nps8410

Embed Size (px)

Citation preview

  • 7/28/2019 nps8410

    1/4

    Solid State Ionics 107 (1998) 213216

    NOx detection using the electrolysis of water vapour in a YSZ cell

    Part I. NOx detection

    a , a b b b*T. Hibino , Y. Kuwahara , T. Otsuka , N. Ishida , T. Oshimaa

    National Industrial Research Institute of Nagoya, 1-1, Hirate-cho, Kita-ku, Nagoya 462, Japanb

    NGK Spark Plug Co., Ltd., 2808 Iwasaki, Komaki-shi, Aichi 485, Japan

    Received 3 October 1997; accepted 4 November 1997

    Abstract

    The reaction of NO with the hydrogen formed by electrolyzing water vapour in a YSZ cell has been applied to detect NO

    with output signal of magnitude in millivolt or milliampere. The experimental apparatus consisted of two YSZ cells,

    PtuYSZuPt, which served as an electrolysis cell and an oxygen sensor, respectively. A mixture of 03000 ppm NO and 3%H O in argon was successively fed to the two cells at 800 8C. In the upstream cell, the hydrogen formed by electrolyzing

    2

    water vapour in the sample gas reacted with NO in the sample gas. In the downstream cell, the electromotive force (EMF)

    value was measured using air as a reference gas. The EMF value of the oxygen sensor was used as a sensor signal, when a

    current of 6.7 mA was applied to the electrolysis cell. The EMF signal decreased with increasing NO concentration in the

    sample gas. Furthermore, the current applied to the electrolysis cell was used as another sensor signal, when the EMF valueof the oxygen sensor was held at 700 mV. The current signal increased with increasing NO concentration in the sample gas.

    1998 Elsevier Science B.V. All rights reserved.

    Keywords: YSZ cell; Water vapour electrolysis; NOx detection

    1. Introduction potentiometric [12] and amperometric ones [13,14].

    In the former, the YSZ cells using metal nitrates as

    NOx sensors are becoming increasingly important sensing electrodes show the EMF value following

    for the improvement of the environmental pollution Nernsts equation. However, the operation tempera-

    such as acid rain and photochemical smog and for ture of the sensor is limited up to the melting point

    the further introduction of lean-burn and diesel of the nitrate electrodes, many of which are lower

    vehicles into the market. So far many kinds of NOx than the highest temperature of the exhaust gases

    sensors have been investigated using oxide semi- from lean-burn or diesel engines. In the latter, when

    conductors [14], metal phytalocyamine [5,6], solid a constant voltage is applied to the YSZ cells, the

    electrolytes [714] and so on. Among these sensors, current based on the electrolysis of NOx is propor-

    those using YSZ can be expected to exhibit the best tional to its concentration. However, the current

    durability. As for such NOx sensors, there has been signal observed is only several mA per 1000 ppm

    NOx.*Corresponding author. Fax: 81 52 911 2422 We have proposed a new NOx sensor in order to

    0167-2738/98/$19.00 1998 Elsevier Science B.V. All rights reserved.P I I S 0 1 6 7 - 2 7 3 8 ( 9 7 ) 0 0 5 3 8 - 9

  • 7/28/2019 nps8410

    2/4

    214 T. Hibino et al. / Solid State Ionics 107 (1998) 213216

    solve the above problems. When water vapour is the oxygen sensor, where the operation temperature

    present in the sample gas, it can be easily elec- was 8008C and air was used as a reference gas. The

    trolyzed to hydrogen in the YSZ cell. Since hydrogen EMF value of the oxygen sensor was measured by an

    is generally reactive to oxidizing gases, it will react electrometer (Hokuto Denko HE-104). As an addi-

    with NOx in the sample gas. This paper deals with tional experiment, the concentrations of NO andthe application of the electrolysis of water vapour to hydrogen in the outlet gas from the electrolysis cell

    the detection of NOx. It is shown that this method were analyzed by an NOx analyzer (Shimazu NOA-

    makes it possible to detect NOx with output signals 7000) and GC (Shimazu GC 8A), respectively.

    in a practical use. Furthermore, it is demonstrated

    that this method can be applied to both potentiomet-

    ric and amperometric sensors. 3. Results and discussion

    Fig. 2 shows the changes in EMF of the oxygen2. Experimental sensor with the current applied to the electrolysis

    cell. The EMF value drastically increased at more

    The experimental apparatus consisted of an elec- than 6 mA, indicating that water vapour is elec-trolysis cell and an oxygen sensor, as shown in Fig. trolyzed to hydrogen. The EMF value, however, was1. YSZ disks were used as solid electrolytes of the strongly dependent on the concentration of NO in thetwo elements. The YSZ powders were purchased sample gas. At the same current, the EMF valuefrom Tosoh. These powders were pressed at 250 became smaller as the concentration of NO in-

    22kg cm and then sintered at 14508C for 10 h in air. creased. In order to elucidate this reason, we mea-The sintered compact was sliced into the form of a sured the gas composition in the outlet gas from thedisk (12 mm diameter and 1.0 mm thickness). Two electrolysis cell. The results are summarized in Fig.platinum electrodes, in the form of a porous film, 3. The formation of hydrogen was depressed in thewere attached to the surfaces of the disk by heating presence of NO. Also, the concentration of NOthe assembly to 9008C for 1 h. decreased at more than 6 mA. These suggest that the

    The electrolysis cell was heated to 8008C. A formed hydrogen reacts with NO in the sample gasmixture of 03000 ppm NO and 3% H O in argon2 as follows:

    was introduced into the upper gas compartment at a21 2 22

    flow-rate of 10 ml min . Direct currents were H O1 2e H 1O , (1)2 2

    applied from a galvanostat (Hokuto Denko HA-501)

    to the cell via two platinum wires. The outlet gas

    from the electrolysis cell was then introduced into

    Fig. 2. Plots of EMF of oxygen sensor against current applied to

    electrolysis cell: solid electrolyte YSZ; operation temperature

    8008C; 03000 ppm NO and 3% H O in argon at a flow-rate of 102

    21Fig. 1. Experimental apparatus consisting of two YSZ cells. ml min .

  • 7/28/2019 nps8410

    3/4

    T. Hibino et al. / Solid State Ionics 107 (1998) 213216 215

    Fig. 4. Change in EMF of oxygen sensor with NO concentration

    in sample gas: operation temperature 8008C; 0500 ppm NO and21

    3% H O in argon at a flow-rate of 10 ml min ; current applied to2

    electrolysis cell was kept at 6.7 mA.

    sensor signal, when a constant current was applied toFig. 3. Concetrations of NO (a) and H (b) in outlet gas from2

    the electrolysis cell. Fig. 4 shows a typical change inelectrolysis cell: experimental conditions are as in Fig. 2.EMF of the oxygen sensor with the concentration of

    NO in the sample gas. In this case, the current

    H 1NO

    H O1 1 /2N . (2) applied to the electrolysis cell was kept at 6.7 mA.2 2 2 The EMF values observed at less than 200 ppm NO

    Probably, this reaction will occur on the platinum drastically decreased from 700 to 200 mV, while

    electrode of the electrolysis cell. The consumption of those observed at more than 200 ppm NO were held

    hydrogen for Eq. (2) will decrease the EMF value of constant at 200 mV. This is due to the fact that the

    the oxygen sensor in the presence of NO. As another hydrogen formed by electrolyzing water vapour at

    reason, it can be considered that the electrolysis of 6.7 mA is entirely consumed by the reaction with

    NO directly occurs at the platinum electrode of the 200 ppm NO in the sample gas. As described in the

    electrolysis cell. The electrolysis of water vapour is following part of this paper, when currents of more

    inhibited during this process, so that the formation of than 6.7 mA are applied to the electrolysis cell, the

    hydrogen is suppressed in the presence of NO. As drastic decreases in EMF are observed at more than

    shown in Fig. 3, however, the decrease in NO 200 ppm NO.

    concentration began at 6 mA, where the voltage In the secondary manner, the current applied to the

    between the two electrodes of the electrolysis cell electrolysis cell was used as another sensor signal,

    was 1.2 V. The electrolysis of NO is thermody- when the EMF value of the oxygen sensor was kept

    namically possible from 0 V. Thus, this consideration at a constant value. Fig. 5 shows the result keeping

    seems to be not suitable for the above reason. the EMF at 700 mV. The applied current almost

    We applied the reaction represented by Eq. (2) to linearly increased with NO concentration. This is due

    detect NO. Two potentiometric and amperometric to additional application of current to the electrolysis

    manners were adopted to observe the sensor signal cell so as to compensate for the amount of hydrogen

    for the NO concentration. In the first manner, the consumed by reacting with NO. The present manner

    EMF value of the oxygen sensor was used as a can be characterized by the fact that the relationship

  • 7/28/2019 nps8410

    4/4

    216 T. Hibino et al. / Solid State Ionics 107 (1998) 213216

    present method disappear when excess oxygen is

    present in the sample gas. In the following part of

    this paper, we will propose an approach for such a

    disadvantage.

    References

    [1] G. Sberveglieri, S. Groppelli, P. Nelli et al., Sensors and

    Actuators B1 (1990) 79.

    [2] M. Akiyama, J. Tamaki, N. Miura, N. Yamazoe, Chem. Lett.

    (1991) 1611.

    [3] N. Imanaka, S. Banno, G. Adachi, Chem. Lett. (1994) 319.

    [4] H. Yamaura, J. Tamaki, N. Miura, N. Yamazoe, Proceedings

    of the 19th Chemical Sensor Symposium, Yokohama, Japan,

    1994, p. 13.

    [5] Y. Sadaoka, Y. Sakai, I. Aso, N. Yamazoe, T. Seiyama, DenkiKagaku 48 (1989) 486.

    [6] T.A. Jones, B. Bott, S.C. Thorpe, Sensors and Actuators 17Fig. 5. Change in current applied to electrolysis cell with NO

    (1989) 467.concentration in sample gas: EMF of oxygen sensor was kept at

    [7] M. Gauthier, A. Chamberland, J. Electrochem. Soc. 124700 mV; other conditions are as shown in Fig. 5.

    (1977) 1579.

    [8] G. Hotzel, W. Weppner, Solid State Ionics 1819 (1986)between the sensor signal and the NO concentration 1223.

    [9] G. Hotzel, W. Weppner, Sensors and Actuators 12 (1987)is linear, although the sensitivity to NO is somewhat449.small at low concentrations.

    [10] S. Yao, Y. Shimizu, N. Miura, N. Yamazoe, Chem. Lett.As described above, the use of electrolysis of(1993) 193.

    water vapour enables one to detect NO with sensor [11] N. Miura, S. Yao, Y. Shimizu, N. Yamazoe, Solid State Ionicssignals of magnitude in millivolt or milliampere. 7071 (1994) 572.

    [12] H. Kurosawa, Y. Yan, N. Miura, N. Yamazoe, Solid StateHowever, there is a large need for the NOx sensorsIonics 79 (1995) 338.capable of monitoring NOx in the exhaust gases

    [13] N. Kato, K. Nakagaki, N. Ina, SAE paper, 960334, 1996.from lean-burn or diesel engines. Such gases gener-[14] N. Kato, Y. Hamada, H. Kurachi, SAE paper, 970858, 1997.

    ally contain more than several % oxygen. It is

    apparent that both the EMF and current signals in the