OOOOWONWONWONWON® User manual for PDS series

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02 ~ 1 ~

User Manual

Portable LCD Digital Storage Oscilloscope

PDS5022S, PDS6042S, PDS6062S, PDS6062T,

PDS7062T, PDS7102T

Page 2: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

OOOOWONWONWONWON® User manual for PDS series

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02 ~ 2 ~

__________________________________________________________

Information contained in this document is subject to change without notice.

© 2011 OWON Technology Ltd. All rights reserved.

Reproduction in any manner is prohibited without the written permission of OWON Technology Ltd.

The Lilliput’s products are under the protection of the patent rights in the U.S.A and other countries,

including those which are applying for.

OWON is the registered trademark of Fujian Lilliput.

Fujian Lilliput Optoelectronics Technology Co. Ltd

OWON Technology Ltd is a company of Lilliput Group.

September 2011 Ver.5.0.2 Document reference: OWON/UK/DOC1/V-01012011

Page 3: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 1

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Warranty

Lilliput warrants this instrument to be free of defects in parts and workmanship for 3

years from date of shipment (a 12 month limited warranty applies to Probes, rechargeable

battery, USB cable and other accessories). The warranty only applies to the original

purchaser and is not transferable to any other third party. If it should become necessary to

return the instrument for service during or beyond the warranty period, Please contact the

UK Customer Service Department at +44 8450 508168 or email to: [email protected]

with your copy of purchase proof. A Return Material Authorization (RMA) must be

issued prior to the return of any product, and RMA number should be visible on the

package of the return and should be quoted all time for any related enquiries. It’s the

sender’s responsibility for Shipping costs, freight, insurance and proper packaging should

be used to prevent damage in transit.

No components of our products are user-serviceable. DO NOT open or remove covers.

Repairs and service may only carry out by our Service Centre or authorized service

engineer. Failure to do so shall void any warranty, stated or implied.

This warranty does not apply to defects resulting from action of the user such as misuse,

improper wiring, any operations outside of its specification, improper maintenance or

repair, or unauthorized modification. We specifically disclaim any implied warranties for

a specific purpose and will not be liable for any direct, indirect, incidental or

consequential damages. Our total liability is limited to repair or replacement of the

product.

Product specifications are subject to change without prior notice.

Calibration and Repair Service

OWON Tec offers repair and calibration services for products we supply.

A UKAS certification is also available through our authorized service provider at

additional cost should you require. Please contact Customer Support & Service

Department for further information, OWON recommends that all Instruments should be

calibrated annually to maintain its performance and accuracy.

Customer Support & Service

OWON Technology Ltd Tel: +44 8450 508168 Email: [email protected]

For latest version of this User’s Manual, Software updates, and other information please visit our

website at: www.owon.co.uk

Page 4: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 2

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Table of Contents

1. Safety Information

1.1 Safety Terms and Symbols……………………………………………………………………4

1.2 Safety information……………………………………………………………………………5

2. Your PDS Oscilloscope

2.1 Overview………………………………………………………………………………………7

3. Use Your Digital Oscilloscope………………………………………………………………….7

3.1 What are in the standard package?..........................................................................................................8

3.2 Front panel and User interface.................................................................................................................9

3.3 Check Your Oscilloscope before first use………………………………………………………………12

3.4 Adjust Your Probe ……………………………………………………………………………………...13

3.4 Set your Probe Attenuation Coefficient………………………………………………………………..14

3.5 Use Your Probe Safely…………………………………………………………………………………15

3.6 Do Self-calibration……………………………………………………………………………………..15

3.7 The Vertical controls…………………………………………………………………………………..16

3.8 The Horizontal controls…………………………………………………………………………….....17

3.9 The Trigger controls…………………………………………………………………………………..18

4.Use more of Your Digital Oscilloscope…………………………………………………………….19

4.1 Use Vertical controls…………………………………………………………………………………..20

4.1.1 CH1/CH2 MENU…………………………………………………………………………………...20

4.1.2 Set Coupling for selected Channel ………………………………………………………………....21

4.1.3 Use Inverted Function …………………………………………………………………………..….22

4.1.4 Use Math Functions ……………………………………………………………………………..…23

4.1.5 Use FFT……………………………………………………………………………………………..25

4.1.6 Use VERTICAL POSITION and VOLTS/DIV Knobs …………………………………………….28

4.2 Use Horizontal controls……………………………..…………………………………………..…….28

4.2.1 Main Time Base ……………………………………………………………………………..……...28

4.2.2 Set up Window ……………………………………………………………………………….……..29

4.2.3 ZOOM IN ………………………………………………………………………………….……….29

4.3 Use Trigger Controls………………………………………………………………………30

4.3.1 Trigger Menu for PDS5022S、、、、PDS6042S、、、、PDS6062S、、、、PDS6062T …………………………30

4.3.2 Trigger Menu for PDS7062T、、、、PDS7102T ………………………………………………………35

Page 5: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 3

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.3.3 Set up a Video Trigger under Single Trigger Mode ………………………………………………38

4.3.4 Alternate trigger ……………………………………………………………………………….......39

4.4 Use MENUS…………………………………………………………………………………………41

4.4.1 Use ACQUIRE …………………………………………………………………………………….41

4.4.2 Use DISPLAY……………………………………………………………………………………....42

4.4.3 Display Format …………………………………………………………………………………….44

4.4.4 Use SAVE/RCL …………………………………………………………………………………… 45

4.4.5 Use UNTILITY …………………………………………………………………………………… 46

4.4.6 Carry out a Self-Calibration ………………………………………………………………………47

4.4.7 Use Automatic Measurement ……………………………………………………………………47

4.4.8 Use CURSOR ………………………………………………………………………………………50

4.4.9 Use Autoscale ………………………………………………………………………………………52

4.5 Use Executive Controls………………………………………………………………………………………….53

4.5.1 AUTOSET

4.5.2 STOP/RUN

5. Quick Start ................................................................................................................................ 55

5.1 Example 1: Measurin a Simple Signal ................................................................................................... 56

5.2 Example 2: Working out the Gain of the Amplifier in the Metering Circuit………………..………….57

5.3 Example 3: Capturing the Single Pulse .................................................................................................. 58

5.4 Example 4: Noise analysis and reduction ............................................................................................... 60

5.5 Example 5: Examine the Phase shift between two related signals ...................................................... 61

5.6 Example 6: Triggering Video Signal outputted from a TV set ............................................................... 62

6. Troubleshooting .......................................................................................................................... 63

7. Appendix A:Specifications ......................................................................................................... 64

8.Appendix B: Accessories ............................................................................................................. 67

9.Appendix C: Maintenance and Service ...................................................................................... 69

10.Appendix D: Using Rechargeable Battery................................................................................ 70

Page 6: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 4

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

1. Safety Information

Use the following safety guidelines to help ensure your own safety and to help protect your Digital

Oscilloscope and working environment from potential damage.

1.1 Safety Terms and Symbols

Safety Symbols

The following symbol may appear on the products:

Hazardous Refer to Manual Protective Grounding Terminal Measurement

Voltage Earth Terminal of Chassis Ground Terminal

Safety Terms

The following symbols and terms may appear on the products and this manual:

---------------------------------------------------------------------------------------------------------------------------

Warning: warning indicates a potential for personal injury, or death.

---------------------------------------------------------------------------------------------------------------------------

Caution: Caution indicates a potential for product damage or other property damage.

---------------------------------------------------------------------------------------------------------------------------

DANGER indicates an injury hazard may be immediately accessible.

WARNING indicates an injury hazard may be not immediately accessible.

CAUTION indicates a hazard to this instrument or other property.

1.2 Safety Information

NO PARTS OF THIS INSTRUMENT ARE USER SERVICEABLE. PLEASE USE ONLY OWON AUTHORIZED SERVICE

CENTRE AND QUAILFIED SERVICE ENGINEER TO CARRY OUT ANY SERVICE FOR YOUR PRODUCT.

PLEASE FOLLOW THE GUIDELINES AND TAKE SAFETY PRECAUTIONS BEFORE STARTING TO USE THIS

INSTRUMENT AND MAKE SURE YOUR USE OF THIS INSTRUMENT MUST BE WITHIN THE RANGE OF ITS

SPECIFICATION.

Page 7: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 5

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Warning:

To avoid fire or electrical shock, when using battery charge as power supply, and

input signal connected is more than 42V peak (30Vrms) or on circuits more than

4800VA, you should

Only use insulated Probes and test lead.

Check the Probes before use and replace it if there are any damages.

Remove Probes, test leads immediately after use.

Disconnect USB/RS232 cable. Do NOT connect to computer when the instrument

is in use.

Not apply high voltage than its rating, use with caution when setting x1 on the

Probe.

NOT use exposed metal BNC or banana plug connectors

NOT insert metal objects into connectors.

To avoid Fire or Personal Injury

Use the proper power cord. Use only the power cord supplied with the product

and certified to use in your country.

Connect or Disconnect Correctly. Please do not connect or disconnect a test

lead/Probe when they are connected to a voltage source.

Ground This product is grounded through the power cord grounding conductor. To

avoid electric shock, it must be grounded properly before making any connection

with input or output terminal of this instrument.

Connect the Probe properly. The voltage level of the Probe ground is at the same

level with this instrument ground. Do not connect the instrument ground to a high

voltage.

Check All Terminals ratings. To avoid fire or shock hazard, check all ratings and

marking on the product you intend to use with this instrument. Refer to the product

manual for further information of their ratings before connecting to the instrument.

Do NOT use without cover or panel. Do not operate the instrument without cover

or panel.

Use the proper fuse. Use only the specified type and rating fuse for this

instrument.

Avoid exposed circuitry. Do not contact any components or parts when the

instrument is powered on.

Page 8: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 6

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Do NOT operate if in any doubt. If you suspect any damages to this instrument

stop operating immediately and contact qualified service engineer to arrange an

inspection.

Use your Oscilloscope in a well-ventilated area. Make sure the instrument

installed with proper ventilation, refer to the user manual for more details.

Do NOT use Oscilloscope in any wet/damp condition or when you are wet.

Do NOT use Oscilloscope in an explosive atmosphere.

Keep Oscilloscope surface clean and dry.

Page 9: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 7

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

2. Your PDS Oscilloscope

2.1 Overview

Models are covered by this manual:

Bandwidth Sample Rate

PDS5022S 25MHz 100MS/s

PDS6042S 40MHz 250MS/s

PDS6062S 60MHz 250MS/s

PDS6062T 60MHz 250MS/s

PDS7062T 60MHz 500MS/s

PDS7102T 100MHz 500MS/s

Record length of 6K points per channel (2.5M points per channel for PDS8202T)

Support cursor readings

20 automatic measurements

Auto scale

Color LCD Display with high resolution and adjustable contrast

(PDS5022s,PDS6042s)

Store up to 4 waveforms;

Auto-settings for a quick and easy measurement

Waveform Math

Built-in FFT

Support both average and peak values of the waveform;

Real-time Digital Oscilloscope

Multiple triggering, including Edge, video and Alternating triggering

USB interface for PC communications

Support USB direct output (PDS8202T only)

Multiple language support (English, Simplified Chinese. French, Russian and

Spanish are optional )

Page 10: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 8

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

3. Use Your Digital Oscilloscope

3.1 What are in the standard package?

Please check the package after receiving your order.

Report any damage up on arrival. Please contact our reseller if you find any

damage of package and instrument on receiving your order and report it to reseller

as soon as possible for a replacement or refund.

Check standard Accessories. A full list of accessories can be found in

Appendix B. Report to reseller immediately if any of them are missing.

3.2 Front Panel and user interface

Our PDS series scope has a very sample Front Panel and user interface to operate.

See picture below for an overview. It consists of LCD display, Menus, Knobs, Push

buttons, inputs and an output of internal standard test signal (1 KHz, 5V).

Fig.1 Front Panel

FRONT PANEL

Main Menu controls:

SAVE/RCL: save or recall waveform. MEASURE: set up and take measurements

ACQUIRE: set up acquisition mode. UTILITY: SYS status and Reset, etc.

CURSOR: measure CURSOR. DISPLAY: set up display type, etc.

Page 11: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 9

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Math Menu: choose MATH for waveform calculations.

Sub-Menu Keys:

F1 to F5 for selecting options when Main Menu is chosen.

Channels controls:

Select and set up Channels including its Vertical range.

Inputs:

Channel inputs, including CH1, CH2 and External

Position Controls:

POSITION: set up and change horizontally windows or time base cursor position.

HORIZONTAL Menu: set up time base window and display.

Trigger Controls:

Set up trigger levels, source, type etc.

Test signal output:

Standard 1 Khz, 5V square wave for adjusting PROBE or input testing.

Fig.2 Menu controls, Knobs and Inputs on Front panel

Page 12: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 10

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

USER INTERFACE

Fig.3 User interface

1. The state of trigger, including:

Auto: Automatic mode and acquire waveform without triggering.

Trig'd: Trigger detected and acquire waveform.

Ready: Pre-triggered data captured and ready for a trigger.

Scan: capture and display the waveform continuous mode.

Stop: data acquisition stopped.

2. Waveform Display area.

3. The purple T cursor indicates the horizontal position for the trigger.

4. Readings for time deviation between the horizontal trigger point and the centre of

LCD screen. It’s 0 when point is in the centre.

5. Readings for the time deviation between the horizontal trigger point and the

centre of selected window. It’s 0 when in the center of selected Window.

6. The current Main Menu selected.

Page 13: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 11

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

7. Options for sub-menu under selected Main Menu.

8. The purple cursor V: current trigger level if it has been set up.

9. Readings for current trigger level.

10. Current Trigger Source.

11. Current trigger type:

Rising edge triggering Falling edge triggering

Video line synchronous triggering Video field synchronous triggering

12. Readings for time base value of current WINDOW.

13. Readings for selected main time based value.

14. Size of selected WINDOW.

15. Coupling mode for CH2.

“—” direct current coupling

“~” AC coupling “ ” indicates GND coupling.

16. Readings for current selected vertical scale (V/D) for CH2.

17. Coupling mode for CH1 :

"–" direct current coupling "~" indicates the AC coupling

" ” indicates GND coupling.

18. Readings for current selected vertical scale for CH1.

19. Readings for current base point positions for CH1 or CH2.

20. The yellow cursor for the grounding zero for CH2, No cursor when CH2 is off.

21. The red cursor for grounding zero for CH1, No cursor when CH1 is OFF.

22. Purple dot lines for cursor measurements when “VOLTAGE” type is selected.

Page 14: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 12

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

3.3 Check your Oscilloscope before first use

A quick start up check is required before using the instrument to make sure the

instrument functions normal according to its specification.

Power on the instrument and Press any key to enter operating mode when prompt

“Press any key entering system”. Select Menu UTILITY and then F2 for “Recall

Factory” settings. The default value for attenuation coefficient of Probe is x10.

Fig.4 Start-up check

Choose a Probe supplied and set it Probe switch to x10, connect it to CH1 input.

Connect Probe tip to 1 Khz 5V test signal output and ground clamp shown as

follows:

Fig.5 Test Signal to Probe for CH1

Press AUTO SET on top of Front panel near Main Menus. The 1 KHz and 5V

signal will be displayed on the screen after few seconds, shown as Fig.6

Fig.6 Screen shot after performing AUTO SET

Page 15: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 13

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Repeat above steps for CH2 to make sure both channels are functional.

3.4 Adjust your Probes

The Probe supplied is a pair of Passive Probe with x1 and x10 attenuation and it is

adjustable. It’s important to adjust the compensation whenever you use x10 for testing

and reduce the chance to have a wrong measurement. Here are the procedures to you

should follow for the adjustment of compensation.

• Set the attenuation coefficient of your Probe to x10, Switch on your instrument and go

to CH1 MENU, than F3 to change the Probe to 10X. Connect your Probe properly to

built-in 1kz 5V test signal output marked with PROBE COMP, Press AUTOSET to

continue.

• Compare the waveform displayed on your with these reference and make sure your

waveform is not overcompensated or under compensated.

Fig.7 Reference Waveforms

• Repeat this if needed. Adjust the Probe using supplied screw until you get a square

wave looked like reference waveform 2.

Fig.8 Adjust to compensate the Probe.

Page 16: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 14

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

3.5 Set correct Attenuation Coefficient for Probes

CAUSTION: The default factory attenuation coefficient set on the instrument for

both Channel 1 and Channel 2 are 10X.

The standard Probes supplied support 1X and 10X. Probes support 100X or 1000x are

available to purchase for an extra cost.

Choosing different attenuation coefficients changes the vertical scale factor of the

display. It’s always important to make sure attenuation coefficient on both the

Channel of the scope and the Probe connected to that channel are matched with the

same settings.

Switch on your Scope and Select the CH1 MENU OR CH2 MENU, under Probe

section, Press F3 to choose Attenuation Coefficient you want to use. Once it is done,

the scope would remember the setting until it is changed.

Check your Probe before connecting to Input Channel and make sure it is set to the

same setting of that Channel on the Oscilloscope. Fig 9 shows you where to change

the settings on your Probe.

Fig.9 Attenuation Switch

CAUSTION: The Probe is limited to work with up to 5mhz signal when using

x1, to use the full bandwidth of the scope, You should always set to use x10.

Page 17: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 15

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

3.6 Use Probe safely

There is a safety ring on the Probe to protect you getting any electric shock, as it is shown

on Fig.10. Please take caution use the Probe.

Fig.10 Finger Guard

Warning:

• Always keep your finger behind the safety ring while any tests are in operation to

avoid electric shock.

• Do NOT touch any metal Part of the Probe including the tip during any

operations or when it is connected to any power source.

• Always connect the Ground terminal of the Probe to the earth before taking any

measurements.

3.7 Do Self-calibration

All our Oscilloscopes have a built-in function called “Do Self Calibration”. This is to set the

Oscilloscope to its best possible condition to help users for a quick and accurate measurement.

You can do this at any time if you wish to, please make sure to run this function whenever the change

of ambient temperature is 5 or over, this is to ensure you always get the best possible measurements

from your instrument.

Before carrying out a self –calibration, please disconnect all Probes or inputs from the

Oscilloscope. Press “UTILITY” then button F3 which is next to “Do Self Cal”. You will

see a Pop-up yellow box to reminder you to remove any Probes and other attachments to

the inputs of Oscilloscope, once they are all done, Press F3 again. The Oscilloscope would

start its self-calibration route and the yellow box remains to show you the progress till the

process is done.

Warning: DO NOT TOUCH OR CONNECT ANYTHING TO THE INPUTS WHEN

IT IS DOING SELF-CALIBRATION. IT COULD CAUSE POTENTIAL DAMAGE TO

THE INSTRUMENT AS WELL AS THE OPERATOR.

Page 18: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 16

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

3.8 The Vertical Controls

As it is shown in Fig.11, there are a few knobs above/under CH1/CH2 Menu for changing

the vertical level of the signal, we call them VERTICAL CONTROLS.

Fig.11 Vertical Controls Area

1. By rotating the Position controls on the top. It moves the CH1 or CH2 trace up

and down on the display, which means the baseline of that channel is changing. If

Coupling in DC mode, you can quickly to measure the DC component of the

signal between trace and the ground signal. Under the AC mode, the DC

component would be filtered out to show a higher sensitivity signal of AC

component.

2. By rotating the knob “VOLTS/DIV”, the vertical scale Factor for corresponding

channel would change between 5mv to 5v(or 50mv to 50v when using x10 etc).

3. Value displayed on the bottom of the screen for each channel changes respectively

when using the Vertical control.

Page 19: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 17

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

3.9 The Horizontal controls

As it is shown in Fig.12, there are few button and knobs on the right side of the front

panel. We call them the “HORIZONTAL CONTROLS”. This is where you can

change the horizontal position of your signal and the time base settings for the

channel.

Fig.12 Horizontal Control Area

1. By rotating the Knob “SEC/DIV”, it changes the time base settings from 5ns to

5s.For PDS5022s, it steps 1-2.5-5, for other PDS models it steps 1-2-5

accordingly.

2. By rotating the Knob “HORIZONTAL POSITION”, it changes the horizontal

position of the signal as well as the displacement of triggering signal if trigger has

been set.

3. By pushing down the button “HORIZONTAL MENU”, you enter in WINDOW

settings, this allows user to set up a window to include the signal which they are

interested in and zoom it in to look more details.

4. Again use Knob “HORIZONTAL POSITION” to move the position and Knob

“SEC/DIV” for the size of WIDNOW, Press F3 next to “Zone Window” to

expend the window on the screen.

5. There are areas on the top and bottom of the screen where values are displayed.

Values representing horizontal position, main time base and the width of the

window would change when Horizontal controls change.

Page 20: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 18

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

3.10 The Triggering Controls

As it is shown in Fig.13, there are four buttons and one knob make up

“TRIGGERING CONTROLS”.

Fig.13 Trigger Control Area

1. TRIG MENU

Press this button to bring up TRIG MODE. This is where you can change trigger Type,

selecting channel, Slope, Coupling and Sensitivity.

2. LEVEL

This knob is to change the trigger level vertically. By rotating the knob, the level of

the trigger cursor “V” for selected channel would move up and down. Trigger level

displayed on the bottom changes accordingly.

3 “SET TO 50% ”

It’s a shortcut to set the trigger level half way 50% of its signal.

3. “FORCE TRIG”

Press this button to force a trigger signal, which is mainly applied to the “Normal" and

"Single” trigger mode.

4. “SET TO ZERO”

Press this to reset the horizontal position of the trigger back to the middle line of the

display.

Page 21: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 19

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4. Use more of your Oscilloscope

After finishing Chapter 3, you now have the basic knowledge of your oscilloscope

and accessories. If you have not read Chapter 3 and don’t have experience about our

oscilloscope, we recommend you to go back and spend few minutes for a start.

In this chapter we more focus on using and setting up your scope ready for a test.

Here is the list of topics we are going to cover in the next 30-40 minutes.

Use Vertical controls

Use Horizontal controls

Use Trigger controls

Use MENUS

Use Executive Controls

Page 22: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 20

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.1 Use Vertical Controls

As you know from last chapter, The VERTICAL CONTROLS includes CH1 MENU,

CH2 MENU, MATH MENU, VERTICA POSITION, VOLTS/DIV for each channel.

Now let’s get into the details of how to use those controls.

4.1.1 CH1/CH2 MENU

CH1 and CH2 are independent in terms of menus and controls, but offer same settings and

functions you can choose. We use CH1 as an example. Press CH1 NENU, you have a

drop down submenu displayed on the right of your screen next to Function keys. By

pressing F1, F2,F3,F4,F5 etc, you can change and select the settings as you can see below:

Fig.14 Channel Setting Menu

Table 4-1 explains more details of these functions and settings.

Function Menu Setting Description

Coupling

AC

DC

Ground

Block the DC component from input signal.

Pass both AC and DC components from input signal

Disconnect input signal.

Channel OFF

ON

Hide and turn off the channel.

Display and turn on the channel.

Probe

1X

10X

100X

1000X

Match this to Probe attenuation factor and have a

accurate reading of vertical scale.

Inverted OFF

ON

Displayed original waveform.

Display inverted waveform.

Table 4-1. Functions for Channel settings

Page 23: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 21

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.1.2 Set Coupling for selected Channel

A square waveform is used as an input in this example.

Press CH1 MENU button to show submenu CH1 SETUP.

• Press F1 next to Coupling and select “AC”, now DC component is blocked from

input signal. See Fig. 15 Coupling “AC”

• Press F1 again to set “DC” mode, both AC and DC components get passed.

Fig.15 AC Coupling Fig.16 DC Coupling

Turn Channel “ON/OFF”

• Press CH1 MENU button to show submenu CH1 SETUP .

• Press F2 next to Channel, choose “OFF”, now Channel 1 input is disconnected,

you see Channel 1 waveform disappears from the screen.

• Press F2 again for “ON”, the input comes in and Channel waveform comes back

on the screen.

Adjust Probe Attenuation settings on Oscilloscope

Probe Attenuation settings on your scope should always match what is on the Probe.

Refer to Table 4-2 for corresponding settings. For example if the attenuation

coefficient on the Probe is 1:1, the setting on your Oscilloscope should be set to 1X.

Attenuation Coefficient of the Probe Corresponding Menu Setting

1:1 1X

10:1 10X

100:1 100X

1000:1 1000X

Table 4-2 Probe attenuation coefficient Vs Oscilloscope Probe settings

To make a change on your oscilloscope,

• Press CH1 MENU button to show submenu CH1 SETUP.

• Press F3 next to Probe to change value, set as 1x as it is shown in Fig.17

Page 24: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 22

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Fig.17 Attenuation coefficient of the Probe

4.1.3 Use Inverted Function

By using this function you mirror the waveform against its base level. You get an

upside down waveform displayed below its base level. Here is how to use the

function.

• Press CH1 MENU to show Submenu CH1 SETUP.

• Press F4 next to Inverted to turn the function ON. You now have an inverted

waveform as it is shown in Fig.19.

• Press F4 again to change to OFF. You now have the waveform back to its original

one as it is shown in Fig. 18.

Fig.18 Original Wave Form Fig.19 Inverted Wave Form

Page 25: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 23

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.1.4 Use Math Functions

There are few Math Functions available in our oscilloscope, to help users calculating the

input signals from CH1 and CH2. Here is the list of Math Functions supported by our

Oscilloscope. See Table 4-3.

Setting Description

CH1-CH2 CH1 subtract Channel 2, Green waveform is the result.

CH2-CH1 CH2 subtract Channel 1, Green waveform is the result.

CH1+CH2 Add Channel 2 to Channel, Green waveform is the result.

FFT Fast Fourier transform on selected channel

Table 4-3

To use Math function, first you have to turn on both Channel 1 and Channel 2 and have

input signals coming in from both channels. Oscilloscope would remind you if any of the

channels is turned off, just go back to Channel settings and turn it on before next.

Press MATH MENU to bring up submenu-WAVE MATH as shown in Fig 20.

Press F1 once, it highlights the CH1-CH2 and a green waveform is displayed on the screen

as a result of this. Press F1 again to cancel the Math, green waveform disappears.

Again, F2 is for CH2-CH1, F3 is for CH1+CH2 etc.

Fig.20 Example of Math CH1+CH2

Page 26: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 24

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.1.5 Use FFT

Users find FFT (fast Fourier transform) is very useful function for analyzing the input signal

on Oscilloscope. This section here is to take you into details of how to use it.

FFT function in this instrument transforms 2048 data points of the time-domain signal into

frequency components mathematically which is made of 1024 points ranging from 0Hz to

Nyquist frequency. Press MATH MENU to bring up submenu WAVE MENU. Press F4

next to FFT to take you into FFT menu. Table 4-4 gives you the details of FFT Menu.

Function Menu Options Description

FFT ON

OFF

Turn on

Turn off

Source CH1

CH2

Do FFT on Channel 1

Do FFT on Channel 2

Window Rectangle

Blackman

Hanning

Hamming

Type of window for FFT.

Format dB

V rms

Display in dB

Display in Vrms

Zoom *1

*2

*5

*10

multiple *1

multiple *2

multiple *5

multiple *10

Table 4-4 FFT submenu

• Press F1 once to turn on FFT, Press again to turn it off. Be aware that only to use FFT

when Horizontal Menu is set for main Timebase NOT on Window setting mode. Again,

the green waveform is displayed as a result of FFT.

• Press F2 to switch the FFT between CH1 and CH2.

• Press F3 to choose WINDOW for FFT, there are four options here, including Rectangle、

Hamming、Hanning and Blackman. We will give details of these options later.

• Press F4 to switch between dB and Vrms.

Page 27: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 25

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

• Press F5 to zoom in/out, options including multiplied by *1, *2, *5, *10.

• Adjust the “Horizontal” knob in horizontal control zone to move the waveform and the

shown frequency of M Pos is the exact frequency of the cursor point in the middle of

spectrum.

• Press F1 to turn off FFT, Press MATH MENU to go back to WAVE MATH menu.

Window options in FFT

There are four options for selecting FFT Window. It is about resolution of frequency and

accuracy of magnitude. Depending on the measurements you are looking for and the

characteristic of your source signal, you can choose different options. Table 4-5 gives you

some basic knowledge on which option you want to choose.

Type Characteristics and use Window

Type

Rectangle

Have best resolution for frequencies measurement, but worst for

magnitude.

Recommend to use for:

• Measure transients or bursts where the signal levels before and after

the event are nearly equal.

• Equal-amplitude sine waves with frequencies that are very close

• Random noise within a broadband with a relatively slow varying

spectrum.

Hamming

Better for frequency measurement than Hanning, and good for

magnitude as well. Recommend to use:

• Measure sine, periodic and narrow banded random noise.

• Work well on transients or bursts where the signal levels before and

after the event are significantly different.

Hanning

Better resolution for measuring magnitude, but OK frequency

measurement. Recommend to use for:

• Measure sine, periodic and narrow banded random noise.

• Work on transients or bursts where the signal levels before and after

the event are significantly different.

Blackman

Best for measuring magnitude, but worst for frequency. Recommend to

use for:

• Measure predominantly single frequency waveforms and look for

higher order harmonics.

Here are examples for measuring sine wave with a frequency of 1Khz under the selection of

Page 28: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 26

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

four different window for FFT.

Fig.21 Rectangle window Fig.22 Hamming window

Fig.23 Hanning window Fig.24 Blackman window

Notes for using FFT:

• Use Zoom function for FFT waveform if it is necessary.

• Use dBVrms for details of multiple frequencies measurement

• Use Linear rms to compare frequencies.

• DC or offset affect magnitude of waveform, Use AC mode on source signal instead of DC.

• Set acquisition mode to “AVERAGE” to reduce effect of random noise

What is Nyquist frequency?

The Nyquist frequency is half the sampling frequency of Digital Oscilloscope. It is sometimes known as

the folding frequency of a sampling of DSO. The aliasing can be avoided if the Nyquist frequency is

greater than the maximum frequency of the signal being sampled.

Page 29: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 27

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

CAUSION::::

While FFT mode is turned on, DO NOT USE THESE SETTINGS:

• Horizontal Window setting.

• Change source channel in CH1/CH2 Setup menu.

• Turn on XY Format.

• Trigger control “SET 50%”

• Auto-scale function.

4.1.6 Use VERTICAL POSITION and VOLTS/DIV Knobs

• The VERTICAL POSITION controls are used to adjust the vertical level of

waveforms, including the captured waveforms and calculated waveforms.

• The VOLTS/DIV controls are for changing the resolution of vertical scale, the

sensitivity of the vertical division steps as 1-2-5. Turning Clockwise to increase

vertical sensitivity and anti-clockwise to decrease.

• When using these controls, the changed value and selected channel are displayed on

the bottom left corner of the screen in yellow. Information disappears few seconds

after user releases the control.

Fig.25 Information displayed for Vertical Position

Page 30: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 28

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.2 Use Horizontal Controls

As you’ve known from the last Chapter, HORIZONTAL CONTROLS includes the

HORIZONTAL MENU, HORIZONTAL POSITION and SEC/DIV. Now our focus

is on details of these controls. Press HORIZONTAL MENU to active TIME MODE

shown as below with description of each function:

Fig.26 Time Mode

4.2.1 Main Time Base

Press the F1 next to Main TimeBase to active, Main Time Base. To set up the main

window, you can use HORIZONTAL POSITION for position and SEC/DIV the time,

available from 5ns to 5s as shown in Fig. 27

Fig.27 Main Time Base

Function

Menu

Setting Description

Main Time

Base

Captured waveform horizontal display is

based on time.

Set Window To define a window of your interests for

Zoom out, between two vertical lines.

Zone

Window

Set full screen view for selected window.

Page 31: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 29

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.2.2 Set up Window

Press F2 next to Set Window to active the function. You will see two yellow vertical lines

appear on the center of your screen (you may see one line if they overlap to each other). Use

HORIZONTAL POSITION to move the lines left and right.Turn SEC/DIV anti-clockwise

to increase the distance between two lines while clockwise to decrease. .

Fig.28 Set Window

4.2.3 Zoom in (Zone Window)

Once the area is selected by using Set Window, Press F3 next to Zone Window, this zooms

in to show the selected area in full screen for you to look into more details of your wave

form. Fig. 29 shows the result of selected area as you’ve see on Fig. 28.

Fig.29 Zone Window

Page 32: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 30

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.3 Use Trigger Controls

You are already familiar with trigger controls after reading Chapter 3. These are TRIG

LEVEL, SET TO 50%, FORCE TRIG, SET TO ZERO and TRIG MENU. We now are

going through each of these controls in details and use them for test.

Your oscilloscope can collect data and plot waveform simultaneously. The waveform would

not be stable and meaningful unless a proper trigger is set up. Very often in the tests, users

are interested in events/data after incoming signal has been triggered. Once your

Oscilloscope detects a trigger, it starts to acquire data and plot waveform from this particular

point, it is what we call- trigger point.

4.3.1 Trigger Menu for PDS5022S、、、、PDS6042S、、、、PDS6062S、、、、PDS6062T

There are two types of trigger for models mentioned above-Edge Trigger and Video

Trigger. Each type has its function menu for users to set them up. Press Trigger Menu to

active your trigger settings. Press F1 next to the display to switch trigger from one type to

another.

Use Edge Trigger

Edge trigger is one of the common triggers you would find on Oscilloscope. The Trigger

happens when the input signal going over a threshold of Voltage on the set direction. You

can choose Rising edge or Falling edge for how you want the trigger to start. Here is the

Edge Trigger Menu shown as Fig. 30.

Fig.30 Edge Trigger Menu

Page 33: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 31

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Function

Menu

Setting Description

Source

CH1

CH2

EXT

EXT/5

Channel 1 as the trigger source.

Channel 2 as the trigger source.

Use external trigger source.

1/5 of the External Trigger Source for increasing range of level.

Slope Rising

Falling

Trigger on the rising edge.

Trigger on the falling edge.

Mode

Auto

Normal

Single

Acquire data and display waveform with or without a trigger.

Acquire data and display waveform when triggering.

Acquire date and display waveform when detecting a trigger and stop

sampling.

Coupling

AC

DC

HF Rjc

LF Rjc

Block the direct current component.

Unblock all components.

Block the high-frequency signal and pass through the low-frequency

component.

Block the low-frequency signal pass through the high -frequency component.

Press to go Down

Press to go Back

HoldOff 100ns~10s Set interval by using TRIG LEVEL control, value range from 100ns to 10s

Holdoff

Reset

Reset Holdoff time to default value (100ns).

Table 4-1 Details of Edge trigger

Use Holdoff function

A time interval before it responses to next trigger, after system triggered the sytem

would ignore the further triggers until the time passes. It provides a chance for user to

check the signal in a short period and helps to check some complex signals, such as AM

waveform etc. Press Function next to Holdoff to activate and use TRIG LEVEL

knob to set time for Holdoff.

Page 34: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 32

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Set up a Rising/Falling for edge trigger

• Applying Sine Waveform to Ch1

• Press TRIG MENU to activate your trigger menu.

• Press F1 next to Type and select Edge.

• Press F2 next to Channel and choose CH1 for source.

• Press F3 next to Slope to choose Rising.

• Press F4 next to Mode to choose Auto.

• Press F5 to go down next.

• Press F2 next to Coupling and choose DC.

• You now have waveform displayed as Fig. 31.

• Press F3 next to Holdoff and use Trig Level to set time.

Fig.31 Waveform when Triggered on Rising Edge

• Press F1 next to Arrow sign to return to previous menu.

• Press F3 next to Slope to choose Falling option.

Fig.32 Wave Form when Triggered on Falling Edge

Page 35: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 33

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Use Video Trigger

Video Trigger is to use with Video format signal (PAL ,NTSC and SECAM). It triggers the

time base on every single line, specified line or field of input Video signal.

In the Trigger mode, Press Function key next to Type to select Video to set up Video

Trigger for input video signal. Fig. 33 shows the set-up menu for Video Trigger.

Fig.33 Video Trigger Sub-menu

Function Menu Setting Description

Source

CH1

CH2

EXT

EXT/5

CH1 as the trigger source

CH2 as the trigger source

External channel as the trigger source.

1/5 of the External Trigger Source for increasing range of level.

Sync

Line

Field

Odd field

Even field

Line NUM

Trigger the time base on the line of input video signal.

Trigger the time base on the field of input video signal.

Trigger the odd fields of input video signal.

Trigger the even fields of input video signal.

Trigger the time base on chosen line of input video signal.

Modulation NTSC/PAL/SECAM Format of video signal.

Move to next submenu

Back to previous menu.

HoldOff 100ns~10s Set interval by using TRIG LEVEL control, value range from 100ns to 10s

HoldOff Reset Reset hold time to default value (100ns).

Table 4-2 Details of Video Trigger Menu

Page 36: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 34

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Set up a Video Trigger

We now take you through an example to set up a video trigger for triggering the field.

• Apply NTSC video to Channel 1.

• Press the TRIG MENU to activate TRIG MODE.

• Press F1 next to Type to select Video for the trigger type.

• Press F2 next to CH SEL and select CH1 for trigger source.

• Press F3 next to Sync and select Field. Now you have the result shown as Fig. 34.

Fig.34 Example of Video Trigger (Sync mode:Field)

Page 37: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 35

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.3.2 Trigger Menu for PDS7062T、、、、PDS7102T

Triggers and their submenus for these models are different with others. There are two

trigger modes Alternating and Single. Each mode has Edge and Video trigger respectively.

Single Trigger

This is used to capture steady waveforms in two channels simultaneously. You can use

either Edge Trigger or Video trigger depending on your application.

Edge Trigger

Here is the Edge Trigger Menu in PDS7062T and PDS7102T.

Fig.35 Edge Trigger Menu

Function Menu Setting Description

Source

CH1

CH2

EXT

EXT/5

CH1 as the trigger source

CH2 as the trigger source

External channel as the trigger source.

1/5 of the External Trigger Source for increasing range of level.

Type Edge Set vertical channel trigger type as Edge trigger.

Slope Rising

Falling

Trigger on the rising edge.

Trigger on the falling edge.

Trig mode

Auto

Normal

Single

Acquire data and display waveform with or without a trigger.

Acquire data and display waveform when triggering.

Acquire date and display waveform when detecting a trigger and stop

sampling.

Coupling

AC

DC

HF Rjc

LF Rjc

Block the direct current component.

Unblock all components.

Block the high-frequency signal and pass through the low-frequency

component.

Block the low-frequency signal pass through the high -frequency component.

Sensitivity 0.2div~1.0div Set sensitivity for trigger

Holdoff 100ns~10s Set interval by using TRIG LEVEL control, value range from 100ns to 10s

Holdoff Reset Reset holdoff time to default value (100ns)

Table 4-3 Details for Edge Trigger menu in Single mode

Page 38: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 36

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Set up an Edge Trigger under Single Trigger mode

Now we take you through to set up a Single Trigger for Channel 1. The source signal is

square waveform on Channel, Type-Edge, Slope-Falling, Mode-Auto and DC for Coupling.

1. Press TRIG MENU to activate Trigger menu.

2. Press F1 next to Type to choose Single.

3. Press F2 next to Source to choose CH1.

4. Press F3 next to Type to choose Edge

5. Press F4 next to Slope to Falling.

6. Press F5 for next submenu and F2 next to Trig mode for AUTO.

7. Press F3 next to Coupling to choose DC.

Here is the result shown below as Fig.36.

Fig.36 Wave Form captured on Falling Edge under Single mode

Page 39: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 37

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Video Trigger

Here is the Video Trigger menu under Single Trigger mode.

Fig.37 Video Trigger Menu

Function

Menu Setting Description

Source

CH1

CH2

EXT

EXT/5

CH1 as the trigger source

CH2 as the trigger source

External channel as the trigger source.

1/5 of the External Trigger Source for increasing range of level.

Type Video Set vertical channel trigger type as video trigger.

Video Type

Sync

Line

Field

Odd field

Even field

Designed Line

Trigger the time base on the line of input video signal.

Trigger the time base on the field of input video signal.

Trigger the odd fields of input video signal.

Trigger the even fields of input video signal.

Trigger the time base on chosen line of input video signal.

Modulation NTSC/PAL/SECAM Format of video signal.

Holdoff 100ns~10s Set interval by using TRIG LEVEL control, value range from 100ns to 10s

HoldOff

Reset Reset holdoff time to default (100ns).

Table 4-4 Details for Video Trigger settings under Single Trigger mode

Page 40: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 38

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.3.3 Set up a Video Trigger under Single Trigger Mode

Now we take your through to set up a Video Trigger in Single Trigger mode. The input

Video signals are NTSC format both Channel 1 and Channel 2.

1. Press TRIG MENU to activate the Trigger menu.

2. Press F1 next to Type for Single.

3. Press F2 next to Source for CH2.

4. Press F3 next to Type for Video.

5. Press F4 next to Sync and select Field. See Fig.38 for result.

Fig.38 Video Triggered waveform (Sync:Field)

6. Press the F2 next to source and change to CH1.

7. Press the F4 next to Sync and change to Line. see Fig. 39 for result

Fig.39 Wave Form Triggered in the Video Line

Page 41: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 39

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.3.4 Alternate trigger

Trigger on none-synchronous signals. One of the common uses of this is to capture two

waveforms with different frequency from Channel 1 and Channel 2. It provides opportunity

for user to look at the different signals to help users on their research and development.

Again Edge trigger and Video trigger are available under this mode. When turning on

Alternating Trigger mode, two “V” arrows appears on the screen, one is in red (CH1) and

the other is in Yellow (CH2).Here are its submenus and settings.

Fig40.Edge trigger options Fig41. Video Trigger options

Set up Alternate Trigger

Now we take this chance to use this trigger for Channel 1 and Channel 2. We apply 10Khz

sine wave to Channel and for CH2 we input a NTSC video signal.

Here is how we set them up:

1. Press TRIG MENU to activate TRIG MODE menu.

2. Press F1 next to Type for Alternating trigger.

3. Press F2 next to CH SEL and choose CH1.

4. Press F3 next to Type and set to Edge.

5. Press F4 next to Slope and set to Rising.

6. Press F5 next to “go down” sign and into submenu.

7. Press F2 next to Coupling and set it to DC.

Page 42: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 40

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

8. Press F3 next to Sensitivity and set to 0.5div.

9. Press F4 next to Holdoff and adjust interval by using Trig level Know.

Now you’ve done all Edge trigger settings for Channel 1, we then move on to Channel 2.

10. Press F1 next to “go up” back to top level.

11. Press F2 next to CH SEL and select CH2.

12. Press F3 next to Type and set to Video.

13. Press F4 next to Sync and set to Line.

14. Press F5 and go to submenu.

15. Press F2 next to Modulation and set it to NTSC.

Now look at your display, two different waveforms are shown as Fig. 42.

Fig.42 Edge trigger on CH1 and Video trigger on CH2

Page 43: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 41

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.4 Use MENUS

On the top of control panel, there are few very important controls used for setting up tests,

storing data, performing measurement etc, including ACQUIRE, DISPLAY,SAVE/RCL,

UTILITY,MEASURE, CURSOR, AUTOSCALE, AUTOSET and RUN/STOP.

4.4.1 Use ACQUIRE

Pressing ACQUIRE to activate ACQU MODE shown as Fig.43.

The description of the Menu is shown as follows:

Fig. 43

Function Menu Setting Description

Sample Normal sampling mode

Peak Detect

Use to capture maximal and

minimal samples. Finding

highest and lowest points over

adjacent intervals.

Average

Use to reduce random noises,

four options are available as

followed.

Averages 4, 16, 64, 128

Four options, indicating the

number of averages.

Table 4-5 ACQU MODE

Here is an example for using Peak Detect function, the input is a standard test square

waveform. You can see the “burrs” on the falling edge of signal is detected with noise when

it is turned on.

Fig.44 Peak Detect mode.

Page 44: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 42

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Fig.45 Normal ACQU Mode Fig.46 Average on with reduced noise on waveform

4.4.2 Use DISPLAY

Press DISPLAY once to activate DISP SETUP menu as shown on Fig. 47

Fig.47

Display Menu

Function Menu Setting Description

Type

Vectors

Dots

The adjacent sampling points are joined by vector form.

Only displaying sampling points captured.

Persist

OFF

1sec

2sec

5sec

Infinite

Options for setting the persistence time for each

sampling point.

Format YT

XY

Show waveform as Voltage against time.

Show waveform as CH1(X) against CH2(Y)

Carry Bitmap

Vectors

Transmit data in Bitmap.

Transmit data in Vector.

Battery ON

OFF

Show Battery level

Turn off Battery level

Table 4-6. Details of Display Menu

Page 45: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 43

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Display Type

Press F1 next to Type to change from Vectors to Dots. Fig.48 and Fig.49 show waveform

in different type.

Fig.48 Waveform Displayed in the Vector Fig.49 Waveform Displayed in Dots

Display Persist

When Persist is turned on, the captured sample points remain on the screen till it is turned

off or time off. There are four options for persistence time. When “Infinite” is chosen, the

sample points stay till it’s off.

Fig.50 Example of Persist is on and set as “Infinite”

Page 46: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 44

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.4.3 Display Format

YT is the common format to display data as Voltage against Time on the screen. It applies

to both Channel 1 and Channel 2. When setting to XY format, CH2 (Y) against CH1(X).

and be ware that certain settings would be changed, for example, the oscilloscope is turned

to sample mode with no trigger applied, the data are displayed as bright dots and the by

default the sampling rate is 1MS/s and remain unchanged till user changes the format back

to YT .

Notes: The following functions would be disabled when XY Format is set.

• Reference or Any calculated wave forms.

• Cursor Function

• All Time base controls

• All Trigger controls

Use XY Format

• Apply input signals to both CH1 and CH2

• Press DISPLAY to activate DISP SET menu.

• Press F1 to set Vectors as Type, F2 to set Infinite for Persist and F3 to set XY as Format

as shown in Fig. 51.

Fig.51 Display in XY Format

Page 47: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 45

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.4.4 Use SAVE/RCL

Our PDS series Digital Oscilloscope is capable to save up to four wave forms. Stored

waveforms can be recalled for review later. To active this function, Press the SAVE/RCL to

bring up WAVE SAVE menu. Details of settings and functions are shown as in Fig.52 and

Table 4-7.

Fig. 52

Function

Menu

Setting Description

Source

CH1

CH2

MATH

Source of wave form to be saved.

Available sources: CH1, CH2,

MATH

WAVE A , B

C , D

Location where selected waveform to

be saved or recalled.

Save Action to save selected waveform to

selected location.

CH(X)

X=A,B,C,D

OFF

ON

Turn Display on/off for selected and

stored waveform.

Table 4-7 SAVE/RCL

Save a waveform

A sine wave form inputs to CH1. Follow these steps to save waveform from CH1.

• Under WAVE SAVE menu, Press F1 next to Source to select CH1.

• Press F2 next to WAVE to assign Location A for this waveform.

• Press F3 next to Save to store the waveform captured from CH1.

Recall a stored waveform

Once the waveform is stored you can recall and display it on the screen. The waveform

would remain there until the next one is stored at the same location, hence, the previous

stored waveform would be overwritten by the new one.

To recall a stored waveform, activate WAVE SAVE menu

• Press F2 next to WAVE to select from one of four locations, the CH(X) would show the

location you’ve chosen.

• Press F4 next to CH(X) to change to ON. Now you will see a stored waveform displayed

on the screen with whatever waveforms are already on there.

Page 48: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 46

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

• Example to save/recall a waveform from location A. Fig. 52

Fig.53 WAVE SAVE/RCL

4.4.5 Use UNTILITY

The function provides user an overview of all current system settings. To activate the

function, Press UNTILITY to bring up FUNCTION menu as shown in Table 4-8.

Fig. 54

Function Menu Setting Description

System Status

Horizontal

Vertical

Trigger

Misc

Display all current Horizontal system settings

including Time base, scale and trigger etc.

Current settings for Vertical scale, Coupling, Probe

and Math etc

Current Trigger settings, type, source and mode etc.

Serial No, firmware version.

Recall Factory Reset system settings to factory default.

Do Self Cal Carry out internal Calibration according to

pre-set procedures.

Language

England

French

German

Chinese

Support multi-languages, choose you

preferred system language.

Table 4-8. UTILITY menu and settings

Page 49: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 47

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.4.6 Carry out a Self-Calibration

We have a built-in Self-Calibration procedure, by performing a Self-Calibration would help

to maintain the accuracy of your oscilloscope under ambient temperature. It is

recommended to carry a self-Calibration if the change of the ambient temperature is up to or

exceeds 5°C.

Notes: DISCONNECT ALL INPUTS BEFORE CARRYING OUT A SELF-CALIBRATION. FAIL TO

DO SO MAY CAUSE DAMAGE TO YOUR OSICLLOSCOPE.

Here are the procedures for carrying a self-calibration.

• Disconnect all inputs, including Probes and wires etc.

• Press UTILITY to activate Function menu.

• Press F3 next to Do Self-Cal.

Now there is a yellow window popping up to ask you for a confirmation. Press F3 again, the

Oscilloscope starts the self-Calibration and yellow window remains till the calibration is

done. PLEASE BE WARE THAT NO CONTACTS TO ANY INPUTS CHANNELS TILL

THE CALIBRATION IS DONE.

4.4.7 Use Automatic Measurement

This is where users would come across every time they use our scopes. It is to take the

measurements by using Oscilloscope. Our oscilloscope does this automatic every time user

press MEASURE button. It is capable to work out 20 different types of measurements, and

there are 4 measurements can be displayed on the screen at one time.

The 20 different types of measurements including: Frequency, Cycle, Mean, Peak-to-Peak

value , Root Mean Square, VmaxVmaxVmaxVmax,,,,VminVminVminVmin,,,,VVVVtoptoptoptop,,,, VbaseVbaseVbaseVbase,,,, VampVampVampVamp,,,, OvershootOvershootOvershootOvershoot,,,, PreshootPreshootPreshootPreshoot,,,,

RiseTimeRiseTimeRiseTimeRiseTime,,,, FallFallFallFall TimeTimeTimeTime,,,, +Width+Width+Width+Width,,,, ----WidthWidthWidthWidth,,,, +Duty+Duty+Duty+Duty,,,, ----DutyDutyDutyDuty,,,, DelayADelayADelayADelayA BBBB and and and and DelayADelayADelayADelayA BBBB .

To use MEASURE, simply Press MEASURE button would activate MEASURE menu.

Press F1 next to Source/Type to allow you to switch between Source and Type.

Source is the signal from the channel which user wants to carry out a measurement, and

Type is one of the 20 measurements user wants Oscilloscope to work out. Under the Type as

you can see in Fig. 57 is what the actual measurement reading is. The reading would show

sign of “?” if there isn’t a measurement can be worked out from the source signal. As in Fig.

57, both Channel1 and Channel2 are chosen as source, measurements are taken for

Frequency, Pk-Pk value, Mean and RMS.

Page 50: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 48

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Fig.57 Measure Menu

Abbreviation Descriptions

Vpp Peak-to-Peak Voltage Value.

Vmax The maximum value of amplitude, which is the highest positive peak value of the chosen source

Vmin The minimum value of amplitude, which is the lowest negative peak value of the chosen source.

Vtop Voltage value for the flat top of the chosen waveform, for measuring square/pulse waveform.

Vbase Voltage value for the flat base of the chosen waveform, for measuring square/pulse waveforms.

Vamp Voltage value between Vtop and Vbase.

Overshoot The value of (Vmax-Vtop)/Vamp, for measuring square/pulse waveforms.

Preshoot The value of (Vmin-Vbase)/Vamp, for measuring square/pulse waveforms.

Average The arithmetic mean value of the chosen source.

Vrms The True Root Mean Square voltage value for the chosen source.

Rise Time Time that the leading edge of the first pulse takes to rise from 10% to 90% of its amplitude.

Fall Time Time that the falling edge of the first pulse takes to fall from 90% to 10% of its amplitude.

+Width The width of the first positive pulse at the point of 50% of its amplitude.

-Width The width of the first negative pulse at the point of 50% of its amplitude.

Delay 1→2 The delay between the two channels at the rising edge.

Delay 1→2 The delay between the two channels at the falling edge.

+Duty +Duty Cycle, it’s (+Width/Period).

-Duty: -Duty Cycle, it’s (-Width/Period).

Freq The number of occurrences of a repeating event per unit time

Period The duration of one cycle.

Table 4-9 Automatic measurements

Page 51: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 49

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

We now take you through an example to use Automatic measurement. Before using them,

all sources of channels have to be switched on. These can be done through CH menu as

mentioned before. Please be ware that this function only applies to waveform captured from

input channels at YT format. It would not work on saved waveforms, calculated waveforms

and waveform in XY or SCAN format. In this example, we apply square wave to Channel 1 and

sine wave to Channel2, and we want to measure the frequency, Peak-to Peak value of Channel 1, and

mean value and RMS value for Channel2. Here is how we set them up:

• Press MEASURE to activate MEASURE MENU.

• Press F1 next to Source/Type to highlight Source.

• Press F2 to choose first source as CH1.

• Press F3 to choose second source again as CH1.

• Press F4 to choose third source as CH2.

• Press F5 to choose fourth source as CH2.

• Press F1 again to highlight Type.

• Press F2 to select Freq. A reading of 1.182Khz appears automatically.

• Press F3 to select Pk-Pk, reading appears immediately under highlighted type.

• Press F4 to select Mean for CH2, reading appears immediately.

• Press F5 to select Cyc RMS, reading appears immediately.

Fig.58 Automatic Measurement

Page 52: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 50

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.4.8 Use CURSOR

Press CURSOR button on the front panel to activate cursor function menu-CURS MEAS

on the screen, there are two types of cursors: Voltage and Time.

Fig.59 CURS MEAS Menu

Functions Setting Description

Type

OFF

Voltage

Time

The cursor is off, no measurement taking.

Turn on horizontal cursors and display their voltage value.

Turn on Vertical cursors and display their time value.

Source CH1/

CH2

Source signal for cursor measurement.

Delta Absolute value of the difference between two cursors.

Cursor 1

Reading of Cursor 1.

Time: read time from start position-the horizontal trigger

point (0s). Voltage: read voltage from ground level(0 v)

Cursor 2

Read the position of Cursor 1 (the Time is read with reference

to the horizontal trigger position and the Voltage is to the

ground point).

Table 4-10. Description of Cursor functions

Page 53: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 51

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

You can adjust the positions of Cursors by using POSITION controls, marked with

CURSOR1 and CURSOR2. Now we take you through two examples to set up and use

cursors for Voltage and Time measurements respectively. An input signal of sine waveform

is used throughout this example.

Voltage Cursor measurement

• Press CURSOR to activate CURS MEAS menu.

• Press F1 to select Voltage as Type, there are two purple horizontal lines appeared on the screen, representing

CURSOR 1 (on the top) and CURSOR 2(on the bottom)

• Press F2 to select CH1 as the source. The results are Delta:46.00v, Cursor1:22.40v, Cursor 2: -23.60v. (See Fig. 60)

Fig.60 Voltage Cursor Measurements

Time Cursor Measurements

• Press CURSOR to activate CURS MEAS menu.

• Press F1 to select Time as Type, there are two purple Vertical lines appeared on the screen, representing CURSOR 1

(on the left) and CURSOR 2(on the right) (Fig. 61)

• Press F2 to select CH1 as the source.

• Move CURSOR1 and CURSOR2 and The results change accordingly as in Fig.

(Delta: Freq: 1.163Khz, 0.86ns. Cursor1 0.000ns, Cursor2 0.860ns)

Fig.61 Time Cursor Measurements

Page 54: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 52

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.4.9 Use Autoscale

This is a very useful function for first time users to carry out a simple and quick test on the

input signal. It tracks the incoming signal and set up all settings to measure the signal

automatically. The settings include Trigger mode, Division for Voltage and Time scale

based on the Type of the signal, the Amplitude of the signal and Frequency of the signal.

Function Menu Setting Instruction

Autoscale

OFF

ON

Turn off Autoscale

Turn on Autoscale

Mode

Vertical

Horizontal

HORI—VERT

Only adjust Vertical scale according to input signal.

Only adjust horizontal scale according to input signal.

Adjust both vertical and horizontal scales according to input signal.

Wave

Display Only one or two cycles in waveform.

Display Multi cycles in waveform

Table 4-11 Details of Autoscale settings

The functions can work on both Channel 1 and Channel 2 simultaneously. Here is an

example for how to use the function:

• Press Autoscale to activate its menu.

• Press F1 next to Autoscale to select ON.

• Press F2 next to mode to select Horizontal- Vertical for Mode item.

• Press F3 to choose and Press F3 to change to . See Fig. 62/63 for

details

Fig.62 Multi cycles mode Fig.63 single cycle mode

Examples of Autoscale

Page 55: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 53

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Notes:

Things you should be ware when turning on Autoscale.

• The symbol is flickering on the top left corner of screen every half second.

• The Oscilloscope is setting Triggers automatically for the incoming signal based on its best approach.

Access from front panel to trigger controls is disabled. A warning message will be displayed on the

bottom left corner of the screen when trying to change submenus of Trigger controls.

• If the display mode is in XY and STOP is turned on, pressing AUTO SET to enter into Autoscale. The

Oscilloscope will be set to YT mode and AUTO triggering.

• Oscilloscope is set as DC coupling with AUTO Triggering. A warning message is displayed when trying

to change these settings.

• The oscilloscope will turn off Autoscale if user is trying to adjust vertical position, voltage division,

trigger level or time scale for CH1 or CH2. To back to Autoscale, Press AUTOSET.

• The Oscilloscope is always on main time base.

• The Oscilloscope is turned into Peak Detection mode if it is in Average mode.

Page 56: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 54

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

4.5 Use Executive controls

These controls include AUTOSET, RUN/STOP.

4.5.1 AUTOSET

It’s a very useful and quick way to apply a set of pre-set functions to the incoming signal,

and display the best possible viewing waveform of the signal and also works out some

measurements for user as well. Table 4-12 gives the details of functions applied to the signal

when using AUTOSET.

Function Items Setting

Acquisition Mode Current

Vertical Coupling DC

Vertical Scale Adjust to the proper division.

Bandwidth Full

Horizontal Level Middle

Horizontal Sale Adjust to the proper division

Trigger Type Current

Trigger Source Show the minimum number of channels.

Trigger Coupling Current

Trigger Slope Current

Trigger Level Mid-point Setting

Trigger Mode Auto

Display Format YT

Table 4-12. Functions of AUTOSET

4.5. 2 RUN/STOP: Enable or disable sampling on input signals.

Notes:

When there is no sampling at STOP state, user is still able to change vertical division or

time base for waveform within a certain range. If the time base is =<50ms, the time base can

be adjusted 4 divisions downwards.

Page 57: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 55

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

5. Quick Start

5.1 Example: Measuring a Simple Signal

The purpose of this example is to measure the frequency and Peak-to-Peak value of a given signal.

Step 1

Go to CH1 MENU and Probe menu to set attenuation coefficient as 10X, turn the switch on the

Probe to 10X as well.

Step 2

Connect the Probe through Channel 1 to the signal source.

Step 3

Press AUTOSET button to let Oscilloscope run through a pre-set functions and apply them to

incoming signal.

Step 4

Take the measurements by going through these settings.

• Press MEASURE to activate measurement function menu.

• Press F1 next to Source to highlight Source.

• Press F2, F3, F4 and F5 one by one to set them to CH1.

• Press F1 again and highlight Type.

• Press F2 and set to Freq.

• Press F3 and set to Period.

• Press F4 and set to Mean.

• Press F5 and set to Pk-Pk.

Now you have your measurement results displayed in Fig.64

Fig.64 Measure Frequency and PK-PK value for a given signal.

Page 58: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 56

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

5.2 Example: Working out the Gain of a Amplifier in a Metering Circuit

The purpose of this example is to work out the Gain of an Amplifier in a Metering Circuit. First we

use Oscilloscope to measure the amplitude of input signal and output signal from the circuit, then

to work out the Gain by using given formulas.

Step 1

Go to Both CH1 MENU and CH2 MENU and then their Probe menu to set attenuation coefficient

as 10X, turn the switch on the Probes to 10X as well.

Step 2

Connect the Probe1 through Channel 1 to its signal source-the input signal in the circuit. And

Probe 2 through Channel 2 to its signal source-the output signal in the circuit.

Step 3

Press AUTOSET button to let Oscilloscope run through a pre-set functions and apply them to

both signals coming into Channel 1 and Channel 2.

Step 4

• Press MEASURE to activate measurement function menu.

• Press F1 next to Source to highlight Source.

• Press F2 to set Source as CH1.

• Press F3 to set Source as CH2.

• Press F1 to highlight Type.

• Press F2 to set Pk-Pk.

• Press F3 to set Pk-Pk.

• Get Pk-Pk readings of Channel 1 and Channel 2. (see Fig.57).

Fig. 57 Measuring input and output signal in a metering circut

Step 5

Calculate the gain for amplifier using this formula.

Gain = Output Signal / Input signal (52.00v/5.200v)=10

Gain (db) = 20×log (gain) =20xlog(10)=20 (db)

Page 59: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 57

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

5.3 Example: Capturing a Single pulse

It’s quite easy to use Digital Oscilloscope to capture non-periodic signal, such as a pulse

and burr etc. But the common problem is how to set up a trigger if you have no knowledge

of the signal? With various functions supported by our Oscilloscope, user can solve this

problem by taking an easy approach. First to run your test using auto trigger to find out the

closest trigger level and trigger type, this helps user to make few small adjustments to

achieve a proper trigger level and mode. Here is how we achieve this.

Step 1

Go to Both CH1 MENU and CH2 MENU and then their Probe menu to set attenuation coefficient

as 10X, turn the switch on the Probes to 10X as well.

Step 2

Connect the Probe1 through Channel 1 to its signal source and adjust VOLTS/DIV and SEC/DIV

controls to set up a proper range of Voltage and time base for the signal.

Step 3

Press ACQUIRE to use ACQUIRE MODE, then F2 to select Peak Detect.

Step 4

Press TRIG MENU to activate Trigger MODE, F1 to select Single, F2 to choose CH1

as a source, F3 to select Edge for the type, F4 to set Rising for the slope and finally adjust

the trigger level to the roughly 50% of the incoming signal.

Step 5

Check the Trigger State Indicator on the top of the screen, if it is not Ready, push down

RUN/STOP and start Acquire, wait for trigger to happen, if not adjust the level repeat

Step 5 again until a sampled signal captured and display on the screen.

By using this approach, a random pulse can be easily captured. For instance, if we want to

find a burst burr of high amplitude, set the trigger level to a slightly higher value of the

average signal level, Press RUN/STOP and wait a trigger, once it happens, a triggered

waveform would be captured. User also can change the Horizontal triggering position to

have more details of how and when the burr happens. (Fig.66 gives you more details).

Fig.66 Capturing a Single pulse

Page 60: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 58

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

5.4 Example: Noise analysis and reduction

Noise is very common inside most of the electronic signal. To find out what’s inside the

noise and reduce the level of noise is very important function our oscilloscope is capable

to offer.

Noise Analysis

The level of noise sometime indicates a failure of electronic circuit. The Peak Detect

functions acts an important role to help you to find out the details of these noise. Here is

how we do it:

Step 1

Apply Square waveform to CH1 as source signal, Press ACQUIRE button to activate

ACQU MODE menu.

Step 2

Press F2 to select Peak Detect function. The signal displayed on the screen containing

some noise, by turning on Peak Detect function and changing time base to slow down the

incoming signal, any peaks or burr would be detected by the function. (See Fig.67).

Fig.67 Signal with Noises

Page 61: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 59

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Reduce the effect of noise from your signal

When focusing on signal itself, the important thing is to reduce the noise level as lower as

possible, this would enable user to have more details about the signal. The Average

function offered by our Oscilloscope can help you to achieve this.

Here are the steps for how to enable Average function.

Step 1

Apply square waveform to CH1 as a source signal. Press ACQUIRE button to bring up

ACQU MODE menu.

Step 2

Press F3 select Average, Press F4 to select number of data points for average. Available

choices of number are 16, 32, 64 and 128. The higher the number, the better result user

will get.

User would see a much reduced random noise level and make it easy to see more details of

the signal itself. As it is in Fig. 68, after applying Average, user can easily indentify the

burrs on the rising and falling edges of some part of the signal (see Fig.68).

]

Fig.68 Reduce Noise level by using Average function.

Page 62: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 60

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

5.5 Example: Examine the Phase shift between two related signals

X-Y mode is a very useful when examining the Phase shift of two related signals. This

example takes you step by step to check out the phase change of the signal after it passes a

specified circuit. Input signal to the circuit and output signal from circuit are used as

source signals.

Step 1

Set up attenuation coefficient to 10X through Probe menu for both CH1 and CH2. Switch

Probes to 10X. For source signals, CH1 takes in input signal to the circuit and CH2 takes

in output signal from circuit.

Step 2

Press AUTOSET, Adjust the VOLTS/DIV for both CH1 and CH2 to the same amplitude

level to get an ellipse.

Step 2

Press DISPLAY to activate DISP SET menu, then F3 to select XY mode. Now the

waveform is displayed as Lissajou’s curve, adjust VOLTS/DIV and VERTICAL

POSITION to get the best possible display, work out the phase shift as in Fig. 69

Fig.69 Lissajous Graph

Sine (q) =A/B or C/D

Where q is the angle of phase shift, A, B, C, and D as them in Fig .69.

The angle of phase shift q =± arcsine (A/B) or ± arcsine (C/D).

The signal must be

centered and are kept in

the horizontal direction.

Page 63: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 61

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

5.6 Example: Triggering Video Signal outputted from a TV set

The Oscilloscope supports both Video Field trigger and Video Line trigger. Video signal

from TV is used as a source signal. The function triggers the input video signal to get a

proper signal displayed for further analysis.

Use Video Field Trigger

Here are how to set up this trigger:

• Press TRIG MENU to bring up the TRIG MODE.

• Press F1 to select Single for type.

• Press F2 to select CH1 as a Source.

• Press F3 to select Video for Type.

• Press F4 to select Field for Sync.

• Adjust VOLTS/DIV, VERTICAL POSITION and SEC/DIV to get the waveform

displayed properly (see Fig.70).

Fig.70 Waveform captured by using Video Field Trigger

Page 64: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 62

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Use Video Line Trigger

Here are how to set up this trigger:

• Press TRIG MENU activate TRIG MODE menu.

• Press F1 to select Single for type.

• Press F2 to select CH1 as Source.

• Press F3 to select Video as type.

• Press F4 to select Line for Sync.

• Adjust controls VOLTS/DIV and VERTICAL POSITION and SEC/DIV to get

the waveform displayed properly.

Fig.71 Waveform captured by using Video Line Trigger

Page 65: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 63

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

6. Troubleshooting

• Oscilloscope is powered on but no Display.

Check power socket/power are connected and make sure all switches are on. If the

problem persists, please contact our Customer Support & Service Team.

• Oscilloscope is powered on, the backlight is on, but can’t see anything as screen is too

bright or dark. (Model PDS5022s, PDS6042s and PDS6062s)

Brightness level of display is wrong. Please adjust the brightness control under the display

to the right level.

• Oscilloscope is powered on and application is running, only two flat lines, but no

waveform displayed.

First to check Probes contact to make sure the connections are OK and the signals are

coming through BNC connector. Second to check the source signal is connected properly

to Probes.

• The Voltage reading is 10 times or 1/10 of the actual value.

Look at the attenuation coefficient for the input channel and the attenuation ration of the

probe, to make sure they are match.

• There is a wave form displayed on the screen, but jumping up and down.

Check your trigger settings to make sure you are using the right source and right choice of

settings.

• Waveform stops and no change whatsoever when pressing RUN/STOP.

Check the Polarity settings in TRIG MODE menu and make sure the trigger level is

within the range of your input signal. Or Press AUTOSET for a quick check.

• The displaying of waveform seems getting slow after increasing AVERAGE value in

ACQU MODE or a longer time is set in the DISP MODE.

It’s normal as the Oscilloscope is working hard on many more data points you’ve included.

Page 66: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 64

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

7. Appendix A: Specifications

Unless otherwise specified, the specifications applied are for PDS series only, and Probes

attenuation set as 10X. It is recommend to leave the Oscilloscope running at least

30minutes before carrying out your measurements. Do a “Self-calibration” if change of

ambient temperature exceeds 5.

Characteristic Instruction

Bandwidth

• 25MHz --PDS5022S

• 40MHz --PDS6042S

• 60MHz --PDS6062S, PDS6062T, PDS7062T

• 100MHz --PDS7102T

Channel 2 + 1 (External)

Acquisition::::

Mode Normal, Peak detect, Averaging

Sample rate

(real time)

• 100MS/s -- PDS5022S

• 250MS/s-PDS6042S, PDS6062S, PDS6062T

• 500MS/s -- PDS7062T, PDS7102T

Input::::

Input coupling DC, AC , Ground

Input impedance

1MΩ±2%,in parallel with20pF±5pF

(PDS5022S, PDS6042S,PDS6062S,PDS6062T)

1MΩ±2%,in parallel with 15pF±3pF(PDS7062T,PDS7102T)

Probe attenuation factor 1X,10X,100X,1000X

Max. input voltage 300V (PK-PK) (DC + AC PK-PK)

Horizontal

System::::

Sampling rate 1S/s~500MS/s

Interpolation (sin x)/x

Record length 6K points per channel.

Scanning speed(S/div)

5ns/div~100s/div, step by 1~2.5~5( PDS5022S)

5ns/div~100s/div, step by 1~2~5

(PDS6042S, PDS6062S, PDS6062T, PDS7062T, PDS7102T)

Sampling rate

/ relay time accuracy

±100ppm

Interval(T)accuracy

(DC~100MHz)

Single:±(1 interval time+100ppm×reading+0.6ns)

Average>16:±(1 interval time +100ppm×reading+0.4ns)

Vertical

system:

A/D converter 8 bits resolution (2 Channels simultaneously)

Sensitivity 5mV/div~5V/div(at input)

Page 67: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 65

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Displacement

±10div(5mV/div~5V/div) (PDS5022S)

±50V(500mV~5V),±2V(5mV~200mV)

(PDS6042S, PDS6062S, PDS6062T, PDS7062T, PDS7102T)

Single bandwidth Full bandwidth

Low Frequency ≥5Hz (at input, AC coupling, -3dB)

Rise time

(at input, Typical)

≤14ns----PDS5022S

≤8.75ns---- PDS60462S

≤5.8ns---- PDS6062S, PDS6062T, PDS7062T

≤3.5 ns----PDS7102T

DC accuracy

±5%(PDS5022S, PDS6042S, PDS6062S, PDS6062T)

±3%(PDS7062T, PDS7102T)

DC accuracy (average)

The voltage difference (V) between any two points on the wave

form after averaging the captured wave forms of more than 16:

± (5% Reading + 0.05 divisions) (PDS5022S, PDS6042S, PDS6062S,PDS6062T

±(3% Reading + 0.05 divisions) (PDS7062T, PDS7102T)

Trigger

Type

Edge Rising、Falling

Video

Line synchronization, Field synchronization, Odd field,

Even field, Any line

Edge trigger Sensitivity Adjustable:0.2div~1.0div

Coupling DC, AC, LF Rjc, HF Rjc

Trigger level

range

Internal ±6div from the center of screen

EXT ±600mV

EXT/5 ±3V

Trigger level

Accuracy

(typical) :

Internal ±0.3div

EXT ±(40mV + 6% of Set Value)

EXT/5 ±(200mV +6% of Set Value)

Horizontal Trigger displacement Pre-trigger : 655 div

Post-trigger : 4 div

Trigger Holdoff range 100ns~10s

Page 68: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 66

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

50% level setting (typical) Input signal frequency ≥50Hz

Trigger sensitivity

(Video trigger, typical)

Internal Peak-peak value of 2div

EXT 400mV

EXT/5 2V

Signal system NTSC, PAL, SECAM (any frequency)

Alternate trigger

CH1 trigger Edge, Video

CH2 trigger Edge, Video

Measurement

Cursor V and T between cursors

Automatic

Peak-to-peak value, average value, root mean

square value, frequency , period,Vmax, Vmin,

Vtop, Vbase, Vamp, Overshoot,

Preshoot, RiseTime, Fall Time, +Width,

-Width, +Duty, -Duty, DelayA B and

DelayA B

Waveform Math +, -, Invert,FFT

Waveform storage 4 waveforms

Lissajou’s figure Bandwidth Full bandwidth

Phase difference ±3 degrees

Probe

compensation

Output voltage (typical) 5V, with the Peak-to-peak value ≥1MΩ

Frequency ( typical) 1KHz square wave

PC interface USB1.1

General Specifications

Display

Model PDS5022S, PDS6042S, PDS6062S PDS6062T,PDS7062T, PDS7102T

Display Type 7.8” color liquid crystal display 8” color liquid crystal display

Display Resolution 640 (horizontal) × 480 (vertical) pixels

Display Color 256 colors 65536 colors, TFT screen

Display Language English, Simplified Chinese

Page 69: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 67

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

Probe compensation

Output voltage (typical) 5V, with the Peak-to-peak value ≥1MΩ

Frequency ( typical) 1KHz square wave

Power

Supply 100-240 VACRMS,50Hz,CAT II

Consumption <15W

Fuse 1A,T class,250V

Battery Optional

Environment

Temperature Working temperature: 0 ~ 40

Storage temperature: -20 ~ +60

Relative Humidity ≤ 90%

Height Operating: 3,000 m

Non-operating: 15,000 m

Cooling Method Normal Air cooling

Mechanical Specifications

Model PDS5022S

PDS6042S

PDS6062S

PDS6062T

PDS7062T

PDS7102T

Weight 1.6kg 1.7kg 1.8kg 1.7kg

Dimension 350mm×157mm× 103mm

*Manufacturer reserves the right to change the specification at anytime without prior notice.

Page 70: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 68

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

8. Appendix B: Accessories

Our Oscilloscope comes with a list of accessories which includes:

A pair of Passive Probes

1xPC suite Software CD (Windows OS only).

1xUSB cable

1x Power Lead (UK or EU)

1xUser's Manual

*Rechargeable battery is optional, NOT included in the standard package, can be purchased separately.

Page 71: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 69

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

9. Appendix C: Maintenance and Service

Keep Oscilloscope and accessories away from dust, liquid and any dump to avoid

damage.

Avoid exposing LCD screen under the direct sunlight and heat.

Disconnect power before cleaning your Oscilloscope.

Use soft wipes to clean any dust from LCD screen, pay extra attention to transparent

screen protector.

Only use soft wipes to clean the surface of Oscilloscope, No any surface cleaners should

be used.

Warning: MAKE SURE YOUR OSCILLOSCOPE IS KEPT IN DRY

CONDITION AT ALL THE TIME, FREE FROM WET/DAMP. ALWAYS CHECK ITS

CONDITION BEFORE POWERING IT ON.

CALIBRATIONS

It’s recommended to calibrate the Oscilloscope every 12 months to maintain its high

performance.

The built-in Self-Calibration can be carried out at anytime if it becomes necessary, to help

your oscilloscope maintain its best possible performance. You can do so if:

• Every 12 months or every 2000 hours whichever comes first.

• The change of ambient temperature exceeds 5 °C.

• After service or repairing.

• Any others if it becomes necessary.

Page 72: OOOOWON User manual for PDS series - OWON Series USER_MANUAL fo… · OOOOWON ® User manual for PDS series V5.02 Email: support@owon.co.uk Tel: +44 8450 508168 ... Portable LCD Digital

User’s Manual: OWON Portable Digital Storage Oscilloscope 70

www.owon.co.uk Email: [email protected] Tel: +44 8450 508168 V5.02

10. Appendix D: Use Rechargeable Battery

All our PDS series Oscilloscope can work on battery, the battery is NOT included in the standard

package. User can turn on the battery indicator if there is a rechargeable battery is installed. The

indicator therefore gives the level of your rechargeable.

Fig 72 Battery level indicator

About Your battery

The lithium battery may not be fully charged when you receive it. Please always make sure to charge

your battery for 12 hours before the first use. (It’s recommended to turn on Oscilloscope while charging

the battery). You battery can last 4-6 hours depending on your usage after fully charged.

, , and indicates the level of your battery. Your battery can still operate up to 5 minutes when

is on.

Note:

When charging battery, please make sure the ambient temperature is within Oscilloscope’s working

range to avoid overheat.

Replace and dispose the rechargeable Battery

The Battery is NOT serviceable. Only qualified service engineer can install and replace the battery if it

is necessary to do so. In case you have to dispose the battery, please follow instruction to protect the

environment.