IONABBased on IMS Technology, Explosives & Narcotic Detector
CONTENTS
I. Introduction of Company
II. IONAB Overview
III. IONAB Demonstration
公司简介
Name of company : IMS TECHNOLOGY Co.,Ltd.
Name of representative : Myung_Won, SHIN
Date of business commencement : June 7, 2010
Business registration number : 113-86-40170
Address(Office) : #1209, 55, 34-Gil, Digital-ro,
Guro-gu, Seoul, Korea
Address(Factory) : #513, 55, 34-Gil, Digital-ro, Guro-
gu, Seoul, Korea
Website : www.imstechnology.co.kr
Main telephone number : +82-2-732-7005
Main fax. Number : +82-2-732-1004
Business Item : Explosives & Narcotics Detector
Introduction
© 2013 by IMS TECHNOLOGY Co.,Ltd. / 4
IMS TECHNOLOGY WORK-ROOM
Workroom
Worktable-1
QC Room
Worktable-2
Shedding
We have a independent manufacturing facility in this building #513.We have ISO 9001 certification(Certificate No.:12-Q-0249-00) and to
produce and maintenance of product by ISO 9001.
History & Results History
KOLAS Test Report (JUN
2010)
IMS Patient-KR
(AUG 2010)
ISO9001 Certification (AUG 2010)
Research Center
(AUG 2010)
Venture Business (NOV 2010)
KTL Test Report (JAN
2011)
CE Certification (MAR 2011)
CHINA Performance Report (MAR 2011)
TAILAND Performance Report
(APR 2011)
Changing Company Name(Old:WOOJU C&T) (APR 2012)
R&D Project The Ministry of Knowledge Economy
Development & Commercialization of analysis equipment using Ion
Mobility
01JUN 2009 ~ 31May 2011 (24 months)
The Ministry of Knowledge Economy
Development of Trace Detector for Narcotics and Explosives
01JAN 2010 ~ 31DEC 2012 (24 months)
The Minister of Environment
Development of in-situ VOCs measuring system using IMS technology
01DEC 2008 ~ 31DEC 2010 (25 months)
Sales & Marketing
KOREA / The Defense Security Command(AUG 2011)
KOREA / 707 Special Force (SEP 2011)
KOREA / Navy HQ (OCT 2011)
CHINA / The Ministry of Public Security (AUG 2010)
CHINA / National Assembly (SEP 2011)
CHINA / Subway station in Beijing (OCT 2011)
THAILAND / Royal Thai Navy (JAN 2010)
PAKISTAN / The border Guard (NOV 2011)
PAKISTAN / The border Guard (DEC 2011)
PAKISTAN / The border Guard (FEB 2012)
Introduction of Com-pany
R&D• IMS Application
Development• Sensor, Module Design• Machine Design• System Development• S/W Configuration
VOCs & Environmental Hazardous Substances• Environmental hazardous
substances : ppb base• VOCs, emissions,
pesticides…
Intelligent lightweight • Compact and easy to carry
modification• Multi-dimensional
management through network and integrated main C/S
Explosives & Narcotics• Explosives & Narcotics :
ppb base• Explosives : TNT, DNT,
C4, RDX, TATP, PETN, SEMTEX, NG…
• Narcotics: cocaine, codeine, heroin, THC, MAMP, AMP, MDMA, MDA, ketamine, GHC….
Early diagnosis of real-time • Respiratory Diagnostics• Lung diseases• Diabetes• Epileptic
R&D
Explosives & Narcotics
Environmental Hazardous Substances
Early diagnosis of real-time
Intelligent lightweight
VISION
Introduction of Com-pany
Main Certification
Patent-KR(10-0971031)
C E(CE-K-0316-11-05-01-
3B)
TNT(KOLAS Test Report)
TNB(KOLAS Test Report)
COCAINE(KOLAS Test Report)
Fall impact Test(KTL Test Report)
CHINA(Performance Re-
port of Public Secu-rity )
THAILAND(Performance Re-
port of EOD)
ISO 9001(12-Q-0249-00)
Research Center(2010112300)
Certificate
© 2013 by IMS TECHNOLOGY Co.,Ltd. / 8
Demonstration on Abroad
We have demonstrated the IONAB at government and other public offices in China since 2010.
IONAB is..
• Detecting and Analyzing device based on IMS technology for trac-
ing hazardous materials, such as explosives, narcotics, etc.
IONAB overview
MAMP, THCCOCAINE…
NARCOTICS
TNT, C4TATP…
EXPLOSIVES
NOXs, SOXsVOLITILE…..
VOCs
Analysis & De-tection Identifi-cation of micro traces of haz-ardous materi-
als
BULKBULK NEUTRONNEUTRON
X-RayX-Ray
Other NuclearOther Nuclear
Standard(Transmission)
Standard(Transmission) Dual EnergyDual Energy BackscatteringBackscattering
ComputedTomographyComputed
Tomography
Thermal NeutronAnalysis
Thermal NeutronAnalysis
(Pulsed) FastNeutron Analysis
(Pulsed) FastNeutron Analysis
(Pulsed) Fast-ThermalNeutron Analysis
(Pulsed) Fast-ThermalNeutron Analysis
Associated alpha particle Time of Flight
Associated alpha particle Time of Flight
Neutron BackscatterNeutron
Backscatter
Gammabackscatter
Gammabackscatter
Gammatransmission
Gammatransmission
ElectroMagneticElectro
MagneticNuclear MagneticResonance(NMR)
Nuclear MagneticResonance(NMR)
Nuclear QuadrupleResonance(NQR)
Nuclear QuadrupleResonance(NQR)
TRACETRACE
Electronic / Chemical
Electronic / Chemical
Ion MobilitySpectrometryIon Mobility
SpectrometryChemi
LuminescenceChemi
LuminescenceElectron
Capture DetectorElectron
Capture Detector
Surface Acoustic Wave
Surface Acoustic Wave Thermo-RedoxThermo-Redox Field Ion
SpectrometryField Ion
Spectrometry
(GC) MassSpectrometry
(GC) MassSpectrometry
BiosensorBiosensor DogsDogs Antigen / AntibodyAntigen / Antibody
CONTRABANDDETECTION
CONTRABANDDETECTION
Technology Classification
IONAB features
IONAB specificationITEM SPECIFICATION
Detector TypeIMS (Ion Mobility Spectrometry)
Ionization Implused Corona discharge
Analysis Materials Gaseous, Fluid and Solid state
Detection ModePositive / Negative ModeThermal(Smear) / Vapor Mode
Calibration Method
Automatic (Negative Mode)Semi-Automatic (Positive Mode)
Calibration Source TNT & Lactic acid
Operating Temperature -20 ~ 55°С
Operating Humidity 0 ~ 95%
Operating Time4 hours(Max. standing time 8H)
Power SupplyAC Input : 100~240V AC, 47~63 HzDC Input : +12V DC, 8A
IONAB features
ITEM SPECIFICATION
Dimensions (cm) W*D*HDetector : 13 * 32 * 18Controller : 15 * 9 * 30
Weight (kg)Detector : 2.5 Controller : 2.6 (Include batteries)
Potable Power Source Rechargeable Li-Ion Battery (2
set)
Warm-up Time Less than 15 min.
Analysis Time Immediately / Less than 5 sec.
Clear-down Time Immediately / Less than 5 sec.
Intake Volume6mL/s (Optimized), 15mL/s (Max.)
Sensitivity (Vapor) ppt
Sensitivity (Particle) Nano-gram
AlarmAlert sound, LCD & LED display
Warranty Pe-riod
IONAB system and parts and devices regarding labor support is up to 1 year.Charged maintenance for prompt and affordable maintenance after 1 year warranty pe-riod expires.
DETECTOR UNIT
CONTROLLER UNIT
EXTERNAL POWER SUP-
PLY UNIT
MOLECULAR SIEVE
25 PIN CA-BLE
POWER CABLE
CALIBRATION SAMPLES
KEYBOARD
SMEAR KIT
12” MONI-TOR
TECHNICAL MANUAL
IONAB Component
IONAB features
Detective materialscan be registered manually over 40 additional materials
Explosives Narcotics
TNT Cocaine
DNT Heroine
C4 Ketamine
RDX THC
TATP Methamphetamine
BLACK POUWDER Amphetamine
HMX MDA
PETN MDMA
TETRYL Diazepam
Ammonium nitrate Sauteralgyl
NG(Nitro Glycerin) Nimetazepam
SEMTEX Codein
• Already established D/B for detection of existing military explosives
• Selectively established D/B for detection of well-known narcotics and psychoactive medicines
• D/B usually can be added and amended By the purpose of users’ convenient operation
• Users can register the unknown materials description in unit by themselves without additional update cost
• LCD Display -Displaying the description of detected material
-Showing up the concentration of it too• LED Indicator• Detecting ALARM
IONAB features
Manufacturer IMS TechnologySmith Detec-
tionImplant Science
Product Name IONAB SABRE 4000 QS-H150
Appearance
Technology IMS IMS IMS
Limit of Detection Nano-gram Nano-gram Nano-gram
Contain Radioactive Source
X O X
Drug Detection O ∆ X
Library Update by user O X X
Auto Calibration O X X
Analysis Time < 5 sec. < 20 sec. < 30 sec.
Battery Time 8 hr. (Standby) 3 ~ 4 hr. 3 ~ 4 hr.
Comparison chart_1
Performance Comparison to competi-tors
IONAB competiveness
Detecting Mode
Calibration
Operating Condition
Operating Time
Dimension & Weight
Analysis TimeIONAB
I company
S company
Performance Comparison to competi-tors
Other units can be detected for only negative or positive mode.
IONAB can select both detection mode and detect explosives (including TATP), narcotics at the same time.
IONAB strong points_1
IONAB strong points
Maximization of detection coverage with alternative detection mode of positive and negative mode
Other units need calibration specimen additionally.
IONAB can be realized of self calibration without additional calibration specimen.
Simple procedure of calibration with just pushing buttons
Other units take more than 5 to 30 seconds for analysis and need a long term of cleaning.
IONAB has the fasted detection and cleaning time only for 5 seconds.
It means the detection time can be reduced up to about 3 times so IONAB can handle more various sam-
ples within same period quickly and perfectly.
The fastest detection time in the world: within 5 seconds
Other units have single battery. So operating & standby time is within 3 hours and lack of continuous op-
erating capability.
IONAB have dual Battery pack (4 hours/set) then it can be operated for 4 hours continuously(Max. standing
time is 8 hours). Also it can be operated with commercial electricity via connecting AC adapter.
IONAB strong points_2
IONAB strong points
The world best performance of battery
Other units have all-in-one structure then more than 5kg weighted unit is too heavy to handle and carry.
IONAB is consisted of separated structure, 2.5kg weighted detector is good to be handled and operated for a long
time.
Maximized portability with separated structure of con-troller and detector
IONAB is using the method of corona discharge then it is stable ionization and maximization efficiency.
Other units are operated with Ni-63 radioactive isotope so it is need the experts and licenses.
No risk of exposure to radioactive
Consumables : smear kit, molecular sieves.
No membrane filter and optional parts.
Actually, extra parts is not need.
We can support consumables 50% lower than competitors’ product.
Durability, Endurance for chemical, Persistence
Infrequent trouble -> low maintenance cost.
IONAB strong points_3
Low maintenance & operating Cost - Consumables, optional parts.
IONAB support quick and perfect service via the united application of manufacturing, R&D and Q&A sys-
tem.
User friendly service system
IONAB strong points
Demonstration_1
IONAB Demonstration
Demonstration_2
IONAB Demonstration
Demonstration_3
IONAB Demonstration
Current Market
Public YES YES YES
Private YES NO
Indus-try
YES NO
Target market
Airport & port Military & Police Customs Post Office Prison Government Office
Convention Subway Hotel & Shopping Mall
Military Industry Nuclear plant Petrochemical plant
Options & Consumables Parts
No. Description Note
1Molecular sieve- Bottle (1L)- Bottle (4L)
2 kinds of molecular sieve
2Smear kit- Mesh (stainless steel)
3Calibration Sample- Positive calibration- Negative calibration
4Battery pack- Li-ion battery pack
DC 12.1V, 11A
5 25pin connector 2 different length
6 Battery quick charger
7 Carrier 2 kinds of carrier
Air Sampler
Sampling paper Recycling kit
IMS Technology Co., Ltd. #1209, kolon science valley, guro-Dong, guro-Gu, Seoul, Korea Tel : +82-2-732-7005 Fax : +82-2-732-1004 E-mail : [email protected]