121
MultiLab CE v3.0 User Guide For the Nova5000

Multilab Para Windows CE V3

Embed Size (px)

DESCRIPTION

Manual para windos de Multilab

Citation preview

Page 1: Multilab Para Windows CE V3

MultiLab CE v3.0

User Guide

For the Nova5000

Page 2: Multilab Para Windows CE V3

Information in this document is subject to change without notice.

© 2011 Fourier Systems Ltd. All rights reserved.

Reproduction of this document in any manner without the prior written consent of Fourier Systems Ltd. is strictly forbidden.

Trademarks used in this text include:

Nova5000, MultiLab, Vernier

Microsoft and Windows are either registered trademarks or trademarks of the Microsoft Corporation in the United States and/or

other countries.

June 2011

P/N BK085 Rev: A

Page 3: Multilab Para Windows CE V3

Customer Support

Fourier Systems is always ready to provide you with any technical support you might need regarding MultiLab, or the Nova5000.

Please contact us as follows:

Toll-free telephone (USA only): 1-866-771-NOVA

e-mail: [email protected]

Web: http://www.fourieredu.com/support.html

Page 4: Multilab Para Windows CE V3

2 Chapter 1 – Introduction to MultiLab

Table of Contents

Appendix A:  What’s New in Multilab 3! ............................................................ 6 

Appendix B:  Chapter 1 Introduction to MultiLab ............................................ 6 

1.1  MultiLab Window Layout ....................................................... 7 

1.2  MultiLab Main Toolbar ........................................................ 11 

1.3  MultiLab Graph Toolbar ...................................................... 12 

1.4  Meter View Toolbar .............................................................. 14 

Appendix C:  Chapter 2 Connecting Sensors to the Nova5000 ........................ 15 

Appendix D:  Chapter 3 Operating MultiLab ................................................... 19 

1.5  3.1 Collecting Data .............................................................. 19 D.1.1.  3.1.1 Launching the MultiLab software ...................... 19 D.1.2.  3.2.2 Connecting the Sensors ...................................... 19 D.1.3.  3.1.3 Setting up the Sensors ........................................ 20 D.1.4.  3.1.3.1 Autodetect and Manual mode sensor settings 20 D.1.5.  3.1. Select Display ....................................................... 27 D.1.6.  3.1.5 Change the X-axis data value ............................ 28 D.1.7.  3.1.6 Turn the graph into a bar graph .......................... 29 D.1.8.  3.1.7 Recording Data ................................................... 30 D.1.9.  3.1.8 Saving your Data ................................................ 31 

1.6  3.2 Viewing Data ................................................................. 32 D.1.10.  3.2.1 Data Map ......................................................... 32 D.1.11.  3.2.2 Graph View ...................................................... 34 D.1.12.  3.2.3 Graph View – Y-Axis ...................................... 35 D.1.13.  3.2.4 Table View ....................................................... 38 D.1.14.  3.2.5 Meter View ...................................................... 40 

1.7  3.3 Opening Data Sets .......................................................... 43 D.1.15.  3.3.1 Opening Saved Data Sets ................................. 43 

1.8  3.4 Analyzing the Data ......................................................... 43 D.1.16.  3.4.1 Reading Data Point Coordinates ...................... 44 D.1.17.  3.4.2 Reading the Difference between Two Coordinate

Values 44 D.1.18.  3.4.3 Using the Analysis Functions ........................... 44 D.1.19.  3.4.4 Manual Curve Fit ............................................. 46 D.1.20.  3.4.5 Statistics ........................................................... 48 

1.9  3.5 Printing ........................................................................... 48 

Page 5: Multilab Para Windows CE V3

1.10  3.6 Exporting/Importing CSV Files .................................... 49 D.1.21.  3.6.1 Exporting CSV Files ........................................ 49 D.1.22.  3.6.2 Importing CSV Files ........................................ 50 

1.11  3.7 Resetting the Logger ..................................................... 51 

Appendix E:  Chapter 4 52 

Appendix F:  Configuring Sensors in MultiLab ............................................... 52 

1.12  4.1 Choosing the Right Setup ............................................. 52 F.1.1.  4.1.1 Sampling Rate ..................................................... 52 F.1.2.  4.1.2 Manual Sampling ................................................ 53 

1.13  4.2 Ohaus Scale .................................................................. 54 F.1.3.  4.2.1 Connecting the scale .......................................... 54 F.1.4.  4.2.2 Configuring the Ohaus scale .............................. 55 F.1.5.  4.2.3 Running an experiment with the Ohaus scale .... 57 

1.14  4.3 Programming Rules and Limitations ............................ 57 F.1.6.  4.3.1 Number of Sampling Points ................................ 57 

1.15  4.4 Setting the Zero Point of a Sensor ................................ 58 

1.16  4.5 Triggering ..................................................................... 60 

1.17  4.6 The Timing Wizard ....................................................... 62 

1.18  4.7 Calibrating Sensors ...................................................... 67 F.1.7.  4.7.1 Hardware Offset Calibration ............................... 67 F.1.8.  4.7.2 Nova5000 Automatic Zero Calibration ............... 68 F.1.9.  4.7.3 pH Temperature Compensation .......................... 68 F.1.10.  4.7.4 MultiLab Sensor Calibration ............................. 68 F.1.11.  4.7.5 Factory Calibration (No Calibration Required). 69 

1.19  4.8 Defining New Sensors ................................................... 70 

1.20  4.9 Changing Preferences .................................................. 71 

Appendix G:  Chapter 5 75 

Appendix H:  The Analysis Wizard .................................................................... 75 

1.21  5.1 Using the Analysis Wizard ............................................ 75 

1.22  5.2 Analysis Functions List ................................................. 76 

5.3 Timing Analysis Wizard .................................................................................... 80 H.1.1.  5.3.1 Working with the Timing Analysis Wizard ....... 80 H.1.2.  5.3.2 Measuring Methods ............................................ 81 H.1.3.  5.3.3 Time Schemes and Calculations ......................... 84 

Page 6: Multilab Para Windows CE V3

4 Chapter 1 – Introduction to MultiLab

H.1.4.  5.3.4 Tips On Using the Timing Wizard and Timing Analysis Wizard. 89 

Appendix I:  Chapter 6 91 

Appendix J:  Working in Graph View .............................................................. 91 

1.23  6.1 The Cursor .................................................................... 91 

1.24  6.2 Autoscale/Graph Properties ......................................... 94 J.1.1.  6.2.1 Autoscale ............................................................. 94 J.1.2.  6.2.2 Manual Scaling .................................................... 94 

1.25  6.2.3 Adjusting the Y-Axis and X-Axis ................................ 95 J.1.3.  6.2.4 Formatting the Graph Colors ............................... 95 

1.26  6.3 Zooming ........................................................................ 96 J.1.4.  6.3.1 Zooming into a Specific Area .............................. 96 

1.27  6.4 Annotations ................................................................... 96 

1.28  6.5 Adding an existing Graph or saved data set to a current Project 97 

1.29  6.6 Smoothing ..................................................................... 98 

1.30  6.7 Prediction Tool ............................................................. 98 

Appendix K:  Chapter 7 100 

Appendix L:  Additional Tool Menu Options ................................................. 100 

1.31  7.1 Unit Settings ............................................................... 100 

1.32  7.2 Graph Title ................................................................. 101 

1.33  7.3 Crop ............................................................................ 101 

1.34  7.4 Manual X-Axis ............................................................ 102 

Appendix M:  Chapter 8 104 

Appendix N:  The Workbook ............................................................................ 104 

1.35  8.1 Working with the Workbook Feature .......................... 104 

1.36  8.2 Opening a Worksheet ................................................. 104 

1.37  8.3 Create Your Own Worksheet ...................................... 106 

Appendix O:  Chapter 9 109 

Page 7: Multilab Para Windows CE V3

Appendix P:  Sensors Supported by the Nova5000 ........................................ 109 

Appendix Q:  Chapter 10 116 

Appendix R:  Adding a Code Resistor to a Custom Sensor ........................... 116 

Appendix S:  Chapter 11 118 

Appendix T:  Nova5000 Data Logger Specifications ...................................... 118 

Page 8: Multilab Para Windows CE V3

6 Chapter 2 – Connecting Sensors to the Nova5000

Appendix A: What’s New in Multilab 3!

Ability to display 2 or 3 different views at one time while logging or analyzing data

Larger menu size, font size, and icon size for better visibility Allows you the option to use your finger to easily navigate the

software, lessening the dependency of the stylus New functions to calculate slope and area on the graph New two y axis view when multiple sensors are connected Support for a Ohaus balance scale via USB New x axis icon for displaying the data from a sensor on the x

axis instead of the time Having the option to change the settings of all the sensor

proprieties from one place. Ability to change the X-axis data values Ability to turn the graph into a bar graph

Appendix B: Chapter 1 Introduction to MultiLab

Fourier Systems has developed a Windows CE version of its renowned MultiLab software, supporting the Nova5000's data logging

capability.

MultiLab is a powerful software tool, letting you collect, display and analyze data, turning the Nova5000 into a true Digital Lab and

engaging students with science.

Using the Nova5000's four sensor ports, you can run up to eight sensors simultaneously from a selection of over 65 Fourier sensors,

20 Vernier sensors, and 8 Data Harvest sensors.

Page 9: Multilab Para Windows CE V3

Chapter 2 – Connecting Sensors to the Nova5000 7

MultiLab's flexible and comprehensive features let you:Collect and display data online, in real-timeDisplay data in graphs, tables, and

metersAnalyze data using the Analysis WizardImport/Export data as CSV filesRun Workbooks to guide students through lab activitiesOpen

video files to watch prerecorded experimentsAnd much more

1.1 MultiLab Window Layout

Due to the new layout of the Multilab software, the interface is simple and easy to understand. Below are some of the different layout

possibilities with the Multilab software.

Main window layout (default layout first visible when launching MultiLab)

Page 10: Multilab Para Windows CE V3

8 Chapter 2 – Connecting Sensors to the Nova5000

Data Map and Graph view

Data Map and Table View

Data Map

Page 11: Multilab Para Windows CE V3

Chapter 2 – Connecting Sensors to the Nova5000 9

Data Map and Meter view

Graph View and Meter View

Table View and Graph View

Page 12: Multilab Para Windows CE V3

10 Chapter 2 – Connecting Sensors to the Nova5000

Data Map, Video View, Graph View, and Table View

Page 13: Multilab Para Windows CE V3

Chapter 2 – Connecting Sensors to the Nova5000 11

1.1.1. Status and Title Bars

The bottom pane of the MultiLab window consists of the Status bar and Title bar.

Status Bar

The Status bar indicates the current status of the MultiLab software.

For example:

Logger ready – MultiLab ready to log data Run – MultiLab is currently logging data

Stopping Logger – MultiLab is stopping data logging Download – MultiLab is currently downloading data

Title Bar The Title bar indicates the name of the file currently opened.

1.2 MultiLab Main Toolbar

The MultiLab toolbar consists of the following buttons for operating the main MultiLab features:

Open file

Save file

Run logger

Page 14: Multilab Para Windows CE V3

12 Chapter 2 – Connecting Sensors to the Nova5000

Stop logger

View / Hide Data Map

Set up logger

Launch video

View / Hide Graph view

View / Hide Table view

View / Hide Meter view

Launch Workbook

Analysis wizard

Linear fit

Derivative

This toolbar is always available no matter what view MultiLab is in.

1.3 MultiLab Graph Toolbar

In Graph view, tap the Gears icon in the lower right corner to enable/disable the following buttons for operating the graph features.

Page 15: Multilab Para Windows CE V3

Chapter 2 – Connecting Sensors to the Nova5000 13

Add annotation

Move annotation

Zoom in

Autoscale

More smoothing

Less smoothing

First cursor

Second cursor

Back cursor

Forward cursor

Graph properties

Change the X-axis data value

Pause/Play predictions

Add prediction

Erase prediction

Turn the graph into a bars graph

This toolbar is not visible in the following modes:

Comment [EM1]: This is a new function it should be explained in detailes

Comment [EM2]: Same as the previous comment

Page 16: Multilab Para Windows CE V3

14 Chapter 2 – Connecting Sensors to the Nova5000

Table view Meters view

Workbook mode

1.4 Meter View Toolbar

The Meter view toolbar is visible when operating in Meter view. Enter

Meter view using the button. The toolbar buttons are as follows:

Analog view

Bar view

Color view

Digital view

Switch between single Meter and split

Meter views See section ??? for a full

Note: Multilab can display up to 3 different views at once and can toggle between them during the experiment

Comment [EM3]: Maybe here is the place to say that we can display concurrently up to 3 different views and we can toggle between them during the experiment

Page 17: Multilab Para Windows CE V3

Chapter 2 – Connecting Sensors to the Nova5000 15

Appendix C: Chapter 2 Connecting Sensors to the

Nova5000

Before you start using the MultiLab software, you should first familiarize yourself with the Nova5000 sensor interface.

Note: You are able to choose from over 65 Fourier sensors, 20 Vernier sensors, and 8 Data Harvest sensors to connect to

the Nova5000. See page 109 for a full list of supported sensors.

Sensors are connected to the 4 (I/O) sockets situated on the top of the Nova5000. These ports are labeled I/O-1 to I/O-4, as shown

below. All four sockets can be used simultaneously.

Page 18: Multilab Para Windows CE V3

16 Chapter 2 – Connecting Sensors to the Nova5000

To connect a sensor to the Nova5000 you must use a Fourier mini-din cable.Plug one end of the cable into the Nova5000 with the arrow

facing down, and the other end into the sensor.

If you are using one sensor only connect it to I/O-1. If you are using two sensors connect them to I/O-1 and I/O-2, and so on.

Note: The mini-din cables are not included with the Nova5000 device and must be purchased separately.

Connecting more than 4 sensors at a time

Page 19: Multilab Para Windows CE V3

Chapter 2 – Connecting Sensors to the Nova5000 17

The Nova5000 will support the connection of up to 8 sensors simultaneously.

In order to connect more than 4 sensors at a time, you will need to use one or more splitter cables.

Using a splitter cable

To connect a splitter cable, first connect to the end of the cable with a single plug with two arrows to one of the Nova5000 I/O ports. The other end of the cable has two plugs which are marked. One is

marked with two arrows and the other is marked with the letter “S”.

The cable marked with the arrows is the main cable input which, when connected to a sensor, is assigned the lower I/O number on the Nova5000. The second plug is marked with the letter S (Split)

indicating that it is the secondary input (the higher I/O number).

Example: If you plug the splitter cable into I/O 1, the cable with the arrows is assigned I/O 1 and the cable with the “S” is assigned I/O 5.

I/O-4 splits into I/O-4 and I/O-5

I/O-3 splits into I/O-3 and I/O-6

Page 20: Multilab Para Windows CE V3

18 Chapter 2 – Connecting Sensors to the Nova5000

I/O-2 splits into I/O-2 and I/O-7

I/O-1 splits into I/O-1 and I/O-8

Note: Before connecting the mini-din cable to the Nova5000 or the sensor sockets, make sure that the mini-din plug is correctly positioned in front of the socket. Connecting the cable in an awkward position might cause damage to the

cable pins.

Connecting Vernier and Data Harvest Sensors

If you are using a Vernier or Data Harvest sensor, first consult Appendix A to ensure that this sensor is supported by the Nova5000.

You must also have the Vernier adapter supplied by Fourier in order to connect Vernier sensors to the Nova5000 sensor interface.

Data Harvest sensors require no adapter and connect directly to the Nova5000.

For MultiLab to recognize third-party sensors, you must define them in the software using the Define New Sensors option. Please refer to

section 4.8

Page 21: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 19

Appendix D: Chapter 3 Operating MultiLab

In this chapter, you'll learn the basics for collecting, viewing and analyzing your data using MultiLab.

1.5 tap3.1 Collecting Data

D.1.1. 3.1.1 Launching the MultiLab software

It is possible to launch the software from two different locations:

Double tap the MultiLab Icon on your desktop. Go to the Start menu; choose Programs, then Science &

Math, then MultiLab.

.

D.1.2. 3.2.2 Connecting the Sensors

Connect between one to four sensors to the Nova5000 sensor inputs. The arrow on the sensor plug should be face down.

Sensors must be added successively, starting with I/O-1 and then I/O-

2 and so on.

Page 22: Multilab Para Windows CE V3

20 Cha

I/O-1 is the input located closest to the power adapter.

All I/O Ports are labeled on the Nova5000.

See Appendix C: Connecting Sensors to the Nova5000 for more details.

D.1.3. 3.1.3 Setting up the Sensors

D.1.4. 3.1.3.1 Autodetect and Manual mode sensor settings

Setting up the sensors is a key step in preparing for data logging. Here, you define the specific sensors you are using and the sampling

rate and sampling size required.

On the Logger menu select Setup or tap the Setup button . The Setup window will open.

On the Sensors tab, the Auto Detect Sensors checkbox should be selected (this is the default setting). The connected sensors should be

automatically displayed in the Input fields.

Page 23: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 21

Note: If you are using more than 4 sensors, you cannot use Auto Detect Sensors. You will have to manually choose each

sensor.

If you are using a sensor which is not automatically detected unselect the Auto Detect Sensors checkbox. You must select the relevant

sensor name from the Input drop-down menu.

List of sensors that are not auto-detected by Multilab:

O2 gas

Smart pulley

Geiger Muller

Flow rate

Rain collector

Drop counter

Volume

3.1.3.2 Sensor Properties

To adjust the Display properties of a specific sensor, tap the Sensor

Properties button adjacent to the Input field.

Page 24: Multilab Para Windows CE V3

22 Cha

In the Sensor properties / Measurement dialog box, you can select the unit of measurement (e.g., choose between (mT) or

(gauss) for a Magnetic Field sensor)

In the Sensor properties / Display Properties dialog box, you can adjust the color of the sensor plot and of the y-axis in the

Graph view

Page 25: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 23

To define the scale of the data that you want displayed (e.g., 0.15 - 0.2 instead of -0.2-0.2 for the Magnetic Fieldy sensor), unselect the Autoscale checkbox and enter the scale in the

Min and Max fields

Page 26: Multilab Para Windows CE V3

24 Cha

To restore the color or scale settings to default, tap Restore Defaults

To adjust the calibration of the sensor tap on the Calibration tab

Note: Not all the sensors will have the Calibration Tab

Note: See section 4.7.4 for more information regarding the Calibration process.

To enable the Set Zero function of, tap the Set Zero Tab

Page 27: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 25

Note: Not all the sensors will have the Set Zero function.

Note: See section 4.4 for more information regarding the Set Zero feature.

Tap OK to confirm your settings and exit the dialog box, or Cancel to exit without saving

3.1.3.3 Setting the Logging Rate

On the Logger Setup box, tap the Rate tab. In the Rate drop-down

menu, select the desired number of samples per second/minute/hour.

Page 28: Multilab Para Windows CE V3

26 Cha

If you select the Manual recording rate, you can manually record each data point. To perform manual logging, first continue to the end of the Setup process. Tap Run to record the first sample. When ready, tap Run again to record the second sample and so on To stop the Manual data logging session, go to the Logger menu and choose Stop Sampling rates: If the sampling rate is set to 100 samples per second or higher, the data will be automatically displayed on the graph once the recording has stopped. If the sampling rate is set to a slower rate, (every 1 minute, every 10 seconds) The Nova5000 will only take a reading at that interval. EXAMPLE: If you are logging every 1 minute, you will not see the data change in real time. You will only see what the reading is at the 1 minute mark, 2 minute mark, 3 minute mark.

Page 29: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 27

3.1.3.4 Set the number of samples Tap the Samples tab and in the drop-down menu, select the relevant

number of samples to record. You will see the maximum recording time based on the number of samples chosen.

On this tab, you can also choose to clear the memory of the data logger before you start recording fresh data. Just select the Clear memory checkbox and tap OK on the upper right of the window.

Note: This will clear all experiment data currently residing in the logger memory – not the MultiLab software data files.

To learn how to use the Triggering function, refer to section 4.5

D.1.5. 3.1. Select Display

From the main MultiLab toolbar, select the display in which you want to view the data recording. You can change views while logging data

or while analyzing data.

Tap one of the following display icons:

Data Set

Page 30: Multilab Para Windows CE V3

28 Cha

Graph

Table

Meter

Multiple views can be displayed simultaneously by tapping them on or off.

The following image shows the Data Map, Graph, and Table views being displayed side-by-side.

MultiLab will only support up to 3 views at a timeand does not support every view combination.

D.1.6. 3.1.5 Change the X-axis data value

To change the X-axis data value, tap the Gears Icon , then the X-

axis icon in the toolbar that is displayed. This will cause the X axis windows to appear.

Page 31: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 29

Here you can choose each sensor from the drop down menu and the select the input or scale you want to represent the X axis.

Here we used temperature as input 1 and ran a data session. You

can see I can select the X axis to represent Time, Temperature, or the Exp 1 data set.

D.1.7. 3.1.6 Turn the graph into a bar graph

To change the line graph (default setting), to a bar graph, tap the

Gears Icon , then the Bars icon .

You will then see the graph change from line graph to bar graph.

Page 32: Multilab Para Windows CE V3

30 Cha

D.1.8. 3.1.7 Recording Data

Note: When MultiLab is recording data, the Nova5000 will not go into Suspend. Instead, the Nova will inform that

there is an experiment in progress and ask it should go into suspend mode.

To start recording data, select Run in the Logger menu or tap Run

on the main toolbar.

Page 33: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 31

As soon as you tap Run, you will see the data being logged in the MultiLab window, in one of the three displays you selected in section

D.1.5 above.

While data is being recorded, you can switch between the three display views: Graph, Table or Meter view.

To manually stop recording data, tap Stop in the Logger menu, or tap

the Stop icon on the main toolbar.

MultiLab will stop recording automatically once the predefined sample size has been reached.

Note: If you accidentally disconnect a sensor during data collection, MultiLab will continue to Run but the collected

data from the point of disconnection will be inaccurate. It is recommended you restart the data collection process.

D.1.9. 3.1.8 Saving your Data

Once recorded, your data is added to the Data Map as a data set (see page 32). However, if you do not save this data, you will lose it upon

exiting MultiLab.

You will be prompted to save open data when exiting MultiLab.

To save your data set, select Save File in the File menu or tap the

Save button in the main toolbar.

This will save all the open data sets in the Data Map under one file name. You cannot save individual data sets.

Note: The default path for files is the Multilab Data folder. This folder is located in \Program Files\Fourier

Systems\Multilab CE\Multilab Data

Page 34: Multilab Para Windows CE V3

32 Cha

To save a file under a different name, select Save As in the File menu.

1.6 3.2 Viewing Data

As previously mentioned MultiLab features three views for viewing recorded data: Graph, Table and Meter. You can easily switch

between these views using the corresponding buttons on the main toolbar.

Once you have completed your experiment you can view the recorded data offline in Graph or Table view.

Meter view has a snapshot mode that will allow you to see the reading at the cursor’s location on the graph.

D.1.10. 3.2.1 Data Map

The Data Map is situated on the left of the MultiLab window and lists the data sets that were recorded in the current MultiLab session. The Data Map is populated by running experiments or opening previously

saved data sets.

A data set contains all data recorded in a particular experiment. Each sensor used in the experiment has its own entry in the data set, and you can toggle which data, if not all, to display in the graph or table.

When you apply analysis functions onto your data, an entry for that specific function is also added to the Data Map.

When using the Data Map, the following actions can be performed:

To display the complete list of sensors measurements for an individual experiment, tap the plus sign (+) next to the experiment entry to

expand its data set To collapse the sensor list under an individual experiment, tap

the minus sign (-) next to the experiment entry

To display a specific data set tap its name in the list, and then tap Show situated at the bottom of the Data Map. You can also

choose to display data from specific sensors in the data set. Simply tap the sensor name and tap Show

Page 35: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 33

To hide a data set tap its name in the list, and then tap Hide situated at the bottom of the Data Map. You can also choose to hide data from specific sensors in the data set. Simply tap the

sensor name and tap Hide

To remove a data set from the Data Map, tap and hold on the data set and select Delete Data

You can also tap and hold on the data set to reveal commands

Display on Graph

Delete Data

Rename

Note: Showing or hiding a data set via the Data Map affects both the Graph and the Table view.

Note: Remember, you can hide this window or enable other window views to better see your data.

Page 36: Multilab Para Windows CE V3

34 Cha

D.1.11. 3.2.2 Graph View

Tap Graph to display the data in Graph view. The default graph display is the data set or sets plotted vs. time.

The graph usually displays all the data sets of a given recording, but you can use the Data Map to remove or hide one or more of the data

sets from the graph.

To change the color of the sensor data, tap the Gears Icon , then

the Graph Properties icon in the toolbar that is displayed. This will cause the Format Graph dialog box to be displayed.

Here you can choose each sensor from the drop down menu and the select the color and scale of the data set.

Autoscale: Multilab will automatically choose the proper scale based on the data it receives.

Visible Lines: The graph will make a solid line through the data points

Markers: The graph will add a mark for each data point.

Page 37: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 35

Color: Allows you to change the data set’s associated color on the graph

Restore defaults: This will set the default values of Autoscale checked, Visible Lines checked, Color of red.

Note: You must choose either Visible Lines or Markers. The

default setting is Visible Lines.

Note: Restore defaults will restore all default values across all data sets.

D.1.12. 3.2.3 Graph View – Y-Axis

One sensor:

If there is only 1 sensor connected to the Multilab software, you will see the Y-axis on the left hand side.

Page 38: Multilab Para Windows CE V3

36 Cha

Two Sensors:

If there are 2 sensors connected, you will see 2 Y-axis, one on the left and one on the right.

The Y-axis on the left represents the 1st sensor, and the y-axis on the right represents the 2nd sensor.

In this case, the Magnetic Field sensor is on the left (1st sensor) and the temperature is on the right (2nd Sensor).

Three or more sensors:

If there are 3 or more sensors connected to Multilab, you will see 2 Y-axis, one on the left and one on the right.

The Y-axis on the left represents the 1st sensor, and the y-axis on the right represents the 2nd sensor.

To see the 3rd ,- 8th sensor, you must tap on data on the graph for the sensor you wish to view that specific Y-axis.

Page 39: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 37

Here we have tapped on the Magnetic Senors data.

Since it is the 1st sensor it’s y-axis will always be displayed on the left.

Here we have tapped on the CO2 Sensors data. Since it is the 2nd sensor, it’s y-axis will be displayed on the right.

Page 40: Multilab Para Windows CE V3

38 Cha

Here we have tapped on the Temperature Sensors data. Since it is the 3rd sensor, it’s y-axis will be displayed on the right.

Tapping each sensors data (Sensors 2-8) will reveal the y-axis on the right. By tapping each sensor data, you will be alternating which y-axis senor data is on the right side of the graph.

D.1.13. 3.2.4 Table View

Tap Table to display the data table.

The data that is displayed in the Table view always matches the data in the Graph view: each sensor Input is represented by a separate

column in the table.

Use the Data Map to change the data that is displayed in the table. You can see below the table view along with other views in Multilab.

Page 41: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 39

Add Manual Column You can manually add a new data set to a data table, even after the

experiment has been conducted. You can then view this data set as a plot line in Graph view:

From the Tools menu, select Add Manual Column.

Page 42: Multilab Para Windows CE V3

40 Cha

In the Insert manual column dialog box, enter the column title and column unit.

On the data table, manually enter data into the newly created column.

In the Data Map, select the new data set and tap Show.

D.1.14. 3.2.5 Meter View

Meter View consists of the following views. Switch between views by tapping the corresponding icon on the Meter toolbar situated below

the Meter View window.

Analog view

Bar view

Color view

Digital view

Switch views

The Switch Views feature allows you to switch between single and multi view.

MultiLab enables you to view data in Meter format in the main MultiLab

window. In Meter view, you will see real-time data readings from the sensors currently connected to MultiLab.

You can see below the Meter view along with other views in Multilab.

Page 43: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 41

Use Meter view to:

View live data without actually recording this data i.e. to monitor the current real-time sensor readings (Snapshot mode)

View data in real-time while it is being recorded by MultiLab (Run mode)

The meters’ scaling automatically matches the graph’s scaling.

While viewing or collecting data in Meter view, you may view data from up to four sensors at a time. If you have more than one sensor connected, switch between these sensors on-the-fly using the drop-

down sensor menu located above the Meter view.

To select the Meter view, tap the Meter icon in the main toolbar.

The Single Meter view or Multiple Meter view window will open, depending on which view you were in last.

Page 44: Multilab Para Windows CE V3

42 Cha

Single Meter view Multiple Meter view

In Single Meter view, you can select the sensor to be displayed via the drop-down menu at the top of the screen. Change the specific

Meter view via the icons at the bottom of the screen.

In Multiple Meter view, tap on one of the four panes within the window and select the sensor to be displayed via the drop-down

menu at the top of the screen. Change the specific Meter view via the icons at the bottom of the screen.

Toggle between Single and Multiple views using the icon.

Snapshot Mode In Snapshot mode, MultiLab continuously displays the real-time data

but the data is not actually saved i.e. a snapshot of the data is

displayed. A small camera icon indicates you are in Snapshot mode, and is located on the bottom left of the Meter window.

To work in Snapshot mode, simply connect your sensor and switch to Meter view. As long as you don’t tap Run, you will remain in Snapshot mode and MultiLab will display the sensor data. MultiLab will display

the data at a sampling rate of one sample per second.

Note: When using a temperature sensor in snapshot mode,

you can change the temperature units from Celsius, Fahrenheit, or Kelvin in the preferences window.

Comment [EM4]: What about the setting of the temperature units from thr logger?

Comment [D5]: I don’t understand what you are asking- DAVE

Page 45: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 43

Note: When using a Rain sensor in snapshotmode, you can change the units from mm to inches in the preference

window. For more information on the Preference window, See Section

4.9

Run Mode To start collecting data while remaining in Meter display, simply tap

Run in the main toolbar. The icon will switch to a small

running man icon , indicating you are now in Run mode. The data is displayed at the sampling rate defined in the Setup wizard.

1.7 3.3 Opening Data Sets

D.1.15. 3.3.1 Opening Saved Data Sets

To open previously saved data sets, select Open File from the File menu and select the desired file from the Open File dialog box.

The saved file, with all the data sets included in it, will be added to the Data Map.

Remember: Once MultiLab is closed, all unsaved data sets will be lost.

1.8 3.4 Analyzing the Data

Once you have collected and viewed your data, you have the option to analyze it using a variety of tools available in MultiLab.

Page 46: Multilab Para Windows CE V3

44 Cha

Appendix E: The sections below detail the various analysis methods. Also

refer to Appendix H: Chapter 5

The Analysis Wizard for further details.

E.1.1. 3.4.1 Reading Data Point Coordinates

When in Graph view, position the cursor (see page 73) on a point on the graph. Its coordinates will be displayed at the bottom of the graph

window.

E.1.2. 3.4.2 Reading the Difference between Two Coordinate Values

First, position one cursor on the first point and then position a second cursor on the second point on the graph. The difference between the

two coordinate values will be displayed at the bottom of the graph window.

E.1.3. 3.4.3 Using the Analysis Functions

A number of analysis functions are available under the Tools menu. These functions can only be applied to data sets that are currently

displayed in the graph window.

When applied, the analysis function will be visible on the graph.

Page 47: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 45

To apply an analysis function:

First ensure the relevant graph is open in the main Graph window.

Use the graph cursors to select the data range to which you want to apply the analysis.

Select Analysis on the Tools menu.

The Analysis menu will open. The following functions are available:

Analysis Wizard - Opens a wizard, guiding you though an extensive list of analysis functions. Linear Fit - Draws a line of linear least square fit y = ax + b Power Curve Fit - Draws a line of power least square fit

Exponential Curve fit - Draws a line of exponential least square fit

Quadratic Curve fit - Draws a line of quadratic least square fit cbxaxy 2

Manual Curve fit - Manually applies a curve fit to your data plot. See section E.1.4 below for more details Derivative - Use Derivative to construct a graph in which each point is the slope of the three consecutive points on the source graph Integral - Use Integral to construct a graph in which each point is the integral of all the preceding points on the source graph. Statistics - Displays statistics for a selected range of data. See section E.1.4 below for more details. Timing Wizard - Measure and calculate many types of time events, including velocity and acceleration, with one or two photo gates. Slope – Calculates and draws the formula for slope at a given point on the graph. Area – Calculates and displays the area on the graph between 2 points along the same data plot

nx

ay

bxaey

Page 48: Multilab Para Windows CE V3

46 Cha

Selecting one of the functions above will apply that function to the graph immediately, and the curve fit equation and correlation

coefficient (R2) will be displayed on the information bar below the graph.

Note: You can also apply the Linear Fit and

Derivative functions by tapping the respective buttons in the main toolbar.

E.1.4. 3.4.4 Manual Curve Fit

You may manually apply a curve fit to your data plot. You have the option to apply linear, quadratic and power curve fits.

Select a plot on the graph and go to the Tools menu, choose Analysis, then Manual curve fit.

Select the relevant X Axis and Y Axis to be displayed on the graph.

Page 49: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 47

The top slide bar will adjust the first number in the equation (A).

The bottom slide bar will adjust the second number in the equation (B).

You can see the equation

= +

The numbers on the left and right of the slide bars are minimum and maximum numbers for adjusting the sidebar. By default, Multilab

will choose the range that fits the data set.

Select the curve fit you wish to apply. Use the icons situated below the graph. The function will then be displayed at the bottom of the

screen.

Manual Curve Fit toolbar

1 – Linear curve fit

2 – Quadratic curve fit

3 – Power curve fits

4 – Add graph to project (save graph)

5 – Exit

6 – Manual Curve Fit equation

Page 50: Multilab Para Windows CE V3

48 Cha

E.1.5. 3.4.5 Statistics

By selecting Statistics from the Tools menu, choose Analysis; you can then display statistics of a selected data set or a range of data.

The statistics include:

Average - The average of all the numbers in the range Median - The median (or middle) data point in the range

Mode - The data point with the highest frequency in the range StDev - The standard deviation

Minimum - The smallest value in the range Maximum - The largest value in the range Sum - Adds all the numbers in the range

Area - The area between the graph and the x-axis in the range Rate - The recording rate

Samples - The number of data points in the range

1.9 3.5 Printing

Assuming you have established a connection between the Nova5000 and a printer, you may choose to print either a graph or a table from

MultiLab.

Select the data set in Data Map and go to the File menu and choose Print. The following dialog box pops up:

To print a graph, simply select the Graph option and tap Print.

Page 51: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 49

To print a table, select the Table option and choose to either print all data in the table or a selection of data, by unselecting the Print all

data checkbox and entering the print range.

1.10 3.6 Exporting/Importing CSV Files

E.1.6. 3.6.1 Exporting CSV Files

Using the Nova5000’s PlanMaker spreadsheet application, you are able to export the raw data contained in any given data set in MultiLab

into a CSV (Comma Separated Values) file.

By doing so, you will be able further manipulate the data in PlanMaker. For example, build bar charts using the data range.

To export a CSV file:

Select the data set in the Data Map that you want to export to PlanMaker. Ensure the data is not hidden.

In the File Menu, select Export CSV file

You can also choose the Tools menu, Export to Planmaker.

The Save As dialog box will open. Enter the name of the CSV file and tap OK.

PlanMaker will launch and the Import options dialog box will pop up. Select the Comma separator option.

The main PlanMaker window will open with the exported data displayed.

Page 52: Multilab Para Windows CE V3

50 Cha

E.1.7. 3.6.2 Importing CSV Files

You can import back into MultiLab, files containing MultiLab data that were saved in CSV format.

To import a CSV file:

From the File menu, select Import CSV file.

In the dialog box that opens, browse to the location of the CSV file and tap OK.

The file will be imported to MultiLab, and you will be able to view the data in Graph and Table view.

Page 53: Multilab Para Windows CE V3

Chapter 3 – Operating MultiLab 51

1.11 3.7 Resetting the Logger

If the MultiLab application is not functioning as expected this could mean the Nova5000 data logger is experiencing operating problems

and requires a reset.

You can reset the Nova5000 by going to the Logger menu and choose Reset Logger. Resetting the logger simply reinitializes the logger and clears the logger memory (i.e. all unsaved experiments

conducted in the current MultiLab session will be deleted).

This operation does not clear the MultiLab data files, however. These files are saved in the MultiLab software folders and can only be

deleted manually.

Page 54: Multilab Para Windows CE V3

52 Chapter 4 – C

Appendix F: Chapter 4

Appendix G: Configuring Sensors in MultiLab

This chapter provides further details on using sensors with MultiLab, including setting up sensors, calibrating sensors, and other important

information.

1.12 4.1 Choosing the Right Setup

G.1.1. 4.1.1 Sampling Rate

The sampling rate should be determined by the frequency of the phenomenon being sampled. If the phenomenon is periodic, sample

at a rate of at least twice the expected frequency. For example, sound recordings should be sampled at the highest sampling rate –

20,000/sec, but changes in room temperature can be measured at slower rates such as once per second or even slower, depending on

the speed of the expected changes.

Note: Sampling at a rate slower than the expected rate can cause frequency aliasing. In such a case, the graph will show a frequency much lower than expected. In the figure below, the higher frequency sine wave was sampled at 1/3 of its frequency. Connecting the sampled points yielded a graph

with a lower, incorrect frequency.

Page 55: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 53

Frequency aliasing

After you have chosen the sampling rate, choosing the number of points will determine the logging period:

Samples / Rate = Logging time.

You can also choose the duration of an experiment first, and then calculate the number of samples:

Samples = Logging time Rate.

G.1.2. 4.1.2 Manual Sampling

Use Manual sampling for:

Recordings or measurements that are not related to time Situations in which you have to stop recording data after each sample

obtained, in order to change your location, or any other logging parameter

Note: During the experiment no changes can be made to the MultiLab sensor configuration.

To start an experiment using manual data logging:

Go to the Rate tab in the Setup Wizard

Select Manual from the Rate drop-down menu

Tap Run once to record the first sample. Tap Run again to record the next data point, and so on

Page 56: Multilab Para Windows CE V3

54 Chapter 4 – C

To stop the Manual data logging session, go to the Logger Menu and select Stop

1.13 4.2 Ohaus Scale

The Nova 5000 is able to support a digital Ohaus scale connected via USB.

G.1.3. 4.2.1 Connecting the scale

1. Plug the scale to an outlet using the provided adapter 2. Plug the scale to one of the Nova's USB ports 3. Turn on the scale and zero it if the scale is not reading 0.00 4. Run the Multilab Software

Note: You must have the Ohaus digital scale on and connected prior to running the Multilab Software. Otherwise

the scale will not be detected

5. Connect any sensor to IO/1. A sensor must be connected for the scale to work properly

6. Go to the Logger menu and choose Setup 7. Mark the checkbox for Ohaus External Scale and tap the

Properties box

Page 57: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 55

Note: You can only select the Ohaus External Scale if a scale is connected and detected by the Multilab software.

G.1.4. 4.2.2 Configuring the Ohaus scale

1. In the properties windows, you can select the scale range and unit

Range: You can choose the scale type, if it is a 0-200

gram scale or a 0-400 gram scale Unit: You can select the unit of measurement from

grams, ounces, pounds, kilograms, or Newtons

Page 58: Multilab Para Windows CE V3

56 Chapter 4 – C

2. Tap the OK button to confirm your settings or CANCEL to exit out without saving.

3. Back to the setup window, tap the rate you wish to log the experiment at.

NOTE: The Ohaus Scale will not work at rates higher than once per second. If you are logging at a rate of more than

every second, the Ohaus scale will report values of identical data. Example: If you have it set a 10 samples per second, the

Ohaus scale will report the same values in batches of 10.

4. Tap the samples tab to configure the amount of samples you wish to take and then tap on the OK button to save your settings.

Page 59: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 57

5. The Ohaus scale can be used in 2 ways, items can either be placed on the plate, or hung from a hook at the bottom of the scale.

G.1.5. 4.2.3 Running an experiment with the Ohaus scale

1. To run an experiment you can tap the run icon or you can tap the Logger menu and choose Run.

2. When you begin logging, you will see the Ohaus Scale show up as another sensor.

1.14 4.3 Programming Rules and Limitations

The following are some rules and limitations you must take into account when programming sensors for use with the Nova5000.

MultiLab integrates all programming limitations automatically and will only allow the programming of settings that comply with the rules

below.

G.1.6. 4.3.1 Number of Sampling Points

Increasing the number of active inputs limits the number of sampling points one can choose. The following condition must always be

satisfied:

Samples Active Inputs < Memory

Page 60: Multilab Para Windows CE V3

58 Chapter 4 – C

The Nova5000’s data logger memory is sufficient for 100,000 samples. However, when sampling at rates faster than 100 samples per second, the memory can store only two experiments of 32,000 samples each.

Selection of 100,000 sampling points will create 2 files of 50,000 points each in the data logger’s memory.

1.15 4.4 Setting the Zero Point of a Sensor

4.4.1 Compatible Sensors

MultiLab enables you to rescale some sensors and to set the current readings to zero for subsequent loggings. This feature applies to

sensors such as:

CO2 Distance

Force Magnetic field

Light Pressure

Spirometer

Note: Not all senors have the Set Zero function

To set the current readings of a sensor to zero:

Tap Setup on the main toolbar to open the Setup Wizard.

Comment [EM6]: It's a part of the sensor proprieties so it should appear in the relevant section and not here/

Page 61: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 59

Tap the Sensor properties icon next to the sensor input you want to set.

Tap the Set Zero tab.

Check the Set the current reading to zero checkbox.

Tap OK, and tap OK again to exit the Setup window.

MultiLab takes a single measurement and sets the reading to zero.

4.4.2 Change the zero point

MultiLab allows you to change the zero point of a sensor. In certain experiments this can be very helpful.

1. Tap Setup on the main toolbar to open the Setup Wizard.

2. Tap the Sensor properties icon next to the sensor input you want to set.

3. Tap the Set Zero tab.

4. Tap the Reset zero button.

5. Tap OK, and tap OK again to exit the Setup window.

4.4.3 Return to the default zero point

After your experiment is finished, you may want to use the senor differently and need to set the sensor back to default.

1. Tap Setup on the main toolbar to open the Setup Wizard.

2. Tap the Sensor properties icon next to the sensor input you want to set.

3. Uncheck the Set the current reading to zero checkbox.

4. Tap OK, and tap OK again to exit the Setup window.

Comment [o7]: Both this section and the next one are missing an introductory paragraph (i.e., “MultiLab allows you to do so-and-so. To do this, you must do the following:”)

Page 62: Multilab Para Windows CE V3

60 Chapter 4 – C

1.16 4.5 Triggering

Triggering allows you to set a set of values that will allow the experiment to begin recording only when those values are met.

Tap Triggering on the Samples tab of the Setup Wizard. This opens the Triggering dialog box:

Select the triggering sensor in the Based on drop-down menu

Choose one of the following from the Type options:

None - Triggering is disabled Above level - Start logging only once the measured value is higher than the trigger level Below level - Start logging only once the measured value is lower than the trigger level Event Recording - This function enables you to record the exact time and date at which a certain phenomenon occurs. The trigger level set for this option is actually a threshold setting. Each time the threshold is crossed, MultiLab will record the exact time and date of the occurrence, and will

Comment [EM8]: It's a part of the sensor proprieties so it should appear in the relevant section and not here/

Page 63: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 61

continue to do so until the desired number of samples has been obtained

Note: The trigger acts on analog measurements only (not on the distance sensor). The trigger condition must be fulfilled

for at least 300S.

Control Level - The control level trigger allows you to create an automatic sense & control system. Tap OK to exit the Triggering dialog box. Tap OK again to exit Setup

Wizard.

4.5.1 Control Level The control level trigger allows you to create an automatic sense & control system. This means that you can connect a sensor measuring a certain phenomenon and set a value on when an event will be triggered. Example: Using a fan that will start operating when the temperature measured by the sensor rises above 30 ˚C). The control level function requires the use of a splitter cable and a control sensor. The cable divides each input into a sensor cable and a controller cable. After setting the control level and starting the data logging process, the sensor will sample and record the data as usual. However, when the measurement from the sensor rises above the predetermined threshold value, the controller cable will send a pulse of 5V to the control sensor, and will continue to do so until the sensor measures a value below the threshold level. When receiving the 5V pulse, the control sensor will close/open a relay capable of switching 110/220V to any load.

Comment [D9]: SCREENSHOTS!!!

Page 64: Multilab Para Windows CE V3

62 Chapter 4 – C

1.17 4.6 The Timing Wizard

MultiLab enables you to measure events such as Time, Velocity and Acceleration using the photogate sensor and the Logger Timing

Wizard.

4.6.1 Working with the Timing Wizard Connect one photogate to I/O-1 of the Nova5000, or two photogates,

one to input I/O-1 and the other to I/O-2, depending on the type of data you require.

Note: There is not a “Photo Gate” sensor entry, the Photo Gate sensor displays as “Voltage 0-5V”

From the Logger menu, select Timing Wizard to open the Timer Module Wizard.

Select one of the options for measurement: Time, Velocity or

Acceleration.

Tap the Timing Wizard – Method tab.

Comment [EM10]: This is a working mode . we should write an introduction describing all the working modes with the multilab: Record a new experiment Open a saved experiment Manual x axis mode Timing wizard module Analysis timing wizard Playing video experiment Workbook mode Data analysis

Page 65: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 63

Select one of the options to select the measuring method.

If required, enter the body’s width in mm (an integer between 0 to 59), or the distance between the gates in cm (an integer between 0 to 99) in the text box (for velocity and acceleration measurements

only).

Tap OK to exit the Timing Wizard and tap Run to enter the timing standby mode.

Timing begins each time a body blocks the photogate in input 1 and ends when a body blocks the photogate in input 2 2 (according to the event method). MultiLab displays the results in a bar graph and in the

data table.

You can record as many measurements as you need. After each event, MultiLab adds the results as a new bar in the graph and as new

raw data in the table.

To exit the timing mode tap Stop on the main toolbar.

4.6.2 Measuring Methods

The Timing Wizard offers you various methods of analyzing the different measurements. In some measurements you will be asked to enter the width of the moving body, or the distance between the two photo gates, to allow for the calculation of velocity and acceleration.

The methods depend on the selected measurement:

Page 66: Multilab Para Windows CE V3

64 Chapter 4 – C

Time

At one gate

Measures the time it takes the body to

cross the photo gate (between blocking and unblocking the infrared

beam)

Between gates

Measures the time it takes the body to

move from one photo gate to the second

photo gate (between blocking the first and blocking the second

infrared beams)

Velocity

At one gate

Measures the time it takes the body to

cross the photo gate (between blocking and unblocking the infrared beam) and returns the

velocity.

Requires you to enter the body’s width in

mm.

Between gates

Page 67: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 65

Measures the time it takes the body to

move from one photo gate to the second

photo gate (between blocking the first and blocking the second infrared beams) and returns the average

velocity.

Requires you to enter the distance between

gates in cm.

Acceleration

Between gates

Measures the crossing time at the first gate, the time it takes the body to move from

one gate to the second gate and the crossing time at the

second gate and returns the average

acceleration.

Requires you to enter the body’s width in

mm.

Page 68: Multilab Para Windows CE V3

66 Chapter 4 – C

4.6.3 Time Schemes and Calculations

TIME MEASUREMENTS Result: Required

ParamaterMeasurementLocation

None Time At one gate

None Time Between gates

w – the body’s width

Velocity At one gate

L – the body’s width

Velocity Between Gates

w – the body’s width

Acceleration Between Gates

Note: After you have finished the experiment, use the Timing Analysis or Analysis Wizard to gather information.

(See Chapter 5)

12 ttt

13 ttt

12; tttt

wv

13; tttt

Lv

t

vva

ttttt

tt

wv

tt

wv

12

1234

342

121

2

;

Page 69: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 67

1.18 4.7 Calibrating Sensors

Though Fourier's sensors are shipped pre-calibrated from the factory, you may choose to manually calibrate the sensors in order to increase their accuracy on-site. This can be done using the various calibration

techniques provided by MultiLab.

Most of the Fourier sensors are linear, i.e. the output level of each sensor changes according to the equation:

Where:

Y - Output of the sensor (voltage level changing from 0 to 5V)

X - Sensor input

a - Sensor gain

b - Sensor offset

The calibration process allows you to control the offset, and in some cases even the gain, of a sensor. You can perform four types of

calibration:

Hardware offset calibration MultiLab Automatic zero calibration

MultiLab Sensor calibration Factory calibration (no manual calibration required)

G.1.7. 4.7.1 Hardware Offset Calibration

Some of the sensors include a screw which controls the sensor offset. To calibrate the sensor, use MultiLab to run an experiment with the sensor and rotate the calibration screw until the sensor shows the

correct measured value (obtain the actual correct value from another source that is known to be accurate).

baXY

Page 70: Multilab Para Windows CE V3

68 Chapter 4 – C

G.1.8. 4.7.2 Nova5000 Automatic Zero Calibration

The Nova5000 data logger is able to automatically calibrate the sensor offset for all analog sensors accurately, quickly, and for every

new experiment conducted.

The calibration method is very simple. Whenever you plug in a sensor, the data logger checks to see if the selected sensor measures a value within 2% of its zero value. If so, MultiLab sets that value as

zero.

To enable this feature, make sure that the sensors are at their zero values when you plug them in. To ensure the most accurate zero

value:

Shorten the Voltage sensor plugs Leave the Current sensor plugs open

Cover the Light, Photogate, and Microphone sensors Insert the pH sensor in a pH 7.0 solution

Unload the Force Transducer Place the Accelerometer on a stationary surface

Place the Temperature probes in ice water Place the Pressure sensor in a 1 ATM (1013 mb) chamber

G.1.9. 4.7.3 pH Temperature Compensation

To compensate a pH sensor for temperature changes, plug the temperature sensor into I/O-1, and the pH sensor in I/O-2. MultiLab

will then display the compensated pH value.

G.1.10. 4.7.4 MultiLab Sensor Calibration

MultiLab enables you to calibrate any of the linear sensors manually. This two point calibration method sets both the gain (slope) and offset

(intercept) of the sensor’s conversion function. The calibration procedure affects MultiLab readings only.

Page 71: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 69

In the Logger menu, tap Calibrate Sensors.

Select a sensor from the Choose Sensor drop-down menu and tap OK.

Enter a distinct real value in each of the Real Value fields (in the unit of the specific sensor) and the corresponding measured values in

each of the Measured Value fields (in the unit of the specific sensor). The measured values are the values displayed by

MultiLab when measuring the two real values.

Tap OK. The calibrated sensor parameters will be saved in MultiLab

To reset to the default calibration for any sensor, repeat steps 1 to 2 above and tap Restore defaults, then tap OK to complete the

process.

G.1.11. 4.7.5 Factory Calibration (No Calibration Required)

All digital sensors that are essentially timers leave the factory fully calibrated, and do not suffer from any accuracy degradation. An

example of such a sensor is the Sonic Ranger distance sensor, which measures the time passed from the transmission of a sound pulse to

its echo reception.

Page 72: Multilab Para Windows CE V3

70 Chapter 4 – C

1.19 4.8 Defining New Sensors

MultiLab enables you to define additional custom sensors. This is a useful tool for when you need the Nova5000 and MultiLab to

communicate with sensors from different sensor vendors, such as Vernier.

Any additional sensor that you would like to connect to the Nova5000 must comply with the following restrictions:

The sensor’s output must be between 0-5 volts. The sensor Transfer Function (sensor output voltage changes vs. the sampled phenomenon changes) must be a linear Transfer Function The sensor must have a code resistor in order to be automatically

identified. If the sensor does not have a code resistor, you will have to work in 8-sensor mode and select the sensor manually

To determine whether or not your sensor has a code resistor, simply connect it to the Nova5000, open MultiLab and enter the Setup

Wizard, then verify that the sensor is displayed in the Sensors tab.

Refer to page 116 to learn how to add a code resistor to your custom sensor.

To define a new sensor From the Logger menu, select Define New Sensor to open the

Define New Sensor dialog box.

Page 73: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 71

Tap Add new sensor.

Enter a sensor name and a measurement unit in the relevant fields.

Enter the sensor output in Voltage, in the Output Value #1 and Value #2 fields.

Enter the corresponding real sensor values, in the unit of the custom sensor, in the Real value fields.

Tap OK. Your sensor has now been defined and is saved in the Sensor list in the Setup Wizard.

Tap Restore defaults to restore the default sensor list, and remove the defined sensors from the Setup Wizard sensor list.

1.20 4.9 Changing Preferences

MultiLab lets the user change meters unit preferences for specific sensors. To open the Preferences dialog, go to the Logger menu and

choose Preferences.

In this section we will go over changing Preferences of the temperature sensor, counter sensor, and drop sensor.

Comment [EM11]: It have to be done from sensor proprieties in the setup menu.

Comment [EM12]: From here you can change only the meter 's units. Only from the sensor properties you can change the sensor's unit when you run an experiment. It's very comfusing…

Page 74: Multilab Para Windows CE V3

72 Chapter 4 – C

Note: From here you can change only the meter 's units. Only from the sensor properties you can change the sensor's

unit when you run an experiment. See section 3.1.3.2 for Sensor Properites.

You may change the preferences for the following sensors:

Temperature

When using any temperature sensor, use the Preference dialog to change the temperature unit to one of the following options:

Celsius Fahrenheit

Kelvin

Counter Sensor Selecting unit of rainfall

By default MultiLab displays daily and accumulated rainfall in mm. When the Rain Collector sensor is connected to the Nova5000, use

Page 75: Multilab Para Windows CE V3

Chapter 4 – Configuring Sensors in MultiLab 73

MultiLab CE to select the unit (mm or inches) in which the rainfall is measured.

Tap Logger on the main toolbar.

Choose Preferences from the drop-down menu.

Ensure Rain 0 – 800m is selected in the Counter Sensor drop-down menu. Then select the desired unit (mm or inches) from the Unit

drop-down menu.

Tap Setup on the main toolbar and unselect the Auto Detect

Sensors checkbox. Select the Rain 0 - 800m sensor from the drop-down menu. Program the data logger’s sample rate and number of

samples.

Tap Run on the main toolbar to start the measurement.

Note: Refer to the Rain Collector data sheet for further steps regarding operation of this sensor with MultiLab CE.

Selecting the Drop Counter

Connect the Drop Counter sensor to the Nova5000’s digital sensor input (starting from I/O-1).

Tap Logger on the main tool bar and select Preferences from the drop-down menu. Ensure that Drop Counter is selected in the

Counter Sensor drop-down menu. Tap OK.

Page 76: Multilab Para Windows CE V3

74 Chapter 4 – C

Tap Setup on the main toolbar and unselect the Auto Detect Sensors checkbox. Select the Drop Counter sensor from the drop-down menu. Program the data logger’s sample rate and number of

samples. Tap OK.

Tap Run on the main toolbar to start the measurement.

Selecting Volume (mL)

The Rain Collector and Drop Counter sensors can be used for measuring the volume in mL. To set these features, use the

Preferences dialog:

Tap Logger on the main tool bar.

Choose Preferences from the drop-down menu.

Select Volume from the Counter Sensor drop-down menu and tap OK.

Tap Setup on the main toolbar and unselect the Auto Detect Sensors checkbox. Select the Volume sensor from the drop-down

menu.

Tap Sensor properties next to the Volume sensor input. Select the Calibration tab and enter the drop volume of the rain drop in the

Drop Volume (mL) field.

Tap OK.

Page 77: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 75

Appendix H: Chapter 5

Appendix I: The Analysis Wizard

The Analysis Wizard will guide you through the extensive mathematical and trigonometric analysis functions available in

MultiLab.

1.21 5.1 Using the Analysis Wizard

To apply an analysis function to an open data set:

First ensure the relevant graph is open in the main Graph window

Use the graph cursors to select the data range to which you want to apply the analysis

Tap the Analysis Wizard button on the main toolbar to start the Analysis Wizard.

Page 78: Multilab Para Windows CE V3

76

Select the desired function from the Functions drop-down menu. The function’s formula is displayed underneath the Functions drop-

down menu.

If you've selected a data set, it will be highlighted in the G1 drop-down menu. However, you have the option of selecting a different data

set.

If the analysis function involves two data sets, select the second data set from the G2 drop-down menu.

Where necessary, you have the option of entering a constant value in the A, B or C fields, relating to constants in the function equation.

Edit the name in the Name field (this is optional – the default name includes both the function’s formula and the data set name).

Edit the Unit field (optional).

Tap OK to apply the function.

1.22 5.2 Analysis Functions List

This section includes a brief description of each of MultiLab’s analysis functions.

In the formulas below, G1 and G2 represent selected data sets, and A, B and C are constants that you can enter.

Page 79: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 77

Absolute CBGAy 1 Draws a line of the absolute values of a data set.

Add 21 BGAGy Draws a line of the addition of two data sets.

Arccosine )arccos( 1BGAy

Draws a line of the arccosine values of a data set (in

radians). Arccosine is the angle whose cosine is 1BG .

The argument 1BG must be between –1 and 1.

Arcsine )arcsin( 1BGAy

Draws a line of the arcsine values of a data set in

radians. Arcsine is the angle whose sine is 1BG . The

argument 1BG must be between –1 and 1.

Cosine )cos( 1 CBGAy

Draws a line of the cosine values of a data set. The

argument CBG 1 must be expressed in radians.

Delta Y )0t(GGy 1n

Draws a line of the difference between the Y-coordinate of every point (Gn) and Y-coordinate of the first point. Use this function to move the data set along the Y-axis so that the point will intersect the Y- axis at

the origin.

Derivative

Draws a line of the slopes of every three consecutive points of a data set. For high recording rates and

small t, this line may be very noisy, so smoothing the data set is recommended before applying the

derivative function.

ratesamplingt

t

yyy nn

n

1,

211

Page 80: Multilab Para Windows CE V3

78

Divide 2

1

BG

AGy

Draws a line of the division of two data sets

Envelope (lower)

Lower envelope of G1 with tolerance of A points

Draws a line that connects the minimum values of a data set. The tolerance defines the minimum distance (in sampling points) between two minima, so that the envelope function is able to ignore random noises.

Envelope (upper)

Upper envelope of G1 with tolerance of A points

Draws a line that connects the maximum values of a data set. The tolerance defines the minimum distance (in sampling points) between two maxima, so that the

envelope function will be able to ignore random noises.

Exp. CAey BG 1

Draws a line of e raised to the power of a data set.

Fourier transform

Discrete Fourier transform of G1.

Draws a line of the amplitudes of the harmonics of

Fourier transform vs. frequency.

Frequency The frequency of G1 (minimum of A points in one

cycle).

Draws a line of the frequency of a periodic data set vs. time. The constant A defines the minimum data points

in one cycle.

Integral tGBAy 1

Draws a line in which each point is the discrete integral of all the preceding points in a data set.

Kinetic energy 21 )(

2

1GAy

Page 81: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 79

Draws a line of the kinetic energy of a data set. The

argument must be the velocity of the body, and the

constant must be the mass of the body.

Linear BAGy 1

Draws a line of a linear displacement of a data set. This function is useful when you want to change the

point of origin of a data set.

Ln )ln( 1BGAy

Draws a line of the natural logarithm of a data set. The

argument 1BG must be positive.

Log )(log 110 BGAy

Draws a line of the logarithm of a data set to base 10.

The argument 1BG must be positive.

Multiply 21 BGAGy Draws a line of the multiplication of two data sets

Quadratic CBGAGy 12

1 Draws a line of the quadratic form of a data set.

Reciprocal (1/X) CBG

Ay

1 Draws a line of the reciprocal values of a data set.

Sine )sin( 1 CBGAy

Draws a line of the sine values of a data set. The

argument CBG 1 must be expressed in radians.

Square (X2) 21)(BGAy

Draws a line of the squares of a data set.

Square root CBGAy 1

1G

1C

Page 82: Multilab Para Windows CE V3

80

Draws a line of the square root values of a data set:

The argument must be greater than or equal to zero.

Subtract 21 BGAGy Draws a line the subtraction of two data sets.

Tan )tan( 1 CBGAy

Draws a line of the tangent values of a data set. The

argument CBG 1 must be expressed in radians.

5.3 Timing Analysis Wizard

The Timing Analysis Wizard is an additional analysis tool, which enables you to easily measure and calculate many types of time events, including velocity and acceleration, with one or two photo

gates.

Connect one photogate to input 1 of the Nova5000 or connect two photogates to input 1 and input 2 of the Nova5000 and perform the

desired experiment. Then let the wizard guide you through the analysis of this experiment.

Using the Analysis Timing Wizard, you can measure sequences of time events at either gate, or between the two gates. MultiLab can

then calculate the velocity and acceleration of these events. Special options make it easy to measure velocities in collisions and the time

period of a pendulum or any other oscillating body.

The Analysis Timing Wizard can handle multiple events. For example, if a body crosses a photo gate several times, applying the wizard will

result in a series of measurements that match the number of crossings.

I.1.1. 5.3.1 Working with the Timing Analysis Wizard

Display the data that you wish to analyze in the Graph window.

12GC

Page 83: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 81

On the Tools menu, select Analysis, then Timing Analysis. This launches the Timing Wizard dialog box.

Select the Time, Velocity or Acceleration measurement.

Select the Timing Wizard – Method tab.

Select one of the measuring methods: At one gate, Between gates, or

Collision (two gates).

If required, enter the body’s width, or the distance between the gates in cm in the Width text box.

Tap OK to display the results.

I.1.2. 5.3.2 Measuring Methods

The Analysis Timing Wizard offers you various methods of analyzing the different measurements. Some methods will require you to enter

Page 84: Multilab Para Windows CE V3

82

the dimension of the moving body, or the distance between the two photo gates, to allow for the calculation of velocity and acceleration.

The methods depend on the selected measurement:

Time

At one gate

Measures the time it takes the body to

cross the photo gate (between blocking and unblocking the infrared

beam)

Between gates

Measures the time it takes the body to

move from one photo gate to the second

photo gate (between blocking the first and blocking the second

infrared beams)

Pendulum

Measures the time period of an oscillating body (the time interval between the first and the third blockings of

the beam)

Velocity

Page 85: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 83

At one gate

Measures the time it takes the body to

cross the photo gate (between blocking and unblocking the infrared beam) and returns the

velocity.

Requires you to enter the body’s width.

Between gates

Measures the time it takes the body to

move from one photo gate to the second

photo gate (between blocking the first and blocking the second infrared beams) and returns the average

velocity.

Requires you to enter the distance between

gates.

Collisions

Measures the crossing time

intervals at each gate and returns the corresponding

velocities.

Requires you to enter the bodies’ width (the

width of the two bodies must be

Page 86: Multilab Para Windows CE V3

84

identical).

Acceleration

At one gate

A card with two flags must be attached to

the moving body (see figure to the left). The

Timing wizard measures the crossing time

intervals of the two flags and returns the

acceleration.

Requires you to enter the flags width.

Between gates

Measures the crossing time at the first gate, the time it

takes the body to move from one gate to the second gate

and the crossing time at the second gate

and returns the average acceleration.

Requires you to enter the body’s width.

I.1.3. 5.3.3 Time Schemes and Calculations

Page 87: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 85

5.3.3.1 TIME MEASUREMENTS

Location Required Parameter 

Result 

At One Gate  None  12 ttt Between Gates  None  13 ttt

Pendulum (one gate)  None  15 ttt

Pendulum (one gate):

5.3.3.2 VELOCITY

Location Required Parameter 

Result 

At One Gate w – The body’s

width  12; tttt

wv

Between Gates L – The distance

between gates 13; ttt

t

Lv

 

t5

t3

t1

t2

t4

Page 88: Multilab Para Windows CE V3

86

Pendulum (one gate) 

w – The bodies’ widths

(must be identical) 

782

562

341

121

;

;

tt

wv

tt

wu

tt

wv

tt

wu

At one gate:

Between gates:

Collision (two gates):

t1 t2

t1

t2

t3

t4

Input 1

Input 2

Page 89: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 87

5.3.3.3 ACCELERATION

Location Required Parameter 

Result 

At One Gate w – The flags’

widths 

t

vva

ttttt

tt

wv

tt

wv

12

1234

342

121

2

;

t1

t

t4

Input 1

t5

t7

Input 2

t8

Page 90: Multilab Para Windows CE V3

88

Between Gates w – The body’s

width 

t

vva

ttttt

tt

wv

tt

wv

12

1234

342

121

2

;

At one gate:

Between gates:

t1

t2

t3

t4

Input 1

t

Page 91: Multilab Para Windows CE V3

Chapter 5 - The Analysis Wizard 89

I.1.4. 5.3.4 Tips On Using the Timing Wizard and Timing Analysis Wizard.

Attach a flag to the moving body When measuring the motion of a moving cart, it is convenient to

attach a vertical flag to the cart (see picture below). You can mount a slotted wooden block on the cart and insert the flag onto the slot, or

use masking tape to attach the flag to one side of the cart.

Use a double flag to measure acceleration at one gate. The width of the two flags must be the same.

t1

t2

t3

t4

Input 1

Input 2

t

Page 92: Multilab Para Windows CE V3

90

Use the cursors Use the cursors to select the graph and data range to which you want

to apply the Timing wizard.

Time resolution The time resolution depends on the sampling rate. Use the table

below to select a rate that meets your needs.

Rate

(samples per second) Resolution

10 0.1 s

25 0.0 4s

50 0.0 2s

100 0.0 1s

500 2 ms

1000 1 ms

Use the Trigger For fast events and high sampling rates use the Trigger tool (see

page 60) to initiate the data logging.

Comment [o13]: This seems less like a tip and more like part of the instructions, no?

Comment [D14]: FIX page 35 reference - DAVE

Page 93: Multilab Para Windows CE V3

Chapter 6 – Working in Graph View 91

Appendix J: Chapter 6

Appendix K: Working in Graph View

While working in Graph view, you have a variety of features at your disposal for viewing and formatting the graph data.

1.23 6.1 The Cursor

You can display up to two cursors on the graph simultaneously.

Use the first cursor to display individual data recording values, to select a curve or to reveal a hidden Y-axis.

Use two cursors to display the difference between two coordinate values or to select a range of data points.

To reveal the icons in graph view, press the Gears Icon

To display the first cursor:

There are 2 methods to display a cursor on the graph.

Page 94: Multilab Para Windows CE V3

92 Chapter 6 – Working in Graph View

Select an individual data point on the graph or tap 1st Cursor

on the graph toolbar. You can drag the cursor with the stylus onto any other point on the plot, or onto a different plot.

You can double-tap the graph in the location you want the cursor.

To display the second cursor:

Tap 2nd Cursor on the graph toolbar or double-tap the graph again.

MultiLab will now display the difference between the two coordinate values.

To remove the cursors:

Tap 1st Cursor a second time.

To remove only the 2nd cursor:

Tap 2nd Cursor a second time.

Moving the cursor:

For finer cursor movements use the forward and backward buttons on the graph toolbar.

It is also possible to tap anywhere on the graph’s plot to move the data point.

When there is only 1 data point selected; the coordinate values of the selected point will appear at the bottom of the graph window.

Page 95: Multilab Para Windows CE V3

Chapter 6 – Working in Graph View 93

When there are 2 data points selected; the differences between the 2 points will appear at the bottom of the graph window.

dX refers to the value between the X axis of the 2 points. dY refers to the value between the Y axis of the 2 points.

Here we can see that that dX value is 3800 ms representing the number of seconds between points 1 and 2. Also the dY value is 0.01

V representing the amount of Voltage between points 1 and 2.

.

Page 96: Multilab Para Windows CE V3

94 Chapter 6 – Working in Graph View

1.24 6.2 Autoscale/Graph Properties

K.1.1. 6.2.1 Autoscale

Tap Autoscale on the graph toolbar to view the full data display. This is useful when you have customized the data scale, and then

wish to return to the full data display.

K.1.2. 6.2.2 Manual Scaling

Tap Format graph on the MultiLab graph toolbar.

Select the axis you want to rescale from the Select plot drop-down

menu.

Uncheck the Autoscale checkbox and enter the desired values in the Min and Max text boxes.

To restore the default scale values, tap Restore defaults.

Page 97: Multilab Para Windows CE V3

Chapter 6 – Working in Graph View 95

Tap OK.

To restore the default scale value while in Graph view, tap Autoscale

.

1.25 6.2.3 Adjusting the Y-Axis and X-Axis

You can press and hold the pointer on any point on the X or Y-axis.

Now drag the pointer up or down to change the scale Y-axis.

Drag the pointer left or right to change the scale of the X-axis.

Tap Autoscale to restore the default scales for all axes.

You can also

K.1.3. 6.2.4 Formatting the Graph Colors

To change the data line’s color:

Page 98: Multilab Para Windows CE V3

96 Chapter 6 – Working in Graph View

Tap Format graph on the graph toolbar.

Select the plot you want to format from the Select plot drop-down menu.

Tap the Color box to open the palette and select the desired color.

Tap OK.

To restore the default plot color, return to the Lines tab, select the relevant plot and tap Restore defaults.

1.26 6.3 Zooming

K.1.4. 6.3.1 Zooming into a Specific Area

To reveal the icons in graph view, press the Gears Icon

Tap Zoom on the graph toolbar and drag the diagonally across the graph to select the area you want to magnify. Remove the stylus to

zoom in to the selected area.

Tap Zoom a second time to disable the Zoom tool.

To revert to the default graph size, tap Autoscale in the graph toolbar.

Note: You can also use the Zoom tool by going to the Logger menu and choose Zoom in or Zoom out.

1.27 6.4 Annotations

You can add annotations (notes) to the graph. An annotation is always connected to a certain data point.

To view annotations:

From the Tools menu, select Show annotations (this option is usually selected by default).

To hide all annotations:

Page 99: Multilab Para Windows CE V3

Chapter 6 – Working in Graph View 97

From the Tools menu, unselect Show annotations to hide all current annotations.

To add an annotation:

Place a cursor on the point to which you want to assign an annotation.

To reveal the icons in graph view, tap the Gears Icon

Tap Add new annotation on the graph toolbar to display the new annotation caption text box.

Enter the text and tap OK.

To move an annotation:

Tap Move annotation on the graph toolbar and drag the annotation text box to the desired location.

To delete an annotation:

Use the Move annotation and then select the annotation you wish to delete. Next, go to the Tools menu and choose Delete

annotation.

1.28 6.5 Adding an existing Graph or saved data set to a current

Project

MultiLab displays new data in the graph window every time you start a new recording. If you want to save a graph that you created to your project, or to update a saved graph with changes you made, use the

Add File tool:

Go to the File menu and choose Add File and then select the file you want to open from the explorer window.

Page 100: Multilab Para Windows CE V3

98 Chapter 6 – Working in Graph View

Note: Add File will only work if you are adding a data set with the same sampling rate

1.29 6.6 Smoothing

The Smoothing tool is very useful in reducing random noises, especially if you want to apply any analysis functions to the data. The smoothing process will create a new plot using every data point from

that data you have selected, and then take the average of its neighboring points.

Use the cursor to select the plot that you want to smooth.

Tap More smoothing on the graph toolbar.

You can repeat the procedure to further smooth the data.

Tap Less smoothing to reduce the amount of smoothing.

1.30 6.7 Prediction Tool

The Prediction tool enables you to draw predictions directly on the graph, prior to displaying the real data.

Make sure you are in Graph view and tap Run on the main toolbar to start recording data

Tap Pause/Continue on the graph toolbar to freeze the graph. The data recording continues, but the plotting of data on the graph

is paused.

Tap Add prediction on the graph toolbar to enable the Add prediction tool. Tap the points on the graph where you predict the

plot will continue. Each point you add to the graph will be connected by a straight line. By tapping Add prediction again, you

can add a second set of predictions, and so on.

Note: Multilab will only allow up to 7 predictions.

Page 101: Multilab Para Windows CE V3

Chapter 6 – Working in Graph View 99

Tap Pause/Continue a second time to resume live data display and to compare your predictions with the real data.

Tap Erase prediction on the graph toolbar to enable the Erase prediction tool. You may erase a set of predictions by tapping any

point within the prediction plot.

Note: You MUST be in prediction mode to erase a prediction

Page 102: Multilab Para Windows CE V3

100

Appendix L: Chapter 7

Appendix M: Additional Tool Menu Options

1.31 7.1 Unit Settings

MultiLab enables you to change the unit prefix and number format of the graph currently open in MultiLab. This function will not change the

actual hard data in any way.

From the Tools menu, select Unit Settings

In the Unit Settings dialog box, choose the plot you want to format from the Select plot drop-down menu.

Select the prefix option you want to use.

Select the desired number of decimal places.

Page 103: Multilab Para Windows CE V3

Chapter 8 – The Workbook 101

To display numbers in scientific format, check the Scientific checkbox.

Tap OK.

1.32 7.2 Graph Title

MultiLab enables you to edit the title of the graph currently open in the Graph view:

From the Tools menu, select Graph Title.

In the Graph Title dialog box, enter the new title to be assigned to the graph currently opened.

Tap OK.

1.33 7.3 Crop

The Crop tool enables you to trim the edges of a data set. Use it to remove unwanted data or to apply manual curve fitting to a desired

range of data points.

The time scale of the trimmed data is shifted so that it will start at 0t .

After applying the Crop tool, the trimmed data set replaces the original set on the graph display and a new icon is added to the Data Map

under cropped data.

To trim all data up to a point Position a cursor on the data point

From the Tools menu, select Crop.

To trim all data outside a selected range Use the cursors to select the range you want to keep.

From the Tools menu, select Crop.

Page 104: Multilab Para Windows CE V3

102

1.34 7.4 Manual X-Axis

Using the Manual X Axis feature, you can now manually input x-axis values during an experiment not just after the experiment as was possible using the Add Manual Column feature (see page 35).

The Manual X Axis feature can be used when performing the Boyle’s Law experiment, for example, for measuring Pressure (Y-Axis) vs. Volume (X-Axis). Manually input the Volume data by adding a new

column in the data table and entering the Volume each time you tap Run to record the pressure.

To use this feature:

Go to the Tools menu and choose Manual X-Axis.

The Manual X Axis item should be selected.

The Insert manual column dialog will pop up. Enter the Column title and Column unit. Tap OK.

Page 105: Multilab Para Windows CE V3

Chapter 8 – The Workbook 103

Attach your sensors to the Nova5000 and enter the Setup window to configure your sensors. You will notice that the sensor rate is already

in Manual mode. When selecting Manual X Axis, the rate is automatically set to Manual. Tap OK when finished.

Tap Run to record the first sample. As soon as you tap Run, the Manual X value dialog will pop up. Enter the value recorded and tap OK. If you don’t wish to record the value at this time, tap Skip Value.

Note: You can edit the X value at any time by selecting the relevant cell and editing the value.

The manual X axis values are displayed in Table view and Graph

view.

To end the Manual X Axis mode, ensure MultiLab CE is in Stop mode. Go to the Logger menu and choose Stop, then go to the

Tools menu and choose Manual X Axis.

Page 106: Multilab Para Windows CE V3

104

Appendix N: Chapter 8

Appendix O: The Workbook

1.35 8.1 Working with the Workbook Feature

The Workbook is an online library of experiment manuals called worksheets that appear in Web-page format. Each worksheet includes an experiment template that automatically configures MultiLab at the push of a button. To begin recording, all you need to do is tap Run.

Every time you run an experiment from a worksheet, MultiLab opens a new project file with the same predefined setup.

You can use Fourier-produced worksheets, or create your own. You can also modify Fourier’s existing worksheets to your own

specifications.

1.36 8.2 Opening a Worksheet

On the Tools menu, select Workbook, then select Open worksheet from the Workbook menu.

Navigate to the folder in which the worksheet is stored.

Tap the file name to open the worksheet.

Page 107: Multilab Para Windows CE V3

Chapter 8 – The Workbook 105

You may need to hide other views to get a full view of the worksheet

by taping one or more of these icons.

View / Hide Data Map

Launch video

View / Hide Graph view

View / Hide Table view

View / Hide Meter view

Launch Workbook

Follow the on-screen instructions and use the scroll bar, hyperlinks and the Back and Forward buttons to navigate within the

document.

Tap Launch to configure MultiLab. The Launch feature may not be available on all workbooks.

Close the Worksheet to return to MultiLab.

Tap Run on the main toolbar to begin recording data.

Page 108: Multilab Para Windows CE V3

106

1.37 8.3 Create Your Own Worksheet

Creating a worksheet consists of two steps. First, create an HTML document using your HTML editor (for example, Nova5000’s

TextMaker).This document should include both the lab manual and the experiment instructions.

The second step is to use the MultiLab software to create a configuration file, which will store the specific settings you wish to

define for the experiment., such as the sensor setup, sampling rate, and graph format.

Creating an HTML document

Open TextMaker on the Nova5000.

In the File menu, choose New.

Create your worksheet, including all necessary content for experiment instruction.

Go to the File menu and choose Save as and select HTML from the File type drop-down menu.

Navigate to My Nova\Program Files\Fourier Systems\MultiLab CE\Experiment documents, enter a name for your file, and tap

Save.

Create a Web page from an existing Word document

On Your PC computer using Microsoft Word:

1. In the File menu, tap Save as.

2. Choose Save as, Web Page (*.html *.html)

3. Copy the file from your computer to the following location on the Nova 5000.

4. My Nova\Program Files\Fourier Systems\MultiLab CE\Experiment documents

Page 109: Multilab Para Windows CE V3

Chapter 8 – The Workbook 107

Note: You can use a USB disk drive to copy the data from the PC computer to the Nova 5000 device..

On the Nova 5000 using TextMaker:

1. In the File menu, tap Save as.

2. Choose Save as, HTML 4.0

3. Save the file in the following location.

4. My Nova\Program Files\Fourier Systems\MultiLab CE\Experiment documents

Create a configuration file Go to the Tools menu and choose Workbook, then Create

workbook.

Select the file you want to open from the Open dialog box and tap OK.

Tap Enter workbook settings to open the Setup Wizard.

Use the setup wizard to preset MultiLab just as with the normal setup

process.

Set the Sensors you wish to have in the workbook along with the sampling rate and number of samples you need.

On the Sensors tab, tap X-axis display properties to enter the X-axis formatting.

Page 110: Multilab Para Windows CE V3

108

When you’ve finished entering the settings, tap OK to update the worksheet.

Page 111: Multilab Para Windows CE V3

Appendix A – Sensors Supported by the Nova5000 109

Appendix P: Chapter 9

Appendix Q: Sensors Supported by the

Nova5000

This appendix provides an up-to-date list of Nova5000-compatible Fourier, Vernier and Data Harvest probeware.

Together with the MultiLab software, these sensors are used for data collection and data analysis.

Please contact Fourier Systems for more information regarding these sensors.

9.1 Compatible Fourier Systems Sensors

P/N Type Range Physics Biology Chemistry

DT138 Acceleration 5g �

AC020A Ammonium Electrode

0.1 to 18,000 ppm � �

AC012A Anemometer

Wind speed: 4 to 280 km/h

Wind direction: 0 to 360

AC019A Calcium

Electrode 0.02 to 40,000

ppm � �

DT040 Carbon Dioxide 350 to 5,000 � �

Comment [o15]: Is this list up-to-date? I see several sensors not mentioned here (Humidity 2%, Wind Speed/Direction, Heat Loss, etc.)

Comment [D16]: Where is the latest list…. The web page must not be updated!

Page 112: Multilab Para Windows CE V3

110 Appendix A – Sensors Supported by the Nova5000

P/N Type Range Physics Biology Chemistry

(CO2) ppm

DT261A Electrostatic

Charge 0.25 uC ;

0.025 uC �

AC018A Chloride Electrode

1.8 to 35,500 ppm � �

DT185A Colorimeter

Three colors; blue, green,

red � �

DT035A Conductivity 0 to 20 mS/cm � � �

DT110 Control Switch

- Closed Normally closed � � �

DT111 Control Switch

- Open Normally open � � �

DT005 Current 2.5 A 2.5 A � �

DT006 Current 250

mA 250 mA � �

DT020-1 Distance 0.2 to 10 m � �

DT293 Drop Counter 0 to 4096 � �

DT189A EKG 0 to 5 V �

DT272 Force 10 N; 50 N �

DT254 Flow Rate 0 – 4.0 m/s � � �

DT116 Geiger Muller 0 to 4096 Bq � � �

DT298A Heart Rate (Exercise) 0 to 200 bpm �

DT155A Heart Rate

(Pulse) 0 to 200 bpm �

DT014 Humidity 0 - 100%, 5%

accuracy � � �

DT009-1 Light 0 to 300 lux � � �

DT009-4 Light Triple-

range 0-600, 0-6000, 0-150,000 lux � � �

Page 113: Multilab Para Windows CE V3

Appendix A – Sensors Supported by the Nova5000 111

P/N Type Range Physics Biology Chemistry

DT156 Magnetic Field 0 to 10 mT, 0

to 0.2 mT �

DT008 Microphone

2.5 V

Frequency range:

35 Hz to 10,000 Hz �

AC017A Nitrate

Electrode 0.1 to 14,000

ppm � �

DT222A Oxygen Adapter

0 to 12.5 mg/L

0 to 25% � �

DT118 Oxygen

Electrode

0 to 12.5 mg/L

0 to 25% � �

DT017 pH Adapter 0 to 14 pH � �

DT018 pH Electrode 0 to 14 pH � �

DT137 Photo Gate 0 to 5 V (Digital) � �

AC008A Potassium Electrode

90 to 39,000 ppm � �

DT015 Pressure 150 to 1150

mbar � � �

DT015-1 Pressure 0 to 700 Kpa � � �

DT015-2 Pressure 0 to 10 Kpa � � �

AC013 Rain Collector 0 to 819 mm � �

DT148A Rotary Motion 128° � �

DT122A Smart Pulley 0 - 99 m/s �

DT171 Soil Moisture 0 to 200 cbar � �

DT320 Sound Level 45 to 110 dB �

DT037A Spirometer (Breathing)

315 L/min

5.25 L/sec

Page 114: Multilab Para Windows CE V3

112 Appendix A – Sensors Supported by the Nova5000

P/N Type Range Physics Biology Chemistry

DT025

Temperature (Thermocouple

) 0 to 1200 °C

� � �

DT027 Temperature

(PT-100) -200 °C to 400

°C � � �

DT029 Temperature -25 °C to 110

°C � � �

DT095A Turbidity 0 to 200 NTU � �

DT001 Voltage 25V 25 V � �

DT002 Voltage 2.5 V 2.5 V � �

DT003 Voltage 5 V 0 - 5 V � �

DT004A Voltage 50 mV 50 mV � �

9.2 Compatible Vernier Sensors Using an adapter available from Fourier Systems, the following

Vernier sensors may be connected to the Fourier sensor interface. To use these sensors, they must first be defined in MultiLab

Note: Only Vernier Analog linear sensors are supported. Digital sensors are not compatible with Fourier data loggers.

Sensor Units Minimum Output Voltage

Minimum Real Value

Maximum Output Voltage

Maximum

Real Value

Accelerometer

3 axis m/s2 0 -51.6

5

63.0

Page 115: Multilab Para Windows CE V3

Appendix A – Sensors Supported by the Nova5000 113

Sensor Units Minimum Output Voltage

Minimum Real Value

Maximum Output Voltage

Maximum

Real Value

Accelerometer

Low g m/s2

0

-51.8

5

62.9

Barometer Hg Hg 0 24.2 5 35.7

Barometer atm atm 0 0.8 5 1.2

Barometer mm Hg mm Hg 0 614.8 5 907.4

Barometer mbar mbar 0 819.5 5 1209.5

CO2 ppm 0 0.0 5 10000.0

Colorimeter % 0 0.0 5 142.9

Current 6A A 0 0.625 5 -0.625

Differential Voltage V 0 6.3 5 -6.3

Dissolved Oxygen mg/L 0 -0.3 5 16.0

Dual Range Force 10N N

0 12.3

5 -12.3

Dual Range Force 50N N

0 61.3

5 -61.3

Flow Rate m/s 0 0.0 5 5.0

Force 10N N 0 12.25 5 -12.25

Force 50N N 0 61.25 5 -61.25

Gas Pressure kPa kPa 0 0.0 5 232.4

Gas Pressure atm atm 0 0.0 5 2.2935

Gas Pressure mm Hg mm Hg

0 0.0

5 1743.2

Light Sensor

0 - 600 Lux lx

0

0.0

5

770.0

Light Sensor

0 – 6000 Lux lx

0

0.0

5

8460.0

Light Sensor

0 – 150,000 Lux lx

0

0.0

5

192120.0

Page 116: Multilab Para Windows CE V3

114 Appendix A – Sensors Supported by the Nova5000

Sensor Units Minimum Output Voltage

Minimum Real Value

Maximum Output Voltage

Maximum

Real Value

Magnetic Field 6.4 mT in gauss

Low gauss

0 -80.6

5 80.6

Magnetic Field

0.3 mT in gauss Hi gauss

0

-3.2

5

4.8

Magnetic Field

0.3 mT in millitesla Low

millitesla

0

-8.1

5

8.1

Magnetic Field

6.4 mT in millitesla Hi

millitesla

0

-0.3

5

0.5

Microphone V 0 0 5 5

ORP mV 0 -559.8 5 1774.6

Oxygen gas % 0 0 5 32.8125

pH pH 0.942 10.01 2.45 4.01

Relative Humidity % 0 -23.8 5 140.7

Salinity ppt 0 0.0 5 81.5

Thermocouple C 0 -188.9 5 1393.0

UVA mW 0 0.0 5 19700.0

UVB mW 0 0.0 5 19700.0

Voltage 6 V V 0 6.25 5 -6.25

X-Long Temp C 0 -53.1 5 238.6

X-Long Temp F 0 -63.5 5 461.5

Direct Connect C 0 -17.8 5 260.0

9.3 Supported Data Harvest Probeware

Data Harvest sensors are able to connect directly to the Fourier sensor interface, without the need for any custom adapters.

Page 117: Multilab Para Windows CE V3

Appendix A – Sensors Supported by the Nova5000 115

To use the following Data Harvest sensors, they must first be defined in MultiLab.

Note: Only Data Harvest Analog linear sensors are supported. Digital sensors are not compatible with Fourier

data loggers.

Sensor Slope Intercept Units

0 Volts

(Min Output

Voltage)

5 Volts

(Max Output Voltage

)

MultiLab Name

Current ± 100 mA

55.807 -139 m/s2 -139

140.035

DH-100mA

Current ± 1 A 0.623

3 -1.559 m/s2 -1.559 1.5575

5 DH-1A

Current ± 10 A 5.547 -13.73 Hg -13.73 14.005 DH-10A

Photogate 0-5 V 1 0 atm 0 5

DH-Photogat

e

pH 0-14 pH 4.340

5 -3.982 mm Hg -3.982

17.7205 DH-pH

Relative Humidity 0-100

% -3.838 13.72 mbar 13.72 -5.47 DH-RH

Light 100 klux

0-100,000 lux 177.5

2 -87.515 ppm -87.515 800.06 DH-Light-

100K

Sound level

40 to 110dBA 118.5

6 -276.66 % -276.66 316.14 DH-

Sound

Page 118: Multilab Para Windows CE V3

116 Appendix B – Adding a Code Resistor to a Custom Sensor

Appendix R: Chapter 10

Appendix S: Adding a Code Resistor to a Custom

Sensor

In order for the Nova5000 to automatically identify a sensor, it must have a code resistor. The figure below shows the Nova5000 input

socket configuration.

You must connect the resistor between the Auto recognition resistance input and the GND input.

Mini DIN female - Nova5000 panel view

Page 119: Multilab Para Windows CE V3

Appendix B – Adding a Code Resistor to a Custom Sensor 117

When you define a custom sensor in MultiLab you must choose one sensor from the list provided. Note its position in the list and look in

the table below to find the resistance.

Here is the list of the uuto recognition resistors and their values:

Position in Define sensor

dialog box drop-down

menu

Resistance (k)

Max Resistance

(k)

Min Resistance (k)

1 372 384.1 360.9

2 410 424.8 396.5

3 455 472.9 437.8

4 508 530.9 486.6

5 573 601.9 544.8

6 653 691.1 615.6

7 754 806.5 703.4

8 888 961.9 814.6

9 1071 1183.0 959.2

10 1337 1524.4 1151.2

Page 120: Multilab Para Windows CE V3

Appendix T: Chapter 11

Appendix U: Nova5000 Data Logger Specifications

Input Auto ID mode:

Up to 4 simultaneous analog inputs

Or

4 digital inputs with automatic sensor identification

8-input (manual) mode: Up to 8 simultaneous analog inputs

Or

Up to 4 simultaneous digital inputs and 4 analog inputs

Outputs 4 Digital Control Outputs

Sampling Capacity: Up to 100,000 Samples

Analog sampling rate: From 1 sample/hour to 20,000 samples/sec Digital sampling rate: >200 kHz

Resolution 12-bit (4096 levels)

Page 121: Multilab Para Windows CE V3

Timer Module Time velocity and acceleration

Range: 0.0014s to 6.5535s Resolution: 0.1mS

Card width: 0 to 59mm Distance between gates: 0 to 99cm

Features

Standalone operation – no PC needed Automatic or manual sensor identification

Triggering method: Programmable or manual Automatic calibration of sensor offsets

Built-in clock and calendar that keeps track of time and date for each data recording

Event recording Up to 10 user defined sensors