Main changes IEC 61010-1:2010 3 rd edition to IEC 61010-1:2010/A1:2016 14 March 2017

Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

Embed Size (px)

Citation preview

Page 1: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

Main changes

IEC 61010-1:2010 3rd edition to IEC 61010-1:2010/A1:201614 March 2017

Page 2: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20162

• IEC 61010-1:2010/A1:2016- DOP: 2016-12-16 (Included in IECEE System 2017-02-16)- IEC test template is not published (Preparing VDE)

• EN 61010-1:2010/A1:2016 (under Low voltage directive: 2014/35/EU)- Estimated dates according to CENELEC Website.- DOP: 2017-11-06 - DOW: expected 2 years later

Date of publication

Page 3: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20163

I. Additions

II. Deletions

III. Replacements

Contents

Page 4: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20164

1. Addition of conformity criterion for tests (cl. 4.1); Tests needed to support a RISK assessment (see Clause 17) are carried out in the combinations of conditions and operations determined during the RISK assessment

If there is any test performed to support risk assessment then the test should be performed in the combinations of conditions and operations determined during the RISK assessment.

I. Additions

Page 5: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20165

2. Clarification for a creepage distance (cl. 6.7.1.3);A CREEPAGE DISTANCE may be split in several portions of different materials and/or have different POLLUTION DEGREES if one of the CREEPAGE DISTANCES is dimensioned to withstand the total voltage or if the total distance is dimensioned according to the material having the lowest CTI and the highest POLLUTION DEGREE.

I. Additions

Page 6: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20166

3. Clarification for TRANSIENT OVERVOLTAGE level for MAINS (cl. 6.7.1.5);NOTE 4 The assumed TRANSIENT OVERVOLTAGE level for MAINS is equal to the required RATED impulse voltage of equipment in Table 443.2 of IEC 60364-4-44:2007/AMD1:2015.

I. Additions

Page 7: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20167

I. Additions

Page 8: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20168

I. Additions

4. Allowance of interpolation of the creepages in table 4 which results in the same values when using Annex K.3 (cl. 6.7.2.1); Linear interpolation of the CREEPAGE DISTANCES is allowed.

Page 9: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:20169

I. Additions

5. Clarification for voltage test procedure (cl. 6.8.1);When verifying a CLEARANCE within equipment , it is necessary to ensure that the specified voltage appears at the CLEARANCE. PROTECTIVE IMPEDANCE, impedances and voltage-limiting devices in parallel with the insulation to be tested may be disconnected.

VDRs, GDTs can be removed during test.

Page 10: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201610

I. Additions

6. Clarification of force measurement in clause 7.3.4 with examples; For the purposes of this subclause, the width of a finger is considered to be 1.2 cm and the width of any other body part is considered to be 5 cm. The contact area will be determined to be the width of the body part multiplied by the width of the moving part, or the cross-sectional contact area of the moving part, if smaller.

Page 11: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201611

I. Additions

7. Clarification for the heating test to allow test at max. rated ambient of the manufacturer (according to 3rd ed, is was only allowed to perform the tests between 15 and 35°C and not at the max. rated ambient when higher) (cl.10.4.1);

Measures are taken to eliminate errors caused by the method of achieving the test ambient temperature (e.g. suitable baffling or enclosure if the test is conducted in an environmental chamber and the forced air movements would cool the exterior of the equipment).

Page 12: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201612

I. Additions

8. Adding of two more mains supply systems in the informative Annex I

Page 13: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201613

I. Additions

9. Adding an example in Annex K.3.2 to show how to calculate the additional transient overvoltage. Important is that this calculation is only valid within primary or between primary and secondary circuits. It can be different when you are in secondary circuits, as in such cases you cannot subtract the mains voltage.

Page 14: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201614

II. Deletions

1. Delete (cl. 4.4.2.2 b));

2. Delete, in item c), the word "control“ (cl. 5.1.5.2);

3. Delete the second sentence of the third paragraph (cl. 7.3.4);

Page 15: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201615

II. Deletions

4. Delete NOTE 2, and renumber NOTE 3 as NOTE 2 (cl.12.2.1.1);

5. Delete, in Figure D .1 h), delete "(see 6.6.1)“ (Annex D);

Page 16: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201616

III. Replacements

1. (Clarification) Replace the conformity statement in cl. 5.1.3 Mains supply; This change is clarification regarding how to measure the power or current consumption: Now measure the max. mean value of a one minute period. To do this correctly you have to measure continuously

2. (Clarification) Replace the first paragraph with the following (cl.5.2);“Warning markings specified in this standard shall meet the following requirements”

Page 17: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201617

III. Replacements

3. (Clarification) Replace the paragraph prior to the conformity statement with the following (cl.5.2);If the instructions for use state that an OPERATOR is permitted to gain access, using a TOOL, to a part or location which in NORMAL USE may present a HAZARD, there shall be a warning marking indicating that the equipment must be placed in a safe state before access. Symbol 14 shall be used for this purpose with the warning text included in the documentation. Additional symbols may be used to indicate the nature of the HAZARD such as symbol 12, 13, or 17 as appropriate. Symbols are the preferred marking method over text warnings. Supplemental text may be provided adjacent to the symbol.

It used to be affixing the warning marking which states disconnect the power of the equipment.

Page 18: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201618

III. Replacements

4. (Clarification) Replace items d) and e) with the following (cl.5.4.2);d) a statement of the range of environmental conditions for which the equipment is designed including:1) indoor or outdoor use,2) altitude,3) temperature, 4) relative humidity, 5) MAINS supply voltage fluctuations, 6) OVERVOL TAGE CATEGORY, except for cord/plug -connected equipment,7) WET LOCATION, if applicable, 8) POLLUTION DEGREE of the intended environment,

e) for equipment RATED for ingress protection according to IEC 60529 the information required in 11.6.1;

Page 19: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201619

III. Replacements

5. (Clarification) Replace the first paragraph with the following (cl.5.4.3);The documentation shall include installation and specific commissioning instructions and, if necessary for safety, warnings against HAZARDS which could arise during installation or commissioning or as a result of improper installation or commissioning of the equipment. Such information includes, if applicable:

6. (Clarification) Replace item f) with the following(cl.5.4.3 );f) requirements and safety characteristics for special external services, for example: maximum and minimum temperature, pressure, or flow of air or cooling liquid;

7. (Clarification) Replace the first paragraph with the following (cl.5.4.5 );Instructions shall be provided to the RESPONSIBLE BODY in sufficient detail to permit safe maintenance, inspection and testing of the equipment, and to ensure continued safety of the equipment after the maintenance, inspection and test procedure.

Page 20: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201620

III. Replacements

8. (Stringent requirements) Replace the text of item a) with the following (cl.6.3.1);a) The a.c . voltage levels are 30 V r.m.s., 42,4 V peak, and the d.c. voltage level is 60 V. For equipment intended for use in WET LOCATIONS, the a.c . voltage levels are 16 V r.m.s ., 22,6 V peak, and the d.c. voltage level is 35 V.

(Clarification) Replace the first two sentences of item a) with the following (cl.6.3.2);a) The a.c. voltage levels are 50 V r.m.s., 70 V peak, and the d.c. voltage level is 120 V. For equipment intended for use in WET LOCATIONS, the a.c. voltage levels are 33 V r.m.s., 46,7 V peak, and the d.c. voltage level is 70 V.

Reduction of the values for the voltage limits in clause 6.3.1 and 6.3.2 to the values of US/CAN standards.

Page 21: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201621

III. Replacements

9. (Clarification) Replace the first paragraph and its conformity statement with the following (cl.6.6.4);TERMINALS for stranded conductors that are intended to be connected during installation, maintenance, or operation of the equipment shall be located or shielded so that there is no possibility of accidental contact between HAZARDOUS LIVE parts of different polarity or between such parts and other ACCESSIBLE parts, even if a strand of a conductor escapes from a TERMINAL. This requirement does not apply to connections that are only to be made at the manufacturing facility.

Conformity is checked by inspection after fully inserting a stranded conductor:a) with the maximum length of insulation removed as specified by the equipment manufacturer, orb) with an 8 mm length of insulation removed if no specification is given by the equipment manufacturer.

With one of the strands free, the free strand shall not touch parts of different polarity or other ACCESSIBLE parts, when bent in every possible direction, without tearing back the insulation or making sharp bends.

Page 22: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201622

III. Replacements

10.(Clarification) Replace the first conformity statement with the following (cl.6.7.2.2.1);Conformity is checked by inspection, and by the a.c. test of 6.8.3.1 or the d.c. test of 6.8.3.2 using the applicable voltage from Table 5 for 1 min.

d.c. hipot test is allowed as alternative way for a.c. hipot test.

11. (Correction) Tables in clause 6.7 (Table 6) and annex K where corrected with the values from Corrigendum 1

Page 23: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201623

III. Replacements

12.(Correction) Replace the first paragraph with the following (cl.6.11.4.2);An equipment circuit -breaker employed as a disconnecting device shall meet the relevant requirements of IEC 60947-2, and be suitable for the application.An equipment switch employed as a disconnecting device shall meet the relevant requirements of IEC 60947-3, and be suitable for the application.

13. (Clarification) Replace the second sentence of the second paragraph with the following sentence (cl. 6.11.4.2);If there is only one device - one switch or one circuit-breaker - symbols 9 and 10 of Table 1 are sufficient if the symbols are marked on or adjacent to the switch or circuit-breaker.

14. (Correction) Replace, in footnote "d Minimum protective measures:" the text of B with the following (cl. Table 12 in cl. 7.3.3);B = Moderate measures; emergency switches, PROTECTIVE BARRIERS or covers removable only with a TOOL, distances (see ISO 13857), or separations (see ISO 13854 or EN 349).

Page 24: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201624

III. Replacements

15.(Correction) Replace the third paragraph with the following (cl. 7.4);Each castor and support foot shall be RATED to support a load not less than its normal load, or the castors and support feet shall be tested according to d) or e) below .

16. (Clarification) Replace item c) as follows (cl. 9.4);It is separated by at least BASIC INSULATION from other circuits having energy values exceeding criteria a) or b) above.

a) is hazardous voltage level, b) is limited energy circuit test requirements.

Page 25: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201625

III. Replacements

17.(Addition) Replace the first paragraph with the following (cl. 11.1);Equipment shall be designed to give protection to OPERATORS and the surrounding area against HAZARDS from fluids and solid foreign objects encountered in NORMAL USE

18. (Change and additions) Replace the title and the text with the following (cl. 11.6);11.6 Equipment RATED with a degree of ingress protection (IP code)

The following sub-clauses are added;Sub-cl.11.6.1 General Including first number of IP code (Solid object), IP rating can be differ depending on the mounting arrangements, or the assembly or operating conditions.Sub-cl.11.6.2 Conditions for testing Preparing test sample, test position depending on movability, Terminals consideration, operating and non-operating condition during the IP test./Sub-cl.11.6.3 Protection against solid foreign objects (including dust) Test and conformity are performed in accordance with IEC 60529. Additionally, EUT is inspected for ingress of foreign objects (deposit on insulation parts, create accumulations can be cause the spread of fire)Sub-cl.11.6.4 Protection against water Test is performed in accordance with IEC 60529

Page 26: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201626

III. Replacements

19.(Change) Replace the text of cl. 11.7.2 Leakage and rupture at high pressure, except the first paragraph, with the following (cl. 11.7.2); National authorities may allow safety to be established by calculation, for example according to the Pressure Equipment Directive (2014/68/EU).

New test procedure and requirements are addressed ex.) new figure 16 (Ratio between test pressure and maximum working pressure) and reference standards, EN 378-2 are introduced. Also, G.2.2 conduct of hydrostatic test for G.2.1 of Annex G Leakage and rupture form fluids under pressure is changed.

Page 27: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201627

III. Replacements

19.(Change) Replace the text of cl. 11.7.2 Leakage and rupture at high pressure, except the first paragraph, with the following (cl. 11.7.2); (Continued)

Page 28: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201628

III. Replacements

20.(Change and additions) Replace the title and text with the following (cl. 12.3); 12.3 Optical radiationEquipment with lamps and lamp systems emitting ultraviolet, visible, or infrared radiation, including light emitting diodes, shall not permit unintentional escape of radiation that could cause a HAZARD.

The previous tile was “Ultraviolet (UV) radiation”, and the requirement was limited to UV radiation. However, this clause changed requires that the radiation sources shall be assessed in accordance with IEC 62471 except for sources considered to be safe (Table 22) or conditionally safe (Table 23).

Page 29: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201629

III. Replacements

20.(Change and additions) Replace the title and text with the following (cl. 12.3); (Continued)

Page 30: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201630

III. Replacements

20.(Change and additions) Replace the title and text with the following (cl. 12.3); (Continued)

Page 31: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201631

III. Replacements

21.(Change and additions) Replace the first two paragraphs with the following (cl. 13.1 Poisonous and injurious gases and substances);Equipment shall not liberate dangerous amounts of hazardous substances in NORMAL CONDITION and in SINGLE FAULT CONDITION .If potentially -hazardous substances are liberated, the OPERATOR shall not be directly exposed to a quantity of the substance that could cause harm.

SFC is added as new requirement. 22. (Change) Replace the title with the following (cl. 14.8);

14.8 Circuits used to limit TRANSIENT OVERVOLTAGES It was “14.8 Circuits or components used as TRANSIENT OVERVOLTAGE limiting devices”The word “Withstand” is removed for Table 21. The conformity statement is changed too.

Page 32: Nemko IEC 61010-1 AMD1 PSES 20170314.pdf - …s3.amazonaws.com/sdieee/2120-Nemko+IEC+61010-1+AMD1+PSES+20170314.pdfwhich results in the same values when using Annex K.3 (cl. 6.7.2.1);

15.03.2017 CHANGES IEC 61010-1:2010 3RD EDITION TO IEC 61010-1:2010/A1:201632

III. Replacements

23.(Correction) Replace Table H.1 (Annex H Conformal coatings requirements); Because according to the previous test specification no coating could pass the test.

24. (Minor corrections in adjustment) There are some minor corrections in adjustment in Normative reference and Bibliography.