Measurement Report (ISO 17025) Agilent N9912A Performance Verification Calibration Facility N/A Test Date 09-Dec-2010 Report Number smoke test on build 16-873S smoke test on build 16 Order Number Customer N/A 106, 110, 111, 231, 235, 236, 302, 308 Options N/A Asset Number / Unique ID A.05.31 Firmware Version US47500037 Serial Number N9912A Model Number N9912A - Performance Verification - Agilent Recommended Test Plan Name Version E.02.11 Test Manager Agilent N7800A Test Management Environment (TME) Software Calibration Application Agilent N7841A Fieldfox RF Analyzer Calibration Application Version E.02.00 As Received Test Session See Individual Test Data Humidity See Individual Test Data Temperature Authorized by N/A Results All measurement and test results guardbanded by the measurement uncertainties. The expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of approximately 95%, unless otherwise stated. Measurement Uncertainty Notes Page 1 of 52 Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID: STP-N9912A-US47500037

Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

  • Upload
    others

  • View
    15

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Measurement Report (ISO 17025)

Agilent N9912A Performance Verification

Calibration Facility

N/A

Test Date 09-Dec-2010Report Number smoke test on build 16-873S

smoke test on build 16Order Number

Customer N/A

106, 110, 111, 231, 235, 236, 302, 308Options

N/AAsset Number / Unique ID

A.05.31Firmware Version

US47500037Serial Number

N9912AModel Number

N9912A - Performance Verification - Agilent RecommendedTest Plan Name

Version E.02.11Test Manager Agilent N7800A Test Management Environment (TME) Software

Calibration Application Agilent N7841A Fieldfox RF Analyzer Calibration Application Version E.02.00

As ReceivedTest Session

See Individual Test DataHumidity

See Individual Test DataTemperature

Authorized by N/A

Results All measurement and test results guardbanded by the measurement uncertainties.

The expanded uncertainty of measurement is stated as the standard uncertainty of

measurement multiplied by the coverage factor k=2, which for a normal distribution

corresponds to a coverage probability of approximately 95%, unless otherwise stated.

Measurement

Uncertainty

Notes

Page 1 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 2: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Test Summary

Model N9912A Serial US47500037 Test Date 09-Dec-2010

Report Number

smoke test on build 16-873S

Test DateTest Name Test Result

PASS 09-Dec-2010Error Terms

Indeterminate 09-Dec-2010Frequency Accuracy

Complete 09-Dec-2010Station Calibration

PASS 09-Dec-2010High Low Power

PASS 09-Dec-2010Displayed Average Noise Level

PASS 09-Dec-2010IF Frequency Response

PASS 09-Dec-2010Noise Sidebands

PASS 09-Dec-2010Absolute Amplitude Accuracy

PASS 09-Dec-2010Frequency Response Preamp Off

PASS 09-Dec-2010Frequency Response Preamp On

PASS: All measured values are within the specification.

Indeterminate: All measured values are within the specification. However, one or more of the measured values plus its

measurement uncertainty exceeds the specification.

Complete: Functional test.

Page 2 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 3: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Test Equipment Summary

Model N9912A Serial US47500037 Test Date 09-Dec-2010

Report Number

smoke test on build 16-873S

Description Serial/Unique ID Trace Number Cal DueModel1.8 GHz Low Pass Filter 10955-0491

Frequency Standard 1 01-Jan-201210MHz Ext Reference

N(m) to N(m) RF Cable 111500C

Power Splitter 111667A

Attenuator Switch Driver 3748A0411511713A

Interconnect cable 111716A

N(m) to N(m) Adapter 11250-1475

Short Termination 11250-3607

Frequency Counter 3546A01585 01-Jan-201253132A

Attenuator Cable 18120-2703

Attenuator Cable 28120-2703

6dB Fixed Attenuator 1 01-Jan-20128491A Opt 006

1 dB Step Attenuator 3246A15579 12489 09-Nov-20118494G

10 dB Step Attenuator 3247A14961 12490 09-Nov-20118496G

Power Meter US38470132 01-Jan-2012E4419A

Signal Generator US44270223 01-Jan-2012E8257D

Power Sensor MY41498147 440530 25-Feb-2011E9304A

Power Sensor MY41496197 431135 25-Oct-2011E9304A H18

Page 3 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 4: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Error Terms

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

1250-3607 Short Termination 1

S11

Last Factory Calibration Run "Wednesday, December 08, 2010"

-0.60

-0.40

-0.20

0.00

0.20

0.40

2.00

301.

90

601.

80

901.7

0

1,20

1.60

1,50

1.50

1,80

1.40

2,10

1.30

2,40

1.20

2,70

1.10

3,00

1.00

3,30

0.90

3,60

0.80

3,90

0.70

4,20

0.60

4,50

0.50

4,80

0.40

5,10

0.30

5,40

0.20

5,70

0.10

6,00

0.00

Frequency (MHz)

Measu

red

Err

or

(dB

)

Measured

Error

Test Limit*

Result

Test

Frequency

(MHz)

Measured

Error

(dBm)

Lower

Limit*

(+/- dB)

Upper

Limit*

(+/- dB)

Maximum Measured Error

5,760.08 0.12 -0.60 0.40 PASS

Measured

Error

(dBm)

Lower

Limit*

(+/- dB)

Upper

Limit*

(+/- dB)

Result

Test

Frequency

(MHz)

Minimum Measured Error

4,560.48 -0.25 -0.60 0.40 PASS

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 4 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 5: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Frequency Accuracy

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result Indeterminate

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

0955-0491 1.8 GHz Low Pass Filter 1

10MHz Ext Reference Frequency Standard 1 01-Jan-2012

53132A Frequency Counter 3546A01585 01-Jan-2012

Measured

Frequency

(Hz)

Calculated

Frequency

(PPM)

Measurement

Uncertainty

(+/- PPM)

Specification

(+/- PPM)

Result

1,000,000,020.90 0.021 5.0012 Indeterminate 2

1,800,000,037.00 0.021 5.0012 Indeterminate 2

Page 5 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 6: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

High Low Power

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

E4419A Power Meter US38470132 01-Jan-2012

E9304A H18 Power Sensor MY41496197 431135 25-Oct-2011

Page 6 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 7: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

High Low Power

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Low Power

-35.0

-30.0

-25.0

-20.0

-15.0

-10.0

-5.0

0.0

2.10

24.0

0

84.0

0

183.

00

392.

00

760.

00

1,08

0.00

1,55

0.00

1,70

0.00

3,10

0.00

3,30

0.00

5,95

0.00

Frequency (MHz)

Po

wer

(dB

m)

Power

Test Limit*

Result

Upper

Limit*

(dB)

Power

(dBm)

Frequency

(MHz)

Lower

Limit*

(dB)

2.10 -24.386 PASS-20.00-30.00

24.00 -24.378 PASS-20.00-30.00

84.00 -23.625 PASS-20.00-30.00

183.00 -24.126 PASS-20.00-30.00

392.00 -22.928 PASS-20.00-30.00

760.00 -26.469 PASS-20.00-30.00

1,080.00 -25.132 PASS-20.00-30.00

1,550.00 -25.475 PASS-20.00-30.00

1,700.00 -25.411 PASS-20.00-30.00

3,100.00 -26.292 PASS-20.00-35.00

3,300.00 -26.971 PASS-20.00-35.00

5,950.00 -31.736 PASS-20.00-35.00

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 7 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 8: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

High Low Power

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

High Power

-2.0

0.0

2.0

4.0

6.0

8.0

10.0

2.10

24.0

0

84.0

0

183.

00

392.

00

760.

00

1,08

0.00

1,55

0.00

1,70

0.00

3,10

0.00

3,30

0.00

5,95

0.00

Frequency (MHz)

Po

wer

(dB

m)

Power

Test Limit*

Result

Upper

Limit*

(dB)

Power

(dBm)

Frequency

(MHz)

Lower

Limit*

(dB)

2.10 4.175 PASS 10.00-2.00

24.00 5.206 PASS 10.00-2.00

84.00 6.258 PASS 10.00-2.00

183.00 5.543 PASS 10.00-2.00

392.00 6.200 PASS 10.00-2.00

760.00 3.435 PASS 10.00-2.00

1,080.00 5.324 PASS 10.00-2.00

1,550.00 5.911 PASS 10.00-2.00

1,700.00 5.949 PASS 10.00-2.00

3,100.00 4.348 PASS 10.00-2.00

3,300.00 3.871 PASS 10.00-2.00

5,950.00 0.692 PASS 10.00-2.00

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 8 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 9: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Displayed Average Noise Level

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

11713A Attenuator Switch Driver 3748A04115

11716A Interconnect cable 1

8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011

8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011

Page 9 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 10: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Displayed Average Noise Level

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Preamp On

-160

-140

-120

-100

-80

-60

10.0

0

1,40

0.00

2,40

0.00

2,99

0.00

3,39

0.00

3,79

0.00

4,19

0.00

4,59

0.00

4,99

0.00

5,39

0.00

5,79

0.00

Frequency (MHz)

DA

NL

(d

Bm

)

Display Average

Noise Level

Measurement

Uncertainty

Specification

Frequency

(MHz)

DANL

(dBm)

Measurement

Uncertainty

(dB)

Specification

(dBm)

Result

10.00 -150.50 0.23 PASS-143.00

100.00 -152.48 0.23 PASS-143.00

500.00 -151.07 0.23 PASS-143.00

1,000.00 -152.86 0.23 PASS-143.00

1,200.00 -151.69 0.23 PASS-143.00

1,400.00 -150.62 0.23 PASS-143.00

1,600.00 -150.51 0.23 PASS-143.00

1,800.00 -149.38 0.23 PASS-143.00

2,000.00 -150.29 0.23 PASS-143.00

2,200.00 -149.04 0.23 PASS-143.00

2,400.00 -147.21 0.23 PASS-140.00

2,650.00 -145.14 0.23 PASS-140.00

2,750.00 -149.54 0.23 PASS-140.00

2,830.00 -149.93 0.23 PASS-140.00

2,910.00 -149.77 0.23 PASS-140.00

2,990.00 -148.79 0.23 PASS-140.00

3,070.00 -148.23 0.23 PASS-140.00

3,150.00 -147.53 0.23 PASS-140.00

3,230.00 -147.31 0.23 PASS-140.00

3,310.00 -146.21 0.23 PASS-140.00

3,390.00 -148.60 0.23 PASS-140.00

3,470.00 -148.39 0.23 PASS-140.00

3,550.00 -147.95 0.23 PASS-140.00

3,630.00 -147.37 0.23 PASS-140.00

3,710.00 -147.34 0.23 PASS-140.00

3,790.00 -147.44 0.23 PASS-140.00

3,870.00 -146.94 0.23 PASS-140.00

Page 10 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 11: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Displayed Average Noise Level

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Frequency

(MHz)

DANL

(dBm)

Measurement

Uncertainty

(dB)

Specification

(dBm)

Result

3,950.00 -146.40 0.23 PASS-140.00

4,030.00 -146.60 0.23 PASS-140.00

4,110.00 -145.93 0.23 PASS-140.00

4,190.00 -145.21 0.23 PASS-140.00

4,270.00 -144.23 0.23 PASS-140.00

4,350.00 -145.59 0.23 PASS-140.00

4,430.00 -145.44 0.23 PASS-140.00

4,510.00 -145.84 0.23 PASS-140.00

4,590.00 -147.32 0.23 PASS-140.00

4,670.00 -146.90 0.23 PASS-140.00

4,750.00 -146.06 0.23 PASS-140.00

4,830.00 -145.66 0.23 PASS-140.00

4,910.00 -145.95 0.23 PASS-140.00

4,990.00 -144.70 0.23 PASS-140.00

5,070.00 -144.53 0.23 PASS-132.00

5,150.00 -144.56 0.23 PASS-132.00

5,230.00 -141.95 0.23 PASS-132.00

5,310.00 -141.11 0.23 PASS-132.00

5,390.00 -141.38 0.23 PASS-132.00

5,470.00 -140.48 0.23 PASS-132.00

5,550.00 -140.19 0.23 PASS-132.00

5,630.00 -139.78 0.23 PASS-132.00

5,710.00 -139.56 0.23 PASS-132.00

5,790.00 -139.34 0.23 PASS-132.00

5,870.00 -138.72 0.23 PASS-132.00

5,950.00 -137.55 0.23 PASS-132.00

6,000.00 -137.16 0.23 PASS-132.00

Page 11 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 12: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

E8257D Signal Generator US44270223 01-Jan-2012

Page 12 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 13: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 0.007 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.03-2,450.00 1.30

PASS-0.05-2,400.00 1.30

PASS-0.08-2,350.00 1.30

PASS-0.11-2,300.00 1.30

PASS-0.11-2,250.00 1.30

PASS-0.09-2,200.00 1.30

PASS-0.06-2,150.00 1.30

PASS-0.02-2,100.00 1.30

PASS 0.01-2,050.00 1.30

PASS 0.01-2,000.00 1.30

PASS 0.01-1,950.00 1.30

PASS-0.01-1,900.00 1.30

PASS-0.03-1,850.00 1.30

PASS-0.05-1,800.00 1.30

PASS-0.06-1,750.00 1.30

PASS-0.05-1,700.00 1.30

PASS-0.03-1,650.00 1.30

PASS 0.00-1,600.00 1.30

PASS 0.01-1,550.00 1.30

PASS 0.01-1,500.00 1.30

PASS 0.01-1,450.00 1.30

PASS-0.01-1,400.00 1.30

PASS-0.02-1,350.00 1.30

PASS-0.03-1,300.00 1.30

PASS-0.03-1,250.00 1.30

Page 13 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 14: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 0.007 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.02-1,200.00 1.30

PASS 0.00-1,150.00 1.30

PASS 0.01-1,100.00 1.30

PASS 0.02-1,050.00 1.30

PASS 0.01-1,000.00 1.30

PASS 0.00-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.06-850.00 1.30

PASS-0.06-800.00 1.30

PASS-0.03-750.00 1.30

PASS 0.00-700.00 1.30

PASS 0.02-650.00 1.30

PASS 0.03-600.00 1.30

PASS 0.03-550.00 1.30

PASS 0.02-500.00 1.30

PASS 0.01-450.00 1.30

PASS-0.02-400.00 1.30

PASS-0.04-350.00 1.30

PASS-0.05-300.00 1.30

PASS-0.02-250.00 1.30

PASS 0.00-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.00-50.00 1.30

PASS-0.01 50.00 1.30

PASS-0.02 100.00 1.30

PASS-0.02 150.00 1.30

PASS-0.01 200.00 1.30

PASS 0.01 250.00 1.30

PASS 0.01 300.00 1.30

PASS 0.01 350.00 1.30

PASS 0.00 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.04 500.00 1.30

PASS-0.05 550.00 1.30

PASS-0.05 600.00 1.30

PASS-0.02 650.00 1.30

PASS 0.00 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.00 800.00 1.30

PASS-0.01 850.00 1.30

PASS-0.02 900.00 1.30

PASS-0.03 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 14 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 15: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 0.007 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.07 1,050.00 1.30

PASS-0.05 1,100.00 1.30

PASS-0.03 1,150.00 1.30

PASS-0.01 1,200.00 1.30

PASS 0.01 1,250.00 1.30

PASS 0.00 1,300.00 1.30

PASS 0.00 1,350.00 1.30

PASS-0.01 1,400.00 1.30

PASS-0.02 1,450.00 1.30

PASS-0.02 1,500.00 1.30

PASS-0.03 1,550.00 1.30

PASS-0.03 1,600.00 1.30

PASS-0.03 1,650.00 1.30

PASS-0.02 1,700.00 1.30

PASS-0.01 1,750.00 1.30

PASS-0.03 1,800.00 1.30

PASS-0.04 1,850.00 1.30

PASS-0.05 1,900.00 1.30

PASS-0.06 1,950.00 1.30

PASS-0.05 2,000.00 1.30

PASS-0.02 2,050.00 1.30

PASS 0.00 2,100.00 1.30

PASS 0.01 2,150.00 1.30

PASS 0.02 2,200.00 1.30

PASS 0.01 2,250.00 1.30

PASS-0.01 2,300.00 1.30

PASS-0.03 2,350.00 1.30

PASS-0.05 2,400.00 1.30

PASS-0.06 2,450.00 1.30

PASS-0.07 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 15 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 16: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 0.050 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.06-2,450.00 1.30

PASS-0.08-2,400.00 1.30

PASS-0.11-2,350.00 1.30

PASS-0.14-2,300.00 1.30

PASS-0.14-2,250.00 1.30

PASS-0.13-2,200.00 1.30

PASS-0.09-2,150.00 1.30

PASS-0.05-2,100.00 1.30

PASS-0.02-2,050.00 1.30

PASS-0.01-2,000.00 1.30

PASS-0.01-1,950.00 1.30

PASS-0.03-1,900.00 1.30

PASS-0.05-1,850.00 1.30

PASS-0.08-1,800.00 1.30

PASS-0.08-1,750.00 1.30

PASS-0.08-1,700.00 1.30

PASS-0.06-1,650.00 1.30

PASS-0.02-1,600.00 1.30

PASS-0.01-1,550.00 1.30

PASS 0.00-1,500.00 1.30

PASS-0.01-1,450.00 1.30

PASS-0.02-1,400.00 1.30

PASS-0.04-1,350.00 1.30

PASS-0.05-1,300.00 1.30

PASS-0.05-1,250.00 1.30

Page 16 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 17: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 0.050 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.05-1,200.00 1.30

PASS-0.02-1,150.00 1.30

PASS 0.00-1,100.00 1.30

PASS 0.00-1,050.00 1.30

PASS 0.00-1,000.00 1.30

PASS-0.02-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.07-850.00 1.30

PASS-0.08-800.00 1.30

PASS-0.05-750.00 1.30

PASS-0.02-700.00 1.30

PASS 0.01-650.00 1.30

PASS 0.02-600.00 1.30

PASS 0.02-550.00 1.30

PASS 0.01-500.00 1.30

PASS 0.00-450.00 1.30

PASS-0.02-400.00 1.30

PASS-0.05-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.00-200.00 1.30

PASS 0.00-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.00-50.00 1.30

PASS-0.01 50.00 1.30

PASS-0.02 100.00 1.30

PASS-0.03 150.00 1.30

PASS-0.01 200.00 1.30

PASS 0.00 250.00 1.30

PASS 0.01 300.00 1.30

PASS 0.01 350.00 1.30

PASS 0.01 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.03 500.00 1.30

PASS-0.05 550.00 1.30

PASS-0.06 600.00 1.30

PASS-0.03 650.00 1.30

PASS-0.01 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.00 800.00 1.30

PASS-0.01 850.00 1.30

PASS-0.01 900.00 1.30

PASS-0.02 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 17 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 18: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 0.050 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.06 1,050.00 1.30

PASS-0.05 1,100.00 1.30

PASS-0.02 1,150.00 1.30

PASS 0.00 1,200.00 1.30

PASS 0.02 1,250.00 1.30

PASS 0.01 1,300.00 1.30

PASS 0.01 1,350.00 1.30

PASS 0.00 1,400.00 1.30

PASS 0.00 1,450.00 1.30

PASS-0.01 1,500.00 1.30

PASS-0.02 1,550.00 1.30

PASS-0.03 1,600.00 1.30

PASS-0.02 1,650.00 1.30

PASS-0.01 1,700.00 1.30

PASS 0.00 1,750.00 1.30

PASS-0.02 1,800.00 1.30

PASS-0.02 1,850.00 1.30

PASS-0.04 1,900.00 1.30

PASS-0.05 1,950.00 1.30

PASS-0.04 2,000.00 1.30

PASS-0.01 2,050.00 1.30

PASS 0.01 2,100.00 1.30

PASS 0.03 2,150.00 1.30

PASS 0.04 2,200.00 1.30

PASS 0.03 2,250.00 1.30

PASS 0.01 2,300.00 1.30

PASS-0.01 2,350.00 1.30

PASS-0.02 2,400.00 1.30

PASS-0.04 2,450.00 1.30

PASS-0.05 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 18 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 19: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 1.000 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-2,450.00 1.30

PASS-0.07-2,400.00 1.30

PASS-0.10-2,350.00 1.30

PASS-0.14-2,300.00 1.30

PASS-0.13-2,250.00 1.30

PASS-0.12-2,200.00 1.30

PASS-0.09-2,150.00 1.30

PASS-0.03-2,100.00 1.30

PASS-0.01-2,050.00 1.30

PASS 0.00-2,000.00 1.30

PASS 0.00-1,950.00 1.30

PASS-0.02-1,900.00 1.30

PASS-0.04-1,850.00 1.30

PASS-0.08-1,800.00 1.30

PASS-0.09-1,750.00 1.30

PASS-0.08-1,700.00 1.30

PASS-0.05-1,650.00 1.30

PASS-0.01-1,600.00 1.30

PASS 0.00-1,550.00 1.30

PASS 0.01-1,500.00 1.30

PASS 0.00-1,450.00 1.30

PASS-0.02-1,400.00 1.30

PASS-0.04-1,350.00 1.30

PASS-0.06-1,300.00 1.30

PASS-0.06-1,250.00 1.30

Page 19 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 20: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 1.000 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.05-1,200.00 1.30

PASS-0.01-1,150.00 1.30

PASS 0.01-1,100.00 1.30

PASS 0.01-1,050.00 1.30

PASS 0.01-1,000.00 1.30

PASS-0.01-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.08-850.00 1.30

PASS-0.09-800.00 1.30

PASS-0.05-750.00 1.30

PASS-0.02-700.00 1.30

PASS 0.01-650.00 1.30

PASS 0.02-600.00 1.30

PASS 0.02-550.00 1.30

PASS 0.02-500.00 1.30

PASS 0.01-450.00 1.30

PASS-0.02-400.00 1.30

PASS-0.06-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.00-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.01-50.00 1.30

PASS-0.01 50.00 1.30

PASS-0.03 100.00 1.30

PASS-0.04 150.00 1.30

PASS-0.01 200.00 1.30

PASS 0.01 250.00 1.30

PASS 0.02 300.00 1.30

PASS 0.02 350.00 1.30

PASS 0.01 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.03 500.00 1.30

PASS-0.06 550.00 1.30

PASS-0.06 600.00 1.30

PASS-0.03 650.00 1.30

PASS 0.00 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.01 800.00 1.30

PASS 0.00 850.00 1.30

PASS-0.01 900.00 1.30

PASS-0.02 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 20 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 21: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 1.000 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.07 1,050.00 1.30

PASS-0.06 1,100.00 1.30

PASS-0.02 1,150.00 1.30

PASS 0.01 1,200.00 1.30

PASS 0.02 1,250.00 1.30

PASS 0.02 1,300.00 1.30

PASS 0.01 1,350.00 1.30

PASS 0.00 1,400.00 1.30

PASS 0.00 1,450.00 1.30

PASS-0.01 1,500.00 1.30

PASS-0.03 1,550.00 1.30

PASS-0.03 1,600.00 1.30

PASS-0.02 1,650.00 1.30

PASS 0.00 1,700.00 1.30

PASS 0.00 1,750.00 1.30

PASS-0.01 1,800.00 1.30

PASS-0.02 1,850.00 1.30

PASS-0.04 1,900.00 1.30

PASS-0.06 1,950.00 1.30

PASS-0.05 2,000.00 1.30

PASS-0.02 2,050.00 1.30

PASS 0.01 2,100.00 1.30

PASS 0.03 2,150.00 1.30

PASS 0.05 2,200.00 1.30

PASS 0.04 2,250.00 1.30

PASS 0.01 2,300.00 1.30

PASS 0.00 2,350.00 1.30

PASS-0.02 2,400.00 1.30

PASS-0.04 2,450.00 1.30

PASS-0.05 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 21 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 22: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 2.200 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-2,450.00 1.30

PASS-0.07-2,400.00 1.30

PASS-0.10-2,350.00 1.30

PASS-0.14-2,300.00 1.30

PASS-0.13-2,250.00 1.30

PASS-0.12-2,200.00 1.30

PASS-0.08-2,150.00 1.30

PASS-0.04-2,100.00 1.30

PASS-0.01-2,050.00 1.30

PASS 0.00-2,000.00 1.30

PASS 0.00-1,950.00 1.30

PASS-0.02-1,900.00 1.30

PASS-0.04-1,850.00 1.30

PASS-0.07-1,800.00 1.30

PASS-0.08-1,750.00 1.30

PASS-0.08-1,700.00 1.30

PASS-0.05-1,650.00 1.30

PASS-0.02-1,600.00 1.30

PASS 0.01-1,550.00 1.30

PASS 0.01-1,500.00 1.30

PASS 0.00-1,450.00 1.30

PASS-0.02-1,400.00 1.30

PASS-0.04-1,350.00 1.30

PASS-0.05-1,300.00 1.30

PASS-0.06-1,250.00 1.30

Page 22 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 23: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 2.200 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-1,200.00 1.30

PASS-0.01-1,150.00 1.30

PASS 0.01-1,100.00 1.30

PASS 0.01-1,050.00 1.30

PASS 0.01-1,000.00 1.30

PASS-0.01-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.07-850.00 1.30

PASS-0.09-800.00 1.30

PASS-0.05-750.00 1.30

PASS-0.02-700.00 1.30

PASS 0.01-650.00 1.30

PASS 0.02-600.00 1.30

PASS 0.03-550.00 1.30

PASS 0.02-500.00 1.30

PASS 0.01-450.00 1.30

PASS-0.02-400.00 1.30

PASS-0.06-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.00-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.00-50.00 1.30

PASS-0.01 50.00 1.30

PASS-0.03 100.00 1.30

PASS-0.03 150.00 1.30

PASS-0.01 200.00 1.30

PASS 0.00 250.00 1.30

PASS 0.02 300.00 1.30

PASS 0.02 350.00 1.30

PASS 0.01 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.04 500.00 1.30

PASS-0.06 550.00 1.30

PASS-0.06 600.00 1.30

PASS-0.04 650.00 1.30

PASS-0.01 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.00 800.00 1.30

PASS 0.00 850.00 1.30

PASS-0.01 900.00 1.30

PASS-0.02 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 23 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 24: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 2.200 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.07 1,050.00 1.30

PASS-0.06 1,100.00 1.30

PASS-0.02 1,150.00 1.30

PASS 0.00 1,200.00 1.30

PASS 0.02 1,250.00 1.30

PASS 0.01 1,300.00 1.30

PASS 0.01 1,350.00 1.30

PASS 0.00 1,400.00 1.30

PASS-0.01 1,450.00 1.30

PASS-0.02 1,500.00 1.30

PASS-0.03 1,550.00 1.30

PASS-0.03 1,600.00 1.30

PASS-0.02 1,650.00 1.30

PASS-0.01 1,700.00 1.30

PASS 0.00 1,750.00 1.30

PASS-0.01 1,800.00 1.30

PASS-0.03 1,850.00 1.30

PASS-0.04 1,900.00 1.30

PASS-0.06 1,950.00 1.30

PASS-0.05 2,000.00 1.30

PASS-0.02 2,050.00 1.30

PASS 0.01 2,100.00 1.30

PASS 0.02 2,150.00 1.30

PASS 0.04 2,200.00 1.30

PASS 0.03 2,250.00 1.30

PASS 0.00 2,300.00 1.30

PASS-0.01 2,350.00 1.30

PASS-0.03 2,400.00 1.30

PASS-0.05 2,450.00 1.30

PASS-0.06 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 24 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 25: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.000 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-2,450.00 1.30

PASS-0.07-2,400.00 1.30

PASS-0.10-2,350.00 1.30

PASS-0.13-2,300.00 1.30

PASS-0.13-2,250.00 1.30

PASS-0.12-2,200.00 1.30

PASS-0.09-2,150.00 1.30

PASS-0.04-2,100.00 1.30

PASS-0.01-2,050.00 1.30

PASS 0.00-2,000.00 1.30

PASS 0.00-1,950.00 1.30

PASS-0.02-1,900.00 1.30

PASS-0.04-1,850.00 1.30

PASS-0.07-1,800.00 1.30

PASS-0.08-1,750.00 1.30

PASS-0.07-1,700.00 1.30

PASS-0.05-1,650.00 1.30

PASS-0.01-1,600.00 1.30

PASS 0.00-1,550.00 1.30

PASS 0.00-1,500.00 1.30

PASS 0.00-1,450.00 1.30

PASS-0.02-1,400.00 1.30

PASS-0.04-1,350.00 1.30

PASS-0.05-1,300.00 1.30

PASS-0.05-1,250.00 1.30

Page 25 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 26: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.000 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-1,200.00 1.30

PASS-0.01-1,150.00 1.30

PASS 0.01-1,100.00 1.30

PASS 0.01-1,050.00 1.30

PASS 0.01-1,000.00 1.30

PASS-0.01-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.07-850.00 1.30

PASS-0.08-800.00 1.30

PASS-0.05-750.00 1.30

PASS-0.01-700.00 1.30

PASS 0.01-650.00 1.30

PASS 0.03-600.00 1.30

PASS 0.04-550.00 1.30

PASS 0.03-500.00 1.30

PASS 0.02-450.00 1.30

PASS-0.02-400.00 1.30

PASS-0.06-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.00-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.00-50.00 1.30

PASS-0.01 50.00 1.30

PASS-0.03 100.00 1.30

PASS-0.03 150.00 1.30

PASS-0.01 200.00 1.30

PASS 0.01 250.00 1.30

PASS 0.02 300.00 1.30

PASS 0.02 350.00 1.30

PASS 0.01 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.04 500.00 1.30

PASS-0.06 550.00 1.30

PASS-0.06 600.00 1.30

PASS-0.03 650.00 1.30

PASS-0.01 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.00 800.00 1.30

PASS 0.00 850.00 1.30

PASS-0.01 900.00 1.30

PASS-0.03 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 26 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 27: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.000 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.07 1,050.00 1.30

PASS-0.06 1,100.00 1.30

PASS-0.03 1,150.00 1.30

PASS 0.00 1,200.00 1.30

PASS 0.02 1,250.00 1.30

PASS 0.01 1,300.00 1.30

PASS 0.00 1,350.00 1.30

PASS 0.00 1,400.00 1.30

PASS-0.01 1,450.00 1.30

PASS-0.02 1,500.00 1.30

PASS-0.03 1,550.00 1.30

PASS-0.03 1,600.00 1.30

PASS-0.02 1,650.00 1.30

PASS-0.01 1,700.00 1.30

PASS-0.01 1,750.00 1.30

PASS-0.02 1,800.00 1.30

PASS-0.03 1,850.00 1.30

PASS-0.04 1,900.00 1.30

PASS-0.06 1,950.00 1.30

PASS-0.05 2,000.00 1.30

PASS-0.02 2,050.00 1.30

PASS 0.01 2,100.00 1.30

PASS 0.02 2,150.00 1.30

PASS 0.03 2,200.00 1.30

PASS 0.02 2,250.00 1.30

PASS 0.00 2,300.00 1.30

PASS-0.01 2,350.00 1.30

PASS-0.03 2,400.00 1.30

PASS-0.05 2,450.00 1.30

PASS-0.06 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 27 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 28: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.400 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-2,450.00 1.30

PASS-0.07-2,400.00 1.30

PASS-0.11-2,350.00 1.30

PASS-0.14-2,300.00 1.30

PASS-0.14-2,250.00 1.30

PASS-0.13-2,200.00 1.30

PASS-0.09-2,150.00 1.30

PASS-0.04-2,100.00 1.30

PASS-0.01-2,050.00 1.30

PASS 0.00-2,000.00 1.30

PASS-0.01-1,950.00 1.30

PASS-0.02-1,900.00 1.30

PASS-0.05-1,850.00 1.30

PASS-0.08-1,800.00 1.30

PASS-0.09-1,750.00 1.30

PASS-0.08-1,700.00 1.30

PASS-0.05-1,650.00 1.30

PASS-0.02-1,600.00 1.30

PASS 0.00-1,550.00 1.30

PASS 0.00-1,500.00 1.30

PASS 0.00-1,450.00 1.30

PASS-0.02-1,400.00 1.30

PASS-0.04-1,350.00 1.30

PASS-0.06-1,300.00 1.30

PASS-0.06-1,250.00 1.30

Page 28 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 29: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.400 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.05-1,200.00 1.30

PASS-0.01-1,150.00 1.30

PASS 0.01-1,100.00 1.30

PASS 0.03-1,050.00 1.30

PASS 0.02-1,000.00 1.30

PASS 0.00-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.08-850.00 1.30

PASS-0.09-800.00 1.30

PASS-0.06-750.00 1.30

PASS-0.02-700.00 1.30

PASS 0.01-650.00 1.30

PASS 0.02-600.00 1.30

PASS 0.03-550.00 1.30

PASS 0.02-500.00 1.30

PASS 0.00-450.00 1.30

PASS-0.03-400.00 1.30

PASS-0.06-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.00-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.01-50.00 1.30

PASS-0.02 50.00 1.30

PASS-0.03 100.00 1.30

PASS-0.04 150.00 1.30

PASS-0.02 200.00 1.30

PASS 0.01 250.00 1.30

PASS 0.02 300.00 1.30

PASS 0.02 350.00 1.30

PASS 0.01 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.04 500.00 1.30

PASS-0.06 550.00 1.30

PASS-0.07 600.00 1.30

PASS-0.04 650.00 1.30

PASS 0.00 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.01 800.00 1.30

PASS 0.00 850.00 1.30

PASS-0.01 900.00 1.30

PASS-0.02 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 29 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 30: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.400 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.07 1,050.00 1.30

PASS-0.06 1,100.00 1.30

PASS-0.02 1,150.00 1.30

PASS 0.00 1,200.00 1.30

PASS 0.02 1,250.00 1.30

PASS 0.02 1,300.00 1.30

PASS 0.01 1,350.00 1.30

PASS 0.00 1,400.00 1.30

PASS-0.01 1,450.00 1.30

PASS-0.02 1,500.00 1.30

PASS-0.03 1,550.00 1.30

PASS-0.03 1,600.00 1.30

PASS-0.02 1,650.00 1.30

PASS 0.00 1,700.00 1.30

PASS 0.00 1,750.00 1.30

PASS-0.01 1,800.00 1.30

PASS-0.02 1,850.00 1.30

PASS-0.04 1,900.00 1.30

PASS-0.07 1,950.00 1.30

PASS-0.05 2,000.00 1.30

PASS-0.02 2,050.00 1.30

PASS 0.01 2,100.00 1.30

PASS 0.03 2,150.00 1.30

PASS 0.04 2,200.00 1.30

PASS 0.03 2,250.00 1.30

PASS 0.01 2,300.00 1.30

PASS-0.01 2,350.00 1.30

PASS-0.03 2,400.00 1.30

PASS-0.05 2,450.00 1.30

PASS-0.06 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 30 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 31: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.800 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.02-2,450.00 1.30

PASS-0.05-2,400.00 1.30

PASS-0.09-2,350.00 1.30

PASS-0.12-2,300.00 1.30

PASS-0.12-2,250.00 1.30

PASS-0.10-2,200.00 1.30

PASS-0.07-2,150.00 1.30

PASS-0.02-2,100.00 1.30

PASS 0.01-2,050.00 1.30

PASS 0.02-2,000.00 1.30

PASS 0.02-1,950.00 1.30

PASS 0.00-1,900.00 1.30

PASS-0.03-1,850.00 1.30

PASS-0.06-1,800.00 1.30

PASS-0.07-1,750.00 1.30

PASS-0.06-1,700.00 1.30

PASS-0.04-1,650.00 1.30

PASS 0.00-1,600.00 1.30

PASS 0.02-1,550.00 1.30

PASS 0.02-1,500.00 1.30

PASS 0.01-1,450.00 1.30

PASS-0.01-1,400.00 1.30

PASS-0.03-1,350.00 1.30

PASS-0.04-1,300.00 1.30

PASS-0.05-1,250.00 1.30

Page 31 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 32: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.800 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.03-1,200.00 1.30

PASS 0.00-1,150.00 1.30

PASS 0.02-1,100.00 1.30

PASS 0.02-1,050.00 1.30

PASS 0.02-1,000.00 1.30

PASS 0.00-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.07-850.00 1.30

PASS-0.08-800.00 1.30

PASS-0.05-750.00 1.30

PASS-0.01-700.00 1.30

PASS 0.02-650.00 1.30

PASS 0.03-600.00 1.30

PASS 0.03-550.00 1.30

PASS 0.03-500.00 1.30

PASS 0.01-450.00 1.30

PASS-0.02-400.00 1.30

PASS-0.06-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.01-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.01-50.00 1.30

PASS-0.02 50.00 1.30

PASS-0.03 100.00 1.30

PASS-0.04 150.00 1.30

PASS-0.01 200.00 1.30

PASS 0.00 250.00 1.30

PASS 0.02 300.00 1.30

PASS 0.01 350.00 1.30

PASS 0.01 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.04 500.00 1.30

PASS-0.07 550.00 1.30

PASS-0.07 600.00 1.30

PASS-0.04 650.00 1.30

PASS-0.01 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.00 800.00 1.30

PASS-0.01 850.00 1.30

PASS-0.02 900.00 1.30

PASS-0.03 950.00 1.30

PASS-0.06 1,000.00 1.30

Page 32 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 33: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 3.800 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.08 1,050.00 1.30

PASS-0.07 1,100.00 1.30

PASS-0.04 1,150.00 1.30

PASS-0.01 1,200.00 1.30

PASS 0.01 1,250.00 1.30

PASS 0.00 1,300.00 1.30

PASS-0.01 1,350.00 1.30

PASS-0.01 1,400.00 1.30

PASS-0.02 1,450.00 1.30

PASS-0.03 1,500.00 1.30

PASS-0.05 1,550.00 1.30

PASS-0.05 1,600.00 1.30

PASS-0.04 1,650.00 1.30

PASS-0.02 1,700.00 1.30

PASS-0.01 1,750.00 1.30

PASS-0.03 1,800.00 1.30

PASS-0.04 1,850.00 1.30

PASS-0.06 1,900.00 1.30

PASS-0.08 1,950.00 1.30

PASS-0.07 2,000.00 1.30

PASS-0.04 2,050.00 1.30

PASS-0.01 2,100.00 1.30

PASS 0.01 2,150.00 1.30

PASS 0.02 2,200.00 1.30

PASS 0.01 2,250.00 1.30

PASS-0.01 2,300.00 1.30

PASS-0.03 2,350.00 1.30

PASS-0.05 2,400.00 1.30

PASS-0.07 2,450.00 1.30

PASS-0.07 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 33 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 34: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 4.280 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-2,450.00 1.30

PASS-0.06-2,400.00 1.30

PASS-0.10-2,350.00 1.30

PASS-0.13-2,300.00 1.30

PASS-0.13-2,250.00 1.30

PASS-0.12-2,200.00 1.30

PASS-0.09-2,150.00 1.30

PASS-0.03-2,100.00 1.30

PASS 0.00-2,050.00 1.30

PASS 0.01-2,000.00 1.30

PASS 0.00-1,950.00 1.30

PASS-0.02-1,900.00 1.30

PASS-0.04-1,850.00 1.30

PASS-0.08-1,800.00 1.30

PASS-0.08-1,750.00 1.30

PASS-0.08-1,700.00 1.30

PASS-0.05-1,650.00 1.30

PASS-0.01-1,600.00 1.30

PASS 0.01-1,550.00 1.30

PASS 0.01-1,500.00 1.30

PASS 0.00-1,450.00 1.30

PASS-0.02-1,400.00 1.30

PASS-0.04-1,350.00 1.30

PASS-0.05-1,300.00 1.30

PASS-0.06-1,250.00 1.30

Page 34 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 35: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 4.280 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.04-1,200.00 1.30

PASS-0.01-1,150.00 1.30

PASS 0.01-1,100.00 1.30

PASS 0.01-1,050.00 1.30

PASS 0.01-1,000.00 1.30

PASS-0.01-950.00 1.30

PASS-0.04-900.00 1.30

PASS-0.07-850.00 1.30

PASS-0.09-800.00 1.30

PASS-0.05-750.00 1.30

PASS-0.02-700.00 1.30

PASS 0.01-650.00 1.30

PASS 0.02-600.00 1.30

PASS 0.03-550.00 1.30

PASS 0.02-500.00 1.30

PASS 0.01-450.00 1.30

PASS-0.02-400.00 1.30

PASS-0.06-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.01-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.01-50.00 1.30

PASS-0.01 50.00 1.30

PASS-0.03 100.00 1.30

PASS-0.04 150.00 1.30

PASS-0.01 200.00 1.30

PASS 0.00 250.00 1.30

PASS 0.02 300.00 1.30

PASS 0.01 350.00 1.30

PASS 0.01 400.00 1.30

PASS-0.01 450.00 1.30

PASS-0.04 500.00 1.30

PASS-0.06 550.00 1.30

PASS-0.06 600.00 1.30

PASS-0.04 650.00 1.30

PASS-0.01 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.01 800.00 1.30

PASS 0.00 850.00 1.30

PASS-0.01 900.00 1.30

PASS-0.03 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 35 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 36: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 4.280 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.07 1,050.00 1.30

PASS-0.07 1,100.00 1.30

PASS-0.02 1,150.00 1.30

PASS 0.00 1,200.00 1.30

PASS 0.01 1,250.00 1.30

PASS 0.01 1,300.00 1.30

PASS 0.00 1,350.00 1.30

PASS-0.01 1,400.00 1.30

PASS-0.01 1,450.00 1.30

PASS-0.02 1,500.00 1.30

PASS-0.04 1,550.00 1.30

PASS-0.04 1,600.00 1.30

PASS-0.02 1,650.00 1.30

PASS-0.01 1,700.00 1.30

PASS 0.00 1,750.00 1.30

PASS-0.02 1,800.00 1.30

PASS-0.02 1,850.00 1.30

PASS-0.04 1,900.00 1.30

PASS-0.07 1,950.00 1.30

PASS-0.06 2,000.00 1.30

PASS-0.03 2,050.00 1.30

PASS 0.00 2,100.00 1.30

PASS 0.02 2,150.00 1.30

PASS 0.03 2,200.00 1.30

PASS 0.02 2,250.00 1.30

PASS 0.00 2,300.00 1.30

PASS-0.02 2,350.00 1.30

PASS-0.03 2,400.00 1.30

PASS-0.05 2,450.00 1.30

PASS-0.07 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 36 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 37: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 4.700 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.06-2,450.00 1.30

PASS-0.08-2,400.00 1.30

PASS-0.11-2,350.00 1.30

PASS-0.14-2,300.00 1.30

PASS-0.14-2,250.00 1.30

PASS-0.13-2,200.00 1.30

PASS-0.10-2,150.00 1.30

PASS-0.04-2,100.00 1.30

PASS-0.02-2,050.00 1.30

PASS-0.01-2,000.00 1.30

PASS-0.01-1,950.00 1.30

PASS-0.03-1,900.00 1.30

PASS-0.05-1,850.00 1.30

PASS-0.09-1,800.00 1.30

PASS-0.09-1,750.00 1.30

PASS-0.08-1,700.00 1.30

PASS-0.06-1,650.00 1.30

PASS-0.02-1,600.00 1.30

PASS 0.00-1,550.00 1.30

PASS 0.00-1,500.00 1.30

PASS-0.01-1,450.00 1.30

PASS-0.03-1,400.00 1.30

PASS-0.05-1,350.00 1.30

PASS-0.06-1,300.00 1.30

PASS-0.06-1,250.00 1.30

Page 37 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 38: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 4.700 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.05-1,200.00 1.30

PASS-0.01-1,150.00 1.30

PASS 0.00-1,100.00 1.30

PASS 0.01-1,050.00 1.30

PASS 0.00-1,000.00 1.30

PASS-0.01-950.00 1.30

PASS-0.05-900.00 1.30

PASS-0.08-850.00 1.30

PASS-0.09-800.00 1.30

PASS-0.06-750.00 1.30

PASS-0.02-700.00 1.30

PASS 0.01-650.00 1.30

PASS 0.02-600.00 1.30

PASS 0.02-550.00 1.30

PASS 0.02-500.00 1.30

PASS 0.00-450.00 1.30

PASS-0.03-400.00 1.30

PASS-0.06-350.00 1.30

PASS-0.07-300.00 1.30

PASS-0.03-250.00 1.30

PASS 0.00-200.00 1.30

PASS 0.01-150.00 1.30

PASS 0.01-100.00 1.30

PASS 0.01-50.00 1.30

PASS-0.01 50.00 1.30

PASS-0.03 100.00 1.30

PASS-0.04 150.00 1.30

PASS-0.02 200.00 1.30

PASS 0.00 250.00 1.30

PASS 0.02 300.00 1.30

PASS 0.02 350.00 1.30

PASS 0.01 400.00 1.30

PASS 0.00 450.00 1.30

PASS-0.03 500.00 1.30

PASS-0.06 550.00 1.30

PASS-0.06 600.00 1.30

PASS-0.04 650.00 1.30

PASS 0.00 700.00 1.30

PASS 0.01 750.00 1.30

PASS 0.01 800.00 1.30

PASS 0.00 850.00 1.30

PASS-0.01 900.00 1.30

PASS-0.02 950.00 1.30

PASS-0.05 1,000.00 1.30

Page 38 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 39: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 4.700 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.07 1,050.00 1.30

PASS-0.06 1,100.00 1.30

PASS-0.02 1,150.00 1.30

PASS 0.00 1,200.00 1.30

PASS 0.02 1,250.00 1.30

PASS 0.01 1,300.00 1.30

PASS 0.01 1,350.00 1.30

PASS 0.00 1,400.00 1.30

PASS-0.01 1,450.00 1.30

PASS-0.02 1,500.00 1.30

PASS-0.04 1,550.00 1.30

PASS-0.03 1,600.00 1.30

PASS-0.02 1,650.00 1.30

PASS-0.01 1,700.00 1.30

PASS 0.00 1,750.00 1.30

PASS-0.02 1,800.00 1.30

PASS-0.03 1,850.00 1.30

PASS-0.04 1,900.00 1.30

PASS-0.07 1,950.00 1.30

PASS-0.06 2,000.00 1.30

PASS-0.03 2,050.00 1.30

PASS 0.00 2,100.00 1.30

PASS 0.02 2,150.00 1.30

PASS 0.03 2,200.00 1.30

PASS 0.02 2,250.00 1.30

PASS 0.00 2,300.00 1.30

PASS-0.02 2,350.00 1.30

PASS-0.04 2,400.00 1.30

PASS-0.06 2,450.00 1.30

PASS-0.07 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 39 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 40: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 5.500 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.09-2,450.00 1.30

PASS-0.07-2,400.00 1.30

PASS-0.05-2,350.00 1.30

PASS-0.04-2,300.00 1.30

PASS-0.02-2,250.00 1.30

PASS 0.00-2,200.00 1.30

PASS-0.02-2,150.00 1.30

PASS-0.03-2,100.00 1.30

PASS-0.06-2,050.00 1.30

PASS-0.09-2,000.00 1.30

PASS-0.10-1,950.00 1.30

PASS-0.08-1,900.00 1.30

PASS-0.06-1,850.00 1.30

PASS-0.05-1,800.00 1.30

PASS-0.04-1,750.00 1.30

PASS-0.04-1,700.00 1.30

PASS-0.05-1,650.00 1.30

PASS-0.07-1,600.00 1.30

PASS-0.06-1,550.00 1.30

PASS-0.05-1,500.00 1.30

PASS-0.04-1,450.00 1.30

PASS-0.03-1,400.00 1.30

PASS-0.02-1,350.00 1.30

PASS-0.01-1,300.00 1.30

PASS 0.00-1,250.00 1.30

Page 40 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 41: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 5.500 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.02-1,200.00 1.30

PASS-0.04-1,150.00 1.30

PASS-0.08-1,100.00 1.30

PASS-0.09-1,050.00 1.30

PASS-0.08-1,000.00 1.30

PASS-0.04-950.00 1.30

PASS-0.03-900.00 1.30

PASS-0.02-850.00 1.30

PASS-0.01-800.00 1.30

PASS 0.00-750.00 1.30

PASS-0.02-700.00 1.30

PASS-0.05-650.00 1.30

PASS-0.07-600.00 1.30

PASS-0.07-550.00 1.30

PASS-0.05-500.00 1.30

PASS-0.02-450.00 1.30

PASS 0.00-400.00 1.30

PASS 0.01-350.00 1.30

PASS 0.01-300.00 1.30

PASS 0.00-250.00 1.30

PASS-0.02-200.00 1.30

PASS-0.04-150.00 1.30

PASS-0.04-100.00 1.30

PASS-0.02-50.00 1.30

PASS 0.01 50.00 1.30

PASS 0.01 100.00 1.30

PASS 0.01 150.00 1.30

PASS 0.00 200.00 1.30

PASS-0.03 250.00 1.30

PASS-0.07 300.00 1.30

PASS-0.06 350.00 1.30

PASS-0.02 400.00 1.30

PASS 0.00 450.00 1.30

PASS 0.02 500.00 1.30

PASS 0.03 550.00 1.30

PASS 0.03 600.00 1.30

PASS 0.02 650.00 1.30

PASS-0.01 700.00 1.30

PASS-0.05 750.00 1.30

PASS-0.08 800.00 1.30

PASS-0.07 850.00 1.30

PASS-0.03 900.00 1.30

PASS 0.00 950.00 1.30

PASS 0.03 1,000.00 1.30

Page 41 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 42: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 5.500 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS 0.03 1,050.00 1.30

PASS 0.02 1,100.00 1.30

PASS 0.01 1,150.00 1.30

PASS-0.03 1,200.00 1.30

PASS-0.03 1,250.00 1.30

PASS-0.04 1,300.00 1.30

PASS-0.03 1,350.00 1.30

PASS 0.00 1,400.00 1.30

PASS 0.01 1,450.00 1.30

PASS 0.02 1,500.00 1.30

PASS 0.02 1,550.00 1.30

PASS 0.00 1,600.00 1.30

PASS-0.03 1,650.00 1.30

PASS-0.06 1,700.00 1.30

PASS-0.06 1,750.00 1.30

PASS-0.05 1,800.00 1.30

PASS-0.02 1,850.00 1.30

PASS 0.00 1,900.00 1.30

PASS 0.02 1,950.00 1.30

PASS 0.02 2,000.00 1.30

PASS 0.02 2,050.00 1.30

PASS-0.01 2,100.00 1.30

PASS-0.07 2,150.00 1.30

PASS-0.10 2,200.00 1.30

PASS-0.10 2,250.00 1.30

PASS-0.11 2,300.00 1.30

PASS-0.08 2,350.00 1.30

PASS-0.06 2,400.00 1.30

PASS-0.03 2,450.00 1.30

PASS-0.02 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 42 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 43: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 5.970 GHz

-1.60

-1.20

-0.80

-0.40

0.00

0.40

0.80

1.20

1.60

-2,4

50.0

0

-1,9

50.0

0

-1,4

50.0

0

-950

.00

-450

.00

100.0

0

600.

00

1,10

0.00

1,60

0.00

2,10

0.00

IF F

req

uen

cy R

esp

on

se (

dB

)

Frequency Offset (kHz)

IF Frequency

Response

Test Limit*

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS-0.05-2,450.00 1.30

PASS-0.03-2,400.00 1.30

PASS-0.02-2,350.00 1.30

PASS 0.00-2,300.00 1.30

PASS 0.02-2,250.00 1.30

PASS 0.04-2,200.00 1.30

PASS 0.03-2,150.00 1.30

PASS 0.01-2,100.00 1.30

PASS-0.02-2,050.00 1.30

PASS-0.05-2,000.00 1.30

PASS-0.06-1,950.00 1.30

PASS-0.04-1,900.00 1.30

PASS-0.02-1,850.00 1.30

PASS-0.01-1,800.00 1.30

PASS 0.01-1,750.00 1.30

PASS 0.00-1,700.00 1.30

PASS-0.02-1,650.00 1.30

PASS-0.03-1,600.00 1.30

PASS-0.03-1,550.00 1.30

PASS-0.02-1,500.00 1.30

PASS-0.01-1,450.00 1.30

PASS 0.01-1,400.00 1.30

PASS 0.01-1,350.00 1.30

PASS 0.01-1,300.00 1.30

PASS 0.02-1,250.00 1.30

Page 43 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 44: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 5.970 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS 0.00-1,200.00 1.30

PASS-0.02-1,150.00 1.30

PASS-0.06-1,100.00 1.30

PASS-0.07-1,050.00 1.30

PASS-0.05-1,000.00 1.30

PASS-0.02-950.00 1.30

PASS-0.01-900.00 1.30

PASS 0.01-850.00 1.30

PASS 0.02-800.00 1.30

PASS 0.01-750.00 1.30

PASS 0.00-700.00 1.30

PASS-0.04-650.00 1.30

PASS-0.05-600.00 1.30

PASS-0.05-550.00 1.30

PASS-0.03-500.00 1.30

PASS 0.00-450.00 1.30

PASS 0.02-400.00 1.30

PASS 0.03-350.00 1.30

PASS 0.02-300.00 1.30

PASS 0.01-250.00 1.30

PASS-0.01-200.00 1.30

PASS-0.03-150.00 1.30

PASS-0.03-100.00 1.30

PASS-0.01-50.00 1.30

PASS 0.01 50.00 1.30

PASS 0.02 100.00 1.30

PASS 0.02 150.00 1.30

PASS 0.01 200.00 1.30

PASS-0.03 250.00 1.30

PASS-0.06 300.00 1.30

PASS-0.05 350.00 1.30

PASS-0.02 400.00 1.30

PASS 0.01 450.00 1.30

PASS 0.02 500.00 1.30

PASS 0.04 550.00 1.30

PASS 0.03 600.00 1.30

PASS 0.02 650.00 1.30

PASS-0.01 700.00 1.30

PASS-0.04 750.00 1.30

PASS-0.07 800.00 1.30

PASS-0.07 850.00 1.30

PASS-0.04 900.00 1.30

PASS 0.00 950.00 1.30

PASS 0.02 1,000.00 1.30

Page 44 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 45: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

IF Frequency Response

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Source Frequency = 5.970 GHz

Result

Test

Limit*

(dB)

IF Frequency

Response

(dB)

Frequency

Offset

(kHz)

PASS 0.02 1,050.00 1.30

PASS 0.02 1,100.00 1.30

PASS-0.01 1,150.00 1.30

PASS-0.03 1,200.00 1.30

PASS-0.04 1,250.00 1.30

PASS-0.04 1,300.00 1.30

PASS-0.04 1,350.00 1.30

PASS-0.01 1,400.00 1.30

PASS 0.02 1,450.00 1.30

PASS 0.02 1,500.00 1.30

PASS 0.01 1,550.00 1.30

PASS 0.00 1,600.00 1.30

PASS-0.04 1,650.00 1.30

PASS-0.06 1,700.00 1.30

PASS-0.07 1,750.00 1.30

PASS-0.06 1,800.00 1.30

PASS-0.02 1,850.00 1.30

PASS 0.00 1,900.00 1.30

PASS 0.01 1,950.00 1.30

PASS 0.02 2,000.00 1.30

PASS 0.01 2,050.00 1.30

PASS-0.03 2,100.00 1.30

PASS-0.07 2,150.00 1.30

PASS-0.10 2,200.00 1.30

PASS-0.11 2,250.00 1.30

PASS-0.11 2,300.00 1.30

PASS-0.09 2,350.00 1.30

PASS-0.05 2,400.00 1.30

PASS-0.02 2,450.00 1.30

PASS-0.02 2,500.00 1.30

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 45 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 46: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Noise Sidebands

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

E8257D Signal Generator US44270223 01-Jan-2012

CW Frequency = 1.00 GHz

Phase Noise

Offset

(kHz)

Measured

Phase Noise

(dBc/Hz)

Measurement

Uncertainty

(dB)

Specification

(dBc/Hz)

Result

PASS 10.00 -91.17 1.78 -85.00

The reported value for phase noise represents both the phase noise of the instrument plus the phase noise of the signal

source used for this measurement. For reported values near the specification limit, the phase noise contribution from the

signal source is negligible. However, for reported values significantly better than the specification limit, the contribution

from the signal source may be significant. For these cases, it can be assumed the instrument's phase noise is better than

the reported value.

Page 46 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 47: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Absolute Amplitude Accuracy

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

11500C N(m) to N(m) RF Cable 1

11500C N(m) to N(m) RF Cable 1

11667A Power Splitter 1

11713A Attenuator Switch Driver 3748A04115

11716A Interconnect cable 1

1250-1475 N(m) to N(m) Adapter 1

8120-2703 Attenuator Cable 1

8120-2703 Attenuator Cable 2

8491A Opt 006 6dB Fixed Attenuator 1 01-Jan-2012

8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011

8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011

E4419A Power Meter US38470132 01-Jan-2012

E8257D Signal Generator US44270223 01-Jan-2012

E9304A Power Sensor MY41498147 440530 25-Feb-2011

E9304A H18 Power Sensor MY41496197 431135 25-Oct-2011

Frequency = 50 MHz

Preamp OFF

Resolution

Bandwidth

(kHz)

Amplitude

Accuracy

(dB)

Measurement

Uncertainty

(+/- dB)

Specification

(+/- dB)

Result

Measured

Amplitude

(dBm)

Span

(kHz)

Nominal

Input Level

(dBm)

PASS-0.089 0.142-5.146 0.80 20.00 1.00-5.00

PASS-0.089 0.159-10.203 0.80 20.00 1.00-10.00

PASS-0.045 0.155-15.016 0.80 20.00 1.00-15.00

PASS-0.073 0.155-20.101 0.80 20.00 1.00-20.00

PASS0.063 0.155-24.611 0.80 20.00 1.00-25.00

PASS0.108 0.155-29.623 0.80 20.00 1.00-30.00

PASS0.101 0.155-34.487 0.80 20.00 1.00-35.00

PASS0.092 0.155-39.552 0.80 20.00 1.00-40.00

PASS-0.084 0.156-45.413 0.80 20.00 1.00-45.00

PASS-0.085 0.156-50.471 0.80 20.00 1.00-50.00

Frequency = 50 MHz

Preamp ON

Resolution

Bandwidth

(kHz)

Amplitude

Accuracy

(dB)

Test

Limit*

(+/- dB)

Result

Measured

Amplitude

(dBm)

Span

(kHz)

Nominal

Input Level

(dBm)

-0.638-35.225 1.10 20.00 1.00-35.00 PASS

-0.736-51.122 1.10 20.00 1.00-50.00 PASS

*Note: Absolute Amplitude Accuracy with the preamplifier ON is not a warranted specification. The test limits may be larger than the

published typical values. The limits have been set to ensure that the hardware is functional and is not defective.

Page 47 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 48: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Frequency Response Preamp Off

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

11667A Power Splitter 1

11713A Attenuator Switch Driver 3748A04115

11716A Interconnect cable 1

8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011

8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011

E4419A Power Meter US38470132 01-Jan-2012

E8257D Signal Generator US44270223 01-Jan-2012

E9304A Power Sensor MY41498147 440530 25-Feb-2011

Page 48 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 49: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Frequency Response Preamp Off

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

-1.6

-1.2

-0.8

-0.4

0.0

0.4

0.8

1.2

1.6

2.00

600.

00

1,20

0.00

1,80

0.00

2,40

0.00

3,00

0.00

3,60

0.00

4,20

0.00

4,80

0.00

5,40

0.00

6,00

0.00

Frequency (MHz)

Fre

qu

en

cy R

esp

on

se (

dB

)

Frequency

Response

Measurement

Uncertainty

Specification

Frequency

(MHz)

Frequency

Response

(dB)

Measurement

Uncertainty

(+/- dB)

Specification

(+/- dB)

Result

2.00 -0.280 0.206 PASS 1.10

100.00 0.095 0.206 PASS 0.90

200.00 0.036 0.206 PASS 0.90

300.00 0.038 0.206 PASS 0.90

400.00 0.081 0.206 PASS 0.90

500.00 0.058 0.206 PASS 0.90

600.00 0.079 0.206 PASS 0.90

700.00 0.099 0.206 PASS 0.90

800.00 0.091 0.206 PASS 0.90

900.00 0.104 0.206 PASS 0.90

1,000.00 0.118 0.206 PASS 0.90

1,100.00 0.119 0.206 PASS 0.90

1,200.00 0.072 0.206 PASS 0.90

1,300.00 0.069 0.206 PASS 0.90

1,400.00 0.116 0.206 PASS 0.90

1,500.00 0.114 0.206 PASS 0.90

1,600.00 0.118 0.206 PASS 0.90

1,700.00 0.140 0.206 PASS 0.90

1,800.00 0.073 0.206 PASS 0.90

1,900.00 0.078 0.206 PASS 0.90

2,000.00 0.115 0.206 PASS 0.90

2,100.00 0.059 0.206 PASS 0.90

2,200.00 0.096 0.206 PASS 0.90

2,300.00 0.080 0.206 PASS 0.90

2,400.00 0.094 0.206 PASS 0.90

2,500.00 0.039 0.206 PASS 0.90

2,600.00 0.088 0.206 PASS 0.90

Page 49 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 50: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Frequency Response Preamp Off

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Frequency

(MHz)

Frequency

Response

(dB)

Measurement

Uncertainty

(+/- dB)

Specification

(+/- dB)

Result

2,700.00 0.086 0.206 PASS 0.90

2,800.00 0.105 0.206 PASS 0.90

2,900.00 0.071 0.206 PASS 0.90

3,000.00 0.064 0.206 PASS 0.90

3,100.00 0.107 0.206 PASS 1.30

3,200.00 0.072 0.207 PASS 1.30

3,300.00 -0.094 0.207 PASS 1.30

3,400.00 0.071 0.207 PASS 1.30

3,500.00 0.015 0.207 PASS 1.30

3,600.00 -0.032 0.207 PASS 1.30

3,700.00 0.025 0.207 PASS 1.30

3,800.00 0.068 0.207 PASS 1.30

3,900.00 0.011 0.207 PASS 1.30

4,000.00 0.115 0.207 PASS 1.30

4,100.00 -0.061 0.229 PASS 1.30

4,200.00 -0.015 0.229 PASS 1.30

4,300.00 0.018 0.229 PASS 1.30

4,400.00 0.073 0.229 PASS 1.30

4,500.00 0.060 0.229 PASS 1.30

4,600.00 0.021 0.229 PASS 1.30

4,700.00 0.016 0.229 PASS 1.30

4,800.00 0.017 0.229 PASS 1.30

4,900.00 -0.027 0.229 PASS 1.30

5,000.00 -0.094 0.229 PASS 1.30

5,100.00 0.023 0.229 PASS 1.50

5,200.00 0.010 0.229 PASS 1.50

5,300.00 0.014 0.229 PASS 1.50

5,400.00 0.017 0.229 PASS 1.50

5,500.00 -0.033 0.229 PASS 1.50

5,600.00 -0.011 0.229 PASS 1.50

5,700.00 0.013 0.229 PASS 1.50

5,800.00 -0.029 0.229 PASS 1.50

5,900.00 0.046 0.229 PASS 1.50

6,000.00 0.085 0.229 PASS 1.50

Page 50 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 51: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Frequency Response Preamp On

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

Temperature 25.00 Celsius

Environmental Conditions

Line Frequency 60.00 HzHumidity 50.00 %

Test Standards and Required Equipment

Cal DueTrace NumberSerial/Unique IDDescriptionModel

11667A Power Splitter 1

11713A Attenuator Switch Driver 3748A04115

11716A Interconnect cable 1

8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011

8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011

E4419A Power Meter US38470132 01-Jan-2012

E8257D Signal Generator US44270223 01-Jan-2012

E9304A Power Sensor MY41498147 440530 25-Feb-2011

Page 51 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037

Page 52: Measurement Report (ISO 17025) Agilent N9912A …na.support.keysight.com/fieldfox/download_files/n9912a_sample_report_e20.pdfMeasurement Report (ISO 17025) Agilent N9912A Performance

Frequency Response Preamp On

Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS

Report Number

smoke test on build 16-873S

-2.0

-1.5

-1.0

-0.5

0.0

0.5

1.0

1.5

2.0

2.00

1,60

0.00

3,40

0.00

5,20

0.00

Frequency (MHz)

Fre

qu

en

cy R

esp

on

se (

dB

)

Frequency

Response

Test Limit*

Result

Test

Limit*

(+/- dB)

Frequency

Response

(dB)

Frequency

(MHz)

2.00 0.048 PASS 1.70

100.00 0.007 PASS 1.70

400.00 0.071 PASS 1.70

700.00 0.071 PASS 1.70

1,000.00 0.106 PASS 1.70

1,300.00 0.089 PASS 1.70

1,600.00 0.077 PASS 1.70

1,900.00 0.145 PASS 1.70

2,200.00 0.118 PASS 1.70

2,500.00 0.092 PASS 1.70

2,800.00 0.129 PASS 1.70

3,100.00 0.098 PASS 1.70

3,400.00 -0.034 PASS 1.70

3,700.00 -0.027 PASS 1.70

4,000.00 0.051 PASS 1.70

4,300.00 -0.081 PASS 1.70

4,600.00 -0.026 PASS 1.70

4,900.00 0.003 PASS 1.70

5,200.00 -0.007 PASS 1.70

5,500.00 -0.047 PASS 1.70

5,800.00 -0.045 PASS 1.70

*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set

to ensure that the hardware is functional and is not defective.

Page 52 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:

STP-N9912A-US47500037