2011‐06‐08
1
Nikon TE2000U
Inverted Fluorescence MicroscopeDépartement de biochimie, UdeM
Nikon TE2000UCondenser Module
• To adjust the heigh of the module
• Polarizer
• Centering screws
Aperture diaphragme (plays on contrast and resolution)
Condenser Turret:•A: brightfield•Ph1: phase contrast (20x)•Ph2: phase contrast (40x)•DIC M: Nomarski (40x with corr. Ring)•DIC H: Nomarski (60x and 100x)
• Condenser lens (O.N.: 0.52)
2011‐06‐08
2
Objective characteristics• Optical Corrections
• Spherical aberrations (PLAN = corrected)
• Chromatic aberrations (APO = corrected for blue, green and red)
• Magnification (here 10x)
• Numerical aperture (here 0,45)
• Dry or immersion (here dry)
• Method(s) of contrast (here Nomarski)
• Working distance in mm (here 4 mm)
• Coverslip thickness to use (here coverslip No 1 ½ of 0,17 mm thick)
• Lens tube length (infini)
NIKON TE2000U TRANSMITED LIGHT MICROSCOPY PARTS
13
14
15
6
5
4
2 ‐ 3
1
9
10
11
8
7
16‐17
18
21
20
7
22
19
12
2011‐06‐08
3
NIKONTE2000UINVERTEDFLUORESCENCEMICROSCOPEPARTSUSEDFORTRANSMITEDLIGHTMICROSCOPY
1. Tungsten lamp2. IR:Infrared filter3. NBC:Neutral Color Balancing filter4. Fielddiaphragm5. Condensermodule6. Stage7. Objectives8. XYmovements9. 1,5Xlens10. Coarse andfinefocusknobs (Donottouch,usetheZ‐motor)11. Opticalpath switchover dial(1‐100%Eyepieces,5‐ 100%Camera)12. Eyepieces13. Eyepiece tube turret (O:open;M:2.5xmagnifier;C:close;B:Bertrandlens)14. FrontPort15. Photomask dial16. D:Diffuser17. ND:Neutral density filter18. Analyser19. Tungsten lamp voltagecontrol20. ZMotor21. Camera22. Dia‐illuminationON/OFFswitch
NIKONTE2000UEPIFLUORESCENCEMICROSCOPYPARTS
28
25
2631
28
22
18
29
24
27
23
30
2011‐06‐08
4
NIKONTE2000UINVERTEDFLUORESCENCEMICROSCOPE
PARTSUSEDFORFLUORESCENCEMICROSCOPY
18. Analyser to remove from light path when observing in fluorescence
22. Dia‐illumination to turn OFF when observing in fluorescence
23. Mercury lamp
24. Uniblitz motorised fluorescence shutter
25. Neutral density filters (ND4 and ND8)
26. Heat absorbing filter (HA)
27. Manuel fluorescence shutter (O: open; C: close)
28. Fluorescence filter cube turret (6 positions)
29. Fluorescence aperture diaphragm (A)
30. Fluorescence field diaphragm (F)
31. UV screen
Starting sequence fordevices
8
6
4
5
1
2
3
7
Fill uplogbook