56
® LEGENDplex™ Mul-Analyte Flow Assay Kit Cat. No. 740233 Non-human Primate (NHP) Adipokine Panel (13-plex) For Cell Culture and Urine Samples Please read the entire manual before running the assay. BioLegend.com

LEGENDplex™ · measurement of the adipokines is critical for abetter understanding of disease progression and immune responses. The LEGENDplexTM NHP Adipokine Panel is a bead-based

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

  • ®

    LEGENDplex™Mul�-Analyte Flow Assay Kit

    For Accurate Quantification of Multiple Human Th(T helper Cell) Cytokines from Cell Culture Supernatant,

    Serum, Plasma and Other Biological Samples

    Please read the entire manual before running the assay

    BioLegend.com

    ®

    LEGENDplex™Mul�-Analyte Flow Assay Kit

    Cat. No. 740233

    Non-human Primate (NHP) Adipokine Panel

    (13-plex)

    For Cell Culture and Urine Samples

    Please read the entire manual before running the assay.

    BioLegend.com

  • For Research Purposes Only. Not for use in diagnostic or therapeutic procedures. Purchase does not include or carry the right to resell or transfer this product either as a stand-alone product or as a component of another product. Any use of this product other than the permitted use without the express written authorization of BioLegend is strictly prohibited.

    It is highly recommended that this manual be read in its entirety before using this product. Do not use this kit beyond the expiration date.

  • biolegend.com 1

    LEGENDplex™ NHP Adipokine Panel

    Table of Contents Page

    Chapter 1: KIT DESCRIPTION..................................................

    Introduction……………………………………………..........................

    PrincipleoftheAssay……………………....……………....….…......

    BeadsUsage...........................................………..……………...

    StorageInformation…………………………………….......…..........

    MaterialsSupplied………………….....……………….................…

    MaterialstobeProvidedbytheEnd-User……...........……...

    Precautions.................................……………………................

    Chapter 2: ASSAY PREPARATION.............................................

    SampleCollectionandHandling…………………………............

    ReagentPreparation…………………………………………...............

    StandardPreparation..........................................................

    SampleDilution…………...........…….......................................

    Chapter 3: ASSAY PROCEDURE..................................................

    PerformingtheAssayUsingaFilterPlate……………….........

    Performing the Assay Using Microtubes, or a 96-Well V- orU-bottomMicroplate…..................................................

    Chapter 4: FLOW CYTOMETER SETUP.......................................

    Setup Procedure for FACSCaliburTM with Dual Laser..........

    Setup Procedure for FACSCaliburTM with a Single Laser.....

    Setup Procedure for BD FACSAriaTM, FACSCantoTM and LSR Series..........................................................................

    Setup Procedure for Other Flow Cytometers ...................

    Chapter 5: DATA ACQUISITION AND ANALYSIS.........................

    DataAcquisition..................................................................

    Data Analysis.....................................................................

    3

    3

    3

    4

    6

    6

    7

    8

    9

    9

    9

    9

    11

    12

    12

    15

    18

    19

    23

    30

    34

    35

    35

    36

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    2

    Chapter 6: FREQUENTLY ASKED QUESTIONS...........................

    Chapter 7: ASSAY CHARACTERIZATION..........................................

    StandardCurve.………………………………………………………........

    37

    41

    41

    41

    42

    42

    43

    44

    45

    46

    49

    52

    53

    AssaySensitivity...……………………………………………………..…..

    Cross-Reactivity……………………………………………………..........

    Accuracy.............................................................................

    LinearityofDilution………………………………………………..........

    Intra-AssayPrecision……………………………………...................

    Inter-AssayPrecision……………………………………...................

    BiologicalSamples…………………………………………….………....

    TROUBLESHOOTING...........................……………………………………...

    PLATE MAP...............……………………………………………………...............

    RACK MAP....................................................................................

  • biolegend.com 3

    LEGENDplex™ NHP Adipokine Panel

    Chapter 1: KIT DESCRIPTION

    Introduction

    Adiposetissue(AT)playsanactiveroleinregulatingmetabolismandenergyhomeostasis. AT expresses and secretes numerous soluble factors, including cytokines,chemokinesandhormones,collectivelyreferredtoasadipokinesherein. Altered levels of these proteins have been associated with a variety of metabolicdiseasesandinflammatory-autoimmuneconditions.Theaccuratemeasurementoftheadipokinesiscriticalforabetterunderstanding of disease progression and immune responses.

    The LEGENDplexTM NHP Adipokine Panel is a bead-based multiplexassay, uing fluorescence–encodedbeadssuitableforuseonvariousflowcytometers.Thispanelallowssimultaneousquantificationof13adipokinesfromRhesus or Cyno-molgus monkey,includingadiponectin,adipsin,RBP4,MCP-1,IL-1β,IP-10,IL-10,IL-8,leptin,IL-6,IFN-γ,resistinandTNF-α.Thispanelprovideshighsensitiv-ity and broad dynamic range. The panel has been validated for use on urine and cell culture supernatant samples. The panel is not suitable for serum or plasma samplesduetodifferentsampledilutionrequirementsfordifferentanalytes.

    The NHP Adipokine Panel isdesignedtoallowflexiblecustomization.Formixandmatch within the panel, please visit www.biolegend.com/legendplex.

    This assay is for research use only.

    Principle of the Assay

    BioLegend’s LEGENDplexTM assays are bead-based immunoassays using the same basic principle as sandwich immunoassays.

    Beadsaredifferentiatedbysizeandinternalfluorescenceintensities.Eachbeadisconjugatedwithaspecificantibodyonitssurfaceandservesasthecapturebeadforthatparticularanalyte.Whenaselectedpanelofcapturebeadsismixedandincubatedwithasamplecontainingtargetanalytesspecifictothecaptureantibodies,eachanalytewillbindtoitsspecificcapturebeads.Afterwashing,abiotinylateddetectionantibodycocktailisadded,andeachdetec-tionantibodyinthecocktailwillbindtoitsspecificanalyteboundonthecap-turebeads,thusformingcapturebead-analyte-detectionantibodysandwiches.Streptavidin-phycoerythrin(SA-PE)issubsequentlyadded,whichwillbindtothebiotinylateddetectionantibodies,providingfluorescentsignalintensitiesinproportiontotheamountofboundanalytes.

    Sincethebeadsaredifferentiatedbysizeandinternalfluorescenceintensityonaflowcytometer,analyte-specificpopulationscanbesegregatedandquanti-fiedbythePEfluorescentsignal.Theconcentrationofaparticularanalyteisdetemined by a standard curve generated in the same assay.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    4

    Beads Usage

    The NHP Adipokine Panelincludestwosetsofbeads.Eachsethasauniquesizethatcanbeidentifiedonflowcytometerbasedontheirforwardscatter(FSC)andsidescatter(SSC)profiles(BeadsAandBeadsB,Figure1).Eachbeadsetcanbefurtherresolvedbasedontheirinternalfluorescenceintensities.TheinternaldyecanbedetectedusingFL3,FL4,orAPCchannel,dependingonthetypeofflowcytometerused.ThesmallerBeadsAconsistsof6beadpopula-tionsandthelargerBeadsBconsistsof7beadpopulations(Figure2-3).

    Usingatotalof13beadpopulationsdistinguishedbysizeandinternalfluo-rescent dye, the NHP Adipokine Panelallowssimultaneousdetectionof13adipokinesinonesampletest.Eachanalyteisassociatedwithaparticularbeadset as indicated in Table 1.

    Figure1.BeadsDifferentiatedbySize

    Beads A = smaller beads

    Beads B = larger beads

    Figure2.BeadsAClassificationbyFL4

    A5 A7 A8

    A4

    A6

    A10

  • biolegend.com 5

    LEGENDplex™ NHP Adipokine Panel

    Figure3.BeadsBClassificationbyFL4

    For Beads usage in the panel, please refer to Table 1 below:

    Table1.BeadsIDandTargetInformation

    Target Bead ID Top Standard Concentrations(ng/mL)

    NHPAdiponectin A4 200

    NHP Adipsin A5 50

    NHPRBP4 A6 50

    NHP MCP-1 A7 10

    NHPIL-1β A8 10

    NHP IP-10 A10 10

    NHP IL-10 B2 10

    NHP IL-8 B3 10

    NHPLeptin B4 10

    NHP IL-6 B5 10

    NHPIFN-γ B6 10

    NHPResistin B7 10

    NHPTNF-α B9 10

    *BeadIDisusedtoassociateabeadpopulationtoaparticularanalyteintheLEGENDplexTMDataAnalysisSoftware.TheassociationofanalyteandbeadIDwillbedefinedduringthegatingstepofthedataanalysis.

    WhenenteringanalyteandbeadIDinfomationduringthegatingstep,alwaysenterinthesequentialorderofthebeadID(e.g,A4,A5,A6...B2,B3,B4...).Please refer to the LEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelpfordetails(www.biolegend.com/legendplex).

    B4 B5

    B6 B7 B3

    B9

    B2

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    6

    StorageInformation

    Recommended storage for all original kit components is between 2°C and 8°C. DONOTFREEZEPre-mixedBeads,DetectionAntibodiesorSA-PE.

    • Oncethestandardshavebeenreconstituted,immediatelytransfercon-tents into polypropylene vials. DO NOT STORE RECONSTITUTED STAN-DARDS IN GLASS VIALS.

    • Uponreconstitution,theleftoverstandardshouldbestoredat≤-70°Cforusewithinonemonth.Avoidmultiple(>2)freeze-thawcycles.Discardanyleftoverdilutedstandards.

    Materials Supplied

    The LEGENDplexTM kit contains reagents for 100 tests listed in the table below. Whenassayedinduplicate,thisisenoughforan8-pointstandardcurveand40samples.

    Kit Components Quantity Volume Part #

    Setup Beads 1: FITC Beads 1 vial 1 mL 77840

    Setup Beads 2: PE Beads 1 vial 1 mL 77842Setup Beads 3: Raw Beads 1 vial 2 mL 77844

    NHP Adipokine Panel Premixed Beads 1bottle 3.5 mL 76601

    NHPAdipokinePanelDetectionAntibodies 1bottle 3.5 mL 76604

    NHP Adipokine Panel Standard Cock-tail,Lyophilized 1 vial lyophilized 76607

    LEGENDplexTM SA-PE 1bottle 3.5 mL 77743LEGENDplexTMAssayBuffer 1bottle 25 mL 77562LEGENDplexTMWashBuffer,20X 1bottle 25 mL 77564Filter plate 1 plate 76187Plate Sealers 4sheets 78101DataAnalysisSoftwareDongle 1 21217NHP Adipokine Panel Manual 1 76599

  • biolegend.com 7

    LEGENDplex™ NHP Adipokine Panel

    Materials to be Provided by the End-User

    • Aflowcytometerequippedwithtwolasers(e.g.,a488nmbluelaseror532nmgreenlaseranda633-635nmredlaser)capableofdistinguishing575 nm and 660 nm oraflowcytometerequippedwithonelaser(e.g.,488nmbluelaser)capableofdistinguishing575nmand670nm.

    Partiallistofcompatibleflowcytometers:

    Flow Cytometer

    Reporter Channel

    ChannelEmission

    ClassificationChannel

    Channel Emission

    Compensa-tionneeded?

    BD FACSCaliburTM(single laser)

    FL2 575 nm FL3 670 nm Yes

    BD FACSCaliburTM(dual laser)

    FL2 575 nm FL4 660 nm No*

    BD FACSArrayTM Yellow 575 nm Red 660 nm No*

    BD FACSCantoTM

    BD FACSCantoTM IIPE 575 nm APC 660 nm No*

    BDTMLSR,LSRIIBD LSRFortessaTM

    PE 575-585 nm APC 660 nm No*

    BD FACSAriaTM PE 575 nm APC 660 nm No*

    *Compensationisnotrequiredforthespecifiedflowcytometerswhensetupproperly,butisrecommendedforconsistentresults.

    Forsettinguptheaboveflowcytometers,pleasefollowtheFlow Cytom-eter Setup guide in this manual or visit: www.biolegend.com/legendplex.

    Forflowcytometersnotlistedhere,theend-userneedstosetupthemachine following similar guidelines. Please refer to Setup Procedure for Other Flow CytometerssectioninChapter4.

    • Multichannelpipettescapableofdispensing5μLto200μL

    • Reagentreservoirsformultichannelpipette

    • Polypropylenemicrofugetubes(1.5mL)

    • Laboratory vortex mixer

    • Sonicatorbath(e.g.,BransonUltrasonicCleanermodel#B200,orequiva-lent)

    • Aluminum foil

    • Absorbent pads or paper towels

    • Plateshaker(e.g.,Lab-LineInstrumentsmodel#4625,orequivalent)

    • Tabletopcentrifuges(e.g.,Eppendorfcentrifuge5415C,orequivalent)

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    8

    Iftheassayisrunusingthefilterplateprovided(recommended),

    • Avacuumfiltrationunit(MilliporeMultiScreen®HTSVacuumManifold,catalog#MSVMHTS00orequivalent).Instructionsonhowtousethevacuum manifold can be found at the supplier’s website.

    • Avacuumsource(minivacuumpumporlinevacuum,e.g.,MilliporeVacuumPump,catalog#WP6111560,orequivalent)

    Iftheassayisruninmicrotubes,V-orU-bottom96-wellplate(optional),

    • 1.1mLpolypropylenemicroFACStubes,in96-tuberack(e.g.,NationalScientificSupplyCo,catalog#TN0946-01R,orequivalent).

    • Centrifugewithaswingingbucketadaptorformicrotiterplatesormicro-tuberacks(e.g.,BeckmanCoulterAllegraTM 6R Centrifuge with MICROPLUS CARRIERadaptorforGH3.8andJS4.3Rotors).

    • 96-well,V-orU-bottom,clearpolypropylenemicroplate(lowproteinbind-ing,e.g.,grenierbio-one,Catalog#651201orequivalent).

    • Precautions

    • Sodiumazidehasbeenaddedtosomereagentsasapreservative.Al-thoughtheconcentrationsarelow,sodiumazidemayreactwithleadandcopperplumbingtoformhighlyexplosivemetalazides.Ondisposal,flushwithalargevolumeofwatertopreventazidebuild-up.

    • Donotmixorsubstitutereagentsfromdifferentkitsorlots.Reagentsfromdifferentmanufacturersshouldnotbeusedwiththiskit.

    • Donotusethiskitbeyonditsexpirationdate.

    • SA-PEandBeadsarelight-sensitive.Minimizelightexposure.

  • biolegend.com 9

    LEGENDplex™ NHP Adipokine PanelChapter 2: ASSAY PREPARATION

    SampleCollectionandHandling

    PreparationofTissueCultureSupernatantorUrineSamples:

    • Centrifuge the sample to remove debris and assay immediately. If not pos-sible,aliquotandstoresamplesat≤-20°C.Avoidmultiple(>2)freeze/thawcycles.

    ReagentPreparation

    PreparationofAntibody-ImmobilizedBeads

    SonicatethePre-mixedBeadsbottlefor1minuteinasonicatorbathandthen vortex for 30 seconds prior to use. If no sonicator bath is available, in-creasethevortexingtimeto1minutetocompletelyresuspendthebeads.

    PreparationofWashBuffer

    • Bringthe20XWashBuffertoroomtemperatureandmixtobringallsaltsintosolution.

    • Dilute25mLof20XWashBufferwith475mLdeionizedwater.Storeun-usedportionsbetween2°Cand8°Cforuptoonemonth.

    StandardPreparation

    1. Priortouse,reconstitutethelyophilizedNHPAdipokinePanelStandardCocktailwith250µLAssayBuffer.

    2. Mix and allow the vial to sit at room temperature for 10 minutes, and then transfer the standard to an appropriately labeled polypropylene microfuge tube. This will be used as the top standard C7.

    3. Label6polypropylenemicrofugetubesasC6,C5,C4,C3,C2andC1,re-spectively.

    4.Add75µLofAssayBuffertoeachofthesixtubes.Prepare1:4dilutionofthe top standard by transferring 25 µL of the top standard C7 to the C6 tube and mix well. This will be the C6 standard.

    5. Inthesamemanner,performserial1:4dilutionstoobtainC5,C4,C3,C2andC1standards(see the tables below).AssayBufferwillbeusedasthe0pg/mLstandard(C0).

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    10

    Tube Serial Dilution

    Assay Buffertoadd(µL)

    Standard to add

    Final Conc.

    (pg/mL)*

    Final Conc.

    (pg/mL)**

    Final Conc.

    (pg/mL)***

    C7 -- -- -- 10,000 50,000 200,000

    C6 1:4 75 25 µL of C7 2,500 12,500 50,000

    C5 1:16 75 25 µL of C6 625 3,125 12,500

    C4 1:64 75 25 µL of C5 156.3 781.3 3,125

    C3 1:256 75 25 µLofC4 39.1 195.3 781.3

    C2 1:1024 75 25 µL of C3 9.8 48.8 195.3

    C1 1:4096 75 25 µL of C2 2.4 12.2 48.8

    C0 -- 75 -- 0 0 0

    *TopstandardconcentrationofMCP-1,IL-1β,IP-10,IL-10,IL-8,Leptin,IL-6,IFN-γ,ResistinandTNF-αis10ng/mL.

    **TopstandardconcentrationofAdipsinandRBP4is50ng/mL.

    ***TopstandardconcentrationofAdiponectinis200ng/mL.

  • biolegend.com 11

    LEGENDplex™ NHP Adipokine Panel

    SampleDilution

    • For cell culture supernatant samples, the levels of analyte can vary greatly from sample to sample. While the samples can be tested without dilu-tions,apreliminaryexperimentmayberequiredtodeterminetheappro-priatedilutionfactor.

    Ifsampledilutionisdesired,dilutionshouldbedonewithcorrespondingfreshcellculturemediumorAssayBuffertoensureaccuratemeasure-ment.

    • Urinesamplesshouldbedilutedtwo-foldwithAssayBufferbeforetesting(e.g.dilute50µLofsamplewith50µLofAssayBuffer).Furtherdilutionisneededifthesampleconcentrationisabovetheupperlimitofthestandard range.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    12

    Chapter 3: ASSAY PROCEDURE

    The LEGENDplexTM assaycanbeperformedeitherinafilterplate,inmicrotubes,orinaV-orU-bottommicroplate.

    • Thein-filterplateassayprocedureishighlyrecommendedduetoitsgoodsample to sample consistency, assay robustness and ease of handling. This procedurerequiresavacuumfiltrationunitforwashing(seeMaterials to beProvidedbytheend-user,Page7).IfyouhaveperformedaLuminex®-basedmultiplexassaybefore,yourlabshouldalreadyhavethevacuumfiltrationunitsetup.

    • Ifthein-filterplateassayprocedureisnotpossible,orifyouprefer,theas-saycanbeperformedinmicrotubesorinaV-orU-bottommicroplate.Forin-tubeassay,werecommendusingmicroFACStubes(seeMaterials to be Providedbytheend-user,Page7).

    Performing the Assay Using a Filter Plate

    • Allowallreagentstowarmtoroomtemperature(20-25°C)beforeuse.

    • Setthefilterplateonaninvertedplatecoveratalltimesduringassaysetupandincubationstepssothatthebottomoftheplatedoesnottouchanysurface. Touching a surface may cause leakage.

    • Keeptheplateuprightduringtheentireassayprocedure,includingthewashing steps, to avoid losing beads.

    • The plate should be placed in the dark or wrapped with aluminum foil for allincubationsteps.

    • Standards and samples should be run in duplicate and arranged on the plateinaverticalconfigurationconvenientfordataacquisitionandanaly-sis(asshowninattachedPLATEMAP,Page52-53).Besuretoloadstan-dardsinthefirsttwocolumns.Ifanautomationdeviceisusedforread-ing,theorientationandreadingsequenceshouldbecarefullyplanned.

    1. Pre-wettheplatebyadding100μLofLEGENDplexTM1XWashBuffertoeach well and let it sit for 1 minute at room temperature. To remove the excess volume, place the plate on the vacuum manifold and apply vacuum. Donotexceed10”Hgofvacuum.Vacuumuntilwellsaredrained(5-10seconds).BlotexcessWashBufferfromthebottomoftheplatebypress-ing the plate on a stack of clean paper towels. Place the plate on top of the inverted plate cover.

  • biolegend.com 13

    LEGENDplex™ NHP Adipokine Panel

    For measuring cell culture supernatant and urine samples:

    • Add25µLofAssayBuffertoallwells.• Add 25 µL of each standard to the standard wells. • Add25µLofeachsampletothesamplewells(SeeSampleDilution)

    2. Vortexthepre-mixedbeadsbottlefor30seconds.Add25μLofthepre-mixedbeadstoeachwell.Thevolumeshouldbe75μLineachwellafterbeadsaddition.(Note:Duringadditionofthebeads,shakethepre-mixedbeadsbottleintermittentlytoavoidbeadsettling).

    3. Seal the plate with a plate sealer. Toavoidplateleaking,donotapplypositivepressuretothesealerwhensealingtheplate.Wraptheentireplate, including the inverted plate cover, with aluminum foil. Place the plate on a plate shaker, secure it and shake at approximate 500 rpm for 2 hours at room temperature.

    4. Do not invert the plate! Place the plate on the vacuum manifold and apply vacuumasbeforeinStep1.Add200µLof1XWashBuffertoeachwell.RemoveWashBufferbyvacuumfiltration.BlotexcessWashBufferfromthebottomoftheplatewithanabsorbentpadorpapertowels. Repeat this washing step once more.

    5. Add25µLofDetectionAntibodiestoeachwell.

    6. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinverted plate cover, with aluminum foil. Place the plate on a plate shaker and shake at approximately 500 rpm for 1 hour at room temperature.

    7. Do not vacuum! Add 25 µL of SA-PE to each well directly.

    8. Sealtheplatewithafreshplatesealer.Wraptheentireplate,includingtheinverted plate cover, with aluminum foil. Place the plate on a plate shaker and shake at approximate 500 rpm for 30 minutes at room temperature.

    9. Repeatstep4above.

    10. Add200µLof1XWashBuffertoeachwell(or150µLifplateistobereadbyanautosampler.Highervolumemayresultinleakingofthefilterplateduring reading. Besuretosetthesamplevolumetobeanalyzedto70uLor less so that sample can be read again if needed).Resuspendthebeadson a plate shaker for 1 minute.

    11. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay (Note:Prolongedsamplestoragecanleadtoreducedsignal)

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    14

    Iftheflowcytometerisequippedwithanautosampler,readtheplatedirectly using the autosampler. The probe height may need to be adjusted when using an autosampler.

    If an autosampler is not available, the samples need to be transferred from thefilterplatetoFACStubesandreadmanually.

    Assay Procedure Summary for Filter PlateAdd 100 μL 1X Wash Bu�er to �lter plate wells

    Vacuum to remove excess bu�er

    Incubate 2 hours, RT, shaking

    Capture beads

    Biotinylated Detection Antibody

    Analytes

    Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking

    Wash 2 times using vacuum �ltration unitAdd 25 μL Detection AntibodiesIncubate 1 hr, RT, shaking

    For cell culture supernatant/urine samples, Add to the plate:25 μL Assay Bu�er to all wells25 μL diluted standard to standard wells25 μL sample to sample wells25 μL pre-mixed beads to all wells

    Wash 2 times using vacuum �ltration unitAdd 200 μL (or 150µL) of 1x Wash Bu�er Read on a �ow cytometer

    BA

    C

    A B C

    A B C

  • biolegend.com 15

    LEGENDplex™ NHP Adipokine Panel

    PerformingtheAssayUsingMicrotubesora96-WellV-orU-bot-tom Microplate

    • Allowallreagentstowarmtoroomtemperature(20-25°C)beforeuse.

    • Determine the number of assays to run. Standards and samples should be runinduplicateandarrangedintheorderconvenientfordataacquisitionand analysis (see below). Ifanautomationdeviceisusedforreading,theorientationandreadingsequenceshouldbecarefullyplanned.

    Ifusingtubes,labelalltubesandarrangetubesonarack(asshowninat-tached RACK MAP, assuming using micro FACS tubes and a 96-microtube rack,e.g.,NationalScientificSupplyCo.,Catalog#:TN0946-01R).A96-mi-crotuberackisrecommendedbecauseitallowspipettingwithamultichan-nelpipette.

    If using a microplate, arrange the standard and samples according to the PLATEMAPattached.

    1. For measuring cell culture supernatant and urine samples:

    • Add25µLofAssayBuffertoalltubes/wells.• Add25µLofeachstandardtothestandardtubes/wells.• Add25µLofeachsampletothesampletubes/wells(See Sample Dilu-tion).• Add25µLofthepre-mixedbeadstoalltubes/wells.• Add25µLDetectionAntibodiestoalltubes/wells.

    Note: Thevolumeshouldbe100μLineachtube/well.Shakebeadsbottleintermittentlyduringtheadditiontoavoidbeadsettling.

    2. Covertheentirerack/platewithaluminumfoiltoprotectthetubes/platefrom light. Shake(approximate1,000rpmforin-tubeassayand600rpmforin-plateassay) on a plate shaker for 2 hours at room temperature.

    3. Withoutwashingthetubes/plate,add25µLSA-PEtoeachtube/wells.

    4. Covertheentirerack/platewithaluminumfoiltoprotectthetubes/platefrom light. Shake(approximate1,000rpmforin-tubeassayand600rpmforin-plateassay)on a plate shaker for 30 minutes at room temperature.

    5. Centrifugethetubes/plateat1,000xg for 5 minutes, using a swinging bucketrotorwithmicroplateadaptor(Please refer to Materials to be Pro-videdbytheend-user,Page7).

    6. Removethesupernatantusingamicrochannelpipette.Becarefulnottoremove beads, but to remove as much liquid as possible.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    16

    7. Add200µLof1XWashBuffertoalltubes/well.Resuspendthebeadsbyvortexing(forin-tubeassay)orshakingonaplateshakerfor1minute(forin-plateassay).Centrifugethetubes/plateat1,000xg for 5 minutes, using a swinging bucket rotor with microplate adaptor. Remove the supernatant.

    8. Add200µLof1XWashBuffertoalltube/well(or 150 µL if plate is to be readwithanautosampler.Highervolumemayresultinleakingofthefil-terplateduringreading.Besuretosetthesamplevolumetobeanalyzedto 70 uL or less so that sample can be read again if needed).Resuspendthebeadsbyvortexing(forin-tubeassay)orpipetting(forin-plateassay).

    9. Readsamplesonaflowcytometer,preferablywithinthesamedayoftheassay(Note:Prolongedsamplestoragecanleadtoreducedsignal).

    Iftheflowcytometerisequippedwithanautosampler,thesamplescanbereadeitherdirectly(forin-plateassay)orafterbeingtransfferedtoa96-wellplate(forin-tubeassay).The probe height may need to be adjusted when using an autosampler.

    If an autosampler is not available, the samples can be read directly in FACS tubesorafterbeingtransferredfromthemicroplatetomicroFACStubes.

  • biolegend.com 17

    LEGENDplex™ NHP Adipokine Panel

    Assay Procedure Summary for Tubes (or V-Bottom Plate)

    Arrange the number of tubes needed on a rack (or set up the plate)

    Incubate 2 hours, RT, shaking

    Without washing, add 25 μL SA-PEIncubate 30 min, RT, shaking

    For cell culture supernatant/urine samples, Add to the tubes/wells:25 μL Assay Bu�er to all tubes/wells25 μL diluted standard to standard tubes/wells25 μL sample to sample tubes/wells25 μL pre-mixed beads to all tubes/wells25 μL Detection Antibodies to all tubes/wells

    Spin down beadsWash one timeAdd 200 μL (or 150µL) of 1x Wash Bu�erRead on a �ow cytometer

    A B

    C

    A

    B

    C

    Capture beads

    Detection Antibody

    Analytes

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    18

    Chapter 4: FLOW CYTOMETER SETUP

    Inordertogeneratereliabledata,theflowcytometermustbesetupproperlybeforedataacquisition.Thefollowingsectionswilladdressmachinesetupfortheflowcytometerslistedbelow.

    ListofFlowCytometersandPossibleConfigurations

    Flow CytometerReporterChannel

    Channel Emission

    ClassificationChannel

    ChannelEmission

    Compensationneeded?

    BD FACSCaliburTM (singlelaser) FL2 575 nm FL3 670 nm Yes

    BD FACSCaliburTM (duallaser) FL2 575 nm FL4 660 nm No*

    BD FACSCantoTM,BD FACSCantoTM II PE 575 nm APC 660 nm No*

    BDTM LSR, LSRII,BD LSRFortessaTM PE

    575-585 nm APC 660 nm No*

    BD FACSAriaTM PE 575 nm APC 660 nm No*

    *Compensationisnotrequiredforthespecifiedflowcytometerswhensetupproperly,butisrecommendedforconsistentresults.

    Forflowcytometersnotlistedhere,theend-userneedstosetupthemachinefollowing similar guidelines. Please refer to Setup Procedure for Other Flow Cytometerssectioninthischapter.

    The setup process typically includes the following steps. Please see the detailed setupprocedurethatfollows,regardingyourspecificinstrument.

    1).Startuptheinstrumentfollowingthemanufacturer’srecommendations.

    2).Createatemplatefordataacquisitionusingyourinstrument’sdata acquisitionsoftware.Atemplateisadocumentorworksheetwithdensityplotsthatallowstheusertoperformmachinesetupanddataacquisition.

    3).SetupthePMTvoltagesofeachchanneltobeusedfordataacquisition using the Setup Beads provided in the kit.

    4).DeterminewhethercompensationisneededbasedontheconfigurationofyoursystemasshowninTable1.Ifcompensationisneeded,perform compensationusingtheSetupBeadsprovidedinthekit.

  • biolegend.com 19

    LEGENDplex™ NHP Adipokine Panel

    Setup Procedure for FACSCaliburTM with Dual Lasers

    For a dual laser FACSCaliburTM,useFL2forreporterandFL4forbeadsclassifica-tion.Ingeneral,thereisnoneedforcompensationbetweenthesechannelsifthe machine is set up properly, following the setup procedure described below.

    1. Start up the Instrument

    Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.

    2. ObtainaTemplateforDataAcquisition

    A template for FACSCaliburTM is a document with density plot that allows theusertoperformmachinesetupanddataacquisition.

    Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplate and proceed to Step 3.

    If a template is not yet available, create a new template by following the instructionsbelow:

    2.1 FromtheBDCellQuestdataacquisitionsoftware,gotoFile→new document.

    2.2 CreateadotplotwithFSC(forwardscatter)forX-axisandSSC(side scatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create twogatesandlabelthemBeadsAandBeadsB(Figure4).

    Figure 4.

    2.3 Create4fluorescentdotplotsasshownbelow(Figure5)withFL2for X-axis,FL4orFL1forY-axis.Forthefluorescentdotplots,gateon BeadsA(dotplotsontheleftpanelbelow)andBeadsB(dotplotson therightpanelbelow).The dot plots should be in log mode.

    02

    00

    40

    06

    00

    80

    01

    00

    0S

    SC

    -H

    00 200 400 600 800 1000FSC-H

    Acquisition Dot Plot

    Beads A

    Beads B

    10

    01

    01

    10

    21

    03

    10

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    01

    01

    10

    21

    03

    10

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    01

    01

    10

    21

    03

    10

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    01

    01

    10

    21

    03

    10

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    Page 1

    Gated on Beads BGated on Beads A

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    20

    2.4 Savethenewdocumentas“LEGENDplexTemplateforFACSCalibur Dual Laser” and proceed to the next step of the setup.

    Figure 5.

    3. Set up the PMT Voltages

    The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannel(FL4/APC)andchannelFL1.TheSetupBeads2:PE

    BeadsareusedtosetupthePMTvoltageofthereporterchannel(FL2/PE)The Setup Beads 1: FITC Beads are not needed for this setup.

    FollowtheinstructionsbelowforsettingupthePMTsettings:

    3.1 Vortex the vial of the Raw Beads for 30 seconds to resuspend the beads.

    3.2 Transfer400μLoftheRawBeadstoafreshFACStube.

    3.3 Settheflowcytometerflowratetolow.Insetupmode,runthe RawBeads.AdjustthesettingsforFSCandSSCsothatbothbead populationsarevisible(Figure6).

    Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadpopulationsafteradjustingsettings.

  • biolegend.com 21

    LEGENDplex™ NHP Adipokine Panel Figure 6.

    3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>200.

    3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate

    (Figure6).

    3.6 AdjusttheFL1settingsothattheFL1signalsforallbeadsare between 1x100 and 1x101(Note:Thisstepisnotrequired,butis

    recommended).

    3.7 AdjusttheFL4settingsothattheFL4signalsforallbeadsarebe- tween 1x101 and 5x103(Figure7).

    Figure 7.

    3.8 Vortex the vial of Setup Beads 2: PE Beads for 30 seconds to resus- pend the beads.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    22

    3.9 Transfer400μLofthePEBeadstoafreshFACStube.

    3.10ReplacetheRawBeadstubefromtheflowcytometerwiththe PE beads tube.

    3.11Settheflowcytometerflowratetolow.Insetupmode,runthePE Beads.Note:PEbeadsareonlyofsmallsize,fallingintheBeadsA

    gate(Figure8).

    3.12AdjusttheFL2settingsothatthemedianfluorescenceintensityof thePEbeadsfallsbetweenthelot-specificrangelabeledonthePE Beadsvial(Figure8,gateR3).

    Figure 8.

    3.13 Save the document again for future use.

    3.14Theflowcytometerisnowreadyforsampleacquisition.

  • biolegend.com 23

    LEGENDplex™ NHP Adipokine Panel

    Setup Procedure for FACSCaliburTM with a Single Laser

    For a single laser FACSCaliburTM,useFL2forreporterandFL3forbeadsclassifi-cation.Compensationisneededtoproperlysetuptheinstrument.

    1. Start up the Instrument

    Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.

    2. ObtainaTemplateforDataAcquisition

    A template for FACSCaliburTM is a document with density plot that allows theusertoperformmachinesetupanddataacquisition.

    Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplate and proceed to Step 3.

    If a template is not yet available, create a new template by following the instructionsbelow:

    2.1 FromtheBDCellQuestdataacquisitionsoftware,gotoFile→new document.

    2.2 CreateadotplotwithFSC(forwardscatter)forX-axisandSSC(side scatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create twogatesandlabelthemBeadsAandBeadsB(Figure9).

    Figure9.

    0200

    400

    600

    800

    1000

    SS

    C-H

    00 200 400 600 800 1000FSC-H

    Acquisition Dot Plot

    Beads A

    Beads B

    10

    010

    110

    210

    310

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    Page 1

    Gated on Beads BGated on Beads A2.3 Create4fluorescentdotplotsasshownbelow(Figure10)withFL2 forX-axis,andFL3orFL1forY-axis.Forthefluorescentdotplots,gate onBeadsA(dotplotsontheleftpanelbelow)andBeadsB(dotplots ontherightpanelbelow).The dot plots should be in log mode.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    24

    Figure 10.

    0200

    400

    600

    800

    1000

    SS

    C-H

    00 200 400 600 800 1000FSC-H

    Acquisition Dot Plot

    Beads A

    Beads B

    10

    010

    110

    210

    310

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    Page 1

    Gated on Beads BGated on Beads A

    2.4 Setallcompensationstozero.

    2.5 Savethenewdocumentas“LEGENDplexTemplateforFACSCalibur Single Laser” and proceed to the next step of the setup.

  • biolegend.com 25

    LEGENDplex™ NHP Adipokine Panel3. SetupPMTVoltagesandCompensation

    The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannelFL3andchannelFL1.TheSetupBeads2:PEBeadsare used to set up the PMT voltage of the reporter channel FL2. All three setupbeadsareneededforsettingupcompensation.

    FollowtheinstructionsbelowforsettingupthePMTsettingsandcompen-sation:

    3.1 Vortex the vial of the Raw Beads for 30 seconds to resuspend the beads.

    3.2 Transfer400μLoftheRawBeadstoafreshFACStube.

    3.3 Settheflowcytometerflowratetolow.Insetupmode,runtheRaw Beads.AdjustthesettingsforFSCandSSCsothatbothbeads populationsarevisible(Figure11).

    Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadpopulations,afteradjustingsettings.

    Figure 11.

    3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>200.

    3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate (Figure11).

    3.6 AdjusttheFL1settingsothattheFL1signalsforallbeadsarebe- tween 1x100 and 1x101 (Figure12).

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    26

    Figure 12.

    3.7 AdjusttheFL2settingssothattheFL2signalsforallbeadsarebe- tween 1x100 and 1x101(Figure12).

    3.8 AdjusttheFL3settingsothattheFL3signalsforallbeadsarebe- tween 1x101 and 5 x 103(Figure12).

    3.9 Vortex the vial of the Setup Beads 1: FITC Beads for 30 seconds to resuspend the beads.

    3.10Transfer200μLoftheFITCBeadstoafreshFACStube.Add200μL ofRawBeadsandmixwell(thiswillgenerateFITC-positiveand FITC-negativepopulationsofbeadsandisneededforproper compensation).

    3.11 In setup mode, run the mixed FITC and Raw Beads.

    3.12OntheFL1vsFL2dotplot(Figure13),thebeadswilldisplayasFITC- negativeandFITC-positivepopulations(indicatedbyanarrow).

    Note:FITCbeadsareonlyofsmallsize,fallingintheBeadsAgate.

    10

    010

    110

    210

    310

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L3-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    10

    010

    110

    210

    310

    4F

    L1-H

    100 101 102 103 104

    FL2-H

    Acquisition Dot Plot

    0200

    400

    600

    800

    1000

    SS

    C-H

    00 200 400 600 800 1000FSC-H

    Acquisition Dot Plot

    Beads A

    Beads B

    Page 1

    Gated on Beads BGated on Beads A

  • biolegend.com 27

    LEGENDplex™ NHP Adipokine Panel Figure 13.

    3.13AdjusttheFL2-%FL1compensationsetting(e.g.,FL2-FL1=20%)so thattheFITC-negativeandFITC-positivepopulationshavesimilar meanFL2fluorescenceintensities(Figure14).

    Figure 14.

    3.14VortexthevialofPEBeadsfor30secondstoresuspendthebeads.

    3.15Transfer200μLofthePEBeadstoafreshFACStube.Add200μLof RawBeadsandmixwell(ThiswillgeneratePE-positiveand PE-negativepopulationsofbeadsandisneededforpropercompen- sation).

    3.16 In setup mode, run the mixed PE and Raw Beads.

    3.17OntheFL1vsFL2dotplot(Figure15),thebeadswilldisplayasPE- negativeandPE-positivepopulations(indicatedbyanarrow).Note: PEbeadsareonlyofsmallsize,fallingintheBeadsAgate.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    28

    Figure 15.

    3.18AdjusttheFL2settingsothatthemedianfluorescenceintensityofthe PEbeadsfallsbetweenthelot-specificrangelabeledonthePEBeads vial(Figure16).

    3.19AdjusttheFL1-%FL2compensationsetting(e.g.,FL1-FL2=1.5%)so thatthePE-negativeandPE-positivepopulationshavesimilarmean FL1fluorescenceintensities(Figure16).

    Figure 16.

    3.20OntheFL3vsFL2dotplot(Figure17),thebeadswilldisplayasPE- negativeandPE-positivepopulations(indicatedbyanarrow).

    3.21 AdjusttheFL3-%FL2compensationsetting(e.g.,FL3-%FL2=40%)so thattheFL3-lowpopulationandthePE-positivepopulationhave similarmedianFL3signal(Figure18).

  • biolegend.com 29

    LEGENDplex™ NHP Adipokine Panel Figure 17.

    Figure 18.

    3.22 Save the document again for future use.

    3.23Theflowcytometerisnowcompensatedandreadyforsample analysis.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    30

    Setup Procedure for BD FACSAriaTM, FACSCantoTM and LSR Series

    ThispartoftheguideappliestoBDdigitalflowcytometersusingFACSDivaTM softwareversion6.0andabove.

    For the BD FACS machines running FACSDivaTM, use the PE channel for reporter andtheAPCchannelforbeadsclassification.Ingeneral,thereisnoneedforcompensationbetweenthesechannelsifthemachineissetupproperlyfollow-ing the setup procedure described below.

    Thissetupprocedureisrequiredunderthefollowingsituations:

    • YouarerunningtheLEGENDplexkitforthefirsttime.

    • It has been over a month since the procedure was last performed.

    • Yourflowcytometerhasbeenservicedsinceyoulastperformedthis procedure.

    This setup process is not needed if you have run this experiment before and haveaccesstoasavedexperimenttemplate(Thesettingswillbesavedinthefinalstepofthissetupprocedureandanysettingssavedcanbeimportedtoanewexperiment.PleaserefertoStep2andStep3.9below).

    1. Start up the Instrument

    Performinstrumentstartupandverificationcheckfollowingthemanufac-turer’srecommendations.

    2. ObtainaTemplateforDataAcquisition

    A template for FACSDivaTM is a worksheet with density plots that allows the usertoperformmachinesetupanddataacquisition.

    If a template is not yet available, create a new template by following the instructions.Afteratemplateiscreated,savethefileinD:\BDExport\Tem-plates\Experiment.DonotchangethenameoftheTemplatesfolder.

    Ifyouhavealreadycreatedatemplatefortheflowcytometer,openthattemplateandproceedtoStep3.Toopenanexistingtemplate,selectEx-periment>NewExperiment.AlistoftemplatessavedinD:\BDExport\Tem-plates\Experimentwillpopup.Selectthedesiredtemplatefromthelist.

    Tocreateanewtemplate,followtheinstructionsbelow:

    2.1 From the BD FACSDivaTMsoftware,gotoExperiment>NewExperi- ment.

  • biolegend.com 31

    LEGENDplex™ NHP Adipokine Panel 2.2 In the global worksheet, open the worksheet. Create a dot plot with

    FSC(forwardscatter)forX-axisandSSC(sidescatter)forY-axis.Be sure to set FSC and SSC to linear mode. Create two gates and label themBeadsAandBeadsB(Figure19).

    Figure19.

    2.3 CreatetwodotplotswithPEforX-axis,APCforY-axis(Figure20), gatedonBeadsA(leftpanelbelow)andBeadsB(rightpanelbelow), respectively.CreateonedotplotwithFITCforX-axis,APCforY-axis, gatedonBeadsAandBeadsB(graphnotshown).The plots should all be in log mode.

    Figure 20.

    2.4 Savethedocumentas“LEGENDplexTemplateforFACSDiva” inD:\BDExport\Templates\Experimentandproceedtothenextstep of setup.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    32

    3. Set up PMT Voltages

    The Setup Beads 3: Raw Beads are used to set up the PMT voltage of the classificationchannelAPC,reporterchannelPE,andFITCchannel.TheSetup Beads 1: FITC Beads and 2: PE Beads are not needed for this setup becausenocompensationisrequiredifthesetupproceduredescribedhereis closely followed.

    FollowtheinstructionsbelowforsettingupthePMTsettings:

    3.1 Vortex the vial of Raw Beads for 30 seconds to resuspend the beads.

    3.2 Transfer400μLoftheRawBeadstoafreshFACStube.

    3.3 Settheflowcytometerflowratetolow.RuntheRawBeads.Adjust thesettingsforFSCandSSCsothatbothbeadpopulationsarevisible (Figure21).

    Pauseandrestartacquisitionfrequentlyduringthesetupprocedure torefreshthebeadspopulationsafteradjustingsettings.

    Figure 21.

    3.4 ContinueadjustingthesettingssothatBeadsAandBeadsBarewell separatedandtheFSCandSSCreadingsare>50(x1000).

    3.5 Move the gates for Beads A and Beads B so that the smaller beads fall into Beads A gate and the larger beads fall into Beads B gate (Figure21).

    3.6 AdjusttheFITCsettingsothattheFITCsignalforthemajorityof beads is between 1x101 and 1x102 (Figure22).

  • biolegend.com 33

    LEGENDplex™ NHP Adipokine Panel Figure 22.

    Beads Beads A+ Beads B

    3.7 AdjustthePEsettingsothatthePEsignalforthemajorityofbeadsis between 1x101 and 1x102(Figure23).

    Figure 23.

    3.8 AdjusttheAPCsettingssothatthetheAPCfluorescenceintensitiesof allbeadpopulationsarebetween1x102 and 5 x 104(Figure23).

    3.9 Save the document again for future use.

    Tosaveyourassay-specificsettings,inthebrowser,right-click CytometerSettingsandselectSavetoCatalog.Namethefile,and thenclickOK.Toimportthesavedsettingforanewexperiment,right clickoncytometersettingsandselectimportsettings.

    3.10Theflowcytometerisnowreadyforsampleacquisition.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    34

    Setup Procedure for Other Flow Cytometers

    Forflowcytometersnotaddressedabove,thesetupprocedurewilldifferfromone to another. It is very important for the end-user to set up the machine following similar guidelines. Inthiscase,machinecompensationbetweenthereporterandbeadsclassificationchannelsisstronglyrecommended.

    Theflowcytometersetupinstructionsarealsopostedonourwebsite:www.biolegend.com/legendplex. Newflowcytometersetupinstructionswillbeadded to on our website when they become available.

  • biolegend.com 35

    LEGENDplex™ NHP Adipokine Panel

    Chapter 5: DATA ACQUISITION AND ANALYSIS

    DataAcquisition

    1. Beforereadingsamples,makesurethattheflowcytometerissetupprop-erly.Forflowcytometersetup,pleasefollowtheFlowCytometerSetupguide in this manual or visit: www.biolegend.com/legendplex.

    2. Createanewtemplateoropenanexistingtemplate(fordetailsonhowtocreateacytometer-specifictemplate,pleaserefertotheFlowCytometerSetupGuide).

    3. Vortex each sample for 5 seconds before analysis.

    4.Settheflowratetolow.Setthenumberofbeadstobeacquiredto2000-2500 on Beads A gate or Beads B gate. Do not acquire too many events (e.g.>10,000totalevents).

    Note: Do not acquire too few or too many beads. Too few beads acquired may result in high CVs and too many beads acquired may result in slow data analysis later.

    Whenreadingsamples,settheflowcytometertosetupmodefirstandwaituntilbeadpopulationisstabilizedbeforeswitchingtoacquisitionmode.

    To simplify data analysis using the LEGENDplexTMDataAnalysisSoftware,read samples in the same order as shown on the PLATE MAP or RACK MAP attachedattheendofthemanual.Foranin-plateassay,readcolumnbycolumn(A1,B1,C1...A2,B2,C2...).Foranin-tubeassay,readrowbyrow(A1,A2,A3,...B1,B2,B3...).

    Whennamingdatafiles,trytousesimplenameswithaconsecutivenum-beringforeasydataanalysis(e.g.forstandards,C0.001,C0.002,C1.003,C1.004,C2.005,C2.006,C3.007,C3.008,...C7.015,C7.016;forsamples,S1.017,S1.018,S2.019,S2.020,S3.021,S3.022…)

    StoreallFCSfilesinthesamefolderforeachassay.Ifrunningmultipleas-says, create a separate folder for each assay.

    5. Proceed to data analysis using LEGENDplexTMDataAnalysisSoftwarewhendataacquisitioniscompleted.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    36

    Data Analysis

    • TheFCSfilegeneratedonaflowcytometershouldbeanalyzedusingBioLegend’s LEGENDplexTMDataAnalysisSoftwareorothercompatibledataanalysissoftware.TheLEGENDplexTMDataAnalysisSoftwarecanbedownloaded here: www.biolegend.com/legendplex.

    • Afterdownloading,installitonaPC(runningWindows7orWindows8)anduseitinconjunctionwiththeDataAnalysisSoftwareDongleincludedin this kit. The dongle has a license key stored in it and is needed to run the software.Tousethedongle,simplyplugitintheUSBportofthecomputeronwhichthedataanalysissoftwareisinstalled,priortolaunchingthesoftware.

    • TheSoftwareDonglehasafixednumberofpoints.Eachanalysiswillconsume a certain number of points. The number of points consumed dependsonassayplexsizeandnumberofsamples.Afterthepointsona dongle are consumed, a new dongle will be needed to run more data analysis.Althoughthesoftwarewillcontinuetobefunctional,thedatawillnotbesaveduntilanewdongleisavailable.Anewdongleisprovidedineach LEGENDplexTMkit.Asavedanalysiscanalwaysbere-analyzedusingthesoftwareregardlessofdonglestatus.

    • Follow the LEGENDplexTMDataAnalysisSoftwareUserGuideandOnlineHelptousethesoftware(www.bioLegend.com/legendplex; or press F1 for online help at any step of the data analysis).

  • biolegend.com 37

    LEGENDplex™ NHP Adipokine Panel

    Chapter 6: FREQUENTLY ASKED QUESTIONS

    Q. WhatisthedifferencebetweenLEGENDplexTM and Luminex®Assays?

    BioLegend’s LEGENDplexTM assays and Luminex®-basedmultiplexassaysare both bead-based immunoassays using the same basic principle of sandwichimmunoassays.Bothsystemsusefluorescence-codedbeadstoachievemultiplexing.Themajordifferenceishowthedataisacquired.LEGENDplexTMassaysusecommonlabflowcytometersandtheirrespec-tivesoftwarefordataacquisition,whereasLuminex®-based assays use dedicatedmachinesandsoftwarefordataacquisition.Therefore,oneofthe advantages of LEGENDplexTM assays is that they can be run on common flowcytometersandnospecializedmachineisneeded.

    Q. Can I use LEGENDplexTM kits on Luminex®machines?

    No. LEGENDplexTMkitsshouldbeusedonaregularflowcytometerandthedatacanbeanalyzedusingtheuser-friendlydataanalysissoftwareavail-ablefordownloadatwww.biolegend.com/legendplex.

    Q. WhatarethecompatibleflowcytometersfortheLEGENDplex™assays?

    In general, LEGENDplexTMassayscanbeusedonmostcommonflowcytom-eters, such as:

    BD FACSCalibur™ BD FACSCanto™ BD FACSCanto™ II BD TM LSR I BD TM LSR II BD LSRFortessaTM BD FACSAria™ BD FACSArrayTM

    PleaserefertotheMATERIALSTOBEPROVIDEDBYEND-USERsectionfordetailsonthechannelconfigurationsofthecytometer.

    Forflowcytometersnotlistedabove,theenduserneedstomakesure

    thatthemachineissetupproperlybeforeuse(refertoSetup Procedure for Other Flow CytometerssectioninChapter4).Inthiscase,machinecompensationbetweenthereporterandbeadsclassificationchannelsisstrongly recommended.

    TheFCSorListmodefilesfromthefollowinginstrumentsarecompatiblewith the LEGENDplex TMDataAnalysisSoftware.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    38

    Vendor Instrument Model VendorInstrument

    Model

    BD

    FACScan™ Sony iCyt Eclipse™

    FACSCalibur™ Life Technologies Attune®

    FACSCanto™ II Miltenyi Biotec MACSQuant®

    LSRFortessa™ Partec PAS

    LSR II Stratedigm S1400

    FACSAria™

    Beckman Coulter

    CyAn™ ADP

    FACSAria™ II FC 500

    Accuri™ C6 Gallios™

    ORFLO Moxi Flow™ MoFlo®Astrios™

    CyteK DxP10™ MoFlo®XDP

    Q. Whatistherightprocedureforrunningtheassay?

    1. Perform the assay as instructed in this kit manual.

    2. Determinethetypeofflowcytometeryouhaveandperformmachinesetup as instructed on Flow Cytometer Setup guide in this manual or visit: www.biolegend.com/legendplex.Openanexistingorcreateamachine-specifictemplatefordataacquisition.

    3. AnalyzethesamplesontheflowcytometerandsavetheFCSfilesina new folder.

    4. Install the LEGENDplexTM DataAnalysisSoftwarealongwiththesoft-waredongleonaPCwithoperatingsystemWindows7orWindows8(32bitor64bit).Makesuretoinstallthecorrectversion(32bitor64bitversion),dependingonyourcomputer’soperatingsystem.

    5. TransferthefoldercontainingtheFCSfilesofyourexperimenttothecomputerwherethedataanalysissoftwareisinstalled.

    6. PerformdataanalysisasinstructedintheDataAnalysisSoftwareUserGuide.

    Q. WhendoIneedtodomachinecompensation?

    Ifaflowcytometerequippedwithasinglelaserisusedforbothreporter(FL2/PE)andclassification(FL3),thencompensationisabsolutelyrequiredto compensate the signal spill over between the two channels, especially from FL2 to FL3. Please use the Setup Beads provided and follow the FLOW CYTOMETER SETUP guide in this manual.

  • biolegend.com 39

    LEGENDplex™ NHP Adipokine Panel Ifaflowcytometerislistedinthecompatibleflowcytometerlistinthis

    manualandisequippedwithtwolasers,oneforreporter(FL2/PE)andtheotherforclassification(FL4/APC),thencompensationisnotrequirediftheFLOW CYTOMETER SETUP guide in this manual is closely followed. How-ever,compensationisrecommendedforconsistentresults.

    Forotherflowcytometersnotlistedinthecompatibleflowcytometerlist, the end-user needs to set up the machine following similar guidelines (refertomanufacturer’smanualforproperinstrumentsetup).Inthiscase,machinecompensationbetweenthereporterandbeadsclassificationchannels is strongly recommended.

    Q. Isthereaspecialsoftwarerequiredfordataanalysis?

    Theflowcytometerrawdatafiles(FCS2.0,3.0)canbeanalyzedusingtheLEGENDplexTMDataAnalysisSoftwareorotherequivalentanalysissoft-ware. Download the LEGENDplexTM DataAnalysisSoftwareforfreehere:www.biolegend.com/legendplex.Asoftwaredonglewithlicensekeyisprovided in the kit.

    Fordataacquisition,nospecialsoftwareisneeded.Justusethedataac-quisitionsoftwarethatcomeswiththeflowcytometer,aslongasthedatageneratedisinFCSformat,whichmeetsFCSconvention2.0,3.0and3.1.

    Q. CanIselecttomeasureonlysomeanalyteswithinapanel?

    Yes.Thetargetswithinapanelarefullycustomizable.Thecustomercanselectanycombinationoftargetswithinapanelandorderacustomizedproduct.Pleaseuseourwebsitetargetselectiontoolfororderingcustom-izedproduct(www.biolegend.com/legendplex).

    Q. CanIselecttomeasureanalytesacrosspanels?

    Cross-panelcustomizationisalsopossible,aslongasthetotalplexsizeisno more than 13 and there is no overlapping beads region among targets selected.Forcross-panelcustomization,[email protected].

    Q. I have run out of a component from the kit. Can I use a similar compo-nentfromadifferentkit?

    TheLEGENDplex™Beads,Standards,DetectionAntibodiesandSA-PEarelot-specificandmustbeusedincombinationwitheachother.Donotmixthesecomponentsfromdifferentkitsorlots.

    Othercomponents,suchasAssayBuffer,WashBuffer,Plate,andPlateSealerarenotlot-specificandcanthereforebeexchangedbetweendiffer-ent kits and lots.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    40

    Q. My standard curve is not linear; can I use the non-linear part of the stan-dardcurveduringanalysis?

    Yes. It is possible to use the non-linear part of the standard curve for calculatingtheresults.TheLEGENDplexTMDataAnalysisSoftwareusesafive-parametercurvefittingalgorithm,whichdeterminestheminimumandmaximumdetectionconcentrationsofeachtargetandreportsthem.Ifsampleconcentrationsfallabovethemaximumdetectableconcentration,thesamplewillhavetobedilutedandreanalyzed.Ifsampleconcentra-tionsfallbelowtheminimumdetectionconcentration,itisconsideredNotDetectablebytheparticularassay.

    Q. Duringdataacquisition,whydothebeadpopulations(definedbyFSCandSSC)sometimesappeartobedispersedorshifted?

    Thisisusuallycausedbyafastflowrateorasuddenchangeinflowrate.Therearethreeflow-throughsettings:low,medium,andhigh.Ifyouareusingalowflowrateandthenchangetomediumandhighflowrate,thesuddenchangeofflowratemaysometimesresultinadispersedorshiftedbeadpopulation.Therefore,itisnotrecommendedtochangeflowrateduringdataacquisition.Thebestwayistorunthesampleinsetupmodeusinganidealflowrate,andoncethepopulationisstable,thenchangeittoacquisitionmode.

    Q. DoestheLEGENDplex™DataAnalysisSoftwarerunonAppleMacintosh?

    Notyet.ThecurrentversionofthesoftwarehastorunonaPCwithoperatingsystemWindows7orWindows8(32bitor64bit).IfyourflowcytometerisconnectedtoaMac,afterdataacquisition,theentirefoldercontainingthedata(FCSfiles)shouldbetransferredtoaPCandanalyzed.

    Q. Whydoeseachkitincludeasoftwaredongle?

    The dongle allows you to use our LEGENDplexTMsoftwarefordataanalysisand each dongle contains 1 million points.

    Q. Iranoutofpointsonmydongle.CanIgetanewoneseparately?Or,canIuseadonglefromonekitforanotherkit?

    Dongle is not sold separately. Contact BioLegend if you need a new dongle. The dongle can be used across kits.

  • biolegend.com 41

    LEGENDplex™ NHP Adipokine Panel

    Chapter 7: ASSAY CHARACTERIZATION

    Standard Curve

    This standard curve was generated using the LEGENDplexTM NHP Adipokine Panelfordemonstrationpurposeonly.Astandardcurvemustberunwitheach assay.

    AssaySensitivity

    Theassaysensitivityorminimumdetectableconcentration(MDC)isthetheoreticallimitofdetectioncalculatedusingtheLEGENDplexTM Data AnalysisSoftwarebyapplyinga5-parametercurvefittingalgorithm.

    Analyte MDC (pg/mL)NHPAdiponectin 39.0

    NHP Adipsin 4.3

    NHPRBP4 3.4

    NHP MCP-1 1.3

    NHPIL-1β 1.3

    NHP IP-10 0.7

    NHP IL-10 2.2

    NHP IL-8 1.2

    NHPLeptin 2.4

    NHP IL-6 2.4

    1

    10

    100

    1000

    10000

    1 100 10000

    MFI

    Concentration (pg/mL)

    Adiponectin Adipsin RBP4 MCP-1 IL-1b IP-10 IL-10 IL-8 Leptin IL-6 IFN-r Resistin TNF-a

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    42

    NHPIFN-γ 1.9

    NHPResistin 2.2

    NHPTNF-α 0.6

    Cross-Reactivity

    Therewasnoornegligiblecross-reactivityfoundfortheanalytesinthispanel.

    Accuracy (Spike Recovery)

    Forspikerecoveryincellculturemedium(CCM),RPMIorDMEMwith10%FCSwasfirstdilutedtwo-foldwithAssayBufferandspikedwithtargetpro-teinsatthreedifferentlevelswithintheassayrange.Thespikedsampleswerethenassayed,andthemeasuredconcentrationswerecomparedwiththe expected values.

    For spike recovery in urine, Rhesus and Cynomolgus monkey urine samples (n=16)werefirstdilutedtwo-foldwithAssayBufferandspikedwithtargetproteinsatthreedifferentlevelswithintheassayrange.Thespikedsam-pleswerethenassayed,andthemeasuredconcentrationswerecomparedwith the expected values.

    Analyte Recovery in CCM Recovery in Urine NHPAdiponectin 96% 96%NHP Adipsin 94% 69%NHPRBP4 70% NANHP MCP-1 92% 109%NHPIL-1β 91% 62%NHP IP-10 89% 91%NHP IL-10 78% 106%NHP IL-8 108% 116%NHPLeptin 94% 94%NHP IL-6 89% 115%

  • biolegend.com 43

    LEGENDplex™ NHP Adipokine Panel

    NHPIFN-γ 110% 131%NHPResistin 84% 103%NHPTNF-α 87% 105%

    NA=NotApplicableduetohighendogenouslevels

    LinearityofDilution

    ForspikelinearityinCCM,RPMIorDMEMwith10%FCSwasfirstdilutedtwo-foldwithAssayBufferandspikedwithaknownconcentrationoftargetproteins.Thespikedsampleswereseriallydiluted1:2,1:4,1:8withAssaybufferandassayed.Themeasuredconcentrationsofseriallydilutedsamples were compared with that of the spiked samples.

    For sprike linearity in urine, Rhesus and Cynomolgus monkey urine samples (n=16)weredilutedtwo-foldwithAssayBufferandspikedwithaknownconcentrationoftargetproteins.Thespikedsampleswereseriallydiluted1:2,1:4,1:8withAssayBufferandassayed.Themeasuredconcentrationsof serially diluted samples were compared with that of the spiked samples.

    Analyte Linearity in CCMLinearity in

    Urine

    NHPAdiponectin 95% 102%

    NHP Adipsin 115% 113%NHPRBP4 117% 100%NHP MCP-1 114% 110%NHPIL-1β 114% 105%NHP IP-10 111% 114%NHP IL-10 121% 103%NHP IL-8 119% 108%NHPLeptin 113% 109%NHP IL-6 111% 103%NHPIFN-γ 120% 96%NHPResistin 125% 110%NHPTNF-α 114% 103%

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    44

    Intra-Assay Precision

    Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedin one assay with 16 replicates for each sample. The intra-assay precision was calculated as below.

    Analyte Sample Mean (pg/mL) STDEV %CV

    NHP Adiponectin

    Sample 1 923.9 108.8 12%Sample 2 3639.8 313.3 9%

    NHP Adipsin

    Sample 1 204.8 8.5 4%Sample 2 787.0 29.0 4%

    NHPRBP4Sample 1 256.5 13.9 5%Sample 2 1024.5 41.9 4%

    NHP MCP-1Sample 1 48.0 2.5 5%Sample 2 199.3 9.1 5%

    NHPIL-1βSample 1 50.4 3.2 6%Sample 2 202.1 11.2 6%

    NHP IP-10Sample 1 65.8 3.7 6%Sample 2 265.7 12.1 5%

    NHP IL-10Sample 1 40.4 4.2 10%Sample 2 143.6 7.8 5%

    NHP IL-8Sample 1 49.4 3.2 7%Sample 2 195.3 10.1 5%

    NHPLeptinSample 1 52.8 3.0 6%Sample 2 201.7 8.8 4%

    NHP IL-6Sample 1 51.7 2.9 6%Sample 2 180.5 8.8 5%

    NHPIFN-γSample 1 47.5 4.7 10%Sample 2 182.5 9.9 5%

    NHPResistinSample 1 52.2 2.4 5%Sample 2 194.7 7.1 4%

    NHPTNF-αSample 1 54.2 3.9 7%Sample 2 206.9 10.4 5%

  • biolegend.com 45

    LEGENDplex™ NHP Adipokine Panel

    Inter-Assay Precision

    Twosampleswithdifferentconcentrationsoftargetproteinswereanalyzedin three independent assays with 3 replicates for each sample. The inter-assay precision was calculated as below.

    Analyte Sample Mean (pg/mL) STDEV %CV

    NHP Adiponectin

    Sample 1 987.4 131.7 13%Sample 2 3554.8 336.5 9%

    NHP Adipsin

    Sample 1 214.9 17.5 8%Sample 2 775.2 45.8 6%

    NHPRBP4Sample 1 274.3 30.7 11%Sample 2 1020.1 80.5 8%

    NHP MCP-1Sample 1 50.9 4.5 9%Sample 2 195.6 14.5 7%

    NHPIL-1βSample 1 52.6 4.9 9%Sample 2 205.2 16.4 8%

    NHP IP-10Sample 1 76.0 10.6 14%Sample 2 290.3 34.0 12%

    NHP IL-10Sample 1 42.4 4.7 11%Sample 2 140.8 10.0 7%

    NHP IL-8Sample 1 58.2 8.1 14%Sample 2 201.1 19.1 9%

    NHPLeptinSample 1 59.5 7.6 13%Sample 2 212.4 18.2 9%

    NHP IL-6Sample 1 55.5 6.2 11%Sample 2 180.2 13.3 7%

    NHPIFN-γSample 1 53.4 7.8 15%Sample 2 185.1 17.6 10%

    NHPResistinSample 1 55.3 5.0 9%Sample 2 193.9 10.6 5%

    NHPTNF-αSample 1 58.6 7.0 12%Sample 2 202.5 16.1 8%

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    46

    Biological Samples

    Urine

    NormalRhesusmonkeyurinesamples(n=8)weretestedforendogenouslevelsofadipokines.Theconcentrationsmeasuredareshownbelow:

    Analyte Range (pg/mL)% of

    Detectable

    Mean of detectable

    (pg/mL)RhesusAdiponectin 3,244-714,352 100% 187,980

    Rhesus Adipsin 104.8-1,159.6 100% 564.0

    RhesusRBP4 665.0 - 201,520 100% 93,739

    Rhesus MCP-1 3.3-13,484 100% 3,594

    RhesusIL-1β 7.4-19.1 100% 11.4

    Rhesus IP-10 3.5 - 226 100% 72.7

    Rhesus IL-10 2.5 - 325.6 100% 60.9

    Rhesus IL-8 ND - 923.9 75% 167.9

    RhesusLeptin ND - 12.8 25% 10.8

    Rhesus IL-6 ND - 190.1 38% 68.9

    RhesusIFN-γ 2.8-31.7 100% 8.5

    RhesusResistin 5.6-24,524 100% 7,185

    RhesusTNF-α ND -7.7 25% 5.4 ND=NotDetectable

    NormalCynomolgusmonkeyurinesamples(n=8)weretestedforendogenouslevelsofadipokines.Theconcentrationsmeasuredareshownbelow:

    Analyte Range (pg/mL)% of

    Detectable

    Mean of detectable

    (pg/mL)Cynomolgus Adiponectin 3,996-979,288 100% 302,385

    Cynomolgus Adipsin 125.4-6,091 100% 1,767

    CynomolgusRBP4 6,013-166,038 100% 106,457

    Cynomolgus MCP-1 228.1-11,386 100% 2,396CynomolgusIL-1β 1.8-9.7 100% 5.6

  • biolegend.com 47

    LEGENDplex™ NHP Adipokine Panel

    Cynomolgus IP-10 2.3 - 189.6 100% 40.5

    Cynomolgus IL-10 19.7 - 270.5 100% 120.4

    Cynomolgus IL-8 ND - 18.0 25% 7.7

    CynomolgusLeptin ND - 31.0 38% 16.5

    Cynomolgus IL-6 ND 0% ND

    CynomolgusIFN-γ ND - 72.0 75% 18.4

    CynomolgusResistin 228.8-13,629 100% 6,818

    CynomolgusTNF-α ND - 9.9 38% 4.9 ND=NotDetectable

    Cell Culture Supernatant

    RhesusPBMC(1x106cells/mL)wereculturedundervariousconditions(PHA,5µg/mL;PMA,20ng/mL;LPS, 1 µg/mL;Ionomycin(I),500ng/mL;IFN-γ,100ng/mL).Supernatantswerecollectedafter3days.Allsamplesweredilutedtwo-foldwithAssayBufferandassayedwiththeLEGEND-plexTMNHPAdipokinePanel.Theresults(allinpg/mL)aresummarizedbelow.

    Analyte (-) PHA PMA+I LPS IFN-γ + LPS

    Rhesus Adiponectin 26.6 ND 24.6 18.8 47.2

    Rhesus Adipsin 4,673 897.8 455.3 935.6 1,045

    RhesusRBP4 21.2 13.1 20.6 19.0 19.8Rhesus MCP-1 1,964 10,259 1,375 12,832 11,026

    RhesusIL-1β ND 54.9 24.7 72.2 91.8Rhesus IP-10 917.9 5,624 1,951 600.3 5,015Rhesus IL-10 7.3 79.2 247.8 75.0 38.9Rhesus IL-8 3,249 12,492 12,907 12,907 12,907RhesusLeptin ND ND ND ND ND

    Rhesus IL-6 ND 2146 54.8 2,225 6,191RhesusIFN-γ ND 1,037 8,374 3.3 >13,977RhesusResistin 106.7 89.6 89.0 103.6 107.4RhesusTNF-α 7.2 70.3 454.9 62.4 888.4ND=NotDetectable

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    48

    CynomolgusPBMC(1x106cells/mL)wereculturedundervariouscondi-tions(PHA,5µg/mL;PMA,20ng/mL;LPS, 1 µg/mL;Ionomycin(I),500ng/mL;IFN-γ,100ng/mL).Supernatantswerecollectedafter3days.Allsamplesweredilutedtwo-foldwithAssayBufferandassayedwiththeLEGENDplexTMNHPAdipokinePanel.Theresults(allinpg/mL)aresumma-rizedbelow.

    Analyte (-) PHA PMA+I LPS IFN-γ + LPS

    Cynomolgus Adiponectin 47.2 15.4 13.8 31.2 17.0

    Cynomolgus Adipsin 7,914 1,506 608.8 1,349 1,648

    Cynomolgus RBP4 18.6 20.6 26.3 21.8 22.5

    Cynomolgus MCP-1 663.4 11,880 2,465 13,350 11,442

    Cynomolgus IL-1β ND 106.7 21.0 100.0 175.9

    Cynomolgus IP-10 2,338 6,778 2,247 522.4 6,548

    Cynomolgus IL-10 10.2 135.9 458.5 170.4 31.7

    Cynomolgus IL-8 1,776 12,907 12,907 12,907 12,907

    Cynomolgus Leptin ND ND ND ND ND

    Cynomolgus IL-6 2.3 3,080 53.1 2,397 6,191

    Cynomolgus IFN-γ ND 126.0 8,909 3.0 >13,977

    Cynomolgus Resistin 129.2 126.3 112.3 125.4 134.1

    Cynomolgus TNF-α 6.3 151.9 728.1 140.9 3,401

    ND=NotDetectable

  • biolegend.com 49

    LEGENDplex™ NHP Adipokine Panel

    TROUBLESHOOTING

    Problem Possible Cause Solution

    Bead popula-tionshiftingupward or downward dur-ingacquisition

    The strong PE signal from high concentra-tionsamplesorstan-dards may spill over to classificationChannel(e.g.,FL3/FL4/APC)and mess up the bead separation.

    OptimizeinstrumentsettingsusingKitSetup Beads, and make appropriate com-pensationbetweenchannels.

    Filter plate will not vacuum or some wells clogged

    Vacuum pressure is insufficientorvacuummanifold does not seal properly.

    Increase vacuum pressure such that 0.2mLbuffercanbesuctionedin3-5seconds. Clean the vacuum manifold and make sure no debris on the manifold. Press down the plate on the manifold to make a good seal.

    Samples have insoluble particlesorsampleistooviscous(e.g.,serumandplasmasamples)

    Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.

    Ifsomewellsarestillcloggedduringwashing, try the following:

    1).Addbuffertoallthewells,pipetteupand down the clogged wells and vacuum again.

    2).Useapieceofcleanwipe,wipetheun-der side of the clogged wells and vacuum again.

    3).Takeathinneedle(e.g.,insulinnee-dle),whileholdingtheplateupward,pokethelittleholeundereachofthecloggedwells and vacuum again. Do not poke too hard or too deep as it may damage the filterandcauseleaking.

    Filter plate was used without pre-wet.

    Pre-wetplatewithwashbufferbeforerunning the assay.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    50

    Insufficientbead count or slow reading

    Beads inappropriately prepared

    Sonicate bead vials and vortex just prior toaddition.Agitatethepre-mixedbeadsintermittentlyinreservoirwhilepipettingthis into the plate.

    Samples cause beads aggregationduetoparticulatematterorviscosity

    Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.

    Beads were lost during washing for in-tube assay

    Make sure beads are spun down by visu-allycheckthepellet(beadsareinlightblueorbluecolor).Beverycarefulwhenremoving supernatant during washing.

    Probemaybepartiallyclogged

    Sample probe may need to be cleaned, or if needed, probe should be removed and sonicated.

    Plate leaked

    Vacuum pressure set too high

    Adjust vacuum pressure such that 0.2 mL buffercanbesuctionedin3-5seconds.Do not exceed 10” Hg of vacuum.

    Plate set directly on table or absorbent tow-elsduringincubationsorreagentadditions

    Set plate on plate holder or raised edge sobottomoffilterisnottouchinganysurface.

    Liquid present on the under side of the plate aftervacuum

    Afterwashing,pressdownplatefirmlyona stack of clean paper towels to dry the underside of the plate.

    Pipettetouchinganddamagedplatefilterduringadditions

    Pipettetothesideofwells.

    High back-ground

    Background wells were contaminated

    Avoidcross-wellcontaminationbychang-ingtipsbetweenpipettingwhenperform-ingtheassayusingamultichannelpipette.

    InsufficientwashesThe background may be due to non-spe-cificbindingofSA-PE.Increasenumberofwashes.

    Debris(FSC/SSC)duringsample acquisi-tion

    Debris or platelet may existinsamplesolution

    Centrifugesamplesbeforeanalyzingsamples. Remove platelet as much as possible.

  • biolegend.com 51

    LEGENDplex™ NHP Adipokine Panel

    Variationbe-tweenduplicate samples

    Beadsaggregation Sonicate and vortex the Beads prior to use.

    Multichannelpipettemay not be calibrated or inconsistent pipet-ting

    CalibratePipette.Ensuregoodpipettingpractice.Primepipettebeforeusemayhelp.

    Plate washing was not uniform

    Make sure all reagents are vacuumed out completely in all wash steps.

    Samples may contain particulatematters.

    Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.

    Low or poor standard curve signal

    The standard was in-correctlyreconstituted,stored or diluted

    Followtheprotocoltoreconstitute,storeand dilute standard. Double check your calculation.

    Wrong or short incuba-tiontime

    Ensurethetimeofallincubationswasappropriate.

    Signals too high, standard curves satu-rated

    PMTvalueforFL2/PEset too high

    MakesurethePMTsettingforthere-porter channel is appropriate

    Plateincubationtimewas too long Useshorterincubationtime.

    Sample read-ings are out of range

    Samples contain no or below detectable levels of analyte

    Make sure the experiment to generate thesamplesworked.Useproperpositivecontrols.

    Samplesconcentrationshigher than highest standard point.

    Dilutesamplesandanalyzeagain.

    Standard curve was saturated at higher end of curve.

    MakesurethePMTsettingforthere-porter channel is appropriate. Use shorter incubationtimeifincubationtimewastoo long

    Missed beads populationsduring reading, ordistributionis unequal

    Sample may cause some beads to ag-gregate.

    Centrifuge samples just prior to assay setup and use supernatant. If high lipid contentispresent,removelipidlayeraftercentrifugation.Samplemayneeddilutionif too viscous.

    Beadspopulationsarenot mixed properly

    Makesureallbeadpopulationsaremixed.and in similar numbers.

  • Tel: 858-768-5800

    LEGENDplex™ NHP Adipokine Panel

    52

    PLA

    TE M

    AP

    (for

    in-p

    late

    ass

    ay)

    1 2

    3 4

    5 6

    7 8

    9 10

    11

    12

    A

    C0

    C4

    Sa

    mpl

    e1

    Sa

    mpl

    e5

    Sa

    mpl

    e9

    Sa

    mpl

    e 13

    Sa

    mpl

    e17

    Sa

    mpl

    e21

    Sa

    mpl

    e25

    Sa

    mpl

    e29

    Sa

    mpl

    e33

    Sa

    mpl

    e 37

    B

    C0

    C4

    Sa

    mpl

    e1

    Sa

    mpl

    e5

    Sa

    mpl

    e9

    Sa

    mpl

    e 13

    Sa

    mpl

    e17

    Sa

    mpl

    e21

    Sa

    mpl

    e25

    Sa

    mpl

    e29

    Sa

    mpl

    e33

    Sa

    mpl

    e37

    C

    C1

    C5

    Sa

    mpl

    e2

    Sa

    mpl

    e6

    Sa

    mpl

    e10

    Sa

    mpl

    e 14

    Sa

    mpl

    e18

    Sa

    mpl

    e22

    Sa

    mpl

    e26

    Sa

    mpl

    e30

    Sa

    mpl

    e34

    Sa

    mpl

    e38

    D

    C1

    C5

    Sa

    mpl

    e2

    Sa

    mpl

    e6

    Sa

    mpl

    e10

    Sa

    mpl

    e 14

    Sa

    mpl

    e18

    Sa

    mpl

    e22

    Sa

    mpl

    e26

    Sa

    mpl

    e30

    Sa

    mpl

    e34

    Sa

    mpl

    e38

    E

    C2

    C6

    Sa

    mpl

    e3

    Sa

    mpl

    e7

    Sa

    mpl

    e11

    Sa

    mpl

    e 15

    Sa

    mpl

    e19

    Sa

    mpl

    e23

    Sa

    mpl

    e27

    Sa

    mpl

    e31

    Sa

    mpl

    e35

    Sa

    mpl

    e39

    F

    C2

    C6

    Sa

    mpl

    e3

    Sa

    mpl

    e7

    Sa

    mpl

    e11

    Sa

    mpl

    e 15

    Sa

    mpl

    e19

    Sa

    mpl

    e23

    Sa

    mpl

    e27

    Sa

    mpl

    e31

    Sa

    mpl

    e35

    Sa

    mpl

    e39

    G

    C3

    C7

    Sa

    mpl

    e4

    Sa

    mpl

    e8

    Sa

    mpl

    e12

    Sa

    mpl

    e 16

    Sa

    mpl

    e20

    Sa

    mpl

    e24

    Sa

    mpl

    e28

    Sa

    mpl

    e32

    Sa

    mpl

    e36

    Sa

    mpl

    e40

    H

    C3

    C7

    Sa

    mpl

    e4

    Sa

    mpl

    e8

    Sa

    mpl

    e12

    Sa

    mpl

    e 16

    Sa

    mpl

    e20

    Sa

    mpl

    e24

    Sa

    mpl

    e28

    Sa

    mpl

    e32

    Sa

    mpl

    e36

    Sa

    mpl

    e40

  • biolegend.com 53

    LEGENDplex™ NHP Adipokine Panel

    RA

    CK

    MA

    P (

    for

    in-t

    ub

    e as

    say)

    1

    2 3

    4 5

    6 7

    8 9

    10

    11

    12

    A

    C0

    C0

    C1

    C1

    C2

    C2

    C3

    C3

    C4

    C4

    C5

    C5

    B

    C6

    C6

    C7

    C7

    Sam

    ple

    1 Sa

    mp

    le

    1 Sa

    mp

    le

    2 Sa

    mp

    le

    2 Sa

    mp

    le

    3 Sa

    mp

    le

    3 Sa

    mp

    le

    4 Sa

    mp

    le4

    C

    Sam

    ple

    5

    Sam

    ple

    5

    Sam

    ple

    6

    Sam

    ple

    6

    Sam

    ple

    7

    Sam

    ple

    7

    Sam

    ple

    8

    Sam

    ple

    8

    Sam

    ple

    9

    Sam

    ple

    9

    Sam

    ple

    10

    Sa

    mp

    le

    10

    D

    Sam

    ple

    11

    Sa

    mp

    le

    11

    Sam

    ple

    12

    Sa

    mp

    le

    12

    Sam

    ple

    13

    Sa

    mp

    le

    13

    Sam

    ple

    14

    Sa

    mp

    le

    14

    Sam

    ple

    15

    Sa

    mp

    le

    15

    Sam

    ple

    16

    Sa

    mp

    le

    16

    E Sa

    mp

    le

    17

    Sam

    ple

    17

    Sa

    mp

    le

    18

    Sam

    ple

    18

    Sam

    ple

    19

    Sa

    mp

    le

    19

    Sam

    ple

    20

    Sa

    mp

    le

    20

    Sam

    ple

    21

    Sa

    mp

    le

    21

    Sam

    ple

    22

    Sa

    mp

    le

    22

    F Sa

    mp

    le

    23

    Sam

    ple

    23

    Sa

    mp

    le

    24

    Sam

    ple

    24

    Sa

    mp

    le

    25

    Sam

    ple

    25

    Sa

    mp

    le

    26

    Sam

    ple

    26

    Sa

    mp

    le

    27

    Sam

    ple

    27

    Sa

    mp

    le

    28

    Sam

    ple

    28

    G

    Sam

    ple

    29

    Sa

    mp

    le

    29

    Sam

    ple

    30

    Sa

    mp

    le

    30

    Sam

    ple

    31

    Sa

    mp

    le

    31

    Sam

    ple

    32

    Sa

    mp

    le

    32

    Sam

    ple

    33

    Sa

    mp

    le

    33

    Sam

    ple

    34

    Sa

    mp

    le

    34

    H

    Sam

    ple

    35

    Sa

    mp

    le

    35

    Sam

    ple

    36

    Sa

    mp

    le

    36

    Sam

    ple

    37

    Sa

    mp

    le

    37

    Sam

    ple

    38

    Sa

    mp

    le

    38

    Sam

    ple

    39

    Sa

    mp

    le

    39

    Sam

    ple

    40

    Sa

    mp

    le

    40

  • LEGENDplex™ Kits are manufactured by BioLegend Inc. 9727 Pacific Heights Blvd.San Diego, CA 92121Tel: 1.858.768.5800Tel: US & Canada Toll-Free: 1.877.Bio-Legend (1.877.246.5343)Fax: 1.877.455.9587Email: [email protected]

    For a complete list of world-wide BioLegend offices and distributors, please visit our website at: biolegend.com

    ®

    76599_V01