Upload
otinanai09
View
6
Download
0
Embed Size (px)
DESCRIPTION
dead space
Citation preview
RESEARCH PAPER
Comparison of two different methods for physiologic dead
space measurements in ventilated dogs in a clinical
setting
Martina Mosing*, Lukas Staub! & Yves Moens"*Faculty of Veterinary Science, The University of Liverpool, Leahurst, Chester High Road, Neston, UK
!MEM Institute for Evaluative Research in Orthopedic Surgery, University of Bern, Switzerland
"Clinic for Anaesthesiology and Perioperative Intensive Care, Veterinary University Vienna, Vienna, Austria
Correspondence: Dr Martina Mosing, Vet Swiss Faculty, University of Zurich, Equine Department, Section of Anaesthesia,
Winterthurerstrasse 260, 8057 Zuerich, Switzerland. E-mail: [email protected]
Abstract
Objective To compare physiologic dead space (VD)
and physiologic dead space to tidal volume (VT)
ratio (VD/VT) obtained by an automated single
breath test for carbon dioxide (CO2) (method SBT)
and a manual calculation (method MC) in venti-
lated healthy dogs.
Study design Prospective clinical study.
Animals Twenty client-owned dogs, ASA I and II
undergoing anaesthesia for clinical purposes.
Methods Following pre-medication, induction of
anaesthesia, and intubation of the trachea, intermit-
tent positive pressure ventilation was commenced.
Mixed expired CO2 partial pressure (P!ECO2) was
measured by two methods: method MC by analysis,
using an infrared capnograph, of the expired gas
collected in a mixing box and method SBT which
calculated it automatically by a device consisting of a
mainstream capnograph and a pneumotachograph.
At four time points arterial partial pressure of CO2
(PaCO2) was measured. Physiologic dead space
variables (VD and VD/VT) were calculated manually
(method MC) or automatically (method SBT) using
the Bohr–Enghoff equation.
Method MC and SBT were compared using Bland–
Altman plots and linear regression. Intra-class
correlation coefficient (ICC) was used to measure
consistency of each method.
Results Four measurement pairs were obtained in
all 20 dogs for method SBT and MC. The bias was
)1.15 mmHg, 7.97 mL and 0.02 for P!ECO2, VD and
VD/VT, respectively. Linear regression analysis
revealed a correlation coefficient (r2) of 0.79,
0.94, and 0.83 for P!ECO2, VD and VD/VT, respec-
tively. The ICC revealed an excellent consistency for
both methods.
Conclusions The single breath test (SBT) can be
used for clinical evaluation of VD and VD/VT in
anaesthetized ventilated dogs.
Clinical relevance Through measuring VD and VD/
VT important information about lung ventilation
can be obtained and the SBT is an easy method to
use for this purpose.
Keywords dead space, dog, single breath test, ven-
tilation.
Introduction
Physiologic dead space (VD) or ‘wasted ventilation’ is
crucial to understand the relationship between
minute volume, tidal volume (VT) and the arterial
partial pressure of CO2 (PaCO2) of a patient during
393
Veterinary Anaesthesia and Analgesia, 2010, 37, 393–400 doi:10.1111/j.1467-2995.2010.00548.x
anaesthesia. Physiologic dead space represents
the sum of the alveolar (VDalv) and the airway dead
space (VDaw). Airway dead space is a more accurate
description for what was previously termed ‘ana-
tomical dead space’ as its size is dynamic and chan-
ges with respiratory variables such as respiratory
rate (fR), VT, airway flow rates and positive pressure
ventilation (Tusman et al. 2009). Calculation of VD
during anaesthesia using the Bohr–Enghoff equation
requires the simultaneous measurement of mixed
expired carbon dioxide tension (P!ECO2) and PaCO2
(Enghoff 1938). Mixed expired CO2 can be sampled
from a Douglas bag or a mixing box incorporated in
the expiratory limb of the anaesthetic circuit (Fig. 1).
More recently an alternative method based on
volumetric capnography became available for clini-
cal use. Volumetric capnography represents the plot
of expired CO2 concentration versus expired volume
(Fig. 2) and is also called single-breath test for
expired CO2 (SBT) (Aitken & Clark-Kennedy 1928).
Figure 2 shows a typical volumetric capnography
curve and illustrates the fact that the expired CO2 is
plotted against the expired volume and demonstrates
the difference between volumetric capnography and
a capnogram where the expired CO2 is plotted
against time. The SBT technique uses information
obtained by a mainstream capnograph and a pneu-
motachograph. The sensors of both devices are
inserted between the endotracheal tube and the
Y-piece of the breathing system. To calculate VD, the
obtained PaCO2 values are entered and an algorithm
is used, making separate measurement of P!ECO2
unnecessary. In humans SBT is used to determine
physiologic dead space variables after pulmonary
embolism, bronchoconstriction and to adapt venti-
lator settings in intensive care units (Fletcher 1990;
Arnold et al. 1995; Olsson et al. 1999; Verschuren
et al. 2004).
The objective of this study was to compare values
for VD and for its ratio to VT (VD/VT) obtained by
manual calculation (method MC) and volumetric
capnography (method SBT) in anaesthetized dogs
undergoing elective surgery in a clinical setting.
Materials and methods
The protocol was discussed and approved by the
head of the department.
Animals
Twenty dogs of different breeds, scheduled for
routine elective surgery, with a median bodyweight
of 33.2 kg (range: 19–55.7 kg) and age of 1.6 years
(range: 0.5–10 years) were included initially in this
study. Ten dogswere females and 10males. Inclusion
criteria were a bodyweight above 15 kg, American
Society of Anesthesiologists classification 1 or 2,
unremarkable lung auscultation at pre-anaesthetic
evaluation, and an arterial line in the dorsal pedal
artery. Eighteen dogs underwent orthopaedic proce-
dures and two soft tissue mass excisions. Seventeen
dogs were positioned in dorsal and three dogs in lat-
eral recumbency during the measurement period.
Pneumo-tachograph
Mainsteamcapnograph
Infraredcapnograph
Pop-offvalve
Ventilator
Sodalime
canister
Fresh gasinlet
Mixing
box
Expiratorylimb
Inspiratorylimb
One-wayvalve
Patient
Figure 1 Graphical illustration of
the arrangement of the circle system,
mixing box, and measurement
devices.
Physiologic dead space measurement in dogs M Mosing et al.
394! 2010 The Authors. Journal compilation
! 2010 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists, 37, 393–400
Anaesthetic technique
All dogs were pre-medicated with acepromazine
(0.02 mg kg)1; Vanastress; Vana, Austria) and
methadone (0.1 mg kg)1; Heptadon; Ebawe Phar-
ma GmbH, Austria) intravenously (IV). Anaesthesia
was induced with propofol given IV to effect (Prop-
ofol ‘Fresenius’; Fresenius Kabi, Austria). An endo-
tracheal tube was placed and connected to an
anaesthetic circle system. Fresh gas flow was 1–2 L
minute)1 oxygen. A 10 L gas mixing box was
placed in the expiratory limb of the anaesthetic
circle system (Fig. 1). Lactated Ringer’s solution
was infused at 10 mL kg)1 hour)1 throughout the
anaesthetic period (Ringer Lactat ‘Fresenius’;
Fresenius Kabi, Austria). Anaesthesia was main-
tained with isoflurane delivered via a precision out-
of-circle vaporiser in six dogs (Isofluran-Baxter;
Baxter AG, Austria) or with a continuous rate
infusion (CRI) of propofol in 14 dogs (10 mg
kg)1 hour)1) delivered via a syringe pump. All dogs
received a fentanyl CRI (0.02 mg kg)1 hour)1;
Fentanyl ‘Janssen’; Janssen-Cilag Pharma, Austria)
IV during anaesthesia. A 22-gauge catheter was
placed in the dorsal pedal artery for invasive arterial
blood pressure measurement and arterial blood
sampling. Intermittent positive pressure ventilation
(IPPV) in a volume-controlled mode was started in
all dogs at the beginning of surgery (Ohmeda 7800
Ventilator, Ohmeda, WI, USA). Delivered tidal vol-
ume and fR were adjusted to keep end-tidal CO2
between 4.66 and 6.65 kPa (35 and 50 mmHg).
Ventilator settings averaged a VT of 12 ± 4 mL kg)1
and a peak airway pressure of 12 ± 1 cmH2O and
were not changed throughout the period of data
collection.
Cardiovascular monitoring consisted of ECG,
invasive arterial blood pressure measurement and
pulse oximetry (HP CMS Monitor, Hewlett Packard,
Germany). All respiratory parameters were mea-
sured using the SBT monitor combination (Capno-
guard 1265 & Ventrak 1550, Novametrix Medical
Systems Inc., CT, USA). The combination consists of
a mainstream capnograph (Capnoguard 1265)
working on the principle of infrared absorption
and a fixed orifice pneumotachograph (Ventrak
1550). The capnograph and differential pressure
sensor were placed between the endotracheal tube
and the Y-piece of the circle system. The SBT
monitors were connected to a laptop and computer
software (Analysis plus!, Novametrix Medical Sys-
tems Inc., CT, USA) which allowed on- and off-line
data analysis (Fig. 1).
The capnograph was calibrated with room air
before each measurement. The accuracy of the
pneumotachograph was verified with a calibration
syringe before and after all measurements. The
system measured and displayed fR, end-tidal CO2,
respiratory volumes and pressures.
Mixed expired CO2 for method SBT was calculated
by the computer by dividing expired CO2 volume by
the tidal volume of a breath (Fig. 2). The P!ECO2
value was entered into the Bohr–Enghoff equation
(Bohr 1887; Enghoff 1938) to calculate VD/VT as
follows:
VD=VT ! "PaCO2 # P!ECO2$=PaCO2 "1$
The corresponding volume of VD in mL is calcu-
lated by multiplying the ratio by the measured tidal
volume (VT):
VD ! VT % ""PaCO2 # P!ECO2$=PaCO2$ "2$
After the start of IPPV respiratory parameters
were stabilised for at least 15 minutes. Data were
collected four times at 15 minute intervals for 1
hour. At each of the four time points (T1–T4) an
arterial blood sample of 0.3 mL was drawn from the
arterial catheter. Blood gases were analysed imme-
diately with a blood gas machine (AVL Compact 3,
Roche Diagnostics GmbH, Germany) calibrated
before each measurement period.
P!ECO2 for method MC was measured by analysing
expired gas sampled from the outlet of the mixing
box in the expiratory limb using a previously
calibrated side stream infrared analyzer (HP CMS
Monitor, Hewlett Packard) (Fig. 1) and retrospective
manual calculation of VD values using equation 1
and 2. The same PaCO2 values were used to
calculate VD with both methods.
Figure 2 Graphical illustration of an example for a single
breath test: the expired CO2 (Exp. CO2) is plotted against
the total expired volume. The volume of CO2 expired over
time represents the mixed expired CO2 (P!ECO2).
Physiologic dead space measurement in dogs M Mosing et al.
! 2010 The Authors. Journal compilation! 2010 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists, 37, 393–400 395
Statistical analyses
The variables P!ECO2, VD and VD/VT measured and
calculated by method MC and SBT were compared
and visualised using Bland–Altman plots and linear
regression (Bland & Altman 1986; Mantha et al.
2000). Intra-subject reproducibility of the four
measurements was evaluated by calculating intra-
class correlation coefficients (ICC). Reproducibility
was interpreted as follows: moderate 0.40–0.59,
good 0.60–0.79, excellent ‡ 0.8. The D’Agostino–
Pearson test was used to assess normal distribution
of the data. For pair-wise comparisons of different
points in time, Bonferroni adjustment was applied.
Statistical analyses were conducted with use of SAS
9.1 (SAS Institute, Cary, NC, USA). The level of
significance was set at 0.05 throughout the study.
Results are presented as mean ± one standard
deviation or medians and ranges depending on the
distribution of data.
Results
In all dogs four paired measurements were obtained
and analysed (80 paired measurements). All mea-
sured respiratory parameters remained within
physiologic limits throughout the anaesthetic period
(Table 1).
The volume per kilogram body mass for VDMC and
VDSBT was 7.60 ± 0.09 and 7.34 ± 0.09 mL kg)1,
respectively. The physiologic dead space ratio was
>50% at all four time points with a range between
46 and 79% for both methods (Table 2). The ICC
was >0.8 in all evaluated variables and revealed an
excellent reproducibility for all VD measurements
obtained by either method
Bland–Altman plots and linear regression revealed
a uniform relationship between the difference and
magnitude of measurements (Figs 3 & 4). Different
levels of bias and lower and upper confidence limits
were observed depending on the investigated vari-
able (Table 3). The bias levels were )1.15 mmHg
()0.15 kPa), 7.97 mL and 0.02 for P!ECO2, VD and
VD/VT, respectively. At most time points method
SBT underestimated P!ECO2 and therefore overesti-
mated both physiological dead space values com-
pared to method MC (Fig. 3).
The correlation coefficients (r2) for P!ECO2, VD and
VD/VT were 0.79, 0.94 and 0.83, respectively
(Table 3).
Discussion
The results of the present study indicate that volu-
metric capnography can be used to obtain physio-
logic dead space values in mechanically ventilated
dogs. Only minor differences were seen for P!ECO2
values compared to the use of a mixing box. Hence
VD and VD/VT were in close agreement as both
methods use the same formula to reveal dead
space parameters using PaCO2 and the aforemen-
tioned P!ECO2. Therefore the minor difference in the
calculated and measured P!ECO2 obtained by the two
methods was the most important finding of
this study. Several clinical studies in humans
confirm the accuracy of SBT to determine P!ECO2
and therefore VD and VD/VT (Fletcher 1987; Eriks-
son et al. 1989; Riou et al. 2004; Blanch et al.
2006). This is the first clinical study in dogs
describing the use of SBT to reveal dead space
parameters.
Physiologic dead space was 7 mL kg)1 and cor-
responds well with the values between 7 mL kg)1 in
spontaneously breathing and 10 mL kg)1 in venti-
lated dogs described in previous studies (Haskins &
Patz 1986; Berdine et al. 1990). In contrast, the
mean ratio of 0.6 for VD/VT for ventilated dogs was
higher than expected. Haskins & Patz (1986) gave a
Table 1 Respiratory parameters of the twenty dogs at the four time points (T1–T4)
Time
(minutes)
VT
(mL)
fR (breaths
per minute)
End-tidal CO2
(kPa)
End-tidal CO2
(mmHg)
PaCO2
(kPa)
PaCO2
(mmHg)
T1 (0) 421 ± 140 10 ± 2 5.72 ± 1.04 43 ± 8 6.06 ± 0.97 46 ± 7
T2 (15) 419 ± 129 10 ± 2 5.81 ± 1.00 44 ± 7 6.04 ± 0.86 45 ± 6
T3 (30) 413 ± 120 10 ± 3 5.92 ± 0.82 44 ± 6 6.17 ± 0.82 46 ± 6
T4 (45) 411 ± 115 10 ± 2 5.97 ± 0.86 45 ± 6 6.33 ± 0.77 48 ± 6
Values are given as mean ± standard deviation.
VT, tidal volume; fR, respiratory rate; PaCO2, arterial partial pressure of CO2.
Physiologic dead space measurement in dogs M Mosing et al.
396! 2010 The Authors. Journal compilation
! 2010 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists, 37, 393–400
ratio of 0.4 for VD/VT in spontaneously breathing
dogs with a mean VT of 18 ± 13 mL kg)1. During
mechanical ventilation VD/VT values up to 0.5 with
a VT of 20 mL kg)1 have been reported in dogs
using a multiple inert gas elimination technique
(Schrikker et al. 1995). In the present study dogs
were ventilated with a VT of 12 ± 4 mL kg)1 and a
peak airway pressure of 12 ± 1 cmH2O. IPPV canTab
le2
Resultsof
respiratorymeasuremen
tsan
dcalculation
sat
thefourtimepo
ints
(T1–T
4),expressedas
mediansan
d(ran
ge)
Tim
e
points
P! ECO
2MC
(kPa)
P! ECO
2MC
(mmHg)
P! ECO
2SBT
(kPa)
P! ECO
2SBT
(mmHg)
VDMC
(mL)
VDSBT
(mL)
VD/V
TMC
VD/V
TSBT
T1
2.39(1.60–3.06)
18(12–23)
2.53(1.73–3.46)
19(13–26)
238(137–475)
242(136–474)
0.59(0.53–0.77)
0.57(0.47–0.77)
T2
2.39(1.33–3.06)
18(10–23)
2.53(1.46–3.46)
19(11–26)
237(165–421)
242(144–415)
0.60(0.51–0.79)
0.59(0.46–0.79)
T3
2.33(1.46–2.93)
17(11–22)
2.53(1.46–3.59)
19(11–27)
239(119–367)
227(108–377)
0.61(0.53–0.77)
0.58(0.49–0.79)
T4
2.39(1.46–2.93)
18(11–22)
2.53(1.46–3.59)
19(11–27)
244(173–386)
237(155–401)
0.61(0.56–0.78)
0.60(0.48–0.78)
ICC
0.845
0.873
0.889
0.887
0.883
0.904
ICC,intraclass
correlatio
nco
efficient;SubscriptedMC,manualcalculatio
n;su
bscriptedSBT,single
breath
test
forCO
2,P! ECO
2,mixedexp
iredCO
2;VD,physiologicdeadsp
ace
;VT,tid
alvolume;VD/V
T,
physiologic
deadsp
ace
fraction.
(a)
(b)
(c)
Figure 3 Bland and Altman scatter plots of (a) the mixed
expired CO2; (b) physiologic dead space and (c) physiologic
dead space ratio, comparing method MC (manual calcu-
lation) and SBT (single breath test); solid line: bias, dashed
lines: 95%-CI of the differences. P!ECO2 = mixed expired
CO2; VD = physiologic dead space; VT = tidal volume VD/
VT = physiologic dead space ratio. For conversion to SI
units, 7.5 mmHg = 1 kPa.
Physiologic dead space measurement in dogs M Mosing et al.
! 2010 The Authors. Journal compilation! 2010 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists, 37, 393–400 397
increase VD/VT by 1) overventilation of parts of the
lung which have a very small blood flow, or 2)
overdistension of the conducting airways (increase
in VDaw) produced by the abnormal pattern of
inflation (Hedenstierna & McCarthy 1975; Tan &
Simmons 1979). The addition of the apparatus dead
space contributes to the increase in VD in anaes-
thetised subjects (Hedenstierna & McCarthy 1975).
This is especially true for VD/VT if small VTs are used
as even small changes have a high impact on the
ratio itself. In the present study a part of the VD
consisted of mechanical or apparatus dead space.
This includes parts of the endotracheal tube extend-
ing beyond the incisors and the SBT sensor itself.
The endotracheal tube connector was kept at the
level of the incisors. The volume of the sensor was
measured by water displacement and added
13.9 mL to the mechanical dead space. Wenzel
et al. (1999) demonstrated that in newborns with a
bodyweight of 2.7 kg an increase in inspired CO2
was apparent 5 minutes after inserting the com-
bined sensor and suggested that therefore the
measurement period should be restricted to a
maximum of 5 minutes in infants. We accounted
for this potential problem by excluding dogs with a
body weight below 15 kg. Nevertheless, the high
physiologic dead space ratio found is more likely as
a result of the the added apparatus dead space than
to an overdistension of the lung tissue and airways
as relatively low peak inspiratory pressures were
used. Evaluation of the changes of VDaw and VDalv
with different airway pressures using the same set
up would be necessary to prove this point.
The data for VD/VT found in this study do not
agree with other recently reported values on
physiologic dead space in dogs (Kudnig et al.
2004, 2006). However there was a major differ-
ence in the measurement methods used. Kudnig
et al. used end-tidal CO2 instead of the P!ECO2 to
calculate the ‘physiologic dead space’ resulting in
(a)
(b)
(c)
Figure 4 Linear regression of (a) the mixed expired CO2;
(b) physiologic dead space and (c) physiologic dead space
ratio comparing method MC (manual calculation) and SBT
(single breath test). Dashed lines represent the 95%
confidence interval and the solid line the fitted regression
line. P!ECO2 = mixed expired CO2; VD = physiologic dead
space; VT = tidal volume, VD/VT = physiologic dead space
ratio. For conversion to SI units, 7.5 mmHg = 1 kPa.
Table 3 Estimation of the difference (bias) and linear
regression (r2) between method SBT (single breath test)
and MC (manual calculation) for all dead space variables,
including 95% confidence limits.
Bias
Lower
limit
Upper
limit r2
P!ECO2 (mmHg) )1.15 )1.60 )0.70 0.79
VD (mL) 7.97 2.76 17.29 0.94
VD/VT 0.02 0.01 0.03 0.83
P!ECO2, mixed expired CO2; VD, physiologic dead space; VD/VT,
physiologic dead space fraction.
Physiologic dead space measurement in dogs M Mosing et al.
398! 2010 The Authors. Journal compilation
! 2010 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists, 37, 393–400
values of 0.11–0.24. The substitution of P!ECO2 by
end-tidal CO2 in the Bohr–Enghoff equation gives
values corresponding more likely with the VDalv
rather than VD (Fletcher 1985) so Kudnig et al.
incorrectly labelled their value as VD when in fact
they calculated the alveolar dead space fraction.
This explains the huge difference in the values in
the aforementioned studies and the study pre-
sented here.
The intra-subject reproducibility of the four mea-
surements was high for both method MC and
method SBT as demonstrated by an excellent ICC.
Therefore the variation around the values of VD for
both methods was small. Reproducibility can be
influenced by several factors including stability of
the baseline after calibration of volumetric capnog-
raphy devices, the blood gas machine, blood gas
sampling, entering data into the software and, most
importantly, analysis of the data. A significant
change in VD over a study period of 1 hour without
changing ventilator settings is not expected. There-
fore the high ICC can be interpreted as confirmation
for the consistency of both methods.
The bias of 8 mL for VD was caused mainly by
data obtained from two dogs and these SBT wave-
forms were inspected more closely. In these dogs
P!ECO2 was underestimated by the SBT by 4
(0.53 kPa) and 4.5 mmHg (0.6 kPa) causing an
overestimation of VD and VD/VT by 40.1 and
29.9 mL and 0.07 and 0.09 respectively.
Re-evaluating the single breath curves of the two
dogs revealed a short elevation in CO2 concentra-
tion in the first phase of the expiration (Fig. 5). The
algorithm of the SBT device only summarises the
area under the curve of phase II (quick increase in
CO2 concentration) and phase III (alveolar plateau)
(manufactures information sheet). The area under
phase I was not included into the P!ECO2 calculation
leading to the underestimation. A small residual
volume between the cuff and the tip of the endo-
tracheal tube (ETT) causes the aberration in the
baseline of the volumetric capnography (Schramel
& Mosing 2009). Carbon dioxide rich gas is trapped
in a small anatomical space between the outer
surface of the ETT distal to the inflated cuff and the
trachea. During exhalation, the trapped CO2-rich
gas is washed out during the first period of
expiration causing an aberration in baseline in
phase I of the expirogram. Thus, if an elevation in
phase I is observed during anaesthesia one can try
to deflate and reinflate the cuff to change the
volume of the space between the ETT and tracheal
wall. In case of persistence of the aberration, dead
space calculations based on SBT curves should be
interpreted with caution.
The handling of the combined monitoring device
for method SBT was convenient in a clinical setting.
With newer devices (NICO, Novametrix Medical
Systems Inc., CT, USA) PaCO2 values are entered
directly into one measurement unit without the
additional need of a laptop, making clinical set-up
even more convenient. The anaesthetist can use the
derived values to adjust ventilator settings during
the anaesthetic period. The classical calculation of
VD with collection of expired air in a mixing box has
several disadvantages in a clinical setting. The
addition of the 10 L-mixing box into the expiratory
limb of the circle system increases the time constant
of the breathing system. This might be a problem
when quick changes of gas composition are neces-
sary in a clinical setting. No such problems were
observed during data collection although only
healthy subjects in a stable plane of anaesthesia
were included into the study. Furthermore an
additional side stream capnograph is necessary at
the outlet of the box to measure P!ECO2 from the
mixing box if dead space parameters are derived by
manual calculation (Fig. 1). For the SBT the PaCO2
value simply has to be entered into the software and
the computer calculates all values, whereas manual
calculations are time consuming and not always
possible during clinical anaesthesia.
Both methods of measurement require arterial
blood gas analysis in order to obtain physiologic
dead space values. This might limit the use for
monitoring dead space variables in very small
animals because of the difficulties in arterial blood
sampling or when blood gas analysis is not available.
Figure 5 Screenshot of a single breath test for CO2 of dog 2
showing irregularities in phase I of the expiration; alter-
ation in baseline marked with arrow given in mmHg. For
conversion to SI units, 7.5 mmHg = 1 kPa).
Physiologic dead space measurement in dogs M Mosing et al.
! 2010 The Authors. Journal compilation! 2010 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists, 37, 393–400 399
At the time of data collection no ethical commit-
tee was in place at the institution. No alteration in
anaesthetic protocol occurred as a result of the
study, and the only potential difference from routine
practice was the addition of the mixing box in the
circuit and the arterial blood sampling. However, if
the study was carried out today, owner consent
would have been obtained.
In summary the SBT is an easy to use semi-
automated tool to estimate physiologic dead space
variables in a clinical setting and shows a high
correlation and low bias with manually calculated
values using the Bohr–Enghoff equation in healthy
ventilated dogs.
References
Aitken RS, Clark-Kennedy AE (1928) On the fluctuation
in the composition of the alveolar air during the
respiratory cycle in muscular exercise. J Physiol 65,
389–411.
Arnold JH, Bower LK, Thompson JE (1995) Respira-
tory deadspace measurements in neonates with
congenital diaphragmatic hernia. Crit Care Med 23,
371–375.
Berdine GG, Johnson JE, Dale D et al. (1990) Inspiratory
flow rate and dead space in dogs. J Appl Physiol 68,
1228–1232.
Blanch L, Romero PV, Lucangelo U (2006) Volumetric
capnography in the mechanically ventilated patient.
Minerva Anestesiol 72, 577–585.
Bland JM, Altman DG (1986) Statistical methods for
assessing agreement between two methods of clinical
measurement. Lancet 1, 307–310.
Bohr C (1887) Uber die Lungenathmung. Centralblatt fur
Physiologie 1, 236–268.
Enghoff H (1938) Volumen inefficax. Bemerkungen zur
Frage des schadlichen Raumes. Uppsala Lak Forhandl
44, 191–218.
Eriksson L, Wollmer P, Olsson CG et al. (1989) Diagnosis
of pulmonary embolism based upon alveolar dead space
analysis. Chest 96, 357–362.
Fletcher R (1985) Deadspace, invasive and non-invasive.
Br J Anaesth 57, 245–249.
Fletcher R (1987) Gas exchange during thoracotomy in
children. A study using the single-breath test for CO2.
Acta Anaesthesiol Scand 31, 391–396.
Fletcher R (1990) Deadspace during anaesthesia. Acta
Anaesthesiol Scand Suppl 94, 46–50.
Haskins SC, Patz JD (1986) Effects of small and large face
masks and translaryngeal and tracheostomy intubation
on ventilation, upper-airway dead space, and arterial
blood gases. Am J Vet Res 47, 945–948.
Hedenstierna G, McCarthy G (1975) The effect of anaes-
thesia and intermittent positive pressure ventilation
with different frequencies on the anatomical and alve-
olar deadspace. Br J Anaesth 47, 847–852.
Kudnig ST, Monnet E, Riquelme M et al. (2004) Cardio-
pulmonary effects of thoracoscopy in anesthetized nor-
mal dogs. Vet Anaesth Analg 31, 121–128.
Kudnig ST, Monnet E, Riquelme M et al. (2006) Effect of
positive end-expiratory pressure on oxygen delivery
during 1-lung ventilation for thoracoscopy in normal
dogs. Vet Surg 35, 534–542.
Mantha S, Roizen MF, Fleisher LA et al. (2000) Comparing
methods of clinical measurement: reporting standards
for Bland and Altman analysis. Anesth Analg 90, 593–
602.
Olsson K, Greiff L, Karlefors F et al. (1999) Changes in
airway dead space in response to methacholine provo-
cation in normal subjects. Clin Physiol 19, 426–432.
Riou Y, Leclerc F, Neve V et al. (2004) Reproducibility of
the respiratory dead space measurements in mechani-
cally ventilated children using the CO2SMO monitor.
Intensive Care Med 30, 1461–1467.
Schramel JP, Mosing M (2010) Aberration in the baseline
of volumetric capnography. Abstracts presented at the
10th World Congress of Veterinary Anaesthesia,
Glasgow. Vet Anaesth Analg 37, Available online DOI:
10.1111/j.1467-2995.2010.00535a.x
Schrikker AC, Wesenhagen H, Luijendijk SC (1995)
Intrapulmonary gas mixing and dead space in artificially
ventilated dogs. Pflugers Arch 430, 862–870.
Tan CSH, Simmons DH (1979) Effect of assisted mechan-
ical ventilation on arterial carbon dioxide tension of
normal anesthetized dogs. Lung 156, 255–264.
Tusman G, Scandurra A, Bohm SH et al. (2009) Model
fitting of volumetric capnograms improves calculations
of airway dead space and slope of phase III. J Clin Monit
Comput 23, 197–206.
Verschuren F, Liistro G, Coffeng R et al. (2004) Volumetric
capnography as a screening test for pulmonary embolism
in the emergency department. Chest 125, 841–850.
Wenzel U, Wauer RR, Schmalisch G (1999) Comparison of
different methods for dead space measurements in ven-
tilated newborns using CO2-volume plot. Intensive Care
Med 25, 705–713.
Received 31 July 2009; accepted 4 January 2010.
Physiologic dead space measurement in dogs M Mosing et al.
400! 2010 The Authors. Journal compilation
! 2010 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists, 37, 393–400