Upload
phamnhan
View
217
Download
0
Embed Size (px)
Citation preview
Keysight TechnologiesM937XA Frequency Upgrade Guide for units 20 GHz and less.
M937XA Frequency Upgrade
Introduction:The purpose of this guide is to show the steps required to do the Frequency upgrade on the M937XA PXI VNA.
Upgrade kits covered in this guide.
1. M9370AU-706
2. M9370AU-709
3. M9370AU-714
4. M9370AU-720
1. M9371AU-709
2. M9371AU-714
3. M9371AU-720
1. M9372AU-714
2. M9372AU-720
1. M9373AU-720
The frequency upgrade kit covered in this guide is the M9370-65702. This kit can be ordered from SPO inventory. K02, K03 and K04 will keep a safety stock of 2 kits at all times.
The M9370-65702 may only be used on upgrades up to 20 GHz. If an upgrade to 26.5 GHz is required, the M9370-65701 kit must be ordered.
For all questions about this kit and its contents. Please contact, Christopher Rich in CSG Customer Support at 707 577 2070 or [email protected]
Kit Contents:This kit contains the following items:
1. 1250-3895 Connector-RF Nut-Spanner 0.32-in-OD 0.07-in-THK Brass Au-plate
2. 1250-4106 Adapter-Coaxial Sparkplug APC-3.5 50Ohm 26.5GHz
3. M9371-80001 Nameplate, M9371A 300kHz-6.5 GHz
4. M9372-80001 Nameplate, M9372A 300kHz-9 GHz
5. M9373-80001 Nameplate, M9373A 300kHz-14 GHz
6. M9374-80001 Nameplate, M9374A 300kHz-20 GHz
Equipment Needed:The equipment needed to perform the upgrade are specific to the PXI VNA.
1. Y1283-60003 SMA and 3.5 mm Connector and Wrench Kit
The kit contains the following items:
1. 1250-3893 Adapter-Coaxial straight Plug APC 3.5 50Ohm 26.5GHz
2. 1250-3895 Connector-RF Nut-Spanner 0.32-in-OD 0.07-in-THK Brass Au-plated
3. 1250-4106 Adapter-Coaxial Sparkplug APC-3.5 50Ohm 26.5GHz
4. 8710-2652 Torque Wrench 5mm-open-end breakaway-style
5. 8710-2656 Tool-Socket for Spanner Nut 0.197-in-HEX 0.330-in-DIA 0.6-in-LG heat treated BeCu Ni-plated
2. T8 Torx driver
The 8710-2652 torque wrench and 8710-2656 Socket are very specific tools for this product and should only be used when replacing connectors on the M937XA PXI VNA.
Test Equipment Needed:Recommended Equipment
1. Frequency Counter 53131A/53232A
2. Power Sensor U8485A-200 (if using a non-USB Power Sensor, a Power Meter is needed)
3. Source N5183B-532 (PSG E8257D-532 also works fine)
4. Splitter 11667B
5. Calkit ECal or equivalent mechanical kit
Cables required for test.
1. 5063-1530 RF Cable Assembly 6 in SMA
2. 8120-5048 Cable A01/A06 1230G
Pictures of Kits and cables:1. Picture of the M9370-65702 Kit
2. Picture of the M9370-65702 Kit with 8710-2652 Torque Wrench and 8710-2656 Tool-Socket for Spanner Nut.
3. Cables used for Service Adjustment
Frequency Upgrade Steps:1. Replace the Connectors on the PXI VNA
a. Using the 8710-2652 Torque Wrench and the 8710-2656 Tool-Socket loosen the locking nut on Port 1 and Port 2.
b. Remove the locking nuts from Port 1 and Port 2
c. Using the 8710-2652 Torque Wrench remove Port 1 and Port 2 connectors.
d. See the pictures below.
Side View:
Front View
Nut and Socket View
Remove the connect with the 8710-2652 Torque Wrench
Connect out and ready for new connect to be inserted
2. Once the connectors have been replaced you are ready to move on to the next steps.
a. Option Enable, the unit will need to have the Model Number changed. To do this please follow these steps.
i. Plug the PXI VNA into the test chassis
ii. Boot the chassis with the PXI controller
iii. Start the PXI VNA application
iv. Select the slot number that matches the unit you are upgrading
v. Go to the Help Menu and select About Network Analyzer
vi. Send the Model Number / Serial Number to [email protected].
vii. An email with the Keyword for the new model number will be sent to you.
viii. Go to the Utility > System > Service menu
ix. Open Option Enable
x. Select Repair from the menu
xi. Check the box (Check if Model shown is incorrect!)
xii. Select Begin Repair
xiii. In the Model Change area use the drop down menu to select the Model number you are upgrading to.
xiv. Add the Keyword to the Hex Keyword and select change Model
xv. Once this is complete the NA Application will restart
b. Now the PXI VNA is upgraded to the new Frequency and you can start the Adjustments.
Adjustments:Service Adjustments are located in Utility > System > Service > Adjustments
All adjustments need to be run, with the exception of Loop Bandwidth Adjustment. This is only required if the unit has option 103 or 104. If the unit does not have these options, you may skip this adjustment. All other adjustments must be run, in order from top to bottom.
Once all the adjustments have been run, the unit can be transferred to the TME Cal Team for TME performance and verification.
myKeysightwww.keysight.com/find/mykeysightA personalized view into the information most relevant to you.
For more information on Keysight Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at:www.keysight.com/find/contactus
Three-Year Warrantywww.keysight.com/find/ThreeYearWarrantyKeysight’s commitment to superior product quality and lower total cost of ownership. The only test and measurement company with three-year warranty standard on all instruments, worldwide.
AmericasCanadaBrazilMexicoUnited States
(877) 894 4414 55 11 3351 7010 001 800 254 2440 (800) 829 4444
Keysight Assurance Planswww.keysight.com/find/AssurancePlansUp to five years of protection and no budgetary surprises to ensure your instruments are operating to specification so you can rely on accurate measurements.
Asia PacificAustraliaChinaHong KongIndiaJapanKorea
1 800 629 485 800 810 0189
800 938 693 1 800 112 929 0120 (421) 345 080 769 0800
Malaysia 1 800 888 848 www.keysight.com/go/qualityKeysight Technologies, Inc.DEKRA Certified ISO 9001:2008Quality Management System
SingaporeTaiwanOther AP Countries
1 800 375 8100
0800 047 866 (65) 6375 8100
Keysight Channel Partnerswww.keysight.com/find/channelpartnersGet the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience.
Europe & Middle EastAustriaBelgiumFinlandFranceGermany
0800 001122 0800 58580 0800 523252 0805 980333 0800 6270999
Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100
www.keysight.com/find/XXXXXX Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353
Opt. 1 (DE)Opt. 2 (FR)Opt. 3 (IT)
United Kingdom 0800 0260637
For other unlisted countries:www.keysight.com/find/contactus
This information is subject to change without notice. ©Keysight Technologies, 2014Published in USA, November 1, 2014XXXX-XXXXXXwww.keysight.com