Upload
dangthuan
View
214
Download
0
Embed Size (px)
Citation preview
LC ConsumablesTips and Tricks
Chromatography Workflow
FY 2019
Solution
Suryadi
Sales Chemical Ultra Standard & Consumable-Chromatography Sipplies
Agilent Technologies
Maret 01, 2019
Chromatography Workflow
SampleCollection
Analysis
SamplePreparation
Detection
Separation
2 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Mobile Phase Reservoir
7 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Care of Mobile Phase
Particulates in the LC solvents or microbial growth in the solvent bottle may plug thesolvent inlet filter, reducing pump performance.
•
•
•
•
If possible, use sterile solvent bottles.
Filter solvents through filters (< 0.45 µm).
Replace the solvents every two days (do not re-filter).
Avoid exposure to direct sunlight (use Aluminum foil, shield from windows) or use amber glass bottles.
• Consider adding 5-10% organic to the aqueous mobile phase to inhibit microbial growth.
8 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Solvent and Buffer Filtration
As internal diameter of connecting capillaries, particle in the column and pore size of columninlet frit are getting smallermust.
and smaller, filtration of mobile phase is becoming an absolute
HPLC use 0.45 m filter
UHPLC use 0.2 m filter
p/n: 3150-0577
9 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
PN Description
5191-4339 Filter membrane, PTFE, 47 mm, 0.20 µm, 100/pk
5191-4336 Filter membrane, PTFE, 47 mm, 0.45 µm, 100/pk
5191-4341 Filter membrane, nylon 47 mm, 0.20 µm, 100/pk
5191-4338 Filter membrane, nylon 47 mm, 0.45 µm, 100/pk
5191-4340 Filter membrane, cellulose 47 mm, 0.2 µm, 100/pk
3150-0576 Filter membrane, regenerated cellulose, 47 mm, 0.45 μm, 100/pk
Solvent Inlet Filter
1.
2.
3.
Remove the solvent inlet filter.
Keep a clean one on hand to compare
Replace the glass filter (or soakconc. H2NO3 for 1 hour.
Flush well with water
in
4.
5. Reinstall the solvent inlet filter
p/n: 5062-8517 p/n: 5041-2168 p/n: 01018-60028
10 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
InfinityLab Solvent Bottles
• Agilent’s new solvent bottles have been
redesigned for easier use and safer
handling, lowering the risk ofthe lab.
accidents in
• The new solvent bottles feature an
ergonomic design
and ergonomically
on both sides of
with a slimmer profileshaped gripping zones
the bottle, promoting
easier and safer usage.
11 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Solvent Bottles
• Borosilicate type 1 ClassA2 glass reservoirs.
• Wash glassware separately from other containers.
• Clean reservoirs – dishwasher, no detergent.
(If more aggressive cleaning is required,
sonicate with 10% formic or nitric acid, then water,
then methanol or acetonitrile, repeat two more time)
• Final rinse with DI water.
• Put in 400 degree muffle furnace for 6 hours.
12 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Key Factor Affecting pH and Concentration of Buffer
Evaporation
14 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
InfinityLab Stay Safe CapWasteMobile Phase and
15 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Stay Safe Cap for Mobile Phase
The safety caps seal the solvent containers so that harmfulvapors do not escape into the laboratory.
The integrated exhaust valve maintains a constant
pressure in the solvent container and allows proper solvent
flow to the HPLC system.
The valve membrane opens when the
HPLC pump is operating and allows air
flow into the bottle, preventing a partial
vacuum inside the solvent container.
to
When the pump stops, the valve membrane
closes and solvent vapors are blocked
from escaping into the laboratory. Venting
valve should be replaced every 6 months
in order to flawless operation.
16 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Open bottle (A) Standard cap (B) Agilent InfinityLab Closed cap (D)
Stay Safe cap (C)
5991-7755EN
17 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Acetonitrile MSDS
19 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Stay Safe Cap for Waste
21 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Routine Maintenance Procedure
•
•
•
•
•
Replace
Replace
Replace
the
the
the
seal and pistons
PTFE frit
cartridge in active inlet valve
Clean
Clean
the outlet ball valve
or replace solvent inlet filter
p/n: 01018-22707
23 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Injection Valve
Standard Autosampler G1329A
54
32
1
Injection valve
1.2.
3.
4.
5.
Stator screws
Stator head
Rotor seal
Isolation seal
Valve body
0100-18490100-1853 0101-1416
0100-1853: Rotor seal (Vespel)0100-1849: Rotor seal (Tefzel),
0101-1416: Rotor seal (PEEK)
25 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Needle and Needle SeatStandard Autosampler G1329A
Needle seat
i.d. 0.17 mm: p/n G1313-87101
i.d. 0.12 mm: p/n G1313-871-3
Needle: p/n G1313-87201
26 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Autosampler Vials
27 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Re-washed Glass Surface
scratched, abraded
and plaques
Re-washed Glass Surface
scratched, abraded
and plaques
To Reuse: Physical Deformations On The Glass Surface
New Glass Surface New Glass Surface
X100
X500
30
smooth and free of
surface scratches
and abrasions
smooth and free of
surface scratches
and abrasions
Amber GlassClear Glass
LC Columns Compartment
31 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Problems!!!
• Peak shape•
•
Peak separation
Peak tailing
• Column blockage• High back pressure
• Column selection• Is it the right column for my application?
32 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
• Delays peak elution time.
• Broadens peaks reducing
efficiency and resolution.
Problem
Effect of Extra-column Volumes
• Decreases peak height and thus S/N.Resolution 0.961mAU
400
Peakwidth 0.037 min300
200
Peakwidth 0.038 min100
0
0.5 1 1.5 2 2.5 min
Resolution 1.902mAU300
250
Peakwidth 0.019 min200
150
100Peakwidth 0.018 min
50
0
0.5 1 1.5 2 2.5 min
33 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
HPLC Instrument Configuration
Design differences between LC instrumentation
– such as power range, delay volume, mixing
behavior, temperature control, extra column
volume and detector cellthe results of analysis.
design – all effect to
Delay volume
gradient mixing behavior
pressure x flow rate
Delay volumeextra column volume
Injection volume
Temperature profileextra column volume
Data rateextra column volume
path-length
35 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Del
ayv
olu
me
Extr
a c
olu
mn
volu
me
Resolution
Retention Time
Sensitivity
Retention Time
Resolution
Retention Time
Resolution
Sensitivity
Retention Time
Resolution
FittingConventional LC
37 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Infinity A-Line Connector
InfinityLab Quick Connect UHPLC columnfittings for truly dead-volume-free fluidic connections
Finger tight up to•
•
Tool-free connection up to 1300 bar
Spring loaded design for easy and truly zero-dead volume connection
600 and 1300 bar
• Removabletypes
and reusable for all column
39 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Zero-dead Volume
Agilent column with InfinityLab Quick Connect
Agilent column with InfinityLab Quick Turn
Too short screw
Competitor column with Swagelog
Dead volume
Competitor column with InfinityLabQuick Connect
40 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Ease of Use
1. Screw the fitting (blue wheel) with lever in
“open” position onto the column – or column
onto the fitting
2. Stop when finger-tight and turn the lever
3. Finished… in a seconds
41 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Finger tight does not mean as far as your finger can go
→ Stop as soon as you can feel the first resistance
Excessive force will damage the fitting!
ressureDetermining The Cause and Correcting High Back
Check pressure with/without column - many pressuredue to blockages in the system or guard column.
problems are
•
•
Remove column - pressure still high?
Remove guard - pressure still high?
If column pressure is high:
• Wash column
Eliminate column contamination and plugged packing
High molecular weight/adsorbed compounds
Precipitate from sample or buffer
•
•
•
• Change frit – Clear plugged frit PREVENT THIS!
42 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
etter Choice
s
el Regulatory statement (if applicable)
The TricksPrevent Technique—The B
• Use column protection
In-line filters
Guard columns
•
•
•
•
Filter samples
Filter buffered mobile phase
•
•
Sample clean-up (i.e. SPE)
Appropriate column flushing
43 January 25, 2019 Title Confidentiality lab
Use Appropriate Guard Column
Guard cartridges should be changed when there is a big increase in backpressure
44 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Detector in LC
• UV-Visible detector
• Diode array detector
• Fluorescence detector
• Refractive index detector
• Evaporative light scattering detector
• Electrical conductivity detector
• Electrochemical detector
• Mass spectrometer
46 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Certified Lamp
By providing higher sensitivity, Agilent lamps can extend
detection capabilities and improve qualification at trace
levels – for more than 2,000 hours of use
47 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)
Caring for your UV detector flow cell
• Many flow cells have quartz windows or fused silicacomponents, so avoid using alkaline solutions with
pH greater than 9.5, which can attack these components
and impair optical performance.
• To prevent crystallization inside the flow cell, always flushbuffers or salt solutions, especially at high pHs.
it with water after running
• When your LC is unused overnight, make sure your flow cell contains at least 10%organic mobile phase to prevent algae growth.
• Some modern flow cells use coated or uncoated fused silica capillaries. You canadversely affect coated capillaries by turning on the UV lamp without solvent flow, so
read the manufacturer ’s recommendations carefully.
48
Flushing The Flow Cell
If your system was shut down unexpectedly during an unattended sequencerun, it is possible that the sample is still trapped in your flow cell.
If the system was stopped overnight with salt buffer, it is important to start the
system flush with a lower flow rate and watch the system pressure. A warm-
water flush using the 60°C setting on your column compartment is an
effective way to clean your UV flow cell.
49
Where can the issue be?
There are 7 basic areas from where the problem may arise:
1.
2.
3.
4.
5.
6.
7.
Injector
Gas Flow/Leaks
GC Column
Detector
Electronics
Gas Supply / Purity of Gases / Filters
Sample Storage - Vials and Closures
But of course it can always be a COMBINATION. A healthy GC can ELIMINATEsome of these problems before they begin to surface.
For example, Peak Tailing
Injector or Column is “Active”
- Reversible adsorption of active compounds like the ones with -OH, -NH, -SH
Flow path problem
-
-
Dead volume, obstruction, poor installation, leaks or
Severe column contamination.
Miscellaneous
- overloading of PLOT columns, co-elution, polarity mismatch between phase, solute or solvent, and some compounds always tail.
*Tip = Inject a light hydrocarbon, should not tail unless there is aflow path problem.
It all begins with “ Maintenance”It is not
always clear when I need to maintain
theinstrument
Sometimes Iforget to do
maintenanceWhen?
Is it time for maintenance? I’m not sure which task I
need toperform to correct or prevent a problem
I’d like some suggestions on a typical maintenance protocol and
best practices
What?
I needresources to improve my
maintenanceskills
I don’t feel comfortable
doing maintenance
How?
5 5 September 2018 Agilent Technologies
How often do I need to do complete routine maintenance?
More OftenLess Often
6 5 September 2018 Agilent Technologies
GC Column Maintenance
1. Bake out the column forthan 2 hours.
no more
2. Cut off around 15-30cm of theinlet end of the column.
3. Cut off more column (repeat asnecessary)
Inlet MaintenanceWhy and when is it needed?• Every analysis means wear and
inlet partstear on
••
Every injection introduces contaminationHigh temperature stress of GC aggravateswear••
•
•
How many samples?What is the sample type?How critical are the results?Do you have a regular standard procedure,SOP for maintenance?
Inlet Maintenance Schedule
Part Many samples Less samples
heavy matrix clean samples
Septum Daily Weekly
Liner Daily Monthly
Gold seal/washer Weekly Monthly
O-ring With liner With liner
Ferrule With column With trimming/exchange
trimming
Split vent trap 3 months Yearly
What are some potential Causes of Split Peaks?
INJECTOR (poor sample introduction)- Injecting the sample twice or injection too slow(manual) or in needle
- Mixed sample solvent (polarity difference)
VOLATILITY- High boilers dropping out on Cold Spots- Transfer line temperature- Unions or fittings not tracking with column temp.
INJECTOR (activity)- Breakdown (not really a split peak, 2 peaks)- Sample degradation in injector
Agilent Inert Flow Path (IFP) Solution Focus
Deactivated Inlet Weldments,
Shell and Transfer LinesDeactivated – FPD, TCD,NPD/FID Jets
Ultra Inert Inlet Liner
UltiMetal Plus treatedFlexible Metal Ferrules
UltiMetal Tubing
Ultra Inert Press-FitUltiMetal Plus treatedCapillary Flow
Technology Devices
Ultra Inert GC ColumnUltra Inert Gold Seal
Gas Clean Gas Purifier
Why Ultra Inert?
Regulatory bodies drive for lower detection limits forincreasingly active and complex samples
Active flow paths causes
• Unreproducible performance
• Waste time, resources and hinder productivity
• Reduce capability of expensive equipment
• Result in sample reruns or reporting an incorrect result
These problems can all be avoided.column selection and maintenance
Let’s start with
Ultra Inert ColumnsPeak shape on DB-5ms UI
4-picoline TMP 1,2-pentanediol
App Note 5989-8665EN
Agilent J&W DB-5ms Ultra Inert
30m x 0.25mm x 0.25µm
5% phenyl Competitor
30m x 0.25mm x 0.25µm
Ultra Inert Inlet Liners
Ultra Inert deactivated inlet liners (Proprietary vapor deposition process) provides higherresponse for sensitive compounds.
Productivity:Touchless packaging with pre-installed o-ring:
quick & easy hassle free installation
Issue: Peak Response Variations
Injector or Column is active/contaminated----
Irreversible adsorption of active compoundsDecomposition of the sample
(-OH, -NH, -SH)
Temperature change or injector discrimination in the inletEvaporation of the sample
Contamination from Liquid Soap on fingers duringcolumn installation
Column:
Carrier:
Injector:
Detector:
Oven:
DB-5ms, 30m x 0.25mm, 0.25um
H2, 60 cm/sec, constant flow
split 1:20, 250CFID, 320C, N2 makeup gas
40C for 0.75 min, 40-325C at 20C/min,
325C for 30 min
Red: Liquid Soap
Blue: system blank
.
Ultra Inert Gold PlatedSeal
o Liner rests on the gold
seal
o The slit allows for the split vent
Cross slit for high split vents (>200ml/min)
o Gold surface important to seat on the
liner
Ultra Inert Gold Seals, Advantages
100 ppb STD
x103
500 ppb STD
x103
8
Ultra Inert Gold SealOmethoateAcephate
14Acephate Omethoate
00
x103x103
8
Siltek SS
1Acephate Omethoate4
OmethoateAcephate
00
x103 x103
8
Standard Gold Seal
1 Acephate Omethoate4
Acephate Omethoate
00
4 4.5 5 5.5 6 6.5 7 7.5 4 4.5 5 5.5 6 6.5 7 7.5
Counts vs. Acquisition Time (min) Counts vs. Acquisition Time (min)
Inert Pathway vs Standard PathwayDrugs
of Abuse
0.25ng on column
App Note 5991-1859EN
Inert Pathway:o Ultra Inert GC/MS column
o Ultra Inert Gold Seal
o Ultra Inert liner
o UltiMetal Plus inlet weldment
o UltiMetal Plus Flex. Ferrules
Bonus Peaks or Ghost Peaks
Contamination in the INJECTOR
••
Carry-over from a backflash or previous sampleSeptum bleed
COLUMN or… Carrier Gas
• Bad tank of gas or traps have expired
NoisyBaseline
FLOW- Contaminated gas- Incorrect detector settings
DETECTOR- Air leak - ECD, TCD- Electronics malfunction
COLUMN- Bleeding if at high temperatures- In detector flame (poor installation)
Safeguarding the Inert GC Flow Path:Gas Purification with Gas Clean CarrierGas
Optimizes your GC/MS performance
• Single combination Filter, optimized to remove
oxygen, moisture and hydrocarbons from
GC/MS
carrier gas
• Increased sensitivity with more accurate data
• Prevents column damage, lower replacement frequency
• Reduces instrument down-time
filter
And now, after lookinghuman
at chromatographic health and safety, here aresome practical health and safety issues
51 9/5/2018 Agilent Confidential
Universal/External
SplitVent Trap
Features:
• Trap analytes in
contamination
sample effluent to prevent
• When testing potentially hazardous samples,
the split vent flow should be vented to an
exhaust hood
• Easy to change cartridge design
• Economical insurance against costly
damage
Ultra Scientific
A Guideline for Standard
> Buffer and Reagent
> Pesticide, PAH, and Lipid Standard
> Inorganic Standard
TitleConfidentiality labelRegulatory statement (if applicable)52 January 25, 2019
Pesticide, PAH, and Lipid Standards
Standar agilent, tersedia dalam Ampul, Vial, danBotol dapat digunakan untuk penggunaan langsungatau untuk penyimpanan jangka panjang. Untukhasil terbaik, standar analitik harus digunakansegera setelah dibuka, dan tidak perlu ditransferuntuk digunakan nanti. Standar Harus disimpandalam botol CERTAN adalah paket inovatif palingaman.Setiap komponen dalam standar referensi sudahdianalisis sebelumnya, dan pelarutnya memilikikualitas terbaik.Standar dikirimkan dengan sertifikat yangmenunjukkan kemurnian senyawa yang tersediauntuk setiap standar. Standar solusi disiapkansecara gravimetri dengan presisi ± 0,5%, dandikirimkan dengan sertifikat yang menunjukkankonsentrasi pestisida dalam standar.
Polynuclear Aromatic Hydrocarbon (PAH) Standards PAH adalah polutan persisten yang tersebar luas. Beberapamemiliki ikatan yang beracun atau karsinogenik. Semuabahan diproduksi di bawah sistem mutu Agilent ISO 9001.Sertifikat yang menunjukkan kemurnian senyawa tersedia untuksetiap standar.
Thank
You
Any Questions
LC & GC Consumable or CrossLabChemical Ultra Standard
[email protected] 1973 4416 Call or WA
86 January 25, 2019 Title Confidentiality label Regulatory statement (if applicable)