2 #61-700 #61-701 #61-702 700 Series 200 Amp Clamp Meters 1 2 3 6 7 8 4 5 1. Non-contact voltage (NCV) (#61-701 and #61-702) With the NCV tab on the tip of the clamp close to an AC voltage, press the NCV button. The NCV LED will light and the beeper will beep. The closer the NCV tab is to AC volt- age, the louder the beep. To differentiate between hot and neutral in an outlet, insert the NCV tab into each slot in the outlet. The beeper will be much louder on the hot side of the outlet than the neutral.

Manual IDEAL Series 700

Embed Size (px)

Citation preview

Page 1: Manual IDEAL Series 700

2

#61-700#61-701#61-702

700 Series200 Amp Clamp Meters

1

2

3

6

7

8

4

5

1. Non-contact voltage (NCV) (#61-701 and #61-702)With the NCV tab on the tip of the clamp close to an ACvoltage, press the NCV button. The NCV LED will light andthe beeper will beep. The closer the NCV tab is to AC volt-age, the louder the beep.

To differentiate between hot and neutral in an outlet, insertthe NCV tab into each slot in the outlet. The beeper will bemuch louder on the hot side of the outlet than the neutral.

ND 2356-1 61-700/1/2 multimeter 1/15/08 11:18 AM Page 1

Page 2: Manual IDEAL Series 700

4

6. Max buttonPress Max prior to a measurement to capture the maximumreading displayed during the measurement. To clear the maxvalue, press the max button again, or press the reset button(#61-700).

7. Rotary switchTurn the rotary switch to the appropriate function and rangefor the measurement.

8. COM port and + portPlug the black lead into the com port and the red lead into the + port for VAC, VDC, Frequency, Resistance, andCapacitance measurements and Diode and Continuitychecks.

WARNING!1. DO NOT UNDER ANY CIRCUMSTANCES EXCEED THESE

RATINGS:• Voltage is not to exceed 600 Volts for #61-700,

and #61-701• Voltage is not to exceed 750VAC or 1000VDC for

#61-702• Resistance, function is not to be performed on circuits

capable of delivering greater than 600 Volts (1000VDC/ 750VAC #61-703).

• Frequency, Capacitance and Continuity functions are not to be performed on circuits capable of delivering greater than 500 Volts.

2. To avoid electrical shock hazards and/or damage to themeter:• Test NCV function on known live wire before using.

(#61-701 and #61-702)• Do not exceed the voltage ratings for the meter. Use

caution when measuring voltage.• Do not use during electrical storms. AC power sources

with inductive loads or electrical storms may result inhigh voltage. High energy transients can damage meterand present a dangerous shock hazard.

• Turn off power to the circuit or device being measuredbefore taking resistance and capacitance measurements.Fully discharge all capacitors before measuring.

The test lead can also be used to differentiate between thehot and neutral. Insert the rod test lead into the + port andsnap it into the test probe holder. Press the NCV button andinsert the probe into each slot of the outlet. The beeper willonly beep on the hot side of the outlet.

2. Test probe holderUse the single test probe holder to make voltage testing easy.

3. Indicator lights (#61-701 and #61-702)HI-V High voltage indicator

In any VAC/VDC range when a voltage greater than 30V is touched, the beeper will beep, and the red Hi-V LED will blink.

In addition to the LED and beeper, the 61-702 will shakewhen connected to high voltage. Even if the LED can’t beseen, or the beeper can’t be heard the shake can be felt.

NCV Non-contact voltage indicator

Continuity indicator

CP Clean power indicator (#61-702)

When the 60Hz AC power is clean, the green LED will be on. If the meter is connected to AC power, and the greenLED is not on, the circuit has more than 5% (±2%) totalharmonic distortion (nominal). Investigation with special-ized equipment may be necessary. Sensitivity: 20VAC on200VAC range; 90V on 750VAC range.

4. Hold buttonPress hold button during measurement to capture the value.To clear the held value, press the hold button again, orpress the reset button. (#61-700)

5. NCV button (#61-701 and #61-702) or Reset button(#61-700)Press the NCV button on the #61-701 and #61-702 to testfor non-contact voltage. On the #61-700, this is a resetbutton to reset the hold and max values.

3

ND 2356-1 61-700/1/2 multimeter 1/15/08 11:18 AM Page 3

Page 3: Manual IDEAL Series 700

6

Lead StorageSilicone test leads are provided with the meter. These testleads stay flexible in very cold weather and can withstand hightemperatures without melting. For convenient lead storage,wrap the leads as shown.

AC Current

Range Resolution Max Accuracy ModelsDisplay

200A 0.1A 199.9 ±(3%+5) 61-700, 61-701,61-702

Frequency response: 50Hz - 60Hz; ±(5%+5) 60Hz - 400Hz.

5

3. Ensure meter is in proper working order before using.Visually inspect meter for damage. Performing a continuity check can verify proper operation. If the meter reading goesfrom overload to zero, this typically means the meter is inproper working order.

4. Visually inspect leads for damage before using. Replace if insulation is damaged or leads appear suspect.

5. Never ground yourself when taking electrical measurements.Do not touch exposed metal pipes, outlets, fixtures etc.Keep your body isolated from ground by using dry clothing,rubber shoes, mats, or any other approved insulating mater-ial. Keep your fingers behind the finger guards on theprobes. Work with others.

6. Before beginning all unknown measurements, set meter to highest possible range.

7. Disconnect the meter from the circuit before turning off any inductor, including motors, transformers, and solenoids.

8. Clean the exterior with clean dry cloth. Do not use liquid.

General Operation

Auto power offTo extend the life of your battery, the meter automatically turnsoff after approximately 60 minutes.

Selecting RangesFor all ranges and functions choose a range just above theexpected value. If display reads “OL” (overrange), select ahigher range. If display shows less than three numbers, selecta lower range for better resolution.

Accuracy SpecificationsAccuracy specifications are based on nominal operating tem-perature of 74°F ±8°, and a relative humidity less than 75%.Accuracy is specified as ± (% of reading + number of leastsignificant digits).

ND 2356-1 61-700/1/2 multimeter 1/15/08 11:18 AM Page 5

Page 4: Manual IDEAL Series 700

87

DC VoltageRange Resolution Max Accuracy Models

Display200mV 0.1mV 199.9 ±(0.5%+1) 61-7002000mV 1mV 1999 ±(0.5%+1) 61-700, 61-702

20V 10mV 19.99 ±(0.5%+1) 61-701200V 0.1V 199.9 ±(0.5%+1) 61-700, 61-702 600V 1V 600 ±(0.5%+1) 61-700

1000V 1V 1000 ±(0.5%+1) 61-702

To Measure Voltage:5. Plug the black test lead into the COM port and the red test

lead into the + port.6. Set the rotary switch to appropriate VAC or VDC range.7. Connect the meter in parallel with the load or circuit.8. Measure the voltage

Frequency

Range Resolution Max Accuracy ModelDisplay

2k~40kHz 1Hz 40.0 ±(0.1%+3) 61-700,61-701

To Measure AC Current:1. Set the rotary switch to the appropriate AAC range.2. With the jaws closed, separate one wire from a bundle using

the long fixed jaw.3. Slide the wire to the corner where the two jaws meet, then

open the jaws to let it in.4. Measure the current in only one wire at a time.

AC & DC Volts

AC VoltageRange Resolution Max Accuracy Models

Display2000mV 1mV 1999 ±(1.2%+3) 61-702

200V 0.1V 199.9 ±(1.2%+3) 61-700, 61-701,61-702

600V 1V 600 ±(2%+5) 61-700, 61-701750V 1V 750 ±(2%+5) 61-702

Frequency response: 50Hz - 500Hz; ±(2%+5) 500Hz - 1KHz;50Hz - 500Hz on 600V and 750V ranges.

ND 2356-1 61-700/1/2 multimeter 1/15/08 11:18 AM Page 7

Page 5: Manual IDEAL Series 700

109

To Measure Frequency:1. Plug the black test lead into the COM port and the red test

lead into the + port.2. Set the rotary switch to the HZ position.3. Connect the meter in parallel with the load or circuit.4. Meter is auto-ranging, and will select 2k 20k or 40k range.5. Measure the frequency.

Resistance (Ohms)

Range Resolution Max AccuracyDisplay

200Ω 0.1Ω 200.0 ±(1% +3)200kΩ 0.1kΩ 200.0 ±(1% +3)

To Measure Resistance:1. Turn the power off to the circuit or device that is to be mea-

sured and discharge all capacitors before attempting a mea-surement.

2. Plug the black test lead into the COM port and the red test lead into the + port.

3. Set the rotary switch to the appropriate Ohms range.4. For correct reading, ensure that the device being tested

contains no voltage.5. Measure the resistance.

Capacitance Testing

Range Resolution Max Accuracy ModelsDisplay

200MFD 0.1MFD 199.9 ±(3% +5) 61-701, 61-702

To Measure Capacitance:1. Disconnect the capacitor from power.2. Short the terminals to discharge the capacitor.3. Disconnect any resistors that might be between the termi-

nals of the capacitor.4. Plug the black test lead into the COM port and the red test

lead into the + port.5. Set the rotary switch to the MFD position.6. Connect the test leads to the capacitor and measure the

capacitance.

Diode Check:To ensure a proper functioning diode, the meter will develop avoltage across the component from a test current. The diodetest function allows measurements of forward voltage dropsacross diode and transistor junctions. 1. Turn off power to the device or circuit that is being tested

and discharge all capacitors.2. Plug the Black test lead into the COM port and the Red test

lead into the + port.3. Set the rotary switch to the position.4. Connect the test leads to the diode. Normally the forward

voltage drop of a good silicone diode is shown between400V and 90V. If the diode under test is defective, “000”(short circuit) or “OL” (non-conductive) is displayed.

ND 2356-1 61-700/1/2 multimeter 1/15/08 11:18 AM Page 9

Page 6: Manual IDEAL Series 700

11

Continuity Check

To Verify Continuity:A continuity test ensures that all circuit connections are intact. 1. Plug the Black test lead into the COM port and the Red test

lead into the + port.2. Set the rotary switch to the position.3. Connect the test leads to the circuit to be measured. The

beeper will sound within 100mS if the resistance of the cir-cuit measured is lower than 100 Ohms.

4. In addition to the beeper, #61-701 and #61-702 have a LEDto verify continuity.

General SpecificationsDisplay: 3-1/2 digit LDC displayPolarity Indication: Automatic, negative indicated,

positive impliedOverrange Indication: “OL” Low Battery Indication: when the battery voltage

drops below operating voltageAuto Power Off: Approx. 60 min.Temperature Coefficient: 0.1 x (specified accuracy) /

°F, <64°F or >82°F Power Requirements: 9V NEDA 1604 Battery Life: 300 hours typical with alkaline

(150 hours #61-702); 150 hourstypical with carbon zinc (75 hours #61-702)

Installation Category: IEC 1010-1, IEC1010-2-032, Cat III 600V (1000V for #61-703)

Maintenance

Battery Replacement:

When the multimeter displays the battery must bereplaced. Disconnect and unplug leads, turn meter off, andremove the battery cover. Replace the battery with a NEDA type1064 9V battery.

Fuse Replacement:(For #61-702)Remove the battery cover and back case. Replace the fuse witha 0.1A 660V (6.3 x 32mm) fast blow fuse.

Warranty StatementThis tester is warranted to the original purchaser against defectsin material and workmanship for two years from the date of pur-chase. During this warranty period, IDEAL INDUSTRIES, INC.will, at its option, replace or repair the defective unit, subject toverification of the defect or malfunction.

This warranty does not cover fuses, batteries or damage fromabuse, neglect, accident, unauthorized repair, alteration, orunreasonable use of the instrument.

Any implied warranties arising out of the sale of an IDEAL prod-uct, including but not limited to implied warranties of mer-chantability and fitness for a particular purpose, are limited to theabove. The manufacturer shall not be liable for loss of use of theinstrument or other incidental or consequential damages,expenses, or economic loss, or for any claim or claims for suchdamage, expenses or economic loss.

State laws vary, so the above limitations or exclusions may notapply to you. This warranty gives you specific legal rights, andyou may also have other rights which vary from state to state.

C US

WarningTo avoid electrical shock, remove test lead before opening thecover. Repairs or servicing not covered in this manual shouldonly be performed by qualified personnel.

IDEAL INDUSTRIES, INC.Sycamore, IL 60178, U.S.A.800-304-3578 Customer Assistancewww.testersandmeters.com ND 2356-2 Made in Taiwan

ND 2356-1 61-700/1/2 multimeter 1/15/08 11:18 AM Page 11

Page 7: Manual IDEAL Series 700

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: 61-700 MODEL: 61-701 MODEL: 61-702 MODEL: 61-704 The Service Information provides the following information: • Precautions and safety information • Specifications • Performance test procedure • Calibration and calibration adjustment procedure • Basic maintenance (replacing the battery)

Form number: TM61700-1-2-4 Revision: 4. Date: Sept 2004

Form number TM61700-1-2-4 Rev 4 September 2004

Page 8: Manual IDEAL Series 700

TABLE OF CONTENTS Title PageIntroduction 1Precautions and Safety Information 1Symbol Table 1Safety Information 2Indicator Functions 2Certifications and compliance 2Specifications 3 General Specification 3 Voltage Specifications 4 Current Specifications 4 Resistance Specifications 4 Frequency Specification 4 Capacitance Specifications 4 Diode / Continuity Specifications 4Performance Verification 5/6Calibration 6/7Calibration Reference Points 8 Replacing the Battery 9

Form number TM61700-1-2-4 Rev 4 September 2004

Page 9: Manual IDEAL Series 700

Page 1 Introduction

Warning To avoid shock or injury, do not perform the verification tests or calibration procedures described in this manual

unless you are qualified to do so.

The information provided in this document is for the use of qualified personnel only.

Caution

The 61-700 series contains parts that can be damaged by static discharge. Follow the standard practices for handling static sensitive

devices.

For additional information about IDEAL INDUSTRIES, INC. and its products, and services, visit IDEAL INDUSTRIES, INC. web site at:

www.idealindustries.com

Precautions and Safety Information Use the meter only as described in the Users Manual. If you do not do so, the protection provided by the meter may be impaired. Read the “Safety Information” page before servicing this product. In this manual, a Warning identifies conditions and actions that pose hazard (s) to the user; a Caution identifies conditions and actions that may damage the meter or the test instruments. The Symbols The symbols used on the Meter and in this manual are explained in Table A. Table A Symbols

Symbol Description Symbol Indicator Lights

Battery HI-V High Voltage Indicator >30 V indicator is on

Cautionary or important information in manual NCV Non-Contact indicator

Danger- Risk of electrical shock Continuity indicator

Double Insulation- Protection Class II CP

Clean power indicator On if power is Clean Off if power has > 5% THD

CAT III IEC Over-voltage Category III

Form number TM61700-1-2-4 Rev 4 September 2004

Page 10: Manual IDEAL Series 700

Page 2 SAFETY Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use the product only as specified.

CAUTION. These statements identify conditions or practices that could result in damage to the equipment or other property.

WARNING. These statements identify conditions or practices that could result in personal injury or loss of life. Specific precautions: Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the product without the covers in place. Electric overload. Never apply a voltage to a connector on the product that is outside the range specified for that connector. Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they are connected to a voltage source. Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or damp conditions. Indicator Functions:

• Hi-V: In ACV, DCV, Resistance, and Continuity/Diode test, the end user will get an audible warning and the Hi-V light will blink if there is Voltage Greater that 30V present at the input terminals.

o In addition to the LED and audible beep, the 61-702 and 61-704 has a tactile vibration when voltage is present.

• NCV: When the NCV button is pressed, the analog to digital circuit is turned off and the NCV circuit is activated. Sensing for the NCV is at the tip of the Clamp Jaw or by the Red Test Lead in the red (+) input terminal. Follow the operation instructions giving in the operator’s manual for proper use of this function. The NCV function is on Ideal Models 61-701,61-702, and 61-704.

• CP: When the 60Hz AC power is clean, with less than 5% Total Harmonics Distortion (THD) on AC voltage, the green CP light will be energized. If there is greater that 5% THD on the line, the green CP light will be de-energized.

Certifications and compliances

Safety Designed to IEC 1010-1, UL3111-1 and CSA specifications 1000V DC Category III 61-702, 61-704 600V DC Category III 61-700, 61-701 750V AC Category III 61-702, 61-704

Input rating

600V AC Category III 61-700, 61-701 CAT III: Distribution level mains, fixed installation. CAT II: Local level mains, appliances, and portable equipment. Over voltage category CAT I: Signal level, special equipment or parts of equipment, telecommunication, electronics.

Form number TM61700-1-2-4 Rev 4 September 2004

Page 11: Manual IDEAL Series 700

Page 3 General specifications

Characteristics Description Display 3½ Digit LCD display Display Count 2000 count, maximum reading 1999 Over range Indication “OL” is displayed Sampling Rate 2.5 time/second Operating Environment: Relative Humidity

0°C to 50°C (32°F to 122°F) 0 ~ 75% RH

Storage Environment: -20°C to 60°C (-4°F to 140°F) at <80 relative humidity Power source: 9V Battery (NEDA 1604) Battery Live: 300 hours typical (alkaline) 61-700 and 61-701

150 hours typical (alkaline) 61-702 and 61-704 Low Battery Indicator: symbol indicates low battery voltage Auto power off Approximately 60 minutes Dimensions 9.7” H X 3.7” W X 1.8” D

247mm H X 94mm W X 46mm D Maximum Cable Size ACA 1¼” (32mm) Weight: Approximately 13.4 oz. or 430g including battery

RANGES and ACCURACY SPECIFICATION Accuracy: Accuracy specifications at 23°C ±5°C (73.4°F ±9°F) less than 75% RH. Temperature Coefficient: 0.1 times the applicable accuracy specification per degree C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F) Electrical Specification: Accuracy are ±(reading plus number of digits) at 23°C ±5°C <75% RH

Form number TM61700-1-2-4 Rev 4 September 2004

Page 12: Manual IDEAL Series 700

Page 4 61-700

Function / Range Ranges Accuracy

200V, 50Hz - 500Hz 1.2% + 3digits AC Voltage 200V, 50Hz - 1KHz 2.0% + 5 digits 600V, 50Hz - 500Hz 2.0% + 5 digits DC Voltage 200mV/2000mV/200V/600V 0.5% + 1 digit AC Current 200A, 50Hz - 60Hz 3.0% + 5 digit 200A, 60Hz - 400Hz 5.0% + 5 digit Frequency 2K ~ 40KHz (Auto-ranging only) 0.1% + 3 digits Resistance 200Ω/200KΩ 1.0% + 3digit Diode Check 1mA ± 0.6mA Not specified Continuity <100Ω on Diode, Continuity Not Specified

61-701

Function / Range Ranges Accuracy

200V, 50Hz - 500Hz 1.2% + 3 digits AC Voltage 200V, 50Hz - 1KHz 2.0% + 5 digits 600V, 50Hz - 500Hz 2.0% + 5 digits DC Voltage 2000mV/20V/ 600V 0.5% + 1 digit AC Current 200A, 50Hz - 60Hz 3.0% + 5 digit 200A, 60Hz - 400Hz 5.0% + 5 digit Capacitance 200MFD 3.0% + 5 digits

Frequency 2K ~ 40KHz (Auto-ranging only) 0.1% + 3 digits Resistance 200Ω/200KΩ 1.0% + 3digit Diode Check 1mA ± 0.6mA Not specified Continuity <100Ω on Diode, Continuity Not specified

61-702, 61-704

Function / Range

Ranges Accuracy

2000mV/200V, 50Hz - 500Hz 1.2% + 3digits AC Voltage 2000mV/200V, 500Hz - 1KHz 2.0% + 5 digits 750V, 50Hz - 500Hz 2.0% + 5 digits DC Voltage 2000mV/200V/1000V 0.5% + 1 digit AC Current 200A, 50Hz - 60Hz 3.0% + 5 digit 200A, 60Hz - 400Hz 5.0% + 5 digit Capacitance 200MFD 3.0% + 5 digits

Frequency 2K ~ 40KHz (Auto-ranging only) 0.1% + 3 digits Resistance 200Ω/200KΩ 1.0% + 3digit Diode Check 1mA ± 0.6mA Not specified Continuity <300Ω on Diode, Continuity Not specified

AC Converter: Average responding, RMS Calibrated to Sine Wave Overload Protection: AC and DC Voltage: not to exceed 600V DC or VAC RMS for the model 61-700 and 61-701, and 1000V DC or 750V AC for the 61-702 and 61-704 Resistance: Not to exceed 500V DC or AC RMS Capacitance, Frequency, Diode Check, Continuity: not to exceed 500V DC or VAC RMS

Form number TM61700-1-2-4 Rev 4 September 2004

Page 13: Manual IDEAL Series 700

Page 5 PERFORMANCE VERIFICATIONS Perform the following analysis; if the meter conforms to the limits listed in Table 1 through 5 the meter is functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure must be performed. Performance Verification Preparation 1. Turn on the calibrator, allow calibrator to warm up. Temperature stabilization should be reached after 30 minutes. 2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5V DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test. 3. Input the values listed in Table 1 through 5 Table 1 AC Voltage Test

Function /Range Input Low

Limit High Limit Model number

V AC 2000mV 1900mV AC @ 50Hz 1874 1926 61-702, 61-704 V AC 2000mV 1900mV AC @ 500Hz 1874 1926 61-702, 61-704 V AC 2000mV 1900mV AC @ 1kHz 1857 1943 61-702, 61-704 V AC 200V 190V AC @ 50Hz 187.4 192.6 61-700, 61-701, 61-702, 61-704 V AC 200V 190V AC @ 500Hz 187.4 192.6 61-700, 61-701, 61-702, 61-704 V AC 200V 190V AC @ 1kHz 185.7 194.3 61-700, 61-701, 61-702, 61-704 V AC 600V 500V AC @ 50Hz 485 515 61-700, 61-701 V AC 600V 500V AC @ 500Hz 485 515 61-700, 61-701 V AC 750V 700V AC @ 50Hz 681 719 61-702, 61-704 V AC 750V 700V AC @ 500Hz 681 719 61-702, 61-704

Table 2 DC Voltage Test

Function /Range Input Low

Limit High Limit Model number

V DC 200mV 190mV 188.9 191.1 61-700 V DC 2000mV 1900mV 1889 1911 61-700, 61-702, 61-704 V DC 20V 19V 18.89 19.11 61-701 V DC 200V 190V 188.9 191.1 61-700, 61-701, 61-702, 61-704 V DC 600V 500V 496 504 61-700, 61-701 V DC 1000V 900V 894 906 61-702, 61-704

Table 3 AC Current Test

Function /Range Input Low

Limit High Limit Model number

A AC 200A 100A AC @ 50Hz 96.5 103.5 61-700, 61-701, 61-702, 61-704 A AC 200A 100A AC @ 400Hz 94.5 105.5 61-700, 61-701, 61-702, 61-704

Form number TM61700-1-2-4 Rev 4 September 2004

Page 14: Manual IDEAL Series 700

Page 6 Table 4 Capacitance, Frequency, and Resistance Test

Function /Range Input Low

Limit High Limit Model number

Hz auto 1KHz @ 10V .996 1.004 61-700, 61-701 Hz auto 10KHz @ 10V 9.96 10.04 61-700, 61-701 MFD 200 100µF 96.5 103.5 61-701, 61-702, 61-704 Ω 200 100Ω 98.7 101.3 61-700, 61-701, 61-702, 61-704 Ω 200K 100KΩ 98.7 101.3 61-700, 61-701, 61-702, 61-704

Table 5 Diode and Continuity Check

Function /Range Test Value Low

limits High Limit Model number

Diode 500mV DC 485 515 61-700, 61-701, 61-702, 61-704

Continuity 40Ω beep on, 150Ω beep off 61-700, 61-701

Continuity

50Ω beep on, 300Ω beep off 61-702, 61-704

CALIBRATION Calibration Preparation 1. Turn on the calibrator, allow calibrator to warm up. Perform calibration at 23±2°C at relative humidity of < 70%. Temperature stabilization should be reached after 30 minutes. 2. Disconnect the test leads and turn the range switch to “OFF”. 3. Remove the screws holding the battery cover and one at the jaw. 4. Remove the case bottom using care not to damage the battery connector and leads to the continuity beeper. (Beeper is attached to the bottom case cover.) 5. Using a calibrated meter ensure the battery measures a minimum of 7.5V DC. If the battery measures under 7.5V DC, replace the battery. Calibration Procedure It is recommended that all IDEAL meters undergo the following calibration procedure on an annual basis. The class of calibrator or equipment should have an accuracy that exceeds, by an expectable ratio the accuracy of this instrument.

Form number TM61700-1-2-4 Rev 4 September 2004

Page 15: Manual IDEAL Series 700

Page 7

V DC Calibration 61-700 (Refer to Figure 1) 1. Set the function / range to 200mV DC. 2. Connect the calibrator to the V and COM inputs on the meter. 3. Output 190.0mV DC. Adjust VR1 (VR 200Ω) until unit display reads 190.0 De-energize source and remove test leads 61-701 (Refer to Figure 2A) 1. Set the function / range to 20V DC. 2. Connect the calibrator to the V and COM inputs on the meter. 3. Output 19.00V DC. Adjust VR1 (VR 200Ω) until unit display reads 19.00 De-energize source and remove test leads 61-702 (Refer to Figure 3A) 61-704 (Refer to Figure 4A) 1. Set the function / range to 2000mV DC. 2. Connect the calibrator to the V and COM inputs on the meter. 3. Output 1.900 VDC.

Adjust VR1 (VR 200Ω) until unit display reads 1900. De-energize source and remove test leads

V AC Calibration: 61-704 only (Refer to Figure 4A) 1. Set the function/range to the 2000mV AC. 2. Connect the calibrator to the V and COM inputs on the meter. 3. Output 1.900 VAC/50Hz

Adjust VR5 (VR 100Ω) until unit display reads 1900V± 1 digit. De-energize source and remove test leads

A AC Calibration: 61-700 (Refer to Figure 1) 61-701 (Refer to Figure 2B; peel front label back to access calibration hole) 61-702 (Refer to Figure 3B; peel front label back to access calibration hole) 61-704 (Refer to Figure 4B; peel front label back to access calibration hole) 1. Set the function / range to the 200A AC 2. Set output of the AC calibrator for 10.00A/60Hz +/- 0.01% and connect it to Coil = 10N = 100.0A

AC. 3. Clamp the jaws to the coil = 10N. 4. Adjust VR2 (VR 2KΩ) for a display reading of 100.0 ±1 digit Calibration of the 61-700 series is complete. Remove all leads from the calibrator and equipment. Return unit to proper operating condition.

Form number TM61700-1-2-4 Rev 4 September 2004

Page 16: Manual IDEAL Series 700

Page 8

Figure 1 Figure 2A Figure 2B

Figure 3A Figure 3B Figure 4A Figure 4B

Form number TM61700-1-2-4 Rev 4 September 2004

Page 17: Manual IDEAL Series 700

Page 9 Battery Replacement (refer to Figure 5) 1. Disconnect the test leads from any circuit under test and turn off meter. 2. Use a Philips head screwdriver to remove the screws on battery cover. 3. Remove battery from the battery compartment. 4. Install new 9V battery (NEDA #1604). An alkaline type is recommended. 5. Install new battery into compartment using care to install to proper polarity. 6. Reinstall battery cover.

Figure 5

Form number TM61700-1-2-4 Rev 4 September 2004