Upload
others
View
13
Download
1
Embed Size (px)
Citation preview
Instruction Manual
This Manual should be made available to personsresponsible for Installation, Operation and Maintenance.
MILTON ROY INDIA (P) LTD.Metering Pumps
LPH
Kg/cm2
M-Pump Model No.........................................................................
Serial No....................................................................................
Capacity.....................................................................................
Pressure....................................................................................
M -Series
4, Rajarajan Street, Visalakshi Nagar,Ekkaduthangal, Chennai - 600 097, India.Telephone : (044)-22254000 - 3Fax : 91-44-22254004 E-mail : [email protected] : www.miltonroyindia.com
Issue Date : 03/2005
WARRANTY Page No.
I. UNPACKING 1
II. LOCATION AND MOUNTING 1
III. ELECTRICAL AND DRIVE 1
IV. LUBRICATION 1
V. PLUMBING CONFIGURATION 1
VI. LIQUID HANDLING ASSEMBLY 1
VII. PUMP OPERATION 3
VIII. MAINTENANCE 4
TROUBLE SHOOTING GUIDE 5
SERIES M DRIVE ASSEMBLY PARTS LIST 7
LIQUID HANDLING ASSEMBLY PARTS LIST 9
TABLE OF CONTENTS
MILTON ROY INDIA (P) LTD.
M - Series Pumps
Milton Roy India (P) Ltd., warrants its metering pump products against defects in workmanship or materials for one
year under normal use from the date of shipment from our warehouse or the warehouse of our agent. All metering
pump components are warranted for one year, except that warranties on equipment and accessories furnished with
the pump but manufactured by others are limited to the warranties offered by the manufacturers of their respective
products. This warranty is not extended to electronic control devices supplied with a Milton Roy India (P) Ltd.,
metering pump. These items are covered by the warranties offered by the manufacturer for Electronic Controls and
Actuators.
All obligations and liabilities under this warranty are limited to refunding, repairing or replacing (at our option), f.o.b,
our plant, such allegedly defective units as are returned to our plant, carrier charges prepaid. Repairs or
replacements are made subject to factory inspection of returned items.
This warranty does not extend to damage by corrosion or erosion. The materials of construction offered are
recommendations subject in all cases to verification and acceptance by the customer. These recommendations,
based on previous Company experience and best available information, do not constitute guarantees against wear
or chemical action.
Expressly excluded from this warranty are defects caused by misuse, abuse, or improper application, employment,
or operation of the unit. Expendable items and damage resulting from unauthorised repair are not covered by this
warranty. No liability for consequential damages or reinstallation labour is accepted. Milton Roy India (P) Ltd., will
not assume responsibility for contingent liability for alleged failure of its products.
RETURNING PUMPS TO THE FACTORY
Pumps will not be accepted for repair without a written consent from the Factory Service Department. Pumps
returned to the Factory for repairs should be clearly labeled to indicate the liquid being pumped. Process liquid
should be flushed from the pump liquid end and oil should be drained from the pump housing before the pump is
shipped.
NOTE
These safety precautions will aid the troubleshooting and repair procedure and preclude serious injury to
repair personnel from hazardous residue in pump liquid end.
All inquiries or parts orders should be addressed to your local Milton Roy India (P) Ltd., representatives or sent to :
MILTON ROY INDIA (P) LTD.
M - Series Pumps
METERING PUMP PRODUCTS TWELVE MONTHLIMITED WARRANTY
MILTON ROY INDIA (P) Ltd.,NO.: 4, RAJARAJAN STREET, VISALAKSHI NAGAR
EKKADUTHANGAL, MADRAS - 600 097.
PHONE NO.: (044) - 22254000 - 3
FAX NO.: 91-44-22254004
I. UNPACKING
Remove tubing, injection check valve, pressure relief
valve and foot valve which are supplied loose. Notify
Factory and Insurance Company immediately if there are
any sings of damage to the metering pump or parts.
II. LOCATION AND MOUNTING
1. Locate the pump in an area that is convenient to
both chemical injection point and electrical
supply. MRIPL chemical metering pumps should
not be subjected to continuous high temperature 0
over 50 C
2. Mount the pump at a convenient operating height
on a firm foundation.
III. ELECTRICAL AND DRIVE
The pump is driven by a 0.5 H.P., 1440 RPM squirrel
cage induction motor. The input power supply is 415V
3 phase 50 C/S. The drive consists of a speed
reduction by means of a worm and worm wheel. The
worm is connected to the motor by a love-joy
coupling.
IV. LUBRICATION
Before operating the pump check oil drain plug for
tightness and fill unit with 250 ml. of SAE 140 Grade
Oil (Servo Gear HP-140 of Indian Oil or equivalent).
Check oil level using dip-stick provided with the filler
plug.
V. PLUMBING CONFIGURTION
1. Figures 1 & 2 show typical chemical metering
pump installations. Note location of injection
check valve which is most important. Refer to
liquid handling assembly instructions section
regarding installation of injection check valve.
2. Back Pressure Requirement
All metering pumps are prone to overpumping
(excessive output) at low discharge pressure
and while pumping into open tanks. To prevent this
condition from occurring it is necessary to maintain
approximately 25 PSI (10 PSI for M-13 & M-14) back
pressure across the pump. This is achieved by a
spring loaded injection check/antisyphon valve.
Consult ALMI factory for injection check/antisyphon
valves with increased back pressure ratings.
3. If chemical level in flooded suction installations
are more than 3 metres consult factory.
4. Fluid Connections - M series metering pumps are
equipped with LDPE flexible tubing for both suction
and discharge. In case you wish to use rigid piping
contact factory for fittings with NPT threads. Following
precautions should be taken.
* Pressure test all plumbing before use to guard
against leaks.
* Prevent thread sealant from getting inside
plumbing.
* Install a ball valve in the suction side of flooded
suction installation to simplify maintenance.
* A foot valve is a must in suction lift installations.
* Do not use teflon tape on flexible tubing
connectors.
* Do not use pipe wrenches on flexible tubing
connectors.
VI. LIQUID HANDLING ASSEMBLY
A. Installing Injection Check Valve
1. The injection check/antisyphon valve should
always be installed as close as possible to the
point of chemical injection at the very end of the
tubing run.
2. Purpose of injection check/antisyphon valve is to
prevent backflow from treated line. It also
provides the back pressure required for the
pump.
3. A 1/2" NPT female fitting with sufficient depth will
accept the injection check/antisyphon valve.
4. In order to ensure correct seating of the ball inside
the check valve, the injection check/antisyphon
valve should be installed vertically upwards.
M-Series Pumps
1
CAUTION
When pumping chemicals make certain that all
tubing is securely attached to the fittings. It is
recommended that tubing or pipe lines be shielded
to prevent possible injury in case of rupture or
accidental damage. Always wear protective face
shield and clothing when working on or near a
chemical metering pump. All M Series pumps are
motor driven mechanical pumps. Your pump has
been equipped with a pressure relief valve supplied
by MRIPL & you have to install the same on the head
of pump before start-up.
CAUTION
Be sure installation does not constitute a cross
connection. Check local plumbing code.
MILTON ROY INDIA (P) LTD.
5. Maximum pump pressure rating is reduced with
spring installed in injection check/antisyphon valve.
Remove spring if pressure at injection point is more
than 20 PSI.
B. Installing Pressure Relief Valve
1. MRIPL PRV consists of a pressure relief valve with a
mounting adaptor.
2. Screw in adaptor on the discharge valve plug. Take
care to see that the Quad seal sits on the taper nozzle.
Do not use teflon tape. Screw in the PRV into the
adapter taking care to see that the quad seal in
position.
C. Connecting Discharge Tubing
1. Cut tubing to length needed for discharge line.
2. Route tubing from injection check valve to chemical
metering pump making sure it does not touch hot
surfaces, sharp surfaces, or is bent so sharply that it
kinks.
3. Slide small end of coupling nut onto tubing.
4. Push tubing over tapped nozzle of discharge valve
housing so that tubing flares out and reaches the
shoulder. If tubing is stiff from cold, dip end in hot
water.
5. Slide down the coupling nut until threads are engaged.
Tighten by hand until tubing is held securely in place.
6. Follow the same procedure for connecting tubing to
injection valve.
D. Connecting Suction Tubing
1. Cut Suction tubing to length such that the foot valve
hangs just above the bottom of the chemical container.
Maximum recommended vertical suction lift 5 ft.
(1.5m).
2. Follow same procedure (see C) in connecting suction
tubing to suction valve and foot valve.
3. If a suction tube straightener is desired, one may be
fabricated from a 3 ft. (1m) piece of 1/2" Schedule 80
PVC pipe.
4. In case of flooded suction use a tubing connector with
1/2 NPT threads. This is available from MRIPL as an
optional accessory.
Excessive force will crack or distort fittings.
DO NOT USE PIPE WRENCH
M-Series Pumps
SUCTION LIFT INSTALLATION FLOODED SUCTION INSTALLATION
Fig. 1 Fig. 2
helpful when pumping
at very low rate
2
MILTON ROY INDIA (P) LTD.
VII PUMP OPERATION
A. Priming
Priming can be done only against atmospheric
pressure when using a suction lift installation. Flooded
suction installations will allow the unit to prime itself
even under pressure.
1. Temporarily remove tubing from injection check
valve and hold the end of tubing so that it is above
pump level.
2. Set pump at 100% stroke and start pump.
3. As soon as chemical is visible through translucent
discharge tubing just past the discharge valve,
stop the pump.
4. Pump is now primed.
5. Reconnect tubing to injection check valve.
Note :
(a) Pump is normally self-priming if suction lift is no
more than 5 ft. (1.5m) and the above steps (A1
thru A3) are followed.
(b) If the pump does not self prime, remove discharge
valve housing and ball and pour water or
chemical slowly into discharge port until it is filled.
Follow steps A2 thru A5 thereafter.
In all cases priming will be assisted by releasing or
reducing pressure on the pump. This is best
accomplished by pulling the lever of the ALMI
Pressure Relief Valve.
B. Output Adjustment
Pump output is adjusted using the knob located at the
top of the pump (fig.3). The lock knob holds the shaft
to prevent charge of setting. Before adjusting the
pump should be running. The dial under the knob
represents "% of total output" and provides accurate
output adjustment above the 10% setting. The knob is
set properly if the line on the knob lines up the zero
M-Series Pumps
Fig. 3
3
MILTON ROY INDIA (P) LTD.
(0) mark on the dial when the knob is turned counter-
clockwise as far as it will go. This is the zero output
position. The maximum output position. The maximum
output position is reached by turning this knob fully
clockwise to 100% setting on dial. Output adjustment
should be done while the unit is running.
C. Output Adjustment knob Setting
With the pump operating, turn the output adjustment knob
counterclockwise until no stroking action is felt when
holding onto the knobs. At the point where stroking action
stops the line on the knob should be directly over the zero
mark on the dial. If alignment is not correct, knob position
on shaft should be adjusted as follows.
1. Loosen 2 Nos. set screws on the knob with suitable
Allen-wrench and remove the knob.
2. Loosen Lock knob.
3. With a griplier adjust shaft until a point is reached
where stroking action just ceases and can go longer
be felt when grasping shaft and stop the pump.
4. Place knob on shaft. Carefully line knob up so that
knob pointer is directly over the zero marking on the
dial.
5. Taking care not to move shaft or knob alignment,
tighten both set-screws.
6. Recheck alignment results.
D. Calibration
Pump output calibration can be easily done if required.
The pump is best calibrated after installation by using a
stopwatch to measure the time required to remove a
known volume of liquid from a tank (fig.4). Proper setting
of the pump's zero output position should be verified
before calibration and is explained is Section B.
VIII MAINTENANCE
A. Lubricant Change
The m series drive train is oil lubricated. Replace oil
every 1000hrs. of running with approximately 250 ml.
of SAE 140 Grade Oil (Servo Gear HP-140 of Indian
Oil or equivalent). Check oil level periodically using
dip-stick and top-up as required.
B. Liquifram Replacement
1. Disconnect suction supply line and shut off to
prevent chemical spill.
2. Operate pump until pump head is empited of
chemical. set output adjustment knob at 100% by
turning knob fully clockwise.
3. Stop the pump.
4. Release pressure from discharge line by pulling
the knob on the PRV.
5. Disconnect discharge line. Be prepared to catch
excess chemical during this procedure.
6. Remove pump head by unscrewing the four/six head
retaining screws holding the head to the spacer.
7. Start the pump. With unit running at 100% output,
unscrew and remove and old Liquidfram.
8. With pump still running, set output adjustment knob
back to 90% for models M-09, M-10, M-11 and M-12,
70% for model M-13 and 50% for Model M-14.
9. Without changing these output settings, screw new
Liquifram in place.
10. Stop the pump. Rotate the shaft at Love-joy coupling
or Motor fan (after removing fanguard) and make sure
diaphragm is fully retracted. Then check Liquifram
alignment by using a straight edge held firmly against
the spacer (take care not to scratch the sealing flange
of Teflon-faced Liquiframs). Centre of Liquifram
should be flush with outer edges of spacer (fig. 5).
11. With new liquifram installed, turn on pump and set
knob at 100% output position. Reinstall pump head
making sure the part number stamped on the face of
the head is in upright position. Tighten the four/six
retaining screws in a criss cross pattern.
M-Series Pumps
4 MILTON ROY INDIA (P) LTD.
DATA SHEET FOR METERING PUMP
MODEL M SERIES
TYPE MECHANICALLY ACTUATED DIAPHRAGM TYPE
MODEL M-09 M-10 M-11 M-12 M-13 M-14
MAX FLOW (IN LPH) 5 10 17 30 50 100
MAX PRESSURE (KG/SQCM) 15 15 12 10 5 3
STROKE PER MINUTE 50 100 144 144 144 144
LINEAR SPEED (M/MIN) 0.4 0.8 1.152 1.152 1.152 1.152
Always wear protective clothing and face shield
when working on or near a chemical metering pump.
TROUBLE SHOOTING GUIDE
Problem Cause Remedy
Solution drips from hole in Liquifram is ruptured Replace Liquifram. See Liquifram Replacement
spacer instructions in maintenance section of this manual
Liquifram does not move Output adjustment knob is set at zero With pump running, rotate output adjustment knob to a higher setting.
Power is off Check power supply.
Solution is too viscous Dilute solution.
Injection into excessive pressure. Check and clean valves if clogged. Check system pressure against maximum injection pressure rating pump. If discharge pressure is within 20 PSI of pump's m a x i m u m rating, remove spring in injection check /antisyphon valve.
Liquifram moves, but no solution Trapped air in pump head Operate pump of 100% output setting. Release is pumped pressure on pump d ischarge. Fol low pr iming
instructions. Operate pump for a few minutes to purge head and valves of air.
Suction lift is too high Reduce suction lift or arrange pump and tank for flooded suction installation.
Loop in suction line permits air trap on Straighten suction tubing.
suction side
Improperly installed, or dirty valve seats Refer liquid Handling Assembly instructions. Clean and reassemble valves.
Air in suction line. Tighten suction valve fitting. Replace if necessary. Check for air leaks throughout suction line.
Pumps air bubbles or air. Suction tubing has leaks or pin holes. Replace suction tubing.
Improperly installed, or dirty valve seats. Refer liquid Handling Assembly instruction. Clean and reassemble valves.
Suction lift is too high. Reduce suction lift or arrange pump and tank for flooded suction installation.
Solution supply is exhausted. Refill solution container.
Foot valve and strainer assembly is clogged Clean foot valve assembly.
Pumps does not start under low Voltage Check your power supply
rated pressure Pressure rise due to discharge line Clean discharge line or dilute solution.
or viscosity of solution.
Clogged injection check/antisyphon Refer Liquid Handling Assembly instruction. Clean and
valve.. reassemble valves.
Suction tubing collapses Foot valve and strainer assembly is Refer Liquid Handling Assembly instruction. Clean and
clogged. reassemble valves. Replace strainer if necessary.
Excessive pump output output adjustment knob is improperly With pump running turn knob counter clockwise until set. pump stops stroking. Line on knob should be directly
over the zero mark on the dial. If setting is off, adjust by following Output Adjustment knob setting instructions in this manual.
Very low pressure at injection point Install injection check/antisyphon valve on pump.
Excessive positive suction head. When flooded suction installation is used, chemical supply should be no more than 20 feet (6 metres) above pump when using standard injection check/ antisypon valve and zero discharge pressure. Alter location of pump or consult factory.
5
KEY NO PART NO DESCRIPTION QTY
1 13531 10-24 X 3/4" PAN HEAD SCREW 12 (M14) & 8 ( ot hers)
2 130420 SPACER ADAPTOR (M-14) ONLY 1
3 131120S SPACER (M-09, M-10, M-11) 1SET
3 131220S SPACER (M-12) 1SET
3 131320S SPACER (M-13, M-14) 1SET
4 130552 GASKET SPACER 1
5 133525 SPRING 1
5A 11531 ADDITIONAL SPRING (M-12) ONLY 1
6 11711 RETAINER PLATE 1
7 130631 LIQUIFRAM SHAFT 1
8 133626 DOWEL 5 X 15 2
9 130713N GEAR PLATE 1
10 131826 BALL BEARING (6200) 1
11 131511 WIDE CAM SPACER 1
12 131444 WORM WHEEL (M-11,M-12,M-13,M-14) 1
12 131344 WORM WHEEL (M-10) 1
12 131244 WORM WHEEL (M-09) 1
13 134026 CAM BEARING (6204) 1
15 131926 BALL BEARING (6201) 1
16 132011N BEARING CAP 1
17 130931S CAM SHAFT ASSY FOR M 1SET
18 131526 BALL BEARING (6000) 1
19 131722 WORM SHAFT (M-11,M-12,M-13,M-14) 1
19 131522 WORM SHAFT (M-10) 1
19 131622 WORM SHAFT (M-09) 1
20 131726 BALL BEARING (6001) 1
21 132331 ADJUSTMENT SHAFT 1
DRIVE ASSEMBLY - M PUMP
22 132311N ADJUSTMENT STOP 1
22A 134052 GROMMET FOR ADJUSTMENT STOP 'M' 1
23 134511 DIAL (M-09,M-10, M-11) 1
23 134911 DIAL (M-12) 1
23 134811 DIAL (M-13, M-14) 1
24 132431S ADJUSTMENT KNOB WITH GRUB SCREW 1 SET
25 132613 HOUSING 1
26 152131S VENT CAP WITH GASKET 1SET
27 132531 LOCK KNOB 'M' 1
27A 132641 LOCK KNOB INSERT 'M' 1
28 132841S DRAIN PLUG WITH O-RING 1SET
29 133721 MOUNTING BASE (STD MOTOR) - 0.5 HP 1
29 133821 MOUNTING BASE (FLP MOTOR) - 0.5 HP 1
30 135531 STUD SS M8 X 35 (FT) 4
31 126231 NUT M8 4
32 11252 OIL SEAL LIQUIFRAM SHAFT 1
33 133120 COUPLING (STD MOTOR) - 0.5 HP 1
33 133220S COUPLING (FLP MOTOR) - 0.5 HP WITH GRUB SC REW 1 SET
34 132952N OIL SEAL 1
34A 133011N OIL SEAL HOLDER 1
35 130811N BEARING RING 1
36 134931 CSK SCREW 3/16" X 1/2" 8
37 134831S ALLEN SCREW M5 X 30 WITH SPRING WASHER 4 SE T
39 M01677S HEX BOLT M6 X 35 WITH NUT AND WASHER 4 SET
39 M01267SHEX BOLT M8 X 45 WITH NUT AND WASHER ( 0.5 HP FLP MOTOR)
4 SET
42 183511 OIL LEVEL INDICATOR 1
43 130713S GEAR PLATE ASSEMBLY M-09 1SET
43 130713D GEAR PLATE ASSEMBLY - M-10 1SET
43 130713T GEAR PLATE ASSEMBLY -M11, M12, M13, M14 1SE T
KEY NO DESCRIPTIONM-09,M-10,M-11.
M-12 M-13 M-14 QTY
2 SPRING 37341 37341 37341 37341 1
3 BALL CERAMIC 10338 10338 10338 10138 3
4 RING SEAL TEFLON 902079 902079 902079 902179N 3
5 SUCTION VALVE SEAT ASSY 902667 902667 902667 902967 2SET
6 QUAD SEAL 905755 905755 905755 905755 2
7 TUBING 10142 I 10142 I 10142 I 10142 I 13FT
8 DISCHARGE VALVE HOUSING ASSY 902767 902767 902767 903067 1SET
9 3/16 X 3/4" PAN HEAD SCREW 13531 13531 13531 NA 4
10 LIQUIFARM (DIAPHRAGM) 30917 31420 31419 25719 1
11 HEAD 900568 900468 900368N 900268 1
12 PRV ADAPTOR 904867N 904867N 904867N 904867N 1
14 Y-STRAINER ASSY 913267S 913267S 913267S 913267S 1SET
15 ANTISYPHON VALVE ASSY 906267 906267 906267 906467 1SET
16 FOOT VALVE ASSY 907067 907067 907067 907167 1SET
17 PRESSURE RELIEF VALVE 904767N 904767N 904767N 904767N 1
18 M6 X 60 HEX BOLT,NUT WITH WASHER NA NA NA M01639S 6 SETS
KEY NO DESCRIPTIONM-09,M-10,M-11
M-12 M-13 M-14 QTY
2 SPRING 37341 37341 37341 37341 1
3 BALL CERAMIC 10338 10338 10338 10138 3
4 RING SEAL TEFLON 902079 902079 902079 902179N 3
5 SUCTION VALVE SEAT ASSY 902667 902667 902667 902967 2SET
6 QUAD SEAL 905755 905755 905755 905755 2
7 TUBING 10142 I 10142 I 10142 I 10142 I 13FT
8 DISCHARGE VALVE HOUSING ASSY 902767 902767 902767 903067 1SET
9 3/16 X 3/4" PAN HEAD SCREW 13531 13531 13531 NA 4
10 LIQUIFARM (DIAPHRAGM) 30917 31420 31419 25719 1
11 HEAD 29611M 27214M 31053M 32234M 1
12 PRV ADAPTOR 904867N 904867N 904867N 904867N 1
14 Y-STRAINER ASSY 913267S 913267S 913267S 913267S 1SET
14A ONLINE STRAINER ASSY 909867S 909867S 909867S 909867S 1SET
15 ANTISYPHON VALVE ASSY 906267 906267 906267 906467 1SET
16 FOOT VALVE ASSY 907067 907067 907067 907167 1SET
17 PRESSURE RELIEF VALVE 904767N 904767N 904767N 904767N 1
18 M6 X 60 HEX BOLT,NUT WITH WASHER NA NA NA M01639S 6 SETS
LIQUID END ASSY FOR PUMPS WITH PP HEAD
LIQUID END ASSY FOR PUMPS WITH PVDF HEAD WITH PVC FITTING
KEY NO DESCRIPTIONM-09,M-10,M-11.
M-12 M-13 M-14 QTY
2 SPRING 37341 37341 37341 37341 1
3 BALL CERAMIC 10338 10338 10338 10138 3
4 RING SEAL TEFLON 902079 902079 902079 902179N 3
5 SUCTION VALVE SEAT ASSY 902632S 902632S 902632S 902932S 2SET
6 QUAD SEAL 905755 905755 905755 905755 2
7 TUBING 10142 I 10142 I 10142 I 10142 I 13FT
8 DISCHARGE VALVE HOUSING ASSY 902732S 902732S 902732S 903032S 1SET
9 3/16 X 3/4" PAN HEAD SCREW 13531 13531 13531 NA 4
10 LIQUIFARM (DIAPHRAGM) 30917 31420 31419 25719 1
11 HEAD 900532 900432 900332N 900232N 1
12 PRV ADAPTOR 904832N 904832N 904832N 904832N 1
14 Y-STRAINER ASSY 913232 913232 913232 913232 1SET
15 ANTISYPHON VALVE ASSY 906232 906232 906232 906432 1SET
16 FOOT VALVE ASSY 907032 907032 907032 907132 1SET
17 PRESSURE RELIEF VALVE 904732N 904732N 904732N 904732N 1
18 M6 X 40 HEX BOLT,NUT WITH WASHER NA NA NA 133431S 6 SETS
KEY NO DESCRIPTIONM-09,M-10,M-11.
M-12 M-13 M-14 QTY
2 SPRING 37341 37341 37341 37341 1
3 BALL CERAMIC 10338 10338 10338 10138 3
4 RING SEAL TEFLON 902079 902079 902079 902179N 3
5 SUCTION VALVE SEAT ASSY 902667 902667 902667 902967 2SET
6 QUAD SEAL 905755 905755 905755 905755 2
7 TUBING 10142 I 10142 I 10142 I 10142 I 13FT
8 DISCHARGE VALVE HOUSING ASSY 902767 902767 902767 903067 1SET
9 3/16 X 3/4" PAN HEAD SCREW 13531 13531 13531 NA 4
10 LIQUIFARM (DIAPHRAGM) 30917 31420 31419 25719 1
11 HEAD 900567 900467 900367N 900267 1
12 PRV ADAPTOR 904867N 904867N 904867N 904867N 1
14 Y-STRAINER ASSY 913267S 913267S 913267S 913267S 1SET
14A ONLINE STRAINER ASSY 909867S 909867S 909867S 909867S 1SET
15 ANTISYPHON VALVE ASSY 906267 906267 906267 906467 1SET
16 FOOT VALVE ASSY 907067 907067 907067 907167 1SET
17 PRESSURE RELIEF VALVE 904767N 904767N 904767N 904767N 1
18 M6 X 60 HEX BOLT,NUT WITH WASHER NA NA NA M01335S 6 SETS
LIQUID END ASSY FOR PUMPS WITH PVC HEAD
LIQUID END ASSY FOR PUMPS WITH SS HEAD