Upload
others
View
4
Download
0
Embed Size (px)
Citation preview
Introduction to UCL Chemistry
Mass Spectrometry Facility
https://www.ucl.ac.uk/chemistry/mass-spec
A mass spectrometer (MS) is an analytical instrument that produces a
beam of gas phase ions from sample (analytes), sorts the resulting
mixture of ions according to m/z ratios using electrical or magnetic fields
(or combination), and provides analog or digital output signals (peaks)
from which the m/z ratio and the intensity (abundance) of each detected
ionic species may be determined.
A mass spectrometer can only analyse IONS (electrically charged
molecules) and they must be in the GASEOUS phase.
Mass spectrometers MEASURE m/z (mass-to-charge ratios); where m is
the total of the masses of the component atoms and z is the number of
charges on the ion (positive or negative) expressed in whole numbers
(1,2,3, etc.).
1. Vaporised sample enters the ion source
2. Unresolved ion beam is accelerated to
the analyser where the ions are
separated either “in space” or “in time”
3. Ions with resolved m/z travel to the
detector
Mass spectrometers are composed of a sample
inlet, ion source, mass analyser, detector,
vacuum system, and computer.
While multiple types of each components can be
mixed and matched, e.g., different ion sources
with different analysers, the ionisation, analysis,
detection sequence must be maintained.
Analysers can be used in single, e.g., Q or TOF,
or in multi-analysers formats, e.g., Q-TOF and
TOF/TOF, with a collision cell incorporated
between the two analysers.
Computers control instrument operation, data
acquisition and data processing.
Block diagram of a mass spectrometer
1 2 3
Accela LC – LTQ mass spectrometer Waters LCT Premier Q-TOF mass spectrometer
The ion trap mass analyser The Q-TOF mass analyser
Waters UPLC – SQD mass spectrometer
The single quadrupole mass analyser
Aqilent LC – 6510 Q-TOF mass spectrometer
The Q-TOF mass analyser
Electron Ionisation (EI) source
The EI source located inside of the instrument’s vacuum chamber and held
at 130°C to maintain the analyte molecules in the vapor phase. Electrons
from the filament are collimated by a magnetic field and traverse the source
in a helical path (to increase efficiency) to the electron trap. The electrons
react with the analyte molecules to produce radical cations.
Reaction: M + e- → M+. + 2e-
Electrospray Ionisation (ESI)
As the droplets shrink the charge
density increases to the point at
which the droplets are no longer
stable and ions are released from the
liquid matrix into the vapor phase.
The choice of ionisation is determined by the polarity of the analyte molecules.
Mass ranges for the different ionisation methods. EI is selected for structural
determination, whereas ESI and MALDI are used for molecules above 1,000 Da.
Energy is added to molecules during ionisation. The distribution of the energy may result in the
breaking of chemical bonds and, consequently, in fragment ion formation. The fragmentation may
be so extensive that no molecular ion is observed.
The form of the molecular ion depends on the mode of ionisation and can include [M]+, [M+H]+ and
other ions, e.g., [M+Na]+.
The base peak represents the most stable ion resulting from the ionisation process and is the most
abundant peak in the spectrum. The intensities of all other are normalised with respect to the base
peak.
Ions, normally of lesser intensities and to the right of each molecular/fragment ion generally
represent isotopic species. Typically, but not always, isotope ions reflect the presence of carbon-
13 (13C).
A mass spectrum is a graphical representation of the m/z values (x-axis) of the ions together with
their intensities (y-axis), collected over a specified mass range.
The atoms of elements are composed of protons, electrons and neutrons. An element is
defined by the number of protons present in the nucleus. However, the number of
neutrons may vary over a small range, yielding products known as isotopes.
Carbon has three isotopes incorporating six, seven or eight neutrons yielding 12C, 13C
and 14C. The presence of the 13C isotope, which occurs with a natural abundance of
~1.1% per carbon atom results in the isotope peaks that are commonly found on the
right-hand side of all ions in mass spectra.
https://www.ucl.ac.uk/chemistry/mass-spec
(c)
(d)
EI mass spectra
[M]+˙
cr621_ei # 7-8 RT: 0.73-0.82 AV: 2 SB: 1 1.99 NL: 4.81E6 T: + c EI Full ms [59.50-800.50]
100 150 200 250 300 350 400 450 500 550 600 650 700 750 800
m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Rela
tive A
bundance
115.0
160.0
174.1
143.0
299.0 301.0
102.0
89.0
254.0
175.1
214.9
302.0 212.9
190.1
Instrument Resolution : 6120
Theoretical Mass (C11H10BrNO4) : 298.9788
Measured Mass : 298.9787
Error : 0.54 ppm
Ionisation mode and polarity
Mass range Electron Ionisation (+) mode
cr621_ei #7-8 RT: 0.73-0.82 AV: 2 SB: 1 1.99 NL: 4.81E6T: + c EI Full ms [59.50-800.50]
60 80 100 120 140 160 180 200 220 240 260 280 300 320 340
m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
100
Re
lative
Ab
un
da
nce
115.0
160.0
174.1
173.0
143.0
116.0
299.0301.0
144.0
102.0
89.062.8
254.0
130.0
175.1159.0
214.974.9
86.9 225.064.9 302.0212.9 226.9
145.0239.091.0 190.1
EXPERIMENTAL
THEORETICAL
288 290 292 294 296 298 300 302 304 306 308 310 312 314
m/z
0
10
20
30
40
50
60
70
80
90
100
0
10
20
30
40
50
60
70
80
90
100
Re
lative
Ab
un
da
nce
299.0301.0
300.0 302.0
299.0301.0
300.0 302.0
NL:3.29E6
cr621_ei#7-8 RT: 0.73-0.82 AV: 2 SB: 1 1.99 T: + c EI Full ms [59.50-800.50]
NL:4.44E5
C 11 H10 BrNO4: C 11 H10 Br1 N1 O4
pa Chrg 1
4 x10
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
5.5
6
6.5
7
7.5
8
8.5
9
9.5
+ESI Scan (rt: 1.373 min) Frag=175.0V Tag_01.d
417.0363
319.0596
833.0650
Counts vs. Mass-to-Charge (m/z)
100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600
[M+H]+
[2M+H]+
OH
O
S
O
O
OH
OH
O
P
OH
OOH
NH
CH3
O
NH2
Exact mass: 417.03634 Da
Chemical formula: C11
H18
N2O
11PS
[M+H]+
Mass error: 0.09 ppm
H+
[M-H3PO4]+
Ionisation mode
and polarity Electrospray ionisation (+) mode
[M-3H]3-
[M-2H]2-
3 x10
0
0.025
0.05
0.075
0.1
0.125
0.15
0.175
0.2
0.225
0.25
0.275
0.3
0.325
0.35
0.375
0.4
0.425
0.45
0.475
0.5
0.525
0.55
0.575
0.6
0.625
0.65
0.675
0.7
0.725
0.75
0.775
0.8
0.825
0.85
0.875
0.9
0.925
0.95
0.975
1
1.025
1.05
1.075 -ESI Scan (rt: 7.115-7.264 min, 4 scans) Frag=20.0V AP1HP1_MS.d
1915.0
2873.0
Counts vs. Mass-to-Charge (m/z)
100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400 2500 2600 2700 2800 2900 3000 3100 3200
m/z
100%
3200 1600 0
5' G A C A T T I C T T T I C A A T I T C 3'
[M-3H]3-
[M-2H]2-
Ionisation mode and polarity Electrospray ionisation (-) mode
S1494
m/z100 200 300 400 500 600 700 800 900 1000
%
0
100
S1494 9 (0.317) 1: TOF MS ES+ 4.79e3241.0
141.0
Instrument Resolution : 13,500
Theoretical Mass (C9H11N4S2+H) : 240.0498
Measured Mass : 240.0499
Error : 0.42 ppm
Ionisation mode and
polarity Electrospray ionisation (+) mode
Type of mass analyser
S1494
m/z222 224 226 228 230 232 234 236 238 240 242 244 246 248 250 252 254 256 258 260 262 264
%
0
100
%
0
100
S1494 (0.036) 1: TOF MS ES+ 8.04e12241.1
242.1 243.1
S1494 10 (0.345) 1: TOF MS ES+ 9.42e3241.0
242.0 243.0
theoretical
experimental
Ionisation mode and
polarity
Type of mass analyser
Ion counts