Japanese Guidance for Ventricular Assist Devices/Total Artificial Hearts *†Takashi Yamane, ‡Shunei Kyo, §Hikaru Matsuda, **Yusuke Abe, **Kou Imachi, ††Toru Masuzawa, ‡‡Takeshi Nakatani, §§Kazuhiro Sase, ***Koichi Tabayashi, †††Setsuo Takatani, ‡‡Eisuke Tatsumi, ‡‡‡Mitsuo Umezu, and §§§Toshie Tsuchiya *National Institute of Advanced Industrial Science and Technology,Tsukuba, Japan; †Pharmaceuticals and Medical Devices Agency,Tokyo, Japan; ‡University of Tokyo Hospital,Tokyo, Japan; §Hyogo University of Health Sciences, Kobe, Japan; **Graduate School of Medicine, University of Tokyo,Tokyo, Japan; ††Graduate School of Engineering, Ibaraki University, Hitachi, Japan; ‡‡National Cardiovascular Center, Suita, Osaka, Japan; §§Medical School, Juntendo University,Tokyo, Japan; ***Graduate School of Medicine,Tohoku University, Sendai, Japan; †††Tokyo Medical and Dental University,Tokyo, Japan; ‡‡‡Department of Advanced Medical Engineering,Waseda University,Tokyo, Japan; and §§§National Institute of Health Sciences,Tokyo, Japan Abstract: To facilitate research and development (R&D) and to expedite the review processes of medical devices, the Ministry of Health, Labor and Welfare (MHLW) and the Ministry of Economy,Trade and Industry (METI) founded a joint committee to establish guidance for newly emerging technology. From 2005 to 2007, two working groups held discussions on ventricular assist devices and total artificial hearts, including out-of-hospital programs, based on previ- ous guidance documents and standards. Based on this discussion, the METI published the R&D Guidelines for innovative artificial hearts in 2007, and in 2008 the MHLW published a Notification by Director regarding the evaluation criteria for emerging technology. Key Words: Guidance—Evaluation criteria—Ventricular assist device—Total artificial heart. Although heart transplantation and artificial heart implantation are two complementary approaches for severe heart disease, the number of donor hearts is limited in Japan.The total number of registrations for heart transplantation was 159 over the past 12 years, while only 70 donors were found in Japan, with an average of five to six a year. Recent innovations in artificial hearts, on the other hand, particularly in reducing the pump size and in developing noncontact bearings, have enabled devices to be implanted and have allowed patients to be discharged home. Artifi- cial hearts are applicable as bridge-to-transplantation (BTT), bridge-to-bridge, bridge-to-recovery (BTR), or destination therapy (DT). Devices for DT, particu- larly those for long-term home care, will be necessary in the near future. However, in Japan there has been a device lag of several years compared with the USA from the application to the approval of medical devices. To this end, the necessity has been recog- nized not only for promotion of research and devel- opment (R&D) for these devices among academia and industry, but also for the establishment of guid- ance for R&D and for clinical evaluation. There have been several previous guidance docu- ments and standards for ventricular assist devices (VADs) and total artificial hearts (TAHs):the Food and Drug Administration (FDA) Guidance for VAD/ TAH (1), the National Institutes of Health Request for Proposal for TAH (2), the American Society for Artificial Internal Organs (ASAIO)--Society of Tho- racic Surgeons (STS) Reliability Recommendation doi:10.1111/j.1525-1594.2010.01091.x Received March 2010; revised June 2010. Address correspondence and reprint requests to Dr. Takashi Yamane, National Intitute of Advanced Industrial Science and Technology (AIST), Namiki1-2-1, Tsukuba, Ibaraki, 305-8564, Japan. E-mail: [email protected] Presented in part at the 17th Congress of the International Society for Rotary Blood Pumps held October 1–3, 2009 in Singapore. Artificial Organs 34(9):699–702, Wiley Periodicals, Inc. © 2010, Copyright the Authors Artificial Organs © 2010, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc. 699

Japanese Guidance for Ventricular Assist Devices/Total Artificial Hearts

Embed Size (px)

Citation preview

Page 1: Japanese Guidance for Ventricular Assist Devices/Total Artificial Hearts

aor_1091 699..702

Japanese Guidance for Ventricular Assist Devices/TotalArtificial Hearts

*†Takashi Yamane, ‡Shunei Kyo, §Hikaru Matsuda, **Yusuke Abe, **Kou Imachi,††Toru Masuzawa, ‡‡Takeshi Nakatani, §§Kazuhiro Sase, ***Koichi Tabayashi,

†††Setsuo Takatani, ‡‡Eisuke Tatsumi, ‡‡‡Mitsuo Umezu,and §§§Toshie Tsuchiya

*National Institute of Advanced Industrial Science and Technology, Tsukuba, Japan; †Pharmaceuticals and Medical DevicesAgency, Tokyo, Japan; ‡University of Tokyo Hospital, Tokyo, Japan; §Hyogo University of Health Sciences, Kobe, Japan;

**Graduate School of Medicine, University of Tokyo, Tokyo, Japan; ††Graduate School of Engineering, Ibaraki University,Hitachi, Japan; ‡‡National Cardiovascular Center, Suita, Osaka, Japan; §§Medical School, Juntendo University, Tokyo,

Japan; ***Graduate School of Medicine, Tohoku University, Sendai, Japan; †††Tokyo Medical and Dental University, Tokyo,Japan; ‡‡‡Department of Advanced Medical Engineering, Waseda University, Tokyo, Japan; and §§§National Institute of

Health Sciences, Tokyo, Japan

Abstract: To facilitate research and development (R&D)and to expedite the review processes of medical devices, theMinistry of Health, Labor and Welfare (MHLW) and theMinistry of Economy, Trade and Industry (METI) foundeda joint committee to establish guidance for newly emergingtechnology. From 2005 to 2007, two working groups helddiscussions on ventricular assist devices and total artificialhearts, including out-of-hospital programs, based on previ-

ous guidance documents and standards. Based on thisdiscussion, the METI published the R&D Guidelinesfor innovative artificial hearts in 2007, and in 2008 theMHLW published a Notification by Director regardingthe evaluation criteria for emerging technology. KeyWords: Guidance—Evaluation criteria—Ventricular assistdevice—Total artificial heart.

Although heart transplantation and artificial heartimplantation are two complementary approaches forsevere heart disease, the number of donor hearts islimited in Japan.The total number of registrations forheart transplantation was 159 over the past 12 years,while only 70 donors were found in Japan, with anaverage of five to six a year. Recent innovations inartificial hearts, on the other hand, particularly inreducing the pump size and in developing noncontactbearings, have enabled devices to be implanted andhave allowed patients to be discharged home. Artifi-

cial hearts are applicable as bridge-to-transplantation(BTT), bridge-to-bridge, bridge-to-recovery (BTR),or destination therapy (DT). Devices for DT, particu-larly those for long-term home care, will be necessaryin the near future. However, in Japan there has beena device lag of several years compared with the USAfrom the application to the approval of medicaldevices. To this end, the necessity has been recog-nized not only for promotion of research and devel-opment (R&D) for these devices among academiaand industry, but also for the establishment of guid-ance for R&D and for clinical evaluation.

There have been several previous guidance docu-ments and standards for ventricular assist devices(VADs) and total artificial hearts (TAHs): the Foodand Drug Administration (FDA) Guidance for VAD/TAH (1), the National Institutes of Health Requestfor Proposal for TAH (2), the American Society forArtificial Internal Organs (ASAIO)--Society of Tho-racic Surgeons (STS) Reliability Recommendation

doi:10.1111/j.1525-1594.2010.01091.x

Received March 2010; revised June 2010.Address correspondence and reprint requests to Dr. Takashi

Yamane, National Intitute of Advanced Industrial Science andTechnology (AIST), Namiki1-2-1, Tsukuba, Ibaraki, 305-8564,Japan. E-mail: [email protected]

Presented in part at the 17th Congress of the InternationalSociety for Rotary Blood Pumps held October 1–3, 2009 inSingapore.

Artificial Organs34(9):699–702, Wiley Periodicals, Inc.© 2010, Copyright the AuthorsArtificial Organs © 2010, International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

699

Page 2: Japanese Guidance for Ventricular Assist Devices/Total Artificial Hearts

for Mechanical Circulatory Support Systems (3), TheAssociation for the Advancement of Medical Instru-mentation Technical Information Report for VAD/TAH (4), the Japanese Society for Artificial OrgansGuidance for Clinical Trials (5), the Reliability Rec-ommendation by the National Clinical Trial InitiativeSubcommittee (6), and International Organizationfor Standardization (ISO) 14708-5 on CirculatorySupport Devices as Active Implants (7).

To facilitate R&D and to expedite the reviewprocesses of medical devices, the Ministry of Health,Labor and Welfare (MHLW) and the Ministry ofEconomy, Trade and Industry (METI) founded ajoint committee to establish guidance for newlyemerging technology. Under the joint committee,two working groups were formed to establish Japa-nese guidance for VAD/TAH. The members of thecommittee represented the Japanese Association forThoracic Surgery and the Japanese Society for Arti-ficial Organs. From 2005 to 2007, the two workinggroups, chaired by Dr. S. Kyo and Dr. H. Matsuda,respectively, held discussions that referred to (i)previous guidance and standards; (ii) the results of aquestionnaire on an out-of-hospital VAD programfor domestic hospitals; and (iii) a survey on out-of-hospital VAD programs through a mission to UShospitals and universities. Based on this discussion,the METI published the R&D Guidelines for inno-vative artificial hearts (8) in 2007 and the MHLWpublished a Notification by Director in 2008 regard-ing the evaluation criteria for emerging technology(9,10).

R&D GUIDANCE

The R&D Guidance, published as METI R&DGuidelines for innovative artificial hearts (8), is com-posed of the following items:

1 Purpose of use: The present guidance covers left/right ventricular assist devices, biventricular assistdevices, and TAHs for bridge and destinationtherapy.

2 Human factors and environments: There shouldbe no adverse events during transportation orexercises. No adverse events in any environmentduring air travel or computed tomography. Noimpingement or necrosis of surrounding tissues.Allowable level of noise/vibration. Usabilityshould be considered for controllers andbatteries. Alarms should satisfy the InternationalElectrotechnical Commission (IEC) 60601-1-8.

3 Fluid dynamic characteristics: Nonpulsatilepumps should satisfy ISO 5198 and pulsatile

pumps ISO 4409. Supplemental information canbe introduced by numerical analyses and/or flowvisualizations.

4 Heating characteristics: There should be noadverse events on surrounding tissues caused byhot spots.The temperature rise should satisfy ISO14708-1.

5 Electronic safety: The device should satisfy IEC60601-1 and ISO 14708-1.

6 Electromagnetic compatibility: The device shouldsatisfy IEC 60601-1-2 and IEC CISPR-11.

7 Control and monitoring: Device should be able tokeep intended flow and flow stability and beimplemented with necessary limiters and withnecessary monitors.

8 Conduits, grafts, artificial heart valves, or heartcuffs: The conduits should be evaluated with ISO7198 and the valves with ISO 5840. The conduits,grafts, or heart cuffs should not induce stenosis orcause air sacks. Junctions should be evaluatedagainst tension, twist, vibration, kink, or leakage.

9 Materials safety: The device materials should beevaluated with ISO 10993-1.

10 Biocompatibility: The device biocompatibilityshould be evaluated with ISO 10993-1, ISO10993-4, and ASTM F1841-97.

11 Animal testing: There should be no severethrombosis caused by the device. The samplesize should be at least six animals and the dura-tion 60 days. Whether a test involving more thaneight animals and 90 days is necessary should beconsidered from the point of international har-monization.

12 Durability (Reliability): The reliability testsshould be conducted with 80% reliability, 60%confidence level (80% is preferable), and a periodof 6 months, to be extended to over 2 years. Testconditions such as flow and pulsatility should bedefined in advance.

Annex A. Supplement for TAH: Common factorsand discrepancies between TAH and VAD.

Annex B. Supplement for durability test: Test con-ditions should reflect the patient’s physiological con-ditions and lifestyle. Handling rules for variousevents should be defined in advance.

Annex C. Supplement for out-of-hospital program:Requests for organizing a medical team for VAD; forestablishing a training system for patients and car-egivers; for establishing discharge criteria; for clarify-ing the emergency system to patients, caregivers, themedical team, and the ambulance system; for arrang-ing a monitoring system; and for arranging a devicemaintenance system.

T. YAMANE ET AL.700

Artif Organs, Vol. 34, No. 9, 2010

Page 3: Japanese Guidance for Ventricular Assist Devices/Total Artificial Hearts

CLINICAL EVALUATION GUIDANCE

The clinical evaluation guidance, published asMHLW Notification by Director (9), is composed ofthe following items:

1 Fundamentals: Human factors and environments2 Nonclinical evaluation

In vitro evaluation (including Annex 1)In vivo evaluation (including Annex 2)

3 Clinical evaluation3.1 Compliance with good clinical practice3.2 Clinical trial protocol

3.2.1 Fundamentals: End points, other treat-ments, control, informed consent,patient management/follow-up, out-of-hospital program (Annex 3), trainingprogram for medical and co-medicalstaff, data safety monitoring board, cri-teria for discontinuing the clinical trial,etc.

3.2.2 Applicable diseases: End-stage cardiacdiseases as defined by the Japanese Cir-culation Society.

3.2.3 Objectives: BTT, BTR, or DT.Inclusion criteria: Patients with end-stage heart disease with extremely lowquality of life (QOL).Exclusion criteria: Severe infection, irre-versible multiorgan failure, pregnancy,complications of chronic pulmonarydisease, pulmonary thrombosis, or pul-monary hypertension, etc.

3.2.4 Sample size and duration: Overseasdata allowed. End points suitable for thefeasibility study or the pivotal study(Annex 4).

3.2.5 Facility qualification (Annex 5)3.2.6 Data acquisition rules3.2.7 Adverse events3.2.8 Safety evaluation3.2.9 Final evaluation (Annex 6)

Annex 1. Durability test (Same as R&D guidancesection 12)

Annex 2. Sample size and duration for animaltesting (Same as R&D guidance section 11)

Annex 3. Out-of-hospital program (Same as R&Dguidance Annex C)

Annex 4.Sample size and duration for clinical trials:Recommendations for clinical trials are around fivecases for a feasibility study and around 15 for a pivotalstudy. The recommendations for end points are 3months for the feasibility study, and, in a pivotal study,6 months for BTT and 12 and 24 months for DT.

Annex 5. Facilities for clinical trials: Institutionscertified as VAD facilities that perform more than100 cardiac surgeries a year.VAD implantation expe-riences for more than five cases. They should haveadequate transplantation facilities or be closely affili-ated with these. Staff members must have finisheda training course and there must be a committeeincluding a cardiologist. The procedure should beswitchable to paracorporeal VAD.

Annex 6. Final evaluation: (the Global Harmoni-zation Task Force [GHTF] SG5/N1R8 and SG5/N2R8) For bridge VAD (BTT and BTR), clinicalevaluation should be conducted (i) when patientsremain alive for more than 6 months with satisfactoryQOL, (ii) in the case of heart transplantation within 6months, or (iii) in the case of recovery within 6months. For DT VAD, clinical evaluation should beconducted (i) when patients remain alive for morethan 12 months with satisfactory QOL, (ii) in the caseof heart transplantation within 12 months, or (iii) inthe case of recovery within 12 months. For TAH,clinical evaluation should be conducted whenpatients remain alive for more than 2 months withsatisfactory QOL.

DISCUSSION

Table 1 shows a comparison between historicalguidance and standards. Regarding reliability testing,the FDA guidance (1) requested that X or moredevices are used with less than Y failures over twicethe intended use for short-term use and over Z yearsfor long-term use.The ASAIO-STS (3) required 80%reliability with at least 60% confidence for 1-yearmission life. The National Clinical Trial InitiativeSubcommittee (6) required the same level beforeInvestigative Device Exemption (IDE).The ISO/DISrequires that a reliability of X with at least Y confi-dence for a Z-year mission life is described.The Japa-nese guidance requires 80% reliability with at least60% (80%) confidence for 6 months, similar to theASAIO-STS recommendation.

Regarding animal testing, the FDA (1) required aduration of 5 months. The ASAIO-STS reduced therequest to eight animals for 90 days before IDE. TheISO/DIS requires only that the sample size and dura-tion are described. The Japanese guidance reducedthe request to at least six animals for 60 days withconsideration of eight animals for 90 days if neces-sary, which is a small reduction from the ASAIO-STSrecommendation.

Regarding clinical trials, the Japanese guidancereferred to practical goals for a feasibility studyand for a pivotal study, although most guidance

JAPANESE GUIDANCE FOR VAD/TAH 701

Artif Organs, Vol. 34, No. 9, 2010

Page 4: Japanese Guidance for Ventricular Assist Devices/Total Artificial Hearts

documents and standards do not refer to practicalperiods. The present guidance is expected to con-tribute to the clinical application of VAD/TAH inJapan.

CONCLUSIONS

The R&D guidance and the clinical evaluationguidance were established through collaborationbetween academia, industry, and government. Suchcollaboration is very important, not only for artificialhearts but also for all newly emerging medicaldevices. Since the present article is an overview pre-sented by the authors and is not an authorized guid-ance by the MHLW or PMDA, it is recommendedthat readers refer to the original guidance.

Acknowledgments: The introduction of guidanceis funded by the METI Introduction Program ofR&D Guidelines for Medical Devices and theMHLW Introduction Program of Evaluation Criteriafor Newly Emerging Medical Devices.

REFERENCES

1. FDA. Guideline for the preparation and content of applica-tions to the Food and Drug Administration for ventricularassist devices and total artificial hearts. F89-33838; 1987.

2. NIH. Phased readiness testing of implantable total artificialhearts. Request for proposal, NHLBI-HV-92-28; 1992.

3. ASAIO-STS. Long-term mechanical circulatory supportsystem reliability recommendation. Ann Thorac Surg 1998;66:1852–9.

4. AAMI. Ventricular assist and heart replacement systems.TIR26:2000.

5. JSAO. Clinical trial guidance for medical devices for seriousheart diseases. 2003.

6. National Clinical Trial Initiative Subcommittee. Long-termmechanical circulatory support system reliability recommen-dation by the National Clinical Trial Initiative Subcommittee.ASAIO J 2009;55:534–42.

7. ISO. Implants for surgery—Active implantable medicaldevices, Part 5: particular requirements for circulatory supportdevices. ISO 14708-5; 2010.

8. METI. R&D Guidelines for active implants 2007: innovativeartificial heart systems. May 2007. Available at: http://www.meti.go.jp/policy/mono_info_service/service/iryou_fukushi/index.html. Accessed August 1, 2010.

9. MHLW. Notification by Director, OMDE, Yakushokuki-hatsuNo. 0404002: evaluation criteria for emerging technology: clini-cal evaluation criteria for next-generation artificial hearts.April 4, 2008. Available at: http://dmd.nihs.go.jp/jisedai/heart/heart_public.pdf. Accessed August 1, 2010.

10. Tatsumi E, Nakatani T, Imachi K, et al. Domestic and foreigntrends in the prevalence of heart failure and the necessity ofnext-generation artificial hearts: a survey by the working groupon establishment of assessment guidelines for next-generationartificial heart systems. J Artif Organs 2007;10:187–94.

TAB

LE

1.C

ompa

riso

nof

hist

oric

algu

idan

cedo

cum

ents

and

stan

dard

sfo

rV

AD

/TA

H

Japa

nese

R&

Dgu

idan

ce(8

)&

Clin

ical

eval

uati

ongu

idan

ce(9

)IS

O14

708-

5(7

)

Rel

iabi

lity

reco

mm

enda

tion

byN

atio

nalC

linic

alTr

ialI

niti

ativ

eSu

bcom

mit

tee

(6)

ASA

IO-S

TS

Rec

omm

enda

tion

(3)

FD

Agu

idan

ce(1

)

Rel

iabi

lity

test

(in

vitr

o)

80%

relia

bilit

yw

ith

60%

(80%

)co

nfide

nce

for

6m

onth

s(t

obe

exte

nded

�2

year

s).

Rel

iabi

lity

Xw

ith

atle

astY

confi

denc

efo

ra

Zye

arm

issi

onlif

e.

Rel

iabi

lity

Rfo

rm

issi

ondu

rati

onT

and

confi

denc

eC

.*Fo

rde

stin

atio

nth

erap

y80

%re

liabi

lity

for

1ye

arw

ith

60%

confi

denc

ebe

fore

IDE

.The

stud

yis

not

ende

dun

tilf

ailu

res

have

been

accu

mul

ated

.

Aca

lcul

ated

80%

relia

bilit

yw

ith

atle

ast

60%

confi

denc

efo

r1

year

mis

sion

life.

Bef

ore

IDE

�8

syst

ems

for

�1

year

.

�X

devi

ces

wit

h�

Yfa

ilure

s.Sh

ort

term

:ove

r�

twic

eth

ein

tend

eddu

rati

on.

Lon

gte

rm:o

ver

�Z

year

s.

Ani

mal

test

(in

vivo

)

At

leas

tsi

xan

imal

sfo

r60

days

(eig

htan

imal

sfo

r90

days

).Sa

mpl

esi

zean

dim

plan

tdu

rati

onap

prop

riat

efo

rde

mon

stra

ting

the

safe

tyan

dpe

rfor

man

cefo

rth

ein

tend

edus

e.

IDE

**:I

na

min

imum

of8

anim

als

for

90da

ysD

urab

ility

and

perf

orm

ance

for

�tw

ice

the

expe

cted

dura

tion

.P

erm

anen

tim

plan

tde

vice

for

�5

mon

ths.

Clin

ical

tria

lSa

mpl

esi

ze:A

roun

d15

case

sfo

ra

pivo

tals

tudy

(aro

und

5ca

ses

for

afe

asib

ility

stud

y).

End

poin

t:3

mon

ths

for

api

lot

stud

y,6

mon

ths

for

BT

T,an

d12

(and

24)

mon

ths

for

DT

for

api

vota

lstu

dy.

ISO

1415

5-1

and

ISO

1415

5-2

PM

A**

*:A

llin

vitr

o,in

vivo

,and

clin

ical

data

repo

rted

byca

tego

ry•

The

num

ber

ofpa

tien

ts;

•T

henu

mbe

rof

pati

ent-

year

s.

Ara

tion

ale

for

the

num

ber

ofpa

tien

tsan

din

stit

utio

nsre

quir

edfo

rth

eco

nclu

sion

sab

out

the

safe

tyan

def

ficac

y(P

relim

inar

yID

Est

udy:

5–10

pati

ents

at1–

2ce

nter

s)

*E

x.1.

Ifei

ghts

yste

ms

for

1ye

arw

ith

nofa

ilure

,80%

confi

denc

efo

r80

%de

vice

sno

tto

fail

for

1ye

ar.E

x.2.

If15

syst

ems

for

1ye

arw

ith

one

failu

re,8

0%co

nfide

nce

for

80%

devi

ces

nott

ofa

ilfo

r1

year

.**

IDE

:Inv

esti

gati

onal

Dev

ice

Exe

mpt

ion.

***

PM

A:P

rem

arke

tApp

rova

l.

T. YAMANE ET AL.702

Artif Organs, Vol. 34, No. 9, 2010