Upload
others
View
3
Download
0
Embed Size (px)
Citation preview
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Maria Pia Toni
Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti
Agenzia Nazionale per le Nuove Tecnologie, l’Energia e lo Sviluppo Economico sostenibile Italy
International Symposium “Standards, Applications and QA in Medical Radiation Dosimetry ”
November 9−12 2010, Vienna, Austria
New brachytherapy standards paradigm shiftNew brachytherapy standards paradigm shift
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
inci
den
ce o
f dam
age
(%)
tum
or
contr
ol (%
)
Dw dose delivered (a.u.)
A lower dose delivered to the target volume could result in lack of tumor
control
Increasing damage
to healthy tissue could result from
exceeding the tolerance levels
uclinic
< 5% (k = 1)
Dw , is the quantity of interest for dosimetry in radiotherapy, this includes either the dosimetry of external radiation beams or the dosimetry of brachytherapy (BT) sources.
The IAEA TRS 398 (2000) recommends a direct traceability chain starting on Dw primary standards:
Rationale for a paradigma shift in BT dosimetry
to avoid additional uncertainties in the measurement of Dw and to optimize the dose delivery in radiation therapy;
to achieve the goal of 5% (k = 1) for the uncertainty of the dose delivered to the target volume at reference conditions, in the clinic.
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
In external beam dosimetry, Dw primary standards are currently operating for dosimetry of external radiation beams used in radiotherapy and the uncertainty on Dw,1 cm is about 2% (k = 1).
In BT dosimetry, no Dw primary standards have been so far made available for dosimetry of the gamma sources used in current BT treatments and the uncertainty on Dw,1 cm is typically from 5% to 8% (k = 1).
Rationale for a paradigma shift in BT dosimetry
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
(a) Brachytherapy sources and (b) transfer standards
calibrated in terms of Reference Air Kerma Rate.
(a) Air Kerma primary standards and secondary standards calibrated in terms of air kerma for the energy
spectrum of brachytherapy sources or (b) Reference Air Kerma Rate
primary standards.
STATE OF ART(a)
(b)
( ) ( )( ) ( ) ( )θ
θθθ ,
,,,
00
rFrgrGrGSrD KW ⋅⋅⎥
⎦
⎤⎢⎣
⎡⋅Λ⋅=&
dose rate constant
Conversion from Reference Air Kerma Rate to Absorbed Dose Rate to water (AAPM TG 43)
(a)
(b)
A significant part of the uncertainty in current BT dosimetry is due to this lack of metrology, because the conversion procedure needed for BT dosimetry is a relevant additional component of uncertainty on the final value of the absorbed dose imparted to the patient.
Rationale for a paradigma shift in BT dosimetry
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
National medical associations and scientific internationa l organizations ─ESTRO, AAPM-BTSC, IAEA, ICRU,
BIPM and
EURAMET─
have ongoing research programs in the field of Brachytherapy and encouraged the metrological community to develop
absorbed dose standards to assure direct traceability
of brachytherapy sources.
The European project “Brachytherapy”
is establishing absorbed
dose primary standards for brachytherapy.
8%
7%
5%3%
9%
68%
gynaecologyprostatebreastbronchushead & neckothers
Main sitesof tumor
treated with Brachytherapy(ESTRO, 2007)
This new approach is expected to improve the accuracy of brachytherapy dosimetry towards a level comparable to that typical of dosimetry in radiotherapy with external beams.
Rationale for a paradigma shift in BT dosimetry
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
The project “Increasing cancer treatment efficacy using 3D Brachytherapy” is one of the 21 European Joint Research Projects in metrology selected for funding in 2008 within the action iMERA plus under the Seventh Framework Programme
for research
and technological development of the European Commission.
The European Commission
co-fund the project according to the
Grant Agreement No. 217257
signed with the European
Association of National Metrology Institutes (EURAMET e.V.).
The project was launched on July 2008
for a period of 36
months
(ending in June 2011), with code T2.J06
and short name
“Brachytherapy”.
The European project “Brachytherapy”
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
This collaborative research project refers to research in ionizing radiation metrology focused on the program “TP2:
Health” in the European
Metrology Research Programme.
It responds to the necessity that:
the dose measurements for brachytherapy treatments be traceable to absorbed-
dose-to-water primary standards;
the uncertainty
on the dose delivered to
the target volume be within international accepted tolerance levels.
Target programmeTP2: HEALTHIn our centre
the Human as theMeasurement Object
The European project “Brachytherapy”
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
PARTICIPANTS
The project puts together the research and measurement capabilities of ten European National Metrology Laboratories
in the field of ionizing
radiation:BEV
(Austria); CMI
(Czech Republic);
ENEA-INMRI
(Italy); ITN
(Portugal);
LNE-LNHB
(France); VSL
(The
Netherlands); NPL
(United Kingdom);
PTB
(Germany); SSM
(Sweden); STUK
(Finland).
Project coordination has been entrusted to the ENEA-INRI, Italy (Dr. Maria Pia Toni).
The European project “Brachytherapy”
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Collaborations for research activities were formalized with:
the company PTW-Freiburg GmbH, Germany
the company IBA Dosimetry
GmbH, Germany
the Linköping
University, Sweden
A scientific collaboration has started with the BRAPHYQS (BRAchytherapy PHYsics Quality assurances System ) group of GEC-
ESTRO as representative of the project end users.
The chair of the BRAPHYQS group, dr. Frank-Andrè
Siebert, gives his contribute to “Brachytherapy”
project meetings.
The European project “Brachytherapy”
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Absorbed dose to water primary standards for brachytherapy sources based on calorimetry (a) or ionometry (b) and Monte Carlo based conversi on procedures
Brachytherapy sources (a) and transfer standards (b) directly calibrated in terms of Dw, 1 cm
(a)
(b)
(a)
(b)
The joint research project “Brachytherapy” developed absorbed dose primary standards for brachytherapy for a more sound metrology and with a target uncertainty on the absorbed dose imparted to the patient in the future brachytherapy practice:
u (Dw
, 1 cm) less than 5% (k=1)
The European project “Brachytherapy”
OBJECTIVES
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Dw
STANDARDSFOR LOW DOSE RATE BRACHYTHERAPY SOURCES
(prescription dose rate from 5 to 20 mGy/min)
Development and commissioning of a set of three independent primary standards
to allow the determination of the DW
at the reference distance of 1 cm from the source, with an uncertainty u(Dw
) ≈2%, k=1.
Standards are based on ionometric methods
(the more
appropriate due the low intensity of the sources): wide-angle and/or variable volume ionisation chambers.
A preliminary comparison exercise has started to assess the standards agreement and identify possible errors (results should be available by February 2011).
New brachytherapy standards for LDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Dw
STANDARDSFOR LOW DOSE RATE
BRACHYTHERAPY SOURCES
(prescription dose rate from 5 to 20 mGy/min)
1 cm
S
Air
Water
S,wK&
LNE-LNHB standard based on a free-in-
air ring-shaped ionisation chamber
and a water equivalent spherical
source phantom assembly.
PTB standard based on a large
extrapolation ionisation chamber
with water equivalent front and back walls.
ENEA standard based on a
large angle and variable volume
ionisation chamber that allows measurements in
graphite phantom.
New brachytherapy standards for LDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
The LNE-LNHB is establishing a two-step procedure to overcome practical difficulties encountered when measuring Dw .
K&
The LNE-LNHB primary standard development in terms of Dw
for 125I brachytherapy seedsI. Aubineau-Lanièce, P. Aviles-Lucas , J.-M. Bordy, B. Chauvenet, D. Cutarella, J. Gouriou, J. Plagnard
1 cm
Water
?wD&
Air
wcolK ,&
1 cm
What we want to assessTheoretical configuration really considered -> a correction factor will be applied to compensate for the lack of backscattering photons because the water sphere is surrounded by air instead of water.
Liquid water is replaced by PMMA -> a correction factor will be applied for the non perfect equivalence of PMMA with liquid water
1. Determine the water at the surface of a 1 cm radius water equivalent spherical phantom placed in air and containing the LDR seed in its centre.
1 cm
Air PMMA
PMMAcolK ,&
The LNE-LNHB Dw
primary standard for 125I BT seeds
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
The LNE-LNHB is establishing a two-step procedure to overcome practical difficulties encountered when measuring Dw .
2. Use a free-in-air circular ring-shaped ionization chamber to measure and apply appropriate MC correction factors to go from to
aircolK ,&
aircolK ,& PMMAcolK ,
&
=wD& .1.)1(
1.1..ge
w
airair −⎟⎠⎞
⎜⎝⎛
ρI.
V ∏i
ikglobalF
The LNE-LNHB Dw
primary standard for 125I BT seeds
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
The PTB Dw
primary standard for 125I BT seeds
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
This factor converts the difference of air kerma at chamber deptThis factor converts the difference of air kerma at chamber depths xhs xii a a and and xi+1xi+1 directly into the water kerma at chamber depth x=0 .directly into the water kerma at chamber depth x=0 .
As in phantom extrapolation measurements are having a nonlinear As in phantom extrapolation measurements are having a nonlinear dependence of dependence of Q(x)Q(x) due to attenuation and scatter, a Montedue to attenuation and scatter, a Monte--Carlo Carlo based conversion factor was introducedbased conversion factor was introduced
New extrapolation concept for the GROVEX IINew extrapolation concept for the GROVEX II
If the thickness of the chamber front wall is equivalent to 1 cm of water, then: )1()0( cmDK W
PhW
&& ≅
The PTB Dw
primary standard for 125I BT seeds
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Measurements with the Grovex II in the PTB reference fields for X-radiation
The PTB Dw
primary standard for 125I BT seeds
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
graphitepolystyrene air lead
dxxdQ
meWxD
aa
)(1)( =
)(1)()(1)()()( νρ
ννρ
ννν eaeaa divKdivBKD Ψ−≅Ψ−−= Ephotons
< 40 keV
[ ] eiia
aa xQxQme
WdKdKii
ΔΨ+−=− +∫∫+
)()(1)()( 1
1
νννννν
∫ ∫+
+
−
=
1
1
)()(
)0(),(
x x
ii
V Vaa
gMCxx
dKdK
KC
νννν
effective measuring volume
Δx
P
x
variable measuring volume
x0
P
r = 30 cm
source P
P0r
source
P
1
2 3
The ENEA-INMRI Dw
primary standard for 125I BT seeds
Potential rings (V/2)
Set-up A
( )[ ] eMC
xxiiMC
xxgiiii
CxQxQCAe
WPK ΔΨ+−=++
+ ),(1),(11
)(1)(
)()()( , PKCPKPD gMC
wgww =≅ )()(0
,0 PDCPD wMC
PPw ⋅=
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Preliminary results by Monte Carlo calculation EGS code (set-up A)
The measurement method applied to the ENEA-INMRI standard was developed by PTB as described in:
T Schneider, B Lange and H-J Selbach, Towards a determination of the absorbed dose to water in water for low-energy photon-emitting brachytherapy seeds, Metrologia 44 (2007) 407–414
T Schneider, A method to determine the water kerma in a phantom for x-rays with energies up to 40 keV, Metrologia 46 (2009) 95–100
ongoing
The ENEA-INMRI Dw
primary standard for 125I BT seeds
N.B.Minimum chamber depth x0 and distances between the
effective measuring volume and the lateral chamber wall are greater than the CSDA range for secondary particles generated in air by photons with E < 40 keV
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
graphitepolystyrene air lead
dxxdQ
meWxD
aa
)(1)( =
)(1)()(1)()()( νρ
ννρ
ννν eaeaa divKdivBKD Ψ−≅Ψ−−=
[ ] eiia
aa xQxQme
WdKdKii
ΔΨ+−=− +∫∫+
)()(1)()( 1
1
νννννν
effective measuring volume
ΔxP
x
variable measuring volume
x0
P
source
source
The ENEA-INMRI Dw
primary standard for 125I BT seeds
Potential rings (0 ÷ V)
P0
P0
)(')()( ,00 ogMC
wgww PKCPKPD ⋅=≅
( )[ ] eMC
xxiiMC
xxgiiii
CxQxQCAe
WPK ΔΨ+−=++
+'
),(1'
),(011
)(1)(
( )( ) ( )∫ ∫
+
+ −=
1
1
0'),(
x x
ii
V Vaa
gMCxx KK
PKC
νν
Ephotons
< 40 keV
1
2
Set-up B
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Dw
STANDARDSFOR HIGH DOSE RATE
BRACHYTHERAPY SOURCES
(prescription dose rate of 200 mGy/min or higher)
Development and commissioning of a set of four independent primary standards
to allow the determination of the Dw at the
reference distance of 1 cm from the source, with an uncertainty u(Dw
) ≈2%, k=1.
The standards are based on calorimetry: water calorimeters and graphite calorimeters.
A preliminary comparison exercise has started to assess the standards agreement and identify possible errors (results should be available by February 2011).
New brachytherapy standards for HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Dw
STANDARDSFOR HIGH DOSE RATE
BRACHYTHERAPY SOURCES
(prescription dose rate of 200 mGy/min or higher)
Thermistor (4x)
Heat sink (Al) Glass cell wall
NPL standard based on a ring-shaped graphite
calorimeter
ENEA-INMRI standard based on a ring-shaped graphite
calorimeter
VSL standard based on water calorimeter
PTB standard based on a water calorimeter
New brachytherapy standards for HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Detector arrangement changed for use with HDR BT sources
possible meassurement positions: +60 mm, +48 mm, +24 mm, -24 mm
stainless steel needle for 1.1 mm source
The PTB Dw
primary standard for 192Ir HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
The PTB Dw
primary standard for 192Ir HDR BT sources
Detector arrangement changed for use with HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
PEEK
cell wall
Thermistor (4x)
Source
entranceconnected to after‐loader
Glass cell wall
Aluminum
heat sink
The VSL Dw
primary standard for 192Ir HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Validation heat transport model
The VSL Dw
primary standard for 192Ir HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Wednesday, 17 November 2010
27
Complete
system of NPL’s new brachytherapy calorimeter
Measurement electronics:
Nanovoltmeters
Power supplies for heaters
Wheatstone bridges for sensors
Vacuum pump
Vacuum housing for HDR brachytherapy graphite calorimeter
Electrical connections of heating and sensing thermistors
The NPL Dw
primary standard for 192Ir HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Wednesday, 17 November 2010
Ring shaped graphite core (2 mm x 5 mm, radius = 2.5 cm)
Sectional view
Vacuum gaps between core, inner graphite tubes and central aluminium tube to control source self-heating
Cylindrical graphite phantom (radius = 10 cm, height = 14 cm) enclosed in vacuum can (not shown)
HDR brachytherapy source will be inserted through aluminium tube
Objective: Direct measurement of absorbed dose to graphite at 2.5 cm due to HDR 192Ir brachytherapy source conversion to quantity of interest: absorbed dose to water at 1 cm
The NPL Dw
primary standard for 192Ir HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Schematic section of the ENEA-INMRI calorimeter (left) with the core in black and the calorimeter assembled (right) in the PMMA housing, (13 cm ext dia, 10 cm int dia ).
medium
top graphite disc with source holder
sou
rce h
old
er
heate
r h
old
er
Bottom graphite disc
j-co
re b
ase
PM
MA
ho
usi
ng
j-co
re b
ase
mediumh
eate
r h
old
er
PM
MA
ho
usi
ng
The ENEA-INMRI Dw
primary standard for 192Ir HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Left: the core and the jacket base fixed to the jacket-core base after embedding the thermistors in the core and in the jacket base. The core is suspended on the jacket base by 3 polystyrene 1 mm dia pins.
The ENEA-INMRI Dw
primary standard for 192Ir HDR BT sources
Right: detail of the core and the 0.3 mm dia NTC thermistor, used in the core and in the jacket.
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
20 c
m
20 cm13 cm
11 c
mThe calorimeter is inserted in a larger graphite phantom (left) to ensure at the core position the full backscattering conditions for 192Ir .
On the right, detail of the 3D heat transfer calculations: irradiation
The ENEA-INMRI Dw
primary standard for 192Ir HDR BT sources
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
MethodologyStar-shaped comparisons
Indirect comparisons by circulating two transfer standard instruments from one NMI to another, for each comparison
Two ionisation well-type chambers of different types have been selected to act as transfer standard instruments: the same chamber models, but different samples, for LDR and HDR comparisons
Each partner will have to determine:
the calibration coefficient of the transfer instruments based on the value determined with its novel primary standard for its own source .
the calibration coefficient of the transfer instruments in terms of to allow an experimental determination of the conversion gamma constant Λ
with an improved uncertainty.
WD&
RK&
Novel LDR&HDR standards comparison exercises
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Transfer standard instruments
PTW 33004
Outer Height 19.05 cm
Outer base diameter 17.8 cm
Outer well diameter 9.3 cm
Inner well diameter 3.2 cm
Weight 2.4 kg
Volume 200 cm3
Potential of HV electrode with respect to collecting electrode
300 V
SI HDR 1000+(ref 90008)
Height 15.6 cm
Diameter 10.2 cm
Weight 2.7 kg
Volume 245 cm3
Potential of HV electrode with respect to collecting electrode
300 V
Novel LDR&HDR standards comparison exercises
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
SOURCES USED FOR THE LDR COMPARISON
BEBIG IsoSeeds® I-125 sources (reference I25.S16)
The IsoSeeds® I-125 sources contain a cylindrical-shaped ceramic saturated with an iodine-125 compound, which is enclosed in a laser- welded titanium tube.
Range Air kerma strength(µGy m²/h)
Apparent activity (mCi)
Apparent activity (MBq)
A25 26.10 – 31.75 20.6 – 25.0 761 - 925
Novel LDR&HDR standards comparison exercises
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Novel LDR&HDR standards comparison exercises
35
Nucletron microSelectron-v1 classic HDR 192Ir source (PTB and NPL)
Nucletron microSelectron-v2 HDR 192Ir source (ENEA-INMRI and VSL)
SOURCES USED FOR THE LDR COMPARISON
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
LDR comparison schedulePTB provided the transfer chambers and carry out the stability checks
The participants will have two weeks to realize the calibration of both transfer chambers.
The instruments will circulate according to the schedule given below.
All the results would be exchanged between participants at the end of January
Measurements at NMI
PTB: from 13 to 26 September 2010 (done)
LNE-LNHB: from 04 to 17 October 2010 (done)
ENEA-INMRI: from 15 to 28 November 2010
PTB
LNE- LNHB
ENEA - INMRI
Stability checks
Novel LDR&HDR standards comparison exercises
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 201037
NPL provided the transfer chambers and carry out the stability checks
Calibration at PTB:7-18 Jun 10 (done)
Calibration at VSL:6-16 Jul 10 (done)
Calibration at ENEA-INMRI:25 Oct - 7 Nov 10 (done)
All the results would be exchanged between participants at the end of January
ENEA- INMRI
(I)
NPL
Stability check
PTB(DE)
NPL(UK)
VSL(NL)
1 2
3
4
56
7
8
Novel LDR&HDR standards comparison exercises
HDR comparison schedule
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Dissemination of the novel BT standards
Dw
BASED METROLOGICAL CHAIN
The final objective of the “Brachytherapy” project is to implementat a metrological chain to assure the traceability
of brachytherapy dosimetry to the new absorbed dose to water standards.
Transfer standard:
well-type ionization chamber;
small volume ionization chambers associated with water phantom
plastic scintillator dosemeters (PSDs) associated with water phantom
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
System with six 0,0System with six 0,088
cmcm33
ionizing chambers for ionizing chambers for measurements in water phantom
Dissemination of the novel BT standards
Well type chambers acting Well type chambers acting as transfer standard as transfer standard (dose (dose calibrators) calibrators)
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Development of the secondary standard based on well-type ionization chamber
(PTW TM33004):
BEV secondary standard for absorbed dose to water quantity for brachytherapy sources was calibrated at the PTB.
Calibration sheet was issued also for NDw quantity with source relevant Λ conversion factor experimentally determined according to the new standard.
Dissemination of the novel BT standards
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
What does it change in treatment plans?
( ) ( )( ) ( ) ( )θ
θθθ ,
,,,
00
rFrgrGrGSrD KW ⋅⋅⎥
⎦
⎤⎢⎣
⎡⋅Λ⋅=&
The updating of the tretment plans for the shift from Air kerma to Absorbed dose to water as reference quantity for calibration of BT source just need to substitute the product (Sk L) with Dw , in the formalism of the AAPM TG 43 protocol:
Dw
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
A three-day scientific Symposium is organized to present the results obtained to the interested Medical Physics community and promote a updated brachytherapy dosimetry protocol based on direct calibrations in terms of Dw .
The symposium is organized together with the project “EBCT” that deals with external beam cancer therapy using ionising radiation and high-intensity ultrasound
Title: "Symposium on developments in physical measurement for cancer therapy".Period: 29.11.2011 to 01.12.2011. Place: PTB, Braunsweigh, Germany
Representative of the association and task group relevant on these topics are expected to contribute to the symposium
Dissemination of the novel BT standards
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
A set of Dw primary standards have been developed within the project “Brachytherapy”.
The dissemination of project results is of direct interest to the international community of radiotherapy centres performing BT clinical treatments.
The possibility given by the present project of basing BT dosimetry on Dw primary standards will potentially increase the accuracy and safety of BT to a level comparable to that typical of radiotherapy with external accelerator beams.
Conclusion
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Conclusion
Based on the current developments, a progress is encouraged for a more direct and accurate traceability chain of BT sources according to the demand for a more sound metrology and an optimization of dose delivery in BT treatments.
No efforts are required at SSDL or clinic level: the same reference chamber (well type ionization chambers) could be still used and to update the software for treatment plan results easy.
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
Organization, late next year, of a key comparison to validate the new Dw BT standards.
Implementation of a Dw based metrological chain to disseminate the new standards througt the SSDL laboratories
Availability of calibrations in terms of DW to brachytherapy centres.
Availability of updated protocols in terms of DW for more accurate dosimetry to brachytherapy centres.
Future work
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010
The research within the EURAMET joint research project “Brachytherapy” receives funding from the European Community’s Seventh Framework Programme, ERA-NET Plus, under the iMERA-Plus Project - Grant Agreement No. 217257.
The author would like to thank the partners of the “Brachytherapy” project for their support to this contribute.
In particular:Isabelle Aubineau-Lanièce
(CEA, LIST, Laboratoire National Henri Becquerel, France),Thorsten Sander
(National Physical Laboratory, United Kingdom),Hans-Joachim Selbach
(Physikalisch-Technische Bundesanstalt, Germany), Jacco de Pooter
(Van Swinden Laboratorium B.V., The Netherlands).
Acknowledgements
INV015, “New brachytherapy standards paradigm shift”, Maria Pia Toni, ENEA-INMRIIAEA International Symposium "Standards, Applications and QA in Medical Radiation Dosimetry (IDOS)“, Vienna, 9-12 November 2010