Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence and Optical Sensors
Lecture 10
1
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Summary
• Fluorescence techniques
• Optical sensor Technology
• Optical detectors
• Case study - Label free DNA detection
2
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence• Fluorescent materials
absorb photons which excite electrons
• When they decay to a ground state they can re-emit photons
• These will have a lower energy
Excited State
Ground State
hν
Blue LightHigh Energy
hν
Green LightLower Energy
3
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence• The shift in λ is called
the “Stokes shift”
• The delay between excitation and emission is lifetime
• Quantum yield is ratio of photons out against number of photons in
Excited State
Ground State
hν
Blue LightHigh Energy
hν
Green LightLower Energy
4
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence• Fluorescence is at the
heart of biological measurement
• Fluorescent spectroscopy of endothelial cells
• Autofluorescence or labelling?
5
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence in Biology
• Some cells express fluorescent proteins like GFP or have other natural fluorescence
• Otherwise they need to be tagged or labelled with a fluorophore
• This is a fluorescent dye, or maybe a quantum dot that is selectively attached
6
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence in Biology• Not just used at a cellular level, most
important use is in microarrays
• Arrays of spots of probes, specific lengths of DNA, proteins or antibodies
• Target molecules from the tested sample are labelled with fluorophores
• They selectively attach to the probes
7
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence in Biology• Not just used at a cellular level, most
important use is in microarrays
• Arrays of spots of probes, specific lengths of DNA, proteins or antibodies
• Target molecules from the tested sample are labelled with fluorophores
• They selectively attach to the probes
7
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence Activated Cell Sorting (FACS)
• Type of flow cytometry
• Cells labelled with specific fluorophores
• Cells in suspension are sprayed in droplets
• Sorted depending on fluorescent label
Laser
Type 1
Flow
ChargedDeflectionPlates
+/−DropletCharging
Computer Control
Camera/Detector
Labelled cells
Waste Type 2
Sheath
8
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Fluorescence Microscopy Techniques
• Spectroscopy - measure different emission wavelengths
• FLIM - lifetime imaging, excite fluorophore with a pulse and measure decay
• TIRF - Total internal reflection fluorescence microscopy, uses near-field effects
• FRET - Förster resonance energy transfer
9
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Reducing Fluorescence• Photobleaching is a process through which a
fluorophore is photochemically altered
• Constant exposure to excitation wavelength eventually reduces fluorescence
• Quenching is reduction of fluorescence through chemical or other stimuli
• Exploited in a number of sensors
10
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Summary
• Fluorescence techniques
• Optical sensor Technology
• Optical detectors
• Case study - Label free DNA detection
11
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Optical Fibre Sensors• Enables use of well developed optical techniques
for bio/chemical sensing
• Optical fibres are low attenuation so can allow remote measurement (not wireless)
• Do not require electrical connection which may be essential in some environments
• Low interference, high sensitivity/bandwidth
12
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Optode and Fibre Sensors• The optode (or sometimes optrode) is a chemical
transducer that is read optically
• They often use an optical fibre or another type of wave guide
• Two types:
‣ Extrinsic - the fibre guides light to/from sensor
‣ Intrinsic - the fibre itself is part of the sensor
13
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Extrinsic Optical Fibre Sensors
• Optical fibres guide light to and from the sensor, which could be chemical or physical
• Detection method could depend on reflectance, absorbance, fluorescence etc.
Sensor
Light Source
LightDetector
Optical Fibres
14
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Oxygen Optode Fibre Sensor
• Ruthenium based fluorophores are quenched by oxygen
• Immobilise the fluorophore in a matrix at the tip of an optical fibre
• Coat with oxygen permeable membrane (PTFE/Teflon)
• Multimode fibres to carry both excitation and emission wavelengths
Fibre
15
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Oxygen Sensor• Results from sensor
using phase modulation of light
• Designed for in-vivo pO2 measurement
• Intensity and phase dependent on O2 concentration
P.A.S. Jorge et al. / Sensors and Actuators B 103 (2004) 290–299 297
difference is expected to be maximum around 188 kHz,which is the ideal modulation frequency [5]. This way, an in-crease in modulation frequency is expected to improve sen-sor performance. In Fig. 5(a) and (b) the phase and the fluo-rescence intensity response of the sensing system to O2/N2saturation cycles can be observed. The phase signal showssome instability indicating the need for SNR improvement.In order to demonstrate the phase insensitivity to opticalpower drift a simple test was performed. With the sensinghead in a 21% O2 atmosphere, the optical power injectedinto the fibre system was changed up to 25%. Fig. 6 showsthe consequence of this variation in the intensity and phaseof the fluorescence signal. Although a significant change in
0 60 120 180 240 300 360 420 480-20
-18
-16
-14
-12
-10
-8(a)
80%
100%
39,1%
60%
20,6%
12,3%8%
0,4%
φ d (d
egre
es)
time (s)
0 60 120 180 240 300 360 420 4800,03
0,04
0,05
0,06
0,07
0,08
0,09
0,10
0,11(b)
100%80%
60%
39,1%
20,6%
12,3%8%
0,4%
Fluo
resc
ence
inte
nsity
(mV)
time (s)
Fig. 8. System response to step variations of O2 concentration level: (a) phase response; (b) fluorescence intensity response.
the intensity response occurs, the phase response remainsessentially unchanged. This confirms the ability of thephase detection scheme to avoid power fluctuations inducederrors.The Stern–Volmer plots obtained from phase measure-
ments with all configurations are clearly non-linear, indicat-ing that the dopant is not homogeneously distributed withinthe silica matrix. Instead, the ruthenium complex occupiesenvironments with different oxygen accessibilities. In thissituation, the standard Stern–Volmer equation (Eq. (1)),based on a single exponential decay, no longer describesaccurately the quenching behavior. Alternatively a dual ex-ponential model, corresponding to two different dominant
16
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Oxygen Sensor• Results from sensor
using phase modulation of light
• Designed for in-vivo pO2 measurement
• Intensity and phase dependent on O2 concentration
P.A.S. Jorge et al. / Sensors and Actuators B 103 (2004) 290–299 297
difference is expected to be maximum around 188 kHz,which is the ideal modulation frequency [5]. This way, an in-crease in modulation frequency is expected to improve sen-sor performance. In Fig. 5(a) and (b) the phase and the fluo-rescence intensity response of the sensing system to O2/N2saturation cycles can be observed. The phase signal showssome instability indicating the need for SNR improvement.In order to demonstrate the phase insensitivity to opticalpower drift a simple test was performed. With the sensinghead in a 21% O2 atmosphere, the optical power injectedinto the fibre system was changed up to 25%. Fig. 6 showsthe consequence of this variation in the intensity and phaseof the fluorescence signal. Although a significant change in
0 60 120 180 240 300 360 420 480-20
-18
-16
-14
-12
-10
-8(a)
80%
100%
39,1%
60%
20,6%
12,3%8%
0,4%
φ d (d
egre
es)
time (s)
0 60 120 180 240 300 360 420 4800,03
0,04
0,05
0,06
0,07
0,08
0,09
0,10
0,11(b)
100%80%
60%
39,1%
20,6%
12,3%8%
0,4%
Fluo
resc
ence
inte
nsity
(mV)
time (s)
Fig. 8. System response to step variations of O2 concentration level: (a) phase response; (b) fluorescence intensity response.
the intensity response occurs, the phase response remainsessentially unchanged. This confirms the ability of thephase detection scheme to avoid power fluctuations inducederrors.The Stern–Volmer plots obtained from phase measure-
ments with all configurations are clearly non-linear, indicat-ing that the dopant is not homogeneously distributed withinthe silica matrix. Instead, the ruthenium complex occupiesenvironments with different oxygen accessibilities. In thissituation, the standard Stern–Volmer equation (Eq. (1)),based on a single exponential decay, no longer describesaccurately the quenching behavior. Alternatively a dual ex-ponential model, corresponding to two different dominant
16
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Intrinsic Optical Sensors
• Optical fibres work through total internal reflection at the core-cladding interface
• Evanescent wave extends beyond this by some small distance, typically less than λ
• This can excite fluorescence in a sensing material coating the fibre
17
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Intrinsic Optical Sensors• Sensing can be distributed along the fibre
• Penetration depth is low so fouling can be a serious problem depending on environment
Sensing Material
Fibre
Sensing zone
18
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Surface Plasmon Resonance
• Evanescent wave can excite plasmons
• These are coherent electron oscillations in a thin metal layer (Au)
• Reflected beam contains information about attachments
19
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Integrated SPR Biosensor 20
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Summary
• Fluorescence techniques
• Optical sensor Technology
• Optical detectors
• Case study - Label free DNA detection
21
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Light Dependent Resistor
• High resistance semiconductor with metal contacts
• Incident light creates electron hole pairs
• This reduces the resistivity of the material
22
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Photomultiplier Tube
• Photons generate electrons at photocathode which are accelerated towards dynode
• Impact excites more electrodes so there is amplification at each dynode
Glass Vacuum Tube
Photon
Electron
Dynodes (electron multipliers)
Anode
Photocathode
Current
Output
23
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Photodiode• p-n or p-i-n junction diode
operated in reverse bias
• Photons absorbed in the depletion region create electron-hole pairs
• These are swept apart by bias voltage leading to a light dependent current
p−type
intrinsic semiconductor
n−type semiconductor
Anode
Cathode
Lig
ht
24
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Charge Coupled Device• Originally developed as a
shift register memory
• CCD is the basis for many digital cameras
• Charge induced by light exposure is stored in potential wells
• Read out by moving charge along a row
25
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
CMOS Camera• Alternative to CCD cameras
• Common in mobile phones
• Arrays of photodiodes with CMOS addressing
• STMicroelectronics imaging division based in Edinburgh
• Originally spun-out from UoE
26
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Single Photon Avalanche Diodes (SPADS)
• Solid-state replacement for PMT
- 50 ps jitter
- 30 ns dead time
- 6 Hz dark count rate
- 20 μm pitch
- 20 V breakdown
- 0.35 μm CMOS
27
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
SPAD - Geiger Mode
IntegrationAv
alan
che Geiger
Reverse Bias Voltage (V)
Curre
nt (A
)
(1) AvalancheBreakdown(2) Quenching
(3) Reset
SPADhν
VD
RL
28
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
SPAD Outputs
Low Intensity Medium Intensity
High Intensity
0V
3.3V
29
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Micro-Fluorimeter
CMOS Backplane
CMOS SPAD Detector Array
Microarray Spots
GaN µLED Array
Filter
30
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Microfluidic Cytometer 31
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Summary
• Fluorescence techniques
• Optical sensor Technology
• Optical detectors
• Case study - Label free DNA detection
32
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Case Study - DNA Based Biosensor
• Sensor using artificial DNA “switches”
• Output mechanism uses FRET
• Control through electrochemical release of ions
33
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Resonant Energy Transfer
• FRET involves energy transfer between two fluorophores with overlapping spectra
• Excitation of the donor leads to emission from the acceptor
Light spectrum
Excitation Emission
Donor Acceptor
Emission Excitation
34
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Resonant Energy Transfer
• The energy transfer is a near field process
• Förster length is between 50 and 100Å
• The acceptor quenches the donor fluorescence
5nm
480 nm
436 nm
CFP YFP
35
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Resonant Energy Transfer
• The energy transfer is a near field process
• Förster length is between 50 and 100Å
• The acceptor quenches the donor fluorescence
5nm
480 nm
436 nm
CFP YFP
5nm
535 nm
FRET
436 nm
480 nm
CFP YFP
35
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Holliday Junction
• Named for Robin Holliday
• 4-way junction between DNA strands
• Artificial but similar to a structure that occurs in replication
36
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
‘OPEN’ Conformation
Holliday Junction
• HJs can switch between two shapes
• This happens when exposed to cations
• These screen phosphate charges leading to collapse
37
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Holliday Junction
• HJs can switch between two shapes
• This happens when exposed to cations
• These screen phosphate charges leading to collapse
‘CLOSED’ Conformation
37
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Sensor Concept• Attach donor and
acceptor fluoro-phores to HJ arms
• Switching detected through observing the FRET signal
• This measures presence of ions Open
No FRETClosed FRET
38
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Sensor Concept
• The emission spectra of the FRET changes with HJ shape
• Emission peak of the donor drops and the acceptor increases when the HJ closes
Low salt No FRET
fluor
esce
nce
wavelength / nm
λ1 λ2
D AA
‘OPEN’
470 nm 517 nm
39
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Sensor Concept
High salt FRET occurs
fluor
esce
nce
wavelength / nm
λ3λ1
D A
‘CLOSED’
470 nm 584 nm• The emission spectra
of the FRET changes with HJ shape
• Emission peak of the donor drops and the acceptor increases when the HJ closes
40
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
HJ FRET Biosensor
• Probe is an incomplete, three strand HJ with fluorophores attached
• Completed with complementary oligonucleotide and switching observed
41
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
HJ Actuation Microsystem
Ag/AgCl Reference Electrode
Ion Switching Film
Interdigitated electrodes for electrochemical control of switching ion (Mg2+) conc.
42
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
HJ Actuation Microsystem
• Pro-HJ probes spotted onto glass cover slips
• Fixed over micro-electrodes with spacer to form fluid chambers
• Complementary and non-complementary oligos in solution
43
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
FRET Measurements
• Measurements made using fluorescence scanner designed for imaging microarrays
• Excitation of the donor and imaging at both the donor and acceptor emission wavelength
• Extraction of intensity data from spots and calculation of FRET ratio (Iacceptor/Idonor)
44
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
FRET Image• Wetted area in centre of
image
• 2 types of spot:
‣ HJ with donors and acceptors
‣ HJs with only donor dye
45
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Initial Results
• Mg2+ ions switched into and out of solution0.0000
1.5000
3.0000
4.5000
6.0000
Mg- Mg+ Mg- Mg+ Mg- Mg+ Mg- Mg+
46
Stewart SmithBiosensors and InstrumentationBeijing University of Posts and
Telecommunications 2019
Case Study Summary• Switching of HJ molecular conformation can be
measured with FRET
• Completion of an HJ with a complementary DNA strand can be detected
• Label free biosensing of DNA with standard microarray measurement
• Switching controlled with microelectrodes
47