Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
LaserCEM® Continuous Emissions Monitoring System
Description The LaserCEM is a complete pre-calibrated multicomponent (NO, NO2, NOx, SO2, CO, HCl, CO2, H2O, H2S,
NH3, N2O, COS, SO3, CH4, HF) laser Infrared Spectrometer.
LPS The LaserCEM features the exclusive LPS Low
Pressure Sampling System enabling efficient
installation and reduced operating cost by
eliminating the need for “heated” sampling system.
OFCEAS The LaserCEM uses the patented OFCEAS
(WO 03031949) IR laser technology for enhanced
specificity, selectivity, accuracy and stability (no
span or zero drift ).
LaserCEM® Key Features
1300 850 862 I www.aquagas.com.au I [email protected] I 3 Wirranina Place - Currumbin - 4223 QLD
True Direct Extractive CEMS (No sample treatment) - 100mbar
sampling pressure removes any risk of chemical adsorption,
desorption and condensation along the sample line.
Interferences Free - OFCEAS technology provides exceptional
selectivity, enabling simultaneous monitoring of
multicomponent without interferences, regardless of the gas
matrix.
Ease of integration - The LaserCEM allows digital (Ethernet,
RS485, RS232, ModBus) analogue and TDR IOs.
No Drift - The zero is contained in the signal, enabling
automated and intrinsic zero drift compensation.
Field proven - The LaserCEM is free of optical moving part and
was designed and built strictly for industrial and on-board
mobile applications.
Low maintenance - The LPS system allows flow low rates
within 3 and 9l/h without affecting the response time and re-
ducing considerably dust and materials build-up.
Applications The LaserCEM is designed for Continuous Emissions Monitoring in a large field of application including gas
and coal fired power station, pharmaceutical and chemical industries, refineries, cement plants and waste
incinerators and more...Highly durable to harsh process conditions , the LaserCEM is a field proven,
reliable, robust, cost-effective and user friendly solution for Continuous Emissions Monitoring.
AP2e The LaserCEM is designed and manufactured by AP2e in France and uses patented analysis and
sampling technologies .
LaserCEM® Specifications
INTEGRATION
Dimensions
Weight
Standard 19” 4U rack unit - 550mm depth
20kg
Flow 3 to 9 lph
Sample Temp 600C
Pressure 1 atm. +/- 100mbar at the
sampling point
Sampling Line No heating required if:
Ambient temp> 10C and H2O<40% vol.
Heating required if:
Ambient temp<10C and H2O>40% vol.
Drift None
Response time 200s max. depending on sample line
Interface 5.7” touch screen
Windows XP / WinProceas
Output Ethernet / RS485 / RS232 / ModBus
Options ATEX - Wall mount version
Power supply 110-240 VAC 50/60 Hz - 200W
Operating
conditions Temp 15-35C Pressure 86-108kPa
Compressed air 1-6 bar (oil free)
LaserCEM® Continuous Emissions Monitoring System
1300 850 862 I www.aquagas.com.au I [email protected] I 3 Wirranina Place - Currumbin - 4223 QLD
GAS RANGES DETECTION LIMIT
SO2 0 to 25ppm 0.22 ppm
NO 0 to 60 ppm 0.09 ppm
HCl 0 to 10 ppm 0.01 ppm
NH3 0 to 10 ppm 0.01 ppm
CO 0 to 60 ppm 0.22 ppm
H2O 0 to 40 % 0.1 %
O2 0 to 25 % 0.05 %
CO2 0 to 20% 0.06 %
SO3 0 to 25 ppm 0.20 ppm
N2O 0 to 100 ppm 0.09 ppm
CH4 0 to 100 ppm 0.11 ppm
NO2 0 to 25 ppm 0.08 ppm
HF 0 to 10 ppm 0.01 ppm