2018 MEDICAL SAFETY POWER SOLUTION GUIDE

Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

2018MEDICAL SAFETYPOWER SOLUTIONGUIDE

Page 2: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

MINMAX has an extensive history and experience with the design of medical safety DC–DC converters and AC–DC

power supplies for demanding applications in both medical and healthcare instrumentation that requires medical

safety and a reinforced insulation system. MINMAX offers a large standard range of high-isolation and reinforced

insulated medical safety DC–DC converters with a power ranging from 1 to 20 W and AC–DC power supplies with

a power ranging from 24 to 60 W.

Given the requirements for medical/healthcare applications for I/O isolation, MINMAX Medical Safety Power

Solutions are rated from 3000 to 5000 VAC and possess reinforced insulation and a low leakage current for operator

protection (2xMOOP) or patient protection (2xMOPP).

All medical safety products meet the latest medical safety standards (ANSI/AAMI ES60601-1 and IEC/EN 60601-1

3rd edition) and are approved for nominal working voltages of 250 Vrms or higher.

MINMAX medical safety DC–DC converters and AC–DC power supplies offer cost-effective power solutions for

demanding medical and healthcare applications in dental chairs, oral care equipment, infusion pumps, medical

assist devices, medical oxygen monitors, medical carts, CT scanning, ultrasound, and many pieces of medical

auxiliary equipment.

History of the Editions of Medical Safety Standard IEC60601

Changes from the 2nd Edition to the 3rd Edition of the IEC60601-1 Standard

Isolation Voltage and Safety Distance in the 3rd Edition of the IEC60601-1 Standard

Leakage Current in the 3rd Edition of the IEC60601-1 Standard

Insulation Type

Structure of the Medical Insulation System

Medical EMC Standards

Cost vs. Risks to Safety

Medical Safety Introduction and Application

MINMAX Medical Safety Power Solutions

09

Successful Application

Medical Safety Products Overview

Medical Safety 1-20W DC-DC Converters

Medical Safety 24-60W AC-DC Power Supplies

Highlighted Performance of Medical Safety Products

15

16

17-22

22-23

02-03

03-04

04-05

05

06

06-07

07-08

08

Medical Safety Power Solutions Engineered by MINMAX

Page 3: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Medical Safety Standard IEC60601 was first published in 1977; the internationally accepted and approved

IEC60601 standard is the fundamental document that addresses many risks and has been developed to help

alleviate safety issues associated with electrical medical and healthcare equipment.

The standard consists of four distinct parts: the base standard (60601-1), collateral standards (60601-1-x),

particular standards (60601-2-x), and performance standards (60601-3-x). The base standard, IEC60601-1, has

been adopted in most major countries as a national standard.

The 3rd edition of the approved IEC60601-1 medical safety standard was first published by the IEC in 2005

(IEC60601:2005). The IEC60601 standard was adopted by the European Union in 2006 and published as

EN60601-1:2006. The 3rd edition standard was also published in 2006 by the USA but is different from the 2nd

edition published by UL. The 3rd edition was published by the American Association for Medical Instrumentation

(AAMI) and published as ANSI/AAMI ES60601:2006. Canada published this medical safety standard in 2008 as

CAN/CSA60601:2008.

The 3rd edition of Medical Safety Standard IEC60601 went into effect at different times depending on

region.

1st EDITION published by IEC1977

2nd EDITION published by IEC1988

3rd EDITION published by IEC2005

EU enforces 3rd

edition on all new and existing products

June 2012

Canada enforces3rd edition for new devices

July 2014

USA enforces3rd edition

for new devices

Jan 2014

History of the Editions of Medical Safety Standard IEC60601

Page 4: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

In Europe, the 2nd edition of the IEC60601-1 standard was withdrawn; all products including both new products

introduced to the market and products already on sale need to be certified under the 3rd edition of the

EN60601-1 standard before the cessation date. The original cessation date for the 2nd edition in the United

States is July 1, 2013. The FDA announced an extension to give US medical device designers a slight reprieve,

setting the updated transition date to December 31, 2013, and the effective date for the released 3rd edition

was January 1, 2014. Unlike the EU, the FDA only requires that new products after this date need to be certified

to the ANSI/AAMI ES60601-1 standard; existing products do not.

In Canada, the original cessation date for the 2nd edition was delayed, as in the USA, with an updated transition

date of June 30, 2014. The effective date for the released 3rd edition was July 1, 2014. However, only new

products after this date need to satisfy the 3rd edition.

All MINMAX medical safety AC–DC power supplies and medical safety DC–DC converters have been certified

to the approved 3rd edition standard (with twice the means of protection for the majority of power supplies)

and also tested against the 2nd edition.

Changes from the 2nd Edition to the 3rd Edition of the IEC60601-1 Standard

History of the Editions of Medical Safety Standard IEC60601

There were three categories of increasing severity:

In the 2nd edition of the IEC60601-1 standard, the guidelines need to be applied when the electrical medical

and healthcare equipment was within the “patient vicinity,” defined as a 6-ft. radius around the patient.

Type "B" (body) :

An applied part that is not in electrical

contact with the patient.

Type "BF" (body floating) :

An applied part that is electrically connected

to the patient but not directly to the patient’s

heart.

Type "CF" (cardiac floating) :

An applied part that is electrically connected

to the heart of the patient.

Page 5: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Insulation test voltages based on 250 VAC working voltage. MOP = Means of protectionMOOP = Means of operation protection MOPP = Means of patient protection

As medical safety AC–DC power supplies and medical safety DC–DC converters have significantly played a

crucial role in certified electrical medical and healthcare equipment, the 3rd edition of the IEC60601-1 standard

introduces new concepts such as the essential performance of equipment and distinguishes protecting the

equipment’s operator and the patient by Means of Operator Protection (MOOP) and Means of Patient Protection

(MOPP) classification in terms of the separation safety distance, insulation schemes, and dielectric strength

requirements.

The 2nd edition of the IEC60601-1 standard only addresses the basic safety issues for protection from any

electrical, mechanical, radiation, and thermal hazards. However, it did not require devices to remain functional;

a fail-safe was adequate and did not take the essential performance of the device into account in a pass/fail result

test. Therefore, the 3rd edition of the IEC60601-1 standard introduces specifications for “essential performance”

that require medical/healthcare equipment to continue functioning throughout the test process.

Isolation Voltage and Safety Distance in the 3rd Edition of the IEC60601-1 Standard

The 3rd edition of the approved IEC60601-1 safety standard specified that the safety distance for the minimum

creepage, air clearance, and isolation voltage must be met in order to avoid risks and to ensure freedom from

dangerous energy shocks due to any electrical shocks and excess energy hazards, transient voltage spikes,

insulation breakdown of the power architecture, mechanical damage, ignition, fires, shorts developing between

PCB tracks and across air gaps, arcing, and ground loops, which comply with the limited leakage current during

normal and single-fault conditions.

The level requirement for the isolation voltage depends on the insulation type, the working voltage, and the degree

of pollution, and the insulation barriers must have undergone a high-voltage test.

From our standpoint at MINMAX, we believe that the power supplies for medical and healthcare equipment should

provide the highest degree of protection; therefore, our medical safety DC–DC converters and AC–DC power

supplies have 2xMOPP/2xMOOP from the primary side to the secondary side (mains AC to low-voltage DC).

This gives medical/healthcare equipment designers flexibility and assurance that there are primary and secondary

reinforced insulation barriers in the medical power architecture to ensure long-term safety.

Changes from the 2nd Edition to the 3rd Edition of the IEC60601-1 Standard

InsulationMOOP MOPP

AirClearance

CreepageDistance

TestVoltage

AirClearance

CreepageDistance

TestVoltage

2.0 mm

4.0 mm

3.2 mm

6.4 mm

1500 VAC

3000 VAC

2.5 mm

5.0 mm

4.0 mm

8.0 mm

1500 VAC

4000 VACDouble or Reinforced(2 x MOP)

Basic (1 x MOP)

The classification determined the mandated or allowed types of levels of isolation, insulation, creepage, air

clearance, and leakage current that operators and patients may come into contact with.

Page 6: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Leakage Current in the 3rd Edition of the IEC60601-1Standard

NC = Normal Conditions

SFC = Single Fault Conditions

* The maximum Earth and enclosure leakage current for patient care equipment in the US is 300 μA.(1) “Enclosure Leakage Current” changed to “Touch Leakage Current” in the 3rd edition of the IEC60601-1 standard

Earth leakage current Current flowing in the earth conductor.

Enclosure leakage current Current flowing to earth via the patient from the enclosure.

Patient leakage current Current flowing to earth via the patient from an applied part.

Patient auxiliary current Current flowing between two applied parts.

Whether the product is considered MOOP or MOPP, the leakage-current requirements must be met.

A further change between the 2nd and 3rd editions is related to the Earth leakage-current requirements.

Leakage-current tests are designed to simulate a human body coming into contact with different parts of the

medical/healthcare equipment. The measured leakage-current values should comply with the acceptable limits.

The leakage current needs to comply with a limit value to avoid risks and ensure freedom from dangerous energy

shocks due to any electrical shock and excess energy hazards, transient voltage spikes, insulation breakdown of

the power architecture, mechanical damage, ignition, fire, shorts developing between PCB tracks, air gaps,

arcing, and ground loops that the operator or patient may come into direct contact during normal and single-

fault conditions.

Isolation Voltage and Safety Distance in the 3rd Edition of the IEC60601-1 Standard

500 µA*

100 µA

100 µA

10 µA

1000 µA

500 µA*

500 µA

50 µA

500 µA*

100 µA

100 µA

10 µA

1000 µA

500 µA*

500 µA

50 µA

500 µA*

100 µA

10 µA

10 µA

1000 µA

500 µA*

50 µA

50 µA

Leakage CurrentTYPE B TYPE CFTYPE BF

NC SFC NC SFC NC SFC

Earth leakage current general

Enclosure leakage current (1)

Patient leakage current

Patient auxiliary current

AC

DC

Clearance Distance

Clearance Distance

Creepage Distance

is the distance between

two points through air

Creepage Distanceis the distance between two points over the surface

Conductor Conductor

Non- conductor

Page 7: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Structure of the Medical Insulation System

The figure below shows the insulation system structure and two insulation barriers that provide two Means of

Protection (MOP) that must be present within medical/healthcare systems.

The five different types of insulation grades are listed below.

Insulation Type

Insulation that is necessary only for the correct functioning of the equipment and doesnot provide any protection against electric shock.

Insulation applied to live parts to provide protection against electric shock.

Independent insulation applied in addition to basic insulation in order to provideprotection against electric shock in the event of a failure of basic insulation.

Insulation comprising both basic insulation and supplementary insulation.

Single insulation system applied to live parts which provide a degree of protectionagainst electric shock equivalent to double insulation.

Functional Insulation

Basic Insulation

Supplementary Insulation

Double Insulation

Reinforced Insulation

Two insulation barriers are required to ensure that the applied part, which includes both patients and operators,

is isolated and protected from dangerous energy shocks due to any electrical shocks and excess energy hazards,

transient voltage spikes, insulation breakdown of the power architecture, mechanical damage, ignition, fire,

shorts developing between PCB tracks, air gaps, arcing, and ground loops and complies with the limited leakage

current during normal and single-fault conditions.

Within the electrical safety area, medical safety approval requires equipment to implement two Means of

Protection (MOP) such that if a failure occurs within one area, a second mechanism safeguards the operator and

/or patient against dangerous energy shocks from any electrical shocks and excess energy hazards, transient

voltage spikes, and insulation breakdown of the power architecture.

MINMAX medical safety AC–DC power supplies and medical safety DC–DC converters have been approved to the

new 3rd edition of the IEC60601-1 standard with a reinforced insulation level. Further, all medical safety AC–DC

power supplies provide 2xMOPP and are suitable as the primary reinforced insulation barrier. All medical safety

DC–DC converters provide 2xMOPP or 2xMOOP and are suitable as the secondary reinforced insulation barrier

of the insulation system structure to ensure the long-term safety of operators/patients.

Page 8: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

* EMC Test Level of Medical Safety DC-DC Converters

MP:Mains Part

LP:Live Part = Non Applied Part Electronic Circuit

AP:Applied Part = Applied Part Electronic Circuit

Medical/Healthcare Insulation System

Structure of the Medical Insulation System

Medical EMC Standards

All MINMAX medical safety AC–DC power supplies and DC–DC converters undergo 4th edition medical EMC

(Emission + Immunity) testing of LIFE-SUPPORTING ME EQUIPMENT to help us provide as much information as

possible during the design-in process to ensure that medical/healthcare equipment comply with EN50155:

2009+AI for EMI (Electromagnetic Interference) and EN60601-1-2: 2015 for EMS (Electromagnetic Susceptibility).

Phenomenon EN 60601-1Reference Clause

ReferenceStandard

MINMAX Test Level ofMedical DC-DC Converters

EMI

EMS

0.15 – 0.5MHz

0.5 – 5.0MHz

5.0 – 30MHz

Quasi-peak79 dBuV

Quasi-peak73 dBuV

Group 1Class A

Group 1Class A

EN 55011

EN 55011

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

IEC 61000-4-2

IEC 61000-4-3

IEC 61000-4-4

IEC 61000-4-5

IEC 61000-4-6

IEC 61000-4-8

80 to 2700MHz : 10V/m385 to 6000MHz : 9-28 V/m

Air Discharge : ±15KVDCContact Discharge : ±8KVDC

Conducted Emission

Radiated Emission

ESD Test

Radiated Immunity (RS)

Electrical Fast Transient (EFT)

Surge Immunity Test

Conducted Immunity (CS)

Power Frequency MagneticField (PFMF)

EN 55011

EN 55011

Quasi-peak73 dBuV

40 dBuV/m

47 dBuV/m

Average66 dBuVAverage60 dBuVAverage60 dBuV

30 – 230MHz

230 – 1000MHz

L1, L2, L1+L2 : ±2KVDC

L1 to L2 : ±2KVDC

0.15 to 80MHz : 10VrmsISM Frequency : 6 Vrms

X, Y, Z axis : 30 A/m

Page 9: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Cost vs. Risks to Safety

Although the 3rd edition of the IEC60601-1 standard offers medical/healthcare equipment manufacturers more

options related to the choice of power supply, the upcoming question of risk vs. cost must be considered, i.e.,

a cheaper power supply with a lower performance to save a few dollars versus a power supply with higher

specifications and safety certification that might cost more but reduce the risk as much as possible.

After all, if you get it wrong in medical device design, it can limit your market, compromise your brand, and

severely delay regulatory approval or worse.

As a result, the specifications of medical safety power supplies that are approved at the IEC/EN 60601-1 standard

and comply with Means of Protection (MOP) is preferred for medical/healthcare equipment manufacturers.

* EMC Test Level of Medical Safety AC-DC Power Supplies

Medical EMC Standards

Voltage Dips

Short Interruptions

Dips : 100% Reduction for 0.5 cycle at 50 Hz

100% Reduction for 1 cycle at 50 Hz

30% Reduction for 25/30 cycle at 50/60 Hz

Interruptions : 100% Reduction for 250/300 cycle at

50/60 Hz

Phenomenon EN 60601-1Reference Clause

ReferenceStandard

MINMAX Test Level ofMedical AC-DC Power Supplies

EMI

EMS

0.15 – 0.5MHz

0.5 – 5.0MHz

5.0 – 30MHz

Quasi-peak66-56 dBuVQuasi-peak

56 dBuVGroup 1Class B

Group 1Class B

EN 55011

EN 55011

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

EN 60601-1-2

IEC 61000-4-2

IEC 61000-4-3

IEC 61000-4-4

IEC 61000-4-5

IEC 61000-4-6

IEC 61000-4-8

80 to 2700MHz : 10V/m385 to 6000MHz : 9-28 V/m

Air Discharge : ±15KVDCContact Discharge : ±8KVDC

Conducted Emission

Radiated Emission

ESD Test

Radiated Immunity (RS)

Electrical Fast Transient (EFT)

Surge Immunity Test

Conducted Immunity (CS)

Power Frequency MagneticField (PFMF)

EN 55011

EN 55011

Quasi-peak60 dBuV

30 dBuV/m

37 dBuV/m

Average56-46 dBuV

Average46 dBuVAverage50 dBuV

30 – 230MHz

230 – 1000MHz

L, N, L+N : ±2KVDC

L to N : ±2KVDC

0.15 to 80MHz : 10VrmsISM Frequency : 6 Vrms

X, Y, Z axis : 30 A/m

EN 60601-1-2 IEC 61000-4-11

Page 10: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Highlighted Performance of Medical Safety Products

Ultra-wide Input Voltage Range

Fully Vacuum Encapsulated to Save Your System

MINMAX medical safety power solutions are fully vacuum encapsulated and use the glue of UL94V-0 grade.

MTBF > 4 Millions HRs

* Example : MIW06-24S12M

* Example : MIW06-24S12M

9–18, 18–36,

36–75VDC

Input VoltageRange

Fully RegulatedOutput Voltage

5, 12, 15,

±12VDC

Page 11: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Reinforced Insulation & 5KVAC Isolation for System Safety

* Example : MIW06-24S12M

Wider Operating Ambient Temperature Range

Wider operating temp. range by latest thermal

management technology and fully vacuum

encapsulated.

Wider Operating Temp. Range

-40 to +95ºC without air-forced fan or heatsink

cooling

Storage Temp. Range

-50 to +125ºC

Operating Humidity

95% rel. H

The 5KVAC I/O isolation with reinforced insulation

and vacuum encapsulated creates a solid electrical

barrier which to protect sensitive circuit load from

noise, electromagnetic disturbances, power bus

fluctuation, avoid the risk and ensure freedom

dangerous energy shock from any electric shock and

excess energy hazards, transient voltage spike,

insulation breakdown of power architecture, mechanical

damage, ignition, fire and short developing between

PCB tracks, air gaps, arcing and ground loop that

provide safety on long-term operation of medical/

healthcare equipment.

* Please refer to the "Isolation Voltage and Safety Distance in the 3rd Edition of the IEC60601-1 Standard & Leakage Current in the 3rd Edition

of the IEC60601-1 Standard & Insulation Type" on Page 04 & 05 & 06 for more information.

* Example : MIW06-24S12M

8mm Creepage & Clearance with 2xMOPP Level

Isolation 5KVAC/60sec with Reinforced Insulation

Low Leakage Current < 2uA

Low I/O Isolation Capacitance 40pF max.

Primary Secondary

Page 12: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

* Example : MIW06-24S12M @ Ta=+25ºC * Example : MIW06-24S12M @ 100% Load & Ta=+25ºC

* Example : MIW06-24S12M @ 100% Load

Green Design for Higher Full Range Efficiency

High efficiency for whole output load, input line

& ambient temp. range by latest green design

technology helps to energy saving, thermal

management, minimize the temp. rise and size

miniaturization.

Green Design for No Min. Load / Dummy Load Requirement

With high stability feedback loop design, the

MINMAX medical safety power solutions may

not oscillate in no-load or light-load condition.

ECO-technology

Green Design for Energy Saving, Minimize Temperature Rise

Ultra low no-load power consumption by latest

green design technology helps to improve and

minimize the temp. rise (avoid thermal problem),

energy saving and prolong the battery life.

Very Low No Load Input Current 5mA @ 24Vin

Very Low No Load Power Consumption 0.12Watt @ 24Vin

Page 13: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Faster start-up time without overshoot ensures

the safety of your system.

The output voltage will still keep with excellent

accuracy, even though the input voltage, output

current and ambient temp. of the system are

unstable .

The output voltage of mainstream products may

undershoot and overshoot obviously during the

load changes. MINMAX medical safety power solutions are still keep with rated output voltage preciously.

Power Your System Precisely

Setting Accuracy ±1% Vom • Line Regulation ±0.5% • Load Regulation ±0.5% • High Transient Response •

Low Temperature Coefficient

Faster Start-up Time without Overshoot

The start-up time of MIW06M decreases from

250mS to 20mS which helps to avoid any

system timing failure caused by long start-up

time.

* Example : MIW06-24S12M * Example : Mainstream/Before

Superior Load Driving Capability

MINMAX medical safety power solutions have

superior load driving capability which can drive

your system during very low voltage and even

zero voltage output without start-up failure.

Page 14: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

No external component needed for conducted

emission meets EN55011 Class A.

Excellent EMC Performance

Excellent EMI performance by upgraded noise filtering technology helps to improve overall system EMI

performance on conduction and radiation emission.

Only few peripheral components needed for

radiated emission meets EN55011 Class B.

Only one E-cap. needed for ±2KV surge immunity by

IEC61000-4-5 with criteria A.

Excellent EMS performance by upgraded noise immunity technology helps to improve overall system EMS

performance on ESD, Surge, EFT, RS, CS and PFMF.

Only one E-cap. needed for ±2KV EFT

immunity by IEC61000-4-4 with criteria A.

* Please refer to the ‘’Medical EMC Standards’’ on Page 07 & 08 for more information.

Lower Ripple & Noise

Small Ripple & Noise for whole output load,

input line & ambient temp. range by upgraded

noise filtering technology helps to reduce the

peripheral components needed and noise

interference.

Page 15: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Protection Functions on Abnormal Operation

Certifications

MINMAX medical safety power solutions are the instantaneous protection for UVP, OCP, OVP & SCP which help

to ensure long-term safety of system operation.

* Example : MIW06-24S12M

Higher Start-up & Shutdown

Precision

OCP byHiccup Mode

Short Circuit Pin

as low as 0.25 Watt by Hiccup Mode

Spike ClampingUnder

VoltageProtection

ShortCircuit

Protection

OverLoad

Protection

OverVoltage

Protection

DESIGN&TECHNICAL SUPPORT

ENVIRONMENTALLY FRIENDLY

Page 16: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Successful Applications

MINMAX • Medical / Healthcare

Healthcare Information System

Dental Equipment CT Scanning

ECG, EEG

Medical Assist System CPAP Machine

Medical Panel PC

Mobile Cart Computer

Page 17: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Medical Safety DC-DC Converters

Medical Safety Products Overview

Medical Safety AC-DC Power Supplies

IEC

/EN

60

60

1-1

3rd

AN

SI/A

AM

I ES

60

60

1-1

Seri

es

Out

put

Po

wer

Inp

ut V

olta

ge

Rang

e(V

DC

)

Out

put

Vo

ltag

e (

VD

C)

Effic

ienc

y

Iso

latio

n(V

AC

)

Op

erat

ing

Tem

p.

Rang

e

Out

put

Reg

ulat

ion

OC

P

SCP

Pro

tect

ion

Leve

l

OV

P

5, 12, 15, ±5, ±12, ±15

5, 12, 15

1W • SIP Package

MAU400

MAU01M

75%

81%

1W

1W

-25~+85°C

-40~+95°C

4.5-5.5,10.8-13.2

4.5-5.5, 10.8-13.2,21.6-26.4

3000VACReinforced

4000VACReinforced

2xMOOP

2xMOPP

4.5-5.5, 10.8-13.2,21.6-26.4

4.5-5.5, 10.8-13.2,21.6-26.4

5, 12, 15,±12, ±15

5, 12, 15,±12, ±15

1-2W • SMD Package

84%

75%

1W

2W

-40~+95°C

-25~+80°C

MSCU01M

MSHU100

4000VACReinforced

4000VACReinforced

2xMOPP

2xMOOP

4.5-5.5, 10.8-13.2,21.6-26.4

9-40, 18-80,36-160

4.5-9, 9-18, 18-36, 36-75

9-18, 18-36,36-75

5, 12, 15,±12, ±15

5, 12,±12, ±15

5, 5.8, 12, 15,±12, ±15

5,12, 15,±12, ±15

2-6W • DIP Package

75%

83%

87%

89%

2W

3W

3.5W

6W

MDHU100

MIHW2000

MIW03M

MIW06M

4000VACReinforced

4000VACReinforced

5000VACReinforced

5000VACReinforced

-40~+75°C

-40~+85°C

-40~+80°C

5, 5.1, 12, ±12, ±15

5, 5.1, 12, 15, 24, ±12, ±15

5, 5.1, 12, 15, 24, ±12, ±15

10-20W • 2"×1" Package

82%

90%

90%

10W

15W

20W

MKW10M

MKW15M

MKW20M

9-18, 18-36,36-75

9-18, 18-36,36-75

9-18, 18-36,36-75

4200VACReinforced

4200VACReinforced

4200VACReinforced

2xMOOP

2xMOPP

2xMOPP

2xMOOP

2xMOOP

2xMOPP

2xMOPP

-25~+80°C

-40~+85°C

-40~+97°C

-40~+95°C

No

Min

. Lo

ad

IEC

/EN

60

60

1-1

3rd

AN

SI/A

AM

I ES

60

60

1-1

OC

P/S

CP

OV

P

Seri

es

Out

put

Po

wer

Inp

ut V

olta

ge

Rang

e(V

AC

)

Out

put

Vo

ltag

e

(V

DC

)

Iso

latio

n(V

AC

)

Effic

ienc

y

Op

erat

ing

Tem

p.

Rang

e

Pack

age

Prot

ectio

n Le

vel

AJM-24

APM-40

AYM-60

24W

40W

60W

4000VACReinforced

4000VACReinforced

85%

85%

88%

-40~+80°C

-40~+80°C

-40~+80°C

2xMOPP

2xMOPP

2xMOPP

PCBChassisDIN-Rail

PCBChassisDIN-Rail

4000VACReinforced

PCBChassisDIN-Rail

85-264

85-264

85-264

5, 12, 15, 24,±12 ±15

5, 9, 12, 15, 24,±12 ±15

5.1, 12, 15, 24, 48

Out

put

Reg

ulat

ion

No

Min

. Lo

ad

Page 18: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Medical Safety DC-DC Converters

MAU400 Series1W

Features

Industrial Standard SIP-7 Package

I/O Isolation 3000VAC with Reinforced Insulation, rated for

300Vrms Working Voltage

Operating Ambient Temp. Range -25°C to +85°C

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 1xMOPP & 2xMOOP per 3rd Edition of

IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

Model Selection Guide

MAU401

MAU402

MAU403

MAU404

MAU405

MAU406

MAU411

MAU412

MAU413

MAU414

MAU415

MAU416

200

80

65

±100

±40

±35

200

80

65

±100

±40

±35

66%

66%

66%

66%

72%

73%

66%

66%

66%

66%

74%

75%

5

12

15

±5

±12

±15

5

12

15

±5

±12

±15

4.5 - 5.5

10.8 - 13.2

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)Model

Number

MAU01M Series1W

*There are different features & spec. by each series.For detailed series datasheet, please refer to www.minmax.com.tw

Features

Industrial Standard SIP-7 Package

I/O Isolation 4000VAC with Reinforced Insulation, rated for

300Vrms Working Voltage

Low Leakage Current < 2µA

Operating Ambient Temp. Range -40°C to +95°C

Short Circuit Protection

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved with CE Marking

Model Selection Guide

MAU01-05S05M

MAU01-05S12M

MAU01-05S15M

MAU01-12S05M

MAU01-12S12M

MAU01-12S15M

MAU01-24S05M

MAU01-24S12M

MAU01-24S15M

200

84

68

200

84

68

200

84

68

79%

80%

81%

79%

81%

79%

76%

79%

79%

5

12

15

5

12

15

5

12

15

4.5 - 5.5

10.8 - 13.2

21.6 - 26.4

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)Model

Number

Page 19: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

4.5 - 5.5

10.8 - 13.2

21.6 - 26.4

MSCU01-05S05M

MSCU01-05S12M

MSCU01-05S15M

MSCU01-05D12M

MSCU01-05D15M

MSCU01-12S05M

MSCU01-12S12M

MSCU01-12S15M

MSCU01-12D12M

MSCU01-12D15M

MSCU01-24S05M

MSCU01-24S12M

MSCU01-24S15M

MSCU01-24D12M

MSCU01-24D15M

200

84

68

±42

±33

200

84

68

±42

±33

200

84

68

±42

±33

76%

80%

83%

80%

84%

76%

79%

80%

79%

80%

76%

80%

80%

80%

80%

5

12

15

±12

±15

5

12

15

±12

±15

5

12

15

±12

±15

ModelNumber

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)

Medical Safety DC-DC Converters

MSCU01M Series

*There are different features & spec. by each series.For detailed series datasheet, please refer to www.minmax.com.tw

1W

Features

Industrial Standard SMD Package

I/O Isolation 4000VAC with Reinforced Insulation, rated for

250Vrms Working Voltage

Low Leakage Current < 2µA

Operating Ambient Temp. Range -40ºC to 95ºC

Short Circuit Protection

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved with CE Marking

Model Selection Guide

MSHU100 Series 2W

Features

Industrial Standard SMD Package

I/O Isolation 4000VAC with Reinforced Insulation, rated for

300Vrms Working Voltage

Low Leakage Current < 2µA

Operating Ambient Temp. Range -25°C to +80°C

Qualified for Lead-free Reflow Solder Process According to

IPC/JEDEC J-STD-020D.1

Tape & Reel Package Available

Water-washable Process Available(option)

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 1xMOPP & 2xMOOP per 3rd Edition of

IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

Model Selection Guide

4.5 - 5.5

10.8 - 13.2

21.6 - 26.4

MSHU102

MSHU104

MSHU105

MSHU108

MSHU109

MSHU112

MSHU114

MSHU115

MSHU118

MSHU119

MSHU122

MSHU124

MSHU125

MSHU128

MSHU129

400

165

133

±83

±66

400

165

133

±83

±66

400

165

133

±83

±66

66%

66%

66%

72%

73%

66%

66%

66%

74%

75%

66%

66%

66%

74%

75%

5

12

15

±12

±15

5

12

15

±12

±15

5

12

15

±12

±15

ModelNumber

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)

Page 20: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

MIHW2022

MIHW2023

MIHW2026

MIHW2027

MIHW2032

MIHW2033

MIHW2036

MIHW2037

MIHW2042

MIHW2043

MIHW2046

MIHW2047

600

250

±125

±100

600

250

±125

±100

600

250

±125

±100

78%

83%

83%

83%

78%

83%

83%

83%

78%

83%

83%

83%

5

12

±12

±15

5

12

±12

±15

5

12

±12

±15

ModelNumber

9 - 40

18 - 80

36 - 160

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)

Medical Safety DC-DC Converters

MDHU100 Series

*There are different features & spec. by each series.For detailed series datasheet, please refer to www.minmax.com.tw

2W

Features

Industrial Standard DIP-16 Package

I/O Isolation 4000VAC with Reinforced Insulation, rated for

300Vrms Working Voltage

Low Leakage Current < 2µA

Operating Ambient Temp. Range -25°C to +80°C

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 1xMOPP & 2xMOOP per 3rd Edition of

IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

Model Selection Guide

MIHW2000 Series3W

Features

Industrial Standard DIP-24 Package

Ultra-wide 4:1 Input Voltage Range

Fully Regulated Output Voltage

I/O Isolation 4000VAC with Reinforced Insulation,

rated for 1000Vrms Working Voltage

Low Leakage Current < 2µA

Operating Ambient Temp. Range -40°C to +85°C

Under-Voltage, Overload and Short Circuit Protection

Conducted EMI EN55011/22 Class A & FCC Level A Approved

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 1xMOPP & 2xMOOP per 3rd Edition of

IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

Model Selection Guide

MDHU102

MDHU104

MDHU105

MDHU108

MDHU109

MDHU112

MDHU114

MDHU115

MDHU118

MDHU119

MDHU122

MDHU124

MDHU125

MDHU128

MDHU129

400

165

133

±83

±66

400

165

133

±83

±66

400

165

133

±83

±66

66%

66%

66%

72%

73%

66%

66%

66%

74%

75%

66%

66%

66%

74%

75%

5

12

15

±12

±15

5

12

15

±12

±15

5

12

15

±12

±15

4.5 - 5.5

10.8 - 13.2

21.6 - 26.4

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)Model

Number

Page 21: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Medical Safety DC-DC Converters

MIW03M Series3.5W

Features

Industrial Standard DIP-24 Package

Wide 2:1 Input Voltage Range

Fully Regulated Output Voltage

I/O Isolation 5000VAC with Reinforced Insulation, rated for

250Vrms Working Voltage

Creepage & Clearance Distance meet 8mm

Low Leakage Current < 2µA

Operating Ambient Temp. Range -40ºC to 97ºC

No Min. Load Requirement

Under-Voltage, Overload/Voltage and Short Circuit Protection

Conducted EMI EN55011 Class A & FCC Level A Approved

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved with CE Marking

Model Selection Guide

MIW06M Series6W

Features

Industrial Standard DIP-24 Package

Wide 2:1 Input Voltage Range

Fully Regulated Output Voltage

I/O Isolation 5000VAC with Reinforced Insulation, rated for

250Vrms Working Voltage

Creepage & Clearance Distance meet 8mm

Low Leakage Current < 2µA

Operating Ambient Temp. Range -40°C to 95°C

No Min. Load Requirement

Under-Voltage, Overload/Voltage and Short Circuit Protection

Conducted EMI EN55011 Class A & FCC Level A Approved

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved with CE Marking

Model Selection Guide

MIW03-05S05MMIW03-05S058MMIW03-05S12MMIW03-05S15MMIW03-05D12MMIW03-05D15MMIW03-12S05MMIW03-12S12MMIW03-12S15MMIW03-12D12MMIW03-12D15MMIW03-24S05MMIW03-24S12MMIW03-24S15MMIW03-24D12MMIW03-24D15MMIW03-48S05MMIW03-48S12MMIW03-48S15MMIW03-48D12MMIW03-48D15M

700600290235

±145±115700290235

±145±115700290235

±145±115700290235

±145±115

83%83%84%84%84%84%83%87%87%87%87%83%86%87%87%86%83%86%85%84%84%

55.81215

±12±15

51215

±12±15

51215

±12±15

51215

±12±15

ModelNumber

4.5 - 9

9 - 18

18 - 36

36 - 75

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)

*There are different features & spec. by each series.For detailed series datasheet, please refer to www.minmax.com.tw

MIW06-12S05M

MIW06-12S12M

MIW06-12S15M

MIW06-12D12M

MIW06-12D15M

MIW06-24S05M

MIW06-24S12M

MIW06-24S15M

MIW06-24D12M

MIW06-24D15M

MIW06-48S05M

MIW06-48S12M

MIW06-48S15M

MIW06-48D12M

MIW06-48D15M

1200

500

400

±250

±200

1200

500

400

±250

±200

1200

500

400

±250

±200

84%

87%

86%

87%

87%

84%

87%

87%

86%

87%

84%

87%

89%

87%

88%

5

12

15

±12

±15

5

12

15

±12

±15

5

12

15

±12

±15

ModelNumber

9 - 18

18 - 36

36 - 75

Iout

(mA)max EfficiencyVin

(VDC)Vout

(VDC)

Page 22: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Medical Safety DC-DC Converters

*There are different features & spec. by each series.For detailed series datasheet, please refer to www.minmax.com.tw

MKW15M Series15W

Features

Industrial Standard 2"× 1" Package

Wide 2:1 Input Voltage Range

Fully Regulated Output Voltage

I/O Isolation 4200VAC with Reinforced Insulation, rated for

300Vrms Working Voltage

Low Leakage Current < 5µA

Operating Ambient Temp. Range -40°C to +85°C

No Min. Load Requirement

Under-voltage, Overload/Voltage and Short Circuit Protection

Conducted EMI EN55011 Class A & FCC Level A Approved

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved with CE Marking

Model Selection Guide

MKW10-12S05M

MKW10-12S051M

MKW10-12S12M

MKW10-12D12M

MKW10-12D15M

MKW10-24S05M

MKW10-24S051M

MKW10-24S12M

MKW10-24D12M

MKW10-24D15M

MKW10-48S05M

MKW10-48S051M

MKW10-48S12M

MKW10-48D12M

MKW10-48D15M

1600

1600

835

±417

±333

2000

2000

835

±417

±333

2000

2000

835

±417

±333

74%

75%

80%

80%

81%

75%

76%

81%

81%

82%

75%

76%

81%

81%

82%

5

5.1

12

±12

±15

5

5.1

12

±12

±15

5

5.1

12

±12

±15

ModelNumber

9 - 18

18 - 36

36 - 75

Iout

(mA)maxVin

(VDC)Vout

(VDC) Efficiency

Model Selection Guide

MKW15-12S05MMKW15-12S051MMKW15-12S12MMKW15-12S15MMKW15-12S24MMKW15-12D12MMKW15-12D15MMKW15-24S05MMKW15-24S051MMKW15-24S12MMKW15-24S15MMKW15-24S24MMKW15-24D12MMKW15-24D15MMKW15-48S05MMKW15-48S051MMKW15-48S12MMKW15-48S15MMKW15-48S24MMKW15-48D12MMKW15-48D15M

3000300012501000625

±625±5003000300012501000625

±625±5003000300012501000625

±625±500

86%86%89%88%88%88%89%88%88%89%89%90%90%89%88%88%88%90%89%89%88%

55.1121524

±12±15

55.1121524

±12±15

55.1121524

±12±15

ModelNumber

9 - 18

18 - 36

36 - 75

Iout

(mA)maxVin

(VDC)Vout

(VDC) Efficiency

MKW10M Series10W

Features

Industrial Standard 2"× 1" Package

Wide 2:1 Input Voltage Range

Fully Regulated Output Voltage

I/O Isolation 4200VAC with Reinforced Insulation,

rated for 1000Vrms Working Voltage

Low Leakage Current < 10µA

Operating Ambient Temp. Range -40°C to +75°C

Under-voltage, Overload and Short Circuit Protection

EMI Emission EN55011/22 Class A & FCC Level A Approved

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 1xMOPP & 2xMOOP per 3rd Edition of

IEC/EN 60601-1 & ANSI/AAMI ES60601-1 Approved

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

Page 23: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Medical Safety DC-DC Converters

MKW20M Series

*There are different features & spec. by each series.For detailed series datasheet, please refer to www.minmax.com.tw

20W

Features

Industrial Standard 2"× 1" Package

Wide 2:1 Input Voltage Range

Fully Regulated Output Voltage

I/O Isolation 4200VAC with Reinforced Insulation,

rated for 300Vrms Working Voltage

Low Leakage Current < 5µA

Operating Ambient Temp. Range -40°C to +80°C

No Min. Load Requirement

Under-voltage, Overload/Voltage and Short Circuit Protection

Conducted EMI EN55011 Class A & FCC Level A Approved

Medical EMC Standard with 4th Edition of EMI EN55011 and

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved with CE Marking

AJM-24 Series 24W

Features

Fully Encapsulated Plastic Case for PCB, Chassis and DIN-Rail

Mounting Version

Universal Input 85-264VAC, 47-440Hz

Protection Class II as per IEC/EN 60536

I/O Isolation 4000VAC with Reinforced Insulation

Operating Ambient Temp. Range -40°C to +80°C

Overload/Voltage and Short Circuit Protection

EMI Emission EN55011/32 Class B & FCC Level B Approved

EMC Immunity EN61000-4-2,3,4,5,6,8,11 Approved

Medical EMC Standard with 4th Edition of EMI EN55011 &

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved

UL508 Safety Approval Specifically for Industrial Application

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

MKW20-12S05MMKW20-12S051MMKW20-12S12MMKW20-12S15MMKW20-12S24MMKW20-12D12MMKW20-12D15MMKW20-24S05MMKW20-24S051MMKW20-24S12MMKW20-24S15MMKW20-24S24MMKW20-24D12MMKW20-24D15MMKW20-48S05MMKW20-48S051MMKW20-48S12MMKW20-48S15MMKW20-48S24MMKW20-48D12MMKW20-48D15M

4000400016701333840

±840±6704000400016701333840

±840±6704000400016701333840

±840±670

86%86%89%88%89%89%89%88%88%89%89%90%90%90%88%88%89%90%89%89%90%

55.1121524

±12±15

55.1121524

±12±15

55.1121524

±12±15

ModelNumber

9 - 18

18 - 36

36 - 75

Iout

(mA)maxVin

(VDC)Vout

(VDC) Efficiency

Model Selection Guide

Medical Safety AC-DC Power Supplies

AJM-24S05

AJM-24S09

AJM-24S12

AJM-24S15

AJM-24S24

AJM-24D12

AJM-24D15

3000

2666

2000

1600

1000

±1000

±800

77%

82%

83%

82%

85%

84%

84%

5

9

12

15

24

±12

±15

ModelNumber

85 - 264

Iout

(mA)maxVin

(VAC)Vout

(VDC) Efficiency

Model Selection Guide

Page 24: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

Medical Safety AC-DC Power Supplies

APM-40 Series

*There are different features & spec. by each series.For detailed series datasheet, please refer to www.minmax.com.tw

40W

Features

Fully Encapsulated Plastic Case for PCB, Chassis and DIN-Rail

Mounting Version

Universal Input 85-264VAC, 47-440Hz

Protection Class II as per IEC/EN 60536

I/O Isolation 4000VAC with Reinforced Insulation

Operating Ambient Temp. Range -40°C to +80°C

Overload/Voltage and Short Circuit Protection

EMI Emission EN55011/32 Class B & FCC Level B Approved

EMC Immunity EN61000-4-2,3,4,5,6,8,11 Approved

Medical EMC Standard with 4th Edition of EMI EN55011 &

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved

UL508 Safety Approval Specifically for Industrial Application

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

AYM-60 Series60W

Features

Fully Encapsulated Plastic Case for PCB, Chassis and DIN-Rail

Mounting Version

Universal Input 85-264VAC, 47-440Hz

Protection Class II as per IEC/EN 60536

I/O Isolation 4000VAC with Reinforced Insulation

Operating Ambient Temp. Range -40°C to +80°C

Overload/Voltage and Short Circuit Protection

EMI Emission EN55011/32 Class B & FCC Level B Approved

EMC Immunity EN61000-4-2,3,4,5,6,8,11 Approved

Medical EMC Standard with 4th Edition of EMI EN55011 &

EMS EN60601-1-2 Approved

Medical Safety with 2xMOPP per 3rd Edition of IEC/EN 60601-1

& ANSI/AAMI ES60601-1 Approved

UL508 Safety Approval Specifically for Industrial Application

UL/cUL/IEC/EN 60950-1 Safety Approval & CE Marking

AYM-60S051

AYM-60S12

AYM-60S15

AYM-60S24

AYM-60S48

10000

5000

4000

2500

1250

84%

87%

87%

87%

88%

5.1

12

15

24

48

ModelNumber

85 - 264

Iout

(mA)maxVin

(VAC)Vout

(VDC) Efficiency

Model Selection Guide

APM-40S05

APM-40S12

APM-40S15

APM-40S24

APM-40D12

APM-40D15

8000

3330

2660

1660

±1660

±1330

81%

84%

85%

84%

84%

85%

5

12

15

24

±12

±15

ModelNumber

85 - 264

Iout

(mA)maxVin

(VAC)Vout

(VDC) Efficiency

Model Selection Guide

Page 25: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

MINMAX Power Solutions

DC-DC Converters • AC-DC Power Supplies1-75W

Ultra-high Isolation family equipped with very high common

mode transient immunity with 15KV/µs qualifies and I/O isolation

4000 to 5000VAC with reinforced insulation, rated for 1000Vrms

working voltage.

SIP Package DC-DC Converters1-2W

DIP Package DC-DC Converters2-6W

2"×1" Package DC-DC Converters10-20W

Chassis & DIN-Rail Mounting DC-DC Converters6-60W

Our General Industrial DC-DC converters are used in every

corner of modern equipment & systems about everywhere.

We offer different electrical specifications and create competitive

advantage performance and cost effective to meet your critical

design.

SMD Package DC-DC Converters1-6W

SIP Package DC-DC Converters1-5W

DIP Package DC-DC Converters1-10W

1"×1" Package DC-DC Converters10-30W

2"×1" Package DC-DC Converters30-50W

Chassis & DIN-Rail Mounting DC-DC Converters6-60W

Switching Regulators0.5-1A

Encapsulated Package AC-DC Power Supplies2-60WThe medical grade DC-DC converters & AC-DC Power Supplies

for demanding applications in both medical and healthcare

instrumentations. All medical products are approved to latest

medical safety standards: ANSI/AAMI ES60601-1,

IEC/EN 60601-1 3rd edition with 2xMOPP/2xMOOP.

SIP Package DC-DC Converters1W

SMD Package DC-DC Converters1-2W

DIP Package DC-DC Converters2-6W

Encapsulated Package AC-DC Power Supplies24-60W

2"×1" Package DC-DC Converters10-20W

Our Railway Certified DC-DC converters are designed for

stringent requirements and harsh environment.

All railway products comply with EN50155 certified and fire

protection EN45545-2 approval.

DIP Package DC-DC Converters3W

2"×1" Package DC-DC Converters10-40W

Quarter Brick DC-DC Converters50-75W

Page 26: Medical Power Solution Guide-EN - DICEL Electronics minmax.pdf · 2018. 6. 14. · MINMAX offers a large standard range of high-isolation and reinforced insulated medical safety DC–DC

for more info., please go to www.minmax.com.tw