11073 in CDA Representing anaesthetic vital sign data in a CDA document, using the 11073 standard. Martin Hurrell Alan Nicol. Acknowledgements. In no particular order thanks to: Terri Monk (IOTA & HL7 Anesthesiology WG)) John Rhoads and Jan Wittenber (HL7 Healthcare Devices WG). - PowerPoint PPT Presentation

Citation preview

Page 1: Martin Hurrell            Alan Nicol

11073 in CDA

Representing anaesthetic vital sign data in a CDA document, using the 11073 standard

Martin Hurrell Alan Nicol

Page 2: Martin Hurrell            Alan Nicol

Acknowledgements

In no particular order thanks to:

Terri Monk (IOTA & HL7 Anesthesiology WG))

John Rhoads and Jan Wittenber (HL7 Healthcare Devices WG)

Page 3: Martin Hurrell            Alan Nicol

Vital Signs Monitoring Setup

Connect the various probes and measurement devices to a patient so that their physiological values can be measured and recorded during an intra-operative procedure.

Scenario

A blood pressure cuff is placed on the left upper arm of the patient and connected to the patient monitors non-Invasive blood pressure module. The non-invasive blood pressure module uses a sphygmomanometric technique to measure systolic and diastolic blood pressure. Blood pressure values are typically measured in millimetres of Mercury (mmHg).

Actors

Patient: A person who is the recipient of healthcare services.Patient Monitor: an electronic medical device that measures and displays a patient's vital signs

Anaesthetic vital signs use case (Setup)

Page 4: Martin Hurrell            Alan Nicol

Vital Signs Monitoring Recording

The context is the measurement and storage of physiological values from a patient during a surgical procedure under anaesthesia. In the following example the values recorded are the patient’s non-invasive blood pressures.

Scenario

Monitoring starts and the first set of values of the Systolic, Mean and Diastolic blood pressures are collected after an initial interval of 5 minutes and at 5 minute intervals thereafter.

Actors

Patient: A person who is the recipient of healthcare services.Patient Monitor: an electronic medical device that measures and displays a patient's vital signs

Anaesthetic vital signs use case (Recording)

Page 5: Martin Hurrell            Alan Nicol

11073 Dim

Term ConceptBlood Pressure Measurement Metric

Module MDSNon-Invasive Blood Pressure ChannelPatient monitor MDS

Page 6: Martin Hurrell            Alan Nicol

Patient monitor/Blood

Pressure Module

11073 Dim

Term ConceptBlood Pressure Measurement Metric

Module MDSNon-Invasive Blood Pressure ChannelPatient monitor MDS

Page 7: Martin Hurrell            Alan Nicol

11073 Dim

Patient monitor/Blood

Pressure Module

Non-Invasive Blood Pressure

Term ConceptBlood Pressure Measurement Metric

Module MDSNon-Invasive Blood Pressure ChannelPatient monitor MDS

Page 8: Martin Hurrell            Alan Nicol

11073 Dim

Patient monitor/Blood

Pressure Module

Non-Invasive Blood Pressure

Blood Pressure Measurement

Term ConceptBlood Pressure Measurement Metric

Module MDSNon-Invasive Blood Pressure ChannelPatient monitor MDS

Page 9: Martin Hurrell            Alan Nicol

11073 to HL7 R-MIM

Page 10: Martin Hurrell            Alan Nicol

11073 to HL7 R-MIM

VMD (Patient Monitor)

Page 11: Martin Hurrell            Alan Nicol

11073 to HL7 R-MIM

VMD (Patient Monitor)

Channel (Non-Invasive Blood

Pressure)

Page 12: Martin Hurrell            Alan Nicol

11073 to HL7 R-MIM

VMD (Patient Monitor)

Channel (Non-Invasive Blood

Pressure)

Metric(Blood Pressure

Measurement)

Page 13: Martin Hurrell            Alan Nicol

11073 to CDA R-MIM

Page 14: Martin Hurrell            Alan Nicol

11073 to CDA R-MIM

VMD (Patient Monitor)

Page 15: Martin Hurrell            Alan Nicol

11073 to CDA R-MIM

VMD (Patient Monitor)

Channel (Non-Invasive Blood Pressure)

Page 16: Martin Hurrell            Alan Nicol

11073 to CDA R-MIM

VMD (Patient Monitor)

Channel (Non-Invasive Blood Pressure)

Metric (Blood Pressure

Measurement)

Page 17: Martin Hurrell            Alan Nicol

11073 to CDA Example (Recording)

<observation><entryRelationship typeCode="COMP">

<observation classCode="OBS" moodCode="EVN"><id extension="67" root="2.16.840.1.113883.6.1" assigningAuthorityName="HL7 GAS WG" displayable="true"/><code code="19229" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_PRESS_CUFF_SYS"/><entryRelationship typeCode="COMP">

<observation classCode="OBS" moodCode="EVN"><effectiveTime value="20080527074500"/><targetSiteCode code="1781" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_UPEXT_ARM_UPPER_L"/><statusCode codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="validated-data"/><value code="3872" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_DIM_MMHG" value="119"/>

</observation></entryRelationship>

</observation>

Page 18: Martin Hurrell            Alan Nicol

11073 to CDA Example (Recording)

<observation><entryRelationship typeCode="COMP">

<observation classCode="OBS" moodCode="EVN"><id extension="67" root="2.16.840.1.113883.6.1" assigningAuthorityName="HL7 GAS WG" displayable="true"/><code code="19229" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_PRESS_CUFF_SYS"/><entryRelationship typeCode="COMP">

<observation classCode="OBS" moodCode="EVN"><effectiveTime value="20080527074500"/><targetSiteCode code="1781" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_UPEXT_ARM_UPPER_L"/><statusCode codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="validated-data"/><value code="3872" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_DIM_MMHG" value="119"/>

</observation></entryRelationship>

</observation>

Channel (Non-Invasive Blood Pressure)

Page 19: Martin Hurrell            Alan Nicol

11073 to CDA Example (Recording)

<observation><entryRelationship typeCode="COMP">

<observation classCode="OBS" moodCode="EVN"><id extension="67" root="2.16.840.1.113883.6.1" assigningAuthorityName="HL7 GAS WG" displayable="true"/><code code="19229" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_PRESS_CUFF_SYS"/><entryRelationship typeCode="COMP">

<observation classCode="OBS" moodCode="EVN"><effectiveTime value="20080527074500"/><targetSiteCode code="1781" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_UPEXT_ARM_UPPER_L"/><statusCode codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="validated-data"/><value code="3872" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_DIM_MMHG" value="119"/>

</observation></entryRelationship>

</observation>

Channel (Non-Invasive Blood Pressure)

Metric (Blood Pressure

Measurement)

Page 20: Martin Hurrell            Alan Nicol

11073 to CDA Example (Setup)

<organizer classCode="CLUSTER" moodCode="EVN"><participant contextControlCode="OP" typeCode="DEV"><participantRole classCode="ROL"><id extension="23" root="2.16.840.1.113883.6.1" assigningAuthorityName="HL7 GAS WG" displayable="true"/><id extension="1.3.2" root="2.16.840.1.113883.7.1" assigningAuthorityName="GE Healthcare" displayable="true"/><playingDevice classCode="DEV" determinerCode="INSTANCE"><code code="4429" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS"/><manufacturerModelName displayName="GE8001i"/><softwareName displayName="1.2.3"/></playingDevice></participantRole></participant>

Page 21: Martin Hurrell            Alan Nicol

11073 to CDA Example (Setup)

<organizer classCode="CLUSTER" moodCode="EVN"><participant contextControlCode="OP" typeCode="DEV"><participantRole classCode="ROL"><id extension="23" root="2.16.840.1.113883.6.1" assigningAuthorityName="HL7 GAS WG" displayable="true"/><id extension="1.3.2" root="2.16.840.1.113883.7.1" assigningAuthorityName="GE Healthcare" displayable="true"/><playingDevice classCode="DEV" determinerCode="INSTANCE"><code code="4429" codeSystem="1.3.6.1.4.1.19376.1.6" codeSystemName="ISO/IEEE 11073" codeSystemVersion="10201" displayName="MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS"/><manufacturerModelName displayName="GE8001i"/><softwareName displayName="1.2.3"/></playingDevice></participantRole></participant>

VMD (Patient Monitor)