65
Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop Mário César dos Santos Caixeirinho Thesis to obtain the Master of Science Degree in Mechanical Engineering Supervisors: Prof. Jorge Manuel Mateus Martins Prof. Paulo José da Costa Branco Examination Committee Chairperson: Prof. Paulo Jorge Coelho Ramalho Oliveira Supervisor: Prof. Jorge Manuel Mateus Martins Member of the Committee: Prof. Carlos Baptista Cardeira June 2017

Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

Modelling and Optimization of a

Magnetorheological Fluid Actuator for Footdrop

Mário César dos Santos Caixeirinho

Thesis to obtain the Master of Science Degree in

Mechanical Engineering

Supervisors: Prof. Jorge Manuel Mateus Martins

Prof. Paulo José da Costa Branco

Examination Committee

Chairperson: Prof. Paulo Jorge Coelho Ramalho Oliveira

Supervisor: Prof. Jorge Manuel Mateus Martins

Member of the Committee: Prof. Carlos Baptista Cardeira

June 2017

Page 2: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an
Page 3: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

Abstract

The main aim of this thesis is structurally optimizing an orthosis applied to footdrop pathology. A

variable impedance actuator wearable leg orthosis is suggested composed of two springs and a

magnetorheological fluid linear actuator to restore patient’s gait. It is designed as the orthosis device

due to its variable damping actuation capacity. Both a hydraulic and an electromagnetic system model

are built to posteriori use in a structural and electromagnetic optimization. This is achieved using a multi-

objective genetic algorithm, minimizing the total volume and actuation power of the actuator.

A final actuator is obtained that is theoretically applicable to the problem. Future work and

improvements are suggested.

Keywords: Footdrop, Genetic Algorithm, Linear Actuator, Magnetorheological Fluid, Structural

Optimization.

Page 4: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an
Page 5: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

Resumo

O objetivo principal desta tese é otimizar a estrutura duma prótese com aplicação na patologia do

Pé Pendente. Uma prótese colocada no exterior da perna composta por um atuador de impedância

variável é sugerida de forma a permitir restaurar a locomoção do doente. Este atuador é composto por

duas molas e um amortecedor linear com um fluido magnetoreológico, sendo este usado devido à sua

capacidade de amortecimento variável. Ambos os modelos dos sistemas hidráulico e eletromagnético

são construídos para utilização posterior na otimização estrutural e eletromagnética. Isto é alcançado

usando como optimizador um algoritmo genético multi-objetivo, minimizando o volume total do atuador

e a potência do atuador.

Um atuador final teoricamente aplicável ao problema é obtido. Trabalho futuro e possíveis melhorias

são apresentados.

Palavras-Chave: Algoritmo Genético, Atuador Linear, Fluido Magnetoreológico, Otimização

Estrutural, Pé Pendente.

Page 6: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an
Page 7: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

vii

Acknowledgments

For all the extensive meetings and patience throughout every step of this work, I would like to thank

firstly my supervisors Professor Jorge Martins and Professor Paulo Branco. I know this wasn’t an easy

road but I’m thankful for all your support and motivation. For inserting me in your work groups to whom

I’m also grateful. Meeting and working with every one of you, made me learn a lot.

Secondly, to my course colleagues and PSEM team members, thank you. I don’t know how to repay

everything you all have done for me, helping me not losing sight of what’s important and comforting me

in the hard times.

Thanks to all my friends who provided me with a normal social life and thus some sanity.

To Nuno Moreira colleague and friend. You were always by my side. I’m sorry if I didn’t help you the

same way. Forever grateful for all you have taught me.

Finally, to my family who had no other choice but to endure. Your love was unconditional and I’m

sorry for all the times I was unfair and short tempered. This thesis is dedicated to you.

Page 8: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

viii

Page 9: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

ix

“All wisdom is from the Lord,

and with him it remains forever”

Sir 1, 1

Page 10: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

x

Page 11: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xi

Table of contents

1. Introduction ............................................................................................................................. 1

2. State of the art ......................................................................................................................... 3

2.1. Dropfoot pathology .......................................................................................................... 3

2.2. Magnetorheological fluid actuator.................................................................................... 5

2.3. NSGA-II Program in MATLAB ........................................................................................ 7

3. Design ...................................................................................................................................... 9

3.1. Geometry ......................................................................................................................... 9

3.2. Magnetic circuit .............................................................................................................. 10

3.3. Air-gap hydrodynamics .................................................................................................. 14

4. Validation ............................................................................................................................... 19

4.1. Tibialis anterior .............................................................................................................. 19

4.2. MRF actuator ................................................................................................................. 20

5. Optimization ........................................................................................................................... 25

5.1. Problem definition .......................................................................................................... 25

5.2. Parameter analysis ........................................................................................................ 27

6. Results ................................................................................................................................... 31

7. Discussion ............................................................................................................................. 39

7.1. Other remarks ................................................................................................................ 39

7.2. Battery usage ................................................................................................................. 40

7.3. Improvements ................................................................................................................ 40

8. Conclusions ........................................................................................................................... 41

9. References ............................................................................................................................ 43

Annex A ........................................................................................................................................... 45

Page 12: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xii

Page 13: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xiii

List of figures

Figure 2.1 Normal gait cycle and evolution states with sequential high-speed captures.................. 3

Figure 2.2 Normal ankle movement of angular position, angular velocity and joint torque .............. 4

Figure 2.3 Dropfoot gait deviations (A) foot-slap after CP and (B) toe drag during mid-swing ......... 4

Figure 2.4 Tibialis anterior replaced by a variable impedance actuator [3] ...................................... 5

Figure 2.5 Linear MRF actuator: (A) Structure [9] and (B) Force vs velocity [5] ............................... 6

Figure 2.6 MF fluid actuation: (A) Valve mode, (B) Direct-shear mode and (C) Squeeze mode ...... 7

Figure 3.1 Linear MRF Actuator ........................................................................................................ 9

Figure 3.2 Actuator geometric variables ........................................................................................... 9

Figure 3.3 Piston with coil and electric analogous of the actuator symmetric magnetic circuit ...... 10

Figure 3.4 B-H curve of MRF-132DG [18] ....................................................................................... 12

Figure 3.5 B-H curve of AISI 416 Annealed Stainless Steel [19] .................................................... 12

Figure 3.6 Static shear stress forces representation between fluid-fluid and fluid-structure .......... 16

Figure 3.7 MRF-132DG electromechanical characteristic curve [18] ............................................. 17

Figure 4.1 Various ankle angle samples from SimMechanics simulation ....................................... 19

Figure 4.2 Sample of the ankle measurements from SimMechanics simulation ............................ 19

Figure 4.3 Linear actuator measurements converted from ankle measurements .......................... 20

Figure 4.4 Geometry dimensions of the MRF linear actuator in [23] .............................................. 21

Figure 4.5 Static model validation ................................................................................................... 21

Figure 4.6 Dynamic model validation .............................................................................................. 22

Figure 4.7 B-H curve of MRF-122EG from [23] ............................................................................... 22

Figure 4.8 MRF-122EG electromechanical characteristic curve from [23] ..................................... 23

Figure 4.9 Geometry measurements of the MRF linear actuator in [3] ........................................... 23

Figure 4.10 Static validation of the model with [3] experimental data ............................................. 24

Figure 5.1 Best result of each population size after 150 generations ............................................. 28

Figure 5.2 Computational time versus population size ................................................................... 28

Figure 5.3 Varying mutation scale on a 1000 individuals’ population after 150 generations .......... 29

Figure 5.4 Mutation scale versus computational time ..................................................................... 30

Figure 6.1 Last Pareto Front ........................................................................................................... 31

Figure 6.2 Best optimized actuator for 110 N after 500 generations .............................................. 32

Figure 6.3 Volume convergence ..................................................................................................... 32

Figure 6.4 Power convergence ....................................................................................................... 33

Figure 6.5 Power volume product convergence.............................................................................. 33

Figure 6.6 Static simulation of the optimized actuator .................................................................... 34

Figure 6.7 Best Pareto Front volume solution after 500 generations.............................................. 34

Figure 6.8 Best Pareto Front power solution after 500 generations ............................................... 35

Figure 6.9 Best optimized actuator for 55 N after 500 generations ................................................ 35

Figure 6.10 Best optimized actuator for 40 N after 500 generations .............................................. 36

Figure 6.11 Best optimized actuator for 55 N with AWG 40 after 500 generations ........................ 36

Figure 6.12 Best optimized actuator for 55 N with AWG 18 after 500 generations ........................ 37

Page 14: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xiv

Figure 7.1 Schematic of piston’s flange future improvement. a) Round tip and b) Ramp tip ......... 40

Page 15: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xv

List of tables

Table 1 – Decision variables ........................................................................................................... 25

Table 2 – Computer system components specification .................................................................. 27

Page 16: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xvi

Page 17: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xvii

Abbreviations

AAFO Active Ankle-foot Orthosis

AFO Ankle-foot Orthosis

AWG American Wire Gauge

CD Controlled Dorsiflexion

CP Controlled Plantarflexion

EMO Evolutionary Multi-Objective Optimization

ER Electrorheological Fluid

GA Genetic Algorithm

MMF Magnetomotive Force

MR Magnetorheological

MRF Magnetorheological Fluid

NGPM A NSGA-II Program in MATLAB

NSGA Non-Dominated Sorting Genetic Algorithm

OAT One-at-a-time

PP Powered Plantarflexion

VIA Variable Impedance Actuator

Page 18: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xviii

Page 19: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xix

List of Symbols

Top/Bottom piston’s superficial area

Cross-sectional area of the coil copper wire

Cross-sectional area of the ferromagnetic steel where the magnetic field passes

, Cross-sectional area of the ferromagnetic steel in section 2

Magnetic density flux (magnetic induction field)

, Magnetic density flux (magnetic induction field) of the MR fluid

, Magnetic density flux (magnetic induction field) of the ferromagnetic steel

Piston inner radius

Gap width of the MRF actuator

Coil copper wire diameter

Cylinder wall thickness

Dynamic force of the MRF actuator

Force density of the MR fluid in the axis

Maximum force of the MRF actuator

Fluid-fluid force density on the MR fluid

Seal friction force of the MRF actuator

Static force of the MRF actuator

Fluid-structure force density on the MR fluid

ℎ Piston flange thickness

Magnetic field

, Magnetic field of the MR fluid

, Magnetic field of the ferromagnetic steel

Electric current

Maximum electric current of the MRF actuator

Number of the term in the sum expression

Piston height

Length of the coil copper wire

Length of the middle magnetic path

, Length of the middle magnetic path in section 2

Number of coil turns

Number of coil turns per column

Number of coil turns per row

Pressure of the MR fluid

Power of the MRF actuator

Maximum power of the MRF actuator

Flow rate of the MR fluid

Flow rate dislocated by the actuator moving piston

Axis of the cylindrical coordinate system

Page 20: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

xx

Piston outer radius

Piston guide radius

Electric resistance of coil copper wire

Magnetic reluctance

, Magnetic reluctance of the MR fluid

, Magnetic reluctance of ferromagnetic steel in section 1

, Magnetic reluctance of ferromagnetic steel in section 2

, Magnetic reluctance of ferromagnetic steel in section 3

Time

Actual functioning temperature of the coil

Reference working temperature of the coil

Maximum applied voltage of the MRF actuator

MR fluid velocity

First term of the MR fluid velocity in the axis

Second term of the MR fluid velocity in the axis

Velocity of the actuator piston

Volume of the MRF actuator

MR fluid velocity in the axis

MR fluid velocity in the axis

MR fluid velocity in the axis

Axis of the cylindrical coordinate system

Temperature coefficient of resistivity

MR fluid-structure static friction coefficient

Reference resistivity of copper

Electric resistivity of copper

Δ Dynamic pressure difference of the MR fluid

Δ Static pressure difference of the MR fluid

MR fluid viscosity

Axis of the cylindrical coordinate system

Magnetic permeability

, Magnetic permeability of the MR fluid

, Magnetic permeability of the ferromagnetic steel

MR fluid mass density

Magnetic normal stress

MR fluid shear stress by fluid-fluid friction

MR fluid shear stress by fluid-structure friction

Φ Magnetic flux

Page 21: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

1

1. Introduction

Nowadays research has evolved greatly in understanding the human body and how it works. There’s

no doubt about how perfectly every part of the body interacts. Although it is a complex system, it rarely

ever has problems in comparison.

Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an unusual

walking pattern. Or in medically terms it’s a reduced or even inexistence activity in the dorsiflexor

muscles affecting the human walking gait. Usually, people who live with this pathology correct their gait

by increasing the knee higher than normal when walking and slapping the foot afterwards on the ground.

Otherwise dragging the foot on the floor. Either way, this unusual gait, in time, damages the rest of the

body.

There are a broad range of different solutions on the market, orthosis or not. None of those solutions

are perfect. The least intrusive are ankle-foot orthoses (AFO). Unfortunately, most of them have a single

locking position or a constant damping coefficient, i.e. non-adaptive during gait, called passive orthoses.

These avoid some body wear but not all. Better solutions are adaptive orthoses that can replicate the

normal human gait called active ankle-foot orthosis (AAFO). Magnetorheological fluid (MRF) dampers

show great potential here.

These dampers are embedded with MR fluid and an electromagnetic system. MR fluids are liquid

solutions with micro magnetisable particles. When subjected to a magnetic field the magnetisable

particles in the fluid aggregate and align in the field direction, increasing the fluid viscosity where the

magnetic field passes. This property allows blocking movement in the presence of a high field passing

through the MRF actuator gap sustaining or damping the acting force.

So, this type of actuators are low power variable damping systems, controllable by applied current.

When applied to dropfoot pathology the variable damping coefficient permits to mimic some states of

the human gait. An actuator model and respective validation are presented.

A structural optimization based on the model is done, where the goal is obtaining the smallest

actuator possible and also the lowest power consumption to achieve the required force. Genetic

algorithm (GA) is chosen to accomplish this optimization.

GA is a computational optimization that mimics biological evolution, employing principles of natural

selection. An evolutionary multi-objective optimization (EMO) is used here to optimize two objective

functions. The algorithm code used is called NGPM (version 1.4) abbreviation of “A NSGA-II Program

in MATLAB” which allows parallel computation, reducing overall process time.

Achieving a wearable and suitable optimized MRF linear actuator orthosis application for dropfoot

pathology is the main goal of this thesis. Important results, conclusions, corrections and future work are

presented.

Page 22: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

2

Page 23: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

3

2. State of the art

In this section are presented the key concepts and all knowledge obtained from previous researches,

explaining where it begun, how it grew and what is being done today. Three major topics were

considered.

All needed concepts around the dropfoot pathology are presented, including different human gait

phases and where the MRF actuator will act.

Secondly, every aspect of the MRF actuator is presented. An important focus is made on the MR

fluids and how they work, since they are responsible for variable damping behavior in these actuators.

Finally, a brief explanation of the optimization algorithm used and some current structural

optimization applications.

2.1. Dropfoot pathology

The human gait is composed off two main phases shown in Figure 2.1, taken from [1, 2] for a 74 Kg

healthy male.

When the foot is on the ground it’s called the stance phase, or else the swing phase. There is also

an alternative designation by states as represented. State 1, represented from heel strike until complete

grounded foot plant, called Controlled Plantarflexion (CP). Second state, until foot is in neutral position,

with leg perpendicular to the floor and all toes on the ground. State 3, then ends on maximum ankle

angular position already with some heel off. These two states form the Controlled Dorsiflexion (CD).

Finally, Powered Plantarflexion (PP) occurs in state 4 with total foot plant lift, reaching toe-off. Foot

swing period happens in states 5 and 6.

Figure 2.1 Normal gait cycle and evolution states with sequential high-speed captures

Angular position, velocity and torque of the ankle during human gait states are shown in Figure 2.2

(data from [1] for a 74 Kg healthy male). Positive ankle angles are considered counter clockwise to the

images as convention.

Page 24: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

4

Figure 2.2 Normal ankle movement of angular position, angular velocity and joint torque

Foot lifts approximately 10° in the end of CD and -20° in the end of PP.

Dropfoot pathology, as already stated, affects human gait because individuals lack or have reduce

activity in the dorsiflexor muscle, better known as tibialis anterior. Two main problems arise: foot slap

and toe drag (Figure 2.3, taken from [3]).

Figure 2.3 Dropfoot gait deviations (A) foot-slap after CP and (B) toe drag during mid-swing

Page 25: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

5

Foot slap occurs after heel strike, due to swing phase attained momentum and body weight. Then

in CP the defective tibialis anterior cannot hold or help damping out the foot, resulting in the usual slap

sound heard from these patient’s gait.

Toe drag on the other hand happens during mid-swing phase which unbalances the body. Normally

the dorsiflexor muscle holds the foot up in the air during this phase.

The use of a variable damping actuator with some attached spring system, also known as a Variable

Impedance Actuator (VIA) is shown in Figure 2.4 and can mimic the behavior of the tibialis anterior.

Figure 2.4 Tibialis anterior replaced by a variable impedance actuator [3]

Two springs are required. One high stiffness spring connected in series with the variable damping

actuator and a low stiffness spring in parallel as shown in the previous figure.

The high stiffness spring and damper will only act in the CP state. This helps control foot damping,

avoiding foot-slap. In contrast, the low stiffness spring is always active supporting the foot weight,

avoiding toe drag.

A linear magnetorheological fluid (MRF) actuator will be used to act as the variable damping

actuator.

The objective is to use an AAFO, in this case a VIA, connected to the foot and leg, helping the

defective tibialis anterior.

2.2. Magnetorheological fluid actuator

As stated in Chapter 1, the goal is modelling and optimizing a linear MRF actuator for dropfoot

orthosis application. This actuator works using MR fluids, which have been studied since World War II.

Research started in 1939 with particle suspensions in low viscosity oils forming an “oil-occluding

fibrous mass” when subjected to an electric field [4], i.e. electrorheological fluids (ER fluids), which was

patented. Almost at the same time, Jacob Rabinow in 1940 at the Bureau of Standards discovered

magnetorheological fluids (MR fluids) and developed the first device applications for it.

Both ER and MR fluids are non-colloidal, i.e. they settle down with time if the fluid is not mixed in

time. The polarizable particles suspensions have a size of micrometers [5]. MR and ER fluids are mostly

different in magnetic saturation values than by the resistance force their polarizable particles can

produce [4].

Page 26: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

6

MR fluids are composed off three main components, the base fluid, magnetic metal particles and

stabilizing additives. Maximum fluid viscosity is due to magnetic saturation of the fluid metal particles

and minimum with the carrier fluid viscosity [6].

The base or carrier fluid normally can be of three different types, i.e. hydrocarbon oils, mineral oils

or silicon oils. This fluid carries the magnetic metal particles, lubricating and damping their movement.

Carrier fluid viscosity must not vary much with temperature so that the magnetorheological effect is only

due to the particles inside it. Higher carrier fluid viscosities are justified by higher head losses when no

magnetic field is present, which is undesirable for most applications [6].

Static and kinetic friction coefficients are similar for low magnetic fields. High magnetic fields show

in contrast a dissimilar dry friction behavior. Nevertheless, kinetic friction coefficient is always higher

than the static in these fluids [6]. Static shear resistance forces are consistent with values from known

magnetic tractive forces [4].

Heat transfer is promoted in the active region and has the same direction as the magnetic field lines.

Magnetic particles in the fluid agglomerate increasing the heat transfer and then when the structure

breaks down (magnetic field disactivated), there is a high increase in the heat transfered by particles

dispersion in the carrier fluid [7]. Eddy currents [8] and joule energy loss [5] in the fluid are negligible.

The increase of the superficial area of the actuator and number of active regions, decreases copper

losses [6].

Recently, MR fluids have been applied in many devices: linear dampers (Figure 2.5), rotary brakes,

vibration dampers, etc.

Figure 2.5 Linear MRF actuator: (A) Structure [9] and (B) Force vs velocity [5]

Page 27: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

7

There are three modes in which they can be applied in devices: valve mode, direct-shear mode and

squeeze mode [5], Figure 2.6 presents these modes (Adapted from [10]).

Figure 2.6 MF fluid actuation: (A) Valve mode, (B) Direct-shear mode and (C) Squeeze mode

High precision MRF actuators have many leakage problems, rubber O-rings are used to seal and

avoid fluid leakage [6]. MR seals are also used but only in low frequency or static system applications

[11].

Actuators with low nickel steel structures are better for minimizing hysteresis effect [6].

2.3. NSGA-II Program in MATLAB

NGPM is built by Lin [12] based on previous NSGA-II algorithm done by Deb [13]. The Non-

Dominated Sorting Genetic Algorithm (NSGA-II) starts by generating an offspring of size equal to the

initial randomly generated parent population through genetic operators (selection, crossover and

mutation). Then parent and offspring population are combined in sets of non-dominated fronts by fast

non-dominated sorting [14]. Individuals are chosen from best fronts to form the new parent population,

having the same initial parent population size. Process repeats until maximum number of generations

defined by the user is reached. Applications using this elitist and crowd comparison operator algorithm

are shown in optimization design of viscoelastic damping structures [15]. It can iterate problems with

many decision variables and is a tool commonly used in structural optimizations. Also, it doesn’t get

stuck on local minimums or maximums.

The solutions are given by a Pareto front (set of optimal solutions), instead of a single point solution,

this gives more options to the decision maker helping him/her decide what is the best solution.

Computational time can be high but it varies a lot from problem to problem and how the user sets the

algorithm parameters.

Page 28: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

8

Page 29: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

9

3. Design

3.1. Geometry

Most common MRF actuators are linear. Three important solid components define it: a cylinder, a

piston and a coil. In the cylinder, the piston will move vertically on its axis, as shown in Figure 3.1 from

[16]. MR fluid will surround the piston inside the cylinder. Finally, around the piston inner radius there is

a copper wire coil where external applied electric current activates the actuator. Piston outer radius ,

piston inner radius , piston flange thickness ℎ, gap width , cylinder wall thickness , piston height

and piston guide radius , define the actuator geometry (Figure 3.2).

Figure 3.1 Linear MRF Actuator

Figure 3.2 Actuator geometric variables

Page 30: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

10

3.2. Magnetic circuit

By passing an electric current throughout the insulated copper wire coil (left side of Figure 3.3) a

magnetic field is created which actuates the MRF actuator. Two holes on top of the piston work as entry

and exit points for the coil.

The applied magnetomotive force ( ) equal to , where is the number of turns, induces a

constant magnetic flux Φ that passes from the piston to the cylinder wall and back, through the MR

fluid, closing the magnetic circuit. All other magnetic flux leakages are neglected.

Considering magnetic reluctances as resistors, the as the electric potential difference and

Φ as the electric current, an electric analogous is shown (right side of Figure 3.3) for the actuator

symmetric magnetic circuit.

Figure 3.3 Piston with coil and electric analogous of the actuator symmetric magnetic circuit

Number of coil turns is estimated knowing the coil housing size and the copper wire diameter :

= = − 2ℎ

− −

(3.1)

In Equation 3.1, is the number of coil turns per column and the number of coil turns per row.

The clearance is a measure given so that the copper wires can exit the coil housing without passing

through the air-gap.

Due to wire heating, copper electric resistance changes as been stated by the Pouillet’s Law:

= ()

(3.2)

,

,

,

,

,

,

Page 31: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

11

Here, is the copper electric resistivity function of the coil working temperature , the length

of it and its cross-sectional area. From [17] a copper wire length approximation is given by:

= 2 + ( − 1)

(3.3)

Electric resistivity changes with temperature so a linear approximation is used near a reference

working temperature equal to 20ºC as stated by Equation 3.4. Here, the reference resistivity at

= equal to 1.72 x 10-8 Ω.m and 3.93 x 10-3 ºC-1 the temperature coefficient of resistivity , both for

annealed copper.

()= [1+ (− )] (3.4)

Nevertheless, for simplification purposes no significant change with varying temperature in the coil

is considered in thus Equation 3.2 reduces to:

= 8 + ∑ ( − 1)

(3.5)

Magnetic reluctances vary with the type of ferromagnetic material where the magnetic density

flux passes, due to its magnetic permeability . Magnetic permeability is not constant for all

values and is defined by the derivative at each point of the material B-H curve. The electric current

passing through the copper wire of the coil generates a magnetic field , which in return due to the

material’s magnetic permeability surrounding it, creates a magnetic induction field . This relation

is seen in Figure 3.4, for the magnetorheological fluid MRF-132DG, and in Figure 3.5 for AISI 416

Annealed Stainless Steel, both used in this work.

Page 32: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

12

Figure 3.4 B-H curve of MRF-132DG [18]

Figure 3.5 B-H curve of AISI 416 Annealed Stainless Steel [19]

Due to the observed nonlinearities (magnetic saturation) on the B-H curves, only the linear working

region is considered. Hence, magnetic permeability of the two magnetic materials is considered

constant, for , ≤ 10 A.m-1 and , ≤ 10

A.m-1.

= (3.6)

Each magnetic reluctance can be calculated as

=

(3.7)

Page 33: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

13

Where is the length of the magnetic path passing through the middle of the material ( was

assumed uniformly distributed) and is the cross-sectional area normal to it. So, for each section of

the magnetic path in Figure 3.3, one has:

, = − ℎ

, (3.8)

, = − ℎ

,[(+ + ) − (+ )]

(3.9)

Also, because the magnetic path that passes through the MR fluid is very small, the area of

associated magnetic reluctance is considered approximately constant and equal to:

, =

,2ℎ (3.10)

For a differential element of magnetic reluctance ,, the cross sectional area , varies with

, [20] as:

, =1

,

,,

(3.11)

Observing that , is the cross-sectional area of a disk and neglecting the initial magnetic path

,, = ,,

= 1

2ℎ,

, =1

2ℎln

(3.12)

Then:

, =1

2ℎ,ln

(3.13)

Finally, Equation 3.14 describes the relation between the produced magnetic flux Φ, the applied

and the magnetic reluctances of the actuator.

Φ =

,+ 2, + , + 2, (3.14)

Magnetic flux is constant with constant current applied on the coil and because , is considered

approximately constant, they are related by the following magnetic relation:

Page 34: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

14

Φ = (3.15)

The magnetic field passing on the MR fluid, ,, can be obtained through the magnetic induction

field using Equation 3.6.

, =Φ

,2ℎ (3.16)

If , ≫ ,, , is reduced to:

, ≈

2 (3.17)

3.3. Air-gap hydrodynamics

The gap between the piston and the inner cylinder wall, called here the “active” region (Figure 3.1),

is analysed next to better understand the MRF actuator hydrodynamics.

Considering the MR fluid a homogeneous and incompressible fluid:

∇∙ = 0 (3.18)

Because it is considered to move only in the direction (Figure 3.2), fluid velocities and are

neglected:

= = 0 (3.19)

Hence:

= 0 (3.20)

That means fluid velocity on the “active” region is constant in the direction.

Considering symmetry in relation to the axis, then:

∂∂

= 0 (3.21)

Using Navier-Stokes equations in cylindrical coordinates (see Annex A) and approximating the MR

fluid to a constant mass density and viscosity fluid:

Page 35: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

15

= −

+

∂∂

(3.22)

For steady-state regime the following equation for the hydrodynamics in the “active” region appears:

+

∂∂

= 0 (3.23)

This means density forces due to MR fluid acceleration

plus pressure

are equal to the

summation of other force densities , which will be explained further, with forces due to head losses

. MRF-132DG used here has a 0.112 (±0.02) Pa.s viscosity [18]. A final equation for the

hydrodynamics of the “active” region is reached:

∂∂

=1

(3.24)

The MR fluid velocity profile in the “active” region can then be estimated solving Equation 3.24,

resulting in:

()=1

2

+ + (3.25)

Then, using the respective boundary conditions, ( = )=

( = + )= 0, constants and can now

be obtained. Therefore:

()= ()+ () (3.26)

With

()=(+ − ) (3.27)

And

()=1

2

− [

− (2+ ) + (+ )] (3.28)

Page 36: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

16

To obtain the pressure gradient /, a fluid flow rate analysis to the “active” region is done:

= ()

(3.29)

Knowing that must be equal to the flow rate dislocated by the moving piston, , with speed

inside the actuator (no fluid accumulation), then:

= = 2 ()

= (3.30)

Where = ( −

).

Finally, Equation 3.31 is reached

= + 4

3 − 3− − 3

( + 2) (3.31)

The pressure gradient

is then given by the one undefined force density plus the gap head

losses between the piston and the cylinder wall.

Pressure drop inside the cylinder occurs due to two major forces independent of fluid velocity

(neglecting head losses): one by shear stress (fluid-fluid friction) and another by fluid-structure

friction , both on the active region, as indicated in Figure 3.6. MR fluid weight is considered

negligible in the “active region” in comparison with these forces since the volume is very small.

Figure 3.6 Static shear stress forces representation between fluid-fluid and fluid-structure

Page 37: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

17

For a static behaviour (actuator blocking the total force) as magnetic field in the air-gap increases,

magnetic particles in the fluid connect more strongly to each other than to the piston or cylinder walls.

Load is supported by since it is lower than . After a certain point of magnetic field increase,

ferromagnetic particles start to grab to the walls more than to each other, > , and

becomes responsible for supporting the load.

Force density done by the MR fluid as seen is then a minimum between fluid and wall shear stress

friction forces:

= (,) (3.32)

Shear stress between the magnetic particles on the fluid when a magnetic field is present is given

by fluid data sheet [18]. The MRF-132DG electromechanical characteristic used in this work is

shown in Figure 3.7.

Figure 3.7 MRF-132DG electromechanical characteristic curve [18]

Comparing with Figure 3.4, in linear regime ( ≤ 10 A/m) it can be stated that:

= 0.3, (3.33)

Using the area parallel to the flow in the middle of actuator active region and the respective volume

of fluid:

=[2ℎ(2+ )]

2ℎ[(+ ) − ]=

(3.34)

Page 38: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

18

The electromagnetic normal stress that attracts the magnetic particles of the fluid into the walls is

given by:

≈1

2,,

(3.35)

Because , ≪ ,. Then wall shear stress is obtained by multiplying by the static

friction coefficient , since is normal to . The largest force acts on the minimum area, so the

piston superficial area on the active region is used instead of the cylinder superficial area. Dividing it by

the volume of fluid passing thought the active region:

=2[2ℎ(2+ )]

2ℎ[(+ ) − ]=,,

(3.36)

Equation 3.32 is then equal to:

= 0.3,

,,,

(3.37)

From Equation 3.31, in static behaviour for constant parameters

is constant and equal to the static

pressure difference ∆ of the MR fluid between the top and bottom of the piston inside the actuator

(no fluid velocity). But since only acts in the “active” region:

Δp = 2ℎ× 0.3,

,,,

(3.38)

In dynamic behaviour, since there are no aligned ferromagnetic particles blocking the whole gap

width , wall shear stress is negligible. So, dynamic pressure difference ∆ of the MR fluid

from Equation 3.31 is given by two components plus head losses. The length where the pressure

difference acts is different for the two components, for head losses and 2ℎ for .

Δp = 2ℎ0.3,

+ 4

3 − 3− − 3

( + 2) (3.39)

The damper strength is known by relating both pressure differences with the moving piston area due

to applied exterior strength plus an added seal friction force . Finally:

= ( −

)∆ + (3.40)

= ( −

)∆ + (3.41)

Page 39: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

19

4. Validation

4.1. Tibialis anterior

Based on a previous modelling work by Geyer [21], was possible to replicate human walk in [22]

using Matlab, Simulink and SimMechanics environment. Simulation samples of the ankle angle were

taken during 10 seconds to reach permanent regime and disregard initial transient results (Figure 4.1).

A sample, after initial transient regime of the walking pattern, was chosen to compare with the known

real data (Figure 2.2). Figure 4.2 shows the simulation measurements of the ankle angle, angular

velocity and torque.

Figure 4.1 Various ankle angle samples from SimMechanics simulation

Figure 4.2 Sample of the ankle measurements from SimMechanics simulation

Page 40: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

20

Only the angular velocity has different values but similar in behavior compared to Figure 4.1. Since

the proposed linear MRF actuator is going to be attached to the leg of the patient, vertically below the

knee, a transformation of this angular measurements is needed to find the linear displacement, velocity

and acting force. After transformation (Figure 4.3), linear displacement and velocity are shown to have

similar behavior to the ankle angular measurements, as expected.

Figure 4.3 Linear actuator measurements converted from ankle measurements

Linear force presented is only related to the tibialis anterior, instead of the whole forces acting on

the human ankle. This is the only force that is important to replicate in the actuator. Tibialis anterior force

exerted on the actuator is approximately 110 N at the end of the CP state, representing the maximum

force the actuator must block. Also, maximum linear velocity achieved by the actuator is around 400

mm/s and the total displacement 30 mm.

4.2. MRF actuator

Previous experimental data from works with MRF linear actuator is used to validate designed model.

First comparison is between the static model and the Costa’s actuator static experimental data [23]

(Figure 4.5). Geometry dimensions of the actuator are shown in Figure 4.4. Actuator contains MRF-

122EG with a 2380 Kg/m3 density and a 0.07(±0.02) Pa.s viscosity . Static tests were made by

finding the current that makes the actuator hold the weight without the piston falling. Weights were slowly

incremented from 1 until 25 Kg.

Page 41: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

21

39

20

10

2

30

2

10

194

Figure 4.4 Geometry dimensions of the MRF linear actuator in [23]

Figure 4.5 Static model validation

Page 42: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

22

Figure 4.6 Dynamic model validation

Model fits the experimental points with a static friction coefficient of 2.4 on Equation 3.36, between

the fluid and the piston flange wall. This value is high and comparable to silicon rubber which has also

a friction coefficient higher than unit. In the presence of high magnetic fields the MR fluid has a solid-

viscous behavior justifying this value. Without current, experimental data shows a 15 N seal friction force

that was added to the model.

The dynamic model fits poorly to the Costa’s actuator dynamic experimental data [23] (Figure 4.6).

A constant offset is observed which suggests a missing load in the model.

Used magnetic permeability and electromechanical characteristic of the fluid are presented in Figure

4.7 and Figure 4.8, respectively.

Figure 4.7 B-H curve of MRF-122EG from [23]

Page 43: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

23

Figure 4.8 MRF-122EG electromechanical characteristic curve from [23]

A smaller actuator was built by Domingues capable of carrying 0.5 Kg with an actuation of 0.8 A [3]

(Figure 4.10). The structure for this prototype was based on the previously presented Costa prototype.

Built model is represented in Figure 4.9. The piston is movable by thin string lines attached to its tips,

so the piston guide radius is considered approximately null. MRF-132DG was the chosen fluid, with a

density of 2950-3150 Kg/m3 [18]. Figures 3.4 and 3.7, in the previous chapter, show fluid B-H curve and

electromechanical characteristic curve, respectively.

4.5

1.5

0.8

0.5

16.6

0.25

0.5

≈ 240

Figure 4.9 Geometry measurements of the MRF linear actuator in [3]

Page 44: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

24

Figure 4.10 Static validation of the model with [3] experimental data

Since no seal friction force was measured, none was introduced in this model. Saturation of the

model around 0.35 A was due to steel magnetic saturation. The steel of the actuator was not specified,

so AISI 416, already stated in chapter 3, was used instead to compare.

Page 45: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

25

5. Optimization

Since the actuator will be used on a patient, the weight is an important aspect that must be

addressed. Also, the additional actuation system must be well thought because it requires a low power

actuation. Low power actuation means smaller and lighter batteries, since the patient must also carry

them. All this conditions justify a multi-objective structural optimization to the actuator explained here.

Nondominated Sorting Genetic Algorithm (NSGA) is a multi-objective genetic algorithm with built in

“binary tournament selection”, “intermediate crossover” and “Gaussian mutation” [12].

“Binary tournament selection” picks pairs of random individuals (solutions), then transforms every

continuous or discrete solution to a corresponding binary value, allowing optimization with both

continuous and discrete decision variables. “Tournaments” between each pair of individuals allow

finding individuals with the best chromosomes for posteriori crossover.

“Intermediate crossover” also known as arithmetic crossover takes pairs of the best individuals from

the parent population and generates one pair of offspring per pair of parents. These offspring are created

using a ratio of random chromosomes from both the parents and building each offspring with these

randomly chosen ratio of chromosomes.

“Gaussian mutation” as the name suggests mutates the offspring population genes by a random unit

Gaussian distribution using a controlled standard deviation specified by the user in the mutation scale

parameter. NSGA also allows for mutation shrinkage (specified by user) controlling the rate at which the

average amount of mutation decreases along the number of generations. This is normally used in

problems that do not get stuck on local solutions, helping to improve the convergence speed.

Software changeable parameters are population size, number of generations, crossover ratio [0,1],

mutation scale (standard deviation of the random mutation number [0,1]) and mutation shrink parameter

(shrinkage rate [0.5,1]).

5.1. Problem definition

Two objective functions, nine constraints and seven decision variables (Table 1) with a top and a

lower boundary for each, formulate this optimization problem.

Table 1 – Decision variables

Piston outer radius

Piston inner radius

Piston flange thickness

Gap width

Piston height

Cylinder wall thickness

Maximum electric current

Minimizing total volume of the actuator (Equation 5.1) and power (Equation 5.2) are the objective

functions. Copper wire electric resistance is obtained from Equation 3.5.

Page 46: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

26

To preserve the normal shape of the actuator two geometric constraints are needed (Equation 5.3

and 5.4). Last one imposes a coil housing around the piston inner radius (see Equation 3.1). Also, from

Costa’s actuator a ratio between flange thickness ℎ and gap lower than 5 is used to avoid probable

gap fluid strangulation (Equation 5.5).

Nonlinear magnetic regime (saturation) is avoided in each section of the actuator (Figure 3.3) by

limiting the maximum magnetic field passing through the MR fluid using Equation 3.16 (Equation 5.6)

and the ferromagnetic steel (Equations 5.7, 5.8 and 5.9).

Maximum (power on) and minimum (power off) force constraints of the linear MRF actuator, are

given by Equation 5.10 and Equation 5.11, respectively (see Equation 3.40 and 3.41). Figure 4.3 gives

maximum force and maximum linear velocity for minimum force constraints. The dynamic force

fixed to 100 N was found to be reasonable for plausible gap size results, since the created model did

not explain totally the dynamic behavior. This value was used although it has no practical sense.

Lastly, decision variables lower and upper bounds are shown in Equations 5.12-18.

The optimization problem is formulated as:

Minimize = (+ + ) (5.1)

Minimize = (5.2)

Subject to > 2 (5.3)

≥ 1 (5.4)

≤ 5 (5.5)

, ≤ 100 kA/m (5.6)

, ≤ 1 kA/m (5.7)

,≤ 1 kA/m (5.8)

,[()()]

≤ 1 kA/m (5.9)

≥ 110 N (5.10)

(= 0, = 400 )≤ 100 N (5.11)

Bounded by 1 ≤ ≤ 50 (5.12)

1 ≤ ≤ 50 (5.13)

0.1 ≤ ≤ 30 (5.14)

0.1 ≤ ≤ 10 (5.15)

1 ≤ ≤ 100 (5.16)

0.1 ≤ ≤ 10 (5.17)

0.01 ≤ ≤ 5 (5.18)

Page 47: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

27

Seal forces are neglected. Copper wire diameter is set to 0.321 mm (AWG 28), the smallest

diameter of wire available easily hand turned into a coil around the piston. Clearance between the coil

end and the flange end of the piston is set to 1 mm (see Equation 3.1). Although not represented in the

following result schematics, the piston guide radius is also fixed and has a value of 0.5 mm, equal to

a thin string line radius. Piston course inside the actuator is not considered in the optimization because

it is constant. It was verified to be around 30 mm from Figure 4.3 and added to every actuator’s

schematic of the optimization results. Finally, the coefficient of friction is set to 2.4 as concluded in

Chapter 4.

5.2. Parameter analysis

To understand and reach the best results without using too much processing time a parameter

analysis is needed.

All the results were obtained running on a computer with the specifications in Table 2. It was found

that early MATLAB versions already have parallel computing inside its functions, so although specific

functions from older versions exist in the NSGA algorithm to force parallel computing, they are useless.

Table 2 – Computer system components specification

Processor Intel® Core™ i5-4430 CPU @ 3.00GHz

Installed memory (RAM) 8,00 GB

Operating System Windows 10 x64

All optimizations end when they reach a stablished number of generations. Initially the population

size must be fixed. To find out the adequate population size, many optimizations are done varying this

parameter and looking at the best result after 150 generations. This technique is called one-at-a-time

(OAT), varying one parameter and fixing all the others. Figure 5.1 shows this evolution. With increased

population size, so the computational time increases exponentially (Figure 5.2). Smaller populations are

better for this reason. Best results converge to a similar solution around 1000 individuals, so this is the

population size used from here on.

NSGA algorithm is implemented with “intermediate crossover” and the crossover ratio is a parameter

established by the user. The crossover ratio is set to 0.9 to use as much genetic background as possible

from the parents.

Page 48: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

28

Pow

er

* V

olu

me [W

.mm

3]

Figure 5.1 Best result of each population size after 150 generations

Tim

e [m

inute

s]

Figure 5.2 Computational time versus population size

Page 49: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

29

To improve solutions and avoid local minimums, after crossover there is some offspring genetic

data that suffers mutation, i.e. some genes are generated randomly in the offspring after the crossover.

The percentage of genetic data mutated in the new offspring is controlled by the mutation scale

parameter as stated. After some optimizations varying this parameter, the best percentage that allows

arising to the best solutions without losing much processing time and getting stuck on local solutions is

found. Figure 5.3 shows the best, worst and mean solution of a population with varying mutation scale

parameter. 5% is clearly the best value, although an increase in computational time is seen in Figure

5.4.

Another mutation parameter is the mutation shrinkage, set to zero until now. As the name suggests,

decreases the mutation percentage from generation to generation. This allows for rapid convergence,

but also getting stuck on local minimums if too high, so the value of this parameter must be well thought.

It varies between 0 and 1, but common values suggested by the manual are between 0.5 to 1, so 0.5 is

chosen.

Pow

er

* V

olu

me [W

.mm

3]

Figure 5.3 Varying mutation scale on a 1000 individuals’ population after 150 generations

Page 50: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

30

Tim

e [m

inute

s]

Figure 5.4 Mutation scale versus computational time

Page 51: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

31

6. Results

As stated in Chapter 5, the goal is to optimize an actuator that can damp and stop a maximum load

of 110 N. 500 generations are used as the stopping criteria for the NSGA algorithm, since there’s no

other stopping criteria implemented.

Figure 6.1 presents the first optimization with all the parameters set. A Pareto Front formed by all

the individuals (solutions) of the population after the 500 generations is obtained. During this process

the solutions converge to the origin, because the objective functions are being minimized, creating a

curve like the one represented.

The best solution for this Pareto Front is given by the minimum product between objective functions.

Figure 6.2 shows a schematic of the best solution size and its respective measures and characteristics

alongside. Proved convergence of the best Pareto Front solution’s volume, power and product of these

two, in every generation is showed in Figure 6.3, 6.4 and 6.5, respectively.

A static simulation of the optimized actuator is presented in Figure 6.6.

Volume 104 [mm3]

0 1 2 3 4 5 6 7 8 90

0.2

0.4

0.6

0.8

1

1.2

500 Generations1000 Individuals

Least Volume

Least Power

Best Solution

Figure 6.1 Last Pareto Front

Page 52: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

32

13.86

6.08

2.66

0.54

8.29

1.19

0.59

0.56

1.06

162

110.32

98.85

5.93 x 103

Figure 6.2 Best optimized actuator for 110 N after 500 generations

Generations0 50 100 150 200 250 300 350 400 450 500

0

1

2

3

4

5

Figure 6.3 Volume convergence

Page 53: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

33

Figure 6.4 Power convergence

Generations0 50 100 150 200 250 300 350 400 450 500

0

5

10

15

20

Figure 6.5 Power volume product convergence

Pow

er

[W]

Page 54: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

34

Current [A]0 0.1 0.2 0.3 0.4 0.5 0.6 0.7

0

50

100

150

Figure 6.6 Static simulation of the optimized actuator

Figure 6.7 and 6.8 present the best volume solution and the best power solution from the last

generated Pareto Front, respectively.

12.81

6.06

2.59

0.53

7.21

1.12

1.07

1.00

1.08

96

110.60

79.12

4.73 x 103

Figure 6.7 Best Pareto Front volume solution after 500 generations

Page 55: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

35

24.02

8.88

6.91

1.46

29.04

3.20

0.09

0.01

1.93

2021

117.54

99.10

7.51 x 105

Figure 6.8 Best Pareto Front power solution after 500 generations

Optionally to a single actuator, two or three actuators with less damping capacity can be set to work

in parallel, decreasing the volume and actuation power. To test this hypothesis, Figure 6.9 and 6.10

present the best volume solution of an optimized actuator with a maximum strength of 55 N and another

of 40 N, respectively.

12.02

4.78

1.33

0.39

3.41

0.57

1.05

1.70

0.62

36

58.92

79.59

1.81 x 103

Figure 6.9 Best optimized actuator for 55 N after 500 generations

Page 56: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

36

8.54

3.88

1.61

0.33

4.33

0.88

0.54

1.60

0.34

30

40.03

59.56

1.30 x 103

Figure 6.10 Best optimized actuator for 40 N after 500 generations

Changing coil wire diameter can also help in reaching smaller volumes and low power actuators.

Previous optimizations have an AWG 28 copper wire diameter (0.321 mm). In Figure 6.11 the best

volume solution for an optimized actuator with maximum strength of 55 N and an AWG 18 copper wire

diameter (1.024 mm) is presented. Also, an AWG 40 copper wire diameter (0.0799 mm) optimized

actuator is presented in Figure 6.12.

9.07

5.24

1.46

0.36

3.85

0.79

2.05

0.20

10.42

374

55.85

48.24

1.26 x 103

Figure 6.11 Best optimized actuator for 55 N with AWG 40 after 500 generations

Page 57: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

37

13.36

4.71

1.73

0.53

7.73

0.79

0.20

2.93

0.07

24

55.08

96.53

5.23 x 103

Figure 6.12 Best optimized actuator for 55 N with AWG 18 after 500 generations

Page 58: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

38

Page 59: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

39

7. Discussion

Results prove the methodology used and how it’s possible to optimize similar linear MRF actuators.

A linear MRF cylindric actuator that can damp a maximum force of 110 N was initially obtained. In Figure

6.2 was presented the new 110 N actuator which can be compared with the initial 250 N actuator (Figure

4.4). Unfortunately, the optimized actuator was still too large in diameter for a leg orthosis application.

So, other solutions were searched and presented.

One first solution was looking in the last Pareto Front population for an individual with the smallest

volume, increasing in return, the actuation power. Pareto Front’s present a set of solutions, instead of

only a single solution, this allows the decision maker to choose what is best for the application in hand.

So, actuator in Figure 6.7 with a total diameter of 28,92 mm and a maximum current of 1 A, is the

smallest actuator of the last population, but it is still too big for the needed application.

Another possible solution was dividing the total strength in parallel in two or three actuators. An

actuator with 55 N of damping force was optimized and the best volume solution presented in Figure

6.9. This actuator has a total diameter of 25.96 mm and a maximum current input of 1.7 A. The same

was made for a 40 N actuator (Figure 6.10), with a total diameter of 19.5 mm reached and a maximum

current of 1.6 A. These solutions approach the plausible maximum diameter size of 25 mm, although

the power increase. Because the difference from two actuators in parallel to three wasn’t substantial,

the two parallel actuators are taken as the better solution.

Finally, the smallest wire possible (AWG 40) for two actuators in parallel was used, expecting a high

decrease in volume. From this optimization (Figure 6.12) a total actuator diameter of 20.44 mm and a

maximum current of 0.2 A was obtained. In this case, although current and diameter values decreased,

the required voltage increased.

Two actuators solution with AWG 28 coil wire is thus pointed out for future improvements,

recognizing that a set of new research lines are specified in the end of this thesis.

7.1. Other remarks

From the optimization point of view and as foreseen, it is important to notice that with decreasing

volume the actuator input power increases significantly. The inverse is also verified and decisions

variables boundaries are reached, like the lowest permittable current for example (Figure 6.8).

Also, important to mention that steel sectional areas stay practically equal to each other’s, which

proves the optimizer is working well avoiding a section magnetic saturation prior to other sections.

Two main constraints control the optimization process. Equation 5.10 forces the optimizer to close

the gap and in contrast Equation 5.11 tries to open it. This is because not only the actuator must work

with acting current, but it must also let the piston move freely without much friction when no current is

applied.

Page 60: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

40

7.2. Battery usage

The goal of this application is a mobile orthosis, so the battery usage must be thought of. From a

simple market study, lithium ion batteries for cellphones are the most used for mobile applications. This

type of battery has around 3.6 V and 3 Ah of capacity each. Other type of batteries have less capacity

and occupy much more space in comparison, to give the same energy.

On average a person walks 7000 steps a day, equivalent to approximately 6 kilometers [24]. Each

step takes about 0.06 seconds during CP (Figure 2.1). For simplification, it’s considered that the actuator

is on maximum current throughout the CP state, which is around 5.1% of the step duration. Considering

70% of battery efficiency due to external factors that may affect the battery performance, the total

number of steps until battery is totally drained, is given by:

=

×× ×

=3×3600

2×1,7×0,06×0.7 = 51 941 ~ 1

It is proven that two actuators can work a full week before the battery has to be recharged again,

by this simple battery life calculation.

7.3. Improvements

An improvement is the addiction of a non-magnetic material in the tip of the piston’s flange to form

a cork like behavior when the magnetic field acts on the MR fluid. Figure 7.1 presents the schematic for

two options of tips. This improvement is expected to increase greatly the damping force of the respective

actuator, because the MR fluid is forced to go through a gradually smaller gap where the next set of

aligned ferromagnetic particles are even more connected than the ones before. The tip must be of non-

magnetic material because otherwise the magnetic flux Φ is strangulated and the fluid loses

connection strength between particles.

Figure 7.1 Schematic of piston’s flange future improvement. a) Round tip and b) Ramp tip

Page 61: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

41

8. Conclusions

This thesis proves that a linear MRF actuator model can be structurally optimized to work as an

orthosis for footdrop pathology patients. The best solution was concluded to be two actuators working

in parallel, having approximately 25 mm of exterior diameter and with a AWG 28 coil wire. One cellphone

lithium ion battery per actuator should be enough to power up the actuators for a full work week.

A validated static model of the actuator has been built and used successfully in the optimization

algorithm. The dynamic model although built and implemented, presents incoherent results when

compared to real experimental data. Future work should focus on correcting the dynamic model by firstly

modelling the actuator without actuation. Both dynamic and static models were created based on a

magnetic and a hydrodynamic air-gap analysis of the actuator.

Constructing and comparing the optimized actuator real data to the model should be the second

priority in future work, for both static and dynamic models. Afterwards, a non-magnetic tip like the ones

presented in the previous chapters should be added to close reinforce the gap strength and test the

desired cork behavior in the MR fluid, repeating the same tests. Other future improvements should be

building a model for low velocities studies, improving actual model robustness.

The best individuals of the Pareto Front after some generations have minimum volume and power,

but as seen the user may want to improve the volume increasing the power of the actuator. In these

cases, because the copper wire diameter is fixed, the temperature increases in the actuator due to Joule

effect and may influence the decision maker to choose bad solutions because the optimizer doesn’t

account for heat generation. Two improvement options arise, or a temperature model is built or the

copper wire diameter is set in the optimization as a discrete decision variable. Both can also work

together instead of implementing just one of the options.

Page 62: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

42

Page 63: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

43

9. References

1. Winter, D.A., Biomechanics & Motor Control of Human Gait. 2nd ed. 1991. 2. Winter, D.A., Biomechanics and Motor Control of Human Movement. 3rd ed. 2004. 3. Domingues, A.R.C. and J.M.M.s. Martins, On the development of an active soft second-skin

orthotic for dropfoot patients. 2014, UTL. Instituto Superior Técnico: Lisboa. 4. Winslow, W.M., Induced Fibration of Suspensions. Journal of Applied Physics, 1949: p. 1137-

1140. 5. Carlson, J.D., D.M. Catanzarite, and K.A. St. Clair, Commercial Magneto-Rheological Fuid

Devices. International Journal of Modern Physics B, 1996: p. 2857-2865. 6. Rabinow, J., The magnetic fluid clutch. Journal of Electrical Engineering, 1948: p. 1167-1167. 7. Shulman, Z.P., V.I. Kordonsky, and S.A. Demchuk, The mechanism of heat transfer in

magnetorheological systems. International Journal of Heat and Mass Transfer, 1979: p. 389-394.

8. Jacob, R., Magnetic fluid torque and force transmitting device. 1951, Google Patents. 9. Dyke, S.J., et al., Modeling and control of magnetorheological dampers for seismic response

reduction. Smart Materials and Structures, 1996: p. 565. 10. Poynor, J.C., Innovative designs for magneto-rheological dampers. 2001, Virginia Tech. 11. Kordonsky, W.I., Magnetorheological effect as a base of new devices and technologies. Journal

of Magnetism and Magnetic Materials, 1993: p. 395-398. 12. Lin, S. NGPM – a NSGA-II program in Matlab v1.4. 2011; Available from:

www.mathworks.com/matlabcentral/fileexchange/31166-ngpm-a-nsga-ii-programin-matlab-v1-4.

13. Deb, K. and J. Sundar, Reference point based multi-objective optimization using evolutionary algorithms, in Proceedings of the 8th annual conference on Genetic and evolutionary computation. 2006, ACM: Seattle, Washington, USA. p. 635-642.

14. Srinivas, N. and K. Deb, Muiltiobjective optimization using nondominated sorting in genetic algorithms. Evol. Comput., 1994: p. 221-248.

15. Xu, C., S. Lin, and Y. Yang, Optimal design of viscoelastic damping structures using layerwise finite element analysis and multi-objective genetic algorithm. Computers & Structures, 2015: p. 1-8.

16. Costa, E. and P.J.C. Branco, Continuum electromechanics of a magnetorheological damper including the friction force effects between the MR fluid and device walls: Analytical modelling and experimental validation. Sensors and Actuators A: Physical, 2009: p. 82-88.

17. Jardineiro, V.L.D., Projecto de Actuador Electromagnético para o Ajuste de Caudal em Contadores de Água com regime de Pré-Pagamento utilizando a norma STS, in Departamento de Engenharia Electrotécnica e de Computadores. 2015, Instituto Superior Técnico. p. 93.

18. LORD. Lord Corporation. 2017; Available from: www.lord.com/. 19. MagWeb. World's Largest Database Soft Magnetic Materials. [cited 2017; Available from:

magweb.us/free-bh-curves/. 20. George P. Gogue & Joseph J. Stupak, J. Theory & Practice of Electromagnetic Design of DC

Motors & Actuators. Available from: www.consult-g2.com/course/chapter3/chapter.html. 21. Geyer, H. and H. Herr, A Muscle-Reflex Model That Encodes Principles of Legged Mechanics

Produces Human Walking Dynamics and Muscle Activities. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2010: p. 263-273.

22. Ferreira, F., Development of a human walking model comprising springs and positive force feedback to generate stable gait, in Dep. of Mechanical Engineering. 2013, IST, Lisbon, Portugal.

23. Costa, E. and P. Branco, Construção de um dispositivo amortecedor magnetoreológico para uma suspensão activa. 2008, UTL, Instituto Superior Técnico: Lisboa.

24. Bumgardner, W. What's Typical for Average Daily Steps? 2017 [cited 2017; Available from: www.verywell.com/whats-typical-for-average-daily-steps-3435736.

25. Appendix B, Navier-Stokes Equations, in Chemically Reacting Flow: Theory & Practice. 2003, John Wiley & Sons, Inc.,. p. 763-773.

Page 64: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

44

Page 65: Modelling and Optimization of a Magnetorheological Fluid Actuator for Footdrop · Footdrop pathology, known for disabling the person’s ability to lift the foot, resulting in an

45

Annex A

Navier-Stokes equations in cylindrical coordinates, constant viscosity [25]

z component

+

+

+

= −

+ + ( + )

(.)

r component

+

+

+

= −

+ −

−2

+ ( + )

(.)

component

+

+

+

+ = −

1

+ −

−2

+ ( + )

1

(.)