Muon g-2 and SLAC Detector TestT-519

Momentum

Spin

e

David HertzogUniversity of Washington

Page 2: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

wa Subset of g-2 Collaboration*

2/20

University of Washington

Cornell

University of Kentucky

University Illinois

University of VirginiaJames Madison

Shanghai Jiao Tong

*38 Institutes, 150 members

Frascati

Naples

Trieste

Pisa

Roma II

Udine

At SLAC

Page 3: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Three sigma is not enough to claim a discovery

BNL E821

2012 e+e- Thy

3.6 s

x10-11

Future Goals

Goal: 140 ppb

Expected Improvement

Expect to turn on in late 2016

a P5 priority

Page 4: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

4

Ring moved to FNAL and Magnet being assembled

g-2

Page 5: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Why SLAC is importantWe measure muon decay to electrons 0 – 3.1 GeVDetectors must measure Energy and its Time and we must optimize them

e+

Calorimeter

Electronics

Events above threshold

Software threshold

5

Page 6: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Muon g-2

• Read out by SiPMs• (pixelated avalanche photodiodes run in Geiger mode)

Compact, non-magnetic calorimeters made of 54 Lead-Fluoride Cherenkov Crystals

Page 7: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Muon g-2

Last November: SLAC T-512

7

• 2.5 – 4 GeV e-

• SiPMs (cooled), augmented by PMTs on sides

• Digitizers for pulse shapes

3.5 GeV

Linearity

Measured resolution with limited size array

Learned how to use the beam effectively

Page 8: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Muon g-2

July 2014: T-519 Test 4x7 array of PbF2 crystals with SiPM readout using 2.2 – 4.5 GeV beam

8

Page 9: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Muon g-2

Calibration System Set Absolute Gains

Tested in Seattle 2 weeks ago and also at SLAC

9

• Laser Filter Wheel 10:90 Splitter– Source Monitor (pin diode) & PMT– Sphere or Diffuser

• 4 monitors (pin diodes, PMT)• Fiber Bundle (60; 800 mm id)

– Front plate + Focusing Lens» Crystals / SiPMs

Pulse shape

pe from laser pulses

Page 10: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Typical beam on detector

0 1 2 3 4 5

electrons per pulse

Usually Good small spot

But, not always so …

Page 11: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Linearity Beam Energy Selection

There remains an ambiguity between “requested” energy and delivered energy and an independent system that can calibrate the beam true energy (i.e., not the detector under “test”)

Page 12: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Now a selection of the good findings in what should be a publishable set of results

dE/E = 2.2%

3.27 GeV

1500 pe/GeV

Resolution 2.7% at 2 GeVEnergy resolution exceeds our goal by a factor of 2

Page 13: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Position scanning using high-quality pencil-like beam and remote XY table

Horizontal Scan (no gain correction)

Comment: Only possible with steady and high-quality beam conditions(takes many hours)

Page 14: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Gain stability very important for g-2Measured over 8 hours: (1.0 ± 1.4) x 10-4 / hour

Data corrected by concurrent laser calibration system

Page 15: Muon g-2 and SLAC Detector Test T-519 Momentu m Spin e David Hertzog University of Washington

Summary• Important for g-2• Beam setup and quality critical• Great results to far

• Hope to come again with complete calorimeter and all onboard electronics and DAQ. Possibly also Straw Tracker system

• Thanks