    US-EPA Equivalent Method

    Met One BAM-1020 is a US-EPA designated beta gauge Instrument used

    world wide.

    EPA Designation vide EQPM-0798-122

    Automatically measures Zero and Reference Mass value in each

    cycle.

    Does not require the use of correction factors. Best accuracy of all beta models.

  • 8/2/2019 Met One BAM-1020

    3/10

    C ECRD Publications

    BAM-1020 Advantages

    Continuously samples and measures ambient air for particulate

    matter.

    45 day continuous operation between filter tape changes.

    Accessories available for measuring PM2.5 and TSP.

    Heated Probe system to avoid moisture condensation in areas

    having high Humidity.

  • 8/2/2019 Met One BAM-1020

    4/10

    C ECRD Publications

    3/31/2012 4

    Theory of Operation Beta rays are measured through clean filter

    tape.

    Air containing particulate is sampled and

    deposited on the filter tape.

    Beta rays are measured through dirty filter

    tape.

    The difference between the two

    measurements is proportional to the mass

    concentration.

  • 8/2/2019 Met One BAM-1020

    5/10

    C ECRD Publications

    3/31/2012 5

    Beta Gauge Advantages

    Time proven, stable technology: beta gauges have been in use

    for over 60 years.

    Beta attenuation technique is simple, easy to use, and requires

    less maintenance than other methods.

    Fewer site trips required, ideal for remote deployment.

    Met One beta gauges are manufactured and supported from the

    United States.

  • 8/2/2019 Met One BAM-1020

    6/10

    C ECRD Publications

    3/31/2012 6

    Beta Gauge Advantages

    Seasonal and geographical correction factors are not needed.

    It is not necessary to calibrate the beta measurement system in

    the field.

    BAMs provide excellent agreement with gravimetric reference

    standards when properly maintained and operated.

    Beta technology is virtually unaffected by the chemical

    composition of the sampled particulate.

  • 8/2/2019 Met One BAM-1020

    7/10C ECRD Publications

    3/31/2012 7

    BAM-1020 Advantages

    Continuously samples and measures ambient air for particulate

    matter.

    Designed to predict the mass concentration measured by manual

    gravimetric samplers with the highest accuracy.

    Does not require the use of correction factors

  • 8/2/2019 Met One BAM-1020

    8/10C ECRD Publications

    3/31/2012 8

    BAM-1020 Advantages

    60 day continuous operation between filter tape changes.

    Designed for long term operation at a fixed installation.

    Powerful integrated 6 channel data logger compatible with a

    wide array of meteorological sensors. Best accuracy of all beta models.

  • 8/2/2019 Met One BAM-1020

    9/10

    3/31/2012 9

    Simplicity Measurement System

    The BAM-1020 measurement systemconsists of a few simple components:

    Carbon 14 Beta source

    Filter tape support

    Photomultiplier tube detector

    Preamp circuitry

  • 8/2/2019 Met One BAM-1020

    10/10C ECRD Publications

    3/31/2012 10

    BAM Comparison to EPA Reference for