Upload
hiroshi-sekiguchi
View
215
Download
2
Embed Size (px)
Citation preview
Non-destructive force measurement in liquid usingatomic force microscope
Hiroshi Sekiguchi, Hideo Arakawa, Takaharu Okajima, Atsushi Ikai*
Laboratory of Biodynamics, Graduate School of Bioscience and Biotechnology, Tokyo Institute of Technology,
4259 Nagatsuta, Midori-ku, Yokohama 226-8501, Kanagawa, Japan
Received 2 September 2001; accepted 21 September 2001
Abstract
Atomic force microscopy (AFM) has been applied to measure inter- or intra-molecular forces acting to hold biological
molecules and structures. For these measurements, it is important to keep the target molecules biologically active on a solid
surface. Besides the strategy for immobilizing them on the surface keeping their biological activities intact, it is crucial to reduce
the force applied to them through the AFM tip to avoid mechanical inactivation of the sample. In this paper, we propose a new
procedure to minimize the effect of contact force. The first step of the procedure is to bring the cantilever tip close to the sample
surface within less than 3 mm, but short of contact with the sample surface. The approximate distance of the tip from the sample
stage is measured using the thermal fluctuation of the cantilever. The second step is a ‘‘compression-free’’ force spectroscopy for
the measurement of protein–protein interactions only, which is possible when the piezo scanner was retracted before the
cantilever starts upward deflection. The interaction force can be measured in the retraction period provided a physical contact is
established between the proteins on the tip and the substrate. This procedure allowed to measure interaction forces between
GroEL and a denatured protein without mechanical deformation. # 2002 Elsevier Science B.V. All rights reserved.
Keywords: Force spectroscopy; Bio-molecule; Thermal fluctuation; Compression-free
1. Introduction
Atomic force microscopy (AFM), invented in 1986
by Binnig et al. [1], has been increasingly used in
biological science not only for imaging but also for
measuring forces [2]. One important precaution to be
considered in the force measurement is how to fix bio-
molecules on a substrate and a probe, securely enough
for measuring force but flexible enough to keep the
molecules biologically active. Various methods for
this purpose have been reported using chemical
cross-linkers, flexible spacer molecules [3], inactive
proteins as cushions [4] and/or self-assembled mono-
layers [5]. Another issue to be addressed in such
experiments is the effect of pushing the molecules
on the substrate and AFM probe in the course of force
measurements. As the AFM probe is pushed onto the
molecule, there is a possibility of denaturing the mole-
cule artificially or adsorbing it to the probe physically.
In reality, however, it has been difficult to apply an
AFM in force measurements without pressing the
sample with the probe for two reasons. First, it is
difficult for us to know how far the probe is from the
sample surface before the cantilever senses the repul-
sive force, and secondly, the probe has to be in contact
with the sample molecule on the substrate for a certain
Applied Surface Science 188 (2002) 489–492
* Corresponding author. Tel.: þ81-45-924-5828;
fax: þ81-45-924-5806.
E-mail address: [email protected] (A. Ikai).
0169-4332/02/$ – see front matter # 2002 Elsevier Science B.V. All rights reserved.
PII: S 0 1 6 9 - 4 3 3 2 ( 0 1 ) 0 0 9 7 7 - 1
period to establish interactions or bond formation. To
keep them in contact, the force curve must include
some period of repulsive, i.e. compressing region.
In this paper, we propose a new method to overcome
these problems. The method consists of two steps. The
first step is to bring a cantilever tip close to the sample
surface within the range of piezo-electric’s movement.
We estimate the distance between the tip and the surface
without prior contact between them by detecting the
power spectrum of the cantilever’s thermal fluctuation
in liquid. We could bring the tip within less than 3 mm to
the surface in this step. The second step of the procedure
is the ‘‘compression-free’’ force spectroscopy measure-
ment where no repulsive force is exerted during the
approaching process to immobilized molecules, but the
interaction force is measured in the retraction process.
The results for a protein pair, GroEL on the substrate
and denatured pepsin on the probe, showed difference
depending on compression-free or not.
By using these procedures, force measurements can
be done almost without touching the sample surface
with the AFM probe.
2. Materials and methods
2.1. AFM
A Multimode Nanoscope IIIa (Digital Instruments,
Santa Barbara, CA) equipped with a K-scanner and
a Signal Access Module, and an SPM-9500-J2
(SHIMADZU, Japan) were used. The deflection signal
was recorded by an A/D converter (ADM-670PCI,
Microscience, Japan) mounted on a commercial PC.
The AFM was operated with a standard software or
with a special software for force spectroscopy mea-
surements, ‘‘Force Curve Software, Ver. 2.54 (SHI-
MADZU, Japan)’’. The latter program enabled us to
control the holding time of the piezo-tube at its
approach end, which was equated with the reaction
time for bio-molecular pairs at the nearest position to
each other.
2.2. Preparation of functionalized substrate and tip
Gold-coated AFM tips (OMCL-TR400-PB, Olym-
pus, Japan) were functionalized with porcine pepsin
(SIGMA, Missouri, USA) through the cross-linker,
sulfo-LC-SPDP (sulfosuccinimidyl-6-[30-(2-pyridyl-
dithio)-propionamido]hexanoate, PIERCE, Illinois,
USA) according to the method in [6]. This protein
is known to lose its native conformation [7] and
interact with GroEL at neutral pH [8].
A drop of a GroEL solution (0.5 mg/ml, 100 ml) in
HEPES buffer (20 mM HEPES, 100 mM KCl, 5 mM
MgCl2, pH ¼ 7:2) was deposited on a freshly cleaved
mica surface and protein molecules were adsorbed on
it for 1 h at room temperature. Then the mica surface
was rinsed with HEPES buffer and kept in the buffer
until use.
2.3. Cantilever’s thermal fluctuation in liquid
The frequency and amplitude of thermal fluctuation
of the cantilever (OMCL-TR400-PS, Olympus, Japan)
in HEPES buffer was measured at various distances
between the tip and the mica surface. The distance was
controlled by a piezo-tube when the distance was less
than 6 mm, and by a step motor at larger distances (i.e.
the maximum movement of piezo was about 6 mm).
Deflection data, sampled at 125 kHz, were Fourier-
transformed into power spectra in order to obtain the
frequency response. The resonance frequency of the
cantilevers used was around 10 kHz in air.
3. Results and discussion
3.1. Spectral shift according to the distance
from surface
To start force spectroscopy measurements, the tip
must be brought close enough to the surface within the
range of the piezo movement. We found that the
distance of the probe from the surface can be estimated
from the spectral shift of the thermal fluctuation of the
cantilever in HEPES buffer.
The thermal fluctuation of the cantilever was mea-
sured without externally vibrating it. Fig. 1 shows the
power spectra of a cantilever at several distances from
the substrate. At larger distances than 10 mm, the
power spectrum of the cantilever fluctuation has a
peak at around 1.5–2.0 kHz (Fig. 1, circles). When the
probe was brought closer to the surface, the peak
shifted to around 1.0 kHz at 3 mm (Fig. 1, squares),
and the peak was hardly observable at 300 nm in
490 H. Sekiguchi et al. / Applied Surface Science 188 (2002) 489–492
distance (Fig. 1, triangles). The spectra were then
analyzed by curve fitting to a Lorentzian equation
[9] given by the expression
PðoÞ ¼ f
ðo2 � o20Þ
2 þ 4g2o2
where P is the power density, o the frequency, o0 the
resonant frequency of the cantilever, f the force
applied from the surrounding particles and g the
damping parameter.
The lines in Fig. 1 represent fitted data and the
values of the parameter are shown in Table 1. It is clear
that when the cantilever is near to the surface, the
resonant frequency (o0) is lower and the damping
parameter (g) is higher. It is noted that at distance of
about 3 mm, these values change drastically. Therefore
it is possible to bring a tip close to the substrate within
3 mm by monitoring the spectrum.
3.2. Compression-free force spectroscopy
Fig. 2 is a schematic drawing of the compression-
free force measurement which is composed of
approaching, holding and releasing processes. The
sample stage piezo starts the approaching process
(Fig. 2a) after the cantilever was brought close to
the surface within 3 mm by the method described
above. For the compression-free measurement, the
stage movement should stop before any indication
of repulsive force. The holding process at the nearest
position to the substrate (Fig. 2b) is important because
Fig. 1. Power spectra of cantilever’s thermal fluctuation at three
different distances from the substrate: 300 nm for triangles, 3 mm
for squares and 10 mm (or more) for circles. Spectra were fitted to a
Lorentzian function for each curve: 0.3 mm for solid, 3 mm for
dotted and 10 mm (or more) for dashed lines. As the distance was
controlled with step motor for 10 mm, the accuracy of the distance
was lower than that of other distances; however, the spectrum did
not change in effect when the distance was longer than 10 mm.
Table 1
Values of the parameter for the Lorentzian equation of the
cantilever’s thermal fluctuation in HEPES buffer
Distance from surface (mm) o0 (kHz) g (kHz)
0.3 1.30 1.31
1 1.52 1.30
3 1.81 1.20
6 2.01 0.93
10 (or more) 2.11 0.90
Fig. 2. Schematic drawing of a compression-free force spectro-
scopy. This method is composed of approaching (a), holding (b)
and releasing (c) processes. Figures on the left represent bio-
molecular pairs and AFM probe.
H. Sekiguchi et al. / Applied Surface Science 188 (2002) 489–492 491
it constitutes the reaction time for the establishment of
biological interactions. When the nearest position of
probe to the sample happens to be within the range for
the pairwise interaction to occur, a positive indication
of interaction is to be obtained (Fig. 2c).
This method was applied to the GroEL system, a
protein machinery that interacts with misfolded pro-
teins and assists them to re-fold correctly [10]. Fig. 3
shows the examples of recorded force curves in a
normal and a compression-free method. Interestingly
the magnitude of force observed in the compression-
free method was less than that observed in the normal
method, suggesting that multiple points or molecules
were reacted when the probe was pushed. The force
curves were also rather reproducible in compression-
free method while the normal method made variety in
profile and the force. This might be caused by the
different kinds and number of interactions happened
in normal method. We believe the compression-free
method gives much better results to study single
molecular event on soft material like bio-molecules
both in quality and reproducibility.
Occasionally we found the cantilever-sensed repul-
sive force during the holding time caused by the drift
of the equipment, but this seemed not damaging the
sample on the probe and the substrate, and not affected
the next compression-free experiment done subse-
quently. This kind of drift is of course not desired
in these experiments and the AFM with less drift is
hoped to be developed.
4. Conclusion
By using the methods of force curve measurement
described above, sample damages as well as tip con-
taminations were reduced so that the force curves
become stable and reliable. Since the contact area
between the tip surface and the substrate was mini-
mized, chances of multiple interactions were also
reduced.
Acknowledgements
This work was supported in part by Grant-in-Aid
to AI from the Japan Society for the Promotion
of Science (Research for the Future Program
#99R167019) and from the Japanese Ministry of
Education, Science, Culture and Sports (Scientific
Research on Priority Areas (B) #11226202).
References
[1] G. Binnig, C.F. Quate, C. Gerver, Phys. Rev. Lett. 56 (1986)
930.
[2] O.H. Willemsen, M.M.E. Snel, A. Cambi, J. Greve, B.G.
Grooth, C.G. Figdor, Biophys. J. 79 (2000) 3267.
[3] O.H. Willemsen, M.M.E. Snel, K.O. Werf, B.G. Grooth, J.
Greve, P. Hinterdorfer, H.J. Gruber, H. Schindler, Y. Kooyk,
C.G. Figdor, Biophys. J. 75 (1998) 2220.
[4] U. Dammer, M. Hegner, D. Anselmetti, P. Wagner, M. Dreier,
W. Huber, H.-J. Guntherodt, Biophys. J. 70 (1996) 2437.
[5] K.A.N.A. Wadu-Mesthrige, G.Y. Liu, Scanning 22 (2000)
380.
[6] T. Osada, S. Takezawa, A. Itoh, H. Arakawa, M. Ichikawa, A.
Ikai, Chem. Senses 24 (1999) 1.
[7] P. McPhine, Biochem. Biophys. Res. Commun. 158 (1989)
115.
[8] K. Aoki, H. Taguchi, Y. Shindo, M. Yoshida, K. Ogasawara,
K. Yutani, N. Tanaka, J. Biol. Chem. 272 (1997) 32158.
[9] J.L. Hutter, J. Bechhoefer, Rev. Sci. Instrum. 64 (1993) 1868.
[10] H.S. Rye, A.M. Roseman, S. Chen, K. Furtak, W.A. Fenton,
H.R. Saibil, A.L. Horwich, Cell 97 (1999) 325.
Fig. 3. Examples of force curves measured by a normal (a) and a
compression-free (b) method. Three typical force curves were
reproduced for each method applied to measure a biological
interaction. Denatured protein (pepsin) attached to the tip was
brought into contact with GroEL on the substrate and binding force
was measured during the retraction period.
492 H. Sekiguchi et al. / Applied Surface Science 188 (2002) 489–492