48
Labwork in photonics. Photometry and sources. 1 Measuring luminance 1 2 Performance of lighting sources 9 3 Photometric characteristics of two objectives 19 4 Science of color 27 lense.institutoptique.fr | Deuxième année | Photonique S7 Engineer - 2 nd year - S1 -Palaiseau Version : September 13, 2018 Year 2018-2019

Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

Labwork in photonics.Photometry and sources.

1 Measuring luminance 1

2 Performance of lighting sources 9

3 Photometric characteristics of two objectives 19

4 Science of color 27

lense.institutoptique.fr | Deuxième année | Photonique S7

Engineer - 2nd year - S1 -PalaiseauVersion : September 13, 2018

Year 2018-2019

Page 2: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

2

Page 3: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

Labwork 1

Measuring luminance

Version: September 13, 2018

Questions P1 to P4 must be prepared in advance

The report must be filed on the website within one week. Its length mustbe 8 pages maximum.

Contents1 Calibration of a visible photodiode . . . . . . . . . . . 1

1.1 Calibration method . . . . . . . . . . . . . . . . . . 21.2 Analysis of the various sources of uncertainty . . . 3

2 Measuring the luminance with the calibrated pho-todiode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.1 Direct measurements with the single photodetector 52.2 Measurements with an optical system and the same

photodetector. Pupil of the system and etendue . . 53 Measurements . . . . . . . . . . . . . . . . . . . . . . . . 6

The goal of this lab session is to learn some commonly used radiometricmeasurement techniques. They are based on simple principles but must beconducted with care. In photometry, measurements rarely have an accu-racy better than 5 to 10 %. Care should be taken during the entire courseof the lab, analyzing as well as possible the sources of uncertainty andtheir influence on the measurements.

1 Calibration of a visible photodiodeThe UDT PIN 10AP photodiode is equipped with a green optical filter inorder to reproduce the spectral response of the human eye (see Figure 1).

1

Page 4: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

2 LABWORK 1. MEASURING LUMINANCE

Visual quantities can therefore be directly measured with this kind of pho-todiode. The area of the detector is S = 1.000 cm2.

Figure 1.1: Relative spectral response of the standard eye for photopic andscotopic vision

To calibrate the photodiode, you will be using a calibrated light sourcewhich gives a known visual intensity for a given continuous electric cur-rent. The nominal values for the current, voltage and visual intensity, aswell as their uncertainties, are written directly on the lamp.

P1 You will need to measure simultaneously the voltage and the currentapplied to the lamp. Give the circuit diagram which ensures the best pre-cision for measuring the current.

1.1 Calibration methodPreparation

To calibrate a detector you have to measure its sensitivity as accurately aspossible. The sensitivity (in µA/lm) is given by :

σ =i

Fv(1.1)

where Fv is the visual flux received by the photodiode and i the resultingphotocurrent.

Page 5: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

1. CALIBRATION OF A VISIBLE PHOTODIODE 3

P2 Recall the definitions and the units of the following quantities char-acterizing a source :

• intensity,

• luminance.

Each quantity is suited for a specific type of source : explain which kindand why. Recall the definitions and the units of these quantities: visualflux, etendue.

P3 Give the expression of the visual flux, Fv, received by a photodetectoras a function of the intensity Iv of a small source, the distance d betweenthe source and the detector and the surface S of the detector.

P4 How do you evaluate the uncertainty on the flux, knowing the inten-sity, the distance and their respective uncertainty?

Measurement

The calibrated light source is connected to a DC power supply which canbe set between 0−250 V. The voltage and current through the lamp shouldbe precisely monitored using a voltmeter and an ammeter. Never exceedthe nominal current and voltage of the lamp! The photodiode will beplaced on a moving mount and the calibrated light source at the end of theoptical bench. By moving the photodiode, you can adjust the illuminanceon the detector.

; M1 Measure the sensitivity σ of the photodiode far from the lightsource (d > 1 m). Work in complete darkness. Try to minimize unwantedreflections and diffusions (stray light) with black stops.

Check your result with a teacher.

1.2 Analysis of the various sources of uncertaintyQ1 Express the relative uncertainty ∆σ

σ as a function of the followingrelative uncertainties:

• ∆FF , the relative uncertainty on the flux received by the photodiode

• and ∆ii , the relative uncertainty on the photocurrent

In the following questions, you will identify the different sources of un-certainty which contribute to these two uncertainty terms.

Page 6: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

4 LABWORK 1. MEASURING LUMINANCE

Relative uncertainty on the flux received by the photodiode

Q2 Express the relative uncertainty ∆FF on the flux received by the pho-

todiode as a function of the following relative uncertainties:

• ∆dd , the relative uncertainty on the distance between the source and

the photodiode

• and ∆II , the relative uncertainty on the visual intensity of the lamp

; M2 Give an estimation of these sources of uncertainty.

Q3 For the visual intensity of the lamp, in addition to the uncertaintydue to its calibration, do you have to take into account the uncertaintyon the measurement of the electrical current flowing through the lamp?Explain how you can estimate this uncertainty.

Relative uncertainty on the photocurrent

Q4 Estimate the relative uncertainty ∆ii on the photocurrent due to the

relative precision of the picoammeter (given on the back side of the picoam-meter).

Relative uncertainty on σ

Q5 Deduce from the previous study an estimation of the uncertainty levelon your measured value of σ. Comment on the relative contribution of thevarious noise sources.

2 Measuring the luminance of an extendedsource with the calibrated photodiode

From the previous measurements, the detector is now calibrated and canmeasure a flux in visual units. In this section, you will measure the visualluminance (Cd/m2) of a very bright white LED source. The fact that thissource is almost lambertian will be assessed in the last part of this lab.

The luminance of a source describes its brightness. A source of highluminance is a very bright primary or secondary light source.

Typical examples of sources and their luminances :

Page 7: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

2. MEASURING THE LUMINANCE WITH THE CALIBRATED PHOTODIODE5

Luminance in Cd/m2

Energy saving lamp (Hg) 5 000Bright Led 2 107

Blue sky (or full moon) 3 000Sun 1, 6 109

Snow (under sun) 104

A light source may cause uncomfortable glare between 3 000 et 10 000 Cd/m2.Above 10000 Cd/m2, glare becomes neutralizing. Even a moderate uncom-fortable glare can impair vision. You will notice that this light source be-comes blinding for a current greater than 1 A.

2.1 Direct measurements with the single photodetec-tor

; M3 In order to control the area of this source, place a diaphragm di-rectly in front of the lamp. Open this diaphragm to its maximum (50 mmdiameter). Use the 0-40V power supply with a current between 1 and 2 A.Pay attention to the settings of the ammeter !. Note the settingsof the power supply (voltage, current), which you should keep the samethroughtout the rest of the labwork.

Q6 We wish to measure the luminance of the source with the single detec-tor calibrated in the previous paragraph. Place the detector at a distanceD = 1 m of the source. Calculate the etendue of the light flux measurementused in this configuration.

; M4 Measure the luminance of the source by placing the photodetec-tor on the bench at 1 m of the source, and then as far as possible from thesource. Comment on your measurements. Estimate the uncertainty onthese measurements by an analysis similar to that of the previous para-graph.

2.2 Measurements with an optical system and the samephotodetector. Pupil of the system and etendue

; M5 Place a lens (focal length f = 120 mm and aperture number N = 3)between the white source and detector, in the 2f−2f imaging configuration.

Q7 Describe the method you used to reach this imaging configuration.

Page 8: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

6 LABWORK 1. MEASURING LUMINANCE

; M6 Measure the light flux received by the detector in this configura-tion. Deduce from this measurement the luminance of the source and theilluminance received in the image plane. You will specify the etendue usedin this configuration (using a simple schematic diagram).

; M7 Change the imaging configuration while checking that the imageof the source remains much larger than the size of the detector. Measureagain the received flux, the luminance of the source and the illuminancereceived in the image plane.

Q8 Comment on these measurements and explain the results.

; M8 Move the lens such that the transverse magnification is far lowerthan 1. Explain your method of adjustment.

; M9 Check that in this case, the detector collects all the flux receivedin the image plane.

; M10 Measure the light flux received by the detector in this configura-tion. Once again, deduce the luminance of the source from this measure-ment.

Q9 Comment on these measurements. Specify the etendue used in thisconfiguration (using a simple schematic diagram). Explain why this mea-suring setup is not suitable for measuring the illumination received in theimage plane.

3 Measurements with a luminance-meter anda spectro-luminancemeter

In this section, your measurements will be compared to those given bytwo commercial instruments for measuring visual luminances of extendedsources (in Cd/m2):

• a luminance-meter LS-100 Minolta

• a spectro-luminancemeter SpectraScan PR 655 PhotoResearch.

Q10 From the manual of the luminance-meter, describe the working prin-ciple of this instrument. Is it the same as any of the methods you usedbefore?

Page 9: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

3. MEASUREMENTS 7

; M11 Measure the luminance of the source with the luminance-meterMinolta. Describe your procedure. Compare with the values obtained inthe previous paragraph.

Q11 Explain with a schematic diagram how the etendue is precisely de-fined (and constant!) in the luminance-meter. Compare to the methodsproposed in the previous paragraph .

In the manual of the spectro-luminancemeter, a simplified operatingprinciple is described (see Figure 1.2).PR-655 / PR-670 User Manual

22

FIGURE 21 – PR-655 / PR-670 SPECTROMETER

The uncorrected or ‘raw’ spectrum is corrected using factors established during the factory calibration of the instrument. These factors include wavelength accuracy correction, spectral distribution correction and photometric correction. The wavelength calibration is performed using a helium spectral line source. The line source provides known spectral emission lines that the software uses to map the dispersion of the grating to the multi-element detector. The helium lines used during the wavelength calibration are 388.6 nm, 447.1 nm, 471.3 nm, 587.6 nm, 667.8 nm, 706.5 nm and 728.13 nm.

Next, the data is corrected by spectral factors. These factors insure that the spectral power distribution (SPD) of the target is and calculated values such as CIE Chromaticity are accurately reported. Lastly, a scaling factor (photometric factor) is applied ensuring the proper photometric levels such as luminance or illuminance are displayed.

Calculations

The corrected spectral data is then used to calculate photometric and colorimetric values including

luminance, CIE 1931 x, y and 1976 u’, v’ chromaticity coordinates, correlated color temperature and

dominant. Following are some of the basic calculations used to generate photometric and colorimetric

parameters:

256 detector

array

Diffracted

light

Diffraction

grating

Objectiv

e lens

Aperture

mirror

Optical

radiation Eyepiece

Input Slit Grating

Incident

light

Figure 1.2: Principle of a PhotoResearch spectro-luminancemeter .

Q12 Describe the measurement made by the spectro-luminancemeter.Compare it to the measurement performed with the luminance-meter.

; M12 Perform a luminance measurement of the source with the spectro-luminancemeter. Compare it to the values obtained in the previous para-graph when using the luminance-meter.

; M13 Display the spectrum and color coordinates. Comment on themeasurements.

Questions E1 à E2 will be answered in the report, after the session.

Page 10: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

8 LABWORK 1. MEASURING LUMINANCE

E1 Synthesize and analyze all your measurement results. In particularcompare the luminance measurements.

E2 Search on the internet which applications use the instruments seenin this lab. Are there any other instruments to perform photometric mea-surements?

Page 11: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

Labwork 2

Performance of lightingsources

Version: September 13, 2018

Questions P1 to P3 must be prepared in advance

The report must be filed on the web site within one week. Its length must be8 pages maximum.

Contents1 Preparing questions . . . . . . . . . . . . . . . . . . . . 102 Flux and luminous efficacy measurement. Princi-

ple and calibration. . . . . . . . . . . . . . . . . . . . . 112.1 Integrating sphere . . . . . . . . . . . . . . . . . . . 112.2 Why is the integrating sphere the ideal instrument

to measure a total flux? . . . . . . . . . . . . . . . 122.3 Description of the detector . . . . . . . . . . . . . . 132.4 Characteristics of the reference source . . . . . . . 13

3 Integrating sphere calibration . . . . . . . . . . . . . 144 Spectrum measurement. Principle, precautions . . . 145 Characterization of some light sources . . . . . . . . 156 Results synthesis . . . . . . . . . . . . . . . . . . . . . . 16

In this session, you will characterize the photometric and colorimetricproperties of 4 light sources. Different measurements will be carried outfor each source:

• Photometric measurements with an integrating sphere. You willmeasure the total visual flux and the luminous efficacy.

9

Page 12: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

10 LABWORK 2. PERFORMANCE OF LIGHTING SOURCES

• Colorimetric measurements with a spectrometer. You will mea-sure the emission spectrum of the light source and the coordinates inthe chromaticity diagram.

The integrating sphere has to be calibrated before performing measure-ments on the different lamps. The lamps under study are commerciallamps (230 V) kept in their original packaging. You will report the rele-vant features provided by the manufacturer on the box of these lamps.

Important Note The interpretation of the measurements must be madeon the spot, during the lab, and if possible controlled by the lab in-structor. All the results will be given with an evaluation of the mea-surement uncertainty.

1 Preparing questionsP1 Write down the definition of the color of an object or the color of alight source.

P2 What is the definition of a Lumen?

P3 What is the maximum luminous efficacy (ratio of the luminous fluxin Lumens to the electrical power supplied) that can be achieved with amonochromatic source? At which wavelength? Can we get a higher valuewith a wide spectrum source?

P4 What is the value of the luminous efficacy of solar radiation?

P5 What is the power factor of an electrical appliance or installation?Why do electrical energy suppliers penalize manufacturers if their powerfactor is too low?

Page 13: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

2. FLUX AND LUMINOUS EFFICACY MEASUREMENT. PRINCIPLE AND CALIBRATION. 11

2 Flux and luminous efficacy measurement.Principle and calibration.

2.1 Integrating sphere

Detector

Lamps

Figure 2.1: Lab Integrating sphere. Note the presence of a small diffusingscreen occulting any direct flux from the source to the detector.

An integrating sphere is a photometric instrument which is designed tomeasure the total flux emitted by either a directional (Laser) or a non-directional (LED, lamp,...) light source. An integrating sphere is a spheri-cal cavity which internal surface is a very good light diffuser (high albedo,ρ, and nearly lambertian). The source under study is placed inside theintegrating sphere (Figure 2.1). Light emitted by the source is submittedto multiple diffusions inside the sphere, leading to an homogenous irra-diance that can be measured by a detector placed at the surface of thesphere. Note the small diffusing screen which aims at occulting any directflux coming from the source on the detector.

; M14 Carefully open the sphere.

; M15 Please, never touch the white paint, which is extremelyfragile!

; M16 Locate the reference lamp, the detector, the lamp holders and thefiber connecting the surface of the sphere to the USB spectrometer.

Page 14: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

12 LABWORK 2. PERFORMANCE OF LIGHTING SOURCES

2.2 Why is the integrating sphere the ideal instrumentto measure a total flux?

Assuming that the surface of the sphere is perfectly Lambertian, thereis a simple relationship between the flux emitted by the source and theilluminance inside the sphere :

E =ρ

(1 − ρ)· 1

4πR2· Flamp,

where E (as Éclairement in french) is the illuminance and R the radiusof the integrating sphere.

This formula is established under ideal conditions, where all the defectsof the sphere (apertures in the inner wall, source, screen...) are neglected.

This formula can be demonstrated as follows : we first recall that thealbedo of a surface is the ratio of the total flux reflected or scattered to thetotal flux received by the surface:

ρ =Fscattered

Freceived

The illuminance of the detector is calculated as follows. By not takinginto account the direct flow from the source to the sphere (which is the casefor the detector which is masked by a cache and receives only indirect flux)and neglecting leaks (holes in the sphere), the flux received at any point bythe surface of the integrating sphere is the total flux emitted by the lampdiffused a large number of times on the sphere. This indirect flux is givenby the expression:

FIndirect = Flamp(ρ+ ρ2 + ρ3 + · · ·

)=

ρ

(1 − ρ)Flamp

If the surface of the sphere is perfectly Lambertian, after the first re-flection (and for all following reflections), the flux received by the sphere isuniformly distributed.

The illuminance of the surface of the sphere due to the indirect lumi-nous flux is simply:

E =FIndirect

SSphere=

1

4πR2FIndirect

where SSphere is the surface of the integrating sphere.The radius of the integrating sphere is 50 cm. A calibrated source is

positioned at the center of the sphere, together with the socket in whichlamps under test will be plugged, see Figure 2.2.

Page 15: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

2. FLUX AND LUMINOUS EFFICACY MEASUREMENT. PRINCIPLE AND CALIBRATION. 13

Calibrated lamp

Do not unplug !!

Lamp sockets

Figure 2.2: Lamp sockets

2.3 Description of the detectorThe detector is placed on the surface of the sphere and the output current,proportional to the irradiance, is measured with a pico-ammeter througha BNC cable. The detector used is a silicon photodiode PIN-10AP with a”V (λ) filter”. Its sensitive surface is a disk of area S = 1.000 cm2. Inci-dent light on the detector creates a current across the device, which can bedirectly measured by a pico-ammeter.

The calibration of this kind of detector with a reference lamp has beenstudied in the lab 11. The detector used during this session has been re-cently calibrated. Its sensitivity is :

σ = (366 ± 13)µA/lm

Knowing the surface, flux and illuminance can be derived. These mea-surements in photometric units are expressed in:

• Lumen (lm), for flux,

• Lux (lx), for illuminance.

2.4 Characteristics of the reference sourceTo calibrate the integrating sphere, we use a reference source with a wellknown total flux for a given current:

REFERENCE LUMINOUS FLUX SOURCE

Current Flux(2, 569 ± 0.004) A (1392 ± 70) lm

The reference source is powered by a regulated power supply (see Fig-ure 2.3). Picoammeter accuracy is given by:

Page 16: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

14 LABWORK 2. PERFORMANCE OF LIGHTING SOURCES

±(0, 5% + 0.004 × 10−3A)

Supply of

the calibrated lamp

Picoammeter

Figure 2.3: Control panel.

3 Integrating sphere calibration

; M17 Press the Start button to turn the reference lamp on. Do notchange the value of the supply current.

; M18 Determine, with its uncertainty, the coefficient κ defined as theratio of the current delivered by the photodiode to the total flux of thesource in the integrating sphere :

κ =iPhotodiode

Flamp

Q1 Assuming that the photodetector measures the illuminance of thesurface of the sphere, deduce an approximate value of the albedo, ρ, ofthe integrating sphere.

4 Spectrum measurement. Principle, precau-tions

The operating diagram of the spectrometer is given on Figure 2.4.

Page 17: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

5. CHARACTERIZATION OF SOME LIGHT SOURCES 15

1. Optical fiber input

2. Entry slit

3. Spatial filter

4, 6. Mirrors

5. Diffraction grating

7. Lens

8. CCD detector

Figure 2.4: Spectrometer description

The digital signal on the computer corresponds to the current given byeach pixel of the CCD-array. A calibration—already done— with the refer-ence lamp allows to convert these data into a spectral irradiance collectedby the fiber (reported in µW/cm2/nm).

Procedure: (An explanatory leaflet is available in the room.)

• Place the fiber either at the output of the integrating sphere (forhigh intensity lamps) or directly in front of the source. Observethe raw signal.

• Set the integration time such that no saturation occurs, whilekeeping a good dynamic range.

• Load the calibration file to display the calibrated irradiance spec-trum in µW/cm2/nm.

• Also display the coordinates of the source in the XYZ chromatic-ity diagram. Choose the ”equal energy” illuminant E.

5 Characterization of some light sourcesWe wish to study different types of commercial lighting lamps supplieddirectly by the mains with 220/240 V E27 base (27 mm diameter diameterthreaded cap). Below is an example of a technical sheet:

Page 18: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

16 LABWORK 2. PERFORMANCE OF LIGHTING SOURCES

Figure 2.5: Example of a technical data sheet for a lighting lamp

Q2 Explain and comment on the characteristics given in the data sheet.

; M19 Measure the total emitted flux (in Lumen), the electrical powerconsumed (in W), the power factor, the spectrum, the color temperatureand the color rendering index of the following sources :

1. A halogen lamp,

2. an energy saving lamp,

3. two commercial LED lamps,

4. one unknown lamp.

Q3 Which source has the best or the lowest luminous efficacy? Explainwhy.

Q4 Which source has the best colour rendering index? Explain why.

Q5 What is the significance of the power factor? Why is it equal to 1 forthe halogen lamp?

Q6 Why should it be close to 1 for any appliance or lighting connected tothe mains?

Q7 Compare source color temperatures. Explain their meaning. Forwhich lamp(s) does this measure have a physical meaning?

6 Results synthesisAnswers to the questions E3 and E4 should be given in the report.

Page 19: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

6. RESULTS SYNTHESIS 17

E1 Synthesize and analyze your measurement results for each lamp.

E2 Compare the studied lamps from an economic and from a colorimetricpoint of view.

E3 Look for information to compare the environmental impact of theirproduction, use and recycling.

E4 Search on the internet if there are more efficient and powerful lamps...still acceptable from a colorimetric point of view.

Page 20: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

18 LABWORK 2. PERFORMANCE OF LIGHTING SOURCES

Page 21: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

Labwork 3

Photometriccharacteristics of twoobjectives

Version : September 13, 2018

Question P1 to P5 must be prepared before the lab !

The report must be filed on the website within one week. Its length mustbe 8 pages maximum.

Contents1 Evaluation of stray light . . . . . . . . . . . . . . . . . . 212 Transmission on the optical axis . . . . . . . . . . . . 233 Objectives aperture numbers . . . . . . . . . . . . . . . 244 Vignetting measurement . . . . . . . . . . . . . . . . . 245 Measurement of stray light directly on the image . . 25

The goal of this lab session is to measure the photometric properties, inthe visible part of the spectrum, of two photographic objectives of 50 mmfocal length. Indicate in your report the serial number of the objec-tives under test. You will perform the same measurements for the twoobjectives, in order to compare them; you will synthesize your results atthe end of your report.

An ideal objective should give an image which exactly reproduces thescene observed, in term of luminance (we do not consider here the effecton the image of the limited spatial resolution of the objective). We recall

19

Page 22: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

20 LAB 3 PHOTOMETRIC CHARACTERISTICS

that the illumination in the image plane of an optical imaging system, foran extended object, is given by:

E = τπLobjet sin(αimg)2 (3.1)

where τ is the transmission factor and sin(αimg) is the image numericalaperture.

Figure 3.1 shows a perfectly black object (zero luminance) on a brightbackground (uniform luminance). In the image plane of the objective, witha given magnification, the same spatial variation of illuminance shouldbe observed. The dashed line shows, with some exaggeration, what isobtained in reality: some stray light in the dark zone (1), a bright zoneattenuated by the transmission of the objective (2) as well as its numeri-cal aperture (3) and vignetting, which significantly reduces the flux on theedge of the field (4).

Scene

x

Luminance

Image

x'

Illuminance

(1) (2) (4)

(3)

Figure 3.1: Image given by a non perfect objective of a black object on abright background.

P1 What is the definition of the f-number of an objective ?

Page 23: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

1. EVALUATION OF STRAY LIGHT 21

P2 For a plane object at infinity, what is the relationship between thef-number and the numerical aperture of the image ?

P3 Find the proof of formula 13.1

P4 The largest f-number (smallest aperture) of the black objective is 22.Write down all the other f-numbers available for this objective.

P5 Describe the causes that can limit the transmission factor, τ of anoptical system.

Equipment

In the last part of this lab, the black objective should be mounted on a USBcamera Ueye, which relies on a CMOS sensor of the following dimensions :6.9 mm × 5.5 nm. The size of the pixels is 5.3 microns × 5.3 microns (1280× 1024 pixels).

Q1 Taking into account the focal length of the objective, calculate thecorresponding object field of view.

The detector is a silicon photodiode with a green filter (giving a spectralresponse similar to the human eye, see Figure 11.1 page 2). The detectorhas an area of S = 1.0 cm2 and is placed inside a tube with a hole at thefront. The axial position of the detector can be changed in order to adjustthe numerical aperture of the system. The measurement itself is done witha pico-ammeter.

The source itself is an integrating sphere with a uniform luminance. A250 Watt lamp is placed inside the sphere (maximum voltage is 10 V).A light trap, useful to measure the stray light, is located at the back of thesphere.

1 Evaluation of the amount of stray light onthe optical axis

Stray light stands for all the undesired light incident on the detector. Inthe case of an objective, it could be light diffused by the metallic frame, an

Page 24: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

22 LAB 3 PHOTOMETRIC CHARACTERISTICS

even number of reflections on the various lenses or stray light due to somedust particles on the optics.

Q2 What is the effect of stray light on the quality of the image given bythe objective ?

In order to measure the level of stray light, we are going to use the lighttrap inside the integrating sphere. Its luminance is roughly 10,000 timessmaller than the rest of the sphere and it can be considered as a perfectlyblack object. To have an independent measurement, we define TP as theratio of the illuminance in the image of the light trap given by the objectiveover the illuminance in the image of the integrating sphere.

Q3 Explain this definition.

; M20 With the help of a luminancemeter, measure the luminance ofthe integrating sphere. What are the operating parameters of the powersupply (current, voltage) ? Is the luminance of the lamp sensitive to theseparameters ?

; M21 How uniform is the luminance of the sphere ? Perform a fewmeasurements to evaluate this uniformity.

; M22 Using the luminancemeter, try to measure the luminance of thelight-trap.

Q4 Comment on this value. Why is the measurement of the light-trapluminance very inaccurate ?

In order to measure the level of stray light induced by the objective, youwill image the light-trap on a small hole and measure the stray light usingthe photodiode (with its green filter) placed right behind it.

; M23 Check that the objective is clean (no dust). Clean the objective ifnecessary (Ask the teacher how to clean it.)

; M24 Adjust the hole exactly in the image plane of the objective and onits optical axis. To do this:

• Find approximately the position of the image plane with a white pa-per screen,

Page 25: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

2. TRANSMISSION ON THE OPTICAL AXIS 23

• Using the low magnification microscope, superpose the hole and theimage of the light trap,

• Then place the photodiode right behind the hole.

• Fine adjust the position in X, Y and Z of the hole, looking for a mini-mum of the incident power on the photodiode.

Q5 Explain with a diagram the method used to find the absolute mini-mum value (as a function of X, Y and Z)

; M25 Measure TP : the ratio of the minimum illuminance (image ofthe light trap, on the optical axis) over the maximum illuminance (outsidethe image of the light trap), for both objectives (the black one and the greyone).

Q6 Why is this measurement very sensitive to the axial position of thehole?

2 Transmission on the optical axisQ7 What is the definition of the transmission factor τ of an objective?

To measure a transmission factor, it will be necessary to measure twoflux: before and after the objective. The etendue of these two measure-ments (with and without the objective) must be exactly the same.

Q8 Explain with a diagram why moving the detector away from the holeallows one to work at constant geometrical extent (with or without theobjective) ? How can you check that your experimental conditions are cor-rect?

You have three brass rings that you can put between the photodiode andthe hole.

Q9 Explain with a schematic the role of these brass rings ? In whichorder should they be inserted ,

; M26 Measure τ with the brass rings and conclude. Repeat this mea-surement and give an uncertainty level.

Page 26: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

24 LAB 3 PHOTOMETRIC CHARACTERISTICS

Q10 Conclude by giving the transmission factor of the objective, with itsuncertainty. Repeat the measurement with the other objective.

3 Objectives aperture numbersWe now want to check the relation between the illuminance and the aper-ture numbers reported on the objectives.

Q11 Explain why the experimental configuration of the former paragraphis not suitable (using a schematic diagram). What should be changed toquantify the effect of the aperture number ? Once again, use a schematicdiagram.

; M27 Carry out the measurement of the current iph as a function ofN - read on the objective aperture ring. Check that the detection hole isalways set very precisely at the center of the image plane. Be careful not toaim the light trap. Draw the graph i = f( 1

N2 ) and compare to the expectedvariation.

Q12 The deviation from the theoretical curve shows that some numbersof aperture do not match those on the ring of the lens diaphragm. What arethese f-numbers? Explain why, for such f-numbers, the real values deviatefrom the expected ones.

We now want to trace the illuminance as a function of the aperture num-bers.

One gives the diameter of the sensor hole (Φ1 = mm) and the sensitivityof the detector:

σ = (359 ± 8)µA/lm

Q13 Trace the curve E = f( 1N2 ) and superimpose the expected theoreti-

cal curve (use the luminance of the sphere measured in Q3). Comment.

4 Vignetting measurementFor a number of applications, it is important to determine the vignettingof objectives, which is the variation of illuminance across the image planeas a function of the field of view of the image.

Page 27: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

5. MEASUREMENT OF STRAY LIGHT DIRECTLY ON THE IMAGE 25

; M28 To characterize the vignetting, place the objective on the USBcamera. Then place the camera just behind the opening of the integratingsphere and realise an image of the interior of the sphere. (Check that thelight-trap is not in the field of view of the objective.)

; M29 Start the control software of the camera » GUIVignettage fromMatlab. Parameters of the camera for this application are automaticallyloaded (Rolling Shutter, fixed gain, no Gamma correction). » GUIVignettagecan trace the illuminance profiles along a diagonal from the image andcompare them when you change the size of the stop. An annex to the man-ual of » GUIVignettage is available in the room.

; M30 Save an image of the zone of the sphere with uniform luminancefor the available f-numbers.

; M31 For each increase of the f-number, your need to increase the in-tegration time of the camera.

Q14 By how much in theory ? Explain the reason for this modification.

Q15 Comment on the results obtained.

Q16 With the help of a simple schematic, explain why the vignetting ismore important when the aperture is fully open.

5 Measurement of stray light directly on theimage

; M32 Exit » GUIVignettage and launch the software» LumiereParasiteGUI.

; M33 Image the light trap with th camera and adjust the focus care-fully.

; M34 The GUI can measure the average value of the signal on a se-lected area of the light trap, then in the vicinity of the trap.

Q17 Compare the values of stray light obtained with the two methods.

Page 28: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

26 LAB 3 PHOTOMETRIC CHARACTERISTICS

Page 29: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

Labwork 4

Science of color

La terre est bleue comme une orange

Paul Eluard. L’amour la poésie.

The report must be deposited on the website within one week, and mustnot exceed 8 pages in length.

At the end of this labwork session you will be able to:

• Define and measure the colorimetric quantities of a light source,

• define the metamerisms (fundamental metamerism and illuminantmetamerism),

• measure the color rendering index (CRI) of a source,

• define and measure the colorimetric quantities of a colored surface,

• describe the phenomenon of chromatic adaptation of vision to a ref-erence white.

Contents1 Overview of the equipment at your disposal . . . . . 282 Quantify the "color" of a source ? . . . . . . . . . . . . 283 Getting started with the CRI Illuminance-meter . . 314 Additive color mixing . . . . . . . . . . . . . . . . . . . 345 Illuminant metamerism and CRI . . . . . . . . . . . . 406 How to define and measure the color of a surface ? . 417 Chromatic adaptation - Balance of whites . . . . . . 45

27

Page 30: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

28 LABWORK 4. SCIENCE OF COLOR

1 Overview of the equipment at your disposalTwo industrial measuring instruments :

• a CRI Illuminance-meter (luxmètre-chromamètre) to measure lightsources,

• a spectrophotometer (spectro-colorimètre) to measure the spec-tral reflectivities of various samples.

A lightbox and a computer screen Controlling their colorimetric param-eters is the concern of part 4. A Matlab graph interfacing windowallows to control the lightbox.

8 lighting sources based on different technologies Paragraph 5 dealswith the comparison of Color Rendering Indices of these sources.

Some colored paper samples They will be used in parts 5 and 6.

2 How to quantify the colored perception ofa light source?

The color of a source or an object is not a physical quantity, it correspondsto a human perception.

However, scientists have tried for centuries to quantify this quantity,and our industrial societies need clean definitions to manufacture productsand also to realize medical diagnostics using e.g. imaging techniques.

The colorimetric quantities, built during the last century, quantify nowa-days the "color" of a light source. They are briefly reminded here. For moredetails you can refer to your colorimetry lectures notes1.

The color perception is "trivariant"; this property is related to the pres-ence of 3 types of cones in the retina of the human eye. It is not possible toquantify a visual perception with only one quantity.

The XYZ tristimuli are three quantities in cd/m2, i.e. visual lumi-nances, defined from the spectral decomposition of the source under studyin the visible spectrum 2. Figure 4.1 recalls their definitions.

1Colorimetry - Bases and applications, 2nd year Lectures notes at IOGS-Palaiseau (2AP)by Hervé Sauer, and Colorimetry, 2nd year Lectures notes at IOGS-St Etienne (2AS) by Math-ieu Hébert

2Note that the solid angle of the CRI Illuminance-meter being calibrated the measuredtristimuli are actually illuminances, in lux.

Page 31: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

2. QUANTIFY THE "COLOR" OF A SOURCE ? 29

Figure 4.1: Definition of the XYZ tristimuli starting from the spectric en-ergetic luminance Lλ(λ) = ∂L

∂λ (λ) of a light source (Lλ(λ) ≥ 0). Courtesy : Hervé

Sauer, Colorimetry - Bases and applications, 2AP.

Note concerning the construction of these three quantities:

• The functions x(λ), y(λ), z(λ) are positive and null. They were definedin 1931, and constitute the CIE 1931 norm, they correspond to linearcombinations of the spectral sensitivities of the sensors of the retina;

• The normalization coefficient is K = 683 lm/W, therefore X, Y , Z arein classical visual units.

• y(λ) has been chosen equal to the sensitivity function V (λ) of the eye,Y is therefore the visual luminance of the light source (in cd/m2).

These quantities are sometimes normalized by the luminance of an idealwhite (but not realizable in practice); i.e. a source the spectrum of whichcontains all wavelengths in equal amount, called "E white". The valuesof the tristimuli then become relative to each other, their relative weightscharacterizing the visual color perception of the source.

For memory Two sources with identical XYZ characteristics are observedwith the same color by a human eye, and vice-versa. This fundamentalproperty is at the basis of the CIE1931 XYZ system.

Representation of the chromatic "coordinates". The color cube.This definition using three quantities can be represented by the position inspace of a point with coordinates (X,Y,Z). In the case of values normalizedto the visual luminance of the E white, Y is restrained to [0, 1] (but notnecessarily X or Y ); the color space is often represented as a cube of side 1

Page 32: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

30 LABWORK 4. SCIENCE OF COLOR

: the color cube. This is obviously an arbitrarily limited representation ofcolor space.

Also, it should be noted that not all the points inside the cube corre-spond to actual visual perceptions, only a fraction of the cube correspondsto physically possible stimuli; this sub-ensemble (mathematically, a conewith vertex (0,0,0)) is called the solid of actual colors and is (partially)shown (with a limited extent) on figure 4.2.

Figure 4.2: Representation in the XYZ space of (a limited part of) the solidof actual colors. Courtesy : Hervé Sauer, Colorimetry - Bases and applications, 2AP.

The chromatic coordinates xy in the Maxwell triangle are also oftenused. They are defined by :

x =X

X + Y + Zand y =

Y

X + Y + Z

They are built using the (conical) projection of the (XYZ) point onto the"diagonal" plane (X + Y + Z = 1) of the cube. They are shown in fig. 4.3.

Page 33: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

3. GETTING STARTED WITH THE CRI ILLUMINANCE-METER 31

0

1

1

1

Maxwell's triangle

Spectrum locus

and purple line

in Maxwell's triangle

Figure 4.3: Definition et representation of the chromatic coordinates xy inthe Maxwell triangle (left). The projection in the Z = 0 plane yields thechromaticity diagram (right). Courtesy : Hervé Sauer, Colorimetry - Bases and applications,

2AP.

The cross-section of the solid of actual colors by the X+Y +Z = 1 planecontains the points of coordinates (xy) that are accessibles by real sources.It is delimited by :

• the spectrum locus (a convex line), the points of which correspondto a particular wavelength of the visible spectrum (monochromaticcolors),

• the purple line (a straight segment), which connects the extremepoints of the spectrum locus.

Many other colorimetric spaces have been defined, starting from theXYZ coordinates (L∗, a∗, b∗ or L∗, u∗, v∗, . . . ). During this labwork session,only the XYZ tristimuli and the xy coordinates in the CIE1931 chromatic-ity diagram will be measured and analyzed.

3 Getting started with the CRI Illuminance-meter

The CRO Illuminance-meter at your disposal (see fig. 4.4) allows you toboth quantify and analyze the spectral components of a source in order todeduce, after some numerical calculations, various quantities such as thecolorimetric coordinates. This type of instrument works on the principle ofa grating spectrometer, completed by some numerical processing.

Page 34: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

32 LABWORK 4. SCIENCE OF COLOR

1

1. Parts Designations and Functions

1. Parts Designations and Functions

1-1  Parts Designations

a USB Connector

c Strap Eyelet

Rear View

Battery Compartment Section

Front View1 Light Receptor

2 Light Selection Ring

0 Tripod Socket

3 Power Button

4 Display Panel

5 Measuring Button

7 Memory Button

6 Menu Button

8 Battery Cover Latch

9 Battery Cover

b Sync Terminal

Bottom View

d Battery Compartment

Figure 4.4: CRI Illuminance-meter KONICA-MINOLTA CL-70F. Schematicfrom the datasheet given by the manufacturer.

; M1 Measure the spectrum of the room light. To do so :

• Switch on the instrument CL-70F KONICA-MINOLTA (Power ButtonÂ).

• Realize the calibration of the instrument by selecting CAL with therotating ring Á , fig. 4.4).

• Rotate the selection ring and select ambiant light (with the pictogramof a sun).

• Select the menu Spectrum on the touch screen.

• Orient the sensor (behind the white diffusive plate) towards the sourceand click on Ä : Measure.

; M2 Save the spectrum on the computer and display it using the Matlabsoftware. To do so :

• Connect the instrument to the computer using the USB cable,

• then launch the application CL70F Utilitaire by using the short-cut on the desktop.

• After the application has opened, select

– the menu Product Setting,– then the tab Tool Box;

Page 35: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

3. GETTING STARTED WITH THE CRI ILLUMINANCE-METER 33

– modify the line Memory Title using the following format2017-MM-DD NamesOrGroupNumber

– then ok to close the window.

• In the instrument, go back to the menu Spectrum and click on Mem-ory, Æ.

• In the application CL70F Utilitaire, choose the menu MemoryData to display the spectra and the colorimetric data on the com-puter screen.

• To access the data (and save them as images) you can click on Preciewand then on Text. Otherwise, you can use the Save command to saveboth images (if you have selected this option) and a (.csv) data file.Please save your files in the MesuresColorimétrie folder on thedesktop, in a RoomLight/ sub-folder for instance.

• Finally, launch Matlab and the »AfficheSpectreCL70F commandto retrieve all the spectral data and to display them.

Q1 What the measuring range of the instrument in terms of wavelength?

Q2 Comment on the global shape of the spectrum. Which technology isthe room light based on ? What does the correlated color temperature ofthis source correspond to ?

The (correlated) color temperature does not entirely characterizethe color of a light source. It is defined as the temperature of the blackbody with closest chromatic coordinates (see fig. 4.5). It is used to qualifylighting sources with different "whites".

Page 36: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

34 LABWORK 4. SCIENCE OF COLOR

Figure 4.5: Locus of CIE1931 chromatic coordinates of a black body versusits temperature (Planckian locus). Courtesy : Mathieu Hébert, "Colorimetry", 2AS.

4 Additive color mixing

4.1 Spectra and colors. Fundamental metamerismMeasure and perception. Comparison of two sources. In this partyou will use the light box equipped with a diffusive opaline (fig. 4.6).

Figure 4.6: Light box with two compartments used in this labwork (left)and control window (right).

This box contains two compartments :

• in the left compartment there are two reference sources (REF1 etREF2); a color filter can be placed in front of them;

• in the right compartment there is a ribbon of trichromic LEDs.

Page 37: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

4. ADDITIVE COLOR MIXING 35

The sources are controlled (by pulse width modulation, PWM) by a mi-crocontroller interfaced via a USB cable to the computer and the Matlabprogram.

; M3 Launch the interface program ManipCouleur. Test the ribbonof LEDs and its connection to the computer by successively driving thevarious sources. The control values must be integers between 0 and 255.

; M4 Measure and analyze the chromatic characteristics of the two ref-erence sources using the CRI Illuminance-meter, for a control value of255.

Q3 One of the reference sources is said to be "colder" that the. Which oneis it ? What is its correlated color temperature (CCT) ? Note the CCT ofthe other source. How does it compare to the CCT of room light ?

; M5 Adjust the control of the ribbon of trichromic LEDs to obtain thesame color (visually) as for REF1. Explain you proceed to equalize thecolors of the two compaetments.

; M6 Measure the chromatic coordinates (x and y) of the two sources.Are they close to each other ?

; M7 Use the CRI Illuminance-meter to adjust more precisely the chro-matic coordinates. Save on the computer the measurements of the trichromicsource that best reproduces the color of the reference source. Check thatyou have modified and labeled correctly your filenames.

Q4 Compare the two spectra.

Fundamental metamerism is the phenomenon demonstrated above.Additive color mixing uses this phenomenon. Indeed, one does not need toperfectly match two spectra to obtain identical colors. Matching the threecolorimetric coordinates XYZ is sufficient.

4.2 Synthesis of colors using the trichromic LEDCharacteristics of each of the three components

; M8 Measure the spectrum and the chromatic characteristics of

Page 38: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

36 LABWORK 4. SCIENCE OF COLOR

• the red component of the trichromic LED at its maximum value (2550 0),

• the green component at its maximum value (0 255 0),

• the blue component at its maximum value (0 0 255),

• the blue component at half maximum (0 0 128),

• of the blue and green components simultaneously at their maximumvalues (0 255 255).

Save these measurements on the computer. Check the names of the vari-ous files.

Q5 Analyze and compare the two measurements performed on the bluecomponent (at its maximum value and at half maximum). Compare inparticular the values of

• the total visual luminance,

• the chromatic coordinates x and y.

and also the overall shape of the spectrum.

; M9 Place the three points corresponding to the three components attheir maximum value on the chromaticity diagram using the program DiagChromaand save your graphic.

; M10 Place the point corresponding to the mixture of blue and greenon the chromaticity diagram.

Q6 Analyze your results. Is the corresponding point aligned with the twopoints of each constitutive component ? Is it in the middle of the associatedsegment ?

Color gamut and triangle or reachable colors

To correctly perform an additive mixture one must work with the 3 chro-matic coordinates. However, not all points (XYZ) inside the solid of actualcolors can be reached. The reachable points define a solid, the so-calledcolor gamut, shown on fig. 4.7. The intercept of the gamut by Maxwell’striangle is a triangle, which projects onto the (x, y) plane and yields theso-called "triangle of reachable colors". In addition, the discreteness of thecontrol values of the cources lead in practice to a finite number of reachablecolors.

Page 39: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

4. ADDITIVE COLOR MIXING 37

Figure 4.7: Representations of the solid of reachable colors (color gamut)inside the color cube (left and center). The solid of reachable colors is ac-tually discretized (left) due to the finite number of control values of thesources. Triangle of reachable colors in the (x, y) plane (right). An interac-tive 3D figure is also available (by typing »SolideXYZ_synth_sRGB ) tohelp you visualize the gamut. Courtesy : Hervé Sauer, Colorimetry - Bases and applications,

2AP; and Mathieu Hébert, Colorimetry, 2AS.

Method of additive color mixing

One first needs to build the matrix of color mixing, starting from the threeXYZ vectors of the red, green and blue components of the trichromic LEDsat their maximal power :

MLEDtri =

XR XG XB

YR YG YBZR ZG ZB

The chromatic coordinates of the (white) light produced by mixing the

three components at their maximal power will thus be :XwhiteYwhiteZwhite

= MLEDtri ×

111

And if you wish to obtain a light with chromatic characteristic

XtargetYtargetZtarget

,

the control values that you need to send to the LEDs should be proportionalto :

M−1LEDtri ×

XtargetYtargetZtarget

Page 40: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

38 LABWORK 4. SCIENCE OF COLOR

; M11 Following this method, use Matlab and the previous measure-ments to calculate the control values that you need to send to the LEDsto obtain exactly the same color as the reference source that you chose inquestion M5.

Matlab syntax, reminder :

• Create a 3 × 3 matrix : M = [ a b c ; d e f; g h i];

• Invert a matrix and multiply it by a vector : inv(M)*V

Q7 Check that the control values are consistent with the ones that youobtained by iterative adjustments in question M5.

; M12 Place an orange filter in front of the reference source REF1 at itsmaximal power. Calculate the control values of the trichromic LED thatlead to the same color. Verify your result experimentally.

; M13 Place the red filter in front of the reference source REF1. Adjustthe control value of REF1 to 10 and perform the same study.

; M14 Place the (xy) coordinates of the two colored sources on the chro-maticity diagram using the DiagChroma routine.

4.3 Reachable colors using the LCD screen; M15 Observe each pixel of the LCD computer screen using the mag-nifying glass at your disposal. (You may take a picture with your smart-phone).

Q8 Describe each pixel of the screen.

; M16 Measure the chromatic characteristics of each RGB componentof the screen. Each pixel is controlled by a value between 0 and 255.Chaque pixel est commandé par une valeur comprise entre 0 et 255. The»Fenetres command opens 3 windows corresponding to the control val-ues (255 0 0), (0 255 0) and (0 0 255). Expand each window maximally toperform a correct measurement.

Q9 By the way, how many colors can be synthesized using this LCDscreen ?

Page 41: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

4. ADDITIVE COLOR MIXING 39

; M17 Plot the triangle of reachable colors of the screen.

Q10 Is the color obtained by filtering the REF1 source with an orangefilter reachable by the screen ?

; M18 Apply the same method as described in paragraph 4.2 to deter-mine the RGB control values of the screen and obtain the same colorimetriccoordinates as those obtained with the red filter (question M13).

; M19 Vérifier le résultat. Utiliser la commande » Fenetre(R,G,B)pour générer une image unie, (Fenetre(255,255,255) affiche un rect-angle blanc).

4.4 Comparison of the two technologiesThe luminance of a pixel does not vary linearly with the control value,which is between 0 and 255. This is the "gamma correction" (see fig. 4.8. γvaries between 1 et 2.4 and can be adjusted for some screens.

0 50 100 150 200 2500

20

40

60

80

100

Numerical value of the software command

Act

ualc

omm

and

ofa

pixe

l(in

%)

Figure 4.8: Shape of the correction law, xγ . Here, γ = 2, 2.

Q11 By the way, why is the correction law linear for the trichromic LEDs?

Page 42: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

40 LABWORK 4. SCIENCE OF COLOR

; M20 Compensate the "gamma correction" by elevating the commandvalues to the power 1

γ before sending the command to the screen and verifythe consistency of your result. The γ power factor of the screen is equal to2,2.

5 Illuminant metamerism and CRI of a light-ing source (subtractive color mixing)

5.1 Illuminant metamerism

; M21 Illuminate successively the color samples on the wall with themercury vapor lamp and then with the desk light.

Q12 What do you observe ?

5.2 IRC

The color rendering index of a source indicates how the source com-pares to a reference source (daylight or incandescent light) as far as visualrendering of colored samples is concerned. The CRI is defined by compar-ison to a reference source of the colorimetric results obtained from 15 cal-ibrated samples. It can be directly calculated from the spectric luminanceof the source.

; M22 Measure the spectrum and the CRI of the mercury vapor lampand of the desk light. The CRI is accessible in the menu CRI of the Illuminance-meter CL-70F and is denoted as Ra.

; M23 Measure the spectra and the CRI of ’white’ lights based on vari-ous technologies : LED, halogen, economical.

Q13 Analyze the results obtained for these various lights, in particularthe link between the shape of the emission spectrum and the measuredCRI.

Page 43: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

6. HOW TO DEFINE AND MEASURE THE COLOR OF A SURFACE ?41

6 How to define and measure the color of asurface ?

The colorimetric characteristics of a surface can be defined or measuredonly for a given illuminant 3.

6.1 DefinitionsThe colorimetric characteristics are most commonly defined for :

• an illuminant D65 i.e. daylight,

• and/or an illuminant A i.e. an incandescent light.

The CIE colorimetric quantities can be calculated from the spectral re-flectivity ρ(λ) of the surface under test.

Let E(λ) denote the spectric illuminance (in (W/m2)/nm) received bythe surface. In the case of a lambertian surface, the spectric luminance ofthe surface under test considered as a secondary source, is given by :

L(λ) = ρ(λ)E(λ)

π(4.1)

The following definitions are easily obtained :

X = G

∫visible

ρ(λ)E(λ)x(λ)dλ,

Y = G

∫visible

ρ(λ)E(λ)y(λ)dλ,

Z = G

∫visible

ρ(λ)E(λ)y(λ)dλ,

where G is a normalization coefficient to get free of the source power :

G =1∫

visible E(λ)y(λ)dλ

Q14 How does one derive equation (4.1) ?

3An "illuminant" is a "normalized" source est (i.e. defined by a norm) that allows a col-orimtric characterization of surfaces under precise lighting conditions. The CIE has definedtens of illuminants, typical of e.g., daylight at certain hours [e.g. D65], incandescence lamps[A], various fluorescent bulbs [e.g. F11], etc...

Page 44: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

42 LABWORK 4. SCIENCE OF COLOR

6.2 Measurement technique

The measurement of the spectric reflectivity is performed in a normalizedconfiguration, by using for instance a diffuse illumination of the surface(realized in practice with an integrating sphere) and by measuring at anangle of 8˚from the normal to the surface, with or without the specularreflection. The measuring device contains :

• a white source, associated to an integrating sphere in the case of dif-fuse lighting,

• a collection system of the flux scattered by the surface under test,

• a two-way spectrometer : one for the measurement, and one for theillumination of the surface.

The schematic of the spectrophotometer KONICA MINOLTA CM2600d isshown in fig. 4.9.

Page 45: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

6. HOW TO DEFINE AND MEASURE THE COLOR OF A SURFACE ?43

Figure 4.9: Schematic illustrating the working principle of the spectropho-tometer. Courtesy : KONICA-MINOLTA datasheet.

The CM2600d illuminates the sample under test with diffuse light usingan integrating sphere, and measures the backscattered light at an angleof 8˚from the normal to the surface of the sample ; the specular reflectionon the (not necessarily Lambertian) sample can be included (or not), cor-responding to the CIE normalized measure configurations:

• di:8˚ (SPI: specular reflection included)

• de:8˚ (SPE: specular reflection excluded).

6.3 Getting started with the spectrophotometer; M24 Switch on the spectrophotometer and launch the SpectraMagicsoftware. Then :

Page 46: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

44 LABWORK 4. SCIENCE OF COLOR

• connect the computer to the device using the Instrument/connexionmenu;

• calibrate the device by following the software instructions. Thereare two steps : to calibrate the dark signal, a light-trap is availableinside the leather pouch. For the calibration of the white, use thecalibrated "reference white" on the support of the spectrophotometer;DO NOT put your fingers on the "reference white" after removing theprotection cap; PROTECT AGAIN the "reference white" with the capafter calibration.

• place the instrument on the surface under test,

• perform the measurement (Echantillon menu)

The various menus are shown in fig. 4.10.

ConnexionCalibrage

Echantillon

Figure 4.10: Window of the spectrophotometer driver SpectraMagic.

6.4 Measurement of colored samples

; M25 Measure the spectral reflectivity of some colored paper sampleson the table. Verify that your measurements are consistent with the aspectof the samples.

Page 47: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

7. CHROMATIC ADAPTATION - BALANCE OF WHITES 45

Q15 Are the colorimetric coordinates different under illumination by il-luminant D65 or illuminant A ? Does it make a difference visually ?

; M26 Perform the measurements for shining surfaces.

Q16 Analyze your results, in particular the differences (if any) betweenthe measurements including the specular reflection (SCI) and those thatexclude it (SCE).

7 Chromatic adaptation - Balance of whites

; M27 Adjust the right side of the light box to obtain a source withchromatic coordinates x = 0, 4 et y = 0, 4. And switch on the source REF1on the left side.

Q17 Which color would you say the right side of the light box is ? Ratherpink, green, blue ?

; M28 Switch off the box and switch on the desk light, so as to illuminatethe left side of the light box. Put a screen between the two compartmentsof the light box so that the desk light does not illuminate directly the rightside but you can still see the two sides simultaneously.

Q18 Has the apparent color of the right side been modified ?

The chromatic adaptation is the phenomenon demonstrated in thisexperiment. Our brain interprets colors differently depending on the aver-age white of the surrounding scene. Cameras do the same during a shot;this phenomenon is called "balance of whites".

Q19 Verify that your observations are consistent with the two graphs infig. 4.11, which gives indications on how a scene is perceived depending onthe white reference that illuminates it (Munsell zones).

Page 48: Labwork in photonics.lense.institutoptique.fr › ressources › DeuxiemeAnnee › ... · P1 You will need to measure simultaneously the voltage and the current applied to the lamp

46 LABWORK 4. SCIENCE OF COLOR

Chromatic adaptation at the point

of coordinates xy=(0.313, 0.329)

(D65 illuminant)

Chromatic adaptation at the point

of coordinates xy=(0.448, 0.407)

(A illuminant)

Indicative limits of color areas

in the xy CIE 1931 chromaticity diagram

under day light chromatic adaptation

Indicative limits of color areas

in the xy CIE 1931 chromaticity diagram

under incandescent light bulb illumination chromatic adaptation

ROUGE = RED

ROUGE ORANGÉ = ORANGE RED

ORANGE = ORANGE

JAUNE ORANGÉ = ORANGE YELLOW

JAUNE = YELLOW

JAUNE VERDÂTRE = GREENISH YELLOW

JAUNE VERT = GREEN YELLOW

VERT JAUNÂTRE = YELLOWISH GREEN

VERT = GREEN

VERT BLEUTÉ = BLUISH GREEN

BLEU VERT = GREEN BLUE

BLEU VERDÂTRE = GREENISH BLUE

BLEU = BLUE

BLEU VIOLET = VIOLET BLUE

VIOLET = VIOLET

POURPRE VIOLET = VIOLET PURPLE

POURPRE= PURPLE

POURPRE ROUGEÂTRE = REDISH PURPLE

POURPRE ROUGE = RED PURPLE

ROUGE POURPRÉ = PURPLISH RED

ROSE = PINK

Wavelengths of some usual spectral lines:cadmium blue line (F’): 480.0nm

'hydrogen blue line (F): 486.1nm

cadmium turquoise line: 508.6nm

mercury green line: 546.1nm

mercury yellow doublet: 577.0nm & 579.1nm

sodium yellow doublet (D): 589.0nm & 589.6nm

He-Ne laser(red):

cadmium red line (C’):

hydrogen red line (C):

632.8nm

643.8nm

656.3nm

Figure 4.11: Courtesy : Hervé Sauer, from the data of the Munsell atlas.