Upload
nguyenkhuong
View
216
Download
2
Embed Size (px)
Citation preview
More Speed and Resolution for Current Methods
A Simple “How –to –Guide” for Easier Method Transfer in theEasier Method Transfer in the
Current Lab Environment
Thomas J. WaegheColumns and ConsumablesAugust 28, 2007
Page 1
Key Points for Successful Transfer
1. Set Your Goals--Speed, Resolution, Both?
2. Choosing The Columng
3. Reducing The Column Length
4. Changing The Column i.d.4. Changing The Column i.d.
5. Adjusting The Injection Volume
6 Transferring to 1 8um Columns - Is and Grad6. Transferring to 1.8um Columns Is and Grad
7. Increase Resolution
8 Speeding It Up8. Speeding It Up
Page 2
Rule of Thumb
Maybe not always 100% Technically Correct but a very good easy alternative
Page 3
Column Choice is Key to Overall Success
• Choose Bonded Phase as Close to Original Method as Possibleas Possible
• Try Alternate Bonded Phases to Solve Peak Placement ProblemsPlacement Problems
• Choose Column Chemistry for Optimum pH Stability and Column Life
Page 4
Bonded Phase Structures Affect Selectivityand pH Ruggedness
StableBond C18 Extend C18
C18C18
and pH Ruggedness
SiO
SiO
C18C18
Eclipse C18 StableBond CN
Page 5
Start Method Development with RRHT Columns: Different ZORBAX RRHT C18 Bonded Phases for Max Selectivity
Eclipse Plus C18
M bil h (69 31) ACN t
1 23
41st choiceBest Resolution& Peak Shape
StableBond SB C18
Mobile phase: (69:31) ACN: waterFlow 1.5 mL/min.Temp: 30 °CDetector: Single Quad ESI positive mode scan Columns: RRHT
1 2 3 4 5
1 234
2nd choiceGood alternate selectivity
Eclipse
SB-C18 Columns: RRHT 4.6 x 50 mm 1.8 um
Sample:1. anandamide (AEA)2 P l it l th l id (PEA)1 4
min1 2 3 4 5
due to non-endcapped
3rd choice Eclipse XDB-C18
2. Palmitoylethanolamide (PEA)3. 2-arachinoylglycerol (2-AG)4. Oleoylethanolamide (OEA)
1 23
4
1 2 3 4 5
3 choiceGood efficiency & peak shapeResolution could be achieved
4th choiceMultiple bonded
Extend-C181 2,3 4
4th choiceResolution not likely,Other choices better, for this separation.
phases for most effective method development.Match to one you are
Page 6
1 2 3 4 5
Match to one you are currently using.
How To “Match” a Column to a ZORBAX RRHT ColumnColumn
General Phase Type Starting ZORBAX Choice
Typical “endcapped” C18 or C8 bonded phases, newer columns Eclipse Plus C18 or C8
Endcapped C18 or C8 columnsEndcapped C18 or C8 columns, older generation Eclipse XDB-C18 or C8
Non-endcapped columns StableBond C18
Older types of columns StableBond C18, C8 etc.
Aqueous “type” columns SB-AQ
CN or Phenyl SB-CN, SB-Phenyl
Page 7
Choose Suitable Column Phase
Mobile Phase pH
Stable Bond pH 1 – 6
Acidic
Stable Bond pH 1 6SB can use increased temperature up to 100°C
IntermediateEclipse Plus /Eclipse XDB + pH 2 – 9
Eclipse can use increased temperature up to 60°C
B i
Extend pH 8 – 11.5
Extend can use increased temperature up to 45°C
Basic
Page 8
The Complete List of RRHT 600 bar Columns – Jan 1Eclipse
Dimensions Eclipse Plus C18 Eclipse Plus C8 XDB-C18 Eclipse XDB-C8 Extend-C184.6 x 150 959994-9024.6 x 100 959964-902 959964-906 928975-902 928975-906 728975-9024.6 x 50 959941-902 959941-906 927975-902 927975-906 727975-9024.6 x 30 959931-902 959931-906 924975-902 924975-906 724975-9024.6 x 20 926975-902 926975-906 726975-9023 0 x 150 959994 302
2 Recommended Starting Choices:
3.0 x 150 959994-3023.0 x 100 959964-302 959964-306 928975-302 928975-306 728975-3023.0 x 50 959941-302 959941-306 927975-302 927975-306 727975-3023.0 x 30 924975-302 924975-306 724975-3023.0 x 20 926975-302 926975-306 726975-3022.1 x 150 959794-9022.1 x 100 959764-902 959764-906 928700-902 928700-906 728700-902
• 959941-902• 959741-902They are Eclipse Plus C18 Columns.2.1 x 50 959741-902 959741-906 927700-902 927700-906 727700-902
2.1 x 30 959731-902 959731-906 924700-902 924700-906 724700-9022.1 x 20 926700-902 926700-906 726700-902Dimensions SB-C18 SB-C8 SB-Phenyl SB-CN SB-AQ Rx-Sil4.6 x 150 829975-902 829975-906 829975-912 829975-905 829975-9144.6 x 100 828975-902 828975-906 828975-912 828975-905 828975-914 828975-9014 6 x 50 827975 902 827975 906 827975 912 827975 905 827975 914 827975 901
C18 Columns.
4.6 x 50 827975-902 827975-906 827975-912 827975-905 827975-914 827975-9014.6 x 30 824975-902 824975-906 824975-912 824975-905 824975-9144.6 x 20 826975-902 826975-9063.0 x 150 829975-302 829975-306 829975-312 829975-3053.0 x 100 828975-302 828975-306 828975-312 828975-305 828975-314 828975-3013.0 x 50 827975-302 827975-306 827975-312 827975-305 827975-314 827975-3013.0 x 30 824975-302 824975-306 824975-3053.0 x 20 826975-302 826975-3062.1 x 150 820700-902 820700-906 820700-912 820700-9052.1 x 100 828700-902 828700-906 828700-912 828700-905 828700-914 828700-9012.1 x 50 827700-902 827700-906 827700-912 827700-905 827700-914 827700-9012.1 x 30 824700-902 824700-906 824700-912 824700-905 824700-9142.1 x 20 826700-902 826700-906
(Main product page)
Page 9
(Main product page)(P/N List)
Fi t L t’ W k E S d G i !First, Let’s Work on Easy Speed Gains!
Page 10
ISOCRATIC ISOCRATIC ELUTION
Page 11
Simplified Method Transfer for Increased Isocratic SpeedIncreased Isocratic Speed
1. Shorten the Column Length
2. Reduce Particle Size
3. Maintain Flow Rate
4. For More Speed….
Wait, Hold the Line - Easy Gains First!
V F i J F Slid A !Very Fast is Just a Few Slides Away!
Page 12
Reducing Column Length and Particle SizeMaintains Resolution While Reducing Assay Time
ColumnLength (mm)
ColumnEfficiency N(5 µm)
ColumnEfficiency N(3 5 µm)
ColumnEfficiency N(1 8 µm)
Analysis Time*
(mm) N(5 µm) N(3.5 µm) N(1.8 µm)150 12,500 21,000 35,000
100 8 500 14 000 23 250Efficiency (N)
AnalysisTime
-
-33%100 8,500 14,000 23,250
75 6000 10,500 17,500
50 4 200 7 000 12 000
( )
Pressure
Time
PeakVolume
-50%
-67%50 4,200 7,000 12,000
30 N.A. 4,200 6,500
15 N A 2 100 2 500
67%
-80%
90%
SolventUsage
15 N.A. 2,100 2,500 -90%
• Reduction in analysis time compared to 150 mm column; all columns 4.6-mm i.d.
• Shorter columns with small particles provide the efficiency of longer
Page 13
p p y gcolumns with larger particles
First, Reduce Column Length
5um 1.8um
Reduce column length by factor of 3
Quite often original method will have more resolution than is actuallyQuite often original method will have more resolution than is actually needed and a reduction by 5 may be possible
3 5 1 83.5 um 1.8um
Reduce column length by factor of 2
Quite often original method will have more resolution than is actually needed and a reduction by 3 is possible
Page 14
Reduced Column Length and Particle Size Reduce Analysis Time
Original methodZORBAX LC column Extend C18
mAU
Both Separations Performed on Same Agilent 1200 SL HPLC
Extend-C184.6 x 150 mm, 5 μm
100
150
200
250
12 5
4
36
min0 2 4 6 8 10
0
50
2 5
mAU
80
100
120 ZORBAX RRHT column Extend C184.6 x 50 mm, 1.8 μm
2 3
4
51 6
3x faster
-20
0
20
40
60
Mobile phase: (70:30) MeOH: 50 mM pyrrolidine buffer Flow = 1.0 mL/min, Temp. : ambient
2 3 5 6
Page 15
min0 2 4 6 8 1020
Reached Flow Rate Limit? Change Column I.D.
Changing Column I.D. Will Allow
Lower Flow Rate to Match Linear velocity
Can improve speed potential – lower initial flow rate used allowsCan improve speed potential – lower initial flow rate used allows higher overheads within the flow rate range of the instrument
Reduce Mobile Phase Use
Can improve sensitivity for same mass loading, but must optimize the flow path volume to minimize dispersion.
Page 16
How conversion works for flow
Flow modification, for columns of different diameters
2⎛ ⎞2 col.
2column2
1 col. FlowDiamDiam.Flow =⎜
⎜⎛
⎟⎟⎠
⎞×
column1.Diam⎜⎝
⎟⎠
ml/minmmmmml/min 21.0
4.62.11.0 i.e.
2=
⎜⎜
⎝
⎛⎟⎠⎞×
Page 17
⎝
Reduce Column Diameterand Flow Rate by Same Factorand Flow Rate by Same Factor
4.6 mm 3 mm
Reduce flow rate by factor of 2.4
Reducing to 3 mm i.d. column allows use of higher linear flow rates as the 1200 SL pump will pump up to 5ml/min.
Compared to 4.6 mm id column at 1 mL/min, you can operate a 3 mm id column at 0 42 mL/min and save solvent or reduce flow rate to MScolumn at 0.42 mL/min and save solvent or reduce flow rate to MS detector, or you can keep the flow rate at 1 mL/min, and have 2.4-fold increase in linear velocity
2.1 mm4.6 mm
Reduce flow rate by factor of 4.8
Page 18
Quick Reference for Changing to Common Column DiametersCommon Column Diameters
Maintains Equivalent Linear Velocity for Different Column IDs:Column Type Column ID Flow Rateyp
Analytical 4.6 mm 1.0 mL/min
Solvent Saver 3.0 mm 0.42 mL/min
N B 2 1 0 21 L/ iNarrow Bore 2.1 mm 0.21 mL/min
MicroBore 1.0 mm 47 μL/min
Capillary 0.5 mm 12 μL/minp y μ
Capillary 0.3 mm 4.2 μL/min
Nano 0.1 mm 472 nL/min
Flow rate column 2 = (diameter column 2)2/(diameter column 1)2 x Flow rate column 1
Nano 0.075 mm 266 nL/min
Page 19
• Maintain equivalent mobile phase linear velocity when scaling down in column diameter.
Changing Both Column Length and Diameter Change from a 4.6 x 250 mm (5 um) to a 3.0 x 100 mm (3.5 um) Column.
1mAU
175
Mobile Phase: 25% methanol in 0.4% Formic Acid
75
100
125
150 ZORBAX SB-C18, 4.6 x 250 mm, 5 mm, 1 mL/min
Solvent Used: 34 mL2
3
4
min0 5 10 15 20 25 30 35
0
25
50
mAU
2 45 6
min5 10 15 20 25 30 35U
100
125
150
175
ZORBAX SB-C18, 3.0 x 100 mm, 3.5 mm, 0.425 mL/minSolvent Used: 5.7 mL, decrease of 83% (decrease in analysis time of 57%)
0
25
50
75
Page 20
min0 5 10 15 20 25 30 35
OK, Let’s Run…More Isocratic SpeedMore Isocratic Speed
• Increase flow rate• If flow rate limit of instrument* is reached, reduce column diameter and
reduce flow rate to maintain linear velocity• If pressure becomes problematic, increase mobile phase/column
temperaturep• Increase flow rate as necessary or desired to within 80-90% of flow or
pressure limit* Flow Rate Limit is 5mL/min for Agilent Binary Pump 10mL/min Agilent Quaternary Pump (P < 200 bar)
Page 21
Flow Rate Limit is 5mL/min for Agilent Binary Pump, 10mL/min Agilent Quaternary Pump (P < 200 bar)
Speeding It Up-How Fast can We Go?Increase Linear Velocity (Flow Rate)
0.0045
Increase Linear Velocity (Flow Rate)
0 0025
0.0030
0.0035
0.0040
5 0
Step 1 (2x Original Flow)Increase the flow rate by 100%
• Reduce run time by 50%
0.0010
0.0015
0.0020
0.0025 5.0 μm
3.5 μmcm
/pla
te)
• Reduce gradient time segments by 50%
-0.0005
0.0000
0.0005
0 0 0 2 0 4 0 6 0 8 1 0 1 2 1 4 1 6
μm 1.8
μm 1mL/minH
ETP
(c
Example0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6
2.0 mL/min
Interstitial linear velocity (ue- cm/sec)
p• 1ml/min > 2ml/min
• 30 minutes > 15 minutes
• 35-65% over 30 mins > over 15 mins
Page 22
Speeding It Up – Faster!Increase Linear Velocity
0.0045 Step 2 (3 x original flow)
Increase Linear Velocity (Flow Rate)
0 0025
0.0030
0.0035
0.0040
5 0
p ( g )
• Increase the flow rate by another 100%
0.0010
0.0015
0.0020
0.0025 5.0 μm
3.5 μmcm
/pla
te) • Reduce original run time to 1/3
• Reduce original Gradient time to 1/3
-0.0005
0.0000
0.0005μm
1.8 μm
1mL/minHET
P (c
Example• 1ml/min > 3ml/min
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6
3.0 mL/minInterstitial linear velocity (ue- cm/sec)
• 30 minutes > 10 minutes
• 35-65% over 30 mins > over 10 mins
Is resolution OK ? Pressure OK? Keep Going!
Page 23
Is resolution OK ? Pressure OK? Keep Going!
Speeding It Up-FASTER!!Increase Linear Velocity
0.0045Step 3 (4 x original flow)
Increase Linear Velocity (Flow Rate)
0 0025
0.0030
0.0035
0.0040
5 0
Step 3 (4 x original flow)
• Increase the flow rate by another 100%
0.0010
0.0015
0.0020
0.0025 5.0 μm
3.5 μmcm
/pla
te)
• Reduce original run time to 1/4.
• Reduce original Gradient time to ¼
-0.0005
0.0000
0.0005μm
1.8 μm
1mL/minHET
P (c
Example1 l/ i 4 l/ i0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6
4.0 mL/min
Interstitial linear velocity (ue- cm/sec)
1ml/min > 4ml/min
• 30 minutes > 7.5 minutes
• 35-65% over 30 mins > over 7.5 mins
I l i OK? P OK? K G i !
Page 24
Is resolution OK? Pressure OK? Keep Going!
Agilent 1200 Rapid Resolution System – SpeedConventional LC → UFLC → RRLC
F= 1.20ml/minT = 40°CAnalysis Time = 11min
High Resolution:4.6mm x 150mm 5.0µm
Conventional LC UFLC RRLCRs = 4
Solvent Cons. = 13.2ml
min0 2 4 6 8 10 12
F = 4.80ml/minT = 40°CAnalysis Time = 1 05min
High Speed:
Rs will be ~60% lower
thoughRs = 2.3
Analysis Time = 1.05minSolvent Cons. = 5.1ml
4.6mm x 50mm 5.0µm
min0 0.2 0.4 0.6 0.8 1
F= 1.00ml/minT = 40°CAnalysis Time = 1.1minSolvent Cons. = 1.1ml
High Speed & Resolution:2.1mm x 50mm 1.8µm
min0.2 0.4 0.6 0.8 10
Max Speed at T = 95oC2.1mm x 50mm 1.8um
F= 2.40ml/minT = 95°CAnalysis Time: 0.4minSolvent Cons. = 1.0ml
PWHH = 197msec
> 20x faster !
Page 25
min0.2 0.4 0.6 0.8 10
8X Faster Analysis on RRHT Columns with Easy Method Transfer – High Pressure not RequiredMethod Transfer High Pressure not Required
Column: ZORBAX Eclipse XDB-C18Mobile Phase: 95% ACN: 5% WaterTemp: AmbientRRHT
Conventional
Temp: AmbientInjection volume: 1 uL
RRHT4.6 x 50 mm 1.8 μm
Flow Rate: 3 mL/minPressure = 229 bar
Conventional4.6 x 150 mm 5 μm
Flow Rate: 1 mL/minP = 37 bar
mAU
60
80
20
40 1.7 min13.5 min
min0 2 4 6 8 10 12 140
Sample: Vitamin E – α, β, γ-tocopherols in gel capE li XDB C18 i d fi t h i f th d
Page 26
Eclipse XDB-C18 is a good first choice for many methods.
Reduce Analysis Time by up to 95% using Rapid Resolution HT ColumnsRapid Resolution HT Columns
4.6 x 250 mm, 5 μm29.65
12
3
4
Rs(1,2) = 4.8N=22680N=21848
min0 5 10 15 20 25 30
4.6 x 100 mm, 3.5 μm12.71
12
3
4
Rs(1,2) = 3.5N=11691
.
4.6 x 30 mm, 1.8 μm 1 mL/min
min0 5 10 15 20 25 30
4 15
12
3
4
Rs(1,2) = 3.3
N=11691
N=65681 mL/min
4.6 x 30 mm, 1.8 μm
min0 5 10 15 20 25 30
4.15
1 2 3
4
s( , )
R (1 2) = 3 1
N=6104
N=6463 4.6 x 30 mm, 1.8 μm2 mL/min
min0 5 10 15 20 25 30
0.5 1 1.5 2 2.52.09
4Rs(1,2) = 3.1
Columns: ZORBAX SB-C18 Mobile Phase: 50% 20 mM NaH2PO4, pH 2.8: 50% ACN Flow Rate: 1 mL/min
N=6460
Page 27
Columns: ZORBAX SB C18 Mobile Phase: 50% 20 mM NaH2PO4, pH 2.8: 50% ACN Flow Rate: 1 mL/min Temperature: RT Detection: UV 230 nm Sample: 1. Estradiol 2. Ethinyl estradiol 3. Dienestrol 4. Norethindrone
When to Stop
• When flow limit of pump is reached
Consider using a smaller i.d. column
Wh i hi i ( 550 b )• When pressure is approaching maximum (e.g. 550 bar)
Remember to allow for increase in pressure at viscosity maximum for mobile phase mixture
Wh l ti i l ti f t• When resolution is no longer satisfactory
Methanol/water 40CACN/water 40C
0 600.801.001.201.40
cosi
ty0.40
0.60
0.80
cosi
ty
0.000.200.400.60
0 20 40 60 80 100 120
%M th l
Visc
0.00
0.20
0 20 40 60 80 100 120
%ACN
Vis
Page 28
%Methanol
Decreased Column Volume May Require Conversion for Injection Volume
Keep Injection volume proportional to column volume
2colcolumn2
1col Inj.Vol.VolumeInj.Vol. =⎜⎜⎛
⎟⎟⎞
× 2 col.column1
1 col. j.Vo .Volume
j.Vo . ⎜⎝
⎟⎟⎠
Zorbax column volume = 3.14 x r2 x L x 0.6 (r and L in cm)
2 col.column1
column21 col. 4
2.04.020 i.e. μlmlmlμl =⎜
⎜⎝
⎛⎟⎟⎠
⎞×
Page 29
column1⎝ ⎠
Reduce injection volume
4.6 mm 3 mm
Reduction to allow for diameter change
= 0.4 x Original
2.1 mm4.6 mm = 0.2 x Original
xReduction to allow for length change
150 mm 50 mm = 0.33 x Original150 mm 50 mm 0.33 x Original
100 mm 50 mm = 0.5 x Original
150 mm 100 mm = 0.67 x Original
e.g. Original 4.6mm x 150mm transferred to 2.1mm x 100mm
= 0.2 x 0.67
100 mm 50 mm 0.5 x Original
Page 30
= 0.13 x original injection volume
Reduce Injection Volume as Column Volume is Reduced
mAU
5075
USP Requirements:L7 column, Rs > 8Tf(5%) < 2.0 for each
4.6 x 150 mm, 5 um5 uL inj.
R = 16.5N/m=73.000
min0 1 2 3 4 5 6
025
50
mAU
f( )N=3000 or 20000/m
Rapid ResolutionmAU
025
5075
p4.6 x 100 mm, 3.5 um3.3 uL inj.R = 17.3
N/m=117,000
Tailing factor for each of these six peaks is <1.3
min0 1 2 3 4 5 6
0
mAU
75
Rapid Resolution HT 4.6 x 50 mm, 1.8
Mobile phase: ( 500:496:4) acetonitrile: water: H3PO4Flow = 2.0ml/min. isocratic Temp: ambient Detection: UV 272nm LC: Agilent 1100
R = 15.8N/m=222000
min0 1 2 3 4 5 6
025
504.6 x 50 mm, 1.8 um1.7 uL inj.
272nm LC: Agilent 1100Sample: “resolution solution”, fenoprofen (peak #1) with gemfibrozil prepared as described in USP
Page 31
• High resolution and exceptional efficiency maintained for low cost updating to fast LC methods
GRADIENT GRADIENT ELUTIONELUTION
Page 32
Simplified Method Transfer for Increased Gradient SpeedIncreased Gradient Speed
1. Shorten the Column Length
2 Adj t G di t Ti b F t2. Adjust Gradient Time by same Factor
3. Maintain Flow Rate
4. For More Speed… Just Wait a Little Longer.
Remember, Easy Gains First!
“Faster” is Just a Few Slides Away!
Page 33
Faster is Just a Few Slides Away!
What is Gradient Elution?Is It More Difficult to Increase Gradient Speed?p
Increasing the solvent strength = Increasing the % organic in the mobile phase
Linear solvent strength gradient = % B per min is a constant
90 90%90
%ACNΔφ = 80%t G = 40 min.50%
70%
gradient time
gradient change
10 } } } } Δφt G = 2%/min.
0 10 20 30 40 min.
30%gradient slope
F 20% h i ACN t i 10 i
} } } }Δt 1 = Δt 2 = Δt 3 = Δt 4
Page 34
For every 20% change in ACN, t is 10 min.
Changing Gradient Time to AffectRetention (k*) and Resolution
100% B
100% B
tg= 5
tg Fk* ∝
0% B
tg= 101/k* = gradient steepness = b
S Δ%B Vm
k ∝
100% B
t = 20
0% B
100% B
tg= 20 ΔΦ = change in volume fraction of B solvent
S = constantF = flow rate (mL/min.)t di t ti ( i )
0% B
0 10 20 30 40
tg= 40tg = gradient time (min.)
Vm = column void volume (mL)
0% B
Group/Presentation TitleAgilent Restricted
Month ##, 200X
0 10 20 30 40Time (min)
000995P1.PPT
How conversion works for time
Run Time or Gradient segment Time Adjustment*
⎛ ⎞2 col.
column21 col. Time
LengthLengthTime =⎜
⎜⎝
⎛⎟⎟⎠
⎞×
1 0⎛ ⎞
column1Length⎜⎝
⎟⎠
.15250150.25 i.e. min
mmmmmin =⎜⎜
⎝
⎛⎟⎠⎞×
*
Page 36
*assumes diameter is equal for columns 1 and 2
Convert run time (and gradient) to shorter column with no flow increaseto shorter column with no flow increase
Flow Rate = As converted
Injection Volume = As convertedInjection Volume = As converted
Run time = Reduced by length change reduction factor
Gradient Times = Reduced by length change reduction factor
150 mm 75 mm
Reduction to allow for length change
= 1/2 x Reduction150 mm
e.g. 18mins 9mins
Page 37
45-90% over 18mins 45-90% over 9mins
Short Columns Reduce Total Gradient Analysis TimeTime
Gradient Separation of Cardiac Drugs
A 4 6 x 150 mm 5 mm B 4 6 x 75 mm 3 5 mm
1
5
5
A. 4.6 x 150 mm, 5 mmEclipse XDB-C8tG= 18 min
B. 4.6 x 75 mm, 3.5 mmEclipse XDB-C8tG= 9 min
1
14 4Run Time 13 min 8 minEquilibration 15 min 7 minTime
4.6 x 150 mm 4.6 x 75 mm
2
23
348 samples/day 96 samples/day
Total Analysis 28 min 15 minTime
Time (min) Time (min)0 2.5 5.0 7.5 1.00 12.5 15.0 0 2.5 5.0 7.5
Page 38
Time to Have Fun…More Gradient Speed
• Reduce Particle Size to Increase Efficiency and Peak Capacity • Decrease Gradient Time and Increase Flow Rate by Same Factor• If Flow Rate limit of instrument* is reached, reduce diameter of column
and reduce flow rate to maintain linear velocity• Repeat Flow Rate Increase to Flow or Pressure LimitRepeat Flow Rate Increase to Flow or Pressure Limit• If pressure becomes problematic, increase mobile phase/column
temperature*
Page 39
* Flow Rate Limit 5mL/min for Agilent Binary Pump, 10mL/min Agilent Quaternary Pump (P < 200 bar)
Faster Gradient Analysis of Cardiac DrugsWhat Changes Did We Make to Accomplish This?What Changes Did We Make to Accomplish This?
Optimized Column and Gradient
5
1
4
Column: Rapid Resolution Eclipse XDB-C8,
4.6 x 50 mm, 3.5 μmMobile Phase: A: 25 mM Na2HPO4, pH 3
Run Time 1.8 minEquilibration 1 min
2
p 3B: MeOH
Gradient: 42 – 90% B in 3 minFlow Rate: 3 mL/minTemperature:35°CSample: Cardiac Drugs
TimeTotal Analysis 2.8 minTime
3
1. Diltiazem2. Dipyridamole3. Nifedipine4. Lidoflazine5. Flunarizine
480 Samples/day
Page 40
Time (min)0.0 0.5 1.0 1.5 2.0 2.5 3.0
Column Efficiency under Gradient Conditions –Concept of Peak Capacity.Standard concept of column efficiency, N (plates), is only appropriate for isocratic conditions. A more useful concept for the case of gradient conditions is Peak Capacity - the number of peaks that can be separated (at a specified resolution) in a given amount of time. It is another measure of column efficiency.
4
2 peaks fit
5
5 μm
1 8 μm
2 peaks fit
Pc = (1+ tG/w)for gradient 1.8 μm
3 peaks fit
for gradient conditions
Rs = +50%
3 peaks fit50% more!!
Page 41
Gradient Resolution and Shorter Columns -Faster Analyses for Complex Samples – Higher Peak Capacity, Pc = (1+ tG/w):
2.1x150mm, 5μmP/N 883700-922 120 peaks
60 min
Detection: UV 214 nm Sample: HSA Tryptic Digestk* = 12.0
min15 20 25 30 35 40 45 50 55
70 min gradient0.2 mL/min
60 min.
min15 20 25 30 35 40 45 50 55
2.1x50mm, 1.8μmP/N 822700-90210 min gradient0.5mL/min
125 peaks10 min.!
k* = 12.9
min3 4 5 6 7 8 9 10
2.1x50mm, 1.8μmP/N 822700-902 156 peaks!k* = 38.6
min5 10 15 20 25
P/N 822700 90230 min gradient0.5mL/min
p25 min
Page 42
min5 10 15 20 25Conditions: Mobile Phase A: Water w/ 0.1% TFA, B: ACN w/0.1% TFA, Gradient 2%B to 50%B, Temperature: 50°C
Peptide Map of Tryptic Digest of BSA run on Agilent RRHT Zorbax SB-C18, 2.1x50mm, 1.8µAgilent RRHT Zorbax SB C18, 2.1x50mm, 1.8µ
Gradient time10min
Peak Capacity10min 347
15 min 398Longer Gradient Times Increase Peak
22.5 min 441
Capacity
Page 43
Peak Capacity as a function of Gradient Time (tg) and Column Length on 1.8 µm RRHT Columnsand Column Length on 1.8 µm RRHT Columns
Gradient Time 50 mm 100 mm 150 mm
10 347 --- ---10 347 --- ---
15 398 --- ---
20 47720 --- 477 ---
22.5 441 --- ---
30 535 54030 --- 535 540
45 --- 587 610
67 5 69467.5 --- --- 694
Tripling column length increases peak capacity by 55%. Increasing gradient time by 225% increase peak capacity by 25%
Page 44
Increasing gradient time by 225% increase peak capacity by 25%
Flow Rate Limitations Due to Pressure? Higher Temperature Can Help!
Page 45
Higher Temperature as an Aid to Faster Operation
Higher Temperature:Temperature should always be considered as a parameter during speed optimizationgProvides more rapid mass transfer:• Improves efficiency – enhances resolution• Decreases analysis time – faster separations with no loss in resolution• Decreases analysis time – faster separations with no loss in resolution
Decreases Mobile Phase Viscosity• Lowers backpressure – allows for higher flow rates, faster separations, greater
efficiency and use of sub 2-micron columnsefficiency and use of sub 2-micron columnsCan change selectivity – optimize resolution
Be wary of on-column decomposition• Faster flow rates shorten analyte residence time at elevated temperature and lead
to less decomposition for labile compounds
Page 46
Elevated Temperature Reduces Pressure – Expands Column Choices – Cardiac drugsColumn Choices Cardiac drugs
4.6 x 250 mm, 3.5 µmR = 4 3
Column: SB-C18, As described below Mobile Phase: A: 0.1% TFA, 5% MeCN, (v/v) B: 0.08% TFA, 95% MeCN (v/v)Sample: 0.1 mg/ml of cardiac drugs Temperature: 70°C Flow: 2 mL/min. gradient Detection: 230,16 nm
Very high N column P=221 bar
Rs= 5.6Rs= 4.3 column
min2 4 6 8 10 12 14 16
4.6 x 150 mm, 1.8 µmP=418 barRs= 4.9
Rs= 3.6
min2 4 6 8 10 12 14 16
4 6 x 50 mm 1 8 µm250 mm150 mm50 mm% B
s
R 3 4 4.6 x 50 mm, 1.8 µmP=164 bar
20.0112.014.0112.5
2012460
17.510.53.560
00012.5Rs= 3.4Rs= 2.4
Page 47
min2 4 6 8 10 12 14 16
INCREASED RESOLUTION! When Do You Need It? How Do You Get It?When Do You Need It? How Do You Get It?
• Complex Samples Large No of Peaks in Short Time Frame• Complex Samples—Large No. of Peaks in Short Time Frame
• Closely Related Compounds
• Changes in Bonded Phase Have Not improved Resolution
• Changes in Mobile Phase Have Not Improved Resolution
• Temperature Has Not Helped Change Selectivity
What Is Left That Will Improve Resolution?
Page 48
Reduce Particle Size and Maintain Column LengthIncreased “N” in Isocratic Separation
mAU
2
2.5
DAD1 A, Sig=254,4 Ref =of f (051119A\SIG10003.D) Column:150x4.6 mm
5µm
Pressure: 93 bar
N: 8213
4.6 x 150, 5um93 barN = 7259R 1 15 20000
25000
30000
pdN 1
∝
p
0 5
0
0.5
1
1.5 Height: 1.25 mAU
S/N: 42.3
Rs = 1.15
tr = 14.9 min (1st epimer)
Nptp: 2.4 10-2 mAU
Rs = 1.15S/N = 42 Height = 1.25Noise = 24uAU
N
5000
10000
15000
20000 pd
min0 5 10 15 20
-1
-0.5
Norm.
DAD1 A, Sig=254,4 Ref=off (051007D\LC_X000001.D)
Norm.
DAD1 A, Sig=254,4 Ref=off (051007D\LC_X000001.D)Norm.
DAD1 A, Sig=254,4 Ref=off (051004D\LC_J0002.D)
Norm.
DAD1 A, Sig=254,4 Ref=off (051004D\LC_J0002.D)
Norm.
DAD1 A, Sig=254,4 Ref=off (051007D\LC_X000001.D)
Norm.
DAD1 A, Sig=254,4 Ref=off (051007D\LC_X000001.D)Norm.
DAD1 A, Sig=254,4 Ref=off (051004D\LC_J0002.D)
Norm.
DAD1 A, Sig=254,4 Ref=off (051004D\LC_J0002.D)1/dp
0.2 0.3 0.4 0.5 0.65000
Column:150 x 4.6 mm 1.8µm
Pressure: 490 bar
N: 28669
Height:1.78
S/N: 43.61.5
2
2.5
3Column:150 x 4.6 mm 1.8µm
Pressure: 490 bar
N: 28669
Height:1.78
S/N: 43.61.5
2
2.5
3Column:150x4.6 mm
3.5µm
Pressure: 165 bar
N: 14862
Height:1 34 mAU1
1.5
2
2.5 Column:150x4.6 mm
3.5µm
Pressure: 165 bar
N: 14862
Height:1 34 mAU1
1.5
2
2.5 Column:150 x 4.6 mm 1.8µm
Pressure: 490 bar
N: 28669
Height:1.78
S/N: 43.61.5
2
2.5
3Column:150 x 4.6 mm 1.8µm
Pressure: 490 bar
N: 28669
Height:1.78
S/N: 43.61.5
2
2.5
3Column:150x4.6 mm
3.5µm
Pressure: 165 bar
N: 14862
Height:1 34 mAU1
1.5
2
2.5 Column:150x4.6 mm
3.5µm
Pressure: 165 bar
N: 14862
Height:1 34 mAU1
1.5
2
2.5 4.6 x 150, 3.5um165 barN = 14862Rs = 1.37
4.6 x 150, 1.8um490 barN = 28669 Rs = 1.80 (+57%)S/N: 43.6
Rs = 1.80
tr = 17.2 (1st epimer)
Nptp: 3 10-2
-1
-0.5
0
0.5
1S/N: 43.6
Rs = 1.80
tr = 17.2 (1st epimer)
Nptp: 3 10-2
-1
-0.5
0
0.5
1Height:1,34 mAU
S/N: 50.7
Rs = 1.37
tr = 15.3 min (1st epimer)
Nptp: 2 10-2 mAU
-1
-0.5
0
0.5
Height:1,34 mAU
S/N: 50.7
Rs = 1.37
tr = 15.3 min (1st epimer)
Nptp: 2 10-2 mAU
-1
-0.5
0
0.5
S/N: 43.6
Rs = 1.80
tr = 17.2 (1st epimer)
Nptp: 3 10-2
-1
-0.5
0
0.5
1S/N: 43.6
Rs = 1.80
tr = 17.2 (1st epimer)
Nptp: 3 10-2
-1
-0.5
0
0.5
1Height:1,34 mAU
S/N: 50.7
Rs = 1.37
tr = 15.3 min (1st epimer)
Nptp: 2 10-2 mAU
-1
-0.5
0
0.5
Height:1,34 mAU
S/N: 50.7
Rs = 1.37
tr = 15.3 min (1st epimer)
Nptp: 2 10-2 mAU
-1
-0.5
0
0.5
s
S/N = 50 Height = 1.34Noise = 20uAU
s ( )S/N = 44 Height = 1.80Noise = 30uAU
Page 49
min101
min101
min0 5 10 15 201
min0 5 10 15 201
min101
min101
min0 5 10 15 201
min0 5 10 15 201
1.8u Particles Reveal More Information and Improve Detection and IntegrationImprove Detection and Integration
7 Impurities7 Impurities4 Impurities
Customer Sample, Translation of Isocratic Impurity Methods, Zoom Critical Time Range (t = 7min)
7 Impurities
All 7 Baseline Separated!
7 Impurities
6 Not Baseline Separated!
4 Impurities
2 Not Baseline Separated!
4.6 x 150, 1.8μm490 barN = 28669
4.6 x 150, 3.5μm165 barN = 14862
4.6 x 150, 5μm93 barN = 7259
R_S = 1.80 (+57%)S/N = 44
R_S = 1.37S/N = 50
R_S = 1.15S/N = 42
Up to 60% higher resolutionwithout loss in sensitivity
Page 50
without loss in sensitivity
1.8u Particle Size Increases Gradient ResolutionIncreased Peak Capacity
Conditions for bothmAU 5Agilent 1100 Zorbax 2.1x150mm Conditions for both experiments
Pumps• Solvent A: H2O + 0.1% TFA
Solvent B: ACN + 0 1% TFA
mAU
250
300
350
17.1
36
15.9
67
59
15.5
55
15.0
65Agilent 1100 Zorbax 2.1x150mm SB C-18, 5µm
Solvent B: ACN + 0.1% TFA• Gradient: 10% to 95% ACN
in 40min, hold for 1min• Flow Rate: 0.4ml/minAutosamplers
100
150
200
15.4
1115.1
5Autosamplers• Injection volume: 3µlThermostatted Column
Comp.T t 50°C
Agilent 1200 RRLC250 2
mAU
079 Zorbax 2.1x150mm
SB C 18 1 8
min13 14 15 16 17 18
50
• Temperature: 50°CDetectors• DAD 2µl cell and 20Hz,
220nm, 150
200
250
2415.1
64 16.7
4615.4
7 215
.79015
. 0 SB C-18, 1.8µm
Ref: Appl. Note 5989-4506 by Edgar Naegele
50
100 15.3
2114
.937
14.7
87
Page 51
min13 14 15 16 17 18
Particle Size - More Peak Capacity with 1.8 um RRHT Columns - Peptide Map of BSARRHT Columns Peptide Map of BSA
673Peak CapacityColumn used
SB C18 2 1 150 1 8
mAU
40
50
Conditions: Columns: as listed, Mobile Phase: A:0.1% TFA in Water B:0.08% TFA in ACN Gradient: 5% B to 60%B in 25 min. Temperature: 80°C Sample: BSA tryptic digest
StartingPressure380 bar673SB-C18, 2.1x150mm, 1.8µ
10
20
30
35% More peak capacity, more resolution
380 bar
min14.5 15 15.5 16 16.5 17 17.5 18 18.5 19
-10
0
mAU
502SB-C18, 2.1x150mm, 3.5µ40
60
Less peak capacity less resolution
StartingPressure105 barPeak Capacity
0
20
Less peak capacity, less resolution
S ll ti l i h k t k it
Page 52
min15 16 17 18 19
-20 Smaller particle size = sharper peaks = greater peak capacity
Tools
Rapid Resolution Compendium CD-ROM
(Please Ask For A Copy – Pub No 5989-5130EN)
• Application Examples
• Technology Overview
S f C f• Rapid Resolution in The Scientific and Trade Press and Conference Posters
• System Configurator (guide to reduce dead volume and extra column effects)
• Method Translator Program
Page 53
Agilent HPLC Method TranslatorA Simple to Use Tool to Move Methods to RRHT
Mini-Demo Method Translator
p
Page 54
SummaryShorter 1 8 m col mns allo faster anal sis ith• Shorter 1.8um columns allow faster analysis with same resolving power as longer 5um and 3.5 um columns
• Change column diameter if necessary to allow for lower solvent consumption and higher accessible linear velocity
• Adjust the injection volume to maintain the same mass loading on the smaller column
• Adjust run time and gradient to allow for the shorter• Adjust run time and gradient to allow for the shorter column to match k* for longer column or to increase k* vs. longer column.S d th fl t til fl• Speed up the flow rate until max flow or max pressure or loss of resolution
Page 55
Questions?
Page 56
Appendix
Page 57
So What Bonded Phase Do I Choose – Column Choice vs. Sample Polarity More Retention Loss in High Aqueous MPs
Very Polar NonModerately PolarPola
Wettability with Low or No Organic for RPLC
Very Polar Non Polar
Moderately PolarPolar
Rx-SIL (HILIC)Zorbax SIL (HILIC)Z b NH2 (HILIC)
SB-CNSB-C3
SB – allBonus-RP
SB – allRx-C18E li XDB CNZorbax NH2 (HILIC)
300SB – all300Extend-C18SB-AqSB-CN
Eclipse XDB-CNSB-PhenylBonus-RPSB-C8SB-C18
Rx-C18Eclipse XDB-CNEclipse XDB-PhenylEclipse XDB-C8Eclipse XDB-C18
Eclipse XDB-CNEclipse XDB-Phenyl Eclipse Plus C18, C8Eclipse XDB-C8Eclipse XDB-C18
Bonus-RPSB C18Eclipse Plus C18, C8
Eclipse XDB C18Eclipse Plus C18, C8
pODS ClassicOriginal ODS
RPLC 0-40% organic (except Eclipse Plus
5-40%)
HILIC: 98-50% organic
RPLC 0 40% organic
RPLC 10-70% organic (except Eclipse Plus
5-70%)RPLC: 50-98% organic
NARP: MeOH < ACN < IPA
• These recommendations provide rough guidance on choosing a column based on sample type, with the examples we have seen. They are arranged in a general order of preference.
• Other columns will work in some cases and could provide a better separation
RPLC 0-40% organic )< CH2Cl2 < acetone
Page 58
• Other columns will work in some cases and could provide a better separation, depending on the actual analytes and excipients present in the sample.
Considerations for Method TransferSituation:You have an isocratic method developed for 4.6 mm x 150 mm, 5 um column. Run time is approximately 15 min.Pump: Agilent 1100 quaternaryA t l St d d t lAutosampler: Standard autosamplerTCC: 1100 standardDetector: 1100 DAD, max. data rate 20 Hz
T i l tti PW 0 05 iTypical setting, PW = 0.05 min.Flow Cell: 13 μL, 10 mm path lengthFlow Rate: 1.0 mL/min.Column temp 23ºCColumn temp. 23ºCGoals: Decrease run time and improve throughput (5X, if possible)
Save solvent usage and waste (implies smaller column id or shorter run at higher flow rate)
• Can anything be done to speed up these methods with existing equipment?
What modifications can be made and which are most important?
Page 59
• What modifications can be made and which are most important?
Assessing the Current Method and PerformanceHow does the current method perform?• Method type: assay (several analytes) or impurity method
(> 6 analytes at trace levels)M th d t I ti di t th d• Method type: Isocratic or gradient method
• Critical or limiting resolution? More than one critical pair?• Method parameters
– Column: length ID particle sizeColumn: length, ID, particle size– Column temperature– Type of column heater– Mobile phases used
B k– Backpressure– Injection volume– Injection precision (RSD)
• Instrumental Parameters– Injector type: autosampler, manual, loop size– Delay Volume (system volume for isocratic sep’ns)– Extracolumn Volume
Page 60
Isocratic method on Conventional ColumnTocopherols
Column: ZORBAX Eclipse XDB-C18Mobile Phase: 95% ACN: 5% WaterTemp: 23ºCRRHT
p
Conventional4.6 x 150 mm 5 μm
Flow Rate: 1 mL/minP = 37 bar
Temp: 23 CInjection volume: 1 uL
RRHT4.6 x 50 mm 1.8 μm
Flow Rate: 3 mL/minPressure = 229 bar
P = 37 barmAU
60
80
Rs ~ 4.4Rs ~ 5.2
20
40 1.7 min13.5 min
Sample: Vitamin E – α, β, γ-tocopherols in gel capE li XDB C18 i d fi t h i f th d
min0 2 4 6 8 10 12 140
Page 61
Eclipse XDB-C18 is a good first choice for many methods.
Efficiency Ranking of Various Column Geometries and Typical Backpressuresand Typical Backpressures
This RRHT column Replaces These Longer Columns50 mm, 1.8 μm 150 mm, 5 μm, 100 mm, 3.5 μm
100 mm, 1.8 μm 250 mm, 5 μm
Page 62
Where to begin? Assess Your Current MethodAssess your current method4.6 x 150 mm, 5 μm column
Questions to ask?What is the mobile phase composition?
1.5 mL/minRT last = 14 minutes
What is the current backpressure?Injection Volume?Data Rate/Peak Width?What is your limiting resolution with current method?What size column can deliver the
l ti d?resolution you need?Can your current instrument be used to apply the shorter column with smaller particle size?Which changes in method parameters are necessary and can you get the same or similar performance and results?
Page 63
Tocopherol method translation
4.6 x 150 mm, 5 µm XDB-C18Viscosity of 95:5 ACN/water at 23ºC is ~0.43 cp.
4.6 x 50 mm, 1.8 um RRHT XDB-C18Column length in shorter dimensions with 1.8 µm particles is 4.6 x 50 mm RRHT0.43 cp.
Flow Rate is 1 mL/minBackpressure is 37 barStandard flow cell (13 µL)
µ pAt 1 mL/min expected backpressure is 79 bar + ~10 bar (a/s and flow cell) or ~90 barExpected run time will be 1/3 of 14 minutes or 4.67 minutesStandard flow cell (13 µL)
Standard 0.17 mm tubing throughoutPeak Width required 0.1 min
4.67 minutesTry 3 mL/min for run time of 1/9 of 14 min. or 1.55 min.Predicted pressure is 238 bar
Response Time = 2 sec or Data Rate = 2.5 Hz is adequate
Limiting resolution will be approximately the same (4.4) or 4.4 x SQRT(13043/12077) = 4.2, IF no band broadening due to extracolumn volume or data rate.Standard DAD or MWD at fastest setting (20Standard DAD or MWD at fastest setting (20 Hz) with 0.17 mm id tubing adequate but not optimumChoose 0.12 mm i.d. tubing and 5 µL flow cell for better results
Page 64
for better results
Agilent Method Translator
Page 65
Adjust Flow resist. factor for 37- 5 = 32 bar 13 uL flow cell and
0.17 mm tubing
Effective N hurt by EC vol.
4.67 min run at 1 mL/min
For isocratic runs the 2nd row must be set to same %B as row 1
Page 66
5 uL flow cell and 0.12 mm id tubing
With 5 uL flow cell and 0.12 mm id tubing
Page 67
Adjust % max. pressure until desired flow
rate
Final method conditionsrate
Adjust to 3 mL/min
Click radioClick radio button to
allow % max pressure
adjustment
Page 68
Isocratic method on Conventional ColumnTocopherols
Column: ZORBAX Eclipse XDB-C18Mobile Phase: 95% ACN: 5% WaterTemp: 23ºCRRHT
p
Solvent used 15 mL
Conventional4.6 x 150 mm 5 μm
Flow Rate: 1 mL/minP = 37 bar
Temp: 23 CInjection volume: 1 uL
RRHT4.6 x 50 mm 1.8 μm
Flow Rate: 3 mL/minPressure = 229 bar
P = 37 barmAU
60
80
Rs ~ 4.4Rs ~ 5.2Solvent used 5.1 mL
20
40 1.7 min13.5 min
min0 2 4 6 8 10 12 140
Sample: Vitamin E – α, β, γ-tocopherols in gel capE li XDB C18 i d fi t h i f th d
Page 69
Eclipse XDB-C18 is a good first choice for many methods.
Choosing the flow cell size
Page 70
Flow Cells for RRLC
13µl Standard Flow Cell: For highest sensitivityHigh-demanding quantitative work, e.g. analytical method development, QA/QCy p , Q Q
2µl Micro Flow Cell: For highest resolutionUltra-fast semi-quantitative work, U t a ast se qua t tat e o ,e.g. Screening Experiments, HT LC/MS/UV
5µl Semi-micro Flow Cell: Best compromise of sensitivity and resolution
Dimension Sensitivity* Resolution*
13 µl / 10 mm +++ + p yFor good quantitative and qualitative results, e.g. Screening, HT LC/MS/UV, Early Formulation Studies
* D d l ti l diti d l di i
5 µl / 6 mm ++ ++2 µl / 3 mm + +++
Page 71
* Depends on analytical conditions and column dimension
Peak Width Setting – Response Time – Data Rate and Sensitivity Set at fastest rate and then y
Don‘t use for > 0.15 sec peak width!
increase rate until peak width
> 0.15 sec
> 0.3 sec
> 0.6 sec
Recommended settings in ultra-fast LC with 50% peak width between 0.15 and 0.6 sec
For 50% peak width between 0.6 and 1.2 sec
> 1.2 sec
> 3 sec
> 6 sec
Notes: • Noise level changes ~ proportional to the
square root of the change in data rate.• For optimum selectivity and sensitivity the
Peak Width should not be chosen smaller> 12 sec
> 24 sec
> 51 sec
Peak Width should not be chosen smaller than necessary.
• For 50% peak width between 0.3 and 0.6 seconds Peak Width of > 0.005 min is recommended, which correspondes to 40Hz
Peak Width = Peak Width at 50% Peak Height
data rate.• Only for peaks narrower than 0.3sec at half
height, Peak Width of > 0.0025min (80Hz data rate) should be used.
• For highest sensitivity in ultra fast LC the
Page 72
• For highest sensitivity in ultra-fast LC the slit can be increased to 8 or 16nm.
Detector Data Acquisition Rates – Effects on Peak Width Resolution and Peak Capacity in UFLC
Data Peak Resolution Peak 1.2
1.4
c
60
70
Peak Width, Resolution and Peak Capacity in UFLC
Rate Width Capacity
80 Hz 0.300 2.25 61
40 Hz 0.329 2.05 560.6
0.8
1
Peak
Wid
ths
/ se
c
20
30
40
50
Peak
Cap
acity
Peak Width [s]
Peak Capacity
20 Hz 0.416 1.71 44
10 Hz 0.666 1.17 28
5 Hz 1.236 0.67 16
0.2
0.4
0 20 40 60 80 100Data Rate [Hz]
0
10
80Hz versus 20Hz Data Rate:– 40% Peak Width => +40% Peak Capacity+ 30% Resolution => + 70% Apparent Column Efficiency08
1
1.2
1.4
dths
/ s
ec
1.5
2
2.5lu
tion
Peak Width [s] pp y
80Hz versus 10Hz Data Rate:– 120% Peak Width => +120% Peak Capacity+ 90% Resolution => +260% Apparent Column Efficiency0.2
0.4
0.6
0.8
0 20 40 60 80 100
Peak
Wid
0
0.5
1 Reso
l
Resolution (4,5)
Page 73
0 20 40 60 80 100Data Rate [Hz]
Data Rate and Slit Width Effect on S/N Ratio (DAD and MWD, VWD data rate)(DAD and MWD, VWD data rate)
S/N can be optimized with data rate Slit width can be increased to improve S/N (2 uL and 5 uL cells)
Page 74
to improve S/N (2 uL and 5 uL cells)