Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
Michelle Rose Ga1
A Thesis submitted in confomiity with
the requirements for the Degree of
MASTER OF APPLIED SCIENCE
in the University of Toronto
O Copyright by M.R. Gd, 1999
Depart ment of Mechanical Engineering
Instihite of Biomedical and Biomaterials Engineering
University of Toronto
National Library l*l of Canada Bibliothèque nationale du Canada
Acquisitions and Acquisitions et Bibliographie Services sewices bibliographiques
395 Wellington Street 395. rue Wellington OttawaON KtAON4 OttawaON K1AûN4 Canada canada
The author has granted a non- exclusive licence allowing the National Library of Canada to reproduce, loan, distribute or seiI copies of this thesis in rnicroform, paper or elec tronic formats.
L'auteur a accordé me Licence non exclusive permettant à la Bibliothèque nationale du Canada de reproduire, prêter, distribuer ou vendre des copies de cetîe thése sous la forme de microficheJfilm, de reproduction sur papier ou sur format électronique.
The author retains ownership of the L'auteur conserve la propriété du copyright in this thesis. Neither the droit d'auteur qui protège cette thèse. thesis nor substantial extracts fiom it Ni la thèse ni des extraits substantiels may be printed or otherwise de celle-ci ne doivent être imprimés reproduced without the author's ou autrement reproduits sans son permission. autorisation.
The development of a glaucoma drainage valve was continued firom an earlier study. The
valve featured an osmotically generated opening/closing pressure, which was expected to
provide a level of intraocular pressure control supenor to that of currently used valves. A
large sale prototype was designed, constructed, mathematically modelled, and tested.
Tests ushg a constant supply pressure revealed an unacceptably long response tirne. In
order to reduce the response time, a second large scale prototype was designed,
constructed, mathematically modeiled, and tested. Testing with a constant supply pressure
c o h e d an adequate reduction in response tirne. The dynamic response of the valve
was then tested by replacing the constant supply pressure with constant supply flows. The
valve did not wmpletely close at low flow levels, a flaw which must be corrected before
development may continue.
My sincere thanks to my advisor, Dr. C. Ross Ethier, for his guidance and incredible
patience during the past two years. There are no adequate words to express my
appreciation.
Thanks to MIE technicians David Esdaile, Paul Kovar, and Jeff Sansorne for a n s w e ~ g
rnany questions and doing a wonderhl job mnstructing my prototypes. Thanks to Brenda
Fung (MIE) and Anne Mitchell (BBME) for helping me with the administrative details.
Financial support for this project was kindly provideci by the Glaucoma Research Society
of Canada and the Naturai Sciences a d Engineering Research Council of Canada.
. . Ab stract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . u
... Acknowledgements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ui Tableofcontents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv ListofSymbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
ListofFigures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . List of Tables xi
Chapter One: Introduction 1 . 1 Overview & Motivation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 The Eye and Glaucoma . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3 Glaucuma Treatment 4
1.3.1 Medication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
. . . . . . . . . . . . . . . . . . . . . . . . . . P-Adrenergic Antagonists (P-blockers) 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . Carûonic Anhydrase Inhibitors (CAI) 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cholinergie Agonists 5 Prostagiandins . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 -2 Trabeculoplasty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 -3 Filtration Surgery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1 .3.4 Drainage Implants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Non-Restrictive Flow Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Restrictive Flow Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1
1.4 Previous work on the Proposed Drainage Impiant . . . . . . . . . . . . . . . . . . . . . . 14 1.4.1 Osmotic Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
1.4.2 Proposed Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
1.4.3 Design Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
1 .4.4 Verification of Valve Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
1.4.5 Verification of Osmotic Valve Characteristics . . . . . . . . . . . . . . . . . . . . . 25
1.4.6Summ ary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Chapter Two: Project Outline
2.1 Hypothesis and Long Term Goals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
2.2 Specific Objectives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3 Valve II Testing Results 80
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3.1 Dextran Retention 80
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3 -2 Valve Il Response Tirne 81
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.3.3 Valve II Dynamic Response 84
Chapter Seven: Conclusions 7.1Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
7.2 Conclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 7.3 Recomrnendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
7.3.1 Valve Design and Experimental Apparatus . . . . . . . . . . . . . . . . . . . . . . . 92
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3 -2 Future Work 93
References
Appendix A . Drawings for Valve 1
Appendk B . Drawings for Valve II
Appendix C . Regession Analysis Output
Appendk D . Deformation Calculations
vii
area coliapsible tube cross sectional area
manometer tube cross sectional area tiltration surface ara virial coefficients
concentration
tube diameter modulus of elasticity
gravitational acceleration
collapsible tube wall thickness membrane permeability
tube Iaw constant
tube length
molecular weight
molecular weight cut-off
pressurtb chamber pressure
supply pressure
net potential diierence
pressure in dextran cuff
volume flow gas constant
flow resistance coliapsible tube radius
Reynolds number
temperature
tune
volume
collapsible tube volume
min
rnL
mL
manorneter tube volume flow across semipermeable membrane fluid velocity dynamic fluid viscosity kinematic fluid viscosity Poisson's ratio osmotic pressure fluid density time constant
d s
N dm2 m2/s - Pa, d g kgh?
min
Figure 1.1. Sagittal section through a human eyebd . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 1.2. Major stages of filtrationsurgery 8
. . . . . . . . . . . . . . . . . . . Figure 1.3 : In situ positioning of the Ahmed Glaucoma Valve 10
Figure 1.4. Exarnples of currently existing glaucoma drainage implants . . . . . . . . . . . 12
Figure 1.5. Two-ceiied system with osmolarity gradient . . . . . . . . . . . . . . . . . . . . . . . 15
Figure 1.6. Schematic drawings of the proposed osrnotic valve . . . . . . . . . . . . . . . . . . 19
Figure 1.7. Sit's (1 996) osmotically controlled valve prototype . . . . . . . . . . . . . . . . . . 20
Figure 1.8 : Experimental apparatus used to determine how closing pressure affects flow
through the collapsible tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Figure 1.9. Collapsible tube flow characteristics (water) . . . . . . . . . . . . . . . . . . . . . . . 23
Figure 1.1 O: Coiiapsible tube flow characteristics (glycerol) . . . . . . . . . . . . . . . . . . . . 24
Figure 1.1 1 : Experimental apparatus used to detennine feasibility of using osmotic pressure as closing pressure source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Figure 1.12. Osrnostic valve flow charateristics (Sit 1996) . . . . . . . . . . . . . . . . . . . . . 26
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 3.1 : Valve 1 33
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 3.2. Valve 1 O dialysis membrane sed 34
Figure 3.3. ValveI-fillingportsandfillingtraas . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 3.4. Valve 1 assembly 36
. . . . . . . . . . . . . . Figure 3.5. Cornparison of Sit's prototype and Valve I O water flow 40
. . . . . . . . . . . . Figure 3.6. Cornparison of Sit's prototype and Valve I w glycerol flow 41
Figure 3.7: Comparison of Sit's (1996) prototype and Valve 1 . glycerol flow. osmotic closingpressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 4 . 1: Manometer configuration 46
. . . . . . . . . . . . . . . . . . . . . . Figure 4.2. Water flux across a semipenneable membrane 48
. . . . . . . . . . . . . . . . Figure 4.3 : Effect of concentration polarization on glycerol flow 49
. . . . . . . . . . . . Figure 4.4. Hypothesized pressure vs t h e relationship in pressure cuff 50
. . . . . . . . . . . . . . . . . . Figure 4.5. Water flux across 14K MWCO dialysis membrane 54
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 4.6. Volume changes in the dextran c& 55
Figure 4.7. Experimental apparatus for determinhg system response tirne . . . . . . . . . 58
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 4.8. Valve 1 time response data 60
. . . . . . . . . Figure 4.9. Log-linear regression curve fit for Valve I time response data 61
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Figure S . 1 : Modifications to Valve 1 66
. . . . . . . . . . . . . . . . . . . . . . . . . . . Figure 5.2. Valve 1 (modified) time response data 68
. . . . . . . . . . . . . Figure 6.1 : Valve II (a) photograph; @) labelled cross-sectional view 71
xi
Figure 6.2. Membrane support structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Figure 6.3. Experimental apparatus for testing dynamic response of Valve Ii . . . . . . . 79
Figure 6.4. Valve II time response data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Figure 6.5. Valve 11 dynarnic tests results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Figure 6.6. The tube law . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
. . . . . . . . Table 4.1 : Cornparison of published values for wia l coefficients of dextran 47
Table 4.2. Cornparison of system time constants for Valve 1 . . . . . . . . . . . . . . . . . . . . 62
Table 5.1 : Comparison of syaem time constants for Valve 1 and Valve 1 (modified) . . 69
. . . . . . Table 6.1 : Cornparison of system tirne constants between Valve I and Valve II 82
Neady 2% of the population over 50 years old suffers fiom glaucoma, the second leading
cause of blindness in Canada (Elolia et al., 1998). In most cases, glaucoma is
characterired by hi@ intraocular pressure (IOP). Kigh IOP Ievels develop when aqueous
humor drainage pathways become p d a l l y blocked causing an increase in resistance to
aqueous humor outfiow. Untreated glaucorna causes heparable damage to the optic
nerve.
Although there is no way of curing this disease, treatrnents have been developed to reduce
IOP and help presewe vision. They range fiom non-invasive topical medications to
drainage implants. Implants, which incorporate sorne form of flow-regulation mechanism,
provide an alternative aqueous humor drainage pathway nom the eye, and are typicaliy
2
used as a last resort in severe cases of glaucoma. Unfominately, implants currently in use
provide much coarser pressure control than is desirable. Specincally, IOP can often
becorne: (i) too high before fluid is ailowed to drain from the eye, risking further damage
to the optic nerve; or (ü) too low before outfiow is restricted, risking softening of the eye
and coliapse of the cornea. An improved Unplant would eliminate this dangerous hysteresis
between maximum and minimum IOPs, providing finer control and mahtaining a d e level
of IOP at al1 tirnes.
In the remainder of this chapter, the eye, glaucoma, glaucoma treatments, and the previous
work done on a new implant design are discussed in more detail.
1.2 THE EYE AND GLAUCOMA
The eye (figure 1.1) is essentiaily a sofbtissue sack that is divided into three chambers:
vitreous, posterior and anterior. Containing no bone or wtilage, the eye's shape is
maintained by the action of fluid pressure. The vitreous chamber is Wed with a
transparent, geClike material called the vitreous body. The posterior and antenor
chambers are fiiled with a transparent liquid called the aqueous humor. In addition to
providing structure, the aqueous humor also supplies the celis of the lens and coma with
nutrients and carries away metabolic waste products. Blood, which is normdy
responsible for this task, is not present in these tissues, since its opaque nature would
prevent light fiom passing through the eye.
3
Aqueous humor, a clear, coiourless liquid, is produced by the ciliary body at a rate of 2-
3pVmui (Guyton and Hall, 1996), and continually drains from the antenor chamber
through a duct known as Schlernrn's Canal, resulting in a relatively constant IOP. The
average valve of IOP in a healthy individual is 1 SmmHg, with a range fiom 12 to 2OrnmHg
(Guyton and Hall, 1996).
Figm 1.1: Sagittal section through a human eyeball. pavson. 19691
Most giaucomas develop when the aqueous humor drainage pathways become partially or
fuUy blocked, increasing resistance to normal outflow. IOP is related to aqueous hurnor
flow (Q) and flow resistance (R) by:
IOP = QR
4
The volume of the eye can be considered constant, therefore the tirne-averaged outflow of
aqueous humor must be equal to the amount produced by the ciliary body. Thus, when
the resistance to aqueous humor outfiow is increased, the result is an elevated IOP. This
elevated pressure causes irreparable darnage to the optic nerve, possibly by inhibithg
blood flow in capillaries of the optic nerve, thus preventing nutnents firom reaching the
individual neurons (Varma and Minckler, 1996). lf the IOP is permitted to remain above
2OmmHg for an extended period of the, vision is progressively impaired and blindness
eventuaiiy results. In extreme cases where IOP can be as high as 60 or 7OrnmH&
blindness rnay ocnir w i t h days or even hours (Guyton and Hall, 1996). in the case of
normal-tension glaucomq even 10P values of 15-2OmmHg can cause optic nerve damage
(Werner, 1 996).
Current medical therapy for al1 forms of glaucoma seeks to lower IOP. This is
accomplished by either reducing aqueous humor production, or reducing resistance to
aqueous humor oudlow. Frequently, these two strategies are combineci. Aithough
darnage to the optic nerve can not be repaired, lowering IOP can help to prevem fkther
damage. Available approaches for the treatment of glaucoma are reviewed below.
1.3.1 MEDICATION
The simplest treatment is the use of medication. Various types of dmgs can be used to
lower IOP. Some of these drug classes and the physiological mechanisms by which they
lower IOP are sumrnarked below.
P-blockers are believed to act on the P,-adrenergic receptors of the epithelid cells of the
ciliary body. Stimulation of the ciliary process is prevented, inhibiting the production of
aqueous humor, and reducing flow through the eye to a basal level.
Carbonic anhydrase catalyses the chernical reaction that fonns bicarbonate. In the ciliary
body, the formation of bicarbonate is iinked to the secretion of sodium ions, which are
required for the production of aqueous humor. CAIs reduce production (i.e. flow) of
aqueous hurnor by inhibiting the formation of biwbonate.
CHOUNERGIC AGONISTS
Cholinergie agonists, in pmiculw, pilocarpine, act on the muscarhic cholinergie receptors
of the ciliary muscle fibres, stimulating contraction. Contraction of the ciliary muscle
produces tension on the scleral spur. This tension is thought to pull open the trabmlar
6
meshwork (the tissue which separates Schlemm's Canal nom the anterior chamber),
reducing resktance to aqueous humor outflow.
PROSTAGLANDNS
Although the major aqueous humor drainage pathway is via the trabecular meshwork and
Schlemm's Canal, flow may also take the uveosclerd route. (Aqueous humor exiting the
anterior charnber by this route passes into the root of the iris, then flows between ciliary
muscle fibres, then percoiates through the sclera). The proaoglandin PGF2, dilates the
intramuscular spaces in the ciiiary body, reducing flow resistance, thus lowering IOP.
Drug therapy is a non-invasive, low risk treatment. However, in many cases, medication
does not sufficiently lower IOP and surgical intervention is required.
The least invasive surgical therapy for giaucoma is argon laser trabeculoplasty (ALT), in
which an argon laser is used to make smaii bums in the trabecular meshwork. It has been
theorized that during the healing process, contraction of the tissue causes the trabecular
meshwork to lose some of its resistance to the flow of aqueous humor.
ALT is not always successfùl and is no longer wmmonly used. In about 50% of cases,
IOP levels become dangerously high within five years of the procedure. In these
situations, a second ALT is generally performed, unless IOP levels have rernained normal
for l e s than a year, in which case filtration surgeiy is recommended.
During filtration surgery, the surgeon attempts to bypass the outfiow pathway altogether
and create an altemate route for aqueous humor outfiow. Figures 1.2 (a) and (b) illustrate
the procedure, and show a cross-sectional view of the resulting drainage pathway,
respectively. An incision is made in the conjunctiva to expose the sclera. Ushg a sharp
blade, the surgeon penetrates through haif of the scleral layer and cuts to fom a
rectangular fiap. A small block of tissue is excised from undemeath the fiap before it is
loosely sutured closed. The conjunctiva is tightly sutured back in its original position.
Aqueous humor seeps out of the chamber through the f'iap's edges. The flow rate can be
adjusted by changing the tightness of the sutures.
This procedure has a poor prognosis for certain types of gfaucoma, and also if the patient
has had surgery penormed on the sclera previously. However, even if the prognosis is
good, the formation of scar tissue may result in failure to maintain a Iow IOP. Laser
treatment to remove the scar tissue is sometixnes effkctive. If this is unsuccessfiii, the
surgeon may attempt another fltration surgery in another quadrant of the eye or
recommend the implantation of a drainage device.
(ii i)
(ii
scleral flap
. . - . . . . . . . . . . . . . . . . . . . . .
1 +excised block
. t Figure L .2(a): Major stages of filtration surgery; (i) penetrating the conjunctiva, (ii)
cutting the half-thickness sclerai flap, (fi) excising a block of scled tissue, (iv) the
sutured flap [ Ritch et al., 19961; (b): Cross-sectional schematic of the sclera after
fiitration surgery, arrows indicate aqueous humor outflow pathway.
A Nowl Glaucoma h m q c Valve
In cases where filtration surgery fails to control IOP or has a poor prognosis, implantable
devices are used. Implants are designed to provide a low-resistance drainage route for
aqueous humor, thus bypassing conventional outflow pathways. Implants direct aqueous
humor outnow frorn the anterior chamber to the space between the sclera and conjunctiva.
From there, the aqueous humor moves through intercellular spaces and is absorbed by
orbital capillaries and lymphatics (Ritch et al., 1996). A wide variety of implants have
been developed in the past 30 years, ranging fiom a simple tube to more complex syaems
involving filtration plates and valves.
Genedy speaking, these drainage systems consist of two sections: an inlet tube and an
outlet structure. The inlet tubes are fairly standard between the various systems: a piece
of silicone tubing with an outer diameter of approximately 0.6 mm and inner diameter of
about 0.3 mm. The proximal end is placed in the antenor chamber. The distai end of the
tubhg is comeaed to a variety of possible structures, such as silicone bands, round
silicone discs, or irregularly shaped plates through which the fluid exits. These structures
are oflen referred to as explants. Figure 1.3 illustrates the in situ positioning of an
implant, and its size relative to the eye.
Figure 1.3: in situ positioning of the Ahrned Glaucoma Valve New Worid Medical,
19961
If aqueous hurnor were perrnitted to flow fieely through an explant, the anterior chamber
would coliapse due to hypotony, that is, low intraocular pressure. Therefore, explants
rnust incorporate some mechanism by which outfiow is controlled. Depending on the
outflow control mechanism in the explant, drainage implants are classified as either non-
restrictive or restrictive flow devices. Current implants belonging to both classifications
are described below, and are illustrated in figure 1.4.
An object implanteci in the body will soon find itself covered with fibrous tissue (Pilliar,
1997). The Molteno drainage implant, introduced in 1969, takes advantage of this
phenornenon to limit aqueous humor outflow. Fibrous tissue foms a "filtration bleb"
around a perforated, aaylic plate explant. The filtration bleb pennits sufficient outflow to
maintain a satisfactorily low IOP (Molteno et al., 1977), yet provides enough flow
resistance to prevent hypotony .
11
The Baerveldt implant works on the same prhciple as Molteno's device, but has holes in
the centre of the acrylic plate explant. When the fiitration bleb foms, the fibrous tissue
goes through the holes, holding the explant f i d y in place.
The major disadvantage associated with these types of explants is that the filtration bleb
takes several weeks to form. This means that there must be a way to temporarily limit the
oudlow during bleb development. To achieve this, the silicone tube leading fiom the
antenor chamber to the explant is often partiaily tied or clamped at the t h e of
implantation, requiring a second surgery several weeks later to release the constriction.
R E S ~ C T T V E FLOW DEVICES
Attempting to avoid shon-term post-operative complications, particularly overdrainage
and hypo tony, implants were designed t hat incorporate flow-restrictive elements.
The Krupin Eye Disk consias of a silicone tube and plate explant. similar to Molteno's
design. However, rather than simply allowing fluid in the tube to drain into the explant,
the tube's distal end is sealed and there are severai horizontal and venicai slits. As IOP
increases, the tube elastically defoms and permits fluid to enter the explant.
The OptiMed implant regulates flow with a conductive polymethyl methacrylate matrix.
Capillary action draws fluid through the matxix as IOP increases.
In the Ahmed implant, fluid must flow between two layers of thin silicone membrane
before leavhg the explant. At low IOP, the membrane layers collapse preventing outflow;
and as IOP increases, the fluid pressure forces the membrane layers apart, openhg the
pathway.
Figure 1.4: Examples of cwrentty existing giaucoma drainage implants: Molteno (A),
Baerveldt (B), ûptimed (C), Ahmed O), Knrpin (E) [Ritch er al., 19961
13
In vivo testing of several ditferent giaucoma drainage implants was performed in rabbit
eyes by Prata et al. (1 995). Their results demonstraîed that none of the devices
maintained their nominal pressure levels when being pefised at the flow rates expected in
human eyes. The implants referred to as "valves" pedormed like fiow restrictors, not as
true valves which open and close at specific pressures. Porter et al. (1997) did an in virro
study of several implants and discovered a wide range in their penormances. The valved
implants had average ciosing pressures that ranged between O. 1 and 12.6mmHg. Only 2
out of thq8 valves tested by Porter el al. succeeded in closing before the pressure
dropped below 9rnmHg. The desired IOP in moa cases is approximately 1SmmHg. Some
surgeons rnay lower IOP to the IO- 12mmHg range, but pressures below this range put the
coma at risk of coilapshg.
The disadvantage with this class of implants is that although they offer post-operative flow
resistance, they do not behave as tnie valves with speci£ic and consistent opening and
closing pressures. This may be an improvement over non-restrictive flow devices, but the
problem of accurately controllhg post-operative IOP remains unaddressed.
A new type of drainage implant was proposed in a study by Sit (1996). This device was
expected to elimlliate the difnculties expenenced with current implants by balancing fluid
and osmotic pressure to control outfiow.
Sit's valve design took the following criteria into consideration:
the valve should have definite opening and closing pressures in order to precisely controI IOP the valve should be fiilly open at aiî pressures above the opening pressure the valve should be simply designed and durable, containing no electronic components and a minimum of moving parts the valve should be compact and biocompatible
The flow through Sit's valve is controlled using osmotic pressure. Therefore, pnor to any
further discussion on Sit's valve, a brief overview of osrnotic pressure is necessary.
Osmosis is the movement of water across a serni-permeable membrane from an area of
high osrnolarity to one of low osmolarity. Osrnolarity refers to the concentration of
dissolved cornponents. For example, a 1 millimolar solution of NaCl, which dissociates
into Na' and CI', has an osmolarity of 2 miliiosmoles.
15
In a twoceiied system such as the one show in figure 1.5, water moves across the serni-
permeable membrane f?om A to B in response to the osmolarity gradient. The result is
higher pressure in B, which drives water fiom B to A Eventually, the trammembrane
osmolarity and pressure dserences reach a point of equilibnurn where the net flux of
water across the membrane is zero. The pressure difference which exists between the two
chambers at this point is referred to as the osmotic pressure dserence.
Figure 1.5: A two-celled system. The celis are separateci with a semi-permeable
membrane and the ceiis contain solutions of drfferent osmolarity. The m w s represent
water crosshg the membrane in response to osmolarity and pressure gradients. Initidly,
(a) the net flux of water is ffom A to B in response to the osmolarity differenœ. resulting
in a higher pressure in B. Eventuaily (b) the pressure difference, hn between the cells
becornes great enough to d u c e the net flux of water between the œlls to zero. The
value of AIi at this point is the osmotic pressure diflerence.
16
The relationship between the osmotic pressure and the solute concentration of a solution
can be described by the equation:
Where the osmotic pressure, 4 is expressed in Pascals; A, are solute-solvent-condition
specific vinal coefficients pa(mUgfi; and c is the solute concentration WrnL]. For ideal
solutions, in which solute-solute interactions can be negleded, this equation can be
simplified to an expression known as the van't Hoff equation:
II = cRT
Where II is again expressed in Pascals; c is the concentration of dissociated species
(osmolarity) between the two chamben [m~=moVm']; R is the gas constant
[8.3 145 1 kgm2/(s2*rnolX)]; and T is the temperature KI. The osmotic pressure gradient,
AII, is determined by calculating II values for both solutions and taking the diference.
Sit's valve design is based on the concept of Stariing resistors, which are used to study
fluid dynarnics in blood vessels. A Starling resistor consists of a collapsible tube that is
attached to ngid tubes at both ends and surrounded by a rîgid chamber. The collapsible
tube is collapsed by pressurizllig the interior of the chamber. An extemal reservoir
supplies the fluid that flows through the collapsible tube.
Using such a device, Lyon et al. (1980) determined relationships between fiow (Q),
chamber pressure (PJ and fluid supply pressure (P,). For low Reynolds number flows, it
was shown that when fluid supply pressures were below the chamber pressure, the
coiiapsible tube remained closed and there was essentially no flow:
When fluid supply pressures were above the chamber pressure, the coiiapsible tube
opened, and fluid passed through the device. The outflow in this case depended on supply
pressure through the relationship:
Where R is the fiow resistance in the tube, dependent on fluid viscosity, tube length and
tube radius.
These relationships between supply pressure and outfiow were deemed ideal for a
glaucoma drainage valve: closed to flow when P,<P,, open to flow when P,>P,. However,
a Starling resistor only exhibits this behaviour if the Reynolds number of the flow is
sufficiently low, that is, less than 1 (Conrad. 1969). Treating aqueous humor as water at
18
37°C and assuming: a viscosity ( v ) as 7.46E-7 m2/s (Roberson & Crowe, 1993), a Bow
rate (Q) as 2p.L/min, and a typical tube diameter @) of 1 mm, the Reynolds number for
aqueous hurnor fiow is computed as:
This Reynolds number was l e s than 1, indicating that a glaucoma drainage implant
incorporating a Starling resistor-type valve could indeed provide a precise opening and
closing pressure. The challenge in adapting the Starling resistor for use in an implant was
to determine an appropnate rnethod of pressurizing the space around the coUapsible tube.
The pressure source had to be sescontaineci and constant.
Sit decided to use osmotic pressure to generate the opening and closing pressure of the
coilapsible tube. Like the Starling resistor, this "osmotic valve" consisted of a collapsible
latex tube housed in a rigid chamber. The collapsible tube was surrounded by a
semipermeable membrane "cuff'. The rigid chamber was filled with water, and the cuff
was filied with a solution of higher osmolarity (figure 1.6a).
Figure 1.6: Schematic drawings of the proposed osmotic valve in the open and closed
positions. h w s indicate fluid flow through the open valve. ' A indiates a space
filied with water, "B" indicates a space filleci with a solution of higher osmolarity; (a)
cross-section, (b) longitudinai cross-section.
This design is comparable to the simple two-celled system discussed in section 1.4.1; the
outer chamber containhg water is ce1 A, and the cuffis ce11 B. The major differences are
that celi B is inside of celi A, rather than the cells being beside each other; and there is a
collapsible tube running through ceIl B. (Note that any fluid passing through the lumen of
the coiiapsible tube does not corne into contact with the solution in cell B.) Following the
osmotic gradient, water moves from A to B. and the resulting pressure increase in B
collapses the latex tube (figure 1.6b). When the supply fluid pressure is just greater than
the osmotic pressure difFerence between A and B, water is forced out of the cuff, allowing
the coiiapsed tube to open and the supply fluid to pass through the lumen. Once the
supply pressure drops to a value below the osmotic pressure gradient, water moves from
A to B and the tube collapses.
1.4.3 DESIGN DETAILS
Sit's device (figure 1.7) incorporated a 12" length of collapsible latex tubing (%" diameter
Penrose Tube, Baxter Medical Supply, Mississauga ON) within a semi-permeable
membrane "cuff' (Spectra Por 2 Dialysis Tubing, 12000-14000 MWCO, 32mm-flat dry
width, Spectmm Medical Technologies, Houston TX). Note: the MWCO or Molecular
Weight Cutoffrating indicates the size of the srnailest molecules that are not permitted to
pass through the membrane, Le. molecules of a molecular weight less than 12000 atornic
rnass units are able to pass through and those greater than 14000 are retained.
Both ends of the collapsible tubing and semipermeable membrane were attached to %"
buMead fittings and secured with wire wraps. This assembly was housed in a @id
plexiglass chamber with holes for the buMead fittings on the ends. Fastened with
machine screws, the chamber iid compressed a rectangular gasket, providing a watertight
seal. Two !4" buMead fittings in the chamber lid served as tilling and bleed ports.
Figure 1.7: Sit's (1996) osmotically controiied vaive prototype.
This osmotic pressure cuntroiled valve was thought to be an improvement over currently
used devices because it would aüow more precise control over IOP. Slit valves in silicone
tubes, cornmonly used in current valved implants, will begin to elastically defom (Le.
open) at pressures lower than their nominal opening pressure. This means that fluid is
passing through the valve before it is supposed to open, making maintenance of a
consistent IOP difncult. In addition, the nominal opening pressure of a slit valve is greatly
dependent on the slit size and elastic properties of the tube. Thus it is not unusual for
"identicai" valves to have dserent opening pressures (Porter et al., 1997). On the other
hand, in the case of Sit's design, it was hypothesized that overcoming the osmotic
pressure difference would be a repeatable event. Thus, it was hoped that this design would
result in a consistent valve opening pressure.
1.4.4 VERIFICATION OF VALVE CHARACTER~STICS
Before testing the complete valve, Sit investigated the behaviour of the coliapsible tube
independently of the osmotic pressure cuE The purpose of this investigation was to
determine how pressure exerted on the coflapsible tube afFeas the flow through it. The
experimental apparatus (figure 1.8) consisted of the valve, without the dextran cufF. a
water reservoir to set the valve's closing pressure; and a fluid supply reservoir.
tId 4 fluid eupply. Pe source
Figure 1.8: Experimental apparatus used to determine how closing pressure affects flow
through the coffapsible tube.
4 &mal Pc eource
valve v
For the first set of experiments, the supply fluid was water. The water reservoir was set to
an arbitrarily chosen height of lOcm above the valve outlet, which translates into a closing
4 collection reeervoir
\
pressure (PJ of 7.4rnmHg. The height of the supply reservoir was incrementaily
>>
,
increased, and supply pressure (PJ and correspondhg flow through the valve were
recorded. The height of the water reservoir was then increased to 20- which translates
into a P, of 14.7mmHg and the experirnents were repeated.
In a second set of experiments, the supply fluid was changed to glycerol. Glycerol has a
viswsity three orders of magnitude higher than water and was expected to provide the
low Reynolds number flow expected in vivo.
When water was used as the supply fluid, results show that the valve aiways dowed a
s m d amount of oudlow (figure 1.9). However, there was a dramatic increase in flow
23
afker P, surpassed P,. Visual observations of the coiiapsible tube indicated that it remaineci
closed when P, was les than P,. However, dong the folded sides of the tube there
appeared to be small channels, through which water evidently passed prior to opening.
Figure 1.9: Enect of (a) 7.4 and (b) 14.7mmHg coiiapsiile tube closing pressures on
water flow at variable supply pressures.
When giycerol was used as the supply fluid, results show that the valve aiiowed no flow
until P, surpasseci P, (figure 1.10). n ie channels observed during the water tests were also
present during the glycerol tests. However, the flow resistance through the narrow
channels was hi& enough to prevent significant glycerol outtlow until P, > P,.
O 1 2 3 4
glycerol outfiow [mus]
O 1 2 3 4 5
glycerol outfiow [mus]
Figure 1.10: Enat of (a) 7.4 and @) 14.7mmHg coilapsible tube dosing prasurrs on
giycerol flow at variable supply pressures.
These preliminary tests, particularly those involving glycerol as the supply fluid, indicated
that flow through a wliapsible tube can indeed be regulated by exerting a closing pressure
on it f?om an extemal source. Further testing involved replacing the extemal pressure
source with the osmotic pressure source.
1.4.5 VERIFKATION OF O S M O ~ C VALVE CHARA~RISTICS
The purpose of Sit's next investigation was to deterrnine the feasibility of using an osmotic
pressure source to set the valve's closing pressure. The experimental apparatus (figure
1.1 1) consisted of: the valve, with the semi-permeable membrane cuff in place; a supply
reservoir; and a penstaltic pump.
\
4 collection rcsavoir valve b
Figure 1.1 1 : Experimentai apparatus used to determine fm'bility of usiag an osmotic
pressure source to set the valve's closing pressure.
26
The d w a s Wed with a 0.224rnM solution of 298 000 MW dextran (Sigma Chernicals,
St. Louis MO). Assurning the validity of using the van't Hoff equation in this situation by
neglecting solute-solute interactions, the vdue 0.224rnM corresponds to an osmotic
pressure of 4.3mmHg. (These values are assumed to have been selected arbitrarily by Sit,
as they do not have any obvious significance.)
Lnitial observations indicated that several hours were required for equilibrium to be
established across the semi-permeable membrane. For this reason, the glycerol supply
reservoir was set at a value of 4.6mrnHg, which is slightly greater than the osmoticdy
generated closing pressure. A peristaltic pump was used to maintain a constant fluid level
in the reservoir, and glycerol flow through the valve was monitored over a 24 hour period.
Figure 1.12: Glycerol £iow through valve using a 4.6mmHg supply pressure and
4.3mmHg osmoticaily generated closing pressure.
The results of this test show that the glycerol flow through the valve steadily increased as
a function of tirne (figure 1.12), supposedly because water moved from the dextran
solution in the cuff into the outer charnber, permitting the collapsed tube to open.
However, at the end of the 24 hour period, visual observations of the coiiapsible tube
indicated that it was completelj open. This was considered uncharacteristic for a tube
under pressure. It was believed that an improper seal caused dextran to leak into the outer
charnber. This elirninated the osmotic pressure difFerence and aliowed a very modest
supply pressure to fully open the coiiapsible tube.
Drainage implants for treatment of glaucoma provide a low-resistance outflow route for
aqueous humor, by-passing conventional drainage pathways. Drainage implants are
classified as "non-restrictive" or "restrictive" depending on their flow control mechanisms.
Non-restrictive flow devices rely on the formation of a filtration bleb to control outflow
and prevent over drainage. Short-term post-implantation hypotony is a chronic problem
with this class of devices, as the filtration bleb requires several weeks to form. Restrictive
flow devices incorporate flow-resistive elements, such as slit valves or conductive
matrices. Although they represent an improvement over non-restrictive devices, none of
the currently existing restrictive flow devices have exact an opening and closing pressure,
and hence they are unable to accurately control post-impiantation IOP.
28
Sit (1996) proposed a glaucorna drainage valve based on the concept of a Starling resistor.
The valve consisted of a flexible tube that is coilapses under pressure. Under low Reynolds
number conditions, the valve was hypothesized to remah closed until the fluid supply
pressure increased to values greater than the pressure used to cuiIapse the flexible tube.
This hypothesis was proven using glycerol as the supply fluid and an extemal pressure
source to coiiapse the flexible tube. In order to make the valve self-contained, the extemal
pressure source was replaced by surrounding the collapsible tube with a semi-permeable
membrane "cuff". The cuff was filied with a dextran solution, and the surrounding
chamber was filled with water. The closing pressure for the collapsible tube was generated
by the osmotic pressure difference between the dextran solution and water. DitFcuIties in
obtaining a watertight seal for the semi-permeable membrane prevented any usehl
experiments from taking place.
Due to the encouraging results obtained using an extemal source to set valve closing
pressure, Sit recommended fbrther investigation using an osmotically generated pressure.
However, modifications to the valve design would be necessary prior to any funher
testing. In particular, a better seal for the serni-permeable membrane was required.
It was hyphesied that :
1. A large sale model of an osmotically controlled valve, such as the one proposed in an
earlier study by Sit (1 996). would: (i) consistently open and close at suppiy pressures
above and below the osmotic pressure difference, respectively and (ü) respond to
pressure changes relatively quickly.
2. A correlation of operating characteristics could be made between a large-scale model
and an actual-sde valve, by applying the principles of similitude.
The long tenn goal was to develop a glaucorna drainage valve whose operating
characteristics are more prediaable than those of currently existing implants. In
particular, we wished to develop a valve that has consistent opening and closing pressures.
2.2 SPECIFIC OBJECTIVES
The initial project objectives were :
1. To construa and test a large-sale m ode1 of an osmoti c pressure controlled valve,
based on Sit's (1996) design and recomrnendations, in order to: (i) ver* that the
A Novtl G l u a m a Dnmige Vaive
30
valve has a consistent opening and closing pressure and (ü) that the valve was capable
of controllhg outflow and maintainhg a constant supply pressure.
2. To apply the prhciples of similitude in order to scde the valve to dimensions
appropriate for implantation. This would require the establishment of dimensionless
numbers using the Buckingham II Theorem. Analysis would focus on the factors
affecthg the buckling of the collapsible tube and resistivity of the serni-permeable
membrane to osmotic fiow.
3. To consmict and test an actual-scale prototype of the valve in order to (i) ver@ that it
had a consistent opening and closing pressure and (ii) enmre that it demonstrated an
adequate dynarnic response' to the characteristic pressure changes of the eye
Unforeseen difficulties due to the slow filtration properties of the dialysis membrane
impeded progress of the project considerably. The project become focused on completing
the first objective, which was segmented into the following tasks:
'For the purposes of this study, "adequate dynamic response" refers to valve opening and closing occurring within an hour as opposed to within a day. An instantaneous dynamic response is not desirable, as the device would respond to short term pressure disturbances, such as rubbing the closed eyelid. A response time of several hours to a day is considered too long, since prolonged exposure to an elevated IOP is thought to be harmfûl to the optic nerve. Therefore, a response time of 1 hour was arbitrarily selected as an adequate value.
31
1. To ident* problems encountered by Sit (1 996) and construct a bench-sale prototype
(Valve 1) of an osmotic pressure controiied valve based on the original design and
recomrnendations.
2. To v e e the operation of Valve I by recreating the tests performed by Sit (1 996)
using extemally and osmotically generated valve closing pressures.
3. To i d e n e aspects of Valve 1 which could be modifieci to facilitate a reduction in
response the , and design a second prototype (Valve II) irnplernenting those changes.
4. To test the operating characteristics of Valve II to ensure that the t h e response had
been adequately reduced.
5. To perform dynamic tests simulating in silu flow conditions and evaluate Valve II's
abiiity to control outnow and maintain a constant supply pressure.
Following a review of Sit's (1996) report, several problems with the original prototype
design were identifid It was suspected that the seal between the dialysis and collapsible
tubing, made with a combination of rubber bands and wire wraps, was not watertight. In
addition, the lack of suppon structure made the dialysis tube awkward to fil1 with the
dextran solution. The rectangular gasket between the chamber waüs and lid did not
provide a reliable seal.
These problems were rectified in the design of the new prototype, Valve 1 (figure 3.1).
Components that worked well initiaily were incorporated in the new design and included:
coliapsible tube (latex penrose tubing, '/;" diameter, 12" length, #20414-050, Baxter
Medical Supply, Mississauga ON)
regenerated cellulose semipermeable membrane (1 2000- 14000 MWCO didysis tubing,
45mm flat width, #2 1 - 1 52-8, Fisher ScientSc, Nepean ON; manufactured by
Membrane Filtration Products, San Antonio TX)
connectors and vents (Nalgene@ barbed bulkhead fittings, %" and '/." diameter,
# 16225-23 2 and # 1633 1 - 1 02 VWR Scientific, Mississauga ON )
Machining of plexiglass components was done in the University of Toronto, Mechanical
Engineering Machine Shop by Mr. Paul Kovar. Complete shop drawings are located in
Appendk A.
Figwe 3.1: (a)Valve 1, the new bench scale glaucoma drainage valve prototype, (b)
labeiied Valve 1 fiont view.
Aithough it was desirable for the dialysis membrane and cdapsible tubing to be
configured as concentric cylinders, there was no need for them to be joined. Therefore,
rather than aîtach the membrane directly ont0 the coiiapsible tube or bukhead fitting,
A Novel Glauanna haimgc Valve
34
cylindrical plexiglass rings were mounted on the inner s d c e s of the chamber endplates,
centred around the bulkhead fittings. The dialysis membrane was pulled over a tightly
fitting rubber O-ring placed around the plexiglass ring. A proper seal was achieved by
securing the membrane with a stainless steel, adjustable hose clamp (CHX-20, SmaU Parts
Inc., Miami Lakes FL). To prevent damage to the membrane, the outer edge of the
plexiglass ring was slightly contoured, and a piece of thin foam was placed around the
inner surface of the hose clamp. DEerent views of the new seal configuration are
illustrated in figure 3.2.
Figure 3.2: Vdve 1 diaiysis membrane seal (a) in situ photograph; @) cross sectional
schematic.
During the assembly of Valve I, it was noticed that the barbed ends of the bulkhead
fittings did not extend beyond the sealing rings. This made the process of pulling the
collapsible tube over the buIkhead f i h g extremely dficult, as there was very Little space
between the bukhead and the sealing ring. To r e m this oversight, the buikhead fittings
35
were extended with short pieces of rigid pipe. The cdapsible tube was originally meant
to be 12" long, as in Sit's (1996) valve. As a result of this last-minute modification, the
collapsible tube was shortened to a length of 9".
Uniike Sit's (1996) prototype, both ends of the dialysis membrane in Valve 1 could be
cunveniently sealed before the dialysis membrane was filied with dextran solution.
Through one of the chamber endplates, narrow tracts extended fiom the outside the
apparatus, and led into the space between the buikhead opening and the plexiglas ring
(figure 3.3). Flow through the tract was controiied by plug valves (HV3-3, #86727,
Hamilton Company, Reno NV) with luer connectors ($35030 and #3503 1. Hamilton
Company, Reno NV) . Dextran solution was injected into the dialysis membrane ushg a
luer-lock syringe attached to one of the valves, while the other served as an a i r bleeding
port. During experiments, pressure was monitored by attaching a pressure iine to one of
the ports.
Figure 3.3: Cross-sectionai M m of
Valve 1 chamber endplate,
iüustrating nIling ports and tracts.
Sit's (1996) experiments to determine the characteristics of flow through a coUapsible
tube using an extemally generated cioshg pressure were duplicated. This was to ensure
that nothing significant had been overlooked in the valve redesign or in recreating
experimentd conditions. Ornitting the dialysis membrane in the valve setup and using a
column of water to fbc the valve closing pressure (P,), the collapsible tube was co~ected
to a fluid supply reservoir of variable height (refer to figure 1.8).
P, was fixeci at 7.4 d g (arbitrarily chosen, Sit. 1996) and water flow through the
coiiapsible tube was recorded as the height of the supply reservoir was incrementaüy
changed. Pc was changed to 14.7 mmHg (arbitrarily chosen, Sit, 1996) and the
experiment was repeated. This set of experiments was then repeated using glycerol as the
supply fluid. Results are given in section 3 -3.1.
With operation of Valve I verified, the externaiiy generated P, was replaced with the
osmotically generated pressure source. The dialysis membrane was attached to the sealing
rings on the chamber endplates as previously described (section 3.1.1). and f led with a
0.224m.M dextran solution (industrial grade 298 000 MW, Sigma Chemicals, St. Louis
MO). The concentration of O.224mM was arbitrarily chosen by Sit and has no apparent
A Nowl Giuicwi Dmkgc Valve
38
significance. The chamber was fïiied with distdied water and a p d a i i y filled fume1 was
left co~ec ted to the chamber vent. The fume1 was Ieft open to the aîmosphere and the
water in it prevented air from being drawn into the chamber. The supply fluid was
glycerol, as used in Sit's experiments, to provide the low Reynolds number flows typically
passing through a glaucoma drainage implant.
Sit's experiments to test the flow characteristics of the osmotically controiied valve were
repeated. According to the van't Hoff equation, a dextran concentration of O.224m.M
produces an osmotic pressure gradient of 4. ImrnHg. The valve's P, was therefore
considered to be 4. lrnrnHg and the glycerol reservoir was positioned to provide a P, of
4.6mmHg (arbitrarily chosen by Sit). Since PB was greater than P , the valve was expected
to open. Results are given in Section 3.3.1.
3.3 RESULTS FOR VALVE 1 TESTING
3.3.1 COUAPSIBLE TUBE BEHAVIOUR
The results from these teas, where closing pressure (P3 was generated by an extemal
source. were compared with those obtained by Sit (1996). The fita set of tests, where the
supply fluid was water, are compared in figure 3 .S. When plotted on supply pressure (P,)
vs flow graphs, the data sets obtained by Sit and fiom Valve 1 do not overlap.
39
A difEerence between the fiow resistance of two valves was observed by cornparhg the
dopes (APJAQ) of the plots in figures 3 3 a ) and (b):
Assuming laminar flow conditions, flow resistance in tubes for viscous fluids is dependent
on fluid dynamic viscosity, tube length and tube cross-sectional area, as given by:
The tube length and cross sectionai a m were not necessarily exactly the same for both
vaives, and during both tests. It was known for certain that the lengths of the collapsible
tubes were difFerent, as discussed in section 3.1.1. The collapsible tube in Valve 1 was
only 9", compared to Sit's 12" tube. However, the difFerence in tube lengths did not
contnbute sigruficantly to the flow resistance difference, because resistance was greuter in
Valve I than in Sit's valve.
Another factor that might explain the dierence in flow resistances is the degree of
longitudinal tension in the coliapsible tube, since this would have an effect on
its cross-sectional area when open. A collapsible tube that was stretched lengthwise
would not open as much as a more relaxed tube. It was suspected that Valve 1's
collapsible tube had a higher degree of longitudinal tension than the tube in Sit's valve,
accounting for the higher flow resistance of Valve 1.
Figure 3.5: Cornparison of the flow characteristics of Sit's (1996) prototype and Valve 1
when water supply pressures were incrementally increased and closing pressures were
set with an extemal pressure source; (a) P, = 7.4 mmHg (b) P, = 14.7 mmHg.
41
Regardless of the flow resistance dieremes, both valves displayed a dramatic increase in
water flow through the tube when P, rose above P,.
O 1 2 3 4 5
glycerol oufflow [mUsJ
O Sit 199ô Valve I
O 1 2 3 4 S
glycerol oufflow (mus]
Figure 3.6: Cornparison of ihe flow characteristics of Sit's (1996) prototype and Vaive 1
when glycerol supply pressures were incrementaiiy increased and ciosing pressures were
set with an estemal pressure source; (a) P, = 7.4 m g ; (b) P, = 14.7 mmHg.
42
The second set of tests, where the supply fluid was giycerol, are compared in figure 3.6.
Both valves exhibited the sarne general behaviour in that neither of hem aiiowed glycerol
flow through the collapsible tube until P, rose above P,. This is the classic behaviour of
low Reynolds number flow through a Starling resistor, and is particularly noticeable Ui
figure 3.6@), where the value of P, was 14.7rnmHg.
A dserence between the flow resistance of two valves was observed by comparing the
slopes (AP jAQ) of the plots in figures 3.6 (a) and @):
Since & is less than %, the coliapsible tube length difEerences may have made a more
significant contribution to the difference in flow resistance than in the previous w e .
However, the tube lengths alone did not account for the different flow resistances, as
illustrateci by the ratio:
In this case, it was suspected that Sit's collapsible tube had a higher degree of longitudinal
tension than the tube P Valve 1. The resulting diierence in the tubes' cross sectionai
areas would account for the remaining 25% diifference in flow resistances.
Glycerol flow through Valve 1 dernonstrateci the characteristic behaviour of a cdapsible
tube under pressure, as described by the previously discussed relationships:
The outflow behaviour of Valve I when the extemally generated P, was replaced with an
osmotic pressure source is discussed in the next section.
3.3.2 Os~onc VALVE BEHAVIOUR
As indicated in figure 3.7, the behaviour of Valve 1 and Sit's prototype differed greatly
when P, was generated osmotically. Sit's valve appeared to open continuously over
several hours, while Valve 1 did not open at dl. In fact, no flow passed through Valve 1
until the glycerol supply level was raised to over 2OmmHg and even then ouMow
remained relatively constant at less than O. 1 mL/min.
Figure 3.7: Cornparison of the fiow characteristics of Sit's (1996) prototype and Valve 1
when glycerol supply pressure was frxed at 4 . 6 d g . A nominal valve closing pressure
of 4.1mmHg was generated osmoticaily with a dextran concentration of O.2NmM.
The faiure for Valve 1 to open at a P, of 4.6mmHg reuiforces Sit's theory that a dextran
solution leak slowly eliminated the osmotic pressure gradient in his valve, reducing P, to
almoa nothing. This explained the dinerence in the behaviour of the two prototypes, but it
did not account for why Valve 1 barely opened even when P, was four times pater than
In the next chapter, possible explanations for the fdure of Valve 1 to open are discussed.
There were three plausible explanations for Valve 1's failure to open:
învaiidity of the van? Hoff osmotic pressure equation, for dextran at the concentration
used in the study,
occurrence of concentration polarkation at the membrane surface, or
extremely low penneability of the dialysis membrane, resulting in long equilibration
Each of the possible explanations for Valve 1's failure to open were investigated.
Procedures and results for each investigation are discussed in the following three sections.
Sit (1996) assumed that the van't Hoff simplification of the osmotic pressure equation was
valid. Since Valve 1 did not open when P, was more than four times greater thm the
calculated osmotically generated P, the question arose as to whether or not the
calcuIation could be trusted.
In order to masure the actual osmotic pressure within the dextran cue a crude
manometer was conaructd. A pressure he, with luer connectors on both ends, was
connected to one of the Hamilton plug valves and run vertically dong the waii beside
Valve 1 (figure 4.1). Osmotic pressure, AII, was taken as the dierence in fluid height
between the dextran in the pressure line and the water in the fùmel.
A Novcl Glutcomp Iiniatgc Vaive
Figure 4.1 : Manometer configuration
The plug vaive was opened and the system was lefi overnight to ensure that equilibrium
wouid be achieved. The following day, the osmotic pressure was measured to be
42mmH& a full order of magnitude larger than the 4.16 mmHg predicted by the van't
Hoff equation. It was concluded that the van? Hoff equation underestimated the osmotic
pressure, and the non-shplified osmotic pressure equation,
would have to be used. Patently, the 0.224rnM dextran solution is not considered ideai
and solute-solute interactions can not be neglected.
A literanire review revealed several studies in which the vinal coefficients (A, values) of
de- (500 000 MW) dissolved in water were determined. Coefficients for dextran of
47
-300 000 MW were unavailable. Using three sets of coefficients, the osmotic pressure
values in table 4.1 were obtahed for a 0.224 mM solution.
Table 4.1: Cornparison of publistied values for virial coeficients (A,) of 500 000 MW
dextran and the calculated osmotic pressure values (n) for a 0.224mM solution.
Wijmans (1 985)
Although these values of II differ significantly, they illustrate that the measured osmotic
pressure of 42 mmHg is within a reasonable range, and that indeed the van? Hoff equation
is insufficient for the purposes of this project. Due to the lack of established Wial
coefficients for 300 000 MW dextran, the manometer was deemed the most reliable
method to determine the osmotic pressure for the experiments during this study.
Another possible cause of the delay in response tirne is the occurrence of concentration
polarkation. When water is forced, by an applied pressure, through a semi-permeable
membrane from a solution of high to low solute concentrations, there is a possibility of
solute accumulation at the membrane as show in figure 4.2. This creates a localized area
of very high solute concentration. Due to this high concentration gradient at the membrane
0.0867 2.98
A,---
89.9 34.5
surface, water will more readily move back across the membrane in response to the
osmotic gradient, in opposition to the applied pressure. The result is a decreased net flux
of water across the membrane. If concentration polahtion occurred in Valve 1, it would
take longer for water to lave the dextnui cuff and thetefore longer for the cdapsed tube
to open.
dextran O
a m a
Figure 4.2: Water flux (mm) amss a semipermeable membrane. (a) Without appIied
extemai pressures, net water flux is into the demm solution. (b) Applied fiuid supply
pressure on the dextran side of the membrane reverses water flux and creates localized
area of hi@ dextran concentration at membrane. so that net flux is reduced by water
rnoving back across the membrane in response to the osmolarity difference.
With the same experimental set up used for the osmotic valve behaviour tests, a test to
determine the extent of concentration polarkation effécts on Valve 1 was perfonned. The
height of the glycerol supply reservoû was incrementaliy increased, then decreased, and
the correspondhg outflow measurements were recordeci at each step. Minimum supply
49
pressure was 18mmHg, at which glycerol outflow just began, and maximum supply
pressure was 6OmmHg, at which point the catalogues used to raise the reservoir became
unstable.
The hypothesis for this experiment was that due to concentration polarization effects, the
coiiapsible tube in Valve 1 would close more easily than it opened. The hysteresis in the
data (figure 4.3) indicated that at any given supply pressure, glycerol outflow was Iower as
P, increased than when it decreased.
O S 10 15 20 25
average glycerol o M o w (mumin]
Figure 4.3: Resuits of test to determine conœntmtion polafization effects on glycerol
flow thmugh Valve I. Glycerol outflow measurements were recorded as supply pressure
was incrementally increased from 18 to 6OmmHg. then incrementally d e c d
50
This higher resistance to glycerol flow as the valve opened confirxned the hypothesis that
concentration polarization was occurring. However, this test was only designed to be a
crude indicator of concentration polarization. The degree to which concentration
polarization actuaily affecteci the operation of the valve was observed more ciearly in a
later experiment, discussed in section 4.3.3.
4.3 MEMBRANE PERMEABILITY & EQW~LIBRIUM TIME CONSTANT
In earlier discussion on Valve 1's P , it was established that when fluid P, was increased to
a level above P, the coilapsible tube opened. This occurred instantaneously when P, was
generated by an extemal source. When P, was generated by the osmotic pressure cuff. it
took time for the water leave the cuff before the collapsed tube opened. If the pressure
inside the cuff were to be measured as a fùnction of time during this process. it would
resemble the curve in figure 4.4.
Figwe 4.4: Hypothesized relationship between pressure inside the dextran &and time
at a constant value of P,, greater than P,.
51
Note the unsp&ed (ie. unknown) t h e scale. The unanswered question was: At what
point in t h e will enough water have crossed the membrane in order to estabîish
equiiibrium between the fluid supply pressure and osmotic pressure? (For the purposes of
this project, the term "equilibrium" refers to the point at which the pressure within the
dextran cuff no longer changes with respect to time.)
It was dficult to plan experiments with the valve since the uctual equilibrium t h e scale
was unknown. Therefore, the valve was mat hematically modelled in order to d e t e d e a
theoreticd equilibrium time scaie. However, before the valve was modelied as a whole
unit, the focus was piaced on the flow across the dialysis membrane.
4.3.1 MODELUNG FLOW ACROSS THE DIALYSE MEMBRANE
Intuitively, the length of tirne required for the valve to reach equilibrium should be
dependent on the rate at which water crosses the dialysis membrane. The water flux
should be dependent on surface area of the membrane, the net potential diifference driving
the water across the membrane, and membrane permeability. The "net potential
difFerenceW can be defined as:
where Apt is actual pressure difEerence between the dextran cuff and outer chamber at any
given point in tirne, and AII is the osmotic pressure difference betwem the dextran cuff
and outer chamber. At equilibrium, Ap, is equal to AIL
The equation goveming volume flow across the membrane is therefore:
where dV1dt is the volume flow rate across the membrane [ d m i n ] , k is membrane
permeability [mUrnin/m2/rnmHg], A is the membrane surface area [m2], and P, is the time
dependent net potentiai dserence [rnmHg].
The volume of the opened collapsible tube, and thus the volume of water required to be
displaced from the cuff, could be calaiiated using the valve's dimensions. Similarly, the
surface area of the membrane could also be caiculated. The net potentiai dEerence over
tirne could be measured using the manometer. A membrane permeability value of 1.5-
1 .7mvmin/m21rnm~g was provided by the manufacturer (Membrane Filtration Products,
San Axitonio TX), however this value was likely determined using distilled water only.
Therefore, the quoted permeability value would neglect the realities of solute interactions
with the membrane, such as pore clogging and concentration polarkation.
In order to ensure the accuracy of any calculation for the equilibrium time scde, the actual
membrane permeability, under Valve 1' s experimental conditions, was verified ushg the
procedure described below.
The actual or effective membrane perrneability was determined by connecting the
manometer and aiiowing the system to reach equilibrium overnight. In the rnoming, a
6mm diameter pipette was attached to the top of the manometer and filled with dextran
solution. Throughout the day, water moved across the dialysis membrane f?om the
dextran side to the water side.
The fluid level in the pipette, which led into the dextran tue was monitored as a fùnction
of t h e . Knowing the cross-seaional area of the pipette, the water flux across the dialysis
membrane was calculated. Taking the dope of a flux vs pressure plot (figure 4.5) and
dividing by the surface area of the membrane resulted in a permeability value of
0.03 137mUmin/rn2/mrnH& which is fifty times less than the value quoted by the
manufacturer.
Figure 4.5: Determination of the water fiw across 14K MWCO diatysis membrane as a
function of driving pressure. Caiculating the slope of this curve and factoring in
membrane mfhce area yields a permestbility value of 0.03 137mWmia/m1/mmHg, whch
is specific to Valve 1 parameters.
The next sep was to mode1 the valve as an entire unit.
The response time of the valve is determined by how quickly the collapsible tube opens
and closes. In order for the volume of the coilapsible tube to change, fiuid must lave the
dextran cuff. As illustrated in figure 4.6, there are two fluid pathways in and out of the
dextran cuff. (i) across the dialysis membrane, and (fi) through the manometer tube.
Figure 4.6: Volume changes in the dextran cuE V, is the volume of giycemI in the
coiiapsibie tube, V, is the volume of water crossing the dialysis membrane. and Vh k the
volume of dextran solution in the manometer tube.
Treating the dextran cuEas a control volume that does not change in shape, and the
dextran solution as an incompressible fluid, the law of conservation of mass applies and
the net volume change mua be zero:
The volume of dextran in the manometer tube is a hnction of its cross-sectional area, &;
and the height of the fluid, h. Height can be expre~sed in terms of the net potential
As discussed in the previous section, the volume of water crossing the dialysis membrane
is a funaion of membrane permeability, k; membrane surface area. 4; and Pt.
Combining Q. 0, and O produced an expression for the change in collapsible tube volume
in terms of net potential difference.
It was assumed t hat Pt changed exponentially with respect to tirne by a function of the
Where r is the equilibrium t h e constant and indicates how quickly the exponential
fiindon decays. The assumption regarding the nature of P, was verified by plotting
experimental data (section 4.3.3) on a log-hear graph, on which the data points formed a
straight h e . Substituthg Pt into 0, and intergrating from time O to =, we obtain:
Where AV, is the change in volume of the collapsible tube as it opened or closed, and P,
is the initial net potential difference. Rearranghg the equation for AV, produced an
expression for the valve's t h e constant, 5:
Parameter values for Valve 1 (opening) were:
The value for AV, was an estimate, based on the observations of the "tapered" shape of
the wllapsible tube when opened. The value for P, was initial net potentiai difference
58
measured when the glycerol P, was raised to 8OmmHg. This P, was generated by placing
the supply reservoir on the shelfabove the workbench. According to these parmeters, a
tirne constant of 1200 minutes was expected.
For scaling purposes, it would be important to know how r would change if the
manometer was removed. Setting 4 to zero resulted in a r of 1160 minutes, a 4%
reduction.
4.3.3 ACTUAL EQUILIBRATION TIME CONSTANT
Although the above calculation may have accurately described the system, the actual
response time was still unknown. Therefore, with the set-up depicted in figure 4.7,
experiments were performed to determine exactly how long the valve does take to reach
Figure 4.7: E.uperimental apparatus for &termining system response t h e .
59
The est step of the experiment was to measure the initial cuff pressure (Am, which was
equal to the valve's osmotically generated closing pressure. At tirne zero, the glycerol
reservoir was raised from its initial OrnmHg position to give a P, of 80 rnrnHg. A
centrifugai pump (Albany Pump Company, Toronto ON) ran continuously to rem the
reservoir and a standpipe maintained a constant fluid level for the duration of the
experiment. Changes in the cuff pressure (ApJ according to the manometer were recorded
as a funaion of time until steady state was reached. The reservoir was then lowered to its
initial position and cuff pressure changes were again recorded until steady state.
Results from three tests to determine Valve 1's response time are show in figure 4.8. Al1
three tests were perfonned without replacing the dialysis membrane or dextran solution.
O 500 t 000 1500 2000 2500 tmie [min]
Figure 4.8: Cornparison of Valve 1 tirne response data h m successive experiments
using the same dialysis membrane, 0.224mM dextran fiIling solution. and glycerol P, of
80mmHg. The descendhg curves represents net ptential merence, Ab-II). while the
vaive is opening, while the ascending curves represents A@$) whik the valve is
closing. The value of AIi was the osmotic pressure gradient measured More each
experiment; and Ap, was the timedependent pressure dinerenœ between the dextran
d a n d the outer chamber. Order of experiments: 0,O.X.
As discussed in section 4.3.2, A(p,-II) was expected to change exponentiall y with respect
to time. This hypothesis was confirmed when the data for each experiment was
transformed to a log-linear sale by ploning ln[A@JI)] vs tirne (figure 4.9). The data was
fit with a regression curve and expressions for A(pt-TZ) as a funaion of time, were
obtained in the fom:
The tirne constant, t = -l/k provided an indication of the equilibration the s d e . Time
constants were the buis for cornparison between the experirnents discussed in this section
and in the following sections.
O 100 200 504 400 500 600 ame [min]
O 200 400 600 800 lm Wnc (min)
Figure 4.9: hg-iinear regression curve fit for Valve 1 time respome data. Daia was h m
experiment "O" from figure 4.8; (a) valve opening; (b) valve closing. Note that
absolute values of A(R-T]n were used for convenience.
62
A complete set of regression cuwe calculations is given in Appendix C, while table 4.2
contains a surnmary of the tirne constants calculated for ail three experiments.
Run #
1 (0)
2 (0)
Table 4.2: Cornparison of system time çonstants for Valve 1. Values are mean * SEM.
3 (x)
Tirne constants ranged fiom 864-1245 minutes when the reservoir was raised (opening the
valve) and tiom 300-1000 minutes when the reservoir was retumed to its initial level
(closing the valve). Observe that the range of r was higher during valve opening, as
water left the dextran cug than during valve closing, as water returned to the cuE This
indicated that there was higher flow resistance at the membrane when water ieft the cuff
than when water retumed to the cuff This phenornenon indicated that concentration
polarization was occuming.
TIME CONSTANT [min]
1245k30 1 3OOi7
On average, the expenmentally determined time constants were lower than the s of 1200
minutes predicted by the mathematical model. R e d that an assumption was made
regarding the volume of the collapsible tube when opened. If the value for AV, was
reduced by 6mL. the estimateci s would be 890 minutes. Since the AV, value used in the
A N m l Glurcoma Dramagc Vdvc
Valve Opening
87W2 1
864k5
Valve Closing
784i10
1 OOm45
63
original calculation was crudely determined, the probability for overestirnation was hi&.
It was concluded that the equation for r would accurately mode1 the system, if a good
estimate of AV, was avaiiable.
It was interesting to note that in successive experiments the time constants increased
significantly. The increasing time constants suggea that the membrane's pores becarne
blocked over time, decreasing pemeability and increasing flow resistance across the
membrane. The exception to this general trend was the 300 minute valve closing tirne
constant for the last run, for which there was no obvious explanation. The entire group of
experirnents was performed using the sarne piece of dialysis tubing, and took a week to
complete. It was possible that by the end of that period the membrane had lost some of its
flow resistance, due to some form of 6bre degradation. Dialysis membrane is intended for
short-term, one-time use, and the week of stress may have taken its toll.
It was very encouraging to observe that, in response to a change in P,, the pressure in the
dextran cuff maintained a constant P,, ifgiven enough time to equilibrate. However, an
equilibrium time constant of 20 hours was considered to be too large a response time sale
for a giaucuma drainage valve. Certainly, the long equilibration time scale made dynarnic
testing of the valve too impractical to attempt.
Membrane fouling over the , resulting in increased time constants, remained an important
A Novci Gluicoau hinage Vaive
64
issue that must be addressed before an osmotically controlled glaucoma valve can be
developed much tiirther. However, for the purposes of this project, the primary issue was
the unacceptably long tirne to reach steady state with a fresh membrane and how to
decrease it. The steps taken to ammplish this objective are discussed in the next chapter.
In order for the osrnotically controlled valve to be considered feasible, the response t h e
had to be reduced. Factors affecthg the response time of Valve 1 were identified, and
where possible, modifications were made in order to reduce response time. In this
chapter, factors affecting response tirne, modifications to Valve 1, and testing of the
modified valve are discussed.
Three factors were identified that would have the most significant effect on the response
tirne: (i) the permeability of the membrane, (ii) the filtration surface area, and (üi) the
volume of fluid required to cross the membrane.
Membrane pemeability is afFected by pore size and pore density. M e r contacthg several
companies it was determineci that 32 mm flat width (dry) dialysis tubing is not available
with a higher MWCO than 14 000 or with a higher pore density than
2 . M OP poresrinch'. Since this was the tubing currently being used, it b e r n e obvious
that in order to change the membrane, an entirely new apparatus would have to be built.
Increasing the fütration surface area could be accomplished by lengt henuig the didysis
tube andor widening the diameter. Either of these changes would require the
construction of a new apparatus.
Since it was desirable to avoid designing a new valve if possible, the focus shifted to
reducing the volume of fiuid to be displaced across the membrane. By shortening the
coilapsible tube, a smaiier volume of fluid would need to be displaced in order to open and
close the valve. This was easily accomplished by extending the buikhead fittings with
pieces of plastic tube and trimming the coUapsible tube (figure 5.1). The volume of the
collapsible tube when open was reduced fiom 22.5mL to 6.5rn.L. During the experiments
performed in the previous chapter, the dialysis tubhg was observed to stretch slightly
when the glycerol reservoir was raised. Therefore, a wire mesh cage was placed around
the dextran cuEin an attempt to reduce deformation and variations in volume.
Figure 5.1 : Modifications to Valve i, a shortened coliapsible tube with wire mesh cage
around the dialysis membrane.
67
The estirnated time constant for the modified valve was calculated to be 336 minutes.
The experhents to determine tirne constant that were discussed in section 4.1.3 were
repeated with the modified valve.
After shonening the collapsible tube in Valve 1, the response time data show in figure 5.2
was obtained. Time constants were calculated for experiments run on the modified valve
(regression analysis given in Appendix C), and compareci with the previously obtained
time constants (table 5.1).
Figure 5.2: Cornparison of Valve 1 (3" collapsible tube) response tirne data h m
successive experiments using the sarne diaiysis membrane, 0.224mM dextran filling
solution. and glycerol P, of 8OmmHg. The descendhg curves repmnt net potential
difference. A@-m. while the valve is opening, the ascending cuve repmnts A@I-n)
while the valve is closing. The value of A I i was the osmotic pressure Merence
rneasured More each experiment; and Ap, was the tirnedependent pressure Merence
between the dextran &and the outer chamber. Order of experiments: O.O.X. Note
ihe unavailabte data for the "valve closing" sîages of the first two nrns. Due to the long
duration of the e.riments, the apparatus occasionaily had to be lefi unattended.
During îhe first two nuis, N~culties with the glycerol supply pump resulted in the
depletion of the glyceml supply. With P, reduaxi to zero. the valve closed before anyone
renrnied.
A Novrl Glaucoma üninage Valve
I TIME CONSTANT [min] II
Table 5.1: Cornparison of system time constants for Valve 1 with the original 9" and
Run #
1 (0)
2 (0)
3 (x)
shortened 3" mllapsible tubes. Values are mean i SEM
For the first two mns with the modified version of Valve 1, the experimentally detennined
time constants were consistent with the mathematical model's prediction of 336 minutes.
As with the original version of Valve 1, the time constant for the last run was significantly
higher than the values obtained for ht two. This was again suspected to be a result of
clogged pores in the dialysis membrane.
Data not available
Although shonening the tube reduced the t h e to reach steady state, the timescale was still
considered too large. Therefore the valve was redesigned to incorporate a different style
of membrane with a higher permeability. The new design is discussed in the next chapter.
C
Valve Opening
9" Tube
87&2 1
864*5
1 245+1 O
Valve Closhg
3" Tube
334k5.6
36W7.0
63 7 î5 .5
9" Tube
784*1 O
1 OOW45
30W7
3" Tube
* *
I
625*5. 1
A new prototype, Valve II (figure 6.1) was designed in an effort to hrther reduce the
system response t h e . Sipnificant changes included:
shorter collapsible tube and larger membrane surface area
100 000 MWCO ultrafiltration membrane
membrane support plates
Plexiglass wmponents for Valve II were constnicted in the MechMcal Engineering
Machine Shop by Mr. Jeff Sansorne. Complete technical drawings and specifications are
located in Appendk B.
pl4 WdV-
cuiiapiible tube
1
o-nng
uMMm&rae
Figure 6.1: Valve II (a) photograph; (b) Iabeiled cross-sectional view
6.1.1 COLLAPSIBLETUBE L E N G ~ 6 MEMBRANE SUWACEAREA
The length of the coilapsible tube was shortened fiom 9" to 1%". By trial and error, it was
determined that a length of 1" is the shortest tube that would fulIy coiiapse. The additional
%" was added for safety.
Since the size of dialysis tubing had to be changed. it was also decided to switch to a flat
(sheet) style of membrane. As a result, the increase in membrane surface area was
constrained only by the size of membranes available. The type of membrane selected (see
next section) was sold in 18"xl8" sheets. An 8%" filtering surface diameter was selected
so that four filters per sheet could be obtained. The actual useable diarneter of the fiiter
was 7%". as the outer %" was compressed between the two halves of the apparatus.
The changes in collapsible tube length and filtration surface area between Valve 1 and
Valve II were expeaed to significantly reduce response time. The anticipateci difference
was estimated using the equilibrium time constant equation, derived in section 4.1.3:
If membrane permeability, manometer tube area, and initial net potential difTerence were
the same for both Valves I and II, the diEerence in time constants could be compared by
the ratio:
Where :
& = 0.02054 m2
%O= 0.0285 m2
AV,,=-22.5 rnL
c AV,=-2.4mL
Comparison of the two equilibrium tirne constants suggested that more than a ten-fol(
reduction in response tirne could be expected due to Valve II dirnensional modifications
alone, i.e.
hcreasing the membrane penneability reduced the estimated time constant for Valve II
even fiirther.
Rat sheet filtration membranes are available with wide variety of pore sises. To decrease
water flw resiaance, and fiirther d u c e Valve II's response tirne, a 100 000 MWCO
ultrafiltration membrane made fiom polyethersulfone was selected (HFK 13 1, Koch
74
Membrane Systems, Wiimington., MA). This pore size would allow a greater water flux
than the previously used diaiysis membrane and would continue to reject 300 000 MW
Dextran.
The clean water flux for this membrane was listed as 200-300 gdonslsq.ft./day, at a
constant pressure of 30psi. For a 0.224mM solution of 300 000 MW dextran, the
manufacturer estimated a flux of 30-50 gallonslsq.tt./day at 30psi (Rice, 1999). This flux
range was converteci to a permeability range of O.OS4543-0.0909O5 mL/min/m2/rnmHg.
For safety, the h u m u m value of 0.054543 mllmin/m2/mmHg was assumed for
dculations.
According to tirne constant equation, s is inversely proportional to permeability.
Therefore, the ratio of time constants between Valves I and II berne:
By changing valve dimensions and membrane pemeability, the time constant predicted by
the mathematical mode1 was reduced by a factor of 23.9 Le. the expected value of s was
lowered nom 1200 minutes to 50 minutes. Note that the prediaed r-reduction factor of
75
23.9 was calculateci using membrane permeability values determineci as water le$ the
dextran solution (valve opening). These k-values took concentration polarization eEects
into account, therefore a predicted r-reduction factor for valve closing could not be
accurately det errnined .
In Valve I, it was observed that increases in glycerol supply pressure caused the dialysis
membrane to stretch. As the membrane stretched, its volume increased, lowering the
pressure inside the cuf This would d o w fluid to pass through the coliapsible tube before
the valve was supposed to open, at supply pressures below P,.
If the colapsible tube opened due to membrane deformatioq it would behave Iike an
elastic valve. This would cause the valve to behave as a flow restrictor, rather than a true
valve with a specific opening and closing pressure. Thus, it was desirable for the
coliapsible tube to open not due to membrane deformation, but excluszvefy due to the
volume changes resulting fiom water crossing the membrane. This would only occur if
the volume within the membrane remained constant at all times.
In Valve Ii, the change in volume required to open the coilapsible tube was approhately
2.4rnL. With the 8%" diameter membrane, a O. lmm centre deformation translated into a
1 -2rnL change in volume (calculations in Appendix D). This was a significant cause for
76
concern, since a relatively indistinguishable deformation had the ability to provide 50% of
the volume change required to open the wilapsible tube. Hence, it was critical to ensure
that membrane deformation was negligible, which made some fonn of membrane support
structure necessary.
In order to prevent membrane deformation, the support stmcture had to be in contact with
the membrane. However, it was critical that this contact area be rninimized so as not to
interfere with fiow across the membrane, thus negating the benefits of possessing a large
filtration area. Patently, the support structure had to be made of a rust-resistant material,
strong enough to resist deformation under normal operating pressures up to 80 mmHg
(baseci on glycerol supply pressures used with Valve 1).
An ideal solution to the problem was to sandwich the membrane between disks of
perforated stainiess steel. Various types of sintered plates and perforated grates were
rejected because they were not avaiiable in the thicknesses necessary to resist deformation.
Eventudly, a pair of ?Af'-thick cast iron drain covers ( M g ' s Plumbing Supply, Toronto
ON) were deemed deformation-proof and treated with TremcladO rut-resistant enarnel.
Intended for floor drains, the covers have a fairly solid surface. To prevent undesirable
resistance to flow across the membrane, disks of expanded metal and fine mesh were used
to provide a 3mm clearance between the drain wvers and membrane (figure 6.2).
Figure 6.2: Membrane support structure. consining of cast iron drain mers and mesh-
covered expandeci metal grate (cross-sectional view).
6.2.1 DDC~~ZAN RETENTION
The solution used to fil1 the model's upper chamber was made 6om industrial grade
300 000 MW dextran. The industrial grade designation means that, although the avenge
weight of the molecules is 300 000, there could be a signifiant number of molecules
falling within in a wide range of values above and below the mean. There was concern
that the smaller dextran molecules would pass through the 100 000 MWCO ultrafiltration
membrane, whose pores were sigrufïcantly larger than those of the previously used 14 000
MWCO dialysis tubing. If a large number of molecules was able to pass through the
membrane, osmotic pressure values in Valve II would not be as high as those obtained
with Valve 1.
To determine ifenough dextran was behg retained by the membrane to provide a
sufficient osmotic gradient, an osmotic pressure test was performed. Valve II was
assembled, and the lower and upper chambers tiUed with distilled water and 0.224rnM
dextran solution, respectively. The lower chamber was co~ec ted to a reservoû of
distilled water (note: the water reservoir replaced the fùnnel used in Valve 1). A pressure
line was co~ec ted to one of the plug valves in the upper chamber, and extended vertically
to serve as a manometer, as with Valve 1. Both charnbers were lefi open to the
atmosphere and sat undisturbed ovemight. The following day, the osmotic pressure
readiig was taken and compared with the typical values obtained with Valve 1. Resuits
are presented in section 6.3.1.
Tests to detemine Valve II response time were performed as previously described
(section 4.1.3).
6.2.3 VALVE II DYNAMIC RESPONSE
Up to this point, valve performance was tested using a constant fluid supply pressure,
which is not representative of in situ conditions. In situ, a valve would be responsible for
the maintenance of a constant pressure (Le. IOP). The following test replaces the constant
presnue supply reservoir with a standpipe and a constantflow supply pump. The
purpose of this investigation was to detertnine whether or not Valve II could maintain a
constant "IOP" at severai d i f ren t supply flow rates.
The experimental apparatus (figure 6.3) for these dynamic response tests was based on
one used by Porter et d (1997), for in vitro testing of actual glaucoma drainage implants.
Comparing the apparatus to the eye, the glycerol flow provided by the pump represents
aqueous humor flow, and the pressure in the standpipe represents IOP. Obviously, Valve
II represents a glaucoma drainage implant, which controls "aqueous humor" outflow in
order to maintain a constant "IOP".
standpipe b 4 manorneter
UI 7'"
Figure 6.3: Experimental apparatus for testing dynamic response of Valve II.
The fira s e p was to record the initial AU. Since the connective tubing and standpipe
were initially empty, P, was OmrnHg and the coUapsible tube was closed. The pump was
tmed on, producing a fiow of 1.8mUs (Re = 0.3 66). Glycerol fiiled the connective
tubing and the standpipe. The fluid level in the standpipe rose until P, was sufficient to
A Novel Giuuxinu DNnigc Vdve
80
open the valve. Glycerol flow, measured at the valve's outlet with a graduated cylinder,
and the change in fluid levels in the manometer and standpipe were recorded as a function
of the . When ail three variables reached steady state, the flow setting on the pump was
increased to 4.0mVs (Re = 0.8 1). Changes in the three variables were again record4 as a
function of tirne, and when steady state was reached, the flow setting on the pump was
decreased to O S d s (Re = 0.102). These three flow settings were selected in order to
represent a range of Reynolds numben below 1.
The osrnotic pressure dEerence obtained in Valve II was 40 mmHg. This was
comparable to the 40-43 mmHg pressures typically obsewed using Valve 1, and indicated
that the majority of dextran molecules were retained by the ultrafiltration membrane. Had
the osmotic pressure gradient in Valve II been significantly lower than that of Valve 1,
dextran concentrations would have had to be increased in order to obtain a sirnilar osmotic
pressure. Othenvise, with two dEerent valve closing pressures. it would have been
difficult to compare the performances of the two valves.
6.3.2 VALVE II RESPONSE
The response t h e results for Valve II are shown in figure 6.4. Tirne constants were
calculated, and compared with those obtained fiom Valve 1 data (table 6.1).
O 50 100 t50 .O 250 350 400 time [min]
Figure 6.4: Cornpanson of Valve II response time data from successive experiments
using the same ultrafiltration membrane. 0.224mM dextran f i lhg solution. and glycerol
P, of 8OmmHg. The descending m e repi'esents net potential merence. while
the valve is openhg the ascending c w e represnts A@-m while the valve is closing.
nie value of was the osmotic pressure gradient measured before each e.uperiment;
and AR was the rimedependent pressure dinerenœ between the de.nran cuff and the
outer chamber. Order of experiments: 0.0 .x+.
A Novcl Glwobau Drunage Vdve
I TIME CONSTANT [min]
Valve Openin
Valve 1
9" Tube 1 3" Tube 1 I I
3 (x) r 1245130 ( 637k5.5 1 60.3I1.1
Valve Closhg
9" Tube 1 3" Tube 1
* Data not available
Table 6.1 : Cornparison of system time constants between Valve 1 (original 9" and
shortened 3" collapsible tubes) and Valve II. Values are mean * SEM.
When the A(pJ) data was ploaed on log-linear axes, some of the data sets produced a
slight curve rather than a araight line. This phenomenon was not observed with data
coiiected using Valve 1. By adding a correction factor, +, to the net potential difference
term, A@JI++), the curves were araightened. Values for 4 were determined based on
the valve which produced the straightest curve, and ranged from 0.1 to 7.OmmHg.
The necessity for correction factors was probably due to the use of a manometer to
measure p,. Although the manometer was a readily avaiiable, inexpensive, and low
maintenance means of pressure measurement, it had a limited dynamic response. When
the glycerol reservoir was raised &om O to 80mrnHg the change in net potential dserence
was Unmediate, but the change in manometer fluid level was not. It took approximately
83
30 seconds for the manometer fluid level to reach a peak valve, due to the flow resistance
in the manorneter tube. The time constants obtained for Valve 11 indicate a rapid
response. Therefore, it was suspected that the net potential difEerence began to change
before Ap, had reached its peak value according to the manometer. Thus, correction
factors were necessary to compensate for the inaccurate initial data. In cornparison, the
response of Valve 1 was slow enough that the manorneter's delay had no noticeable effea.
Valve II demowtrated a significantly faster response than both versions of Valve 1. This is
consistent with earlier predictions that: (i) a ten-fold reduction in response tirne would be
achieved by shortening the coiiapsible tube and increasing the membrane surface area, and
(ü) that increasing membrane permeability would contribute to further response t h e
reduction. These combined changes reduced s's between Valve 1 and Valve 11 by a
factors of 16-24 for valve opening (which wrresponded to the predicted weduction
factor of Z3.9), and 40-50 for valve closing.
Valve II time constants did not appear to increase with successive uses of the
ultrafiltration membrane, hdicating that no significant amount of pore clogging occurred.
Due to the larger pore diameter and Valve II's smaller required water flux, the pores of
the ultrafiltration membrane had less probability of becoming clogged (compared to Valve
1). therefore more experiments would be required for any trend to become apparent.
A NovcI G f w c a ~ Dmhgc Valve
84
The relatively rapid response time demonstrated by Valve II made it practical to consider
performing dynamic tests.
The results from the dynamic tests provided some interesting information regarding the
behaviour of Valve II (figure 6.5). The fluid level in the manometer was expressed as the
net potential difference, A(p,-II), as in previous experiments. The fluid level in the
standpipe, representing supply pressure or 'TOP" was expressed as A(p,-LI). Ideally, at
aeady state for each of the tests, A@,-II) should equal to A(p&).
During the tests, A(pJ1) quickly recovered fiom changes to Bow settings and a consistent
pressure was maintained around the coilapsible tube. This indicated an osmotic pressure
source works well as Valve II's self-contained closing pressure generator.
O 100 200 300 400 500 600 700 time [min]
- read left - read lefl - read right
set supply flow change su ply change supply at 1.8mUs 1 h to 4mbs fiow to O.5mUs
Figure 6.5: Results h m Uuee consecutively run dyaamic tests on Valve II. Changes in
ouifiow. net potentiai Merence. A@,-II); and supply net potential W e ~ n c e , A@,-II)
were monitored as a constantomte giycerol flow was pumped through the valve. m e n
the variables reached steady m e . the experiment was repeated for a new flow rate.
A SP
h & A L) &
P A A
9
1- I b o e a œ ~ a I I
C I S m -
œ m - * - - .
1 . œ m O 8.0 O
A 4 a 0 6 3 O
n t f
h -L
œ - & a O
O m
a 1 1 t . 1 a I a I a 1 1
8 u 1 I . v w 1 8 u m u
Throughout the first test (0-220 minutes), A@,-II) was approxirnately qua1 to A@JI),
and both variables were zero at aeady state. This indicated that at a constant flow of
1.8rnL/s, Valve II does an excellent job of maintainhg a constant supply pressure of AII.
5
-4
# œ
-3
-2
m
- - 1
I D
O
During the second test (220-450 minutes), A@,-a is consistently 5mmHg above A(prII).
At this higher flow rate, 4.OmUs, the cdapsible tube was nearly fully open. Although not
A Novcl G h c m u Dnmrgc Vaivc
evident on the pressure vs. t h e plot, the fluid level in the standpipe was actualiy
oscillating by *2mmHg. This observation was explained using the "tube law".
Figure 6.6: The tube law. Graphical represeniation of the relationship between
transwall pressure and cuUapsibIe tube compression. ~ececioglu et al.. 198 11
The tube law describes the non-linear relationship between transwall pressure and
coilapsible tube compression. The general expression for tube volume as a iùnction of
pressure is given as:
87
Where V is the volume of the collapsible tube as it changes shape, Vo is the maximum
volume of the collapsible tube, p is the pressure within the tube, and p, is the pressure
outside of the tube. The constant K is determined by tube materials and dimensions,
includiig: Poisson's ratio (v), Young's modulus (E), tube radius (R), and wall thickness
(hl.
As iliustrated by the varying slopes on the plot in figure 6.6, the collapsible tube is fairly
stiffat VNo < 0.27. relatively compliant at 0.27 < VN, < 0.92, and becornes very stifYat
VN, > 0.92. During the second stage of the experiment. it was suspected that VN, had
reached its upper limit for cornpliance. The collapsible tube resisted inflation, allowing
only some of the glycerol supply to pass through. The "extra" glycerol accumulated in the
standpipe, and the supply pressure increased untü the latex yielded. The "extra" glycerol
drained and the process repeated itself, approxirnately every 3 seconds. This phenornenon
indicated that the collapsible tube itself plays a significant role in determining the upper
limit of an osmotic valve's operating range.
During the third test (450-610 minutes). A(p,-11) was consistently lower than A(p,-II). At
steady aate. A@,Q was oscillating by *2.5mmHg, at an average value 8mmHg below
A@JI). In this situation, the wliapsible tube would be expected to fuUy close, reducing
outflow to zero until A@,-m was equal or greater than A@,-II). Although the collapsible
tube appeared fuily closed, there were srnail channels dong its sides that remained open.
A Novd Glaucomn hinage Vdvc
These channels were large enough to d o w the glycerol to dnp slowiy at a rate of
0.5mUs. The size of these channels depends on factors sUch as tube thickness and
material properties. In order for the vaive to close, a tube would have to be selected that
would have srnalier side channels when fùlly collapsed. This phenornenon indicated again
that the coliapsible tube itself plays a signifiant role in determining the valve's operating
chmct eristics .
These results suggested that Valve II would only maintain "IOP at AI1 if the glycerol
supply flow was set within a certain range. Flow rates of OSmUs and 4.0mUs were
clearly outside of this range, while 1.8mUs was within the range. Since the arnount of
aqueous humor production in the eye varies throughout the day, it is crucial that a
drainage device be able to maintain a consistent 10P regardless of flow vdume. The
challenge would be to select a coliapsible tube such that: (i) the valve's upper flow limit is
larger than the highest expected aqueous humor flow; and (ü) flow through the side
channels is negligible.
Valve 1 was designed and constructed, taking Sit's (1996) recommendations into account.
Significant aiterations to Sit's original design included: improvement of the diaiysis
membrane seal, addition of dextran f i h g pons, and cylindrical construction of the outer
shell.
Operation of Valve 1 was verified by setting the valve closing pressure with an extemai
water reservoir, and o b s e ~ n g the outflow as a function of supply pressure. DifEerences
between the flow resistances of the two prototypes was likely due to differences in
collapsible tube characteristics, specifically, length and longitudinal tension.
Valve 1 did not open as expected when the collapsible tube had an osmotically generated
closing pressure. An investigation reveaied that the van't Hoff equation underestirnated
the actual osrnotic pressure of the dextran solution. The actual osmotic pressure was
determined using a manometer and supply pressures were raised accordingly. In addition,
it was determined that the amai diaiysis membrane permeability was 50 tirnes lower than
the manufacturer's quoted value. Further testing revealed that Valve 1 had an extremely
slow response tirne, with time constants (5) ranging between 870 and 1250 minutes d u ~ g
valve opening, and from 740 to 1 O00 minutes during valve closing. These experimental
90
values were consistent with the expected values detemiined using a mathematical model.
The volume of the fully open coUapsible tube has a great impact on the value of T. In an
attempt to reduce response the, the collapsible tube in Valve 1 was shortened.
Experimental values for s ranged from 330 to 640 minutes dunng valve opening. This
was a signincant reduction, but it was not enough.
Valve II was constructed in an effort to fùnher reduce response time. It featured a
coilapsible tube even shorter than in the modified version of Valve I, a membrane with a
higher permeability, and a larger membrane surface area. Response tirne was drarnatically
reduced, with values of r ranging between 47 and 60 minutes during valve opening, and
from 18 to 20 minutes during valve closing.
Dynamic testing of Valve II indicated that osmotic pressure works well as a self-contained
source of pressure. However, it was observed that the collapsible tube irnposed significant
limitations on the operation of the valve. At higher supply flows, the collapsible tube
opened to a certain volume and then appeared to resist further inflation. The result was an
oscillating TOP" with an average value 5rnmHg above II. At lower supply fiows, the
coUapsible tube was not compietely collapsed. Glycerol passed through the side channels,
and the oscillating T O P had an average value 8mrnHg below II. These observations
suggested that selecting an appropnate collapsible tube would be the greatest challenge in
constmcting a useable osmotically controiled glaucoma valve.
As discussed in chapter two, the objective of this project wax
To construa and test a large-scale mode1 of an osmotic pressure controiied valve,
based on Sit's (1996) design and recommendations, in order to: (i) ver@ that the
valve has a consistent opening and closing pressure and (ü) veriS, that the valve
capable of controliing outflow and maintainhg a constant supply pressure.
Valves 1 and II both opened at supply pressures greater than the osmotically generated
closing pressure, and closed at supply pressures less than the closing pressure. However,
as observed in the dynamic experiments using Valve II, being "open" or "closed" did not
necessarily mean that a constant supply pressure of AI1 could be maintainecl.
Valve II's ability to control outtlow and maintain a constant supply pressure was
dependent on the supply flow rate. When the valve was open, the collapsible tube
appeared to resist fûrther expansion beyond a certaii? point. Thus, at high supply flow
rates, fluid accumulated in the standpipe, and the supply pressure oscillated, having an
average value above the set closing pressure, An. When the valve was closed,
incompressible channels remained dong the sides of the coUapsible tube. Thus at low
flow rates, fluid did not accumulate in the standpipe, and the supply pressure oscillated,
having an average value below AII. At mid-range flow rates, supply pressure was
maintained at AIX.
1.3.1 VALVE DESIGN AND EXPERTMENTAL APPARATUS
The ha that a supply pressure of AI1 could not be maintained at low flow rates
constitutes a major flaw in the valve design. Side channel formation in a coilapsible tube is
unavoidable. However, the use of a thimer-wded tube may reduce the radius of the
channels and hence increase resistance to flow. Altematively, rather than a round tube, it
rnay be worihwhile to consider using two separate pieces of latex that have been sealed
dong the sides. In this configuration, there would be no folds in the latex and side
channels would not form.
Pnor to any fkther testing, it is suggested to replace the manometer with a strain gauge-
based pressure transducer. Eiirninating the manometer would simplify the mathematicai
mode1 of the device. More importantly, the response of a strain gauge would be faster and
readings could be recorded by a datalogger during the lengthy experiments. It would also
be useful to place a strain gauge-based pressure transducer at the base of the standpipe in
the dynamic test apparatus in order to accurately record pressure oscillation.
1.3.2 FUTURE WORK
As it is, the valve can not maintain a supply pressure of AII at low fiow rates. Outflow,
via the side channels in the collapsible tube, must be reduced to a neghgible amount if the
valve is to operate properly. Mer this is accomplished in the large-scale prototype,
design of a scaled-dom valve may begin.
Cheryan, Munir. WZtrajîZtration H d b o o k . Technomic Publishing Company, Inc. Lancaster, Pemsylvania, 1986.
Conrad , William A Pressure-Flow Relationships in Collapsible Tubes. IEEE Transactions on Biomedical Engineering BME16(4): 284-295, 1969.
Davson, H. [ed]. The Eye. Academic Press, New York, New York, 1969.
Edsman, K. and Sundeloef, L.D. Interaction Virial Coefficients in Some Mixed Polymer Solutions. Polymer 29(3), 53 5-540, 1988.
Ethier, C.R. Personal Communication. 1997- 1999.
Fatt, lrving and Weissman, Barry A. PhysoZogy ofthe Eye: An ln@&ction to the Vegetutive Functions. Butterwort h-Heinemann, Boston, Massachusetts 1 992.
Feher, Joseph J. and Ford, George D. A Simple Student Laboratory on Osmotic Flow, Osmotic Pressure and the Reflection Coefficient. American J w m d of PhysioIogy 13 ( 1 ) : S 10420, 1995.
Francis, Brian A, et al. Characteristic of Giaucoma Drainage Implants during Dynarnic and Steady-state Flow Conditions. Ophthalmology. 105(9): 1708- 17 14.
Freedman, JefEey. Clinical Experience With the Molteno Dual-Chamber Single Plate Implant. Ophthulmic Surgery. 23(4): 238-241, 1992.
Gaube, Johann. Pfennig, Andreas. Stumpf. Matthais. Vapor-Liquid Equilibrium in Bhary and Ternary Aqueous Solutions of Poly(ethy1ene glycol) and Dextran. Journal of Chernical and Engtneering Data 38(1): 163- 166, 1993.
Guyton, Arthur, C. And Hall, John E. Textbook of Medicai Physiology. W .B. Saunders Company, Philadelphia, Pennsylvania, 1996.
Hitchings, R.A. et al. Use of One-piece Valved Tube and Variable Surface Area Explant for Glaucoma Drainage Surgery. Ophthalmoogy 94(9): 1079- 1083, 1987.
Heuer, Dale K. et al. Which is Better? One or Two? A Randomized Clinid Trial of Single-plate versus Double-plate Molteno Implantations for Glaucomas in Aphakia and Psuedo phakia. Ophthalmology W(l0): 1 5 1 2- 1 5 1 9, 1 992.
Jonsson, G. Boundary Layer Phenomena During Ultrafiltration of Dextran and Whey Protein Solutions. Desdination 5 1 : 6 1-77, 1984.
Joseph, N.H. et al. A One-Piece Drainage System for Glaucoma Surgery. Tr~72sactiuns of the Ophthulmologrcul Societies of the United Kingdom 1 O5 : 6 5 7-663, 1 9 86.
Kececioglu, 1. McClurken, M.E. Kamm, R.D. Shapko, AH. Steady, supercritical flow in cullapsible tubes. Part 1. Experimental Observations. Journal of Fluid Mechanics. 109: 367-389, 1981.
Knipin, Theodore, et al. Valve Implants in Filtering Surgery. Amerzcm J m m I of Ophtholmology 8 1(2): 232-235, 1976.
Krupin, Theodore, et al. A Long Krupin-Denver Valve Implant Attached to a 180 O
Scleral Explant for Glaucoma Surgery. Ophhzlmology 95(9): 1 174- 1 180, 1988.
Krupin, Theodore, et al. Knipin Eye Valve with Disk for Filtration Surgery. OphthaZmology 1 O 1 (4): 65 1-657, 1 994.
Lyon, Carol K. et al. Flow through Collapsible Tubes at low Reynolds Numbers: Applicability of the Waterfdl Model. Circulation Research 47(1): 68-73, 1980.
McClurken, M.E. Kececioglu, 1. Kamm, R.D. Shapiro, AH. Steady supercritical Bow in coilapsible tubes. Part 2. Theoretical Studies. Jmml of Fluid Mecbics . 1 O9: 39 1 - 415, 1981.
Normann, Richard A. Principles ojBioimtnrmentation. John Wiley and Sons, New York, New York, 1988.
Pilliar, R.M. MMS-420 Biomateriais. Course Notes and Lectures, 1997.
Porter, Jeffiey M., et al. In Vitro Flow Testing of Glaucoma Drainage Devices. Ophthalnmlogy 1 O4(l O): 1 70 1 - 1707, 1 997.
Prata, Joao A. et al. In Vitro and In Vivo flow Characteristics of Glaucoma Drainage Implants. Ophrhalmology 1 O2(6): 894-904, 1995.
Pusch, Wolfgang. Osmotic Pressure of De- Tl 0 Solutions. Desaiimtion 68: 69-73, 1988.
Rice, Wdliam. Koch Membrane Systems R&D. Personal Communication. 1998- 1999.
Ritch, Robert. Shields, M. Bruce. Krupin, Theodore. The Glmcomas. Mosby-Year Book Inc, St. Louis, Missouri, 1996.
Roark, Raymond J. Young, Warren C. Fornulas for Stress and S W n . McGraw-Hill Book Company, New York, NewYork, 1975.
Robenon, John A. and Crowe, Clayton T. Engineering FluÎdMechanics. Houghton Mifflin Company, Boston, Massachusetts, 1993.
S herwood, Lauralee. Humun Physiology: From Cells to Systems. West Pub iishing Company, Mimeanapolis, Minnesota, 1 993.
Shields, M. Bruce. Textbook of Glmcoma. Williams & Wilkins, Baltimore, Maryland, 1987.
Sit, Arthur J. A Novel Glaucoma Drainage Implant: Preliminary Report. Department of Mechanical and Industrial Engineering, University of Toronto, Toronto, Ontario. 1996.
Skalak, Richard. Chie% Shu. Handbook of Bioengineering. McGraw-W Book Company, New York, New York, 1987.
Thomas, John V. Glaucoma Surgery. Mosby-Year Book Inc, St. Louis , Missouri, 1992.
Werner, Eüiot B. Normal-Tension Glaucoma The Glaucomas. Mosby-Year Book Inc, St. Louis, Missouri, 1996.
White, T.C. Clinicd Results of Glaucoma Surgery Using the White Glaucoma Pump Shunt. Anncrls of OphthImology 24(10): 365-373, 1992.
Wijmans, J.G. Nakao, S. Van den Berg, J. W.A. Troelstra, FR. Smolders, C.A. Hydrodynamic Resiaance of Concentration Polarization Boundary Layers in Ultrafiltration. Journal ofMembrane Science 22: 1 17- 13 5, 1985.
Varma, Rohit. Minckler, Don S. Anatomy and Pathophysiology of the Retina and Optic Nerve. The GIaucumm. Mosby-Year Book Inc, St. Louis, Missouri, 1996.
APPENDE A . Drawings for Valve I
Large Outlet Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A l
Small Outlet Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2
Sealing Ring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A3
Endplate (Outer Shell) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A4
Outer Shell with Enclplate Attached . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A5
APPENDIX B . Drawings for Valve iI
Cross Section O AssembIed Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B1
BoxTop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
BoxSides . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B3
Topplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B4
TopWall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . BS
Bottom Wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B6
Bottom Plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B7
APPENDIX C . Regression Analysis Output
Valve 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cl
Valve 1 (modified) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C7
ValveII . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cl1
REGRESSION ANALYSE
VALVE: Valve l(9" tube) RUN: #î, valve opening
m: 48.5rnrnHg DATE: August 26 1998
O 50 100 150 200 250 300 350 400 time [min]
Regression Output: TlME CONSTANT: Constant 3.290436 Std Err of Y Est 0.020476 TAU: -870.2 R Squared 0.9751 92 LOW TAU: -850.0 No. of Observations 47 HlGH TAU: -891.4 Degrees of F reedom 45 +/- 20.7
X Coefficient(s) -0.001 1494 36 Std Err of Coef. 2.73221 E-05
REGRESSION ANALYSE
VALVE: Valve I (9" tube) RUN: #1, valve dosing
46.5mmHg DATE: August 26 1998
O 200 400 600 800 1000 time [min] l2Oo I
Regression Output: TlME CONSTANT: Constant 2.56421 9 Std E r of Y Est 0.01 91 31 TAU: -784.3 R Squared 0.997992 LOW TAU: -774.6 No. of Observations 15 HlGH TAU -794.1 Degrees of Freedom 13 +/- 9.8
X Coefficient(s) -0.001 27508 Std Err of Coef. 1 S864E-05
REGRESSlON ANALYSIS
VALVE: Valve 1 (9" tube) RUN: #2, valve opening
46.0mmHg DATE: August 29 1998
O 100 200 300 400 500 600 time [min]
Regression Output: TlME CONSTANT: Constant 3.409326 Std Err of Y Est 0.007339 TAU: -864.1 R Squared 0.998614 LOW TAU: -859.0 No. of Observations 42 HlGH TAU -869.2 Degrees of Freedom 40 +/- 5.1
X Coefficient(s) -1.1 573E-03 Std Err of Coef. 6.81723E-06
VALVE: Valve 1 (9" tube) RUN: #2, valve closing Cr 46.0mmHg
DATE: August 29 1998
O 200 400 600 800 1000 i time [min] I
i
Reg ression Output: TlME CONSTANT: Constant 2.73057 Std EIT of Y Est 0.064339 TAU: -789.0 R Squared 0.97663 LOW TAU: -762.0 No, of Observations 21 HlGH TAU: -818.1 Degrees of Freedom 19 +/- 28 .O
X Coefficient(s) -0.001 267367 Std Err of Coef. 4.497694E-05
REGRESSION ANALYSE
VALVE: Valve 1 (9" tub) RUN: #3, valve opening
f l 46.2mmHg DATE: Septernber 1 1998
O 1 O0 200 300 400 500 600 1 tirne [min] i
Regression Output: TlME CONSTANT: Constant 3.357364 Std ER of Y Est 0.01 9366 TAU: -1244.8 R Squared 0.9821 38 LOW TAU: -121 5.9 No. of Observations 34 HlGH TAU -1275.2 Degrees of Freedom 32 +/- 29.7
X Coefficient(s) -0.00080331 6 Std Err of Coef. 1.91 51 E-OS
REGRESSION ANALYSIS
VALVE: Valve I (3" tube) RUN: #1, valve opening
& 48.3mmHg DATE: Odober 27 1998
Regression Output: T1ME CONSTANT: Constant 3.397354 Std Err of Y Est 0.033966 TAU: -334.2 R Squared 0.991691 LOW TAU: -328.7 No. of Obsewations 32 HlGH TAU -339.9 Degrees of Freedom 30 +/- 5 -6
X Coefficient(s) -0.002991 8335 Std Err of Coef. 4.999989E-O5
VALVE: Valve 1 (3" tube) RUN: #Ki, valve opening a 52.3rnmHg
DATE: November 4 1998
250 500 tima [min]
Regression Output: TlME CONSTANT: Constant 3.524992 Std Err of Y Est 0.01 1536 TAU: -637.6 R Squared 0.997099 LOW TAU: -632.1 No. of Observations 40 HlGH TAU: -643.2 Degrees of f reedom 38 +/- 5.6
X Coefficient(s) -0.OOi 56833592 -637.61 9 Std Err of Coef. 1 -37231 03E-05
REGRESSION ANALYSIS
VALVE: Valve 1 (3" tube) RUN: #3, valve opening
AT 48.3mmHg DATE: November 16 1998
250 500 time [min]
Regression Output: TlME CONSTANT: Constant 3.493055 Std EIT of Y Est 0.045696 TAU: -360.7 R Squared 0.98899 LOW TAU: -353.8 No. of Observations 31 HlGH TAU -367.9 Degrees of Freedom 29 +/- 7.1
X Coeffcient(s) -0.0027721 4986 -360.731 Std Err of Coef. 5.431 502E-05
VALVE: Valve 1 (3" tube) RUN: #3, valve dosing
& 48.1mmHg DATE: November 16 1998
Regression Output: Constant Std Err of Y Est R Squared No. of Observations Degrees of Freedom
TlME CONSTANT:
TAU: 652.0 LOW TAU: 657.2 HlGH TAU: 646.9
+/- 5.1
X Coefficient(s) 0,001 5337208 Std Err of Coef. 1.205788E-05
VALVE: Valve II RUN: #2, valve opening
+ 40.76 + 5.0 = 45.76mmHg DATE: June 2 1999
-2 1 1 1 I
t I 9
1
O 50 100 150 200 250 time [min]
Regression Output: TlME CONSTANT: Constant 2.6561 59 Std Err of Y Est 0,185416 TAU: -53.6 R Squared 0.972393 LOW TAU -52.0 No. of Observations 33 HlGH TAU -55.2 Degrees of Freeâom 3 1 +/- 1 .S
X Coefficient(s) -0.01 8666831 -53.571 Std Err of Coef. 0.000564904
VALVE: Valve Il RUN: H, valve opening AT++: 42.3 + 2.7 = 45.02mmHg
DATE: June 1 1999
ta 150 2m 1 time [min] 1 l
! 1
Regtession Output: TlME CONSTANT: Constant 2.8261 68 Std Err of Y Est 0.082899 TAU: -59.6 R Squared 0.990731 LOW TAU: -58.4 No. of Observations 24 HlGH TAU: -60.9 Degrees of Freedom 22 +P 1 -2
X Coefficient(s) -0.01 676991 2 -59.6306 Std Err of Coef. 0.0003458183
REGRESSION ANALYSE
VALVE: Valve II RUN: W , valve ciosing A%++ 40.0 + (-2.4) = 37.6 mmHg
DATE: June 2 1999
O 10 M 30 40 50 time [min]
Regression Output: TlME CONSTANT: Constant 2.631 86 Std Err of Y Est 0.057422 TAU: -1 8.6 R Squared 0.994045 LOW TAU: -1 8.2 No. of Observations 14 HlGH TAU -19.0 Degrees of Freedom 12 +/- 0.4
X Coefficient(s) -0.0538841 65 -1 8.5583 Std Err of Coef. 0.0012039121
REGRESSION ANALYSIS
VALVE: Valve il RUN: #3, valve opening
39.1 + 7.4 = 46.5mmHg DATE: June 3 1999
r 1 l I i
O 50 400 150 tOO 250 #KI tirne [min]
Regression Output: TlME CONSTANT: Constant 2.638358 Std Err of Y Est 0.1 11 882 TAU: -60.3 R Squared 0.991312 LOW TAU: -59.2 No. of Observations 28 HlGH TAU: -61.4 Degrees of Freedom 26 +/- 1.1
X Coefficient(s) -0.01 6597389 -60.2504 Std Err of Coef. 0.000304729
REGRESSION ANALYSIS
VALVE: Valve II RUN: #3, valve closing M++: 39.1+ 0 = 39.1 mmHg DATE: June 3 1999
i f O 1 O 20 30 40 54 60
time (min] ! I l i I i
Regression Output: TlME CONSTANT: Constant 2.720793 Std Err of Y Est O. 1 323 TAU: -1 9.9 R Squared 0.983501 LOW TAU: -1 9.3 No. of Observations 19 HlGH TAU -20.6 Degrees of Freedom 17 +/- 0.6
X Coefficient(s) -0.05021 71 6 -1 9.91 35 Std Er of Coef. 0.001 57749
VALVE: Valve il RUN: M, valve opening M+@ 40.1 + 6.4 = 46.5mmHg
DATE: June 4 1999
I
-1 ! 1 1 T r
1 L ,
O 20 40 613 8Q 100 120 140 time [min]
Regtession Output: TlME CONSTANT: Constant 2.640947 Std Err of Y Est 0.07874 TAU: -46.8 R Squared 0.98481 4 LOW TAU: -45.6 No. of Observations 25 HIGH TAU: 48.1 Degrees of Freedom 23 +/- 1.2
X Coefficierlt(s) -0.021 358329 -46.8201 Std Err of Coef. 0.000553022
REGRESSION ANALYSIS
VALVE: Valve Il RUN: #4, valve dosing AT* 40.1 + (-2.5) = 37.6mmHg DATE: June 4 1999
Regression Output: TIME CONSTANT: Constant 2.66055 Std Err of Y Est 0.1 û6595 TAU: -1 8.0 R Squared 0.9861 58 LOW TAU: -1 7.6 No. of Observations 28 HlGH TAU: -1 8.4 Degrees of Freedom 26 +/- 0.4
X Coefficient(s) -0.05561 1 343 -1 7.981 9 Std En of Coef. 0.001 2921 385
APPENDIX D - DEFORMATION CALCULATIONS
CALCULATIONS FOR SUPPORT PLATE DEFORMATION
General deformation equation for a solid round plate, firmly supported around perirneter (Roark & Young, 1975):
where:
a = plate radius [O. 107% ml E = Young's modulus [8.53e10 Pa] q = applied pressure [9800 Pa] t = plate thickness [0.002 ml v = Poission's ratio [0.3] r = point of interest dong plate radius [ml
(e.g. at centre of plate, r = O; at outside edge, r = a) y = vertical deformation at r [ml
Rewrite y(r) in terms of known variables:
Substitute known values and simplify:
y(r) = -9.87e-5 + 0.016941r2 - 0.72688r4
Treat y(r) as a constant, and rearrange to standard quadratic fom:
Let x = ? and solve for x using quadratic formula: 1
Apply boundary condition: r = 0.10795, y = O
General equation for volume o f rotation around y-axis:
Substitute equation for r(y) detemineci above:
Integrating and setting y, = -O. lmm (arbitrary small value):
V = 1.204045~-6m3 = 1.2mL
Therefore it can be concluded that small membrane deformations result in large volume changes.