Upload
violeta-grigoras
View
221
Download
0
Embed Size (px)
Citation preview
8/10/2019 Iyerppt
1/35
1
Perspective on Use of Statistical Tools
in Pharmaceutical Manufacturing
Karthik Iyer (CQE, CSSBB)
Senior Policy Advisor
CDER/OC/OMPQMarch 9th, 2012
AOAC Conference
* This presentation reflects the views of the author and should not
be construed to represent FDAs views or policies.
8/10/2019 Iyerppt
2/35
2
Agenda
Enforcement Action Examples
CGMP References
ASTM Standards
Conclusions
8/10/2019 Iyerppt
3/35
3
Recent warning letters and other compliance issues
Examples involving
Incorrect application of sampling plans
Equipment changes and process
capability
Container closure - determining baseline
defect rates
Recall exampleapplication of ASTM
E2709
8/10/2019 Iyerppt
4/35
4
1. Warning Lettersampling plans
Firm using sampling plans incorrectly Pooled X vials, used only 1 reportable value, but used
n=X in sampling plan.
.based your lot or batch acceptance/rejectioncriteria on a single reportable value averaged from apooled sample.
For .., you are collecting 3 pooled samples (each pool= 10 vials). This equates to a lot disposition action on3 reportable values with corresponding AQL of X%and LQ of X% respectively. This is not equivalent toan X or X plan as claimed in your SOP.
8/10/2019 Iyerppt
5/35
5
1. Warning Lettersampling plans
Response to 483 indicated firm did not know
how to use and interpret sampling plans
correctly. Firm did not understand concepts of
Acceptable Quality Level (AQL) and Limiting
Quality (LQ) and Operating Characteristic
Curve (OC) of a specific sampling plan.
8/10/2019 Iyerppt
6/35
6
2. Warning Letterequipment comparability and process
capability
Four (4) tablet products, various strengths
Initial process qualification used a single-sided tablet
press. During routine production, however, theseproducts were also being manufactured using a
double-sided tablet press.
Compression using the double sided press was not
qualified.
Firms response to the FDA 483 attempted to show
statistical equivalence between the single and double
sides presses.
8/10/2019 Iyerppt
7/35
7
2. Warning Letterequipment comparability and process
capability
The firms written response referenced the Cpk values for processesusing a double-sided tablet press and the single-sided tablet press.
FDA evaluation of the FDA 483 response
The Cpk value alone was not an appropriate metric to demonstratestatistical equivalence. Cpk analysis requires a normal underlyingdistribution and a demonstrated state of statistical process control. Thefirm did not address these issues in their response.
Statistical equivalence between the two presses could have beenshown by using either parametric or non-parametric (based ondistribution analysis) approaches and comparing means and variances.Neither of these approaches was employed. Firm did not use theproper analysis to support their conclusion that no significant differencesexisted between the two compression processes.
8/10/2019 Iyerppt
8/35
8
2. Warning Letterequipment comparability and process
capability
Issues
Data did not support proper statistical
conclusions. Firm did not understand underlying
assumptions required to conduct Process
Capability calculations.
Firm did not conduct proper statistical
analysis to demonstrate equivalence between
two operations.
8/10/2019 Iyerppt
9/35
9
3. Warning Lettercontainer closure quality and
baseline defect rates
ProductLVP in a dual chamber bag Numerous complaints of leaks, bursts, and premature activation
during 2 year period.
Root cause - variability in the film thickness that influenced thesealing of the bags. Bags have two seals and their strength (orweakness) relative to each other led to different failure modes.
Critical defects compromising sterility and stability.
Poor history for the supplier of this container closure
system. Incoming acceptance activities, as well as in-process
and finished product release activities, were foundinadequate.
8/10/2019 Iyerppt
10/35
10
3. Warning Lettercontainer closure quality and
baseline defect rates
Issues In responding to the 483, the firm equated customer
complaints to true manufacturing defect rate. They did
not understand that market incident data may nottrack with the quality of the product prior to release.
The proposed sampling plans to identify these knownpotential defects were not based on appropriatestatistics.
Firm did not understand sampling plan used for lotrelease.
Firm could not justify using a riskier sampling plan(higher probability of accepting bad product).
8/10/2019 Iyerppt
11/35
11
Application of ASTM E2709 - Standard Practice for Demonstrating Capability
to Comply with an Acceptance Procedure.
Tablets, Q value: 70%
Background: Firm was having recall issues due to dissolution failures on
stability. Dissolution data was analyzed using ASTM 2709. Sample data
below shown for 2 lots (Each row is a different lot). If evaluated correctly,
these lots would have been flagged as high risk for failure.
Unit
1
Unit
2
Unit
3
Unit
4
Unit
5
Unit
6
Unit
7
Unit
8
Unit
9
Unit
10
Unit
11
Unit
12 Mean SD RSD
USP -
PASS
or
FAIL
ASTM E2709
Probability @
95%
confidence
96% 72% 82% 74% 102% 70% 97% 63% 71% 78% 74% 60% 78% 14% 17% Pass 0.14%
77% 73% 90% 95% 92% 59% 73% 94% 60% 72% 62% 85% 78% 13% 17% Pass 0.14%
8/10/2019 Iyerppt
12/35
12
CGMP References
211.110(a) & (b)
211.165(d)
211.180(e)
Preamble for 21 CFR 210, 211
8/10/2019 Iyerppt
13/35
13
Key elements in these requirements
Control procedures
Monitor the output
Performance Variability in the characteristics of in-process
material and the drug product
.derived fromprevious acceptable processaverage and process variability estimates
(where possible)
.determined by...suitable statistical
procedures(where appropriate)
8/10/2019 Iyerppt
14/35
14
Statistics
Can sample tablets at any stage of process and analyze for: Weight.
Content Uniformity.
Dissolution.
Other critical quality attributes and or parameters of interest.
Can make decisions at any stage of process with respect to: Ability for a lot to pass USP UDU and or Dissolution tests in the future.
(ASTM E2709)
Confidence in sampling. (ASTM E2334 & ASTM E122)
Capability and Performance analysis. (ASTM E2281) Statistical Process Control Charts. (Monitor Variation, ASTM E2587)
Following tools illustrate making inferences about untested units ona particular attribute, variable and or parameter with respect tosample size and an associated confidence.
8/10/2019 Iyerppt
15/35
15
Voluntary Consensus Standards:
US Government Agencies OMB Circular A119
Federal Participation in the Development and Use ofVoluntary Consensus Standards and in Conformity
Assessment Activities (Rev. Feb 10, 1998) directs agencies to use voluntary consensus
standards in lieu of government-unique standardsexcept where inconsistent with law or otherwiseimpractical
intended to reduce to a minimum the reliance byagencies on government-unique standards.
http://www.whitehouse.gov/omb/circulars/a119/a119.html
http://www.whitehouse.gov/omb/circulars/a119/a119.htmlhttp://www.whitehouse.gov/omb/circulars/a119/a119.html8/10/2019 Iyerppt
16/35
16
Sample Size Effect on RSD Limit @ 95%Confidence / 95%Probability
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
85.1 87.5 90 92.5 95 97.5 100 102.5 105 107.5 110 112.5 114.9
Sample Mean
RS
n=10
n=20
n=30
n=100
n=500
n=1000
Sample Size Effect on RSD Limit @ 99% Confidence / 95% Probability
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
85.1 87.5 90 92.5 95 97.5 100 102.5 105 107.5 110 112.5 114.9
Sample Mean
RSD
n=10
n=20
n=30
n=100
n=500
n=1000
ASTM E2709Standard Practice for Demonstrating
Capability to Comply with an Acceptance
Procedure
One tool to analyze
Uniformity of Dosage
Units
8/10/2019 Iyerppt
17/35
17
ASTM E2709 ExplanationStandard Practice for Demonstrating Capability
to Comply with an Acceptance Procedure
Slide shows the relationship between sample size andtolerance for variability. As sample size increases, so doesthe tolerance for variability.
The analysis was performed using ASTM E2709-10. The
RSD limits on the y-axis represent the maximum variability alot can possess to ensure with 95 or 99% confidence thatthere is at least a 95 or 99% probability a lot will comply withthe USP Uniformity of Dosage Units test based upon a givensample size, confidence level, and sample mean.
For example: If you sampled 30 units and had a samplemean of 95%, then the maximum RSD value for those 30units would be ~3.0% to be 95% confident that there is atleast a 95% probability a future sample from the lot wouldpass the USP UDU test.
8/10/2019 Iyerppt
18/35
18
ASTM E2334Setting an Upper Confidence Bound For a Fraction or
Number of Non-Conforming items, or a Rate of Occurrence
for Non-conformities, Using Attribute Data, When There is a
Zero Response in the Sample
Confidence vs Sample Size
0
10
20
30
40
50
60
70
80
90
100
6 10 12 24 30 100 150 300 500 1000
Sample Size
Confidence
Pd=1%
Pd=0.5%
Pd=0.065%
8/10/2019 Iyerppt
19/35
19
ASTM E2334 ExplanationSetting an Upper Confidence Bound For a Fraction or
Number of Non-Conforming items, or a Rate of Occurrence
for Non-conformities, Using Attribute Data, When There is a
Zero Response in the Sample
Slide shows the relationship between Confidence andSample Size. As sample size increases, so doesconfidence demonstrated.
The analysis was performed using ASTM E2334-09.Keeping the maximum percent defective constant (1,0.5, and 0.065%) a line was generated to show howsample size effects the confidence demonstrated inhaving no more than the maximum percent defective. A
zero response was assumed (that is zero defects in thesample) and a binomial distribution was used. For example: If you desire a percent defective of no more than
0.5% and sample 30 units, then you are only ~15% confidentthat your lot has no more than 0.5% defects.
8/10/2019 Iyerppt
20/35
20
ASTM E2334Setting an Upper Confidence Bound For a Fraction or
Number of Non-Conforming items, or a Rate of Occurrence
for Non-conformities, Using Attribute Data, When There is a
Zero Response in the Sample
Sample Size vs Percentage of Non Conformities
0
5
10
15
20
25
30
35
40
10 20 30 50 500 1000
Sample Size
Percen
tageNonConforming
99%
Confidence
95%
Confidence
8/10/2019 Iyerppt
21/35
21
ASTM E2334 ExplanationSetting an Upper Confidence Bound For a Fraction or
Number of Non-Conforming items, or a Rate of Occurrencefor Non-conformities, Using Attribute Data, When There is a
Zero Response in the Sample
Slide shows the relationship between the upperconfidence bound on percent defects and sample size.
As sample size increases the upper confidence bound
on percent defects decreases. The analysis was performed using ASTM E2334-09.
Keeping the confidence level constant (95 and 99%) aline was generated to show how sample size effects theupper confidence bound percent defects. A zero
response was assumed (that is zero defects in thesample) and a binomial distribution was used. For example: If you want to be 99% confident that there is no
more than 1% defective units in your lot, then you must sample~460 units with a zero response.
8/10/2019 Iyerppt
22/35
22
ASTM E122Standard Practice for
Calculating Sample Size to Estimate, With Specified
Precision, the Average for a Characteristic of a Lot or
Process
Estimate Precision vs Sample Size
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
9.00
10.00
10 20 30 50 500 1000
Sample Size
P
recision
Range
=1.0
=3.0
=5.0
=6.0
=10.0
8/10/2019 Iyerppt
23/35
23
ASTM E122 explanationStandard Practice for
Calculating Sample Size to Estimate, With SpecifiedPrecision, the Average for a Characteristic of a Lot or
Process
Slide shows the relationship between sample
size and precision. As sample size increases,
so does your estimate precision. The analysis was done using ASTM E122-09.
Lines were generated using different sample
sizes to show the effect it has on your estimate
precision. For example: If you sampled 30 units and your sigma
value was 6, then your sample average is ~ +/-3.5%
of your true population average.
8/10/2019 Iyerppt
24/35
24
ASTM E2281Standard Practice for
Process and Measurement Capability Indices
Confidence in Manufacturing Capability
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
6 10 12 20 24 30 50 100 250 500 1000
Sample Size
ManufacturingCapability
95% Confidence
99% Confidence
8/10/2019 Iyerppt
25/35
25
ASTM E2281 ExplanationStandard Practice for
Process and Measurement Capability Indices
Slide shows the relationship between a reportedProcess Capability Index (Cpk(3.14)) andsample size. As sample size increases, so does
the reported Cpk. When reporting a Cpk, a lower 95 or 99%
confidence bound should always be the valuereported. As this value accounts for the samplesize in which the C
pkwas estimated.
For example: If you sampled 30 units and estimateda Cpkof 3.14, then the value reported should be ~2.5(that is I am 99% confident that the Cpkfor myprocess is at least 2.5). The analysis was done using
ASTM E2281-08.
8/10/2019 Iyerppt
26/35
26
ASTM E2281Standard Practice for
Process and Measurement Capability Indices
Confidence in Manufacturing Performance
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
6 10 12 20 24 30 50 100 250 500 1000
Sample Size
Manufa
cturingPerformanc
e
95% Confidence
99% Confidence
8/10/2019 Iyerppt
27/35
27
ASTM E2281 ExplanationStandard Practice for
Process and Measurement Capability Indices
Slide shows the relationship between a reportedProcess Performance Index (Ppk(2.79)) andsample size. As sample size increases, so does
the reported Ppk. When reporting a Ppk, a lower 95 or 99%
confidence bound should always be the valuereported. As this value accounts for the samplesize in which the P
pkwas estimated.
For example: If you sampled 30 units and estimateda Ppkof 2.79, then the value reported should be ~2.2(that is I am 99% confident that the Ppkfor my processis at least 2.2). The analysis was done using ASTME2281-08.
8/10/2019 Iyerppt
28/35
28
SPC (Statistical Process Control) Charts are acollection of very effective statistical-graphicaltools which can be used to:
Understand and diagnose your data.
Track performance to identify problems, or shifts inperformance (good or bad).
Control or adjust the process to maintain desired
performance. Can be applied for data based on Incoming, In-
process, or Lot release samples.
ASTM E2587Standard Practice for Use of Control Charts inStatistical Process Control
8/10/2019 Iyerppt
29/35
29
9181716151413121111
103
102
101
100
Sample
Sample
Mean __
X=101.868
UCL=103.048
LCL=100.689
9181716151413121111
4.5
3.0
1.5
0.0
Sample
SampleRange
_R=2.046
UCL=4.326
LCL=0
1
66
5
1
55
5
5
52
2
Voice of the ProcessIs the process in a state of control?
ASTM E2587Standard Practice for Use of Control Charts in Statistical Process ControlVariable X-bar-R chart
8/10/2019 Iyerppt
30/35
30
ASTM E2587Standard Practice for Use of Control Charts in Statistical Process Control
Chart is used to detect special causes ofvariation during manufacturing.
Control is determined against standard 8rules established by Dr. Walter Shewhart.
Preceding chart is called X-bar-Range withSubgroup size of 5 tablets (each point isan average of 5 individual results).
Control limits reveal true variability of theprocess.
8/10/2019 Iyerppt
31/35
31
ASTM E2587Standard Practice for Use of Control Charts in Statistical Process ControlAttribute nP chart
28252219161310741
12
10
8
6
4
2
0
Sample
SampleCount
__NP=4.9
UCL=11.38
LCL=0
Pass/Fail test for incoming tablet bottles (nP chart)
The tablet bottle is either pass or fail (Binary response)
Sample 100 bottles per incoming lot
Here we assume the failure rate is 5 bottles/100 bottlesNeed to know historical defect rate from supplier
8/10/2019 Iyerppt
32/35
32
Conclusions
What have we covered?
Enforcement action examples and CGMP references.
Use of statistics to quantify relationship betweenconfidence associated with attribute, variable and or
parameter of interest with respect to the sample size
collected.
To make inferences on untested units.
Can be applied to In-coming, In-process, or Finished
samples.
Can be used for real time manufacturing and or
annual/periodic product reviews.
8/10/2019 Iyerppt
33/35
33
Conclusions
Other Statistical Tools
Sampling plans
Do they describe Consumers Risk?
How are true defect rates calculated to use a particular
sampling plan?
Confidence, Prediction, and Tolerance Intervals
Is the correct statistical tool being applied for the
right analysis?
Do the tools help answer questions about
product quality and process performance?
8/10/2019 Iyerppt
34/35
34
Conclusions
The specific statistical tools and analysisdepends on what variables, attributes,parameters are being used to monitor process
performance and product quality. The preceding example is just one set of
statistical methods available to monitor processperformance and product quality.
Statistics is a tool to elicit information to confirmthat a specific manufacturing process isproducing quality product.
8/10/2019 Iyerppt
35/35
35
Acknowledgements
Alex ViehmannCDER/OPS
Grace McNallyCDER/OC