Upload
julia-yakovleva
View
219
Download
2
Embed Size (px)
Citation preview
Microfluidic enzyme immunosensors with immobilised protein A andG using chemiluminescence detection
Julia Yakovleva a, Richard Davidsson b, Martin Bengtsson c, Thomas Laurell c,Jenny Emneus b,*
a Department of Chemical Enzymology, Faculty of Chemistry, M.V. Lomonosov Moscow State University, Moscow 119899, Russiab Department of Analytical Chemistry, Lund University, P.O. Box 124, SE-221 00 Lund, Sweden
c Department of Electrical Measurements, Lund Institute of Technology, Lund University, P.O. Box 118, SE-221 00 Lund, Sweden
Received 22 August 2002; received in revised form 25 January 2003; accepted 9 March 2003
Abstract
Affinity proteins were covalently immobilised on silicon microchips with overall dimensions of 13.1�/3.2 mm, comprising 42
porous flow channels of 235 mm depth and 25 mm width, and used to develop microfluidic immunosensors based on horseradish
peroxidase (HRP), catalysing the chemiluminescent oxidation of luminol/p -iodophenol (PIP). Different hydrophilic polymers with
long flexible chains (polyethylenimine (PEI), dextran (DEX), polyvinyl alcohol, aminodextran) and 3-aminopropyltriethoxysilane
(APTS) were employed for modification of the silica surfaces followed by attachment of protein A or G. The resulting
immunosensors were compared in an affinity capture assay format, where the competition between the labelled antigen and the
analyte for antibody-binding sites took place in the bulk of the solution. The formed immunocomplexes were then trapped by the
microchip affinity capture support and the amount of bound tracer was monitored by injection of luminol, PIP and H2O2. All
immunosensors were capable of detecting atrazine at the sub-mg l�1 level. The most sensitive assays were obtained with PEI and
DEX polymer modified supports and immobilised protein G, with limits of detection of 0.006 and 0.010 mg l�1, and IC50 values of
0.096 and 0.130 mg l�1, respectively. The protein G based immunosensors were regenerated with 0.4 M glycine�/HCl buffer pH 2.2,
with no loss of activity observed for a storage and operating period of over 8 months. To estimate the applicability of the
immunosensors to the analysis of real samples, PEI and DEX based protein G microchips were used to detect atrazine in surface
water and fruit juice, spiked with known amounts of the atrazine, giving recovery values of 87�/102 and 88�/124% at atrazine
fortification levels of 0.5�/3 and 80�/240 mg l�1, respectively.
# 2003 Elsevier Science B.V. All rights reserved.
Keywords: Microfluidics; Immunosensor; Protein G and A; Hydrophilic polymers; Immobilisation; Chemiluminescence
1. Introduction
Downsizing of analytical equipment is nowadays a
rapidly developing area and has resulted in the concept
known as the micro total analysis system (mTAS) (Reyes
et al., 2002). Integrated miniaturised assay systems have
been applied to the analysis in many different biological
areas, including genomics (Sanders and Manz, 2000),
proteomics (Wilson and Nock, 2002), multianalyte
immunoassays (Silzel et al., 1998), cancer diagnostics
(Askari et al., 2001) and others.
In respect to the immunoassay field, various micro-
fluidic devices have been used to adopt the conventional
immunoassays (ELISA) at microscale, including glass
chips with embedded microbeads (Sato et al., 2001) or
polydimethylsiloxane chips with microfluidic channels
(Eteshola and Leckband, 2001). Miniaturised immu-
noassays make extensive use of electrokinetic phenom-
ena, e.g. electroosmotic flow (McCreedy, 2000; Kricka,
2001) for fluid movement through the microchip chan-
nels, and capillary electrophoresis (CE) as a separation
technique (Dolnik et al., 2000; Hashimoto et al., 2000).
A number of CE-based enzyme immunoassays has been
reported (Koutny et al., 1996; Chiem and Harrison,
* Corresponding author. Tel.: �/46-46-222-4820; fax: �/46-46-222-
4544.
E-mail address: [email protected] (J. Emneus).
Biosensors and Bioelectronics 19 (2003) 21�/34
www.elsevier.com/locate/bios
0956-5663/03/$ - see front matter # 2003 Elsevier Science B.V. All rights reserved.
doi:10.1016/S0956-5663(03)00126-X
1998; Wang et al., 2001), where glass or silica micro-
fluidic devices were used to conduct the assay procedure.
Another attractive biosensing material is porous silicon
with high surface area, which has been applied for thedevelopment of enzyme microreactors (Laurell et al.,
1996; Thust et al., 1999) and microfluidic immunosen-
sors with immobilised antibodies (Yakovleva et al.,
2002).
Immunosensors explore the principles of heteroge-
neous immunoassay and thus require one of the specific
immunoreaction components to be immobilised on the
solid phase. Different assay formats can be established,including the immobilised antibodies (Yakovleva et al.,
2002) or antigen (Brecht et al., 1998) format. An
alternative immunoassay format to overcome the regen-
eration problems typically encountered in the immobi-
lised antibody format is based on affinity proteins such
as protein G and A. The feature of protein A from
Staphylococcus aureus and protein G from group G
Streptococci to selectively bind the Fc region of a widerange of immunoglobulins (IgG) with no interference to
the antigen binding sites, make them very attractive as
affinity-capture supports for immunobiosensing appli-
cations (Valat et al., 2000).
Though a large number of assays employing protein
A or G have been reported within recent years, most of
them in fact use the affinity proteins as a specific
support for oriented antibody immobilisation (Ander-son et al., 1997; Quinn et al., 1999; Garcinuno et al.,
2000; Pulido-Tofino et al., 2000; Sheikh and Mulchan-
dani, 2001). The affinity capture format, however,
implies that competition between the tracer and analyte
for a corresponding antibody takes place in the bulk of a
solution, and the resulting immunocomplexes are cap-
tured by the affinity support downstream (Gonzalez-
Martinez et al., 1998; Penalva et al., 1999; Bjarnasson etal., 2000; Burestedt et al., 2000). Detection of bound
label in the flow systems reported in literature was
performed fluorometrically (Penalva et al., 1999) or by
biosensor (Nistor et al., 2002). Alternatively, the systems
based on monitoring of unbound label used fluoro-
metric (Gonzalez-Martinez et al., 1998; Bjarnasson et
al., 2000), biosensor (Burestedt et al., 2000) or direct
electrochemical (Kronkvist et al., 1997; Wang et al.,2001) detection. In most cases the affinity proteins were
immobilised on the beads and packed into the column.
In this work, we report the development of micro-
fluidic immunosensors based on affinity protein A and
G immobilised on silicon microchip surfaces by different
chemistries, and chemiluminescence (CL) detection.
Among various techniques applied for label detection
in microfluidic immunoassays, including fluorescence(Chiem and Harrison, 1998; Bernard et al., 2001; Hayes
et al., 2001), electrochemical (Lacher et al., 2001; Wang
et al., 2001) and thermal lens microscope (Sato et al.,
2000), one of the most promising techniques is CL.
Application of CL detection technique offers high
sensitivity and is especially suitable at microscale since
no external light source is needed (Hashimoto et al.,
2000). A similar system was recently published based onimmobilised antibodies (Yakovleva et al., 2002). The
aim of the present study was to optimise the affinity
protein immobilisation on the silicon microchip sur-
faces, to evaluate the assay performance in terms of
stability and sensitivity, and to investigate the imple-
mentation of the immunosensors to the analysis of real
samples.
2. Materials and methods
2.1. Materials
Silicon microchips were fabricated by anisotropic wet
etching technique, as described previously (Laurell et al.,
1996; Drott et al., 1997). The microchips had an overall
dimension of 13.1�/3.2 mm, comprising 42 porous flowchannels, each of those were 235 mm deep and 25 mm
wide. The porous surface structure of the channels was
achieved by anodising in a 1:1 mixture of 40% hydro-
fluoric acid and 96% ethanol with a current density of 50
mA cm�2 for 5 min.
Affinity purified polyclonal anti-atrazine IgG fraction
from sheep serum was generously provided by Dr
Ramadan Abuknesha (King’s College University ofLondon, UK). Protein G (recombinant, E. coli ) was
purchased from CalBiochem (La Jolla, CA, USA).
Protein A, polyethylenimine (PEI, MW 750 000, 50%
w/v aqueous) dextran (DEX, MW 64 000�/76 000, clin-
ical grade), horseradish peroxidase (HRP) type VI-A,
glutaraldehyde (GA) 25% v/v aqueous solution grade I,
N ,N ?-dicyclohexylcarbodiimide (DCC), 1-(3-dimethyla-
minopropyl)-3-ethylcarbodiimide (EDC), N -hydroxy-succinimide (NHS), dimethylformamide (DMF),
succinic acid anhydride (SAA), and 3-aminopropyl-
triethoxysilane (APTS) were purchased from Sigma
(St. Louis, MO, USA). Polyvinyl alcohol (PVA, MW
89 000�/98 000, 99% hydrolysed), ammonia, hydrochlo-
ric acid, hydrogen peroxide and toluene were from
Merck (Darmstadt, Germany). Luminol 97% and p-
iodophenol (PIP) 99% were obtained from Aldrich(Milwaukee, MI, USA). Lysine-monohydrate was sup-
plied from Carl Roth Co. (Karlsruhe, Germany), and
dimethylsulfoxide (DMSO) was from Fisher Co. (New
Jersey, USA). The atrazine hapten derivative, used for
tracer synthesis, 6-{{4-chloro-6-[(1-methylethyl) amino]-
1,3,5-triazin-2-yl}amino}hexanoic acid (iPr/Cl/
(CH2)5COOH) was synthesised according to Goodrow
et al. (1990). Atrazine was obtained from Riedel-deHaen (Seelze, Germany). All other chemicals were of
analytical grade. Microcon YM-10 micro-concentrators
were obtained from Amicon, Inc. (Beverley, MA, USA).
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/3422
Milli-Q water was obtained from a Milli-Q purification
water system (Millipore, Bedford, USA).
2.2. Buffers and solutions
The carrier buffer was 0.05 M Tris�/HCl, pH 7.4 and
the substrate buffer was 0.05 M Tris�/HCl, pH 9.0.
When not in use, the microchips were kept in 0.1 M
Tris�/HCl buffer, pH 7.0. Borate buffer (BB) was 0.01 M
sodium tetraborate, pH adjusted to 7.0 with 0.5 M boricacid. The regeneration solution was 0.4 M glycine�/HCl,
pH 2.2. The substrate solution was a mixture of 50 ml of
luminol (50 mM in DMSO), 340 ml of PIP (150 mM in
DMSO), 8.5 ml of 30% w/w H2O2 and 50 ml of substrate
buffer. Working solution of the atrazine-HRP enzyme
tracer was prepared from a 1 mg l�1 stock solution by
diluting 10 000 times with the carrier buffer. Calibration
solutions of atrazine of 0.001, 0.01, 0.1, 1, and 10 mg l�1
were prepared from a 5 mg l�1 water stock solution by
diluting with carrier buffer.
2.3. Immobilisation of protein G and protein A on silicon
microchips
Protein G was immobilised on silicon microchip
surfaces modified with four different hydrophilic poly-
mers: (1) PEI; (2) DEX; (3) PVA, and (4) aminodextran
(AMD), while protein A was immobilised only on
surfaces modified with PEI and DEX and APTS, see
Fig. 1. Silanisation with APTS followed by GA activa-
tion (Weetall, 1976) was used as a reference technique.
Prior to any immobilisation, the silicon microchipswere cleaned in a mixture of 25% NH3, 30% H2O2 and
H2O (1:1:5, by volume) at 100 8C for 5 min, followed by
another treatment in a mixture of 37% HCl, 30% H2O2
and H2O (1:1:5, by volume) at 100 8C for 5 min. The
microchips were then carefully rinsed with water,
ethanol and acetone, and dried under a stream of air.
2.3.1. Immobilisation via APTS
The cleaned microchips were silanised in 10% APTS
in sodium-dried toluene (Weetall, 1976). To remove air
bubbles from the microchip pores, vacuum was appliedfor 1�/3 min. The reaction mixture was refluxed for 1 h
at room temperature (RT) in a sealed vessel, protected
from moisture with a drying tube filled with silica gel.
After removal of the silanisation solution, the micro-
chips were rinsed several times with toluene, acetone and
finally rinsed with BB.
The amino-groups on the APTS silanised surface was
activated in 2.5% v/v GA in BB for 60 min at RT, andfinally rinsed with Milli-Q water in order to remove
traces of GA to avoid cross-linking after addition of the
protein solution. The proteins were attached to the
surface by immersing the microchip in 0.5 mg ml�1
protein A solution in BB and reacted overnight at �/
4 8C under stirring. After 12 h the remaining aldehyde
groups were blocked in 10 mg ml�1L-lysine in BB. The
microchips were then carefully rinsed and stored in 0.1M Tris�/HCl buffer, pH 7.0 at �/4 8C until use.
2.3.2. Immobilisation via PEI
The cleaned microchips were immersed in 0.5% v/v
solution of PEI in BB under stirring at RT overnight,
and then rinsed with BB. The amino-groups of the
adsorbed PEI were activated in 2.5% v/v GA in BB for 2
h at RT under stirring. After careful washing with Milli-
Q water and BB, the microchips were placed in 0.5 mg
ml�1 protein G or protein A in BB, and the reaction
was allowed to proceed overnight at �/4 8C. Thefollowing blocking reactions were performed as de-
scribed in Section 2.3.1.
2.3.3. Immobilisation via DEX
Polyaldehyde dextran was obtained by partial oxida-
tion of DEX with sodium periodate according to a
previously described protocol (Penzol et al., 1998).
Briefly, 60 mg DEX was dissolved in 3 ml of Milli-Q
water and then 87 mg of solid sodium periodate was
added during stirring. The reaction mixture was kept atRT for 2 h and protected from light. Finally the oxidised
DEX was dialysed extensively against 3 l of Milli-Q
water at �/4 8C.
The cleaned microchips were silanised, as described in
Section 2.3.1. The dried microchips were immersed in a
fresh solution of polyaldehyde DEX and then allowed to
react overnight at �/4 8C. After attachment of DEX the
microchips were carefully rinsed with Milli-Q water andput in a 0.5 mg l�1 solution of protein G or protein A
dissolved in BB. The protein-coupling step via aldehyde
groups on DEX was then performed as described in
Section 2.3.1.Fig. 1. Structures of the hydrophilic polymers used for silicon
microchips surface modification.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/34 23
2.3.4. Immobilisation via PVA
The cleaned microchips were silanised, as described in
Section 2.3.1, and the amino groups of APTS-activated
microchips were then reacted in 2.5% v/v GA in BB for 1h at RT to introduce aldehyde groups on the surface,
followed by rinsing with BB.
A 10 mg ml�1 aqueous solution of PVA was prepared
according to the following protocol; 0.5 g PVA was
suspended in 50 ml of Milli-Q water and allowed to wet
for 5 h. The suspension was heated in a water bath for
approximately 10 min until the solution became trans-
parent and then the pH was adjusted to 1.0 withconcentrated HCl. The polymer was immobilised by
immersing the microchips in the PVA solution, while
stirring for 30 min at RT. After the removal of the
immobilisation solution, the microchips were treated
with 5% GA (pH adjusted to 1.0 with concentrated HCl)
for another 30 min to cross-link the polymer matrix
layer and introduce aldehyde-groups on the surface. The
PVA�/GA-activated microchips were then thoroughlyrinsed with Milli-Q water and put in a 0.5 mg ml�1
protein G solution in BB. The protein attachment was
then carried out as described in Section 2.3.1.
2.3.5. Immobilisation via AMD
AMD was prepared by oxidation of DEX followed by
reductive amination, using a modified protocol reported
earlier (Piehler et al., 1996). The oxidation was carried
out using 20 ml of 0.36 mM DEX aqueous solution and0.15 M sodium periodate. The reaction mixture was
stirred for 2 h at RT, after that the oxidised DEX was
purified by dialysis against 3 l of Milli-Q water for 2
days at �/4 8C.
A 100-fold molar excess (0.7 mmol) of ammonium
chloride was added to the dialysed DEX solution, and
pH was adjusted to 6, followed by heating to 100 8C. A
50-fold molar excess (0.35 mol) of sodium cyanobor-ohydride was added to the reaction mixture in several
portions at 100 8C during 3 days. The AMD product
was purified by dialysis against Milli-Q water for 2 days
at �/4 8C, and then filtered through a 0.45 mm nitro-
cellulose filter. A 10-fold excess of methanol (by volume)
was added to the filtrate (20 ml) and the pH was
adjusted to 4.0 to precipitate AMD. Finally the product
was filtered on a 0.45 mm nitrocellulose filter and dried.To immobilise AMD, the pre-cleaned microchips
were first silanised with APTS, according the protocol
described in Section 2.3.1. The introduced amine groups
were then derivatised to carboxylic groups by succinila-
tion according to the following protocol (Piehler et al.,
1996). The APTS-functionalised microchips were im-
mersed in 10 ml of BB, and 1 g of solid SAA was added.
The pH of the solution was maintained at 6.0 bycontinuous addition of 1 M NaOH, while the reaction
mixture was stirred and SAA dissolved, i.e. after 1 h.
The reaction was allowed to proceed for another hour
and after which the microchips were carefully rinsed
with BB and dried.
Finally, the AMD polymer was immobilised by
immersing the SAA-treated microchips in 15 ml 10 mgml�1 AMD aqueous solution, with pH adjusted to 3
with 2 M HCl, followed by addition of 50 mg of EDC.
The reaction mixture was stirred overnight at RT. The
AMD coated microchips were then carefully rinsed with
Milli-Q water and activated with 5% v/v GA in BB for 2
h at RT under stirring (the GA concentration was higher
than that applied for the PEI activation in order to
cross-link the AMD layer and provide free aldehydegroups capable of conjugating to the proteins). After
washing with BB to remove the excess of GA, the
microchips were immersed in 2 ml of 0.7 mg ml�1 of
protein G solution. The protein immobilisation was
performed as described in Section 2.3.1.
2.4. Synthesis of enzyme tracer
The enzyme tracer was synthesised by attachment ofthe iPr/Cl/(CH2)5COOH atrazine derivative to HRP by
the NHS ester method (Giersch, 1993). The hapten (1
mg), NHS (1.7 mg) and DCC (6.2 mg) were mixed with
130 ml of DMF. The mixture was agitated overnight at
RT. Then the activated ester solution was added drop-
wise to a solution containing 1 mg HPR in 0.5 ml of 0.13
M NaHCO3. The coupling reaction was performed for 3
h, after which the tracer was purified by dialysis against3 l of PBS (five changes of buffer) using a Slide-A-Lyzer
cassette (Pierce, Rockford, IL, USA) with a molecular
cut-off of 10 kDa.
2.5. System set-up and assay procedures
The scheme of the microfluidic system is shown in
Fig. 2a. A CMA Microdialysis 100 syringe pump(CMA/Microdialysis, Solna, Sweden) was used to de-
liver the 0.05 M Tris�/HCl pH 7.4 carrier buffer, at
either 40 or 50 ml min�1. For regeneration of the
immunosensor surface the flow was switched to 0.4 M
glycine�/HCl buffer via a Gilson Minipuls 2 peristaltic
pump (Villiers-le-Bell, France). A Rheodyne six-port
valve (Berkeley, CA, USA) equipped with a 4 ml loop
was used for injecting the samples. The microchipimmunosensor was incorporated in the flow system via
a specially designed flow cell unit (Fig. 2b) made of plexi
glass with the inlet and outlet tubing glued into holes
drilled in the top cover. A thin transparent silicon
rubber membrane was placed between the microchip
and the top cover to prevent leakage. The CL signal was
monitored via a photomultiplier tube (PMT, model no.
HC135-01 UV to visible, Hamamatsu Photonics K. K.,Japan), aligned right above the microchip flow cell with
the surface of immobilised affinity proteins exposed.
The microchip flow cell and the PMT was placed in a
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/3424
holder to assure correct positioning of the two units and
the whole detection was performed in a ‘black box’ to
shut-out light from the surroundings, as seen in Fig. 2a.
The total volume of the flow system was 20 ml with
connecting PEEK tubing (i.d. 0.25 mm, Alltech, Deer-
field, IL, USA). Data acquisition was performed with a
homemade software.
The assay procedure is shown in Fig. 3 and is based
on the principle of affinity capture competitive immu-
noassay, according to the following steps: enzyme
tracer, atrazine standard or a sample, and anti-atrazine
antibody were mixed off-line and injected directly, or
after off-line pre-incubation for 20�/30 min, into the
system, at 40 ml min�1. The microchip was rinsed with
the carrier buffer for 2 min at 40 ml min�1 to remove
unbound immuno-complexes. Next, the substrate mix-
ture (luminol/PIP/H2O2) was injected at 50 ml min�1
and the enzyme-catalysed chemiluminescent reaction
taking place on the microchip surface was monitored
via the PMT. To complete the assay cycle, removing the
immunocomplex from the affinity proteins, the carrier
buffer was switched to 0.4 M glycine�/HCl buffer pH 2.2
via the Gilson Minipuls 2 peristaltic pump at 50 ml
min�1 for 2�/3 min. The total operation time for oneassay cycle, including rinsing, three repetitive injections
of substrate and regeneration step, was approximately
10 min.
2.6. Analysis of fortified surface water and orange juice
samples
Spiked surface water samples, from Hoje a, a river
that passes through a very active agricultural region in
Lund, Sweden, were applied to the DEX modified
Fig. 2. (a) Scheme of the microfluidic immunosensor manifold: a syringe pump and a peristaltic pump were used for carrier buffer and regeneration
solution at flow rates of 40 and 50 ml min�1, respectively. The sample, containing the enzyme tracer, atrazine standard and antibody, pre-mixed off-
line, was injected through a six-port injection valve, and the chemiluminescent signal was detected by a PMT placed above the flow cell, containing
the microfluidic immunosensor. (b) Detailed view of the plexi glass microchip flow cell and a magnified image of the microchip channel network.
Fig. 3. Scheme of the affinity capture competitive immunosensor format performed in the system shown in Fig. 2.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/34 25
protein G immunosensor. The surface water samples
were filtered through a 0.45 mm nitrocellulose filter
(Millipore) and analysed without any pH adjustments.
Undiluted sample and samples diluted 1:2 and 1:4 withthe carrier buffer were used to construct a calibration
curve for atrazine and to evaluate any potential matrix
effects. To estimate the assay recoveries, the surface
water was spiked with known amount of atrazine to give
the analyte concentrations of 0.5, 1 and 3 mg l�1 and
analysed by DEX modified protein G immunobiosen-
sor. Three assays were performed in triplicate and the
respective CL signals were used to estimate atrazineconcentration by interpolation in the calibration curve,
constructed the same day as the sample measurement.
Recovery values and standard deviations (S.D.) were
calculated for each sample and expressed as a percen-
tage ratio.
In a similar way, orange juice samples from commer-
cial sources were tested for matrix effects. The juice
samples were centrifuged at 10 000 rpm for 5 min, thesupernatant was filtered through a glass microfibre filter
GF/A, (Cat No 1820 047, Whatman, Maidstone,
England). The carrier buffer was used to dilute the
samples to give a juice to carrier buffer ratio of 1:50,
1:100, 1:200, 1:400 and 1:500. Calibration curves for
atrazine were obtained using the analyte standards in
carrier buffer and in diluted orange juice. To determine
the recoveries of atrazine, the juice samples were spikedto give concentrations of 80, 160 and 240 mg l�1
atrazine, and assayed by the PEI modified protein G
immunosensor. Three assays were performed in tripli-
cate and used to estimate the atrazine concentration in
the spiked sample and to calculate the recovery values
and S.D. for each sample. The recovery (R ) values were
calculated according to the formula:
R�CIA � CIA;0
Cspiked
where CIA is the analyte concentration in the spiked
sample evaluated by interpolation in the standard curve,
CIA,0 is the concentration of analyte in the sample
without spiking and Cspiked is the calculated concentra-
tion of the analyte spiked into each sample.
3. Results and discussion
The development of highly sensitive, precise, repro-
ducible, rapid, and cost-efficient assay techniques, im-
plying a fast analysis of a large number of samples, is
general goal that would fulfil the growing demands on
environmental monitoring. Miniaturisation of assays isnow an increasingly useful approach, which allows not
only to detect minute amounts of analyte, but also to
build up array systems for multiplex analysis, and, in
addition, offers experimental simplicity of handling
(Auroux et al., 2002).
Recently we reported microfluidic immunosensors
based on the same approach as presented here, butwith immobilised antibodies, displaying sensitivities in
the low ng l�1 range, thus exceeding those of conven-
tional batch and flow immunoassays previously re-
ported for this analyte (Yakovleva et al., 2002). Our
results indicated that the immunosensor performance
was largely a function of the immobilisation chemistry
used on the silica surface. A comparison of the attach-
ment of antibodies to APTS functionalised silica sup-port and to a hydrophilic layer of PEI, revealed an
appreciable improvement of the assay sensitivity using
the latter method. Significant discrepancies were also
found between the stability, sensitivity, and reproduci-
bility of the antibody coatings formed by the use of
different molecular weight of PEI covalently attached or
physically adsorbed on the silica surface, thus implying
the importance of optimising the immobilisation chem-istry of the sensing surface for the development of
sensitive, stable, and reusable immunosensors.
The immobilised antibody format is, however, af-
fected by the reduction of antigen binding capacity and
hence decreased sensitivity with time due to exposure of
the antibody to harsh regeneration conditions as re-
quired for the desorption of the antigen after each assay
cycle. The present work extends the application of thesesilicon microchips to the affinity capture assay format as
shown in Fig. 3, using protein A or protein G
immobilised in hydrophilic matrices of various poly-
mers, and the same anti-atrazine antibodies as in our
previous work (Yakovleva et al., 2002). This format
extends the flexibility of the assay, because a whole
range of antibodies with different specificity can be
bound and dissociated from protein A and G coatedsurfaces. Thus, the employment of the affinity capture
proteins allows the development of generic assays,
potentially applicable to any analyte.
In this work, different immobilisation chemistries for
protein G and A on silicon microchips were investigated
and the resulting immunosensors evaluated in terms of
stability, reusability and sensitivity, and finally applied
for the analysis of surface water and fruit juice samples.
3.1. Hydrophilic matrices for immobilisation of affinity-
capture proteins
Long flexible hydrophilic polymers are known to
provide a beneficial microenvironment for the immobi-
lisation of various biological macromolecules. Hydro-
philic matrices have been put in use to achieve an
improved performance of batch immunoassays likeELISA (Gregorius et al., 1995), SPR sensor chips (Lofas
et al., 1993), and piezoelectric biosensors (Liu et al.,
2001).
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/3426
Four different hydrophilic polymers were employed
and compared in this work for immobilisation of the
affinity capture proteins: (i) PEI (MW 750 000 Da); (ii)
DEX (MW 64 000�/76 000 Da); (iii) PVA (MW 89 000�/
98 000, (iv) AMD, and finally (v) APTS used as a
reference to estimate the efficiency of the hydrophilic
polymer modification.
Matrices based on DEXs have for many years been
used in size-exclusion chromatography, and recently
DEX coatings have also been recognised as efficient
supports for immobilisation of antibodies and other
proteins. Carboxymethyldextran (Howell et al., 1998)and periodate-oxidised DEX (Penzol et al., 1998) are the
most widely used DEX linkers for surface modification.
Utilisation of DEX as a coupling matrix for protein A
has been shown to improve the reusability under acidic
regeneration conditions and increase the sensitivity,
binding capacity and specificity (Lu et al., 2000). The
topology of DEX coatings on a silica surface is
dependent on the polymer molecular weight, and thedensity of DEX clusters decreases with the increase of
molecular weight (Tasker et al., 1996).
Antibody immobilisation in a hydrophilic PEI layer
has been reported in a number of publications (Babacan
et al., 2000; Yakovleva et al., 2002), where improved
sensor stability and reusability in comparison with
conventional APTS-functionalised supports were
achieved. The ability of PEI to form a thin polymerlayer irreversibly bound to the surface of silicon dioxide
by direct adsorption makes this polymer very attractive
for fast and efficient surface modification procedure for
antibody immobilisation.
PVA matrix is widely applied for the immobilisation
of enzymes (Rossi et al., 1999) and cells (Khoo and
Ting, 2001), however, the use of this polymer as a
functionalised support in immunoanalytical applicationsis limited by a few examples (Barbosa et al., 2000). The
stabilising effect of PVA on the activity of immobilised
antibodies was reported for long-term storage of ad-
sorbed IgG (Dankwart et al., 1998).
AMD is a less common polymeric support for
immobilisation of biospecific ligands, mainly due to
the elaborate protocol of preparation. Nonetheless,
several applications of AMD for optical immunosensorsbased on immobilised antigens were reported in litera-
ture, demonstrating the lowest levels of non-specific
binding and the maximal binding capacity of AMD-
modified immunosurfaces (Piehler et al., 1996; Brecht et
al., 1998).
Considering the particular feasibility of hydrophilic
polymers to improve the performance of immobilised
biomolecules, we applied different coupling matrices totether affinity proteins to the microchip surface, and
performed a comparative study of stability and sensi-
tivity of the resulting microfluidic immunosensors.
Protein A and G in combination with PEI, DEX,
PVA, AMD polymers or conventional APTS support
were used to establish the most efficient immobilisation
procedure.
3.2. Optimisation of assay procedure
A systematic study of assay parameters (immunor-
eagent concentrations, nature of affinity protein, on-
chip incubation and off-line preincubation time of
immunoreagents) that might affect the antibody�/anti-
gen reaction, and also the binding of immunocomplexes
was performed.
The amount of antibodies is an important factorinfluencing the sensitivity of a competitive immunoas-
say, and lower antibody concentrations should be used
to facilitate a true competition and reach the most
sensitive assay (Hock, 1997). Optimal working antibody
concentrations for APTS, PEI and DEX modified
immunosensors with immobilised protein A and G
were determined by recording the antibody dilution
curves, and summarised in Table 1. As seen from thesedata, protein G immunosensors result in lower optimal
working antibody concentrations compared with pro-
tein A, with DEX modified protein G sensor giving the
lowest value. Table 1 also shows the parameters of
competitive affinity capture assays performed with
atrazine as a model analyte. The values of calibration
curve midpoint (IC50) are consistent with the optimal
antibody concentration obtained for immunosensorswith immobilised protein A and G, displaying the
most sensitive atrazine calibration curve with DEX-
protein G as an affinity capture support. The presented
data also indicate the effectiveness of PEI modification
of silica surface for immobilisation of protein A, as
follows from the comparison of IC50 values for APTS-
and PEI-modified sensors with immobilised protein A.
Based on the optimal antibody conditions, theimmunosensors with protein G were chosen for further
experiments due to lower reagents consumption and
superior assay sensitivity. For additional optimisation of
affinity protein immobilisation procedure we also pre-
pared protein G immunosensors modified with PVA
and AMD. Optimal antibody concentrations for PVA
and AMD modified immunosensors were 5 and 25 nM,
respectively, however, AMD sensor required the appli-cation of higher tracer concentration (9 nM).
In the next step, different protein G immunosensors
were used to investigate the immunoreagent incubation
time. The assays were performed with or without on-
chip incubation of antibody and labelled antigen, thus
influencing the kinetics of immunocomplex affinity
capture by protein G (Fig. 4a and b). The on-chip
incubation of the immunocomplex did not have aconsiderable influence on the binding response obtained
with the PEI and DEX sensors (Fig. 4a), whereas 2�/5
min incubation was required for stable and intense
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/34 27
response using the PVA and AMD sensors (Fig. 4b).
The importance of on-chip incubation for the PVA and
AMD sensors may indicate the steric hindrance arising
from densely cross-linked polymer layers, which inter-
feres with the binding of the large immunocomplex.The effect of off-line pre-incubation of the antibody,
enzyme tracer and analyte on the assay performance was
investigated using the protein G based-immunosensors
and the optimised on-chip incubation conditions. Atra-
zine calibration curves were run by 20 min pre-incuba-
tion of the components or by immediate injection of the
immunoreagents to the microfluidic system. As seen in
Table 2, the assay parameters varied dramatically for
PEI and DEX based assays, leading to a much better
limit of detection (LOD) and curve midpoint (IC50)
values without any pre-incubation compared with the
PVA based assay. Off-line pre-incubation for 20 min
resulted, however, in a significant drop in sensitivity for
the PEI and DEX based assays. Taking into considera-
tion that antibody affinity constant toward the labelled
antigen is usually higher than that toward the target
analyte (Hatzidakis et al., 2002) one can assume that the
equilibrium for the formation of immunocomplex dur-
ing the off-line pre-incubation will shift toward the
binding of enzyme tracer, thus leading to a decrease in
assay sensitivity. For the PVA microchip, the observed
sensitivity was practically equal for both 20 min pre-
incubation and immediate injection of the components.
This fact may account for the influence of immunocom-
plex on-chip incubation, which is necessary to obtain the
intense and reproducible CL signals. Thus, in the case ofPVA microchip the equilibrium shift toward the forma-
tion of immunocomplex with enzyme tracer may occur
during the on-chip incubation step. As a result, the
observed sensitivity for PVA microchip is lower than
that for PEI and DEX microchips, and not dependent of
off-line pre-incubation of the components.
3.3. Regeneration and short- and long-term stability
Regeneration of immunosurfaces is a key factor inimmunosensor development, which implies, in regard to
affinity capture surfaces, the desorption of bound
immunocomplex without affecting the activity of im-
mobilised proteins or damaging the bonds between the
protein, modifying polymer layer, and silica surface.
In comparison with immobilised antibodies, protein
A and G are known to require more drastic conditions
for desorption of bound material, whereas the stabilityof immobilised affinity proteins is much higher than for
antibodies (Gonzalez-Martinez et al., 1998; Gonzalez-
Martinez et al., 1999; Quinn et al., 1999). Based on our
previous findings (Yakovleva et al., 2002), the best
recovery of immobilised antibody layer was achieved by
Table 1
Optimal antibody concentrations and curve midpoint values (IC50) for different protein A and G immunosensors
Microchip immunosensor APTS-protein A PEI-protein A DEX-protein A PEI-protein G DEX-protein G
[Ab] (nM) 5 5 5 4.2 3.1
IC50 (mg l�1) 0.4719/0.132 0.1349/0.013 �/ 0.0889/0.012 0.0729/0.005
Optimal antibody concentrations were determined from antibody dilution curves as values, corresponding to 70% binding of enzyme tracer. IC50
values were determined from atrazine calibration curves, obtained at optimal antibody concentration. Tracer concentration was 4.5 nM in all
experiments. The conditions were as follows: carrier buffer 50 mM Tris�/HCl (pH 7.4), substrate buffer 50 mM Tris�/HCl (pH 9.0), regeneration
solution 0.4 M glycine�/HCl (pH 2.2), carrier flow rate 40 ml min�1, substrate flow rate 50 ml min�1.
Fig. 4. Effect of on-chip incubation time on the immunocomplex binding response. A mixture of antibody and enzyme tracer taken at optimal
concentrations (Table 1 and Section 3.2) was injected to the microfluidic system (4 ml). The mixture was either allowed to pass through the affinity
capture surface (40 ml min�1) without incubation (peaks 1) or the carrier flow was stopped when the mixture reached the immunosensing surface, and
incubated for 2�/5 min (peaks 2). The flow cell was washed for 2 min, and a CL signal after triplicate substrate injection (50 ml min�1) was registered.
(a) Shows the peak pattern for protein G immunosensors modified with DEX and PEI. (b) Shows the peak pattern for protein G immunosensors
modified with AMD and PVA.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/3428
the application of acidic pH, and, especially by using
glycine�/HCl buffer (0.4 M, pH 2.2). The same buffer
was used in this work to remove the bound immuno-
complex from protein A and G sensing surfaces.
For comparison of the reusability and thus short-term
stability of immunosurfaces based on different poly-
mers, the immunocomplex binding response was mon-
itored after several repetitive binding-desorption cycles.
Fig. 5a and b show the regeneration profiles given by
different immunosensors based on protein A and G,
respectively. Each binding cycle was accompanied by
three substrate injections in order to monitor the signal
reproducibility, which can be impaired by dissociation
of antibody�/affinity protein complex. As seen in Fig.
5a, the protein A coatings displayed poor reproduci-
bility due to the elution of bound immunocomplex
within one binding cycle (triplicate substrate injection)
by the carrier flow. The protein G coatings were more
reproducible, with the best short-term stability exhibited
by PEI and DEX modified immunosensors (Fig. 5b).
Such behaviour of protein A and protein G may be
explained by different affinity of these proteins to the
sheep IgG used in this work (Hermanson et al., 1992).
Since protein A coatings did not display sufficient
reproducibility and stability within regeneration experi-
ments, further studies were performed with protein G
immunosensors. Estimation of the immunosensor long-
term stability after long periods of storage and use was
based on the following criteria: (i) the ability to maintain
a stable zero analyte dose signal and (ii) the reproduci-
bility of the whole calibration curve and its parameters.
Fig. 6 shows the immunocomplex binding response
obtained on different days for protein G immunosensors
modified with PEI, DEX, and PVA. As seen from these
data, the immunocomplex binding response observed
for PEI and DEX based immunosensors is consistent
throughout a period of 260 days. No decrease of surface
binding capacity was registered for these immunosen-
sors, and in addition, a minor activation effect was
observed after long period of storage (between day 30
and 180). The PVA immunosensor displayed a gradual
decrease of binding response already within a 2-week
Table 2
Influence of off-line pre-incubation on assay parameters using PEI, DEX and PVA modified protein G immunosensors
Immunosensor No off-line pre-incubation Off-line pre-incubation for 20 min
PEI DEX PVA PEI DEX PVA
LOD (mg l�1) 0.006 0.005 0.065 0.15 0.17 0.17
IC50 (mg l�1) 0.089/0.02 0.089/0.03 0.859/0.31 1.229/0.16 0.969/0.11 0.749/0.08
Slope 0.599/0.08 0.419/0.12 0.629/0.28 0.969/0.11 1.199/0.08 1.299/0.16
Atrazine calibration curves with or without off-line pre-incubation were run on the same day. The curve midpoint (IC50) and a slope were extracted
from a four-parameter logistic equation used to fit the sigmoidal curve. LOD values were defined as the analyte concentration corresponding to 90%
tracer binding. Antibody concentrations for PEI, DEX and PVA modified immunosensors were 4.2, 3.1 and 5 nM, respectively. On-chip incubation
for PVA immunosensor was 2 min. PEI and DEX immunosensors were used without on-chip incubation. Tracer concentration and flow conditions
were the same as in Table 1.
Fig. 5. Short term stability of immunosensors. The intensity of CL signal after three consecutive cycles of binding (injection of antibody and enzyme
mixture; 4 ml; 40 ml min�1) and desorption (a flow of 0.4 M glycine�/HCl, pH 2.2; 3 min; 50 ml min�1) is plotted vs. time. (a) Shows the
immunocomplex binding response for protein A immunosensors. On-chip incubation for 2 min, antibody and tracer conditions were as in Table 1.
(b) Shows the immunocomplex binding response for protein G immunosensors. No on-chip incubation was performed for PEI and DEX modified
sensors, antibody and tracer conditions were as in Table 1. On-chip incubation for 2 min was applied for PVA and AMD modified sensors, antibody
concentrations were 5 and 25 nM, respectively. Tracer concentration was 5 nM.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/34 29
period, thus showing much lower stability than PEI and
DEX based immunosensors.
The monitoring of atrazine calibration curve varia-
bility with time for DEX and PEI modified immuno-
sensors containing immobilised protein G is presented in
Fig. 7a and b, respectively. Atrazine calibration curves
coincide very well for the immunosensor modified with
DEX (Fig. 7a), whereas for PEI more dramatic increase
of the IC50 value and thus decrease of sensitivity was
observed after 215 days of storage and use (Fig. 7b).
Inter-assay precision for both PEI and DEX based
immunosensors (R.S.D. within the linear range of the
calibration curve) was below 15% over 8-month period,
thus evidencing that both immunosensors are capable of
long-term use.
An interesting observation is that PEI and DEX
modified protein G immunosensors exhibited similar
and very high long-term stability, although the PEI
modification is performed by physical adsorption of thepolymer on the silica surface, while DEX modification is
based on covalent attachment of the polymer to the
microchip surface. Immobilisation of antibodies on a
PEI layer performed in our previous work (Yakovleva et
al., 2002) resulted in immunosensor lifetime of 1 month,
whereas this work demonstrates that the significant
prolongation of PEI based immunosensor lifetime may
be achieved by making use of the affinity capture assayformat.
3.4. Analytical performance
Investigation of the analytical characteristics of the
atrazine assay involved a comparison between different
hydrophilic polymers used for modification of silicon
microchip surface and immobilisation of protein G. The
parameters of the calibration curves obtained with PEI-,
DEX-, PVA-, and AMD-protein G immunosensors are
given in Table 3. As seen from these data, application ofdifferent matrices for protein G immobilisation signifi-
cantly influenced the assay performance, where the PEI-
and DEX modified immunosensors were found to be the
most sensitive. A correlation was found between the
optimal antibody conditions (Table 1 and Section 3.2)
and the sensitivity of the calibration curve. The PEI and
DEX microchip modification procedures resulted in the
lowest optimal immunoreagent concentrations and mostsensitive atrazine calibration curves, thus being the most
suitable choices for affinity protein immobilisation.
Fig. 6. The long-term stability of PEI, DEX, and PVA based
immunosensors with immobilised protein G. Immunocomplex binding
responses at zero analyte doze were measured on different days and
plotted vs. time over. Average signals from three repetitive binding-
desorption cycles are shown. The conditions were the same as in Fig. 5.
Fig. 7. Variability of atrazine calibration curves obtained with protein G immunosensors modified with PEI and DEX. Calibration curves
constructed on different days were fit to a four parameter logistic equation: y�/(A�/D)/[1�/(x/C)B]�/D, where A and D are the maximal and minimal
signals of the assay, respectively, B corresponds to the slope of the sigmoid and C is the concentration of atrazine resulting in 50% inhibition of
binding of the enzyme-tracer to antibody (IC50). (a) Atrazine calibration curves using the DEX based immunosensor. (b) Atrazine calibration curves
using the PEI based immunosensor. Each curve is an average of three replicates, performed on each day. (�/) Represents the IC50 values (right y-
axis) plotted vs. time (top x-axis), extracted from the corresponding four-parameter curves. Antibody and tracer conditions were as in Table 1. No
on-chip incubation or off-line pre-incubation was performed.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/3430
Despite the different analytical characteristics dis-
played by the protein G immunosensors, the LODvalues of all assays with exception for the PVA-based
were below the threshold level for single pesticide
content in drinking water according to official European
Union (EU) requirements (0.1 mg l�1). PEI and DEX
modified protein G immunosensors were chosen for
further analysis of surface water and fruit juice samples,
considering their good stability (Figs. 6 and 7) and that
the obtained IC50 values were around 0.1 mg l�1.
3.5. Analysis of spiked surface water samples
Surface water collected from a small river in an
agricultural region in southern Sweden was used for
this study. In order to investigate the matrix effect,
surface water samples were spiked with atrazine to give
different analyte concentrations between 0 and 10 mg
l�1. Crude surface water and surface water diluted with
50 mM Tris�/HCl buffer at various ratios were used toprepare the spiked samples and construct atrazine
calibration curves, see Fig. 8. No calibration curve
could be obtained with atrazine standards prepared in
crude surface water, the assay resulted in very low light
intensity, indicating strong matrix influence. Water
sample diluted 1:2 in carrier buffer lead to the calibra-
tion curve being shifted to the range of higher analyte
concentrations, with zero analyte dose signals approxi-mately 1.5 times lower than that obtained in the carrier
buffer. Almost identical calibration curves were ob-
served for Tris�/HCl and surface water diluted with
buffer 1:4.
Surface water was fortified with atrazine added at 0.1,
1 and 3 mg l�1 and then diluted 1:4 with carrier buffer,
i.e. after dilution the obtained concentrations fell into
the assay working range. The mean recovery data for thethree spiked samples were 87�/102% as shown in Table
4, thus confirming that 1:4 sample dilution in the carrier
was sufficient to minimise interference for surface water
matrix, however, as seen in Table 4, the precision,
especially for higher concentrations, was rather low.
These finding must be taken into account for assay
optimisation in any future work.
3.6. Analysis of spiked fruit juice samples
Food matrix effects are well-known phenomena inimmunoassays, and hence the influence of orange juice
on the performance of the DEX-immunosensor was
investigated in order to estimate the applicability to
food analysis. The atrazine recovery and matrix effects
were tested in orange juice in a similar way as for surface
water. Orange juice was diluted by various factors (1:10,
1:50, 1:200, 1:400 and 1:500) with carrier buffer. The
influence of juice dilution on the zero analyte dose signalis shown in Fig. 9. As can be seen, at least 1:400 dilution
is required in order to eliminate the matrix effects. The
obtained results indicate that the presented protein G
Table 3
Assay characteristics for protein G immunosensors modified with different hydrophilic polymers
Hydrophilic polymer PEI DEX PVA AMD
LOD (mg l�1) 0.006 0.010 0.202 0.051
IC50 (mg l�1) 0.0969/0.018 0.1309/0.027 0.8569/0.307 1.6049/0.597
Dynamic range (mg l�1) 0.014�/1.120 0.014�/0.822 0.173�/4.580 0.077�/13.58
R.S.D.intra (%) 4.98 5.62 10.84 5.75
R.S.D.inter (%) 15.33 11.45 13.26 23.10
The LOD was calculated as atrazine concentration corresponding to 90% tracer binding. The dynamic range was defined by the analyte
concentration inhibiting the maximum signal by 20�/80%, and the IC50 value was estimated as the analyte concentration that inhibits 50% of tracer
binding. The intra-assay precision (R.S.D.intra) was calculated as the average R.S.D. between three replicates within one selected calibration curve,
using all points of the assay dynamic range. The inter-assay precision (R.S.D.inter) was calculated as the R.S.D. between three different averaged
assays performed on different days, including all points (N�/3) in the dynamic range of the assays. Antibody conditions for PEI and DEX modified
immunosensors were as in Table 1, without on-chip incubation. Antibody concentrations for PVA and AMD modified immunosensors were 5 and 25
nM, respectively, on-chip incubation 2 min. Tracer concentration for PEI, DEX and PVA sensors was 4.5 nM, for AMD sensor 9 nM. No off-line
pre-incubation was performed in all experiments. Flow conditions were as in Table 1.
Fig. 8. The surface water matrix influence on the DEX immunosensor.
Atrazine calibration curves were constructed using surface water at
various dilutions or the carrier buffer (50 mM Tris�/HCl, pH 7.4)
fortified with atrazine at 0.01, 0.1, 1 and 10 mg l�1. The plotted signals
are averages of three measurements. The conditions are as in Fig. 5.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/34 31
immunosensors with chemiluminescent detection are
very sensitive to interference from the complex vegetable
matrices.
In order to estimate atrazine recoveries, orange juice
was spiked with atrazine at 80, 160 and 240 mg l�1 to
give analyte concentrations within the assay working
range after a 1:400 dilution in the carrier buffer. The
spiked samples were analysed using the calibration curve
obtained the same day. Table 5 shows the results
obtained for each fortified juice sample determined in
triplicate with recoveries between 88�/124%. In this case,
however, better precision was obtained since R.S.D.
values did not exceed 7%. For environmental samples
the US EPA recommendation is that the recovery
should fall between 70�/120% (Krotzky and Zeeh,
1995) and based on this the sensor is suitable for
atrazine quantification in juice.
4. Conclusions
Microfluidic immunosensors based on affinity pro-
teins immobilised on silicon microchip surfaces and CL
detection were developed. Various hydrophilic polymer
coatings were employed for immobilisation protein A
and G and used for comparative study to establish the
best activity, reusability and sensitivity of the sensor.
Affinity capture assay based on protein A or G
represents a generic tool, applicable for a wide range
of analytes.
Of the two affinity proteins used in this study, protein
G resulted in better reusability and long-term stability of
the immunosensors compared with protein A. The
matrix of the hydrophilic polymer was an other para-
meter influencing the stability and sensitivity of the
assay, where the PEI and DEX modified microchips
offered the most sensitive detection. The other polymers
used here, PVA and AMD, resulted in worse reusability
and sensitivity of protein G immunosensors. Addition-
ally, the assay characteristics were found to be substan-
tially dependent on the incubation conditions, where
off-line pre-incubation of the tracer, analyte and anti-
body led to an appreciable loss in sensitivity. Based on
the optimised immobilisation chemistry and assay con-
ditions, we developed microfluidic immunosensors for
determination of atrazine, with calibration curve mid-
points around the EU threshold lever for drinking water
(0.1 mg l�1) and long-term stability over a period of 8
months.
Due to the high sensitivity provided by PEI and DEX
coated protein G immunosensors, they were applied for
atrazine determination in surface water and juice
samples. Surface water matrix effects could be mini-
Table 4
Recoveries in surface water fortified with atrazine
Fortification concentration (mg l�1) Found concentration (mg l�1) S.D. Recovery (%9/R.S.D.)
0.5 0.510 0.028 1029/15
1 0.880 0.279 889/32
3 2.623 0.684 879/26
Surface water was first spiked with known amounts of atrazine and then diluted 1:4 with 50 mM Tris�/HCl buffer (pH 7.4) and analysed using the
DEX immunosensor. Averages of three replicates obtained by interpolation into a calibration curve constructed the same day as the measurement
was performed. Antibody, tracer and flow conditions were as in Table 1, no on-chip or off-line incubation was performed.
Fig. 9. The orange juice matrix influence on the PEI immunosensor is
shown. Tracer signals in the presence of zero analyte doze are plotted
at different dilutions of juice sample with 50 mM Tris buffer, pH 7.4.
Averages of three measurements are shown.
Table 5
Recoveries in orange juice fortified with atrazine
Fortification concentration (mg l�1) Found concentration (mg l�1) S.D. Recovery (%9/R.S.D.)
80 99.1 6.9 1249/7
160 163.5 10.8 1029/7
240 210.0 13.0 889/6
Orange juice was first spiked with known amounts of atrazine and then diluted 1:400 with 50 mM Tris�/HCl buffer (pH 7.4) and analysed using the
PEI immunosensor. The calibration curve and sample assay performed in triplicate the same day were used to calculate the average recoveries and
R.S.D. of the samples. Antibody, tracer and flow conditions were as in Table 1, no on-chip or off-line incubation was performed.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/3432
mised by a 1:4 dilution of the analysed sample, and the
recovery of atrazine in surface water spiked with
concentrations between 0.5 and 3 mg l�1 was 87�/
102%. The application to the analysis of orange juicesamples required at least 1:400 dilution, with recoveries
between 88 and 124% for juice spiked with concentra-
tions between 80 and 240 mg l�1 atrazine. The total
assay time was 10 min including, sample injection, three
substrate injections and regeneration.
High sensitivity, versatility, suitability for point-of-
care analysis, high sample throughput and possibility to
build up array systems are important advantages ofmicrofluidic immunosensors, which make these kind of
devices very promising for the future application in
many fields, including potential multiple biological
information and environmental multiplex analysis using
channel arrays with different sensing elements immobi-
lised. For the future we anticipate further development
of a multi-assay platform, making use of different
immobilised biorecognition elements, such as receptors,antibodies, enzymes and cells.
Acknowledgements
The authors acknowledge financial support from the
European Commission (INCO Copernicus project ER-
BIC15-CT98-0910), the Swedish Foundation for Strate-
gic Environmental Research (MISTRA), the Swedish
Council for Forestry and Agricultural Research (SJFR),and the Swedish Research Council (Vetenskapsradet).
Authors also wish to thank Dr Ramadan Abuknesha
(King’s College University of London, UK) for provid-
ing us with anti-atrazine antibodies and Dr Oleg
Kolyasnikov (Department of Chemical Enzymology,
Faculty of Chemistry, Moscow State University) for
useful discussions. Sven Hagg (Department of Analy-
tical Chemistry, Lund University) is acknowledged forproviding us with a home made software for data
acquisition.
References
Anderson, G.P., Jacoby, M.A., Ligler, F.S., King, K.D., 1997.
Effectiveness of protein A for antibody immobilization for a fiber
optic biosensor. Biosens. Bioelectron. 12 (4), 329�/336.
Askari, M., Alarie, J.P., Moreno-Bondi, M., Vo-Dinh, T., 2001.
Application of an antibody biochip for p53 detection and cancer
diagnosis. Biotechnol. Prog. 17 (3), 543�/552.
Auroux, P.-A., Iossifidis, D., Reyes, D.R., Manz, A., 2002. Micro total
analysis systems. 2. Analytical standard operations and applica-
tions. Anal. Chem. 74 (12), 2637�/2652.
Babacan, S., Piravnik, P., Letcher, S., Rand, A.G., 2000. Evaluation of
antibody immobilization methods for piezoelectric biosensor
application. Biosens. Bioelectron. 15 (11�/12), 615�/621.
Barbosa, G.H., Santana, E.M., Almeida, A.M., Araujo, A.M.,
Faitibello-Filho, O., Carvallo, L.B., Jr, 2000. The use of filter
paper plasticised with polyvinyl alcohol-glutaraldehyde in ELISA.
Braz. J. Med. Biol. Res. 33 (7), 23�/827.
Bernard, A., Michel, B., Delamarche, E., 2001. Micromosaic immu-
noassays. Anal. Chem. 73 (1), 8�/12.
Bjarnasson, B., Chimuka, L., Onnerfjord, P., Eremin, S., Jonsson, J.-
A., Johansson, G., Emneus, J., 2000. Enzyme flow immunoassay
using a protein G column for the screening of triazine herbicides in
surface and waste water. Anal. Chim. Acta 426 (2), 197�/207.
Brecht, A., Klotz, A., Barzen, C., Gauglitz, G., Harris, R.D., Quigley,
G.R., Wilkinson, J.S., Sztajnbok, P., Abuknesha, R., Gascon, J.,
Oubina, A., Barcelo, D., 1998. Optical immunoprobe development
for multiresidue monitoring in water. Anal. Chim. Acta 362 (1),
69�/79.
Burestedt, E., Nistor, C., Schagerlof, U., Emneus, J., 2000. An enzyme
flow immunoassay that uses beta-galactosidase as the label and a
cellobiose dehydrogenase biosensor as the label detector. Anal.
Chem. 72 (17), 4171�/4177.
Chiem, N.H., Harrison, D.J., 1998. Microchip systems for immunoas-
say: an integrated immunoreactor with electrophoretic separation
for serum theophilline determination. Clin. Chem. 44 (3), 591�/598.
Dankwart, A., Muller, J., Hock, B., 1998. Stabilisation of enzyme
immunoassay for atrazine. Anal. Chim. Acta 362 (1), 35�/45.
Dolnik, V., Liu, S., Jovanovich, S., 2000. Capillary electrophoresis on
microchip. Electrophoresis 21 (1), 41�/54.
Drott, J., Laurell, T., Rosengren, L., Lindstrom, K., 1997. Porous
silicon as the carrier matrix in microstructured enzyme reactors
yielding high enzyme activities. J. Micromach. Microeng. 7 (1), 14�/
23.
Eteshola, E., Leckband, D., 2001. Development and characterisation
of an ELISA assay in PMDS microfluidic channels. Sens. Actuat. B
72 (129), 133.
Garcinuno, R.M., Fernandez, P., Perez-Conde, C., Guttierez, A.M.,
Camara, C., 2000. Development of a fluoroimmunosensor for
theophylline using immobilized antibody. Talanta 52 (5), 825�/832.
Giersch, T., 1993. A new monoclonal-antibody for the sensitive
detection of atrazine with immunoassay in microtiter plate and
dipstick format. J. Agric. Food Chem. 41 (6), 1006�/1011.
Gonzalez-Martinez, M.A., Morais, S., Puchades, R., Maquieira, A.,
Marco, M.-P., Barcelo, D., 1998. Automation of a heterogeneous
enzyme immunoassay for atrazine. Comparison of three immobi-
lization supports. Fresenius J. Anal. Chem. 361 (2), 179�/184.
Gonzalez-Martinez, M.A., Puchades, R., Maquieira, A., 1999. On-line
immunoanalysis for environmental pollutants: from batch assays
to automated sensors. Trends Anal. Chem. 18 (3), 204�/218.
Goodrow, M.H., Harrison, R.O., Hammock, B.D., 1990. Hapten
synthesis, antibody development, and competitive inhibition en-
zyme-immunoassay for s-triazine herbicides. J. Agric. Food Chem.
38 (4), 990�/996.
Gregorius, K., Mouritsen, S., Elsner, H.I., 1995. Hydrocoating: a new
method for coupling biomolecules to solid phases. J. Immunol.
Methods 181 (1), 65�/73.
Hashimoto, M., Tsukagoshi, K., Nakajima, R., Kondo, K., Arai, A.,
2000. Microchip capillary electrophoresis using on-line chemilumi-
nescence detection. J. Chromatogr. A 867, 271�/279.
Hatzidakis, G.I., Tsatsakis, A.M., Krambovitis, E.K., Spyros, A.,
Eremin, S.A., 2002. Use of L-lysine fluorescence derivatives as
tracers to enhance the performance of polarization fluoroimmu-
noassays. A study using two herbicides as model antigens. Anal.
Chem. 74 (11), 2513�/2521.
Hayes, M.A., Polson, N.A., Phayre, A.N., Garcia, A.A., 2001. Flow-
based microimmunoassay. Anal. Chem. 73 (24), 5896�/5902.
Hermanson, G.T., Mallia, A.K.P., Smith, P., 1992. Peptide antigens,
antibodies, and immunoglobulin binding proteins. In: Immobilized
Affinity Ligand Techniques. Academic Press, San Diego, pp. 210�/
251.
Hock, B., 1997. Antibodies for immunosensors. Anal. Chim. Acta 347
(1�/2), 177�/186.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/34 33
Howell, S., Kenmore, M., Kirkland, M., Badley, R.A., 1998. High
density immobilization of an antibody fragment to a carboxy-
methylated dextran-linked biosensor surface. J. Mol. Recognit. 11
(1�/6), 200�/203.
Khoo, K.M., Ting, Y.P., 2001. Polyvivyl alcohol as an immobilisation
matrix*/a case of gold biosorption. Water Sci. Technol. 43 (11),
17�/23.
Koutny, L.B., Schmalzing, D., Taylor, T.A., Fuchs, M., 1996.
Microchip electrophoretic immunoassay for serum cortisol. Anal.
Chem. 68 (1), 18�/22.
Kricka, L.J., 2001. Microchips, microarrays, biochips and nanochips:
personal laboratories for the 21st century. Clin. Chim. Acta 307
(1�/2), 219�/223.
Kronkvist, K., Lovgren, U., Svensson, J., Edholm, L.E., 1997.
Competitive flow injection enzyme immunoassay for steroids using
a post-column reaction technique. J. Immunol. Methods 200 (1�/2),
145�/153.
Krotzky, A.J., Zeeh, B., 1995. Immunoassays for residue analysis of
agrochemicals: proposed guidelines for precision, standardization
and quality control. Pure Appl. Chem. 67 (12), 2065�/2088.
Lacher, N.A., Garrison, K.E., Martin, R.S., Lunte, S.M., 2001.
Microchip capillary electrophoresis/electrochemistry. Electrophor-
esis 22 (12), 2526�/2536.
Laurell, T., Drott, J., Rosengren, L., Lindstrom, K., 1996. Enhanced
enzyme activity in silicon integrated enzyme reactors utilizing
porous silicon as the coupling matrix. Sens. Actuat. B 31 (3),
161�/166.
Liu, Y.C., Wang, C.M., Hsiung, K.P., 2001. Comparison of different
protein immobilization methods on quartz crystal microbalance
surface in flow injection immunoassay. Anal. Biochem. 299 (2),
130�/135.
Lofas, S., Johnsson, B., Tegendal, K., Ronnberg, I., 1993. Dextran
modified gold surfaces for surface plasmon resonance sensors:
immunoreactivity of immobilized antibodies and antibody-surface
interaction studies. Coll. Surf. B: Biointerf. 1 (2), 83�/89.
Lu, H.-C., Chen, H.-M., Lin, Y.-S., Lin, J.W., 2000. A reusable and
specific protein A-coated piezoelectric biosensor for flow injection
immunoassay. Biotechnol. Prog. 16 (1), 116�/124.
McCreedy, T., 2000. Fabrication techniques and materials commonly
used for the production of microreactors and micro total analytical
systems. Trends Anal. Chem. 19 (6), 396�/401.
Nistor, C., Rose, A., Farre, M., Stoica, L., Wollenberg, U., Ruzgas,
T., Barcelo, D., Gorton, L., Emneus, J., 2002. In-field monitoring
of cleaning efficiency in waste water treatment plants using two
phenol-sensitive biosensors. Anal. Chim. Acta 456 (1), 3�/17.
Penalva, J., Gonzalez-Martinez, M.A., Puchades, R., Maquieira, A.,
Marco, M.-P., Barcelo, D., 1999. Immunosensor for trace deter-
mination of Irgarol 1051 in sea water using organic media. Anal.
Chim. Acta 387 (3), 227�/233.
Penzol, G., Armisen, P., Fernandez-Lafuente, R., Rodes, L., Guisan,
J.M., 1998. Use of dextrans as long and hydrophilic spacer arms to
improve the performance of immobilized proteins acting on
macromolecules. Biotechnol. Bioeng. 60 (4), 518�/523.
Piehler, J., Brecht, A., Geckeler, K.E., Gauglitz, G., 1996. Surface
modification for direct immunoprobes. Biosens. Bioelectron. 11
(6�/7), 579�/590.
Pulido-Tofino, P., Barrero-Moreno, J.M., Perez-Conde, C., 2000.
Flow-through fluoroimmunosensor for isoproturon determination
in agricultural foodstuff. Evaluation of antibody immobilization
on sold support. Anal. Chim. Acta 417 (1), 85�/94.
Quinn, J., Patel, P., Fotzpatrick, B., Manning, B., Dillon, P., Daly, S.,
O’Kennedy, R., Alcoter, M., Lee, H., Morgan, M., Lang, K., 1999.
The use of regenerable, affinity ligand-based surfaces for immuno-
sensor applications. Biosens. Bioelectron. 14 (6), 587�/595.
Reyes, D.R., Iossifidis, D., Auroux, P.-A., Manz, A., 2002. Micro total
analysis systems. 1. Introduction, theory and technology. Anal.
Chem. 74 (12), 2636�/2636.
Rossi, A., Morana, A., Lernia, I.D., Di Tombrino, A., De Rosa, M.,
1999. Immobilization of enzymes on spongy polyvinyl alcohol
cryogels: the example of beta-galactosidase from Aspergillus
oryzae . Ital. J. Biochem. 48 (2), 91�/97.
Sanders, G., Manz, A., 2000. Chip-based microsystems for genomic
and proteomic analysis. Trends Anal. Chem. 19 (6), 364�/378.
Sato, K., Tokeshi, M., Odake, T., Kimura, H., Ooi, T., Nakao, M.,
Kitamori, T., 2000. Integration of an immunosorbent assay system:
analysis of a secretory human immunoglobulin A on polystyrene
beads in a microchip. Anal. Chem. 72 (6), 1144�/1147.
Sato, K., Tokeshi, M., Kimura, H., Kitamori, T., 2001. Determination
of carcinoembyonic antigen in human sera by integrated bead-bed
immunoassay in a microchip for cancer diagnosis. Anal. Chem. 73
(6), 1213�/1218.
Sheikh, S.H., Mulchandani, A., 2001. Continuos flow fluoro-immu-
nosensor for paclitaxel measurement. Biosens. Bioelectron. 16 (9�/
12), 647�/652.
Silzel, J.W., Cercek, B., Dodson, C., Tsay, T., Obremski, R.J., 1998.
Mass-sensing, multianalyte microarray immunoassay with imaging
detection. Clin. Chem. 44 (9), 2036�/2043.
Tasker, S., Mathijs, G., Davies, M.C., Roberts, C.J., Schacht, E.H.,
Tendler, S.J.B., 1996. Molecular resolution imaging of dextran
monolayers immobilized on silica by atomic force microscopy.
Langmuir 12 (26), 6436�/6442.
Thust, M., Schoning, M.J., Schroth, P., Malkoc, U., Dicker, C.I.,
Steffen, A., Kordos, P., Luth, H., 1999. Enzyme immobilisation on
planar and porous silicon substrates for biosensor applications. J.
Mol. Catal. B 7 (1�/4), 77�/83.
Valat, C., Limoges, B., Huet, D., Romette, J.-L., 2000. A disposable
protein A-based immunosensor for flow-injection assay with
electrochemical detection. Anal. Chim. Acta 404 (2), 187�/194.
Wang, J., Ibanetz, A., Chatrach, M.P., Escarpa, A., 2001. Electro-
chemical enzyme immunoassays on microchip platforms. Anal.
Chem. 73 (21), 5323�/5327.
Weetall, H., 1976. Covalent coupling methods for inorganic support
materials. Methods Enzymol. 44, 134�/138.
Wilson, D.S., Nock, S., 2002. Review: functional protein microarrays.
Curr. Opin. Chem. Biol. 6 (1), 81�/85.
Yakovleva, J., Davidsson, R., Lobanova, A., Bengtsson, M., Eremin,
S., Laurell, T., Emneus, J., 2002. Microfluidic enzyme ummunoas-
say using silicon microchip with immobilized antibodies
and chemiluminescence detection. Anal. Chem. 74 (13), 2994�/
3004.
J. Yakovleva et al. / Biosensors and Bioelectronics 19 (2003) 21�/3434