The world leader in serving science Migrating From He to H 2 Carrier Gas in GC-MS Analysis of Volatile and Semi-Volatile Organic Compounds in Water Jessie Butler Alexander N. Semyonov

Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

Embed Size (px)

Citation preview

Page 1: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

1The world leader in serving science

Migrating From He to H2 Carrier Gasin GC-MS Analysis

of Volatile and Semi-Volatile Organic Compounds in Water

Jessie ButlerAlexander N. Semyonov

Page 2: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

2

Helium and Profitability of Drinking Water Testing Labs

N2

from

Gen

erat

or

H2

from

Gen

erat

or

H2

fro

m C

ylin

de

r

UH

PH

efr

om C

ylin

der

Page 3: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

3

Carrier Gases: Physical Properties

Property He H2 N2 Ar

Molecular Mass, Da 4 2 28 40

Density, kg/m3 0.18 0.09 1.25 1.78

Diffusion Coefficient, cm2/s 0.55 0.60 0.15 0.20

Viscosity, Pa•s × 106 23.2 10.3 20.9 27.0

uopt, cm/s 20 40 12 <10

Carrier Gas Density: Less is Better

Diffusion Coefficient: More is Better**

Dynamic Viscosity : Less is Better

He H2

He H2

* Crystal Lattice Density Analogy only, O.Fryazinov; A.Pasko; V.Adzhiev Computer-Aided Design 2011, 43(3) 265-277** On the RHS of the van Demmter curve, that is.

*

Page 4: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

4

Thermo Scientific Position on Hydrogen with GC-MS

• Hydrogen Kit• Required to Run with H2

• Required for H2 Specifications• Includes Hydrogen Sensor• Requires 300 L/s Turbo Pump

• Upgrades in the Field• All existing ISQ GC-MS Systems

are either already H2 capable or are upgradeable with 300 L/s Turbo and H2 Kit

• IF you decide to use your existing Thermo Scientific™ DSQ™, DSQ II & ITQ GC-MS with H2, you do so at your own risk.• Thermo Scientific™

Hydrogen Systems• ISQ™ Single Quadrupole GC-MS• TRACE™ 1300 Series GC• TSQ™ 8000 GC-MS/MS• Explosion Tested

& Safety Certified

Page 5: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

5

ISQ GC-MS Hydrogen Carrier Installation Specs

• ISQ Single Quadrupole GC-MS• Guaranteed installation specifications with hydrogen carrier gas• Choice of either He or H2 on installation spec sign-off• PCI/NCI same S/N!

* Requires Hydrogen Kit for H2.

In EI mode, 1 μL of 1 pg/μL octafluoronaphthalene (OFN) will produce the following minimum signal to noise for m/z 272 when scanning from 50 – 300 u:

1,500:1 100:1

In PCI mode, 1 μL of 100 pg/μL benzophenone will produce the following minimum signal to noise for m/z 183 when scanning from 80 – 230 u using methane reagent gas:

300:1 300:1

In NCI mode, 2 μL of 100 fg/μL of OFN will produce the following minimum signal to noise for m/z 272 when scanning from 50 – 300 u using methane reagent gas:

600:1 600:1

Page 6: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

6

Hydrogen Purity Specifications

• GC-MS Hydrogen Generator• O2 < 0.01 ppm• H2O < 1 ppm• THC N/A

• Hydrogen Grade 5.0 UHP• O2 < 1 ppm• H2O < 3 ppm• THC < 0.1 ppm

• Hydrogen FID Fuel Grade• O2 < 1 ppm• H2O < 3 ppm• THC < 0.5 ppm

• Hydrogen Grade 6.0 UHP• O2 < 0.1 ppm• H2O < 0.5 ppm• THC < 0.1 ppm

Page 7: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

7

Hydrogen Generators: Choose the Best

Page 8: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

8

Understanding Hydrogen Plumbing

• Tubing• Stainless steel preferred• 1/8” pre-cleaned• New tubing is preferred

over used with He

• Gas Filter, triple stage

• Filter – only need with tanks• For removal of water,

oxygen and hydrocarbons

What you need to know about plumbing up hydrogen to your GC

Page 9: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

9

Limits on Hydrogen Flow and Column Selection

1 mL/min hydrogen on 0.25 mm × 30 m column

1 mL/min hydrogen on 0.18 mm × 20 meter column

Page 10: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

10

Tuning Report: He vs. H2

He H2

Page 11: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

11

EPA Methods 524 and 525

• EPA Method 524• Purge & Trap• BFB Tuning Criteria• 5 mL sample volume• Calibration (0.4 – 200 ppb)

• EPA Method 525• Direct 1 µL liquid injection• DFTPP Tuning Criteria• 1 liter sample volume• Calibration ( 0.1 – 10 ppm)

Volatile and Semi-volatile Organics

Page 12: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

12

EPA Method 8270

• EPA Method 8270• Direct 1 µL liquid injection• Calibration (1 – 200 ppm)• More than 120 compounds• Surrogates = 40 ppm• Internal Standards = 40 ppm

• Goals• Improve peak shape and

resolution• Improve method runtime

(<15 minutes)

Semi-volatile Organics

Page 13: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

13

DFTPP (Decafluorotriphenylphosphine) Spectrum

50 100 150 200 250 300 350 400 450 500m/z

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

198.00

442.00

127.01

110.0077.05

68.97255.0151.03

206.06107.02 129.02 275.00186.05

244.0581.03 224.04 423.01167.09 296.01 364.92156.0557.03 323.02 334.00 402.97391.69 469.66 490.76197.99

127.02

442.01

110.0151.04 77.0768.97

255.00206.07129.01107.01

186.06 275.01224.05168.0581.03 244.05 423.00295.97148.02 364.9857.02 323.03 334.03 403.18383.08 453.36 487.59

NL: 2.73E6dftpp4#1315 RT: 5.28 AV: 1 SB: 1 5.24 T: {0,0} + c EI Full ms [35.00-500.00]

NL: 4.04E6DFTPP12#1310 RT: 5.27 AV: 1 SB: 1 5.24 T: {0,0} + c EI Full ms [35.00-500.00]

Helium

Hydrogen

Page 14: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

14

BFB (1-bromo-4-fluorobenzene)

40 60 80 100 120 140 160 180 200 220 240 260 280 300m/z

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

95

174176

75

5074

6851 9437 7638 6149 9663 177888757 14379 105 157 173117 135 14812841 207111 161 182 21119598 203 217 233 260 267251236224 285243 300281271 294

95

174176

75

50

74

94 9651 681777649 8840 61 6257 81 14397 130117 147 208193157133 211173104 178 261201165 234125113 290285227 237 268249218187 273 293255

NL: 1.92E62152012_a001#3712-3715 RT: 12.42-12.43 AV: 4 SB: 1 12.37 F: {0,0} + c EI Full ms

NL: 5.25E5blk11#2927-2930 RT: 8.31-8.32 AV: 4 SB: 1 8.27 F: {0,0} + c EI Full ms

Helium

Hydrogen

Page 15: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

15

BFB H2 Tuning Report in Target Software

Page 16: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

16

Thermo Scientific Target Software – Still Viable Product

Page 17: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

17

Ion Ratio Stability for DFTPP

Page 18: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

18

BFB Ion Ratio Stability

m/z Criteria run01 run02 run03 run04 run05 run06 run07 run08 run09 run10 run11

50 15‐40% 28 26 26 24 27 26 25 26 28 25 25

75 >30% 52 51 48 48 51 52 49 48 48 50 53

95 100% 100 100 100 100 100 100 100 100 100 100 100

96 5‐9% 8.0 7.3 7.9 7.7 8.2 7.2 7.7 7.4 7.8 7.5 7.7

173 < 2% of 174 0.4 0.6 0.7 1.5 0.4 0.4 0.8 0.7 0.5 1.1 0.3

174 >50% 76 75 69 69 75 70 72 72 71 74 72

175 5‐9% of 174 7.3 7.4 7.4 5.9 7.6 8.5 6.4 7.7 6.8 6.4 6.7

176 95‐101% of 174 96 95 100 96 93 96 98 95 96 101 96

177 5‐9% of 176 5.1 6.3 6.6 5.2 6.4 6 5.1 5.5 6.6 6.5 6

Page 19: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

19

EPA Method 8270 Performance MixRT: 4.03 - 8.70

4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 6.2 6.4 6.6 6.8 7.0 7.2 7.4 7.6 7.8 8.0 8.2 8.4 8.6Time (min)

0

10

20

30

40

50

60

70

80

90

100

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

RT: 6.90AA: 22134292BP: 184.15

RT: 5.86AA: 15396601BP: 198.03

RT: 7.97AA: 18017012BP: 235.06

RT: 5.53AA: 12295161BP: 265.91

RT: 5.27AA: 19434464BP: 197.98

RT: 6.20AA: 20663815BP: 184.13

RT: 7.02AA: 19668080BP: 165.11

RT: 4.90AA: 9725003BP: 265.85

NL:2.15E7TIC MS ICIS DFTPP

NL:2.69E7TIC MS ICIS DFTPP11

Helium

HydrogenPeak 1 pentachlorophenolPeak 2 DFTPPPeak 3 benzidinePeak 4 p,p’-DDT

Page 20: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

20

Method Parameters: EPA Method 524.2

• Purge & Trap: 5 mL sample • Oven

• Inlet• S/SL: 200°• Split : 40:1

• Carrier: 0.7 mL/min H2

• Column: TG VMS 0.18 mm x 20 m x 1.0 µm

• MS

50° 1 min. 30°/min150° 0 min 20°/min200° 0 min 30°/min300° 0 min

Source 275°EC 25 uampsFS 0.8-1.61.6- 15 min

47-30035-300

Scan Speed 1,687 u/s

VOCARB Purge 11 minDry Purge

1 min

Desorb 240° 2 min

Page 21: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

21

Method Parameters: EPA Method 525.2

• Standards• Prepared in ethyl acetate• Calibration curve

(0.1, 0.5, 1.0, 2.0, 5.0, 10 ug/mL)

• Oven:

• Inlet• S/SL: 325°C• Split Mode: 15:1

• Carrier: 1 mL/min H2

• Column• TG-5 SilMS 20m x 0.18mm x

0.36 µm

• MS

50° 1 min 30°/min150° 0 min 20°/min200° 0 min 30°/min300° 0 min 20°/min

Source 325°EC 15 uampsFull Scan 35-500Scan speed

4,650 u/ s

Page 22: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

22

Method Parameters: EPA Method 8270D

• Standards• Prepared in ethyl acetate• Calibration curve

(1.0, 2.0, 5.0, 10, 50, 100, 200 ug/mL)

• Oven:

• Inlet• S/SL: 325°C• Split Mode: 15:1

• Carrier: 1 mL/min H2

• Column• TG-5 SilMS 20m x 0.18mm x

0.36 µm

• MS

50° 1 min. 30°/min150° 0 min 20°/min200° 0 min 30°/min300° 0 min 20°/min350° 0.5 min

Source 325°EC 15 uampsFull Scan 35-500Scan speed

4,650 u/ s

Page 23: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

25

Linear Fit of Low Level PAHs in SIM (0.05 to 10 ppm)

Compound %RSD Compound %RSD

naphthalene 6.1 flouranthene 6.7

2-methylnaphthalene 8.7 benzo(a)anthracene 8.2

1-methylnaphthalene 7.7 chrysene 5.9

acenaphthylene 8.5 benzo(b)fluoranthene 8.6

acenaphthene 8.5 benzo(k)fluoranthene 9.2

flourene 8.5 benzo(a)pyrene 7.8

phenanthene 3.8 indeno(1,2,3,c,d)pyrene 5.1

anthracene 7.0 dibenz(a,h)anthracene 5.4

pyrene 3.6 benzo(g,h,i)perylene 6.0

acenaphthene-d10 4.6 phenanthene-d10 4.5

chrysene-d12 9.7 perylene-d12 7.4

Page 24: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

26

Cal Curves: Naphthalene & Benzo(g,h,i)perylene

Page 25: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

27

Low Level PNAs 50 ppb (+/- 50%) Accuracy

10

20

30

40

50

60

70

80

0 2 4 6 8 10 12 14 16 18

Page 26: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

28

8270: Comparison of Hydrogen to Helium Run Times

C:\chem\...\level8.d\level8 12/5/2012 2:04:03 PM

RT: 0.00 - 20.01

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

6.83

3.8510.24

9.198.987.904.56

12.3510.276.26 7.645.186.97 11.198.065.65

8.4111.163.88 4.61 11.449.763.34 7.014.2610.79 12.558.84

3.31

2.25 2.981.80 2.51

12.708.80 13.21 13.849.447.58 12.43 14.6212.1310.717.46 10.656.63

8.59 13.969.56 13.158.066.14 10.899.0111.27 15.37

9.61 18.0815.857.22 12.575.36 15.335.34 10.02

6.04 6.76

18.71

3.8415.92

2.82

11.32 18.913.24 16.42 17.112.65 4.17

NL:5.35E8TIC MS level8

NL:1.12E8TIC MS 03091102_110309112925

Hydrogen

Helium

Page 27: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

29

EPA Method 8270 MDLs

Page 28: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

30

F:\524\20ppbB_130208153106 2/8/2013 3:31:06 PM

RT: 0.79 - 11.91

1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5 10.0 10.5 11.0 11.5Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

7.40

9.28

8.95

7.268.67 9.038.477.83 9.68

5.89 8.087.2011.56

9.374.07

11.258.39 9.77

2.456.62

6.462.231.81 6.215.05 5.32 5.683.923.861.67 4.604.133.33

4.262.16 2.841.09

4.956.714.45 10.561.46

2.8911.112.97 6.78 9.94 10.19

NL:8.77E7TIC MS 20ppbB_130208153106

Performance: EPA Method 524 - 20 ppb Standard

Page 29: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

31

Gases at 200 ppt

RT: 0.80 - 1.81 SM: 7B

0.80 0.85 0.90 0.95 1.00 1.05 1.10 1.15 1.20 1.25 1.30 1.35 1.40 1.45 1.50 1.55 1.60 1.65 1.70 1.75 1.80Time (min)

0

50

100

0

50

100

0

50

100

0

50

100

Rel

ativ

e A

bund

ance

0

50

100

0

50

100

RT: 0.93AA: 9279BP: 85

RT: 1.04AA: 47618BP: 50

RT: 1.09AA: 19284BP: 62

RT: 1.29AA: 21231BP: 94

RT: 1.38AA: 14294BP: 64RT: 1.08

AA: 5325BP: 64

RT: 1.47AA: 16876BP: 101

NL: 7.88E3m/z= 85-86 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC

NL: 2.67E4m/z= 50-51 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC

NL: 1.51E4m/z= 62-63 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC

NL: 1.30E4m/z= 94-95 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC

NL: 9.91E3m/z= 64-65 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC

NL: 1.11E4m/z= 101-102 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC

dichlorodifluoromethane

chloromethane

vinylchloride

bromomethane

chloroethane

trichlorofuoromethane

Page 30: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

32

Accuracy of 400 ppt - Passed +/- 50%

‐0.1

6E‐16

0.1

0.2

0.3

0.4

0.5

0.6

0 20 40 60 80 100 120 140

methylenechloride

Page 31: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

33

Moving To Hydrogen on the GC Side

• Hydrogen Sensor is a Must• In GC oven• Possible on the hydrogen generator• May also be required at the ceiling of the lab

• Expect Lower Inlet Pressure• Due to viscosity differences of H2 and He• Move to a smaller id column

Page 32: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

34

Moving to Hydrogen on the MS Side

• NO MS hardware change required to meet H2 Installation Specs• Maintain good vacuum

• Extended performance turbomolecular pumps

• 9.8 x 10 -6 Torr (1 ml/min hydrogen flow)

• Higher initial background• Minimized with stainless steel pre-cleaned tubing

• UHP Grade 5.0 or better Hydrogen Source

• Bake out source at 350°C for one hour with filament on

• CI Effect on some compounds• Require linear or quadratic fit

• Reduce flow rate of H2 into MS

• Pressure dependency:• Minimize solvent vapor with smaller id columns

ISQ Off-Axis Single Quadrupole GC-MS

Page 33: Migrating From He to H Carrier Gas in GC-MS Analysis … Thermo WCTO… · • Thermo Scientific™ your own risk. Hydrogen Systems • ISQ™ Single Quadrupole GC-MS • TRACE™

35

Thank you for your attention!

Questions?

Stay connected with us!

Twitter@ChromSolutions

Chromatography Solutions Bloghttp://chromblog.thermoscientific.com/blog

YouTubehttp://www.youtube.com/ChromSolutions

Facebookhttp://www.facebook.com/ChromatographySolutions

Pinteresthttp://pinterest.com/chromsolutions/