Steve Hollis1, Dave Strand2, Patrick Wheeler3
1Amgen, Inc., Cambridge, MA, 2Protasis, Corp., Marlboro, MA, 3Advanced
Chemistry Development Inc., Toronto, ON, Canada
Integrating Automation and Data Management
in Open-Access NMR
Better tools are always required to smooth the pathway from sample submission to
data acquisition, to knowledge, while integrating as many legacy systems as
possible. Amgen, Protasis and ACD present here a web-centric, enterprise-wide
automation software - NMR Visor, which supports complete NMR scheduling and
acquisition using IconNMR and standard Bruker auto-sampler hardware. NMR
data is automatically passed to ACD Automation Server where it is processed
using automated structure verification, and the results are distributed directly to the
submitter. Structures can be included with sample submission in standard
commercial formats and metadata is accessible through Active Directory or LDAP
services. Processed data, NMR report documents and structure searchable
databases are constructed for easy access to the chemists using web based
software tools.
Making Life Easier for Our Chemists
Chemists can submit crude reaction mixtures directly to the separations
group for automated purification and LC/MS/NMR characterization
Can submit up to 250 mg/sample dissolved in up to 2.5 mL of DMSO
Desired component should be > 50% purity by initial HPLC screen
Can submit single samples or entire library of up to 24 compounds
2
3
Automating NMR Data Collection and Analysis
for HTPP and Sample Registration at Amgen
HTPP Process is an automated mass-directed purification procedure that aids
chemists in their efforts to design and synthesize new therapeutic agents.
Chemists employ parallel synthesis to make focused libraries which are purified
and characterized using LC-MS and 1H NMR using Flow-NMR techniques.
Purified materials are provided in Sample-Bank vials and processed NMR data is
distributed to chemists in a convenient electronic format for e-Notebook records.
Design Synthesis ScreeningPurification Characterization Registration
HTPP/Singleton
Purification Service DAS Separations
Med ChemD S
P C
R
S Biology
Crude reaction mixtures
are submitted
MS-Directed
Purification
1. Product Vial
2. Solubility Vial
3. NMR Plate
4. LC/MS QC Plate
Output
General Procedure for the HTPP-NMR Process
Sample collection is triggered by Mass Spec and the selected product is
dried, weighed and formatted into Sample Bank and solubility vials and
two 96-well plates are prepared in automation using a Tecan Evo robot.
4
Mass Directed PurificationUV
MS
Purification Sample Drying Sample Reformatting
Genovac Tecan EvoWaters LC-MS
2 Vials per Sample
LC-MS QC Plate
NMR Plate0.5 mmol / well
2 Plates per Library
NMR Workflow
HTPP Library and Singleton Plates
• Metadata collected from separation process is organized by
Pipeline Pilot - saved as a SD file.
• SD files are downloaded to the One-Minute NMR computer on
the Protasis flow-NMR system – file used to run the NMR.
• SD Plate map is deconstructed into single-sample sdf’ s which
are embedded in the raw NMR data tree after data acquisition.
• Structures are included in mol format within the SD file.
• Completed data is copied to a sweep-folder for automated
processing by ACD/Labs Automation Server software.
5
Flow-NMR is reliable and reproducible with minimal downtime in practice: more
than 5000 plate samples are run per year with this Protasis system.
Samples are dissolved in 25 mL of dmso-d6 (20 mM) covered and placed on LEAP.
Samples and solvent are kept dry with closed system, using dry N2 and He blanket.
Greatly reduced sample requirements - 15 mL injections used to load samples.
HTPP Flow-NMR System - using a 500 MHz Bruker AVIII
with a Robotic Sampling Station
Protasis NMR
Probe with
10 mL flow-cell
LEAP
Pump/Solvent
Conditioner
Vial or Plate
Injector
Loop
25 mL
syringe
DMSO-d6
6
HTPP-NMR and Singleton Plates
Pipeline Pilot is used to download sample metadata for direct NMR
submissions. SD file is used to collect and process the NMR data.
7
HTPP-NMR and Singleton Plates
SD file from Pipeline Pilot is used to initiate the NMR data acquisition - assembles the combined metadata for each plate into a single file.
8
Running HTPP/Singleton-NMR in Automation
9
NMR is Setup by Shooting a Barcode
Reads
SD File
from
Pipeline Pilot
• SD file is imported directly
into One-Minute NMR
• Generates sample queue
• Click “Queue Run” to start the
NMR for the entire plate
SD File
Click Run
One-Minute NMR
9
HTPP-NMR and Singleton Plates
One-Minute NMR software reads each sample from the SD plate file “one
line at a time” and sends a single-sample sdf file to the NMR spectrometer.
• The individual sdf file is a simple formatted text file – it is saved in the
Bruker data tree in the same location where the fid is stored.
• When the data collection is complete the spectrometer software copies
the NMR file to a remote server in a sweep directory named AS_data.
• The appearance of the data file in the sweep directory triggers ACD/Labs
Automation Server to process, distribute and archive the data in a
searchable NMR database that is distributed to the NMR users.
Example of a Single-Sample sdf file – 117442-03-12.sdf
10
Automation Server Output - Processed-Data Folder
11
12
Processed Spectra are deposited in esp Folder
ACD/Spectrus Processor for processing and viewing spectra.
13
Example Output: Plate-Database Files (nd9)- in Table View
Click on spectrum to start processor
Browse
with
scroll tab
14
Plate-Database File - in Plate View
Click on a well - shows the
processed spectrum
Click on this spectrum -
starts the ACD-processor
15
Example Report – four page pdf
Data summary
for notebook in
journal format
16
Four Page Report – Expansions pages 2-3
17
Page 4 - Contains NMR Summary
NMR reports are deposited in the chemist’s reports folder and a copy is
sent to the inbox of their Electronic Notebook.
Review of the pdf report can often be a suitable for final review without the
need for spectrum reprocessing.
The HTPP process also distributes LC/MS data generated on all HTPP-
Singleton samples – data is sent directly to each users directory on LC/MS
file server.
18
LC/MS Reports Generated for Each Sample
Chromatogram with Mass Spec
Improving Open-Access NMR
• Use automated data processing features that were developed
for the HTPP/Singleton process in Open-Access NMR.
• Build new web tools to simplify the sample submission process:
- Structures can be added to sample submission to take advantage
. of the ACD processing tools.
- Auto-processing - peak-picking, scaling, integration and plotting
. all are improved when the structure is included with the data.
• NMR Visor - New sample-submittal/tracking software from Protasis
provides real time scheduling and tracking of all NMR submissions
over the network from the chemists desktop computers.
• Integrates with existing AS processing system and Bruker Iconnmr.
Building with Automation Server
19
NMR Visor – a New Tool for Open-Access NMR
NMR VisorScheduling Tool
Data processing uses
existing HTPP/Singleton
workflow
Runs
NMR
Logs in Sample
with Structure
ACD/Labs
Automation ServerOpen-Access NMR Network
Chemist
Reports
to e-LN
Spectrometer
Processing Done by ACD/Labs
• Integrated NMR output
• Creates pdf reports
• Structures included
• NMR database created
• Can search by structure
20
NMR Visor: Open-Access Submission and Tracking Tool
Can draw or attach structures, track projects, and more…
21
NMR Visor Architecture
22
NMR
Spectrometer Two
Spectrometer Three
Spectrometer N
Location Two
NMR Visor Architecture
Browser-based Web application
Attractive JavaScript Front-End
Distributed Across Enterprise
Secure Enterprise Network
Secure WebAPI Server & Oracle SQL Database
Secure, Remote Server(s)
Real-time Instrument Controller
Schedule/Monitor Experiments
NMR
RTIC
23
NMR Visor Architecture – Web Site
Enterprise Web Application
– JavaScript Framework• MVC Architecture
– Rich, responsive graphics
– Browsers supported:• IE 11
– Operating Systems:• Windows
• Mac
• Linux
– Devices supported:• Desktops
• Notebooks
COC1=CC2=C(C=CN=C2C=C1)
C(C3CC4CCN3CC4C=C)O
C20H24N2O2
324.41676
24
NMR Visor Architecture – Web API Server
Application DB Server
– WebAPI• REST Architecture
• Microsoft C# Controllers
– Oracle SQL Standard Edition
• Local or remote
• Consistency constraints
• Custom Fields
– Active Directory • Secure LDAP Authentication
• Windows Authentication
WebAPIC# REST
Controller
SQLsamples
LDAPdirectory
File
Serverspectra
CloudVPN
25
NMR Visor Architecture - Instrument Controller
Real-time scheduler
– Custom Priorities
– Overnight Samples
– Experiment Chains
Windows Application
– Runs on Spectrometer Computer (with Topspin)
– Network Interface
– Connects to DB
– Shared Drop Folder
26
Getting to the Answer – “Did I Make What I Thought?”
Submit Acquire Report
27
NMR Visor Starts by Showing Spectrometer Status
Submit a sample here
28
Login Windows Domain
Authenticate against
Amgen active directory
29
Select NMR Spectrometer
Add Sample ID and Choose Experiment Details with Dropdown Menus
The Project field will automatically
populate with the last Project used.
The Experiment field will automatically
populate with the last experiment that you
entered if it’s on the same spectrometer.
30
Add Structure with Plugin Tool
Draw freehand,
Drag ‘n Drop or
Copy/Paste
From ChemDraw or ELN
31
Add More Experiments or Additional Samples (optional)
32
Submit Sample
33
Put Sample in Holder and Confirm when Ready
After submitting, the samples are in the ‘Pending’ state.
Walk to the instrument and place the sample in the holder
Confirm sample – after placing in auto-sampler
34
Monitor the Sample Queue from Your Desk
• Sample is waiting to be run on the instrument
35
Review Completed Samples
Displays samples that have completed within the last 7 days.
If a user is logged in, list is filtered by that user.
Filtered by site, spectrometer, chemist.
36
Cancel a Sample in the Queue
Shows samples in the ‘Pending’ and ‘Waiting’ states.
Buttons:
Moves samples/experiments to the ‘Cancelled’
state. Cancelled samples/experiments are
shown on the “Resubmit Failed” tab.
Permanently deletes the samples/experiments.
Role restrictions:
• Users with the NMR Visor ‘Admin’ role will see everyone samples listed here.
• Users with the NMR Visor ‘Submitter’ role will see only their own samples here.
37
Archive Sample Context
Spectra saved on spectrometer
Sample record stored in secure database (remote or local)
Molecule structure saved in multiple formats (.sk2, .mol, .cdx, .smi)
Spectra transferred to ACD AS for Processing
– Bruker AU script support
Quinine 01080910482D 25 28 0 0 0 0 999 V2000 -0.7145 -0.8250 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 -1.4289 -0.4125 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 -1.4289 0.4125 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 -0.7145 0.8250 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 0.0000 0.4125 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 0.0000 -0.4125 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 0.7145 -0.8250 0.0000 N 0 0 0 0 0 0 0 0 0 0 0 0 0.7145 0.8250 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 1.4289 0.4125 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 1.4289 -0.4125 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 1.4289 2.8875 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 2.8578 2.0626 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 2.8578 2.8876 0.0000 C 0 0 2 0 0 0 0 0 0 0 0 0 2.5559 2.5857 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 3.5723 3.3001 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 2.1434 3.3001 0.0000 C 0 0 1 0 0 0 0 0 0 0 0 0 2.1434 1.6501 0.0000 N 0 0 2 0 0 0 0 0 0 0 0 0 2.5560 0.9357 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 4.2868 2.8876 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 1.4289 2.0625 0.0000 C 0 0 2 0 0 0 0 0 0 0 0 0 0.7144 1.6500 0.0000 C 0 0 2 0 0 0 0 0 0 0 0 0 0.0000 2.0625 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0 -2.1435 0.8250 0.0000 O 0 0 0 0 0 0 0 0 0 0 0 0 1.4289 1.2375 0.0000 H 0 0 0 0 0 0 0 0 0 0 0 0 -2.8579 0.4125 0.0000 C 0 0 0 0 0 0 0 0 0 0 0 0 4 3 1 0 0 0 0 3 23 1 0 0 0 0 3 2 2 0 0 0 0 1 2 1 0 0 0 0 5 4 2 0 0 0 0 6 5 1 0 0 0 0 1 6 2 0 0 0 0 7 10 2 0 0 0 0 6 7 1 0 0 0 0 8 5 1 0 0 0 0 21 8 1 0 0 0 0 10 9 1 0 0 0 0 8 9 2 0 0 0 0 20 21 1 0 0 0 0 11 20 1 0 0 0 0 20 17 1 0 0 0 0 16 11 1 0 0 0 0 12 13 1 0 0 0 0 16 13 1 0 0 0 0 13 15 1 1 0 0 0 16 14 1 6 0 0 0 14 18 1 0 0 0 0 15 19 2 0 0 0 0 17 18 1 6 0 0 0 21 22 1 1 0 0 0 20 24 1 1 0 0 0 17 12 1 0 0 0 0 23 25 1 0 0 0 0M END
38
Conclusions
Flow and tube NMR offer complementary advantages. A common
processing and reporting approach allows each sample format to be
used to full advantage.
Structures are captured during sample submission so they are available
for downstream processing and reporting.
Structural graphics add visual information to improve the final report,
helping chemists to confirm and interpret spectra and prevent errors.
ACD/Labs software has proven to be a versatile and reliable tool for
processing NMR spectra in automation – using both flow and
tube-based NMR techniques.
SQLsamples
An integrated NMR sample history database is
fully integrated into the sample management
system to improve automation and provide NMR
use statistics.
39
I Made What I Thought I Made!
40
Acknowledgements
Wolfgang Goetzinger, Amgen
Ashley Nichols, Amgen
Paul Krolikowski, Amgen
Grace Bi, Amgen
Larry Miller, Amgen
Peter Grandsard, Amgen
Craig Milling, Protasis
Igor Teslya, ACD/Labs
Dave Adams, ACD/Labs
41