Document No. 031972 Revision 01 October 2003 LC/MS Nitrogen Generator MSQ10LA

MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

Embed Size (px)

Citation preview

Page 1: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

Directions for Use

Document No. 031972 Revision 01

October 2003

LC/MS Nitrogen Generator

MSQ10LA

Anuta
New Stamp
Page 2: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 1

©2003 by Dionex Corporation All rights reserved worldwide. Printed in the United States of America.

This publication is protected by federal copyright law. No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or transmitted into any human or computer language, in any form or by any means, electronic, mechanical, magnetic, manual, or otherwise, or disclosed to third parties without the express written permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088-3603 U.S.A. The MSQ10LA Nitrogen Generator is manufactured for Dionex by:

Peak Scientific Instruments Fountain Crescent Inchinnan Business Park Inchinnan, Renfrewshire PA4 9RE Scotland, UK

DISCLAIMER OF WARRANTY AND LIMITED WARRANTY THIS PUBLICATION IS PROVIDED “AS IS” WITHOUT WARRANTY OF ANY KIND. DIONEX CORPORATION DOES NOT WARRANT, GUARANTEE, OR MAKE ANY EXPRESS OR IMPLIED REPRESENTATIONS REGARDING THE USE, OR THE RESULTS OF THE USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY, RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX CORPORATION RESERVES THE RIGHT TO REVISE THIS PUBLICATION AND TO MAKE CHANGES FROM TIME TO TIME IN THE CONTENT HEREINOF WITHOUT OBLIGATION OF DIONEX CORPORATION TO NOTIFY ANY PERSON OR ORGANIZATION OF SUCH REVISION OR CHANGES.

TRADEMARKS

Molex is a registered trademark of Molex, Inc. Swagelok is a registered trademark of Swagelok Co.

PRINTING HISTORY Revision 01, October 2003

The products of Dionex Corporation are produced under ISO 9001 accredited quality management systems.

Published by Technical Publications, Dionex Corporation, Sunnyvale, CA 94086.

Page 3: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 2

CONTENTS 1 INTRODUCTION..................................................................................................................................... 3

2 UNPACKING AND INSTALLATION................................................................................................... 3

3 ELECTRICAL CONNECTION.............................................................................................................. 3

4 AIR CONNECTION................................................................................................................................. 4

5 PRINCIPLE OF OPERATION............................................................................................................... 4

6 COMMISSIONING.................................................................................................................................. 5

7 ROUTINE INSPECTION ........................................................................................................................ 6

8 TROUBLESHOOTING ........................................................................................................................... 8

9 TECHNICAL SPECIFICATIONS ....................................................................................................... 10

10 PARTS LIST ........................................................................................................................................... 11

MAINTENANCE RECORD LOG ................................................................................................................. 14

IMPORTANT NOTICE TO USERS These instructions must be read thoroughly and understood before installation and operation of your LC/MS Nitrogen Generator. Please follow the appropriate safety standards for handling gases as determined by the laws and regulations of your country. Please save the product packaging for storage or future shipment of the generator.

Page 4: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 3

1 Introduction The MSQ10LA is a self-contained nitrogen generator designed specifically for use with Liquid Chromatography/Mass Spectrometry (LC/MS) analytical instruments as a source of carrier gas. The generator has been designed to operate totally independent of an external air supply and only requires an electrical supply to deliver high volume, high pressure, clean, dry, pure nitrogen. 2 Unpacking and Installation Although every precaution has been taken for safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage. ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND DIONEX. After unpacking and a visual inspection, the unit should be placed in a ventilated area away from direct sunlight. Care should be taken not to obstruct the ventilation holes on the sides of the unit or the fan outlet at the rear. The generator should be placed on a steady and level base. It has been designed to fit under most workbenches. Alternatively, the unit may be placed on the workbench adjacent to the machine or in any location convenient to the user. A guide for installation is included at the end of the document. Performance of the generator (as with any sophisticated equipment) is affected by ambient temperature. Continuous operation in ambient temperatures exceeding 25°C will lead to a reduction in capacity and prolonged operation in temperatures exceeding 35°C will shorten the life of the unit. Note should also be taken of the proximity of air conditioning outlets. These can sometimes give rise to “pockets” of air with high relative humidity. Operation of the generator within such a pocket could adversely affect its performance.

3 Electrical Connection Important Electrical Notice This unit is classified as SAFETY CLASS 1 equipment. THIS UNIT MUST BE EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency. EARTH/GROUND (E): Green & Yellow LIVE (L): Brown Neutral (N): Blue Fuse The generator protection fuse is in the pull-out drawer of the mains inlet IEC Euroconnector located on the bottom right-hand side of the cabinet, adjacent to the off/on switch. The fuse is rated at 230V ±10% (50/60Hz ac). The drawer also holds a spare. Connect the generator to a single-phase supply using the power cord provided, or the power cord supplied with the equipment from Dionex for the country of use.

Page 5: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 4

4 Air Connection The MSQ10LA generator has its own built-in air compressor and therefore does not require an external compressed air source. The generator has one bulkhead connection to the rear of the unit. This is supplied with a 6mm push-in fitting. This is the nitrogen outlet fitting and should be coupled to the application. There is no drain on this machine. Any moisture liberated by the filter-separator is discharged through the high capacity exhaust system, where the sudden reduction in pressure causes instant evaporation. The water vapor is safely removed from the unit by the ventilation system. 5 Principle of Operation The MSQ10LA Nitrogen Generator utilizes a pressure swing adsorption (PSA) method to extract pure nitrogen from air. This is where unwanted gases can be selectively adsorbed from compressed air into a porous carbon molecular sieve material (CMS). Dionex instruments utilize a unique single-column system where the column is alternately pressurized and vented under a finely tuned timing cycle. The rates of pressurization and venting are accurately set, which guarantees higher purity than can be achieved with a similarly sized traditional 2-column system.

10

14

Nitrogen GeneratorStandard System

Parts List1. Compressor

14. Bulkhead Connection15. Internal Gauge Rail

9. Nitrogen Storage Tank10. Pressure Regulator11. Internal Pressure Gauge

13. Front Panel Gauge

8.

7.

15.

12

116.

9.

13

2. 3.

7. CMS Column8. Non-return Valve

2. Cooling Coil

4. Filter / Separator5. Exhaust Valve6. Exhaust Silencer

3. Cooling Fan (if fitted)

16. 16.

16. Compressor Cooling Fan (if fitted)

1.

5. 4.

12. Flow Regulator

Air is drawn into the system by the compressor (1) and passed via the cooling coil (3) and the filter/separator (4) into the CMS column (7). Oxygen molecules in the air are trapped by the sieve. However, the molecules of nitrogen pass straight through and are collected in the nitrogen storage tank (9). After a time interval, the compressor is stopped and the exhaust valve (5) opens, allowing the sieve column to vent to atmosphere. The trapped oxygen is liberated and escapes to atmosphere via the exhaust valve (5) and the silencer (6). The generated nitrogen in the storage tank is then regulated to the correct pressure and flow rate. After another time interval, the exhaust valve shuts and the compressor starts. This cycle runs continuously.

Page 6: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 5

6 Commissioning With the generator installed as described earlier, connect power to the unit and turn it on. Disconnect the nitrogen outlet connection to allow the generator to vent to atmosphere until the unit is stabilized. At Switch On, the exhaust valve will open and the generator will commence its Venting Cycle. This is to allow venting of any residual pressure in order that the compressor does not start against pressure. The Vent Cycle may last up to 90 seconds. At the end of the vent cycle, the compressor should be heard to run and the normal operating cycle will begin. Pressure should begin to build, and the gauge on the front panel should indicate 75 psig after approximately 10 minutes. The generator has been pre-set at the factory to give 10 liters ATP per minute at 85 psig. Once the pressure in the nitrogen receiver exceeds that setting, the generator will stabilize and produce pure nitrogen. Maximum purity will be achieved after around 45 minutes. At this time, the generator can be re-connected to the application. The generator is designed to deliver up to 10 liters of nitrogen at 85 psig with an oxygen content of 2% or less. Should the demand for nitrogen be less than 10 liters, or should the demand stop, the generator will continue to operate without any problems. The generator is protected from over-pressure and its normal operating cycle ensures frequent venting. Timer Setting The cycle timer has been set at the factory and should not require further adjustment. Adjusting the timer will affect the volume and purity of the delivered nitrogen; therefore, the timer should NEVER be adjusted without reference to the factory. The normal settings for this generator are 175 seconds ON, 85 seconds OFF. These settings may vary slightly between machines.

Page 7: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 6

7 Routine Inspection Due to the simplicity of the design and the small number of moving parts, the MSQ10LA Nitrogen Generator will have a long and trouble-free life. However, as with all scientific and technical equipment, the generator should be regularly inspected by a competent person and the following points noted. Every 12 Months Note: All routine maintenance filter elements are contained in Dionex P/N 061492. Compressor Inlet Filter Element

1. Switch the generator OFF and disconnect it from the electrical supply. Remove the fuse from the Euro-socket to prevent the generator being inadvertently switched on by another person.

2. Remove the 5-off X-head screws securing the lower front panel. Remove the panel to expose

the compressor.

3. The filter is located on top of the compressor and faces toward the rear, as shown below.

The cover is removed by twisting it one-quarter turn and pulling clear. The element simply lifts out. Part Number 02-4187 Reassembly is the reverse procedure.

Inlet Filter

Page 8: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 7

Coalescing Filter and Exhaust Silencer

1. Switch the generator OFF and disconnect it from the electrical supply. Remove the fuse from the Euro-socket to prevent the generator being inadvertently switched on by another person.

2. Pull the generator clear of the bench (if located under) to a place where access is possible to

all four sides.

3. Remove the 8-off X-Head screws securing the lid. Remove the lid.

4. Remove the coalescing filter bowl shown below by holding down the black slide and twisting the bowl one-quarter turn. The element simply unscrews.

1. Fit new element and clean, and re-fit bowl. 2. Unscrew and remove the green plastic cover of the exhaust silencer. Remove the 10mm nut

and prize off the element.

1. Fit new element and re-fit plastic cover.

2. Replace all covers and switch on. Allow 4 hours to recover purity before reconnecting to application.

Page 9: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 8

8 Troubleshooting The PSA method for nitrogen production is the most reliable of the popular methods. Many thousands of PSA systems are currently operating worldwide and have given many years of trouble-free operation. Provided the CMS material is protected from oils and oxygen enrichment, the purity of the nitrogen produced will remain consistent over the lifetime of the machine. Problems associated with the generator will be confined to the compressed air or control systems and will probably show up through loss of capacity. Reference should be made to the following Fault Finding Chart.

InsuficientN2 Flow

CompressorCycling On/Off

Sieve Vents onExhuast Cycle

Check TimerOperating

Check TimerOperation Fit New Timer

Yes

No

Ok

Yes

No

Ok

Not Running

Check Power Repair WiringFault

Ok

No Power

Check Power outto Solenoid Valve

No

Yes

Check ExhaustValve

Replace FaultyValve

CheckCompressor Inlet

Filter

clean

DirtyFit New Filter

Check Pressure@ Filter Bowl

Check Systemfor Leaks Repair Leaks

Check GaugeRail Pressure Adjust Pressure

< 100 psi

>110 psi No

Yes

ServiceCompressor

>100 psi

< 100 psi

Check TimerSettings

Page 10: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 9

Compressor Requirements The intake filter requires periodic replacement. Initial inspection should be after 500 hours operation. Most compressor-related problems can be prevented by keeping the filter clean. A dirty inlet filter will decrease compressor performance and may shorten its life. The compressor head surfaces and motor casing become very HOT during operation. Do not touch these parts until the unit has been switched off and allowed to cool. Switch the generator OFF and disconnect it from the electrical supply. Remove the fuse from the Euro-socket to prevent the generator being inadvertently switched on by another person. This is a non-lubricated compressor and should never be oiled. Oiling this compressor will cause damage. Compressor Replacement Remove the generator to a clean dry place with sufficient room to work.

1. Remove the lower front panel. 2. Disconnect the braided hose from cone fitting where it is secured to the right-hand side panel.

Take care not to disturb the fit between the cone fitting and the bracket assembly.

3. Disconnect the Molex® connector at the front of the compressor.

4. Tip the generator onto its back and remove the 4-off nuts securing the compressor to the base. Support the weight of the compressor with one hand while removing the last nut.

5. Lift the compressor clear of the generator.

6. Remove and clean all the fittings from the old compressor and reassemble onto the new

machine using PTEF tape and sealing compound. (Swagelok™ Teflon Free Pt No MS-TFS-50) is recommended.)

Reassembly is the reverse procedure.

Page 11: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 10

9 Technical Specifications General Details

Minimum Operating Ambient Temperature 5 oC (41oF)

Maximum Operating Ambient Temperature 30 oC (86oF)

Inlet Conditions (Free of oil and bulk moisture)

Minimum Air Inlet Pressure N/A

Maximum Air Inlet Pressure N/A

Minimum Air Inlet Flow Rate N/A

Outlet Gas

Maximum Pressure Drop (Outlet-Inlet) δP N/A

Maximum Gas Outlet Pressure 85 psig

Maximum Gas Outlet Flow for Specified Purity 10 Liters/min (ATP)

Pressure Dew point -75oC (-103oF) (1.4ppm @ ATP)

Start up time for Purity 45 minutes

Particles 0.01um

O2 level (maximum) 2.0%

Electrical Requirements

@230V ac ± 10% (50/60Hz ac) 4.2 Amps

Fuse (220V 50/60Hz ac) 10 Amps

Electrical Connection IEC Euroconnector

Plug Type NEMA Code N6/15

General

Dimensions H x W x D cm (inches)

63 x 43 x 41

(24.6 x 17 x 16)

Weight Kg (lbs)

47.5 (104.5)

Shipping Weight Kg (lbs)

69 (152)

Page 12: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 11

10 Parts List Recommended Spares

MSQ10LA Service Spares

The following parts are in Dionex P/N 061492

Item Description Interval Part No. No. Off

1 Compressor Inlet Filter Element 12 Months 02-4187 1

2 Coalescing Filter Element 12 Months 02-4509 1

3 Exhaust Element 12 Months 02-4336 1

MSQ10LA Maintenance Spares

Item Description Part No. No. Off

1 Braided Hose 02-1100 1

2 Internal Gauge 02-1021 1

3 Exhaust Valve 02-4289 1

4 Non-Return Valve 02-1135 1

5 Timer 04-1019 1

6 Contactor 04-4439 1

7 Cooling / Ventilation Fan 04-1021 2

8 Flow Controller 02-4209 1

9 Pressure Regulator 02-1110 1

10 Panel Gauge 02-1020 2

11 Safety Valve 02-4544 1

12 Compressor Anti-vibe Mounts 06-1020 4

13 Compressor 06-5513 1

14 Back Pressure Relief Valve 02-1213 1 Additional parts can be ordered from: Peak Scientific Instruments Fountain Crescent Inchinnan Business Park Inchinnan, Renfrewshire PA4 9RE Scotland, UK

Contact information: +44-141-812-8100 +44-141-812-8200 fax [email protected]

Page 13: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 12

14

15.

10 11

12

16.

1.

9.

6. 5.

13

7.

8.

11. Internal Pressure Gauge

Nitrogen GeneratorStandard System

14. Bulkhead Connection15. Internal Gauge Rail

13. Front Panel Gauge

16. Compressor Cooling Fan (if fitted)

12. Flow Regulator

16.

4.

Parts List1. Compressor

9. Nitrogen Storage Tank10. Pressure Regulator

7. CMS Column8. Non-return Valve

2. Cooling Coil

4. Filter / Separator5. Exhaust Valve6. Exhaust Silencer

3. Cooling Fan (if fitted)

2. 3.

NL ESwitch

Fuse

Fan Motor(s)

MSQ10LA Electrical Diagram

Terminal Rail

Compressor

Timer

Valve

Exhaust

Page 14: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 13

80.00 mm

800 mm

Pow

er O

utle

t Pos

ition

(Max

imum

Dis

tanc

e LH

S)

Pow

er O

utle

t Pos

ition

(Max

imum

Dis

tanc

e R

HS)

Rea

r Ven

tilat

ion

Gap

Cle

aran

ce fo

r Mai

nten

ance

875 mm

900 mm

Not

es

1. A

cces

s is

requ

ired

thro

ugh

Fron

t Pan

els

& Li

d fo

rm

aint

enan

ce.

2. P

lug

Pow

er C

ord

into

a c

onve

nien

t out

let w

ith s

uffic

ient

slac

k to

allo

w th

e un

it to

be

pulle

d fo

rwar

for a

cces

s.3.

Allo

w s

uffic

ient

sla

ck in

the

Gas

Lin

e to

allo

w th

ege

nera

tor t

o be

pul

led

forw

ard

for m

aint

enan

ce

MSQ

10LA

Inst

alla

tion

Not

es

Nitr

ogen

Pre

ssur

e

110 mm

625 mm

I O

410 mm

Gap

requ

ired

for P

ower

Cor

d

Page 15: MSQ10LA LC/MS Nitrogen Generator Manualtools.thermofisher.com/content/sfs/manuals/Man-031972-MSQ10LA... · permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, ... operation

MSQ10LA Nitrogen Generator

Instructions for Use

MSQ10LA issue 1 14

11 Maintenance Record Log

Model- MSQ10LA Serial Number:

Work Done

Remarks Date Name