TRANSASIA BIO - MEDICALS LTD.Transasia House, 8 Chandivali Studio Road, Andheri (E), Mumbai - 400 072Tel.: (022) 4030 9000, Fax : (022) 2857 3030 Email : [email protected] Website : www.transasia.co.in Twitter :https://twitter.com/transasia_1
ERBA Diagnostics Mannheim GmbHMallaustrasse 69-73 68219 Mannheim, GermanyTelephone : (+49) 621 8799770 Fax : (+49) 621 8799688Email : [email protected] Website : www.erbamannheim.com
Physical Specifications
Dimensions
Length : 47.25 inch 3.93 ft
Height : 27.50 inch 2.29 ft
Width : 26.00 inch 2.16 ft
Weight : 90 kgs
Power Supply
Voltage : 100 – 240 Volts
Frequency : 50 / 60Hz
Current : 5.5 A
Operational Temperature : 18 - 30°C
Humidity : 0 - 80 % (Non-condensing)
General Specifications
Self-check at startup : Yes
Start-up : < 3 minutes
Assay per plate : Up to 12
Number of ELISA plates : 4
Sample capacity : 120
Cups for pre dilution : 120
Reagent capacity : 8 Reagent bottles, 56 Reagent tubes
Perform serial dilution : Yes
Reader Specifications
Dynamic range : 0.0 - 3.0 O.D
Wavelength range : 400 - 800 nm
Filter slots : 405 nm, 450 nm, 492 nm, 620 nm & 2 optional
Light source : Halogen Lamp
Reading mode : Single / dual
Reading speed : 5 seconds
Precision : ± 0.5 % and ± 0.005 O.D at 1.5 O.D
Accuracy : ± 1 % and ± 0.005 O.D at 1.5 O.D
Washer Specifications
Manifold configuration : 16 needles - 8 aspiration & 8 dispensing Precision : < 5 % at 300 µl
Wash cycle : Programmable
Residual volume : < 5 µl ( 1 µl in V wells, flat < 3 µl)
Soak time : Programmable
Washer buffer capacity : 2 × 2 L
Waste water container : 1 × 2 L
Probe clean container : 1 × 2 L
T E C H N I C A L S P E C I F I C A T I O N S
Incubator Specifications
Number of incubators : 4
Temperature range : 37°C or R.T
Temperature accuracy : ± 1°C across plate @37°C
Incubator time : Programmable
Time to set temperature : < 8 min
Temperature monitoring : Yes
Pipetting Specifications
Number of probes : 1
Multi-dispensing : Yes
Minimum volume dispensing accuracy : 4 µl
Precision Sample & Reagent : < 5 µl is ± 1 %
Sample tube
minimum volume : 200 µl
Micro sample tube
minimum volume : 100 µl
Process Security
Liquid level sensing : Yes
Universal barcode system : Hand held barcode reader reads upto 30 alphanumeric character
Languages Read : Code 39, Code 128, Interleaved 2 of 5
Level detection system : Conductive
Strip verification : Yes
Buffer level sensing : Yes
Alarms : Yes
Software
Operating System : Multitasking (Windows XP Professional)
LIS interface : RS 232, bi-directional
USB Port : Yes
Graphic load interface : Yes
Graphical work list : Yes descriptive
Data processing : Quantitative & Qualitative
Multi password access : Yes
Ability to archive data : Yes
Specifications are subject to change without prior notification.
· introduces a level of automation that is in a class by itself.MAGO 4s
· a true walk away system-Compact bench top model to perform all steps of ELISAMAGO® 4
· Ability to perform multiple assays
· Teflon coated steel probe for dispensing samples and reagents
· Continuous probe clean- Maintained by internal hydraulics automatically.
· Multi-shot dispensing- Allows precise pipetting of small sample volumes.
· Liquid level sensing & mechanism reliability- Warning flag if reagent & sample is insufficient.
· Inbuilt independent incubation system for each plate
· 16 channel manifold with top of aspiration, no overflow
· Modular ELISA reader – Single & Dual wavelength reading
· Easy to use-Advanced graphic user interface with color coding process in each step of ELISA
· Unique graphic interface guides users at every step
· Installation base globally- Proven instrument technology from decades
Fully Automated 4 plate ELISA processor
Open system for :
Automation with…………Quality, speed, reliability & high throughput
ELISA
Fully Automated 4 Plate ELISA Processor
multiple solutions . ..
infinite possibilities
TBM
/IM
M/M
AG
O4
S/B
RO
/23
07
20
15
/V1
Probe cleaning to take care of carry over:
= The probe is internally coated with Teflon® which
reduces the adhesion of substances to the surface
= Needle is automatically cleaned internally and externally with probe cleaning solution.
= Allows precise pipetting and dispensing accurately even at small sample volumes.
Maximum flexibility for ELISA processing
Removable reagent rack to reduce setup time:
= can test up to 120 patient specimens and
report up to 384 results.
= One reagent rack is provided containing 64 positions, 8 positions for 60ml bottles & 56 postions for 5 ml test
tubes
= Reagent rack is removable so that once loaded can be
stored and reused in subsequent runs of the same
type.
Probe with liquid sensing mechanism reliability :
= Can dispense reagents sequentially to 24 wells without
stopping to refill
= Multi-shot dispensing minimizes assay drift, ensuring
consistent results across the plate
= uses the metallic probe to handle liquids.MAGO 4s
Internal ELISA Washer:
= employs a unique 16 channel automatic MAGO 4s well washer, which rapidly performs three wash cycles of each well within a strip.
= Multipoint aspiration to achieve minimal residual volume
= ELISA washing with high efficiency
Incubator with assay temperature control and probe indicator panel:
= has four plate block which is individually temperature controlled
= Maintains homogenous temperature across the plate
= Dark protective cover protects samples, reagents and reactions from exposure to common environmental contaminations
Internal ELISA Reader:
= Measurement range: 0.000 - 3.000 O.D.
= Wavelength range: 400 to 800 nm
= Accuracy: ± 1% and ± 0.005 O.D. at 1.5 O.D.
= Precision: ± 0.5% and ± 0.005 O.D. at 1.5 O.D.
= Reading speed: 5 seconds per plate
Dedicated drawer:
= A built-in compartment for high volume bottles
= There are two 2L bottles for wash solution, one 2L bottle for probe clean and one 2L bottle for waste collected from the wash manifold
= Separate built-in compartment for compressor and tubes
Each step can be viewed on the screen along with color change when the test is in progress
ü software MAGO® 4s application intended to execute diagnostic sessions
ü User guiding operations (defining sessions, loading, handling liquids, unloading)
ü Exchange data with the [HOST]
ü Multilingual and intuitive software
ü Preprogrammed assay protocols
ü Self-check system
ü Warning alarms
ü Out of range flags
ü Archiving / Downloading capability
ü Custom/User programmable channels
ü Bidirectional LIS communication available
ü Remote diagnosis capability
ü Possibility to export results in excel format
Real-Time ELISA software
Main screen
Sample outlay screen Test report
Test Report
ELISA plate outlay screen
Reagents outlay screen
Validation Format