7/29/2019 MilliQ water
1/16
Milli-Q Advantage A10Water Puriication SystemsUser-adapted ultrapure water
7/29/2019 MilliQ water
2/16
2
7/29/2019 MilliQ water
3/16
GIVING YOU THE ADVANTAGE
As a scientist today, you ace many challenges. Youre responsible or deliveringhigh quality technical resultsbut now at an even aster pace. You also have tocomply with a growing number o quality and regulatory standards, while workingin a laboratory where space may be limited. Resulting rom on-going dialogueswith scientists like you, Merck Millipore continues to develop water systems thatimprove your work environment.
The Milli-Q Advantage system has been designed especially or you. Thissystem delivers high quality water adapted to todays lab environment.
3
7/29/2019 MilliQ water
4/16
A NEW APPROACH
Advanced technology meets total exibility.
The Milli-Q Advantage system simplifes your daily lie in the lab by utilizing two
distinct components:
The Milli-Q production unit is a compact system that can be placed either on thebench, under the bench or on a wallwherever it best suits you.
The Q-POD (Quality-Point-o-Delivery) unit provides fnal polishing at the point-o-use, delivering ultrapure water adapted to your specifc needs.
The revolutionary Q-POD technology empowers you with convenient and exibledispensing options. Up to three Q-POD units can be used with each Milli-QAdvantage system at dierent locations within the same laboratory.
Can you envision a laboratory environment
where the ultrapure water source is exible
enough to support a variety o daily activities?
4
7/29/2019 MilliQ water
5/16
IMPROVING THE WAY
...YOU PURIFY
Appropriate water quality
Depending on your application, specifc contaminantscan aect your results.The Milli-Q Advantage system oers a unique rangeo fnal purifers or complete peace o mind.
...YOU CONTROL
Required inormation
The Milli-Q Advantage system provides relevant dataabout the water quality and system status.
The Q-POD units multicolor graphic displayallows easy interaction with the system and providesinormation about water quantity and quality.
The primary graphic screen on the Milli-Qproduction unit details the systems operation andperormance or managing routine maintenance andtroubleshooting procedures.
...YOU OPERATE
Flexible water delivery
Working both intuitively and precisely, the Q-POD unit
is always within your reach. Select a very precise volumethrough the autofll key or manually press the plunger.
5
7/29/2019 MilliQ water
6/16
6
7/29/2019 MilliQ water
7/16
7
7/29/2019 MilliQ water
8/16
DESIRED WATER QUALITY
You just choose the appropriate media that remove
the specic contaminants which may aect your
results depending on the application.
For example, the Millipak flter with the MilliporeExpress membrane (0.22 m) produces particulate-and bacteria-ree ultrapure water or analyticalapplications such as spectrophotometry, spectroscopyand chromatography. The BioPak ultraflter, while
removing particulates and bacteria, producespyrogen- and nucleaseree ultrapure water orbiochemical applications.
PRACTICALITY CONVENIENT DELIVERY
Daily perormance is acilitated by having ultrapure
water and inormation conveniently delivered at the
point-o-use.
EASY DATA ACCESS
Following your own workfow, relevant data is ac-
cessible conveniently.
See essential inormation on the Q-POD unit screenat any time.All critical inormation, including water quality,system status and warnings, is summarized on themulticolor graphic display. The water quality statusalso can be printed directly rom the Q-POD unit.
Control system use and maintenance on the waterproduction unit screen.The main graphic screen on the production unit displays
details o the systems operation and perormance.Graphics assist the user in perorming specifc tasks,including maintenance procedures.
Protect access to critical inormation.An ID code and password ensure that only thedesignated user can access felds containing criticalinormation, such as water quality set points.
8
7/29/2019 MilliQ water
9/16
ENHANCED FLEXIBILITY
Your work is enhanced through convenient and
adaptable dispensing.
Adjust the Q-POD unit arm (rotation and height)or all commonly-used laboratory glassware andplasticware.
Set the optimized water quality beore delivery bypressing the water recirculation button on the base othe Q-POD unit.
Select the desired volume by simply using the + and keys and then pressing the central key. The lastdelivered volume is recorded to acilitate re-use.
Move the Q-POD dispenser rom the arm or varyingwater delivery:
low-ow
medium-ow
high-ow (up to 2 l/min)
9
7/29/2019 MilliQ water
10/16
QUALITY
OPTIMIZED PURIFICATION
The most reliable ultrapure water quality depends onbuilding an optimized purifcation sequence.
REQUIRED PURIFICATION STEPS...
You need a exible system that can match thediversity o your applications. This is possible with avariety o purifcation media!
Pure water, ideally rom an Elix purifcation system,enters the Q-Gard pretreatment pack (1) which ischosen based on the eed water source.
Pretreated water then passes into a dual wavelengthUV lamp (2), which ensures organic molecule oxidationand bacteria destruction.
Next, the Quantum polishing cartridge (3) removesionic and organic contaminants below trace levels tomatch the water quality required or your application.
...UP TO THE POINT OF DELIVERY
The ultrapure water produced by the system recirculatesthrough a loop up to the Q-POD unit, where the fnalpurifcation step required or your particular applicationoccurs
UNDER CONTROL
The measurement o both organic and ionic qualityo produced water occurs at the outlet o the system,
through the appropriate calibrated meters:
Accurate Total Oxidizable Carbon (TOC) monitor (4)
High-sensitivity resistivity cell (5)
Both the Q-Gard and Quantum cartridges enable ull traceability.
10
7/29/2019 MilliQ water
11/16
4
31
2
5
TOC
11
7/29/2019 MilliQ water
12/16
ALLOWING SENSITIVE IONIC DETECTION
Milli-Q Advantage systems high precision resistivity meters have specic
eatures to ensure that the value displayed on the screen is meaningul:
Patented cell design with coaxial electrodes to warrant cell constant stabilityFlow-through design to make sure that the measure is representative o the actualionic concentration in waterLow cell constant (0.01 cm-1) to ensure optimum measurement accuracy o lowionic contamination as required by ASTM norm D 1125-95 (1999)Temperature measurement with a 0.1 C resolution or proper report o temperature-compensated resistivity, as recommended in ASTM norm D 1125-95 (1999)Automatic warning messages i the resistivity measure is compromised by a deectDesign allows TOC suitability test to be perormed as required by USP ( 645).
12
7/29/2019 MilliQ water
13/16
TOTAL CONTROL
Double monitoring allows control over both ionic andorganic contaminants that can impact your results.
PREVENTING ORGANIC BREAKTHROUGH
The proper measurement o TOC levels is key or
confrming that the systems organic removal processis operating within specifcations.
The integrated TOC meter accurately monitors rom 1to 999 ppb. The design takes into account USP suitability requirements.
TOC measurements are perormed automatically on acontinuous basis during production and intermittentlyduring periods o non-use. This lets you check theorganic content o the water regularly, while avoidingthe risk o your work being compromised by anundetected organic breakthrough.
The TOC monitor uses a 0.5 ml quartz cell to captureultrapure water.
When the A10 UV lamp is on, photocatalyticoxidation o organic compounds occurs.
The end product o organic oxidation is carbon dioxide,which dissolves in water and causes conductivity toincrease. This change in conductivity (temperature-compensated to 25 C) is monitored continually by thetitanium electrodes in the TOC monitor.
A set o algorithms confrms complete oxidationand calculates the carbon level associated with thisconductivity change.
Zeit
conductivity
valve
lamp
time
fll
initial measure
o conductivityoxidation & fnal
conductivity measure
purge
open
on o
closed
on
open
solenoid valve
thermistor
quartz cell
0,5 ml
titanium electrodes
185/254 nm
UV lamp
13
7/29/2019 MilliQ water
14/16
ADVANCED TECHNICAL
SUPPORTMerck Millipore application specialists provideinormation about system use and application insightsas well as how to select the best services related toyour particular situation.
COMPREHENSIVE SERVICE PROGRAM
Choose the services you need rom Merck Milliporescomprehensive service program.
InstallationTechnical and scientifc assistancePreventive maintenance visitsTroubleshooting visitsCustomized user trainingVerifcation and/or calibration o monitoring devicesPharmacopeia suitability testsValidation supportMaintenance plans
QUALIFICATION EXPERTISE
Merck Millipores Qualifcation Program acilitates
laboratory validation procedures.
Validation support is provided by trained MerckMillipore Field Service Support Engineers usingcalibrated equipment and Qualifcation Workbooks.
With more than 10 years experience in water systemqualifcation services, Merck Millipore can assist youin complying with regulatory standards applicable toyour industry.
14
7/29/2019 MilliQ water
15/16
PRODUCT SPECIFICATIONS
Product specifcations Milli-Q Advantage A10 System
Product Water Specifcations
Parameter Value
Resistivity 18.2 M.cm at 25 C
TOC 5 ppb
Particulates* (> 0.22 m/ml) < 1 particulate/ml
Bacteria* < 0.1 cu/ml
Pyrogens (endotoxins)* < 0,001 EU/ml
RNases* < 0,01 ng/ml
DNases* < 4 pg/
Flow Rate up to 2 l/min
*Test conditions with the appropriate Q-POD fnal polisher. These values are typical and may vary
depending on the nature and concentration o contaminants in the eed water.
Milli-Q Advantage A10 System Specifcations
Parameter Value
Production unit dimensions (H x W x D)497 x 332 x 360 mm(19.5 x 13 x 14.2 in)
Q-POD delivery unit dimensions (H x D)579 x 230 mm(22.8 x 9 in)
Production unit weight 15 kg (33 lb)
Production unit operating weight 19 kg (41.9 lb)
Q-POD delivery unit weight 5 kg (11 lb)
Q-POD delivery unit operating weight 5.5 kg (12.1 lb)
Distance rom production unit to Q-POD 2.9 m (9.5 t)
Dispenser tubing length 80 cm (31.5 in)
Electric power cable length 2.9 m (9.5 t)
Electric power supply voltage 100 230 V +/- 10 %
Electric power supply requency 50 60 Hz
Feed water connector: GazMilli-Q Advantage A10 main unit data connection: Ethernet (RJ45)Q-POD data connection: Parallel port (25-pin D-Sub)
ALL-INCLUSIVE SYSTEM CERTIFICATES
To assist you in ollowing industry requirements, Milli-Q Advantage systems aredelivered with specifc Certifcates o Quality and Calibration or temperature,resistivity and TOC meters. Merck Millipores manuacturing site is ISO 9001v.2000 and ISO 14001 certifed.
15
7/29/2019 MilliQ water
16/16
Lit. No. PB0001EN00 01/12. Copyright 2012, EMD Millipore, Billerica, MA, U.S.A. Merck Millipore, A10, Milli-Q, Q-POD, BioPak,
Quantum, Millipak, Merck Millipore Express, Elix and Q-Gard are registered trademarks o Merck KGaA, Darmstadt, Germany. The M
mark and Merck Millipore are tr ademark o Merck KGaA. USP is a registered trademark o the United States Pharmacopeial Convention.
ISO is a registered trademark o the International Organization or Standardization. All rights reserved. Photographs: BHL Production.
Design: Sysaxe.
ORDERING INFORMATION
To order a Milli-Q Advantage A10 system and associated consumables, pleasecontact your local Merck Millipore subsidiary, or visit our dedicated web page at:www.millipore.com/advantage
16