manua siemes.pdf

Embed Size (px)

Citation preview

  • 8/10/2019 manua siemes.pdf

    1/461

    SOMATOM DefinitionApplication Guide

    ProtocolsPrinciplesHelpful Hints

    Software Version syngoCT 2007A

  • 8/10/2019 manua siemes.pdf

    2/461

    To improve future versions of this application guide we

    would greatly appreciate your questions, suggestions,and comments. The information presented in thisapplication guide is for illustration only and is notintended to be relied upon for instruction in the prac-tice of medicine.

    Any healthcare practitioner reading this information is

    reminded that they must use their own learning, train-ing, and expertise in dealing with their individualpatients. This material is no substitute for that dutyand is not intended by Siemens Medical Solutions Inc.,to be used for any purpose in that regard.

    The drugs and doses mentioned are consistent with

    the approval labeling for uses and/or indications of thedrug. The treating physician bears the sole responsibil-ity for the diagnosis and treatment of patients, includ-ing drugs and doses prescribed in connection withsuch use.

    The pertaining operating instructions must always be

    strictly followed when operating a SOMATOM Defini-tion scanner. The statutory source for the technicaldata are the corresponding data sheets.

    The available options depend on the system configura-tion. Please contact your local Siemens representativefor further information.

    We would like to express our sincere gratitude to themany customers who contributed valuable input.

    Special thanks go to Alexander Zimmermann, AndreasSchaller, Heike Theessen, Herbert Bruder, Rainer Rau-pach, and Michele Sohlden, for their valuable assis-tance.

    Please contact us:International CT Application Hotline:

    Tel. no. +49-1803-112244

    email: [email protected]

    Editors: Christiane Bredenhller and Ute Feuerlein

  • 8/10/2019 manua siemes.pdf

    3/461

    Overview

    3

    User Documentation 14

    Scan and Reconstruction 20

    Dose Information 56

    Workflow Information 86

    Application Information 116

    Contrast Medium 142

    Head 152

    Neck 180

    Shoulder 202

    Thorax 212

    Abdomen 242

    Spine 268

    Pelvis 286

    Upper Extremities 300

    Lower Extremities 306

    Cardiac CT 314

    Vascular 316

    Specials 370

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 8/10/2019 manua siemes.pdf

    4/461

  • 8/10/2019 manua siemes.pdf

    5/461

    Overview

    5

  • 8/10/2019 manua siemes.pdf

    6/461

    6

    Contents

    User Documentation 14- Application Guides 14

    - SOMATOM Operator Manual 15

    - syngoCT Operator Manual 16

    - System Owner Manual 18

    - Online Help 18

    Scan and Reconstruction 20

    Concept of Scan Protocols 20

    Scanner Features 22

    - Straton-Tube 22

    - z-Sharp Technology 24

    - UFC detector 26

    - Shaped Filter for Dose Reduction 27

    Scan Modes 29

    - Sequential Scanning 29

    - Spiral Scanning 29

    - Dynamic Multiscan 30

    - Dynamic Serioscan 30

    - Cardiac Scanning with Dual Source CT 30

    Scan Set-up 32

    - Acquisition, Slice Collimation and

    Slice Width 33

    - Recon Increment 36

    - Pitch and Pitch adaptation 36

    - Kernels 39

    - Extended FoV 45

    - Topo Length 46

    - Feed in/Feed out 46 Image Filters 47

    Head Imaging 49

    - Head Modes 49

    - Neuro Modes 50

    - Automatic Bone Correction 51

    - Positioning 52- Automatic Isocenter Adaptation 53

  • 8/10/2019 manua siemes.pdf

    7/461

  • 8/10/2019 manua siemes.pdf

    8/461

    8

    Contents

    - SOMATOM LifeNet online 120

    Image Converter 128 Report Template Configuration 131

    File Browser 132

    Camtasia 136

    - Key features 136

    - Additional Important Information 140

    Patient Protocol 141

    Contrast Medium 142

    The Basics 142

    IV Injection 145 Bolus Tracking 146

    Test Bolus using CARE Bolus 148

    Test Bolus 149

    Head 152 Overview 152

    - General Hints 154

    - Head Kernels 155

    Scan Protocols 156

    - HeadRoutine 156

    - HeadNeuro 160- HeadRoutineSeq 162

    - HeadNeuroSeq 166

    - InnerEar 168

    - Sinus 172

    - SinusVol 174

    - Orbit 176- Dental 178

    Neck 180

    Overview 180

    - General Hints 182

    - Body Kernels 183

    Scan Protocols 186

  • 8/10/2019 manua siemes.pdf

    9/461

    Contents

    9

    - NeckRoutine 186

    - NeckVol 190- NeckThorax 194

    - NeckThorAbd 198

    Shoulder 202

    Overview 202

    - General Hints 203

    - Body Kernels 204

    Scan Protocols 206

    - ShoulderRoutine 206

    - ShoulderVol 210

    Thorax 212

    Overview 212

    - General Hints 214

    - Body Kernels 215 Scan Protocols 218

    - ThoraxRoutine 218

    - ThoraxVol 220

    - ThoraxHR 224

    - ThoraxSeqHR 226

    - ThoraxECGSeqHR 228- LungLowDose 230

    - LungCARE 234

    - ThorAbd 238

    Abdomen 242

    Overview 242

    - General Hints 244

    - Body Kernels 246

    Scan Protocols 248

    - AbdomenRoutine 248

    - AbdomenVol 252

    - AbdMultiPhase 256

    http://-/?-http://-/?-
  • 8/10/2019 manua siemes.pdf

    10/461

  • 8/10/2019 manua siemes.pdf

    11/461

  • 8/10/2019 manua siemes.pdf

    12/461

    12

    Contents

    - Trauma 374

    - TraumaVol 375- PolyTrauma 376

    - HeadTrauma 379

    - HeadTraumaSeq 380

    - Additional Important Information 381

    TestBolus Protocol 382

    - Test Bolus 382

    Radiation Therapy 384

    Radiation Therapy Planning 384

    - Benefits 387 Workflow 391

    Scan Protocols 392

    - Overview 392

    - RT_Head 393

    - RT_Thorax 394

    - RT_Breast 395- RT_Abdomen 396

    - RT_Pelvis 398

    - Additional Important Information 400

    Children 402 Overview 402

    - General Hints 406

    - Head Kernels 410

    - Body Kernels 411

    Scan Protocols 412

    - HeadRoutine 412- HeadRoutineSeq 416

    - InnerEar 420

    - Sinus 424

    - Orbit 428

    - NeckRoutine 430

    - ThoraxRoutine 434

    - ThoraxSeqHR 438

    - AbdomenRoutine 440

  • 8/10/2019 manua siemes.pdf

    13/461

    Contents

    13

    - SpineRoutine 444

    - Extremity 446- HeadAngio 448

    - CarotidAngio 450

    - BodyAngio 454

    - NeonateBody 458

  • 8/10/2019 manua siemes.pdf

    14/461

    14

    User Documentation

    Welcome to SOMATOM Definition - The World's

    First Dual Source CT

    As a SOMATOM Definition user you will have access tocutting-edge technology, comprehensive imagingapplications, and a wide range of powerful workflowsolutions.

    The SOMATOM CT user documentation is intended forhealthcare practitioners, helping you to optimally useyour CT scanner in daily clinical work and cutting edgeresearch. The following components of the user docu-mentation will support you to fully utilize the SOMA-TOM Definition's unprecedented image quality whileproviding increased imaging speed.

    The applications available on your SOMATOM Defini-tion depend on your system configuration.

    Application Guides

    The Application Guides complete the SOMATOM Defi-nition user documentation. The Application Guides areavailable in English, German, French, Spanish, Japa-nese, and Chinese. They consist of a scanner-specificguide, a cardiac guide, and one for the post-processingapplications:

    The SOMATOM Definition Application Guidecon-tains:

    Protocols

    Principles

    Helpful Hints

  • 8/10/2019 manua siemes.pdf

    15/461

    User Documentation

    15

    The Cardiac CT Application Guidecontains:

    ProtocolsPrinciples

    Helpful Hints

    syngoCalcium Scoring

    syngoInSpace 4D CT

    syngoCirculation

    The Clinical Applications Application Guidecon-tains:

    syngo3D

    syngoColonography CT

    syngoDental CTsyngoPulmo CT

    syngoLungCARE CT

    syngoVolume Calculation

    syngoDynamic Evaluation

    syngoNeuro DSA CT

    syngoNeuro Perfusion CTsyngoBody Perfusion CT

    In addition to the Application Guides, the SOMATOMOperator Manual, the syngoCT Operator Manual, the

    System Owner Operator Manual, and the Online Helpare also supplied.

    SOMATOM Operator Manual

    In this manual, you will find a description of the hard-ware components of the system and its operation. TheRelease Note is also included.

  • 8/10/2019 manua siemes.pdf

    16/461

    16

    User Documentation

    syngoCT Operator Manual

    The syngoCT Operator Manuals are a workflow orien-tated description for operation of the system softwareand its frequently used functions. They contain step-by-step instructions from registration to examinationas well as evaluation and documentation of your

    examination results.For further information about basic software opera-tion, please refer to the corresponding syngoCT Oper-ator Manual:

    syngoCT Operator Manual Volume 1:

    syngoSecurity Package

    Siemens Virus Protection

    Basics

    Life

    Patient Browser

    Data Set ConversionCamtasia

    Savelog

    syngoCT Operator Manual Volume 2:

    Preparations

    Examination

    MPPS

    HeartView CT

    Care Bolus CT

    Viewing

    Filming

  • 8/10/2019 manua siemes.pdf

    17/461

    User Documentation

    17

    syngoCT Operator Manual Volume 3:

    syngo3D

    syngoDental CT

    syngoVolume Calculation

    syngoDynamic Evaluation

    syngoCT Operator Manual Volume 4:syngoLung CARE CT

    syngoPulmo CT

    syngoNeuro Perfusion CT

    syngoBody Perfusion CT

    syngoCT Operator Manual Volume 5:

    syngoCalcium Scoring

    syngoCirculation

    syngoNeuro DSA CT

    syngoCT Operator Manual Volume 6:syngoInSpace4D CT

    syngoColonography CT

  • 8/10/2019 manua siemes.pdf

    18/461

    18

    User Documentation

    System Owner Manual

    In this manual, you will find details and preconditionsfor the operation of the CT scanner.

    Online Help

    The Online Help explains the operation of the systemssoftware. It is available in English, German, French,Spanish, Japanese, and Chinese.

  • 8/10/2019 manua siemes.pdf

    19/461

    User Documentation

    19

  • 8/10/2019 manua siemes.pdf

    20/461

    20

    Scan and Reconstruction

    Concept of Scan ProtocolsThe scan protocols for adults and children are definedaccording to body regions Head, Neck, Shoulder,Thorax, Abdomen, Pelvis, Spine, Upper Extremities,Lower Extremities, Vascular, RT, Specials,Private

    and Cardiac.

    The protocols for special applications are described inthe Application Guide "Clinical Applications"or in thecase of a Heart View examination, in the ApplicationGuide "Cardiac CT".

    The general concept is as follows: All protocols withouta suffix are standard spiral modes. For example,"Sinus" means the spiral mode for the sinus.

    The suffixes of the protocol names are as follows:

    "Seq": for sequence studies

    "Routine": for routine studies

    "ECG": use an ECG-gated or -triggered mode

    "Vol": use the 3D Recon workflow

    "Neuro": for neurological examinations with a specialmode

  • 8/10/2019 manua siemes.pdf

    21/461

    Scan and Reconstruction

    21

    The prefixes of the protocol names are as follows:

    "RT": for radiation therapy planning studies

    "DS": for Dual Source CT scans

    The availability of scan protocols depends on the sys-tem configuration.

  • 8/10/2019 manua siemes.pdf

    22/461

    22

    Scan and Reconstruction

    Scanner Features

    Straton-Tube

    The SOMATOM Definition CT system is equipped withtwo STRATON X-ray tubes.

    These newly developed X-ray tubes offer significantlyreduced cooling times for shorter interscan delays andincreased power reserves. The full X-ray power of160 kW provides the needed dose for patients with alarge body habitus.

    Example for a one tube mode is the ThoraxRoutineprotocol (120 kV, 100 mAs, 0.5 s rot, 64 x 0.6 mm,pitch factor 1.4):

    a scan range of 300 mm can be covered in 6.58 s

    dose can be increased up to 170 mAs without areduction of table feed.

  • 8/10/2019 manua siemes.pdf

    23/461

    Scan and Reconstruction

    23

    Conventional anode cools down slowly after exposure

    STRATON accumulates very little heat during the expo-sure, which is dissipated quickly

    Slow heat exchange

    Anode within vacuum

    Cooling oil

    Anode in direct contactwith cooling oil

    Cooling oil

    Fastest heat exchange

  • 8/10/2019 manua siemes.pdf

    24/461

    24

    Scan and Reconstruction

    z-Sharp Technology

    The unique STRATON X-ray tube utilizes an electronbeam that is accurately and rapidly deflected, creat-ing two precise focal spots alternating 4,608 timesper second. This doubles the X-ray projections reach-ing each detector element. The two overlapping pro-

    jections result in oversampling in the z-direction,known as z-Sharp Technology. The resulting mea-surements interleave half a detector slice width,doubling the scan information without a corre-sponding increase in dose.

    The purpose of this technique is two-fold: First, by

    doubling the sampling, the slice width can bereduced. With SOMATOM Definition it is possible toreconstruct 0.6 mm slices at any pitch (1.5) withbest image quality. With overlapping 0.6 mm slices,a z-axis resolution below 0.4 mm is obtained. Sec-ond, the improved sampling completely removes, atany pitch, the so-called windmill artifacts frequently

    visible in Multislice CT images in the vicinity of sharpcontrasts in axial direction. Of course, since the doseis distributed over the two overlapping measure-ments, there is no dose penalty.

  • 8/10/2019 manua siemes.pdf

    25/461

    Scan and Reconstruction

    25

    z-Sharp is used with all spiral modes in 0.6 mm colli-

    mation.

    Alternating focal spots of Straton Tube

    Alternating focal spots

    STRATON tube

    64-slice DAS

    Over-

    sampling

    Anode

    Heat

    Cooling oil

    Cathode

  • 8/10/2019 manua siemes.pdf

    26/461

    26

    Scan and Reconstruction

    UFC detector

    Siemens proprietary, high-speed Ultra Fast Ceramic(UFC) detector enables a virtually simultaneous read-out of two projections for each detector elementresulting in up to a 64-slice acquisition.

  • 8/10/2019 manua siemes.pdf

    27/461

    Scan and Reconstruction

    27

    Shaped Filter for Dose Reduction

    SOMATOM Definition provides an additional filter forall cardiac, head, and pediatric modes, which will beapplied automatically. The shaped filter reduces thedose by approx. 20%. It is recognizable by the reducedCTDIvolwhich is shown on the Routinetab card.

    Shaped filters

    ts

    Additionalshaped filter

    Permanentshaped filter

  • 8/10/2019 manua siemes.pdf

    28/461

    28

    Scan and Reconstruction

    Head:

    The scan field of view is limited to 300 mm as before,which is also shown graphically on the topogram.

    Cardiac:The scan field of view of 300 mm will be displayed bythe system for cardiac scanning. You can also recon-struct with a field of view greater than 300 mm. Inthis case the temporal resolution will slowlydecrease.

    Children:For pediatric examinations, in particular, it is neces-sary to keep the dose low. We therefore apply theshaped filter in all pediatric protocols. You can alsoreconstruct with a field of view greater than300 mm. In this case the temporal resolution willslowly decrease.

    ts

  • 8/10/2019 manua siemes.pdf

    29/461

    Scan and Reconstruction

    29

    The shaped filter is not applied in the following scan

    protocols: HeadAngio, HeadTrauma, HeadTrau-maSeq, and ChestPainECG. In these cases you canalso use a scan field of view of 500 mm.

    Scan Modes

    Sequential Scanning

    This is an incremental, slice-by-slice imaging mode in

    which there is no table movement during data acquisi-tion. A minimum interscan delay between each acqui-sition is required to move the table to the next sliceposition.

    Spiral Scanning

    Spiral scanning is a continuous volume imaging mode.The data acquisition and table movement are per-formed simultaneously for the entire scan duration. Atypical range can be acquired in a single breath hold.

    Each acquisition provides a complete volume data set,

    from which images with overlapping slices can bereconstructed at any slice position.

  • 8/10/2019 manua siemes.pdf

    30/461

    30

    Scan and Reconstruction

    Dynamic Multiscan

    Multiple continuous rotations at the same table posi-tion are performed for data acquisition. Normally, it isapplied for fast dynamic contrast studies, such assyngoBody Perfusion CTor syngoNeuro PerfusionCT. The increment represents the time between recon-

    structed images, for example, if the increment is1.0 sec., then one image will be reconstructed everysecond. The image order can be defined on the Recontab card.

    Dynamic Serioscan

    Dynamic serial scanning is performed without tablefeed. Dynamic serio can be used for dynamic evalua-tion such as Test Bolus. The image order can bedefined on the Recontab card.

    SOMATOM Definition permits the use of two X-raysources and two detectors at the same time.

    SOMATOM Definitions heart rate independent tempo-ral resolution of 83 ms allows a very robust scanningtechnique with any heart rate, for example, in coro-nary visualization and in functional analysis of theheart.

    Cardiac Scanning with Dual

    Source CT

  • 8/10/2019 manua siemes.pdf

    31/461

    Scan and Reconstruction

    31

    Clinical benefits

    Reliable imaging of all heart rates without betablockers.

    Industrys highest heart rate independent temporalresolution of 83 ms.

    Down to 42 ms temporal resolution for accurate

    functional imaging using bi-segment reconstruc-tion.

    Fast cardiac acquisition within the shortest breath-hold time.

    Improved plaque delineation and higher accuracy of

    in-stent imaging. Ability to scan arrhythmic patients.

    Highly accurate coronary lesion measurement.

    50% lower dose at typical heart rates compared withtodays most dose-efficient, Single Source CT scan-

    ners. The possibility to reconstruct the end-systolic phase.

    Makes patient positioning much easier due to the78 cm gantry bore, field of view, and 200 cm scanrange.

  • 8/10/2019 manua siemes.pdf

    32/461

    32

    Scan and Reconstruction

    Optimal heart rate occurs in the diastolic phase of the

    heart beat. The faster the heart rate, the shorter thediastolic phase becomes. With a Single Source CT scan-ner, the X-ray tube/detector system has to obtain pro-jection data through 180 degrees to take an image.

    With Dual Source CT, each of the two tube/detectorcombinations only needs to travel 90 degrees to

    acquire an image. Based on a 0.33 s rotation time, thisconcept provides an unprecedented temporal resolu-tion of 83ms irrespective of the heart rate.

    Scan Set-up

    Scans can be set up simply by selecting a predefinedexamination protocol. To repeat any mode, just clickthe chronicle with the right mouse button for repeat.To delete it, select cut. Each range name in the chron-icle can be easily changed before loading the scan pro-tocol.

    Multiple ranges can be run either automatically withauto range, which is denoted by a bracket connectingthe two ranges, or separately with a pauseinbetween.

    Up to 19 scan ranges can be combined to auto range.

  • 8/10/2019 manua siemes.pdf

    33/461

    Scan and Reconstruction

    33

    Acquisition, Slice Collimation and

    Slice Width

    Slice collimation is the slice thickness resulting fromthe effect of the tube-side collimator and the adaptivedetector array design. In Multislice CT, the Z-coverageper rotation is given by the product of the number of

    active detector slices and the collimation.

    Slice width is the FWHM (full width at half maximum)of the reconstructed image.

    With SOMATOM Definition, you select the slice collima-tion together with the slice width desired. The slicewidth is independent of pitch, i.e. what you select isalways what you get. Actually, you do not need to beconcerned about the pitch value any more; the soft-ware does it for you.

    By doubling the sampling, the slice width can bereduced. With SOMATOM Definition, it is possible toreconstruct 0.6 mm slices at any pitch ( 1.5) with bestimage quality. The improved sampling almost com-pletely removes, at any pitch, the so-called windmillartifacts frequently visible in Multislice CT images inthe vicinity of sharp contrasts in the axial direction. Ofcourse, since the dose is distributed over the two over-lapping measurements, there is no dose penalty. The

    Acq.(Acquisition) is displayed on the Examinationtask card.

    Acquisition = No. slices/ rotation x width of one slice

  • 8/10/2019 manua siemes.pdf

    34/461

    34

    Scan and Reconstruction

    RT images are reconstructed during acquisition (real-

    time display). Select Autoreconon the Reconstruc-tiontask card for each job, otherwise, you will have topress Reconfor each recon job separately.

    Spiral Mode

    *depends on the rotation time and scan protocol

    Acquisi-

    tion

    Collima-

    tion

    Slice width z-

    Sharp

    DS*

    64 x 0.6 0.6 0.6, 0.75,

    1, 1.5, 2, 3,

    4, 5, 6, 7,

    8, 10 mm

    X X

    20 x 0.6

    (Neuro

    mode)

    0.6 0.6, 0.75,

    1, 1.5, 2, 3,

    4, 5, 6, 7,

    8, 10 mm

    X

    24 x 1.2 1.2 1.5, 2, 3, 4,5, 6, 7, 8,

    10 mm

    X

  • 8/10/2019 manua siemes.pdf

    35/461

    Scan and Reconstruction

    35

    Sequence Mode

    *depends on the rotation time and scan protocol

    Acquisition Collimation Slice width DS*

    24 x 1.2 1.2 1.2, 2.4,

    4.8, 7.2,

    9.6,

    14.4 mm

    X

    12 x 1.2

    (Neuro

    mode)

    1.2 2.4, 4.8,

    7.2, 14.4

    6 x 3 3 3, 6, 9 mm X

    1 x 5

    (Neuro

    mode)

    5 5 mm

    1 x 10 10 10 mm

  • 8/10/2019 manua siemes.pdf

    36/461

    36

    Scan and Reconstruction

    Recon Increment

    The recon increment is the distance between thereconstructed images in Z direction. When the chosenincrement is smaller than the slice thickness, theimages are created with an overlap. This technique isuseful for reducing partial volume effect, giving you

    better detail of the anatomy and high quality 2D and3D post-processing.The increment can be freely adapted from0.1 - 10 mm.

    Pitch and Pitch adaptation

    You select the slice collimation together with the slicewidth desired. The slice width is independent of pitch,i.e. what you select is what you get. You no longer

    need to be concerned with the algorithm: the softwaredoes it for you.

    The pitch factor can be freely adapted from 0.45 - 2.0.Pitch values with a step width of 0.05 can be selectedfor all modes.

    Pitch= feed per rotation z-coverage

    Z-coverage = detector rows x collimated slice width

    Feed/Rotation = table movement per rotation

  • 8/10/2019 manua siemes.pdf

    37/461

    Scan and Reconstruction

    37

    Data from scans with pitch factors between 1.5 and 2

    can be reconstructed with a slice thickness 7 mm(collimation 1.2 mm) and a slice thickness 5 mm (col-limation 0.6 mm) only.

    For cardiac scanning SOMATOM Definition automati-cally adapts the pitch to the heart rate. Prior to thepatient's scan, the system monitors the heart rate and

    automatically calculates the best pitch value for thesubsequent scan. SOMATOM Definition increases thepitch with higher heart rates, resulting in a faster tablespeed and a corresponding reduction of dose expo-sure. The higher the heart rate, the less time isrequired for imaging the heart, and consequently the

    lower the dose needed.

    Diagram of the taken pitch, depending on the heartrate

    ts

    bpm (beats per minute)

    Pmono-segmentPbi-segment

    Pitchfactor

  • 8/10/2019 manua siemes.pdf

    38/461

    38

    Scan and Reconstruction

    To adapt the pitch automatically, the Estimated Heart

    Rate(Est.HR) is set to auto on the Triggertab card, bydefault.The heart rate can also be selected manuallyvia the drop-down menu.

    Observe the patients heart rate and select the corre-sponding entry. The system will choose a pitch valuefor the selected heart rate.

  • 8/10/2019 manua siemes.pdf

    39/461

    Scan and Reconstruction

    39

    Kernels

    There are five different types of kernels: Hstands forHead, Bstands for Body, Cstands for ChildHead, and Sstands for Special Application.

    The image sharpness is defined by the numbers: thehigher the number, the sharper the image; the lower

    the number, the smoother the image.

    The endings "s" or "f" indicate the mode: "s" stands forstandard mode, "f" for fast mode, and "h" for high res-olution modes.

    Head Kernels

    For soft tissue head studies, the standard kernel isH40s; softer images are obtained with H30s or H20s,H10s, sharper images with H50s. The kernels H21s,H31s, H41s yield the same visual sharpness as H20s,H30s or H40s respectively. The image appearance,however, is more acceptable due to a "fine-grained"noise structure; quite often, the low contrast detect-ability is improved by using H31s, H41s instead ofH30s, H40s.

    In emergency examinations, kernels H22s, H32s, andH42s can be used because they allow fast reconstruc-tion (FR) and easy patient positioning (50 cm FoV). Toensure best performance, special online bone correc-tion (PFO) is not used.

  • 8/10/2019 manua siemes.pdf

    40/461

    40

    Scan and Reconstruction

    High Resolution head studies should be performed

    with H60s, H70h (for example, for dental and sinuses).It is essential to position the area of interest in the cen-ter of the scan field.

    Kernel Description

    H10f, H10s very smooth

    H20f, H20s smoothH21f, H21s smooth +

    H22f, H22s smooth FR

    H30f, H30s medium smooth

    H31f, H31s medium smooth +

    H32f, H32s medium smooth FRH37f, H37s medium

    H40f, H40s medium

    H41f, H41s medium +

    H42f, H42s medium FR

    H45f, H45s medium

    H47f, H47s medium

    H48f, H48s medium

    H50f, H50s sharp

    H60f, H60s medium

    H70h very sharp

  • 8/10/2019 manua siemes.pdf

    41/461

    Scan and Reconstruction

    41

    Body Kernels

    As standard kernels for body tissue studies B30s orB40s are recommended; softer images are obtainedwith B20s or B10s (extremely soft). The kernels B31sor B41s have about the same visual sharpness as B30sor B40s respectively. The image appearance, however,is more acceptable due to a "fine-grained" noise struc-

    ture. Quite often, the low contrast detectability isimproved by using B31s, B41s instead of B30s, B40s.

    For higher sharpness, as is required, for example, inpatient protocols for cervical spine, shoulder, extremi-ties and thorax, the kernels B50s, B60s, B70s, B75h,B80s are available.

    The kernels B25s (smooth++) and B26s (HeartViewsmooth) offer the resolution of a standard kernel forbody tissue studies implying an advanced noise reduc-tion algorithm. The noise level will be reduced to val-ues comparable with an extremely soft kernel butkeeping the standard sharpness at contours. B25 and

    B26 improve the image quality of, for example, MIPswithout the drawbacks of the loss of spatial resolutionby simply using an extremely soft kernel.

  • 8/10/2019 manua siemes.pdf

    42/461

    42

    Scan and Reconstruction

    For heart reconstructions, as required, for example, in

    MIPs for coronary CTAs, the kernel B26s is available.

    Kernel Description

    B10f, B10s very smooth

    B18f, B18s very smooth

    B20f, B20s smooth

    B25f, B25s smooth ++

    B26s HeartView smooth

    B30f, B30s medium smooth

    B31f, B31s medium smooth +

    B35f, B35s HeartView medium

    B36f HeartView mediumB40f, B40s medium

    B41f, B41s medium+

    B45f, B45s medium

    B46f, B46s HeartView sharp

    B50f, B50s medium sharp

    B60f, B60s sharp

    B70f, B70s very sharp

    B75h, B75f very sharp

    B80f, B80s ultra sharp

  • 8/10/2019 manua siemes.pdf

    43/461

    Scan and Reconstruction

    43

    ChildHead Kernels

    For head scans of small children, the kernels C20s,C30s (for example, for soft tissue studies) and C60s(for example, provided for sinuses) should be choseninstead of the "adult" head kernels H20s, H30s, andH60s.

    Special Applications

    Select S80s and S90s for special patient protocols andHigh Resolution bone studies (S90s).

    Note:

    For 3D studies only use kernel B10s with at least 50%overlapping for image reconstruction.

    For vascular studies use B25s and for cardiac studiesuse kernel B26s for high resolution with reducednoise.

    Do not use kernels for body parts other than whatthey are designed for.

    Kernel Description

    C20f, C20s smooth

    C30f, C30s medium smooth

    C60s sharp

    Kernel Description

    S80f, S80s Shepp-Logan with notch filter

    S90f, S90s Shepp-Logan without notch filter

  • 8/10/2019 manua siemes.pdf

    44/461

    44

    Scan and Reconstruction

    Topogram Kernels

    Kernel Description

    T10s smooth

    T20s standard

    T21s standard

    T80s sharp

    T81s sharp

    T90s ultra sharp

  • 8/10/2019 manua siemes.pdf

    45/461

    Scan and Reconstruction

    45

    Extended FoV

    SOMATOM Definition provides an extended field ofview. The range can be individually adapted by theuser from 50 cm up to 78 cm.

    To use this feature you have to select the extendedFoVcheckbox on the Recontab card. The default set-

    ting is 65 cm, but can be modified.

    Extended FoVcan be used with each scan protocol.

    The extended FoV value should be adapted carefullyto the exact patient size in order to achieve best possi-ble image quality outside the standard scan field.

  • 8/10/2019 manua siemes.pdf

    46/461

  • 8/10/2019 manua siemes.pdf

    47/461

    Scan and Reconstruction

    47

    Image FiltersIf you use kernels, the images are reconstructed againwith the selected kernel value. If you use image filters,the images are not reconstructed again and the resultis much quicker.

    Three different filters are available:

    "LCE": The Low-contrast enhancement filterenhances low-contrast detectability and reduces theimage noise.

    Similar to reconstruction with a smoother kernel

    Reduces noise Enhances low-contrast detectability

    Adjustable in four steps

    Automatic post-processing

    Image taken without

    the LCE filter

    Image taken with the LCE

    filter

  • 8/10/2019 manua siemes.pdf

    48/461

    48

    Scan and Reconstruction

    "HCE": The High-contrast enhancement filter

    enhances high-contrast detectability. It increases theimage sharpness, similar to reconstruction with asharper kernel.

    Increases sharpness

    Faster than raw data reconstruction

    Enhances high-contrast detectability

    Automatic post-processing

    "ASA": Advanced Smoothing Algorithmreduces noise in soft tissues while edges with highcontrast are preserved.

    Reduces noise without blurring edges Enhances low-contrast detectability

    Individually adaptable

    Automatic post-processing

    Image taken without

    the HCE filter

    Image taken with the

    HCE filter

  • 8/10/2019 manua siemes.pdf

    49/461

    Scan and Reconstruction

    49

    Head Imaging

    Head Modes

    For a given (eff.) mAs, the CTDI values (16 cm CTDIphantom) are lower for a lower tube voltage.

    kV Increase of

    mAs

    Reduction of

    CTDI vol

    80 30% 15%

    100 20% 13%120 15% 12%

  • 8/10/2019 manua siemes.pdf

    50/461

    50

    Scan and Reconstruction

    Neuro Modes

    In addition to the standard collimations, SOMATOMDefinition provides a special mode that is optimized forNeuro applications. Excellent low contrast, and detailresolution are achieved.

    For sequence scans, a 12 times 1.2 mm acquisition and

    a single 5 mm slice with optimized dose utilization andimage quality can be acquired. A 20 x 0.6 acquisitionmode is provided for spiral scans. Both approachesshow a minimized partial volume effect, i.e. low levelof artifacts in the base of the skull or near vertebralbodies, as 0.6 mm detector rows are used and the nar-

    row collimation reduces scattered radiation.Three scan protocols are predefined for adults and chil-dren:

    HeadNeurousing an acquisition of 20 x 0.6 mmand

    HeadNeuroSequsing an acquisition of 1 x 5 mm.

    SpineNeurousing an acquisition of 20 x 0.6 mm

    We recommend using these special protocols for dedi-cated Neuro examinations.

    For fast standard examinations such as rule out ofhemorrhage or ischemia, the "Routine" protocol shouldbe used.

  • 8/10/2019 manua siemes.pdf

    51/461

    Scan and Reconstruction

    51

    Automatic Bone Correction

    The head protocols provide significant improvementsregarding image quality for heads. An automatic bonecorrection algorithm has been included in the standardimage reconstruction. Using a new iterative tech-nique, typical artifacts arising from the beam harden-

    ing effect, for example, Hounsfield bar, are minimizedwithout additional post-processing. This advancedalgorithm produces excellent images of the posteriorfossa, but also improves head image quality in general.Bone correction is activated automatically for bodyregion "Head". The reconstruction algorithm for "Head"also employs special adaptive convolution kernels that

    help to improve the sharpness to noise ratio. More pre-cisely, anatomic contours are clearly displayed whilenoise is suppressed at the same time without causinga blurring of edges.

    Head image withoutcorrection.

    Head image withcorrections.

  • 8/10/2019 manua siemes.pdf

    52/461

    52

    Scan and Reconstruction

    Positioning

    In order to optimize image quality versus radiationdose, scans in the "Head" body region are providedwith a maximum scan field of 300 mm with respect tothe isocenter. No recon job with a field of view exceed-ing those limits will be possible. Therefore, patient

    positioning has to be performed accurately to ensurethe skull is centered in the isocenter using the intersec-tion of the lasers.

    correct positioning wrong positioningof the head of the head

    For trauma examinations of the head two protocols areprovided which you will find in the Specialsfolder:

    HeadTrauma

    HeadTraumaSeq.

    The scan protocols enable you to utilize the full 50 cmFoV, resulting in easier patient positioning for traumaexaminations. To ensure the highest performance,dedicated online bone correction (PFO) is not used.

  • 8/10/2019 manua siemes.pdf

    53/461

    Scan and Reconstruction

    53

    Automatic Isocenter Adaptation

    Automatic adaptation of the table height allows you toposition the object exactly in the isocenter. To achievebest image quality, it is necessary to position thepatient exactly in the isocenter. You see the isocenterin the topogram graphical display and can adapt the

    table height after scanning the topogram. As the tablemoves, the graphic position is also updated.

  • 8/10/2019 manua siemes.pdf

    54/461

    54

    Scan and Reconstruction

    The table position can be adapted in two ways:

    Click on the Feed Up, Feed Downbuttons on theRoutinetab card.

    Enter the new table height manually on the Routinetab card.

    Note:

    Adaptation is only possible with lateral topograms.

    If the table has been vertically repositioned based onthe active field of view in a topogram, the topogramis still valid for scan planning and reconstructions.

    Hint:We recommend using the automatic isocenteradaptation, especially for head examinations, because

    of the image quality.

  • 8/10/2019 manua siemes.pdf

    55/461

    Scan and Reconstruction

    55

  • 8/10/2019 manua siemes.pdf

    56/461

    56

    Dose Information

    CTDIW

    and CTDIVol

    The average dose in the scan plane is best described bythe CTDIWfor the selected scan parameters. The CTDIWis measured in dedicated plastic phantoms 16 cmdiameter for head and 32 cm diameter for body (as

    defined in IEC 60601 2 44). For scan modes with z-Sharp the CTDI100 is calculated using the single num-ber of tomographic sections (not doubled by z-Sharp)to remain within the terms of IEC 60601-2-44. The z-coverage with and without z-Sharp is the same and sois the dose. This dose index gives a good estimation ofthe average dose applied in the scanned volume, aslong as the patient size is similar to the size of therespective dose phantoms.

    Since the body size can be smaller or larger than32 cm, the CTDIWvalue displayed can deviate from thedose in the scanned volume.

    The CTDIWdefinition and measurement are based onsingle axial scan modes. For clinical scanning, i.e.scan-ning of entire volumes in patients, the average dosewill also depend on the table feed between axial scansor the feed per rotation in spiral scanning. The dose,expressed as the CTDIW, must therefore be corrected by

    the pitch factor of the spiral scan or an axial scan seriesto describe the average dose in the scanned volume.

  • 8/10/2019 manua siemes.pdf

    57/461

  • 8/10/2019 manua siemes.pdf

    58/461

    58

    Dose Information

    The CTDIw value does not provide the entire informa-

    tion of the radiation risk associated with CT examina-tion. For this purpose, the concept of the EffectiveDose was introduced by ICRP (International Commis-sion on Radiation Protection). The effective dose isexpressed as a weighted sum of the dose applied notonly to the organs in the scanned range, but also to the

    rest of the body. It could be measured in whole bodyphantoms (Alderson phantom) or simulated withMonte Carlo techniques.

    The calculation of the effective dose is rather compli-cated and has to be performed by sophisticated pro-grams. These have to take into account the scan

    parameters, the system design of the individual scan-ner, such as X-ray filtration and gantry geometry, thescan range, the organs involved in the scanned rangeand the organs affected by scattered radiation. Foreach organ, the respective dose delivered during theCT scanning has to be calculated and then multipliedby its radiation risk factor. Finally, the weighted organ

    dose numbers are added up to get the effective dose.

    The concept of effective dose allows the comparison ofradiation risk associated with different CT or X-rayexams, i.e. different exams associated with the sameeffective dose would have the same radiation risk forthe patient. It also allows a comparison of the applied

    X-ray exposure to the natural background radiation,for example, 2 3 mSv per year in Germany.

  • 8/10/2019 manua siemes.pdf

    59/461

  • 8/10/2019 manua siemes.pdf

    60/461

    60

    Dose Information

    Effective mAsIn sequential scanning, the dose (Dseq) applied to thepatient is the product of the tube current-time (mAs)and the CTDIw per mAs:

    In spiral scanning, however, the applied dose (Dspiral)is influenced by the conventional mAs (mA x Rot Time)and additionally by the pitch factor. For example, if aMultislice CT scanner is used, the actual dose appliedto the patient in spiral scanning will be decreasedwhen the pitch factor is greater than 1, and increasedwhen the pitch factor is less than 1 (for constant mA).Therefore, the dose in spiral scanning has to be cor-rected by the pitch factor:

    To simplify this task, the concept of the "effective" mAswas introduced with SOMATOM Multislice scanners.

    The effective mAs takes into account the influence ofpitch on both the image quality and dose:

    Dseq

    = DCTDIw

    x mAs

    Dspiral=(DCTDIwx mA x Rot Time)

    Pitch factor

    Effective mAs = mAsPitch factor

  • 8/10/2019 manua siemes.pdf

    61/461

  • 8/10/2019 manua siemes.pdf

    62/461

  • 8/10/2019 manua siemes.pdf

    63/461

    Dose Information

    63

    CARE Dose 4DCARE Dose 4D is an automated exposure control,which ensures constant diagnostic image quality overall body regions at the lowest possible dose.

    CARE Dose 4D combines three different adaptationmethods to optimize image quality at the lowest doselevel:

    Automatic adaptation of the tube current to thepatient size

    Automatic adaptation of the tube current to the

    attenuation of the patients long axis, the so-called z-axis.

    Automatic adaptation of the tube current to theangular attenuation profile measured online foreach single tube rotation, the so-called angle modu-lation.

    Based on a single a.p. or lateral topogram, CARE Dose4D determines the adequate mAs level for every sec-tion of the patient. Based on these levels, CARE Dose4D modulates the tube current automatically duringeach tube rotation according to the patients angularattenuation profile. Thus, the best distribution of dose

    along the patients long axis and for every viewingangle can be achieved.

  • 8/10/2019 manua siemes.pdf

    64/461

    64

    Dose Information

    Based on a user defined Image Quality Reference

    mAs, CARE Dose 4D automatically adapts the (eff.)mAs to the patient size and attenuation changeswithin the scan region. With the setting of the ImageQuality Reference mAsyou can adjust image quality(image noise) to the diagnostic requirements and theindividual preference of the radiologist.

    Note: The Image Quality Reference mAs should notbe adjusted to the individual patient size!

  • 8/10/2019 manua siemes.pdf

    65/461

    Dose Information

    65

    How does CARE Dose 4D work?

    CARE Dose 4D combines two types of tube currentmodulation:

    1.Axial tube current modulation:

    Based on a single topogram (a.p. or lateral) the atten-

    uation profile along the patients long axis is measuredin the direction of the projection and calculated for theperpendicular direction by a sophisticated algorithm.

    Example of lateral and a.p. mA and mAs profile evalu-ated from an a.p. topogram.

    Based on these attenuation profiles, axial tube currentprofiles (lateral and a.p.) and the resulting eff. mAs forevery table position are calculated. The correlationbetween attenuation and tube current is defined by ananalytical function which results in the optimum doseand image noise in every slice of the scan.

    mA,mAs

    mAs meanmAs lateralmA a.p.

    max

    table position (mm)

  • 8/10/2019 manua siemes.pdf

    66/461

    66

    Dose Information

    This correlation is based on a clinical assessment of

    diagnostical image quality. The result was that forgood diagnostical image quality at reasonable lowdose level the image noise does not need to be con-stant for all body sizes and body regions.

    For every type of protocol a typical X-ray attenuation ofa standard sized patient is internally stored and a pre-

    defined Quality reference mAs value is related to thisreference X-ray attenuation. You can modify the imagequality by adjusting the Quality reference mAs.

    For an advanced adjustment, the slope of the right andthe left branch of the curve may be adjusted in theExamination Configurationdialog box to achieve a

    stronger or weaker increase or decrease of eff. mAswith body size or body region.

    weak average strong

    slim/decrease A B C

    obese/increase F E D

    Reduced dose level rbased on topogram

    Average X-ray attenuation (lo/l) resp. Body Size

    Quality ref. mAs

    Constant Image Noise

    ReferenceA

    ttenuation

    D

    Clinical reasonable mAs

    E

    F

    BC

    A

    eff.

    mAs

  • 8/10/2019 manua siemes.pdf

    67/461

    Dose Information

    67

    In the example, the reference attenuation is that for athorax protocol (600) and a Quality reference mAs of100 is selected. Based on this pair of values the eff.mAs is adapted to the body size according to the rela-tion shown in the graph. The resulting eff. mAs of thisexample for different body size and body regions islisted in the table.

    Marker StandardPatient

    (adult/75kg)

    ObesePatient

    X-rayAttenuation

    eff.mAs

    1 Neck 100 45

    2 Neck 360 80

    3 Thorax 600 100

    4 Thorax 820 110

    5 Abdomen 1000 120

    6 Abdomen 1950 150

  • 8/10/2019 manua siemes.pdf

    68/461

    68

    Dose Information

    2.Angular tube current modulation:

    Based on the above described axial eff. mAs profile,the tube current is modulated during each tube rota-tion. Therefore the angular attenuation profile is mea-sured automatically during the scan and the tube cur-rent is modulated accordingly in real time to achievean optimum distribution of the X-ray intensity for

    every viewing angle.

    Tube current profile along a spiral scan with CAREDose 4D: Combination of tube current adaptationalong the body axis and tube current modulation overthe rotation angle. Every single oscillation corre-sponds to half a tube rotation. The variation of the

    oscillation's mean value over the table position showsthe variation of eff. mAs.

    tubecurre

    ntmA

    table position (mm)

  • 8/10/2019 manua siemes.pdf

    69/461

    Dose Information

    69

    CARE Dose 4D allows you to reduce the patient dose

    significantly (locally up to nearly 70%) while keepingthe diagnostic image quality constant or even improv-ing it and enabling longer scan ranges.

    The mAs value displayed in the user interface and inthe patient protocol is the mean (eff.) mAs value forthe scan range.

    The mAs value recorded in the images is the local (eff.)mAs value.

    Special Modes of CARE Dose 4D

    For certain examination protocols, CARE Dose 4D usesmodified tube current modulation, to meet specificconditions, for example:

    for Adult Head protocols the tube current is adaptedto the variation along the patients long axis and notto the angular attenuation profile.

    for Extremities, NeuroPCT and BodyPCT, and otherspecial protocols (indicated as CARE Dose), onlyangular tube current modulation is supported.

    for Cardio protocols the mAs/rot setting is adjusted tothe patient size and not modulated during the scan,except if ECG pulsing is switched on.

  • 8/10/2019 manua siemes.pdf

    70/461

    70

    Dose Information

    Scanning with CARE Dose 4D

    If the settings of Image Quality Reference mAs are cor-rectly predefined*, no further adjustment of the tubecurrent is required to perform a scan.

    CARE Dose 4D automatically adapts the tube current todifferent patient sizes and anatomic shapes, but it

    widely ignores metal implants.

    Note: Otherwise the magnification of the topogramwould be distorted which would lead to an under-estimation or overestimation of the required eff.mAs.

    For an accurate mAs adaptation to the patients sizeand body shape with CARE Dose 4D, the patient shouldbe carefully centered in the scan field.

    When using protocols with CARE Dose 4D for bodyregions other than those they are designed for, theimage quality should be carefully evaluated.

    As CARE Dose 4D determines the (eff.) mAs for everyslice of the topogram, a topogram must be obtainedfor use of CARE Dose 4D.

    *For Siemens scan protocols of SW version syngo CT2007A, the settings of CARE Dose 4D are already pre-defined but may be changed to meet the customerspreference of image quality (image noise).

  • 8/10/2019 manua siemes.pdf

    71/461

    Dose Information

    71

    Outside the topogram range, CARE Dose 4D will con-

    tinue the scan with the last available topogram infor-mation. Without a topogram, CARE Dose 4D cannot beswitched on. Repositioning of the patient on the tableand excessive motion of the patient must be avoidedbetween the topogram and the scan. If two topogramsof the same projection exist for one scan range, the

    last acquired will be used for determining the (eff.)mAs. If a lateral and a.p. topogram exist for one scanrange, both will be used for determining the (eff.)mAs.

  • 8/10/2019 manua siemes.pdf

    72/461

    72

    Dose Information

    After the topogram has been scanned, the (eff.) mAs

    value in the Routinetab card displays the mean (eff.)mAs estimated by CARE Dose 4D based on the topo-gram*. After the scan has been performed this value isupdated to the mean (eff.) mAs that was applied. Thevalues may differ slightly due to the online modulationaccording to the patients angular attenuation profile.

    *When tuning the CARE Dose 4D parameter setting tothe individual preference for image quality, we rec-ommend keeping track of this value and comparingit with the values used without CARE Dose 4D.

  • 8/10/2019 manua siemes.pdf

    73/461

    Dose Information

    73

    The Quality reference mAsvalue is displayed on the

    Scantab card. This defines the overall image quality ofthe scan protocol currently being used. This value canbe adapted for each protocol according to the usersindividual requirements of image quality. Here you canalso view the effective mAs value that the system isgoing to use for the current scan range.

    You can also deselect CARE Dose4D on this tab card.

  • 8/10/2019 manua siemes.pdf

    74/461

  • 8/10/2019 manua siemes.pdf

    75/461

  • 8/10/2019 manua siemes.pdf

    76/461

    76

    Dose Information

    if the automatic dose increase for obese patients (or

    patient sections) has to be more moderate than thepreset (choose obese: weak increase), resulting inmore image noise and a lower dose for thoseimages.

    if the automatic dose decrease for slim patients (orpatient sections) has to be stronger than the preset

    (choose slim: strong decrease), resulting in moreimage noise and a lower dose for those images.

    if the automatic dose decrease for slim patients (orpatient sections) has to be more moderate than thepreset (choose slim: weak decrease), resulting inless image noise and a higher dose for those images.

    On the Patienttab card you can adjust the image qual-

    ity (for more information see chapter How does CAREDose 4D work).

    Note: Changing this adaptation strength effects allprotocols!

  • 8/10/2019 manua siemes.pdf

    77/461

    Dose Information

    77

    Activating and Deactivating

    CARE Dose 4D may be activated or deactivated for thecurrent scan in the Scan tabcard. If CARE Dose 4D isactivated as default, the Image Quality ReferencemAsvalue is set to the default value of the protocol.After deactivating CARE Dose 4D, the Image Quality

    Reference mAsis dimmed and the (eff.) mAs valuehas to be adjusted to the individual patients size! IfCARE Dose 4D is switched on again, the Image QualityReference mAs is reactivated. Note that the last settingof the Image Quality Reference mAs or the (eff.) mAswill be restored when you switch from and back toCARE Dose 4D usage. The default activation state of

    CARE Dose 4D may be set in the Scan Protocol Man-ager. CARE Dose 4D must be selected (column CAREDose type). The corresponding column for activatingCARE Dose 4D is called CARE Dose (4D), with possibledefault onor off.

  • 8/10/2019 manua siemes.pdf

    78/461

    78

    Dose Information

    If you are unsure about the correct Image Quality Ref-

    erence mAs value, follow this simple procedure:

    Enter the (eff.) mAs value used for that type of pro-tocol without CARE Dose 4D.

    There is a simple way of ascertaining what eff. mAsCARE Dose 4D will use along the scan range: When

    the topogram is complete shrink the scan range toit's minimum. As you move this small box over thetopogram you can see how the eff. mAs displayed inthe Routine and Scantab card varies along thepatient's body.To achieve a certain eff. mAs at a patient's particularbody region you can move the small scan range to

    this position and then adjust the Quality referencemAs so that the displayed eff. mAs value is asdesired. After resizing the scan range to the rangefor the examination, carefully observe the displayedmean eff. mAs. After the subsequent scan is com-pleted inspect the image quality to ensure that the

    chosen Quality reference mAs is the right value. With that setting perform the first scan and carefully

    inspect the image quality. In that first step the dosemay not be lower than without CARE Dose 4D butwill be well adapted to the patients attenuation,resulting in improved image quality.

    Starting from that setting, reduce the Image QualityReference mAs step by step to meet the necessaryimage quality level.

    Store the scan protocol with the adapted image qual-ity reference mAs.

  • 8/10/2019 manua siemes.pdf

    79/461

    Dose Information

    79

    Additional Important Information

    For ideal dose application it is very important to posi-tion the patient in the isocenter of the gantry.

    Example of an a.p. topogram:

    Patient is positioned in the isocenter - optimal doseand image quality

    Patient is positioned too high - increased mAs

    Patient is positioned too low - reduced mAs andincreased noise

    Detector

    X-ray tube

    Patient

    (centered)

    Patient(centered)

    X-ray tube

    Detector

    X-ray tube

    Detector

    Patient

    (not centered)

    Patient

    X-ray tube

    (not centered)

    Detector

  • 8/10/2019 manua siemes.pdf

    80/461

    80

    Dose Information

    Three different table positions are displayed in the fol-

    lowing patient protocols according to the topograms:

    The first scan was performed in the isocenter. Here areasonable total mAs value was achieved.

    The second topogram scan was performed abovethe isocenter. Here the mAs value was increased.

    The third topogram scan was performed below theisocenter. The mAs value was significantly lower andthe resulting images had increased image noise.

  • 8/10/2019 manua siemes.pdf

    81/461

    Dose Information

    81

    Clinical Benefits of CARE Dose 4D

    The same scan protocol can be used for slim andobese patients.

    An optimal image quality in every slice can beachieved at the lowest possible dose level.

    A patient dose reduction of up to 67% locally can beachieved depending on the scan protocol.

  • 8/10/2019 manua siemes.pdf

    82/461

    82

    Dose Information

    100kV ProtocolsThe system offers a spectrum of four kV settings(80 kV, 100 kV, 120 kV, and 140 kV) for individualadaptation of the patient dose in pediatric scans andfor optimization of the contrast-to-noise ratio in con-

    trast-enhanced CT angiographic studies.

    In contrast enhanced studies, such as CT angiographicexaminations, the contrast-to-noise ratio for a fixedpatient dose increases with decreasing tube voltage.As a result, to obtain a given contrast-to-noise ratio,the patient dose can be reduced by choosing lower kV-

    settings. This effect is even more pronounced forsmaller patient diameters. It can be demonstrated byphantom measurements using small tubes filled withdiluted contrast agent embedded in plastic phantomswith different diameters. The iodine contrast-to-noiseratio for various kV-settings is depicted in the followingtable as a function of the phantom diameter. Com-pared with a standard scan with 120 kV, the same con-trast-to-noise ratio in a 24 cm phantom, correspondingto a slim adult, is obtained with 0.5 times the dose for80 kV (1.5 times the mAs) and 0.7 times the dose (1.1times the mAs) for 100 kV. Ideally, 80 kV should beused for the lowest patient dose. In practice, however,

    the use of 80 kV for larger patients is limited by theavailable mA-reserves of the X-ray generator.

    In these patients, 100 kV is a good compromise andthe preferable choice for CTA examinations.

  • 8/10/2019 manua siemes.pdf

    83/461

    Dose Information

    83

    Iodine contrast-to-noise ratio as a function of thephantom diameter for kV-settings at a constant dose(CTDIw in these phantoms).

    80 kV

    120 kV

    100 kV

    140 kV

    IodineSignal/

    Noise

    Phantom Diameter, cm

  • 8/10/2019 manua siemes.pdf

    84/461

    84

    Dose Information

    Results of iodine contrast, noise and dose measure-ments for different kV settings and phantom sizes. Rel-

    ative dose numbers (CTDIw in the respective phan-tom) and mAs settings needed for a certain contrast-to-noise ratio in the center of the 16 cm and 24 cmplastic phantoms. The required dose for the same con-trast-to-noise ratio is significantly lower for lower kVvalues.

    80 kV 100 kV 120 kV 140 kV

    Relative dose,

    24 cm

    0.49 0.69 1.0 1.49

    Relative dose,

    16 cm

    0.44 0.68 1.0 1.43

    Relative mAs

    setting

    150 110 100 100

  • 8/10/2019 manua siemes.pdf

    85/461

    Dose Information

    85

  • 8/10/2019 manua siemes.pdf

    86/461

    86

    Workflow Information

    WorkStream4D

    Recon Jobs

    In the Recontab card, you can define up to eightreconstruction jobs for each range with different

    parameters either before or after you acquire the data.When you click on Recon, these jobs are performedautomatically in the background. If you want to addmore than eight recon jobs, simply click the icon for analready completed recon job in the chronicle with theright mouse button and select delete recon job.

    Another recon job will now become available on theRecontab card.

    Note: What you delete is just the job from the display,not the images that have been reconstructed. Oncereconstructed, these completed recon jobs stay in thebrowser until they are deleted from the local database.

    You can also reconstruct images for all scans per-formed by not selecting any range in the chronicleprior to clicking Recon.

    Another item you will find in the right mouse menu iscopy/replace recon parameters. This function isavailable for spiral scans only.

    The main goal is to support the transfer of volumeparameters between oblique recon jobs of rangeswhich mainly cover the same area, for example, twospiral scans with/without contrast media.

  • 8/10/2019 manua siemes.pdf

    87/461

    Workflow Information

    87

    3D Recon

    3D Recon allows you to perform oblique and/ or dou-ble-oblique reconstructions in any user-defined direc-tion directly after scanning.

    No further post-processing or data loading is needed.The high-quality SPO (spiral oblique) images are calcu-

    lated using the systems raw data.

    Key Features

    Reconstruction of axial, sagittal, coronal, andoblique / double oblique segments

    3 planning segments in the 3 standard orientations

    (coronal, axial, sagittal) Image types for planning MPR Thick (3 mm),

    MIP Thin (10 mm)

    Field of view and reference image definition possiblein each planning segment

    Asynchronous reconstruction (several reconstruc-tion jobs are possible in the background, axial andnon-axial)

    Workstream 4D performs reconstructions on thebasis of CT raw data

    If the raw data is saved you can start the 3D recon-struction on your syngo CT Workplace.

  • 8/10/2019 manua siemes.pdf

    88/461

    88

    Workflow Information

    Workflow Description

    WorkStream 4D improves your workflow whenevernon-axial images of a CT scan are required, for exam-ple, examinations of the spine.

    3D reconstructions are possible:

    spiral scan is needed

    as soon as one scan range is finished and at least oneaxial reconstruction job has been performed (RTimages).

    Select a new recon job and mark Recon Job Type 3Don the Recontab card. The first recon job that is suit-able for the 3D reconstruction is used as Availableplanning volumes.

  • 8/10/2019 manua siemes.pdf

    89/461

    Workflow Information

    89

    Additional Important Information

    Pitch factor for 3D Recon

    For reconstruction of 3D recon jobs the maximumpitch factor is 1.5.If the pitch factor is > 1.5 a message window informsyou that this 3D recon job cannot be started and may

    be deleted. In this case use the standard 3Dtask cardwith an axial image series for reconstruction.

  • 8/10/2019 manua siemes.pdf

    90/461

    90

    Workflow Information

    Three planning segments in perpendicular orienta-

    tions will appear in the upper screen area. You canchoose between MPR Thick(3 mm) and MIP Thin(10 mm) as the image type for your planning volumeusing the relevant buttons.

    In each segment you will find a pink rectangle whichrepresents the boundary of the result images. The

    image with the right down marker represents the fieldof view (FoV) of the result images (viewing direction).

    Right

    down

    marker

    Reference lines

  • 8/10/2019 manua siemes.pdf

    91/461

    Workflow Information

    91

    The rectangle with the grid represents the reference

    image (topogram) which is added to the topogramseries including the reference lines after reconstruc-tion.

    Topographics

    indicator

    Recon area

    Reference lines

  • 8/10/2019 manua siemes.pdf

    92/461

    92

    Workflow Information

    Preview Image

    A preview of the actual FoV is now available.

    After pressing the button Preview Imagethe actualFoV to be reconstructed will be displayed.

    Clicking again on the button will deactivate the pre-view image and display the whole reference image

    again.

    Double clicking into the FoV image will activate ordeactivate the Preview Imagefunction as well.

    If the Preview Imagefunction is active and you moveor rotate the box, or change the recon begin and end

    position, the Preview image in the FoV segment will beupdated accordingly.

  • 8/10/2019 manua siemes.pdf

    93/461

    Workflow Information

    93

    Depending on the desired resultant images, choose

    axial, coronal, sagittal, or oblique recon axis.

    1. Sagittal/Coronal Reconstructions

    Adjust the field of view size to your needs.

    It is only possible to reconstruct images with asquared matrix.

    2. Oblique/Double-oblique Reconstructions

    If you want to define the orientation of the resultimages independently of the patients axis:

    Enable theFree View Modeand rotate the refer-ence lines in the three segments until the desired

    image orientation is displayed. The vertical and hor-izontal lines are always perpendicular to each other.With the default orientation button you can reset theimage orientation at any time.

    It is only possible to reconstruct images with asquared matrix.

    Set the field of view to the active segment by clickingthe Set FoV button. The result images will then beorientated as in the FoV segment. You can adjust theextension perpendicular to the field of view in thesame way in the other two segments.

  • 8/10/2019 manua siemes.pdf

    94/461

  • 8/10/2019 manua siemes.pdf

    95/461

    Workflow Information

    95

    Additional Information

    As soon as you define a new recon range, all reconranges will be shown in the topo segment. The twonumbers on the right-hand side at the beginning ofeach recon range indicate the recon job the rangebelongs to. The first number stands for the scan range,the second number stands for the recon job to which

    the range belongs. If no recon job is pending, only thescan ranges are shown in the topo segment. Only onenumber on the right-hand side at the beginning ofeach scan range indicates which scan the rangebelongs to.

    If the first recon job is saved as an Obliquerecon job,

    RTD images are displayed after scanning and theExaminationtask card is automatically switched to3D reconstruction

    Patient Browser:for each double oblique recon job, one series isadded in the Patient Browser.

    If Auto Reference Linesis selected the correspond-ing reference image is added to the 3D recon series.

    All reconstructions are performed in the background

    Do not use high resolution images

    Do not use extended FoV

    If no entry is selected in the chronicle, all open

    reconstructions are automatically reconstructed. If Autoreconis selected on the Recontab card, this

    recon job (axial and oblique) will be automaticallyreconstructed after scanning.

  • 8/10/2019 manua siemes.pdf

    96/461

  • 8/10/2019 manua siemes.pdf

    97/461

    Workflow Information

    97

    Case Examples

    Some scan protocols are supplied with predefinedoblique reconstructions. These protocols aremarked with the suffix VOL.

    Coronal and sagittal reconstruction of the spine:

    Scan a topogram

    Plan your axial spiral scan range Reconstruction of the spiral images (RT images)

    Select Recon job Typesagittal/coronal

    Select the axial image segment

    Press button Set FoV

    Adjust the FoV to your needs

    Define your desired reconstruction parameters(for example, image type SPO)

    Start reconstruction

    Repeat the reconstruction steps for the otherorientation (sagittal/coronal)

  • 8/10/2019 manua siemes.pdf

    98/461

    98

    Workflow Information

    Oblique reconstruction of the sinuses:

    Scan a topogram Plan your axial spiral scan range

    Reconstruction of the spiral images(RT images)

    Select Recon job Typeoblique

    Select the sagittal image segment

    Enable Free Mode Rotate the reference lines until the best view of

    the sinuses is displayed in one of the other seg-ments

    Select this segment and press the Set FoVbutton

    Adjust the FoV to your needs

    Define your desired reconstruction parameters(e.g., image type SPO)

    Start reconstruction

  • 8/10/2019 manua siemes.pdf

    99/461

  • 8/10/2019 manua siemes.pdf

    100/461

    100

    Workflow Information

    Workflow

    Patient Position

    A default patient position can be linked and stored toeach scan protocol. The SIEMENS default protocols are

    already linked to a default patient position (Head first -supine).

    If a scan protocol is selected and confirmed in thePatient Model Dialog, the linked patient positionstays active until the user changes it, even if a scan pro-tocol with a different patient position is selected.

    Auto Reference Lines

    The Auto Reference linessettings defined in thePatient Model Dialogcan be linked and saved to eachscan protocol.

    If a scan protocol is selected and confirmed in thePatient Model Dialog, the linked Auto Referencelinessettings stay active until the user changes them,even if a scan protocol with different Auto Referencelinessettings is selected.

  • 8/10/2019 manua siemes.pdf

    101/461

  • 8/10/2019 manua siemes.pdf

    102/461

    102

    Workflow Information

    Study Continuation

    An existing study can be continued at a later time.

    To load an existing study:

    Select the desired study in the Patient Browser.

    Select Registerfrom the Patientdrop-down menu.

    The patient data is loaded in the Registrationdialogbox.

    The previous scan protocol is already preselected, butit is also possible to select any desired scan protocol.

    After the patient has been registered, the patient isloaded into the Examinationcard.

    The ranges already scanned are listed. The followingchronicle entry is shown between the ranges alreadyscanned and the new ranges: Exam Continue.

    If you want to continue a contrast media study, the sys-

    tem asks you if the next scan should be continued as anon-contrast scan instead.

    If you want to continue as a non-contrast scan, thechronicle entries for the new scan range is indicatedas a non-contrast scan. (No injector symbol isshown.)

    If you continue a study as a contrast study, the chroni-cle entries of the new scan range are indicated as acontrast scan. (An injector symbol is shown.)

  • 8/10/2019 manua siemes.pdf

    103/461

  • 8/10/2019 manua siemes.pdf

    104/461

    104

    Workflow Information

    Examination Job Status

    You can obtain an overview of all recon jobs by clickingon the Recon task icon in the status bar or selectingTransfer Examination Job statusin the patientmain menu in the Patient Browser.

    The Examination Job Statusdialog box will appear

    where all recon jobs (completed, queued and inprogress) are listed. You can stop, restart, and deleteeach job by clicking the corresponding button. To givea selected job a higher priority click urgent.

    The column Typeshows you which kind of reconstruc-tion is queued.

    Two types are displayed:

    Reconall recon jobs from the Recontask card on the syngoCT Workplace

    Auto 3Dall 3D reconstructions which you have sent via Auto

    post-processingautomatically into the 3Dtask card.These jobs will be deleted from the job list as soon asthe patient is closed in the 3Dtask card.

  • 8/10/2019 manua siemes.pdf

    105/461

    Workflow Information

    105

    Auto Load in 3D and Post-Processing

    Presets

    You can activate the Auto load in 3Dfunction on theExaminationtask card/Auto Taskingand link it to arecon job, for example, the 2ndrecon job with thinnerslice width in some of the examination protocols. If the

    post-processing type is chosen from the pull-downmenu, the reconstructed images will be loaded auto-matically into the 3Dtask card on the syngoAcquisi-tion Workplacewith the corresponding post-process-ing type.

    On the 3Dtask card you can create parallel and radial

    ranges for Multi-Planar-Reconstruction (MPR) and ThinMaximum-Intensity-Projection (MIP Thin), which canbe linked to a special series.

    For example, if you always perform sagittal MPRs for aspine examination, as soon as you load a spine exami-nation into the3Dtask card, select the image type

    (MPR), orientation, and open the Range Parallel func-tion. Adapt the range settings (image thickness, dis-tance between the images etc.) and click the link but-ton and save your settings. You now have a predefinedpost-processing protocol linked to the series descrip-tion of a spine examination.

  • 8/10/2019 manua siemes.pdf

    106/461

    106

    Workflow Information

    The same can be done for VRT presets. In the main

    menu under Type > VRT Definition, you can link VRTpresets with a series description.

    Some of the scan protocols, primarily for Angio exami-nations, are already preset in the protocol with Autoload in 3D. If you prefer not to have this preset, dese-lect the Auto load in 3D and save your scan protocol.

    Some of the scan protocols are preset in the protocolwith links to a post-processing protocol. If you prefernot to have this preset, please delete the Range Parallelpreset or overwrite them with your own settings.

    Scan Protocol Creation

    You can modify or create your scan protocols in twodifferent ways:

    by editing/saving scan protocols

    via scan protocol manager.

    Edit/Save scan protocols

    If you want to modify an existing protocol or create anew one, for example, you want to have two Abdo-menRoutine Protocolswith different slice widths, werecommend you do this directly on the Examinationtask card.

  • 8/10/2019 manua siemes.pdf

    107/461

    Workflow Information

    107

    User-specific scan protocols can be saved with the fol-

    lowing basic procedure:

    Register a patient, you can choose any patient posi-tion in the Patient Model Dialog.

    Select an existing scan protocol in the Patient ModelDialog.

    Modify the scan protocol, change parameters, add

    new ranges etc. to adapt the new protocol to yourrequirements.

    Scan your patient as usual.

    Check that all parameters are as you desire.

    Select Edit/Save Scan Protocolin the main menu.

    Select the folder where you want the new protocol

    to appear and the scan protocol name in the pop-updialog box.

    You can either use the same name to overwrite theexisting scan protocol or enter a new name, whichwill create a new protocol name and will not alterany of the existing protocols already stored.

    If you want to save an "old" protocol again, you mayhave to modify the protocol name. The old protocol(with the old name) must be cancelled explicitly.

  • 8/10/2019 manua siemes.pdf

    108/461

    108

    Workflow Information

    Additional important information:

    You can save your scan protocol at any time duringthe examination.

    It is recommended that you save your own scan pro-tocol under a new name in order to avoid overwrit-ing the default scan protocol.

    Do not use special characters or blanks. Allowed areall numbers from 0 to 9, all characters from A to Zand a to z and explicitly the _ (under-score), but nocountry-specific characters, for example, , , , , .

    Do not rename scan protocol files at Windows level.This will lead to inconsistencies.

    You can now save your own scan protocols in anypre-defined folder. The organ characteristics willbelong to the scan protocol, not to the region.

    In the Patient Model Dialog, the modified scan pro-tocols are marked by a dot in front of the protocol.

    Scan Protocol Manager

    If you want to modify special parameters for all exist-ing scan protocols or you want to modify the folderstructure we recommend doing this in the Scan Proto-

    col Manager.Call up Options > Configuration...and click on theScan Protocol Managericon in the configurationpanel. The Scan Protocol Manageris opened and allprotocols are loaded. Multiple selection of the rowsworks in the usual way via the left mouse click

    together with the Ctrl key.

  • 8/10/2019 manua siemes.pdf

    109/461

  • 8/10/2019 manua siemes.pdf

    110/461

  • 8/10/2019 manua siemes.pdf

    111/461

    Workflow Information

    111

    Define a protocol as the emergency protocol. Select

    the desired scan protocol with the right mouse but-ton. Select item Set as new emergency protocol.The selected protocol is marked with a red cross.- Open the Scan Protocol Manager(Options > Con- figuration).- Mark the protocol, with a right mouse click and

    select "Set as new emergency protocol". A red crosswill appear in front of the protocol name.- Save the changes. From now on the new protocol is used as the default when you select the emergency button in the Patient Registrationdialog box.

  • 8/10/2019 manua siemes.pdf

    112/461

    112

    Workflow Information

    Change the structure of the protocol tree. You can

    sort all customers protocols as you want to. Selectthe desired protocols, press the right mouse button.You will find the items:- Cut/Copy-Paste.To change the order of your protocol tree, you can

    drag & drop the protocols into the desired position.Or use the items Move Up and Move Downfrom thecontext menu to change the sorting of your protocoltree.

    With the Filter function you can reduce the numberof displayed scan protocols and Recon jobs.

    for reducing the displayed Recon jobs you canchoose between:All axial Recon jobsAll sagittal and coronal Recon jobsAll oblique Recon jobs

    for reducing the displayed scan protocols selectModified Scan Protocols only

    for displaying all scan protocols and Recon jobsselect: Off.

    Display of actions:

    Invalid protocols or parameters are marked in yel-low.

    Changed, but not yet saved protocols and/orparameters are marked in green.

  • 8/10/2019 manua siemes.pdf

    113/461

    Workflow Information

    113

    Table with all possible states of tree nodes

    State IconChild protocols

    Adult protocols

    region

    SIEMENS protocol

    USER protocol

    changed USER protocol

    invalid USER protocol

    SIEMENS emergency protocol

    scan protocol entry

    autorange - start

    autorange - middle

    autorange - end

    recon job

  • 8/10/2019 manua siemes.pdf

    114/461

    114

    Workflow Information

    Additional Important Information:

    With the Find/Replacefunction you can easily insertan API command for all protocols as needed.

    Also all Auto Tasking actions, for example, transfer toconfigured network nodes can be set within oneaction. With the Function button you can set your

    transfer actions depending on the slice width. Sorting the scan protocols will ensure that all organ

    characteristics belong to the protocol, so it does notmatter which folder you choose.

    For security reasons it is not possible to make anychanges in the Siemens protocols, but you can copy

    or drag & drop these protocols into a Customerfolder.

    You can save your changes with the items Save/Saveas/Save all.

    You can delete individual Recon jobs by selecting the

    Recon job in question and selecting Delete from thecontext menu.

    Only those Scan protocols that belong to the USERcategory can be deleted. Only complete scan proto-cols can be deleted. It is not possible to delete scanprotocol entries or scan protocol recon jobs.

    The item Set to defaultsin the main menu/ contextsensitive menuwill reset all your changes to theSiemens default values.

  • 8/10/2019 manua siemes.pdf

    115/461

    Workflow Information

    115

    A message will inform you if not all scan protocols

    have been saved when you close the Scan ProtocolManager.

    You can configure the displayed columns and theirposition with View configure columns.

    System/Runoffers the tool Restore Default Scan

    Protocolswhich allows you to remove user-specificscan protocols and restore the Siemens default set-tings.

    The main menu itemEditoffers save/delete ScanProtocols.

    System/Runor the main menu item View in the

    Scan Protocol Manageroffers the tool List ScanProtocolswhich generates an HTML table of allavailable scan protocols. This list can be printed orsaved to floppy disk (File/Save As).

  • 8/10/2019 manua siemes.pdf

    116/461

  • 8/10/2019 manua siemes.pdf

    117/461

    Application Information

    117

    Key Features

    SOMATOM LifeNet offline (All users):

    General Information about your system and configu-ration

    Access to E-training

    SOMATOM LifeNet online (in combination with a Sie-mens Remote Service connection):

    Newsticker archive and FAQ (frequently asked ques-tions) section

    Free trial software order and installation

    Download of information, manuals, and scan proto-cols

    A contact function providing an easy and fast inter-face with Siemens including the possibility to attachup to two DICOM images

  • 8/10/2019 manua siemes.pdf

    118/461

    118

    Application Information

    SOMATOM LifeNet offline

    Start SOMATOM LifeNet by clicking LifeNeton the Lifetask card. For syngoMultiModality Workplace, callup Options> SOMATOM LifeNetfrom the mainmenu.

    With SOMATOM LifeNet you can access various infor-

    mation about your hardware and software environ-ment.

    Under System Information,for example, you will findinformation such as software version or a scan secondcounter. Under Customer Informationyou can enteryour contact data.

  • 8/10/2019 manua siemes.pdf

    119/461

  • 8/10/2019 manua siemes.pdf

    120/461

  • 8/10/2019 manua siemes.pdf

    121/461

    Application Information

    121

    Downloading files

    Each download is performed in the background andeven if you disconnect your SOMATOM LifeNet onlinesession and start work with the CT scanner, it will con-tinue the process until downloading is completed. Dueto bandwidth restrictions it is only possible to performone download at a time.

    Depending on your connection speed downloading oflarger files like, for example, the Advanced Applica-tion Trainingmay take quite some time. Shuttingdown the system in between will interrupt the down-load.

    Downloaded PDFs can be found offline underOptions > File Browser > H:\SiteData/Manuals.

  • 8/10/2019 manua siemes.pdf

    122/461

    122

    Application Information

    The E-training is automatically installed as soon as it is

    downloaded and can be started within the SOMATOMLifeNet offline section under SOMATOM Educate > E-Training.

    For syngoAcquisition Workplaces, you can down-load and install scan protocols.

    To install the downloaded protocols open the ScanProtocol Managervia Options > Configuration.SelectImport Scan Protocolsin the menu Scan Proto-col. You can select which protocols you want to importand which not from a selection box which opens,allowing you to choose the protocols you want to copyto your Customer Folder. These protocols are then

    ready to use for your next examination.

  • 8/10/2019 manua siemes.pdf

    123/461

    Application Information

    123

    All new protocols, including those you did not choose

    to import to the Customer Folderwill also be stored inthe Siemens Folderso that they can be copied later.

    All downloaded scan protocols are named "DL_.." sothey can be easily identified when mixed with pre-installed protocols.

    Also, see the chapterScan Protocol Manager for moreinformation.

  • 8/10/2019 manua siemes.pdf

    124/461

    124

    Application Information

    Contact Function

    Contacting Siemens via Email is possible directly fromyour scanner and even DICOM images can be attachedto your message. To attach an image, please first selectthe images on your Viewingtask card first and thenexport these images to H:\SiteData\Offline as a Dicomimage.

    After exporting the image, open the Siemens Extranetand choose Contact. After entering your message youcan easily attach the image by selecting the imagesfrom the File Browser with the shortcut CTRL and C andpaste it with the shortcut CTRL and V into the Extranet.

  • 8/10/2019 manua siemes.pdf

    125/461

    Application Information

    125

    Every patient image is made anonymous before send-

    ing. Because the SOMATOM LifeNet window is alwaysin the foreground we recommend to restore/minimizeit to be able to switch between both screens, the FileBrowser and the SOMATOM LifeNet window.

    Trial Order and Installation

    As a SOMATOM CT user you can request trial clinicalsoftware directly from the scanner. The requested soft-ware will be made available and installed automati-cally through our Siemens Remote Services connec-tion.

    After you have accessed the SOMATOM LifeNet onlinearea, you can choose system-specific trial softwareunder SOMATOM Expand > Trial Licenses. After click-ing on Order Trialand confirming a license agreementfor trial-use software, you will enter your contact dataand then submit the license request.

  • 8/10/2019 manua siemes.pdf

    126/461

  • 8/10/2019 manua siemes.pdf

    127/461

    Application Information

    127

    Forwarding information via Email

    This service enables those customers who do not havea printing device connected to their CT Scanner, butwould like to have the information provided in SOMA-TOM LifeNet sent to their regular email account or for-warded to a contact person of interest as an email.

    Information about current courses and clinical trainingprograms, as well as CT accessories can be sent fromthe scanner to any email account. This information canthen be printed or a quote requested from your localSiemens representative.

  • 8/10/2019 manua siemes.pdf

    128/461

    128

    Application Information

    Image ConverterThe CT application Common DICOM Adapter can con-vert different DICOM data sets to formats provided byother CT vendors.

    You will find the converter in the Applicationsmenu

    of the Patient Browser.

  • 8/10/2019 manua siemes.pdf

    129/461

    Application Information

    129

    In the pop-up window you can select the application

    for which you want to convert images.

  • 8/10/2019 manua siemes.pdf

    130/461

    130

    Application Information

    Split-Up Multi-Phase series

    1.Select Split-up series according to trigger delayfrom the section Split-up Heartview 4D series, ifyou intend to split up a series of multi-phase heartreconstructions, according to their different heartphases.

    Or2.Select Split-up series according to slice positionto

    split up a series according to the different slice posi-tions it contains.

    After conversion you can load these data sets into theapplication of your choice.

  • 8/10/2019 manua siemes.pdf

    131/461

    Application Information

    131

    Report TemplateConfiguration

    Under Options > Configurationyou will find theReport Template Configuration.

    With the Report Configuration you can edit basic infor-mation, for example, clinic information for your reporttemplates for the applications:

    syngo Calcium Scoring

    syngoColonography CT

    syngoLung CARE CT

    Additionally you can insert your logo and select whichreference data you want to use.

  • 8/10/2019 manua siemes.pdf

    132/461

    132

    Application Information

    File BrowserThe File Browser provides you with a secure means ofaccessing and managing data in a private folder, whichis a well defined part of the computer file system. Thisuser partition is strictly separated from the system

    operating file system.

    The user partition is shared read only and may be usedfor transferring data from the scanner to other com-puters, for example, transferring DICOM images(export to offline), transferring AVIs, or accessing files(PDF files) which were downloaded via SOMATOM

    LifeNet.

    Key Features

    Copy images and files to the CD Burn folder.

    Raw data transfer.

    Access to all created reports and movies (AVI files).

    Access to the offline folder. Access to downloaded files.

    Open the File Browser via main menu itemOptions > File Browser.

    The File Browser contains special folders for CT applica-tions. The created reports and movies are saved withinthese folders.

    If an external PC is connected you can access youroffline data on the external PC for post-processing.

  • 8/10/2019 manua siemes.pdf

    133/461

    Application Information

    133

    Raw data transfer:

    Raw data sets can be transferred.

    First configure the directory to which you want totransfer the raw data:

    Open the File Browser.

    Choose the desired directory from the navigation

    tree to the left of the File Browser. Select the item Set as Export Rootfrom the CT Data

    Transferdrop-down menu.

    Select the raw data files you want to transfer in thecontent area of the Patient Browser.

    From the browser, open the Patientmain menu and

    select the item Transfer CT Da