Oligonucleotide analysis by Linear MALDI-TOF mass spectrometry Stephen J. Hattan , Kenneth C. Parker, Marvin L. Vestal; SimulTOF Systems, 60 Union Ave, Sudbury, MA - The SimulTOF 100 MALDI-TOF mass spectrometer is a bench-top linear system that is well suited for oligonucleotide, intact protein, and peptide analyses. -Using high energy ion acceleration and post acceleration in the detector, the SimulTOF 100 provides sensitivity and dynamic range for detection for ON, protein and peptides - 5000 Hz laser and 5mm / sec scan speed allows for routine high throughput analysis - Aquisition parameters for current analysis -laser f = 3000 hz -scan speed = 3 mm / sec - laser power = 18 μ joule - Signal average = 500 shots / spectrum Oligonucleotide Ladder Positive ion mode 60mer 50mer 35mer 40mer 30mer 10mer 25mer 20mer 15mer Oligonucleotide Ladder Negative ion mode 10mer 35mer 30mer 40mer 50mer 25mer 20mer 15mer 60mer 5.00E-05 1.00E-04 1.50E-04 2.00E-04 2.50E-04 3.00E-04 3.50E-04 0 100000 200000 300000 400000 500000 600000 700000 Neg mode Pos mode -4.5 -4 -3.5 -3 2.5 3.5 4.5 5.5 Neg mode Pos mode Concentration mol / L Signal intensity (Area) Log Concentration Log Signal Sample matrix preparation Normal preparation (pH ~ 4.0) 3-hydroxypicolinic acid (3-HPA) -Stock solution at 50mg/mL in 1:1 (acetonitrile:H2O) diammonium citrate -Stock solution at 50mg/mL in H 2 O Working matrix solution combine 3-HPA and diammonium Citrate stock in 9:1 ratio. 9 parts HPA : 1 part diammonium Citrate Acidic preparation (pH ~ 3.0) 3-hydroxypicolinic acid (3-HPA) -Stock solution at 50mg/mL in 1:1 (acetonitrile:H 2 O 0.1% TFA) diammonium Citrate -Stock solution at 50mg/mL in H 2 O Working-matrix solution combine 3-HPA and diammonium Citrate stock in 9:1 ratio. 9 parts HPA : 1 part diammonium Citrate Sample deposition 1) Mix working matrix solution with sample 1:1 to give ~ 0.1 – 1 pmol / uL [sample] 2) Spot matrix on sample plate and allow to dry then overlay sample on matrix crystals Matrix assisted laser desorption/ionization (MALDI) mass spectrometry is a powerful tool for performing oligonucleotide (ON) analyses. MALDI-TOF mass spectrometry is routinely applied to the analysis of ON for the investigation of naturally occurring anomalies like single nucleotide polymorphisms (SNPs), alternative splicing, methylation and for purposes of quality control in synthetic production. Despite its popularity, MALDI analysis of ON is more particular to analytical conditions in comparison to MALDI analysis of proteins and peptides. In working with synthetic, purified, single stranded DNA samples, even subtle differences in analytical formulation, spotting procedure and instrument acquisition parameters can have a profound impact on analytical success. Introduction Methods Mass Spectrometry 3-hydroxypicolinic acid (3-HPA) Oligonucleotide structure has both hydrophilic and hydrophobic character Effects of sample preparation pH HPA matrix at pH 4 (normal preparation) tends to crystallize into rings with matrix and sample depositing on around the perimeter of the spot HPA matrix with pH 3 (acidic preparation) crystallized into large shards with marked drop-off in signal intensity Subtle change in pH has a dramatic effect on crystal morphology. This change in morphology may impact ON incorporation and ionization? 3-HPA has pKa of ~4, therefore lower pH will keep the molecule protonated. Results are confirmed with preps from other labs. Other observations - Drying time similar for both preparations ~4-5min - Sample deposition method (co-mingled or overlay) had no effect on results Outside lab pH ~ 4.0 pH ~ 3.0 Our lab Our lab Outside lab Outside lab Our lab Results Sample analysis - With proper sample preparation SimulTOF 100 is easily capable of detecting ON with 100s of bases -Regardless, the sample preparation is still prone to having signal generate from discreet locations in the sample (hotspots) Signal generation still prone to “hotspots” 100mer 90mer 85mer 75mer 80mer 70mer With proper sample preparation analysis of ON spanning a large mass range is possible Sample Acquisition -despite the “hotspot” phenomenon thorough interrogation of sample produces reasonably quantitative signal across an order of magnitude in mass range and sample concentration -Results are consistent for both positive and negative modes of instrument operation Conclusions -Success of ON analysis by MALDI mass spectrometry is highly dependent on proper sample preparation -Subtle changes in 3-HPA matrix preparation formulation can have drastic impact on crystal morphology -Subtle changes in ON sample formulation (pH) can have a drastic impact on signal intensity -Analyses using 3-HPA matrix are prone to “hotspots” in sample signal generation -Adequate sample interrogation can still produce quantitative signal intensity over a large range in sample mass and concentration -Both the quality and quantity of signal are similar in positive and negative modes of instrument operation MP08 - 158 References 1)L. Haff, P. Juhasz, S. Martin, M. Roskey, I. Smirnov, W. Stanick, M. Vestal and K. Waddell; “Oligonucleotide analysis by MALDI-MS.” Analusis 1998; 26: 26-30 2)S. Sauer; “The essence of DNA sample preparation for MALDI mass spectrometry.” J. Biochem. Biophys. Methods. 2007; (70): 311-318 3)S. Sauer, D. Lechner, I. G. Gut; Mass Spec and Genomic Analysis. Kluwer Academic Publishers, Norwell, MA (2001):49-55

Oligonucleotide Ladder Positive ion mode

  • Upload
    sinjin

  • View
    30

  • Download
    1

Embed Size (px)

DESCRIPTION

The SimulTOF 100 MALDI-TOF mass spectrometer is a bench-top linear system that is well suited for oligonucleotide , intact protein, and peptide analyses. - PowerPoint PPT Presentation

Citation preview

Page 1: Oligonucleotide  Ladder Positive ion mode

Oligonucleotide analysis by Linear MALDI-TOF mass spectrometryStephen J. Hattan , Kenneth C. Parker, Marvin L. Vestal; SimulTOF Systems, 60 Union Ave, Sudbury, MA

-The SimulTOF 100 MALDI-TOF mass spectrometer is a bench-top linear system that is well suited for oligonucleotide, intact protein, and peptide analyses.

-Using high energy ion acceleration and post acceleration in the detector, the SimulTOF 100 provides sensitivity and dynamic range for detection for ON, protein and peptides

-5000 Hz laser and 5mm / sec scan speed allows for routine high throughput analysis

-Aquisition parameters for current analysis-laser f = 3000 hz -scan speed = 3 mm / sec-laser power = 18 μ joule-Signal average = 500 shots / spectrum

Oligonucleotide LadderPositive ion mode

60mer

50mer35mer 40mer30mer

10mer

25mer20mer

15mer

Oligonucleotide Ladder Negative ion mode 10mer

35mer30mer 40mer 50mer25mer20mer

15mer

60mer

5.00E-05 1.00E-04 1.50E-04 2.00E-04 2.50E-04 3.00E-04 3.50E-040

100000

200000

300000

400000

500000

600000

700000

Neg modePos mode

-4.4 -4.2 -4 -3.8 -3.6 -3.4 -3.2 -32.5

33.5

44.5

55.5

6

Neg modePos mode

Concentration mol / L

Sign

al in

tens

ity (A

rea)

Log Concentration

Log

Sign

al

Sample matrix preparation Normal preparation (pH ~ 4.0)

3-hydroxypicolinic acid (3-HPA) -Stock solution at 50mg/mL in 1:1 (acetonitrile:H2O)   

diammonium citrate -Stock solution at  50mg/mL in H2O

Working matrix solution combine 3-HPA and diammonium Citrate stock in 9:1 ratio.9 parts HPA : 1 part diammonium Citrate

Acidic preparation (pH ~ 3.0)3-hydroxypicolinic acid (3-HPA)

-Stock solution at 50mg/mL in 1:1 (acetonitrile:H2O 0.1% TFA)   diammonium Citrate

-Stock solution at  50mg/mL in H2OWorking-matrix solution

combine 3-HPA and diammonium Citrate stock in 9:1 ratio.9 parts HPA : 1 part diammonium Citrate

Sample deposition1) Mix working matrix solution with sample 1:1 to give ~ 0.1 – 1 pmol / uL [sample]2) Spot matrix on sample plate and allow to dry then overlay sample on matrix crystals

Matrix assisted laser desorption/ionization (MALDI) mass spectrometry is a powerful tool for performing oligonucleotide (ON) analyses. MALDI-TOF mass spectrometry is routinely applied to the analysis of ON for the investigation of naturally occurring anomalies like single nucleotide polymorphisms (SNPs), alternative splicing, methylation and for purposes of quality control in synthetic production. Despite its popularity, MALDI analysis of ON is more particular to analytical conditions in comparison to MALDI analysis of proteins and peptides. In working with synthetic, purified, single stranded DNA samples, even subtle differences in analytical formulation, spotting procedure and instrument acquisition parameters can have a profound impact on analytical success.

Introduction

Methods

Mass Spectrometry

3-hydroxypicolinic acid (3-HPA)

Oligonucleotide structure has both hydrophilic and hydrophobic character

Effects of sample preparation pHHPA matrix at pH 4 (normal preparation) tends to crystallize into rings with matrix and sample depositing onaround the perimeter of the spotHPA matrix with pH 3 (acidic preparation) crystallized into large shards with marked drop-off in signal intensity

Subtle change in pH has a dramatic effect on crystal morphology. This change in morphology may impact ON incorporation and ionization? 3-HPA has pKa of ~4, therefore lower pH will keep the molecule protonated. Results are confirmed with preps from other labs.

Other observations - Drying time similar for both preparations ~4-5min - Sample deposition method (co-mingled or overlay) had no effect on resultsOutside lab

pH ~ 4.0

pH ~ 3.0

Our lab

Our lab

Outside lab

Outside lab

Our lab

Results

Sample analysis-With proper sample preparation SimulTOF 100 is easily capable of detecting ON with 100s of bases-Regardless, the sample preparation is still prone to having signal generate from discreet locations in the sample(hotspots)

Signal generation still prone to “hotspots”

100mer

90mer

85mer75mer80mer

70mer

With proper sample preparationanalysis of ON spanninga large mass range is possible

Sample Acquisition-despite the “hotspot” phenomenon thorough interrogation of sample produces reasonably quantitative signal across an order of magnitude in mass range and sample concentration-Results are consistent for both positive and negative modes of instrument operation

Conclusions-Success of ON analysis by MALDI mass spectrometry is highly dependent on proper sample preparation-Subtle changes in 3-HPA matrix preparation formulation can have drastic impact on crystal morphology -Subtle changes in ON sample formulation (pH) can have a drastic impact on signal intensity-Analyses using 3-HPA matrix are prone to “hotspots” in sample signal generation-Adequate sample interrogation can still produce quantitative signal intensity over a large range in sample mass and concentration -Both the quality and quantity of signal are similar in positive and negative modes of instrument operation

MP08 - 158

References 1)L. Haff, P. Juhasz, S. Martin, M. Roskey, I. Smirnov, W. Stanick, M. Vestal and K. Waddell; “Oligonucleotide analysis by MALDI-MS.” Analusis 1998; 26: 26-30 2)S. Sauer; “The essence of DNA sample preparation for MALDI mass spectrometry.”J. Biochem. Biophys. Methods. 2007; (70): 311-3183)S. Sauer, D. Lechner, I. G. Gut; Mass Spec and Genomic Analysis. Kluwer Academic Publishers, Norwell, MA (2001):49-55