1
Neurotransmitters in the Striatum Tissues from Animal Brain: UPLC-MS/MS Method Validation and Application to Animal Models of Parkinson’s Disease Changyu Quang 1 , Brett D. Dunbar 1 ,Carina Witters-Teulings 1 , Nichole R. Myers 1 , William C. Nethero 1 , Melissa J. Beck 1 , Michael D. Schlosser 1 , and Kuldip D. Dave 2 1 WIL Research Laboratories, LLC, 140 7 George Road, Ashland, Ohio, USA 2 ; The Michael J. Fox Foundation for Parkinson’s Research, 498 Seventh Avenue, 18 th New York, NY, USA 60 th ASMS Conference Results and Discussion l Bioanalytical method validation High performance liquid chromatography with tandem mass spectrometry detection (HPLC-MS/MS) has an extensive history as a quantitative tool for small molecules in biological matrices. The quantitative determination of endogenous (i.e., naturally occurring) compounds in biological samples of limited size (rat or mouse striatum tissues) is more complicated, both analytically and from method validation point of view. We have previously shown that UPLC-MS/MS provides sensitive and selective measurements of dopamine, serotonin and metabolites in the animal striatum tissues homogenates 1 . To demonstrate the validity of a bioanalytical method, we have shown that charcoal treated (2X) animal (rat and mouse) brain homogenates were suitable control matrices for the study samples (striatum tissue homogentates) 1 . Dopamine, serotonin and their metabolites are small polar molecules and the use of more polar pentaflurophenylpropyl (PFP) stationary phase provided adequate retentions without using ion- pairing reagents. While monitoring phospholipids using an MRM transition m/z 184/184, we observed a phosphatidylcholine peak at an extremely high abundance in tissue homogenates samples, but not in calibration standards. Figure 3 shows that phosphatidylcholine derivatives can be completely eliminated using a Supelco HybrideSPE filter plate, which improves the PFP column performance and method robustness significantly. Additionally, Whatman filter plates were used for processing samples for UPLC-MS/MS analysis of DOPAC and HVA in negative ion MRM mode because of low recoveries and instability of DOPAC when using Supelco lipids plates. Meanwhile, for the analysis of acidic DOPAC and HVA, the best UPLC-MS/MS conditions were obtained on a Waters HSS T3 column (2.1x50 mm) using 0.1 % acetic acid in water and acetonitrile mobile phases. Note that Whatman filter plates which partially remove phosphatidylcholine derivatives were used for processing all mouse tissue homogenates due to limited sample volume. With the suitable control matrices, the above UPLC-MS/MS method for endogenous neurotransmitters was validated following FDA Guidance on Bioanalytical Method Validation. Validation results are summarized in Table 2 through Table 5 for the 6 neurotransmitters in mouse and rat brain homogenates, respectively. Method precision and accuracy were acceptable based on the WIL Research SOP established criteria. Other validation parameters including matrix effects, recovery, and matrix and stock stabilities were also assessed during our method validation efforts. Method application to study sample analysis The LC-MS/MS method described above was used for the quantitative analysis of 6 neurotransmitters in animal striatum tissue samples collected from a large number of animal models of Parkinson’s disease at various ages. Figure 4 shows representative MRM chromatograms of mouse striatum homogenate with endogenous levels of neurotransmitters. Figure 5 presents typical calibration standard curves in duplicates of the six neurotransmitters prepared in 0.1% formic acid. Figure 6 shows the internal standard (IS) responses of stable isotope labeled neurotransmitters. The IS responses were relatively consistent throughout analysis batch runs for calibrators (in 0.1% formic acid), QCs (in surrogate matrix), and study samples, indicating that matrix effects were insignificant and that reliable concentrations could be obtained with the method. UPLC-MS/MS analysis results for all study samples collected so far in the on-going studies are shown in Figure 7 (mouse striatum tissues) and Figure 8 (rat striatum tissues). Overall, the endogenous concentration levels in study samples are within the calibration ranges of the assay for 5 neurotransmitters (DA, 5-HT, 5-HIAA, DOPAC, and HVA). The concentrations of NE were detectable, but below the lower limit of quantitation (LLOQ) in most of the study samples. As dopamine metabolism is studied in a separate mouse model study, Figure 9 shows the analysis data of mouse brain homogenates from a mouse model study. In this separate study, the homogenates of mouse whole brain tissues were analyzed for DA and its metabolites (DOPAC and HVA). Unfortunately, large variations of neurotransmitter concentrations were observed among study samples even though the described assay demonstrated good precision and accuracy. The temporal and heterogenic nature of neurotransmitters in the animal brain locations and their chemical instability may play significant roles in assay variations. Therefore, it is very important to improve the precision and accuracy both in sample collection and sample processing steps when dealing with a large number of animal study samples. Conclusions and future work A simple, sensitive, specific, and robust UPLC-MS/MS method was validated with good linearity, precision and accuracy for the determination of endogenous dopamine, serotonin and their acidic and basic metabolites in animal striatum tissues. The method has been used successfully to assay samples from animal models of Parkinson’s disease and from a separate animal model study. Recently, we have extended the work to include other classes of neurotransmitters such as amino acids, acetylcholine, g-aminobutyric acid (GABA), and monoamines (Figure 10) in animal brain tissues and in microdialysis fluids using UPLC-MS/MS with new generations of mixed-mode chromatographic stationary phases. References l 1. C. Quang et al., 59 th Conference on Mass Spectrometry and Allied Topics (2011), WP 670. 2. Ewa Balkowice-Iskra et al., Acta Neurobiol. Exp. 2007, 67:379-388 3. E. H. Stotz et al., Brain Research, 606 (1993) 267-272 4. Hitoshi Nohta et al., Analytical Sciences, August 1993, Vol. 9, 537-540. 5. Mark M. Kushnir et al., Clinical Chemistry 48(2), 2002 323-331 6. Paivi Uutela et al., Anal. Chem. 81 (2009) 8417-8425 Acknowledgments The authors acknowledge the excellent technical support provided by Alison Johnson, Robyn Eggleston, Ryan Scavinski, and other staff of WIL Research Laboratories, LLC. Novel Aspect l Validated UPLC-MS/MS methods for determining dopamine, serotonin, and their metabolites in the striatum tissues from animal models of Parkinson’s disease. Abstract l Simultaneously determining dopamine, serotonin and their metabolites in animal striatum tissues has been a challenge as their concentrations are low, sample size is limited, and many polar endogenous compounds may interfere with their analyses. Because of its unique features of sensitivity, selectivity, and speed (3S), ultra performance liquid chromatography with tandem mass spectrometry detection (UPLC-MS/MS) is increasingly used for the quantitation of biological compounds including small polar neurotransmitters. In this presentation, we describe UPLC-MS/ MS method validation and application to the analysis of dopamine, serotonin and their metabolites in the striatum tissues from animal models of Parkinson’s disease. The striatum tissue samples were surgically removed from euthanized animals and immediately flash frozen with liquid nitrogen and stored frozen at approximately –70°C until processing. The tissue samples were homogenized in 0.1% formic acid aqueous solution in a ratio of 1:9 [weight (mg) : mL]. In the positive ion multiple reaction monitoring (MRM) mode for norepinephrine (NE), dopamine (DA), serotonin (5-HT), and 5-hydroxyindole-3-acetic acid (5-HIAA), the sample homogenates were filtered through a Supelco HybrideSPE plate and separated on a Restek Ultra II PFP Propyl column (1.9 mm, 50 x 2.1 mm) using mobile phases of 0.1% formic acid in water and in acetonitrile with a 4-minute gradient elution. In the negative ion MRM mode for 3,4-dihydoxyphenylacetic acid (DOPAC) and homovanillic acid (HVA), the homogenates were filtered through a Whatman filter plate and separated on a Waters HSS T3 column (1.7 mm 50 x 2.1 mm) using mobile phases of 0.1% acetic acid in water and in acetonitrile in a 4-minute run time. Stable isotope labeled internal standards were used for the 6 endogenous analytes to minimize the matrix effects on their quantitation. The calibration standards were prepared in 0.1% formic acid aqueous solution with ranges from 6.00 to 1200 ng/mL for DA and 5-HT, and from 12.0 to 2400 ng/mL for NE, 5-HIAA, DOPAC, and HVA. The method has been validated using quality control samples prepared in a surrogate matrix, charcoal treated (2X) animal brain homogenates. Finally the validated method has been routinely used to analyze study samples from various animal models. Introduction l Research in neuroscience and neurodegenerative diseases has been translational in nature, i.e., laboratory animals have been used for investigating biological processes, studying the causes of diseases, and testing new therapeutic means of treating and preventing them. The capability of producing transgenic, knockout, mutation cell lines of animal species has given researches many models linked to human diseases. As small-molecule biomarkers are commonly monitored in neurochemistry studies, reliable bioanalytical methods with high sample throughput are needed for the quantitation of the endogenous biomarkers in biological matrices. We describe our continuing efforts of developing and validating a UPLC-MS/MS method for determining dopamine, serotonin, and their metabolites in striatum tissues from animal (i.e., rat and mouse) models of Parkinson’s disease 1 . Neurotransmitters have been analyzed by using liquid chromatography coupled with electrochemical detection (LC-ECD) 2,3 , fluorescence detection (LC-FL) 4 or lately electrospray tandem ionization mass spectrometric (ESI-MS/MS) detection 5-6 . The lack of method specificity and sensitivity of LC-ED and FC-FL methods resulted in differences in concentrations of various neurotransmitters reported in literature. The analysis time per sample injection is typically long (20–30 minutes) using the nonspecific LC-ED and LC-FL methods. We describe here a simple, specific, and sensitive UPLC-MS/MS method to quantify dopamine, serotonin and their acidic and basic metabolites in small striatum tissues. XIC of +MRM (17 pairs): 146.1/87.0 amu from Sample 72 (7 025 WIL-784002 30527 1 Right BRSH-2 1) of I5-784002_add analyte.wiff (T... Max. 7.4e5 cps. 1 2 3 4 5 6 7 8 9 Time, min 0.0 5.0e5 1.0e6 1.5e6 2.0e6 2.5e6 3.0e6 3.5e6 4.0e6 4.5e6 5.0e6 5.5e6 6.0e6 6.3e6 Intensity, cps 4.82 4.21 GABA 5-HT DA ACh Glu Asp 5-HIAA PCh iso-ACh XIC of +MRM (9 pairs): 152.0/107.0 amu from Sample 36 (HBT BLK IS) of Supelco LRM.wiff (Turbo Spray) Max. 7920.0 cps. 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Time, min 0.0 2.0e5 4.0e5 6.0e5 8.0e5 1.0e6 1.2e6 1.4e6 1.6e6 1.8e6 2.0e6 Intensity, cps XIC of +MRM (9 pairs): 152.0/107.0 amu from Sample 32 (HBT BLK IS) of Supelco LRM.wiff (Turbo Spray) Max. 4440.0 cps. 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Time, min 0.0 1.0e5 2.0e5 3.0e5 4.0e5 5.0e5 6.0e5 7.0e5 8.0e5 Intensity, cps 1.52 2.86 3.28 2.11 1.32 2.63 1.63 1.91 2.31 3.02 1.22 3.09 3.47 0.72 0.55 2.54 1.78 0.82 0.16 3.71 0.41 0.38 0.89 phosphatidylcholine m/z 184 XIC of +MRM (8 pairs): 152.0/107.0 amu from Sample 37 (10 034 WIL-784001 7877 8 Left BRH-1 1) of I5-0010.wiff (Turbo Spray) Max. 2.2e4 cps. 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Time, min 0.0 2.0e4 4.0e4 6.0e4 8.0e4 1.0e5 1.2e5 1.4e5 1.6e5 1.8e5 2.0e5 2.1e5 Intensity, cps 0.83 XIC of -MRM (4 pairs): 167.0/123.0 amu from Sample 39 (11 034 WIL-784001 7877 8 Left BRH-1 1) of I5-0011.wiff (Turbo Spray) Max. 1.9e5 cps. 0.5 1.0 1.5 2.0 2.5 3.0 3.5 Time, min 0.0 2.0e4 4.0e4 6.0e4 8.0e4 1.0e5 1.2e5 1.4e5 1.6e5 1.8e5 2.0e5 Intensity, cps 1.63 0.28 3.47 NE DA 5-HIAA 5-HT DOPAC HVA I5-784001#7p.rdb (NE): "Quadratic" Regression ("1 / (x * x)" weighting): y = -3.42e-008 x^2 + 0.00552 x + 0.101 (r = 0.9845) 0 500 1000 1500 2000 Analyte Conc. / IS Conc. 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 Analyte Area / IS Area I5-784001#7p.rdb (Dopamine): "Quadratic" Regression ("1 / (x * x)" weighting): y = -1.76e-007 x^2 + 0.00929 x + -0.00368 (r = 0.9997) 0 200 400 600 800 1000 1200 Analyte Conc. / IS Conc. 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 Analyte Area / IS Area I5-784001#7p.rdb (Serotinin): "Quadratic" Regression ("1 / (x * x)" weighting): y = 2.68e-007 x^2 + 0.0117 x + -0.00588 (r = 0.9993) 0 200 400 600 800 1000 1200 Analyte Conc. / IS Conc. 0 2 4 6 8 10 12 14 15 Analyte Area / IS Area I5-784001#7p.rdb (5-HIAA): "Quadratic" Regression ("1 / (x * x)" weighting): y = 1.72e-007 x^2 + 0.0301 x + -0.0128 (r = 0.9996) 0 500 1000 1500 2000 Analyte Conc. / IS Conc. 0 10 20 30 40 50 60 70 75 Analyte Area / IS Area I5-784001#8n.rdb (DOPAC): "Quadratic" Regression ("1 / (x * x)" weighting): y = 1.93e-008 x^2 + 0.00166 x + -0.00303 (r = 0.9967) 0 500 1000 1500 2000 Analyte Conc. / IS Conc. 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 Analyte Area / IS Area I5-784001#8n.rdb (HVA): "Quadratic" Regression ("1 / (x * x)" weighting): y = -1.08e-008 x^2 + 0.00254 x + -0.000964 (r = 0.9960) 0 500 1000 1500 2000 Analyte Conc. / IS Conc. 0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 Analyte Area / IS Area NE (12-2400 ng/mL) DA (6-1200 ng/mL) 5-HT (6-1200 ng/mL) 5-HIAA (12-2400 ng/mL) DOPAC (12-2400 ng/mL) HVA (12-2400 ng/mL) IS Peak Area vs Index (NE) 0 20 40 60 80 100 120 140 160 180 200 220 240 Index 0.0 1.0e4 2.0e4 3.0e4 4.0e4 5.0e4 6.0e4 7.0e4 8.0e4 IS Peak Area, counts IS Peak Area vs Index (Dopamine) 20 40 60 80 100 120 140 160 180 200 220 240 Index 0.0 2.0e4 4.0e4 6.0e4 8.0e4 1.0e5 1.2e5 1.4e5 1.6e5 IS Peak Area, counts IS Peak Area vs Index (Serotinin) 20 40 60 80 100 120 140 160 180 200 220 240 0.0 5000.0 1.0e4 1.5e4 2.0e4 2.5e4 3.0e4 3.5e4 IS Peak Area, counts IS Peak Area vs Index (5-HIAA) 20 40 60 80 100 120 140 160 180 200 220 240 0.0 2000.0 4000.0 6000.0 8000.0 1.0e4 1.2e4 1.4e4 1.6e4 1.8e4 2.0e4 2.2e4 2.4e4 2.6e4 IS Peak Area, counts IS Peak Area vs Index (DOPAC) 10 20 30 40 50 60 70 80 90 100 110 120 Index 0.0 5.0e4 1.0e5 1.5e5 2.0e5 2.5e5 3.0e5 3.5e5 4.0e5 4.5e5 IS Peak Area, counts IS Peak Area vs Index (HVA) 10 20 30 40 50 60 70 80 90 100 110 120 Index 0.0 5000.0 1.0e4 1.5e4 2.0e4 2.5e4 3.0e4 3.5e4 4.0e4 4.5e4 IS Peak Area, counts NE-d6 DA-d4 5-HT-d4 5-HIAA-d5 DOPAC-d5 HVA-d4 Mouse Brain Homogenate Conc. (ng/mL) Sample number Figure 9. Concentrations (ng/mg tissue) of neurotransmitters in mouse brain tissues collected from study animals. Figure 8. Concentrations (ng/mg tissue) of neurotransmitters in mouse brain tissues collected from study animals. Figure 7. Concentrations (ng/mg tissue) of neurotransmitters in mouse striatum tissues collected from animals of Parkinson’s disease models at various ages. Figure 6. Batch internal standard responses of stable isotope labeled compounds: NE-d6, DA-d4, 5-HT-d4, 5-HIAA-d5, DOPAC-d5, and HVA-d4. () calibrators, () quality control samples, and () study samples. Figure 5. Calibration standard curves of NE (12-2400 ng/mL), DA (5-1200 ng/mL), 5-HIAA (12-2400 ng/mL), 5-HT (6-1200 ng/mL), DOPAC (12-2400 ng/ mL), and HVA (24-2400 ng/mL). The curves were prepared in aqueous 0.1% formic acid solution. Figure 4. UPLC-MS/MS chromatograms of a mouse striatum homogenate (Top) in positive ion MRM mode and (Bottom) in negative ion MRM mode. Stable isotope-labeled internal standards were added to the sample. Figure 3. Representative UPLC-MS/MS chromatograms of mouse striatum homogenates. (Top) the homogenate sample without phospholipids removal, (Bottom) the sample extract filtered with a HybridSPE phospholipids plate. Figure 10. UPLC-MS/MS chromatogram of mouse striatum homogenate. Chromatographic column: Scherzo SS-C18 (150x3 mm, 3µm). Mobile phases: A: 0.5% formic acid, B: 0.5% formic acid and 50 mM ammonium formate in acetonitrile/water (50/50, v/v). Figure 2. Chemical structures and metabolic pathways of (A) dopamine, (B) serotonin, and their acidic and basic metabolites. Figure 1. Dopamine signaling in Parkinson’s disease (Source: Cell Signaling Technology, www.cellsignal.com). Table 4. Calibration Standard Statistics (Rat Striatum Assay) DA (ng/mL) 6.00 12.0 60.0 120 300 600 1080 1200 Mean 5.90 12.5 58.7 120 301 598 1090 1190 %RSD 1.60 3.30 2.50 2.30 1.70 3.40 1.20 1.40 %RE -1.67 4.17 -2.17 0.00 0.33 -0.33 0.93 -0.83 n 10 10 10 10 10 10 10 10 DOPAC (mg/mL) 12.0 24.0 120 240 600 1200 2160 2400 Mean 11.8 24.9 117 239 599 1220 2130 2420 %RSD 3.1 5.7 5.8 4.2 2.9 5.2 2. 8 2.4 %RE -1.67 3.75 -2.50 -0.42 -0.17 1.67 -1.39 0.83 n 9 9 9 9 9 9 9 9 HVA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00 Mean 11.9 24.7 118 238 602 1220 2150 2400 %RSD 1.9 4.4 5.1 3.9 6.6 2.0 2. 5 2.9 %RE -0.83 2.92 -1.67 -0.83 0.33 1.67 -0.46 0.00 n 11 11 11 11 11 11 11 11 5-HT (ng/mL) 6.00 12.00 60.00 120.00 300.00 600.00 1080.00 1200.00 Mean 6.03 12.0 59.0 120 303 608 1090 1190 %RSD 2.8 6.1 3.5 2.8 3.6 2.2 1. 5 1.5 %RE 0.50 0.00 -1.67 0.00 1.00 1.33 0.93 -0.83 n 11 11 11 10 11 11 11 11 5-HIAA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00 Mean 12.0 24.3 118 241 601 1210 2140 2410 %RSD 1.5 3.4 2.6 2.4 2.4 3.3 2. 9 2.7 %RE 0.00 1.25 -1.67 0.42 0.17 0.83 -0.93 0.42 n 11 11 11 11 11 11 11 11 NE (ng/mL) 12.0 24.0 120 240 600 1200 2160 2400 Mean 11.9 24.7 119 238 602 1210 2170 2380 %RSD 1.9 5.1 4.0 3.9 3.2 3.4 1. 7 1.9 %RE -0.83 2.92 -0.83 -0.83 0.33 0.83 0.46 -0.83 n 10 10 10 10 10 10 10 10 Table 5. Precision and Accuracy of Quality Controle Samples (Rat Striatum Assay) 6.00 18.0 600 960 6.00 18.0 600 960 Mean 5.67 17.4 583 937 6.33 17.6 588 942 %RSD 3.4 5.0 3.3 3.1 5.5 4.9 4. .8 %RE -5.5 -3.3 -2.8 -2.4 5.5 -2.2 -2.0 -1.9 n 4 10 10 10 5 11 1 1 12.0 36.0 1200 1920 12.0 36.0 1200 1920 Mean 12.9 36.3 1180 1870 11.7 35.4 1200 1920 %RSD 9.0 4.5 4.2 3.5 4.5 4.8 3. .2 %RE 7.5 0.8 -1.7 -2.6 -2.5 -1.7 0.0 0.0 n 5 11 11 11 4 10 1 0 12.0 36.0 1200 1920 12.0 36.0 1200 1920 Mean 12.3 38.9 1240 2000 11.6 36.1 1170 1860 %RSE 13.6 11 5.3 4.1 10.9 6.5 5. .2 %RE 2.5 8.1 3.3 4.2 -3.3 0.3 -2.5 -3.1 n 5 9 9 9 6 11 11 11 DA (ng/mL -HT (ng/mL) 5-HIAA (ng/mL E (ng/mL) DOPAC (ng/mL VA (ng/mL) Tables 4–5. Validation results of rat homogenate assay Table 2. Calibration Standard Statistics (Mouse Striatum Assay) DA (ng/mL) 6.00 12.0 60.0 120 300 600 1080 1200 Mean 6.02 11.9 59.8 119 305 604 1060 1210 %RSD 1.00 4.20 4.00 4.70 2.70 2.40 2.60 2.90 %RE 0.33 -0.83 -0.33 -0.83 1.67 0.67 -1.8 .83 n 10 10 10 10 10 10 1 0 DOPAC (mg/mL) 12.0 24.0 120 240 600 1200 2160 2400 Mean 12.0 24.0 119 237 611 1210 2150 2400 %RSD 1.8 5.3 4.0 4.7 3.8 3.9 2. .4 %RE 0.00 0.00 -0.83 -1.25 1.83 0.83 -0.46 0.00 n 9 9 9 9 9 9 9 9 HVA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00 Mean 12.0 24.1 120 239 603 1210 2120 2440 %RSD 2.0 6.0 1.9 4.2 3.4 3.0 2. .0 %RE 0.00 0.42 0.00 -0.42 0.50 0.83 -1.8 .67 n 9 9 9 9 9 9 9 9 5-HT (ng/mL) 6.00 12.00 60.00 120.00 300.00 600.00 1080.00 1200.00 Mean 5.99 12.2 58.7 121 304 603 1070 1200 %RSD 2.2 4.9 2.7 3.3 3.8 1.9 4. .8 %RE -0.17 1.67 -2.17 0.83 1.33 0.50 -0.93 0.00 n 11 11 11 10 11 11 1 1 5-HIAA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00 Mean 12.1 23.8 119 238 610 1210 2150 2400 %RSD 1.8 4.0 3.2 5.8 3.6 3.4 2. .2 %RE 0.83 -0.83 -0.83 -0.83 1.67 0.83 -0.4 .00 n 9 9 9 9 9 9 9 9 NE (ng/mL) 12.0 24.0 120 240 600 1200 2160 2400 Mean 12.1 23.9 121 236 605 1210 2160 2400 %RSD 1.4 3.7 3.4 3.8 1.8 2.1 2. .0 %RE 0.83 -0.42 0.83 -1.67 0.83 0.83 0.0 .00 n 9 9 9 9 9 9 9 9 Table 3. Precision and Accuracy of Quality Controle Samples (Mouse Striatum Assay) 6.00 18.0 600 960 6.00 18.0 600 960 Mean 5.59 18.1 574 911 6.27 19.2 572 907 %RSD 7.5 11.2 6.6 6.5 8.5 10.1 6.1 6.2 %RE -6.8 0.6 -4.3 -5.1 4.5 6.7 -4.7 -5.5 n 6 10 10 10 6 11 1 1 12.0 36.0 1200 1920 12.0 36.0 1200 1920 Mean 13.3 37.9 1190 1830 10.7 33.8 1170 1880 %RSD 6.4 8.5 4.5 5.4 9.4 24.7 7.3 6.1 %RE 10.8 5.3 -0.8 -4.7 -10.8 -6.1 -2. 2.1 n 4 9 9 9 4 9 9 9 12.0 36.0 1200 1920 12.0 36.0 1200 1920 Mean 12.6 38.3 1250 1960 11 36.2 1200 1880 %RSE 7.8 11 5.2 5.3 12.9 8.5 6. .8 %RE 5.0 6.4 4.2 2.1 -8.3 0.6 0. 2.1 n 5 9 9 9 5 9 9 9 DA (ng/mL) 5-HIAA (ng/mL E (ng/mL) DOPAC (ng/mL VA (ng/mL) 5-HT (ng/mL) Tables 2–3. Validation results of mouse homogenate assay. Table 1. Animal models of Parkinson’s disease Evaluated at three ages (4, 8, and 12 months). Experimental Section l Calibration standards and quality control samples: Stock solutions of neurotransmitters in Figure 2 were prepared in ice-cold 0.1% formic acid at a concentration of 1.00 mg/mL, and the stocks were subaliquoted into tubes (0.50 mL each) and stored at -80°C. Calibration standards were prepared from the stock solutions with ice-cold 0.1% formic acid at various concentrations on the day of the analysis. Quality control samples were prepared in charcoal treated (2X) rat and mouse brain homogenates spiked with neurotransmitters at four concentration levels (LLQC, LQC, MQC, HQC). Quality control samples were subaliquoted into tubes and stored at -80°C. Preparation of study samples: Striatum tissues were surgically removed from euthanized animals. After weighing, tissue samples were immediately flash frozen with liquid nitrogen and stored frozen at approximately -80°C until sample processing. During analysis, tissue samples were homogenized on wet ice with ice-cold 0.1% formic acid using a Kontes microgrinder at a ratio of 1:9 [mg (weight): mL]. For the analysis of rat striatum samples, aliquots of 50.0 mL of homogenates were added to a 96-well sample plate (2-mL well size), followed by 50 mL of 0.1% formic acid solution containing internal standards. The samples were further diluted with 350 mL of ice cold 0.1% formic acid (total 450 mL). Aliquots (200 mL) of the diluted samples were transferred and filtered through a Supelco HybridSPE® plate for the positive ion analysis (NE, DA, 5-HT, and 5-HIAA). Second portions (200 mL) of the diluted samples were filtered through a Whatman filter plate for the negative ion analysis (DOPAC and HVA). Because of limited sample size for mouse striatum tissues (6–28 mg), Whatman filter plates were used to prepare the diluted homogenates for both positive ion and negative ion modes of analysis. Dopamine (a catecholamine), serotonin and their metabolites are known to be unstable in the presence of light, oxygen, metal ions, and low or high pH, as well as sensitive to temperature. Samples should be processed on ice quickly in 0.1% formic acid. Proper collection, processing, and storage of tissue samples are important for generating accurate results. Triple-stage quadrupole mass spectrometer (API-4000) coupled with Waters Acquity UPLC® systems were used for UPLC-MS/MS analysis. UPLC conditions (the positive ion mode): LC System: Waters Acquity UPLC® System Column: Restek Ultra™ PFP-Propyl 50 x 2.1 mm, 1.9 µm Column Temperature: 60°C Autosampler Temperature: 4°C Mobile Phase A: 0.1% Formic acid in water Mobile Phase B: 0.1% Formic acid on acetonitrile Strong Wash: 0.1% Formic acid in acetonitrile Weak Wash: 0.1% Formic acid in water Seal Wash: 10% Acetonitrile UPLC conditions (the negative ion mode): LC System: Waters Acquity UPLC® System Column: Waters HSS T3, 50 x 2.1 mm, 1.7 µm Column Temperature: 60°C Autosampler Temperature: 4°C Mobile Phase A: 0.1% Acetic acid in water Mobile Phase B: 0.1% Acetic acid on acetonitrile Strong Wash: 0.1% Acetic acid in acetonitrile Weak Wash: 0.1% Acetic acid in water Seal Wash: 10% Acetonitrile UPLC gradient program Time (min) A% B % Flow rate (mL/min) Initial 100 0 0.40 0.50 100 0 0.40 2.60 70 30 0.40 2.70 10 90 1.00 3.30 10 90 1.00 3.31 100 0 1.00 4.00 100 0 1.00 Mass spectrometric detection parameters Compound Name Precursor Ion m/z Product Ion m/z Ionization Mode NE 152 107 Positive DA 154 91 Positive 5-HT 177 115 Positive 5-HIAA 192 146 Positive NE-d6 158 111 Positive DA-d4 158 95 Positive 5-HT-d4 181 118 Positive 5-HIAA-d5 197 150 Positive UPLC gradient program Time (min) A% B % Flow rate (mL/min) Initial 100 0 0.40 2.50 75 25 0.40 2.60 5 95 1.00 3.20 5 95 1.00 3.21 5 95 1.00 3.30 100 0 1.00 4.00 100 0 1.00 Mass spectrometric detection parameters Compound Name Precursor Ion m/z Product Ion m/z Ionization Mode DOPAC 167 123 Negative HVA 181 122 Negative DOPAC-d5 172 128 Negative HVA-d5 186 127 Negative

Neurotransmitters in the Striatum Tissues from Animal Brain ...USA-WilResearch/...Neurotransmitters in the Striatum Tissues from Animal Brain: UPLC-MS/MS Method Validation and Application

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Neurotransmitters in the Striatum Tissues from Animal Brain ...USA-WilResearch/...Neurotransmitters in the Striatum Tissues from Animal Brain: UPLC-MS/MS Method Validation and Application

Neurotransmitters in the Striatum Tissues from Animal Brain: UPLC-MS/MS Method Validation and Application to Animal Models of Parkinson’s Disease

Changyu Quang1, Brett D. Dunbar1,Carina Witters-Teulings1, Nichole R. Myers1, William C. Nethero1, Melissa J. Beck1, Michael D. Schlosser1, and Kuldip D. Dave2

1 WIL Research Laboratories, LLC, 1407 George Road, Ashland, Ohio, USA2; The Michael J. Fox Foundation for Parkinson’s Research, 498 Seventh Avenue, 18th New York, NY, USA

60th ASMS Conference

Results and Discussion lBioanalytical method validation

High performance liquid chromatography with tandem mass spectrometry detection (HPLC-MS/MS) has an extensive history as a quantitative tool for small molecules in biological matrices. The quantitative determination of endogenous (i.e., naturally occurring) compounds in biological samples of limited size (rat or mouse striatum tissues) is more complicated, both analytically and from method validation point of view. We have previously shown that UPLC-MS/MS provides sensitive and selective measurements of dopamine, serotonin and metabolites in the animal striatum tissues homogenates1. To demonstrate the validity of a bioanalytical method, we have shown that charcoal treated (2X) animal (rat and mouse) brain homogenates were suitable control matrices for the study samples (striatum tissue homogentates)1. Dopamine, serotonin and their metabolites are small polar molecules and the use of more polar pentaflurophenylpropyl (PFP) stationary phase provided adequate retentions without using ion-pairing reagents. While monitoring phospholipids using an MRM transition m/z 184/184, we observed a phosphatidylcholine peak at an extremely high abundance in tissue homogenates samples, but not in calibration standards. Figure 3 shows that phosphatidylcholine derivatives can be completely eliminated using a Supelco HybrideSPE filter plate, which improves the PFP column performance and method robustness significantly. Additionally, Whatman filter plates were used for processing samples for UPLC-MS/MS analysis of DOPAC and HVA in negative ion MRM mode because of low recoveries and instability of DOPAC when using Supelco lipids plates. Meanwhile, for the analysis of acidic DOPAC and HVA, the best UPLC-MS/MS conditions were obtained on a Waters HSS T3 column (2.1x50 mm) using 0.1 % acetic acid in water and acetonitrile mobile phases. Note that Whatman filter plates which partially remove phosphatidylcholine derivatives were used for processing all mouse tissue homogenates due to limited sample volume.

With the suitable control matrices, the above UPLC-MS/MS method for endogenous neurotransmitters was validated following FDA Guidance on Bioanalytical Method Validation. Validation results are summarized in Table 2 through Table 5 for the 6 neurotransmitters in mouse and rat brain homogenates, respectively. Method precision and accuracy were acceptable based on the WIL Research SOP established criteria. Other validation parameters including matrix effects, recovery, and matrix and stock stabilities were also assessed during our method validation efforts.

Method application to study sample analysisThe LC-MS/MS method described above was used for the quantitative analysis of 6 neurotransmitters in animal striatum tissue samples collected from a large number of animal models of Parkinson’s disease at various ages. Figure 4 shows representative MRM chromatograms of mouse striatum homogenate with endogenous levels of neurotransmitters. Figure 5 presents typical calibration standard curves in duplicates of the six neurotransmitters prepared in 0.1% formic acid. Figure 6 shows the internal standard (IS) responses of stable isotope labeled neurotransmitters. The IS responses were relatively consistent throughout analysis batch runs for calibrators (in 0.1% formic acid), QCs (in surrogate matrix), and study samples, indicating that matrix effects were insignificant and that reliable concentrations could be obtained with the method.

UPLC-MS/MS analysis results for all study samples collected so far in the on-going studies are shown in Figure 7 (mouse striatum tissues) and Figure 8 (rat striatum tissues). Overall, the endogenous concentration levels in study samples are within the calibration ranges of the assay for 5 neurotransmitters (DA, 5-HT, 5-HIAA, DOPAC, and HVA). The concentrations of NE were detectable, but below the lower limit of quantitation (LLOQ) in most of the study samples. As dopamine metabolism is studied in a separate mouse model study, Figure 9 shows the analysis data of mouse brain homogenates from a mouse model study. In this separate study, the homogenates of mouse whole brain tissues were analyzed for DA and its metabolites (DOPAC and HVA). Unfortunately, large variations of neurotransmitter concentrations were observed among study samples even though the described assay demonstrated good precision and accuracy. The temporal and heterogenic nature of neurotransmitters in the animal brain locations and their chemical instability may play significant roles in assay variations. Therefore, it is very important to improve the precision and accuracy both in sample collection and sample processing steps when dealing with a large number of animal study samples.

Conclusions and future workA simple, sensitive, specific, and robust UPLC-MS/MS method was validated with good linearity, precision and accuracy for the determination of endogenous dopamine, serotonin and their acidic and basic metabolites in animal striatum tissues. The method has been used successfully to assay samples from animal models of Parkinson’s disease and from a separate animal model study. Recently, we have extended the work to include other classes of neurotransmitters such as amino acids, acetylcholine, g-aminobutyric acid (GABA), and monoamines (Figure 10) in animal brain tissues and in microdialysis fluids using UPLC-MS/MS with new generations of mixed-mode chromatographic stationary phases.

References l1. C. Quang et al., 59th Conference on Mass Spectrometry and Allied Topics (2011), WP 670.2. Ewa Balkowice-Iskra et al., Acta Neurobiol. Exp. 2007, 67:379-3883. E. H. Stotz et al., Brain Research, 606 (1993) 267-2724. Hitoshi Nohta et al., Analytical Sciences, August 1993, Vol. 9, 537-540.5. Mark M. Kushnir et al., Clinical Chemistry 48(2), 2002 323-3316. Paivi Uutela et al., Anal. Chem. 81 (2009) 8417-8425

Acknowledgments

The authors acknowledge the excellent technical support provided by Alison Johnson, Robyn Eggleston, Ryan Scavinski, and other staff of WIL Research Laboratories, LLC.

Novel Aspect lValidated UPLC-MS/MS methods for determining dopamine, serotonin, and their metabolites in the striatum tissues from animal models of Parkinson’s disease.

Abstract lSimultaneously determining dopamine, serotonin and their metabolites in animal striatum tissues has been a challenge as their concentrations are low, sample size is limited, and many polar endogenous compounds may interfere with their analyses. Because of its unique features of sensitivity, selectivity, and speed (3S), ultra performance liquid chromatography with tandem mass spectrometry detection (UPLC-MS/MS) is increasingly used for the quantitation of biological compounds including small polar neurotransmitters. In this presentation, we describe UPLC-MS/MS method validation and application to the analysis of dopamine, serotonin and their metabolites in the striatum tissues from animal models of Parkinson’s disease. The striatum tissue samples were surgically removed from euthanized animals and immediately flash frozen with liquid nitrogen and stored frozen at approximately –70°C until processing. The tissue samples were homogenized in 0.1% formic acid aqueous solution in a ratio of 1:9 [weight (mg) : mL]. In the positive ion multiple reaction monitoring (MRM) mode for norepinephrine (NE), dopamine (DA), serotonin (5-HT), and 5-hydroxyindole-3-acetic acid (5-HIAA), the sample homogenates were filtered through a Supelco HybrideSPE plate and separated on a Restek Ultra II PFP Propyl column (1.9 mm, 50 x 2.1 mm) using mobile phases of 0.1% formic acid in water and in acetonitrile with a 4-minute gradient elution. In the negative ion MRM mode for 3,4-dihydoxyphenylacetic acid (DOPAC) and homovanillic acid (HVA), the homogenates were filtered through a Whatman filter plate and separated on a Waters HSS T3 column (1.7 mm 50 x 2.1 mm) using mobile phases of 0.1% acetic acid in water and in acetonitrile in a 4-minute run time. Stable isotope labeled internal standards were used for the 6 endogenous analytes to minimize the matrix effects on their quantitation. The calibration standards were prepared in 0.1% formic acid aqueous solution with ranges from 6.00 to 1200 ng/mL for DA and 5-HT, and from 12.0 to 2400 ng/mL for NE, 5-HIAA, DOPAC, and HVA. The method has been validated using quality control samples prepared in a surrogate matrix, charcoal treated (2X) animal brain homogenates. Finally the validated method has been routinely used to analyze study samples from various animal models.

Introduction lResearch in neuroscience and neurodegenerative diseases has been translational in nature, i.e., laboratory animals have been used for investigating biological processes, studying the causes of diseases, and testing new therapeutic means of treating and preventing them. The capability of producing transgenic, knockout, mutation cell lines of animal species has given researches many models linked to human diseases. As small-molecule biomarkers are commonly monitored in neurochemistry studies, reliable bioanalytical methods with high sample throughput are needed for the quantitation of the endogenous biomarkers in biological matrices. We describe our continuing efforts of developing and validating a UPLC-MS/MS method for determining dopamine, serotonin, and their metabolites in striatum tissues from animal (i.e., rat and mouse) models of Parkinson’s disease1.

Neurotransmitters have been analyzed by using liquid chromatography coupled with electrochemical detection (LC-ECD)2,3, fluorescence detection (LC-FL)4 or lately electrospray tandem ionization mass spectrometric (ESI-MS/MS) detection5-6. The lack of method specificity and sensitivity of LC-ED and FC-FL methods resulted in differences in concentrations of various neurotransmitters reported in literature. The analysis time per sample injection is typically long (20–30 minutes) using the nonspecific LC-ED and LC-FL methods. We describe here a simple, specific, and sensitive UPLC-MS/MS method to quantify dopamine, serotonin and their acidic and basic metabolites in small striatum tissues.

XIC of +MRM (17 pairs): 146.1/87.0 amu from Sample 72 (7 025 WIL-784002 30527 1 Right BRSH-2 1) of I5-784002_add analyte.wiff (T... Max. 7.4e5 cps.

1 2 3 4 5 6 7 8 9Time, min

0.0

5.0e5

1.0e6

1.5e6

2.0e6

2.5e6

3.0e6

3.5e6

4.0e6

4.5e6

5.0e6

5.5e6

6.0e66.3e6

Inte

nsity

, cps

4.824.21

GABA

5-HT

DAACh

Glu

Asp

5-HIAA

PCh

iso-ACh

XIC of +MRM (9 pairs): 152.0/107.0 amu from Sample 36 (HBT BLK IS) of Supelco LRM.wiff (Turbo Spray) Max. 7920.0 cps.

0.5 1.0 1.5 2.0 2.5 3.0 3.5Time, min

0.0

2.0e5

4.0e5

6.0e5

8.0e5

1.0e6

1.2e6

1.4e6

1.6e6

1.8e6

2.0e6

Inte

nsity

, cps

XIC of +MRM (9 pairs): 152.0/107.0 amu from Sample 32 (HBT BLK IS) of Supelco LRM.wiff (Turbo Spray) Max. 4440.0 cps.

0.5 1.0 1.5 2.0 2.5 3.0 3.5Time, min

0.0

1.0e5

2.0e5

3.0e5

4.0e5

5.0e5

6.0e5

7.0e5

8.0e5

Inte

nsity

, cps

1.52 2.86 3.282.111.32 2.631.63 1.91 2.31 3.021.22 3.09 3.470.720.55 2.541.780.820.16 3.710.410.38 0.89

phosphatidylcholine m/z 184

XIC of +MRM (8 pairs): 152.0/107.0 amu from Sample 37 (10 034 WIL-784001 7877 8 Left BRH-1 1) of I5-0010.wiff (Turbo Spray) Max. 2.2e4 cps.

0.5 1.0 1.5 2.0 2.5 3.0 3.5Time, min

0.0

2.0e4

4.0e4

6.0e4

8.0e4

1.0e5

1.2e5

1.4e5

1.6e5

1.8e5

2.0e52.1e5

Inte

nsity

, cps

0.83

XIC of -MRM (4 pairs): 167.0/123.0 amu from Sample 39 (11 034 WIL-784001 7877 8 Left BRH-1 1) of I5-0011.wiff (Turbo Spray) Max. 1.9e5 cps.

0.5 1.0 1.5 2.0 2.5 3.0 3.5Time, min

0.0

2.0e4

4.0e4

6.0e4

8.0e4

1.0e5

1.2e5

1.4e5

1.6e5

1.8e5

2.0e5

Inte

nsity

, cps

1.63

0.28

3.47

NE

DA

5-HIAA

5-HT

DOPAC

HVA

I5-784001#7p.rdb (NE): "Quadratic" Regression ("1 / (x * x)" weighting): y = -3.42e-008 x^2 + 0.00552 x + 0.101 (r = 0.9845)

0 500 1000 1500 2000Analyte Conc. / IS Conc.

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0

11.0

12.0

13.0

Anal

yte

Area

/ IS

Are

a

I5-784001#7p.rdb (Dopamine): "Quadratic" Regression ("1 / (x * x)" weighting): y = -1.76e-007 x^2 + 0.00929 x + -0.00368 (r = 0.9997)

0 200 400 600 800 1000 1200Analyte Conc. / IS Conc.

0.0

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0

11.0

Anal

yte

Area

/ IS

Are

a

I5-784001#7p.rdb (Serotinin): "Quadratic" Regression ("1 / (x * x)" weighting): y = 2.68e-007 x^2 + 0.0117 x + -0.00588 (r = 0.9993)

0 200 400 600 800 1000 1200Analyte Conc. / IS Conc.

0

2

4

6

8

10

12

1415

Anal

yte

Area

/ IS

Are

a

I5-784001#7p.rdb (5-HIAA): "Quadratic" Regression ("1 / (x * x)" weighting): y = 1.72e-007 x^2 + 0.0301 x + -0.0128 (r = 0.9996)

0 500 1000 1500 2000Analyte Conc. / IS Conc.

0

10

20

30

40

50

60

7075

Anal

yte

Area

/ IS

Are

a

I5-784001#8n.rdb (DOPAC): "Quadratic" Regression ("1 / (x * x)" weighting): y = 1.93e-008 x^2 + 0.00166 x + -0.00303 (r = 0.9967)

0 500 1000 1500 2000Analyte Conc. / IS Conc.

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

Anal

yte

Area

/ IS

Are

a

I5-784001#8n.rdb (HVA): "Quadratic" Regression ("1 / (x * x)" weighting): y = -1.08e-008 x^2 + 0.00254 x + -0.000964 (r = 0.9960)

0 500 1000 1500 2000Analyte Conc. / IS Conc.

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Anal

yte

Area

/ IS

Are

a

NE (12-2400 ng/mL) DA (6-1200 ng/mL)

5-HT (6-1200 ng/mL) 5-HIAA (12-2400 ng/mL)

DOPAC (12-2400 ng/mL) HVA (12-2400 ng/mL)

IS Peak Area vs Index (NE)

0 20 40 60 80 100 120 140 160 180 200 220 240Index

0.0

1.0e4

2.0e4

3.0e4

4.0e4

5.0e4

6.0e4

7.0e4

8.0e4

IS P

eak

Area

, cou

nts

IS Peak Area vs Index (Dopamine)

20 40 60 80 100 120 140 160 180 200 220 240Index

0.0

2.0e4

4.0e4

6.0e4

8.0e4

1.0e5

1.2e5

1.4e5

1.6e5

IS P

eak

Area

, cou

nts

IS Peak Area vs Index (Serotinin)

20 40 60 80 100 120 140 160 180 200 220 240Index

0.0

5000.0

1.0e4

1.5e4

2.0e4

2.5e4

3.0e4

3.5e4

IS P

eak

Area

, cou

nts

IS Peak Area vs Index (5-HIAA)

20 40 60 80 100 120 140 160 180 200 220 240Index

0.0

2000.0

4000.0

6000.0

8000.0

1.0e4

1.2e4

1.4e4

1.6e4

1.8e4

2.0e4

2.2e4

2.4e42.6e4

IS P

eak

Area

, cou

nts

IS Peak Area vs Index (DOPAC)

10 20 30 40 50 60 70 80 90 100 110 120Index

0.0

5.0e4

1.0e5

1.5e5

2.0e5

2.5e5

3.0e5

3.5e5

4.0e5

4.5e5

IS P

eak

Area

, cou

nts

IS Peak Area vs Index (HVA)

10 20 30 40 50 60 70 80 90 100 110 120Index

0.0

5000.0

1.0e4

1.5e4

2.0e4

2.5e4

3.0e4

3.5e4

4.0e4

4.5e4

IS P

eak

Area

, cou

nts

NE-d6 DA-d4

5-HT-d4 5-HIAA-d5

DOPAC-d5 HVA-d4

Mou

se B

rain

Hom

ogen

ate

Con

c. (n

g/m

L)

Sample number

Figure 9. Concentrations (ng/mg tissue) of neurotransmitters in mouse brain tissues collected from study animals.

Figure 8. Concentrations (ng/mg tissue) of neurotransmitters in mouse brain tissues collected from study animals.

Figure 7. Concentrations (ng/mg tissue) of neurotransmitters in mouse striatum tissues collected from animals of Parkinson’s disease models at various ages.

Figure 6. Batch internal standard responses of stable isotope labeled compounds: NE-d6, DA-d4, 5-HT-d4, 5-HIAA-d5, DOPAC-d5, and HVA-d4. (●) calibrators, (▲) quality control samples, and (■) study samples.

Figure 5. Calibration standard curves of NE (12-2400 ng/mL), DA (5-1200 ng/mL), 5-HIAA (12-2400 ng/mL), 5-HT (6-1200 ng/mL), DOPAC (12-2400 ng/mL), and HVA (24-2400 ng/mL). The curves were prepared in aqueous 0.1% formic acid solution.

Figure 4. UPLC-MS/MS chromatograms of a mouse striatum homogenate (Top) in positive ion MRM mode and (Bottom) in negative ion MRM mode. Stable isotope-labeled internal standards were added to the sample.

Figure 3. Representative UPLC-MS/MS chromatograms of mouse striatum homogenates. (Top) the homogenate sample without phospholipids removal, (Bottom) the sample extract filtered with a HybridSPE phospholipids plate.

Figure 10. UPLC-MS/MS chromatogram of mouse striatum homogenate. Chromatographic column: Scherzo SS-C18 (150x3 mm, 3µm). Mobile phases: A: 0.5% formic acid, B: 0.5% formic acid and 50 mM ammonium formate in acetonitrile/water (50/50, v/v).

Figure 2. Chemical structures and metabolic pathways of (A) dopamine, (B) serotonin, and their acidic and basic metabolites.

Figure 1. Dopamine signaling in Parkinson’s disease (Source: Cell Signaling Technology, www.cellsignal.com).

Table 4. Calibration Standard Statistics (Rat Striatum Assay)DA (ng/mL) 6.00 12.0 60.0 120 300 600 1080 1200

Mean 5.90 12.5 58.7 120 301 598 1090 1190%RSD 1.60 3.30 2.50 2.30 1.70 3.40 1.20 1.40%RE -1.67 4.17 -2.17 0.00 0.33 -0.33 0.93 -0.83

n 10 10 10 10 10 10 10 10DOPAC (mg/mL) 12.0 24.0 120 240 600 1200 2160 2400

Mean 11.8 24.9 117 239 599 1220 2130 2420%RSD 3.1 5.7 5.8 4.2 2.9 5.2 2.8 2.4%RE -1.67 3.75 -2.50 -0.42 -0.17 1.67 -1.39 0.83

n 9 9 9 9 9 9 9 9HVA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00

Mean 11.9 24.7 118 238 602 1220 2150 2400%RSD 1.9 4.4 5.1 3.9 6.6 2.0 2.5 2.9%RE -0.83 2.92 -1.67 -0.83 0.33 1.67 -0.46 0.00

n 11 11 11 11 11 11 11 115-HT (ng/mL) 6.00 12.00 60.00 120.00 300.00 600.00 1080.00 1200.00

Mean 6.03 12.0 59.0 120 303 608 1090 1190%RSD 2.8 6.1 3.5 2.8 3.6 2.2 1.5 1.5%RE 0.50 0.00 -1.67 0.00 1.00 1.33 0.93 -0.83

n 11 11 11 10 11 11 11 115-HIAA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00

Mean 12.0 24.3 118 241 601 1210 2140 2410%RSD 1.5 3.4 2.6 2.4 2.4 3.3 2.9 2.7%RE 0.00 1.25 -1.67 0.42 0.17 0.83 -0.93 0.42

n 11 11 11 11 11 11 11 11NE (ng/mL) 12.0 24.0 120 240 600 1200 2160 2400

Mean 11.9 24.7 119 238 602 1210 2170 2380%RSD 1.9 5.1 4.0 3.9 3.2 3.4 1.7 1.9%RE -0.83 2.92 -0.83 -0.83 0.33 0.83 0.46 -0.83

n 10 10 10 10 10 10 10 10

Table 5. Precision and Accuracy of Quality Controle Samples (Rat Striatum Assay)

6.00 18.0 600 960 6.00 18.0 600 960Mean 5.67 17.4 583 937 6.33 17.6 588 942

%RSD 3.4 5.0 3.3 3.1 5.5 4.9 4. .8%RE -5.5 -3.3 -2.8 -2.4 5.5 -2.2 -2.0 -1.9

n 4 10 10 10 5 11 1 1

12.0 36.0 1200 1920 12.0 36.0 1200 1920Mean 12.9 36.3 1180 1870 11.7 35.4 1200 1920

%RSD 9.0 4.5 4.2 3.5 4.5 4.8 3. .2%RE 7.5 0.8 -1.7 -2.6 -2.5 -1.7 0.0 0.0

n 5 11 11 11 4 10 1 0

12.0 36.0 1200 1920 12.0 36.0 1200 1920Mean 12.3 38.9 1240 2000 11.6 36.1 1170 1860%RSE 13.6 11 5.3 4.1 10.9 6.5 5. .2%RE 2.5 8.1 3.3 4.2 -3.3 0.3 -2.5 -3.1

n 5 9 9 9 6 11 11 11

DA (ng/mL -HT (ng/mL)

5-HIAA (ng/mL E (ng/mL)

DOPAC (ng/mL VA (ng/mL)

Tables 4–5. Validation results of rat homogenate assay

Table 2. Calibration Standard Statistics (Mouse Striatum Assay)DA (ng/mL) 6.00 12.0 60.0 120 300 600 1080 1200

Mean 6.02 11.9 59.8 119 305 604 1060 1210%RSD 1.00 4.20 4.00 4.70 2.70 2.40 2.60 2.90%RE 0.33 -0.83 -0.33 -0.83 1.67 0.67 -1.8 .83

n 10 10 10 10 10 10 1 0DOPAC (mg/mL) 12.0 24.0 120 240 600 1200 2160 2400

Mean 12.0 24.0 119 237 611 1210 2150 2400%RSD 1.8 5.3 4.0 4.7 3.8 3.9 2. .4%RE 0.00 0.00 -0.83 -1.25 1.83 0.83 -0.46 0.00

n 9 9 9 9 9 9 9 9HVA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00

Mean 12.0 24.1 120 239 603 1210 2120 2440%RSD 2.0 6.0 1.9 4.2 3.4 3.0 2. .0%RE 0.00 0.42 0.00 -0.42 0.50 0.83 -1.8 .67

n 9 9 9 9 9 9 9 95-HT (ng/mL) 6.00 12.00 60.00 120.00 300.00 600.00 1080.00 1200.00

Mean 5.99 12.2 58.7 121 304 603 1070 1200%RSD 2.2 4.9 2.7 3.3 3.8 1.9 4. .8%RE -0.17 1.67 -2.17 0.83 1.33 0.50 -0.93 0.00

n 11 11 11 10 11 11 1 15-HIAA (ng/mL) 12.00 24.00 120.00 240.00 600.00 1200.00 2160.00 2400.00

Mean 12.1 23.8 119 238 610 1210 2150 2400%RSD 1.8 4.0 3.2 5.8 3.6 3.4 2. .2%RE 0.83 -0.83 -0.83 -0.83 1.67 0.83 -0.4 .00

n 9 9 9 9 9 9 9 9NE (ng/mL) 12.0 24.0 120 240 600 1200 2160 2400

Mean 12.1 23.9 121 236 605 1210 2160 2400%RSD 1.4 3.7 3.4 3.8 1.8 2.1 2. .0%RE 0.83 -0.42 0.83 -1.67 0.83 0.83 0.0 .00

n 9 9 9 9 9 9 9 9

Table 3. Precision and Accuracy of Quality Controle Samples (Mouse Striatum Assay)

6.00 18.0 600 960 6.00 18.0 600 960Mean 5.59 18.1 574 911 6.27 19.2 572 907

%RSD 7.5 11.2 6.6 6.5 8.5 10.1 6.1 6.2%RE -6.8 0.6 -4.3 -5.1 4.5 6.7 -4.7 -5.5

n 6 10 10 10 6 11 1 1

12.0 36.0 1200 1920 12.0 36.0 1200 1920Mean 13.3 37.9 1190 1830 10.7 33.8 1170 1880

%RSD 6.4 8.5 4.5 5.4 9.4 24.7 7.3 6.1%RE 10.8 5.3 -0.8 -4.7 -10.8 -6.1 -2. 2.1

n 4 9 9 9 4 9 9 9

12.0 36.0 1200 1920 12.0 36.0 1200 1920Mean 12.6 38.3 1250 1960 11 36.2 1200 1880%RSE 7.8 11 5.2 5.3 12.9 8.5 6. .8%RE 5.0 6.4 4.2 2.1 -8.3 0.6 0. 2.1

n 5 9 9 9 5 9 9 9

DA (ng/mL)

5-HIAA (ng/mL E (ng/mL)

DOPAC (ng/mL VA (ng/mL)

5-HT (ng/mL)

Tables 2–3. Validation results of mouse homogenate assay. Table 1. Animal models of Parkinson’s disease Evaluated at three ages (4, 8, and

12 months).

Experimental Section lCalibration standards and quality control samples: Stock solutions of neurotransmitters in Figure 2 were prepared in ice-cold 0.1% formic acid at a concentration of 1.00 mg/mL, and the stocks were subaliquoted into tubes (0.50 mL each) and stored at -80°C. Calibration standards were prepared from the stock solutions with ice-cold 0.1% formic acid at various concentrations on the day of the analysis. Quality control samples were prepared in charcoal treated (2X) rat and mouse brain homogenates spiked with neurotransmitters at four concentration levels (LLQC, LQC, MQC, HQC). Quality control samples were subaliquoted into tubes and stored at -80°C.

Preparation of study samples: Striatum tissues were surgically removed from euthanized animals. After weighing, tissue samples were immediately flash frozen with liquid nitrogen and stored frozen at approximately -80°C until sample processing. During analysis, tissue samples were homogenized on wet ice with ice-cold 0.1% formic acid using a Kontes microgrinder at a ratio of 1:9 [mg (weight): mL]. For the analysis of rat striatum samples, aliquots of 50.0 mL of homogenates were added to a 96-well sample plate (2-mL well size), followed by 50 mL of 0.1% formic acid solution containing internal standards. The samples were further diluted with 350 mL of ice cold 0.1% formic acid (total 450 mL). Aliquots (200 mL) of the diluted samples were transferred and filtered through a Supelco HybridSPE® plate for the positive ion analysis (NE, DA, 5-HT, and 5-HIAA). Second portions (200 mL) of the diluted samples were filtered through a Whatman filter plate for the negative ion analysis (DOPAC and HVA). Because of limited sample size for mouse striatum tissues (6–28 mg), Whatman filter plates were used to prepare the diluted homogenates for both positive ion and negative ion modes of analysis.

Dopamine (a catecholamine), serotonin and their metabolites are known to be unstable in the presence of light, oxygen, metal ions, and low or high pH, as well as sensitive to temperature. Samples should be processed on ice quickly in 0.1% formic acid. Proper collection, processing, and storage of tissue samples are important for generating accurate results.

Triple-stage quadrupole mass spectrometer (API-4000) coupled with Waters Acquity UPLC® systems were used for UPLC-MS/MS analysis.

UPLC conditions (the positive ion mode):LC System: Waters Acquity UPLC® SystemColumn: Restek Ultra™ PFP-Propyl 50 x 2.1 mm, 1.9 µmColumn Temperature: 60°CAutosampler Temperature: 4°CMobile Phase A: 0.1% Formic acid in waterMobile Phase B: 0.1% Formic acid on acetonitrileStrong Wash: 0.1% Formic acid in acetonitrileWeak Wash: 0.1% Formic acid in waterSeal Wash: 10% Acetonitrile

UPLC conditions (the negative ion mode):LC System: Waters Acquity UPLC® SystemColumn: Waters HSS T3, 50 x 2.1 mm, 1.7 µmColumn Temperature: 60°CAutosampler Temperature: 4°CMobile Phase A: 0.1% Acetic acid in waterMobile Phase B: 0.1% Acetic acid on acetonitrileStrong Wash: 0.1% Acetic acid in acetonitrileWeak Wash: 0.1% Acetic acid in waterSeal Wash: 10% Acetonitrile

UPLC gradient programTime (min)

A% B % Flow rate (mL/min)

Initial 100 0 0.400.50 100 0 0.402.60 70 30 0.402.70 10 90 1.003.30 10 90 1.003.31 100 0 1.004.00 100 0 1.00

Mass spectrometric detection parameters Compound

NamePrecursor Ion

m/zProduct Ion

m/zIonization

ModeNE 152 107 PositiveDA 154 91 Positive

5-HT 177 115 Positive5-HIAA 192 146 PositiveNE-d6 158 111 PositiveDA-d4 158 95 Positive

5-HT-d4 181 118 Positive5-HIAA-d5 197 150 Positive

UPLC gradient programTime (min)

A% B % Flow rate (mL/min)

Initial 100 0 0.402.50 75 25 0.402.60 5 95 1.003.20 5 95 1.003.21 5 95 1.003.30 100 0 1.004.00 100 0 1.00

Mass spectrometric detection parameters Compound

NamePrecursor Ion

m/zProduct Ion

m/zIonization

ModeDOPAC 167 123 Negative

HVA 181 122 NegativeDOPAC-d5 172 128 Negative

HVA-d5 186 127 Negative