56
BC-2800 Hematology Analyzer

Mindray BC-2800 Training

Embed Size (px)

DESCRIPTION

Mindray Hematology Analyzer

Citation preview

Page 1: Mindray BC-2800 Training

BC-2800 Hematology Analyzer

Page 2: Mindray BC-2800 Training

7.5“ color LCD , 19 parameters and 3 histograms

3-part differential of WBC

automatically cleaning sample probe

Automatic diluting, lyzing, mixing, rinsing and clog-clearing

Count speed: 30tests/hr

3 counting mode : all parameters 、 WBC/HGB 、 RBC/PLT

Automatic flushing, zapping

remain reagents alarm

expired date alarm

build-in recorder

bar code scanner and keyboard option

Up to 10000 sample results (including histograms) can be stored

BC-2800 Specification

Page 3: Mindray BC-2800 Training

BC-2800 Installation---Environment

Environment

The environment should be as free as possible from dust, mechanical vibrations, loud

noises and electrical interference.

Avoid proximity to brush-type motors, flickering fluorescent lights and electrical

contacts that regularly turn on and off.

Avoid placing the instrument in direct sun or in front of a source of heat.

Page 4: Mindray BC-2800 Training

BC-2800 Installation---Space

Space

Comfortable working height

At least 40cm on each side is the preferred access to perform service procedures.

At least 10cm behind for cabling and ventilation

There is enough space for reagents.

 

Page 5: Mindray BC-2800 Training

BC-2800 Installation---Power

Power

A properly grounded socket supplying 220 or 110 volts

The fluctuation should be in the range of ±10%.

The power frequency should be 50/60Hz.

Power consumption is less than 180VA.

 

 

Page 6: Mindray BC-2800 Training

Select some samples in the sample tabular review menu, and then p

ress [F5] to access “extension function” menu.

[PRINT] print the current displayed content

[HELP] help information

[MENU] exit menu

Extension function: “repetition” and “trend graph”

“trend graph” function only open to administrator

Extension function in tabular review menu

Page 7: Mindray BC-2800 Training

Password: 2826

Procedure: press F5 then press F1

Trend graph analysis mean: it is to monitor one patient’s illness trend.

Trend graph

Page 8: Mindray BC-2800 Training

Value of lyse: 0.5mL(whole blood) , 0.36mL(pre-diluted)

Blood needed:

whole blood: 13μL

pre-diluted: 20μL

Count time : less than 2 min / test

Aperture size : 80μm ( diameter ), 70mm ( depth )

Ratio of whole blood Capillary

WBC/HGB 1:308 1:400

RBC/PLT 1:44862 1:44402

Volume of volumetric testing: 300μL ×2times

BC-2800 sample and dispense

Page 9: Mindray BC-2800 Training

Range of linear and display

Parameter

Linear Display

WBC 0 - 99.9 ×109/L 0.0-499.9 ×109/L

RBC 0 -9.99×1012/L 0.0-9.99 ×1012/L

HGB 0 - 300g/L 0- 300g/L

MCV ------ 0-250 fL

PLT 0 - 999×109/L 0-2999 ×109/L

Page 10: Mindray BC-2800 Training

%100Mean

DiffCV

Parameter linear Repetition(CV%)

WBC4.0-7.5×109/L ≤ 3.0

7.6-15.0×109/L ≤ 2.0

RBC3.0-4.5×1012/L ≤ 2.0

4.5-6.5×1012/ L ≤ 1.5

HGB 110 – 180g/L ≤ 1.5

MCV 70.0 – 100.0fL ≤ 1.0

PLT100-200 ×109/L ≤ 5.0

201 - 500×109/L ≤ 4.0

Repetition

Page 11: Mindray BC-2800 Training

AC Input :

AC 220V±30V 50/60Hz±1Hz

Consumption :

150 VA

Fuse :

250 V 2A

Power Supply

Page 12: Mindray BC-2800 Training

Without any operation in 10 minutes, the system will enter the

saver program.

LCD display turns off

Power indicator is still light.

Any key exists the saver program.

Screen Saver

Page 13: Mindray BC-2800 Training

Power indicator

Start key

7.5” color LCD

Keypad

Record

Sample probe

Front Panel

Page 14: Mindray BC-2800 Training

Power switch and fuse

Equipotential terminal

Diluent coonector

Rinse connector

Lyse connector

Waste connector

Rear panel

Page 15: Mindray BC-2800 Training

Keyboard

Power or floppy

Transition

Printer or upgrade interface

Scanner interface

Left side

Page 16: Mindray BC-2800 Training

Display featureTitle Count Mode System

StatusSystem time

Window

Error Messages Reagent Status

MENU On-line Help

Page 17: Mindray BC-2800 Training

1 ) >99tests : remain reagents and remain waste container are enough for 99tests

2 ) 0 test : any one of the remain reagents or remain waste container is not enough for 1 test. The warning message will be show in error area. Actual tests : show the actual tests number

Remain reagents

Page 18: Mindray BC-2800 Training

注意: PS/2 标准键盘操作的区别

用 PS/2 标准键盘执行模式切换功能键是: [Ctrl + A]

Keypad

Page 19: Mindray BC-2800 Training

Menu tree

Count

Revi ew

Cal i brat i on Manual

Auto

Qual i ty control

Password

Sel f - test

Log

Status

Servi ce Mai ntenance

Confi g

Pri nt

Shutdown

Control s

X-B anal ysi s

Sett i ngs

Setup

Auto - f resh bl ood

Page 20: Mindray BC-2800 Training

Diluent Rinse LyseE-Z

cleanser

Normal start-up

46.5ml( clean & once background test)

16.5ml 0.5ml ---

75ml( clean & twice background test

)26ml 1ml ---

All parameters(whole blood)

28.5ml 9.5ml 0.5ml ---

All parameters(pre-diluted)

30.5ml9.5ml 0.36ml ---

Normal shut down 13.5ml 5.5 ml --- 1ml

Reagents consumption

Page 21: Mindray BC-2800 Training

The last shut down statusProcedure

Normal shutdown clean tubing and background testingAbnormal shutdown( Directly shutdown )

Reset motors, check software, clean tubing and background test

Package or empty tubing Prime tubing, clean tubing several times, background test.

1. Switch on BC-2800 , and the power indicator will be light. 2. The Initializing procedure will continue about 80s. 3. Self test procedure.

4 The system will enter “count” menu. If there are errors during startup, the error message will be displayed on left bottom of screen.

Power on and shut down

Page 22: Mindray BC-2800 Training

Background test

Parameter Acceptable limit

WBC 0.3 109 / L

RBC 0.03 1012/ L

HGB 1 g / L

HCT 0.5 %

PLT 10 109 / L

Page 23: Mindray BC-2800 Training

1. Only ID mode

2. All information

3. Batch information mode

Sample information input

Page 24: Mindray BC-2800 Training

1. Press [MODE] to select the prediluted mode (any mode preceded by PB). 2. Press [DILUENT] to enter the Add Diluent screen3. Prepare a clean sample cup under the sample probe and press [START] to dispense 1.6ml diluent to the cup.

4. Take out the sample cup after the sample probe rose up and press [ENTER] to exit the Add Diluent screen. 5. Collect 20μL capillary blood specimen and immediately add it to the sample cup and shake to mix them.

Preparation of pre-diluted blood

Page 25: Mindray BC-2800 Training

[F1] is to select the setting items.

[SETUP]-[Settings]

Page 26: Mindray BC-2800 Training

Reagent Valid volume

Diluent ( 0-30.0 ) L

Rinse ( 0-30.0 ) L

Lyse ( 0-999 ) mL

Waste ( 0-99.9 ) L

[SETUP]-[Settings]--Reagents Remain

Page 27: Mindray BC-2800 Training

[SETUP]-[Settings]—Print & Communication

Page 28: Mindray BC-2800 Training

[SETUP]-[Settings]—Print & Communication

Printing Device

Format1 Format2 Format3 Format4

Recorder Parameters + histograms

Parameters Parameters+ histograms

Parameters

Printer Whole page with histograms

Whole page without

histograms

Whole page with histograms

Whole page without

histograms

Page 29: Mindray BC-2800 Training

[SETUP]-[Settings]—Date & time

Page 30: Mindray BC-2800 Training

[Setup]-[Settings]---[Gain]

Page 31: Mindray BC-2800 Training

[Setup]-[Settings]---[Gain]--WBC

• too small WBC gain that should be increased

• too large WBC gain that should be decreased.

Page 32: Mindray BC-2800 Training

[Setup]-[Settings]---[Gain]--RBC

• When the difference between the actual MCV analysis result and the expected result exceeds 6%, need to change the RBC gain.

• For example, assuming the expected MCV result is 90.0fL, while the actual analysis result is 82.0fL, the gain factor as below:

098.10.82

0.90=

ActualMCV

VExpectedMC

Page 33: Mindray BC-2800 Training

[Setup]-[Settings]---[Gain]--HGB

• adjust the HGB gain to change the HGB blank voltage, which usually should be set between 4.3 ~ 4.5V.

Page 34: Mindray BC-2800 Training

[Setup]-[Settings]--count

Page 35: Mindray BC-2800 Training

[Setup]-[Settings]--Reference RangePatient Group Sex Age

General Not specified Not specified

Not specified > 13 years

Male adults Man > 13 years

Female adults Woman > 13 years

Pediatric patients

Not specified>= 1 month and <= 13

years

Neonates Not specified <1 month

Page 36: Mindray BC-2800 Training

[Setup]-[Settings]--other

Page 37: Mindray BC-2800 Training

[Service]-[Maintenance]--Prime

Page 38: Mindray BC-2800 Training

[Service]-[Maintenance]—Zap & Flush

Page 39: Mindray BC-2800 Training

[Service]-[Maintenance]—Probe cleanser clean

• If your analyzer is to run 24 hours a day, you should perform this procedure every 3 days.

• If you follow the shutdown procedure to turn off your analyzer everyday, you should perform this procedure every week.

• To make sure this analyzer functions normally, every time the accumulated analyzed samples reach 200, a dialog box will pop up to remind the user to perform the Probe cleanser cleaning

procedure.

Page 40: Mindray BC-2800 Training

“Lyse test” is to check the injected volume for lyse is correct or not. When shall we use this function?When the WBC result or its histogram is abnormal, we should run this function.

Screw off these two screws

[Service]-[Maintenance]—Lyse test

Page 41: Mindray BC-2800 Training

Remove the Count bath shield cover

The injected lyse must over the second mark

2ml lyse will be dispensed to the count bath.

[Service]-[Maintenance]—Lyse test

Page 42: Mindray BC-2800 Training

[Service]-[Maintenance]—Drain

• Drain bath

Empty the count bath, and prime it again

• Drain Tubing

Empty all tubing, then turn off the equipment

Page 43: Mindray BC-2800 Training

[Service]-[Maintenance]—Clean

• Clean bathUse diluent to clean count bath

• Clean wipe blockAfter being used for a long time, the bottom of the probe wipe

block may be contaminated by blood.

When the accumulated analyzed samples reach 4,000, a dialog box will

pop up to remind the user to clean the wipe block

Page 44: Mindray BC-2800 Training

[Service]-[Maintenance]—Prepare to ship

• Empty the tubing

• Wash the tubing

• Empty the tubing

• Turn off the equipment

Page 45: Mindray BC-2800 Training

[Service]-[System test]

Page 46: Mindray BC-2800 Training

The available self-test items are divided into four groups, the

Tubing, Machine, Valve and Circuit.

[Service]-[self-test]

Page 47: Mindray BC-2800 Training

[Service]-[self-test]—[Test tubing]

Count Time WBC( S) & Count Time RBC( S) It measures the duration of a WBC or RBC countAperture(v) It measures the voltage (v) over the aperture. Vacuum It checks whether the vacuum system functions normally. Pressure It checks whether the system flushes the aperture at a normal press

ure. Filter It checks whether the filter functions normally.

Page 48: Mindray BC-2800 Training

[Service]-[self-test]—[Machine]

Page 49: Mindray BC-2800 Training

[Service]-[self-test]—[Valve]

Page 50: Mindray BC-2800 Training

It is to monitor that WBC, RBC and PLT signals can be properly converted

into digital signals or not.

[Service]-[self-test]—[Circuit]

Page 51: Mindray BC-2800 Training

The recorded events are divided into four groups, All, Para set, Error and Other

(including setting discriminators, system self-test and updating system software)

This analyzer can save maximum 100 log files and once the maximum number has been reached, the newest log will automatically cover the oldest one.

[Service]-[Log]

Page 52: Mindray BC-2800 Training

Every screen displays 13 items

It only shows BC-2800 current configuration, no change.

[Service]-[Configuration]

Page 53: Mindray BC-2800 Training

The printing tasks are queued in this screen

[Service]-[Print management]

Page 54: Mindray BC-2800 Training

[Calibration]---[Manual]

Password: 2826

Valid range:75%---125%

Page 55: Mindray BC-2800 Training

[Calibration]---[Auto]

Password: 2826

Valid range:75%---125%

Page 56: Mindray BC-2800 Training

[Calibration]---[Fresh blood]

Password: 2826

Valid range:75%---125%