LED Source Tests
RSL August 2012
http://www.oceanoptics.com/technical/LLSLightSource.pdf
Blu Loop
LLS Cool White – see the last page of:
http://www.oceanoptics.com/Products/bluloop.asp
t
600 um core fiber, goes to small integrating sphere; plan is to swap at source
fiber going to source(s); P/N QP-600-2-SR
integrating sphere; M/N FOIS-1
SEI M/N SR-3500 spectroradiometer
SR-3500 input optical fiber
zoomed sphere/SR-3500 fiber end (it is inside the sphere)
test plan• Warm up: take a dark, turn on source, take one lite (an average of
10 scans) every 5 min for 3 – 4 hours• Repeatability: repeat with the other LED source by changing the
600 um patch fiber at the source end (also covers warm-up)• Do this for a couple of days (tests repeatability of source level and
effect of decoupling and recoupling the SMA connector at the source/fiber junction)
• Turn on both sources, leave them on several days, swap fiber back and forth, acquiring data every 2 h or so
• Move SEI bare fiber to NPR to get a calibration• If time and data look good, move sources to
temperature/humidity chamber and measure levels