europe Community (only applicable to EC market)

Anke’s representative in Europe Community: B-K Medical A/S

Anke’s representative in Europe Community in short: B-K Medical

Addres: World Headquarters, Mileparken 34, DK 2730 Herlev, Danmark

Contact phone: (45) 44528100 Fax: (45) 44528199

Page 5: Openmark 5000 User Manual

Equipment Manual

5

Duties and Responsibilities of Anke’s representative in the Europe Community (only applicable to EC market)

B-K Medical as Anke authorized representative in EC market will perform the following duties and

responsibilities on behalf of Anke:

B-K Medical shall keep the latest technical files and product changes information for the CE marked

products (such files shall be kept for at least 5 years after the last batch of products). If you have any

questions, please contact B-K Medical and you will get satisfactory answers.

If there is any accident of the CE marked products you purchased from Anke, please notify B-K medical

immediately and B-K medical will contact Anke and assist to investigate the causes and report to the

authority in specified time period.

Please notify B-K Medical of all information (including complaints) related to CE marked products of Anke

sold on the EC market. B-K Medical will forward your information to Anke and give you satisfactory

answers.

Product Safety Classification: (According to IEC60601-1)

Class I, Type B, Ordinary equipment

Disinfection: See section 1.4.1

It is forbidden to operate this system in the environment with inflammable anesthetic gas, oxygen and

nitrous oxide.

This system works continuously.

Important Declaration of Anke

The content in this guide is subject to change without notification.

The copyright is reserved by Anke. Any reproduction in any form without permission is forbidden.

User of the system please translates the warning signs and forbidden signs in local official language

and makes them into warning boards, and hangs them on the entrance of the scan room.

Interpretation right of this manual is reserved by Anke.

Page 6: Openmark 5000 User Manual

Equipment Manual

6

Chapter 1 Safety Precautions

This equipment has strong magnetic field, high voltage, heavy current and mechanical movement

components. Please follow the following safety precautions to safeguard personnel safety.

1.1 Warning Signs and Prohibition Signs

Warning: magnetic field

This equipment uses permanent magnet that keeps the same

magnetic field strength, regardless of the system being on or

not.

Warning: non-ionizing radiation

This equipment generates high frequency electromagnetic

field during operation.

Prohibition: no access for person with pacemaker

Person with pacemaker is forbidden to enter the scan room.

Otherwise, that will result in serious and fatal injury.

Prohibition: no access for person with metal implants

Person with metal implant such as metal implant or foreign

metal matter in eyeball is forbidden to enter the scan room.

Otherwise, that will result in serious and fatal injury.

Prohibition: no metallic pieces or watches

Forbidden: unfixed ferromagnetic articles

Unfixed ferromagnetic articles such as pens, keys, glasses,

ornaments, hair pins and coins are forbidden to take into scan

room.

Page 7: Openmark 5000 User Manual

Equipment Manual

7

Forbidden: ferromagnetic tools

Ferromagnetic tools are forbidden to take into scan room.

Forbidden: ferromagnetic carts

Ferromagnetic carts are forbidden to take into scan room.

Forbidden: ferromagnetic containers

Any ferromagnetic container is forbidden to take into scan

room.

Forbidden: magnetic sensitive articles

Magnetic sensitive articles such as credit card, disk, magnetic

record medium and watch, hearing aides, camera, and mobile

phone are forbidden to take into scan room. Data and property

may be damaged.

The above warning Signs and Forbidden signs should be hung on obvious places at the entrance of scan

room.

1.2 External Signs

Alternative current

N Connection point for the neutral conductor

Protective Earth (Ground)

Earth (Ground)

Attention, consult accompanying documents

Operating instructions

Comply with WEEE Directive

Off (Power: disconnection from the mains)

Page 8: Openmark 5000 User Manual

Equipment Manual

8

On (Power: connection to the mains)

“Off” (only for a part of equipment)

“On” (only for apart of equipment)

Equipotentiality

Dangerous Voltage

Serial Number

Manufacturer

Authorized Representative

Type B applied part

Notified body’s code

非专业人员勿动No User Repairable Parts

It is forbidden to operate the equipments

tagged this sign without permission by

manufacturer.

MR专用插座Socket For MR

Any other equipment is forbidden to use the

socket for MR without permission by

manufacturer.

小心拖动Move With Care

When pulling or pushing patient table,

operator should put hand on handle of

bedstead or knob of bedplate, don’t put your

hands between bedstead and bedplate or the

shell of magnet for fear of crash.

小心拖动机箱,防止机柜倾倒

Pull Drawers CarefullyCabinet May Lose Balance

Keep balance of gravity center of component

when moving equipment in test and repair.

Safeguard personal and equipment in safety.

Page 9: Openmark 5000 User Manual

Equipment Manual

9

警告1.避免激光直射病人眼部Prompt patient to close eyes and mustn'tstare at laser beam.2.移动病床时,小心病床夹手Patient table is prone to clip bodyparts of patient during moving.

When positioning, please prompt patient to

close eyes and mustn’t stare at laser beam.

Various gaps on the patient table is prone to

clip body parts, such as finger, hair, clothes,

etc.

Nameplate of Product Manufacturer: Shenzhen Anke High-Tech

Co., Ltd.

Model Number: ASM-050P

Trade Name: OPENMARK 5000

Trade Mark:

Power Supply: 380V 3N~

Power Frequency: 50Hz

Input Power: 15kVA

Classification:

1.3 Safety Prompts Used in This Manual

In this manual, the prompts of safety will be in following three methods.

Prompts that It may cause personal injury or casualty.

Caution!

Warning!

Page 10: Openmark 5000 User Manual

Equipment Manual

10

Prompts that it may cause equipment damage.

Prompts that it is very important, but may not cause injury or equipment damage.

1.4 Safety Precautions

1.4.1 Avoiding Cross Infection between Patients.

The coils of this equipment do not be allowed to touch patient’s skin. Every scan patient should use

throughway bed sheet and head cushion sheet.

1.4.2 Pre-screening of Patients and MR Worker

Patients are forbidden to make MR examine and not permitted to enter scan room if their bodies have

cardiac pacemaker, metal foreign body in eyeball, metal tubes and clamps leaved in body after operation

and ferromagnetic substance articles with larger mass. Traumatic patients should be ensured before

examine that there is no metal articles in their body. Patients in emergency treatment and restless patients

who can’t repose for 5 minutes can’t be allowed to make MR examine. The patients with little

ferromagnetic substance, the body part where metal ferromagnetic foreign matter lies in is not applicable

for MR examine. Pay special attention to women within 12 weeks’ gestation. Only trained personal and

physician are permitted to enter scan room.

Warning!

For fear of cross infection between patients, please

don’t touch patient’s skin with the coils of

equipment.

Attention!

Page 11: Openmark 5000 User Manual

Equipment Manual

11

1.4.3 Procedure of Monitoring Patient

User must establish patient monitoring procedure, which ensures that all patients can receive routine

wardship at least. This system always runs in normal operating mode, it is not necessary to consider of

running status and monitoring status under first level and second level controlled operating modes.

The recommended routine monitoring includes opening intercom subsystem when scanning, confirming

that in normal condition intercom subsystem is in the state of operator “listening”. Operator should closely

observe patient’s status through observation window. Medical care personnel or relative of patient should

leave in scan room to care children or serious patients.

1.4.4 Emergency Medical Procedures

User must regulate and carry out concrete emergency medical procedures according to concrete patient

conditions and the existence of magnetic field to ensure in-time medical treatment to patients if they are

attacked or injured during MR examine.

The emergency medical measures must include the requirement how operator pull the Patient Table out

from magnet cavity in emergency.

1.4.5 Excessive Acoustic Noise

According to regulation of clause 26g) in IEC60601-2-33 special requirement for medical MR equipment

(Hereinafter called IEC60601-2-33), the measured max A-weighted effective SPL (Sound Pressure Level)

of this equipment is less than 99 dB (A) in any normal running condition. That in the control desk is less

than 45dB (A).

In spite of the noise in magnetic cavity of this equipment being lower than the regulation in IEC60601-2-33,

the operator is still responsible to tell the patient that there will be noise in the scan process to make

necessary mind preparation. You can jam up patient’s ears with one-off medical absorbent cotton to reduce

disturbance of noise to patients.

Warning!

Pre-screening of patients must be performed

before MR examine. Patients who are not

applicable for MR examine will be injured in scan

room.

Page 12: Openmark 5000 User Manual

Equipment Manual

12

Operator is responsible to alarm patient who is sensitive to noise, such as baby, pregnant woman, elders.

Take special protection for these patients and pay attention to their acceptable sound pressure level.

For newborn baby or premature baby who can not use normal ear protection or any other protection,

operator should have training experience of handling protection reasonably.

Pay seriously attention to unconscious patient as anesthesia patient, ever though their protection is under

normal sound pressure level.

1.4.6 Controlled Access Area

User is responsible to obey the local legal requirement about entering controlled access area. It is necessary

to determine and establish a controlled access area around the MR equipment to ensure that the magnetic

field strength at any point out of this area is not more than 0.5 mT.

The controlled access area of this equipment is scan room (5mX5m), its center is at the center of magnetic

machine.

In the process of installation, the controlled access area is restrained in scan room. For the personal and

equipment safety, the warning and forbidden signs provided by our company should be hung up on the

obvious place of entrance of scan room (See section 1.1 for the content). User must obey the above

regulations and is responsible to prompt patients and other relative personnel to pay attention to.

Unapproved patients are not allowed to enter scan room. He who is permitted to enter should read warning

and prohibitive Signs outside the gate of scan room carefully and make sure that there is no metal implant

and active implant in body and taking no ferromagnetic substance articles before entering scan room.

Warning!

Persons without permission may ignore warning

signs and maybe injure in scan room.

Attention!

Jam up patient’s ears before scanning.

Page 13: Openmark 5000 User Manual

Equipment Manual

13

Only special non magnetic tools equipped with this machine can be used in routine inspection and

maintenance in scan room, such as 8" non magnetic adjustable wrench, 6" non magnetic “-“ screw driver, 8"

non magnetic “+” screws driver and 10" non magnetic “-“ screw driver. Other tools and equipment are not

allowed to use in scan room with permission by maintenance personnel of our Company.

Peripheral equipment such as monitoring device, life-holdout device and emergency nursing device are

forbidden to take into scan room, for these devices may be disturbed by radio frequency field or magnetic

field and may disturb the normal functions of this equipment.

1.4.7 Radio Frequency magnetic field

With the purpose of avoiding risk rising by partial RF over heat, operator must pay attention to:

Alarm patient take down all metal, metallic clothes or metallics from their body. Metal ferromagnetic

foreign matter, electric matter and implants are forbidden to exit in sensitive area of RF transmitting coil.

Use transdermal drug with caution which would bring subcutaneous burn.

It should be told to patient that skin touching may create partial body galvanic circle. Alarm patient avoid

touching from skin to skin, such as: touching between inside thigh and thigh, crus and crus, hand and hand,

hand and body, ankle and ankle.

Patient wearing moist clothes is forbidden to enter RF magnetic field.

Alarm patient not touch the RF transmitting coil surface and cables or wiring around of RF coils by body or

limbs.

Operator is just allow to use ECG poles and leads compatible with MRI system and within usage period.

Warning!

Unapproved tools and equipments maybe lose

control in scan room.

Warning!

Unapproved peripheral equipment may disturbed

by radio frequency field or magnetic field and

can’t work normally; or may disturb the normal

functions of this equipment.

Page 14: Openmark 5000 User Manual

Equipment Manual

14

Operator should keep a strict filtration of the patient who takes sedative, lose consciousness of all body or

partial, to avoid risk caused by that patient can not warn operator overheat or tissue damage.

Operator should be responsible to make sure there is no unconnected receiving coil or cable inside RF

transmitting coil while scanning.

1.4.8 Operating Modes

Any switching time of gradient under normal operating mode of this equipment is more than 500

microseconds, according to the regulations of clause 51.102 in IEC60601-2-33, the detected time rate of

change (dB/dt) of maximum magnetic field is always less than 20T/s.

In any 15 minutes under normal operating mode of this equipment, according to the requirements of clause

51.103 in IEC60601-2-33, the detected SAR in any part of human body is always less than 1.5W/kg.

In clauses of 51.102, 51.103 of IEC60601-2-33, the operating modes of MR equipment include three

modes: normal operating mode, first level controlled operating mode and second level controlled operating

mode according to the difference of time rate of change of magnetic filed and specific absorption rate

(SAR).

This equipment is always running in normal operating mode in any case.

1.4.9 Occupational Exposure

According to clause 6.8.2 kk) of international standard IEC60601-2-33, there is no evidence supports that

static magnetic field and time dependent field have cumulative effect on human body.

1.4.10 Auxiliary Equipment

Any application of auxiliary equipment to the patient, such as physiological monitoring, gating equipment

and RF transmit coils, should be used under the user guide. Any untested and unapproved auxiliary

equipments are not permitted to use. You should strictly follow the discretion by manufacture when using

marked compatible auxiliaries with MR equipment.

Warning!

The use of untested and unapproved auxiliaries of

MR equipment such as physiological monitoring

device, gate control device and RF coil may cause

burn or other injury to patient.

Page 15: Openmark 5000 User Manual

Equipment Manual

15

1.4.11 Fire Protections

User should be responsible for discussing fireproofing affairs and formulating emergency fire measures with

local fire department. Scan Room, Operating Room and Machine Room are forbidden smoking and should be

equipped with necessary fire extinguishing apparatus.

1.4.12 Artifacts

Steps and links in MRI are various and imaging methods are relative complex, so that there are so many kinds

of artifacts. Some artifacts are caused by intrinsic methods or the equipment itself, so that they are called

system artifacts; some are caused by patient’s autokinetic movement or normal physiological motions, they

are called motion artifacts; some are caused by ferromagnetic matters, they are called metal artifacts; some

are caused by hardware fault, they are called fault artifacts.

To reduce the occurrence probability of artifacts and their amplitude and to distinguish the influence of

artifacts is an important subject in MRI field. You should take serious consideration on it.

Some common artifacts and countermeasures are introduced in the following:

1.4.12.1 System Artifacts

Warning!

Nonobservance of the instruction made by

manufacturer when using compatible auxiliaries

with MR equipment, especially the instruction

about riding positions of conductive electrodes,

may injure patient.

Warning!

User should be responsible for discussing

fireproofing affairs and formulating fire

emergency measures with local fire department.

Smoking may cause fire.

Page 16: Openmark 5000 User Manual

Equipment Manual

16

1.4.12.1.1 Aliasing Artifacts

Aliasing artifacts mainly occur in the phase encoding direction because most sequences use over-sampling

anti-wrap technique in frequency encoding direction. When the FOV of scanned image is smaller than the

signal generating area, signals from outside of FOV in the phase encoding direction will appear in the

imaging area after Fourier transform, so that aliasing artifacts occur.

There are many methods to eliminate aliasing artifacts as follows:

Increase FOV. Increasing FOV can eliminate aliasing artifacts, but it declines the resolution of images at

the same time.

Select FOV factors more than 1. FOV factor of 1.5 or 2.0 can reduce the probability of occurrence of

aliasing artifacts because the actual FOV in phase encoding direction is become 1.5 or 2.0 times

compared with the original FOV, which equals to increasing FOV in phase encoding direction. Adding

FOV factor, the resolution of images will remain unchanged but scan time and SNR will increase. So

that you should jointly select NAV to keep scan time and SNR unchanged.

1.4.12.1.2 Truncation Artifacts

When the selected phase encoding number is less than 256, K space is not filled completely: only the central

part of K space is filled, while the data at the edge part of phase encoding direction in K space still remain

zero. In process of Fourier transform imaging, discontinuities between zero and filled data will lead to the

occurrence of artifacts. This kind of artifacts is called truncation artifacts.

Incompleteness of K space filling is not completely equivalent to the lost of high frequency of images, so that

truncation artifacts can be considered as the RESULT of losing high frequency of images.

Truncation artifacts mainly appear in the part where signals change fierily with the form of ringing artifacts

on the two sides of the special part of the image.

The larger phase encoding number is, the smaller truncation artifacts are; the smaller phase encoding number

is, the more serious truncation artifacts are. Taking the relation between phase number and scan time,

resolution and SNR into account, you should balance factors in all directions to reasonably select phase

encoding number.

1.4.12.1.3 Partial volume Effect

Every pixel of image corresponds to a certain actual voxel of body, but the fine structure of corresponded

actual voxel can’t be displayed clearly, which is called partial volume effect.

Because of this theory, the display effect for an image with thickness of 4 mm is better than that of image with

thickness of 10 mm in the same position.

You should reasonably select slice thickness according to actual requirement so as not only to get satisfying

SNR of images but also to reduce the influence of partial volume effect as much as possible.

Page 17: Openmark 5000 User Manual

Equipment Manual

17

1.4.12.1.4 Cross-talking Artifacts

In multi-slice and multi-angle scan, when the defined slices cross in imaging area, black strip will appear on

the images corresponding to the crossing area, which is called cross-talking artifacts.

You should place the cross area to the out side of patient’ body or uninteresting area in the location stage to

eliminate the influence of artifacts on diagnosis.

1.4.12.1.5 Inhomogeneity

The deviation of scanned part to magnet center or coil center can lead to the reduction of homogeneity of the

images.

You should make sure that the interesting part is located in the center of the coil and the coil is located in the

center of magnet as much as possible in the locating process to keep the homogeneity of the image.

Additionally, the absolute value of slice offset should not be set too large.

1.4.12.2 Motion Artifacts

1.4.12.2.1 Motion Artifacts

Any movement and motion can cause artifacts in data acquisition process of Fourier transform imaging. The

data acquisition is actually the process of filling K space. As we known, every pinot in K space contains the

information of the whole image, so that the motion in data acquisition process can influence the whole image

and cause motion artifacts.

Motion artifacts usually appear in the phase encoding direction. To eliminate motion artifacts, you should not

only perform pre-discriminating of patients to avoid scanning unqualified patients, but also prompt patients

not to move in scanning. Pre-saturation is an effective method to reduce or eliminate motion artifacts, i.e. imposing selective

pre-saturation RF pulse to saturate motion tissues, and then exciting the imaging area and acquiring data.

Thus the saturated motion tissues can’t generate MR signals any longer.

Selecting proper parameters and reducing scan time also can reduce the risk of occurrence of motion artifacts.

1.4.12.2.2 Physiological Motion Artifacts

Physiological motions include breath, heartbeat, pulsate, swallowing and so on.

Breath artifacts are motion artifacts caused by the normal breath physiological motion. You should use

abdomen belt to restrict patient’s breath amplitude and increase average to restrict the influence of breath

artifacts.

Heartbeat artifacts are motion artifacts caused by the normal physiological motion of heart. Because the

heart’s motion is periodic, you can adopt ECG gate control method to restrict heartbeat artifacts.

Page 18: Openmark 5000 User Manual

Equipment Manual

18

Pulsate artifacts are motion artifacts caused by pulsate of trunks, another normal physiological motion. You

can select encoding exchange to change the direction of pulsate artifacts so that improve the quality of image

of interesting area.

Swallow artifacts are motion artifacts caused by swallowing motion of the patient. When scanning

nasopharynx or cervical part, operator should prompt patient to restrict swallow as much as possible in

scanning.

You can consider using pre-saturation to eliminate the above motion artifacts according imaging slices.

1.4.12.2.3 Flow Artifacts

The flow of blood and cerebrospinal fluid can cause artifacts on images, which is called flow artifacts.

You can select sequences with flow compensation function to eliminate artifacts caused by flow within slice.

The influence of flow perpendicular to the slice to images is complex to analyze. For SE sequences, if

scanning only one slice, the blood vessels perpendicular to the slice represent low signals, because the flow of

blood lead to difference between blood excited by 90 degrees pulse and blood excited by 180 degrees pulse;

if performing multi-slice scan, the situation becomes extremely complex: the blood may be excited by 90

degrees pulse in a certain slice, and then be excited by 180 degree in another slice, the echo lies on the

acquisition window of the third slice, now, the blood vessels represent low signal. So that a common situation

is that the blood vessels represents high signal in one slice and represent low signal in another slice.

You can impose pre-saturation belt upon or under the imaging area to saturate the signal of blood vessels

perpendicular to and flow to imaging slice to eliminate the flow artifacts caused by the fluctuate of blood

signal.

1.4.12.3 Metal Artifacts

Ferromagnetic metal foreign matters in the imaging area and at the edge of imaging area can cause artifacts

on the image, which is called metal artifacts.

Metal foreign matters include small ferromagnetic articles, such as hair grips, buttons, brooches, pins,

brassiere and false teeth; various ornaments containing ferromagnetic material, such as hair dye, eye black

and lipstick; internal metal implant, such as medial metal clamps, fixing steel plate, tags with metal material

and intrauterine device.

The cause of occurrence of metal artifacts is destroying of the homogeneity of main magnetic filed, which

makes the image deformed and the signal lost.

The seriousness of metal artifacts is related to the selection of sequences. GR sequence is very sensitive to

metal artifacts. SE and FSE sequences are not so sensitive to metal artifacts compared with GR sequence.

Page 19: Openmark 5000 User Manual

Equipment Manual

19

To eliminating metal artifacts, you should make good efforts to pre-discriminate patients to prevent

unqualified patients from scan and you should require patients to be scanned to take way all removable

ferromagnetic articles before entering scan room.

1.4.12.4 Fault Artifacts

1.4.12.4.1 K Space Partial Data Artifacts

Only partial data is acquired for K space because of hardware failures in scanning, others remain zero.

Artifacts caused in this case are called K space partial data artifacts.

K space partial data artifacts represent different forms according to the occurrence time of hardware failures.

If the failure occurs at the final stage of data acquisition, when acquired data has covered the center of K

space, the artifacts represent as the artifacts in phase encoding direction increase, the images become dim; if

the failure occurs in the beginning stage of data acquisition, when the data haven’t covered the center of K

space, only edge part is shown on image, other parts are noise.

The cause of this failure is gradient amplifier or RF power amplifier is overloaded for the unexpected

reasons. You should reset them first and then scan again to eliminate the artifacts.

1.4.12.4.2 K Space Point Artifacts

In data acquisition, if the system acquires an unexpected signal not belonging to normal MR signals, the

disturbance of parallel strips with equal intervals will appear on the image. Because this disturbance is caused

by an unexpected signal from a point in K space, we call artifacts caused by this reason K space point

artifacts.

The interval of parallel strips on image depends on the position of the point generating unexpected signal: the

closer the point to the center of K space is, the larger the interval of parallel strips is, the more outer the point

to the center of K space, the smaller the interval of parallel strips is (more closer). If more than one

unexpected signals are acquired, the grid artifacts will appear on the images.

The cause of unexpected signal usually is that the power source for shielded room strikes fire, or filter is in

failure. If the occurrence probability is still larger after you turn off the power source for shielded room, you

should contact maintenance personnel of this equipment.

1.4.12.4.3 RF Artifacts

When the radio shielding is abnormal, some lines or strips artifacts will appear along the phase encoding

direction in some fixed positions on the image, which is called RF disturbance artifacts.

If RF disturbance artifacts appear on the image, you should check whether there is unapproved electronic

equipment is operating in shielded room, whether there are cables or wires pass through the door of shielded

room and whether the door of shielded room is closed well. If the above items are all normal, please contact

maintenance personnel of this equipment.

Page 20: Openmark 5000 User Manual

Equipment Manual

20

1.4.12.4.4 Power Frequency Artifacts

If there is power frequency equipment such as high-voltage transmission lines, high power transformer and

electromotor near the equipment, power frequency magnetic filed may influence the equipment and cause

power frequency disturbance artifacts.

The feature of power frequency disturbance is that there are two artifacts along the two sides of phase

encoding direction, most part of which lap over the normal image, disturbing display of the image. The

distance between the center of artifact and the center of normal image is related to the setting of TR. Because

the stability of power frequency disturbance is much weaker than stability of pulse sequence, the situation of

power frequency may be more complex.

You must find out disturbance source and get rid of this source to eliminate power frequency disturbance.

1.4.12.4.5 Saturation Artifacts

Saturation artifacts are artifacts caused by saturated signal deriving from hardware failure. The main features

are as follows: the image is dim and inhomogeneous, and background noise increase obviously. You can scan

again to eliminate the artifacts. If the artifacts still can’t be eliminated after re-scan, you should contact

maintenance personnel of this equipment.

1.4.12.4.6 DC Artifacts

In data acquisition, the DC component of the system can cause a bright spot in the center of the image, which

is called DC artifacts.

Some sequences may cause central bright spot artifacts in one time average scan, which is normal in this

equipment. You can eliminate DC artifacts by selecting even average. If the bright spot still appears after you

select even average, you should contact maintenance personnel of this equipment.

1.4.12.4.7 Frequency Setting Error

When unexpected cause lead to error in setting frequency, the system can’t acquire signal and the acquired

images are full of noise.

You should check hardware first, and then tune frequency again if the hardware is in good condition. If the

RF Power Amplifier or Gradient Power Amplifier is failure, you should reset it and then tune frequency

again.

1.4.13 Recommended training

Only personnel trained by our company and passed examine are allowed to operate this equipment.

Misoperation by untrained personnel will cause equipment trouble; qualified operators can only carry out

applications in the guide, don’t run other program random, or it may cause equipment trouble.

Page 21: Openmark 5000 User Manual

Equipment Manual

21

Operator should strictly follow the procedures of regulations about startup and shutdown equipment in this

guide, don’t leave operation room when scanning. If the equipment is not started up and shut down

according to normal procedures, or can’t scan, please contact our company immediately.

You must follow the concrete operation instructions, warnings, cautions and attentions in every chapter

during system routine operation, or else, any deviation may cause safety accidents.

Generally speaking, this system is not permitted to connect with other systems or components. If the

connection is necessary, it must be approved by technical personnel of our company. It is forbidden to

supply power of this system to other system or components in any case.

Caution!

Unsuitable cable connection may cause part or

whole damage of the equipment.

Warning!

The high voltage and heavy current in some

components in the cabinet of this equipment may

injure people. These components can be treated by

service personnel of our company only.

Attention!

Untrained personnel are forbidden to operate this

system. No person is allowed to run the

unapproved programs, for it may cause

unpredictable equipment trouble.

Attention!

Make sure to obey the operation instructions,

warnings, cautions and attentions in every chapter,

or safety accidents may occur.

Page 22: Openmark 5000 User Manual

Equipment Manual

22

1.4.14 Quality assurance

To keep system’s normal performance and get excellent image, carry on daily system check and system

calibration regularly of MRI.

1.4.13.1 The system needs check and record the result before scanning the patient after system’s starting

every day.

Methods: Put the square sample in the middle of the body coil and to the normal scanning position advanced

in magnet. On APEX interface, scan scheme choose the head–oriented image, pre-scanning, choose the

frequency, Shimming field, and flip angle. Open the spectrometer RINMR interface input the following

orders under P mode, receive Shimming field, and flip angle.

●frequency: order SF request: f±5kHz(f is the original frequency)

●uniformity field: order x offset request<0.004

order y offset request<0.004

order z offset request<0.004

●flip angle: record in file D:\APEX\WORKING\DLLFILES\RFA_5MSA_90. Request ±3.

1.4.13.2 System testing

Test system’s self performance every year, includes: phase place stability, resolution ration, geometric

distortion

● SNR and phase place stability

Put the square sample in the middle of the body coil and to the normal scanning position advanced in

magnet. After tuning the receiving coil, choose SE0_SNR sequence, then choose testing SNR and phase

place, using default parameter, average time is 30, scan and get the results.

*SN R:S/N>8000

*Phase place stability:R M S O F PH A SE<0.05

● Image high-contrast spatial resolution

Select the body coil, place the test phantom on the center of the body coil, and make sure the phantom is

arranged toward the small holes. Then scan, measuring by eye, find the shortest row with the smallest holes.

Page 23: Openmark 5000 User Manual

Equipment Manual

23

The number of visible 1mm holes is more than 7.

● Image geometric distortion

Select the body coil. Place the test phantom in the center of the coil and scanning.

Call out the TRAN, CORO, and SAGI images created during the space resolution test. Measure the

distance between the center of circle and each bright spot on circle respectively.

%10<−

o

oi

RRR

o

opp

RRRR

2221 −+

<10%

1.4.15 Maintenance

All user maintenance must be done by appropriately trained personnel. Individual components of MRI

system contain no user-serviceable parts. Maintenance by untrained personnel may present an electric

shock hazard. See chapter 4 of equipment maintenance guide for further information.

1.4.16 Electromagnetic interference

Using this system may disturb electronic equipment nearby, or maybe the running of electronic equipment

nearby can disturb this system’s running. Keep a appropriate distance from other electronic equipment

while running this system, in case there is still disturbance, please contact the manufactory.

1.4.17 Compatibility technical specifications

Page 24: Openmark 5000 User Manual

Equipment Manual

24

With the purpose of ensure normal running of peripheral equipment test, please refer to chapter2.5 of this

volume for compatibility technical specification of this system.

1.4.18 Environmental protection

This system is electrical product that would bring harm to human and environment if it is disposed of as

domestic garbage at the end of usage period. It must be delivered to recycling center which is appointed to

handle these products. Please contact local government, domestic garbage disposal, its agent or

manufactory for more information of garbage disposal, recovery, recycling station for this kind of

electrical product.

1.4.19 Precautions for system safety

The multi-purpose portable socket-outlet of this system is only designed to supply power for the parts of

this system. The maximum output of the multi-purpose portable socket-outlet is 2.2kVA/220VAC. For

system safety, please make sure that all protective earth conductors of this system are connected with the

common protective earth conductor of the mains when installed, and the scan room is only used for patient

environment.

Warning!

Don’t connect an additional multiple portable

socket-outlet to this system.

Warning!

Don’t place the multiple portable socket-outlet on

the floor.

Page 25: Openmark 5000 User Manual

Equipment Manual

25

1.5 Alarm system

Please seriously pay attention to the alarm signals and indictors during operating this MRI system. The alarm

system of this MRI system is described as follows:

Item Alarm condition Priority Alarm signal frequency Indicator

color

1 Can not connect to database Low Fault to connect to

database

Constant Yellow

2 Hard disk’s capacity is not

enough

Low Hard disk’s capacity is

not enough

Constant Yellow

3 Coil tuning reflection is too

strong

Low Coil tuning reflection is

too strong

Constant Yellow

4 TE/TR is too short Low TE/TR is too short,

please increase TE/TR

to XXX

Constant Yellow

5 Make filter post processing

during scanning

Low Can not make filter post

processing during

Constant Yellow

Caution!

Connecting electrical equipment which has not

been supplied as a part of this system to the

multiple portable socket-outlet must comply with

the IEC60601-1-1, Otherwise that may result in

personnel injury.

Caution!

Connecting the computer or other equipment

which has been supplied as a part of this system

directly to the wall outlet may result in personnel

injury.

Page 26: Openmark 5000 User Manual

Equipment Manual

26

scanning

6 Signal is too low when the

gain gets up to the limit

Low Signal is too low, gain

has got up to the limit

Constant Yellow

7 Tuning channel does not

work

Low Tuning channel fault constant Yellow

1.6 Manufacturer’s Responsibility

The MRI system is sold with the following understandings that the manufacturer, its agent, and

representative do not accept any responsibility for personal injury or death, or equipment damage due to

improper use (authorized or unauthorized persons).

Page 27: Openmark 5000 User Manual

Equipment Manual

27

Chapter 2 Structure and Components of the Equipment

The components of OPENMARK 5000 MRI equipment, the usage of components and units, brief of

operating principle and key performance index are described in this chapter.

2.1 Structure of the Equipment

The site of MRI equipment consists of Scan Room, Engine Room, Operating Room and Air Conditioner

Room. Scan Room has radio frequency (RF) shielding effect: it can shield the disturbance of external radio

wave to MRI signal when closing the door of the Scan Room. There are magnet, gradient coils, transmitting

coil, RF Preamplifier, receiving coils, magnet shell and Patient table in the Scan Room; there is a

transmission plate on the wall facing Engine Room; the filters and radio channels are mounted on the

transmission plate; the transmission plate is connected to cabinet in Engine Room with cables. In Engine

Room, there is cabinet, auxiliary power supply (APS), isolated electric voltage regulator for the whole

equipment; there are spectrometer, gradient power amplifier (PA), and RF power amplifier (PA) and tuning

box in cabinet. There are magnet temperature control unit and lighting power supply for Scan Room in

auxiliary power supply (APS). Operating Room is the place for operator to carry out routine work, there are

operating desk, uninterrupted power supply (UPS) in it, and there are computer, intercom subsystem, ECG

gate controller (optional) and other options on Console. There are air conditioner and blower fans in Air

Conditioner Room which blow wind to Scan Room for keeping temperature of Scan Room in comfortable

arrange for patients.

Structure of the OPENMARK 5000 MRI Equipment

UPS

Printer &Options

Camera

Computer

Display

Intercomsubsystem

Spectrometer

TuningBox

RF PA

GradientAmplifier

APS AirConditioner

ElectricVoltageRegulator

Operating Room Engine Room Scan Room

Air Conditioner Room

BlowerFans

Magnet and Coils

RFPreamplifier

Page 28: Openmark 5000 User Manual

Equipment Manual

28

Signal flow of Open 5000 MRI equipment is as follows: the computer-controlled spectrometer generates

special pulse sequence according the requirements of user, the RF signal of pulse sequence is amplified by

RF power amplifier and transmitted to transmitting coil to generate RF magnetic field; the gradient signal of

pulse sequence is amplified by gradient power amplifier, and transmitted to gradient coils to generate

gradient magnetic field. The objective examined generates magnetic resonance (MR) signal under the

action of RF magnetic field and gradient magnetic field. The MR signal is received by receiving coil and

amplified and transmitted to spectrometer for synchronous demodulation and data acquisition, the acquired

data is transmitted to computer to rebuild MR images, MR images are transmitted to display for showing

and analyzing.

Signal flow of OPENMARK 5000 MRI Equipment

The power distribution of whole system is as shown in the following page.

The input of isolated electric voltage regulator is 380 V AC, and its output are 5 channels voltage: channel 1

supplies power to UPS, intercom subsystem, camera, printer, ECG gate Controller, in which UPS is

specially for computer; channel 2 supplies power to spectrometer and tuning box, in which tuning box

powers Patient table locating system in Scan Room; channel 3 supplies power to gradient PA; channel 4

supplies power to RF PA; channel5 supplies power to RF Preamplifier.

The input of APS is 220V AC, and its output are 3 channels of voltage : channel 1 powers Scan Room

lighting; channel 2 powers wall sockets on the wall of Scan Room; channel 3 powers to magnet

temperature.

Page 29: Openmark 5000 User Manual

Equipment Manual

29

Air conditioner and blower fans are powered with 220V and 380V AC separately.

P

o

w

er

D

is

tr

ib

ut

io

n

C

h

ar

t

of OPENMARK 5000 MRI Equipment

2.1.1 Scan Room

Scan Room is patient environment and controlled access area of approach is here too. Unapproved persons

are not allowed to enter Scan Room. For fear of serious consequences, operator should prompt the patients

to read the warning signs and inhibit signs at the door of Scan Room before they enter, and read the warning

signs in the front face of magnet in Scan Room carefully.

Besides magnet and Patient Table, the components used in patient environment include bed cushions,

various receiving coils and various one-off medical sheets, caps, cotton piece.

Warning!

All outputs of isolation electric voltage regulator

are special source for this system and you are not

allowed to connect other equipment to them without

permission.

UPS

Printer &Options

Camera

Computer

Display

Intercom

Subsystem

`

Spectrometer

TuningBox

RF PA

GradientPA

APS

RFPreamplifier

AirConditione

r

ElectricVoltage

Regulator

Operating room Engine Room Scan Room

Air Conditioner Room

Blower

Fans

Magnet and Coils

~380V3Φ4W

~380V3Φ4W~220V

1Φ3W

~220V1Φ3W

~220V1Φ3W

~220V1Φ3W

~220V1Φ3W

~220V1Φ3W

~36V1Φ2W

~220V1Φ3W

+15V1Φ2W

+48V

1Φ2W

~380V/3Φ4W

WallSocket

Lighting

~220V1Φ2W

+36V1Φ2W

1Ω Grounding

+24V1Φ2W

Page 30: Openmark 5000 User Manual

Equipment Manual

30

The environmental requirements to Scan Room are as follows: temperature: 22±3℃, relative humidity:

40%-80%.

Any Fault of air-condition in air-condition room would affect the temperature in scan room, so the operator

is advised to monitor the scan room; temperature usually.

2.1.2 Engine Room

Engine Room is equipped with door lock, and untrained persons are not allowed to enter it. The trained

operator only can enter the Engine Room to start up and shut down the equipment. The environmental

requirements to Engine Room are as follows: temperature: 21±5℃; relative humidity: 40%-80%; air: dust

free.

2.1.3 Air Conditioner Room

The air conditioner is equipped with door lock, and untrained persons are not allowed to enter it. The trained

operator only can enter Air Conditioner Room to start up and shut down the air conditioner for Scan Room

and to handle fortuitous event.

2.1.4 Operating Room

Operating Room is the space for operator to do routine work. Untrained persons are forbidden to operate

this system. Environmental requirements to Operating Room are as follows: temperature: 21±5℃;

Relative humidity: 40%-80%.

Warning!

There are components with high voltage and heavy

current in cabinet and transmission plate, untrained

persons are forbidden to enter Engine room.

Warning!

There are components with high voltage and heavy

current on air conditioner equipment, untrained

persons are forbidden to enter air conditioner room.

Page 31: Openmark 5000 User Manual

Equipment Manual

31

2.2 Components of the Equipment

The key components of 5000 MRI equipment include magnet, Patient Table, cabinet, transmission plate,

Console, receiving coils, etc.

Key components are introduced as follows:

Key Components of OPENMARK 5000 MRI Equipment

2.2.1 Magnet

OPENMARK 5000 Magnet is made from rare earth permanent magnet mater NdFeB, after magnetizing,

Once magnetized, excepting accident, NdFeB will keep magnetism permanently. This magnetic will keep

being regardless the system is running or not.

Sketch map of edge magnetic field of magnet shows the three-view drawing of the distribution of edge

magnetic along space contour line, the unit of spatial distance is mm, and the strength of magnetic field is

Gs.

User should operate the equipment according to safety precautions in section 1.4 for fear of personal injury.

Page 32: Openmark 5000 User Manual

Equipment Manual

32

Sketch Map of Edge Magnetic Filed of Magnet of OPENMARK 5000

2.2.2 Patient Table

Patient table is used to support patients, and its max load capacity is 135Kg. The Patient table consists of

bedstead, bedplate and floor guide rails. The bedstead can move forth and back in the plane, and bedplate

can move linearly. Through the cross cursor of locating light and distance display, you can move the

scanning part of the patient to the center of magnet. After scanning, you can move the Patient Table out of

the center of magnet manually. You should pull out the Patient table with hand from magnet in case of

unexpected conditions.

For fear of accidental injury, operator should abide by the followings when operating beside of Patient

Table: 1. Operator should prompt patients to close eyes and mustn’t stare at the cross laser cursor generated

by locating light during locating operation; 2. Operator should put the hands on the upper side of Patient

table and on the handle of grip brake, don’t put hands under the Patient Table, or else the Patient table and

shell may crush the fingers.

Warning!

Please prompt patient to close eyes and mustn’t

stare at locating laser beam during locating

operation.

Please put your hands on upper side of patient table

and on the handle of grip brake, not be crashed by

patient table.

Page 33: Openmark 5000 User Manual

Equipment Manual

33

2.2.3 Receiving Coils and RF Preamplifier

Receiving coils receives the electromagnetic wave radiated by stimulated protons in selected part of patient.

Operator should select proper coil from coils for head, neck, chest, abdomen, spinal column, joints, breast

etc.

The signal of receiving coil feeds to RF Preamplifier though the socket on shell of magnet. Please make

sure that coil plug is reliably connected with socket on the shell of magnet. The socket on the shell of

magnet is forbidden to use for other purpose, it is only use for coil plug connection. RF Preamplifier

amplifies MR signal. The amplified MR signal is transmitted to the spectrometer for process.

2.2.4 Cabinet

The cabinet is the most centralized place for electronic units. There are high voltage and heavy current in it,

so that there are protective doors both in front of and behind the cabinet. Untrained personnel are forbidden

to open the protective doors. In normal condition, switches on electric voltage regulator control the power

to all components in cabinet and operator is not necessary to open the protective doors.

2.2.4.1 Spectrometer

Spectrometer is the real-time control unit of OPENMARK 5000 MRI equipment, which can acquire MR

signal in real time through running the programmed pulse sequence. Spectrometer can also receive external

trigger signal to synchronize pulse sequence with external events, or output synchronized signal to external

to realize synchronous operation with external equipment.

Warning!

In normal condition, turnon and turnoff of

spectrometer are controlled by switch on the

cabinet of electric voltage regulator. Please don’t

move the switch of spectrometer without

permission

Warning!

There are units with high voltage and heavy current

in the cabinet, untrained personnel are forbidden to

open protective doors.

Page 34: Openmark 5000 User Manual

Equipment Manual

34

Spectrometer consists of system control unit, timing controller, gradient waveform generator, RF unit and

data acquisition unit.

The power switch of spectrometer is in the back of spectrometer. In normal condition, the switch of electric

voltage regulator controls the turnon and turnoff of the spectrometer. Please don’t turn the switch in the

back of the spectrometer without permission. See section 3.2 for concrete operation procedure.

2.2.4.2 Tuning Box

The main functions of tuning box are as follows:

(1) Automatically recognize the type of receiving coil (50 ohms single coil or quadrature coils).

(2) Output test signal for tuning receiving coils.

(3) Receive the parameters of tuned voltage delivered by computer, and transform it to voltage signal

thought D/A transformation, and then apply this voltage to receiving coils.

(4) Provide power supply for locating circuit

The turnon and turnoff of tuning box are controlled by electric voltage regulator; see section 3.2 for

concrete operation procedure. In normal condition, please don’t carry out any operation on tuning box.

2.2.4.3 Gradient Power Amplifier

Gradient power amplifier establishes gradient magnetic field with gradient coils mounted within the shell of

magnet.

Gradient power amplifier can separately put out three channels of current to drive three channels of gradient

coils and establish gradient magnetic field along three orthogonal directions in space.

Turnon and turnoff of gradient power amplifier are controlled by electric voltage regulator. See section 3.2

for concrete operation procedure. In normal condition, it is not necessary to carry out any operation on

gradient power amplifier.

Warning!

There are high voltage and heavy current in gradient

power amplifier. Please don’t touch any component

of output.

Page 35: Openmark 5000 User Manual

Equipment Manual

35

2.2.4.4 RF Power Amplifier

RF power amplifier establishes RF magnetic field with transmitting coils mounted within the shell of

magnet.

There is RF radiation to patients when RF power amplifier is working normally. Even if output with max

power, the SAR (specific absorption rate) is always much lower than damage level to patients and

operators.

Turnon and turnoff of RF power amplifier are controlled by electric voltage regulator. See section 3.2 for

concrete procedure. In normal condition, please don’t carry out any operation on RF power amplifier.

Warning!

There is high voltage and heavy current in RF power

amplifier. Please don’t touch any component of

output.

Caution!

Output connectors of RF power amplifier are in the

back of cabinet. Without permission, operator is

forbidden to loose connectors, or it may damage

amplifier.

Warning!

Any change of buttons, switches and connectors on

gradient power amplifier may cause system damage

or life risk.

Page 36: Openmark 5000 User Manual

Equipment Manual

36

2.2.5 Auxiliary Power Supply (APS)

APS is mounted in Engine Room and there is magnet temperature control unit and lighting power supply for

Scan Room in it.

This APS is powered by special cable. Without permission, any one is forbidden to loose the cable in the

back of APS. After installation of the equipment, the APS should always be in working status. The power

supply to APS should never be cut down unless fire and other natural hazard with force majeure occur.

Magnet temperature control unit controls temperature of magnet to keep the temperature of magnet stable.

The front panel of this APS shows average temperature of magnet. In normal condition, the temperature is

stably in a range of a constant±0.1 ℃. If the average temperature changes sharply, which indicates that the

magnet temperature control part or display part may be in trouble, you should contact maintenance

personnel of our company in time.

The lighting power supply for Scan Room is connected to Scan Room through the filters on the

transmission plate. Scan Room is lighted by 36VDC incandescent lamps.

2.2.6 Transmission Plate

The transmission plate is mounted on the wall of Scan Room and is used to connect external signals and

internal signals of Scan Room. The MR signal induced by receiving coil is very weak so that the Scan Room

is equipped with radio shielding. For fear of RF leakage, all signals coming in and out of Scan Room must

be filtered (except for RF output and input signal).

Filers are mounted on the transmission plate. This system uses various types of filter, and different type of

filter can’t replace each other in use.

There are protective doors with locks at both internal and external side of transmission plate. There are high

voltage and heavy current passing through the transmission plate. For fear of accident, untrained personnel

are forbidden to open protective doors of transmission plate or change the connection on the plate.

Warning!

Without permission, none is allowed to loose the

cable connection in the back of APS. The power

supply to APS should never be cut off unless fire

and natural hazard with force majeure occur.

Page 37: Openmark 5000 User Manual

Equipment Manual

37

2.2.7 Console

The console is the working interface for operator. UPS, computer and intercom subsystem as well as

options such as printer, camera is on or under the console.

The console is dedicated for this equipment. You are not allowed to place other articles beyond the

components supplied by our company on the console, say nothing to place heavy on the edge of the console.

There is a special power outlet on the console, whose output voltage is 220V~, 50/60Hz. the maximum

output power is 1kVA. This outlet powers only for intercom subsystem, printer (optional), camera

(optional) and other optional devices. Without permission, any other equipment is forbidden to use this

special outlet.

Warning!

There are high voltage and heavy current passing

through the transmission plate. Without

permission, any one is forbidden to open the

protective doors and to touch connection poles.

Warning!

Using special power outlet on console without

permission may cause personal injury or equipment

damage.

Warning!

Unequipped articles of this equipment are not

allowed to place on the console.

Page 38: Openmark 5000 User Manual

Equipment Manual

38

UPS only powers computer, any other equipment is forbidden to connect to UPS. The function of UPS is to

continue to power computer after unexpected power cut, so that computer can complete normal exiting

operation, protecting file system from damage.

AS the central control unit of the whole MRI equipment, computer controls and coordinates running of

every parts of the equipment, undertakes many key tasks such as managing database, reconstructing

original data acquired from scan to image data, and displaying and analyzing images. Refer to computer

user manual for computer’s running environment.

The computer adopts DELL workstation, which has outstanding performance and strong capability of

numerical calculation. Double hard disks system is equipped with the computer, which obviously improved

the reliability of the equipment.

Computer has many kinds of data transmission interfaces. One of network interfaces of computer connects

to spectrometer to transmit sequence, parameter and control spectrometer, and get original data from

spectrometer to hard disk of computer. The second network interface, used as the interface between the

system and camera, supports DICOM3.0 standard and requires the IP address of camera to be

192.9.200.232. RS232 interface of computer connects to tuning box, transmits tuning and controlling

parameters to receiving coil, and recognize type of receiving coil, etc. There are several USB interfaces on

computer that can connect plug-and-play devices. Printer is connected to the computer by one USB

interface.

The computer used in the system has stronger power of image processing. The computer can display the

reconstructed images on screen for doctors to analyze and diagnose, with various kinds of operation such as

zoom in, zoom out, flip, gray scale transform, etc. The computer is equipped with graphic accelerating

processor, which can process images more quickly. The computer uses 19" or above, high resolution, color

display. All operations to the software are carried out through keyboard and mouse.

The computer has equipped with CDROM and DVD/CDRW driver, which can read and save data. See Vol.

Ⅱof this guide for data saving. The computer is dedicated for this equipment. It is forbidden to set up and

run unapproved software on the computer or connect unapproved network link, or else, it may damage

system software.

Caution!

It is forbidden to set up and run unapproved

software on the computer or connect unproved

network link. Or else, it may cause damage of the

system software.

Page 39: Openmark 5000 User Manual

Equipment Manual

39

Turn on and turn off the computer are controlled by panel switch of electric voltage regulator and UPS;

before shutdown of computer, you mustn’t turn off the power supply of computer, or else, the file system

may be damaged. In case of sudden power cut, you must follow the operation regulation in section 3.2 to

exit the software system under support of the UPS.

Intercom subsystem is a semiduplex intercom device, which is used to talk between operator and patient. In

normal condition, patient can speak to operator through microphone mounted in magnet cavity; when

operator pushes down the button of communication on intercom system, he can speak to patient in the

magnet cavity. This Intercom subsystem also can play background music.

2.2.8 Three-Phase Isolated Electric Voltage Regulator

Three-phase isolated electric voltage regulator is the main power supply of the equipment and is installed in

Engine Room.

Input voltage of three-phase isolated electric voltage regulator is three phase, 380VAC (line-to-line),

50/60Hz; rated input power of it is required more than 15kVA; connection method of it is three-phase

four-wire system.

Electric voltage regulator is the power control center of the whole equipment. It controls the power

distribution to each device. It controls turnon and turnoff of each device through the switches on the panel.

You must operate electric voltage regulator according to operation regulation. See section 3.2 for details.

The type of replaceable fuse of electric voltage regulator is RT18, and rated value is T6AL/500V, which

complies with criterion IEC269.

There is high voltage and heavy current in three-phase isolated electric voltage regulator, the power switch

of its own is in the back of it and the power switches for other devices are in front panel of it. Untrained

personnel are forbidden to operate switches on electric voltage regulator.

Caution!

In normal turn off computer or in case of sudden

power cut, you must follow the operation

regulation to exit the software system under support

of the UPS, instead of turn off computer power

source directly.

Page 40: Openmark 5000 User Manual

Equipment Manual

40

2.2.9 Camera

It only permits to connect the camera qualified by our company.

2.3 Brief Introduction of Operating Principle

Before scanning, you should carry out equipment’s preparation, which includes powering devices

(Console, cabinet, gradient PA, RF PA and RF Preamplifier) and Software log-on, etc.

After equipment’s preparation, you should carry out patient’s preparation, including pre-discriminating

patients, letting patients to read safety precautions, registering patients, installing coils, positioning patient

and tuning coils, etc.

Operator can start to scan patient after equipment’s preparation and patient’s preparation. You should select

pulse sequence and parameters, after you confirm your selection and start this scan; the computer

downloads the corresponding pulse sequence and parameters into spectrometer and then run this pulse

sequence in spectrometer. Pulse sequence produces the output of gradient and RF pulse in predestined

timing. After amplified, those pulses establish the gradient and RF magnet field in magnet cavity, which

excite the protons in selected part of patient’s body. The excited protons begin to relax after the RF filed

finishes. The weak signal emitted by protons in the relaxation process is received by receiving coils. This

weak signal is amplified by RF Preamplifier and delivered to the receiving channel of spectrometer. The

pulse sequence controls the spectrometer to acquire data at the special time segment of protons relaxation.

The acquired data contain the space encoding information and special physical and chemical information of

scanned part of the patient. These data is transferred to computer and saved in the hard disk. After scan,

through rebuilding the data you can get images containing special physical and chemical information, saved

in the hard disk. Operator can handle and display the image through mouse according to different desires.

You can do almost any other work during data acquisition: When a scan begins, the operator can select

another pulse sequence and change parameters for next scan, and start up many new scans, instead of

waiting for the end of the current scan. When current scan finish, the system can automatically start up the

next scan according to the starting up order; you can load images, adjust gray scale, analyze images, add

Warning!

Misoperation to three-phase isolation electric

voltage regulator may cause personal injury or

equipment damage.

Without permission, it is forbidden to change fuse,

no speaking of using other type of fuse.

Page 41: Openmark 5000 User Manual

Equipment Manual

41

notations, print films, etc; you also can register the next patient. Note: You can’t enhance images during

scanning.

2.4 Transporting and Storing of the Components

This equipment is immobility type. Once finishing installation and commissioning, no one is allowed to

move any fixed unit or fixed component of this equipment.

The environmental requirements in transportation and storage of the equipment are as follows:

⑴ Range of Temperature: -40℃-55℃

⑵ Range of Relative Humidity: ≤95%.

⑶ Atmospheric Pressure: 860hPa~1060hPa.

⑷ No acidic or alkaline matter or corrosives around components.

⑸ When storage, there is no other ferromagnetic substance to be allowed within 2.5 meters around the

magnet.

2.5 Key Performance Index

2.5.1 Magnet

Type: Double column type permanent magnet

Strength of Static Magnet Filed: 0.51T± 5%

Homogeneity of Static Magnet Filed: 1.7ppm/30cm, 2.5ppm/40cm×40cm×34cm, ellipsoid

Shimming: passive

Control precision of magnet temperature: ±0.1℃

2.5.2 Gradient System

Strength of Gradient Field: Max 30mT/m.

Slew Rate of Gradient: Max 100T/m/s.

dB/dt (time rate of change of magnetic field): always less than 20T/s, under normal using condition.

Nonlinearity of Gradient: less than 5%@40×40×34cm, ellipsoid

2.5.3 RF System (Spectrometer included)

Center Frequency: 21.29MHz

RF Power: max 6KW

Page 42: Openmark 5000 User Manual

Equipment Manual

42

SAR (specific absorption rate): always less than 1.5W/kg in any 15minutes under normal using condition of

any part of human body.

Receiving Coils: head coil, chest coil, abdomen coil, neck coil, spinal column coil, breast coil, knee coil,

wrist coil, shoulder coil and ankle coil.

Other: RF Self Shielding.

2.5.4 Computer System

Computer: DELL workstation, 2GB EMS, 160×2GB HD, 19” and above Display, keyboard/mouse

External Storage: standard DVD-RW.

Network Capability: Network interface, can link up network to PACS system based on workstation or PC.

2.5.5 Patient System

Patient Table: dynamoelectric Patient Table, width 85cm, max 135kg patient load

Height of patient space: 40cm.

Locating: Cross laser locating light, LCD display.

Position limit of Patient table: mechanical limit.

Patient Monitoring: intercom subsystem and observation window

2.5.6 Software Configuration

This equipment is equipped with special MRI software produced by Anke.

2.5.7 External Power supply

Requirements for External Power supply for Equipment: 380V 3N~,50Hz,15kVA.

2.5.8 Image Performance Index

Image display: the maximum grey scale is not less than 256, display matrix is 256×256, 512×512,

1024×1024.

SNR: According to the requirements of clause 4.2 in YY/T0482-2004, refer to following table 1 for system

image SNR.

Coil Head coil

14"body coil

17"body coil

20"body coil

Neck coil

Breast coil

Knee coil

Shoulder coil

Ankle coil Wrist coil

SNR ≥200 ≥120 ≥120 ≥120 ≥100 ≥100 ≥120 ≥100 ≥160 ≥300

Page 43: Openmark 5000 User Manual

Equipment Manual

43

Table 1 Image SNR

Image geometric distortion: According to the requirements of clause 4.3 in YY/T0482-2004, the

geometric distortion of system image should be not more than 5%.

Image uniformity: According to the requirements of clause 4.4 in YY/T0482-2004, uniformity of system

image should be not less than 75%.

High contrast spatial resolution: According to the requirements of clause 4.7 in YY/T0482-2004, high

contrast spatial resolution of system should be not more than 0.5mm.

Low contrast resolution: low contrast resolution should be not worse than distinguishing circular hole

with 2mm diameter and 0.5mm depth.

Slice thickness: According to the requirements of clause 4.8 in YY/T0482-2004, the thinnest slice is 1mm,

classical thickness is 5mm, 7mm. thickness error should be no more than ±1mm.In addition, there

are many kinds of thickness to choose, such as 2mm、3mm、4mm、5mm、6mm、8mm、9mm、10mm.

Positioning and Slice gap: According to the requirements of clause 4.9 in YY/T0482-2004, slice

positioning error should be no more than ±3mm, slice gap error should be no more than ±1mm or less

than 20%, use the larger one of these.

Image aspect ratio: image aspect ratio of circle and square model should be between 90% and 110%.

Page 44: Openmark 5000 User Manual

Equipment Manual

44

Chapter 3 Routine Operation Guide

Some general operations are described in this chapter such as startup, shutdown, patient preparations, usage

of receiving coils, operations of Console, operations of Patient Table and Center Locating. See Vol.Ⅱ for

MRI software operations.

3.1 Routine Scan Procedure Reference

Serial Number Process

1 Review of Safety Precaution

2 Startup

3 Patient Preparations

4 Patient Registration

5 Tuning

6 Locating Scan

7 Scan

8 Displaying and Analyzing Images

9 Print films

10 Shutdown

Before daily scanning, it is highly recommended to put the phantom in the center of magnet and tune the

MR system with abdomen coil in order to perform essential adjustment and ensure the quality of images.

3.2 Start-up and Shutdown

In normal condition, the power on and power off the system are controlled by the voltage regulator and

UPS. In order to avoiding damaging the software system, you can not directly power off the voltage

regulator or UPS until exiting the software system according to operation regulation, even if in case of

sudden power cut, you must exit the software system under the support of UPS according to operation

regulation in this section before turn off the voltage regulator and UPS.

Start-up Steps:

Page 45: Openmark 5000 User Manual

Equipment Manual

45

1) Turn on the switch of voltage regulator in engine room, the “POWER” indicator and “STOP” indicator is

lighted on when the voltage regulator is in good connection, then press “STOP” bottom and the “STOP”

indicator is off, that means the voltage regulator is connected to mains power supply successfully.

It displays normal output voltage on the screen if the voltage regulator works in normal condition, and no

any anomaly displaying.

2) Waiting for 10 seconds when voltage regulator entering normal operation condition, Press down the

“START” indicator on the front panel, and the “START” indicator is on automatically. You can hear the

sound of AC contactor taking actions and voltage regulator works.

3) Confirm that voltage regulator works in normal condition, turn on “Console K1”, “Cabinet K2” in order,

and wait for a sounds “di da di….” from spectrometer, then continue to turn on “Gradient K3”, “RF

Amplifier K4” and “RF Preamplifier K5” in order. All of the system parts have been powered on, and all

their power indicators are lighted on.

4) Turn on air conditioner and set the proper temperature (21°±5℃)

Check: after step3, the green indicator in the front panel of RF Power Amplifier is on, indicating that power

source is connected. A few seconds later, yellow indicator flickers, after a few second flicker, the green

indicator of “POWER” and “READY” is on, indicating that RF Power Amplifier is ready to work.

Check: after step 3, the green indicator in the front panel of RF Power Amplifier is on, indicating that power

source is connected. a few seconds later, the green indicator of “POWER” and NORMA:L” is on, indicating

that Gradient Power Amplifier is ready to work., If other is displayed, indicating that something must be

wrong.

5) Turn on the switch of the UPS under the console. After powering on computer system, the display on

console shows the startup and self-check information of the computer. This process may take several

Console

Preamplifier

RF PA

Gradient

Cabient

Start

Stop

Page 46: Openmark 5000 User Manual

Equipment Manual

46

minutes. When Microsoft Windows XP starts up and self-check finish, the following information appears

on the screen:

User’s Name:

Password:

Now the self-check and startup finish and the system are ready to log on.

6) Type user’s name and password and then press [enter]. The system switch to main interface of

Windows XP; double click RINMR icon on desktop;After the interface of RINMR appears, then double

click shortcut icon of MRI software to enter MRI system.

If any display different from the above description appears, which indicates the failure of the startup, please

enter troubleshooting procedure, see chapter 5 for details.

Shut down steps: (The following steps must be carried out even in case of sudden power cut)

1) Exit MRI software system;

2) Exit RINMR;

3) Press and hold down Ctrl and ALT key, press DEL key at the same time, choose “Shutdown” in prompt

menu and choose “Shutdown” in shutdown menu.

4) Turn the UPS off after the computer is powered off automatically.

5) In machine room, pull down the five switches on the front panel of voltage regulator to turn off them in

the following order: “Console”, “Cabinet”, “RF Amplifier”, “RF Preamplifier”, and “Gradient”.

6) Press down “STOP” indicator on the front panel of voltage regulator, “START” indicator is off and

“STOP” indicator is on. AC connector makes break sound. (This step is elided in case of sudden power

cut).

7) On the back of voltage regulator, Pull down the triple switch marked “Input Switch” to turn off the power

of voltage regulator.

8) Shut down the air conditioner in machine room and lock the door of machine room if not starting up any

longer in the current day.

It is recommended starting up in every morning before work and shutting down after whole day’s scan.

Emergency is exception.

3.3 Console Operations

There are a set of computer, intercom subsystem, printer (optional), camera (optional), etc on the control

desk; UPS and Talk/Listen foot switch under the desk.

Page 47: Openmark 5000 User Manual

Equipment Manual

47

The equipment is equipped with high quality color monitor, which only needs to adjust brightness, contrast

and geometric position. If the brightness and contrast of monitor are satisfied, please don’t adjust them any

longer.

This equipment adopts standard computer keyboard.

Mouse is used to move cursor and make selection in menu. In general, press left button to select and press

right button to display submenu.

Please refer to their own operating manuals of intercom subsystem, UPS and other options such as

printer, camera, etc.

See relative chapters in vol. Ⅱ for keyboard and mouse operations for MRI software.

3.4 Patient Table Operation

3.4.1 Pre-screening of patients

The patients must be pre-screened before they enter scan room to perform further scanning.

The patients who may have life risk induced by strong magnetic field, including patients with cardiac

pacemakers, metal foreign object in eyeball, metal tubes and clips leaved in the body after surgical or larger

mass of ferromagnetic substance in body, are forbidden to make MR examination or even enter scan room.

Traumatic patients must make sure that there is no metal in body before examine.

Patients with warding devices, including patients impending death and in emergency treatment, are not

applicable to make MR examine.

For the patients who need to cautiously treat for the special physical environment, including patients with

less ferromagnetic substance in body, and patients in pregnancy within 12 weeks, doctors should carefully

ask about the medical history of patients, weight the benefits getting from and risks deriving from MR

examine. Only when the benefits are far more than the risks can the MR examination be carried out.

Patients who may can’t endure routine scan examine for the sake of patient’s own cause, including patients

in mania, patients who can’t repose five minutes, children within 6 years old, if they are really need to make

MR examination, sedative is necessary before examination, and must be accompanied by family numbers

during scanning , judged and analyzed by doctors.

Warning!

Pre-discrimination on patients must be done

before MR examination. Inapplicable patients for

MR examine may be injured in scan room.

Page 48: Openmark 5000 User Manual

Equipment Manual

48

3.4.2 Patient Table Introduction

Patient table is used to support patients, and its max load capacity is 135Kg. The Patient table can be

operated dynamoelectric and manually. In any case, the patient table can be moved by slightly push. After

scanning or come across accident, patient table can be immediately moved out manually.

There are two patient table control and display panels laying lying on two side of shell’s top cover, that are

symmetrical from left to right while you face to the magnet, and they have the same function. The operator

can use the left on or right one, but can not operate both of them at the same time.

You can operate patient table through patient table control and display panel, which controls the movement,

positioning, display of patient table. The number on the patient table control screen shows relative position of

patient table, the unit is “mm”. The number will decrease if the patient table moves into magnet, and the

number will increase if the patient table moves out of magnet.

3.4.3 Manual Patient Table Operation

Operator moves the patient table slowly by pulling and pushing the knob at the end of bedplate. And patient

table can move inside or outside of magnet.

Press the “SET” key to tune on the cross cursor of positioning light, finely adjust the position of coil by

pulling the patient table to make cross cursor laps over the center of the coil or the interesting part; press

“SET” key again to turn off position light; push patient table to the center of magnet while the display

number decrease to zero, Now the center of coil or interesting part is in center of the magnet.

Emergency condition, grasp the grip brake at the end of bedstead to pull the patient table out from magnetic

center quickly.

Please do not energize too much when operation patient table manually, do not quickly move patient table,

it must decelerate when patient table moving forwards/back limiting position, to avoid stopping patient

table suddenly so as to frighten patient and mar patient table.

3.4.4 Electric Patient Table Operation

Forward: Press continuous move forward or move forward button, patient table move forward automatically.

Attention!

Grasp the grip brake at the end of

bedstead to pull the patient table out from

magnetic center quickly in emergency

Page 49: Openmark 5000 User Manual

Equipment Manual

49

Back: Press continuous move back or move back button, patient table move back automatically.

Stop: Press Stop button or release the move forward or move back button.

Press the “SET” key to tune on the cross cursor of positioning light, finely adjust the position of coil by

moving patient table under electric control to make cross cursor laps over the center of the coil or the

interesting part; press “SET” key again to turn off position light; push patient table to the center of magnet

while the display number decrease to zero, Now the center of coil or interesting part is in center of the

magnet.

It is forbidden to press two or more than two button at the same time, it would cause damage for patient

table.

It is forbidden to operate the two control panels at the two side of magnet shell.

Please do not switch the moving direction suddenly while operating the electric patient table, to avoid

damage.

3.4.5 Patient’s laying position

Lying on back or abdomen is available; Lying the head of patient in the both side of the patient table is also

available, incorrect patient table operate will lead to unexpected harm for patient.

Right Lying Posture

Page 50: Openmark 5000 User Manual

Equipment Manual

50

Wrong Lying Posture

3.4.6 Safety Precautions on Patient table Operation

In patient table operation, the following places on the patient table, shown as in the fig, must be paid more

attention for safety.

Various gaps on the patient table are prone to clip the objects, such as finger, hair, clothes, etc. The

positioning light is a laser cross cursor with minimal power. Please avoid laser beam irradiate patient’s eyes

in any case.

The Places Must Be Paid More Attention For Safety. The gaps between patient table and shell. The gaps between bedstead and bedplate. The gaps between bedstead and the support of patient table. The salient of shell. Danger zone which means movement range of patient table. Position light.

Incorrect operations may injure patients or operators or damage equipment accidentally. Please pay

attention to following conditions:

Please don’t use too much strength when operating patient table, the moving speed of patient table

shouldn’t be too quick. The table should slow down before approaching the limited positions of back or

front, right or left, to avoid sudden stop, which may frighten patients or damage patient table.

When patient goes to or leaves patient table, the patient’s center of gravity should lie upon the part of

bedplate which is within the bedstead supporting area, don’t lie upon part of bedplate beyond bedstead

supporting area, or else, the bedplate may turn over.

When pulling or pushing patient table, operator should put hand on handle of bedstead or knob of

bedplate. Don’t put your hands between bedstead and bedplate or the shell of magnet for fear of crash.

If patient doesn’t lie on the table in correct position, the head, legs or other parts of patient’s body may

crash on magnet when the table is moving.

Patient’s clothes may be clipped. Operator should prompt and help patients to do with their clothes.

Long hair or plait of patient may be clipped by the gaps on the table. Patients with long hair or long plait

should wear one-off head cap.

After selecting and placing proper table cushions and coils, operator should lay one-off bed sheet on the

cushion for patient, and then let patient to lay on it. To avoid cross infection between patients, operator

should change the sheet after scan.

Page 51: Openmark 5000 User Manual

Equipment Manual

51

The positioning light is a laser cross cursor with minimal power. To avoid laser beam irradiate patient’s

eyes, operator should prompt patient to close eyes and mustn’t stare at laser beam when carrying out

positioning operation.

3.5 Usage of Coils

3.5.1 Head Coil

Head coil consists of upper and lower two parts. Put the lower part on Patient table, the lead of coil facing

coil socket; lay cushion and one-off bed sheet, lay head cushion on the lower part of head coil; patient wear

one-off medical cap .

Warning!

Various gaps on the patient table is prone to clip

objects, such as finger, hair, clothes, etc

Warning!

When positioning operation, please prompt patient

to close eyes and mustn’t stare at laser beam.

Upper Part of Coil

Lower Part of Coil

Warning!

Don't stand in danger zone when patient table is

moving.

Page 52: Openmark 5000 User Manual

Equipment Manual

52

Warning!

Please don’t clip cushion or any part of patient’s

body when inserting the upper part of head coil

into the lower part.

Let the patient wear one-off medical cap and lay on back on patient table, aligning head to center of lower

part of head coil, put head on cushion on lower part of head coil; align the interface of upper and lower parts

of head coil, hold two sides of the upper part of head coil with two hands and press it down tightly.

You can let patient move slightly to position the head on the center of the coil. Insert the plug of head coil

in the socket, and then move the patient table according to requirements in section 3.4 to perform center

locating.

Pull out patient table after scan, the operator holds and presses down the handle behind the coil with thumb

to divide the coil into two parts, and then let patient sit up.

3.5.2 Neck Coil

Neck coil consists of upper and lower two parts. Put the lower part of neck coil on patient table with the lead

of the coil facing coil socket; lay bed cushion, neck cushion and head cushion ; place the neck cushion on

the lower part of coil. Lay one-off bed sheet on bed cushion, and neck cushion sheet and head cushion sheet,

then align patient’s neck to lower part of coil and let patient lie on back, placing neck on the neck cushion on

the lower part of neck coil, and place the head on head cushion. Let the “V” shape opening on the upper part

of neck coil face patient’s head; hold two sides of the upper part of neck coil with two hands, align interface

of upper and lower parts of neck coil and insert the upper part into the lower part tightly, the neck coil will

lock itself after inserting. Insert plug of neck coil into socket and move patient table according to

requirement in section 3.4.to perform center locating.

Upper Part of Coil

Lower Part of Coil

Page 53: Openmark 5000 User Manual

Equipment Manual

53

Warning!

Please don’t clip cushion or sheet or any part of the

patient’s body when you insert the upper part into

lower part of neck coil

Pull out the patient table after scan, the operator should hold and press down handles in the two sides of

neck coil to separate the two parts, take away the upper part of neck coil and then let the patient sit up.

3.5.3 Body Coil

Body coil is used in chest or abdomen scan. We provide various body coils with different sizes according

to scan parts and figures of patients. These body coils have the similar shape as shown in the following fig.

And the ways of operation are same.

Upper Part of Coil

Warning!

You should align the two sides of the neck coil

when inserting (or separate) the upper part into

(from) the lower part and act on strength evenly at

the same time, or else, uneven strength may

damage coil interface.

Page 54: Openmark 5000 User Manual

Equipment Manual

54

Warning!

Please don’t clip patient’s arms when you insert the

upper part of body coil into the lower plate!

Caution!

Please act on strength evenly with two hands at the

two sides of upper part of body coil when you insert

(depart) it into (from) the lower plate. Uneven

strength may damage the interface of the body coil.

Before using, place the lower plate of the coil on the patient table, and then lay well bed cushion and one-off

bed sheet, and then let patient lay on back on patient table, position the scanning part to the center of body

coil, Insert the upper part into the two sockets on the lower plate. Patient can put two arms either in or out of

body coil.

The lead of the body coil faces socket for coil. Insert the plug of body coil and move patient table according

to section 3.4 to perform center locating operation.

Pull out patient table after scan, operator should hold the two sides of upper part of body coil and pull it up

and then let the patient sit up.

3.5.4 Knee Coil

Knee coil is used in scanning single knee joint, elbow joint and other joint. The coil consists of upper and

lowers two parts. Put the lower part on patient table, the lead of the coil facing coil socket; lay bed cushion,

coil cushion, one-off bed sheet and head cushion sheet, and then position the joint part to the center of the

coil; align the pin to the socket on the lower part of the coil, and then insert upper part into lower part; insert

knee coil plug into the socket and then move patient table according to section 3.4.to perform center

locating operation.

Lower Plate of Coil

Upper Part of Coil

Page 55: Openmark 5000 User Manual

Equipment Manual

55

Pull out the patient table after scan,

the operator should hold and press

down handles in the two sides of the

coil to separate the two parts, take

away the upper part of knee coil and

then let the patient sit up.

3.5.5 Spinal Coil

Because of its shape, spinal coil is also called flat coil. Put the spinal coil at the proper position on the

patient table, the lead of the coil facing coil socket. Lay bed cushion, one-off table sheet and head cushion

sheet, and then let patient lay on back on patient table, place head on head cushion, position the scanning

part to the center of the coil, you can let patient move slightly on the patient table to adjust the position;

insert the coil plug into the socket, and move patient table according to section 3.4 to perform center

locating operation.

Warning!

Please don’t clip cushion or sheet or part of patient’s

body when you insert the upper part to lower part of

knee coil!

Warning!

Please align two sides and act on strength evenly

when you insert or (separate) the upper part into

(from) the lower part of knee coil. Uneven strength

may damage interface of the coil.

Lower Part of Coil

Page 56: Openmark 5000 User Manual

Equipment Manual

56

Pull out patient table after scan and then let the patient sit up.

3.5.6 Breast Coil

Put the breast coil at the proper position on the patient table, the lead of the coil facing coil socket; lay bed

cushion, breast coil cushion and one-off table sheet, and then let the patient lie prone, two breasts naturally

fall in two cavities on the breast coil. Insert breast coil plug into the socket, and move the patient table

according to section 3.4 to perform center locating.

Pull out the patient table after scan and then let patient sit up.

3.5.7 Usage of Coil Plug

Every receiving coil has a coil plug, insert it in the coil socket to form a complete receiving loop. The coil

socket and coil plug are shown in the following fig. The coil socket is below the locating display panel on

the lower magnet

Page 57: Openmark 5000 User Manual

Equipment Manual

57

The pins in plugs of each coil are deferent. Please pay attention to the direction of the plug when you insert

the plug to the coil socket. The two columnar pins (or two columnar holds) must align to columnar holes (or

column pins). Please don’t insert incorrectly, or else it may damage the plug or socket, and cause bad

contact and debase the image quality.

Chapter 4 Equipment Maintenance Guide

Basic routine maintenance works on the whole equipment and the main units are described in this chapter.

4.1 Monitoring Equipment Status

Equipment status includes temperature of scan room and magnet, which can indicate the general status of

magnet. It is recommended to monitor the above status every day when starting up.

4.1.1 Monitoring Scan Room Temperature

Scan room is the external

environment of magnet. The

change of temperature of scan

room will finally influence

temperature of magnet and impact the stability of central frequency. Fierce temperature change in scan

room will influence uniformity of the magnet. Further more, the magnet temperature control subsystem

can’t work normally if the temperature of scan room is higher than 28 ℃, so the operator should monitor

temperature of scan room before starting up the equipment. Please check the work status of air conditioner

in air conditioner room if the temperature in scan room is higher than 28℃ or lower than 20℃.

4.1.2Monitoring Magnet Temperature

Warning!

The temperature of scan room above 28℃ will

hazard the stability of the equipment.

Page 58: Openmark 5000 User Manual

Equipment Manual

58

The stability of magnet

temperature is closely related to the

stability of central frequency of the

equipment. This equipment is

equipped with magnet temperature

control subsystem, the control unit is in

Auxiliary Power Supply (APS),

and the value of temperature is

displayed by digitron in the front panel of APS. There is a switch button in the front panel of APS: When

button TP pressed down, the digitron displays average temperature of magnet. In normal condition, the

average temperature is in a range of “constant ±0.1”. When the display of temperature changes fiercely, the

magnet temperature control or display units will be in trouble, you should contact us for technical support in

time.

4.2 Maintaining Equipment

The maintenance of this system should be jointly done by manufacturer and user. In this section, the items,

precautions and period of maintenance undertook by user are regulated. Other maintenance done by the

manufacturer, including the maintenance on electronic units such as auxiliary power supply, spectrometer,

computer and electric voltage regulator and readjustment and calibration of basic system parameters, are

carried out once a year.

The maintenance persons of user must be trained and authorized by manufacturer or others under the guide

of these persons. The maintenance persons of user must abide by the requirements of safety precautions in

chapter 1 and requirements in this section in the process of maintenance. The user should be responsible for

the equipment damage or personal injury deriving from user’s misoperations against relative regulations.

Manufacturer can provide necessary material and information to qualified maintenance persons in repairing

the repairable devices or units appointed by manufacturer.

Maintenance persons should power off the system anytime when maintain the units and devices of this

equipment except temperature control system Operators are not allowed to enter scan room wearing

Warning!

If the change of average temperature (TP )is larger

than 0.4℃ , it may hazard the stability of the

equipment, and you should contact the MRI service

department of our company.

Page 59: Openmark 5000 User Manual

Equipment Manual

59

high-heel shoes, and are forbidden to take water in unsealed containers into scan room, machine room, air

conditioner room and operating room.

4.2.1 Maintaining the Shell of Magnet

Shell of magnet is the surface that patients and operators touch frequently, so it should be cleaned and

disinfected frequently too. Maintenance persons should wipe the shell at least one time a week with

absorbent cotton soaked in alcohol. Maintenance persons should read carefully the safety precautions in the

signs at the door of scan room before entering the scan room.

4.2.2 Maintaining Patient Table

Patient table is mechanical moving unit. The dirt and foreign matters on the guide rails may influence the

stationarity and obstruct patient table from moving, so that the place where patient table passing should

keep clear. Maintenance persons can wipe the shell of magnet cavity with absorbent cotton soaked in

alcohol, wipe guide rails with gauze soaked in little water. You can wipe wheels of patient table with gauze

soaked in alcohol if necessary.

Foot brake of patient table should keep its reliability: when it is in free status, the patient table can’t move;

when it is pressed up by the foot, the

patient table can move back and

forth easily. Otherwise, the foot

brake should be repaired.

The control panel of the patient table

must keep clear and dry, because

there are electrical devices within the

control panel. Please clean it just

by dry gauze.

You mustn’t take water into scan room when maintaining patient table.

Warning!

Any maintenance on units and devices of this

equipment should be carried out under the power

off status. It is forbidden to take water in unsealed

containers into scan room, machine room, air

conditioner room and operating room.

Page 60: Openmark 5000 User Manual

Equipment Manual

60

Patient table maintenance should be carried out at least twice every week.

4.2.3 Maintaining the Cushions

Although bed cushions don’t touch patients directly, they may contaminate dirt and germs after used for a

long-term, so that they need regular scrubbing. You can wipe cushions with gauze soaked in alcohol. The

cushion maintenance should be carried out at least twice every week.

4.2.4 Maintaining Receiving Coils

4.2.4.1 Placing Receiving Coils

You should place a four-floor wooden shelf in the room according to the size of scan room and put coils on

the shelf as much as possible. Please put the frequently-used and seldom-used coils separately according to

routine use status. Please keep clean and tidy in scan room.

Please don’t crash or pull cables and plugs of coils during placing and using them. Please be careful to take

and put coils when moving them, you must not fall them to the ground or crash them with other articles.

4.2.4.2 Maintaining Cables, Plugs and Sockets of Coils

The cables, plugs of coils and coil sockets on the magnet shell are key components of the circuit. The bad

contact or damage of cable may cause serious consequences such as SNR decreasing, tuning failure or no

signal, so they should be protected specially. Please be careful when you insert or pull out plugs: keep

horizontally inserting or pulling out, and avoid unevenly stress acting on plugs and sockets. The working

coil cable should be minded specially too: for fear of harmful effect, please don’t act stress on or stamp it.

You should avoid crashing or stamping coil cables or plugs in routine operation.

You shouldn’t crash, stamp or pull cables and plugs of coils during placing and using coils. Please take and

put coils carefully and gently when moving them, you must not fall them to the ground or crash them with

other articles.

4.2.4.3 Maintaining the Body of Receiving Coils

Although the body of receiving coils don’t touch patient directly, after using for a long-term, they may

contaminate dirt, so they need regular scrubbing. You should wipe them with gauze soaked in alcohol.

Maintenance of body of receiving coils should be carried out at least one time every week.

4.2.4.4 Maintaining the Interface of Receiving Coils

Some receiving coils are split type. You should join the upper part and lower part together through coil

interface when using. The compactness of interface directly influence the performance of the coil, the bad

connection will induce SNR declines.

Besides acting on two sides with even strength when you are inserting (pulling out) coils, you should clean

the pins on interface of coils with gauze soaked in alcohol: gently wipe the surface of copper pins on coil

Page 61: Openmark 5000 User Manual

Equipment Manual

61

interface to remove the copper rust to get the best receiving effect. Cleanout of copper pins on coil interface

should be carried out at least one time every season.

4.2.5 Maintaining Phantoms

You should put the phantoms on the wooden shelf as much as possible that is used to place coils. Please

place the frequently-used coils and seldom- used coils separately according to routine use status. Please

keep clear and tidy in scan room.

Please be careful to take and put phantoms when using or placing samples, you mustn’t fall them to the

ground or crash them with other articles.

4.2.6 Maintaining Console

4.2.6.1 Maintaining Keyboard

The keyboard used in this system is not of waterproof type, please don’t wipe with water. You can wipe

keyboard with gauze soaked in little alcohol under power off status.

4.2.6.2 Maintaining Mouse

The mouse used in this system is of mechanical type. After used a period of time, it may contaminate dirt on

the roll ball and may be not sensitive. You should clean up the mouse with gauze soaked in little alcohol

under power off status: according to the indication in mouse body, open the mouse, and take out roll ball,

use gauze soaked in little alcohol to wipe roll ball and roll axis gently, finally install the mouse well.

4.2.6.3 Maintaining Display

The display used in this system can’t be washed by water. You should wipe display under power off status:

gently wipe the display screen with lens tissue, wipe the body of display with gauze soaked in alcohol.

4.2.6.4 Maintaining Body of the Control Desk

You can wipe the body with gauze soaked water under power off status.

The above maintenance on control desk should be carried out at least one time a week.

Page 62: Openmark 5000 User Manual

Equipment Manual

62

Chapter 5 Troubleshooting

The potential trouble in routine operations and corresponding resolutions are described in this chapter.

Trouble Possible Cause Resolvent Remark

Neither “OFF”/

“STOP” nor “ON” /

“START” indicator is

on when the triple

switch on the back of

electric voltage

regulator turns on.

Power cut in external

transmission system or

trouble in power supply

Check power supply

circuit and contact local

power supply.

Units in cabinet have

no electric power

1. “ON” indicator in the

front panel is off.

2. The five switches on front

panel are not turned on.

3. The power source plugs of

the units are dropped out.

4. The switches of these

units are turned off.

Press down “ON”

indicator.

Turn on five switches in

order.

Insert plugs tightly

Turn them on.

Power on is controlled

by the five switches in

the front panel of

Electric voltage

regulator.

In normal condition,

the power source

switches of units

should be on “ON”

position.

The digitron

displaying position of

patient table is not on.

The cable supplying power

from tuning box to patient

table has dropped out.

Screw tightly again.

Figure of laser

locating is not

accurate.

There are dirt and waste on

guides rails;

Bad connection between

sensing truckles and guide

rails;

Remove foreign matters;

Inform our company;

Patient table makes

abnormal sound in

operation

There are foreign matters on

the moving direction, which

have increased the resistance

of Patient table in moving.

Remove foreign matters

Inform our company;

Page 63: Openmark 5000 User Manual

Equipment Manual

63

The wheel of patient table

may be failed

It’s hard to push or

move patient table

Ditto;

Two hands act on strength

unevenly.

Foot brake is failed;

The weight of patient lays

unevenly on the table

Ditto;

Two hands act on

strength evenly.

Inform our company;

patient lays in the center

of the table

The movement range

exceeds limit in

right-left direction

Spacer pin is failed. Inform our company

The system can’t start

to scan

Tuning Box is not powered

on;

Power on Tuning Box

No signal 1. Plug of the receiving

coil is not connected

well;

2. RF amplifier is not

powered;

3. RF amplifier is

overloaded;

Connect cable well;

Power on RF amplifier.

Reset RF amplifier

The first cause is the

most common.

The signal of

pre-scan is too weak.

1. Cable of coil is not

connected well;

2. RF amplifier is

overloaded;

Connect cable well;

Reset RF amplifier

The first cause is the

most common.

SNR is bad. 1. The contact points of

receiving coil are not

connected well;

2. The receiving coil is not

tuned well;

Clean out contract

points;

Retune the receiving

coil;

Image is saturated. Gain adjustment is not

normal.

Rescan; If the images remain

abnormal after rescan,

Page 64: Openmark 5000 User Manual

Equipment Manual

64

which indicates that the

hardware in the system

is failed.

Or software error.

Scanning has

stopped at the last

second for a long

time, and the scan

can’t finish normally.

The cable between

spectrometer and main

computer is not connected

well.

Reconnect cable. Exit

RINMR and main

interface, log off the

current user and then log

on again.

Main window close

down

Software proceeding

conflicts

Wait for 1 minute and

then enter main window

again.

Locating is failed. Software error Exit main window, wait

for 1 minute and then

enter main window

again, rescan.

The system can’t

print films

Network cables connected to

computer or network cards

are failed.

Shut down main

computer and camera,

and then reset them

respectively.

Images have no

encoding

1. Gradient amplifier is

not powered;

2. Gradient amplifier is

overloaded.

Power on gradient

amplifier

Reset gradient amplifier .

Images are deformed. There are ferromagnetic

articles on or in the patient’s

body, such as false teeth,

hair grips, brassiere, buckle,

metal zips, contraceptive

ring, etc.

Take away the

removable

ferromagnetic articles.

Interface shows the

load sequence error

or spectrometer error,

Software is in trouble Exit APEX and RINMR,

and wait for 1 minute,

then start up RINMR and

Page 65: Openmark 5000 User Manual

Equipment Manual

65

Caution!

If the trouble mentioned above can’t be

resolved or other trouble occurs, please

inform service department of our company or

our agency in your region. Only service

engineers of our company or trained

maintenance men can repair the equipment

according to relative repair regulation.

Without authorization, operators are

forbidden to disassemble, replace units and

components of the system or carry out

and the system can’t

work normally

enter APEX again.

System halted Software or operating

system is failed.

Exit APEX, RINMR and

all applications, then

press and hold down

[Ctrl] and [Alt] key, and

press [Del] at the same

time, choose [Shutdown]

in prompt menu and

choose [Shutdown] in

shutdown menu.

After the power source

of computer turns off

automatically, wait for 1

minute, and then start up

the computer again.

You must never turn

off computer power

source optionally, or

else, it may damage file

system.

Page 66: Openmark 5000 User Manual

Equipment Manual

66

Chapter 6 Guidance and Manufacturer’s Declaration

– IEC 60601-1-2

Guidance and manufacturer’s declaration –electromagnetic emissions

The ASM-050P is intended for use in the electromagnetic environment specified below. The customer or the user of the ASM-050P should assure that it is used in such an environment. Emission Test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11

Group 2 The ASM-050P shall emit RF energy to fulfill its intended functions. Its RF emissions are likely to cause interference in nearby electronic equipment.

RF emissions CISPR 11

Class A The ASM-050P must be used only in a shielded location with a minimum RF shielding effectiveness and, for each cable that exits the shielded location, a minimum RF filter attenuation of 80 dB from 10MHz to 20MHz, 100 dB from 20MHz to 80MHz, 80 dB from 80MHz to 100MHz. (minimum 100dB at 20MHz, minimum 80dB at 80MHz.)

The ASM-050P when installed in such a location is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Harmonic Emissions IEC 61000-3-2

Not applicable

Voltage Fluctuation/ Flicker emissions IEC 61000-3-3

Not applicable

NOTE: It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the minimum specification.

Table 6-1 – Electromagnetic Emissions

Warning!

Use of ACCESSORIES, transducers and cables

other than those specified may result in increased

EMISSION and/or decreased IMMUNITY of the

ASM-050P.

Page 67: Openmark 5000 User Manual

Equipment Manual

67

Guidance and manufacturer’s declaration –electromagnetic immunity

The ASM-050P is intended for use in the electromagnetic environment specified below. The customer or the user of the ASM-050P should assure that it is used in such an environment. Immunity Test IEC 60601-1-2

Test Level (kV) Compliance Level (kV)

Electromagnetic Environment-Guidance

Electrostatic Discharge (ESD) IEC 61000-4-2

±2,±4,±6 contact ±2,±4,±8 air

±2,±4,±6 ±2,±4,±8

Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst IEC 61000-4-4

±1 differential mode ±2 common mode

±1 ±2

Mains power quality should be that of a typical commercial and/or hospital environment.

Surge IEC 61000-4-5

±1 differential mode ±2 common mode

±1 ±2

Voltage dips, short interruptions and voltage variations on power supply input lines. IEC 61000-4-11

<5% UT (>95% dip in UT) for 0.5 cycle. 40% UT (60% dip in UT) for 5 cycles. 70% UT (30% dip in UT) for 25 cycles. <5% UT (>95% dip in UT) for 5 seconds.

<5% UT 40% UT 70% UT <5% UT

Mains power quality should be that of typical commercial and/or hospital environment. If the user of the ASM-050P requires continued operation during power mains interruptions, it is recommended that the ASM-050P be powered from an uninterruptible power supply or a battery.

Power Frequency (50/60Hz) Magnetic field IEC61000-4-8

3A/m Not applicable Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

Table 6-2 – Electromagnetic Immunity ( ESD, EFT, Surge, Dips and Magnetic Field)

Page 68: Openmark 5000 User Manual

Equipment Manual

68

Guidance and manufacturer’s declaration –electromagnetic immunity

The ASM-050P is intended for use in the electromagnetic environment specified below. The customer or the user of the ASM-050P should assure that it is used in such an environment. Immunity test IEC 60601-1-2 test

level Compliance level

Electromagnetic environment-guidance

Conducted RF IEC 61000-4-6

Radiated RF IEC 61000-4-3

3Vrms 150 kHz to 80 MHz

3V/m 80 MHz to 2,5 GHz

3Vrms

3V/m

Portable and mobile RF communications equipment should be used no closer to any part of the ASM-050P, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

d=1,17 x √P

d=1,17 x √P 80 MHz to 800MHz

d=2,33 x √P 800MHz to 2,5 GHz

Where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). b

Field strengths from fixed RF

transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b

Interference may occur in the vicinity

of equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ASM-050P is used exceeds the applicable RF compliance level above, the ASM-050P should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the ASM-050P.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Page 69: Openmark 5000 User Manual

Equipment Manual

69

Table 6-3 – Electromagnetic Immunity (RF Radiated and conducted)

Recommended separation distances between portable and mobile RF communications equipment and the ASM-050P

The ASM-050P is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ASM-050P can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ASM-050P as recommended below, according to the maximum output power of the communication equipment.

Rated maximum

output power of

transmitter w

Separation distance according to frequency of transmitter m

150 kHz to 80 MHz

d = 1,17 x √P

80 MHz to 800 MHz

d = 1,17 x √P

800 MHz to 2,5 GHz

d = 2,33 x √P

0,01 0,117 0,117 0,233 0,1 0,37 0,37 0,74 1 1,17 1,17 2,33 10 3,7 3,7 7,4 100 11,7 11,7 23,3

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer.

NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Table 6-4 - Recommended separation distances between portable and mobile RF communications equipment and the ASM-050P

Page 70: Openmark 5000 User Manual

Equipment Manual

70

Shenzhen Anke High-Tech Co., Ltd

Part Name: Serial Number: Power Supply Frequency: Power Voltage: Input Power:

Appendix A:

Guide for Temperature Control System

Manufacturer: Full Name of Manufacturer: Shenzhen Anke High-Tech Co., Ltd.

Abbreviation of Manufacturer: Anke

Address of Manufacturer: 26 Yanshan Road, Shekou, Shenzhen, Guangdong 518067, P.R.China

Post Code: 518067

Tel: (86) (755) 2668 8889

Fax: (86) (755) 2669 5307 Website: www.anketech.com

Product Safety Classification: (According to IEC60601-1)

Class I, Type B, Ordinary equipment

Disinfection: See section 1.4.1

It is forbidden to operate this system in the environment with inflammable anesthetic gas, oxygen and

nitrous oxide.

This system works continuously.

Important Declaration of Anke

The content in this guide is subject to change without notification.

The copyright is reserved by Anke. Any reproduction in any form without permission is forbidden.

User of the system please translates the warning signs and forbidden signs in local official language and

makes them into warning boards, and hangs them on the entrance of the scan room. Interpretation right of this manual is reserved by Anke.

External Sign

This system is a device whose power supply is mains power, the label above record its rated voltage, power

supply frequency, input power, etc..

Page 71: Openmark 5000 User Manual

Equipment Manual

71

1. General MRI magnet temperature control subsystem is used to control temperature of MRI magnet, to stabilize the

average temperature of magnet on a given level so as to stabilize the central frequency of MRI system.

Temperature control system (TCS) consists of temperature control and lighting two parts.

2. Panel Instruction 2.1 Front Panel On the top of front panel, there is a temperature display window, which is used to display the

temperature of magnet channel. There is also a LED which lights if the temperature exceeds

In the middle part of front panel, there is a set of buttons, which is used to select the channel of

temperatures to be displayed: select TA to display temperature of the top part of magnet; select TB to

display temperature of the upper pole plate; select TC to display temperature of the lower pole plate;

select TD to display temperature of the bottom part of magnet; select Tave to display average

temperature of the above four channels. It is recommended to select [Tave] in common condition. [Reset]

button is used to release the status in which all buttons are pressed down.

There are three LED indicators to display heating status: [Upper] indicator, [Lower] indicator and

[Working] indicator. [Upper] indicator and [Lower] indicator indicate the heating status of upper and

lower pole plates respectively: if the indicator is lighting, it indicates that the corresponding pole plate is

now heated. [Working] indicator indicates the working status of temperature control subsystem: if the

indicator is lighting, it indicates that the subsystem is heated normally; if the indicator is off, it indicates

that the subsystem is in abnormal status.

[Power] switch is used to tune on or off the power of APS.

There is a small cover on the front panel, which is used by manufacturer to adjust some parameters.

2.2 Back Panel Ra, Rb, Rc and Rd are the input ports of temperature sensing signals of corresponding channels from

magnet; Rp is input port of sensing signals from magnet temperature protection.

[Blanking] is the input port of control signal to enable or disenable the work of temperature control part.

[Lighting] is the output port of lighting power (+36VDC).

[Heater1、Heater2] are output ports of heating power (+48VDC) for upper and lower magnetic poles.

[Heating Switch] is used to enable or disenable the work of temperature control unit. It is always in

“ON” status in common condition to keep the control on magnet temperature.

[AC220V] is the input of main power supply of APS.

3. System principle Temperature control system (TCS) consists of temperature control board (TCB), drive board, RSP1000-48

switch power supply, mains transformer, mains filter, etc..

The tested points temperature would be tested and adjusted after through operational amplifier which signal

imported sensor RA、RB、RC、RD and RP form the socket on control board. And then send to display drive

Page 72: Openmark 5000 User Manual

Equipment Manual

72

through SW switch.

We can get upside temperature and bottom temperature of magnet from amplifying the upside and bottom

sensor signal of magnet after weighting average. Then DIFF the temperature voltage and reference voltage to

get the error signal which can drive heating control signal of upside and bottom magnet.

The average temperature of whole magnet can be selected and sent to display.

If the VMOS can generate interference during switching, M1, M2 can be force to cut off by BLANKING

signal while scanning. AC input power supply of switch power supply is controlled by relay.

Normally, protect signal RP is transmit to comparator, output a low level signal from STOP, that NORMAL

indicator is lighted on indicate temperature in normal condition, and power on relay to connect to switch

power supply. When it is over heat, there is a high level signal from STOP, buzzing alarm and over heat

indicator is on. Switch power supply is cut off. When the temperature is so low, there is also buzzing alarm,

but switch power supply is still connected to heat. Low temperature alarm can be cut off by short post JS.

Signal flowing fig. of this system is shown as below:

Rt

Connector

FUSE F1A/250V

Q4

FUSE F1A/250V

N

L

RL2

43

12

FUSE F1A/250V

VCC

Q3

DC48V

Optocoupler

FUSE

SW

FILTER

EMI

PE

TTRNSFMR

1

3

4

2 56

7

CONTROL

BOARD

SMPS

HEATING BOARD

DL-2D1

USL 125mA/AC250V

R19

SP-500-48FUSE F3A/250V

RVV 3X1mm**2

G2R-1-12VDC

R3-11

4. Specifications 4.1 Environment

Shielding room temperature: 16~26℃, less than 2℃/h Change Rate

Input voltage: 220ACV

Power supply Frequency: 50MHz

Page 73: Openmark 5000 User Manual

Equipment Manual

73

Input power: 1500VA

4.2 Heating Power: <500W

4.3 Stability of Magnet Average Temperature: <±0.1℃

4.4 Stability of Magnet Center Field:

±10PPM/10MIN for short term

±50PPM/WEEK for long term

4.5 Temperature Display

Four channels sensor temperature and magnet average temperature are available to choose.

Temperature available display range: 0~40℃

Temperature precision: ±0.2℃

4.6 Fuse

Fuse of USL125,125mA/250V is used in this system.

5. Operating Instruction 5.1 Lighting Part:

Lighting subsystem uses DC power supply (+36Vswitch power). The ON or OFF of DC power supply are

controlled by [Light Power] on the front panel.

5.2 Temperature Control Part:

Normal Working Condition

After commissioning of this system, the temperature control subsystem should be in continuous working

status, you must never turn off the temperature control power supply. Generally speaking, you should

select [Tave], or itinerantly display temperatures of every channel.

Abnormal Working Condition

Lower Limit Temperature Alarm: The APS will make alarm sound if the average temperature is

lower than 25℃, but it can continue heat norm ally. If the subsystem is re-started up (for example because

of power cut), the alarm will be released when the temperature rises up to above 25℃.

Upper Limit Temperature Alarm: the APS will make alarm sound if the average temperature is more than

upper limit (42℃ generally). The [Working] indicator is off, the subsystem stops heating. The temperature

exceeding indicator on the front panel is on, which indicates that the subsystem is in critical failure status.

You should immediately turn off the heating power supply in the back panel

6. Maintenance All user maintenance must be done by appropriately trained personnel. Individual components of MRI

system contain no user-serviceable parts. Maintenance by untrained personnel may present an electric shock

hazard. See chapter 4 of equipment maintenance guide for further information.

7. Electromagnetic interference

Page 74: Openmark 5000 User Manual

Equipment Manual

74

Using this system may disturb electronic equipment nearby, or maybe the running of electronic equipment

nearby can disturb this system’s running. Keep a appropriate distance from other electronic equipment

while running this system, in case there is still disturbance, please contact the manufactory.

8. Environmental protection This system is electrical product that would bring harm to human and environment if it is disposed of as

domestic garbage at the end of usage period. It must be delivered to recycling center which is appointed to

handle these products. Please contact local government, domestic garbage disposal, its agent or manufactory

for more information of garbage disposal, recovery, recycling station for this kind of electrical product.

Appendix B:

Page 75: Openmark 5000 User Manual

Equipment Manual

75

Guide for Ankle Joint Supporter

1. Structure of Ankle Joint Supporter

Fig.1 Structure of Ankle Joint Supporter As shown in fig.1, ankle joint supporter consists of foot pedal, knob, angle mark, etc. It is usually matched

with ankle coil and locating plate of supporter.

The black points of angle mark the division value, which is 1º. That is to say, turn the knob 1 black point in

clockwise (counter-clockwise), the foot pedal will turn 1º in the same direction. In normal scanning, it is

recommended to get one image at every 6º or 5º.

2. Scanning with Ankle joint supporter

Fig.2. Locating Ankle Joint Supporter

The steps of performing ankle joint motion analyzing scan:

1. Place the locating plate of ankle joint supporter on the end of patient table;

Foot Pedal

Knob

Angle Mark

Ankle Coil

Locating Plate for Ankle Joint Supporter

Locating Zone for Supporter Locating Zone for Supporter

Rotation Axis

Capped Edge of Foot Pedal

Page 76: Openmark 5000 User Manual

Equipment Manual

76

2. Place the ankle joint supporter on the proper position according the difference of scanning right or

left ankle joint;

3. Place the ankle coil on the soleplate of the supporter, the end face tightly sticking on the rotation

axis;

4. Let the patient lay on the patient table, put the foot to be scanned on the foot pedal of the supporter

and sticking on the pedal tightly;

5. Turn the foot pedal clockwise (anticlockwise) to the limiting position (the limiting position refers

to the position that if you continue to turn the pedal in the same direction, the patient will feel that

his foot can’t stick on the pedal tightly any longer).

6. To scan the first group of images on this limiting position: on APEX interface, tune coil, determine

scan slices, sequence and parameters;

7. Press [Scan] button to scan;

8. When finishing scan, turn the pedal a certain angle (6ºor 5ºis recommended) in the reverse

direction, then scan the next group images;

9. Repeat step 7 -8 to get a series of images, till the pedal reach the other limiting position;

Notice:

1. The patient’s cooperation is extremely important to get satisfying motion analyzing images. You

shall prompt the patient to stick the foot sole on the pedal tightly and keep still when scanning;

2. 2. The moving toward the right and left direction can seriously influences the quality of images

and continuity of motion playback. It recommended putting the patient’s foot against the same

capped edge of foot pedal in the whole process of scan